Starting the Application.......................................................................................................................................10
Exiting the Application ........................................................................................................................................10
General Settings .................................................................................................................................................63
Signal Characteristics ...............................................................................................................64
Standard .........................................................................................................................64
Frequency .......................................................................................................................64
Frequency Band..............................................................................................................64
Link Gate and Marker .....................................................................................................90
Import/Export of IQ Data ....................................................................................................................................91
4 Remote Control .................................................................................................. 132
Description of commands ................................................................................................................................132
FORMat Subsystem .........................................................................................................................................183
SENSe Subsystem ...........................................................................................................................................190
STATus Subsystem..........................................................................................................................................199
SYSTEM Subsystem .......................................................................................................................................203
UNIT Subsystem...............................................................................................................................................211
Status Reporting Registers ..............................................................................................................................212
Description of the Status Registers.........................................................................................214
All plants and locations of the Rohde & Schwarz group of companies make every effort to keep the
safety standard of our products up to date and to offer our customers the highest possible degree of
safety. Our products and the auxiliary equipment required for them are designed and tested in
accordance with the relevant safety standards. Compliance with these standards is continuously
monitored by our quality assurance system. The product described here has been designed and tested
in accordance with the EC Certificate of Conformity and has left the manufacturer’s plant in a condition
fully complying with safety standards. To maintain this condition and to ensure safe operation, observe
all instructions and warnings provided in this manual. If you have any questions regarding these safety
instructions, the Rohde & Schwarz group of companies will be happy to answer them.
Furthermore, it is your responsibility to use the product in an appropriate manner. This product is
designed for use solely in industrial and laboratory environments or in the field and must not be used in
any way that may cause personal injury or property damage. You are responsible if the product is used
for an intention other than its designated purpose or in disregard of the manufacturer's instructions. The
manufacturer shall assume no responsibility for such use of the product.
The product is used for its designated purpose if it is used in accordance with its product documentation
and within its performance limits (see data sheet, documentation, the following safety instructions).
Using the product requires technical skills and a basic knowledge of English. It is therefore essential
that the product be used exclusively by skilled and specialized staff or thoroughly trained personnel with
the required skills. If personal safety gear is required for using Rohde & Schwarz products, this will be
indicated at the appropriate place in the product documentation.
Observe
product
documentation
Supply
voltage
ON/OFF
Weight
indication for
units >18 kg
Standby
indication
Symbols and safety labels
Danger of
electric
shock
Warning!
Hot
surface
PE terminal Ground
Direct
current
(DC)
Alternating
current (AC)
Direct/alternating
current (DC/AC)
Ground
terminal
Device fully
protected by
double/reinforced
insulation
Attention!
Electrostatic
sensitive
devices
1171.0000.42-03.00 Sheet 1
Page 8
Safety Instructions
Observing the safety instructions will help prevent personal injury or damage of any kind caused by
dangerous situations. Therefore, carefully read through and adhere to the following safety instructions
before putting the product into operation. It is also absolutely essential to observe the additional safety
instructions on personal safety that appear in relevant parts of the product documentation. In these
safety instructions, the word "product" refers to all merchandise sold and distributed by the Rohde &
Schwarz group of companies, including instruments, systems and all accessories.
Tags and their meaning
DANGER
WARNING
CAUTION This tag indicates a hazard carrying a low risk of minor or moderate
ATTENTION
NOTE
These tags are in accordance with the standard definition for civil applications in the European
Economic Area. Definitions that deviate from the standard definition may also exist in other economic
areas or military applications. It is therefore essential to make sure that the tags described here are
always used only in connection with the related product documentation and the related product. The
use of tags in connection with unrelated products or documentation can result in misinterpretation and
thus contribute to personal injury or material damage.
This tag indicates a definite hazard carrying a high risk of death or
serious injury if not avoided.
This tag indicates a possible hazard carrying a medium risk of death or
(serious) injury if not avoided
injury if not avoided.
This tag indicates the possibility of incorrect use that can cause damage
to the product.
This tag indicates a situation where the user should pay special attention
to operating the product but which does not lead to damage.
.
Basic safety instructions
1. The product may be operated only under
the operating conditions and in the
positions specified by the manufacturer. Its
ventilation must not be obstructed during
operation. Unless otherwise specified, the
following requirements apply to
Rohde & Schwarz products:
prescribed operating position is always with
the housing floor facing down, IP protection
2X, pollution severity 2, overvoltage
category 2, use only in enclosed spaces,
max. operation altitude 2000 m above sea
level, max. transport altitude 4500 m above
sea level.
Unless specified otherwise in the data
sheet, a tolerance of ±10% shall apply to
the nominal voltage and of ±5% to the
nominal frequency.
2. Applicable local or national safety
regulations and rules for the prevention of
accidents must be observed in all work
performed. The product may be opened
only by authorized, specially trained
personnel. Prior to performing any work on
the product or opening the product, the
product must be disconnected from the
supply network. Any adjustments,
replacements of parts, maintenance or
repair must be carried out only by technical
personnel authorized by Rohde & Schwarz.
Only original parts may be used for
replacing parts relevant to safety (e.g.
power switches, power transformers,
fuses). A safety test must always be
performed after parts relevant to safety
have been replaced (visual inspection, PE
conductor test, insulation resistance
measurement, leakage current
measurement, functional test).
3. As with all industrially manufactured goods,
the use of substances that induce an
allergic reaction (allergens, e.g. nickel)
such as aluminum cannot be generally
excluded. If you develop an allergic
reaction (such as a skin rash, frequent
sneezing, red eyes or respiratory
difficulties), consult a physician immediately
to determine the cause.
1171.0000.42-03.00 Sheet 2
Page 9
Safety Instructions
4. If products/components are mechanically
and/or thermically processed in a manner
that goes beyond their intended use,
hazardous substances (heavy-metal dust
such as lead, beryllium, nickel) may be
released. For this reason, the product may
only be disassembled, e.g. for disposal
purposes, by specially trained personnel.
Improper disassembly may be hazardous to
your health. National waste disposal
regulations must be observed.
5. If handling the product yields hazardous
substances or fuels that must be disposed
of in a special way, e.g. coolants or engine
oils that must be replenished regularly, the
safety instructions of the manufacturer of
the hazardous substances or fuels and the
applicable regional waste disposal
regulations must be observed. Also
observe the relevant safety instructions in
the product documentation.
6. Depending on the function, certain products
such as RF radio equipment can produce
an elevated level of electromagnetic
radiation. Considering that unborn life
requires increased protection, pregnant
women should be protected by appropriate
measures. Persons with pacemakers may
also be endangered by electromagnetic
radiation. The employer/operator is
required to assess workplaces where there
is a special risk of exposure to radiation
and, if necessary, take measures to avert
the danger.
7. Operating the products requires special
training and intense concentration. Make
certain that persons who use the products
are physically, mentally and emotionally fit
enough to handle operating the products;
otherwise injuries or material damage may
occur. It is the responsibility of the
employer to select suitable personnel for
operating the products.
8. Prior to switching on the product, it must be
ensured that the nominal voltage setting on
the product matches the nominal voltage of
the AC supply network. If a different voltage
is to be set, the power fuse of the product
may have to be changed accordingly.
9. In the case of products of safety class I with
movable power cord and connector,
operation is permitted only on sockets with
earthing contact and protective earth
connection.
10. Intentionally breaking the protective earth
connection either in the feed line or in the
product itself is not permitted. Doing so can
result in the danger of an electric shock
from the product. If extension cords or
connector strips are implemented, they
must be checked on a regular basis to
ensure that they are safe to use.
11. If the product has no power switch for
disconnection from the AC supply, the plug
of the connecting cable is regarded as the
disconnecting device. In such cases, it
must be ensured that the power plug is
easily reachable and accessible at all times
(corresponding to the length of connecting
cable, approx. 2 m). Functional or
electronic switches are not suitable for
providing disconnection from the AC
supply. If products without power switches
are integrated in racks or systems, a
disconnecting device must be provided at
the system level.
12. Never use the product if the power cable is
damaged. Check the power cable on a
regular basis to ensure that it is in proper
operating condition. By taking appropriate
safety measures and carefully laying the
power cable, ensure that the cable cannot
be damaged and that no one can be hurt by
e.g. tripping over the cable or suffering an
electric shock.
13. The product may be operated only from
TN/TT supply networks fused with max.
16 A (higher fuse only after consulting with
the Rohde & Schwarz group of companies).
14. Do not insert the plug into sockets that are
dusty or dirty. Insert the plug firmly and all
the way into the socket. Otherwise, this can
result in sparks, fire and/or injuries.
15. Do not overload any sockets, extension
cords or connector strips; doing so can
cause fire or electric shocks.
16. For measurements in circuits with voltages
V
> 30 V, suitable measures (e.g.
rms
appropriate measuring equipment, fusing,
current limiting, electrical separation,
insulation) should be taken to avoid any
hazards.
17. Ensure that the connections with
information technology equipment comply
with IEC 950/EN 60950.
18. Unless expressly permitted, never remove
the cover or any part of the housing while
the product is in operation. Doing so will
expose circuits and components and can
lead to injuries, fire or damage to the
product.
1171.0000.42-03.00 Sheet 3
Page 10
Safety Instructions
19. If a product is to be permanently installed,
the connection between the PE terminal on
site and the product's PE conductor must
be made first before any other connection
is made. The product may be installed and
connected only by a license electrician.
20. For permanently installed equipment
without built-in fuses, circuit breakers or
similar protective devices, the supply circuit
must be fused in such a way that suitable
protection is provided for users and
products.
21. Do not insert any objects into the openings
in the housing that are not designed for this
purpose. Never pour any liquids onto or into
the housing. This can cause short circuits
inside the product and/or electric shocks,
fire or injuries.
22. Use suitable overvoltage protection to
ensure that no overvoltage (such as that
caused by a thunderstorm) can reach the
product. Otherwise the operating personnel
will be endangered by electric shocks.
23. Rohde & Schwarz products are not
protected against penetration of water,
unless otherwise specified (see also safety
instruction 1.). If this is not taken into
account, there exists the danger of electric
shock for the user or damage to the
product, which can also lead to personal
injury.
24. Never use the product under conditions in
which condensation has formed or can form
in or on the product, e.g. if the product was
moved from a cold to a warm environment.
25. Do not close any slots or openings on the
product, since they are necessary for
ventilation and prevent the product from
overheating. Do not place the product on
soft surfaces such as sofas or rugs or
inside a closed housing, unless this is well
ventilated.
26. Do not place the product on heatgenerating devices such as radiators or fan
heaters. The temperature of the
environment must not exceed the maximum
temperature specified in the data sheet.
27. Batteries and storage batteries must not be
exposed to high temperatures or fire. Keep
batteries and storage batteries away from
children. Do not short-circuit batteries and
storage batteries.
If batteries or storage batteries are
improperly replaced, this can cause an
explosion (warning: lithium cells). Replace
the battery or storage battery only with the
matching Rohde & Schwarz type (see
spare parts list). Batteries and storage
batteries must be recycled and kept
separate from residual waste. Batteries and
storage batteries that contain lead, mercury
or cadmium are hazardous waste. Observe
the national regulations regarding waste
disposal and recycling.
28. Please be aware that in the event of a fire,
toxic substances (gases, liquids etc.) that
may be hazardous to your health may
escape from the product.
29. The product can be very heavy. Be careful
when moving it to avoid back or other
physical injuries.
30. Do not place the product on surfaces,
vehicles, cabinets or tables that for reasons
of weight or stability are unsuitable for this
purpose. Always follow the manufacturer's
installation instructions when installing the
product and fastening it to objects or
structures (e.g. walls and shelves).
31. Handles on the products are designed
exclusively for personnel to hold or carry
the product. It is therefore not permissible
to use handles for fastening the product to
or on means of transport such as cranes,
fork lifts, wagons, etc. The user is
responsible for securely fastening the
products to or on the means of transport
and for observing the safety regulations of
the manufacturer of the means of transport.
Noncompliance can result in personal injury
or material damage.
32. If you use the product in a vehicle, it is the
sole responsibility of the driver to drive the
vehicle safely. Adequately secure the
product in the vehicle to prevent injuries or
other damage in the event of an accident.
Never use the product in a moving vehicle if
doing so could distract the driver of the
vehicle. The driver is always responsible for
the safety of the vehicle. The manufacturer
assumes no responsibility for accidents or
collisions.
33. If a laser product (e.g. a CD/DVD drive) is
integrated in a Rohde & Schwarz product,
do not use any other settings or functions
than those described in the product
documentation. Otherwise this may be
hazardous to your health, since the laser
beam can cause irreversible damage to
your eyes. Never try to take such products
apart, and never look into the laser beam.
1171.0000.42-03.00 Sheet 4
Page 11
Informaciones de seguridad
Por favor lea imprescindiblemente antes de
la primera puesta en funcionamiento las
siguientes
Informaciones de seguridad
El principio del grupo de empresas Rohde & Schwarz consiste en tener nuestros productos siempre al
día con los estandards de seguridad y de ofrecer a nuestros clientes el máximo grado de seguridad.
Nuestros productos y todos los equipos adicionales son siempre fabricados y examinados según las
normas de seguridad vigentes. Nuestra sección de gestión de la seguridad de calidad controla
constantemente que sean cumplidas estas normas. El presente producto ha sido fabricado y
examinado según el comprobante de conformidad adjunto según las normas de la CE y ha salido de
nuestra planta en estado impecable según los estandards técnicos de seguridad. Para poder preservar
este estado y garantizar un funcionamiento libre de peligros, el usuario deberá atenerse a todas las
informaciones, informaciones de seguridad y notas de alerta. El grupo de empresas Rohde & Schwarz
está siempre a su disposición en caso de que tengan preguntas referentes a estas informaciones de
seguridad.
Además queda en la responsabilidad del usuario utilizar el producto en la forma debida. Este producto
solamente fue elaborado para ser utilizado en la industria y el laboratorio o para fines de campo y de
ninguna manera deberá ser utilizado de modo que alguna persona/cosa pueda ser dañada. El uso del
producto fuera de sus fines definidos o despreciando las informaciones de seguridad del fabricante
queda en la responsabilidad del usuario. El fabricante no se hace en ninguna forma responsable de
consecuencias a causa del mal uso del producto.
Se parte del uso correcto del producto para los fines definidos si el producto es utilizado dentro de las
instrucciones de la correspondiente documentación de producto y dentro del margen de rendimiento
definido (ver hoja de datos, documentación, informaciones de seguridad que siguen). El uso del
producto hace necesarios conocimientos profundos y conocimientos parciales del idioma inglés. Por
eso se deberá tener en cuenta de exclusivamente autorizar para el uso del producto a personas peritas
o debidamente minuciosamente instruidas con los conocimientos citados. Si fuera necesaria
indumentaria de seguridad para el uso de productos de R&S, encontrará la información debida en la
documentación del producto en el capítulo correspondiente.
Símbolos y definiciones de seguridad
Ver
documentación de
producto
Informaciones
para
maquinaria
con uns peso
de > 18kg
Peligro de
golpe de
corriente
¡Advertencia!
Superficie
caliente
Conexión a
conductor
protector
Conexión
a tierra
Conexión
a masa
conductora
¡Cuidado!
Elementos de
construcción
con peligro de
carga
electroestática
El aparato está
protegido en su
totalidad por un
aislamiento de
doble refuerzo
potencia EN
MARCHA/PARADA
Indicación
Stand-by
Corriente
continua
DC
Corriente
alterna AC
Corriente
continua/alterna
DC/AC
1171.0000.42-03.00 Sheet 5
Page 12
Informaciones de seguridad
Tener en cuenta las informaciones de seguridad sirve para tratar de evitar daños y peligros de toda
clase. Es necesario de que se lean las siguientes informaciones de seguridad concienzudamente y se
tengan en cuenta debidamente antes de la puesta en funcionamiento del producto. También deberán
ser tenidas en cuenta las informaciones para la protección de personas que encontrarán en el capítulo
correspondiente de la documentación de producto y que también son obligatorias de seguir. En las
informaciones de seguridad actuales hemos juntado todos los objetos vendidos por el grupo de
empresas Rohde & Schwarz bajo la denominación de „producto“, entre ellos también aparatos,
instalaciones así como toda clase de accesorios.
Palabras de señal y su significado
PELIGRO Identifica un peligro directo con riesgo elevado de provocar muerte o
lesiones de gravedad si no se toman las medidas oportunas.
ADVERTENCIA Identifica un posible peligro con riesgo medio de provocar muerte o
lesiones (de gravedad) si no se toman las medidas oportunas.
ATENCIÓN Identifica un peligro con riesgo reducido de provocar lesiones de
gravedad media o leve si no se toman las medidas oportunas.
CUIDADO Indica la posibilidad de utilizar mal el producto y a consecuencia
dañarlo.
INFORMACIÓN Indica una situación en la que deberían seguirse las instrucciones en el
uso del producto, pero que no consecuentemente deben de llevar a un
daño del mismo.
Las palabras de señal corresponden a la definición habitual para aplicaciones civiles en el área
económica europea. Pueden existir definiciones diferentes a esta definición en otras áreas económicas
o en aplicaciones militares. Por eso se deberá tener en cuenta que las palabras de señal aquí descritas
sean utilizadas siempre solamente en combinación con la correspondiente documentación de producto
y solamente en combinación con el producto correspondiente. La utilización de las palabras de señal
en combinación con productos o documentaciones que no les correspondan puede llevar a
malinterpretaciones y tener por consecuencia daños en personas u objetos.
Informaciones de seguridad elementales
1. El producto solamente debe ser utilizado
según lo indicado por el fabricante referente
a la situación y posición de funcionamiento
sin que se obstruya la ventilación. Si no se
convino de otra manera, es para los
productos R&S válido lo que sigue:
como posición de funcionamiento se define
principialmente la posición con el suelo de la
caja para abajo , modo de protección IP 2X,
grado de suciedad 2, categoría de
sobrecarga eléctrica 2, utilizar solamente en
estancias interiores, utilización hasta 2000 m
sobre el nivel del mar, transporte hasta
4.500 m sobre el nivel del mar.
A menos que se especifique otra cosa en la
hoja de datos, se aplicará una tolerancia de
±10% sobre el voltaje nominal y de ±5%
sobre la frecuencia nominal.
2. En todos los trabajos deberán ser tenidas en
cuenta las normas locales de seguridad de
trabajo y de prevención de accidentes. El
producto solamente debe de ser abierto por
personal perito autorizado. Antes de efectuar
trabajos en el producto o abrirlo deberá este
ser desconectado de la corriente. El ajuste,
el cambio de partes, la manutención y la
reparación deberán ser solamente
efectuadas por electricistas autorizados por
R&S. Si se reponen partes con importancia
para los aspectos de seguridad (por ejemplo
el enchufe, los transformadores o los
fusibles), solamente podrán ser sustituidos
por partes originales. Despues de cada
recambio de partes elementales para la
seguridad deberá ser efectuado un control de
seguridad (control a primera vista, control de
conductor protector, medición de resistencia
de aislamiento, medición de medición de la
corriente conductora, control de
funcionamiento).
1171.0000.42-03.00 Sheet 6
Page 13
Informaciones de seguridad
3. Como en todo producto de fabricación
industrial no puede ser excluido en general
de que se produzcan al usarlo elementos
que puedan generar alergias, los llamados
elementos alergénicos (por ejemplo el
níquel). Si se producieran en el trato con
productos R&S reacciones alérgicas, como
por ejemplo urticaria, estornudos frecuentes,
irritación de la conjuntiva o dificultades al
respirar, se deberá consultar inmediatamente
a un médico para averigurar los motivos de
estas reacciones.
4. Si productos / elementos de construcción son
tratados fuera del funcionamiento definido de
forma mecánica o térmica, pueden generarse
elementos peligrosos (polvos de sustancia
de metales pesados como por ejemplo
plomo, berilio, níquel). La partición elemental
del producto, como por ejemplo sucede en el
tratamiento de materias residuales, debe de
ser efectuada solamente por personal
especializado para estos tratamientos. La
partición elemental efectuada
inadecuadamente puede generar daños para
la salud. Se deben tener en cuenta las
directivas nacionales referentes al
tratamiento de materias residuales.
5. En el caso de que se produjeran agentes de
peligro o combustibles en la aplicación del
producto que debieran de ser transferidos a
un tratamiento de materias residuales, como
por ejemplo agentes refrigerantes que deben
ser repuestos en periodos definidos, o
aceites para motores, deberan ser tenidas en
cuenta las prescripciones de seguridad del
fabricante de estos agentes de peligro o
combustibles y las regulaciones regionales
para el tratamiento de materias residuales.
Cuiden también de tener en cuenta en caso
dado las prescripciones de seguridad
especiales en la descripción del producto.
6. Ciertos productos, como por ejemplo las
instalaciones de radiación HF, pueden a
causa de su función natural, emitir una
radiación electromagnética aumentada. En
vista a la protección de la vida en desarrollo
deberían ser protegidas personas
embarazadas debidamente. También las
personas con un bypass pueden correr
peligro a causa de la radiación
electromagnética. El empresario/usario está
comprometido a valorar y señalar areas de
trabajo en las que se corra un riesgo
aumentado de exposición a radiaciones para
evitar riesgos.
7. La utilización de los productos requiere
instrucciones especiales y una alta
concentración en el manejo. Debe de
ponerse por seguro de que las personas que
manejen los productos estén a la altura de
los requerimientos necesarios referente a
sus aptitudes físicas, psíquicas y
emocionales, ya que de otra manera no se
pueden excluir lesiones o daños de objetos.
El empresario lleva la responsabilidad de
seleccionar el personal usuario apto para el
manejo de los productos.
8. Antes de la puesta en marcha del producto
se deberá tener por seguro de que la tensión
preseleccionada en el producto equivalga a
la del la red de distribución. Si es necesario
cambiar la preselección de la tensión
también se deberán en caso dabo cambiar
los fusibles correspondientes del prodcuto.
9. Productos de la clase de seguridad I con
alimentación móvil y enchufe individual de
producto solamente deberán ser conectados
para el funcionamiento a tomas de corriente
de contacto de seguridad y con conductor
protector conectado.
10. Queda prohibida toda clase de interrupción
intencionada del conductor protector, tanto
en la toma de corriente como en el mismo
producto. Puede tener como consecuencia el
peligro de golpe de corriente por el producto.
Si se utilizaran cables o enchufes de
extensión se deberá poner al seguro, que es
controlado su estado técnico de seguridad.
11. Si el producto no está equipado con un
interruptor para desconectarlo de la red, se
deberá considerar el enchufe del cable de
distribución como interruptor. En estos casos
deberá asegurar de que el enchufe sea de
fácil acceso y nabejo (según la medida del
cable de distribución, aproximadamente
2 m). Los interruptores de función o
electrónicos no son aptos para el corte de la
red eléctrica. Si los productos sin interruptor
están integrados en construciones o
instalaciones, se deberá instalar el interruptor
al nivel de la instalación.
1171.0000.42-03.00 Sheet 7
Page 14
Informaciones de seguridad
12. No utilice nunca el producto si está dañado el
cable eléctrico. Compruebe regularmente el
correcto estado de los cables de conexión a
red. Asegure a través de las medidas de
protección y de instalación adecuadas de
que el cable de eléctrico no pueda ser
dañado o de que nadie pueda ser dañado
por él, por ejemplo al tropezar o por un golpe
de corriente.
13. Solamente está permitido el funcionamiento
en redes de distribución TN/TT aseguradas
con fusibles de como máximo 16 A
(utilización de fusibles de mayor amperaje
sólo previa consulta con el grupo de
empresas Rohde & Schwarz).
14. Nunca conecte el enchufe en tomas de
corriente sucias o llenas de polvo. Introduzca
el enchufe por completo y fuertemente en la
toma de corriente. Si no tiene en
consideración estas indicaciones se arriesga
a que se originen chispas, fuego y/o heridas.
15. No sobrecargue las tomas de corriente, los
cables de extensión o los enchufes de
extensión ya que esto pudiera causar fuego
o golpes de corriente.
16. En las mediciones en circuitos de corriente
con una tensión de entrada de U
> 30 V se
eff
deberá tomar las precauciones debidas para
impedir cualquier peligro (por ejemplo
medios de medición adecuados, seguros,
limitación de tensión, corte protector,
aislamiento etc.).
17. En caso de conexión con aparatos de la
técnica informática se deberá tener en
cuenta que estos cumplan los requisitos de
la EC950/EN60950.
18. A menos que esté permitido expresamente,
no retire nunca la tapa ni componentes de la
carcasa mientras el producto esté en
servicio. Esto pone a descubierto los cables
y componentes eléctricos y puede causar
heridas, fuego o daños en el producto.
19. Si un producto es instalado fijamente en un
lugar, se deberá primero conectar el
conductor protector fijo con el conductor
protector del aparato antes de hacer
cualquier otra conexión. La instalación y la
conexión deberán ser efecutadas por un
electricista especializado.
20. En caso de que los productos que son
instalados fijamente en un lugar sean sin
protector implementado, autointerruptor o
similares objetos de protección, el circuito de
suministro de corriente deberá estar
protegido de manera que usuarios y
productos estén suficientemente protegidos.
21. Por favor, no introduzca ningún objeto que
no esté destinado a ello en los orificios de la
caja del aparato. No vierta nunca ninguna
clase de líquidos sobre o en la caja. Esto
puede producir corto circuitos en el producto
y/o puede causar golpes de corriente, fuego
o heridas.
22. Asegúrese con la protección adecuada de
que no pueda originarse en el producto una
sobrecarga por ejemplo a causa de una
tormenta. Si no se verá el personal que lo
utilice expuesto al peligro de un golpe de
corriente.
23. Los productos R&S no están protegidos
contra el agua si no es que exista otra
indicación, ver también punto 1. Si no se
tiene en cuenta esto se arriesga el peligro de
golpe de corriente para el usario o de daños
en el producto lo cual también puede llevar al
peligro de personas.
24. No utilice el producto bajo condiciones en las
que pueda producirse y se hayan producido
líquidos de condensación en o dentro del
producto como por ejemplo cuando se
desplaza el producto de un lugar frío a un
lugar caliente.
25. Por favor no cierre ninguna ranura u orificio
del producto, ya que estas son necesarias
para la ventilación e impiden que el producto
se caliente demasiado. No pongan el
producto encima de materiales blandos como
por ejemplo sofás o alfombras o dentro de
una caja cerrada, si esta no está
suficientemente ventilada.
26. No ponga el producto sobre aparatos que
produzcan calor, como por ejemplo
radiadores o calentadores. La temperatura
ambiental no debe superar la temperatura
máxima especificada en la hoja de datos.
1171.0000.42-03.00 Sheet 8
Page 15
Informaciones de seguridad
27. Baterías y acumuladores no deben de ser
expuestos a temperaturas altas o al fuego.
Guardar baterías y acumuladores fuera del
alcance de los niños. No cortocircuitar
baterías ni acumuladores. Si las baterías o
los acumuladores no son cambiados con la
debida atención existirá peligro de explosión
(atención celulas de Litio). Cambiar las
baterías o los acumuladores solamente por
los del tipo R&S correspondiente (ver lista de
piezas de recambio). Las baterías y
acumuladores deben reutilizarse y no deben
acceder a los vertederos. Las baterías y
acumuladores que contienen plomo,
mercurio o cadmio deben tratarse como
residuos especiales. Respete en esta
relación las normas nacionales de
evacuación y reciclaje.
28. Por favor tengan en cuenta que en caso de
un incendio pueden desprenderse del
producto agentes venenosos (gases, líquidos
etc.) que pueden generar daños a la salud.
29. El producto puede poseer un peso elevado.
Muévalo con cuidado para evitar lesiones en
la espalda u otras partes corporales.
30. No sitúe el producto encima de superficies,
vehículos, estantes o mesas, que por sus
características de peso o de estabilidad no
sean aptas para él. Siga siempre las
instrucciones de instalación del fabricante
cuando instale y asegure el producto en
objetos o estructuras (por ejemplo paredes y
estantes).
31. Las asas instaladas en los productos sirven
solamente de ayuda para el manejo que
solamente está previsto para personas. Por
eso no está permitido utilizar las asas para la
sujeción en o sobre medios de transporte
como por ejemplo grúas, carretillas
elevadoras de horquilla, carros etc. El
usuario es responsable de que los productos
sean sujetados de forma segura a los medios
de transporte y de que las prescripciones de
seguridad del fabricante de los medios de
transporte sean tenidas en cuenta. En caso
de que no se tengan en cuenta pueden
causarse daños en personas y objetos.
32. Si llega a utilizar el producto dentro de un
vehículo, queda en la responsabilidad
absoluta del conductor que conducir el
vehículo de manera segura. Asegure el
producto dentro del vehículo debidamente
para evitar en caso de un accidente las
lesiones u otra clase de daños. No utilice
nunca el producto dentro de un vehículo en
movimiento si esto pudiera distraer al
conductor. Siempre queda en la
responsabilidad absoluta del conductor la
seguridad del vehículo. El fabricante no
asumirá ninguna clase de responsabilidad
por accidentes o colisiones.
33. Dado el caso de que esté integrado un
producto de laser en un producto R&S (por
ejemplo CD/DVD-ROM) no utilice otras
instalaciones o funciones que las descritas
en la documentación de producto. De otra
manera pondrá en peligro su salud, ya que el
rayo laser puede dañar irreversiblemente sus
ojos. Nunca trate de descomponer estos
productos. Nunca mire dentro del rayo laser.
1171.0000.42-03.00 Sheet 9
Page 16
Page 17
Certified Quality System
DIN EN ISO 9001 : 2000
DIN EN 9100 : 2003
DIN EN ISO 14001 : 2004
DQS REG. NO 001954 QM UM
QUALITÄTSZERTIFIKAT CERTIFICATE OF QUALITY CERTIFICAT DE QUALITÉ
Sehr geehrter Kunde,
Sie haben sich für den Kauf eines
Rohde& Schwarz-Produktes entschieden. Hiermit erhalten Sie ein nach
modernsten Fertigungsmethoden
hergestelltes Produkt. Es wurde nach
den Regeln unseres Managementsystems entwickelt, gefertigt und
geprüft.
Das Rohde& Schwarz Managementsystem ist zertifiziert nach:
DIN EN ISO 9001:2000
DIN EN 9100:2003
DIN EN ISO 14001:2004
Dear Customer,
you have decided to buy a Rohde&
Schwarz product. You are thus assured of receiving a product that is
manufactured using the most modern
methods available. This product was
developed, manufactured and tested
in compliance with our quality management system standards.
The Rohde& Schwarz quality management system is certified according to:
DIN EN ISO 9001:2000
DIN EN 9100:2003
DIN EN ISO 14001:2004
Cher Client,
vous avez choisi d‘acheter un produit
Rohde& Schwarz. Vous disposez
donc d‘un produit fabriqué d‘après
les méthodes les plus avancées. Le
développement, la fabrication et les
tests respectent nos normes de gestion qualité.
Le système de gestion qualité de
Rohde& Schwarz a été homologué
conformément aux normes:
DIN EN ISO 9001:2000
DIN EN 9100:2003
DIN EN ISO 14001:2004
1171.0200.11-02.00
Page 18
Page 19
Customer Support
Technical support – where and when you need it
For quick, expert help with any Rohde & Schwarz equipment, contact one of our Customer
Support Centers. A team of highly qualified engineers provides telephone support and will
work with you to find a solution to your query on any aspect of the operation, programming
or applications of Rohde & Schwarz equipment.
Up-to-date information and upgrades
To keep your Rohde & Schwarz equipment always up-to-date,
please subscribe to our electronic newsletter at
or request the desired information and upgrades via email from your Customer Support
Center (addresses see below).
Feedback
We want to know if we are meeting your support needs. If you have any comments please
email us and let us know CustomerSupport.Feedback@rohde-schwarz.com.
USA & Canada
East Asia
Rest of the World
Monday to Friday (except US public holidays)
8:00 AM – 8:00 PM Eastern Standard Time (EST)
Tel. from USA 888-test-rsa (888-837-8772) (opt 2)
From outside USA +1 410 910 7800 (opt 2)
Fax +1 410 910 7801
E-mail Customer.Support@rsa.rohde-schwarz.com
Monday to Friday (except Singaporean public holidays)
8:30 AM – 6:00 PM Singapore Time (SGT)
Tel. +65 6 513 0488
Fax +65 6 846 1090
E-mail Customersupport.asia@rohde-schwarz.com
Monday to Friday (except German public holidays)
08:00 – 17:00
Tel. from Europe +49 (0) 180 512 42 42
From outside Europe +49 89 4129 13776
Fax +49 (0) 89 41 29 637 78
Central European Time (CET)
1171.0200.29-01.00
E-mail CustomerSupport@rohde-schwarz.com
Page 20
Page 21
12
Address List
Headquarters, Plants and Subsidiaries
Headquarters
ROHDE& SCHWARZ GmbH & Co. KG
Mühldorfstraße 15 · D-81671 München
P.O.Box 80 14 69 · D-81614 München
Please refer to our homepage: www.rohde-schwarz.com
◆ Sales Locations
◆ Service Locations
◆ National Websites
12
1171.0200.42-02.00
Page 22
Page 23
R&S FSP-K93 General Information
1General Information
The Rohde & Schwarz FSP-K93 application extends the functionality of the R&S FSP spectrum
analyzer to enable WiMAX and WiBro TX-measurements according to the standard
1. IEEE 802.16-2004/Cor 1-2005, IEEE 802.16e-2005 OFDMA physical layer mode . The short form
‘IEEE 802.16e-2005 OFDMA’ is used in this document, to reference this standard.
2. IEEE 802.16-2004/Cor 1-2005, IEEE 802.16e-2005 based WiBro. The short form ‘WiBro’ is used in
this document, to reference this standard.
The standard is selected in the General Settings Panel, Standard field.
This manual supports the user in working with R&S FSP-K93. It aids the preparation, execution and
evaluation of a measurement and gives many helpful hints and examples.
For the user wanting to make a quick start to using R&S FSP-K93, the Quick Start Guide section below
works step-by-step through an ordinary measurement. The remainder of this section describes all of the
basic information about how the R&S FSP-K93 application works, without covering measurements in
detail. A detailed description of all measurement modes, settings and results can be found in section 2.
Section 4 covers remote control operation of R&S FSP-K93.
This section covers the following subjects:
Introduction to R&S FSP-K93 measurements
• Installation
• Starting the application
• Exiting the application
• Quick start guide – allows the user to get up-and-running in minimum time
The use of an R&S FSP spectrum analyzer with its high sensitivity enables the accurate and reproducible
TX-measurement of a WiMAX Device Under Test (DUT) according to the standards specified for the device:
• Modulation formats:
– IEEE 802.16-2004
- QPSK
- 16QAM
- 64QAM
• Modulation measurements:
– Constellation diagram
– I/Q offset and I/Q imbalance
– Carrier and symbol frequency errors
– Modulation error (EVM) per carrier or symbol
– Amplitude response and group-delay distortion (spectral flatness)
• Amplitude statistics (CCDF) and crest factor
• Transmit spectrum mask
• Adjacent Channel Power (Absolute and Relative)
• FFT, also over a selected part of the signal, e.g. preamble
• Payload bit information
• Capture time selectable up to 50 ms, multiple sweeps possible for large number of bursts
Installation
From the analyzer select firmware update.
Press the SETUP hardkey followed by NEXT, FIRMWARE UPDATE and finally the FIRMWARE
UPDATE softkey. Following the instructions displayed.
Once the installation has completed the analyzer will reboot.
Once the option has been installed it needs to be activated:
Start up the analyzer.
Press the SETUP hardkey, followed by the GENERAL SETUP softkey and then the OPTIONS softkey.
A list of the options currently activated is displayed.
Press the INSTALL OPTION softkey. A Dialog is displayed allowing the option key to be entered.
1308.5523.42 8 E-1
Page 25
R&S FSP-K93 General Information
Enter the option key supplied with the R&S FSP-K93 software.
When a valid option key has been supplied a dialog will be displayed explaining that a reboot is required
to complete this operation. Select OK in this dialog and the instrument will be rebooted.
When the analyzer starts after the reboot a new hotkey will be displayed at the bottom of the display
labelled WIMAX. In addition an entry for the R&S FSP-K93 option will be displayed in the FIRMWARE OPTIONS dialog.
1308.5523.42 9 E-1
Page 26
General Information R&S FSP-K93
Starting the Application
Power up the R&S FSP spectrum analyzer. When R&S FSP-K93 is correctly installed there will be a
hotkey labelled WIMAX at the bottom of the screen. Press the WIMAX hotkey to start R&S FSP-K93.
Note that if the spectrum analyzer is powered down whilst R&S FSP-K93 is active, then when the
spectrum analyzer is powered up again it will start up in the R&S FSP-K93 application.
Exiting the Application
To exit the R&S FSP-K93 option, press the SPECTRUM hotkey at the bottom of the screen. This will
cause the option to exit and the spectrum analyzer to be activated.
1308.5523.42 10 E-1
Page 27
R&S FSP-K93 General Information
Quick Start Guide
This section helps the user to quickly become familiar with R&S FSP-K93 by working step-by-step
through an ordinary measurement. (Refer to section 2 for a detailed reference guide.)
or this example a DUT using IEEE 802.16-2004 will be used. The DUT will be connected to the analyzer
F
using the RF input of the analyzer. The DUT will generate a signal modulated using 64QAM 2/3.
Setting up the Measurement
Start the R&S FSP-K93 application.
Press the GENERAL SETTINGS softkey to open the General Settings view.
Select the Frequency field and enter the desired frequency to measure. Note that when a valid frequency
is entered the Channel No field updates.
Select the Frequency Band in which the Signal to be analyzed is located. The target band is either one of
1
the bands given as example in the 802.16-2004 standard
or an unspecified band.
Select either the Channel Bandwidth (BW) or Sampling Rate (Fs) according to the characteristics of the
2
signal to be analyzed. The second parameter is derived from the first according to the standard
Select the
TTG =CP [Cyclic Prefix] to useful time ratio according to the characteristics of the signal to
bg
.
be analyzed.
Switch Off the Auto Level Field. In this example the level detection measurement will be executed
manually.
All other settings in this view are sufficient for this example.
1
B.3.2 Wireless MAN-OFDM/OFDMA PHY symbol and performance parameters.
2
8.3.2.2 Derived Parameter definitions
1308.5523.42 11 E-1
Page 28
General Information R&S FSP-K93
Press the DEMOD SETTINGS softkey to open the Demod Settings view
Select the Link Mode of the Bursts to be analyzed.
In case Modulation Detection Mode is set to All , All modulation formats of the bursts will be analyzed.
This is useful, to get an overview of the signal content.
Close the Demod Settings by pressing the WIMAX hotkey.
1308.5523.42 12 E-1
Page 29
R&S FSP-K93 General Information
Performing Level Detection
Connect the DUT to the RF input of the spectrum analyzer. (see Fig. 1)
.
.
.
S
PRESET
AL
C
ETUP
S
COPY
H
PECTRUM ANALYZER 9kHz . . . 3GHz
REV
P
N
Fig. 1 Preparation for level detection
EXT
FREQ
S
PAN
MKR
MKR
7
89
4
5
123
0
.
SC
E
ENTER
ANCEL
C
T
RACE
L
INES
D
ISP
FILE
M
6
-
G
EN OUTPUT 50
AX 0V DC
AF OUTPUT
A
MPT
B
W SWEEP
M
KR
M
EAS TRIG
F
CTN
Hz
G
s
V
dBm
-
m
M
Hz
s
Bm
d
mV
>
Hz
k
>
s
B
d
V
n
H
z
s
n
V
B..
d
BACK
KEYBOARDPROBEPOWER
EXT MIXER
R
FINPUT
0
5
MA
+
30 dBm / 50V DC
X
M
ADE IN GERMANY
R
F
INPUT
DUT
Start
the Level Detection measurement by pressing the AUTO LVL hotkey.
During the level detection measurement the text "Running ..." is displayed in the Status Bar at the bottom
of the screen.
After successful level detection, the Status Bar will display “Measurement Complete”, the signal level field
for the selected input will be updated to show the detected signal level and the Magnitude Capture Buffer
(Screen A) will be updated to show the zero span trace obtained during the measurement sequence.
Performing the main measurement
After level detection has been successfully completed the main measurement can be started. The set-up for
the main measurement is the same as that for the level detection measurement.
Start the measurement by pressing the RUN SGL hotkey.
During the measurement, the text "Running..." is displayed in the Status Bar at the bottom of the screen.
Measurement results are updated once the measurement has completed. The results are displayed in
graphical form. The display can be toggled to a tabular list of measurement points by pressing the
DISPLAY Softkey.
1308.5523.42 13 E-1
Page 30
General Information R&S FSP-K93
Navigation
This section deals with navigation within the option. Navigation here is taken to mean all forms of interaction
with the option except for remote control. The different methods of interacting with the option are:
• Hotkeys
• Softkeys
• Hardkeys
• Numeric Keypad
• Roll-key
• Cursor Keys
• External Keyboard
• Mouse
Hotkeys
Hotkeys are allocated to the seven keys at the bottom edge of the screen. On initial start-up of the R&S
FSP-K93 option, the hotkeys provided are shown in Fig. 2. These hotkeys are present at all times once
the option has been started.
SPECTRUM
WIMAX
AUTO LVL
RUN SGLRUN CONT
REFRESH
SCREEN B
Fig. 2 Initial Hotkey menu
A keystroke activates the associated hotkey. An activated hotkey changes colour to green, as shown.
AUTO LVLAUTO LVL
1308.5523.42 14 E-1
Page 31
R&S FSP-K93 General Information
These hotkeys perform the following operations:
SPECTRUM
The SPECTRUM hotkey exits the R&S FSP-K93 option & returns to the spectrum
analyzer with all previous settings restored.
A
RUN SGL
REFRESH
SCREEN [A|B]
IMAX
W
UTO LVL
The WIMAX hotkey returns the user to the main measurement menu of R&S FSP-K93,
where measurement results can be seen. All settings views and dialogs are removed
from the display, and the default softkey menu is displayed
The WIMAX hotkey remains green whenever R&S FSP-K93 is active
The AUTO LVL hotkey starts an automatic level detection measurement.
If another measurement is running then the running measurement will be aborted
before the automatic level detection measurement is started. If a continuous
measurement is running when the AUTO LVL hot-key is pressed then the continuous
measurement shall resume after the automatic level detection has completed
Pressing the AUTO LVL hotkey whilst an automatic level detection measurement is
running causes the measurement to be stopped (aborted).
The RUN SGL hotkey starts the selected measurement in single sweep mode.
If another measurement is running, such as a continuous sweep measurement, the
running measurement will be aborted before the single sweep measurement is started.
Pressing the RUN SGL hotkey whilst a single frequency range measurement is running
causes the measurement to be stopped (aborted).
The RUN CONT hotkey starts the selected measurement in continuous sweep mode
If another measurement is running then the running measurement will be aborted
before the continuous sweep measurement is started.
Pressing the RUN CONT hotkey whilst a continuous sweep measurement is running
causes the measurement to be stopped (aborted).
The REFRESH hotkey updates the current measurement results with respect to the
current gate settings.
The REFRESH hotkey is only available for measurement results which are affected by
the gate settings (Spectrum FFT, PVT and CCDF). The REFRESH hotkey only
becomes available when the gate settings are modified after a measurement result has
been obtained.
The SCREEN [A|B] hotkey selects the specified screen as the active screen.
In full screen mode pressing the SCREEN [A|B] hotkey will display the specified screen.
After Pressing the SCREEN [A|B] hotkey the label displayed in the hotkey is changed
e.g. after pressing the SCREEN A hotkey the label of the hotkey is changed to
SCREEN B. The label indicates which screen will become the active screen after the
hotkey is pressed.
1308.5523.42 15 E-1
Page 32
General Information R&S FSP-K93
Softkeys
Settings Softkeys
The softkeys are assigned to the nine keys on the right-hand side of the display. These enable quick
access to all of the parameter settings and measurement screens of the R&S FSP-K93 option. Each of
the top two softkeys, when pressed, brings up a settings view for a group of parameters. These
softkeys are always available (except when using Save/Recall and Print manager or controlling
markers) and are as follows:
Fig. 3 Main Softkeys
Each of these groups of settings is described in detail in the Measurements & Settings section of this
manual.
1308.5523.42 16 E-1
Page 33
R&S FSP-K93 General Information
Other Softkeys
All other softkeys have different functions depending on the instrument state. Therefore, the labels (text)
on the softkeys will vary to reflect their current function. The state of the softkeys is indicated by
different appearances and colours, as follows:
SOFTKEY
LABEL 1
SOFTKEY
LABEL 2
Softkey available (normal state)
SOFTKEY
LABEL 3
SOFTKEY
LABEL 4
SOFTKEY
LABEL 5
SOFTKEY
LABEL 6
SOFTKEY
LABEL 7
SOFTKEY 8
VAL1
VAL2
Softkey active (green)
Softkey active and dialog
displayed (red)
No softkey available
Softkey disabled => function not available
(without 3D Frame)
Toggle softkey (current value of parameter
highlighted in green)
Fig. 4 Setup of the softkey area
A softkey in its normal state, where its function is available, is coloured grey with a 3D border.
A softkey that is disabled, because its function is not
available, is coloured grey without a 3D border.
Softkeys may become disabled because of the state of the instrument or because other settings disable
the function associated with the softkey.
An active softkey (highlighted in green) is used when the softkey selects an item or view. For example,
the GATE SETTINGS softkey will be highlighted green when the Gate Settings pop-up dialog is
displayed.
A toggle softkey is used to change the value of a parameter that has only two states. Each press of the
softkey toggles the value of the parameter. The current parameter value is highlighted in green in the
lower half of the softkey label. For example, in the measurement results view, the DISPLAY softkey will
have either LIST or GRAPH highlighted in green depending on whether the results are currently
displayed as a list of measurement points or graphical trace(s).
When no function is assigned to a softkey then no softkey label will be shown.
1308.5523.42 17 E-1
Page 34
General Information R&S FSP-K93
Hardkeys
ardkeys allow quick access to the desired parameter and various functions. The hardkeys supported
H
by the R&S-FSP-K93 option are as follows (other hardkeys do nothing):
FREQ Hardkey
AMPT Hardkey When the AMPT hardkey is pressed the General Settings view is displayed (if it is
MKR Hardkey When the MKR hardkey is pressed the main Marker softkey menu is displayed (if it
MKR-> Hardkey When the MKR-> hardkey is pressed the Marker extension softkey menu is
SWEEP Hardkey When the SWEEP hardkey is pressed the General Settings view is displayed (if it is
MEAS Hardkey
TRACE Hardkey
LINES Hardkey
DISP Hardkey
FILE Hardkey
PRESET
Hardkey
HCOPY Hardkey
When the FREQ hardkey is pressed the General Settings view is displayed (if it is
not already being displayed) and the Frequency parameter is selected.
not already being displayed) and the relevant Signal Level parameter for the
selected signal input is selected.
is not already being displayed).
displayed (if it is not already being displayed).
not already being displayed) and the Capture Time parameter is selected.
When the MEAS hardkey is pressed the Main softkey menu is displayed (if it is not
already being displayed).
When the TRACE hardkey is pressed the General Settings view is displayed (if it is
not already being displayed) and the Burst Count parameter is selected.
When the LINES hardkey is pressed the Limit Lines softkey menu is displayed (if it
is not already being displayed) and the first limit in the results summary is selected.
Note this hard-key only functions when the results summary is displayed
When the DISP hardkey is pressed the Display softkey menu is displayed (if it is not
already being displayed).
When the FILE hardkey is pressed, the Save & Recall softkey menu is displayed,
allowing the save & recall of settings and/or measurement results of the R&S FSPK93 option.
When the PRESET hardkey is pressed the R&S FSP-option is exited and a preset
will be performed. Note that all options (including R&S FSP-K93) shall also be
preset.
When the HCOPY hardkey is pressed the print manager softkey menu is displayed,
allowing selection of the items to be printed.
1308.5523.42 18 E-1
Page 35
R&S FSP-K93 General Information
External Keyboard
he external keyboard is optional. The keys on the external keyboard that can be used to interact with
T
the R&S FSP- option are as follows:
Number keys 0 to 9
Decimal point (“.”) Inserts a decimal point “.” at the cursor position.
Minus key(“-”) Changes the sign of the mantissa or exponent of a numeric parameter. A “-“ is
inserted at the cursor position in the case of an alphanumeric parameter.
ESC key Aborts the entry before it has been terminated. The previous value is restored.
Closes the entry field after termination of input.
Closes pop-up dialogs.
ENTER key Terminates the input of dimension quantities. The new value is set.
Invokes the input of parameters or immediately sets the new value.
Selects the highlighted item in drop-down menus.
Left and Right
Cursor Keys are
used to:
Up and Down
Cursor keys are
used to:
CTRL keys Used to activate hotkeys. Each of the seven hotkeys is allocated a different
Navigate between individual parameters within the setting views and some of the
pop-up dialogs.
Navigate between the individual items within drop-down menus.
Move the cursor left & right inside the entry window to reach a particular position
in the string during alphanumeric entry.
Navigate between individual parameters within the setting views and some of the
pop-up dialogs.
Navigate between the individual items within drop-down menus.
Increment or decrement the value of a parameter during numeric entry.
function (F) key. To access these hotkeys press CTRL and the corresponding F
key together (see Fig. 5 ):
Function Keys Used to activate softkeys. Each of the nine softkeys is allocated a different function (F)
key. To access these softkeys the corresponding F key, as shown below:
F1
F2
F3
F4
F5
F6
F7
F8
F9
SOFTKEY 1
SOFTKEY 2
SOFTKEY 3
SOFTKEY 4
SOFTKEY 5
SOFTKEY 6
SOFTKEY 7
SOFTKEY 8
SOFTKEY 9
Fig. 6 Quick access to softkeys
1308.5523.42 19 E-1
Page 36
General Information R&S FSP-K93
Mouse
he mouse can be used to select individual parameters within the settings views or data entry dialogs
T
and to activate hotkeys and softkeys. It can also be used to select values from a drop-down list.
Selecting & Editing Parameters
Parameters are set either by numeric or alphanumeric entry or by simple selection from a list of
possible values (a drop-down list is used to select an “enumerated” value) or by using checkboxes to
turn a parameter setting on and off.
In all cases, the parameter has to be selected by placing focus on it and then editing has to be enabled
before its value can be changed.
The rollkey and cursor keys on the front panel are provided for navigation and selection of parameters.
The numeric keypad, rollkey and cursor keys on the front panel and an external keyboard (optional) are
provided for the entry of parameter values.
Numeric Keypad
The numeric keypad is provided for entry of numeric parameters. It contains
s
89
7
4
56
12
0
.
ESC
CANCEL
ENTER
OK
GHz
V
-dBm
MHz
dBm
the following keys:
m
s
• Number keys 0 to 9
mV
mV
Starts editing of the selected parameter. This enables a new value to be
>s
kHz
3
-
dB
Hz
dB..
BACK
>V
ns
nV
entered for a parameter directly without having to press ENTER first. The digit
will be displayed as the first digit of the newly entered value.
Inserts a digit at the cursor position when editing an alphanumeric parameter.
• Decimal point
Inserts a decimal point “.” at the cursor position.
• Sign key (“-“)
Changes the sign of the mantissa or exponent of a numeric parameter. A “-“ is
inserted at the cursor position when editing an alphanumeric parameter.
• Unit keys (GHz/-dBm, MHz/dBm, kHz/dB and Hz/dB)
• Provides the numeric value entered with the selected unit and sets the
parameter to that value.
The unit keys are all assigned the value “1” for dimensionless quantities or for
level entries (e.g. in dB). The unit keys thus assume the function of an ENTER
key.
• BACK key
Deletes the character to the left of the cursor with alphanumeric entry.
• ESC/CANCEL key
Aborts the entry of a new parameter value. The previous value is restored.
Closes pop-up dialogs.
• ENTER key
Enables editing of the selected parameter (using numeric keys or rollkey).
Finishes the editing of a parameter value. The new value is set.
For an alphanumeric value, the new value is set to that displayed (using the
current unit if applicable).
In a drop-down menu, the parameter is set to the currently selected value in
the list.
1308.5523.42 20 E-1
Page 37
R&S FSP-K93 General Information
Rollkey
The rollkey has various functions:
• In a settings view, the rollkey can be used to navigate between individual
parameters (the parameter selected being highlighted).
• In drop-down menus, the rollkey can be used to navigate between the
individual values for the parameter.
• During numeric entry, the parameter is incremented (by turning
clockwise) or decremented (by turning counter-clockwise) at a defined
step size (depending on the parameter).
• In setting views and data entry dialogs pressing the rollkey invokes the
input of parameters or immediately sets the new value, i.e. pressing the
rollkey is like pressing the ENTER key.
• In drop-down menus, pressing the rollkey selects the relevant item.
Cursor Keys
The keys
• Navigate between individual parameters within the setting views and
some of the pop-up dialogs.
• Navigate between the individual values within drop-down menus.
• Move the cursor left & right inside the entry window to reach a particular
position in the string during alphanumeric entry.
The keys
• Navigate between individual parameters within the setting views and
some of the pop-up dialogs.
• Navigate between the individual items within drop-down menus.
• Increment or decrement the value of a parameter during numeric entry.
and are used to:
and are used to:
Selection of a parameter within a settings view
Selection using rollkey Press GENERAL SETTINGS softkey for example (General settings view is
displayed).
Rotate the rollkey until reaching the required parameter.
Turning the rollkey clockwise selects parameters in the upward direction,
turning it counter-clockwise selects parameters in the downward direction.
Example: Selecting Signal Level (General settings)
Signal Level (RF)
Note: When the Signal Level parameter is selected its label is
highlighted blue.
When the desired parameter is reached press the rollkey to edit the
parameter.
1308.5523.42 21 E-1
Page 38
General Information R&S FSP-K93
Selection using cursor keys
Cursor
,,
Within a list of parameters, the Down and Right both move to the next
item (down) in the list and the Up and Left keys both move to the
previous item (up) in the list.
Within a table of parameters, the cursor keys move the cursor in the
irection indicated.
d
To start editing the parameter, either press the ENTER key on the
numeric keypad, or press the rollkey.
For numeric parameters, editing can also be started by entering the
new value directly from the numeric keypad without pressing the
ENTER key first.
Example: Selecting PDSU Modulation
PDSU Modulation BPSK
or
until obtaining the required parameter.
Selection using mouse
Selection using external
keyboard
ENTER
OK
PDSU Modulation BPSK
BPSK
QPSK
16 QAM
64 QAM
Note: When ENTER is pressed, a drop-down menu is
displayed, which contains all the available settings to
which the Demodulator parameter can be set.
Use the mouse to move the cursor to the parameter and press the left
mouse button to select the parameter.
To start editing the parameter, either press the ENTER key on the
numeric keypad, or press the rollkey.
For numeric parameters, editing can also be started by entering the new
value directly from the numeric keypad without pressing the ENTER key
first.
Select parameter using the cursor keys (in the same way as using the
cursor keys on the front panel).
To start editing the parameter, either press the ENTER key on the
numeric keypad, or press the rollkey.
For numeric parameters, editing can also be started by entering the new
value directly from the numeric keypad without pressing the ENTER key
first.
1308.5523.42 22 E-1
Page 39
R&S FSP-K93 General Information
Entry of a numeric value
Once a parameter has been selected (see above), a new value for a numeric parameter can be entered in a
number of ways. With the exception of entry via the number keys, to start editing the parameter, either press
the ENTER key on the numeric keypad, or press the rollkey before following the instructions below.
If an error occurs, for example, the entered value is out of range, then the new value will not be
accepted for the parameter setting.
Entry using number keys
(numeric keypad)
Example: To enter 200 MHz
Entry using cursor keys
Enter required value using the number keys.
2
0
0
MHz
dBm
Frequency200 MHz
Note: The parameter is not set to the new value until either one
of the unit keys on the numeric keypad, the ENTER or
the rollkey is pressed.
If the new value is not valid, then a message box is displayed and the
entered value will be replaced with a valid value. For example, when a
value above the maximum allowed is entered, then the maximum value
allowed will be shown in the entry box. The parameter will still be ready
for editing so that another value can be entered if desired.
Cursor
or until obtaining the required value.
The application prevents the minimum and maximum values of the
parameter from being exceeded and displays an “Out of range“
message box if attempted.
N.B The cursor keys increment/decrement a parameter value in large
steps.
Example: Cursor down to 100MHz
Frequency100 MHz
Note: Each change of the parameter value takes place
immediately. No other keys need to be pressed.
1308.5523.42 23 E-1
Page 40
General Information R&S FSP-K93
Entry using rollkey
Entry using external
keyboard
Terminating the entry
Rotate the rollkey until reaching the required value.
Turning the rollkey clockwise increases the value, turning it counterclockwise decreases the value.
he application prevents the minimum and maximum values of the parameter
T
from being exceeded and displays an “Out of range“ message box if
attempted.
N.B. The rollkey increment/decrement a parameter value in small steps
Example: Rotate to 200 MHz
Frequency200 MHz
Note: Each change of the parameter value takes place
immediately. No other keys need to be pressed.
Enter value using number keys 0 – 9 in the same way as for using the
number keys on the numeric keypad (see above).
Press one of the unit keys on the numeric keypad.
The unit is entered in the parameter’s edit box and the new parameter
value is set immediately.
Press the ENTER key (on numeric keypad or external keyboard) or press
the rollkey.
Correcting the entry
Aborting the entry
The new parameter value is set immediately.
Note:Pop-up dialogs, where used, do not close automatically.
They can be closed by pressing the ESC key.
In both cases, if the new value is not valid, then a message box is displayed
and the entered value will be replaced with a valid value. For example, when
a value above the maximum allowed is entered, then the maximum value
allowed will be shown in the entry box. The parameter will still be ready for
editing so that another value can be entered if desired.
Position the cursor to the right of the digit which is to be deleted using the
cursor keys
or .
Press the BACK key. The digit to the left of the cursor is deleted.
Enter new digits. Each digit is inserted to the left of the cursor, the other
digits are shifted right.
Press the ESC key during parameter editing.
The original parameter value is restored. The new entry is deleted.
If a pop-up dialog is displayed, press the ESC key again.
The entry window is closed, the original value remains active.
1308.5523.42 24 E-1
Page 41
R&S FSP-K93 General Information
Entry of an enumerated value
Once a parameter has been selected (see above), a new value for an enumerated parameter can be
entered in a number of ways. To start editing the parameter, either press the ENTER key on the
numeric keypad, press the rollkey or left click with the mouse on the drop down button before following
the instructions below.
Note: When the rollkey or ENTER is pressed, a drop-down menu is displayed, which contains all
the available settings that may be selected for the parameter.
Selection of setting using
cursor keys
Cursor
Press ENTER on external keyboard or numeric keypad, or press
rollkey to select the desired setting of parameter.
Note: Currently selected setting of the parameter is highlighted blue.
Pressing ENTER sets the new setting of the parameter
or
immediately.
until obtaining the required setting.
Selection of setting using
rollkey
Selection of setting using
mouse
Selection of setting using
external keyboard
Rotate the rollkey until reaching the required setting.
Press rollkey to select setting.
Example: Select Mode parameter.
Note: Currently selected setting of the parameter is highlighted
blue.
Pressing the rollkey sets the new setting of the parameter
immediately.
When the parameter is selected and ready for editing, select a
new setting using the mouse by left-clicking on the new setting
from the drop-down list. The new setting of the parameter is set
immediately.
Select setting using cursor keys.
Press ENTER to set the parameter to the new value.
1308.5523.42 25 E-1
Page 42
General Information R&S FSP-K93
Entry of a checkbox
A checkbox is used for parameter settings that are either On or Off (Boolean settings). A checkmark ()
appears in the box when the setting is On; the checkbox is empty when the setting is Off.
Once a parameter has been selected (see above), a new value for a Boolean parameter can be entered
in a number of ways. Because Boolean parameters are very simple, it is not necessary to press the
NTER key on the numeric keypad or to press the rollkey in order to edit them.
E
Toggle between the two
states of a checkbox using
rollkey
Toggle between the two
states of a checkbox using
numeric keypad
Press the rollkey to toggle between the two states.
Example: Turn Auto Level setting to Off
press
ENTER
OK
Note: Thecheckbox is empty when the settings is Off
Press the ENTER key to toggle between the two states.
Example: Turn Auto Level setting to On
ENTER
OK
Toggle between the two
states of a checkbox using a
mouse
Toggle between the two
states of a checkbox using
external keyboard
Left-click on the checkbox to toggle between the two states.
Press ENTER to toggle between the two states.
1308.5523.42 26 E-1
Page 43
R&S FSP-K93 General Information
Status Bar & Title Bar
Title Bar
The title bar is visible at the very top of the display when R&S FSP-K93 is active and no settings views
are displayed.
Fig. 7 Title Bar
The centre of the title bar shows the WiMAX standard applicable to the current measurement.
Status Bar
The main status bar is displayed at the bottom of the display, just above the hotkeys.
When a parameter in a settings view is selected, the status bar will display the minimum and maximum,
settings for the selected parameter (see Fig. 8).
MIN:<XX.XX>MAX:<XX.XX>
Fig. 8 Status Bar
When a parameter whose value is enumerated or Boolean in type is selected in any dialog, the status
bar will show “N/A“ displayed for the minimum and maximum, since the minimum and maximum values
are “Not Applicable.”
At other times, the status bar shows the current measurement status along with detailed information
about the progress through any running measurement.
The status bar is also used to display warning and error messages to the user. In order to highlight
these messages, warning messages are displayed with a blue background and error messages with a
red background. Refer to Section 6 for a list of warning and error messages.
1308.5523.42 27 E-1
Page 44
General Information R&S FSP-K93
Save/Recall
This section of the user manual describes the Save/Recall facility of the option.
he FILE hardkey brings up the Save/Recall softkey menu. Any settings views on
FILE
T
display when the save/recall softkey menu is displayed shall be closed.
Fig. 9 Save/Recall softkey menu
The save/recall facility provided by R&S FSP-K93 is exactly the same as that provided by the host
analyzer. Refer to the user manual for the spectrum analyzer for details of the save/recall facility
operation.
The save/recall facility in R&S FSP-K93 provides the following items that can be saved and/or recalled:
• Current Settings
• WIMAX Results
• User Limits
• IQ Data
To close the save/recall softkey menu and return to the main R&S FSP-K93 softkey menu, press the
WIMAX hotkey.
All user settings provided by R&S FSP-K93
All current trace and table results
All limit lines and table limit values.
Allows the raw IQ trace results to be stored. When recalled the data is
reprocessed to generate results
1308.5523.42 28 E-1
Page 45
R&S FSP-K93 General Information
Printing
This section of the user manual describes print facility of the option
H
COPY
The HCOPY hardkey brings up the print softkey menu. Any settings views on display
hen the print softkey menu is displayed shall be closed.
w
Fig. 10 Print softkey menu
The print facility provided by R&S FSP-K93 is exactly the same as that provided by the host analyzer.
Refer to the user manual for the spectrum analyzer for details of the print facility operation.
To close the print softkey menu and return to the main R&S FSP-K93 softkey menu, press the WIMAX
hotkey.
1308.5523.42 29 E-1
Page 46
Measurements & Settings R&S FSP-K93
2Measurements & Settings
This section contains a detailed description of all measurement modes, settings & results. It covers the
following subjects:
• Measurement modes
• Running measurements
• Measurement results
• General settings
• Demodulation settings
• Gate settings
• Marker settings
• General hints about measurements
Measurements
R&S FSP-K93 provides two main measurement types:
• IQ Measurements
• Frequency Sweep Measurements
IEEE 802.16e-2005 WiBro Measurements
Current restrictions3to the IEEE802.16e-2005 WiBro Signal to be analyzed
1. Standard
IEEE 16-2004/Cor1-2005, IEEE 16e-2005, OFDMA Physical Layer Mode, 1 October 2004/28
February 2006.
IEEE 16-2004/Cor1-2005, IEEE 16e-2005 based WiBro.
2. FFT sizes
The FFT sizes 128, 512, 1024 and 2048 are supported.
3. Subframes
a. Either a DL-Subframe or a UL-Subframe can be anaylyzed at one point in time.
b. One DL-PUSC zone per DL-Subframe is supported.
c. One DL-FUSC zone immediately following the first mandatory DL-PUSC zone is supported [DLSubframe].
d. One UL-PUSC zone per UL-Subframe is supported.
e. The Subframes have to be separated by sufficient power off gaps.
4. Segments
The first mandatory DL-PUSC zone must contain exclusively one segment. The assignment of
subchannel groups to this segment is not restricted.
5. Symbols
The signal to be analyzed must contain at least two OFDMA data symbols. More OFDMA data
symbols will improve the accuracy of the measurement results.
6. Burst
Each OFDMA symbol of a zone must contain at least one active burst. I.e. each OFDMA Symbol of
3
Most of these restrictions will be removed in future releases!
1308.5523.42 30 E-1
Page 47
R&S FSP-K93 Measurements & Settings
the zone must comprise subchannels belonging to an active burst. At the moment, the
one/segment must be covered by active bursts. I.e. no gaps at the beginning, mid or end of the
z
zone/segment.
7. UL Subframes
In case of UL-Subframe analysis, the Frame Number must be constant for the signal to be analyzed!
IEEE802.16e-2005 WiBro Measurement Results
For these measurement results, the minimum, mean and maximum is taken over the analyzed zones of
the current capture buffer content.
EVM measurements
• EVM Data and Pilots
EVM over Data and Pilot Symbols for the analyzed zone or the analyzed segment.
• EVM Data
EVM over Data Symbols for the analyzed zone or the analyzed segment.
• EVM Pilots
EVM over Pilot Symbols of the analyzed zone or the analyzed segment.
IQ constellation measurements
• IQ Offset
Power at spectral line 0 normalized to the total transmitted power.
Center Frequency Error
• Difference between measured and reference center frequency
Symbol Clock Error
• Difference between measured and reference symbol clock relative to the system sampling rate.
Power Measurement
• Crest Factor
peak to average power ratio of the analyzed zone
R&S FSP-K93 Graphical OFDMA Measurement Results
Power vs Time measurements
• Capture Buffer
Power profile of the capture buffer data being analyzed
• Full Subframe
min mean max power profile of the subframe containing the zone to be analyzed.
• Rising Falling
min mean max profile of the rising respective falling edge of the subframe containing the zone to be
analyzed.
1308.5523.42 31 E-1
Page 48
Measurements & Settings R&S FSP-K93
EVM measurements
• EVM vs Symbol ALL
EVM versus the symbols of the analyzed zone. In case the capture buffer contains more than one
analyzed zones, the corresponding result graphs will be appended. The min mean max statistics is
performed over carriers.
• EVM vs Carrier ALL
EVM versus physical carriers. The min mean and max statistics is performed over the symbols of all
analyzed zones in the capture buffer.
Spectrum Measurements
• Spectrum Flatness
Average energy of the constellations for the physical carriers. The limit lines are displayed according
to the requirements of the standard. The min mean and max statistic is performed over the analyzed
zones in the capture buffer.
• Spectrum Flatness Difference
Absolute difference of adjacent carriers. The limit lines are displayed according to the requirements
of the standard. The min mean and max statistic is performed over the analyzed zones in the
capture buffer.
• Spectrum IEEE
Spectrum Mask measurement according to the IEEE standard.
• Spectrum ETSI
Spectrum Mask measurement according to the ETSI standard.
• Spectrum ACP ABS/REL
• Spectrum FFT
Constellation Measurements
• Constellation Diagramm
Complex Constellation Diagramm of the modulation symbols. The modulation symbols belong to the
Bursts of the analyzed zone. The different modulation formats are assigned unique colors. These
colors are also used for the Bitstream measurement. With the [CONSTELL SELECTION] Softkey it
is possible to suppress unwanted information.
Statistic Measurements
• CCDF Cumulative Complementary Distribution Function
Complementary Probability Distribution for the capture buffer sample exceeding the average power.
• Bit Stream
The modulation symbols define bit sequences. The mapping is defined in the standard. The
modulation symbols of the bursts from the analyzed zone are displayed. The different modulation
formats are assigned unique colors. These colors are also used for the Constellation Diagram
Measurement. The selection panel allows select highlight the data the user is interested in.
• Burst Summary
Provides information about the bursts from the analyzed zone. I.e. Modulation of the burst, Power
Boosting information, EVM of the burst. Each analyzed zone will produce a set of burst results.
1308.5523.42 32 E-1
Page 49
R&S FSP-K93 Measurements & Settings
SELECTION
*)
GENERAL
SETTINGS
DEMOD
SETTINGS
*)
CCDF
BIT
STREAM
BURST
SUMARY
BIT
GRAPH
LIST1 LIST2
GENERAL
SETTINGS
SETTINGS
LIST1 LIST2
SPECTRUM
CONSTELL
STATISTICS
MANAGER
GATING
ON OFF
GATE
SETTINGS
IMPORT
EXPORT
SUPPORT
GRAPH
LIST1 LIST2
DEMOD
GRAPH
PVT
EVM
FILE
(avilable for FRAME
CONFIG Selecton)
LIST1 LIST2
RECALL
SEND TO
SMU
NEW FOLDER
COPY
PASTE
RENAME
DELETE
FILE
GRAPH
LIST1 LIST2
GRAPH
5
IST ZONE
L
URST
B
OPY
C
ZONE
INSERT
ZONE
NEW
ZONE
NEW
SEGMENT
DELTE
ZONE
FILE
MANAGER
GRAPH
LIST1 LIST2
IST ZONE
L
URST
B
OPY
C
BURST
INSERT
BURST
NEW
ZONE
NEW
BURST
DELTE
BURST
FILE
MANAGER
5
GENERAL
SETTINGS
DEMOD
SETTINGS
*)
CONSTELL
CONSTELL
SELCTION
GATING
ON OFF
GATE
SETTINGS
SEM
SETTINGS
ACP
SETTINGS
IMPORT
EXPORT
SUPPORT
FLAT GRDEL
S
SETTINGS
SUBFRAME
GENERAL
SETTINGS
DEMOD
SETTINGS
*)
FLATNESS
FLATNESS
DIFFERENCE
SPECTRUM
MASK
SPECTRUM
FFT
ACP
ABS REL
ENERAL
G
ETTINGS
EMOD
D
*)
FULL
RISINGS
FALLING
GENERAL
SETTINGS
SETTINGS
EMV VS
SYMBOL
EMV VS
CARRIER
FREQ PHASE
Y-AXIS/
DEMOD
*)
ERROR
DIV
O
ON OFF
GATING
ON OFF
GATE
SETTINGS
SEM
SETTINGS
ACP
SETTINGS
IMPORT
EXPORT
SUPPORT
ATING
G
NOFF
GATE
SETTINGS
IMPORT
EXPORT
SUPPORT
GATING
GATE
SETTINGS
SAME
SETTINGS
EXPORT
SUPPORT
Fig. 11 IEEE 802.16e-2005 WiBro menu
1308.5523.42 33 E-1
Page 50
Measurements & Settings R&S FSP-K93
IQ measurements
The following measurements results are obtained in IQ measurement mode:
• Power vs Time (PVT)
• EVM vs Symbol
• EVM vs Carrier
Phase vs Preamble
•
• Frequency vs Preamble
• Spectrum Flatness
• Spectrum Group Delay
• Spectrum Flatness Adjacent Carrier Power Difference
• Spectrum FFT
• Constellation vs Symbol
• Conditional Cumulative Distribution Function (CCDF)
• Bit Stream
For each of the above results the Magnitude Capture Buffer display is available.
Fig. 12 Magnitude Capture Buffer Results
The Magnitude Capture buffer shows the complete range of captured data for the last sweep. All analyzed
bursts are identified with a green bar at the bottom of the Magnitude Capture buffer. Only those bursts which
matched the required criteria are marked with a green bar.
All IQ measurements process the same signal data and as such all IQ measurement results are available after
a single IQ measurement execution.
IQ measurements can be run in split screen mode (allowing both the Magnitude Capture Buffer Display
and the selected IQ measurement results to be displayed simultaneously) or in full screen mode (with
either the Magnitude Capture Buffer Display or the selected IQ measurement results displayed).
IQ measurements may be performed for RF or Baseband input.
1308.5523.42 34 E-1
Page 51
R&S FSP-K93 Measurements & Settings
Power vs Time (PVT)
PVT
The PVT measurement results are selected by pressing the PVT softkey in the main
measurement softkey menu.
The PVT results display shows measured input data, or over complete bursts displayed
within the gating lines if gating is switched on. The results are displayed as a single
burst.
Pressing the FULL SUBFRAME softkey displays the results in a single graph with all
subframe data being displayed.
Fig. 13 PVT Results – Full Subframe
Remote: CONF:BURS:PVT:SEL FULL
1308.5523.42 35 E-1
Page 52
Measurements & Settings R&S FSP-K93
Pressing the RSINING FALLING softkey displays the results in two separate
graphs, the left hand side showing the start and the right hand side showing the end
of a subframe.
Fig. 14 PVT Results – Rising/Falling
Remote: CONF:BURS:PVT:SEL EDGE
1308.5523.42 36 E-1
Page 53
R&S FSP-K93 Measurements & Settings
EVM vs Symbol
EVM VS
SYMBOL
The EVM vs Symbol measurement results are selected by pressing the EVM softkey in
the main measurement softkey menu followed by the EVM VS SYMBOL Softkey:
Fig. 15 EVM vs Symbol
The EVM vs Symbol results display shows the EVM measured over the full range of
the measured input data. The results are displayed on a per-symbol basis, with blue
vertical lines marking the boundaries of each burst. Note that burst boundary lines are
only displayed if the number of analyzed bursts is less than 250. The minimum,
average and maximum traces are displayed.
For the trace labelled with EVM a commonly used EVM definition is applied, which is the
square root of the momentary error power normalized by the averaged reference power.
For details of this measurements please refer to chapter 3.
Remote: CONF:BURS:EVM:ESYM:IMM
1308.5523.42 37 E-1
Page 54
Measurements & Settings R&S FSP-K93
Y AXIS/
D
IV
Pressing Y AXIS/ DIV softkey displays a pop-up dialog which allows the settings of the
Y-Axis to be controlled.
Fig. 16 EVM vs Symbol - Y-Axis Scaling
Remote: DISP2:TAC1:Y:SCAL:AUTO
EVM vs Symbol Y Axis
Auto Scaling
Per Division
Unit
DISP2:TRAC1:Y:SCAL:PDIV
Auto Scaling specifies whether the Y-Axis scale should be calculated
manually or automatically.
When the Auto Scaling parameter is switched OFF then the setting of
the Per Division setting is used to calculate the Y-Axis scale.
Remote: DISP2:TAC1:Y:SCAL:AUTO
Per Division specifies the scaling is to be used when Auto Scaling is
switched OFF.
Remote: DISP2:TRAC1:Y:SCAL:PDIV
Unit allows Y-Axis unit to be specified.
1308.5523.42 38 E-1
Page 55
R&S FSP-K93 Measurements & Settings
EVM vs Carrier
EVM VS
CARRIER
The EVM vs Carrier measurement results are selected by pressing the EVM softkey
in the main measurement softkey menu followed by the EVM vs CARRIER softkey.
Y AXIS/
DIV
Fig. 17 EVM vs Carrier Results
The EVM vs Carrier results display shows the all EVM values recorded on a percarrier basis over the full set of measured data. The minimum, average and
maximum traces are displayed.
The scaling of the Y-Axis can be modified to allow the results to be scaled to an
optimum level.
Remote: CONF:BURSt:EVM:ECAR:IMM
Pressing Y AXIS/ DIV softkey displays a pop-up dialog which allows the settings of
the Y-Axis to be controlled. The settings provided are the same as for the EVM vs
Symbol measurement screen.
Remote: DISP:WIND2:TRAC:Y:SCAL:AUTO ON
DISP:WIND2:TRAC1:Y:SCAL:PDIV 2
1308.5523.42 39 E-1
Page 56
Measurements & Settings R&S FSP-K93
Error vs Preamble
ERROR
REQ PHASE
F
The Error vs Preamble results are selected by pressing the EVM softkey in the main
measurement softkey menu followed by the ERROR FREQ/PHASE softkey. Subsequent
presses of the ERROR FREQ/PHASE softkey toggle the Error measurement between
Frequency Error vs Preamble and Phase Error vs Preamble.
Y AXIS/
DIV
Fig. 18 Error vs Preamble Results
The Error vs Preamble results display shows the error values recorded over the
preamble part of the burst. A minimum, average and maximum trace are displayed.
The results display either relative frequency error or phase error.
The scaling of the Y-Axis can be modified to allow the results to be scaled to an
optimum level.
Remote: CONF:BURS:PRE:SEL FREQ
CONF:BURS:PRE:IMM
Pressing Y AXIS/ DIV softkey displays a pop-up dialog which allows the settings of
the Y-Axis to be controlled. The settings provided are the same as for the EVM vs
Symbol measurement screen.
Remote: DISP:WIND1:TAC1:Y:AUTO
DISP:WIND1:TRAC1:Y:PDIV
1308.5523.42 40 E-1
Page 57
R&S FSP-K93 Measurements & Settings
Spectrum Flatness/Spectrum Flatness Group Delay
FLATNESS
FLAT GRDEL
he Spectrum Flatness measurement results are selected by pressing the
T
SPECTRUM softkey in the main measurement softkey menu followed by the
FLATNESS FLAT/GRDEL softkey. Subsequent presses of the FLATNESS
FLAT/GRDEL softkey toggle the Flatness measurement between Spectrum Flatness and
Spectrum Flatness Group Delay.
Fig. 19 Spectrum Flatness Results
The Spectrum Flatness results display shows either the Spectrum Flatness or the
Group Delay values recorded on a per-carrier basis over the full set of measured data.
A minimum, average and maximum trace are displayed for each of the result types.
When Spectrum Flatness results are selected an upper and lower limit line
representing the limits specified for the selected standard are displayed. An overall
pass/fail status is displayed for the obtained (average) results against these limit lines.
Remote: CONF:BURS:SPEC:FLAT:SEL FLAT
CONF:BURS:SPEC:FLAT:IMM
1308.5523.42 41 E-1
Page 58
Measurements & Settings R&S FSP-K93
Flatness Difference
FLATNESS
DIFFERENCE
The Flatness Difference results are selected by pressing the EVM softkey in the main
measurement softkey menu followed by the FLATNESS DIFFERENCE softkey.
Fig. 20 Flatness Difference Results
The Flatness Difference results display shows the adjacent carrier power difference
the preamble part of the burst. A minimum, average and maximum trace are
displayed.
Remote: CONF:BURS:SPEC:FLAT:SEL DIFF
CONF:BURS:SPEC:FLAT:IMM
1308.5523.42 42 E-1
Page 59
R&S FSP-K93 Measurements & Settings
Spectrum FFT
SPECTRUM
FFT
The Spectrum FFT measurement results are selected by pressing the SPECTRUM
softkey in the main measurement softkey menu followed by the SPECTRUM FFT
softkey.
Fig. 21 Spectrum FFT Results (example)
The Spectrum FFT results display shows the Power vs Frequency results obtained
from a FFT performed over the range of data in the Magnitude Capture buffer which
lies within the gate lines. If the gate start or gate length are altered then the results can
be updated to reflect these changes by pressing the REFRESH hotkey.
Remote: CONF:BURS:SPEC:FFT:IMM
1308.5523.42 43 E-1
Page 60
Measurements & Settings R&S FSP-K93
Constellation vs Symbol
CONSTELL
The Constellation diagram is selected by pressing the CONSTELL softkey in the main
measurement softkey menu.
Fig. 22 Constellation vs Symbol Results (IEEE 802.16-2004)
The Constellation display shows the Inphase and Quadrature phase results over the
full range of the measured input data. The ideal points for the selected modulations
scheme are displayed for reference purposes.
The amount of data displayed in the Constellation results display can be reduced by
selecting the carrier or carriers for which data is to be displayed.
Remote: CONF:BURS:CONS:CSYM:IMM
1308.5523.42 44 E-1
Page 61
R&S FSP-K93 Measurements & Settings
Pressing CONSTELL SELECTION softkey displays a pop-up dialog which allows
he displayed results to be filtered. The results may be filtered by any combination
t
of modulation, burst, symbol or carrier. The results will be updated as soon as any
change to the constellation selection parameters are made.
Fig. 23 Constellation vs Symbol - Constell Section
Conditional Cumulative Distribution Function (CCDF)
C
CDF
The CCDF measurement results are selected by pressing the STATISTICS softkey in
the main measurement softkey menu followed by the CCDF softkey.
Fig. 24 CCDF Results
The CCDF results display shows the probability of an amplitude within the gating
lines exceeding the mean power measured between the gating lines. The X-axis
displays power relative to the measured mean power. If the gate start or gate length
are altered then the results can be updated to reflect these changes by pressing the
REFRESH hotkey.
Remote: CONF:BURS:STAT:CCDF:IMM
1308.5523.42 46 E-1
Page 63
R&S FSP-K93 Measurements & Settings
Bit Stream
B
IT
STREAM
The Bit Stream measurement results are selected by pressing the STATISTICS softkey in
the main measurement softkey menu followed by the BIT STREAM softkey.
Fig. 25 Bit-Stream Results
The bit stream results display shows the demodulated data stream. These results are
grouped by burst and symbol. When no other dialogs are on display the results can be
scrolled through using the cursor keys or scroll wheel.
Remote: CONF:BURS:STAT:BSTR:IMM
1308.5523.42 47 E-1
Page 64
Measurements & Settings R&S FSP-K93
Pressing BIT SELECTION softkey displays a pop-up dialog which allows the
isplayed results to be filtered. The results may be filtered by any combination of
d
modulation, symbol or burst. The results will be updated as soon as any change to
the bit selection parameters are made.
Fig. 26 Bit-Stream - Bit Selection
Remote: ---
1308.5523.42 48 E-1
Page 65
R&S FSP-K93 Measurements & Settings
Burst Summary
BURST
SUMMARY
The Burst Summary measurement results are selected by pressing the STATISTICS
softkey in the main measurement softkey menu followed by the BURST SUMMARY
softkey.
Fig. 27 Burst-Summary Results
The burst summary results display shows the summary data for the analyzed bursts.
When no other dialogs are on display the results can be scrolled through using the
cursor keys or scroll wheel.
Remote: CONF:BURS:STAT:BSUM:IMM
1308.5523.42 49 E-1
Page 66
Measurements & Settings R&S FSP-K93
Frequency sweep measurements
he following measurements results are obtained in frequency sweep mode:
T
• Spectrum Emission Mask
• Spectrum ACP/ACPR
The frequency sweep measurements use different signal data to IQ measurements and as such it is not
possible to run an IQ measurement and then view the results in the frequency sweep measurements and
vice-versa. Also because each of the frequency sweep measurement use different settings to obtain signal
data it is not possible to run a frequency sweep measurement and view the results in another frequency
sweep measurement.
All frequency sweep measurements are run in full screen mode. Frequency sweep measurements are
only available when RF input is selected.
Spectrum Mask
The Spectrum Emission Mask measurement results are selected by pressing the
SPECTRUM softkey in the main measurement softkey menu followed by the
SPECTRUM MASK softkey.
Fig. 28 Spectrum Mask Results (example)
The Spectrum Emission Mask results display shows power against frequency. A limit
line representing the spectrum mask is displayed and an overall pass/fail status is
displayed for the obtained results against this limit line. Above the graph is a table
displaying the results for each of the individual frequency ranges.
For each of the specified frequency ranges the highest power value is marked on the
trace with a marker.
Remote: CONF:BURS:SPEC:MASK:SEL IEEE | ETSI
CONF:BURS:SPEC:MASK:IMM
The Spectrum Emission Mask measurement can be configured from the SEM
settings view:
1308.5523.42 50 E-1
Page 67
R&S FSP-K93 Measurements & Settings
The SEM SETTINGS softkey brings up the SEM Settings view.
SEM according to specifies how the Spectrum Emission
Mask settings and limits are applied. This parameter
provides the following settings:
TTA Standard – Settings and limits are as specified in the
standard (WiBro only)
User – Settings and limits are configured via an XML file
When User settings are specified, File Name shows the
name of the loaded XML file. Clicking the arrow switches to
the File Manager to locate an XML file, and automatically
selects SEM According To:User.
When using TTA/ETSI/IEEE standards, File Name reflects
the name of the built-in configuration.
1308.5523.42 51 E-1
Page 68
Measurements & Settings R&S FSP-K93
Link Direction
Link Direction allows the settings and limits for the relevant
link direction (Uplink or Downlink) to be displayed.
Power Class
The SEM settings allow different set of limits depending on
the signal level. By default the Power Class is set to Auto,
which selects the limits automatically based on the
measured signal level. If required, the Power Class can be
set by hand, to force a set of limits to be used.
SEM Configuration
The SEM configuration shows the settings and limits applied over specified frequency ranges around
the TX channel. The settings displayed are dependent on the selected Link Direction and Power Class.
Fig. 30 SEM Configuration
1308.5523.42 52 E-1
Page 69
R&S FSP-K93 Measurements & Settings
Spectrum ACP
ACP
RELABS
The Spectrum ACP measurement results are selected by pressing the SPECTRUM
softkey in the main measurement softkey menu followed by the ACP REL/ABS
softkey. Subsequent presses of the ACP REL/ABS softkey toggle the measurement
between relative and absolute results display.
Fig. 31 Spectrum ACPR Results (example)
The Spectrum ACPR (Adjacent Channel Power (Relative)) is similar to the Spectrum
Mask measurement, and provides information about leakage into adjacent channels.
The results show the relative power measured in the three nearest channels either side
of the measured channel. This measurement is the same as the Adjacent Channel
Power measurement provided by the Spectrum Analyzer. . If the Sweep Count (Mask/ACP) parameter in the General Settings view is set to any value other than 1
then the measurement is performed over the specified number of sweeps. When the
measurement is performed over multiple sweeps a max hold trace is displayed as well
as an average trace.
By default the ACP Settings are derived from the Channel Bandwidth setting of the
General Settings Panel.
Remote: SENS:POW:ACH:MODE REL | ABS
CONF:BURS:SPEC:ACPR:IMM
The Spectrum ACP measurement can be configured from the ACP settings view:
1308.5523.42 53 E-1
Page 70
Measurements & Settings R&S FSP-K93
ACP
ETTINGS
S
The ACP SETTINGS softkey brings up the ACP Settings view.
Fig. 32 ACP Settings view
Noise Correction
No of Channels
Noise Correction - when selected enables noise reduction
to be performed on the signal.
Remote: SENS:POWer:NCOR
No of Channels specifies then number of channels adjacent
to the transmit channel to be measured.
If No of Channels is set to 0 then only the transmit channel
is measured.
Remote: SENS:POW:ACH:ACP
1308.5523.42 54 E-1
Page 71
R&S FSP-K93 Measurements & Settings
TX/ACP Channel BW
he TX/ACP Channel BW settings specify the bandwidth of specific channels.
T
TX Channel Bandwidth
TX Channel Bandwidth specifies the bandwidth of the
transmit channel to be measured by the ACP
measurement.
Adjacent Channel Bandwidth
Alternate 1,2,3 & 4 Channel Bandwidth
Remote:
Adjacent Channel Bandwidth specifies the bandwidth of the
first channel before and after the transmit channel to be
measured.
Remote:
SENS:POW:ACH:BAND:CHAN
SENS:POW:ACH:BAND:ACH
Adjacent Channel Bandwidth specifies the bandwidth of the
specified channel before and after the transmit channel to
be measured.
Remote:
SENS:POW:ACH:BAND:ALT<1..11>
ACP Channel Spacing
The ACP Channel Spacing settings specify the spacing of specific channels.
Adjacent Channel Spacing
Adjacent Channel Bandwidth specifies the spacing of the
first channel before and after the transmit channel to be
measured.
Alternate 1,2,3 & 4 Channel Spacing
Remote:
Adjacent Channel Bandwidth specifies the spacing of the
specified channel before and after the transmit channel to
be measured.
Remote:
SENS:POW:ACH:SPAC:ACH
SENS:POW:ACH:SPAC:ALT<1..4>
1308.5523.42 55 E-1
Page 72
Measurements & Settings R&S FSP-K93
Automatic Level Detection
utomatic level detection allows the Signal Level and Reference Level settings to be calculated from
A
the input signal.
Automatic level detection can be performed in one of two ways:
Pressing the AUTO LVL hot-key
Setting Auto Level in the General Settings view to ON.
Running the Automatic Level Detection by pressing the AUTO LVL hot-key allows the level detection to
be run at any time. Once the Automatic Level Detection has completed the trace is displayed in the
Magnitude Capture Buffer display.
Running the Automatic Level Detection by Setting Auto Level in the General Settings view to ON means
that an Automatic Level Detection measurement is executed before each main measurement sweep.
When running the Automatic Level Detection this way no trace for the Automatic Level Detection
measurement is displayed. This way of running the Automatic Level Detection will cause measurements
to run more slowly and as such it is recommended that Automatic Level Detection is only run in this way
when the speed of measurement is not important or when running continuous measurements are being
run with a signal which varies in level from one sweep to the next.
1308.5523.42 56 E-1
Page 73
R&S FSP-K93 Measurements & Settings
Running measurements
To start a measurement, press the RUN SGL hotkey (single) or RUN CONT hotkey (continuous).
A single measurement will complete once the requested number of bursts have been obtained or a
single sweep has been completed, depending on the measurement settings. When a measurement is
completed in continuous mode then a new measurement will be started.
Note that if one measurement is started whilst another measurement is in progress, for example, a
single measurement is started whilst a continuous measurement is in progress, then the first
measurement will be aborted and the new measurement started immediately.
During a measurement, the text "Running..." is displayed in the Status Bar at the bottom of the screen.
After successful completion of a single measurement, the Status Bar will display “Measurement Complete”.
1308.5523.42 57 E-1
Page 74
Measurements & Settings R&S FSP-K93
Measurement results
The tabular section below the title bar shows the overall measurement settings used to obtain the
current measurement results.
The settings summary includes the following information:
• Frequency
• N
• Zone/Seg
• Signal Level
• Sweep Mode
• Modulation
• Ref Level
• External Att
• Trigger Mode/Offset
FT
F
The frequency of the measured input signal.
Shows the FTT size. IEEE 802.16e – 2005 WiBro only
The zone and segment being analyzed. IEEE 802.16e – 2005 WiBro.
The expected mean signal level for the input signal.
Shows “Single” for a single measurement and “Continuous” for a continuous
measurement.
The Demodulator setting from the Demod Settings panel is shown. In case the
Modulation Analysis Scope is set to ALL, ALL is displayed.
The internal instrument reference level used for the measurement.
The attenuation (positive values) or gain (negative values) applied to the
signal externally (i.e. before the RF- or IQ- connector of the spectrum
analyzer).
e.g.
External Att = 10 dB means that before the RF-connector of the R&S a 10 dB
attenuator is used
External Att = -20 dB means that before the RF-connector of the R&S FSP a
amplifier with 20 dB gain is used
A combined display of the trigger source and any associated offset applied.
IEEE 802.16e – 2005 WiBro
• Zone Offset/Length
1308.5523.42 58 E-1
A combined display of the offset and length of the analyzed zone. IEEE
802.16e – 2005 WiBro.
Page 75
R&S FSP-K93 Measurements & Settings
Results Summary
he results summary table is displayed for IQ measurements when the display mode is set to List. This
T
table shows the overall measurement results and provides limit checking for result values in accordance
with the selected standard.
Result values which are within the limit as specified by the standard are displayed in green. Result
values which are outside of the limits specified by the standard are displayed in red with a ‘*’ to the left.
Results which have no limits specified by the standard are displayed in white. Limit values are displayed
in white (not bold).
The results for WiBro standard are displayed on two tables. The first table is displayed by pressing the
DISPLAY LIST 1 softkey.
When this table is displayed then the softkey text changes to DISPLAY LIST 2. Pressing this softkey
displays the second results table.
When this table is displayed the softkey text is changed to DISPLAY GRAPH. Pressing the softkey
again will return the user back to the trace display.
First WiBro results table. This table displays results that are related to the subframes of the analysed
signal.
Fig. 34 Subframe Result Summary Table for WiBro
1308.5523.42 59 E-1
Page 76
Measurements & Settings R&S FSP-K93
Second WiBro results table. This table displays results that are related to the Zones/Segments of the
nalysed signal.
a
Fig. 35 Zone/Segment Result Summary Table for WiBro
For more details regarding how the results are calculated see section 3.
1308.5523.42 60 E-1
Page 77
R&S FSP-K93 Measurements & Settings
The limit values in the table of results can be modified.
LINES
Pressing the LINES hardkey allows limits to be selected and modified
Fig. 36 Editing Limit Values
Limits are modified for the currently selected modulation scheme. Each modulation scheme may have
its own set of user defined limits.
DEFAULT
CURRENT
Pressing the DEFAULT CURRENT softey resets all limits for the current
modulation scheme to those values specified in the selected standard.
The results displayed in this table are for the entire measurement. If a specific number of bursts has
been requested which requires more than one sweep then the results summary update at the end of
each sweep shows the results so far. The number of bursts measured and the number of bursts
requested are displayed to show the progress through the measurement.
If more than one burst is evaluated (several analyzed bursts in the capture buffer or with the help of
Overall Burst Count), the Min / Mean / Max columns show the minimum, mean or maximum values of
the burst results.
1308.5523.42 62 E-1
Page 79
R&S FSP-K93 Measurements & Settings
General Settings
This section of the user manual describes the General Settings view where all settings related to the
overall measurement can be modified, that is the Signal Characteristics, Data Capture, Trigger IQ, Input
and Advanced settings.
GENERAL
SETTINGS
The GENERAL SETTINGS softkey brings up the General Settings view.
Fig. 37 General Settings view for WiBro standard
The parameters within the General settings view are logically grouped together into:
• Signal Characteristics
• Level Settings
• Data Capture Settings
• Trigger Settings
• IQ Settings
• Input Settings
• Advanced Settings
Any parameters that are not available for editing will have a grey background. This
usually occurs when one parameter setting makes another parameter invalid, for
example if the Trigger Mode is Free Run then none of the parameters below this
(Trigger Offset, Power Level and Auto Power Trigger Level) have any meaning, so
these parameters are greyed out.
When a particular parameter is selected within the General Settings view the status bar
changes to display information about the valid settings for the selected parameter.
1308.5523.42 63 E-1
Page 80
Measurements & Settings R&S FSP-K93
Signal Characteristics
he Signal Characteristics settings are the general settings concerning the physical attributes of the
T
signal to be measured.
Standard
Standard is used to select the WiMAX standard to be
measured against. This is necessary to ensure that the
measurements are performed according to the specified
standard with the correct limit values and limit lines.
This parameter provides a list of settings. The list provided
depends on the standards which have been installed:
For R&S FSP-K93 the only available Standard is IEEE
802.16e-2005 WiBro.
Remote: CONF:STAN
Frequency
Frequency Band
FFT Size N
FFT
Frequency specifies the Center Frequency of the signal to be
measured.
Remote: SENS:FREQ:CENT
Frequency Band specifies the relationship between the
Channel Bandwidth and the Sample Rate parameters.
For R&S FSP-K93 the Frequency Band is always "Unspecified".
Channel Bandwidth specifies the bandwidth of the channel to
be measured.
For R&S FSP-K93 the Channel Bandwidth is fixed at 8.75 MHz.
Remote: SENS:BAND:CHAN 1.75MHZ
Fs/BW acc. to Standard
The FS/BW setting allows a user-defined Channel Bandwidth
to Sampling Rate relationship.
For R&S FSP-K93 the ratio is fixed at: n = 8/7.
IEC/IEEE-bus command:
CONF:WIM:FSBW:AUTO ON
CONF:WIM:FSBW:FS 2.5
CONF:WIM:FSBW:BW 1.5
Sample Rate
G = Tg/Tb
Sample Rate specifies the sample rate used for IQ
measurements.
For R&S FSP-K93 the Sampling Rate is fixed at 10 MHz.
Note: IQ data is captured at 32 MHz, and reduced to 10 MHz
Remote: TRAC:IQ:SRAT 20000HZ
G = Tg/Tb specifies the guard time ratio
For R&S FSP-K93 the ratio is fixed at 1/8.
Remote: CONF:WIM:IGR 4
during signal analysis.
1308.5523.42 65 E-1
Page 82
Measurements & Settings R&S FSP-K93
Level Settings
he level settings are the general settings concerning the level and attenuation of the signal to be
T
measured.
Auto Level
Auto Level selects whether the reference level for
measurements is measured automatically (ON) or entered
manually by the user (OFF).
When Auto Level is set to ON, R&S FSP-K93 will measure
the reference level automatically at the start of each
measurement sweep. This ensures that the reference level is
always set at the optimal level for obtaining accurate results
but will result in slightly increased measurement times.
Remote: CONF:POW:AUTO ON
Signal Level (RF)
Signal Level (RF) specifies the expected level of the RF
input signal.
The Signal Level (RF) is updated after an automatic level
detection measurement has been executed when RF input
is selected.
The Signal Level (RF) parameter is not editable when Base
band input is selected because Signal Level (RF) only
make sense for an RF signal.
Remote: CONF:POW:EXP:RF -331
Signal Level (Baseband)
Ext Att
Signal Level (Baseband) specifies the expected level of the
Baseband input signal.
The Signal Level (Baseband) is updated after an automatic
level detection measurement has been executed when
Baseband input is selected.
The Signal Level (Baseband) parameter is not editable
when RF input is selected because Signal Level (Baseband) only make sense for a Baseband signal.
Remote:
Ext Att specifies the external attenuation or gain applied to
the RF signal. A positive value indicates attenuation, a
negative value indicates gain.
All displayed power level values will be shifted by this value.
Remote:
DISP:WIND:TRAC:Y:RLEV:IQ 1
DISP:WIND:TRAC:Y:RLEV:OFFS 0
1308.5523.42 66 E-1
Page 83
R&S FSP-K93 Measurements & Settings
Data Capture Settings
he Data Capture settings specify how much data is to be captured and measured.
T
Capture Time
Capture Time specifies the time (and therefore the amount of
data) to be captured in a single measurement sweep.
Remote: SENS:SWE:TIME
Capture Count
Capture Count specifies whether a specified number of
subframes are to be captured and analyzed.
When Capture Count is set to OFF then data analysis shall be
performed on a single measurement sweep. When Capture Count is set to ON then data analysis may be performed over a
number of consecutive sweeps until the required number of
subframes have been captured and analyzed.
Remote: SENS:ZONE:COUN:STAT ON
Subframes to Analyze
Sweep Time
Subframes to Analyze specifies the number of subframes to
be measured.
If the number of subframes specified are not contained in a
single measurement sweep then measurement sweeps will
continue to be performed until the requested number of
subframes have been captured.
The Subframes to Analyze parameter is not editable when Capture Count is set to OFF.
Remote: SENS:ZONE:COUN?
Sweep Time specifies whether the sweep time for the
spectrum ACP & Mask measurements should be calculated
automatically or specified manually.
If the sweep time auto check box is switched off the sweep
time to be used can be specified.
Remote: SENS:SWE:ACPR:TIME 2s
Sweep Count
Sweep Count specifies the number of sweeps which should
be performed for Spectrum ACP/ACPR and Spectrum Mask
measurements.
Remote: SENS:SWE:COUNt 1
1308.5523.42 67 E-1
Page 84
Measurements & Settings R&S FSP-K93
Trigger Settings
he Trigger settings group contains all the settings related to the triggering of a measurement sweep.
T
Trigger Mode
Trigger Mode is the source of the trigger for the measurement
sweep.
The possible values for the Trigger Mode are:
Free Run the measurement sweep starts immediately.
External the measurement sweep starts when the external
trigger signal meets or exceeds the specified
external trigger level.at the input connector EXT TRIGGER/GATE on the rear panel.
Power the measurement sweep starts when the signal
power meets or exceeds the specified power trigger
Remote: TRIG:SEQ:MODE IMM
Trigger Offset
level.
External Trigger Level
Auto Power Trigger Level
Trigger Offset specifies the time offset between the trigger signal
and the start of the sweep. A negative value indicates a pre-trigger.
The Trigger Offset parameter is not editable when Trigger Mode is set to Free Run because this indicates that the
measurement sweep should trigger immediately and as such
a trigger delay or pre-trigger would not be appropriate.
Remote: TRIG:SEQ:HOLD -10us
External Trigger Level specifies the trigger level when a
external trigger is selected.
The External Trigger Level parameter is editable only when
Trigger Mode is set to External.
Remote: TRIG:SEQ:LEV 1.4V
Auto Power Trigger Level selects whether the level for the
power trigger is measured automatically (ON) or entered
manually by the user (OFF).
When Auto Power Trigger Level is set to ON, R&S FSP-K93 will
measure and determine the power trigger level automatically at
the start of each measurement sweep. This ensures that the
power trigger level is always set at the optimal level for obtaining
accurate results but will result in a slightly increased
measurement times.
The Auto Power Trigger Level parameter is editable only
when Trigger Mode is set to Power.
Remote: TRIG:SEQ:LEV:POW:AUTO ON
1308.5523.42 68 E-1
Page 85
R&S FSP-K93 Measurements & Settings
Trigger Level (RF)
Trigger Level(RF) Specifies the trigger level when a power
trigger is selected and Baseband input is not selected.
The Trigger Level parameter is editable only when Trigger Mode is set to Power.
The Trigger Level (RF) parameter is not editable when
Baseband input is selected because Trigger Level (RF) only
makes sense for an RF signal.
Remote: TRIG:SEQ:LEV:IFP 0DBM
Trigger Level (Baseband)
Trigger Level (Baseband) Specifies the trigger level when a
power trigger is selected and Baseband input is selected.
The Trigger Level parameter is editable only when Trigger Mode is set to Power.
The Trigger Level (Baseband) parameter is not editable
when RF input is selected because Trigger Level (Baseband) only makes sense for a Baseband signal.
Remote: TRIG:SEQ:LEV:IFP 500mV
Ext.Trigger Lvl
Ext. Trigger Lvl specifies the trigger level when an external
trigger is selected.
The Ext. Trigger Level parameter is editable only when
Trigger Mode is set to External.
Remote: TRIG:SEQ:MODE EXT 1.4V
IQ Settings
The IQ settings contains settings related the Inphase and Quadrature phase of the input signal.
Swap IQ
Swap IQ allows selection between normal and inverted I/Q
modulation. The settings for this parameter are:
OFF Normal I/Q modulation.
ON I and Q signals are interchanged.
The Swap IQ parameter is editable only when IQ Path is set
to I + j*Q.
Remote: SWAP ON
1308.5523.42 69 E-1
Page 86
Measurements & Settings R&S FSP-K93
Advanced Settings
he Advanced Settings contains settings which control details of how the instrument operates and how
T
measurements are performed.
The advanced settings are displayed on the right hand side of the general settings view, and are only
displayed when the Advanced Settings checkbox is checked.
Fig. 39 Advanced Settings for WiBro
Input Settings
The Input settings group contains settings related to the input source of the signal to be measured. Note
that this group is only available when the B71 option is installed.
Auto Level
Auto Level selects whether the reference level for
measurements is measured automatically (ON) or entered
manually by the user (OFF).
The Auto Level advanced settings is the same as the Auto Level in the Signal Characteristics. It is repeated in the
Advanced Settings for convenience.
Remote: SENS:SWE:ACPR:TIME:AUTO OFF
1308.5523.42 70 E-1
Page 87
R&S FSP-K93 Measurements & Settings
Auto Track Time
Auto Track Time specifies the sweep time used for the auto
level measurements.
Remote: CONF:POW:AUTO:SWE:TIME 100MS
Ref Level
Ref Level specifies the reference level to use when running
measurements.
When the Ref Level is modified the Signal Level parameter is
updated accordingly (depending on the currently selected
standard and measurement type).
The Ref Level is editable only when Auto Level is switched off.
Remote:
DISP:WIND:TRAC:Y:SCAL:RLEV?
RF Att
El Att
Rf Att specifies the mechanical attenuation to be applied to the
input RF signal.
Remote:
The El Att parameters specify the settings for the electronic
attenuator.
The El Att parameters are:
Mode Specifies whether the electronic attenuator should be
State Specifies whether the electronic attenuator should be
Settings Specifies the attenuation value the electronic
When the mode is set to Auto then the state and settings fields
will be automatically calculated by the option and so these
fields will not be editable.
When the state is switched off the setting field is non-editable.
When the frequency is set such that the electronics attenuator
cannot be used then all the Electronic Attenuator fields are readonly.
Remote:
INP:ATT 5DB
controlled manually (Manual) or by the option (Auto)
switched ON or OFF.
attenuator should be switched to.
INP:EATT:AUTO ON
1308.5523.42 71 E-1
Page 88
Measurements & Settings R&S FSP-K93
RSSI, CINR avg parameter
RSSI, CINR avg parameter specifies the mean and standard
deviation for the RSSI and CINR results
See [IEEE Std 802.16-2004] “8.3.9.2 RSSI mean and standard
eviation”, “8.3.9.3 CINR mean and standard deviation”.
d
Remote: CONF:WIMAX:AVER
List Results Unit
List Results Unit specifies the units for the results in the
results summary table.
Remote: UNIT:TABL DB | PCT
SMU Address
SMU Address specifies the IP address of an R&S SMU Signal
generator. This can be used to upload or download settings in
order to keep the R&S SMU and R&S FSP synchronised.
Remote: SYST:COMM:TCP:ADDR 192.168.1.1
1308.5523.42 72 E-1
Page 89
R&S FSP-K93 Measurements & Settings
Demod Settings
his section of the user manual describes the Demod (Short for Demodulation) Settings view for the
T
IEEE 802.16e-2005 WiBro standard where the settings associated with the signal modulation can be
modified.
DEMOD
SETTINGS
The DEMOD SETTINGS softkey brings up the Demod Settings view.
The Demod settings are logically grouped together into:
• DemodSettings – contains the demodulation settings which are used to analyse the signal.
• Frame Global – contains the general frame settings used in configuring and anlysing the signal.
• Frame Config – contains the Zone and Burst setting that specifies a Frame.
When a particular parameter is selected within the Demod Settings view the status bar changes to
display information about the valid settings for the selected parameter. The parameters available in the
Demod setting view are dependent on the currently selected standard.
1308.5523.42 73 E-1
Page 90
Measurements & Settings R&S FSP-K93
Channel Estimation Range
Allows that user to specify the Channel Estimation settings used in measuring a signal.
Downlink (DL)
Downlink (DL) specifies how channel estimation is
performed for downlink signals. The following values are
supported:
Preamble Only
Preamble and Payload [Payload comprises Pilots and Data]
Payload Only
Remote: SENS:DEM:CEST
Uplink (UL)
Uplink (UL) specifies how channel estimation is performed
for uplink signals. This parameter is read only and is for
information purposes only as it indicates that channel
estimation for uplink signals is always performed in payload
[payload comprises pilots and data] only.
Remote: TRAC:PIL DET
Bursts To Analyze
Displays the modulations to be analysed.
Modulation Analysis Scope
Demodulator
Modulation Analysis Scope specifies which modulation
formats will be analyzed. This parameter is read only and is
for information purposes only as it indicates that all
detected modulation schemes will be analyzed.
Remote:
Demodulator is read only and is for information purposes
only as it indicates the modulation scheme detected in the
analyzed signal. For signals using multiple modulation
schemes then the modulation scheme with the highest data
rate is displayed. This parameter also indicates which limits
are applied to the EVM results in the table or results.
Remote: SENS:DEM:FORM:BAN ‘QPSK1/2’
SENS:DEM:FORM:[BCON]:AUT
Tracking Settings
The tracking settings allow various errors in measurement results to be compensated for.
1308.5523.42 74 E-1
Page 91
R&S FSP-K93 Measurements & Settings
Phase
Phase is used to specify whether or not the measurement
results should be compensated for phase error.
When Phase is set to ON then the measurement results will be
compensated for phase error on a per-symbol basis.
Remote: SENS:TRAC:PHAS ON
Timing
Timing is used to specify whether or not the measurement
results should be compensated for timing error.
When Timing is set to ON then the measurement results will be
compensated for timing error on a per-symbol basis.
Remote: SENS:TRAC:TIME ON
Level
Level is used to specify whether or not the measurement
results should be compensated for changes in DUT gain over
time.
When Timing is set to ON then the measurement results will be
compensated for gain on a per-symbol basis.
Remote: SENS:TRAC:LEV ON
Pilots for Tracking
Pilots for Tracking is used to specify how the pilot sequence,
for is determined for frame synchronisation. The following
values are supported:
According to Standard The pilot sequence is computed
Detected The pilot sequence detected in the
Remote: SENS:TRAC:PIL DET
according to the standard.
signal is used.
Advanced
This section contains the more advanced setting used in analysing the signal.
FFT Start offset rel to CP center
FFT start offset rel to CP center: The application determines the
optimal FFT starting point. However this parameter allows you to
add an offset to the optimal FFT starting point determined by the
Application. The parameter ranges from -100 %, …, 100 % of the
Cyclic Prefix [CP] length. The default setting is 0 %. In this case
the optimal FFT starting point, determined by the application, is
used.
Remote:
SENS:FFT:OFFS
1308.5523.42 75 E-1
Page 92
Measurements & Settings R&S FSP-K93
Frame Global
This section of the user manual describes the Frame Global Settings view where the common settings
associated with the frame to be analyzed can be modified.
DEMOD
SETTINGS
The DEMOD SETTINGS softkey brings up the Demod Settings view. Selecting the
Frame Global tab displays the Frame Global settings view. Use the left/right arrow to
navigate between the tabs.
Fig. 41 Frame Global Settings view
1308.5523.42 76 E-1
Page 93
R&S FSP-K93 Measurements & Settings
Frame
Analyze Using
Use … for analysis
The option specifies whether the analysis should be
performed using the current user-defined configuration, or
whether automatic determination of the frame configuration
should be performed.
Predefined MAP Uses the current user-defined configuration
Signal MAP Performs an additional initial sweep to
Automatic detection of the frame configuration requires
several measurement parameters to be correctly configured:
Frequency, Sample Rate / Channel Bandwidth, FFT size,
Guard Time, IDCell and Preamble Index (if not in the range 0 - 96).
The current release supports Signal MAP determination for
DL signals only.
Note: Loading a .WIMAX file automatically sets "Predefined
Remote: DEM:FORM:AUT PRED
MAP"
determine the configuration from the signal.
File Name
IDCell
File Name indicates the name of the loaded .WIMAX
settings file. If no file is loaded the ‘….’ Is displayed. To
load a setting file, select the parameter and press the icon
displayed. The File Manager dialog is opened.
Remote: MMEM:LOAD:FRAM:STAT
IDCell affects the definition of the Preamble sequence [in
conjunction with the used segments according to the Used Subchannel Bitmap selection], the pilot sequence of the first
zone and the subchannel definition of the first zone.
Remote: CONF:WIM:DLSF:IDC 0
1308.5523.42 77 E-1
Page 94
Measurements & Settings R&S FSP-K93
DL Subframe
Preamble Mode
Preamble Mode determines whether how the preamble
index is calculated. The values supported are:
Auto The Preamble index is automatically calculated
according to the IDcell and Used Subchannel Bitmap
settings
User The preamble index can be specified manually. I.e.
the preamble pattern is chosen according to the
standard using this setting.
Preamble Index
Remote: :CONF:WIM:DLSF:PMOD Preamble Mode
Preamble Index: The Preamble mode has been set to User.
The Preamble Index defines the preamble pattern –
according to the standard – to be used.
Remote: :CONF:WIM:DLSF:PIND Preamble Index
Used Subchannel Bitmap
Used Subchannel Bitmap assigns subchannel groups to
segments. See [802.16e-2005].360 Table 268a Subchannel
index of the six subchannel groups. This table defines how
many subchannels will be available for the Burst List
belonging to the corresponding segment of a DL-PUSC
zone [with Use All Subchannels being false]. In the Burst
List - corresponding to the segment – this selection controls
the height of the white area in the burst map.
Remote:
CONF:WIMAX:DLSF:SEG
1308.5523.42 78 E-1
Page 95
R&S FSP-K93 Measurements & Settings
UL Subframe
UL Control Region Length
UL Control Region Length specifies the length, in symbols,
of a control region to be found at the start of any UL zone.
The control region will not be analysed, but the length must
be specified to allow synchronisation.
Remote: CONF:WIM:ULSF:CRL
Frame Number
Frame Number specifies the number of the frame to be
analyzed. In case of UL-Subframe analysis, the Frame
Number must be constant for the signal to be analyzed!
Allocated Subchannel Bitmap
Remote:
Allocated Subchannel bitmap specifies which subchannels
are to be analyzed. This parameter is read only and is for
information purposes only as all subchannels are used in
the UL-Subframe.
Remote: ---
CONF:WIMAX:ULSF:SEG
1308.5523.42 79 E-1
Page 96
Measurements & Settings R&S FSP-K93
Frame Config (IEEE 802.16e-2005 WiBro)
This section of the user manual describes the Frame Config Settings view where the settings
associated with the frame configuration can be modified.
DEMOD
SETTINGS
The DEMOD SETTINGS softkey brings up the Demod Settings view. Selecting the
Frame Config tab displays the Frame Config settings view. Use the left/right arrow to
navigate between the tabs.
Fig. 42 Frame Configuration Settings view
The Frame Configuration Settings view comprises
• the Zone/Segment List
• the Burst List
• the Display Pane. I.e. according to the focus being on the Zone/Segment List or the Burst List, the
content of the selected List is visually displayed.
1308.5523.42 80 E-1
Page 97
R&S FSP-K93 Measurements & Settings
Zone/Segment List
Fig. 43 Definition of the Zone Offset [red arrows] and the Zone Length [green arrows] parameters
from the Zone/Segment List
Zone/Segment List parameters:
1. ID
Provides a unique ID for the zone. This parameter is read only. If there are any errors in the
zone/segment configuration then the background of this item will be red.
2. Bursts
Pressing the roll-key or ENTER on tyhis field causes the Burst List to become selected and the Burst
Map to be displayed.
3. Analyze
Defines the Zone/Segment to be analyzed. Note that only one sone/segment can be marked for
analysis at any given time.
4. Zone
Defines the zone type. The following types are supported:
DL-PUSC
DL-FUSC
UL-PUSC
5. Segment
Defines the segment for a DL-PUSC zone.
Note: in case of DL-FUSC, this parameter has no effect.
1308.5523.42 81 E-1
Page 98
Measurements & Settings R&S FSP-K93
6. Length [Symb]
efines the length in OFDMA Symbols of the zone.
D
The zone Length must be a multiple of the number of symbols per slot. In case this condition is not
hold a ‘Zone/Burst configuration Error’ is displayed in the status bar and the measurement won’t
start.
7. Offset [Symb]
Defines the offset in OFDMA Symbols from the start of the subframe.
Note I: The first DL-PUSC zone starts with 1.
Note II: The first UL-PUSC zone starts with 0.
8. PermBase
This parameter is required for the computation of the subchannel – physical carrier assignment
permutation.
9. PRBS_ID
Is required for the computation of the pilot sequence.
Burst List
Fig. 44 Definition of the Burst Offset [red arrows] and the Burst Length [green arrows] parameters
from the Burst List.
1308.5523.42 82 E-1
Page 99
R&S FSP-K93 Measurements & Settings
Burst List parameter:
1. ID
Provides a unique ID for the burst. This parameter is read only. The background colour of this
parameter reflects the modulation of the burst. If there are any errors in the burst configuration then
he background of this item will be red.
t
2. Modulation
Defines the modulation format of the burst. Identical modulation formats use the same colour in the
Burst Map.
3. No. of Subch.
Defines the number of logical Subchannels occupied by the burst.
4. No. of Symb.
Defines the number of OFDMA symbols occupied by the burst.
The No. of Symb. must be a multiple of the number of symbols per slot. In case this condition is not
hold a ‘Zone/Burst configuration Error’ is displayed in the status bar and the measurement won’t
start.
5. Offset Subch
Defines the offset in logical subchannels from subchannel 0.
6. Offset Symb
Defines the offset in OFDMA symbols. The start of the zone defines symbol 0.
The Offset Symb. must be a multiple of the number of symbols per slot. In case this condition is not
hold a ‘Zone/Burst configuration Error’ is displayed in the status bar and the measurement won’t
start.
7. Power[dB]
Defines the boosting factor of the burst.
8. Burst Type
Defines the burst type from the protocol layer perspective
Zone/Segment Map
Fig. 45 Zone/Segment Map
The Zone/Segment map is displayed at the bottom of the Frame Config dialog when the zone/segment
list has focus. The Zone/Segment map displays a graphical view of the Zone/Segment list. The
currently selected zone/segment is highlighted with a blue border. The zone/segment selected for
analysis is shown in green.
1308.5523.42 83 E-1
Page 100
Measurements & Settings R&S FSP-K93
Burst Map
Fig. 46 Burst Map
The Burst map is displayed at the bottom of the Frame Config dialog when the burst list has focus. The
Burst map displays a graphical view of the Burst list. The currently selected burst in the Zone/Segment
list is highlighted in the Burst map. The background color of each burst in the diagram corresponds to
the modulation format of the burst. Any burst that is incorrectly configured will be displayed with a red
background. In example if the burst definition exceeds the available time x subchannel area [indicated
white in the graphic above].
Fig. 47 Burst Map with Burst C2 exceeding the available amount of logical subchannels. The area
causing the problem is highlighted red.
In this case the measurement won’t start unless the frame definition is correct. In the example above
this might be achived by.
a. reducing the number of subchannels for burst C2 in the Burst List or
b. assigning more subchannels to the zone/segment to which burst C2 belongs or
c. increase
1308.5523.42 84 E-1
N.
FFT
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.