The R&S OSP-B157W8 or R&S OSP-B157W8PLUS and R&S OSP-B157WX are standard modules for the open switch platform R&S OSP.
The R&S OSP-B157WX serves as a frequency extension for the base module
R&S OSP-B157W8 or R&S OSP-B157W8PLUS. While the extension forwards lower
frequency signals to the base module, it can bypass higher frequency signals to an
appropriate measurement channel.
Both modules are typically used in the automatic test system R&S TS8997 that performs regulatory conformance tests of wireless short range devices (SRDs). The tests
cases are focused on the 2.4 GHz and 5 GHz frequency bands for Wi-Fi and ISM
applications (industrial, scientific and medical). These tests are performed according to
the standards FCC 15.407, ETSI EN 300 328, ETSI EN 301 893, ETSI EN 302 502,
ETSI EN 300 440 and others.
Intended use
The R&S OSP-B157W8, R&S OSP-B157W8PLUS and R&S OSP-B157WX are designated for the development, production and verification of electronic components and
devices in industrial and laboratory environments. Use the modules only for their designated purpose. Observe the operating conditions and performance limits stated in the
data sheets.
Preface
Designation "module" vs. "switch unit"
This manual uses the following nomenclature:
●
The R&SOSP-B157W8 is referred to as "base module".
●
The R&SOSP-B157W8PLUS is referred to as "base module".
●
The R&SOSP-B157WX is referred to as "extension module".
●
The R&SOSP is referred to as "switch unit".
Integration
●
The base module R&S OSP-B157W8 or R&S OSP-B157W8PLUS is (or must be)
integrated in an R&S OSP150 or R&S OSP120 switch unit.
●
The extension module R&S OSP-B157WX is (or must be) integrated in an
R&S OSP120 switch unit (not in an R&S OSP150).
Except for one optional additional single-slot module in slot A11 of the switch unit, no
additional modules can be installed in a single switch unit beside one R&S OSPB157W8, one R&S OSP-B157W8PLUS or one R&S OSP-B157WX. For all information
related to the R&S OSP switch units, refer to their operating manual.
5User Manual 1178.5795.02 ─ 03
Page 6
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Frequency ranges
Basically, the modules are designed to cover the 2.4 GHz and 5 GHz frequency bands.
However, some of the above mentioned regulatory ETSI and FCC tests include measuring spurious emissions up to 18 GHz or even up to 40 GHz.
Therefore, the frequency coverage of the modules is as follows:
●
The R&S OSP-B157W8 operates in a frequency range up to 6 GHz. The
R&S OSP-B157W8PLUS operates in a frequency range up to 7.5 GHz. Both base
modules offer a special by-path for spurious emissions measurements up to
18 GHz. The integrated power meter and signal conditioning is designed for inband use only.
The lower end of the frequency range of the R&S OSP-B157W8 and R&S OSPB157W8PLUS depends on the selected paths and application. It starts, for example, approximately at 100 MHz, if internal power measurements are involved, or
approximately at 1 GHz, if directional couplers are involved.
●
The R&S OSP-B157WX extends the spurious emissions frequency range up to
40 GHz. Signals to be measured in this frequency range are not forwarded to the
DUT input ports of the base module, but bypassed directly to the signal analyzer.
Preface
Key features of the base module R&S OSP-B157W8
The 6 GHz module R&S OSP-B157W8 with up to 8 channels is based on a printed RF
switch board in solid-state relay (SSR) architecture. It allows flexible operation of the
connected DUT (up to 8 ports) and measurement instruments.
The module features:
●
Signal conditioning by integrated attenuators, couplers and combiners (without
amplifiers or filters)
●
RF switching for wireless test cases up to 6GHz
●
RF switching for spurious emission measurements up to 18 GHz (exception: if a
frequency extension module R&S OSP-B157WX is present, these signals are handled within the extension module)
●
Power measurements in the 2.4 GHz and 5 GHz bands with specific evaluations
for wireless standards
●
Built-in data acquisition board for control of the signal paths and for collection and
A/D conversion of the power measurements
●
Multiple trigger input and output options
●
External clock input
●
Video input
●
LAN interface for remote control of the module operation by the test software
R&S WMS32
Key features of the base module R&S OSP-B157W8PLUS
In addition to the features of the ".02" version R&S OSP-B157W8, the ".05" version of
the module features:
●
Frequency range for signalling up to 7.5GHz
6User Manual 1178.5795.02 ─ 03
Page 7
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
●
Additional step attenuator R2 for an extended attenuation range, see Figure 4-3
Each of the 8 configurable channels in the R&S OSP-B157W8 and R&S OSPB157W8PLUS has the following integrated semiconductor components:
●
An individually programmable attenuator
●
An analog power detector for synchronous RMS power measurement with high
sampling rate and burst detection
●
An A/D converter for the power measurement results
●
Solid-state switches, which enable:
–Switching the (vector) signal generator output to the calibration port or to the
DUT ports
–Switching the measurement paths from DUT ports to a spectrum analyzer / test
receiver or to the power measurement
–Signal conditioning via the integrated couplers, directional couplers, attenuators
and combiners
Key features of the frequency extension module R&S OSP-B157WX
Preface
The 40 GHz module R&S OSP-B157WX is especially designed to provide RF switching for spurious measurements up to 40 GHz.
It uses coaxial RF architecture and features:
●
Electromechanical RF relay switches for flexible operation of up to 8 DUT channels
●
The module can forward incoming signals on each DUT channel to the connected
DUT input of the R&S OSP-B157W8 or R&S OSP-B157W8PLUS. This forwarding
is used for measurements at frequencies up to 6 GHz, which the R&S OSPB157W8 can handle, or up to 7.5 GHz, which the R&S OSP-B157W8PLUS can
handle.
●
Alternatively, incoming signals on one single DUT channel can be switched to the
module's [Rx] port and on to a connected signal analyzer, receiver or other measurement instrument. This switching state is typically used for measurements at
frequencies above 6 GHz or 7.5 GHz and up to 40 GHz, which the base modules
cannot handle. (Signals for spurious emission measurements up to 18 GHz are
also switched to the instrument connected to the extension module's [Rx] port.)
●
If the R&S OSP-B157WX forwards the DUT's signals, it also switches the signal
analyzer path from its [RxOUT] port (input port) to its [Rx] port. This [Rx] port is
connected to the signal analyzer or test receiver, while the [Rx] port of the
R&S OSP-B157W8 or R&S OSP-B157W8PLUS is connected to the [RxOUT] port
(input port) of the R&S OSP-B157WX.
●
Additionally, you can add the signal of a companion device to one of the DUT paths
via its internal switches and a directional coupler. The type of companion device
depends on the DUT. For example, if your DUT is a Wi-Fi transceiver, the companion device can be a wideband radiocommunication tester R&S CMW500. In this
example, it would simulate a wireless router that maintains a bidirectional connection with the transceiver.
7User Manual 1178.5795.02 ─ 03
Page 8
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Screenshots
Sample screenshots in this documentation are used to illustrate as much as possible of
the functions provided by the modules and of potential interdependencies between
parameters. Note that:
●
The values in these screenshots do not necessarily represent realistic application
situations.
●
Do not consider the values as recommended by Rohde&Schwarz.
●
The screenshots shown can differ, depending on your particular equipment and
configuration.
Preface
8User Manual 1178.5795.02 ─ 03
Page 9
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
2Documentation Overview
This section provides an overview of the R&S OSP-B157W8, R&S OSP-B157W8PLUS
and R&S OSP-B157WX user documentation. Unless specified otherwise, you find the
documents on the R&S OSP product page at:
www.rohde-schwarz.com/manual/osp
2.1Getting Started Manual
Introduces the modules and describes how to set up and start working with them.
Includes safety instructions, key features, preparation, mounting, setup and basic operation. A printed version is delivered with the modules.
Documentation Overview
Help
2.2User Manual
Contains the description of all functions of the modules. It includes the contents of the
getting started manual and provides explanations, step-by-step procedures, figures
and examples. The user manual supports you during your first experience with the
product, from installation and configuration to various measurements. The content also
provides an introduction to remote control, a complete description of the remote control
commands with programming examples, and information on maintenance, instrument
interfaces and error messages.
The user manual is available for download or for immediate display on the Internet.
2.3Operating Manual of the Switch Unit
Contains the description of all instrument modes and functions of the R&S OSP Open
Switch and Control Platform (switch unit). It is available for download on the Internet.
2.4Help
The R&S OSP-B157W8 and R&S OSP-B157W8PLUS are operated either by direct
bus commands or by the software R&S WMS32, which is part of the R&S RF Test
Suite. The R&S OSP-B157WX can also be operated by the software R&S OSP Panel.
The software-embedded help content in the R&S RF Test Suite provides quick, context-sensitive access to information for the software packages R&S EMC32,
R&S AMS32 and R&S WMS32. Refer to the R&S WMS32 part of this help content for
application of the modules and switch units in the R&S TS8997 test system. The help
content also describes templates for test cases that you can use in tests according to
9User Manual 1178.5795.02 ─ 03
Page 10
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
various FCC and ETSI standards. Examples are FCC 15.407, ETSI EN 300 328 or
ETSI EN 301 893.
The software, which requires a license, is available at www.rohde-schwarz.com/soft-
ware/emc32.
2.5Basic Safety Instructions
Contains safety instructions, operating conditions and further important information.
The printed document is delivered with the switch unit.
2.6R&S OSP on the Internet
The website www.rohde-schwarz.com/product/osp provides:
●
Key facts, features and options for the family of R&SOSP switch units
●
Rohde&Schwarz contacts for information, quotes, demos and calibration services
●
Technical documentation (operating manual for R&SOSP switch units)
●
Brochures and Data Sheets
●
Application Notes
●
Drivers
●
The current firmware version with OSA and release notes
●
The R&S OSP Panel software with release notes
●
Related documents and articles
●
News and information on software updates
●
Service Manual (only for registered users)
Documentation Overview
R&S OSP on the Internet
For operation of the modules, refer to the documentation of the R&S WMS32 software.
The documentation is available as a help function in the R&S RF Test Suite, which
requires a license and comprises the software packages R&S EMC32, R&S AMS32
and R&S WMS32. Also, refer to the related release notes.
2.6.1Brochures and Data Sheets
The data sheet contains the technical specifications of the R&S OSP. It also lists the
options and their order numbers, and optional accessories.
The brochure provides an overview of the instrument and deals with the specific characteristics.
See www.rohde-schwarz.com/brochure-datasheet/osp.
10User Manual 1178.5795.02 ─ 03
Page 11
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
2.6.2Application Notes
Application notes and application cards deal with special applications or background
information on particular topics.
See www.rohde-schwarz.com/application/osp.
2.6.3Drivers
Up-to-date instrument drivers are available for download.
See www.rohde-schwarz.com/driver/osp.
2.6.4Release Notes and Open Source Acknowledgment (OSA)
The release notes list new features, improvements and known issues of the current
firmware or software version and describe the installation.
Documentation Overview
R&S OSP on the Internet
The open source acknowledgment document provides verbatim license texts of the
used open source software.
See www.rohde-schwarz.com/firmware/osp and www.rohde-schwarz.com/software/
osp.
2.6.5Service Manual
Describes handling failed modules, module replacement, troubleshooting and special
remote control commands for service purposes. The document also contains spare
part lists and mechanical drawings.
The service manual ("Classified Service Document") is available for Rohde & Schwarz
personnel, only.
11User Manual 1178.5795.02 ─ 03
Page 12
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
3Preparing for Use
●Unpacking and Checking........................................................................................ 12
●Mounting the Base Module..................................................................................... 12
●Mounting the Extension Module..............................................................................17
●Installing or Updating the Firmware........................................................................ 21
●Putting into Operation............................................................................................. 22
3.1Unpacking and Checking
If your delivery includes a module that is not mounted in a switch unit, be especially
careful with unpacking the module, which is not protected in a casing.
Check the delivery for completeness according to the included shipment list. If the
delivery is incomplete, contact Rohde & Schwarz.
Retain the original packing material. If the instrument needs to be transported or shipped later, you can use the material to protect the control elements and connectors.
Preparing for Use
Mounting the Base Module
Mounting modules into switch units
●
If you have received one or more modules that are already integrated in switch
units, you can skip the chapters Mounting the Base Module and Mounting the
Extension Module. Proceed with Chapter 3.5, "Putting into Operation",
on page 22.
●
If you have received one or more modules that are not yet integrated in switch
units, proceed with chapters Mounting the Base Module and Mounting the Exten-
sion Module.
3.2Mounting the Base Module
Mounting the base module R&S OSP-B157W8 or R&S OSP-B157W8PLUS into an
R&S OSP150 or R&S OSP120 switch unit requires the space of two slots: rear slots
A12 and A13 together with front slots A12F and A13F. Typically, the rear slot A11
remains free, but optionally it can remain occupied by a single-slot module (not operational). Only one base module fits into one R&S OSP150 or R&S OSP120 switch unit.
12User Manual 1178.5795.02 ─ 03
Page 13
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Risk of electric shock
Opening a switch unit implies the risk of touching current-carrying components and suffering an electric shock, if the device is not properly disconnected from grid power.
Before opening the casing, ensure that the switch unit is disconnected from the power
supply by removing the plug from the rear AC power connector.
Observe the safety instructions for units with a removable casing.
Carefully follow the step-by-step instructions for module replacement to avoid injury
and to help ensure safe operation.
Risk of injuries
When removing the rear feet, the switch unit can slip out of its casing.
To avoid the risk of personal injuries and any damage, put the switch unit onto its front
handles before removing the rear feet and taking off the casing.
Preparing for Use
Mounting the Base Module
When mounting the casing back onto the switch unit, take care not to pinch your fingers. Also pay attention not to damage or pull off protruding parts, especially cables.
Screw the rear feet back on immediately after mounting the casing. Do not move the
switch unit with the rear feet missing.
Removing an existing (old) R&S OSP-B157 module
If you want to install the R&S OSP-B157W8 or R&S OSP-B157W8PLUS in a switch
unit, which is occupied by an existing module (for example R&S OSP-B157), first
remove the existing module.
Remove the existing module in the reverse order of the mounting procedure, described
in the service manual of the R&S OSP (1505.3896.82), for example chapter "Replacing
the Option OSP-B157".
Mounting procedure for the R&S OSP-B157W8 or R&S OSP-B157W8PLUS
For mounting the module into an R&S OSP150 or R&S OSP120 switch unit:
1. Turn off the switch unit.
13User Manual 1178.5795.02 ─ 03
Page 14
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
2. Unplug the power supply cable.
3. Disconnect all other cables, too.
4. If the switch unit holds any modules, remove these modules and their ribbon
cables, except for a module in the rear slot A11.
5. If the rear slot A11 of the switch unit holds a module, we recommend removing it.
(Optionally you can leave it there, but you cannot operate it due to missing control
interfaces.)
6. If you have removed a module from rear slot A11, close the open slot with a blind
plate, using the same screws that fixed the module.
7. If blind plates cover unused front slots A12F or A13F of the switch unit, remove
these blind plates.
8. If blind plates cover unused rear slots A12 or A13 of the switch unit, remove these
blind plates.
9. If a blind plate covers an unused rear slot A11 of the switch unit, leave it there.
Preparing for Use
Mounting the Base Module
10. Place the switch unit on its front handles.
11. Optionally, if at the switch unit's feet there are yellow stickers that you want to save,
remove the blue rubber bumpers from the feet.
12. Screw off the switch unit's feet.
Removing these screws releases the switch unit's casing.
13. Push the switch unit's casing upwards.
14. Remove the casing from the switch unit.
15. Lay down the switch unit, with its open side facing upwards.
16. Unplug the CAN bus cable (W3) from connector X8 on the switch unit's main board
and from the CAN connector on the switch unit's rear panel.
17. Remove the CAN connector from the switch unit's rear panel.
18. If your switch unit is an R&S OSP120:
●Unplug the LAN cable (W4) from the RJ45 connector on the processor board
and from the LAN connector on the switch unit's rear panel.
●Remove the LAN connector from the switch unit's rear panel.
19. Remove the cable holder of the CAN (and LAN) cable.
20. Remove the rear panel strut between slot A12 and slot A13.
21. Screw the rear plate of the module to the switch unit's rear panel (10 countersunk
screws, type Torx M2.5x8, included in the delivery; step 34 uses 3 more of these
screws).
22. Insert the guide cage into the switch unit's free space, with the protruding flange
(item (2) in Figure 3-1) directed towards the switch unit's power supply and control
board, as in Figure 3-3.
14User Manual 1178.5795.02 ─ 03
Page 15
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Figure 3-1: Guide cage for holding the module in the switch unit
1 = 2 holes for accessing 2 screw hole positions below
2 = Protruding flange
3 = 3 screw holes
4 = Guiding structure elements, into which you slide the module in step 26
23. Screw the guide cage to the switch unit's base plate (5 cylinder-head screws, type
Torx M2.5x6, included in the delivery), items (1) and (3) in Figure 3-1 or (5) and (6)
in Figure 3-3.
Preparing for Use
Mounting the Base Module
24. Connect the 4-pole connector at the short end of cable W20 (4 wires, red/black/
black/gray) to the power input connector X201 of the module's RF board, item (9)
in Figure 3-3.
25. Connect the 4-pole connector at the long end of cable W20 (2 wires, brown/black)
to the power input connector X203 of the module's data acquisition board, item (10)
in Figure 3-3.
26. In the open front panel of the switch unit, place the module's rear end just into the
guiding structure elements of the guide cage, as shown in Figure 3-3.
Figure 3-2: Placing the base module in the opened switch unit, before inserting it
27. Do not yet insert the module into the switch unit.
15User Manual 1178.5795.02 ─ 03
Page 16
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
28. Insert the single end of cable W20 (with the 12-pole connector) through the open
front panel of the switch unit while passing the guide cage on its right-handside.
As a result, the cable enters the switch unit through a narrow gap between the front
panel and the guide cage.
29. Connect the 12-pole connector of cable W20 to connector X12 on the switch unit's
control board, item (8) in Figure 3-3.
30. Push the module 80% of its length into the switch unit.
31. Fix the middle part of cable W20 by using the cable holder close to the front of the
switch unit, item (11) in Figure 3-3.
32. Push the module all the way into the switch unit, until the module's cover panel is
level with the switch unit (item (2) in Figure 3-3).
Preparing for Use
Mounting the Base Module
Figure 3-3: Opened switch unit with guide cage mounted and base module inserted
1 = Guide cage, mounted in the switch unit
2 = Front plate of the inserted module
3 = Rear end of the inserted module
4 = Secondary power cable for the control board, already existing in the switch unit
5 = 2 screws that you can reach through holes in the upper plate of the guide cage, fixing its left side
6 = 3 screws, fixing the protruding flange at the right side of the guide cage (underneath the secondary
power cable)
7 = Cable W20 (here shown as partly visible through the guide cage)
8 = 12-pole connector of cable W20, connected to connector X12 of the switch unit's control board
9 = 4-pole connector at the short end of cable W20 (4 wires), plugged into connector X201 on the
module's RF board (see detail at left-hand side)
10 = 4-pole connector at the long end of cable W20 (2 wires), plugged into connector X203 on the mod-
ule's data acquisition board
16User Manual 1178.5795.02 ─ 03
Page 17
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
11 = Cable holder for fixing cable W20
12 = Detail: identifying the connector for the short end (9) of cable W20
13 = Detail: view from the front of the switch unit (front panel not shown) through the empty guide cage
towards the rear panel
33. Screw the module's front plate to the switch unit's front panel (10 countersunk
screws, type Torx M2.5x6, included in the delivery).
34. Screw the module's rear plate, which you have already attached to the switch unit's
rear panel (step 21), to the module (3 countersunk screws, type Torx M2.5x8, included in the delivery).
35. Place the switch unit on its front handles.
36. Shift the casing over the unit.
37. Push the switch unit's casing down to the stop.
38. Screw on the switch unit's feet.
Tightening these screws fixes the switch unit's casing to the unit.
Preparing for Use
Mounting the Extension Module
39. If you have removed the blue rubber bumpers from the switch unit's feet, reattach
them.
40. Lay down the unit.
41. Reconnect the power cable to the unit.
3.3Mounting the Extension Module
Mounting the frequency extrension module R&S OSP-B157WX into an R&S OSP120
switch unit requires the space of two slots: rear slots A12 and A13 together with front
slots A12F and A13F. Typically, the rear slot A11 remains free, but optionally it can
remain occupied by a single-slot module (not operational). Only one R&S OSPB157WX module fits into one R&S OSP120 switch unit.
Risk of electric shock
Opening a switch unit implies the risk of touching current-carrying components and suffering an electric shock, if the device is not properly disconnected from grid power.
Before opening the casing, ensure that the switch unit is disconnected from the power
supply by removing the plug from the rear AC power connector.
Observe the safety instructions for units with a removable casing.
Carefully follow the step-by-step instructions for module replacement to avoid injury
and to help ensure safe operation.
17User Manual 1178.5795.02 ─ 03
Page 18
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Risk of injuries
When removing the rear feet, the switch unit can slip out of its casing.
To avoid the risk of personal injuries and any damage, put the switch unit onto its front
handles before removing the rear feet and taking off the casing.
When mounting the casing back onto the switch unit, take care not to pinch your fingers. Also pay attention not to damage or pull off protruding parts, especially cables.
Screw the rear feet back on immediately after mounting the casing. Do not move the
switch unit with the rear feet missing.
Preparing for Use
Mounting the Extension Module
Removing an existing (old) R&S OSP-B157 module
If you want to install the R&S OSP-B157WX in a switch unit, which is occupied by an
existing module (for example R&S OSP-B157), first remove the existing module.
Remove the existing module in the reverse order of the mounting procedure, described
in the service manual of the R&S OSP (1505.3896.82), for example chapter "Replacing
the Option OSP-B157".
Mounting procedure for the R&S OSP-B157WX
For mounting the module into an R&S OSP120 switch unit:
1. Turn off the switch unit.
2. Unplug the power supply cable.
3. Disconnect all other cables, too.
4. If the switch unit holds any modules, remove these modules and their ribbon
cables, except for a module in the rear slot A11.
5. If the rear slot A11 of the switch unit holds a module, we recommend removing it.
(Optionally you can leave it there, but you cannot operate it due to missing control
interfaces.)
6. If you have removed a module from rear slot A11, close the open slot with a blind
plate, using the same screws that fixed the module.
7. If a blind plate covers the unused rear slot A11 of the switch unit, leave it there.
18User Manual 1178.5795.02 ─ 03
Page 19
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
8. If you have removed one or two modules from rear slots A12 or A13, close the
open slots with blind plates, using the same screws that fixed the modules.
9. If blind plates cover unused rear slots A12 or A13 of the switch unit, leave them
there.
10. If blind plates cover unused front slots A12F or A13F of the switch unit, remove
these blind plates.
11. Place the switch unit on its front handles.
12. Optionally, if at the switch unit's feet there are yellow stickers that you want to save,
remove the blue rubber bumpers from the feet.
13. Screw off the switch unit's feet.
Removing these screws releases the switch unit's casing.
14. Push the switch unit's casing upwards.
15. Remove the casing from the switch unit.
Preparing for Use
Mounting the Extension Module
16. Lay down the switch unit, with its open side facing upwards.
17. Unplug the CAN bus cable (W3) from connector X8 on the switch unit's main
board, but not from the CAN connector on the switch unit's rear panel.
18. Unplug the LAN cable (W4) from the RJ45 connector on the processor board, but
not from the LAN connector on the switch unit's rear panel.
19. Remove the cable holder of the CAN and LAN cable.
20. Screw the adapter plate (item (1) in Figure 3-4), which comes separately, to the
base plate of the switch unit (12 cylinder-head screws, type Torx M2.5x6, included
in the delivery).
Figure 3-4: Overview of the parts for the fixation of module R&S
OSP-B157WX
19User Manual 1178.5795.02 ─ 03
Page 20
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
1 = Adapter plate with 11 threaded insert nuts (dark gray) and 12 screws (yellow) that fix the adapter
plate to the base plate of the switch unit
2 = Ground plate of the module, here in an isolated presentation for better view of the 11 screws (yel-
low) that fix the module to the adapter plate
3 = Top view of the complete module with the same 11 screws highlighted as in the isolated presenta-
tion (2)
21. Insert the module into the switch unit from the front side (item (3) in Figure 3-5).
Preparing for Use
Mounting the Extension Module
Figure 3-5: Inserting the module R&S OSP-B157WX
1 = 12 screws that fix the adapter plate to the base plate of the switch unit
2 = 11 threaded insert nuts that carry the module
3 = Inserting the module
4 = 11 screws that fix the module's ground plate to the adapter plate
5 = 10 countersunk screws that fix the module's cover panel to the front panel of the switch unit
22. Push the module all the way into the switch unit, until the module's cover panel is
level with the switch unit.
23. Screw the ground plate (item (2) and (3) in Figure 3-4) to the adapter plate (11 cylinder-head screws, type Torx M2.5x6, included in the delivery).
24. Screw the module's front plate to the switch unit's front panel (item (5) in Fig-
ure 3-5, 10 countersunk screws, type Torx M2.5x6, included in the delivery).
25. Connect the cables (included in the delivery) as follows between the interface
board of the module R&S OSP-B157WX and the main board of the switch unit:
Figure 3-6: Cable connections for controlling the frequency extension module
Left= Interface board of the module R&S OSP-B157WX
Right= Main board of the switch unit
1= 6-wire power cable
2 and 3 = 26-pin ribbon cables
X...= Connector numbers, see labels on the boards
26. Reconnect the CAN cable (W3) to connector X8 on the switch unit's main board).
Preparing for Use
27. Reconnect the LAN cable (W4) to the RJ45 connector on the switch unit's processor board.
28. Place the switch unit on its front handles.
29. Shift the casing over the unit.
30. Push the switch unit's casing down to the stop.
31. Screw on the switch unit's feet.
Tightening these screws fixes the switch unit's casing to the unit.
32. If you have removed the blue rubber bumpers from the switch unit's feet, reattach
them.
33. Lay down the unit.
34. Reconnect the power cable to the unit.
3.4Installing or Updating the Firmware
For installing or updating the firmware of the R&S OSP-B157W8 or R&S OSPB157W8PLUS, use the R&S WMS32 software package in the R&S RF Test Suite.
Click "Self Check" (labeled (1) in Figure 3-7) to open this dialog:
21User Manual 1178.5795.02 ─ 03
Page 22
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Figure 3-7: Firmware installation or update in R&S WMS32
Preparing for Use
Putting into Operation
Proceed as follows:
1. Click "Self Check".
2. Select the "R&S OSP-B157W8" or "R&S OSP-B157W8PLUS" entry in the device
list of the "Self Check" dialog and click "Test device".
3. In the "Firmware Update" tab of the next dialog, click "Select Firmware File" and
browse to the correct file.
4. Click "Start Firmware Upload".
The firmware is installed or updated automatically.
During this process, the power LED flashes green.
When the installation or update is completed, the base module restarts and is
ready for use.
3.5Putting into Operation
For using the base module and the extension module in the R&S TS8997 Regulatory
Test System for Wireless Devices, mount the switch units into the test system's 19''
rack. This procedure is described in the operating manual of the R&S OSP and in the
mounting instructions supplied with the R&S ZZA-211 rack adapter (order number
1096.3260.00).
22User Manual 1178.5795.02 ─ 03
Page 23
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Risk of injury when stacking instruments
A stack of instruments can tilt over and cause injury if not stacked correctly. Furthermore, the instruments at the bottom of the stack can be damaged due to the load
imposed by the instruments on top.
Observe the following instructions when stacking instruments:
●
Never stack more than three instruments. If you need to stack more than three
instruments, install them in a rack.
●
The overall load imposed on the lowest instrument must not exceed 500N.
●
It is best if all instruments have the same dimensions (width and length).
If you need to stack smaller instruments on the top, the overall load imposed on the
lowest instrument must not exceed 250 N.
●
If the instruments have foldable feet, fold them in completely.
Preparing for Use
Putting into Operation
Risk of injury if feet are folded out
The feet can fold in if they are not folded out completely or if the instrument is shifted,
which can cause damage or injury.
●
Fold the feet completely in or out to ensure stability of the instrument. Never shift
the instrument when the feet are folded out.
●
When the feet are folded out, do not work under the instrument or place anything
underneath.
●
The feet can break if they are overloaded. The overall load on the folded-out feet
must not exceed 100 N.
F
max
23User Manual 1178.5795.02 ─ 03
Page 24
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Risk of instrument damage due to insufficient airflow in a rack
If you mount several instruments in a rack, you need an efficient ventilation concept to
ensure that the instruments do not overheat. Insufficient airflow for a longer period can
disturb the operation and even cause damage.
For putting the modules into operation after they are mounted inside a switch unit and
optionally inside a rack, proceed as described in the operating manual of the
R&S OSP. This manual also contains valuable information on the switch unit's setup,
on remote operation, on working with the operating system, and much more.
Unlike all other modules for R&S OSP switch units, the R&S OSP-B157W8 and
R&S OSP-B157W8PLUS are not equipped with on-board memory to store the switching configuration of the module's signal paths. Therefore, the hosting switch unit cannot
recognize the module. Instead, the module's LAN interface must be connected to the
control computer and the switching configuration must be saved as a "Hardware
Setup" in the R&S WMS32 software. In this software, the switching configuration of the
R&S OSP-B157W8 and R&S OSP-B157W8PLUS is already implemented.
Preparing for Use
Putting into Operation
For operating the R&S OSP-B157W8 or R&S OSP-B157W8PLUS and the R&S OSPB157WX, see Chapter 6, "Operating the Modules", on page 34.
24User Manual 1178.5795.02 ─ 03
Page 25
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Use RF connector adapters as port savers
SMA and K-type RF connectors have a limited lifetime. It is typically limited to 500 connector mating cycles, due to normal wear. After this number of mating cycles, the connectors must be replaced.
●
The SMA connectors on the 8 DUT ports at the front panel of the R&S OSPB157W8 and R&S OSP-B157W8PLUS are soldered in. Therefore, their replacement is a costly and time-consuming procedure, which must be handled by
Rohde & Schwarz service.
To save the lifetime of these RF connectors, use the 2.92 mm K-type RF connector
adapters as port savers on the original connectors. A set of 8 adapters is included
in the delivery. Fix the adapters with a torque of 0.8 Nm to 1.1 Nm.
Leave the port saver adapters on the ports and connect your DUT channels to
these adapters, until the lifetime of the adapters is reached, due to wear. Then
replace the worn port saver adapters for new ones.
Preparing for Use
Putting into Operation
Figure 3-8: Port saver adapter
●
If any SMA or K-type connectors of the R&S OSP-B157WX are worn, replacing
these connectors is much more simple, because they are not soldered in.
Refer to the switch unit's operating manual for more operation information regarding,
for example, power supply, connection, grounding, ambient conditions, ventilation, system integration, switching on or reaching the currentless state.
25User Manual 1178.5795.02 ─ 03
Page 26
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
4Tour of the Base Module
The base module R&S OSP-B157W8 or R&S OSP-B157W8PLUS must be mounted in
an R&S OSP150 or R&S OSP120 switch unit.
The module is remotely operated via Ethernet (LAN), independent of the switch unit, by
the test software R&S WMS32, which must be installed on your control computer.
Operation is described in Chapter 6, "Operating the Modules", on page 34 and in the
operating manual of the R&S OSP switch units (see chapter on R&S EMC32 software).
On the front panel, the R&S OSP-B157W8 or R&S OSP-B157W8PLUS has the connectors for up to eight DUT channels and for the test instruments:
Less frequently used connectors are on the rear panel:
Figure 4-2: R&S
For an overview of the connectors, see Table 4-1.
As the LAN and CAN connectors on the rear panel of the switch unit are disabled, use
the RJ45 LAN connector of the module R&S OSP-B157W8 or R&S OSPB157W8PLUS itself (top center in Figure 4-2) for remote control.
The R&S OSP-B157W8 and R&S OSP-B157W8PLUS operate according to the following circuit diagram:
OSP-B157W8 mounted in R&S OSP150 (rear panel view)
26User Manual 1178.5795.02 ─ 03
Page 27
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Tour of the Base Module
Figure 4-3: Circuit diagram of the R&S OSP-B157W8 and R&S OSP-B157W8PLUS RF board
Orange, right= Front DUT and instrument ports
Orange, left= Instrument or trigger ports, and data acquisition board
Yellow= 1-to-2 switches
Red= Power detectors
Green= 1-to-n switches, attenuators or bi-directional couplers
Blue= Splitters/combiners
Pink= 1-to-2 switches or splitters/combiners in the 18 GHz by-path
Light green= Attenuators or splitters/combiners
(*), yellow highlighted = Step attenuator R2, only available in R&S OSP-B157W8PLUS, controlled together
with step attenuator R1033 in the companion path
With this RF circuitry, the R&S OSP-B157W8 or R&S OSP-B157W8PLUS can operate
flexibly in a test system in a multitude of ways.
●
Incoming signals on each DUT channel can be handled as follows:
–Signals can be individually attenuated.
–Signals can be switched to one or more instrument ports of the module.
–The power level can be measured by internal detectors.
●
Incoming signals on each instrument port can be handled as follows:
–Signals can be individually attenuated.
–Signals can be switched to one or more DUT ports of the module.
–The power level can be measured by internal detectors.
27User Manual 1178.5795.02 ─ 03
Page 28
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Connecting Devices to the Base Module
The exact behavior of the module is test-case-specific and depends on the applicable
test standard. This flexibility allows setting up and handling complex test and measurement sequences with the connected devices.
4.1Connecting Devices to the Base Module
The base module has various external connectors on its front and rear panel: SMA for
RF signals, BNC for analog and digital trigger control, D-Sub 9 for digital trigger control
and LAN for Ethernet access. The details of these connections are described in
Table 4-1.
Observe maximum input power and no DC voltage
All front connectors and some rear connectors have the following limitations:
●
A maximum tolerable input power level (dBm)
●
No tolerance for a direct current (DC) voltage share in the input signal
Tour of the Base Module
If the input signals exceed these limits, the internal circuits of the module can be damaged.
Do not exceed the dBm levels and "0 VDC" limits that are indicated on the front and
rear panels.
Depending on the switch unit that you use, the following additional connectors are
available:
●
If you use an R&S OSP150 switch unit, its rear panel has a power supply connector with fuse.
●
If you use an R&S OSP120 switch unit, its front panel has a DVI video monitor
interface and two USB 2.0 ports (all inactive in this setup). Its rear panel has a
power supply connector with fuse.
For information on the various connectors of the switch unit and for replacing the fuse,
refer to the operating manual of the R&S OSP.
Connect the power supply connector to an isolated ground receptacle of your mains
grid. The supply voltage must be between 110 VAC and 240 VAC with frequencies
ranging from 50 Hz to 60 Hz. Within these ranges, the device is automatically adapted
to the supply voltage. To turn the power on or off, put the AC power switch into position
I (on) or 0 (off).
The following table describes all other external connectors of the base module.
28User Manual 1178.5795.02 ─ 03
Page 29
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Connecting Devices to the Base Module
Table 4-1: External connectors overview of the R&S OSP-B157W8 or R&S OSP-B157W8PLUS
PanelLabelDirectionSignalTypeValue range
Tour of the Base Module
FrontDUT1 ...
DUT8
FrontCOMPBiDirCompanion device, max.
FrontX1BiDirCompanion device extension
FrontRxOut (**)Spectrum analyzer / test
RearTrig 4InTrigger input Ch4Digital BNCHigh: 3.3 V, low: 0 V
RearTrig 3OutTrigger output Ch3Digital BNCHigh: 3.3 V, low: 0 V
RearLANBiDirEthernetDigital RJ45
RearANTOutAntenna outputRF SMA1 GHz to 18 GHz
RearVIDEOInVideo input, max. 3 VAnalog SMADC to 1 MHz
RearGENInSignal generator, max.
RearVGENInVector signal generatorRF SMA1 GHz to 18 GHz
RearTrig 2OutTrigger output Ch2Analog BNCHigh: 3.3 V, low: 0 V
High: 3.3 V, low: 0 V
If you use the base module R&S OSP-B157W8 or R&S OSP-B157W8PLUS in combination with the extension module R&S OSP-B157WX, make the following changes to
the connections listed above:
●
(*)
Connect the ports of your DUT to the frequency extension instead of the base
module.
And connect the DUT ports of the base module to the [OUT1] to [OUT8] ports of
the frequency extension module, as in Figure 5-4.
●
(**)
Connect the signal analyzer or test receiver to the [Rx] port of the frequency extension instead of the base module.
And connect the [Rx] port of the base module to the [RxOUT] port of the frequency
extension module, as in Figure 5-4.
29User Manual 1178.5795.02 ─ 03
Page 30
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
5Tour of the Extension Module
The frequency extension module R&S OSP-B157WX must be mounted in an
R&S OSP120 switch unit.
This switch unit has front connectors for a USB keyboard, USB mouse and DVI display
for manual operation. However, we recommend connecting the switch unit via Ethernet
(LAN) to your control computer for remote operation by the software R&S OSP Panel.
Operation is described in Chapter 6, "Operating the Modules", on page 34 and in the
operating manual of the R&S OSP switch units.
On the front panel, the R&S OSP-B157WX has the connectors for up to eight DUT
channels with corresponding forwarding (loop-through) output ports and the connectors
for the test instruments:
The module has no rear connectors, but just blind plates. The only connectors on the
rear panel are the switch unit's power supply, LAN interface and CAN bus port.
Figure 5-2: R&S
Use the switch unit's LAN connector for remote control
In contrast with the base module, the frequency extension module has no LAN connector of its own.
For controlling the R&S OSP-B157WX, use the RJ45 LAN connector on the rear panel
of the switch unit.
Hence, for controlling a combination of both the R&S OSP-B157W8 or R&S OSPB157W8PLUS and R&S OSP-B157WX in an R&S TS8997 test system, use two LAN
cables to connect the modules to the test systems' Ethernet switch.
OSP-B157WX mounted in R&S OSP120 (rear panel view)
For an overview of the connectors, see Chapter 5.1.
30User Manual 1178.5795.02 ─ 03
Page 31
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Connecting Devices and a Base Module to the Extension Module
The R&S OSP-B157WX operates according to the following circuit diagram:
Figure 5-3: Circuit diagram of the R&S OSP-B157WX RF board
Orange, right= 8 DUT input and output ports (front connectors)
Orange, left= 3 instrument ports (also front connectors)
Blue squares= SPDT (1-to-2) relays
Blue upright rectangles = SP6T (1-to-6) relays
Blue horizontal rectangle = Directional coupler
Tour of the Extension Module
With this RF circuitry, the R&SOSP-B157WX is operated in the test system as follows:
●
Incoming signals on each DUT channel are forwarded to the connected DUT input
of the base module R&S OSP-B157W8 (supporting up to 6 GHz) respectively
R&S OSP-B157W8PLUS (supporting up to 7.5 GHz).
●
Incoming signals on one single DUT channel at frequencies above 6 GHz respectively above 7.5 GHz are switched to the module's [Rx] port.
Optionally, the control system can switch an [RxOUT] signal (instead of a DUT) or a
[X1] signal (companion device, in addition to a DUT) to the module's [Rx] port.
Signal paths for different frequency ranges
You must define the signal paths in such a way that the test system performs out-ofband measurements up to 40 GHz via the R&S OSP-B157WX.
5.1Connecting Devices and a Base Module to the Extension Module
We recommend placing the R&S OSP-B157WX on top of the base module or mounting the two modules in the same arrangement into a 19'' rack.
31User Manual 1178.5795.02 ─ 03
Page 32
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Connecting Devices and a Base Module to the Extension Module
Observe maximum input power and no DC voltage
The R&S OSP-B157WX typically forwards RF signals from a DUT to the R&S OSPB157W8 or R&S OSP-B157W8PLUS. Hence, if you connect the R&S OSP-B157WX to
the base module, be aware of the following limitations for the front ports:
●
A maximum tolerable input power level (dBm)
●
No tolerance for a direct current (DC) voltage share in the input signal
If the forwarded signals exceed these limits, the internal circuits of the base module
can be damaged.
Do not exceed the dBm levels and "0 VDC" limits that are indicated on the front (and
rear) panels of the R&S OSP-B157W8 or R&S OSP-B157W8PLUS.
Connecting the R&S OSP-B157WX
1. Connect the port [OUT1] of the R&S OSP-B157WX to the port [DUT1] of the
R&S OSP-B157W8 or R&S OSP-B157W8PLUS.
Tour of the Extension Module
2. Proceed in the same way for the remaining channels 2 through 8, as required in
your test setup.
3. Connect the port [DUT1] of the R&S OSP-B157WX to channel 1 of your DUT.
4. Proceed in the same way for the remaining channels 2 through 8, as required in
your test setup.
5. For your companion device (which is connected to the [COMP] port of the base
module), connect the port [X1] of the R&S OSP-B157WX to the port [X1] of the
R&S OSP-B157W8 or R&S OSP-B157W8PLUS.
6. Connect the port [RxOUT] of the R&S OSP-B157WX to the port [Rx] of the
R&S OSP-B157W8 or R&S OSP-B157W8PLUS.
Figure 5-4: Connected modules, upper = R&S
R&S OSP-B157W8PLUS
OSP-B157WX, lower = R&S OSP-B157W8 or
32User Manual 1178.5795.02 ─ 03
Page 33
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Connecting Devices and a Base Module to the Extension Module
7. Connect the port [Rx] of the R&S OSP-B157WX to the input port of your measure-
ment instrument, for example a spectrum analyzer or test receiver.
8. On the rear panel of the R&S OSP-B157WX, connect the switch unit's RJ45 [LAN]
port for remotely controlling the R&S OSP-B157WX as described in the operating
manual of the R&S OSP switch units. Note that this usage is different from the
base module, where you must use the module's LAN port instead of the switch
unit's LAN port. Do not use the CAN port for controlling the R&S OSP-B157WX.
9. Connect the power supply connector to an isolated ground receptacle of your
mains grid. The supply voltage must be between 110 VAC and 240 VAC with frequencies ranging from 50 Hz to 60 Hz. Within these ranges, the device is automatically adapted to the supply voltage. To turn the power on or off, put the AC power
switch into position I (on) or 0 (off).
Table 5-1: External connectors overview of the R&S OSP-B157WX
PanelLabelDirectionSignalTypeValue range
Tour of the Extension Module
FrontDUT1 ...
DUT8
FrontOUT1 ...
OUT8
FrontX1BiDirCompanion device (from the
FrontRxOutSpectrum analyzer / test
FrontRxOUTInSpectrum analyzer / test
FrontUSB (2x, only R&S
OSP120)
FrontMONITOR DIGITAL
(only R&S OSP120)
RearLANBiDirEthernetDigital RJ45
RearCAN
BiDirDevice under test, channel 1
to 8, max. 30 dBm
BiDirForwarding DUT1 to 8 to the
base module
base module)
receiver
receiver (from the base module)
disabled
disabled
disabled
RF K-Type1 GHz to 40 GHz
RF SMA1 GHz to 6 GHz
RF SMA1 GHz to 6 GHz
RF SMA1 GHz to 40 GHz
RF SMA1 GHz to 6 GHz
Note that "RxOUT" is - in spite of its name - an input port for signals coming from the
"Rx" port of the R&S OSP-B157W8 or R&S OSP-B157W8PLUS base module. The
label "RxOUT" was chosen due to the connection with the Rx output port of the base
module.
33User Manual 1178.5795.02 ─ 03
Page 34
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
6Operating the Modules
Observe the safety instructions in Chapter 3.5, "Putting into Operation",
on page 22.
Risk of instrument damage due to inappropriate operating conditions
An unsuitable operating site or test setup can damage the instrument and connected
devices. Before switching on the instrument, observe the information on appropriate
operating conditions provided in the data sheet. In particular, ensure the following:
●
All fan openings are unobstructed and the airflow perforations are unimpeded. A
minimum distance of 10 cm to other objects is recommended.
●
The instrument is dry and shows no sign of condensation.
●
The instrument is positioned as described in the following sections.
●
The ambient temperature does not exceed the range specified in the data sheet.
●
Signal levels at the input connectors are all within the specified ranges.
●
Signal outputs are connected correctly and are not overloaded.
Operating the Modules
Calibration
The base and extension modules mounted in the switch units (and optionally in a rack)
are remotely controlled via Ethernet (LAN).
●
The module R&S OSP-B157W8 or R&S OSP-B157W8PLUS with pre-configured
signal paths is operated in the test software R&S WMS32 (V10.60 or later).
●
The module R&S OSP-B157WX must be configured by you and connected as in
Figure 5-4 to be operated in R&S WMS32.
Manual operation of the modules or remote control operation via the R&S OSP Panel
software tool is not possible.
We recommend that you Use RF connector adapters as port savers for the DUT ports
of the module R&S OSP-B157W8 and R&S OSP-B157W8PLUS.
The modules R&S OSP-B157W8 or R&S OSP-B157W8PLUS and R&S OSP-B157WX
are delivered with a factory calibration, conforming to the standards of the German
national accreditation body. The calibration data is saved in the modules.
34User Manual 1178.5795.02 ─ 03
Page 35
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Configuring Devices and Signal Paths
If, instead of factory mounting, the modules are mounted by the Rohde & Schwarz service, the calibration is preserved.
For a later recalibration, including the power detectors in the R&S OSP-B157W8 or
R&S OSP-B157W8PLUS, we recommend that you send your switch unit (with the
module) to the Rohde & Schwarz service.
Alternatively, for your own module calibration, you must switch each path individually
and determine the path attenuation with your generator and measuring instrument.
However, this calibration does not include the power detectors in the R&S OSPB157W8 or R&S OSP-B157W8PLUS.
6.2Configuring Devices and Signal Paths
The base module R&S OSP-B157W8 or R&S OSP-B157W8PLUS and the frequency
extension module R&S OSP-B157WX are designed only for use in the R&S TS8997
regulatory conformance test system.
Operating the Modules
To configure the modules for this system, the R&S EMC32 software in the R&S RF
Test Suite offers a "WMS Configuration" wizard.
Insert the USB flash drive, which is included in the delivery, into a free USB port of
your computer.
35User Manual 1178.5795.02 ─ 03
Page 36
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Configuring Devices and Signal Paths
●
Have the address for each of your devices available. (The type of address depends
on the device's interface and its protocol, which can be IP, GPIB or USB / RS232.)
Obtaining valid IP addresses
If your network does not support DHCP or if you choose to disable dynamic TCP/IP
configuration, you must assign valid address information before connecting the instrument to the LAN. To obtain a valid IP address, contact your network administrator, as
connection errors can affect the entire network.
If required, see Chapter 10.2, "Resolving Network Connection Failures", on page 71.
In the wizard dialog, proceed as follows:
1. In the list of devices, enter an arbitrary ("speaking") name and the address for each
of your devices.
For example, if your spectrum analyzer has the IP address 10.113.0.104, enter
it as TCPIP::10.113.0.104::inst0::instr
2. For each device that is not present in your setup, select "<none>".
Operating the Modules
3. Check the checkbox "Import OSP-B157WX attenuation files".
4. Click "select source Folder".
5. Select the folder on the USB flash drive that contains the calibration data of your
R&S OSP-B157WX.
6. Confirm your selection to import the files.
7. In the field "Hardware Setup", enter a name for your hardware setup.
8. Check the checkboxes "Replace existing DeviceList entries" and "Replace existing
attenuation tables".
Note: We do not recommend using the same computer for a different test system,
too. If you let the wizard replace existing device entries and tables, these changes
can corrupt existing test templates required for another test system.
If you still wish to use the same computer for several test applications, proceed as
described in Chapter 6.2.1, "Creating Separate Profiles for Multiple Applications",
on page 37.
9. Click "Create".
The wizard creates the device entries in the device list, the required signal paths
and attenuation tables and a hardware setup.
10. When the wizard procedure is completed, restart your computer.
The path configuration is now stored and available for remote control operation.
The final preparatory steps are creating a DUT configuration and a test template as
described in the documentation (help function) of the R&S RF Test Suite software, sec-
tion "WMS Measurements (WMS32)".
36User Manual 1178.5795.02 ─ 03
Page 37
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Configuring Devices and Signal Paths
The R&S TS8997 can now be used with the R&S OSP-B157W8 or R&S OSPB157W8PLUS and R&S OSP-B157WX, see Chapter 6.3, "Operation Overview",
on page 39.
Use the R&S OSP Panel software to register the R&S OSP as described in the operat-
ing manual of the switch unit. See the following chapters:
●
"Connecting R&SOSP and PC"
●
"Configuring/Connecting R&S OSP Panel".
Once registered and connected, all modules mounted in the switch unit are displayed
in the following dialog:
Note that the R&S OSP-B157WX is listed twice in the R&S OSP Panel dialog, as the
module occupies two slots in the switch unit and uses two internal interfaces.
Also note that the base modules R&S OSP-B157W8 and R&S OSP-B157W8PLUS are
not recognized by the R&S OSP Panel software.
6.2.1Creating Separate Profiles for Multiple Applications
If you want to run more than one test application on the same computer, we recommend creating an additional profile for each application.
To do so, proceed as follows:
1. Go to the R&S EMC32 program icon in the Windows start menu.
2. Drag this program icon to your desktop while pressing the [Alt] key.
This creates a program link with a new icon on the desktop.
3. Right-click the new program link icon.
4. Select "Properties" in the context menu.
37User Manual 1178.5795.02 ─ 03
Page 38
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Configuring Devices and Signal Paths
5. In the dialog that comes up, select the "Shortcut" tab.
6. Enter a space and the name of the new profile (for example "Profile") as an
address parameter at the end of the "Target" field:
Operating the Modules
Figure 6-2: EMC32 Properties dialog
7. Confirm this setting with "OK".
When you start the software from this program link, the software runs under the new
profile.
Within this profile, you can create a "Device List" that is independent of device lists in
other profiles. You can also configure independent folder paths for each profile:
Figure 6-3: Defining folder paths for your new profile
Using individual profiles for different test applications, each, allows using the same
computer for several applications without interfering settings.
38User Manual 1178.5795.02 ─ 03
Page 39
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
6.3Operation Overview
The base modules R&S OSP-B157W8 and R&S OSP-B157W8PLUS and the frequency extension module R&S OSP-B157WX are designed only for use in the
R&S TS8997 regulatory conformance test system.
Operation of the modules is controlled remotely via LAN from the R&S WMS32 wireless measurement software, which is part of the R&S RF Test Suite. (This suite also
includes the software packages R&S EMC32 electromagnetic compatibility measurement and R&S AMS32 antenna and over-the-air (OTA) measurement.)
The following chapters provide more information for operating the modules.
●Operating the Base Module.................................................................................... 39
●Operating the Extension Module.............................................................................40
6.3.1Operating the Base Module
Use the R&S WMS32 software to control the base module R&S OSP-B157W8 or
R&S OSP-B157W8PLUS. Manual operation of the module or control via the R&S OSP
Panel software tool is not possible.
Operating the Modules
Operation Overview
R&S WMS32 uses the pre-configured signal paths in the device setup. This feature
allows calling signal paths directly from the user interface.
Figure 6-4: Dialog for base-module settings in the R&S RF Test Suite (V10.60 and later)
For more information on operation and remote control functions, refer to the help function of the software R&S WMS32, which is part of the R&S RF Test Suite.
39User Manual 1178.5795.02 ─ 03
Page 40
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
6.3.2Operating the Extension Module
Use the R&S WMS32 software to control the frequency extension module R&S OSPB157WX. Manual operation of the module or remote control via the R&S OSP Panel
software tool is not possible.
Path configuration is a precondition for operating the module. R&S WMS32 takes over
the path configuration into the device setup. This feature allows calling signal paths
directly from the user interface of R&S WMS32.
For more information on operation and remote control functions, refer to the help function of the software R&S WMS32, which is part of the R&S RF Test Suite.
6.4SCPI Commands
Optionally, you can use SCPI commands for direct communication with the modules,
as described in Chapter 8, "Remote Control Commands", on page 42.
Operating the Modules
SCPI Commands
40User Manual 1178.5795.02 ─ 03
Page 41
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
7Measurements
The modules R&S OSP-B157W8 or R&S OSP-B157W8PLUS and R&S OSP-B157WX
are dedicated for operation in the regulatory test system R&S TS8997. This system
investigates the conformance of short range devices (SRDs) to FCC and ETSI standards. The modules serve for switching the signal paths between a DUT and the
required test instruments. Also, the signals can be conditioned and their power levels
can be measured in the R&S OSP-B157W8 or R&S OSP-B157W8PLUS. This functionality is used in a broad range of tests.
Table 7-1: Typical test cases
Measurements
Test case2.4 GHz band
(ETSI EN 300 328)
Carrier frequency accuracy
RF output power
Transmit power control (TPC)
Spectrum power density
Duty cycle, TX sequence, TX gap
Dwell time, minimum frequency occupation, hopping sequence
(only for frequency hopping DUTs)
Hopping frequency separation
Medium utilization (MU) factor
Adaptivity
Occupied channel bandwidth
Transmitter unwanted emissions in the out-of-band domain
Transmitter unwanted emissions in the in-band domain
Transmitter unwanted emissions in the spurious domain
Receiver spurious emissions
••
••
•
•
•
•
••
••
•
••
•
•
5 GHz band
(ETSI EN 301 893)
•
•
• 1)
• 1)
Receiver blocking
Dynamic frequency selection (DFS)
••
•
1) For spurious emission measurements above 18 GHz, the R&S OSP-B157WX is
used to make the required direct RF connection to the spectrum analyzer.
These measurements and test cases, together with their configuration, are described
in the documentation (help function) of the R&S RF Test Suite software, section "WMS
Measurements (WMS32)".
41User Manual 1178.5795.02 ─ 03
Page 42
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
8Remote Control Commands
This chapter describes the remote control commands for direct communication with the
modules R&S OSP-B157W8 or R&S OSP-B157W8PLUS and R&S OSP-B157WX via
LAN interface and SCPI commands.
SCPI commands
Operation of the R&S OSP does not require that you send any SCPI commands. This
function is an additional feature for test and debug purposes. Using SCPI commands
requires knowledge of the R&S OSP remote control commands. For details, refer to
the operating manual of the R&S OSP.
For normal operation of the modules, use the graphical user interface of the
R&S WMS32 software.
Service-related RC commands are only available to Rohde & Schwarz personnel.
These commands are described in the service manual (1178.6504.02) of the
R&S OSP-B157W8, R&S OSP-B157W8PLUS and R&S OSP-B157WX.
Remote Control Commands
For direct SCPI communication with the modules, proceed as follows:
●
If you want to communicate with the module R&S OSP-B157W8 or R&S OSP-
B157W8PLUS:
1. Connect the module's LAN connector (Figure 4-2) to your control computer or to
your local area network.
2. For SCPI command exchange, use one of the following tools:
●The scripting tool R&S Forum
(For getting started, select "Settings" > "Instruments" > "Add", select the interface type "VXI11" and enter the TCPIP address of your module. Board number
and instance are typically "0". See Chapter 6.2, "Configuring Devices and Sig-
nal Paths", on page 35)
●The Measurement & Automation eXplorer "LabVIEW MAX" from National
Instruments (NI)
●A terminal emulator such as PuTTY
(However, this only permits sending individual SCPI command lines, no
scripts.)
●Or a similar software tool
Example:
For example, to activate the path from port1 (DUT1) to the analyzer port (Rx) in the
module R&S OSP-B157WX, send the following SCPI command to the R&S OSP:
:CONFigure:PORT1:MODE ANAlyze
●
If you want to communicate with the module R&S OSP-B157WX:
1. Connect the switch unit's LAN connector to your control computer.
42User Manual 1178.5795.02 ─ 03
Page 43
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
2. Start the R&S OSP Panel software.
3. Select "Utility" > "Read/Write…".
4. In the dialog that appears, enter your SCPI command in the upper text field.
5. If the command is a setting, click the "Write" button.
6. If the command is a query, click the "Read" button.
For the R&S OSP-B157WX, only use SCPI commands from the operating manual of
the R&S OSP.
The SCPI commands available for the R&S OSP-B157W8 or R&S OSP-B157W8PLUS (not R&S OSP-B157WX) are described in the following chapters.
Individual status query, returns the contents of the IST flag in decimal form. The IST
flag is the status bit that is sent during a parallel poll.
Parameters:
<ISTflag>0 | 1
Rohde&Schwarz,OSP,1527.1144.02/200153,10.60.0
Usage: Query only
44User Manual 1178.5795.02 ─ 03
Page 45
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
*OPC
Operation complete, sets bit 0 in the event status register, when all preceding commands have been executed. This bit can be used to initiate a service request.
The query writes a "1" into the output buffer, when all preceding commands have been
executed. This bit is used for command synchronization.
*OPT?
Option identification, queries the options included in the instrument. For a list of all
available options and their description, refer to the data sheet.
Parameters:
<Options>The query returns a list of options. The options are listed at fixed
positions in a comma-separated string. A zero is returned for
options that are not installed.
Usage: Query only
Remote Control Commands
Common Commands
*PRE <Value>
Parallel poll register enable, sets the PRE register to the indicated value. The query
returns the contents of the PRE register in decimal form.
Parameters:
<Value>Range: 0 to 255
*PSC <Action>
Power on status clear, determines whether the contents of the ENABle registers are
preserved or reset when the instrument is switched on. Thus, a service request can be
triggered when the instrument is switched on, if the status registers ESE and SRE are
suitably configured. The query reads out the contents of the "power-on-status-clear"
flag.
Parameters:
<Action>0 | 1
0
The contents of the status registers are preserved.
1
Resets the status registers.
*RCL <Number>
Recall, loads the instrument settings from an intermediate memory.
The settings can be stored to this memory using the command *SAV with the associated number.
45User Manual 1178.5795.02 ─ 03
Page 46
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Parameters:
<Number>Identifies the intermediate memory.
*RST
Reset, sets the instrument to a defined default status. Same as SYST:PRES.
Usage: Setting only
*SAV <Number>
Save, stores the current instrument settings in an intermediate memory.
The settings can be recalled using the command *RCL with the associated number.
Parameters:
<Number>Specifies the number under which the settings are stored in the
intermediate memory.
Remote Control Commands
Common Commands
*SRE <Contents>
Service request enable, sets the SRE register to the indicated value. This command
determines, under which conditions a service request is triggered.
Parameters:
<Contents>Contents of the SRE register in decimal form. Bit 6 (MSS mask
bit) is always 0.
Range: 0 to 255
*STB?
Status byte query, reads the contents of the status byte in decimal form.
Usage:
*TRG
Trigger, generates an immediate manual trigger signal and triggers all actions that are
waiting for a trigger event.
This common command is equivalent to INITiate:IMMediate and complements the
commands of the TRIGger subsystem.
Query only
Usage:
Event
46User Manual 1178.5795.02 ─ 03
Page 47
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
*TST?
Self-test query, initiates a self-test of the instrument and returns an error code.
Parameters:
<ErrorCode>0
No errors occurred.
1
An error occurred.
Usage: Query only
*WAI
Wait to continue, prevents servicing of the subsequent commands until all preceding
commands have been executed and all signals have settled. See also *OPC / command synchronization.
Sets the signal path from the selected port to the attenuator.
Remote Control Commands
Configuration Setting Commands
Suffix:
{n}
.
1 to 8
Sets the port number.
If you do not specify the port, the software automatically
addresses port 1.
Parameters:
<numeric_value>Attenuation level
Range: 0 to 31.75
Default unit: dB
CONFigure:PORT{n}:MODE <portmode>
Switches the signal path from the selected port to the analyzer.
Suffix:
{n}
.
1 to 8
Sets the port number.
If you do not specify the port, the software automatically
addresses port 1.
Parameters:
<portmode>ADI
Switches the direct 18 GHz signal path to the analyzer.
ANA
Switches the 6 GHz signal path to the analyzer.
PDI
Switches the direct signal path (with 3 dB switch) to the power
detector.
SUM
Switches the sum signal path for all selected ports together.
To select the ports for the sum signal measurement, you must
send the command again for each port that should be added to
the SUM path.
CONFigure:GENSignal:MODE <genmode>
Sets the signal path from the generator to the selected output.
Parameters:
<genmode>ABY
Signal path from the signal generator to the analyzer bypass.
50User Manual 1178.5795.02 ─ 03
Page 51
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
ANA
Signal path from the signal generator to the analyzer.
ANT
Signal path from the signal generator to the antenna.
SELF
Signal path from the signal generator to the self-test.
CONFigure:GENerator ON | OFF
Enables or disables the signal generator input, or queries the signal generator input
state.
CONFigure:VGENerator ON | OFF
Enables or disables the vector signal generator input, or queries the signal generator
input state.
Remote Control Commands
Configuration Setting Commands
CONFigure:BYPass:MODE <bypassmode>
Configures the dynamic frequency selection (DFS) interferer bypass path for the analyzer or the signal generator.
Parameters:
<bypassmode>ANA
Sets the DFS interferer bypass path for the analyzer.
GEN
Sets the DFS interferer bypass path for the signal generator.
OFF
Disables the DFS interferer bypass path.
CONFigure:BYPass:ATTenuator <attenuation_level>
Sets the attenuation level of the the attenuator in the Dynamic Frequency Selection
(DFS) interferer bypass path.
Parameters:
<attenuation_level>Range: 0 to 31.75
Default unit: dB
CONFigure:EXTCompanion <extcompmode>
Configures the path for the companion device.
51User Manual 1178.5795.02 ─ 03
Page 52
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
schaltet entweder das companion device auf Port 4 für geringe Dämpfung zwischen
DUT und Companion (wird für non LBT devices benötigt) oder auf X1 falls die WX verwendet wird, oder aus falls er nicht benötigt wird. In Stellung aus ist der Port abgeschlossen und terminiert.
Parameters:
<extcompmode>DUT4
Switches the path from the companion device ([COMP]) port to
the [DUT4] port. This setting enables low attenuation between
DUT and companion device, for example required for non-LBT
devices.
EXT
Switches the path from the [COMP] port to the [X1] port. This
setting is required if you use the extension module R&S OSPB157WX.
OFF
Disables any path from the [COMP] port, if it is not required. In
this setting, the port is closed and terminated.
Remote Control Commands
Configuration Setting Commands
CONFigure:COMPanion:MODE <compmode>
Sets the mode of the companion device.
Parameters:
<compmode>ON
Enables normal operation of the companion device.
SELF
Enables the self-test mode for the companion device.
Sets the attenuation level of the the attenuator in the companion device's signal path.
The available range of attenuation levels depends on the base unit version (.02 or .05).
Parameters:
<attenuation_level>Range: 0 to 31.75 for the R&S OSP-B157W8 (version .02),
0 to 63.50 for the R&S OSP-B157W8PLUS (.05)
Default unit: dB
CONFigure:CLOCk <clockselect>
Selects the clock source.
Parameters:
<clockselect>C250hz
Activates the internal 250 MHz clock.
C252mhz
Activates the internal 252 MHz clock.
52User Manual 1178.5795.02 ─ 03
Page 53
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
CEXTernal
Enables using an external clock.
CONFigure:TRIGger:INPut{m}?
Returns the state of the trigger input on the selected trigger channel.
Remote Control Commands
Measurement Setting Commands
Suffix:
{m}
.
1 to 4
Selects the channel number of the input trigger.
Example: If the query CONF:TRIG:INP1? returns "0", the trigger has not
been released and the input level at trigger port 1 is 0 V (or at
least <2.5 V). If the query returns "1", the trigger input level is
>2.5 V.
Usage: Query only
CONFigure:TRIGger:OUTPut{m}
<mode>
Sets the trigger output on the selected trigger channel to the output level of 3.3 V. If the
trigger output command is not sent, the output level is 0 V.
The query returns the trigger output state.
Suffix:
{m}
.
1 to 3
Selects the channel number of the output trigger.
Parameters:
<mode>1 or ON
Enables the trigger (turns it on, until parameter 0 or OFF is
sent).
0 or OFF
Disables the trigger (turns it off, until parameter 1 or ON is sent).
8.4Measurement Setting Commands
In most measurement settings commands, you can optionally select the sensor channel (channels 1 through 8) by adding, for example, SENSe5: in front of your command.
If you do not specify the sensor channel, the software automatically addresses sensor
channel 1.
Configures the power meter for a sum-power measurement.
OFF
Disables the sensor. To activate it again, send either the SINGle
or the SUM command.
[SENSe{1..8}:][SAMPling:]TIME? <time>
Specifies the duration of the measurement for the selected sensor.
54User Manual 1178.5795.02 ─ 03
Page 55
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Parameters:
<time>Range: 1.0e-3 to 100.0
*RST: 1
Default unit: s
Usage: Query only
[SENSe{1..8}:][SAMPling:]RATE? <sampling_rate>
Specifies the sampling rate for the selected sensor.
Parameters:
<sampling_rate>Range: 10e6 to 10e7
*RST: 10e6
Usage: Query only
Remote Control Commands
Measurement Setting Commands
[SENSe{1..8}:][SAMPling:]DIVFrequency
Specifies the divider value for setting the sampling rate of the selected sensor. The
sampling rate is the clock frequency divided by this value.
Parameters:
<sampling_div>Range: 25 to 256
*RST: 250
[SENSe{1..8}:][SAMPling:]DIVStorage <storage_div>
Specifies the divider interval for setting the storage data reduction of the selected sensor.
For example, a divider interval of 100 determines, that only 1% of the original sampling
data is saved.
Depending on the [SENSe{1..8}:][SAMPling:]MODE setting, the MEAN, MAXI-MUM or LAST SAMPLE out of the 100 original samples is stored.
Parameters:
<storage_div>Range: 0 to 65355
*RST: 0
<sampling_div>
[SENSe{1..8}:][SAMPling:]MODE <sampmode>
For the selected sensor, the storage mode specifies an operator for transforming a set
of samples from one divider interval into a single value.
Parameters:
<sampmode>SAMPle
Stores the value of the current sample at saving time, hence the
last value in a divider interval.
55User Manual 1178.5795.02 ─ 03
Page 56
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
MAX
Stores the maximum value of all samples between two saving
times (during the divider interval).
MEAN
Stores the arithmetic mean value of all samples between two
saving times (during the divider interval).
[SENSe{1..8}:][SAMPling:]COUNt <count>
Specifies the number of samples to be acquired by the selected sensor. The acquisition is stopped when this number is reached.
Parameters:
<count>Range: 0 to 100,000,000
*RST: 0
[SENSe{1..8}:][SAMPling:]PRECount <precount>
Remote Control Commands
Measurement Setting Commands
The precount is a negative trigger offset that specifies the required number of samples
in the selected sensor, before a trigger event is accepted. Hence, this number of samples must be there, otherwise a trigger cannot be released.
Parameters:
<precount>Range: 0 to 100,000,000
*RST: 0
[SENSe{1..8}:][SAMPling:]OPERation <opermode>
Specifies the mode of operation for the selected sensor.
Parameters:
<opermode>ONCE
Stops the acquisition after a single measurement.
LOOP
Continues acquiring data and writing it into a ring buffer, until the
measurement is stopped by ABORt.
[SENSe{1..8}:]TRIGger{1..4}:SET <trigset>
Sets the response to a trigger input for starting to sample the selected sensor.
56User Manual 1178.5795.02 ─ 03
Page 57
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Remote Control Commands
Measurement Setting Commands
Suffix:
{1..4}
Parameters:
<trigset>This parameter set is a group of three boolean decisions, for
Example:
.
Selects the trigger input port, see Figure 4-2. If not selected, the
default port is trigger 1.
example "1,1,0" or "ON,OFF,ON". The first boolean decision
switches the first parameter "ENABLE" on or off, the second
decision switches "NTR" on or off, the third decision switches
"PTR" on or off. These three parameters are described below.
ENABLE
Enables starting to sample at either a positive or a negative transition of the trigger input signal (or none of them, or both),
depending on the boolean decision for "NTR" and "PTR".
NTR
Enables starting to sample at a neagitve transition of the trigger
input signal.
PTR
Enables starting to sample at a positve transition of the trigger
input signal.
The command SENS:TRIG:SET ON, OFF, ON defines to start
sampling the default sensor channel 1, when a transition with a
positive slope is detected (edge trigger) on the default input trigger port 1.
[SENSe{1..8}:]STATus?
Returns the sampling status message of the selected sensor.
Parameters:
<status_message>ERRor
The sensor is in error mode.
IDLE
The sensor is idle and ready for the next measurement.
WAITing
The sensor is on hold, waiting for an INIT command or a trigger.
Sets the threshold level below peak (not absolute) for detecting a burst in the selected
sensor.
Query only
57User Manual 1178.5795.02 ─ 03
Page 58
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
(For the sum-power detector, see SUMPowerdetector:BURSt:SEARchlevel.)
Parameters:
<searchlevel>Specifies the numeric value of the threshold level.
[W | DBM]Specifies the unit for the threshold level: "W" = watts, "DBM" =
dB.
We recommend using "DBM", which sets the threshold level in
relation (dB) to the highest value in the measurement.
Range: -200 to 200
*RST: 30
Default unit: dB
Remote Control Commands
Measurement Setting Commands
[SENSe{1..8}:]CORRection:ATTenuation
Sets the attenuation value for a correction of the selected sensor.
Parameters:
<attenuation>Range: 0 to 31.75
*RST: 0
Default unit: dB
[SENSe{1..8}:]CORRection:BFGain <bfgain>
Sets the beamforming (BF) gain value for a correction of the selected sensor.
While the gain is defined in dBi, the software uses dB for calculation.
While antenna gain takes effect for a single port, only, beamforming gain is relevant for
all ports.
Parameters:
<bfgain>Range: -200 to 200
*RST: 0
Default unit: dB
[SENSe{1..8}:]CORRection:GAIN <gain>
<attenuation>
Sets the gain value for a correction of the selected sensor.
While the gain is defined in dBi, the software uses dB for calculation.
Parameters:
<gain>Range: -200 to 200
*RST: 0
Default unit: dB
[SENSe{1..8}:]FREQuency <frequency>
Specifies the DUT frequency, hence, the frequency of the measured input signal, for
the selected sensor.
58User Manual 1178.5795.02 ─ 03
Page 59
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Parameters:
<frequency>Range: 0.0 to 110.0e9
*RST: 1.0e9
Default unit: Hz
Inspite of the available range setting (0 Hz to 110 GHz), the calibration effectively limits
the range to within 120 MHz to 6 GHz.
[SENSe{1..8}:]VIDeo:MIN
Sets the currently measured video input signal value of the selected sensor as the minimum value. For the video port, see Figure 4-2.
[SENSe{1..8}:]VIDeo:MAX
Sets the currently measured video input signal value of the selected sensor as the
maximum value. For the video port, see Figure 4-2.
Remote Control Commands
Measurement Commands
8.5Measurement Commands
In most measurement commands, you can optionally select the sensor channel (channels 1 through 8) by adding, for example, SENSe3: in front of your command. If you do
not specify the sensor channel, the software automatically addresses sensor channel 1.
Aborts the current measurement and returns to the passive "idle" state or to the "wait
for trigger" state (trigger engaged, but not released).
Usage:
Setting only
59User Manual 1178.5795.02 ─ 03
Page 60
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
INITiate[:IMMediate]
Starts a measurement for all sensors that are activated by [SENS{1..8}:]MEAS.
With the optional addition of :IMM (equivalent to *TRG), the measurement starts immediately.
Without :IMM, the measurement starts upon a trigger event, if a trigger is activated by
[SENS{1..8}:]TRIG{1..4}:SET.
Usage: Setting only
[SENSe{1..8}:]BURSt:COUNt?
Returns the number of detected bursts.
Remote Control Commands
Measurement Commands
Usage:
[SENSe{1..8}:]BURSt:LEVel?
Returns an array with the detected time-domain power (TDP) levels of each burst
(hence, one value per burst, while typically there may be hundreds or thousands of
bursts in one measurement).
Usage:
[SENSe{1..8}:]BURSt:MAX?
Returns the time-domain power (TDP) levels of the burst that has the maximum level
within a measurement.
Usage:
[SENSe{1..8}:]BURSt:STARttime?
Returns an array with the start times of all detected bursts within a measurement.
Usage:
Query only
Query only
Query only
Query only
[SENSe{1..8}:]BURSt:ONTime?
Returns an array with the "burst on" times of all detected bursts within a measurement.
Together with the burst start times, the "burst on" times allow reconstructing the
envelope curve without overloading the software interface with millions of points.
Usage:
[SENSe{1..8}:]DCYCle?
Returns the duty cycle, calculated over all bursts within the measurement.
Query only
60User Manual 1178.5795.02 ─ 03
Page 61
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Note: Since the DUTs are "load based", hence the individual durations of the bursts
depend on the actual data transfer, calculating the duty cycle requires a statistical evaluation of a number of bursts that is sufficient to do statistics.
Usage: Query only
[SENSe{1..8}:]MU?
Returns the medium utilization (MU) factor, calculated from the measured values.
MU = duty cycle * maximum power / measured power
Usage: Query only
[SENSe{1..8}:]RAW:DATA?
Returns the measured raw data of all recorded samples.
Remote Control Commands
Measurement Commands
Usage:
[SENSe{1..8}:]RAW:MIN?
Returns the minimum level in the raw data samples (values from the ADCs).
Usage:
[SENSe{1..8}:]RAW:MAX?
Returns the maxmum level in the raw data samples (values from the ADCs).
Usage:
[SENSe{1..8}:]RMS?
Returns the RMS value (root mean squared) over all samples in the measurement,
without burst evaluation.
Usage:
[SENSe{1..8}:]TRACe:DATA?
Query only
Query only
Query only
Query only
Returns all level result data of the recorded trace.
The values are returned in dBm, hence with correction factors (as opposed to the data
in RAW:DATA).
Usage:
[SENSe{1..8}:]TRACe:MIN?
Returns the minimum level result of the recorded trace.
Query only
61User Manual 1178.5795.02 ─ 03
Page 62
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
The values are returned in dBm, hence with correction factors (as opposed to the data
in RAW:MIN).
Usage: Query only
[SENSe{1..8}:]TRACe:MAX?
Returns the maximum level result of the recorded trace.
The values are returned in dBm, hence with correction factors (as opposed to the data
in RAW:MAX).
Starts a sum-power measurement for all detector channels that are activated by
[SENS{1..8}:]MEAS.
To select the channels, the command needs to be resent for each channel that should
be added to the SUM measurement.
With the optional addition of :IMM, the measurement starts immediately.
Without :IMM, the measurement starts upon a trigger event, if a trigger is activated by
[SENS{1..8}:]TRIG{1..4}:SET.
Usage:
Setting only
62User Manual 1178.5795.02 ─ 03
Page 63
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
SUMPowerdetector:STATus?
Returns the status message of the sum-power sensor sampling.
Parameters:
<status_message>ERRor
The sum-power sensor is in error mode.
IDLE
The sum-power sensor is idle and ready for the next measurement.
WAITing
The sum-power sensor is on hold, waiting for an SUMP:INIT
command.
MEASure
The sum-power sensor is currently measuring.
READy
The sum-power sensor has completed a measurement.
Remote Control Commands
Sum-Power Detector Commands
Usage:
SUMPowerdetector:TRACe:DATA?
Returns all level result data of the recorded trace.
The values are returned in dBm, hence with correction factors.
Usage:
SUMPowerdetector:TRACe:MAX?
Returns the maximum level result of the recorded trace.
The value is returned in dBm, hence with correction factors.
Usage:
SUMPowerdetector:TRACe:MIN?
Returns the minimum level result of the recorded trace.
The value is returned in dBm, hence with correction factors.
Usage:
Query only
Query only
Query only
Query only
SUMPowerdetector:TRACe:RMS?
Returns the RMS value (root mean squared) of all level results of the recorded trace,
without burst evaluation.
The value is returned in dBm, hence with correction factors.
Usage:
Query only
63User Manual 1178.5795.02 ─ 03
Page 64
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
SUMPowerdetector:BURSt:COUNt?
Returns the number of detected bursts.
Usage: Query only
SUMPowerdetector:BURSt:LEVel?
Returns an array with the detected time-domain power (TDP) levels of each burst
(hence, one value per burst, while typically there may be hundreds or thousands of
bursts in one measurement).
Usage: Query only
SUMPowerdetector:BURSt:MAX?
Returns the time-domain power (TDP) levels of the burst that has the maximum level
within a sum-power measurement.
Remote Control Commands
Sum-Power Detector Commands
Usage:
SUMPowerdetector:BURSt:STARttime?
Returns an array with the start times of all detected bursts within a sum-power measurement.
Usage:
SUMPowerdetector:BURSt:ONTime?
Returns an array with the "burst on" times of all detected bursts within a sum-power
measurement.
Together with the burst start times, the "burst on" times allow reconstructing the
envelope curve without overloading the software interface with millions of points.
Sets the threshold level below peak (not absolute) for detecting a burst in the sumpower detector.
Query only
Query only
Query only
Parameters:
<searchlevel>Specifies the numeric level value.
64User Manual 1178.5795.02 ─ 03
Page 65
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
[W | DBM]Specifies the unit for the threshold level: "W" = watts, "DBM" =
dB.
We recommend using "DBM", which sets the threshold level in
relation (dB) to the highest value in the sum-power measurement.
Range: -200 to 200
*RST: 30
Default unit: dB
SUMPowerdetector:DCYCle?
Returns the duty cycle, calculated over all bursts within the sum-power measurement.
Note: Since the DUTs are "load based", hence the individual durations of the bursts
depend on the actual data transfer, calculating the duty cycle requires a statistical evaluation of a number of bursts that is sufficient to do statistics.
Remote Control Commands
List of Commands
Usage:
Query only
SUMPowerdetector:MU?
Returns the medium utilization (MU) factor, calculated from the measured values.
The instrument does not need periodic maintenance. Only cleaning the instrument is
recommended.
When ordering spare parts, follow the instructions in the Service Manual and the safety
instructions. Find order numbers for spare parts in the service manual, which also
includes further information on troubleshooting and repair.
Mounting the modules into a switch unit is described in Chapter 3, "Preparing for Use",
on page 12.
Due to normal wear, the SMA and K-type RF connectors of the modules have a limited
lifetime of typically 500 connector mating cycles. If these connectors are worn, replace
them.
For the hard-to-replace DUT ports of the base module R&S OSP-B157W8 or
R&S OSP-B157W8PLUS, we recommend that you Use RF connector adapters as port
savers. Once the port savers are worn, replace them with new connector adapters.
Maintenance
9.1Replacing the Power Supply Fuses
The R&S OSP is protected by two fuses, located in the fuse holder below the power
switch on the rear panel.
Use only fuses of the type IEC 60127-2 T4.0A/250V (order no. 0020.7600.00) for all
nominal AC supply voltages specified in the data sheet.
Risk of electric shock
The fuse is part of the main power supply. Therefore, handling the fuse while power is
on can lead to electric shock. Before opening the fuse holder, make sure that the
instrument is switched off and disconnected from all power supplies.
Always use fuses supplied by Rohde & Schwarz as spare parts, or fuses of the same
type and rating.
To replace the fuses
1. Switch off the instrument.
2. Disconnect the power cable.
3. Open the lid of the AC power connector, using, for example, a small screwdriver.
4. Lift the fuse holder out of its slot.
5. Remove both fuses.
68User Manual 1178.5795.02 ─ 03
Page 69
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
6. Insert the new fuses.
7. Push the fuse holder back into its slot.
8. Close the lid.
9. Reconnect the power cable.
9.2Cleaning
Risk of electric shock
If moisture enters the casing, for example if you clean the instrument using a moist
cloth, contact with the instrument can lead to electric shock. Before cleaning the instrument other than with a dry cloth, make sure that the instrument is switched off and disconnected from all power supplies.
Maintenance
Cleaning
Instrument damage caused by cleaning agents
Cleaning agents contain substances such as solvents (thinners, acetone, etc.), acids,
bases, or other substances. Solvents can damage the front panel labeling, plastic
parts, or screens, for example.
Never use cleaning agents to clean the outside of the instrument. Use a soft, dry, lintfree dust cloth instead.
Risk of instrument damage due to obstructed fans
If the instrument is operated in dusty areas, the fans become obstructed by dust or
other particles over time. Check and clean the fans regularly to ensure that they always
operate properly. If the instrument is run with obstructed fans for a longer period, the
instrument overheats, which can disturb the operation and even cause damage.
1. Clean the outside of the instrument using a soft, dry, lint-free dust cloth.
2. Check and clean the fans regularly to ensure that they always operate properly.
3. Clean the touchscreen as follows:
a) Apply a small amount of standard screen cleaner to a soft cloth.
b) Wipe the screen gently with the moist, but not wet, cloth.
c) If necessary, remove any excess moisture with a dry, soft cloth.
69User Manual 1178.5795.02 ─ 03
Page 70
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
9.3Storing and Packing
The storage temperature range of the instrument is given in the data sheet. If the
instrument is to be stored for a longer period of time, it must be protected against dust.
Repack the instrument as it was originally packed when transporting or shipping. The
two protective foam plastic parts prevent the control elements and connectors from
being damaged. The antistatic packing foil avoids any undesired electrostatic charging
to occur.
If you do not use the original packaging, use a sturdy cardboard box of suitable size
and provide for sufficient padding to prevent the instrument from slipping inside the
package. Wrap antistatic packing foil around the instrument to protect it from electrostatic charging.
Maintenance
Storing and Packing
70User Manual 1178.5795.02 ─ 03
Page 71
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Resolving Network Connection Failures
10Troubleshooting
10.1Status LEDs
The front panel of the base module has 2 LEDs. labeled "Power" and "Ready".
The "Power" LED can show the following states:
●
Green: normal operation
●
Flashing green: busy (measurement or update)
●
Red: error, proceed as follows:
–First try a restart.
–If the LED is still red, install a firmware update.
–If the LED is still red, contact Rohde & Schwarz service.
Troubleshooting
The "Ready" LED can show the following states:
●
Yellow: trigger is set
●
Green: trigger is released
10.2Resolving Network Connection Failures
Risk of network connection failure
Network cables and cable connectors of poor quality, or failures in the autonegotiation
process, can cause network connection failures.
If the network connection to the instrument fails, check the network infrastructure and
contact your network administrator.
Several issues may cause failures in the network connection to the instrument. This
section lists the most likely reasons and the recommended solutions.
Common reasons for network connection failures
●
Network connecting cables and cable connectors of poor quality
●
Incompatibility between the network interface of the base module and certain
switches or routers available on the market
●
An invalid IP address assigned to the instrument
71User Manual 1178.5795.02 ─ 03
Page 72
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Resolving Network Connection Failures
Risk of network failure
Consult your network administrator before performing the following tasks:
●
Connecting the instrument to the network
●
Configuring the network
●
Changing IP addresses
●
Exchanging hardware
Errors can affect the entire network.
Possible solutions
●
Check the network infrastructure. Exchange connecting cables if obvious damage
is visible.
●
Observe the link status LED on the connected network device. The link status LED
is located next to the LAN connector. If a link failure is detected, connect the instrument to a different device port or to a different network device.
●
Check whether the LAN interface and the required LAN services are enabled.
●
Check whether the IP address of the instrument is within the network's address
range. IP addresses that are set manually may be invalid.
Troubleshooting
72User Manual 1178.5795.02 ─ 03
Page 73
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
Glossary: Abbreviations
Symbols
3GPP: 3rd Generation Partnership Project, a telecommunication industry consortium
B
BER: Bit error rate, the number of bit errors divided by the total number of transferred
bits
C
C: Common terminal of a relay (see NC and NO)
CTIA: A nonprofit wireless communications industry organization in the United States,
formerly known as Cellular Telecommunications Industry Association
D
dB: Decibel, 10 times the common (decadic) logarithm of the ratio of two power quanti-
ties, or of the ratio of the squares of two field amplitude quantities
Glossary: Abbreviations
dBm: Decibel milliwatt, the power level based on a power ratio in decibels (dB) of the
measured power, referenced to 1 mW
DFS: Dynamic frequency selection, automatic selection of wireless network frequency
channels with low interference, for example in WLAN access points
DHCP: Dynamic host configuration protocol
DL: Downlink, typically the transmission path from a base station to a mobile phone
DUT: Device under test, for example a mobile communication device in a development
or production environment
Duty cycle: The fraction of time in which a signal is active within a total signal period
time
E
EIRP: Equivalent isotropically radiated power
EMC: Electromagnetic compatibility (with both EMI and EMS requirements)
EMI: Electromagnetic interference: Ability of an electrical appliance to avoid disturbing
its environment by emitting an interfering signal
EMS: Electromagnetic susceptibility: Property of an electrical appliance to tolerate a
disturbance of a particular level without showing any faults
73User Manual 1178.5795.02 ─ 03
Page 74
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
ERP: Effective radiated power
F
FCC: Federal Communications Commission, a US government agency for telecommu-
nications regulation
L
LAN: Local area network within a limited space, such as an office building. LAN con-
nects computers or electronic equipment with processing capability, using network
media technologies such as Ethernet (over electric or fiber-optic cables) or wireless
LAN (Wi-Fi).
LBT: Listen before talk, a scheme in which a device or system checks for eventual
presence of other signals in its transmission channel, before starting to transmit
LTE: Long Term Evolution, a mobile communication technology standard
Glossary: Abbreviations
M
MIMO: Multiple-input and multiple-output, relating to wireless communication technolo-
gies that use several transmitting and receiving antennas for enhancing a radio link
N
NC: Normally closed position of a relay terminal. When the relay is not activated, termi-
nal NC is connected to the common terminal C.
NO: Normally open position of a relay terminal. When the relay is activated, terminal
NO is connected to the common terminal C.
O
OTA: Over-the-air (tests)
P
PER: Packet error rate
R
RF: Radio frequency, electromagnetic oscillation in the frequency range of around
3 kHz to 300 GHz
RMS: Root mean square, a statistical measure defined as the square root of the arithmetic mean of the squares of the original values
Rx: Receive direction, received signals, reception
S
SCPI: Standard commands for programmable instruments
74User Manual 1178.5795.02 ─ 03
Page 75
R&S®OSP‑B157W8 R&S®OSP‑B157W8PLUS R&S®OSP‑B157WX
SMA connector: Sub-Miniature Assembly, a coaxial RF connector, version A (stan-
dard)
SNR: Signal-to-noise ratio is the ratio of the level of a desired signal to the level of
background noise (measured in quantities of power). SNR > 1 (or SNR > 0 dB) indicates more signal than noise.
SRD: Short range device, typically with an effective radiated power level below 20 dBm
(100 mW). Examples are near-field identification devices (RFID), car keys, or equipment that uses Wi-Fi or Bluetooth for communication across a distance of a few
meters.
T
TIS: Total isotropic sensitivity, the spatial integral of a receiver-antenna system's aver-
aged sensitivity
Glossary: Abbreviations
TRD: Also written as "Trd", short for transducer
TRP: Total radiated power, the sum of all RF power radiated by an antenna (source