Rohde&Schwarz FSMR3, FSMR26, FSMR43, FSMR50 Quick Start Guide

R&S
®
FSMR
1313.9698.02
Measuring Receiver
Getting Started
03
The Getting Started describes the following R&S®FSMR models and options:
& R&S FSMR3 1313.9200.03 & R&S FSMR26 1313.9200.26 & R&S FSMR431313.9200.43 & R&S FSMR50 1313.9200.50
The software contained in this product makes use of several valuable open source software packages. For information, see the "Open Source Acknowledgment" on the user documentation CD-ROM (included in delivery). Rohde & Schwarz would like to thank the open source community for their valuable contribution to embed-
ded computing.
09 Rohde & Schwarz GmbH & Co. KG
© 20 81671 Munich, Germany Printed in Germany – Subject to change – Data without tolerance limits is not binding.
®
R&S
is a registered trademark of Rohde & Schwarz GmbH & Co. KG.
Trade names are trademarks of the owners.
The following abbreviations are used throughout this manual:
®
FSMR is abbreviated as R&S FSMR.
R&S
1
Risk of injury and instrument damage
The instrument must be used in an appropriate manner to prevent personal injury or instrument damage.
Do not open the instrument casing.
Read and observe the "Basic Safety Instructions" delivered as
printed brochure with the instrument.
Read and observe the safety instructions in the following sections.
Note that the data sheet may specify additional operating conditions.
Keep the "Basic Safety Instructions" and the product documentation
in a safe place and pass them on to the subsequent users.
Riesgo de lesiones y daños en el instrumento
El instrumento se debe usar de manera adecuada para prevenir descargas eléctricas, incendios, lesiones o daños materiales.
No abrir la carcasa del instrumento.
Lea y cumpla las "Instrucciones de seguridad elementales"
suministradas con el instrumento como folleto impreso.
Lea y cumpla las instrucciones de seguridad incluidas en las
siguientes secciones. Se debe tener en cuenta que las especificaciones técnicas pueden contener condiciones adicionales para su uso.
Guarde bien las instrucciones de seguridad elementales, así como
la documentación del producto, y entréguelas a usuarios posteriores.
Safety Instructions Instrucciones de seguridad Sicherheitshinweise Consignes de sécurité
1171.1307.42 - 05
2
Gefahr von Verletzungen und Schäden am Gerät
Betreiben Sie das Gerät immer ordnungsgemäß, um elektrischen Schlag, Brand, Verletzungen von Personen oder Geräteschäden zu verhindern.
Öffnen Sie das Gerätegehäuse nicht.
Lesen und beachten Sie die "Grundlegenden Sicherheitshinweise",
die als gedruckte Broschüre dem Gerät beiliegen.
Lesen und beachten Sie die Sicherheitshinweise in den folgenden
Abschnitten; möglicherweise enthält das Datenblatt weitere Hinweise zu speziellen Betriebsbedingungen.
Bewahren Sie die "Grundlegenden Sicherheitshinweise" und die
Produktdokumentation gut auf und geben Sie diese an weitere Benutzer des Produkts weiter.
Risque de blessures et d'endommagement de l'appareil
L'appareil doit être utilisé conformément aux prescriptions afin d'éviter les électrocutions, incendies, dommages corporels et matériels.
N'ouvrez pas le boîtier de l'appareil.
Lisez et respectez les "consignes de sécurité fondamentales"
fournies avec l’appareil sous forme de brochure imprimée.
Lisez et respectez les instructions de sécurité dans les sections
suivantes. Il ne faut pas oublier que la fiche technique peut indiquer
des conditions d’exploitation supplémentaires.
Gardez les consignes de sécurité fondamentales et la
documentation produit dans un lieu sûr et transmettez ces documents aux autres utilisateurs.
1171.1307.42 - 05

Customer Support

Technical support – where and when you need it
For quick, expert help with any Rohde & Schwarz equipment, contact one of our Customer Support Centers. A team of highly qualified engineers provides telephone support and will work with you to find a solution to your query on any aspect of the operation, programming or applications of Rohde & Schwarz equipment.
Up-to-date information and upgrades
To keep your instrument up-to-date and to be informed about new application notes related to your instrument, please send an e-mail to the Customer Support Center stating your instrument and your wish. We will take care that you will get the right information.
Europe, Africa, Middle East
North America
Latin America
Asia/Pacific
China
Phone +49 89 4129 12345
customersupport@rohde-schwarz.com
Phone 1-888-TEST-RSA (1-888-837-8772)
customer.support@rsa.rohde-schwarz.com
Phone +1-410-910-7988
customersupport.la@rohde-schwarz.com
Phone +65 65 13 04 88
customersupport.asia@rohde-schwarz.com
Phone +86-800-810-8228 / +86-400-650-5896
customersupport.china@rohde-schwarz.com
1171.0200.22-06.00
R&S FSMR Tabbed Divider Overview

Tabbed Divider Overview

Contents
Safety Instructions Certificate of Quality EU Certificate of Conformity List of R&S Representatives
Manuals for Measuring Receiver R&S FSMR
Tabbed Divider
Chapter 1: Putting into Operation
Chapter 2: Getting Started
Chapter 3: Menu Overview
Index
1166.6204.62 RE E-1
R&S FSMR Documentation Overview

Documentation Overview

The documentation of the R&S FSMR consists of base unit manuals and option manuals. All manu­als are provided in PDF format on the CD-ROM delivered with the instrument. Each software option available for the instrument is described in a separate software manual.
The base unit documentation comprises the following manuals and documents:
Quick Start Guide
Operating Manual
Service Manual
Release Notes
Apart from the base unit, these manuals describe the models and options of the R&S FSMR Measu­ring Receiver that are listed in the documentation overview of the Operating Manual. All other options are described in separate manuals. These manuals are provided on the CD-ROM. For an overview of all options available for the R&S FSMR visit the R&S FSMR Measuring Receiver Internet site. This manual is delivered with the instrument in printed form. It provides the information needed to set up and start working with the instrument. Basic operations and basic measurements are described. The function descriptions are provided in the Operating Manual. The Quick Start Guide includes gen­eral information (e.g. Safety Instructions) and the following chapters:
Quick Start Guide
Chapter 1 describes the control elements and connectors on the front and rear panel as well
as all procedures required for putting the R&S FSMR into operation and integra­tion into a test system.
Chapter 2 gives an introduction to typical measurement tasks of the R&S FSMR which are
explained step by step.
Chapter 3 describes the operating principles, the structure of the graphical interface and of-
fers a menu overview.
Chapter 10 contains an index for the quick start guide.
1166.6204.62 0.1 E-1
R&S FSMR Contents - Preparing for Operation
Contents - Chapter 1 " Preparing for Operation "
1 Preparing for Operation...................................................................................... 1.1
Description of Front and Rear Panel Views ...................................................................................1.1
Front View ................................................................................................................................1.1
Rear View...............................................................................................................................1.11
Getting Started with the Instrument .............................................................................................1.16
Preparing the Instrument for Operation..................................................................................1.16
Setting Up the Instrument.......................................................................................................1.16
Standalone Operation ..................................................................................................1.16
Safety Instruction for Instruments with Tiltable Feet ....................................................1.17
Rackmounting ..............................................................................................................1.17
EMC Safety Precautions ........................................................................................................1.18
Connecting the Instrument to the AC Supply .........................................................................1.18
Switching the Instrument On/Off ............................................................................................1.18
Switching On the Instrument........................................................................................1.19
Startup Menu and Booting ...........................................................................................1.19
Switching Off the R&S FSMR ......................................................................................1.19
Operating the R&S FSMR with the Removable Hard Disk Option R&S FSMR-B18 ...1.19
Power-Save Mode........................................................................................................1.20
Recalling the Most Recent Instrument Settings .....................................................................1.20
Function Test ..................................................................................................................................1.20
Properties Specific to Windows XP ..............................................................................................1.21
Connecting an External Keyboard................................................................................................1.22
Connecting a Mouse ......................................................................................................................1.23
Connecting an External Monitor ...................................................................................................1.24
Connecting a Printer ......................................................................................................................1.25
Selecting a Printer ..................................................................................................................1.25
Installation of Plug&Play Printers ...........................................................................................1.28
Installation of Non-Plug&Play Printers....................................................................................1.28
Local Printer.................................................................................................................1.30
Configuring a Network Printer ................................................................................................1.34
Connection of USB Devices ..........................................................................................................1.36
Installing Windows XP Software...................................................................................................1.38
Windows XP Software Tested for the Instrument ..................................................................1.38
1313.9723.12 I-1.1 E-1
Contents - Preparing for Operation R&S FSMR
Fig. 1-1 Front View
1313.9723.12 I-1.2 E-1
R&S FSMR Front View

1 Preparing for Operation

Chapter 1 describes the controls and connectors of the Spectrum Analyzer R&S FSMR by means of the front and rear view. Then follows all the information that is necessary to put the instrument into operation and connect it to the AC supply and to external devices.
A more detailed description of the hardware connectors and interfaces can be found in chapter 8. Chapter 2 provides an introduction into the operation of the R&S FSMR by means of typical examples of configuration and measurement; for the description of the concept for manual operation and an overview of menus refer to chapter 3.
For a systematic explanation of all menus, functions and parameters and background information refer to the reference part in chapter 4.
For remote control of the R&S FSMR refer to the general description of the SCPI commands, the instrument model, the status reporting system, and command description in chapter 5 and 6.

Description of Front and Rear Panel Views

Front View

1
Display Screen see Chapter 3
2
Softkeys see Chapter 3
3
Keypad for data input
0...9 input numbers
see Chapter 3
. input decimal point
change sign
ESC – close input field (for uncompleted or CANCEL already closed inputs, the original
entry is kept)
– erase the current entry in input field
(beginning of an input)
– close message window (status, error
and warning messages)
ENTER close the data input.
BACK – erase last character input for
uncompleted input
– restore previous input (undo)
1313.9723.12 1.1 E-2
Front View R&S FSMR
Fig. 1-1 Front View
1313.9723.12 1.2 E-2
R&S FSMR Front View
Keypad for data input
GHz s The units keys close the data
-dBm V input and define the multipli- cation factor for each basic unit. MHz ms For dimension-less or dBm mV alphanumeric inputs, the units keys have weight 1.
kHz µs They behave, in this case, like the dB µV ENTER key.
Hz ns dB.. nV
see Chapter 3
4
FREQ Set frequency axis
SPAN Set span
AMPT Set level indication and configure
RF input.
MKR Select and set standard marker and
delta marker functions.
MKR-> Change instrument settings via
markers
MKR Select further marker and delta FCTN marker functions
see Chapter 4
5
BW – Set resolution bandwidth, video
bandwidth and sweep time,
– Set coupling of these parameters
SWEEP Select sweep
MEAS Select and set power measurements
TRIG Set trigger sources
see Chapter 4
1313.9723.12 1.3 E-2
Front View R&S FSMR
Fig. 1-1 Front View
1313.9723.12 1.4 E-2
R&S FSMR Front View
6
Key group for entering data and for cursor movement
Cursor keys – Move the cursor within the input
fields and tables.
Vary the input value. Define the direction of movement
for the roll-key.
Roll-key – Vary input values.
Move markers and limits. Select letters in the help line editor. Move cursor in the tables Close data input (ENTER)
see Chapter 3
7
Connector for the power sensors NRP-Zx
8
Output connector for an external noise source s. Chapter 3
9
CF Card option R&S FSU-B18
1313.9723.12 1.5 E-2
Front View R&S FSMR
Fig. 1-1 Front View
1313.9723.12 1.6 E-2
R&S FSMR Front View
10
AF OUTPUT
Volume control
Head phone connector
PROBE POWER Power supply and coded socket
(+15 V/ -12 V) for accessories
USB
RF INPUT RF input see Chapter 8
NOTICE!
The maximum DC voltage is 50 V, the maximum power is 1 W (=^ 30 dBm) at 10 dB attenuation.
AUDIO INPUT Audio input with 50 J or 1 MJ input impedance
NOTICE!
The input can handle a maximum power of 1 W or a peak voltage of 20 V.
POWER REF Reference source 50 MHz, 1mW
NOTICE!
USB interfaces
s. Chapter 8
The maximum DC voltage is 0.
11
TRACE Select and activate traces and detectors
LINES Set limit lines
DISP Configure display
FILE – Save and recall instrument data
– Configuration of memory media and data
1313.9723.12 1.7 E-2
see Chapter 4
Front View R&S FSMR
Fig. 1-1 Front View
1313.9723.12 1.8 E-2
R&S FSMR Front View
12
13
14
15
HCOPY
Menu-change keys
NEXT Change to side menu
PREV Call main menu
Hotkeys see Chapter 3
ON/STANDBY switch see Chapter 1
Configure and start a print job see Chapters 1 and 4
see Chapter 3
16
17
18
Define general configuration see Chapter 4
SETUP
Record correction data see Chapter 4
CAL
Call default settings see Chapter 4
1313.9723.12 1.9 E-2
Rear View R&S FSMR
20
100- 240 VAC
3.1- 1.3 A
21
625
SCPI
LPT
22
23
24
MONITOR COM
19
625
IEC 2
SCPI
LAN
IEC 2
OUT
I / Q DATA OUT
VIDEO
IF OUT
REF OUT
640 MHz
404.4 MHz
REF OUT
AUX CONTROL
MOUSE
GATE IN
EXT TRIGGER/
SWEEP
25
26
27
28
29
IF OUT
34
33
REF IN
20.4 MHz
32
1...20 MHz
31
30
Fig. 1-2 Rear View
1313.9723.12 1.10 E-2
R&S FSMR Rear View
Rear View
19
nd
IEC/IEEE bus-connector see Chapter 8
2
20
100 - 2 40 VAC
3. 1 - 1 . 3 A
Power switch and AC power connector see Chapter 1
I 0
21
22
23
24
625
SCPI
COM
MONITOR
LPT
IEC/IEEE bus-connector see Chapter 8
Parallel interface connector
see Chapter 8
(printer connector)
Connector for a serial interface
see Chapter 8
(9-pin socket; COM)
Connector for an external monitor see Chapter 8
1313.9723.12 1.11 E-2
Rear View R&S FSMR
20
100- 240 VAC
3.1- 1.3 A
21
625
SCPI
LPT
22
23
24
MONITOR COM
19
625
IEC 2
SCPI
LAN
IEC 2
OUT
I / Q DATA OUT
VIDEO
IF OUT
REF OUT
640 MHz
404.4 MHz
REF OUT
AUX CONTROL
MOUSE
GATE IN
EXT TRIGGER/
SWEEP
25
26
27
28
29
IF OUT
34
33
REF IN
20.4 MHz
32
1...20 MHz
31
30
Fig. 1-2 Rear View
1313.9723.12 1.12 E-2
R&S FSMR Rear View
25
MOUSE
26
Connector for a PS/2 mouse see Chapter 8
USB
27
AUX CONTROL
28
EXT TRIG /
GATE IN
29
Connector for USB see Chapter 8
Connector to control an external generator ((only with option FSP-B10)
Input connector for an external trigger or an
see Chapter 8
external gate signal
SWEEP
Output connector
see Chapter 8
During a sweep saw toth voltage is output which is proportional to frequency
30
Reserved for options
31
REF IN
1...20 MHz
REF OUT
1313.9723.12 1.13 E-2
REF IN Input connector for an external
reference (1 to 20 MHz)
REF OUT Output connector for an internal
reference (10 MHz)
Rear View R&S FSMR
20
100- 240 VAC
3.1- 1.3 A
21
625
SCPI
LPT
22
23
24
MONITOR COM
19
625
IEC 2
SCPI
LAN
IEC 2
OUT
I / Q DATA OUT
VIDEO
IF OUT
REF OUT
640 MHz
404.4 MHz
REF OUT
AUX CONTROL
MOUSE
GATE IN
EXT TRIGGER/
SWEEP
25
26
27
28
29
IF OUT
34
33
REF IN
20.4 MHz
32
1...20 MHz
31
30
Fig. 1-2 Rear View
1313.9723.12 1.14 E-2
R&S FSMR Rear View
32
20.4 MHz
33
34
IF OUT
LAN
IF OUT
404.4 MHz
IF 20.4 MHz OUT Output connector for
see Chapter 8
20.4 MHz IF
IF 404.4 MHz OUT Output connector for
404.4 MHz IF
LAN-Interface
Reserved for options
1313.9723.12 1.15 E-2
Getting Started with the Instrument R&S FSMR

Getting Started with the Instrument

The following section describes how to activate the instrument and how to connect external devices such as printer and monitor.
Chapter 2 explains the operation of the instrument using simple measurement examples.
Optional notice heading
Prior to switching on the instrument, make sure that the following conditions are fulfilled:
The instrument cover is in place and tightly screwed on
Fan
Signal levels at the inputs are within specified limits
Signal outputs are connected correctly and not overloaded.
Any non-compliance may cause damage to the instrument
openings are not obstructed

Preparing the Instrument for Operation

Take the instrument out of the packaging and check whether the
items listed in the packing list and in the lists of accessories are all included.
Remove the two protective covers from the front and rear of the
R&S FSMR and carefully check the instrument for damage.
remove protective caps
Should the instrument be damaged, immediately notify the carrier
and keep the box and packing material.
For further transport or shipment of the R&S FSMR, the original
packing should be used. It is recommended to keep at least the two protective covers of the front and rear panels in order to prevent damage to the controls and connectors.

Setting Up the Instrument

Standalone Operation
The instrument is designed for use under general laboratory conditions. The ambient conditions required at the site of operation are as follows:
The ambient temperature must be in the range specified in
Wrist strap with cord
Buildin g ground
Ground connection of operational site
Heel strap Floor mat
the data sheet.
All fan openings must be unobstructed and the air flow at
the rear panel and at the side-panel perforations must be unimpeded. The distance to the wall should be at least 10 cm.
he
T
mounting surface should be flat.
To avoid damage of electronic components of the DUT
due to electrostatic discharge on manual touch, protective measures against electrostatic discharge are recommended.
1313.9617.12 1.16 E-2
R&S FSMR Getting Started with the Instrument
instrument be guaranteed and reliable operation be ensured. With the feet out, the total
Safety Instruction for Instruments with Tiltable Feet
The feet must be fully folded in or out. Only in this way can the stability of the
load for the feet must not exceed 500 N (own weight and additional units put onto the instrument). These units must be secured against slipping (e.g. by locking the feet of the unit at the top side of the enclosure).
<500N
When shifting the instrument with the feet out, the feet might collapse and fold in. To avoid injuries, the instrument must therefore not be shifted with the feet out.
The instrument can be operated in any position.
Rackmounting
For rack installation, ensure that the air flow at the side-panel perforations and the air exhaust at the rear panel are not obstructed.
The instrument may be installed in a 19" rack by using a rack adapter kit (Order No. see data sheet). The installation instructions are part of the adapter kit.

EMC Safety Precautions

In order to avoid electromagnetic interference (EMI), the instrument may be operated only with all covers closed. Only adequately shielded signal and control cables may be used (see recommended accessories).
1313.9617.12 1.17 E-2
Getting Started with the Instrument R&S FSMR

Connecting the Instrument to the AC Supply

The R&S FSMR is equipped with an AC voltage selection feature and will automatically adapt itself to the applied AC voltage (range: 100 to 240 V AC, 40 to 400 Hz). External voltage selection or adaptation of the fuses are not necessary. The AC power connector is located on the rear panel (see below).
Connect the instrument to the AC power source using the AC
I
o
Power connector
power cable delivered with the instrument.
As the instrument is designed according to the regulations for safety class EN61010, it must be connected to a power outlet with earthing contact.

Switching the Instrument On/Off

Do not power down during booting.
Such a switch-off may lead to corruption of the hard disk files
AC power switch on the rear panel
P o we r s w i t c h
Power connector
I 0
ON/STANDBY switch on the front panel
ON STANDBY
Power switch
Position I = ON
In the I position, the instrument is in standby mode or in ope
ration, depending on the position of the ON/STANDBY
key at the front of the instrument.
Note: The AC power switch may remain ON continuously.
Switch i n g t o O FF i s o n ly required w h e n t h e instrument must be completely removed from the AC power source.
Position O = OFF
The 0 position implies an all-pole disconnection of the
instrument from the AC power source.
Standby switch
The ON/STANDBY switch activates two different operating modes indicated by coloured LEDs:
Operation ON - ON/STANDBY is depressed
The green LED (ON) is illuminated. The instrument is ready for operation. All modules within the instrument are supplied with power.
Shock hazard
In standby mode, the AC power voltage is present within the instrument
STANDBY - ON/STANDBY switch is not pressed.
The yellow LED (STANDBY) is illuminated. Only the power supply is supplied with power and the quartz oven is maintained at normal operating temperature.
1313.9617.12 1.18 E-2
R&S FSMR Getting Started with the Instrument
Switching On the Instrument
In order to switch on the R&S FSMR, set the power switch on the rear panel to position I.
Set the R&S FSMR to operating mode by pressing the ON/STANDBY key on the front panel. The
green LED must be illuminated.
Startup Menu and Booting
After switching on the instrument, a message indicating the installed BIOS version (e.g. Analyzer BIOS Rev. 1.2) appears on the screen for a few seconds.
Subsequently Windows XP is booted first and after that the instrument firmware will boot. As soon as the boot process is finished the instrument will start measuring. The settings used will be the one that was active when the instrument was previously switched off, provided no other device configuration than
FACTORY had been selected with STARTUP RECALL in the FILE menu.
Switching Off the R&S FSMR
Switch the ON/STANDBY key on the front panel to standby mode by pressing it once.
The R&S FSMR will write the current instrument settings to disk before performing a Windows XP shutdown. At the end of the shutdown procedure the power supply will be switched to STANDBY mode.
The yellow LED must be illuminated.
Only when removing the R&S FSMR completely from the AC power source:
Set the power switch at the rear panel to position 0.
Operating the R&S FSMR with the Removable Hard Disk Option R&S FSMR-B18
The R&S FSMR-B18 option allows the instrument to be operated with differently configured hard disks. Additional hard disks are therefore available as option FSU-B19.
Type and source of hazardous situation
Switch off the instrument before you remove and exchange the hard disk. Exchanging the hard disk during operation can produce severe data errors.
1313.9617.12 1.19 E-2
Function Test R&S FSMR
ttings is not possible if the instrument is switched off
Power-Save Mode
Display:
The R&S FSMR offers the possibility of switching on a power-save mode for the screen display. The backlighting will be switched off if no entry is made on the front panel (key, softkey or hotkey as well as spinwheel) during the selected response time.
In order to switch on the power-save mode:
1. Call the DISPLAY - CONFIG DISPLAY submenu to configure the screen display: Press DISP key Press CONFIG DISPLAY softkey
2. Activate the save mode
Press DISPLAY PWR SAVE softkey.
The softkey is highlighted in colour, thus indicating that the power-save mode is on. At the
same time the data entry for the delay time is opened.
3. Define the delay time
Enter the required response time in minutes and confirm the entry using the ENTER key.
The screen will be blanked out after the selected time period has elapsed.
Hard disk:
A power-save mode is preset for the built-in hard disk which is automatically closed down 15 minutes after the last access.

Recalling the Most Recent Instrument Settings

The R&S FSMR stores its current instrument settings onto the hard disk every time it is switched off via the ON/STANDBY key. After each power-on, the R&S FSMR is reloaded with the operational parameters which were active just prior to the last power-off (STANDBY or AC power OFF) or were set with STARTUP RECALL (see Chapter 4 "Saving and Recalling Data Sets").
Storing the current instrument se using the POWER ON switch at the rear panel or when unplugging the mains cord. After power-on the instrument settings stored previously on the hard disk will be loaded in this case.

Function Test

After turning on the AC power, the R&S FSMR will display the following message on the display screen:
Rohde & Schwarz GmbH & Co. KG
After appearance of the above message, a selftest of the controller hardware is performed. Subsequently, the Windows XP controller boots and the measurement screen will appear.
The system self-alignment is activated via CAL key, CAL TOTAL softkey. The individual results of the self-alignment (PASSED / FAILED) can be displayed in the CAL menu (CAL RESULTS). With the aid of the built-in selftest functions (SETUP key, SERVICE, SELFTEST soft keys), the
functional integrity of the instrument can be verified and/or defective modules can be localized.
Analyzer BIOS Vx.y
1313.9617.12 1.20 E-2
R&S FSMR Properties Specific to Windows XP

Properties Specific to Windows XP

The instrument contains the Windows XP operating system. To ensure that the instrument software functions properly, certain rules must be adhered to when using the operating system..
Possible impairment of the instrument function
The instrument is equipped with the Windows XP operating system. It is thus possible to install commercial off-the-shelf software on the instrument. The use and installation of commercial software may impair the instrument functions. We recommend that you only execute programs tested by Rohde & Schwarz with regard to their compatibility to the instrument software. The program packages that have been tested are listed in section Windows XP Software Tested for the Instrument.
Install only service packs approved by Rohde & Schwarz to avoid impairment of instrument functions that could lead to the necessity to repair the instrument.
The user is especially warned against using Service Packs of Windows XP Home or of the Professional Edition, since these Service Packs are not compatible with Windows XP Embedded.
The drivers and programs used on the instrument under Windows XP have been adapted to the instrument. Existing instrument software must only be modified with update software released by Rohde & Schwarz.
Login
Windows XP requires that users identify themselves by entering a user name and password in a login window. The instrument provides a factory-installed auto login function, i.e. login is carried out automatically in the background. The ID used for auto login has administrator rights. As user name and
password instrument (lowercase) are set.
Administrator ID
The ID used by the instrument to perform auto login has administrator rights.
Windows XP Service Packs
The Windows XP operating system on the instrument is supplied with pre-installed XP service packs (SP1 and additional updates/fixes) that are necessary and suitable for operating the instrument.
Opening the Windows XP Start Menu
You can open the Windows XP Start menu by pressing the Windows key (
CTRL+ESC. From the start menu, you can navigate to the submenus by using the mouse or the arrow keys. To return to the measurement screen, activate the R&S Analyzer Interface button in the task bar.
You can also open the Windows Start menu with the OPEN START MENU softkey in the SETUP menu.
) or the key combination
1313.9617.12 1.21 E-2
Connecting an External Keyboard R&S FSMR

Connecting an External Keyboard

Connect the keyboard only when the instrument is switched off (STANDBY). Otherwise, proper functioning cannot be ensured due to interactions with the firmware.
The R&S FSMR allows an external PC keyboard to be connected to one of the USB interfaces on the front panel or on the rear panel.
USB
Keyboards and mouse devices conform to USB Standards 1.1 or 2.0 are suitable for connection to the USB interface.
1313.9617.12 1.22 E-2
R&S FSMR Connecting a Mouse

Connecting a Mouse

To make Windows XP operation easier, the R&S FSMR allows a mouse to be connected to one of the USB interfaces interface on te front panel or on the rear panel.
USB
Microsoft and Logitech mouse types are supported.
The recommended keyboard PSP-Z2 is equipped with a trackball for mouse control. Connecting an additional mouse will cause interface conflicts and lead to malfunctions of the instrument.
After connection the mouse is automatically recognized. Special settings such as mouse cursor speed etc, can be performed in the Windows XP menu START - SETTINGS - CONTROL PANEL - MOUSE. Chapter 8 contains the interface description for the connectors.
1313.9617.12 1.23 E-2
Connecting an External Monitor R&S FSMR

Connecting an External Monitor

The monitor may only be connected when the instrument is switched off (STANDBY). Otherwise, the monitor may be damaged.
Do not modify the screen driver (display type) and display configuration since this will severely affect instrument operation.
The instrument is equipped with a rear-panel MONITOR connector for the connection of an external monitor.
MONITOR
After connecting the external monitor the instrument needs to be rebooted in order to recognize the monitor. After that the measurement screen is displayed on both the external monitor and the instrument. Further settings are not necessary.
1313.9617.12 1.24 E-2
R&S FSMR Connecting a Printerr

Connecting a Printer

A printer can be connected while the instrument is running. The R&S FSMR allows two different printer configurations for printing a hardcopy to be created plus
switchover between these two configurations. The DEVICES table in the HCOPY menu shows the
available selection of installed printers (see section 4.4 "Documentation of Measurement Results").
The interfaces for connecting printers are on the rear panel:
LPT
USB
Chapter 8 contains the interface description for the connectors.
Selecting a Printer
Before a hardcopy can be printed, the printer has to be selected from the "HCOPY" menu.
In the following example, an HP DeskJet 660C printer that was preinstalled for LPT1 is selected as
DEVICE2 for hardcopies of the screen content.
Press the HCOPY key. The HCOPY menu will open.
HCOPY
DEVICE
SETUP
DEVICE
1 2
COLORS
DEVICE
1 2
Press the DEVICE 1/2 softkey.
Device 2 will become the active output
unit.
Note:
If the printer is to be operated as device 1, this step can be omitted.
1313.9617.12 1.25 E-2
Connecting a Printer R&S FSMR
ting a printer (Name),
The list closes and the selected printer
DEVICE
SETUP
Press the DEVICE SETUP softkey. The HARDCOPY SETUP table opens and
displays the selection of output formats. The
current selection "Clipboard" is
highlighted and marked with a dot in the option button.
Use the cursor key
to move the
selection bar to "Printer" and press ENTER.
Windows for selec printing to file (Print to File) and selecting printout orientation (Orientation) are displayed.
Use the cursor key to set the selection
bar to "Name" and press ENTER .
The list of available printer types appears.
Use the cursor key
/ or the spinwheel to move the selection bar to the "HP DeskJet 660C" printer and press
ENTER.
appears in the "Name" field.
Note:
If the desired printer is not available in the selection list, its driver must first be installed. For further information, see sections "Installation of Plug&Play Printers", "Installation of Non-Plug&Play Printers" and "Installation of Network Printers".
1313.9617.12 1.26 E-2
R&S FSMR Connecting a Printerr
A hardcopy of the screen contents will be
VARIATION
Press the cursor key or turn the
spinwheel until the “Close” button is
reached.
Further settings can still be made:
“Print to File” redirects printing to a file. In
this case, the system prompts you for a file name when printing is started.
The selection is activated by pressing
ENTER or the spinwheel.
“Orientation” is used to switch between
portrait and landscape format.
To change the selection, open the list by
pressing ENTER and select the desired
orientation with the cursor key
/ . To
close the list, press ENTER again.
The “Close” button is used to complete the
setup.
Press ENTER as soon as the “Close”
button is available.
The dialog closes. Printing will now be performed according to the selected settings.
PRINT
SCREEN
1313.9617.12 1.27 E-2
Start printing
Press the PRINT SCREEN softkey.
printed.
Connecting a Printer R&S FSMR
Mouse
The factory setting for DEVICE 2 is “Clipboard”. In this case, the printout will be copied to the Windows XP clipboard which is supported by most Windows applications. The contents of the clipboard can be pasted directly into a document via EDIT – PASTE.
Table 1-1 shows the factory settings for the two output devices.
Table 1-1 Factory settings for DEVICE 1 and DEVICE 2 in the HCOPY menu
Setting Selection in
configuration table
Output device DEVICE WINDOWS METAFILE CLIPBOARD
Output PRINT TO FILE YES ---
Orientation ORIENTATION --- ---
Setting for DEVICE 1 Setting for DEVICE 2
Installation of Plug&Play Printers
The installation of Plug&Play printers under Windows XP is quite simple: After the printer is connected and switched on, Windows XP automatically recognizes it and installs its driver, provided the driver is included in the XP installation. If the XP printer driver is not found, Windows XP prompts you to enter the path for the corresponding installation files. In addition to pre-installed drivers, a number of other printer drivers can be found in directory D:\I386.
When you install printer drivers that are not pre-installed on the instrument, the installation process will prompt you to specify the path containing the new driver. This can be a memory stick or USB CD-ROM drive (see section “Connection of USB Devices”).
Installation of Non-Plug&Play Printers
The dialogs below can be controlled either from the front panel or via the mouse and keyboard (see sections “Connecting a Mouse” and “Connecting a Keyboard”). and PC keyboard are absolutely essential for configuring network printers.
A new printer is installed with the INSTALL PRINTER softkey in the HCOPY menu.
Press the HCOPY key. The HCOPY menu will open.
Press the NEXT key to open the side
HCOPY
NEXT
DEVICE
SETUP
DEVICE
1 2
COLORS
menu.
1313.9617.12 1.28 E-2
R&S FSMR Connecting a Printerr
INSTALL PRINTER
Press INSTALL PRINTER to open the
Printers and Faxes dialog window.
Select Add Printer in the list using the
spinwheel.
Highlight the selected item with CURSOR
RIGHT and press ENTER or the
spinwheel to confirm the selection.
The Add Printer Wizard is displayed.
Select NEXT with the spinwheel and press
the spinwheel for confirmation.
Local or Network Printer can be selected.
To install a local printer, select Local
printer attached to this computer with the
spinwheel. Press the spinwheel for confirmation and continue with the “Local Printer” section.
To install a network printer, select A
network printer or a printer attached to another computer. Press the spinwheel
for confirmation and continue with the “Network Printer” section.
1313.9617.12 1.29 E-2
Connecting a Printer R&S FSMR
Local Printer
In the following example, a Star LC24 printer is connected to the LPT1 interface and configured as
DEVICE2 for hardcopies of screen contents. The Add Printer Wizard has already been opened as
described in the section “Starting the Add Printer Wizard” .
To select the USB interface, open the list of
ports by clicking the spinwheel. Select the printer port with spinwheel/arrow keys and confirm by pressing the spinwheel. The selection list is closed again.
o
select the LPT connector, the selection
T
list need not be opened.
Place the cursor on the Next button and
confirm by pressing the spinwheel.
The “Install Printer Software” dialog is
opened.
Select the desired manufacturer (“Star”) in
the Manufacturer table using the up / down keys.
1313.9617.12 1.30 E-2
R&S FSMR Connecting a Printerr
Go to the Printers list with the spinwheel.
Select the desired printer type (Star LC24-
200 Colour) using the up / down keys and
confirm with ENTER.
Note: If the desired printer type is not in the list, the respective driver is not installed yet. In this case click the HAVE DISK button with the mouse key. You will be prompted to insert a disk with the corresponding printer driver. Press OK and select the desired printer driver.
The printer name can be changed as
required in the Printer name entry field
(max. 60 characters). A PC keyboard is required in this case.
Use the spinwheel to select Yes or No for
the default printer.
Choose the desired status with the up
/down keys.
Confirm with ENTER.
The Printer Sharing dialog is opened.
Exit the dialog with ENTER.
The Print Test Page dialog is opened.
1313.9617.12 1.31 E-2
Connecting a Printer R&S FSMR
Exit the dialog with ENTER.
The Completing the Add Printer Wizard
dialog is opened.
Check the displayed settings and exit the
dialog with ENTER.
The printer is installed. If Windows finds the required driver files, the installation is completed without any further queries.
If Windows cannot find the required driver files, a dialog is opened where the path for the files can be entered.
Select the Browse button with the
spinwheel and confirm with by pressing the spinwheel.
The Locate File dialog is opened.
1313.9617.12 1.32 E-2
R&S FSMR Connecting a Printerr
Turn the spinwheel to select the directory
and path D:\I386 and press it to confirm the selection.
If the selected item is not printed on a blue
background, it must be marked with the cursor up / down keys before it can be activated by pressing the spinwheel.
Select the driver file with the spinwheel
and confirm by pressing the spinwheel.
The file is included in the Files Needed
dialog.
Note:
If the desired file is not in the D:\I386 directory, a disk with the driver file is needed. In this case, exit the dialog with ESC and repeat the selection starting from the "Files needed" dialog.
Select the OK button with the spinwheel
and press the spinwheel to confirm.
The installation is completed.
Finally, the instrument must be configured for printout with this printer using the softkeys DEVICE SETUP and DEVICE 1/2 in the hardcopy main menu (see section "Selecting a printer").
1313.9617.12 1.33 E-2
Connecting a Printer R&S FSMR
Configuring a Network Printer
To select a network printer, click the option
"A network printer or a printer attached to another computer".
Continue with Next.
Click Browse for a printer and then Next.
A list of selectable printers is displayed.
Mark the desired printer and select it with
OK.
1313.9617.12 1.34 E-2
R&S FSMR Connecting a Printerr
Confirm the subsequent prompt to install a
suitable printer driver with "OK". The list of available printer drivers is
ayed.
displ The manufacturers are listed in the left­hand table, the available printer drivers in the right-hand table.
Select the manufacturer from the
Manufacturers table and then the printer driver from the Printers table.
Note: If the desired type of output device is not shown in the list, the driver has not yet been installed. In this case, click the "HAVE DISK" button. You will be prompted to insert a disk with the corresponding printer driver. Insert the disk, select "OK" and then choose the desired printer driver.
Click Next.
If one or more printers have already been installed, this window queries whether the printer last installed is to be used as the default printer for the Windows XP
applications. The default selection is No.
Start the printer driver installation with
Finish.
Finally, the instrument has to be configured for printout with this printer using the softkeys DEVICE SETUP and DEVICE 1/2 in the hardcopy menu (see section "Selection of a Printer").
1313.9617.12 1.35 E-2
Connection of USB Devices R&S FSMR

Connection of USB Devices

Up to two USB devices can be directly connected to the analyzer via the USB interface on the rear of the R&S FSMR. This number can be increased as required by interconnecting USB hubs. Owing to the wide variety of available USB devices, the R&S FSMR can be expanded with almost no limitations. The following list shows a selection of USB devices suitable for the R&S FSMR:
Power Sensor R&S NRP-Z11 or R&S NRP-Z21 (Adapter Cable R&S NRP-Z4 required)
Pendrive (memory stick) for easy data transfer from/to the PC (e.g. firmware updates)
CD-ROM drive for easy installation of firmware applications
PC keyboard for entering comments, file names, etc
Mouse for easy operation of Windows dialogs
Printer for documentation of measurement results
Modem for remote control of the R&S FSMR over great distances
The installation of USB devices is quite simple under Windows XP since all USB devices are Plug&Play. Apart from the keyboard and the mouse, all USB devices can be connected to or disconnected from the R&S FSMR while the instrument is running.
After the instrument is connected to the USB interface, Windows XP automatically searches for a suitable device driver.
If Windows XP does not find a suitable driver, you will be prompted to specify a directory where the driver software can be found. If the driver software is on a CD, a USB CD-ROM should first be connected to drive to the R&S FSMR.
As soon as the connection between the R&S FSMR and the USB device is interrupted, Windows XP will again recognize the modified hardware configuration and will deactivate the corresponding device driver.
Example:
Connecting a pendrive (memory stick) to the R&S FSMR:
1. After the pendrive is connected to the USB interface, Windows XP will recognize the newly connected hardware:
2. Windows XP installs the corresponding driver. After successful installation, XP signals that the unit is ready for operation:
3. The pendrive is now available as a new drive and is displayed in Windows Explorer:
1313.9617.12 1.36 E-2
R&S FSMR Connection of USB Devices
The pendrive can be used as a normal drive to load or save files.
4. If the pendrive is no longer required or if files are to be transferred to another computer, the pendrive is simply disconnected. Windows XP will then deactivate the driver. If the corresponding drive is still selected in Explorer, an error message will be displayed indicating that the drive is no longer available.
1313.9617.12 1.37 E-2
Installing Windows XP Software R&S FSMR

Installing Windows XP Software

Windows XP Software Tested for the Instrument
The driver software that is used and the system settings of Windows XP have been fine-tuned to support the measurement functions of the R&S FSUP. Thus, flawless instrument function is ensured only if software and hardware tested or offered by Rohde & Schwarz are used.
Using other software or hardware may cause the functions of the R&S FSMR to perform improperly or fail.
The following program packages have been successfully tested for compatibility with the measurement instrument software:
R&S FS-K3 – Software for measuring noise factor and gain
R&S FS-K4 – Software for measuring phase noise
R&S Power Viewer (virtual power meter for displaying results of the power sensors of the R&S NRP-
Zxx series)
Windows XP Remote Desktop
FileShredder – For reliable deletion of files on the hard disk
Symantec Norton AntiVirus – Virus-protection software
McAfee Virusscan
1313.9617.12 1.38 E-2
R&S FSMR Contents - Getting Started
Contents – Chapter 2 "Getting Started"
2 Getting Started .................................................................................................... 2.1
Instrument overview ........................................................................................................................2.1
Setup the instruments......................................................................................................................2.1
Select the Preset state .....................................................................................................................2.2
Measurement Examples...................................................................................................................2.3
Example 1: RF Power Measurements with a Power Sensor ...................................................2.4
Measurement .................................................................................................................2.4
Main Receiver Functions................................................................................................2.4
Measurement Setup.......................................................................................................2.4
Measurement Sequence – RF Power Measurements ...................................................2.4
Example 2: Low RF Level Measurements with Tuned RF Level Mode ...................................2.7
Measurement .................................................................................................................2.7
Main Receiver Functions................................................................................................2.7
Measurement Setup.......................................................................................................2.7
Measurement Sequence – Tuned RF Level Measurements .........................................2.7
Example 3: AM Modulation and Modulation Distortion Measurements..................................2.13
Measurement ...............................................................................................................2.13
Main Receiver Functions..............................................................................................2.13
Measurement Setup.....................................................................................................2.13
Measurement Sequence – AM Modulation Measurements .........................................2.13
Example 4: Audio Measurements with the Audio Analyzer mode..........................................2.17
Measurement ...............................................................................................................2.17
Main Receiver Functions..............................................................................................2.17
Measurement Setup.....................................................................................................2.17
Measurement Sequence – Audio Measurements........................................................2.17
1166.3311.12 I-2.1 E-1
R&S FSMR Instrument overview

2 Getting Started

This chapter provides a fast introduction to operation by guiding the user step by step through measurement examples .
Before starting any measurement with the FSMR, please note the instructions given in chapter 1 of the users manual for putting the instrument into operation. In chapter 3 you will find detailed information on customizing the instrument and the display.
For a systematic explanation of all menus, functions and parameters and background information refer to the reference part in chapter 4.

Instrument overview

The R&S FSMR is a very versatile instrument that combines many measurement capabilities required for calibration, troubleshoot and development into a single box instrument. The following functions are included:
RF Power Meter, for highest accuracy RF power measurement.
RF Level Meter, for measurements of low level signals.
Modulation Analyzer, for AM/FM and 4M modulated signals.
Audio Analyzer, for Audio Signal level and distortion, using the Audio input.
Frequency Counter, fastest frequency measurements with „mHz“-resolution.
Spectrum Analyzer, for high performance spectral analysis capabilities.
The R&S FSMR still offers an easy-to-use operation which does not require expert knowledge to perform the basic measurements.
This product note is designed to explain the functions and operation of the Measuring Receiver R&S FSMR. In the following descriptions, each step is explained in detail so that the instrument can be immediately used without the need for learning all of the available functions.

Setup the instruments

Most of the following examples use the same test setup. To perform the measurements, in addition to the Measuring Receiver R&S FSMR a signal generator and a power sensor is required. The examples in this paper will require a single, stable RF test signal at lower RF frequencies (for example 100 MHz). The generator shall offer analog modulation capabilities (at least AM modulation) at a variable modulation rate from 1 kHz to 3 kHz and 30 % modulation depth. For the audio measurement examples a LF generator is required. Most of the R&S signal generators are equipped with a modulation generator which is sufficient for these measurements (for example R&S SML, R&S SMIQ). The power sensor used in the examples shall be one of the sensors from the R&S NRP series. There are many different sensors available, a terminating average power sensor will best fit to the requirements (for example R&S NRP-Z21)
1166.3311.12 2.1 E-2
Select the Preset state R&S FSMR

Select the Preset state

All of the following examples assume the standard settings for the measuring receiver. These are set with the PRESET key. The PRESET must be set to Factory default values and measuring Receiver Mode.
Check the PRESET setting with the following steps:
SETUP
NEXT
PRESET
RECEIVER
FILE
STARTUP
RECALL
1. Define Preset instrument mode
Press the SETUP key.
The setup menu is opened.
Press the menu change key NEXT.
The submenu is opened.
Press the PRESET RECEIVER softkey.
The following presets will force the instrument into the Measuring Receiver mode.
2. Define the Startup Recall
Press the FILE key.
The file menu is displayed.
In the FILE menu press the STARTUP RECALL softkey.
A table with available instrument recall states is displayed.
In the STARTUP table, highlight the entry named “FACTORY”
and enable it with the Enter key as STARTUP RECALL.
PRESET
3. Reset the instrument.
Press the PRESET key.
The main receiver menu is opened.The receiver mode is set.
1166.3311.12 2.2 E-2
R&S FSMR Measurement Examples
Fig.2-1 Display after selecting the default setup in receiver mode
The main default parameters are listed in the following table:
Table 2-1 Default parameters after preset in measuring recveiver mode
Parameter Parameter Name Value
Receiver frequency Frequency 100 MHz
Reference Level Ref Level Manual -20 dBm
RF attenuation RF ATT Auto
Preamplifier Preamp Off
Demodulation Demod FM
Detector Det. +/- peak/2
Measurement time Meas Time 100 ms
Trigger Trigger Free run

Measurement Examples

All of the following examples assume the standard settings for the measuring receiver. These are set with the PRESET key The described measurement applications are:
Measurement of the RF power of a signal using the Power Meter mode.
Measurement of low level signals using the Tuned RF Level mode.
Measurement of AM modulated signals and modulation distortion with the Modulation Analyzer mode.
Measurement of audio signals with the Audio Analyzer mode.
1166.3311.12 2.3 E-2
Measurement Examples R&S FSMR

Example 1: RF Power Measurements with a Power Sensor

Measurement
The FSMR includes a power meter functionality. Measurements of RF Power will be performed in the power meter mode with highest accuracy. The FSMR will automatically handle all required correction factors and compensate the measurements. The FSMR does support the following Rohde&Schwarz power sensors:
o R&S NRP-Z11 10 MHz to 8 GHz / 200 pW to 200 mW o R&S NRP-Z21 10 MHz to 18 GHz / 200 pW to 200 mW o R&S NRP-Z51 DC to 18 GHz / 1 uW to 100 mW o R&S NRP-Z55 DC to 40 GHz / 1 uW to 100 mW
Main Receiver Functions
The power meter functions are available in the Measuring Receiver POWER METER-mode. All required settings and evaluations can be reached from the POWER METER main menu.
Measurement Setup
Fig.2-2 RF Power measurement setup
The following settings are used on the generator:
1. PRESET the generator.
2. Set RF Frequency to 1 GHz.
3. Set the output level to 0 dBm.
Measurement Sequence – RF Power Measurements
Connect the Power sensor with the Signal Generator output and with the Measuring Receiver power sensor connector on the front panel.
The following steps are performed:
1. Zero the power sensor.
2. Connect the power sensor to the DUT.
3. Set the RF frequency.
4. Measure the RF Power
1166.3311.12 2.4 E-2
R&S FSMR Measurement Examples
1. Zero the Power Sensor
PRESET
PWR METER
FREQ
Reset the instrument.
Press the PRESET key.
The main receiver menu is opened. The receiver mode is set.
Switching to Power Meter Mode
Press the PWR METER hotkey.
The power meter menu is displayed.
Set the measuring frequency
Press the FREQ key on the front panel.
The measuring receiver frequency menu is displayed.
RECEIVER
FREQUENCY
PWR METER
ZERO
In the frequency menu press the RECEIVER FREQUENCY softkey.
The frequency entry is displayed. Enter the measuring frequency
with the number keys and the appropiate unit key. The frequency will be used by the instrument to calculate the calibration factors for the power sensor head.
In the RECEIVER FREQUENCY entry window key in “1000 MHZ”.
Zero the power sensor
Press the PWR METER hotkey.
The power meter menu is displayed.
In the POWER METER menu press the ZERO softkey.
In the message box, confirm CONTINUE and press ENTER.
The zeroing of the sensor will be performed.
After the zeroing is completed, a message will appear on the screen.
1166.3311.12 2.5 E-2
Measurement Examples R&S FSMR
Fig.2-3 RF power measurement after Zeroing
2. Measure the RF power in Watt
Connect the power sensor to the signal generator RF output
PWR METER
DISPLAY LOG LIN
Linear display – Change the Unit to Watt
Press the PWR METER hotkey.
The POWER METER menu is displayed.
In the POWER METER menu press the DISPLAY LOG / LIN
softkey.
The RF power unit is changed from dB to Watt.
Fig.2-4 RF power measurement in UNIT Watt
1166.3311.12 2.6 E-2
R&S FSMR Measurement Examples

Example 2: Low RF Level Measurements with Tuned RF Level Mode

Measurement
The main functionality of a measuring receiver is the RF level measurement. Measurements of low power RF signals will be performed in the Tuned RF Level mode with highest accuracy. The FSMR will automatically handle all required settings.
Main Receiver Functions
The measurement of RF levels over a wide input level range is performed in three different level ranges. To reach the highest possible accuracy, a calibration is performed with a power sensor. The FSMR does offer fully automated calibration procedures in the Tuned RF Level mode. The following example will guide through the calibration steps. All functions for Tuned RF level measurements can be reached from the RF LEVEL main menu.
Measurement Setup
Fig.2-5 Tuned RF Level measurement setup
The following settings are used on the generator:
1. PRESET the generator.
2. Set RF Frequency to 300 MHz.
3. Set the output level to 0 dBm.
Measurement Sequence – Tuned RF Level Measurements
Connect the Power sensor with the Signal Generator output and with the Measuring Receiver power sensor connector on the front panel.
The following steps are performed:
1. Calibrate the FSMR with the power meter.
2. Change the output level and perform further Range calibrations.
3. Average the results for enhanced stability.
4. Measure an unstable source (unstable frequency)
1166.3311.12 2.7 E-2
Measurement Examples R&S FSMR
1. Calibrate the setup- absolute power calibration
PRESET
RF LEVEL
FREQ
Reset the instrument.
Press the PRESET key.
The main receiver menu is opened. The receiver mode is set.
Switching to Tuned RF Level Mode
Press the RF LEVEL hotkey.
The Tuned RF level menu is displayed.
Set the measuring frequency
Press the FREQ key on the front panel.
The measuring receiver frequency menu is displayed.
RECEIVER
FREQUENCY
RF LEVEL
CAL ABS
POWER
In the frequency menu press the RECEIVER FREQUENCY softkey.
The frequency entry is displayed. Enter the measuring frequency
with the number keys and the appropiate unit key.
In the RECEIVER FREQUENCY entry window key in “300 MHZ”.
Connect the power sensor to the signal generator RF output
Calibrate the Tuned RF Level – Absolute Power calibration
Press the RF LEVEL hotkey.
The RF LEVEL menu is displayed.
In the RF LEVEL menu press the CAL ABS POWER softkey.
The calibration of the receiver will be performed. A message box with instructions about the setup appears. The first step is a power measurement with the power sensor connected to the source.
In the message box, confirm CONTINUE and press ENTER.
In the next step of the calibration the receiver is connected to the
source.
Connect FSMR RF input to the signal generator RF output
In the next message box, confirm CONTINUE and press ENTER.
1166.3311.12 2.8 E-2
R&S FSMR Measurement Examples
RECAL
The FSMR is now calibrated in one RF level range (the upper level range does handle RF levels from -10 dBm to +30 dBm). Measurements in this level range can now performed with full accuracy.
Fig.2-6 Tuned RF Level measurement after calibration
RF LEVEL
(MANUAL)
2. Range to range calibration- relative calibration
Change the signal generator output level
The following settings are used on the generator:
Change the signal generator RF Level in 1 dB steps down. Observe the FSMR screen for the yellow RECAL message to appear (at levels about -10 dBm). This is the point where the automated Range-to-Range calibration can be performed.
The FSMR is now ready for a RF level range change. The calibration of the level difference due to the range change will be performed.
Press the RF LEVEL hotkey.
The RF LEVEL menu is displayed.
In the RF LEVEL menu press the RECAL (MANUAL) softkey.
The RECAL procedure is performed. Do not change any signal
generator settings at this time.
A message box appears. The instrument will perform a
measurement in the calibrated range, then change the RF settings (RF attenuator, IF gain) and then measure again. After the recalibration measurements with high accuracy are available in the RF level range 2 (covers levels from -10 dBm to -50 dBm).
1166.3311.12 2.9 E-2
Measurement Examples R&S FSMR
RECAL
Change the signal generator output level further down
The following settings are used on the generator:
Change the signal generator RF Level in 10 dB steps down. Observe the FSMR screen for the yellow RECAL message to re­appear again (at levels about -50 dBm).
RF LEVEL
As soon as the RECAL flag appears, the FSMR is ready for a next RF level range change. The calibration of the level difference due to the range change will be performed.
Press the RF LEVEL hotkey.
The RF LEVEL menu is displayed.
(MANUAL)
In the RF LEVEL menu press the RECAL (MANUAL) softkey.
The RECAL procedure is performed.
A message box appears. The instrument will perform a
measurement in the calibrated range, then change the RF settings (RF attenuator, IF gain) and then measure again. After the recalibration measurements with high accuracy are available in the RF level range 3 (covers levels from -50 dBm to noise level).
The FSMR is now calibrated in all RF level ranges (for RF levels from -140 dBm to +30 dBm). Measurements in all level ranges can now be performed with full accuracy. No further calibration is required, the Autorange function will automatically adapt the level setting of the FSMR to the input signal.
Fig.2-7 Tuned RF Level measurement at low levels
1166.3311.12 2.10 E-2
R&S FSMR Measurement Examples
3. Average the results for enhanced stability
Change the signal generator output level further down
The following settings are used on the generator:
Change the signal generator RF Level to -110 dBm).
The FSMR is now performing a calibrated measurement at – 110 dBm. The reading will be flickering due to a low signal to noise ratio. For a more stable reading, the FSMR does offer several functions.
The stability of a low level measurement is mainly depending on the
RF LEVEL
measurement time. A longer acquisition of the signal does allow for a more accurate measurement. The averaging of several single acquisitions will lead to an effective longer measurement time.
Press the RF LEVEL hotkey.
The RF LEVEL menu is displayed.
AVERAGING
[10]
In the RF LEVEL menu press the MANUAL AVG softkey.
The AVERAGE will be switched on. The number of measurements
can be entered in the AVERAGING COUNT box which appears on top of the screen. The default value is set to 10 Averages.
The instrument will now perform an averaged measurement. The
average function is a floating average over the number of measurements chosen with the Average Count entry (default value:
10).
Fig.2-8 Tuned RF Level measurement with Averaging
1166.3311.12 2.11 E-2
Measurement Examples R&S FSMR
4. Measure an unstable source (unstable in frequency)
The simulation of unstable sources requires a signal which is unstable in the frequency domain. This signal can be simulated with active frequency modulation on the signal generator.
The following settings are used on the generator:
1. PRESET the generator.
2. Set RF Frequency to 100 MHz.
3. Set the output level to 0 dBm.
4. Select FM Modulation, 3 kHz deviation.
5. Set the Modulation frequency to 1 kHz.
PRESET
Reset the instrument.
Press the PRESET key.
The main receiver menu is opened. The receiver mode is set.
Switching to Tuned RF Level Mode
RF LEVEL
Press the RF LEVEL hotkey.
RECEIVER
FREQUENCY
DETECTOR
WIDE
The Tuned RF level menu is displayed.
Set the measuring frequency
In the RF LEVEL menu press the RECEIVER FREQUENCY
softkey.
The frequency entry is displayed. Enter the measuring frequency
with the number keys and the appropriate unit key.
In the RECEIVER FREQUENCY entry window key in “100 MHZ”.
The measurement of an unstable signal requires the measuring receiver to use a wide bandwidth. The FSMR offers a wideband detector mode to acquire wideband signals with high accuracy
In the RF LEVEL menu press the DETECTOR WIDE softkey.
The wideband detector mode acquire signals with the full acquisition
bandwidth. The acquisition bandwidth (DEMOD BW) of the FSMR can be set in the BW menu, the preset value is 12.5 kHz. The bandwidth must be set wide enough to capture the maximum frequency deviation of the input signal. In this example, the frequency deviation is set to 3 kHz and the default bandwidth is 12.5 kHz.
Note: The wideband detection is not only suitable for unstable signals,
it can also be used to accurately measure modulated signals
Next steps:
The following steps will be performed as described in the previous example (Tuned RF Level).
1166.3311.12 2.12 E-2
R&S FSMR Measurement Examples

Example 3: AM Modulation and Modulation Distortion Measurements

Measurement
The FSMR includes a modulation analyzer. Measurements like AM, FM or PM modulation will be performed in the modulation analyzer mode. The instrument offers standard measurements like modulation depth, deviation and modulation frequency, but also more sophisticated function like modulation distortion or audio frequency response are available.
Main Receiver Functions
The modulation analyzer functions are available in the Measuring Receiver DEMOD-mode. All required settings and evaluations can be reached from the DEMOD main menu.
Measurement Setup
Fig.2-9 AM modulation measurement setup
The following settings are used on the generator:
1. PRESET the generator.
2. Set RF Frequency to 100 MHz.
3. Set the output level to - 30 dBm.
4. Select the AM Modulation, 30% modulation depth.
5. Set the Modulation frequency to 1 kHz.
Measurement Sequence – AM Modulation Measurements
Connect the Signal Generator output with the Measuring Receiver RF Input.
The following measurement steps are performed:
1. Measure the AM modulation depht.
2. Relative Audio measurements
3. Measure the Total harmonic distortion in %.
1166.3311.12 2.13 E-2
Measurement Examples R&S FSMR
PRESET
DEMOD
AM
FILTER
1. Measure the AM modulation depth
Reset the instrument.
Press the PRESET key.
The main receiver menu is opened. The receiver mode is set.
Switching to AM demodulation
Press the DEMOD hotkey.
The analog demodulation menu is displayed.
In the DEMOD menu press the AM softkey.
The modulation frequency and the AM modulation depth with
peak detection measurement results are displayed in the lower window of the screen. The upper window shows the RF Frequency and the RF input level.
Limiting the bandwidth - Audio Filters
Press the DEMOD hotkey.
The analog demodulation menu is displayed.
In the DEMOD menu press the FILTER softkey.
LOW PASS
15 KHZ
The audio filter menu is displayed. To suppress unwanted
broadband noise or harmonics of the demodulated signal, the bandwidth of the measurement can be limited with highpass and lowpass filters.
In the FILTER menu press the LOW PASS 15 KHZ softkey.
The following screen is displayed:
Fig.2-10 AM Modulation measurement in the DEMOD mode
In the FILTER menu press the LOW PASS 15 KHZ softkey again
to switch the filter off.
1166.3311.12 2.14 E-2
R&S FSMR Measurement Examples
DEMOD
2. Relative Audio measurements – audio frequency response
Press the DEMOD hotkey.
The analog demodulation menu is displayed.
In the DEMOD menu press the RELATIVE softkey.
RELATIVE
The RELATIVE menu is displayed.In the relative menu, only the
detectors which have been switched on in the DETECTOR menu are available for relative measurements.
In the RELATIVE menu press the +- PEAK/2 softkey.
+- PEAK/2
The actual measurement value of the detector will be saved as a
reference value and the result display will change to a relative reading (indicated with K 0.00 %).
Change the modulation frequency to 50 kHz
The following settings are used on the generator:
1. Select the AM Modulation, 50 kHz modulation frequency.
Relative display – Change the Unit to dB
RELATIVE
Press the DEMOD hotkey.
RELATIVE
DB %
The analog demodulation menu is displayed.
In the DEMOD menu press the RELATIVE softkey.
The RELATIVE menu is displayed.
In the RELATIVE menu press the +- PEAK/2 softkey.
The actual result display will change from a relative reading
indicated in % to a dB reading.
Fig.2-11 Relative measurement in the demod mode
1166.3311.12 2.15 E-2
Measurement Examples R&S FSMR
DEMOD
3. Measure the harmonic distortion - Choosing the detector
Press the DEMOD hotkey.
In the DEMOD menu press the DETECTOR SELECTION softkey.
The DETECTOR SELECTION menu is displayed
DETECTOR
SELECTION
The instrument is equipped with a wide range of detectors for
measuring modulation signals. The peak detectors are a good choice for capturing the highest positive or negative peak within the measurement time, while the RMS and Average detectors are the best choice for measuring noise and residual modulation. The THD & SINAD detector measure the distortion of the demodulated audio
DISTORTION
& SINAD
signal. All detectors can be used in parallel.
In the DETECTOR SELECTION menu press the DISTORTION &
SINAD softkey.
AMPT
Total Harmonic Distortion – Change the Unit to %
Press the AMPT hardkey on the frontpanel.
The amplitude settings menu is displayed.
UNIT
In the AMPLITUDE menu press the UNIT softkey.
The measuring receiver UNIT menu is displayed.
DIST UNIT
% DB
In the UNIT menu press the DIST UNIT % / DB softkey.
The actual result display for the THD measurement will change from
a “DB” reading to a “%” reading.
Fig.2-12 THD measurement in UNIT %
1166.3311.12 2.16 E-2
R&S FSMR Measurement Examples

Example 4: Audio Measurements with the Audio Analyzer mode

Measurement
The FSMR includes an audio analyzer functionality. Measurements of audio signals will be performed in the audio analyzer mode with highest accuracy. The FSMR will automatically handle all required settings.
Main Receiver Functions
The audio analyzer functions are available in the Measuring Receiver AUDIO-mode. All required settings and evaluations can be reached from the AUDIO main menu.
Measurement Setup
Fig.2-13 Audio measurement setup
Connect the signal generator LF (low frequency) output with the FSMR Audio input.
The following settings are used on the generator:
1. PRESET the generator.
2. Set LF output to ON .
3. Set the LF output level to 1 volt.
4. Set the LF output frequency to 1 kHz.
Measurement Sequence – Audio Measurements
The following steps are performed:
1. Measure the audio level and frequency.
2. Measure the frequency response of the audio filters.
3. Measure the influence of the input impedance.
1166.3311.12 2.17 E-1
Measurement Examples R&S FSMR
PRESET
AUDIO
Reset the instrument.
Press the PRESET key.
The main receiver menu is opened. The receiver mode is set.
Switching to Audio Mode
Press the AUDIO hotkey.
The audio menu is displayed. The audio measurement results are displayed in parallel, no further settings like frequency is required.
Fig.2-14 Audio measurement after Preset, Mode Audio
AUDIO
4. Relative Audio measurements
Press the AUDIO hotkey.
The audio menu is displayed.
In the AUDIO menu press the RESULT ABS REL softkey.
RESULT
ABS REL
The relative measurement function is enabled.
The actual measurement value of the detector will be saved as a
reference value and the result display will change to a relative reading (indicated with K 0.00 dB). The unit can be changed with the REL UNIT DB % softkey in the audio menu.
1166.3311.12 2.18 E-1
R&S FSMR Measurement Examples
AUDIO
Limiting the bandwidth - Audio Filters
Press the AUDIO hotkey.
The audio menu is displayed.
In the AUDIO menu press the FILTER softkey.
FILTER
The audio filter menu is displayed. To suppress unwanted
broadband noise or harmonics of the signal, the bandwidth of the measurement can be limited with higpass and lowpass filters.
In the FILTER menu press the LOW PASS 15 KHZ softkey.
LOW PASS
15 KHZ
Change the LF output frequency to 15 kHz
The following settings are used on the generator:
Set the LF output frequency to 15 kHz.
The audio frequency is now changed to the bandwidth of the active 15
kHz low pass filter. In the following measurement the frequency response of the audio low pass filter is measured.
Fig.2-15 Audio measurement in relative mode
1166.3311.12 2.19 E-1
R&S FSMR Contents - Manual Operation
Contents - Chapter 3 "Manual Operation"
3 Manual Operation ................................................................................................ 3.1
The Screen ........................................................................................................................................3.1
Diagram Area ...........................................................................................................................3.2
Indications in the Diagram Area .....................................................................................3.3
Full Screen .....................................................................................................................3.9
Split Screen....................................................................................................................3.9
Softkey Area...........................................................................................................................3.10
Hotkey Area............................................................................................................................3.11
Calling and Changing the Menus..................................................................................................3.11
Setting Parameters.........................................................................................................................3.12
Numeric Keypad.....................................................................................................................3.12
Spinwheel and Cursor Keys ...................................................................................................3.13
Selection and Setting of Parameters via Keys or Softkeys ....................................................3.14
Editing of Numeric Parameters ..............................................................................................3.17
Entry of Alphanumeric Parameter ..........................................................................................3.20
Editing with External Keyboard ....................................................................................3.20
Editing with Help Line Editor ........................................................................................3.21
Selection and Setting of Parameters via Tables ....................................................................3.24
Menu Overview Receiver Mode .......................................................................................................27
FREQENCY Key .......................................................................................................................27
SPAN Key..................................................................................................................................28
AMPT Key .................................................................................................................................29
BW Key .....................................................................................................................................30
SWEEP/MEAS Keys .................................................................................................................31
Hotkey Menus ...........................................................................................................................32
RF LEVEL Key ................................................................................................................32
PWR METER Key...........................................................................................................33
DEMOD Key....................................................................................................................34
AUDIO Key......................................................................................................................35
Menu Overview Spectrum Analysis Mode ......................................................................................36
FREQUENCY Key.....................................................................................................................36
SPAN Key..................................................................................................................................37
AMPT Key .................................................................................................................................38
MEAS Key .................................................................................................................................39
BW Key .....................................................................................................................................40
SWEEP Key ..............................................................................................................................41
Menu Overview another Keys ..........................................................................................................42
MKR Key ...................................................................................................................................42
MKR-> Key................................................................................................................................43
MKR FCTN Key.........................................................................................................................44
TRIG Key...................................................................................................................................45
TRACE Key ...............................................................................................................................46
LINES Key.................................................................................................................................47
1166.3363.12 I-3.1 E-1
Contents - Manual Operation R&S FSMR
DISP Key...................................................................................................................................48
FILE Key....................................................................................................................................49
CAL Key ....................................................................................................................................50
SETUP Key ...............................................................................................................................51
HCOPY Key...............................................................................................................................52
Hotkey Menu .............................................................................................................................53
LOCAL Menu.............................................................................................................................53
Menu Overview Ext. Generator Control ..........................................................................................54
Menu Overview Option Network Mode............................................................................................55
1166.3363.12 I-3.2 E-1
R&S FSMR The Screen
3 Manual Operation
Chapter 3 provides an overview of the operating concept and the basic steps of manual operation of the R&S FSMR. This includes a des cription of the s creen, of the control of menus and of the setting of parameters. An overview of the menus will be listed at the end of this chapter.
The functions of the menus are described in detail in Chapter 4. Chapter 2 contains a short introduction on step-by-step simple measurements. The remote control of the instrument is described in Chapters 5, 6 and 7
The operation of the spectrum analyzer is menu-controlled via keys, hotkeys and softkeys. The setting of the ins trument and test parameters in the m enus is made either directly via s oftkeys or by entry of values in entry fields and by selection in tables . The operating mode and the s creen mode is selected via the hotkeys. If required, data entry windows and tables are superimposed on the screen.

The Screen

The screen informs continuously on the r esults and parameters of the selected measuring functions. It shows the as signment of the s oftkeys and m enus, which are required for setting the measuring parameters. The display of test results, the s oftkey labeling and the type of menu depend on the selected measuring function. The screen is subdivided into three areas:
Diagram area hotkey area softkey area
Fig. 3-1 Subdivision of screen
1166.3363.12 3.1 E-1
The Screen R&S FSMR

Diagram area

Softkey area
Hotkey area
Diagram Area
logo titel /date and time hardware settings marker/deltamarker
reference/ max level
Status information
This area contains the m easuring diagrams and other measured-value information as well as the par ameters and status information which are important for analysis of the results. In addition, message fields, entry windows and tables may be shown in this area.
This area contains the instrument functions which can be s elected via the softkeys. The softkey area is not superimposed by other graphics.
This area contains the available oper ating modes and s creen modes. The hotkey area is not superimposed by other graphics.
Enhancement label
Trace-info
Measureme nt window
level-axis labelling
x-axis labelling
Fig. 3-2 Subdivision of the R&S FSMR screen in analyzer mode (without measuring diagram)
1166.3363.12 3.2 E-1
R&S FSMR The Screen
Marker
Limit line Limit line
Deltamarker
Fig. 3-3 Measuring diagram
Indications in the Diagram Area
The following graphic elements are displayed in the diagram area:
General indications
Indication of the logo
Logo
Screen title
Indication of selected screen title
Date / time
Indication of date and time
Hardware settings
Ref
Offset
Att
RBW
VBW
SWT
Indication of the reference level
Indication of the offset of reference level.
Indication of the set RF attenuation.
Indication of the set resolution bandwidth. If the bandwidth does not correspond to the value of the automatic coupling, a green asterisk "*" is prefixed to the field.
Indication of the set video bandwidth. If the bandwidth does not correspond to the value of the automatic coupling, a green asterisk "*" is prefixed to the field.
Indication of the set sweep time. If the sweep time does not correspond to the value of the automatic coupling an asterisk "*" is prefixed to the field. The colour of the asterisk turns red as soon as the sweep time falls below the value of the automatic coupling.
1166.3363.12 3.3 E-1
The Screen R&S FSMR
Marker/deltamarker This label displays the position of the last selected marker or
deltamarker in the x and y-directions and the marker/deltamarker index. The square brackets contain the curve which the marker is assigned to and the active measuring function of the marker indicated. The measuring function of the markers in the second field is indicated by the following abbreviations:
FXD reference fixed active PHN phase noise measurement active CNT frequency counter active TRK signal track active NOI noise measurement active MOD measurement of AM modulation depth active TOI TOI measurement active (3
Limit check Indication of the result of the limit check.
X-axis labelling
10 MHz/DIV
Center 1.2345678901234 GHz
Start 1.2345678901234 GHz
Span 1.2345678901234 GHz
Stop 1.2345678901234 GHz
Trigger 1.234 ms
Display of the x-scaling.
The distance between two grid lines is displayed in this label.
The set center frequency or start frequency is displayed in this label depending on w hether the k eys FREQ/SPAN or the softkeys START/STOP were last pressed. If span = 0 Hz, the center frequency is always displayed.
The set frequency range (SPAN) or the s top frequency is displayed, depending on whether the k eys FREQ/SPAN or the s oftkeys START/STOP were last pressed. If span = 0 Hz, the trigger moment (PRETRIGGER) is displayed.
rd
order intercept)
1166.3363.12 3.4 E-1
R&S FSMR The Screen
Status information
#SMPL
UNCAL
OVLD / IFOVL
LOUNL / EXREF
OVEN
Trace info:
The status information on the lef t side of the diagram hint at ir regularity (e.g. UNCAL)
"#SMPL" indicates that the relation Span / RBW is higher than 125 while the RMS detector is activated. In this case, a stable signal evaluation is no longer possible due to an ins ufficient number of A/D c onverter samples.
reduce span or increase RBW
"UNCAL" is indicated under the following circumstances:
correction data are switched off (menu CAL, CAL CORR OFF). switch on CAL CORR ON or PRESET
no valid correction data. T his may occur after a c old start of the
instrument following a firmware update.
record correction data
Sweep time too s hort for current instrument settings (span,
resolution bandwidth, video bandwidth). increase sweep time
OVLD is indicated when the input mixer is overloaded.
Increase input attenuation
IFOVL is indicated when overload occurs in the IF signal path after the input mixer.
Increase reference level
LOUNL is indicated when an error occurs in the frequency processing of the instrument.
EXREF is indicated when the analy zer is configured for use of an external reference signal, but no r eference signal is detected at the corresponding input.
OVEN is indicated when the c rystal oscillator (option R&S FSMR-B4) has not yet reached its operating temperature. This indication vanishes after a few minutes after switch on.
Every active measurement curve (trace BLANK) is allocated trace information of two or three lines at the left of the diagram. The trace information has the same colour as the measurement curve. The information on the currently selected trace is displayed in inverse video (see TRACE - SELECT TRACE softkey).
1166.3363.12 3.5 E-1
The Screen R&S FSMR
<n> <detector> <*> <mode> <trace math>
Example:
1 PK * CLRWR 1-2
n = trace number (1 ... 3)
detector = selected detector AP: AUTOPEAK detector PK: MAX PEAK detector MI: MIN PEAK detector SA: SAMPLE detector AV: AVERAGE detector RM: RMS detector QP: QUASIPEAK detector
* = indicates that the selected detector does not correspond
to that of the automatic coupling.
mode = indication of sweep mode
CLRWR: CLEAR/WRITE MAXH: MAX HOLD MINH: MIN HOLD AVG: AVERAGE VIEW: VIEW
Trace math = trace math active 1 - 2 trace 1 - trace 2 1 - 3: trace 1 - trace 3
1166.3363.12 3.6 E-1
R&S FSMR The Screen
Instrument settings (Enhancement Labels)
*
Indication of user instrument settings which influence the measuring result and w hich are not im mediately obvious when viewing the measured values.
The current instrument setting does not c orrespond to the one which
applied when one of the dis played curves had been s tored. This occurs under the following circumstances:
The instrument setting is modified while a m easurement is being
performed.
The instrument setting is modified in SINGLE SWEEP mode
subsequent to the end of the sweep and no new sweep is started.
The instrument setting is modified after setting the trace to VIEW.
A / B
SGL
GAT
TRG
LVL
FRQ
PRN
75 The input impedance of the instrument is set to 75 Ω.
EXT
PA
Entry fields:
Entry window
Tables
The display is retained until the c ause is eliminated by the user. I.e., either a new sweep is started (SINGLE SWEEP mode), or the tr ace of interest is switched to BLANK.
Identification for screen A / B. W hen screen A / B is activated for the entry of test parameters, this label is highlighted.
The sweep is set to SINGLE SWEEP.
The frequency sweep is controlled via the EXT TRIG/GATE input of
the instrument.
The instrument is not triggered in FREE RUN mode.
A level offset 0 dB has been set.
A frequency offset 0 Hz has been set.
A printer output is active.
The instrument is configured for operation with external reference
The RF preamplification is switched on (option B23 or B25).
The data entry window is superimposed in the left upper corner of the diagram area, if required. It covers the display of the title and the time. The field is used to enter numeric or alphanumeric device parameters.
The tables are superimposed in the diagram area, if required. They are used for displaying and configuring device parameters.
1166.3363.12 3.7 E-1
The Screen R&S FSMR
Message windows:
Message field
System messages
Traces:
Message fields provide notes on measurements, e.g. results of the limit check (PASS/ FAIL). These notes are no error messages, which are indicated as system messages.
They can be masked out by pressing the ESC key.
System messages indicate warnings and error messages.
Message without action field:
These system messages contain only arbitrary information. They hint at events which are of interest for the us er but do not affect the measurement or functioning of the instrument.
They are deleted either automatically after a pr edefined time has passed (3 seconds) or upon any keystroke or mouse click.
Message with action field:
These system messages require a dec ision to be tak en by the us er. They are not deleted until any action has been s elected. Deletion of the message initiates the action selected and appropriate measures to be taken. The action field consists of one (OK), two (OK / CANCEL) or three (arbitrary) buttons. The user may select one of the buttons using the cursor keys and initiate the associated action by means of the unit keys. The ESC key is used to acknowledge the message without releasing any action.
1 to 3
Limit lines
The R&S FSMR provides two display modes:
Full Screen: 1 window, the measurement is performed in the active diagram.
Split Screen: 2 windows, the measurements are performed in both diagrams.
Up to 3 traces in each measurement diagram can be displayed simultaneously.
Limit lines are used to mark level curves or spectral distributions which must not be exceeded or dropped below.
1166.3363.12 3.8 E-1
R&S FSMR The Screen
Full Screen
In the full-screen mode, the settings and measurements are performed in the ac tive visible window. All indications on the s creen refer to this window. The designation (SCREEN A or SCREEN B) is inserted as enhancement label A or B on the right diagram margin. Switching between the w indows is by means of SCREEN A/B hotkey. The current measurement is terminated when its window is blanked out. Switching from split-screen to full-screen mode is performed in menu DISP.
Split Screen
In Split Screen mode, the screen is divided into two halves.
Fig.3-4 Subdivision of the screen in Split Screen mode
The upper half is assigned Screen A, the low er one Sc reen B. T he settings for measurement can be selected independently for both s creens. E.g., a s pectrum may be displayed in Screen A and a time amplitude in the time range is displayed in Screen B.
The indications which are valid only for one window appear in the m argin of the as sociated diagram. Indications which are valid for the two windows are displayed between the diagrams.
The window for entry of the m easuring parameters or the m arker operation is selected using the SCREEN A/B hotkey. The measurements are simultaneously performed in the two windows irrespective of the currently active one. Switching from full-screen to split-screen mode is performed in menu DISP.
1166.3363.12 3.9 E-1
The Screen R&S FSMR

Softkey Area

The softkeys are assigned to the nine keys on the right side of the display The setup of the softkey area is independent of the operating mode. It consists of the following graphic elements:
Softkey menu:
Fig.3-5 Setup of the softkey area
The softkeys have different functions depending on the instrument state. Therefore, their labeling can be varied. The labeling of all softkeys which call a submenu includes a arrow. The function and current state of the softkeys is indicated in the label by different texts and colors. The color assignment is factory-set as follows:
SOFTKEY LABEL 1
SOFTKEY LABEL 2
SOFTKEY LABEL 3
SOFTKEY LABEL 4
SOFTKEY LABEL 5
SOFTKEY LABEL 6
SOFTKEY LABEL 7
SOFTKEY LABEL 8
SOFTKEY LABEL 9
indication of submenu (arrow)
softkey active (green)
softkey function not available (without 3D frame)
softkey active, opens a data entry field (red)
indication of side menu
Table 3-1 Factory-set color assignment of soft keys
Softkey color Meaning
gray Softkey switched off
green Softkey switched on
red Softkey switched on and data entry active
These colors can be changed by the user as desired in the DISP - CONFIG DISPLAY menu.
A softkey is switched on or off by pressing the r espective hardkey (see following section "Setting the Parameters").
Softkeys are masked out, too, if the functionality which they represent is not available. A dis tinction has to be made between two cases:
If an instrument function depends on an option, and if this option is not fitted, the associated softkey is
masked out,.
If the instrument function is not available temporarily due to specific settings, the softkey is displayed
without the 3D frame.
The label
on the r ight lower corner of the softkey area indicates that a s ide menu is available. T he
side menu is called by pressing the NEXT key.
1166.3363.12 3.10 E-1
R&S FSMR Calling and Changing the Menus

Hotkey Area

Hotkeys are allocated to the eight k eys on the bottom margin of the screen. They change between modes and the active diagrams. The menu only shows the hotkeys which are actually used:
SPECTRUM RF LEVEL
PWR METER
DEMOD
AUDIO
A keystroke activates the associated hotkey. An activated hotkey changes its frame:
MRECEIVER

Calling and Changing the Menus

The operation of the s pectrum analyzer is menu-controlled via k eys and s oftkeys. Various softkey menus are displayed depending on the instrument status. The individual menus constitute the so-called menu tree. The top menu (the root of the menu tree) is always called by means of a keystroke. Arrows at the lower edge of the softkey area indicate whether a supplementary menu can be entered or not.
Softkeys with an arrow allow for branching into further menus (so-called submenus): The field " lower right side of the softkey area indicates that this menu has a side menu.
The menu change keys on the front panel below the softkey area allow for switching between the main menu and the side menus and submenus.
NEXT
The NEXT key calls the side menu.
" at the
PREV
The PREVIOUS key returns to the next higher menu.
Several menus provide for automatic change, i.e., return to the next higher menu is caused automatically after pressing a softkey.
Selection of a submenu is always effected via a softkey.
SOFTKEY
The labeling of all softkeys which call a submenu includes a arrow.
1166.3363.12 3.11 E-1
Setting Parameters R&S FSMR

Setting Parameters

Parameters are set either by simple selection (selection parameters) or by (alpha)numeric entries in data entry windows or tables.
The numeric keypad on the f ront panel, an ex ternal keyboard (optional), a s pinwheel and the c ursor keys are provided for the entry of instrument parameters in an entry window or in a table.
The external keyboard is optional. If it is not fitted, the help line editor is called automatically for entry of alphanumeric parameters. The help line editor provides for selection of individual letters and a number of special characters which are copied into the actual entry window.

Numeric Keypad

The numeric keypad is provided for entry of numeric
s
7
4
89
56
12
0
.
GHz
-dBm
MHz
dBm
3
-
kHz
dB
Hz
dB..
parameters. It contains the following keys:
V
Number keys 0 to 9
ms
Decimal point
mV
mV
µs
µV
Inserts a decimal point "." at the cursor position.
Sign key
Changes the sign of the mantissa or exponent of a numeric parameter. A "-" is inserted at the cursor position in case of an alphanumeric parameter.
nV
ns
Unit keys (GHz/-dBm, MHz/dBm, kHz/dB and Hz/dB)
– Provide the numeric value entered with the selected unit
and terminate the entry. The unit k eys are all as signed the value "1" for
BACK
ESC
CANCEL
ENTER
dimensionless quantities or for level entries (e.g., in dB). The unit k eys thus assume the f unction of an ENTER key. The same applies for an alphanumeric entry.
– Open and close the selection windows of tables.
BACK key
– Deletes the character left to the c ursor with numeric
entry.
– Allows for toggling between the current and the previous
values subsequent or prior to entry (UNDO function).
ESC/CANCEL key
– Aborts the entr y before it has been ter minated. The
previous value is restored.
Closes the entry field after termination of input. Closes system messages.
ENTER key
– Terminates the input of dimensionless quantities. The
new value is set.
Note: The ENTER keys assumes the func tion of the
Hz key for frequency input, and the func tion of
µ
s(kHz) key for time input.
the
1166.3363.12 3.12 E-1
R&S FSMR Setting Parameters

Spinwheel and Cursor Keys

The spinwheel and the cursor keys are arranged besides the numeric keypad.
The spinwheel has various functions:
With numeric entry, the ins trument parameter is incremented (turning
clockwise) or decremented (turning counterclockwise) at a defined step size.
In tables, the s pinwheel can be us ed to s hift the cursor horizontally or
vertically when no entry window is open. The direction (horizontal/ vertical) is switched over using the cursor keys.
The spinwheel is used with the help- line editor to s elect the individual
letters.
It can be used to shift markers, display lines, limit lines etc.
Pressing the spinwheel terminates the input of parameters.
In tables, the c ursor keys are used to shift the cursor between the lines and columns of the table.
The keys reach a particular position in the string.
The keys
– increase or decrease the value of a parameter for numeric input . – switch between editing line and help line editor for alphanumeric input.
and
and
are used to s hift the c ursor inside the entr y window to
1166.3363.12 3.13 E-1
Setting Parameters R&S FSMR

Selection and Setting of Parameters via Keys or Softkeys

The selection of parameters and their settings is effected by means of a k ey, a s oftkey or in a table depending on the hierarchical level of the menu they are assigned to. Selec tion and s etting of parameters in tables is described in section "Selection and Setting of Parameters in Tables" .
Selection via key
Most keys of the netw ork analyzer are used to enter menus where the s election and the settings are made. Few settings can be m ade directly by means of a keystroke, only.
Example: Call up of preset settings
¾ Press PRESET key
The spectrum analyzer is brought into a pr edefined
initial state.
PRESET
Selection via softkey
1. The softkey is active or inactive.
2. The softkey acts like a toggle switch, each pressing changes the active selection.
In most cases the s election is made by pressing the respective softkey. There are various alternatives of making the selection:
Example: Switching on/off the frequency counter
¾ Press MKR key.
¾ Press SIGNAL COUNT softkey.
Each time the softkey is pressed, the marker info list
is switched on or off. If the softkey is active (= marker info list on), it is illuminated.
SIGNAL
COUNT
SIGNAL
COUNT
Example: Selection of the reference (internal or external)
¾ Press SETUP key.
¾ Press GENERAL SETUP softkey, the GENERAL
SETUP submenu is opened.
¾ Press REFERENCE INT/EXT softkey.
With each pressing, the c heckmark on the s oftkey
changes from INT (internal reference) to EX T (external reference) and vic e versa. When in the active setting the softkey menu item is illuminated.
REFERENCE INT EXT
REFERENCE INT EXT
1166.3363.12 3.14 E-1
R&S FSMR Setting Parameters
3. Various softkeys act like selection switches. Only one softkey may be active at a time.
Example: sweep setting
¾ Press SWEEP key.
¾ Press CONTINUOUS SWEEP softkey.
The continuous sweep is thus set. The CONTINOUS SWEEP softkey is colored (factory-set: green). The
second alternative, a s eries of n s weeps according to the trigger definition, can be s elected via the SINGLE SWEEP softkey in the s ame menu. Only one of the two softkeys can be ac tive at a time, the softkeys thus act like selection switches.
CONTINUOUS
SWEEP
SINGLE
SWEEP
4. The softkey is used to select the parameter, the setting is made in an alphanumeric data entry window.
Example: SWEEP COUNT parameter
¾ Press SWEEP key
¾ Press SWEEP COUNT softkey.
The window for entering the num ber of sweeps for the SINGLE SWEEP mode is opened. T he softkey is colored (factory-set: red). (Data entry is described in the next section).
SWEEP COUNT
CONTINUOUS
SWEEP
SINGLE
SWEEP
SWEEP COUNT
SPAN
100
1166.3363.12 3.15 E-1
Setting Parameters R&S FSMR
4. The softkey is used to select the parameter, the setting is made in an (alpha)numeric data entry window. The softkey function is
Example: parameter MARKER
¾ Press MRK key.
¾ Press MARKER 1 softkey.
switched on. To switch off the function, the softkey has to be pressed again.
The window for entering the m arker frequency is opened. The softkey is colored (factory set: red). Marker 1 is switched on and the peak search is started.
MARKER 1
¾ Press MARKER 2 softkey.
The entry window for the m arker frequency of marker 2 is opened. T he softkey is colored (red), marker 2 is switched on, and the MARKER 1 softkey turns green.
MARKER 1
MARKER 2
¾ Press MARKER 1 softkey again.
MARKER 1
SPAN
42.03456788 MHz
DELTA MARKER 2
SPAN
1.746729018 GHz
5. The softkey selects the parameter, the setting is made in a selection table.
Marker 1 is switched off.
Example: Selection of the sorting criteria of a file list
¾ Press FILE key.
¾ Press FILE MANAGER softkey.
¾ Press SORT MODE softkey.
A selection table is displayed. The softkey is colored (factory-set: red). (operation see below).
SORT MODE
SORT MO
by NAME by DA TE/ TIM E by EX TEN SIO N
DE
1166.3363.12 3.16 E-1
R&S FSMR Setting Parameters
A
p
g

Editing of Numeric Parameters

The entry of numeric values is always made in a data entr y window, which is displayed automatically after selection of the parameter.
Head line with
START FREQUENCY
10.2457535 GHz
RT FREQUENCY OUT OF RANGE
ST
The headline indicates the nam e of the instrument parameter, which has been s elected. The entry is made in the editing line. Subs equent to calling the entry window, the current parameter value including the unit is displayed left-justified in the editing line. Status and error messages which refer to the current entry are displayed in the third and (optionally) fourth line.
Note: Entry windows may be represented transparent (cf. DISPLAY - CONFIG DISPLAY menu)
Entry of a numeric value
¾ Call data entry window (cf. selection of parameters) The editing line indicates the current value
Example: center frequency (frequency-sweep mode)
FREQ
Entry via number keys
¾ Enter required value via number keys.
Example:
2
arameter name
Editing line with parameter value and unit
Status and messa
CENTER FREQUENCY
1.5 GHz
e line
Entry via cursor keys
0
0
MHz dBm
mV
ms
CENTER FREQUENCY
200 MHz
¾ Cursor or until obtaining the required value.
Example:
CENTER FREQUENCY
200 MHz
1166.3363.12 3.17 E-1
Setting Parameters R&S FSMR
Entry via spinwheel
¾ Rotate the spinwheel until reaching the required value.
The variation step size increases with increasing
rotational speed.
Turning the s pinwheel clockwise increases the value,
turning it counterclockwise decreases the value.
Example:
CENTER FREQUENCY
1.75 GHz
Terminating the entry
Note: When the value is modified by means of the
spinwheel or the c ursor keys the new value is immediately set.
¾ Press one of the unit keys The unit is entered in the entry window and the new
setting is accepted by the instrument. For dimensionless quantities: ¾ Press the ENTER key or the spinwheel The new setting is accepted by the instrument.
A few data entry windows close automatically whereas
others like the entr y window for the s tart frequency
remain open even af ter termination of the entr y. They
can be closed by pressing the ESC key.
If an error occurs, a c orresponding error message is
displayed in the s tatus line of the entry window, e.g., "Out of range", "Value adjusted", etc. In this case, the new value is not accepted for the instrument setting.
Example:
0
Hz dB.
nV
ns
CENTER FREQUENCY
0 Hz out of range
Correcting the entry
Deleting an entry
¾ Position the c ursor beside the digit w hich is to be
.
deleted using the cursor keys
or
¾ Press the BACK key. The entry left to the c ursor is
deleted.
¾ Enter new numbers. The number is inserted to the left
of the cursor, the other numbers are shifted right.
1166.3363.12 3.18 E-1
R&S FSMR Setting Parameters
Restoring the original value
¾ Press the BACK key For numerous parameters, the data adm inistration of
the instrument stores the previously valid parameter
value in addition to the c urrent value. T he BACK key
can be us ed to toggle betw een these two values. This
applies for terminated entries as long as the data entry
window is displayed.
Aborting the entry ¾ Press ESC key
The original parameter value is restored. The new
entry is deleted.
¾ Press ESC key again The entry window is closed, the or iginal value r emains
active.
or
¾ Press any key or any softkey (even the s oftkey which
has opened the entry window). The entry is aborted and the entr y window is closed.
The original value remains active.
1166.3363.12 3.19 E-1
Setting Parameters R&S FSMR
A
NDPA
A

Entry of Alphanumeric Parameter

A help-line editor or an ex ternal keyboard (optional) are provided for the entr y of alphanumeric instrument parameters. The spinwheel and the exponent key have no function with alphanumeric entry. All unit keys assume the function of an ENTER key.
The entry is always made in a data entr y window which is displayed automatically upon selection of the parameter. The editing line c omprises 60 c haracters. Up to 256 c haracters may be enter ed. If a text exceeds 60 c haracters the contents is shifted automatically 20 c haracters left or right when the lef t or right margin of the editing line is touched by the cursor.
Editing with External Keyboard
COMMENT CHANNEL
B
Entry of text ¾ Select parameter.
SS-FILTERTEST 23
The data entry is active automatically upon calling the
data entry window. The cursor is positioned at the
beginning of the previous entry.
input lin
message line
¾ Press the required character on the keyboard.
The character is inserted prior to the cursor.
¾ Enter further characters
Correcting the entry ¾ Delete the entry using the DELETE key or
BACKSPACE key.
Terminating the entry
Aborting the entry ¾ Press the ESC key on the front panel
¾ Press the ENTER key of the external keyboard
The data entry window is closed and the new value is accepted for the instrument.
or
¾ any softkey.
The data entry window is closed, the original value is
retained.
1166.3363.12 3.20 E-1
R&S FSMR Setting Parameters
Editing with Help Line Editor
If the external keyboard is not fitted, the help line editor is called automatically with entry of alphanumeric parameters. The help line editor is an extension of the alphanum eric entry window. It c ontains the complete alphabet with uppercase and lowercase letters as well as special characters in two lines of 52 characters, each. Individual letters and a series of special characters can be s elected in the help line editor and copied into the entry line.
The help line editor comes in two versions:
Version 1:
CALIBRATION FILE
C:\MEASDATA\STANDARD.CAL
BCDEFGHIJKLMNOPQRSTUVWXYZÄÖÜ!?"$%/(){[]}\+*#~'-_=.:
A
abcdefghijklmnopqrstuvwxyzäöüß,;<>|@µ²³^°1234567890
The cursor keys
or toggle between entry in the editing line and selection in the help line editor.
Entry of text
¾ Select parameter.
The data entry is active automatically upon calling the data entry window. The cursor is positioned on the left in the editing line.
input line
message line
help line editor
Correcting the entry
Terminating the entry
Aborting the entry
¾ Position the cursor in the help line editor using the cursor key.
The cursor marks the first letter of the editor.
¾ Locate the cursor on the character which is to be enter ed into the
or the spinwheel.
editing line using the cursor keys
or
¾ Press the ENTER key or the spinwheel
The character is entered in the editing line.
¾ Locate the cursor in the editing line using the cursor key
.
¾ Position the cursor after the character which is to be deleted using
the cursor keys
and .
¾ Press the BACK key.
The entry left to the cursor is deleted.
¾ Locate the cursor in the editing line using the cursor key
.
¾ Press any unit key or the spinwheel.
The data entry window is closed, the new entry is accepted by the instrument.
¾ Press the ESC key.
The data entry window is closed, the previous entry is retained.
1166.3363.12 3.21 E-1
Setting Parameters R&S FSMR
Version 2:
The entry area consists of two parts:
¾ the editing line ¾ the character selection field
The cursor keys field .
Entry of text
or are used to toggle between entry in the editing line and the character selection
¾ Select parameter.
The data entry is active automatically upon calling the data entry window. In tables, the help line editor is accessed by the cursor key
In the editing line, the cursor is positioned at the beginning of the current entry.
¾ Position the cursor to the character selection field using the cursor
key The cursor highlights the first letter of the editor.
¾ Position the cursor to the character to be enter ed in the editing line
by means of the cursor keys
¾ Press the ENTER key or the spinwheel.
The character will be entered in the editing line.
.
.
or or the spinwheel.
Correcting the entry (Version 1)
¾ Move to the character << in the character selection field using the
spinwheel.
¾ Position the cursor behind the character to be deleted by moving
and pressing the spinwheel on << and >>.
¾ Move to the BACK field using the spinwheel and press the
spinwheel. The entry to the left of the cursor in the editing line will be deleted.
1166.3363.12 3.22 E-1
R&S FSMR Setting Parameters
Correcting the entry (Version 2)
Terminating the entry (Version 1)
Terminating the entry (Version2)
Interrupting the entry
¾ Position the cursor to the editing line by means of the cursor key
.
¾ Position the cursor after the character to be deleted by means of
the cursor keys
¾ Press the BACK key.
The entry to the left of the cursor will be deleted.
¾ Select the EXIT field using the spinwheel and press the spinwheel.
The data entry window will be closed; the new entry will be accepted by the instrument.
¾ Move the cursor to the editing line using the cursor key
¾ Press one of the unit keys or the spinwheel.
The data entry window will be closed; the new entry will be accepted by the instrument.
¾ Press the ESC key.
The data entry window will be closed; the previous entry will be retained.
and or with the spinwheel.
.
1166.3363.12 3.23 E-1
Setting Parameters R&S FSMR

Selection and Setting of Parameters via Tables

The spectrum analyzer uses numerous tables for display and configuration of instrument parameters.
The tables differ considerably in the num ber of lines, columns and inscriptions. The basic steps of operation for the selection and setting of parameters are, however, the same for all tables. Shown below is the typical entry of parameters into a table.
Note:
Most of the tables are coupled to a softkey menu which provides further functions for editing table entries such as deletion of tables , copying of lines or columns, marking of table elements , restoring default states.
The definition of individual tables and the operation of particular editing functions can be looked up in the description of the corresponding softkey menu.
1. Activating the table
2. Selection of the parameter
If the menu has only one table, the latter is activated
automatically subsequent to c alling this menu in most cases and the m arking cursor is positioned to the top field of the left column.
If the m enu contains various tables, the table of interest
must be selected using the s oftkey which is labeled w ith the title of the table.
marking cursor
NAME COMPATIBLE LIMIT CHECK TRACE MARGIN
GSM22UP off 1 0 dB LP1GHz on 2 0 dB LP1GHz off 1 0 dB MIL461A off 2 -10 dB
LIMIT LINES
Selection of the par ameter (or the s etting) is made using the marking cursor.
¾ Press the cursor keys to move to the wanted field.
or
¾ Rotate the spinwheel until the wanted field is marked. The cursor
keys are used to specify the direction of the spinwheel movement (horizontal or vertical)
When shifting the cursor, elements may be s kipped which can
not be edited. Table elements, which can not be s elected are indicated by a different color.
1166.3363.12 3.24 E-1
R&S FSMR Setting Parameters
3. Editing the marked parameter
a) Toggling between two
states
¾ Press the ENTER key or the spinwheel.
The parameter /the setting has been selected.
The selected parameter can be edited the way described below:
NAME CO MPATIBLE LIMIT CHEC K TRACE MARGIN
GSM22UP off 1 0 dB LP1G Hz on 2 0 dB LP1G Hz off 1 0 dB MIL461A off 2 -10 dB
checkmark
LIMIT LINES
TRACE
1
2
3
MARGIN
-10 dB
If an elem ent of a table c an be s witched on and off only, the unit keys are used to toggle between these two states.
¾ Press one of the unit keys.
The table elem ent is switched on and pr ovided with a checkmark. (√).
¾ Press one of the unit keys once more.
The table element is switched off.
or
¾ Press one of the unit keys.
The table element is switched on, "on" is displayed.
¾ Press one of the unit keys once more.
The table element is switched off, "off" is displayed
1166.3363.12 3.25 E-1
Setting Parameters R&S FSMR
b) Opening a data entry
window
c) Opening a selection
table
If a table entr y consists of an ( alpha-) numeric value, the corresponding entry window will be opened af ter selecting the numeric value by pressing the ENTER key or the spinwheel.
Note 1: For numeric instrument parameters, the editing
operation may be s tarted by entering any number or letter on the fr ont panel or on the external keyboard. In this case, the data entr y window is opened automatically.
Note 2: For alphanumeric table fields for which the help line
editor version 2 is required, the help line editor will be opened by pressing the CURSOR DOWN key after ENTER.
If a table entr y may have various states (e.g., colors from a color pallet, fixed filter bandwidths, etc.), a table indic ating all possible states is displayed with selection. The current state is and marked by a checkmark and by the cursor.
¾ Set the cursor to the desired setting. ¾ Press one of the unit keys.
The setting is switched on and m arked (). The selection table closes and the value is transferred to the original table. The cursor is positioned automatically on the nex t table element.
Abortion of entry ¾ Press the ESC key.
The current entry/ selection is aborted and the original setting is restored.
Scrolling
Some tables contain more entries than c an be dis played on one screen page. In this case, a margin of the table, w hose slider shows the current position in the text.
¾ Actuate the PAGE UP or PAGE DOWN softkeys.
The table is paged forward or back by one page.
¾ Press cursor key
The table is scrolled up or down by one line.
scrollbar is displayed at the r ight
or .
1166.3363.12 3.26 E-1
R&S FSMR Menu Overview

Menu Overview Receiver Mode

The following sections gives a graphical overview of the R&S FSMR menus. Side menus are marked by an arrow directed to the left/right, submenus by an arrow showing upwards. The menus appear in the order corresponding to the arrangement of keys on the front panel. The available hotkeys and the LOCAL menu appearing during the remote control of the instrument are also displayed.
The functions of menus are described in detail in Chapter 4. The IEC/IEEE-bus command associated with each softkey is indicated. In addition, the softkey list at the of Chapter 6 gives the assignment of IEC/IEEE-bus commands to softkeys.

FREQ Key

1166.3363.12 3.27 E-3