Rohde&Schwarz FSP-K90 User Manual

Page 1
Test and Measurement Division
Software Manual
WLAN 802.11A TX Tests
Application Firmware R&SFSP-K90
1300.6850.02
Printed in Germany
1300.6789.42-03 1
Page 2
®
is a registered trademark of Rohde & Schwarz GmbH & Co. KG.
R&S
Trade names are trademarks of the owners.
1300.6789.42 2 E-2
Page 3
R&S FSP-K90 Contents
Contents
1 General Information ............................................................................................ 1.1
Introduction to R&S FSP-K90 & Wireless LAN Measurements ...................................................1.2
Installation.........................................................................................................................................1.2
Starting the application ...................................................................................................................1.4
Exiting the application .....................................................................................................................1.4
Quick Start Guide .............................................................................................................................1.5
Setting up the measurement.......................................................................................................1.5
Performing Level Detection.............................................................................................1.7
Performing the main measurement ............................................................................................1.7
Navigation .........................................................................................................................................1.8
Hotkeys .......................................................................................................................................1.8
Softkeys ....................................................................................................................................1.10
Hardkeys...................................................................................................................................1.11
External Keyboard ....................................................................................................................1.13
Mouse .......................................................................................................................................1.14
Selecting & Editing Parameters ................................................................................................1.14
Numeric Keypad ...........................................................................................................1.15
Rollkey ..........................................................................................................................1.16
Cursor Keys ..................................................................................................................1.16
Selection of a parameter within a settings view............................................................1.17
Entry of a numeric value ...............................................................................................1.19
Entry of an enumerated value.......................................................................................1.21
Entry of a checkbox ......................................................................................................1.22
Status Bar & Title Bar ...............................................................................................................1.23
Title Bar.........................................................................................................................1.23
Status Bar .....................................................................................................................1.23
Save/Recall......................................................................................................................................1.24
Printing ............................................................................................................................................1.25
2 Measurements & Settings .................................................................................. 2.1
Measurements ..................................................................................................................................2.1
IQ measurements .......................................................................................................................2.3
EVM Vs Symbol ..............................................................................................................2.4
EVM Vs Carrier ...............................................................................................................2.5
Error Vs Preamble ..........................................................................................................2.6
Constellation Vs Symbol .................................................................................................2.7
Constellation Vs Carrier ..................................................................................................2.8
Frequency sweep measurements ..............................................................................................2.9
Spectrum Emission Mask ...............................................................................................2.9
Spectrum ACPR............................................................................................................2.12
Automatic Level Detection ........................................................................................................2.13
Running measurements ................................................................................................................2.13
Measurement Results ....................................................................................................................2.14
1300.6789.42.42 3 E-3
Page 4
Contents R&S FSP-K90
Results Summary .....................................................................................................................2.15
General Settings .............................................................................................................................2.16
Signal Characteristics ...............................................................................................................2.17
Standard........................................................................................................................2.17
Frequency .....................................................................................................................2.17
Channel No ...................................................................................................................2.17
Auto Level .....................................................................................................................2.17
Ext Att............................................................................................................................2.18
Signal Level (RF) ..........................................................................................................2.18
Data Capture Settings ..............................................................................................................2.18
Capture Time ................................................................................................................2.18
Overall Burst Count.......................................................................................................2.18
No of Burst to Analyze ..................................................................................................2.18
Trigger Settings ........................................................................................................................2.19
Trigger Mode.................................................................................................................2.19
Trigger Offset ................................................................................................................2.19
Power Level (RF) ..........................................................................................................2.19
Auto Power Trigger Level .............................................................................................2.19
IQ Settings ................................................................................................................................2.20
Swap IQ ........................................................................................................................2.20
Advanced Settings ....................................................................................................................2.21
Auto Level .....................................................................................................................2.21
Auto Level time .............................................................................................................2.22
Ref Level .......................................................................................................................2.22
RF Att ............................................................................................................................2.22
El Att..............................................................................................................................2.22
YIG Filter .......................................................................................................................2.23
Sample Rate .................................................................................................................2.23
High Dynamic................................................................................................................2.23
Gate Settings ............................................................................................................................2.24
Delay .............................................................................................................................2.24
Length ...........................................................................................................................2.24
Link Gate and Marker ...................................................................................................2.24
Demod Settings ..............................................................................................................................2.25
Burst To Analyze Settings ........................................................................................................2.26
Burst Type.....................................................................................................................2.26
Demodulator .................................................................................................................2.26
Equal Burst Length .......................................................................................................2.26
No of Data Symbols ......................................................................................................2.26
Min No of Data Symbols ...............................................................................................2.27
Max No of Data Symbols ..............................................................................................2.27
Chan Est in Preamb and Payl.......................................................................................2.27
Tracking Settings ......................................................................................................................2.28
Phase ............................................................................................................................2.28
Timing ...........................................................................................................................2.28
Level..............................................................................................................................2.28
Markers............................................................................................................................................2.29
Adjusting Markers .....................................................................................................................2.29
Marker Zoom.............................................................................................................................2.30
Toggle Marker Display..............................................................................................................2.30
Assigning Markers to Traces ....................................................................................................2.31
Display Settings..............................................................................................................................2.32
1300.6789.42 4 E-3
Page 5
R&S FSP-K90 Contents
3 Measurements in Detail ...................................................................................... 3.1
Abbreviations....................................................................................................................................3.1
Signal Processing of the IEEE802.11a application .......................................................................3.1
4 Remote Control ................................................................................................... 4.1
escription of commands ...............................................................................................................4.1
D
Notation.......................................................................................................................................4.1
SCPI Commands...............................................................................................................................4.4
ABORt Subsystem............................................................................................................................4.4
CALCulate: Subsystem....................................................................................................................4.4
CALCulate:BURSt Subsystem....................................................................................................4.4
CALCulate:LIMit Subsystem .......................................................................................................4.5
CALCulate:MARKer Subsystem...............................................................................................4.12
CALCulate:MARKer:FUNCtion Subsystem ..............................................................................4.15
Configure Subsystem ....................................................................................................................4.17
DISPlay Subsystem ........................................................................................................................4.21
FETCh Subsystem.........................................................................................................................4.23
INPut Subsystem ............................................................................................................................4.28
INSTrument Subsystem .................................................................................................................4.30
MMEMory Subsystem ....................................................................................................................4.31
SENSe Subsystem..........................................................................................................................4.32
STATus Subsystem........................................................................................................................4.38
TRACe Subsystem .........................................................................................................................4.42
Trigger Subsystem .........................................................................................................................4.45
Unit Subsystem ..............................................................................................................................4.46
Status Reporting Registers ...........................................................................................................4.47
Description of the Status Registers ..........................................................................................4.49
5 List of Warnings & Error Messages................................................................... 5.1
6 Index..................................................................................................................... 6.1
1300.6789.42.42 5 E-3
Page 6
Page 7
Before putting the product into operation for
the first time, make sure to read the following

Safety Instructions

All plants and locations of the Rohde & Schwarz group of companies make every effort to keep the safety standard of our products up to date and to offer our customers the highest possible degree of safety. Our products and the auxiliary equipment required for them are designed and tested in accordance with the relevant safety standards. Compliance with these standards is continuously monitored by our quality assurance system. The product described here has been designed and tested in accordance with the EC Certificate of Conformity and has left the manufacturer’s plant in a condition fully complying with safety standards. To maintain this condition and to ensure safe operation, observe all instructions and warnings provided in this manual. If you have any questions regarding these safety instructions, the Rohde & Schwarz group of companies will be happy to answer them.
Furthermore, it is your responsibility to use the product in an appropriate manner. This product is designed for use solely in industrial and laboratory environments or in the field and must not be used in any way that may cause personal injury or property damage. You are responsible if the product is used for an intention other than its designated purpose or in disregard of the manufacturer's instructions. The manufacturer shall assume no responsibility for such use of the product.
The product is used for its designated purpose if it is used in accordance with its product documentation and within its performance limits (see data sheet, documentation, the following safety instructions). Using the product requires technical skills and a basic knowledge of English. It is therefore essential that the product be used exclusively by skilled and specialized staff or thoroughly trained personnel with the required skills. If personal safety gear is required for using Rohde & Schwarz products, this will be indicated at the appropriate place in the product documentation.
Observe product documentation
Supply voltage ON/OFF
Weight indication for units >18 kg
Standby indication
Symbols and safety labels
Danger of electric shock
Warning! Hot surface
PE terminal Ground
Direct current (DC)
Alternating current (AC)
Direct/alternating current (DC/AC)
Ground terminal
Device fully protected by double/reinforced insulation
Attention! Electrostatic sensitive devices
1171.0000.42-03.00 Sheet 1
Page 8
Safety Instructions
Observing the safety instructions will help prevent personal injury or damage of any kind caused by dangerous situations. Therefore, carefully read through and adhere to the following safety instructions before putting the product into operation. It is also absolutely essential to observe the additional safety instructions on personal safety that appear in relevant parts of the product documentation. In these safety instructions, the word "product" refers to all merchandise sold and distributed by the Rohde & Schwarz group of companies, including instruments, systems and all accessories.
Tags and their meaning
DANGER
WARNING
CAUTION This tag indicates a hazard carrying a low risk of minor or moderate
ATTENTION
NOTE
These tags are in accordance with the standard definition for civil applications in the European Economic Area. Definitions that deviate from the standard definition may also exist in other economic areas or military applications. It is therefore essential to make sure that the tags described here are always used only in connection with the related product documentation and the related product. The use of tags in connection with unrelated products or documentation can result in misinterpretation and thus contribute to personal injury or material damage.
This tag indicates a definite hazard carrying a high risk of death or serious injury if not avoided.
This tag indicates a possible hazard carrying a medium risk of death or (serious) injury if not avoided
injury if not avoided.
This tag indicates the possibility of incorrect use that can cause damage to the product.
This tag indicates a situation where the user should pay special attention to operating the product but which does not lead to damage.
.
Basic safety instructions
1. The product may be operated only under the operating conditions and in the positions specified by the manufacturer. Its ventilation must not be obstructed during operation. Unless otherwise specified, the following requirements apply to Rohde & Schwarz products: prescribed operating position is always with the housing floor facing down, IP protection 2X, pollution severity 2, overvoltage category 2, use only in enclosed spaces, max. operation altitude 2000 m above sea level, max. transport altitude 4500 m above sea level. Unless specified otherwise in the data sheet, a tolerance of ±10% shall apply to the nominal voltage and of ±5% to the nominal frequency.
2. Applicable local or national safety regulations and rules for the prevention of accidents must be observed in all work performed. The product may be opened only by authorized, specially trained personnel. Prior to performing any work on
the product or opening the product, the product must be disconnected from the supply network. Any adjustments, replacements of parts, maintenance or repair must be carried out only by technical personnel authorized by Rohde & Schwarz. Only original parts may be used for replacing parts relevant to safety (e.g. power switches, power transformers, fuses). A safety test must always be performed after parts relevant to safety have been replaced (visual inspection, PE conductor test, insulation resistance measurement, leakage current measurement, functional test).
3. As with all industrially manufactured goods, the use of substances that induce an allergic reaction (allergens, e.g. nickel) such as aluminum cannot be generally excluded. If you develop an allergic reaction (such as a skin rash, frequent sneezing, red eyes or respiratory difficulties), consult a physician immediately to determine the cause.
1171.0000.42-03.00 Sheet 2
Page 9
Safety Instructions
4. If products/components are mechanically and/or thermically processed in a manner that goes beyond their intended use, hazardous substances (heavy-metal dust such as lead, beryllium, nickel) may be released. For this reason, the product may only be disassembled, e.g. for disposal purposes, by specially trained personnel. Improper disassembly may be hazardous to your health. National waste disposal regulations must be observed.
5. If handling the product yields hazardous substances or fuels that must be disposed of in a special way, e.g. coolants or engine oils that must be replenished regularly, the safety instructions of the manufacturer of the hazardous substances or fuels and the applicable regional waste disposal regulations must be observed. Also observe the relevant safety instructions in the product documentation.
6. Depending on the function, certain products such as RF radio equipment can produce an elevated level of electromagnetic radiation. Considering that unborn life requires increased protection, pregnant women should be protected by appropriate measures. Persons with pacemakers may also be endangered by electromagnetic radiation. The employer/operator is required to assess workplaces where there is a special risk of exposure to radiation and, if necessary, take measures to avert the danger.
7. Operating the products requires special training and intense concentration. Make certain that persons who use the products are physically, mentally and emotionally fit enough to handle operating the products; otherwise injuries or material damage may occur. It is the responsibility of the employer to select suitable personnel for operating the products.
8. Prior to switching on the product, it must be ensured that the nominal voltage setting on the product matches the nominal voltage of the AC supply network. If a different voltage is to be set, the power fuse of the product may have to be changed accordingly.
9. In the case of products of safety class I with movable power cord and connector, operation is permitted only on sockets with earthing contact and protective earth connection.
10. Intentionally breaking the protective earth connection either in the feed line or in the
product itself is not permitted. Doing so can result in the danger of an electric shock from the product. If extension cords or connector strips are implemented, they must be checked on a regular basis to ensure that they are safe to use.
11. If the product has no power switch for disconnection from the AC supply, the plug of the connecting cable is regarded as the disconnecting device. In such cases, it must be ensured that the power plug is easily reachable and accessible at all times (corresponding to the length of connecting cable, approx. 2 m). Functional or electronic switches are not suitable for providing disconnection from the AC supply. If products without power switches are integrated in racks or systems, a disconnecting device must be provided at the system level.
12. Never use the product if the power cable is damaged. Check the power cable on a regular basis to ensure that it is in proper operating condition. By taking appropriate safety measures and carefully laying the power cable, ensure that the cable cannot be damaged and that no one can be hurt by e.g. tripping over the cable or suffering an electric shock.
13. The product may be operated only from TN/TT supply networks fused with max. 16 A (higher fuse only after consulting with the Rohde & Schwarz group of companies).
14. Do not insert the plug into sockets that are dusty or dirty. Insert the plug firmly and all the way into the socket. Otherwise, this can result in sparks, fire and/or injuries.
15. Do not overload any sockets, extension cords or connector strips; doing so can cause fire or electric shocks.
16. For measurements in circuits with voltages V
> 30 V, suitable measures (e.g.
rms
appropriate measuring equipment, fusing, current limiting, electrical separation, insulation) should be taken to avoid any hazards.
17. Ensure that the connections with information technology equipment comply with IEC 950/EN 60950.
18. Unless expressly permitted, never remove the cover or any part of the housing while the product is in operation. Doing so will expose circuits and components and can lead to injuries, fire or damage to the product.
1171.0000.42-03.00 Sheet 3
Page 10
Safety Instructions
19. If a product is to be permanently installed, the connection between the PE terminal on site and the product's PE conductor must be made first before any other connection is made. The product may be installed and connected only by a license electrician.
20. For permanently installed equipment without built-in fuses, circuit breakers or similar protective devices, the supply circuit must be fused in such a way that suitable protection is provided for users and products.
21. Do not insert any objects into the openings in the housing that are not designed for this purpose. Never pour any liquids onto or into the housing. This can cause short circuits inside the product and/or electric shocks, fire or injuries.
22. Use suitable overvoltage protection to ensure that no overvoltage (such as that caused by a thunderstorm) can reach the product. Otherwise the operating personnel will be endangered by electric shocks.
23. Rohde & Schwarz products are not protected against penetration of water, unless otherwise specified (see also safety instruction 1.). If this is not taken into account, there exists the danger of electric shock for the user or damage to the product, which can also lead to personal injury.
24. Never use the product under conditions in which condensation has formed or can form in or on the product, e.g. if the product was moved from a cold to a warm environment.
25. Do not close any slots or openings on the product, since they are necessary for ventilation and prevent the product from overheating. Do not place the product on soft surfaces such as sofas or rugs or inside a closed housing, unless this is well ventilated.
26. Do not place the product on heat­generating devices such as radiators or fan heaters. The temperature of the environment must not exceed the maximum temperature specified in the data sheet.
27. Batteries and storage batteries must not be exposed to high temperatures or fire. Keep batteries and storage batteries away from children. Do not short-circuit batteries and storage batteries. If batteries or storage batteries are improperly replaced, this can cause an explosion (warning: lithium cells). Replace
the battery or storage battery only with the matching Rohde & Schwarz type (see spare parts list). Batteries and storage batteries must be recycled and kept separate from residual waste. Batteries and storage batteries that contain lead, mercury or cadmium are hazardous waste. Observe the national regulations regarding waste disposal and recycling.
28. Please be aware that in the event of a fire, toxic substances (gases, liquids etc.) that may be hazardous to your health may escape from the product.
29. The product can be very heavy. Be careful when moving it to avoid back or other physical injuries.
30. Do not place the product on surfaces, vehicles, cabinets or tables that for reasons of weight or stability are unsuitable for this purpose. Always follow the manufacturer's installation instructions when installing the product and fastening it to objects or structures (e.g. walls and shelves).
31. Handles on the products are designed exclusively for personnel to hold or carry the product. It is therefore not permissible to use handles for fastening the product to or on means of transport such as cranes, fork lifts, wagons, etc. The user is responsible for securely fastening the products to or on the means of transport and for observing the safety regulations of the manufacturer of the means of transport. Noncompliance can result in personal injury or material damage.
32. If you use the product in a vehicle, it is the sole responsibility of the driver to drive the vehicle safely. Adequately secure the product in the vehicle to prevent injuries or other damage in the event of an accident. Never use the product in a moving vehicle if doing so could distract the driver of the vehicle. The driver is always responsible for the safety of the vehicle. The manufacturer assumes no responsibility for accidents or collisions.
33. If a laser product (e.g. a CD/DVD drive) is integrated in a Rohde & Schwarz product, do not use any other settings or functions than those described in the product documentation. Otherwise this may be hazardous to your health, since the laser beam can cause irreversible damage to your eyes. Never try to take such products apart, and never look into the laser beam.
1171.0000.42-03.00 Sheet 4
Page 11
Informaciones de seguridad
Por favor lea imprescindiblemente antes de la primera puesta en funcionamiento las siguientes
Informaciones de seguridad
El principio del grupo de empresas Rohde & Schwarz consiste en tener nuestros productos siempre al día con los estandards de seguridad y de ofrecer a nuestros clientes el máximo grado de seguridad. Nuestros productos y todos los equipos adicionales son siempre fabricados y examinados según las normas de seguridad vigentes. Nuestra sección de gestión de la seguridad de calidad controla constantemente que sean cumplidas estas normas. El presente producto ha sido fabricado y examinado según el comprobante de conformidad adjunto según las normas de la CE y ha salido de nuestra planta en estado impecable según los estandards técnicos de seguridad. Para poder preservar este estado y garantizar un funcionamiento libre de peligros, el usuario deberá atenerse a todas las informaciones, informaciones de seguridad y notas de alerta. El grupo de empresas Rohde & Schwarz está siempre a su disposición en caso de que tengan preguntas referentes a estas informaciones de seguridad.
Además queda en la responsabilidad del usuario utilizar el producto en la forma debida. Este producto solamente fue elaborado para ser utilizado en la industria y el laboratorio o para fines de campo y de ninguna manera deberá ser utilizado de modo que alguna persona/cosa pueda ser dañada. El uso del producto fuera de sus fines definidos o despreciando las informaciones de seguridad del fabricante queda en la responsabilidad del usuario. El fabricante no se hace en ninguna forma responsable de consecuencias a causa del mal uso del producto.
Se parte del uso correcto del producto para los fines definidos si el producto es utilizado dentro de las instrucciones de la correspondiente documentación de producto y dentro del margen de rendimiento definido (ver hoja de datos, documentación, informaciones de seguridad que siguen). El uso del producto hace necesarios conocimientos profundos y conocimientos parciales del idioma inglés. Por eso se deberá tener en cuenta de exclusivamente autorizar para el uso del producto a personas peritas o debidamente minuciosamente instruidas con los conocimientos citados. Si fuera necesaria indumentaria de seguridad para el uso de productos de R&S, encontrará la información debida en la documentación del producto en el capítulo correspondiente.
Símbolos y definiciones de seguridad
Ver documen­tación de producto
Informaciones para maquinaria con uns peso de > 18kg
Peligro de golpe de corriente
¡Advertencia! Superficie caliente
Conexión a conductor protector
Conexión a tierra
Conexión a masa conductora
¡Cuidado! Elementos de construcción con peligro de carga electroestática
El aparato está protegido en su totalidad por un aislamiento de doble refuerzo
potencia EN MARCHA/PARADA
Indicación Stand-by
Corriente continua DC
Corriente alterna AC
Corriente continua/alterna DC/AC
1171.0000.42-03.00 Sheet 5
Page 12
Informaciones de seguridad
Tener en cuenta las informaciones de seguridad sirve para tratar de evitar daños y peligros de toda clase. Es necesario de que se lean las siguientes informaciones de seguridad concienzudamente y se tengan en cuenta debidamente antes de la puesta en funcionamiento del producto. También deberán ser tenidas en cuenta las informaciones para la protección de personas que encontrarán en el capítulo correspondiente de la documentación de producto y que también son obligatorias de seguir. En las informaciones de seguridad actuales hemos juntado todos los objetos vendidos por el grupo de empresas Rohde & Schwarz bajo la denominación de „producto“, entre ellos también aparatos, instalaciones así como toda clase de accesorios.
Palabras de señal y su significado
PELIGRO Identifica un peligro directo con riesgo elevado de provocar muerte o
lesiones de gravedad si no se toman las medidas oportunas.
ADVERTENCIA Identifica un posible peligro con riesgo medio de provocar muerte o
lesiones (de gravedad) si no se toman las medidas oportunas.
ATENCIÓN Identifica un peligro con riesgo reducido de provocar lesiones de
gravedad media o leve si no se toman las medidas oportunas.
CUIDADO Indica la posibilidad de utilizar mal el producto y a consecuencia
dañarlo.
INFORMACIÓN Indica una situación en la que deberían seguirse las instrucciones en el
uso del producto, pero que no consecuentemente deben de llevar a un daño del mismo.
Las palabras de señal corresponden a la definición habitual para aplicaciones civiles en el área económica europea. Pueden existir definiciones diferentes a esta definición en otras áreas económicas o en aplicaciones militares. Por eso se deberá tener en cuenta que las palabras de señal aquí descritas sean utilizadas siempre solamente en combinación con la correspondiente documentación de producto y solamente en combinación con el producto correspondiente. La utilización de las palabras de señal en combinación con productos o documentaciones que no les correspondan puede llevar a malinterpretaciones y tener por consecuencia daños en personas u objetos.
Informaciones de seguridad elementales
1. El producto solamente debe ser utilizado según lo indicado por el fabricante referente a la situación y posición de funcionamiento sin que se obstruya la ventilación. Si no se convino de otra manera, es para los productos R&S válido lo que sigue: como posición de funcionamiento se define principialmente la posición con el suelo de la caja para abajo , modo de protección IP 2X, grado de suciedad 2, categoría de sobrecarga eléctrica 2, utilizar solamente en estancias interiores, utilización hasta 2000 m sobre el nivel del mar, transporte hasta
4.500 m sobre el nivel del mar. A menos que se especifique otra cosa en la hoja de datos, se aplicará una tolerancia de ±10% sobre el voltaje nominal y de ±5% sobre la frecuencia nominal.
2. En todos los trabajos deberán ser tenidas en cuenta las normas locales de seguridad de
trabajo y de prevención de accidentes. El producto solamente debe de ser abierto por personal perito autorizado. Antes de efectuar trabajos en el producto o abrirlo deberá este ser desconectado de la corriente. El ajuste, el cambio de partes, la manutención y la reparación deberán ser solamente efectuadas por electricistas autorizados por R&S. Si se reponen partes con importancia para los aspectos de seguridad (por ejemplo el enchufe, los transformadores o los fusibles), solamente podrán ser sustituidos por partes originales. Despues de cada recambio de partes elementales para la seguridad deberá ser efectuado un control de seguridad (control a primera vista, control de conductor protector, medición de resistencia de aislamiento, medición de medición de la corriente conductora, control de funcionamiento).
1171.0000.42-03.00 Sheet 6
Page 13
Informaciones de seguridad
3. Como en todo producto de fabricación industrial no puede ser excluido en general de que se produzcan al usarlo elementos que puedan generar alergias, los llamados elementos alergénicos (por ejemplo el níquel). Si se producieran en el trato con productos R&S reacciones alérgicas, como por ejemplo urticaria, estornudos frecuentes, irritación de la conjuntiva o dificultades al respirar, se deberá consultar inmediatamente a un médico para averigurar los motivos de estas reacciones.
4. Si productos / elementos de construcción son tratados fuera del funcionamiento definido de forma mecánica o térmica, pueden generarse elementos peligrosos (polvos de sustancia de metales pesados como por ejemplo plomo, berilio, níquel). La partición elemental del producto, como por ejemplo sucede en el tratamiento de materias residuales, debe de ser efectuada solamente por personal especializado para estos tratamientos. La partición elemental efectuada inadecuadamente puede generar daños para la salud. Se deben tener en cuenta las directivas nacionales referentes al tratamiento de materias residuales.
5. En el caso de que se produjeran agentes de peligro o combustibles en la aplicación del producto que debieran de ser transferidos a un tratamiento de materias residuales, como por ejemplo agentes refrigerantes que deben ser repuestos en periodos definidos, o aceites para motores, deberan ser tenidas en cuenta las prescripciones de seguridad del fabricante de estos agentes de peligro o combustibles y las regulaciones regionales para el tratamiento de materias residuales. Cuiden también de tener en cuenta en caso dado las prescripciones de seguridad especiales en la descripción del producto.
6. Ciertos productos, como por ejemplo las instalaciones de radiación HF, pueden a causa de su función natural, emitir una radiación electromagnética aumentada. En vista a la protección de la vida en desarrollo deberían ser protegidas personas embarazadas debidamente. También las personas con un bypass pueden correr peligro a causa de la radiación electromagnética. El empresario/usario está
comprometido a valorar y señalar areas de trabajo en las que se corra un riesgo aumentado de exposición a radiaciones para evitar riesgos.
7. La utilización de los productos requiere instrucciones especiales y una alta concentración en el manejo. Debe de ponerse por seguro de que las personas que manejen los productos estén a la altura de los requerimientos necesarios referente a sus aptitudes físicas, psíquicas y emocionales, ya que de otra manera no se pueden excluir lesiones o daños de objetos. El empresario lleva la responsabilidad de seleccionar el personal usuario apto para el manejo de los productos.
8. Antes de la puesta en marcha del producto se deberá tener por seguro de que la tensión preseleccionada en el producto equivalga a la del la red de distribución. Si es necesario cambiar la preselección de la tensión también se deberán en caso dabo cambiar los fusibles correspondientes del prodcuto.
9. Productos de la clase de seguridad I con alimentación móvil y enchufe individual de producto solamente deberán ser conectados para el funcionamiento a tomas de corriente de contacto de seguridad y con conductor protector conectado.
10. Queda prohibida toda clase de interrupción intencionada del conductor protector, tanto en la toma de corriente como en el mismo producto. Puede tener como consecuencia el peligro de golpe de corriente por el producto. Si se utilizaran cables o enchufes de extensión se deberá poner al seguro, que es controlado su estado técnico de seguridad.
11. Si el producto no está equipado con un interruptor para desconectarlo de la red, se deberá considerar el enchufe del cable de distribución como interruptor. En estos casos deberá asegurar de que el enchufe sea de fácil acceso y nabejo (según la medida del cable de distribución, aproximadamente 2 m). Los interruptores de función o electrónicos no son aptos para el corte de la red eléctrica. Si los productos sin interruptor están integrados en construciones o instalaciones, se deberá instalar el interruptor al nivel de la instalación.
1171.0000.42-03.00 Sheet 7
Page 14
Informaciones de seguridad
12. No utilice nunca el producto si está dañado el cable eléctrico. Compruebe regularmente el correcto estado de los cables de conexión a red. Asegure a través de las medidas de protección y de instalación adecuadas de que el cable de eléctrico no pueda ser dañado o de que nadie pueda ser dañado por él, por ejemplo al tropezar o por un golpe de corriente.
13. Solamente está permitido el funcionamiento en redes de distribución TN/TT aseguradas con fusibles de como máximo 16 A (utilización de fusibles de mayor amperaje sólo previa consulta con el grupo de empresas Rohde & Schwarz).
14. Nunca conecte el enchufe en tomas de corriente sucias o llenas de polvo. Introduzca el enchufe por completo y fuertemente en la toma de corriente. Si no tiene en consideración estas indicaciones se arriesga a que se originen chispas, fuego y/o heridas.
15. No sobrecargue las tomas de corriente, los cables de extensión o los enchufes de extensión ya que esto pudiera causar fuego o golpes de corriente.
16. En las mediciones en circuitos de corriente con una tensión de entrada de U
> 30 V se
eff
deberá tomar las precauciones debidas para impedir cualquier peligro (por ejemplo medios de medición adecuados, seguros, limitación de tensión, corte protector, aislamiento etc.).
17. En caso de conexión con aparatos de la técnica informática se deberá tener en cuenta que estos cumplan los requisitos de la EC950/EN60950.
18. A menos que esté permitido expresamente, no retire nunca la tapa ni componentes de la carcasa mientras el producto esté en servicio. Esto pone a descubierto los cables y componentes eléctricos y puede causar heridas, fuego o daños en el producto.
19. Si un producto es instalado fijamente en un lugar, se deberá primero conectar el conductor protector fijo con el conductor protector del aparato antes de hacer cualquier otra conexión. La instalación y la conexión deberán ser efecutadas por un electricista especializado.
20. En caso de que los productos que son instalados fijamente en un lugar sean sin protector implementado, autointerruptor o similares objetos de protección, el circuito de suministro de corriente deberá estar protegido de manera que usuarios y productos estén suficientemente protegidos.
21. Por favor, no introduzca ningún objeto que no esté destinado a ello en los orificios de la caja del aparato. No vierta nunca ninguna clase de líquidos sobre o en la caja. Esto puede producir corto circuitos en el producto y/o puede causar golpes de corriente, fuego o heridas.
22. Asegúrese con la protección adecuada de que no pueda originarse en el producto una sobrecarga por ejemplo a causa de una tormenta. Si no se verá el personal que lo utilice expuesto al peligro de un golpe de corriente.
23. Los productos R&S no están protegidos contra el agua si no es que exista otra indicación, ver también punto 1. Si no se tiene en cuenta esto se arriesga el peligro de golpe de corriente para el usario o de daños en el producto lo cual también puede llevar al peligro de personas.
24. No utilice el producto bajo condiciones en las que pueda producirse y se hayan producido líquidos de condensación en o dentro del producto como por ejemplo cuando se desplaza el producto de un lugar frío a un lugar caliente.
25. Por favor no cierre ninguna ranura u orificio del producto, ya que estas son necesarias para la ventilación e impiden que el producto se caliente demasiado. No pongan el producto encima de materiales blandos como por ejemplo sofás o alfombras o dentro de una caja cerrada, si esta no está suficientemente ventilada.
26. No ponga el producto sobre aparatos que produzcan calor, como por ejemplo radiadores o calentadores. La temperatura ambiental no debe superar la temperatura máxima especificada en la hoja de datos.
1171.0000.42-03.00 Sheet 8
Page 15
Informaciones de seguridad
27. Baterías y acumuladores no deben de ser expuestos a temperaturas altas o al fuego. Guardar baterías y acumuladores fuera del alcance de los niños. No cortocircuitar baterías ni acumuladores. Si las baterías o los acumuladores no son cambiados con la debida atención existirá peligro de explosión (atención celulas de Litio). Cambiar las baterías o los acumuladores solamente por los del tipo R&S correspondiente (ver lista de piezas de recambio). Las baterías y acumuladores deben reutilizarse y no deben acceder a los vertederos. Las baterías y acumuladores que contienen plomo, mercurio o cadmio deben tratarse como residuos especiales. Respete en esta relación las normas nacionales de evacuación y reciclaje.
28. Por favor tengan en cuenta que en caso de un incendio pueden desprenderse del producto agentes venenosos (gases, líquidos etc.) que pueden generar daños a la salud.
29. El producto puede poseer un peso elevado. Muévalo con cuidado para evitar lesiones en la espalda u otras partes corporales.
30. No sitúe el producto encima de superficies, vehículos, estantes o mesas, que por sus características de peso o de estabilidad no sean aptas para él. Siga siempre las instrucciones de instalación del fabricante cuando instale y asegure el producto en objetos o estructuras (por ejemplo paredes y estantes).
31. Las asas instaladas en los productos sirven solamente de ayuda para el manejo que solamente está previsto para personas. Por eso no está permitido utilizar las asas para la sujeción en o sobre medios de transporte como por ejemplo grúas, carretillas elevadoras de horquilla, carros etc. El usuario es responsable de que los productos sean sujetados de forma segura a los medios de transporte y de que las prescripciones de seguridad del fabricante de los medios de transporte sean tenidas en cuenta. En caso de que no se tengan en cuenta pueden causarse daños en personas y objetos.
32. Si llega a utilizar el producto dentro de un vehículo, queda en la responsabilidad absoluta del conductor que conducir el vehículo de manera segura. Asegure el producto dentro del vehículo debidamente para evitar en caso de un accidente las lesiones u otra clase de daños. No utilice nunca el producto dentro de un vehículo en movimiento si esto pudiera distraer al conductor. Siempre queda en la responsabilidad absoluta del conductor la seguridad del vehículo. El fabricante no asumirá ninguna clase de responsabilidad por accidentes o colisiones.
33. Dado el caso de que esté integrado un producto de laser en un producto R&S (por ejemplo CD/DVD-ROM) no utilice otras instalaciones o funciones que las descritas en la documentación de producto. De otra manera pondrá en peligro su salud, ya que el rayo laser puede dañar irreversiblemente sus ojos. Nunca trate de descomponer estos productos. Nunca mire dentro del rayo laser.
1171.0000.42-03.00 Sheet 9
Page 16
Page 17
Certified Quality System
DIN EN ISO 9001 : 2000 DIN EN 9100 : 2003 DIN EN ISO 14001 : 2004
DQS REG. NO 001954 QM UM
QUALITÄTSZERTIFIKAT CERTIFICATE OF QUALITY CERTIFICAT DE QUALITÉ
Sehr geehrter Kunde,
Sie haben sich für den Kauf eines Rohde& Schwarz-Produktes entschie­den. Hiermit erhalten Sie ein nach modernsten Fertigungsmethoden hergestelltes Produkt. Es wurde nach den Regeln unseres Management­systems entwickelt, gefertigt und geprüft. Das Rohde& Schwarz Management­system ist zertifiziert nach:
DIN EN ISO 9001:2000 DIN EN 9100:2003 DIN EN ISO 14001:2004
Dear Customer,
you have decided to buy a Rohde& Schwarz product. You are thus as­sured of receiving a product that is manufactured using the most modern methods available. This product was developed, manufactured and tested in compliance with our quality manage­ment system standards. The Rohde& Schwarz quality manage­ment system is certified according to:
DIN EN ISO 9001:2000 DIN EN 9100:2003 DIN EN ISO 14001:2004
Cher Client,
vous avez choisi d‘acheter un produit Rohde& Schwarz. Vous disposez donc d‘un produit fabriqué d‘après les méthodes les plus avancées. Le développement, la fabrication et les tests respectent nos normes de ges­tion qualité. Le système de gestion qualité de Rohde& Schwarz a été homologué conformément aux normes:
DIN EN ISO 9001:2000 DIN EN 9100:2003 DIN EN ISO 14001:2004
1171.0200.11-02.00
Page 18
Page 19
Customer Support
Technical support – where and when you need it
For quick, expert help with any Rohde & Schwarz equipment, contact one of our Customer Support Centers. A team of highly qualified engineers provides telephone support and will work with you to find a solution to your query on any aspect of the operation, programming or applications of Rohde & Schwarz equipment.
Up-to-date information and upgrades
To keep your Rohde & Schwarz equipment always up-to-date, please subscribe to our electronic newsletter at
http://www.rohde-schwarz.com/www/response.nsf/newsletterpreselection
or request the desired information and upgrades via email from your Customer Support Center (addresses see below).
Feedback
We want to know if we are meeting your support needs. If you have any comments please email us and let us know CustomerSupport.Feedback@rohde-schwarz.com.
USA & Canada
East Asia
Rest of the World
Monday to Friday (except US public holidays) 8:00 AM – 8:00 PM Eastern Standard Time (EST)
Tel. from USA 888-test-rsa (888-837-8772) (opt 2) From outside USA +1 410 910 7800 (opt 2) Fax +1 410 910 7801
E-mail Customer.Support@rsa.rohde-schwarz.com
Monday to Friday (except Singaporean public holidays) 8:30 AM – 6:00 PM Singapore Time (SGT)
Tel. +65 6 513 0488 Fax +65 6 846 1090
E-mail Customersupport.asia@rohde-schwarz.com
Monday to Friday (except German public holidays) 08:00 – 17:00
Tel. from Europe +49 (0) 180 512 42 42 From outside Europe +49 89 4129 13776 Fax +49 (0) 89 41 29 637 78
Central European Time (CET)
1171.0200.29-01.00
E-mail CustomerSupport@rohde-schwarz.com
Page 20
Page 21
12
Address List
Headquarters, Plants and Subsidiaries
Headquarters
ROHDE& SCHWARZ GmbH & Co. KG Mühldorfstraße 15 · D-81671 München P.O.Box 80 14 69 · D-81614 München
Plants
ROHDE& SCHWARZ Messgerätebau GmbH Riedbachstraße 58 · D-87700 Memmingen P.O.Box 16 52 · D-87686 Memmingen
ROHDE& SCHWARZ GmbH & Co. KG Werk Teisnach Kaikenrieder Straße 27 · D-94244 Teisnach P.O.Box 11 49 · D-94240 Teisnach
ROHDE& SCHWARZ závod Vimperk, s.r.o. Location Spidrova 49 CZ-38501 Vimperk
ROHDE& SCHWARZ GmbH & Co. KG Dienstleistungszentrum Köln Graf-Zeppelin-Straße 18 · D-51147 Köln P.O.Box 98 02 60 · D-51130 Köln
Subsidiaries
R&S BICK Mobilfunk GmbH Fritz-Hahne-Str. 7 · D-31848 Bad Münder P.O.Box 20 02 · D-31844 Bad Münder
ROHDE& SCHWARZ FTK GmbH Wendenschloßstraße 168, Haus 28 D-12557 Berlin
ROHDE& SCHWARZ SIT GmbH Am Studio 3 D-12489 Berlin
R&S Systems GmbH Graf-Zeppelin-Straße 18 D-51147 Köln
GEDIS GmbH Sophienblatt 100 D-24114 Kiel
HAMEG Instruments GmbH Industriestraße 6 D-63533 Mainhausen
Fax +49 (89) 41 29-121 64
info.rs@rohde-schwarz.com
Phone +49 (83 31) 1 08-0
info.rsmb@rohde-schwarz.com
Phone +49 (99 23) 8 50-0
info.rsdts@rohde-schwarz.com
Phone +420 (388) 45 21 09
Fax +49 (22 03) 49 51-229
info.rsdc@rohde-schwarz.com
service.rsdc@rohde-schwarz.com
Phone +49 (50 42) 9 98-0
info.bick@rohde-schwarz.com
Phone +49 (30) 658 91-122
info.ftk@rohde-schwarz.com
info.sit@rohde-schwarz.com
Phone +49 (22 03) 49-5 23 25
Fax +49 (22 03) 49-5 23 36
info.rssys@rohde-schwarz.com
Phone +49 (431) 600 51-0
Phone +49 (89) 41 29-0
+49 (83 31) 1 08-1124
Fax +49 (99 23) 8 50-174
Fax +420 (388) 45 21 13
Phone +49 (22 03) 49-0
Fax +49 (50 42) 9 98-105
Fax +49 (30) 655 50-221
Phone +49 (30) 658 84-0
Fax +49 (30) 658 84-183
Fax +49 (431) 600 51-11
sales@gedis-online.de
Phone +49 (61 82) 800-0
Fax +49 (61 82) 800-100
info@hameg.de
Locations Worldwide
Please refer to our homepage: www.rohde-schwarz.com
Sales Locations
Service Locations
National Websites
12
1171.0200.42-02.00
Page 22
Page 23
R&S FSP-K90 General Information

1 General Information

The Rohde & Schwarz R&S FSP-K90 application extends the functionality of the R&S FSP spectrum analyzer to enable Wireless LAN TX-measurements according to IEEE standard 802.11a.
This manual supports the user in working with R&S FSP-K90. It aids the preparation, execution and evaluation of a measurement and gives many helpful hints and examples.
For the user wanting to make a quick start to using R&S FSP-K90, the Quick Start Guide section below works step-by-step through an ordinary measurement. The remainder of this section describes all of the basic information about how the R&S FSP-K90 application works, without covering measurements in detail. A detailed description of all measurement modes, settings and results can be found in section 2. Section 4 covers remote control operation of R&S FSP-K90.
This section covers the following subjects:
Introduction to R&S FSP-K90 & Wireless LAN measurements
Installation
Starting the application
Exiting the application
Quick start guide – allows the user to get up-and-running in minimum time
Navigation
Save/recall – saving & recalling user settings & measurement results
Printing
1300.6789.42 1.1 E-3
Page 24
General Information R&S FSP-K90

Introduction to R&S FSP-K90 & Wireless LAN Measurements

The use of an R&S FSP spectrum analyzer with its high sensitivity enables the accurate and reproducible TX-measurement of wireless LAN Device Under Test (DUT) according to the standards specified for the device:
Modulation formats:
- BPSK
- QPSK
- 16QAM
- 64QAM
Modulation measurements:
- Constellation diagram
- Constellation diagram per OFDM carrier
- I/Q offset and I/Q imbalance
- Carrier and symbol frequency errors
- Modulation error (EVM) per OFDM carrier or symbol
- Amplitude response and group-delay distortion (spectral flatness)
Transmit spectrum mask
Capture time selectable up to 4.06 ms, multiple sweeps possible for large number of bursts

Installation

From the analyzer select firmware update. Press the SETUP hardkey followed by NEXT,
FIRMWARE UPDATE and finally the FIRMWARE UPDATE softkey. Following the instructions
displayed.
Once the installation has completed the analyzer will reboot.
Once the option has been installed it needs to be activated:
Start up the analyzer.
Press the SETUP hardkey, followed by the GENERAL SETUP softkey and then the OPTIONS
softkey. A list of the options currently activated is displayed.
Press the INSTALL OPTION softkey. A Dialog is displayed allowing the option key to be entered.
1300.6789.42 1.2 E-3
Page 25
R&S FSP-K90 General Information
Enter the option key supplied with the R&S FSP-K90 software.
When a valid option key has been supplied a dialog will be displayed explaining that a reboot is
required to complete this operation. Select OK in this dialog and the instrument will be rebooted
When the analyzer starts after the reboot a new hotkey will be displayed at the bottom of the display
labelled WLAN. In addition an entry for the R&S FSP-K90 option will be displayed in the FIRMWARE OPTIONS dialog.
1300.6789.42 1.3 E-3
Page 26
General Information R&S FSP-K90

Starting the application

Power up the R&S FSP spectrum analyzer. When R&S FSP-K90 is correctly installed there will be a hotkey labelled WLAN at the bottom of the screen. Press the WLAN hotkey to start R&S FSP-K90.
Note that if the spectrum analyzer is powered down whilst R&S FSP-K90 is active, then when the spectrum analyzer is powered up again it will start up in the R&S FSP-K90 application.

Exiting the application

To exit the R&S FSP-K90 option, press the SPECTRUM hotkey at the bottom of the screen. This will cause the option to exit and the spectrum analyzer to be activated with the same settings as were set when the R&S FSP-K90 option was activated.
1300.6789.42 1.4 E-3
Page 27
R&S FSP-K90 General Information

Quick Start Guide

This section helps the user to quickly become familiar with R&S FSP-K90 by working step-by-step through an ordinary measurement. (Refer to section 2 for a detailed reference guide.)
or this example a DUT using IEEE 802.11a will be used. The DUT will be connected to the analyzer
F using the RF input of the analyzer. The DUT will generate a signal modulated using 16QAM

Setting up the measurement

Start the R&S FSP-K90 application.
Press the GENERAL SETTINGS softkey to open the General Settings view.
Select the Frequency field and enter the desired frequency to measure. Note that when a valid
frequency is entered the Channel No field updates.
Switch Off the Auto Level Field. In this example the level detection measurement will be executed
manually.
All other settings in this view are sufficient for this example.
Press the DEMOD SETTINGS softkey to open the Demod Settings view
1300.6789.42 1.5 E-3
Page 28
General Information R&S FSP-K90
Enter the modulation scheme used in the Demodulator field.
Bursts to Analyze
Burst Type
Demodulator
Direct Link Burst
16QAM
Equal Length Burst
Min No of Data Symbols
Max No of Data Symbols
Close the Demod Settings by pressing the WLAN hotkey.
1
1366
1300.6789.42 1.6 E-3
Page 29
R&S FSP-K90 General Information
Performing Level Detection
Connect the DUT to the RF input of the spectrum analyzer. (see Fig. 1-1 )
RF
DUT
Fig. 1-1 Preparation for level detection
the Level Detection measurement by pressing the AUTO LVL hotkey.
Start
During the level detection measurement the text "Running ..." is displayed in the Status Bar at the bottom of the screen.
After successful level detection, the Status Bar will display “Measurement Complete”, the signal level field for the selected input will be updated to show the detected signal level and the Magnitude Capture Buffer (Screen A) will be updated to show the zero span trace obtained during the measurement sequence.

Performing the main measurement

After level detection has been successfully completed the main measurement can be started. The set-up for the main measurement is the same as that for the level detection measurement. Start the measurement by pressing the RUN SGL hotkey.
During the measurement, the text "Running..." is displayed in the Status Bar at the bottom of the screen.
Measurement results are updated once the measurement has completed. The results are displayed in graphical form. The display can be toggled to a tabular list of measurement points by pressing the DISPLAY softkey
1300.6789.42 1.7 E-3
Page 30
General Information R&S FSP-K90

Navigation

This section deals with navigation within the option. Navigation here is taken to mean all forms of interaction with the option except for remote control. The different methods of interacting with the option are:

Hotkeys

Softkeys
Hardkeys
Numeric Keypad
Roll-key
Cursor Keys
External Keyboard
Mouse
Hotkeys
Hotkeys are allocated to the seven keys at the bottom edge of the screen. On initial start-up of the K90 option, the hotkeys provided are shown in Fig. 1-2 . These hotkeys are present at all times once the option has been started.
SPECTRUM
AUTO LVLWLAN
RUN SGL RUN CONT
SCREEN B
Fig. 1-2 Initial Hotkey menu
A keystroke activates the associated hotkey. An activated hotkey changes colour to green, as shown.
AUTO LVL AUTO LVL
1300.6789.42 1.8 E-3
Page 31
R&S FSP-K90 General Information
These hotkeys perform the following operations:
PECTRUM
S
The SPECTRUM hotkey exits the R&S FSP-K90 option & returns to the spectrum analyzer with all previous settings restored.
LAN
W
AUTO LVL
RUN SGL
RUN CONT
The WLAN hotkey returns the user to the main measurement menu of R&S FSP-K90, where measurement results can be seen. All settings views and dialogs are removed from the display, and the default softkey menu is displayed
The WLAN hotkey remains green whenever R&S FSP-K90 is active
The AUTO LVL hotkey starts a automatic level detection measurement.
If another measurement is running then the running measurement will be aborted before the automatic level detection measurement is started.
Pressing the AUTO LVL hotkey whilst an automatic level detection measurement is running causes the measurement to be stopped (aborted).
The RUN SGL hotkey starts the selected measurement in single sweep mode.
If another measurement is running, such as Calibration or a fixed frequency measurement, the running measurement will be aborted before the single sweep measurement is started.
Pressing the RUN SGL hotkey whilst a single frequency range measurement is running causes the measurement to be stopped (aborted).
The RUN CONT hotkey starts the selected measurement in continuous sweep mode
If another measurement is running then the running measurement will be aborted before the continuous sweep measurement is started. Pressing the RUN CONT hotkey whilst a continuous sweep measurement is running causes the measurement to be stopped (aborted).
REFRESH
The REFRESH hotkey updates the current measurement results to reflect the current measurement settings. The REFRESH hotkey is available for all IQ measurement. The REFRESH hotkey becomes available only when IQ data is available.
SCREEN [A|B]
The SCREEN [A|B] hotkey selects the specified screen as the active screen. In full screen mode pressing the SCREEN [A|B] hotkey will display the specified screen. After Pressing the SCREEN [A|B] hotkey the label displayed in the hotkey is changed e.g. after pressing the SCREEN A hotkey the label of the hotkey is changed to SCREEN B. The label indicates which screen will become the active screen after the hotkey is pressed.
1300.6789.42 1.9 E-3
Page 32
General Information R&S FSP-K90

Softkeys

Settings Softkeys
The softkeys are assigned to the nine keys on the right-hand side of the display. These enable quick access to all of the parameter settings and measurement screens of the K90 option. Each of the top two softkeys, when pressed, brings up a settings view for a group of parameters. These softkeys are always available (except when using Save/Recall and Print manager or controlling markers) and are as follows:
GENERAL
SETTINGS
DEMOD
SETTINGS
DISPLAY
GRAPH
LIST
EVM
SPECTRUM
CONSTELL
General settings
Trigger and IQ settings
D
emod settings -specifies details about the type of
burst to measure
- Signal characteristics, Data Capture,
Fig. 1-3 Main softkeys
Each of these groups of settings is described in detail in the Measurements & Settings section of this manual.
Other Softkeys
All other softkeys have different functions depending on the instrument state. Therefore, the labels (text) on the softkeys will vary to reflect their current function. The state of the softkeys is indicated by different appearances and colours, as follows:
1300.6789.42 1.10 E-3
Page 33
R&S FSP-K90 General Information
SOFTKEY
LABEL 1
SOFTKEY
LABEL 2
SOFTKEY
LABEL 3
SOFTKEY
LABEL 4
SOFTKEY
LABEL 5
SOFTKEY
LABEL 6
SOFTKEY
LABEL 7
SOFTKEY 8
VAL1
VAL2
Softkey available (normal state)
Softkey active (green)
Softkey active and dialog displayed (red)
No softkey available
Softkey disabled => function not available (without 3D Frame)
Toggle softkey (current value of parameter highlighted in green)
Fig. 1-4 Setup of the softkey area
A softkey in its normal state, where its function is available, is coloured grey with a 3D border.
A softkey that is disabled, because its function is not
available, is coloured grey without a 3D border. Softkeys may become disabled because of the state of the instrument or because other settings disable the function associated with the softkey.
An active softkey (highlighted in green) is used when the softkey selects an item or view. For example, the CARRIER SELETION softkey will be highlighted green when the Carrier Selection Dialog is displayed
A toggle softkey is used to change the value of a parameter that has only two states. Each press of the softkey toggles the value of the parameter. The current parameter value is highlighted in green in the lower half of the softkey label. For example, in the measurement results view, the DISPLAY softkey will have either LIST or GRAPH highlighted in green depending on whether the results are currently displayed as a list of measurement points or graphical trace(s).
When no function is assigned to a softkey then no softkey label will be shown.

Hardkeys

Hardkeys allow quick access to the desired parameter and various functions. The hardkeys supported by the K90 option are as follows (other hardkeys do nothing):
FREQ Hardkey
When the FREQ hardkey is pressed the General Settings view is displayed (if it is not already being displayed) and the Frequency parameter is selected.
1300.6789.42 1.11 E-3
Page 34
General Information R&S FSP-K90
AMPT Hardkey
When the AMPT hardkey is pressed the General Settings view is displayed (if it is not already being displayed) and the relevant Signal Level parameter for the selected signal input is selected.
MKR Hardkey
When the MKR hardkey is pressed the main Marker softkey menu is displayed (if it is not already being displayed).
MKR-> Hardkey
When the MKR-> hardkey is pressed the Marker extension softkey menu is displayed (if it is not already being displayed).
SWEEP Hardkey
When the SWEEP hardkey is pressed the General Settings view is displayed (if it is not already being displayed) and the Capture Time parameter is selected.
MEAS Hardkey
When the MEAS hardkey is pressed the Main softkey menu is displayed (if it is not already being displayed).
TRACE Hardkey
When the TRACE hardkey is pressed the General Settings view is displayed (if it is not already being displayed) and the Burst Count parameter is selected.
DISP Hardkey
When the DISP hardkey is pressed the Display softkey menu is displayed (if it is not already being displayed).
FILE Hardkey
When the FILE hardkey is pressed, the Save & Recall softkey menu is displayed, allowing the save & recall of settings and/or measurement results of the R&S FSP-K90 option.
PRESET Hardkey
When the PRESET hardkey is pressed the K90 option is exited and a preset will be performed. Note that all options (including R&S FSP-K90) shall also be preset.
HCOPY Hardkey
When the HCOPY hardkey is pressed the print manager softkey menu is displayed, allowing selection of the items to be printed.
1300.6789.42 1.12 E-3
Page 35
R&S FSP-K90 General Information

External Keyboard

he external keyboard is optional. The keys on the external keyboard that can be used to interact with
T the R&S FSP-K90 option are as follows:
Number keys 0 to 9
Decimal point (“.”) Inserts a decimal point “.” at the cursor position.
Minus key (“-“) Changes the sign of the mantissa or exponent of a numeric parameter. A “-“ is inserted at the cursor position in the case of an alphanumeric parameter.
ESC key Aborts the entry before it has been terminated. The previous value is restored. Closes the entry field after termination of input. Closes pop-up dialogs.
ENTER key Terminates the input of dimension quantities. The new value is set. Invokes the input of parameters or immediately sets the new value. Selects the highlighted item in drop-down menus.
Left and Right Cursor Keys are used to: Navigate between individual parameters within the setting views and some of the pop-up dialogs. Navigate between the individual items within drop-down menus. Move the cursor left & right inside the entry window to reach a particular position in the string during alphanumeric entry.
Up and Down Cursor keys are used to: Navigate between individual parameters within the setting views and some of the pop-up dialogs. Navigate between the individual items within drop-down menus. Increment or decrement the value of a parameter during numeric entry.
CTRL keys Used to activate hotkeys. Each of the seven hotkeys is allocated a different function (F) key. To access these hotkeys press CTRL and the corresponding F key together (see Fig. 1-5 ):
CTRL + F1 CTRL + F2 CTRL + F3 CTRL + F4 CTRL + F5 CTRL + F7
AUTO LVLNOISESPECTRUM
RUN CONT
SCREEN BRUN SGL
Fig. 1-5 Quick Access to Hotkeys
1300.6789.42 1.13 E-3
Page 36
General Information R&S FSP-K90
Function Keys
sed to activate softkeys. Each of the nine softkeys is allocated a different function (F) key. To access
U these softkeys the corresponding F key, as shown below:
SOFTKEY 1
S
OFTKEY 2
SOFTKEY 3
SOFTKEY 4
S
OFTKEY 5
S
OFTKEY 6
SOFTKEY 7
S
OFTKEY 8
SOFTKEY 9
F2
F
F
F5
F6
F
F8
F
F1
3
4
7
9
Fig. 1-6 Quick access to softkeys

Mouse

The mouse can be used to select individual parameters within the settings views or data entry dialogs and to activate hotkeys and softkeys. It can also be used to select values from a drop-down list.

Selecting & Editing Parameters

Parameters are set either by numeric or alphanumeric entry or by simple selection from a list of possible values (a drop-down list is used to select an “enumerated” value) or by using checkboxes to turn a parameter setting on and off.
In all cases, the parameter has to be selected by placing focus on it and then editing has to be enabled before its value can be changed.
The rollkey and cursor keys on the front panel are provided for navigation and selection of parameters.
The numeric keypad, rollkey and cursor keys on the front panel and an external keyboard (optional) are provided for the entry of parameter values.
1300.6789.42 1.14 E-3
Page 37
R&S FSP-K90 General Information
Numeric Keypad
The numeric keypad is provided for entry of numeric parameters. It contains
s
89
7
4
56
12
0
.
ESC
CANCEL
ENTER
OK
GHz
V
-dBm
MHz
dBm
kHz
3
dB
Hz
-
d
B..
BACK
the following keys:
ms
Number keys 0 to 9
m
m
V
V
Bs BV
ns nV
Starts editing of the selected parameter. This enables a new value to be entered for a parameter directly without having to press ENTER first. The digit will be displayed as the first digit of the newly entered value. Inserts a digit at the cursor position when editing an alphanumeric parameter.
Decimal point
Inserts a decimal point “.” at the cursor position.
Sign key (“-“)
Changes the sign of the mantissa or exponent of a numeric parameter. A “-“ is inserted at the cursor position when editing an alphanumeric parameter.
Unit keys (GHz/-dBm, MHz/dBm, kHz/dB and Hz/dB)
Provides the numeric value entered with the selected unit and sets the
parameter to that value.
The unit keys are all assigned the value “1” for dimensionless quantities or for level entries (e.g. in dB). The unit keys thus assume the function of an ENTER key.
BACK key
Deletes the character to the left of the cursor with alphanumeric entry.
ESC/CANCEL key
Aborts the entry of a new parameter value. The previous value is restored.
Closes pop-up dialogs.
ENTER key
Enables editing of the selected parameter (using numeric keys or rollkey). Finishes the editing of a parameter value. The new value is set. For an alphanumeric value, the new value is set to that displayed (using the current unit if applicable).
In a drop-down menu, the parameter is set to the currently selected value in the list.
1300.6789.42 1.15 E-3
Page 38
General Information R&S FSP-K90
Rollkey
The rollkey has various functions:
In a settings view, the rollkey can be used to navigate between individual parameters (the parameter selected being highlighted).
In drop-down menus, the rollkey can be used to navigate between the individual values for the parameter.
During numeric entry, the parameter is incremented (by turning clockwise) or decremented (by turning counter-clockwise) at a defined step size (depending on the parameter).
In setting views and data entry dialogs pressing the rollkey invokes the input of parameters or immediately sets the new value, i.e. pressing the rollkey is like pressing the ENTER key.
In drop-down menus, pressing the rollkey selects the relevant item.
Cursor Keys
The keys
Navigate between individual parameters within the setting views and some of the pop-up dialogs.
Navigate between the individual values within drop-down menus.
Move the cursor left & right inside the entry window to reach a particular
position in the string during alphanumeric entry.
The keys
Navigate between individual parameters within the setting views and some of the pop-up dialogs.
Navigate between the individual items within drop-down menus.
Increment or decrement the value of a parameter during numeric entry.
and are used to:
and are used to:
1300.6789.42 1.16 E-3
Page 39
R&S FSP-K90 General Information
Selection of a parameter within a settings view
Selection using rollkey Press GENERAL SETTINGS softkey for example (General settings view
is displayed).
Rotate the rollkey until reaching the required parameter.
urning the rollkey clockwise selects parameters in the upward direction,
T turning it counter-clockwise selects parameters in the downward direction.
Example: Selecting Signal Level (General settings)
ignal Level (RF)
S
Note: When the Signal Level parameter is selected its label is
highlighted blue.
When the desired parameter is reached press the rollkey to edit the parameter.
Selection using cursor keys
Cursor
, , Within a list of parameters, the Down and Right both move to the next item (down) in the list and the Up and Left keys both move to the previous item (up) in the list. Within a table of parameters, the cursor keys move the cursor in the direction indicated.
To start editing the parameter, either press the ENTER key on the numeric keypad, or press the rollkey.
For numeric parameters, editing can also be started by entering the new value directly from the numeric keypad without pressing the ENTER key first.
Example: Selecting Demodulator (Demod settings)
or
until obtaining the required parameter.
ENTER
OK
Note: When ENTER is pressed, a drop-down menu is displayed,
which contains all the available settings to which the Demodulator parameter can be set.
1300.6789.42 1.17 E-3
Page 40
General Information R&S FSP-K90
Selection using mouse
Selection using external keyboard
Use the mouse to move the cursor to the parameter and press
he left mouse button to select the parameter.
t
To start editing the parameter, either press the ENTER key on
the numeric keypad, or press the rollkey.
For numeric parameters, editing can also be started by entering the new value directly from the numeric keypad without pressing the ENTER key first.
Select parameter using the cursor keys (in the same way as
using the cursor keys on the front panel).
To start editing the parameter, either press the ENTER key on
the numeric keypad, or press the rollkey.
For numeric parameters, editing can also be started by entering the new value directly from the numeric keypad without pressing the ENTER key first.
1300.6789.42 1.18 E-3
Page 41
R&S FSP-K90 General Information
Entry of a numeric value
Once a parameter has been selected (see above), a new value for a numeric parameter can be entered in a number of ways. With the exception of entry via the number keys, to start editing the parameter, either press the ENTER key on the numeric keypad, or press the rollkey before following the instructions below.
If an error occurs, for example, the entered value is out of range, then the new value will not be accepted for the parameter setting.
Entry using number keys (numeric keypad) Example: To enter 200 MHz
Entry using cursor keys
Enter required value using the number keys.
2
0
0
MHz
dBm
Note: The parameter is not set to the new value until either one
of the unit keys on the numeric keypad, the ENTER or the rollkey is pressed.
If the new value is not valid, then a message box is displayed and the entered value will be replaced with a valid value. For example, when a value above the maximum allowed is entered, then the maximum value allowed will be shown in the entry box. The parameter will still be ready for editing so that another value can be entered if desired.
Cursor
or until obtaining the required value.
The application prevents the minimum and maximum values of the parameter from being exceeded and displays an “Out of range“ message box if attempted.
N.B The cursor keys increment/decrement a parameter value in large steps.
Example: Cursor down to 100MHz
Note: Each change of the parameter value takes place
immediately. No other keys need to be pressed.
1300.6789.42 1.19 E-3
Page 42
General Information R&S FSP-K90
up dialogs, where used, do not close automatically.
Entry using rollkey
Entry using external keyboard
Rotate the rollkey until reaching the required value.
Turning the rollkey clockwise increases the value, turning it counter clockwise decreases the value.
he application prevents the minimum and maximum values of the
T parameter from being exceeded and displays an “Out of range“ message box if attempted.
N.B The rollkeys increment/decrement a parameter value in small steps
Example: Rotate to 200MHz
Note: Each change of the parameter value takes place
immediately. No other keys need to be pressed.
Enter value using number keys 0 – 9 in the same way as for using the
number keys on the numeric keypad (see above).
Terminating the entry
Correcting the entry
Aborting the entry
Press one of the unit keys on the numeric keypad.
The unit is entered in the parameter’s edit box and the new parameter value is set immediately.
Press the ENTER key (on numeric keypad or external keyboard) or
press the rollkey.
The new parameter value is set immediately.
Note: Pop-
They can be closed by pressing the ESC key.
In both cases, if the new value is not valid, then a message box is displayed and the entered value will be replaced with a valid value. For example, when a value above the maximum allowed is entered, then the maximum value allowed will be shown in the entry box. The parameter will still be ready for editing so that another value can be entered if desired.
Position the cursor to the right of the digit which is to be deleted using
the cursor keys
Press the BACK key. The digit to the left of the cursor is deleted.
Enter new digits. Each digit is inserted to the left of the cursor, the
other digits are shifted right.
Press the ESC key during parameter editing.
The original parameter value is restored. The new entry is deleted.
If a pop-up dialog is displayed, press the ESC key again.
or .
The entry window is closed, the original value remains active.
1300.6789.42 1.20 E-3
Page 43
R&S FSP-K90 General Information
Entry of an enumerated value
Once a parameter has been selected (see above), a new value for an enumerated parameter can be entered in a number of ways. To start editing the parameter, either press the ENTER key on the numeric keypad, press the rollkey or left click with the mouse on the drop down button before following the instructions below.
Note: When the rollkey or ENTER is pressed, a drop-down menu is displayed, which contains all
the available settings that may be selected for the parameter.
Selection of setting using cursor keys
Cursor
Press ENTER on external keyboard or numeric keypad, or press
rollkey to select the desired setting of parameter.
Note: Currently selected setting of the parameter is highlighted blue. Pressing ENTER sets the new setting of the parameter
or
immediately.
until obtaining the required setting.
Selection of setting using rollkey
Selection of setting using mouse
Selection of setting using external keyboard
Rotate the rollkey until reaching the required setting.
Press rollkey to select setting.
Example: Select Mode parameter.
Note: Currently selected setting of the parameter is highlighted
blue. Pressing the rollkey sets the new setting of the parameter immediately.
When the parameter is selected and ready for editing, select a
new setting using the mouse by left-clicking on the new setting from the drop-down list. The new setting of the parameter is set immediately.
Select setting using cursor keys.
Press ENTER to set the parameter to the new value.
1300.6789.42 1.21 E-3
Page 44
General Information R&S FSP-K90
Entry of a checkbox
A checkbox is used for parameter settings that are either On or Off (Boolean settings). A checkmark () appears in the box when the setting is On; the checkbox is empty when the setting is Off.
Once a parameter has been selected (see above), a new value for a Boolean parameter can be entered in a number of ways. Because Boolean parameters are very simple, it is not necessary to press the
NTER key on the numeric keypad or to press the rollkey in order to edit them.
E
Toggle between the two states of a checkbox using rollkey
Toggle between the two states of a checkbox using numeric keypad
Press the rollkey to toggle between the two states.
Example: Turn Auto Level setting to Off
press
ENTER
OK
Note: The checkbox is empty when the settings is Off
Press the ENTER key to toggle between the two states.
Example: Turn Auto Level setting to On
ENTER
OK
Toggle between the two states of a checkbox using a mouse
Toggle between the two states of a checkbox using external keyboard
Left-click on the checkbox to toggle between the two states.
Press ENTER to toggle between the two states.
1300.6789.42 1.22 E-3
Page 45
R&S FSP-K90 General Information

Status Bar & Title Bar

Title Bar
The title bar is visible at the very top of the display when FSP-K90 is active and no settings views are displayed.
Fig. 1-7 Title Bar
The centre of the title bar shows the wireless LAN standard applicable to the current measurement
Status Bar
The main status bar is displayed at the bottom of the display, just above the hotkeys.
When a parameter in a settings view is selected, the status bar will display the minimum and maximum, settings for the selected parameter (see ).
MIN:<XX.XX> MAX:<XX.XX>
Fig. 1-8 Status Bar
When a parameter whose value is enumerated or Boolean in type is selected in any dialog, the status bar will show “N/A“ displayed for the minimum and maximum, since the minimum and maximum values are “Not Applicable.”
At other times, the status bar shows the current measurement status along with detailed information about the progress through any running measurement.
The status bar is also used to display warning and error messages to the user. In order to highlight these messages, warning messages are displayed with a blue background and error messages with a red background. Refer to Section 5 for a list of warning and error messages.
1300.6789.42 1.23 E-3
Page 46
General Information R&S FSP-K90

Save/Recall

This section of the user manual describes the Save/Recall facility of the option.
F
ILE
The FILE hardkey brings up the Save/Recall softkey menu. Any settings views on display
hen the save/recall softkey menu is displayed shall be closed.
w
Fig. 1-9 Save/Recall softkey menu
The save/recall facility provided by FSP-K90 is exactly the same as that provided by the host analyser. Refer to the user manual for the spectrum analyzer for details of the save/recall facility operation.
The save/recall facility in FSP-K90 provides the following items that can be saved and/or recalled:
Current Settings
All Traces
Raw IQ Data
To close the save/recall softkey menu and return to the main FSP-K90 softkey menu, press the WLAN hotkey.
All user settings provided by FSP-K90
All current trace results
Allows the raw IQ trace results to be stored. When recalled the data is reprocessed to generate results
1300.6789.42 1.24 E-3
Page 47
R&S FSP-K90 General Information

Printing

This section of the user manual describes print facility of the option
HCOPY
The HCOPY hardkey brings up the print softkey menu. Any settings views on display when the print softkey menu is displayed shall be closed.
Fig. 1-10 Print softkey menu
The print facility provided by R&S FSP-K90 is exactly the same as that provided by the host analyser. Refer to the user manual for the spectrum analyzer for details of the print facility operation.
To close the print softkey menu and return to the main R&S FSP-K90 softkey menu, press the WLAN hotkey.
1300.6789.42 1.25 E-3
Page 48
Page 49
R&S FSP-K90 Measurements & Settings

2 Measurements & Settings

This section contains a detailed description of all measurement modes, settings & results. It covers the following subjects:
Measurement modes
Running measurements
Measurement results
General settings
Demodulation settings
Gate settings
Marker settings
General hints about measurements

Measurements

R&S FSP-K90 provides two main measurement types:
IQ Measurements
Frequency Sweep Measurements
1300.6789.42 2.1 E-3
Page 50
Measurements & Settings R&S FSP-K90
1300.6789.42 2.2 E-3
Page 51
R&S FSP-K90 Measurements & Settings

IQ measurements

he following measurements results are obtained in IQ measurement mode:
T
EVM Vs Symbol
EVM Vs Carrier
Phase Vs Preamble
Frequency Vs Preamble
Constellation Vs Symbol
Constellation Vs Carrier
For each of the above results the Magnitude Capture Buffer display is available.
Fig. 2-1 Magnitude Capture Buffer Results
The Magnitude Capture buffer shows the complete range of captured data for the last sweep.
All IQ measurements process the same signal data and as such all IQ measurement results are available after a single IQ measurement execution.
IQ measurements can be run in split screen mode (allowing both the Magnitude Capture Buffer Display and the selected IQ measurement results to be displayed simultaneously) or in full screen mode (with either the Magnitude Capture Buffer Display or the selected IQ measurement results displayed).
1300.6789.42 2.3 E-3
Page 52
Measurements & Settings R&S FSP-K90
EVM Vs Symbol
E
VM VS
SYMBOL
The EVM Vs Symbol measurement results are selected by pressing the EVM softkey in the main measurement softkey menu followed by the EVM Vs SYMBOL softkey.
Fig. 2-2 EVM Vs Symbol
The EVM Vs Symbol results display shows the minimum, average and maximum EVM measured over the full range of the measured input data. The results are displayed on a per-symbol basis, with blue vertical lines marking the boundaries of each burst. Note that burst boundary lines are only displayed if the number of analyzed bursts is less than 250.
The scaling of the Y-Axis can be modified to allow the results to be scaled to an optimum level.
Y AXIS/
DIV
Pressing Y AXIS/ DIV softkey displays a pop-up dialog which allows the step size for the Axis to be defined.
1300.6789.42 2.4 E-3
Page 53
R&S FSP-K90 Measurements & Settings
EVM Vs Carrier
E
VM VS
CARRIER
The EVM Vs Carrier measurement results are selected by pressing the EVM softkey in the main measurement softkey menu followed by the EVM Vs CARRIER softkey.
Fig. 2-3 EVM Vs Carrier Results
The EVM Vs Carrier results display shows the all EVM values recorded on a per-carrier basis over the full set of measured data. An average trace is also displayed. Note the results are restricted to carriers –14 to +14 (inclusive).
The scaling of the Y-Axis can be modified to allow the results to be scaled to an optimum level.
Y AXIS/
DIV
Pressing Y AXIS/ DIV softkey displays a pop-up dialog which allows the step size for the Axis to be defined.
1300.6789.42 2.5 E-3
Page 54
Measurements & Settings R&S FSP-K90
Error Vs Preamble
ERROR
REQ PHASE
F
The Error Vs Preamble results are selected by pressing the EVM softkey in the main measurement softkey menu followed by the ERROR FREQ/PHASE softkey. Subsequent presses of the ERROR FREQ/PHASE softkey toggle the error measurement between Frequency Error Vs Preamble and Phase Error Vs Preamble.
Fig. 2-4 Error Vs Preamble Results
The Error Vs Preamble results display shows the error values recorded over the preamble part of the burst. A minimum, average and maximum trace are displayed. The results display either relative frequency error or phase error.
The scaling of the Y-axis can be modified to allow the results to be scaled to an optimum level.
Y AXIS/
DIV
Pressing the Y AXIS/ DIV softkey displays a pop-up dialog which allows the settings of the Y-axis to be controlled. The settings provided are the same as for the EVM Vs Symbol measurement sceen.
1300.6789.42 2.6 E-3
Page 55
R&S FSP-K90 Measurements & Settings
Constellation Vs Symbol
CONSTELL
The Constellation diagramm is selected by pressing the CONSTELL softkey in the main measurement softkey menu followed by the CONSTELL softkey.
Fig. 2-5 Constellation Vs Symbol Results
The Constellation display shows the Inphase and Quadrature phase results over the full range of the measured input data. The ideal points for the selected modulations scheme are displayed for reference purposes.
The amount of data displayed in the Constellation results display can be reduced by selecting the carrier or carriers for which data is to be displayed.
CARRIER
SELECTION
Pressing CARRIER SELECTION softkey displays a pop-up dialog which allows the carrier for data display to be selected. Either a specific carrier number or pilots only can be selected. Selecting All Carriers allows all the results to be displayed. Note that carrier selection is restricted to carriers –14 to +14 (inclusive).
1300.6789.42 2.7 E-3
Page 56
Measurements & Settings R&S FSP-K90
Constellation Vs Carrier
C
ONSTELL
VS CARRIER
The Constellation Vs Carrier measurement results are selected by pressing the
CONSTELL softkey in the main measurement softkey menu followed by the CONSTELL VS CARRIER softkey.
Fig. 2-6 Constellation Vs Carrier Results
The Constellation Vs Carrier results display shows the Inphase and Quadrature phase results over the full range of the measured input data plotted on a per-carrier basis. The magnitude of the Inphase and Quadrature part is shown on the Y-axis, both are separated by different colors (I-> yellow, Q->blue). Note the results are restricted to carriers –14 to +14 (inclusive)
1300.6789.42 2.8 E-3
Page 57
R&S FSP-K90 Measurements & Settings

Frequency sweep measurements

he following measurements results are obtained in frequency sweep mode:
T
Spectrum Mask
Spectrum ACPR
he frequency sweep measurements use different signal data to IQ measurements and as such it is not
T possible to run an IQ measurement and then view the results in the frequency sweep measurements and vice-versa. Also because each of the frequency sweep measurement use different settings to obtain signal data it is not possible to run a frequency sweep measurement and view the results in another frequency sweep measurement.
All frequency sweep measurements are run in full screen mode.
Spectrum Emission Mask
SPECTRUM
MASK
The Spectrum Emission Mask measurement results are selected by pressing the
SPECTRUM softkey in the main measurement softkey menu followed by the SPECTRUM MASK softkey.
Fig. 2-7 Spectrum Emission Mask Results
The Spectrum Mask results display shows power against frequency. The span of the results is related to the specified sample rate. A limit line representing the spectrum mask specified for the selected standard is displayed and an overall pass/fail status is displayed for the obtained results against this limit line. If the Sweep Count (Mask/ACP) parameter in the General Settings view is set to any value other than 1 then the measurement is performed over the specified number of sweeps. When the measurement is performed over multiple sweeps a max hold trace is displayed as well as an average trace.
1300.6789.42 2.9 E-3
Page 58
Measurements & Settings R&S FSP-K90
The Spectrum Emission Mask measurement can be configured from the SEM settings
iew:
v The SEM SETTINGS softkey brings up the SEM Settings view.
Fig. 2-8 ACP Settings view
IEC/IEEE-bus command: SENS:POW:SEM USER|IEEE|ETSI SENS:POW:SEM:CLASs
SEM according to
File Name
SEM according to specifies how the Spectrum Emission Mask settings and limits are applied. This parameter provides the following settings:
ETSI – Settings and limits are as specified in the standard
IEEE – Settings and limits are as specified in the standard
User – Settings and limits are configured via an XML file
1300.6789.42 2.10 E-3
Page 59
R&S FSP-K90 Measurements & Settings
SEM Configuration
The SEM configuration shows the settings and limits applied over specified frequency ranges around the TX channel. The settings displayed are dependent on the selected Link Direction and Power Class
Fig. 2-9 SEM Configuration
1300.6789.42 2.11 E-3
Page 60
Measurements & Settings R&S FSP-K90
Spectrum ACPR
S
PECTRUM
ACPR
The Spectrum Mask measurement results are selected by pressing the
SPECTRUM softkey in the main measurement softkey menu followed by the SPECTRUM ACPR softkey.
Fig. 2-10 Spectrum ACPR Results
The Spectrum ACPR (Adjacent Channel Power (Relative)) is similar to the Spectrum Mask measurement, and provides information about leakage into adjacent channels. The results show the relative power measured in the three nearest channels either side of the measured channel. This measurement is the same as the Adjacent Channel Power measurement provided by the Spectrum Analyzer.
1300.6789.42 2.12 E-3
Page 61
R&S FSP-K90 Measurements & Settings

Automatic Level Detection

utomatic level detection allows the Signal Level setting to be calculated from the input signal.
A
Automatic level detection can be performed in one of two ways:
Pressing the AUTO LVL hot-key Setting Auto Level in the General Settings view to ON
Running the Automatic Level Detection by pressing the AUTO LVL hot-key allows the level detection to be run at any time. Once the Automatic Level Detection has completed the trace is displayed in the Magnitude Capture Buffer display.
Running the Automatic Level Detection by Setting Auto Level in the General Settings view to ON means that an Automatic Level Detection measurement is executed before each main measurement sweep. When running the Automatic Level Detection this way no trace for the Automatic Level Detection measurement is displayed. This way of running the Automatic Level Detection will cause measurements to run more slowly and as such it is recommended that Automatic Level Detection is only run in this way when the speed of measurement is not important or when running continuous measurements are being run with a signal which varies in level from one sweep to the next.

Running measurements

To start a measurement, press the RUN SGL hotkey (single) or RUN CONT hotkey (continuous).
A single measurement will complete once the requested number of bursts have been obtained or a single sweep has been completed, depending on the measurement settings. When a measurement is completed in continuous mode then a new measurement will be started.
Note that if one measurement is started whilst another measurement is in progress, for example, a single measurement is started whilst a continuous measurement is in progress, then the first measurement will be aborted and the new measurement started immediately.
During a measurement, the text "Running..." is displayed in the Status Bar at the bottom of the screen. After successful completion of a single measurement, the Status Bar will display “Measurement Complete
1300.6789.42 2.13 E-3
Page 62
Measurements & Settings R&S FSP-K90

Measurement Results

Measurement results
The tabular section below the title bar shows the overall measurement settings used to obtain the current measurement results.
Fig. 2-11 Overall measurement settings summary
The settings summary includes the following information:
Frequency - The frequency of the measured input signal. Note that no value is displayed in this field when Baseband input is selected.
Sweep Mode - Shows “Single” for a single measurement and “Continuous” for a continuous measurement.
Burst Type - The type of burst being analyzed.
Signal Level - The expected mean signal level for the input signal
Trigger Mode - The trigger source used to start the measurement sweep
Modulation - Shows either the settings of the Demodulator or PSDU Modulation to
Analyze from the Demod Settings view depending on which parameter is active.
External Att - The attenuation (positiv values) or gain (negative values) applied to the signal external (ie. before the RF- or IQ- connector of the spectrum analyzer) eg. External Att = 10 dB means that before the RF-connector of the R&S FSP a 10 dB attenuator is used. External Att = -20 dB means that before the RF-connector of the R&S FSP a amplifier with 20dB gain is used.
Trigger Offset - The trigger delay (positive value) or pre-trigger time (negative value) applied to the trigger.
No Of Data Symbols - shows the minimum and maximum number of data symbols that a burst may have if it is to be considered in results analysis.
1300.6789.42 2.14 E-3
Page 63
R&S FSP-K90 Measurements & Settings

Results Summary

he results summary table is displayed for IQ measurements when the display mode is set to List. This
T table shows the overall measurement results and provides limit checking for result values in accordance with the selected standard.
Fig. 2-12 Result Summary Table
Result values which are within the limit as specified by the standard are displayed in green. Result values which are within the limit but inside the margin area around the limit are displayed in yellow. Result values which are outside of the limits specified by the standard are displayed in red with a ‘*’ to the left. Results which have no limits specified by the standard are displayed in white. Limit values are displayed in white (not bold).
The results displayed in this table are for the entire measurement. A specific number of bursts have been requested which requires more than one sweep then the results summary update at the end of each sweep showing the results so far. The number of bursts measured and the number of bursts requested are displayed to show the progress through the measurement.
1300.6789.42 2.15 E-3
Page 64
Measurements & Settings R&S FSP-K90

General Settings

This section of the user manual describes the General Settings view where all settings related to the overall measurement can be modified, that is the Signal Characteristics, Data Capture, Trigger and IQ settings.
GENERAL
SETTINGS
The GENERAL SETTINGS softkey brings up the General Settings view.
Fig. 2-13 General Settings view
The parameters within the General settings view are logically grouped together into:
Signal Characteristics
Data Capture Settings
Trigger Settings
IQ Settings
Advanced Settings
Any parameters that are not available for editing will have a grey background. This usually occurs when one parameter setting makes another parameter invalid, for example if the Trigger Mode is Free Run then none of the parameters below this (Trigger Offset, Power Level and Auto Power Trigger Level) have any meaning, so these parameters are greyed out.
When a particular parameter is selected within the General Settings view the status bar changes to display information about the valid settings for the selected parameter.
1300.6789.42 2.16 E-3
Page 65
R&S FSP-K90 Measurements & Settings

Signal Characteristics

he Signal Characteristics settings are the general settings concerning the level, attenuation and
T frequency of the signal to be measured.
Standard
Standard is used to select the wireless LAN standard to be measured against. This is necessary to ensure that the measurements are performed according to the specified standard with the correct limit values and limit lines.
This parameter provides a list of settings. The list provided depends on the standards which have been installed.
Frequency
Frequency specifies the Center Frequency of the signal to be measured.
When the Frequency is modified the Channel No parameter is updated accordingly.
The Frequency parameter is not editable when Base band input is selected because Frequency only make sense for a RF signal.
Channel No
Auto Level
Channel No specifies the channel to be measured.
When the Channel No is modified the Frequency parameter is updated accordingly.
The Channel No parameter is not editable when Base band input is selected because Channel No only make sense for a RF signa.l
Auto Level selects whether the reference level for measurements is measured automatically (ON) or entered manually by the user (OFF).
When Auto Level is set to ON, R&S FSP-K90 will measure the reference level automatically at the start of each measurement sweep. This ensures that the reference level is always set at the optimal level for obtaining accurate results but will result in slightly increased measurement times.
1300.6789.42 2.17 E-3
Page 66
Measurements & Settings R&S FSP-K90
Ext Att
Ext Att specifies the external attenuation or gain applied to the RF signal. A positive value indicates attenuation, a negative value indicates gain.
All displayed power level values will be shifted by this value.
Signal Level (RF)
Signal Level (RF) specifies the expected level of the RF input signal.
The Signal Level (RF) is updated after an automatic level detection measurement has been executed when RF input is selected.
The Signal Level (RF) parameter is not editable when Base band input is selected because Signal Level (RF) only make sense for an RF signal.

Data Capture Settings

The Data Capture settings specify how much data is to be captured and measured.
Capture Time
Capture Time specifies the time (and therefore the amount of data) to be captured in a single measurement sweep.
Overall Burst Count
Overall Burst Count specifies whether a specified number of bursts are to be captured and analyzed.
When Overall Burst Count is set to OFF then data analysis shall be performed on a single measurement sweep. When Overall Burst Count is set to ON then data analysis may be performed over a number of consecutive sweeps until the required number of bursts has been captured and analyzed.
No of Burst to Analyze
No of Burst to Analyze specifies the number of bursts to be measured.
If the number of bursts of the specified type are not contained in a single measurement sweep then R&S FSP­K90 will continue to perform measurement sweeps until the requested number of bursts of the specified type have been captured. The No of Burst to Analyze parameter is not editable when Overall Burst Count is set to OFF.
1300.6789.42 2.18 E-3
Page 67
R&S FSP-K90 Measurements & Settings

Trigger Settings

he Trigger settings contains all the settings related to the triggering of a measurement sweep.
T
Trigger Mode
Trigger Mode is the source of the trigger for the measurement sweep.
The possible values for the Trigger Mode are:
Free Run - the measurement sweep starts
immediately.
External - triggering via a TTL signal at the input
connector EXT TRIGGER/GATE on the rear panel.
Power - the measurement sweep starts when the
signal power meets or exceeds the specified power trigger level.
Trigger Offset
Power Level (RF)
Auto Power Trigger Level
Trigger Offset specifies the time offset between the trigger signal and the start of the sweep. A negative value indicates a pre-trigger.
The Trigger Offset parameter is not editable when Trigger Mode is set to Free Run because this indicates that the measurement sweep should trigger immediately and as such a trigger delay or pre-trigger would not be appropriate.
Power Level (RF) Specifies the trigger level when a power trigger is selected and RF input is selected.
The Power Level parameter is editable only when Trigger
Mode is set to Power.
Auto Power Trigger Level selects whether the level for the
power trigger is measured automatically (ON) or entered manually by the user (OFF).
When Auto Power Trigger Level is set to ON, R&S FSP­K90 will measure and determine the power trigger level automatically at the start of each measurement sweep. This ensures that the power trigger level is always set at the optimal level for obtaining accurate results but will result in a slightly increased measurement times.
The Auto Power Trigger Level parameter is editable only when Trigger Mode is set to Power.
1300.6789.42 2.19 E-3
Page 68
Measurements & Settings R&S FSP-K90

IQ Settings

he IQ settings contains settings related the Inphase and Quadrature phase of the input signal.
T
Swap IQ
Swap IQ allows selection between normal and inverted I/Q modulation. The settings for this parameter are: OFF Normal I/Q modulation. ON I and Q signals are interchanged.
1300.6789.42 2.20 E-3
Page 69
R&S FSP-K90 Measurements & Settings

Advanced Settings

he Advanced Settings contain settings that control details about how the instrument operates and how
T measurements are performed.
The advanced settings are displayed on the right-hand side of the general settings view, and are only displayed when the Advanced Settings checkbox is checked.
Fig. 2-14 Advanced Settings
Auto Level
Auto Level selects whether the reference level for measurements is measured automatically (ON) or entered manually by the user (OFF).
The Auto Level advanced setting is the same as the Auto Level in the Signal Characterstics. It is repeated in the Advanced Settings for convenience.
1300.6789.42 2.21 E-3
Page 70
Measurements & Settings R&S FSP-K90
Auto Level time
Auto Level time specifies the sweep tim e to be used for the auto level measurement.
Ref Level
Ref Level specifies the reference level to use when running measurements.
When the Ref Level is modified, the Signal Level parameter is updated accordingly (depending on the currently selected standard and measurement type).
The Ref Level is editable only when Auto Level is switched off.
RF Att
El Att
The RF Att parameters specify the settings for the meachanical attenuator.
The RF Att is editable only when Auto Level is switched off.
When Auto Level is switched on, the RF Att setting is coupled to the Ref Level setting. When Auto Level is switched off, the RF Att setting is independent from the Ref Level setting and must be specified separately.
The El Att parameters specify the settings for the electronic attenuator.
The El Att parameters are:
Mode – Specfies whether the electronic attenuator should be controlled manually (Manual) or by the Auto option
State – Specfies whether the electronic attenuator should be switched ON or OFF.
Settings – Specifies the attenuation value the electronic attenuator should be switched to
When the mode is set to Auto, the state and setting fields will be automatically calcuated by the option, i.e. fields will not be editable.
When the state is switched off, the setting field is non­editable.
When the frequency is set such that the electronic attenuator cannot be used or the electronic attenuator is not installed, all Electronic Attenuator fields are read-only.
1300.6789.42 2.22 E-3
Page 71
R&S FSP-K90 Measurements & Settings
YIG Filter
The YIG Filter parameter specifies how the YIG filter is used during measurements.
The YIG Filter parameters are:
Mode – Specfies whether the YIG filter should be controlled manually (Manual) or by the Auto option
State – Specfies whether the YIG filter should be switched ON or OFF.
When the mode is set to Auto, the state field will be automatically calcuated by the option, i.e. this field will not be editable.
Sample Rate
Sample Rate specifies the sample rate used for IQ measurements
The Sample Rate is editable only when Standard is set to IEEE 802.11a.
High Dynamic
The High Dynamic parameter is only available when the B72 option is installed, and the sample rate is set between
20.4 MHz and 40.8 MHz.
When the High Dynamic parameter is set to ON then measurements are performed using a wide filter in the normal signal path rather than using the B72 option.
The advantage of setting the High Dynamic parameter ON is a higher resolution (because the "normal" signal path uses a 14 bit ADC). The disadvantage is that all signals left or right the spectrum of interest are folded into the spectrum itself.
1300.6789.42 2.23 E-3
Page 72
Measurements & Settings R&S FSP-K90

Gate Settings

he gate settings settings specify the characteristics of the gate to be applied to the measurement.
T Note that when the gate settings dialog is displayed the measurement view automatically swaps to the Magnitude Capture buffer display in order to allow the gate to be set the correct part of the sweep.
Delay
Delay specifies the start point of captured data to be used in results calculation, i.e. the delay from the start of the captured data.
Delay can be specified in the following ways:
Time
Samples
When the delay is specified in one unit, the other unit field is updated accordingly.
When a value is supplied for the gate delay (in any unit) the gate delay line in the Magnitude Capture Buffer (marker with GD) is displayed in the new position.
Length
Link Gate and Marker
Length specifies the amount of captured data to be used in results calculation.
Length can be specified in the following ways:
Time
Sample
When the length is specified in one unit, the other unit field is updated accordingly.
When a value is supplied for the gate length (in any unit) the gate length line in the Magnitude Capture Buffer (marker with GL) is displayed in the new position.
Link Gate and Marker allows the position of the marker and the gate lines to be linked.
When Link Gate and Marker is switched ON the marker is positioned half way between the gate start and the gate end. The marker position alters when the gate is modified, and the gate lines move with the marker when the marker position is altered.
1300.6789.42 2.24 E-3
Page 73
R&S FSP-K90 Measurements & Settings

Demod Settings

This section of the user manual describes the Demod (Short for Demodulation) Settings view where the settings associated with the signal modulation can be modified.
DEMOD
SETTINGS
The DEMOD SETTINGS softkey brings up the Demod Settings view.
Fig. 2-15 Demod Settings view
The Demod settings are logically grouped together into:
Burst to analyze settings
Tracking settings
When a particular parameter is selected within the Demod Settings view the status bar changes to display information about the valid settings for the selected parameter.
1300.6789.42 2.25 E-3
Page 74
Measurements & Settings R&S FSP-K90

Burst To Analyze Settings

he burst to analyze settings specify the characteristics of the bursts to be considered in the
T measurement results. Only those burst which meet the criteria specified in this group will be included in measurement analysis
Burst Type
Burst Type specifies the type of burst which should be included in measurement analysis. Only one burst type can be selected for measurement results.
The settings for this parameter depend on which wireless LAN standard has been set for the Standard parameter in the General Settings view.
Demodulator
Demodulator is used to specify the modulation to be applied to the measured data. If the captured data uses a different modulation scheme than that specified by this parameter then the results will be of limited use.
The settings for this parameter depend on which wireless LAN standard has been set for the Standard parameter in the General Settings view.
Demodulator is only available when Use Signal Symbol Field Content is set to OFF.
Equal Burst Length
No of Data Symbols
Equal Burst Length allows burst with a range of data symbols or bursts with a specific number of data symbols to be selected for measurement analysis.
When Equal Burst Length is set to ON then the next parameter in the group becomes No of Data Symbols and only burst with exactly the number of symbols specified in No of Data Symbols are considered for measurement analysis.
When Equal Burst Length is set to OFF then the next parameter in the group becomes Min No of Data Symbols and bursts with a number of data symbols within the range specified by the Min No of Data Symbols and Max No of Data Symbols parameters take part in measurement analysis.
No of Data Symbols specifies the number of data symbols a burst must have for it to be considered in measurement analysis.
No of Data Symbols is only available when Equal Burst Length is set to ON.
1300.6789.42 2.26 E-3
Page 75
R&S FSP-K90 Measurements & Settings
Min No of Data Symbols
Min No of Data Symbols specifies the minimum number of data symbols a burst must have for it to be considered in measurement analysis.
Min No of Data Symbols is only available when Equal Burst
ength is set to OFF.
L
Max No of Data Symbols
Max No of Data Symbols specifies the maximum number of data symbols a burst must have for it to be considered in measurement analysis.
Max No of Data Symbols is only available when Equal Burst Length is set to OFF.
Chan Est in Preamb and Payl
Chan Est in Preamb and Payl specifies how accurately the EVM results should be calculated.
When Chan Est in Preamb and Payl is set to OFF then the EVM results are calculated in accordance with the selected standard. In this case channel estimation is done in the the preamble only.
When Chan Est in Preamb and Payl is set to ON then the EVM results are calculated more accurately. In this case channel estimation is done in both preamble and payload.
1300.6789.42 2.27 E-3
Page 76
Measurements & Settings R&S FSP-K90

Tracking Settings

he tracking settings allow various errors in measurement results to be compensated for.
T
Phase
Phase is used to specify whether or not the measurement results should be compensated for phase error.
When Phase is set to ON then the measurement results will be compensated for phase error on a per-symbol basis.
Timing
Timing is used to specify whether or not the measurement results should be compensated for timing error.
When Timing is set to ON then the measurement results will be compensated for timing error on a per-symbol basis.
Level
Level is used to specify whether or not the measurement results should be compensated for level error.
When Level is set to ON then the measurement results will be compensated for level error on a per-symbol basis.
1300.6789.42 2.28 E-3
Page 77
R&S FSP-K90 Measurements & Settings

Markers

This section of the user manual describes the Marker facility of the option.
he MKR hardkey brings up the marker softkey menu. Any settings views on display
MKR
T when the marker softkey menu is displayed shall be closed.
Fig. 2-16 Marker Menu
From the marker menu it is possible to adjust the marker position, use the marker to zoom in & out and switch the marker display off.

Adjusting Markers

The marker can be adjusted by pressing the marker softkey in the marker softkey menu
MARKER 1
The MARKER softkey displays the Marker pop-up dialog. The contents of the marker pop-up depends on the type of graph to which the marker being adjusted belongs.
As soon as an field in the marker pop-up dialog is adjusted then the marker position in the trace will update, along with the results displayed for the marker.
1300.6789.42 2.29 E-3
Page 78
Measurements & Settings R&S FSP-K90

Marker Zoom

zoom facility is available for markers belonging to the following result traces:
A
Magnitude Capture
Constellation Vs Symbol
Constellation Vs Carrier
MARKER
ZOOM
The MARKER ZOOM softkey displays the Marker Zoom pop-up dialog which contains the magnification factor for the zoom.
The maximum magnification depends on the type of results trace.
UNZOOM
The UNZOOM softkey cancels the marker zoom.

Toggle Marker Display

Markers can be toggled by pressing the Marker soft-key as follows:
MARKER 1
MARKER 1
MARKER 1
Pressing the MARKER softkey when the marker is not displayed (softkey has grey background) causes the Marker pop-up to be displayed and the marker to be switched on.
Pressing the MARKER softkey when the Marker pop-up is displayed (softkey has red background) causes the marker to be switched off.
Pressing the MARKER softkey when the Marker is displayed but the Marker pop-up is not displayed (softkey has green background) causes the Marker pop-up to be displayed.
All markers in the active screen can be switched off by pressing the MARKER OFF softkey.
1300.6789.42 2.30 E-3
Page 79
R&S FSP-K90 Measurements & Settings

Assigning Markers to Traces

n result graphs which have more than one trace (e.g. EVM vs. Symbol) the marker can be assigned to
I any of the displayed traces.
MKR->
Pressing the MKR-> hardkey displays the Marker Extension softkey menu.
MKR->TRACE
Fig. 2-17 Marker Extension Softkey Menu
Pressing the MKR->TRACE displays the Select Trace pop-up dialog. This allows the ID of the trace to which the marker is to be attached to be supplied.
1300.6789.42 2.31 E-3
Page 80
Measurements & Settings R&S FSP-K90

Display Settings

The layout of the display can be controlled using the display menu
he DISPLAY hardkey brings up the display softkey menu. Any settings views on display
DISPLAY
T when the display softkey menu is displayed shall be closed.
Fig. 2-18 Display Menu
The display menu allows the display to be changed between split and full screen display for IQ measurement results. Note that for frequency sweep measurement results (Spectrum Mask and Spectrum ACPR) the display is always full screen.
The active screen can be selected by pressing the SCREEN A/SCREEN B hotkey. A screen needs to be made active in order to control the markers it displays.
In full screen mode the SCREEN A/SCREEN B hotkey also toggles which screen is displayed.
1300.6789.42 2.32 E-3
Page 81
R&S FSP-K90 Measurements & Settings
Fig. 2-19 Display Menu
The display menu allows the display to be changed between split and full screen display for IQ measurement results. Note that for frequency sweep measurement results (Spectrum Mask and Spectrum ACPR) the display is always full screen.
The active screen can be selected by pressing the SCREEN A/SCREEN B hotkey. A screen needs to be made active in order to control the markers it displays.
IIn full screen mode the SCREEN A/SCREEN B hotkey also toggles which screen is displayed.
1300.6789.42 2.33 E-3
Page 82
Page 83
R&S FSP-K90 Measurements in Detail
g
=

3 Measurements in Detail

This section provides a more detailed explanation of the measurements provided by R&S FSP-K90 and provides help for using R&S FSP-K90 to measure the characteristics of specific types of DUT.

Abbreviations

kla,
EVM error vector magnitude of subcarrier k
k
EVM
error vector magnitude of current packet
symbol at symbol l of subcarrier k
signal gain
f frequency deviation between Tx and Rx
l symbol index
]_,1[ Symbolsnofl
symbolsnof _ number of symbols of payload
H channel transfer function of subcarrier k
k
k
K modulation dependant normalization factor
mod
channel index ]32,31[=k
relative clock error of reference oscillator
klr,
subcarrier k of symbol l

Signal Processing of the IEEE802.11a application

This description gives a rough view of the IEEE802.11a application signal processing. Details are disregarded in order to get a concept overview.
A diagram of the interesting blocks is shown in figure 1. First the RF signal is downconverted to the IF
frequency
After bandpass filtering, the signal is sampled by an Analog to Digital Converter (ADC) at a sampling
rate of
=
f 80 MHz which is a multiple of the Nyquist rate (20 MHz). The subsequent digital downconverter
2s
shifts the IF signal to the complex base band. In the next step the base band signal is filtered by a FIR
filter. To get an idea, the rough transfer function is plotted in the figure. This filter fulfills two tasks: first it
suppresses the IF image frequency, secondly it attenuates the aliasing frequency bands caused by the
subsequent downsampling. After filtering, the sequence is sampled down by the factor of 4. Thus the
sampling rate of the downsampled sequence
the digital part is implemented in an ASIC.
=
f 20.4 MHz. The resulting IF signal )(tr
IF
=
f 81.6 MHz. This digital sequence is resampled to the new sampling frequency of
1s
)(ir is the Nyquist rate of =
is shown on the left-hand side of the figure.
IF
f 20 MHz. Up to this point
3s
1300.6789.42 3.1 E-3
Page 84
Measurements in Detail R&S FSP-K90
=
-j
·kT
IF
S2
e
rIF(t)
~ ~ ~
DC
A
fs1= 81.6 MHz
Resampler
f
= 80 MHz
s2
FIR
16.4 MHz
H
( f )
IR
F
f
0
r(i)
4
f
=20MHz
3
s
packet search :
1.coarse timing
2.fine timing
LS
timing
H
f
k
payload window
coarse
(LS)
frequency
compensation
F
FT
N = 64
ilot
p table
r
,k
l
a
l,k
stimation
e
of
gain, frequency, time
g
l
d
f
,
est
r
l
l
user defined
compensation
ull
f
compensation
r'
l,k
estimate
data symbols
data
a
l,k
channel
estimation
pilots + data
r''
1
l,k
H
k
measurement
of
arameters
PL)
(
H
k
H
k
p
Fig. 3-1
In the lower part of the figure the subsequent digital signal processing is shown. In the first block the
packet search
coarse timing
on cyclic repetition within the LS after
[1]-[3]. Furthermore a coarse estimate
metric in [6]. This can easily be understood because the phase of
frequency offset. As the frequency deviation
is performed. This block detects the Long Symbol (LS) and recovers the timing. The
is detected first. This search is implemented in the time domain. The algorithm is based
64
N samples. Numerous treatises exist on this subject, e.g.
1
fˆ
of the Rx-Tx frequency offset f is derived from the
coarse
*
)()(
Nirir + is determined by the
fcan exceed half a bin (distance between neighbor sub-
carriers) the preceding Short Symbol (SS) is also analyzed in order to detect the ambiguity.
After the coarse timing calculation the time estimate is improved by the fine timing calculation
)LS(
achieved by first estimating the coarse frequency response
ˆ
H , with ]26,26[=k denoting the
k
. This is
channel index of the occupied sub-carriers. First the FFT of the LS is calculated. After the FFT
calculation the known symbol information of the LS sub-carriers is removed by dividing by the symbols.
ˆ
H
The result is a coarse estimate
of the channel transfer function. In the next step the complex
k
channel impulse response is computed by an IFFT. Next the energy of the windowed impulse response
(the window size is equal to the guard period) is calculated for every trial time. Afterwards the trail time
of the maximum energy is detected. This trial time is used to adjust the timing.
Now the position of the LS is known and the starting point of the useful part of the first payload symbol
can be derived. In the next block this calculated time instant is used to position the payload window
.
Only the payload part is windowed. This is sufficient because the payload is the only subject of the
subsequent measurements.
In the next block the windowed sequence is compensated
by the coarse frequency estimate
ˆ
f
.
coarse
This is necessary because otherwise inter channel interference (ICI) would occur in the frequency
domain.
1
In this paper the hat generally describes an estimate. Example: xˆis the estimate of x.
1300.6789.42 3.2 E-3
Page 85
R&S FSP-K90 Measurements in Detail
(
)
=
=
The transition to the frequency domain is achieved by an FFT
symbol-wise for every of the
y
b
ith
w
klr,
the symbol index
symbolsnof _ symbols of the payload. The calculated FFTs are described
]_,1[ symbolsnofl = and
of length 64. The FFT is performed
the channel index ]32,31[ !=k .
In case of an additive white Gaussian noise (AWGN) channel the FFT is described by [4], [5]
timing(
commom(
)
klklkl
+
)
phasephasej
kll
,
neHgaKr
+=
kl
,,mod,
with
the modulation-dependant normalization factor
the symbol kla
of sub-carrier k at symbol l ,
,
the gain lg at the symbol l in relation to the reference gain
K ,
mod
at the long symbol (LS),
1=g
the channel frequency response kH at the long symbol (LS),
the common phase drift
the phase
phase of sub-carrier k at symbol l caused by the timing drift (see below),
,kl
phase of all sub-carriers at symbol l (see below),
)timing(
the independent Gaussian distributed noise samples
)common(
l
n
.
l
k
,
The common phase drift in equation (1) is given by
common(
)
dlTfNNphase
+= 2
lrestsl
(1)
(2)
with
80=sN being the number of Nyquist samples of the symbol period,
64
N being the number of Nyquist samples 64
f being the (not yet compensated) frequency deviation,
rest
•ld
being the phase jitter at the symbol l .
In general, the coarse frequency estimate
remaining frequency error
f represents the not yet compensated frequency deviation in klr
rest
fˆ (see figure 1) is not error-free. Therefore the
coarse
N of the useful part of the symbol,
.
,
Consequently the overall frequency deviation of the device under test (DUT) is calculated by
ˆ
fff +=
. Remark: The only motivation for dividing the common phase drift in equation (2)
restcoarse
into two parts is to be able to calculate the overall frequency deviation of the DUT.
d
The reason for the phase jitter
in equation (2) may be different. The nonlinear part of the phase
l
jitter may be caused by the phase noise of the DUT oscillator. Another reason for nonlinear phase jitter
may be the increase of the DUT amplifier temperature at the beginning of the burst. Please note that
besides the nonlinear part the phase jitter
caused by the not yet compensated frequency deviation
that the measurement of the phase starts at the first symbol
frequency response
H in equation (1) represents the channel at the long symbol of the preamble.
k
Consequently the not yet compensated frequency deviation
d
also contains a constant part. This constant part is
l
f . To understand this, please keep in mind
rest
1=l of the payload. In contrast the channel
f produces a phase drift between the
rest
long symbol and the first symbol of the payload. Therefore this phase drift appears as a constant value
(“DC value”) in
d
.
l
Referring to the IEEE802.11a measurement standard Chapter 17.3.9.7 “Transmit modulation accuracy
test” [6], the common phase drift
phase must be estimated and compensated from the pilots.
)common(
l
1300.6789.42 3.3 E-3
Page 86
Measurements in Detail R&S FSP-K90
=
=
M
Therefore this “symbol wise phase tracking” (Tracking Phase) is activated as the default setting of the
R&S FSP-K90.
Furthermore the timing drift in equation (1) is given by
)timing(
kl
,
s
lkNNphase
=2
(3)
with
being the relative clock deviation of the reference oscillator. Normally a symbol-wise timing jitter
is negligible and thus not modelled in equation (3). There may be situations where the timing drift has to
be taken into account. This is illustrated by an example: In accordance to [6] the allowed clock deviation
of the DUT is up to
assumed. From equations (1) and (3), it results that the phase drift of the highest sub-carrier
the last symbol
=
20 ppm. Furthermore a long packet with 400_ =symbolsnof symbols is
max
26
k in
symbolsnofl _= is 93 degrees. Even in the noise-free case, this would lead to symbol
errors. The example shows that it is actually necessary to estimate and compensate the clock deviation,
which is accomplished in the next block.
Referring to the IEEE802.11a measurement standard [6], the timing drift
phase is not part of the
)timing(
,
kl
requirements. Therefore the “time tracking” (Tracking Time) is not activated as the default setting of the
R&S FSP-K90.
The time tracking option should rather be seen as a powerful analysing option.
In addition the tracking of the gain
reference gain
function
)L(ˆS
H is calculated. This makes sense since the sequence
k
channel transfer function
gain at the symbol
at the time instant of the long symbol (LS). At this time the coarse channel transfer
1
g
H before estimating the symbols. Consequently a potential change of the
k
l (caused, for example, by the increase of the DUT amplifier temperature) may lead
to symbol errors especially for a large symbol alphabet
g in equation (1) is supported for each symbol in relation to the
l
' is compensated by the coarse
)L(ˆS
klr,
of the MQAM transmission. In this case the
estimation and the subsequent compensation of the gain is useful.
Referring to the IEEE802.11a measurement standard [6], the compensation of the gain
g is not part of
l
the requirements. Therefore the “gain tracking” (Tracking Gain) is not activated as the default setting of
the R&S FSP-K90.
How can the parameters above be calculated? In this application the optimum maximum likelihood
algorithm is used. In the first estimation step the symbol-independent parameters
f and
rest
are
estimated. The symbol dependent parameters can be neglected in this step i.e. the parameters are set
1=
g and 0=
to
l
d
. Referring to equation (1) the log likelihood function
l
2
_
~
(
rest
1
eq.(1)
)~,
symbolsnof
l
=
1
k
21,7,7,21
=
ˆ
,,
()
)LS(
eHarfL
=
kklkl
)common(
~~
hasephasepj
+
2
)timing(
,
kll
with
eq.(2)
~
~
)common(
)timing(
eq.(3)
2
2
sl
~
lTfNNhasep
=
restsl
~
lkNNhasep
=
(4)
2
In this paper the tilde generally describes an estimate. Example: x~is the trial parameter of x.
1300.6789.42 3.4 E-3
Page 87
R&S FSP-K90 Measurements in Detail
~
~
~
must be calculated as a function of the trial parameters
the minimum of the log likelihood function are used as estimates
nown pilot symbols
k
In the second step for every symbol
re read from a table.
a
kla,
l the log likelihood function
f~ and
est
r
. The trial parameters leading to
fˆ and
rest
ˆ
. In equation (4) the
eq.(1)
~
)
,~(
2
21,7,7,21
=
k
~
ˆ
,,
~
()
)L(
S
eHgardgL
=
klklklll
)common(
ˆ
+
h
asephasepj
2
)timing(
kll
,
with
eq.(2)
~
ˆ
is calculated as a function of the trial parameters
)common(
2
eq.(3)
)timing(
2
sl
g
and
d
l
minimum of the log likelihood function are used as estimates
ˆ
ˆ
lkNNhasep
=
. Finally, the trial parameters leading to the
l
ˆ
g
and
l
~
dlTfNNhasep
+=
lrestsl
ˆ
d
.
l
This robust algorithm works well even at low signal to noise ratios with the Cramer Rao Bound being
reached.
After estimation of the parameters, the sequence
is compensated in the compensation blocks.
klr,
In the upper analysing branch the compensation is user-defined i.e. the user determines which of the
parameters are compensated. This is useful in order to extract the influence of these parameters. The
resulting output sequence is described by
r,' .
kl
In the lower compensation branch the full compensation is always performed. This separate
compensation is necessary in order to avoid symbol errors. After the full compensation the secure
estimation of the data symbols
transfer function
)LS(
ˆ
H calculated from the LS. Usually an error free estimation of the data symbols can be assumed.
k
H must be removed. This is achieved by dividing the known coarse channel estimate
k
ˆ
a
is performed. From equation (1) it is clear that first the channel
kl
,
)PL(
In the next block a better channel estimate
using all
symbolsnof _ symbols of the payload (PL). This can be accomplished at this point because the
phase is compensated and the data symbols are known. The long observation interval of
symbols (compared to the short interval of 2 symbols for the estimation of
ˆ
H of the data and pilot sub-carriers is calculated by
k
symbolsnof _
)L(
S
ˆ
H ) leads to a nearly error-
k
free channel estimate.
In the following equalizer block
compensated sequence is described by
)L(
S
ˆ
H (from the long symbol) or the nearly error-free channel estimate
k
equalization. In case of using the improved estimate
r
' is compensated by the channel estimate. The resulting channel-
kl
,
r
'' . The user may either choose the coarse channel estimate
kl
,
)L(
S
ˆ
H a 2 dB reduction of the subsequent EVM
k
)L(
S
ˆ
H (from the payload) for
k
measurement can be expected.
)L(
S
ˆ
According to the IEEE802.11a measurement standard [6], the coarse channel estimation
H (from
k
the long symbol) has to be used for equalization. Therefore the default setting of the R&S FSP-K90 is
equalization from the coarse channel estimate derived from the long symbol.
In the last block the measurement variables
are calculated. The most important variable is the error
vector magnitude
1300.6789.42 3.5 E-3
Page 88
Measurements in Detail R&S FSP-K90
Symbolsnof
k
1
Symbolsnof
_
_
l
''
=
1
2
aKrEVM =
klkl
,mod,
(5)
f the sub-carrier
o
f the current packet. Furthermore the packet error vector magnitude
k o
26
is derived by averaging the squared
1
52
EVM versus k . Finally the average error vector magnitude
k
EVMEVM
=
26
k
=
)0(
k
_
1
_
packetsnof
of all
counter
is calculated by averaging the packet
=
EVM
equivalent to the so-called “RMS average of all errors
commandment (see [6], Chapter 17.3.9.7).
2
k
acketsnof
p
=
packetsnof _
Error ” of the IEEE802.11a measurement
2
1
counterEVMEVM
)(
detected packets. This parameter is
RMS
(6)
(7)
1300.6789.42 3.6 E-3
Page 89
R&S FSP-K90 Remote Control

4 Remote Control

Description of commands

This section specifies all the remote control commands specific to the R&S FSP-K90 option. Only those commands provided for this option are specified. For details of remote control commands provided by the host analyzer please refer to the analyzer user manual.

Notation

In the following sections, all commands implemented in the instrument are first listed in tables and then described in detail, arranged according to the command subsystems. The notation is adapted to the SCPI standard. The SCPI conformity information is included in the individual description of the commands.
Table of Commands
Command: In the command column, the table provides an overview of the
commands and their hierarchical arrangement (see indentations).
Parameter: The parameter column indicates the requested parameters together with
their specified range.
Unit: The unit column indicates the basic unit of the physical parameters.
Comment: In the comment column an indication is made on:
– whether the command does not have a query form,
– whether the command has only one query form
– whether the command is implemented only with a certain option of the instrument
Indentations The different levels of the SCPI command hierarchy are represented in
the table by means of indentations to the right. The lower the level, the further the indentation to the right. Please note that the complete notation of the command always includes the higher levels as well.
Example:
follows:
Individual description The individual description contains the complete notation of the
command. An example for each command, the *RST value and the SCPI information are included as well.
SENSe:FREQuency:CENTer
SENSe
:FREQuency
:CENTer
is represented in the table as
first level second level third level
Upper/lower case notation Upper/lower case letters are used to mark the long or short form of the
key words of a command in the description (see Section 3.5.2). The instrument itself does not distinguish between upper and lower case letters.
1300.6798.42 4.1 E-2
Page 90
Remote Control R&S FSP-K90
Special characters | A selection of key words with an identical effect exists for several
ommands. These keywords are indicated in the same line; they are
c separated by a vertical stroke. Only one of these keywords needs to be included in the header of the command. The effect of the command is independent of which of the keywords is used.
Example:
The two following commands with identical meaning can be created. They set the frequency of the fixed frequency signal to 1kHz:
SENSe:FREQuency:CW 1E3 = SENSe:FREQuency:FIXed 1E3
A vertical stroke in parameter indications marks alternative possibilities in the sense of "or". The effect of the command is different, depending on which parameter is used.
Example: Selection of the parameters for the command
DISPlay:FORMat FULL | SPLit
If parameter FULL is selected, full screen is displayed, in the case of SPLit, split screen is displayed.
SENSe:FREQuency:CW|:FIXed
[ ] Key words in square brackets can be omitted when composing the
header (cf. Section 3.5.2, Optional Keywords). The full command length must be accepted by the instrument for reasons of compatibility with the SCPI standards. Parameters in square brackets can be incorporated optionally in the command or omitted as well.
{ } Parameters in braces can be incorporated optionally in the command,
either not at all, once or several times.
Description of parameters Due to the standardisation, the parameter section of SCPI commands
consists always of the same syntactical elements. SCPI has therefore specified a series of definitions, which are used in the tables of commands. In the tables, these established definitions are indicated in angled brackets (<...>) and will be briefly explained in the following (see also Section 3.5.5, "Parameters").
<Boolean> This keyword refers to parameters which can adopt two states, "on" and
"off". The "off" state may either be indicated by the keyword OFF or by the numeric value 0, the "on" state is indicated by ON or any numeric value other than zero. Parameter queries are always returned the numeric value 0 or 1.
<numeric_value>
1300.6789.42 4.2 E-3
Page 91
R&S FSP-K90 Remote Control
<num> These keywords mark parameters which may be entered as numeric
alues or be set using specific keywords (character data).The following
v keywords given below are permitted:
MINimum This keyword sets the parameter to the smallest possible value.
AXimum This keyword sets the parameter to the largest
M possible value.
DEFault This keyword is used to reset the parameter to its default value.
UP This keyword increments the parameter value.
DOWN This keyword decrements the parameter value.
The numeric values associated to MAXimum/MINimum/DEFault can be queried by adding the corresponding keywords to the command. They must be entered following the quotation mark.
Example: SENSe:FREQuency:CENTer? MAXimum returns the maximum possible numeric value of the center frequency as result.
<arbitrary block program data>
This keyword is provided for commands the parameters of which consist of a binary data block.
1300.6798.42 4.3 E-2
Page 92
Remote Control R&S FSP-K90

ABORt

SCPI Commands

ABORt Subsystem

The ABORt subsystem provide a mechanism by which running measurements can be aborted.
COMMAND PARAMETERS UNIT COMMENT
ABORt
This causes the current measurement, which is being performed, to be aborted.
Example:ABOR” The R&S FSP-K90 option will attempt to abort the current
active measurement.
Characteristics: *RST value: -
SCPI: Conforming

CALCulate: Subsystem

CALCulate:BURSt Subsystem

COMMAND PARAMETERS UNIT COMMENT
:CALCulate<1|2>
:BURSt
[:IMMediate]
CALCulate<1|2>:BURSt[:IMMediate]
This command forces the IQ measurement results to be recalculated according to the current settings.
Example: "CALC:BURS" Forces an update of the IQ results.
Characteristics: *RST value: -
SCPI: Device specific
- -
1300.6789.42 4.4 E-3
Page 93
R&S FSP-K90 Remote Control

CALCulate:LIMit Subsystem

COMMAND PARAMETERS UNIT COMMENT
:CALCulate<1|2> :LIMit<1 to 8> :BURSt :ALL
:RESult? :FERRor [:AVERage] :RESult? :MAXimum :RESult?
:SYMBolerror [:AVERage] :RESult? :MAXimum :RESult? :IQOFfset [:AVERage] :RESult? :MAXimum :RESult? :EVM :ALL [:AVERage] :RESult? :MAXimum :RESult? :DATA [:AVERage] :RESult? :MAXimum :RESult? :PILot [:AVERage] :RESult? :MAXimum :RESult? :SPECtrum :Mask :CHECk :X? :Y? FAIL?
<numeric_value>
--
<numeric_value>
-­<numeric_value>
<numeric_value>
-­<numeric_value>
--
<numeric_value>
-­<numeric_value>
--
<numeric_value>
-­<numeric_value>
--
<numeric_value>
-­<numeric_value>
--
<numeric_value>
-­<numeric_value>
--
--
--
--
S
S
S
S
HZ
HZ
DB
DB
DB|PCT
DB|PCT
DB|PCT
DB|PCT
HZ DBM
query only
query only
query only
query only
query only
query only
query only
query only
query only
query only
query only
query only
query only
query only query only query only
1300.6798.42 4.5 E-2
Page 94
Remote Control R&S FSP-K90
CALCulate<1|2>:LIMit<1>:BURSt:ALL
This command sets or returns all the limit values. The results are input or output as a list of values separated by ’,’ in the following (ASCII) format:
average frequency error>, <max frequency error>,
< <average symbol error>, <max symbol error>, <average IQ offset>, <maximum IQ offset>, <average EVM all bursts>, <max EVM all bursts>, <average EVM data carriers >, <max EVM data carriers > <average EVM pilots >, <max EVM pilots >
Note that the units for the EVM results are specified with the UNITS:EVM command
Example:CALC:LIM:BURS:ALL? All limit values are returned
Characteristics: *RST value: -
SCPI: Device Specific
CALCulate<1|2>:LIMit<1>:BURSt:ALL:RESULT
This command returns all the limit results (PASSED | FAILED). The results are output as a list of result strings separated by ’,’ in the following (ASCII) format:
<average frequency error>, <max frequency error>, <average symbol error>, <max symbol error>, <average IQ offset>, <maximum IQ offset>, <average EVM all bursts>, <max EVM all bursts>, <average EVM data carriers >, <max EVM data carriers > <average EVM pilots >, <max EVM pilots >
Note that the units for the EVM results are specified with the UNITS:EVM command
Example:CALC:LIM:BURS:ALL:RES? All limit values are returned.
Characteristics: *RST value: -
SCPI: Device Specific
CALCulate<1|2>:LIMit<1>:BURSt:FERRor[:AVERage]
This command sets the average frequency error limit in Hertz
Example:CALC:LIM:BURS:FERR 10000” The average frequency error limit is set to 10000 Hz.
Characteristics: *RST value: -
SCPI: Device Specific
CALCulate<1|2>:LIMit<1>:BURSt:FERRor[:AVERage]:RESult
This command returns the average frequency error limit result (PASSED | FAILED)
Example: “CALC:LIM:BURS:FERR:RES?” Average frequency error limit result is returned.
Characteristics: *RST value: -
SCPI: Device Specific
1300.6789.42 4.6 E-3
Page 95
R&S FSP-K90 Remote Control
CALCulate<1|2>:LIMit<1>:BURSt:FERRor:MAXimum
This command sets the maximum frequency error limit in Hertz
xample:CALC:LIM:BURS:FERR:MAX? Maximum frequency error limit is
E returned.
Characteristics: *RST value: -
SCPI: Device Specific
CALCulate<1|2>:LIMit<1>:BURSt:FERRor:MAXimum:RESult
This command returns the maximum frequency error limit result (PASSED | FAILED)
Example:CALC:LIM:BURS:FERR:MAX:RES?” Maximum frequency error limit result is returned.
Characteristics: *RST value: -
SCPI: Device Specific
CALCulate<1|2>:LIMit<1>:BURSt:SYMBolerror[:AVERage]
This command sets the average symbol error limit in Hertz
Example:CALC:LIM:BURS:SYMB 10000” The average symbol error limit is set
to 10 kHz.
Characteristics: *RST value: -
SCPI: Device Specific
CALCulate<1|2>:LIMit<1>:BURSt:SYMBolerror[:AVERage]:RESult
This command returns the average symbol error limit result (PASSED | FAILED)
Example:CALC:LIM:BURS:SYMB:RES? Average symbol error limit result is returned.
Characteristics: *RST value: -
SCPI: Device Specific
CALCulate<1|2>:LIMit<1>:BURSt: SYMBolerror:MAXimum
This command sets the maximum symbol error limit in Hertz.
Example:CALC:LIM:BURS:SYMB:MAX?” Maximum symbol error limit is
returned.
Characteristics: *RST value: -
SCPI: Device Specific
CALCulate<1|2>:LIMit<1>:BURSt:SYMBolerror:MAXimum:RESult
This command returns the maximum symbol error limit result (PASSED | FAILED)
Example:CALC:LIM:BURS:SYMB:MAX:RES?” Maximum symbol error limit result is returned.
Characteristics: *RST value: -
SCPI: Device Specific
1300.6798.42 4.7 E-2
Page 96
Remote Control R&S FSP-K90
CALCulate<1|2>:LIMit<1>:BURSt:IQOFfset[:AVERage]
his command sets the average IQ Offset error limit in dB
T
xample:CALC:LIM:BURS:IQOF –10.0 Average IQ Offset error limit is set
E to –10.0 dB.
Characteristics: *RST value: -
SCPI: Device Specific
CALCulate<1|2>:LIMit<1>:BURSt:IQOFfset[:AVERage]:RESult
This command returns the average IQ Offset error limit result (PASSED | FAILED)
Example:CALC:LIM:BURS:IQOF:RES?” Average IQ Offset error limit result is returned. Characteristics: *RST value: -
SCPI: Device Specific
CALCulate<1|2>:LIMit<1>:BURSt:IQOFfset:MAXimum
This command sets the maximum IQ Offset error limit in dB
Example:CALC:LIM:BURS:IQOF:MAX 15.0” Maximum IQ Offset error limit is to –15.0 dB.
Characteristics: *RST value: -
SCPI: Device Specific
CALCulate<1|2>:LIMit<1>:BURSt:IQOFfset:MAXimum:RESult
This command returns the maximum IQ Offset error limit result (PASSED | FAILED)
Example:CALC:LIM:BURS:IQOF:MAX:RES:?” Maximum IQ Offset error limit result is returned.
Characteristics: *RST value: -
SCPI: Device Specific
CALCulate<1|2>:LIMit<1>:BURSt:EVM:ALL[:AVERage]
This command sets the average Error Vector Magnitude Limit in dB. This is a combined figure that represents the pilot, data and the free carrier.
Example:CALC:LIM:BURS:EVM:ALL –25.0” Average EVM for all carrier limit is set to –25.0 dB.
Characteristics: *RST value: -
SCPI: Device Specific
1300.6789.42 4.8 E-3
Page 97
R&S FSP-K90 Remote Control
CALCulate<1|2>:LIMit<1>:BURSt:EVM:ALL[:AVERage]:RESult
This command returns the average Error Vector Magnitude Limit result (PASSED | FAILED). This is a combined figure that represents the pilot, data and the free carrier.
Example:CALC:LIM:BURS:EVM:ALL:RES?” Average EVM for all carrier limit result is returned.
haracteristics: *RST value: -
C
SCPI: Device Specific
CALCulate<1|2>:LIMit<1>:BURSt:EVM:ALL:MAXimum
This command sets the maximum Error Vector Magnitude Limit in dB. This is a combined figure that represents the pilot, data and the free carrier.
Example:CALC:LIM:BURS:EVM:ALL:MAX?” Maximum EVM for all carrier limit is returned.
Characteristics: *RST value: -
SCPI: Device Specific
CALCulate<1|2>:LIMit<1>:BURSt:EVM:ALL:MAXimum:RESult
This command returns the maximum Error Vector Magnitude Limit result (PASSED | FAILED). This is a combined figure that represents the pilot, data and the free carrier.
Example:CALC:LIM:BURS:EVM:ALL:MAX:RES?” Maximum EVM for all carrier limit result is returned.
Characteristics: *RST value: -
SCPI: Device Specific
CALCulate<1|2>:LIMit<1>:BURSt:EVM:DATA[:AVERage]
This command sets the average Error Vector Magnitude limit summary for the data carrier in dB.
Example:CALC:LIM:BURS:EVM:DATA –30.0” Average EVM for data carrier limit is set to –30.0 dB.
Characteristics: *RST value: -
SCPI: Device Specific
CALCulate<1|2>:LIMit<1>:BURSt:EVM:DATA[:AVERage]:RESult
This command returns the average Error Vector Magnitude limit result summary (PASSED | FAILED) for the data carrier in dB.
Example:CALC:LIM:BURS:EVM:DATA:RES?” Average EVM for data carrier limit result is returned.
Characteristics: *RST value: -
SCPI: Device Specific
1300.6798.42 4.9 E-2
Page 98
Remote Control R&S FSP-K90
CALCulate<1|2>:LIMit<1>:BURSt:EVM:DATA:MAXimum
This command sets the maximum Error Vector Magnitude limit summary for the data carrier in dB.
xample:CALC:LIM:BURS:EVM:DATA:MAX? Maximum EVM for data burst limit is
E
returned. Characteristics: *RST value: -
SCPI: Device Specific
CALCulate<1|2>:LIMit<1>:BURSt:EVM:DATA:MAXimum:RESult
This command returns the maximum Error Vector Magnitude limit result summary (PASSED | FAILED) for the data carrier in dB.
Example:CALC:LIM:BURS:EVM:DATA:MAX:RES?” Maximum EVM for data carrier limit result is returned.
Characteristics: *RST value: -
SCPI: Device Specific
CALCulate<1|2>:LIMit<1>:BURSt:EVM:PILot[:AVERage]
This command sets the average Error Vector Magnitude limit summary for the pilot carriers in dB.
Example:CALC:LIM:BURS:EVM:PIL –8.0” Average EVM for pilot carrier limit is set to –8.0 dB.
Characteristics: *RST value: -
SCPI: Device Specific
CALCulate<1|2>:LIMit<1>:BURSt:EVM:PILot[:AVERage]:RESult
This command returns the average Error Vector Magnitude limit result summary (PASSED | FAILED) for the pilot carriers in dB.
Example:CALC:LIM:BURS:EVM:PIL:RES?” Average EVM for pilot carrier limit result is returned.
Characteristics: *RST value: -
SCPI: Device Specific
CALCulate<1|2>:LIMit<1>:BURSt:EVM:PILot:MAXimum
This command sets the maximum Error Vector Magnitude limit summary for the pilot carriers in dB.
Example:CALC:LIM:BURS:EVM:PIL:MAX?” Maximum EVM for pilot carrier limit is returned.
Characteristics: *RST value: -
SCPI: Device Specific
1300.6789.42 4.10 E-3
Page 99
R&S FSP-K90 Remote Control
CALCulate<1|2>:LIMit<1>:BURSt:EVM:PILot:MAXimum:RESult
This command returns the maximum Error Vector Magnitude limit result summary (PASSED | FAILED) for the pilot carriers in dB.
Example:CALC:LIM:BURS:EVM:PIL:MAX:RES?” Maximum EVM for pilot carrier limit result is returned.
haracteristics: *RST value: -
C
SCPI: Device Specific
CALCulate<1|2>:LIMit<1>:SPECtrum:MASK:CHECk:X
This command returns the X-value at the maximum overstepping of the spectrum mask limits.
Example: CALC:LIM:SPEC:MASK:CHEC:X? Returns the frequency at the maximum overstepping.
Characteristics: *RST value: -
SCPI: Device Specific
CALCulate<1|2>:LIMit<1>:SPECtrum:MASK:CHECk:Y
This command returns the Y-value at the maximum overstepping of the spectrum mask limits
Example:CALC:LIM:SPEC:MASK:CHEC:Y? “ Returns the power at the maximum overstepping.
Characteristics: *RST value: -
SCPI: Device Specific
CALCulate<1|2>:LIMit<1>:FAIL?
This command queries the result of the limit check of the limit line indicated in the selected measurement window. It should be noted that a complete sweep must have been performed for obtaining a valid result. A synchronization with *OPC, *OPC? or *WAI should therefore be provided. The result of the limit check responds with 0 for PASS and 1 for FAIL. Note that no limit lines are displayed in screen A and as such all CALCulate1:LIMit:FAIL? Command will return 0. The index for LIMit for specific limit lines is as follows:
Index Limit
1 to 2 These indexes are not used
3 ETSI Spectrum Mask limit line
4 to 5 These indexes are not used
6 IEEE Spectrum Mask limit line
Example: “ "
INIT;*WAI CALC2:LIM1:FAIL?
Starts a new sweep and waits for its end.
Queries the result of the check for limit line 1 in screen B. Characteristics: *RST value: -
SCPI: Device Specific
1300.6798.42 4.11 E-2
Page 100
Remote Control R&S FSP-K90

CALCulate:MARKer Subsystem

he CALCulate:MARKer subsystem checks the marker functions in the R&S FSP-K90 option.
T
CALCulate 1: Screen A CALCulate 2: Screen B
Note: Currently there is only 1 marker is available and it is always on.
COMMAND PARAMETERS UNIT COMMENT
CALCulate<1|2>:
MARKer<1>
:TRACe :X :Y :SYMBol :CARRier : AOFF [:STATe]
<numeric_value> <numeric_value> <numeric_value> <numeric_value> <numeric_value>
<Boolean>
CALCulate<1|2>:MARKer<1>:TRACe
HZ | S | DB PCT | DB
This command assigns the selected marker to the indicated measurement curve in the selected measurement window. This command is only available for the following result displays: Constellation versus Carrier EVM versus Symbol PVT Full Burst PVT Rising / Falling
Example: "CALC2:MARK:TRAC 2" Assigns marker 1 in screen B to trace 2.
"CALC2:MARK:TRAC 5" Assigns marker 1 in screen B to trace 5.
Characteristics: *RST value: 1
SCPI: Device Specific
CALCulate<1|2>:MARKer<1>:X
This command positions the selected marker to the indicated inphase (constellation Vs symbol), frequency (spectrum FFT, spectrum mask), time (magnitude capture buffer, auto level, PVT full burst, PVT rising / falling), power (CCDF), sub-carrier (constellation Vs carrier, EVM Vs carrier, spectrum flatness) or symbol (EVM Vs symbol) in the selected measurement window. This command is query only for the following result displays: Constellation Vs Symbol Constellation Vs Carrier
Example: "CALC:MARK:X 2ms" Positions marker 1 in screen A to time 2 ms.
Characteristics: *RST value: -
SCPI: Device Specific
1300.6789.42 4.12 E-3
Loading...