The software contained in this product uses several valuable open source software packages. For information, see the "Open
Source Acknowledgment" on the user documentation CD-ROM (included in delivery) or at https://www.rohde-
schwarz.com/en/firmware/osp.
Rohde & Schwarz would like to thank the open source community for their valuable contribution to embedded computing.
General Safety Instructions
Please make sure to observe the instructions of the following sections so that you
cannot cause damage to the instrument or endanger people. This is of particular
importance when you use the instrument for the first time. Also observe the general
safety instructions at the beginning of this manual.
The following topics will help you to get familiar with the instrument and perform the
first steps:
●Front Panel Tour R&S OSP120orFront Panel Tour R&S OSP130 or Front Panel
Tour R&S OSP150
● Rear Panel Tour
● Putting the Instrument into Operation
2.1 Front Panel Tour R&S OSP120
This chapter gives an overview of the front panel controls and connectors of the
R&S OSP120 and gives all information that is necessary to put the instrument into
operation and connect external devices. Notes on reinstallation of the instrument
software appear at the end of the chapter.
Chapter 3.3, R&S OSP Panel Functions, of this manual provides an introduction to the
operation of the instrument by means of the control program OSP Panel. For a
description of the operating concept and an overview of the instrument’s capabilities
refer to the Instrument Functions (chapter 5).
The front panel of the R&S OSP120 consists of STANDBY switch, Status LEDs and
connectors. Brief explanations on the function of these items and the rear panel can
be found on the next pages.
R&S OSP Preparing for Use
Operating Manual 1505.3896.12 - 15 16
2.1.1 Status LEDs and Standby Key
The status LEDs light to indicate the following instrument states:
PWR: shows standby (yellow LED) and ready state (green LED).
LAN: shows if data are received via LAN network; seeRemote Control (chapter 6).
CLK: is flashing if data are transferred via CAN bus.
On an R&S OSP150, the logic is inverted: the CLK indicator goes off shortly during
CAN bus activity.
The STANDBY key serves the following purpose:
Toggle between standby and ready state (indicated by the yellow and green PWR
LED, respectively).
Single Universal Serial Bus connectors of type A (master USB), used to connect a
keyboard or flash drive. All front panel USB connectors comply with standard USB 2.0;
refer to the "Specifications".
USB Connection
The length of passive connecting USB cables should not exceed 1 m. The maximum
current per USB port is 500 mA. It is recommended to use double-shielded USB
cables.
The status LEDs and the standby toggle switch are located in the bottom right corner
of the front panel. See also chapter 2.5.9.
2.1.2 Front Panel Connectors
2.1.2.1 USB Connectors
R&S OSP Preparing for Use
Operating Manual 1505.3896.12 - 15 17
2.1.2.2 Monitor Connector
An external monitor with a digital interface can be connected to the MONITOR
DIGITAL connector on the front panel of the R&S OSP120. An external monitor,
together with a keyboard, is a prerequisite for manual intervention in order to define
the LAN configuration or in case of service to have access to the Linux Operating
system.
Monitor connection:
Before the external monitor is connected, the instrument must be switched off
(standby mode) to prevent damage to the monitor and the R&S OSP120. After
connection, the external monitor is detected when the instrument is started. The Linux
Operating system menu of the R&S OSP is then displayed on the external monitor.
Further settings are not required.
It is recommended to use a double-shielded monitor cable equipped with ferrites at
each end.
2.2 Front Panel Tour R&S OSP130
This chapter gives an overview of the front panel controls and connectors of the
R&S OSP130 and gives all information that is necessary to put the instrument into
operation and connect external devices. Notes on reinstallation of the instrument
software appear at the end of the chapter.
The chapter Getting Started of this manual provides an introduction to the operation of
the instrument by means of the control program OSP Panel. For a description of the
operating concept and an overview of the instrument’s capabilities refer to the
Instrument Functions (chapter 5). The graphical user interface on the instrument is
explained in chapter 5.3.
The front panel of the R&S OSP130 has the same STANDBY key and Status LEDs as
the R&S OSP120. See the corresponding sections in chapter 2.1.1 for details. There is
no connector for an external monitor.
R&S OSP Preparing for Use
Operating Manual 1505.3896.12 - 15 18
2.2.1 Navigation Keys
In the top row next to the screen there are three keys to quickly access different
hierarchy levels in the menus.
pressing the key MENU activates the top level menu.
pressing the key BACK moves up in hierarchy by one menu level.
pressing the key HOME leaves the menu and displays the Main menu screen
In the top row next to the screen there are three keys to quickly access different
hierarchy levels in the menus.
With the cursor keys the focus is moved within the screen elements; see a detailed
description in chapter 5.3.
In the top row next to the screen there are three keys to quickly access different
hierarchy levels in the menus.
pressing the key FUNCTION executes a switching action, if applicable.
pressing the key OK confirms a selection, for example for going to the next menu
level.
pressing the key STATUS displays status information for the chosen item (device or
module) or changes the selection of a switch for path configuration.
The OFF key is not used at the moment.
The ON key is not used at the moment.
Pressing the RESET key leaves the menu and displays the Main menu screen.
The LOCAL key is not used at the moment.
Error in supply voltage, please contact service.
General error condition. Please note any additional information what could have led to
this condition and contact R&S if the error is reproducible.
The navigation keys allow access to the various functions of the graphical user
interface. They can be subdivided into three groups.
2.2.2 Status Keys
The status keys serve the following purposes:
2.2.3 Status Indicators
The status indicators have the following meaning:
R&S OSP Preparing for Use
Operating Manual 1505.3896.12 - 15 19
General warning condition. Not used at the moment.
Indicates communication via the CAN interface.
Indicates communication via the CAN interface as well.
2.2.4 Front Panel USB Connectors
Single Universal Serial Bus connectors of type A (master USB), used to connect a
keyboard or flash drive. All front panel USB connectors comply with standard USB 2.0;
refer to the "Specifications".
USB Connection
The length of passive connecting USB cables should not exceed 1 m. The maximum
current per USB port is 500 mA. It is recommended to use double-shielded USB
cables.
2.3 Front Panel Tour R&S OSP150
This chapter gives an overview of the front panel controls of the R&S OSP150. The
front panel of the R&S OSP150 is similar to the front panel of the R&S OSP120 having
the Standby key and the three status LEDs but does not have any connector. See
chapter 2.1.1 for the description of the front panel elements.
Please notice that the operation of the CLK LED is different to OSP120 and OSP130.
2.4 Rear Panel Tour
This section gives an overview of the rear panel connectors of the R&S OSP120. The
rear panels of the R&S OSP130 and of the R&S OSP150 are almost identical to it.
The rear connectors and interfaces are described in detail in the complete operating
manual. The following connectors are available on the instruments:
R&S OSP Preparing for Use
Operating Manual 1505.3896.12 - 15 20
LAN connector (RJ-45) is used to integrate the instrument to a Local Area Network,
8-pin LAN connector RJ-45 used to connect the R&S OSP120 or OSP130 to a Local
Area Network (LAN). Refer to Remote Operation in a LAN.
LAN Connection
Depending on the connection (a non-dedicated network connection or dedicated
connection to a single controller) a standard RJ-45 cable or cross-over RJ-45 cable is
required. See chapter 2.9.1 for further information.
It is recommended to use double-shielded LAN cables of category 6 (SSTP).
primarily for remote control purposes. This connector is not available on the
R&S OSP150.
CAN bus connector (D-Sub, 9 pin) is a control port to connect one or several extension
units R&S OSP150.
The R&S OSP owns three slots which can be configured with the options available for
the R&S OSP. The above configuration shows the following options:
Option R&S OSP-B101: Relay module consisting of six coaxial relays of SPDT type;
the RF coaxial connectors are SMA type
Option R&S OSP-B102: Relay module consisting of two coaxial relays of SP6T type;
the RF coaxial connectors are SMA type
Option R&S OSP-B103: I/O module with 16 Bit input / output ports; the connectors are
D-Sub types, 25 pin, female for output and male for input
In case of the options which can be installed in the R&S OSP, the connectors related
to each option are described in detail in chapter 9.2, R&S OSP Module Interfaces.
2.4.1 LAN Connector
R&S OSP Preparing for Use
Operating Manual 1505.3896.12 - 15 21
2.4.2 CAN Bus Connector
9-pin connector D-Sub male used to connect the R&S OSP120 or R&S OSP130 to
extension units R&S OSP150. Refer to Instrument Functions in chapter 5.
Maximum input levels
The maximum input levels and voltages of the input connectors at the front and rear
panel must not be exceeded.
Supply Voltage over CAN bus
In some CAN bus applications the R&S OSP120, OSP130 or OSP150 is required to
deliver the supply voltage to an external device. An example for this is the connection
of the fiber-optic extender R&S OSP-Z104. Using the cable OSP-Z106 is
recommended for this case.
However, when connecting an R&S OSP150 to an OSP120 or OSP130, this supply
voltage line must not be present. Using the connecting cables R&S OSP-Z101 or
R&S OSP-Z102 is recommended. Not observing these precautions may damage the
power supplies in the R&S OSP.
See also chapter 9.1.2.1 for the pinout of the CAN bus connector.
The mains connector is located at the bottom left corner of the rear panel.
Next to the connector is a mains fuse, and above the fuse the mains switch. With the
positions “0” the instrument is disconnected from mains, and in position “1” it is
connected. See chapters 2.5.6 and 2.5.7 for more information.
2.4.3 Mains Switch and Connector
R&S OSP Preparing for Use
Operating Manual 1505.3896.12 - 15 22
2.5 Putting the Instrument into Operation
Instrument setup
Before turning on the instrument, please make sure that the following conditions are
fulfilled:
Instrument covers are in place and all fasteners are tightened.
Fan openings are unobstructed.
Signal levels at the input connectors are all within the specified ranges.
Signal outputs are correctly connected and not overloaded.
The instrument is dry and shows no condensation.
Non-observance may cause damage to the instrument or other devices in the test
setup.
This section describes the basic steps to be taken when setting up the R&S OSP for
the first time.
2.5.1 Unpacking the Instrument and Checking the Shipment
Remove the instrument from its packaging and check the equipment for completeness
using the delivery note and the accessory lists for the various items.
First, pull off the polyethylene protection pads from the instrument's rear feet and then
carefully remove the pads from the instrument handles at the front.
Pull off the corrugated cardboard cover that protects the rear of the instrument.
Carefully unthread the corrugated cardboard cover at the front that protects the
instrument handles and remove it.
Check the instrument for any damage. If there is damage, immediately contact the
carrier who delivered the instrument. In this case, make sure not to discard the box
and packing material.
It is advisable to keep the original packing material in order to prevent control elements
and connectors from being damaged in case the instrument is to be transported or
shipped at a later date.
R&S OSP Preparing for Use
Operating Manual 1505.3896.12 - 15 23
2.5.2 Instrument Setup
Allow for sufficient air supply in the rack.
Make sure that there is sufficient space between the ventilation holes and the rack
casing.
The R&S OSP is designed for use under laboratory conditions, either on a bench top
or in a rack. The general ambient conditions required at the operating site are as
follows:
The ambient temperature must be in the ranges specified for operation and for
compliance with specifications (see "Specifications").
All fan openings including the rear panel perforations must be unobstructed. The
distance to the wall should be at least 10 cm.
2.5.3 Bench Top Operation
If the R&S OSP is operated on a bench top, the surface should be flat.
The instrument is used in horizontal position, standing on its feet.
2.5.4 Mounting in a 19" Rack
The instrument can be mounted in 19" racks using a ZZA-211 adapter (order number
1096.3260.00). Please note the mounting instructions supplied with the rack adapter.
2.5.5 EMI Protective Measures
In order to avoid electromagnetic interference (EMI), the instrument may only be
operated when it is closed and with all shielding covers fitted. Only appropriate
shielded signal and control cables may be used.
R&S OSP Preparing for Use
Operating Manual 1505.3896.12 - 15 24
2.5.6 Connecting the Instrument to the AC Supply
To turn the power on or off, press the AC power switch to position I (On) or 0 (Off).
See also Replacing Fuses.
After power-on, the R&S OSP instrument is in ready state after about 30 seconds.
The STANDBY key at the front panel of the instrument is used to toggle between
standby and ready state. The AC power switch can be permanently on.
Switching off is required only if the instrument must be completely removed from the
AC power supply but may be used in order to reduce power consumption when not in
use.
Extension units
If communication between OSP120 or OSP130 and any extension unit OSP150
connected via CAN bus cannot be established properly, try to turn on the extension
unit(s) first before turning on the OSP120 or OSP130.
Shock hazard
For fuse replacement, ensure that the instrument is switched off and disconnected
from the power supply by removing the plug from the AC power connector.
The R&S OSP is automatically adapted to the AC supply voltage supplied. The supply
voltage must be between 100 V and 240 V with frequencies ranging from 50 Hz to
60 Hz (see also the tolerances quoted in the "Specifications"). The mains connector is
located in the lower left corner of the rear panel.
► Connect the instrument to the AC power source using the AC power cable
delivered with the instrument.
The maximum power consumption of the instrument depends on the installed options.
The typical power consumption is also listed in the "Specifications".
The R&S OSP is protected by two fuses located in the fuse holder below the AC power
switch; see Replacing Fuses.
2.5.7 Power on and off
The mains connector is located at the bottom left corner of the rear panel.
2.5.8 Replacing Fuses
The instrument is protected by two fuses (IEC 127- T4.0H/250V, stock no.
0020.7600.00) located in the fuse holder below the AC power switch on the rear panel.
To replace the fuses
R&S OSP Preparing for Use
Operating Manual 1505.3896.12 - 15 25
1. Open the lid of the AC power connector.
After switching on the AC power, the R&S OSP is in ready mode after about
30 seconds.
Press the STANDBY key on the front panel briefly to switch the R&S OSP from the
standby to ready state or vice versa.
Shock hazard
The instrument is still power-supplied while it is in standby mode.
Instrument damage caused by cleaning agents
Cleaning agents contain substances that may damage the instrument, e.g. solvent-
containing cleaning agents may damage the front panel labeling or plastic parts. Never
use cleaning agents such as solvents (thinners, acetone, etc), acids, bases, or other
substances.
2. Lift the fuse holder out of its slot.
3. Exchange the fuses.
4. Put the fuse holder back in its slot and close the lid.
2.5.9 Standby and Ready State
The STANDBY toggle switch is located in the bottom right corner of the front panel.
In standby state, the left PWR LED is yellow. The standby power only supplies the
power switch circuits. In this state it is safe to switch off the AC power and disconnect
the instrument from the power supply.
After Power On or when changing from Standby to Ready state, the left PWR LED is
immediately switched to green and all modules are power-supplied. Please note that
the Linux operating system of R&S OSP takes about 30 seconds to start up. After this
time the R&S OSP is ready for operation.
2.6 Maintenance
The R&S OSP does not require any special maintenance. Make sure that the air vents
are not obstructed. The outside of the instrument is suitably cleaned using a soft, nonfluffy dust cloth.
For our support center address and a list of useful R&S contact addresses refer to the
pages at the beginning of this manual.
R&S OSP Preparing for Use
Operating Manual 1505.3896.12 - 15 26
2.6.1 Storing and Packing
A USB flash drive can be connected to one of the USB Connectors on the front panel
of the R&S OSP120 or OSP130.
The flash drive is detected automatically when it is connected.
A keyboard can be connected to one of the USB Connectors on the front panel of the
R&S OSP120.
The keyboard is detected automatically when it is connected. The default input
language is English – US. Refer to chapter 2.10.1 for changing the keyboard
properties.
Keyboard configuration
The keyboard configuration already is pre-configured.
Operating the R&S OSP120 does not require a keyboard. Usually all essential
functions can be controlled via LAN interface.
The R&S OSP can be stored at the temperature range quoted in the data sheet. When
it is stored for a longer period of time the instrument should be protected against dust.
The original packing should be used, particularly the protective caps at the front and
rear, when the instrument is to be transported or dispatched. If the original packing is
no longer available, use a sturdy cardboard box of suitable size and carefully wrap the
instrument to protect it against mechanical damage.
2.7 Connecting External Accessories
The LAN interface at the rear panel of the R&S OSP120 or OSP130 is used for remote
control of the instrument:
●A LAN connection can be established in order to remotely control the instrument
from an external PC (seeConnecting a LAN Cable).
In addition the R&S OSP120 provides interfaces for monitor connection and USB
connection:
●An external monitor shows the menus of the Linux Operating System, if any
access or setup within the Linux Operating System should be necessary.
● A keyboard simplifies the entry of data (seeConnecting a Keyboard).
● A flash drive supports the firmware update (see Connecting a USB Flash Drive).
2.7.1 Connecting a USB Flash Drive
2.7.2 Connecting a Keyboard
R&S OSP Preparing for Use
Operating Manual 1505.3896.12 - 15 27
2.7.3 Connecting a Monitor
A standard monitor can be connected to the DVI-D connector on the front panel of the
R&S OSP120.
Monitor configuration
There is no particular configuration of the monitor required.
Operating the R&S OSP120 does not require a monitor. Usually all essential functions
can be control via LAN interface.
Monitor connection
The monitor must be connected while the instrument is switched off (in standby mode).
Otherwise correct operation can not be guaranteed.
A LAN cable can be connected to the LAN connector on the rear panel of the
R&S OSP120 or OSP130. To establish a LAN connection proceed as follows:
Refer to chapter Assigning an IP Address and learn how to avoid connection errors.
Connect an appropriate LAN cable to the LAN port. Use a commercial RJ-45 cable to
establish a non-dedicated network connection, or a cross-over RJ-45 cable to
establish a dedicated connection between the instrument and a single PC.
The monitor displays the menus of the Linux Operating System which is integral part of
the R&S OSP120. The monitor together with keyboard is required if any setups /
changes have to be done in the Linux operating system.
2.7.4 Connecting a LAN Cable
Dedicated vs. non-dedicated network connections
There are two methods to establish a LAN connection of the R&S OSP:
●A non-dedicated network (Ethernet) connection from the instrument to an existing
network made with an ordinary RJ-45 network cable. The instrument is assigned an
IP address and can coexist with a computer and with other hosts on the same
network.
●A dedicated network connection between the instrument and a single computer
made with a cross-over RJ-45 network cable. The computer must be equipped
with a network adapter and is directly connected to the instrument. The use of
hubs, switches, or gateways is not needed, however, data transfer is still made
using the TCP/IP protocol.
R&S OSP Preparing for Use
Operating Manual 1505.3896.12 - 15 28
2.8 Starting the R&S OSP and Shutting Down
Standby state
It is recommended to switch the R&S OSP to standby state before disconnecting it
from the AC supply.
To start the R&S OSP, proceed as follows:
●Make sure that the instrument is connected to the AC power supply and switch
the power switch (see chapter 2.5.7) on the rear panel to position I (On). All
modules of the R&S OSP are powered (the left PWR LED is green) and after
about 30 seconds the instrument is in ready state.
●If necessary, press the STANDBY toggle switch (see chapter 2.5.9) on the front
panel to switch the instrument to ready state (the left PWR LED is green).
●In ready state, the instrument already has booted the Linux Operating System
(see chapter 2.10) and started the R&S OSP application. Independent of the last
setup, the R&S OSP always starts with the default conditions, i.e. all relays in
reset condition.
To shut down the R&S OSP, proceed as follows:
●Press the STANDBY key, which will shut down the Linux operating system and
set the instrument to standby state.
●If desired, set the AC power switch to position 0 (Off).
2.9 Remote Operation in a LAN
A LAN connection is used to integrate the R&S OSP (not the OSP150) into a
home/company network. The LAN connection is required for:
● Remote control operation of the R&S OSP120 or OSP130.
● Manual control of the R&S OSP from a remote computer using the "OSP Panel"
application.
To establish the connection proceed as follows:
●Assign an IP address to the R&S OSP following the directions below and connect
the instrument to the network as described in chapter 2.7.4, Connecting a LAN
Cable.
2.9.1 Assigning an IP Address
There are two different modes for the OSP LAN configuration. The OSP as it is
delivered works in the LAN configuration as described in chapter 2.9.1.1.
With OSP firmware version 2.51 onwards, the OSP LAN configuration can be setup as
described in chapter 2.9.1.2.
R&S OSP Preparing for Use
Operating Manual 1505.3896.12 - 15 29
The actual LAN configuration can be read out from the OSP by the following SCPI
Valid IP addresses
If your network does not support DHCP, or if you choose to disable dynamic TCP/IP
configuration, you must assign valid address information before connecting the
R&S OSP to the LAN. Contact your network administrator to obtain a valid IP address,
because connection errors can affect the entire network.
command:
SYSTem:NETWork:MODe?
2.9.1.1 Default mode of LAN Configuration
The actual LAN configuration is set to mode -> DHCP_AUTO
Depending on the network capacities, the IP address information for the R&S OSP120
or R&S OSP130 can be obtained in different ways.
●If the network supports dynamic TCP/IP configuration using the Dynamic Host
Configuration Protocol (DHCP), all address information can be assigned
automatically.
●If the network does not support DHCP, or if the instrument is set to use alternate
TCP/IP configuration, a static IP address is used.
By default, the R&S OSP is configured either to use:
●A dynamic TCP/IP configuration and obtain all address information automatically.
This means that it is safe to establish a physical connection to the LAN without
any previous R&S OSP configuration.
(Priority 1)
●A user defined static IP address. This address can be defined (but must not) in
addition to the default address. See chapter 3.3.3 for information how to configure
the user defined IP address via the OSP Panel application.
(Priority 2)
●A static default IP address. Per default, the R&S OSP is set to the IP address
192.168.48.147. See also chapter 3.1.
(Priority 3)
Please note the priorities shown above which is related to IP address handling done
by the OSP.
Reading the IP address
When using the dynamic TCP/IP configuration, the actual IP address information will
be displayed when booting the R&S OSP. Proceed as follows:
Connect the R&S OSP via a commercial RJ-45 cable to your network supporting
DHCP
R&S OSP Preparing for Use
Operating Manual 1505.3896.12 - 15 30
Switch on the R&S OSP120 with Monitor and Keyboard connected (not applicable to
the OSP130)
At the end of the booting process, the IP address will be displayed as shown below for
the OSP120 and the OSP130. The values of the IP address shown are only examples.
Press Okay button or the ENTER key on the keyboard (OK key for the R&S OSP130)
for confirmation and to continue the operation.
Manual TCP/IP configuration
If your network does not support DHCP, proceed as follows:
1. Connect the R&S OSP120 or OSP130 via a cross-over RJ-45 cable to your
computer (or network not supporting DHCP).
2. Set the network configuration of your computer to work with a static IP address.
The following setup is recommended:
3. Switch on the R&S OSP and wait until boot-up is finished.
The R&S OSP now can be accessed under its default IP address from your computer.
Loading...
+ 195 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.