Application Firmware Module FSE-K11
Measurement according to Transmission Standards
1057.3392.02
Printed in the Federal
Republic of Germany
1057.3440.42-031
Page 2
Supplement to FSE-K10 and FSE-K11 Manuals
K10/K11 Limit Lines
For FSE options K10 and K11 (GSM mobile station and GSM base stati on anal yzer ), many internal limit
lines are used which are deri ved from the applicable s tandards. To make these limit lines access ible to
the user, they are available as data sets. There are 6 data sets:
K10_XGSM.LIA Limit lines for mobiles, standards EGSM, PGSM and RGSM
K10_DCS.LIA Limit lines for mobiles, standards DCS1800 (GSM 1800)
K10_PCS.LIA Limit lines for mobiles, standards PCS1900 (GSM 1900)
K11_XGSM.LIA Limit lines for base stations, standards EGSM, PGSM und RGSM
K11_DCS.LIA Limit lines for base stations, standards DCS1800 (GSM 1800)
K11_PCS.LIA Limit lines for base stations, standards PCS1900 (GSM 1900)
To load the data sets, proceed as follows:
1. Copy the desired data sets (one or several) from the CD to disk or analyzer.
2. Press the hardkey RECALL, and enter path (for example A:\) via softkey EDIT PATH
(see operating manual, section 2.7.3).
3. Select menu item “Lines“ via softkey ITEMS TO RECALL
(this is the default setting; see operating manual, section 2.7.3.1).
4. Select the desired data set via softkey DATA SET LIST. This will cause all limit lines of the selected
data sets to be copied to the limit line directory of the instrument.
From the GSM software, the limit lines can be accessed and edited via the softkey EDIT in the
righthand side menus of the following measurements:
•
Power vs time
•
Modulation spectrum
•
Transient spectrum or
•
Spurious
Outside GSM measurements, the limit lines are accessible in the analyzer mode of the instrument.
Rohde & Schwarz makes every effort to keep the safety standard of its products up to date and to offer
its customers the highest possible degree of safety. Our products and the auxiliary equipment required
for them are designed and tested in accordance with the relevant safety standards. Compliance with
these standards is continuously monitored by our quality assurance system. This product has been
designed and tested in accordance with the EC Certificate of Conformity and has left the manufacturer’s
plant in a condition fully complying with safety standards. To maintain this condition and to ensure safe
operation, observe all instructions and warnings provided in this manual. If you have any questions
regarding these safety instructions, Rohde & Schwarz will be happy to answer them.
Furthermore, it is your responsibility to use the product in an appropriate manner. This product is
designed for use solely in industrial and laboratory environments or in the field and must not be used in
any way that may cause personal injury or property damage. You are responsible if the product is used
for an intention other than its designated purpose or in disregard of the manufacturer's instructions. The
manufacturer shall assume no responsibility for such use of the product.
The product is used for its designated purpose if it is used in accordance with its operating manual and
within its performance limits (see data sheet, documentation, the following safety instructions). Using
the products requires technical skills and knowledge of English. It is therefore essential that the
products be used exclusively by skilled and specialized staff or thoroughly trained personnel with the
required skills. If personal safety gear is required for using Rohde & Schwarz products, this will be
indicated at the appropriate place in the product documentation.
Observe
operating
instructions
Supply
voltage
ON/OFF
Weight
indication for
units >18 kg
Standby
indication
Symbols and safety labels
Danger of
electric
shock
Direct
current
(DC)
Warning!
Hot
surface
PE terminal Ground
Alternating
current (AC)
Direct/alternating
current (DC/AC)
Ground
terminal
Device fully
protected by
double/reinforced
insulation
Attention!
Electrostatic
sensitive
devices
1171.0000.42-02.00 Sheet 1
Page 10
Safety Instructions
Observing the safety instructions will help prevent personal injury or damage of any kind caused by
dangerous situations. Therefore, carefully read through and adhere to the following safety instructions
before putting the product into operation. It is also absolutely essential to observe the additional safety
instructions on personal safety that appear in other parts of the documentation. In these safety
instructions, the word "product" refers to all merchandise sold and distributed by Rohde & Schwarz,
including instruments, systems and all accessories.
Tags and their meaning
DANGER
WARNING
CAUTION This tag indicates a safety hazard with a low potential of risk for the user
ATTENTION
NOTE
These tags are in accordance with the standard definition for civil applications in the European
Economic Area. Definitions that deviate from the standard definition may also exist. It is therefore
essential to make sure that the tags described here are always used only in connection with the
associated documentation and the associated product. The use of tags in connection with unassociated
products or unassociated documentation can result in misinterpretations and thus contribute to personal
injury or material damage.
This tag indicates a safety hazard with a high potential of risk for the
user that can result in death or serious injuries.
This tag indicates a safety hazard with a medium potential of risk for the
user that can result in death or serious injuries.
that can result in slight or minor injuries.
This tag indicates the possibility of incorrect use that can cause damage
to the product.
This tag indicates a situation where the user should pay special attention
to operating the product but which does not lead to damage.
Basic safety instructions
1. The product may be operated only under
the operating conditions and in the
positions specified by the manufacturer. Its
ventilation must not be obstructed during
operation. Unless otherwise specified, the
following requirements apply to
Rohde & Schwarz products:
prescribed operating position is always with
the housing floor facing down, IP protection
2X, pollution severity 2, overvoltage
category 2, use only in enclosed spaces,
max. operation altitude max. 2000 m.
Unless specified otherwise in the data
sheet, a tolerance of ±10% shall apply to
the nominal voltage and of ±5% to the
nominal frequency.
2. Applicable local or national safety
regulations and rules for the prevention of
accidents must be observed in all work
performed. The product may be opened
only by authorized, specially trained
personnel. Prior to performing any work on
the product or opening the product, the
product must be disconnected from the
supply network. Any adjustments,
replacements of parts, maintenance or
repair must be carried out only by technical
personnel authorized by Rohde & Schwarz.
Only original parts may be used for
replacing parts relevant to safety (e.g.
power switches, power transformers,
fuses). A safety test must always be
performed after parts relevant to safety
have been replaced (visual inspection, PE
conductor test, insulation resistance
measurement, leakage current
measurement, functional test).
3. As with all industrially manufactured goods,
the use of substances that induce an
allergic reaction (allergens, e.g. nickel)
such as aluminum cannot be generally
excluded. If you develop an allergic
reaction (such as a skin rash, frequent
sneezing, red eyes or respiratory
difficulties), consult a physician immediately
to determine the cause.
1171.0000.42-02.00 Sheet 2
Page 11
Safety Instructions
4. If products/components are mechanically
and/or thermically processed in a manner
that goes beyond their intended use,
hazardous substances (heavy-metal dust
such as lead, beryllium, nickel) may be
released. For this reason, the product may
only be disassembled, e.g. for disposal
purposes, by specially trained personnel.
Improper disassembly may be hazardous to
your health. National waste disposal
regulations must be observed.
5. If handling the product yields hazardous
substances or fuels that must be disposed
of in a special way, e.g. coolants or engine
oils that must be replenished regularly, the
safety instructions of the manufacturer of
the hazardous substances or fuels and the
applicable regional waste disposal
regulations must be observed. Also
observe the relevant safety instructions in
the product documentation.
6. Depending on the function, certain products
such as RF radio equipment can produce
an elevated level of electromagnetic
radiation. Considering that unborn life
requires increased protection, pregnant
women should be protected by appropriate
measures. Persons with pacemakers may
also be endangered by electromagnetic
radiation. The employer is required to
assess workplaces where there is a special
risk of exposure to radiation and, if
necessary, take measures to avert the
danger.
7. Operating the products requires special
training and intense concentration. Make
certain that persons who use the products
are physically, mentally and emotionally fit
enough to handle operating the products;
otherwise injuries or material damage may
occur. It is the responsibility of the
employer to select suitable personnel for
operating the products.
8. Prior to switching on the product, it must be
ensured that the nominal voltage setting on
the product matches the nominal voltage of
the AC supply network. If a different voltage
is to be set, the power fuse of the product
may have to be changed accordingly.
9. In the case of products of safety class I with
movable power cord and connector,
operation is permitted only on sockets with
earthing contact and protective earth
connection.
10. Intentionally breaking the protective earth
connection either in the feed line or in the
product itself is not permitted. Doing so can
result in the danger of an electric shock
from the product. If extension cords or
connector strips are implemented, they
must be checked on a regular basis to
ensure that they are safe to use.
11. If the product has no power switch for
disconnection from the AC supply, the plug
of the connecting cable is regarded as the
disconnecting device. In such cases, it
must be ensured that the power plug is
easily reachable and accessible at all times
(length of connecting cable approx. 2 m).
Functional or electronic switches are not
suitable for providing disconnection from
the AC supply. If products without power
switches are integrated in racks or systems,
a disconnecting device must be provided at
the system level.
12. Never use the product if the power cable is
damaged. By taking appropriate safety
measures and carefully laying the power
cable, ensure that the cable cannot be
damaged and that no one can be hurt by
e.g. tripping over the cable or suffering an
electric shock.
13. The product may be operated only from
TN/TT supply networks fused with max.
16 A.
14. Do not insert the plug into sockets that are
dusty or dirty. Insert the plug firmly and all
the way into the socket. Otherwise this can
result in sparks, fire and/or injuries.
15. Do not overload any sockets, extension
cords or connector strips; doing so can
cause fire or electric shocks.
16. For measurements in circuits with voltages
V
> 30 V, suitable measures (e.g.
rms
appropriate measuring equipment, fusing,
current limiting, electrical separation,
insulation) should be taken to avoid any
hazards.
17. Ensure that the connections with
information technology equipment comply
with IEC 950/EN 60950.
18. Never remove the cover or part of the
housing while you are operating the
product. This will expose circuits and
components and can lead to injuries, fire or
damage to the product.
1171.0000.42-02.00 Sheet 3
Page 12
Safety Instructions
19. If a product is to be permanently installed,
the connection between the PE terminal on
site and the product's PE conductor must
be made first before any other connection
is made. The product may be installed and
connected only by a skilled electrician.
20. For permanently installed equipment
without built-in fuses, circuit breakers or
similar protective devices, the supply circuit
must be fused in such a way that suitable
protection is provided for users and
products.
21. Do not insert any objects into the openings
in the housing that are not designed for this
purpose. Never pour any liquids onto or into
the housing. This can cause short circuits
inside the product and/or electric shocks,
fire or injuries.
22. Use suitable overvoltage protection to
ensure that no overvoltage (such as that
caused by a thunderstorm) can reach the
product. Otherwise the operating personnel
will be endangered by electric shocks.
23. Rohde & Schwarz products are not
protected against penetration of water,
unless otherwise specified (see also safety
instruction 1.). If this is not taken into
account, there exists the danger of electric
shock or damage to the product, which can
also lead to personal injury.
24. Never use the product under conditions in
which condensation has formed or can form
in or on the product, e.g. if the product was
moved from a cold to a warm environment.
matching Rohde & Schwarz type (see
spare parts list). Batteries and storage
batteries are hazardous waste. Dispose of
them only in specially marked containers.
Observe local regulations regarding waste
disposal. Do not short-circuit batteries or
storage batteries.
28. Please be aware that in the event of a fire,
toxic substances (gases, liquids etc.) that
may be hazardous to your health may
escape from the product.
29. Please be aware of the weight of the
product. Be careful when moving it;
otherwise you may injure your back or other
parts of your body.
30. Do not place the product on surfaces,
vehicles, cabinets or tables that for reasons
of weight or stability are unsuitable for this
purpose. Always follow the manufacturer's
installation instructions when installing the
product and fastening it to objects or
structures (e.g. walls and shelves).
31. Handles on the products are designed
exclusively for personnel to hold or carry
the product. It is therefore not permissible
to use handles for fastening the product to
or on means of transport such as cranes,
fork lifts, wagons, etc. The user is
responsible for securely fastening the
products to or on the means of transport
and for observing the safety regulations of
the manufacturer of the means of transport.
Noncompliance can result in personal injury
or material damage.
25. Do not close any slots or openings on the
product, since they are necessary for
ventilation and prevent the product from
overheating. Do not place the product on
soft surfaces such as sofas or rugs or
inside a closed housing, unless this is well
ventilated.
26. Do not place the product on heatgenerating devices such as radiators or fan
heaters. The temperature of the
environment must not exceed the maximum
temperature specified in the data sheet.
27. Batteries and storage batteries must not be
exposed to high temperatures or fire. Keep
batteries and storage batteries away from
children. If batteries or storage batteries are
improperly replaced, this can cause an
explosion (warning: lithium cells). Replace
the battery or storage battery only with the
1171.0000.42-02.00 Sheet 4
32. If you use the product in a vehicle, it is the
sole responsibility of the driver to drive the
vehicle safely. Adequately secure the
product in the vehicle to prevent injuries or
other damage in the event of an accident.
Never use the product in a moving vehicle if
doing so could distract the driver of the
vehicle. The driver is always responsible for
the safety of the vehicle; the manufacturer
assumes no responsibility for accidents or
collisions.
33. If a laser product (e.g. a CD/DVD drive) is
integrated in a Rohde & Schwarz product,
do not use any other settings or functions
than those described in the documentation.
Otherwise this may be hazardous to your
health, since the laser beam can cause
irreversible damage to your eyes. Never try
to take such products apart, and never look
into the laser beam.
Page 13
Por favor lea imprescindiblemente antes de
la primera puesta en funcionamiento las
siguientes informaciones de seguridad
Informaciones de seguridad
Es el principio de Rohde & Schwarz de tener a sus productos siempre al día con los estandards de
seguridad y de ofrecer a sus clientes el máximo grado de seguridad. Nuestros productos y todos los
equipos adicionales son siempre fabricados y examinados según las normas de seguridad vigentes.
Nuestra sección de gestión de la seguridad de calidad controla constantemente que sean cumplidas
estas normas. Este producto ha sido fabricado y examinado según el comprobante de conformidad
adjunto según las normas de la CE y ha salido de nuestra planta en estado impecable según los
estandards técnicos de seguridad. Para poder preservar este estado y garantizar un funcionamiento
libre de peligros, deberá el usuario atenerse a todas las informaciones, informaciones de seguridad y
notas de alerta. Rohde&Schwarz está siempre a su disposición en caso de que tengan preguntas
referentes a estas informaciones de seguridad.
Además queda en la responsabilidad del usuario utilizar el producto en la forma debida. Este producto
solamente fue elaborado para ser utilizado en la indústria y el laboratorio o para fines de campo y de
ninguna manera deberá ser utilizado de modo que alguna persona/cosa pueda ser dañada. El uso del
producto fuera de sus fines definidos o despreciando las informaciones de seguridad del fabricante
queda en la responsabilidad del usuario. El fabricante no se hace en ninguna forma responsable de
consecuencias a causa del maluso del producto.
Se parte del uso correcto del producto para los fines definidos si el producto es utilizado dentro de las
instrucciones del correspondiente manual del uso y dentro del margen de rendimiento definido (ver
hoja de datos, documentación, informaciones de seguridad que siguen). El uso de los productos hace
necesarios conocimientos profundos y el conocimiento del idioma inglés. Por eso se deberá tener en
cuenta de exclusivamente autorizar para el uso de los productos a personas péritas o debidamente
minuciosamente instruidas con los conocimientos citados. Si fuera necesaria indumentaria de
seguridad para el uso de productos de R&S, encontrará la información debida en la documentación del
producto en el capítulo correspondiente.
Símbolos y definiciones de seguridad
Ver manual
de
instrucciones
del uso
Informaciones
para
maquinaria
con uns peso
de > 18kg
Peligro de
golpe de
corriente
¡Advertencia!
Superficie
caliente
Conexión a
conductor
protector
Conexión
a tierra
Conexión
a masa
conductora
¡Cuidado!
Elementos de
construción
con peligro de
carga
electroestática
El aparato está
protegido en su
totalidad por un
aislamiento de
doble refuerzo
potencia EN
MARCHA/PARADA
Indicación
Stand-by
Corriente
continua
DC
Corriente
alterna AC
Corriente
continua/alterna
DC/AC
1171.0000.42-02.00 página 1
Page 14
Informaciones de seguridad
Tener en cuenta las informaciones de seguridad sirve para tratar de evitar daños y peligros de toda
clase. Es necesario de que se lean las siguientes informaciones de seguridad concienzudamente y se
tengan en cuenta debidamente antes de la puesta en funcionamiento del producto. También deberán
ser tenidas en cuenta las informaciones para la protección de personas que encontrarán en otro
capítulo de esta documentación y que también son obligatorias de seguir. En las informaciones de
seguridad actuales hemos juntado todos los objetos vendidos por Rohde&Schwarz bajo la
denominación de „producto“, entre ellos también aparatos, instalaciones así como toda clase de
accesorios.
Palabras de señal y su significado
PELIGRO Indica un punto de peligro con gran potencial de riesgo para el
ususario.Punto de peligro que puede llevar hasta la muerte o graves
heridas.
ADVERTENCIA Indica un punto de peligro con un protencial de riesgo mediano para el
usuario. Punto de peligro que puede llevar hasta la muerte o graves
heridas .
ATENCIÓN Indica un punto de peligro con un protencial de riesgo pequeño para el
usuario. Punto de peligro que puede llevar hasta heridas leves o
pequeñas
CUIDADO Indica la posibilidad de utilizar mal el producto y a consecuencia
dañarlo.
INFORMACIÓN Indica una situación en la que deberían seguirse las instrucciones en el
uso del producto, pero que no consecuentemente deben de llevar a un
daño del mismo.
Las palabras de señal corresponden a la definición habitual para aplicaciones civiles en el ámbito de la
comunidad económica europea. Pueden existir definiciones diferentes a esta definición. Por eso se
debera tener en cuenta que las palabras de señal aquí descritas sean utilizadas siempre solamente en
combinación con la correspondiente documentación y solamente en combinación con el producto
correspondiente. La utilización de las palabras de señal en combinación con productos o
documentaciones que no les correspondan puede llevar a malinterpretaciones y tener por
consecuencia daños en personas u objetos.
Informaciones de seguridad elementales
1. El producto solamente debe ser utilizado
según lo indicado por el fabricante referente
a la situación y posición de funcionamiento
sin que se obstruya la ventilación. Si no se
convino de otra manera, es para los
productos R&S válido lo que sigue:
como posición de funcionamiento se define
principialmente la posición con el suelo de la
caja para abajo , modo de protección IP 2X,
grado de suciedad 2, categoría de
sobrecarga eléctrica 2, utilizar solamente en
estancias interiores, utilización hasta 2000 m
sobre el nivel del mar.
A menos que se especifique otra cosa en la
hoja de datos, se aplicará una tolerancia de
±10% sobre el voltaje nominal y de ±5%
sobre la frecuencia nominal.
2. En todos los trabajos deberán ser tenidas en
cuenta las normas locales de seguridad de
trabajo y de prevención de accidentes. El
producto solamente debe de ser abierto por
personal périto autorizado. Antes de efectuar
trabajos en el producto o abrirlo deberá este
ser desconectado de la corriente. El ajuste,
el cambio de partes, la manutención y la
reparación deberán ser solamente
efectuadas por electricistas autorizados por
R&S. Si se reponen partes con importancia
para los aspectos de seguridad (por ejemplo
el enchufe, los transformadores o los
fusibles), solamente podrán ser sustituidos
por partes originales. Despues de cada
recambio de partes elementales para la
seguridad deberá ser efectuado un control de
1171.0000.42-02.00 página 2
Page 15
Informaciones de seguridad
seguridad (control a primera vista, control de
conductor protector, medición de resistencia
de aislamiento, medición de medición de la
corriente conductora, control de
funcionamiento).
3. Como en todo producto de fabricación
industrial no puede ser excluido en general
de que se produzcan al usarlo elementos
que puedan generar alergias, los llamados
elementos alergénicos (por ejemplo el
níquel). Si se producieran en el trato con
productos R&S reacciones alérgicas, como
por ejemplo urticaria, estornudos frecuentes,
irritación de la conjuntiva o dificultades al
respirar, se deberá consultar inmediatamente
a un médico para averigurar los motivos de
estas reacciones.
4. Si productos / elementos de construcción son
tratados fuera del funcionamiento definido de
forma mecánica o térmica, pueden generarse
elementos peligrosos (polvos de sustancia
de metales pesados como por ejemplo
plomo, berilio, níquel). La partición elemental
del producto, como por ejemplo sucede en el
tratamiento de materias residuales, debe de
ser efectuada solamente por personal
especializado para estos tratamientos. La
partición elemental efectuada
inadecuadamente puede generar daños para
la salud. Se deben tener en cuenta las
directivas nacionales referentes al
tratamiento de materias residuales.
5. En el caso de que se produjeran agentes de
peligro o combustibles en la aplicación del
producto que debieran de ser transferidos a
un tratamiento de materias residuales, como
por ejemplo agentes refrigerantes que deben
ser repuestos en periodos definidos, o
aceites para motores, deberan ser tenidas en
cuenta las prescripciones de seguridad del
fabricante de estos agentes de peligro o
combustibles y las regulaciones regionales
para el tratamiento de materias residuales.
Cuiden también de tener en cuenta en caso
dado las prescripciones de seguridad
especiales en la descripción del producto.
6. Ciertos productos, como por ejemplo las
instalaciones de radiación HF, pueden a
causa de su función natural, emitir una
radiación electromagnética aumentada. En
vista a la protección de la vida en desarrollo
deberían ser protegidas personas
embarazadas debidamente. También las
personas con un bypass pueden correr
peligro a causa de la radiación
electromagnética. El empresario está
comprometido a valorar y señalar areas de
trabajo en las que se corra un riesgo de
exposición a radiaciones aumentadas de
riesgo aumentado para evitar riesgos.
7. La utilización de los productos requiere
instrucciones especiales y una alta
concentración en el manejo. Debe de
ponerse por seguro de que las personas que
manejen los productos estén a la altura de
los requerimientos necesarios referente a
sus aptitudes físicas, psíquicas y
emocionales, ya que de otra manera no se
pueden excluir lesiones o daños de objetos.
El empresario lleva la responsabilidad de
seleccionar el personal usuario apto para el
manejo de los productos.
8. Antes de la puesta en marcha del producto
se deberá tener por seguro de que la tensión
preseleccionada en el producto equivalga a
la del la red de distribución. Si es necesario
cambiar la preselección de la tensión
también se deberán en caso dabo cambiar
los fusibles correspondientes del prodcuto.
9. Productos de la clase de seguridad I con
alimentación móvil y enchufe individual de
producto solamente deberán ser conectados
para el funcionamiento a tomas de corriente
de contacto de seguridad y con conductor
protector conectado.
10. Queda prohibida toda clase de interrupción
intencionada del conductor protector, tanto
en la toma de corriente como en el mismo
producto ya que puede tener como
consecuencia el peligro de golpe de corriente
por el producto. Si se utilizaran cables o
enchufes de extensión se deberá poner al
seguro, que es controlado su estado técnico
de seguridad.
11. Si el producto no está equipado con un
interruptor para desconectarlo de la red, se
deberá considerar el enchufe del cable de
distribución como interruptor. En estos casos
deberá asegurar de que el enchufe sea de
fácil acceso y nabejo (medida del cable de
distribución aproximadamente 2 m). Los
interruptores de función o electrónicos no
son aptos para el corte de la red eléctrica. Si
los productos sin interruptor están integrados
en construciones o instalaciones, se deberá
instalar el interruptor al nivel de la
instalación.
1171.0000.42-02.00 página 3
Page 16
Informaciones de seguridad
12. No utilice nunca el producto si está dañado el
cable eléctrico. Asegure a través de las
medidas de protección y de instalación
adecuadas de que el cable de eléctrico no
pueda ser dañado o de que nadie pueda ser
dañado por él, por ejemplo al tropezar o por
un golpe de corriente.
13. Solamente está permitido el funcionamiento
en redes de distribución TN/TT aseguradas
con fusibles de como máximo 16 A.
14. Nunca conecte el enchufe en tomas de
corriente sucias o llenas de polvo. Introduzca
el enchufe por completo y fuertemente en la
toma de corriente. Si no tiene en
consideración estas indicaciones se arriesga
a que se originen chispas, fuego y/o heridas.
15. No sobrecargue las tomas de corriente, los
cables de extensión o los enchufes de
extensión ya que esto pudiera causar fuego
o golpes de corriente.
16. En las mediciones en circuitos de corriente
con una tensión de entrada de Ueff > 30 V se
deberá tomar las precauciones debidas para
impedir cualquier peligro (por ejemplo
medios de medición adecuados, seguros,
limitación de tensión, corte protector,
aislamiento etc.).
17. En caso de conexión con aparatos de la
técnica informática se deberá tener en
cuenta que estos cumplan los requisitos de
la EC950/EN60950.
18. Nunca abra la tapa o parte de ella si el
producto está en funcionamiento. Esto pone
a descubierto los cables y componentes
eléctricos y puede causar heridas, fuego o
daños en el producto.
19. Si un producto es instalado fijamente en un
lugar, se deberá primero conectar el
conductor protector fijo con el conductor
protector del aparato antes de hacer
cualquier otra conexión. La instalación y la
conexión deberán ser efecutadas por un
electricista especializado.
20. En caso de que los productos que son
instalados fijamente en un lugar sean sin
protector implementado, autointerruptor o
similares objetos de protección, deberá la
toma de corriente estar protegida de manera
que los productos o los usuarios estén
suficientemente protegidos.
21. Por favor, no introduzca ningún objeto que
no esté destinado a ello en los orificios de la
caja del aparato. No vierta nunca ninguna
clase de líquidos sobre o en la caja. Esto
puede producir corto circuitos en el producto
y/o puede causar golpes de corriente, fuego
o heridas.
22. Asegúrese con la protección adecuada de
que no pueda originarse en el producto una
sobrecarga por ejemplo a causa de una
tormenta. Si no se verá el personal que lo
utilice expuesto al peligro de un golpe de
corriente.
23. Los productos R&S no están protegidos
contra el agua si no es que exista otra
indicación, ver también punto 1. Si no se
tiene en cuenta esto se arriesga el peligro de
golpe de corriente o de daños en el producto
lo cual también puede llevar al peligro de
personas.
24. No utilice el producto bajo condiciones en las
que pueda producirse y se hayan producido
líquidos de condensación en o dentro del
producto como por ejemplo cuando se
desplaza el producto de un lugar frío a un
lugar caliente.
25. Por favor no cierre ninguna ranura u orificio
del producto, ya que estas son necesarias
para la ventilación e impiden que el producto
se caliente demasiado. No pongan el
producto encima de materiales blandos como
por ejemplo sofás o alfombras o dentro de
una caja cerrada, si esta no está
suficientemente ventilada.
26. No ponga el producto sobre aparatos que
produzcan calor, como por ejemplo
radiadores o calentadores. La temperatura
ambiental no debe superar la temperatura
máxima especificada en la hoja de datos.
1171.0000.42-02.00 página 4
Page 17
Informaciones de seguridad
27. Baterías y acumuladores no deben de ser
expuestos a temperaturas altas o al fuego.
Guardar baterías y acumuladores fuera del
alcance de los niños. Si las baterías o los
acumuladores no son cambiados con la
debida atención existirá peligro de explosión
(atención celulas de Litio). Cambiar las
baterías o los acumuladores solamente por
los del tipo R&S correspondiente (ver lista de
piezas de recambio). Baterías y
acumuladores son deshechos problemáticos.
Por favor tirenlos en los recipientes
especiales para este fín. Por favor tengan en
cuenta las prescripciones nacionales de cada
país referente al tratamiento de deshechos.
Nunca sometan las baterías o acumuladores
a un corto circuito.
28. Tengan en consideración de que en caso de
un incendio pueden escaparse gases tóxicos
del producto, que pueden causar daños a la
salud.
29. Por favor tengan en cuenta que en caso de
un incendio pueden desprenderse del
producto agentes venenosos (gases, líquidos
etc.) que pueden generar daños a la salud.
30. No sitúe el producto encima de superficies,
vehículos, estantes o mesas, que por sus
características de peso o de estabilidad no
sean aptas para él. Siga siempre las
instrucciones de instalación del fabricante
cuando instale y asegure el producto en
objetos o estructuras (por ejemplo paredes y
estantes).
31. Las asas instaladas en los productos sirven
solamente de ayuda para el manejo que
solamente está previsto para personas. Por
eso no está permitido utilizar las asas para la
sujecion en o sobre medios de transporte
como por ejemplo grúas, carretillas
elevadoras de horquilla, carros etc. El
usuario es responsable de que los productos
sean sujetados de forma segura a los medios
de transporte y de que las prescripciones de
seguridad del fabricante de los medios de
transporte sean tenidas en cuenta. En caso
de que no se tengan en cuenta pueden
causarse daños en personas y objetos.
32. Si llega a utilizar el producto dentro de un
vehículo, queda en la responsabilidad
absoluta del conductor que conducir el
vehículo de manera segura. Asegure el
producto dentro del vehículo debidamente
para evitar en caso de un accidente las
lesiones u otra clase de daños. No utilice
nunca el producto dentro de un vehículo en
movimiento si esto pudiera distraer al
conductor. Siempre queda en la
responsabilidad absoluta del conductor la
seguridad del vehículo y el fabricante no
asumirá ninguna clase de responsabilidad
por accidentes o colisiones.
33. Dado el caso de que esté integrado un
producto de laser en un producto R&S (por
ejemplo CD/DVD-ROM) no utilice otras
instalaciones o funciones que las descritas
en la documentación. De otra manera pondrá
en peligro su salud, ya que el rayo laser
puede dañar irreversiblemente sus ojos.
Nunca trate de descomponer estos
productos. Nunca mire dentro del rayo laser.
1171.0000.42-02.00 página 5
Page 18
Page 19
FSE-K11Presetting the FSE/FSIQ
1 Getting Started
This chapter is mainly intended for operators using the GSM BTS analyzer mode for the first time.
Operators are guided step by step through the most frequent base station tests.
The chapter describes FSE/FSIQ setup, relevant settings and individual measurements.
Layout and content are application-oriented and provide important information on FSE/FSIQ as well as
the GSM/DCS1800/PCS1900 system.
It is recommended to read through this chapter and perform the operations on the FSE/FSIQ at the
same time. T his ensures a simple and rapid introduction to the GSM BTS firmware func tions so that
correct and useful measurement results are achieved.
Some of the steps are provided with reference num bers, eg , relating to a m ore detailed information
on the opposite page.
For more detailed information on the individual menus and softk eys kindly refer to the reference part in
chapter 2.
Note:The descriptions of some of the functions are provided with notes pointing out that some
functions of FSE-K11 are available only if Vector Signal Analysis Option FSE-B7 is
installed. These notes apply only to FSE, since FSIQ c ontains this option in the basic
version.
Power attenuator))
Fig. 1-1 BTS with separate RX input and TX output (connections for measurem ents in the RX band
shown as dotted lines)
Ext. reference signal
Ext.
TX
trigger
/ RX
Power
attenuator
Trigger
Fig. 1-2 Test setup for BTS with combined RX/TX input/output
Caution:Another test setup is required for measurements in the RX band (see sections 1.1.2 and
1.1.3)
À Connect the frame trigger output of the BTS to the rear-panel BNC connector:
EXT TRIG GATE
(external trigger or gate input) of the FSE/FSIQ. W ith the aid of this external trigger signal FSE/FSIQ is
able to trigger exactly to any slot even if several slots of the frame are occupied.
1057.3392.021.2E-2
Page 21
FSE-K11Presetting the FSE/FSIQ
À Connect the antenna output (or TX output) of the BT S via a suitable power attenuator to the RF input of
FSE/FSIQ (except for measurements in the RX band; special test setup).
1057.3992.021.3E-2
Page 22
FSE-K11Connecting the FSE/FSIQ
1.1.2 Test Setup for Measurements in the RX Band (Spurious and
Modulation Spectrum)
RX-band test setup for BTSs with combined RX/TX input/output:
Low noise
amplifier
BTS
TX
DiplexerDiplexer
RX
Antenna
signal
RX
Bandstop
(optional)
TX
50 Ohms Load
RX-band test setup for BTSs with separate RX input and TX output:
(optional)
RF
FSE
50 Ohms Load
BTS
RX
TX
Low noise
Amplifier
(optional)
FSE
1.1.3 Test Setup for Measuring Spurious in the TX and outside the TX/RX
Band (<>TX Band):
Antenna
BTS
Signal
Attenuator
(20dB)
Bandstop
(optional)
FSE
1057.3392.021.4E-2
Page 23
FSE-K11Presetting the FSE/FSIQ
Additional information
The stringent requirements for measurements of spurious in the RX band on BTSs using a
combined RX input/TX output demand adequate suppression of the carrier signal(s) (max. carr ier
level < -10 dBm at the FSE/FSIQ input), eg with the aid of bandstop filters and/or diplexers. Since
the requirements for G SM phase 1 are more severe than those f or phase 2, an external low-noise
preamplifier may be required in addition because of the higher noise floor of FSEB, FSEM and
FSEK compared to FSEA.
With BTSs using a
To obtain maximum sensitivity, 0 dB attenuation is selected when RX BAND is set on the
FSE/FSIQ.
Softkey
values the insertion loss of the preceding diplex ers and bandstop filters), which is then taken into
account in the FSE display.
The attenuation of the external attenuator entered under Settings is ignored when the RX band is
measured.
RX BAND GAIN
separate
RX input, the TX channel is sufficiently suppressed at the RX input.
defines the gain of the external pream plifier (or in the case of negative
Caution: The level at the FSE input of +20 dBm with a 0 dB input attenuator must under no
circumstances be exceeded.
The following warning is output when the RX-BAND softkey is pressed on the screen.
Caution:
Connect bandpass or bandstop to suppress TX-band!
Notes on amplifier:
The noise figure of the preamplifier should be as low as possible (LF < approx. 6 dB) and the gain in
the range 15 to 25 dB so that the noise floor of FSE/FSIQ (approx. 15 to 17 dB with FSEA30,
approx. 18 to 21 dB with FSEB/M/K30) can be neglected.
Make sure that the preamplifier is not overdriven by insufficiently suppressed TX signals (for this
reason the amplifier gain should not be too high).
1057.3992.021.5E-2
Page 24
FSE-K11Measuring the Phasen/Frequency Error
1.2 Presetting the FSE/FSIQ
Step 1
CONFIGURATION
MODE
SETUP
EXT REF
FREQUENCY
REF
INT EXT
CONFIGURATION
MODE
SETUP
GSM BTS
ANALYZER
SETTINGS
Setting reference frequency
À Press SETUP key.
À Press softkey
EXT. REF FREQUENCY
and enter the frequency depending on
the available frequency standard.
À Switch softkey
REF INT/EXT
to EXT
(external reference frequency).
Step 2
Selecting the standard
À Press the
BTS ANALYZER
after the other and then
The menu for selecting the various
standards is displayed.
MODE
key and softkeys
and
SETTINGS
.
one
GSM
BTS TEST
SETTINGS
P-GSM 900
E-GSM 900
GSM 1800
(DCS 1800)
GSM 1900
(PCS 1900)
R-GSM 900
PHASE I
PHASE II
PHASE II+
1057.3392.021.6E-2
Page 25
FSE-K11Presetting the FSE/FSIQ
9
Additional information
Step 1
To attain the required frequency accuracy in the frequency error measurement
ERROR)
2,...16 has to be applied to the FSE/FSIQ.
a highly accurate (deviation < 1*10
-
) external reference frequency n x 1 MHz with n = 1,
(PHASE/FREQ
1057.3992.021.7E-2
Page 26
FSE-K11Measuring the Phasen/Frequency Error
Step 3
P-GSM 900
E-GSM 900
R-GSM 900
DCS 1800
GSM 1800
PCS 1900
GSM 1900
with GSM900 and DCS1800:
PHASE I
PHASE II
PHASE II+
À Select standar d P-GSM900, E-G SM900,
DCS 1800, PCS 1900 or R-GSM,
depending on the BTS to be tested.
À Select between Phase I, II or II+. With
some standards, the softkeys are not
available (PCS 1900) or partly available
because no phases or not all of them
are defined.
Press
to return to the SETTINGS
menu.
Step 4
Entering the external attenuation
EXTERNAL
ATTEN
À Press the
EXTERNAL ATTEN
softkey.
À Enter the external attenuation. (For
recommended attenuation see ô)
1057.3392.021.8E-2
Page 27
FSE-K11Presetting the FSE/FSIQ
Additional information
Step 4
ô
The following values are recomm ended for the external attenuator to ensure that the FSE/FSIQ
RF input is protected and the sensitivity of the FSE/FSIQ is not reduced too much.
The FSE/FSIQ level display is automatically corrected by the entered value.
Max. powerRecommended ext. attenuation
≥ 55 to 60 dBm
≥ 50 to 55 dBm
≥ 45 to 50 dBm
≥ 40 to 45 dBm
≥ 35 to 40 dBm
≥ 30 to 35 dBm
≥ 25 to 30 dBm
≥ 20 to 25 dBm
< 20 dBm 0 dB
35 to 40 dB
30 to 35 dB
25 to 30 dB
20 to 25 dB
15 to 20 dB
10 to 15 dB
5 to 10 dB
0 to 5 dB
1057.3992.021.9E-2
Page 28
FSE-K11Measuring the Phasen/Frequency Error
Step 5
Selecting the transmission channel
ARFCN /
FREQUENCY
À Press the
softkey.
ARFCN / FREQUENCY
ARFCN
ARFCN/
AUTOSELECT
FREQUENCY
À Press softkey
AUTOSELECT
ARFCN
With
ARFCN, ARFCN
or
FREQUENCY
selected:
.
À Enter the transmission channel number
of your base station. í
FSE/FSIQ will tune to this frequency in
compliance with the selected standard.
With
ARFCN AUTOSELECT
, FSE/FSIQ
automatically searches for the active
transmission channel on condition that
frequency hopping is not active.
FREQUENCY
With
selected:
À Enter the transmission frequency of your
base station or any other frequency (eg
the IF of a BTS module).
Frequencies of the following range are
permissible:
1.8MHz frequency (maximum device
frequency - 1.8MHz).
À Press
to return to the SETTINGS
menu.
1057.3392.021.10E-2
Page 29
FSE-K11Presetting the FSE/FSIQ
Additional information
Step 5
í
Enter one of the possible channel numbers 1 to 124 for the P-GSM frequency range. Enter a
number between 975 and 1023 or 0 for the extended GSM band. For PCN (DCS1800) the
numbers 512 to 885 are available, for PCS1900 numbers 512 to 810. For R-GSM channel
numbers 955 to 1023 and 0 are available in addition to the numbers of the P-GSM band (1 to 124).
To protect the FSE/FSIQ, the settings made in the PO WER menu are automatically checked to
make sure that the input level at the FSE/FSIQ does not exceed 27 dBm.
To prevent the occurrence of an overload, a warning is displayed for those settings which would
lead to the threshold level being exceeded (
POWER, POWER CTRL LEVEL)
and the set value is ignored.
POWER CLASS, EXTERNAL ATTEN, OUT PUT MS
1057.3992.021.13E-2
Page 32
FSE-K11Measuring the Phasen/Frequency Error
or
STATIC PWR
CTRL LEVEL
DYNAM PWR
CTRL LEVEL
LIMIT/PWR
COUPLED
SIGNAL
POWER
LIMIT LINE
REF POWER
Step 7
With reduced level of the BTS enter the
static and/or the dynamic power control
level in addition. (The default setting is 0)
û:
À Enter the desired static power control
level N (integer, N = 0, 1, 2 to 6). û.
À Enter the desired dynamic power control
level N (integer, N = 0, 1, 2 to 15). ø
The coupling between power and that of
the power class as well as between the
power actually applied and the limits
selected with respect to this power can be
suppressed (softkey
COUPLED
).
À In this case press softkey
POWER
and specify the actual output
power of the BTS (external attenuation is
taken into account).
With softkey
LIMIT LINE REF POWER
the power determining the selection of
limits can be selected.
LIMIT/PWR
SIGNAL
SLOT NO.
SLOT NO.
SLOT NO.
MANUAL
SLOT NO.
AUTOSELECT
.
.
.
Step 8
À Press softkey
À Select the desired slot number 0 to 7
using
SLOT NO. MANUAL
automatic search with
SELECT.
The
FSE/FSIQ thus knows in which slot
the measurement should be performed.
SLOT NO.
or start the
SLOT NO. AUTO
ù
1057.3392.021.14E-2
Page 33
FSE-K11Presetting the FSE/FSIQ
Additional information
Step 7
û
Static Power Control LevelRelative power referred to
00 dB
1-2 dB
2-4 dB
3-6 dB
4-8 dB
5-10 dB
6-12 dB
ø
Max. 16 dynamic power control levels are specif ied in addition to the static power control levels.
maximum output power
(GSM/DCS/PCS)
They reduce the base station level referred to the maximum output power of the base station and
the static power control level in up to 16 steps of 2 dB (dynamic power control level N = 0 to 15,
attenuation = N × 2 dB).
Step 8
ù
A precondition for automatic search is that only one slot is active.
(For an automatic detection of the slot the fram e trigger of the base station must be c onnected to
the external trigger input of the unit and the trigger source set to
EXTERN
, see above).
1057.3992.021.15E-2
Page 34
FSE-K11Measuring the Phasen/Frequency Error
Step 9
Selecting the current midamble
MIDAMBLE
MIDAMBLE
AUTO
TSC_0
TSC_1
TSC_2
TSC_3
TSC_4
TSC_5
TSC_6
TSC_7
À Press softkey
MIDAMBLE .
The table of training sequences
(
m
idamble)
is displayed. î
In the position AUTO (default s etting) FSE
uses the training sequence (midamble)
according to the slot number , for instanc e
TSC_2 for slot number 2.
À If you wish to use another training
sequence, select it accordingly.
1057.3392.021.16E-2
Page 35
FSE-K11Presetting the FSE/FSIQ
Additional information
Step 9
î
Table for the midamble used in the GSM system:
NAMEPATTERN(Bit No.: 61 - 86)
TSC_00 0 1 0 0 1 0 1 1 1 0 0 0 0 1 0 0 0 1 0 0 1 0 1 1 1
An external
trigger signal (frame trigger of BTS) is
required for almost all measurements.
The frame trigger pulse is output once
for each frame. In the ideal case it
comes with bit 0 in slot 0 of the frame. ü
But this depends on the BTS to be
tested. For this reason trigger
adjustment is provided (described later).
The default setting for the trigger is
therefore EXTERN. The default value for
the external trigger level is 1.4 V but, if
required, the level can be adapted by
pressing
EXTERN
and then entering a
new value in the data entry field. The
polarity of the trigger signal can be set
using softkey
SLOPE POS NEG
. Default
setting: positive slope.
À Press softkey
TRIGGER ADJUST
the trigger reference. Automatic
(preferably) or manual adjustment can
be performed.
to set
SLOT
ADJUST
AUTO SLOT
VIDEO
ADJUST
TRIGGER
LEVEL
SLOPE
POS NEG
1057.3392.021.18E-2
Page 37
FSE-K11Presetting the FSE/FSIQ
Additional information
Step 10
External frame tr igger
ü
1.4V
T
0 dB
Slot No. 0 Slot No. 1 Slot No. 2
Transition from bit 73 to 74
Delay between burst edge and transitionbit 13/14 dof midamble of slot N (where N=0,1,2...7):
T=<283 .23 us> +N *57 6.92 us
Slot No. 3
Slot No. 4
Slot No. 5 Slot No. 6
Slot No. 7
Slot No. 0
1057.3992.021.19E-2
Page 38
FSE-K11Measuring the Phasen/Frequency Error
Either automatic adjustment:
AUTO FRAME
ADJUST
AUTO SLOT
ADJUST
FRAME
FINE
Only one slot can be active.
À Press softkey
AUTO FRAME ADJUST
FSE/FSIQ sets the time reference
between trigger signal and midamble ü
and takes this referenc e into account for
all measurements .
AUTO ADJUST
acts
upon all slots.
If only the slot used at present is to be
adjusted:
À Press softkey
AUTO SLOT ADJUST
and adjust the time referenc e of the slot
used at present.
or manual adjustment:
(if automatic adjustment is not possible,
eg with bursts without midamble or when
option FSE-B7 is not installed.)
À Press softkey
FRAME COARSE
and set
the time reference of the external trigger
to the frame with the aid of the
spinwheel (the -20 dB point of the rising
burst edge should be in the display
center. This point is marked by
crosshairs made up of the vertical
screen center line and the -20 dB line).
.
FRAME
FINE
or
SLOT
ADJUST
À Press softkey
FRAME FINE
and adjust
the time reference of all slots in the
same way.
10
If the slots have a different length (156
or 157 bits), a fine adjustment can be
performed for each slot with the aid of
SLOT ADJUST
.
FSE/FSIQ is now set as required by the
BTS to be measured and
measurements can be started.
1057.3392.021.20E-2
Page 39
FSE-K11Presetting the FSE/FSIQ
Additional information
10
Fig 1-3Correcting adjustment for FRAME FINE
Caution:The manual adjustment to the -20 dB point of the trigger edge can yield only an
approximate time reference since the exact r eference to the midamble is missing. The
automatic adjustment produces the correct time reference to the midamble ( depending
on the steepness of the burst edge the -20 dB point can be up to approx . 7
from the screen center).
µ
s away
1057.3992.021.21E-2
Page 40
FSE-K11Measuring the Phasen/Frequency Error
1.3 Measuring the Phase/Frequency Error
Note:The phase and frequency error can only be measured if the Vector Signal Analyzer option
FSE-B7 is integrated.
Step 1
PHASE/FREQ
ERROR
NO. OF
BURSTS
À Press the softkey
ERROR
.
PHASE/FREQ
Step 2
À Press softkey
the number of measurements to be
carried out.
Default setting:
The number of bursts defined in the
selected standard is set with SET TO
STANDARD (in the phase/frequency error
measurement: 20 for all standards).
Press
menu.
NO. OF BURSTS
NO. OF BURSTS
to return to the SETTINGS
and set
= 1.
1057.3392.021.22E-2
Page 41
FSE-K11Presetting the FSE/FSIQ
Additional information
BTS setting:
At least 1 slot on the selected ARFCN must be active. All slots may be occupied.
The measurement can also be performed on BTSs with active SFH (slow frequency hopping).
FSE/FSIQ measures at the set r eceive frequency (ARFCN). If a number of bursts > 1 is set, only
results are considered at which FSE/FSIQ is able to synchronize. Thus even if the carrier frequency
of the slot to be measured changes cyclically in successive frames, only slots with the set slot
number at the selected ARFCN will be considered in the measurement.
However, this will considerably slow down the measurement sequence (approx. by the factor 8
against a non-hopping BTS). This disadvantage can be avoided if a trigger signal is pr ovided for the
FSE/FSIQ (triggering not on every frame but only if the selected slot is at the set ARFCN).
1057.3992.021.23E-2
Page 42
FSE-K11Measuring the Phasen/Frequency Error
Step 3
SINGLE
CONTINUOUS
or
À Start the measurement sequence with
SINGLE
When the measurement is completed,
an overview of the numeric modulation
errors is displayed in result window A f or
the 147 useful bits. ô
or
CONTINUOUS
.
1057.3392.021.24E-2
Page 43
FSE-K11Presetting the FSE/FSIQ
Additional information
Step 3
ô
The information on instantaneous frequency and power and the following numerical results are
displayed in the window. It also contains the corresponding limit values and pass/fail information.
•
Max. hold and average value of peak phase error
•
Max. hold and average value of rms phase error
• Max. hold and average value of frequency error
The phase error versus time is displayed in window B over the 147 useful bits of the normal burst.
• Trace 2: Max Hold (maximum positive phase error of all measured bursts)
• Trace 3: Min Hold (maximum negative phase error of all measured bursts)
1057.3992.021.25E-2
Page 44
FSE-K11Measuring the Carrier Power versus Time
1.4 Measuring the Average Carrier Power
Step 1
CARRIER
POWER
NO. OF
BURSTS
SYNC TO
MIDAMBLE
NO. OF
BURSTS
SET
MANUAL
SET TO
STANDARD
À Press softkey
CARRIER POWER.
Step 2
À Press softkey
the desired number of measurements.
The number of bursts defined in the
selected standard is set with SET TO
STANDARD (in the carrier power
measurement: 1 for all standards).
NO. OF BURSTS
and set
Step 3
Possible only if option FSE-B7 is
installed:
In the default setting, softkey
MIDAMBLE
is active and thus the
synchronization to the burst midamble is
set. This setting is of advantage if the
burst to be measured contains a
midamble and a bit-accurate time
reference is required. (softkey colour:
green)
SYNC TO
À Disable
SYNC TO MIDAMBLE
for bursts
without midamble or if maximum
measurement s peed is required (s oftk ey
colour: grey) ô.
1057.3392.021.26E-2
Page 45
FSE-K11Measuring the Carrier Power versus Time
Additional information
BTS setting:
At least three neighbouring slots of a TDMA frame of equal power must be active.
If the BTS supports SFH (slow frequency hopping), SFH should be selected.
FSE/FSIQ measures at the set r eceive fr equency (ARFCN). If a num ber of bursts > 1 has been set,
only measurement results will be considered at which FSE/FSIQ is able to synchronize. Thus even if
the carrier frequency of the slot to be measured changes cyclically in successive frames, only slots
with the set slot number at the selected ARFCN will be considered in the measurement.
However, this will considerably slow down the measurement sequence (approx. by the factor 3
compared to a BTS without SFH). This disadvantage can be avoided if a trigger signal is provided
for the FSE/FSIQ (triggering not on every frame but only if the selected slot is at the set ARFCN).
Step 3
ô
Explanation concerning SYNC TO MIDAMBLE:
If option FSE-B7 is built in and the softkey is active, FSE/FSIQ uses the vector signal analyzer to
synchronize to the midamble. Power measurement is performed in the scalar spectrum analyzer
mode by means of the time reference obtained by th signal demodulation.
Thus an accurate and updated tim e reference is ensured (even with a jittered trigger puls e). If the
softkey is not active, the vector analyzer mode is off and FSE/FSIQ uses the time reference set
using ADJUST TRIGGER.
1057.3992.021.27E-2
Page 46
FSE-K11Measuring the Carrier Power versus Time
Step 4
Setting the measurement bandwidth
MEAS
BANDWITH
CONDITIONS
NORM EXTR
MEAS MAX
OUTPUT PWR
À Press softkey
A table indicating the settable
bandwidths is displayed on the screen.
Selecting STANDARD (default setting)
sets the measurement bandwith
specified in the standard.
Or:
Select the measurem ent bandwidth 3 kHz
or 1 MHz using the up/down keys and
press the Enter key.
MEAS BANDWITH.
Step 5
À The tolerance range specified in the
standards for the measurement is
selected with
adapted to the measurement conditions.
Default setting is NORM.
CONDITIONS
and thus
Step 6
À The reference level at power control
level 0 (static and dynamic) is
determined with
PWR
and checked for the permissible
tolerance against the level entered with
softkey
CLASS
OUTPUT POWER
in menu
MEAS MAX OUTPUT
or
POWER
BTS TEST SETTINGS
.
Step 7
INC ST A T
PWR CTRL
1057.3392.021.28E-2
À The average carrier power is measured
at incremented static power control
levels (1, 2, 3, 4, 5, 6) and checked for
compliance with limit values. To do so
the BTS level must be reduced each
time by one step in interactive mode and
softkey
then to be pressed for the
measurement.í
INCR STAT PWR CTRL
has
Page 47
FSE-K11Measuring the Carrier Power versus Time
T
Additional information
Step 7
í
Upon completion of the measurement (after pressing seven times
corresponding previous level reduction of BTS) a results table similar to the one below is displayed:
accordingly for the measurements at
incremented dynamic power control
levels (at a specific s tatic power control
level) and press softkey
PWR CTRL
fifteen times. ÷. After this
INC DYNC
has been done for static power control
level 6, the display is switched from splitscreen to full-screen output of the table.
À The table can be scrolled using the
spinwheel or keys Ù and Ú.
or
Step 9
With the previous procedure m any power
control levels have to be gone through
even if only one power control level is to
be measured.
SGL MEAS
ON OFF
STATIC PWR
CTRL LEVEL
MEAS SGL
PWR LEVEL
CLEAR SGL
RESULT TAB
and
DYNAM PWR
CTRL LEVEL
If single power control levels are to be
measured, proceed as follows:
À Switch softkey
SGL MEAS
to
ON.
À Select the power control levels to be
measured using sof tkeys
CTRL LEVEL
and
DYNAM PWR CTRL
STATIC PWR
LEVEL.
À Start the power measurement of this
power control level using sof tkey
SGL PWR LEVEL
.
MEAS
The results obtained can be cleared from
the table using softkey
RESULT TAB
.
CLEAR SGL
The entries to be examined can be
selected with the cursor keys and the
spinwheel.
1057.3392.021.30E-2
Page 49
FSE-K11Measuring the Carrier Power versus Time
Additional information
Step 8
÷
An example of the table is given below in which the static power control level has been
incremented once (to 1), then all dynamic power control levels (to 15) and again the static power
control level (to 6) have been incremented.
Example: Measurement of the dynamic power control level for static power level 1.
The dynamic power control level can be incremented for all static power control levels or only for
the required ones. The measurement sequence is completed when the
softkey is actuated again at static power control level 6 or when
again at static power control level 6 and dynamic power control level 15.
1057.3992.021.31E-2
INC STAT PWR CTRL
INC STAT PWR CTRL
is actuated
Page 50
FSE-K11Measuring the Carrier Power versus Time
1.5 Measuring the Carrier Power versus Time ô
Step 1
POWER
VS TIME
À Press softkey
POWER VS TIME.
Step 2
NO. OF
BURSTS
and
SYNC TO
MIDAMBLE
NO. OF
BURSTS
SET
MANUAL
SET TO
STANDARD
À Press softkey
NO. OF BURSTS
and set
the number of bursts to be measured.
The number of bursts defined in the
selected standard is set with SET TO
STANDARD
(1 for all standards).
Step 3
Possible only if option FSE-B7 is
installed
À Activate softk ey
and set synchronization to the burst
midamble if the burst to be measured
contains a midamble and a bit- accurate
time reference is required.
À Switch off
bursts without midamble or if m aximum
measurement speed is required. í
SYNC TO MIDAMBLE
SYNC TO MIDAMBLE
for
1057.3392.021.32E-2
Page 51
FSE-K11Measuring the Carrier Power versus Time
Additional information
BTS setting:
A single timeslot should be active on the transceiver (TRX) to be measured, all other timeslots of the
TDMA frame should be set to IDLE.
The measurement can also be performed if the BTS is in the SFH mode if option FSE-B7 is
installed and
To accelerate the measurement, switch off SFH on the BTS.
ô
Explanation concerning SYNC TO MIDAMBLE:
If option FSE-B7 is built in and the softkey is active, FSE/FSIQ uses the vector signal analyzer to
synchronize to the midamble. Power measurement is performed in the scalar spectrum analyzer
mode by means of the time reference obtained by th signal demodulation.
Thus an accurate and updated tim e reference is ensured (even with a jittered trigger puls e). If the
softkey is not active, the vector analyzer mode is off and FSE/FSIQ uses the time reference set
using ADJUST TRIGGER.
SYNC TO MIDAMBLE
is active in FSE/FSIQ.
Step 3
í
FSE/FSIQ measures at the set r eceive fr equency (ARFCN). If a num ber of bursts > 1 has been set,
only measurement results will be considered at which FSE/FSIQ is able to synchronize (only if
softkey
measured changes cyclically in successive frames, only slots with the set slot number at the
selected ARFCN will be considered in the measurement.
However, this will considerably slow down the measurement sequence (approx. by the factor 3
compared to SFH). This disadvantage can be avoided if a trigger signal is provided for the
FSE/FSIQ (triggering not on every frame but only if the selected slot is at the s et ARFCN). In this
case, the measurem ent can be perf ormed at s low frequency hopping without synchronization to the
midamble.
SYNC TO MIDAMBLE
is activated). Thus even if the carrier frequency of the slot to be
1057.3992.021.33E-2
Page 52
FSE-K11Measuring the Carrier Power versus Time
Step 4
MEAS
BANDWITH
REF MEAS
AUTO USER
START
REF MEAS
Setting the measurement bandwidth
À Press softkey
MEAS BANDWITH.
A table indicating the settable
bandwidths is displayed on the screen.
Selecting STANDARD (default setting)
sets the measurement bandwith
specified in the standard.
Or:
À Select the measurement bandwidth 3
kHz or 1 MHz using the up/down keys
and press the Enter key.
Step 5
REF MEAS AUTO / USER
select if a preview is to be perfor m ed prior
to each measurement for the c orrect level
adjustment of FSE/FSIQ or if the preview
is to be started exclusively by the user
START REF MEAS
using
Default setting is
REF MEAS AUTO
permits to
.
.
FULL BURST
RISINGFALLING
or
or
RESOLUTION
HIGH
or
Step 6
À Select the resolution for the display of
bursts: ÷
•
FULL BURST
displayed
•
HIGH RESOLUTION
bits are displayed with 1 dB level
resolution.
•
RISING:
edge with high time resolution.
•
FALLING
edge with high time resolution.
•
FALLING EDGE:
fallenden Burstflanke mit hoher
Zeitauflösung.
: The complete burst is
: The 147 useful
Display of the rising burst
: Display of the falling burst
Darstellung der
1057.3392.021.34E-2
Page 53
FSE-K11Measuring the Carrier Power versus Time
T
m
m
Additional information
Step 6
÷
30
20
10
0
-10
-20
-30
-40
34 dBm
34 dBm
21.7 dBm
21.7 dBm
1MAX
2MIN
3AVG
XG_BNU
RBW 1 MHzMax/Ref Lvl
VBW 1 MHz
SWT 800 Îs
LIMIT CHECK : PASSED
RF Att 20 dB
Mixer-20 dB
UnitdB
A
LN
TRG
1SA
2SA
3SA
EX
-50
-60
Date: 6.JUL.1999 13:11:02
Fig. 1-4 Display of burst for
FULL BURST
80 Îs/Center 935.2 MHz
1057.3992.021.35E-2
Page 54
FSE-K11Measuring the Carrier Power versus Time
Step 7
SINGLE
or
CONTINOUS
À Press softkey
measurement sequence or
CONTINUOUS for a continuous
measurement sequence.
The bursts are displayed according to
the selection made at step 4.
SINGLE
to start a single
1057.3392.021.36E-2
Page 55
FSE-K11Measuring the Carrier Power versus Time
This page is left intentionally empty to ensure that the user’s notes are always opposite to the additional
information.
1057.3992.021.37E-2
Page 56
FSE-K11Measuring the Spectrum due to Modulation
/
1.6 M easur ing the Spectrum due to Modulation
Note:The test setup is described in section 1.1.
Step 1
MODULATION
SPECTRUM
À Press softkey
SPECTRUM.
ô
MODULATION
Step 2
NO. OF
BURSTS
ARFCN
±1.8MHz
ARFCN ±/
TX BAND
or
RX BAND
GSM 900
.
NO. OF
BURSTS
SET
MANUAL
SET TO
STANDARD
oror
or
TX BAND
RX BAND
RX BAND
DCS1800
with GSM900
DCS1800
À Press softkey
NO. OF BURSTS
and
determine the number of bursts to be
measured.
The number of bursts defined in the
selected standard is set with SET TO
STANDARD
.
This depends on the selected frequency
band. í
Step 3
À Select the frequency range for the
measurement:
TX-BAND, RX-BAND
ARFCN ± 1.8 MHz/TX BAND
possible only with GSM/DCS phase 1).
Caution: The test setup is changed for
ARFCN ± 1.8 MHz
or combined band
. (RX band
measurement in RX band.
,
RX BAND
with PCS 1900
Step 4
SINGLE
FREQ SWEEP
or
1057.3392.021.38E-2
CONT
FREQ SWEEP
À Start the measurement in the frequency
range (spectral display) with
FREQ SWEEP
FREQ SWEEP
(single sweep) or
(continuous sweep). í
SINGLE
CONT
Page 57
FSE-K11Measuring the Spectrum due to Modulation
Additional information
BTS setting:
Set all timeslots of a T DMA f r ame to the same power and modulate them with pseudo-random data.
(Exception: timeslot 0 may also be modulated with normal BCCH data).
SFH (slow frequency hopping ) should be switched off.
Step 1
ô
The measurem ent is performed as prescribed by excluding the midamble of 50 to 90 % of the bur s t.
(Precondition: correct setting of Adjust Trigger in the SETTINGS menu; step 1
If required, the setting can be checked and corrected using the
ADJUST TRIGGER
ù).
softkey.
The measurement can be performed as a spectral display measurement or, depending on the
standard, as a zero-span measurement with subsequent output of tabulated numerical results.
Step 4
í
A spectral display is not possible in the combined band
SINGLE FREQ SWEEP, CONT FREQ SWEEP
40
30
20
10
0
-10
-20
-30
-40
-50
-60
Fig. 1-5 Spectral display (
SINGLE FREQ SWEEP
ARFCN ± 1.8 MHz/TX-BAND
cannot be operated).
) in the range
ARFCN ± 1.8 MHz
(softkeys
1057.3992.021.39E-2
Page 58
FSE-K11Measuring the Spectrum due to Modulation
Step 5
START
LIST
À Start the strictly standard-conforming
zero-span measurement with
LIST:
÷
START
1057.3392.021.40E-2
Page 59
FSE-K11Measuring the Spectrum due to Modulation
Additional information
Step 5
÷
Fig. 1-6 Result display for zero-span measurement (
START LIST
) in the range
ARFCN ±1.8 MHz
1057.3992.021.41E-2
Page 60
FSE-K11Measuring the Spectrum due to Modulation
1.7 Measurement of Spectrum due to Transients
Note:The test setup is described in section 1.1.
Step 1
TRANSIENT
SPECTRUM
NO. OF
BURSTS
MEAS
BANDWITH
NO. OF
BURSTS
SET
MANUAL
SET TO
STANDARD
À Press softkey
TRANSIENT
SPECTRUM.
Step 2
À Press softkey
the number of bursts to be measured.
The number of bursts defined in the
selected standard is set with SET TO
STANDARD
Default setting is
NO. OF BURSTS
.
NO OF BURSTS
and set
=1
Step 3
À Set the measurement bandwidth with
MEAS BANDWIDTH
the settable bandwidths is displayed on
the screen. Selecting STANDARD sets
the measurement bandwith specified in
the standard.
Selection is performed by means of the
up/down keys and the Enter key.
. A table indicating
Step 4
SINGLE
FREQ SWEEP
oder
CONT
FREQ SWEEP
or
START
LIST
1057.3392.021.42E-2
ô
À Press softkey
or
CONT FREQ SWEEP
SINGLE FREQ SWEEP
The measurement is started in the
spectral range.
Alternatively:
À Press softkey
START LIST
A zero-span measurement is started ÷.
. í
.
Page 61
FSE-K11Measuring the Spectrum due to Modulation
Additional information
BTS setting:
The measurement can be performed with or without SFH (slow frequency hopping) as required.
The FSE/FSIQ measurement routine is able to handle the two cases.
According to the standards, the slots of a TDMA frame can be measured at two different levels
depending on whether the BTS supports dynamic power control or not.
Step 4
ô
The measurement can be performed as an overview measurement in the frequency range with
spectrum display or as a zero span measurement and subsequent output of tabulated numerical
results depending on the standard. In both cases a maximum off set frequency of ±1.8 MHz is used
for the measurement.
í
Fig. 1-7 Display of spectrum due to transients of a GSM base station (meas urement in fr equency
range). Limit lines are inserted, the pass/fail info is displayed.
÷
Fig. 1-8 Display of transient spectrum list (zero-span measurement). Results , limits and pass/fail
information for the offs et frequencies prescribed by the standards and a general pas s/fail
information are also displayed.
1057.3992.021.43E-2
Page 62
FSE-K11Measuring the Spurious
1.8 Measurement of Spurious
Note:The test setup is described in section 1.1.
Step 1
SPURIOUS
SPURIOUS
START LIST
SGL STEP
SHOW
CONT LIST
SPURIOUS
SGL STEP
START
LIST
SWEEP
COUNT
TX BAND
<> TX BAND
RX BAND
TX BAND
±2.0 MHZ
SELECT
STEP
À Press softkey
SPURIOUS
.
SWEEP
COUNT
Step 2
In the submenu which can be opened by
means of soft key
SWEEP COUNT
number of sweeps for averaging in the RX
band or the number of sweeps for
determining the peak values within and
outside the TX band. The SWEEP
number required by the standard can be
set with SET TO STANDARD in the
submenu.
Press
to return to the SETTINGS
menu.
set the
1057.3392.021.44E-2
Page 63
FSE-K11Measuring the Spurious
Additional information
BTS setting:
The BTS is to be configured depending on the selected frequency band (TX, <>TX or RX band):
GSM900, DCS1800 phase 1:
TX band and <>TX band
If the BTS supports SFH (slow frequency hopping), SFH should be selected (change between B, M
and T channel). Switch every second timeslot to full power.
RX band:
Switch all slots to full power.
GSM900/DCS18000 phase 2:
The BTS is to be configured depending on the selected frequency band (TX, <>TX or RX band):
TX band:
Switch on a TRX with maximum power on all slots.
Switch off SFH.
<>TX band:
Switch every second timeslot to full power. If the BTS supports SFH, SFH should be selected.
RX-Band:
Switch all timeslots of the BTS to full power. Switch off SFH (slow frequency hopping).
1057.3992.021.45E-2
Page 64
FSE-K11Measuring the Spurious
Step 3
TX BAND
6(/(&7
67(3
TX SUPPR
ON OFF
or
<> TX BAND
or
RX BAND
À Determine the measurem ent range with
TX BAND, <> TX BAND
RX Band) or
RX BAND
(outside TX and
. The test setup
has to be modified depending on the
frequency band used (see setion 1.1.3).
Other subranges can be selected or
exluded in the used band using softkey
SELECT STEP
.
Measurement in <>TX Band:
À For BTSs with an output power of
approx. > 35 dBm, the carrier signal(s)
must be suppressed by at least 25 dB to
give FSE/FSIQ sufficient sensitivity for a
large measurement bandwidth
(≥ 300 kHz) and also to prevent
harmonics generated by FSE/FSIQ from
being measured. The suppression can
be performed using a bandstop filter or a
highpass filter if harmonics are
measured. In this case T X SUPPR ON
has to be activated in addition. FSE then
increases its sensitivity by 20 dB
(reference level reduced by 20 dB).
BTS SFH
ON OFF
SWEEPTIME
STD AUTO
or
BTS SFH
ON OFF
Step 4
À If the BTS is to be meas ured in the SFH
mode (slow frequency hopping), switch
BTS SHF
to ON. (Condition: ARFCN =
M channel). ô
Step 5
À Select between two settings for the
analyzer sweep time using
SWEEPTIME
long in some cases) required by the
selected standard can be set with
The minimum sweep time of the
analyzer is set with
. The sweep time (very
STD
AUTO
.
.
1057.3392.021.46E-2
Page 65
FSE-K11Measuring the Spurious
Additional information
Step 4
ô
The measurement time is extended as required (by the factor 3 compared to SFH OFF). For
measurements in the TX band the B and T channels are skipped in addition to the ARFCN channel.
1057.3992.021.47E-2
Page 66
FSE-K11Measuring the Spurious
Step 6
COSITING
À If a cositing BTS is to be measured for
GSM900/DCS1800 phase 2, activate the
softkey
COSITING
í.
Step 7
START
LIST
Alternativ:
START LI ST
SGL STEP
and
CONT SGL
STEP LIS T
À If a complete measurement is to be
performed in the selected TX, <>TX or
RX band, press softkey
START LIST
.
Alternative:
À For a
step-by-step
measurement of the
individual frequency bands (TX and <>
TX band), press softkeys START LIST
SGL STEP and CONT SGL LIST:
After the measurement has been
completed, a tabulated overview of results
with limits and pass/fail information is
displayed. ÷
1057.3392.021.48E-2
Page 67
FSE-K11Measuring the Spurious
T
Additional information
Step 6
í
A cositing BTS can send and receive in the GSM band and the DCS1800 band.
If measurements are per formed in the RX band of the system used (defined under
SETTINGS), the RX band of the other system is also measured.
Fig. 1-9 Example of a spurious list
In addition to the general information at the top of the table including the general PASS/FAIL
information, all frequency ranges, the m aximum level measured in the range and the applicable
range limit plus status information are listed.
A frequency range is defined by an identical RBW/VBW setting of the FSE/FSIQ and the same
LIMIT.
If the margin or the limit of a frequency range is exceeded, spurious frequencies with levels are
listed in addition (with numbers).
1057.3992.021.49E-2
Page 68
FSE-K11Sequence of Operations of FSE-K11 Measurements
2 General Information on Option FSE-K11
This application allows the following measurements to be performed on GSM900, DCS1800 or
PCS1900 base stations in line with standards (GSM 11.20, GSM 11.20-DCS, GSM11.21 and J-STD007 Air Interface) in a very convenient way:
•
PFE Phase/frequency error
•
CPW Carrier power
•
PVT Power versus time
•
MOD Spectrum due to modulation
•
TRA Spectrum due to transients
•
SPU Spurious emissions
This chapter gives basic information on the sequence of operations, settings and messages of the
instrument. Details on the individual m easurements will be found in chapter 3. The information given
under "General Remarks" is not needed for normal test operation. It is intended to provide a better
understanding of instrument settings and operations.
Cold Start
Warm Start
Opening FSE-K11
Changing within FSE-K11
A complete reset to default values is made.
The previous status is assumed to a large extent. This applies to all
parameter values and to the selected measurement. The settings
specific to the type of measurement and band are made. The default
measurement type (LIST/SINGLE) is selected. The result history is
cleared, so the results of previous measurements are no longer
available after a warm start !
Re-opening FSE-K11
Action: Exiting FSE-K11 via the mode m enu, changing to non-FSE-K11
mode, re-opening FSE-K11.
The response is the same as with a warm start except that the result
history remains stored.
Changing between FSE-K10 and FSE-K11
Action: Switching between FSE-K10 and FSE-K11 via the mode menu.
The response is the same as with a cold start.
Changing the standard
Each change of standard causes a complete r eset to the default values
valid for the standard in question. The reset is performed without any
warning to the user.
Changing the phase
Changes of phase are perform ed with as few modific ations as possible.
If problems due to incom patibility of output data are detected during a
phase change, the responsible group of parameters is reset to default
values. In such case a tempor ary message is output inf orming the user
of the modifications made.
With IEC/IEEE-bus control, the response is the same as with a change
of standard !
Possible problems and resulting actions:
Change of
frequency group:
1057.3440.412-1E-3
No problems are to be expected.
Page 69
Sequence of Operations of FSE-K11 MeasurementsFSE-K11
Level group:If the power class or the power control level of the
previous phase does not exist for the new phase,
the parameters assigned to the level group are set
to default values.
Warning: "OutputPower & ExtAtten set to Default!"
Changing the measurement
type / band
Measurement type
and band:
Changing the measurement type
When the measur ement type is changed (eg from SPU to MOD), initial
actions (basic settings) specific to the selected measurement type are
performed once. A def ault band or the band last active is selected. LIST
measurement (for SPU, TRA, MOD) or SINGLE measurement (PFE,
PVT) can be selected as the measurement mode. In the case of the
CPW m enu being opened, the measurement last active is selected as
the measurement mode.
Changing the band
If the band is changed, no visible responses take place as a rule.
There are no other settings or initial actions performed.
The response is the same as with a change of
measurement type (see below).
2.1 Sequence of Operations of FSE-K11 Measurements
Measurements performed with FSE-K11 can be divided into 2 types according to their sequence of
operations:
Measurements with and without pre-measurement
a
b
In the pre-measurement, the currently applied signal level, which is essential for the main measurement,
is determined if requir ed for the selected m easurement. T his applies to all relative meas urements ( see
below) which are based on a reference value. If the measured s ignal level is below a minimum value
(expected value –20 dB), the measurement is interrupted and a warning is output. Otherwise the m ain
measurement is performed.
On opening a measurement menu, presettings are first performed which set the FSE/FSIQ to the
default state required for the m easurem ent type in question. On starting the m easurem ent, the settings
required for the selected measurement are carried out. Then the measurement is performed, if
necessary in several consecutive steps. At the end of the measurement, post-settings (such as the
activation of markers) are made if necessary and results are displayed.
Pre-measurement===>Main measurement
PreSettings[PostSettings]PreSettingsPostSettings
[Sequence data]
Main measurement
PreSettingsPostSettings
[Sequence data]
1057.3440.412-2E-3
Page 70
FSE-K11Sequence of Operations of FSE-K11 Measurements
The following overview lists the types of FSE-K11 measurement in accordance with the above
description:
MeasurementModePre-
KXX PFE.....XX---
KXX CPW.....XXXX
KXX CPI.....XXXX
KXX PVT.....XX---Also without pre-measurement, user-selectable;
KXX MOD RXSINGLE,
KXX MOD RXLISTXXXX
KXX MOD Tx,
ARFCN ±1.8 MHz
KXX MOD Tx,
ARFCN ±1.8 MHz
KXX SPU xxx.....XXXX
CONT
SINGLE,
CONT
LISTXXXXBandwidth correc tion necessary (see below)
measurement
withoutwith
XX---
Sequence dataRemarks
normally with 2 pre-measurem ents
XX---Bandwidth correction necessary (see below)
Measurements of the spectrum due to modulation require a reference value for the limit lines to be
measured at 30 kHz bandwidth. To evaluate the actual carrier power (ie the signal power measured at
infinite bandwidth) a bandwidth-dependent correction value is added to the pre-m easured value: this is
at present 8 dB for 30 kHz (following GSM 05.50). The corrected value is also used for selecting the limit
lines.
The following diagram shows the sequenc e of operations of a m easurement perf ormed with FSE-K11.
This diagram applies to all measurem ents except for the CPW /CPI meas urement, which is shown in a
separate diagram (for sequence of operations see also chapter 3).
The measurement type is selected in the main menu. Prior to this, general settings are performed if
necessary, with most parameters also being acc essible from the side menus of the measur ements. On
opening a measurement menu, general settings specific to the measurement in question are made.
These settings are made only once. In the basic menu for a given measurement, the results of the
measurement last perfor med (table) are displayed in the full-screen mode (MOD , TRA, SPU, CPW ) on
opening the menu. For the CPI, PVT and PFE measurements, on the other hand, the screen settings for
the main measurement are activated immediately (for CPI the table cannot be displayed in the fullscreen mode). For MOD and SPU the band can be selected. T he table contents change depending on
the selected band.
The measurem ent start initiates the (f irst) pre-m easurem ent. After com pletion of the pre-m easurem ent,
the measured signal level is evaluated. If the pre-measurement is succ essful, either the second premeasurement is performed with optimum sensitivity (for PVT) or the m ain measurem ent is carried out
with level control orientated to the pre-measurement. The result is displayed at the end of the
measurement.
1057.3440.412-3E-3
Page 71
Sequence of Operations of FSE-K11 MeasurementsFSE-K11
0
Main menu
1
General
test
parameters
2
3
Selection of
measurement
4
Default setting
for preset
parameters
5
Measurement
menu
Setting of
measurement-specific
parameters (ARFCN, ...
Menu up
10
6
Display of
table
8
Optional:
band selection
7
MOD/
SPU
nein
9
Start
ja
10.210.1
Fig. 2-1Sequence of operations upon selecting a measurement
1057.3440.412-4E-3
Page 72
FSE-K11Sequence of Operations of FSE-K11 Measurements
Explanation of Legend:
0T he m ain menu of FSE-K11 is called: general settings (SETT INGS) or a m easurem ent can be
selected.
1General settings for GSM measurement (ARFCN, attenuation, power, trigger, ...).
2Main m enu: one of the FSE-K11 measurements PFE-CPW/CPI-PVT-MOD-T RA-SPU can be
selected.
3Selection of measurement type via main menu.
4FSE/FSIQ assumes default setting for the selected measurement.
5,6,7,8 In the basic menu for a given measurem ent, the results of the measurement last perfor med
(table) are displayed in the full-screen mode ( MOD, TRA, SPU, CPW ) on opening the m enu.
For the CPI, PVT and PFE measurem ents, on the other hand, the s creen settings for the m ain
measurement are activated immediately (for CPI the table cannot be displayed in the full-
screen mode). For MOD and SPU the band can be selected. The table contents change
depending on the selected band.
1057.3440.412-5E-3
Page 73
Sequence of Operations of FSE-K11 MeasurementsFSE-K11
10.110.2
9
Start
11
13.1
No
Carrier
detected
No
MOD/TRA/
PVT
ja
12
(First) pre-
measurement
13
Carrier power
sufficient
ja
NO
see detail
chart
14
PVT?
Yes
15
Second pre-
measurement
16
Hauptmessung
17
Display of results
(table / trace
No
see detail
chart
see detail
chart
Fig. 2-2Sequence of operations upon starting a measurement (overview)
1057.3440.412-6E-3
Page 74
FSE-K11Sequence of Operations of FSE-K11 Measurements
Explanation of Legend:
9Start of measurement with ’SINGLE’, ’CONTINUOUS’ or ’START LIST’.
11For the measurements TRA/PVT and MOD (not RX band), the (first) pre-measurement is
performed; for the measurements PFE, CPW/CPI, MOD-Rx and SPU the main measurement is
started immediately.
12The sequence of operations of a pre-measurement is presented in detail in a separate diagram
(see "General sequence of operations of a measurement").
This sequence chart applies to both pre-measurements and main measurements.
13Sufficient carrier power: the measured signal level (incl. bandwidth correction) must not be
more than 20 dB below the signal level to be expected for the settings made.
13.1An error message is output temporarily and the IEC/IEEE status bit set.
14, 15After the first pre-measurement, a second pre-measurement is carried out for PVT (see
"Sequence of operations of PVT measurement in detail") to determine the exact reference
value.
The second pre-measurement is performed with optimum sensitivity.
16Main measurement with level control orientated to pre-measurement. The sequence of
operations is presented in detail in a separate chart (see "Gener al sequence of oper ations of a
measurement") .
This sequence chart applies to both pre- and main measurements.
17The result is displayed as a table or a trace.
1057.3440.412-7E-3
Page 75
Sequence of Operations of FSE-K11 MeasurementsFSE-K11
10
From main menu
100
Start
101
AUTO / pre-
measurement
ja
102
(Frist)
pre-measzrement
No
107
Determination
and storage of
level last pre-
measurement
Yes
104
No carrier
detected
106
USER pre-
measurement?
No
No
103
Carrier power
sufficient
g
ja
105
Second pr e-
measurement
108
Main measurement
with settings of
(previos) pre-
measurement
109
Display of
trace including
limit lines
Fig. 2-3Sequence of operations of PVT measurement in detail
1057.3440.412-8E-3
Page 76
FSE-K11Sequence of Operations of FSE-K11 Measurements
The above chart presents the PVT measurement in detail. The sequence of oper ations is deter mined by
the setting for the pre-measurement –
With AUTO, the PVT measurement is performed as follows:
1st pre-measurement - 2nd pre-measurement - main measurement.
The settings for the fir st pre-m easurem ent (with m ax. 2 measured burs ts) ar e m ade in acc ordance with
the expected level. The settings for the second pre-measurement are made in accordance with the
signal level determined in the first pre- measurement, and level measurement is performed with max.
20 burs ts. The signal level determined in the second pre-m easurement is us ed as a reference for the
limit lines used in the subsequent main measurement.
Two-stage pre-measurement results in highly accurate level determination even in cases where the
signal level is considerably below the expected value since, thanks to the follow-up level control, it is
ensured that the measurement is always performed with optimum sensitivity.
With the USER setting, the pre-measurem ents are skipped, and the level determined in the last premeasurement is used both for setting the operating point and as a reference f or the limit lines. If no level
has been determined, the nominal value (expected signal level) is used.
This setting allows fast switching between the various burst views (Rising, Falling, Full, HighRes).
Actuating the Start RefMeas key will trigger a sequence of two pre-measurements without any main
measurement following.
REF MEAS AUTO or REF MEAS USER.
1057.3440.412-9E-3
Page 77
Sequence of Operations of FSE-K11 MeasurementsFSE-K11
19
ZeroSpan vs FrequSweep,
Trigger, Trace, Detector, ...
Yes
Settings:
InpAtt & RefLvl,
20
Sync to
Midamble
nein
No
Time reference
30
31
required?
200
See detail
Sync to midamble
PFE / PVT & CPW
32
Trigger
FreeRun
Yes
ja
33
34
35
36
Sweep and data
acquisition
n
37
Sweep end at
tirgger/gate
34
Waiting for trigger
event
No
39
Pre-
measurements
result or final
result
Yes
38
No
Number of Bursts
attianed and
SingleSweep
or List?
Fig. 2-4General sequence of operations of a measurement
1057.3440.412-10E-3
Page 78
FSE-K11Sequence of Operations of FSE-K11 Measurements
The sequence of operations is the same for each m easurement (the above sequence c hart applies to
both pre-measurements and main measurements):
19The settings needed for the intended measurement are made.
20Synchronization to midamble is optional for PVT and CPW/CPI and mandatory for PFE.
This choice is possible only if option FSE-B7 is fitted.
30-39Measurement without digital demodulator (ie without option FSE-B7).
31,33All measurements (except transient and spurious) require a time reference.
32,33If FreeRun is not selected, FSE-K11 waits for the (external) trigger event here too.
34FSE-K11 waits for the trigger event, then starts or continues the sweep.
35The sweep is enabled.
36Measured data are collected until the end of the sweep or, with gated sweep, during a specific
gate time.
37It is checked if the sweep is terminated.
38It is checked if the number of preset s weeps has been attained AND if the measurement is of
the SINGLE or the LIST type.
39The m easurement is completed if the preset number of sweeps has been attained and the
measurement is of the SINGLE or the LIST type.
If digital demodulation is active, there are s ome dif ferenc es concerning tr igger logic as com pared to the
sequence of operations shown in the chart "General sequence of operations of a measurement".
The sequence of operations with digital demodulation active is shown below for PFE and PVT/CPW:
1057.3440.412-11E-3
Page 79
Sequence of Operations of FSE-K11 MeasurementsFSE-K11
200
Sync to midamble
201
202
Trigger
FreeRun
206
Burst found?
Yes
No
Yes
No
203
204
205
Sweep and data
aaquisition
207
Burst not
found
203
Waiting for
trigger event
208
No. of Burst > 1
No
Yes
209
Sync found?
Yes
No
Measurement completed
210
Sync not
found
212
Trace and
Limit check
222
211
No. of Burst > 1
No
213
Max. No. of
Bursts and
SINGLE
Yes
Yes
No
Fig. 2-5PFE - Sequence of operations of synchronization in detail
1057.3440.412-12E-3
Page 80
FSE-K11Sequence of Operations of FSE-K11 Measurements
PFE – sequence of operations:
200-204 Waiting for trigger event or FreeRun (not meaningful !).
205Collection of measured data.
206Search for bursts in a predefined time window (compare settings for result buffer).
207,
208In case of an error, a message is output and the IEC/IEEE st atus bit 'Burst not Found' is set.
In the special case of "NofBur sts == 1", the error condition is ignored however dur ing further
processing.
209Search for sync pattern in the selected burst.
210,
211In case of an error , a message is output and the IEC/IEEE status bit 'Sync not Found' is set.
In the special case of "NofBursts == 1", however, the err or condition is ignored during f urther
processing.
212Evaluation of trace with numerical indication of the measured frequency error and phase
error.
213If the predefined number of bur sts has been attained and the m easurem ent has been star ted
with 'SINGLE', it ends at this point.
1057.3440.412-13E-3
Page 81
Sequence of Operations of FSE-K11 MeasurementsFSE-K11
200
Sync to midamble
251
252
Trigger
FreeRun
256
Sync f ound?
Yes
No
Yes
No
253
254
255
Sweep and data
acquisition
257
Sync not
found
253
Waiting for
trigger event
No. ofBurst > 1
258
Yes
No
259
Trace and
Limit check
222
Measurement compledet
260
Max No. of bursts
and SINGLE?
Yes
No
Fig. 2-6PVT/CPW - sequence of operations of synchronization in detail
1057.3440.412-14E-3
Page 82
FSE-K11Sequence of Operations of FSE-K11 Measurements
PVT/CPW – sequence of operations:
The sequence of operations for PVT/CPW differs from that for PFE in that no burst search is performed.
Only the sync pattern (items 256 and 257) is searched within a limited range of the recorded data
stream (compar e "Settings") to establish a high-precision time ref erence. In the case of an error, the
procedure is the same as for PFE, including the special case of "NofBursts == 1".
200-254 Waiting for trigger event or FreeRun (not meaningful !).
255Collection of measured data.
256Search for sync pattern in the selected burst.
257,258 In case of an error, a m essage is output and the IEC/IEEE st atus bit 'Sync not Found' is set.
In the special case of "NofBur sts == 1", the error condition is ignored however dur ing further
processing.
259Evaluation of trace with limit check.
260If the predefined number of bur sts has been attained and the m easurem ent has been star ted
with 'SINGLE', it ends at this point.
1057.3440.412-15E-3
Page 83
Sequence of Operations of FSE-K11 MeasurementsFSE-K11
1
Main menu
2
3
MEAS MAX
OUTPUT PWR
4
Sta:= Dyn:= 0
Measurement (0,0)
5
13
Half-screen display
of table
12
Messung mit
Ctrl Lvl (Sta,Dyn)
6
All Pwr Ctrl
levels?
No
7
Max. Dyn Ctrl
Level?
ja
10
INC STATIC
PWR LEVEL
11
Sta := Sta + 1
Dyn := 0
Yes
No
14
Full-screen display
of table
No
8
INC DYNAM
PWR LEVEL
Yes
9
Dyn := Dyn + 1
Fig. 2-7Sequence of operations of CPW measurement in detail
1057.3440.412-16E-3
Page 84
FSE-K11Sequence of Operations of FSE-K11 Measurements
The sequence of operations of a CPW m easurement ( BTS/K11) is shown above. The m easurement is
initiated by pressing the ’MEAS MAX OUTPUT PWR’ key. Then either the static or the dynamic power
control level can be increm ented as required. Each increm entation of the static power control level will
reset the dynamic power control level to 0 (zero). If the initializing k ey ’MEAS MAX OUTPUT PW R’ is
actuated again during the current session, the session is aborted (measurement status: ABORTED).
Entries by the user are dependent on the availability of keys ’INC STATIC PWR LEVEL’ and ’INC
DYNAM PWR LEVEL’: if the maxim um value of the dynamic power control level is attained, the ’INC
DYNAM PWR LEVEL’ key is inhibited so that only the static power control level can be increased. If the
maximum value is attained here too, pressing the key again causes the session to be term inated and
the total result to be displayed on the full screen.
With the ’full-screen table’ status, both INC keys are inhibited, and a new session can be started only by
pressing the ’MEAS MAX OUTPUT PWR’ key.
Unlike the CPW measurement, the CPI m easurement allows any single carrier-power m easurements.
In contrast to all other measurements performed with FSE-K11, single CPI measurements are not
related to one another. There is no total result for the session. The CPI menu can be exited and
activated any time without this having any effect on the table contents of the CPI measurement.
CPI measurements are intended m ainly for developm ent engineers in the lab who m odify parameter s of
their workstation and want to investigate the effect s thereof on the carrier power. CPW m easurem ents,
on the other hand, are aimed at verifying compliance with acceptance test specifications of the GSM
standard.
CPI measurements, therefor e, are only used for evaluating absolute levels; no relative c ompar isons are
made between the levels obtained with different power control level settings. Operator control is
described in the additional information following the description of the softkey menus.
Explanation of Legend:
Selection of CPW measurement, activation of main menu (deactivation of CPI measurement if necessary)
3,4Start of measurement with maximum power in line with standard (CTRL LEVEL 0.0).
5,13End of sweep, result is entered into table.
6Measurement completed?
7-12If a power control level is incremented, the next sweep is started.
14End of measurement, switchover to full screen.
1057.3440.412-17E-3
Page 85
Sequence of Operations of FSE-K11 MeasurementsFSE-K11
2.1.1 Errors and Failures During Measurements
There are three main sources of error in GSM measurements:
Carrier signal missing:
Causes:Wrong operating frequency (ARFCN), SFH active, test line defective or incorrect
attenuation, wrong power class
Effects:Pr e-m easurem ent:inadequate power meas urem ent (eg average meas urement with
SFH), measurement is aborted.
Sweep stops (in measurements with midamble synchronization).
Main measurement:s weep stops when midamble synchronization (PFE, PVT, CPW)
is active and number of bursts > 1. When signal is applied again,
the sweep continues.
Trigger missing:
In all triggered measurements: sweep stops (indicated by trigger LED).
The following is recommended for the trigger source:
•
Optimal solution:External trigger. This allows all measurements without restrictions.
•
FreeRun:Allowed for SPU and TRA if not possible otherwise
(MOD-Tx without gating also possible).
Burst missing:
Causes:SFH active, wrong frame trigger
Effects:In triggered measurements, limit lines may be exceeded (PVT);
sweep stops (in measurements with midamble synchronization).
2.1.2 Abort of Measurements
a) Abort by User
In manual operation, each running measurem ent can be aborted by pressing the start k ey again (ie the
key with which the measurement was started).
Starting another measurement likewise caus es an abort of the running m easurem ent. This also applies
to switchover from SINGLE to CONTINUOUS or LIST, for example.
all
Measurements are also aborted on exiting the menu in question (New! From 1.90 for
or modifying a test parameter (even in side menus).
measurements)
All aborts, except for spurious single-step measurements, are without warning. Only if – during a running
spurious single-step measurem ent – the user attempts to res tart this measurem ent or to start the LIST
measurement, is he queried to confirm the (re)start before it is made.
For remote control, a special ABO RT comm and is available which causes the current measurem ent to
be aborted (see manual for IEC/IEEE-bus control).
1057.3440.412-18E-3
Page 86
FSE-K11Sequence of Operations of FSE-K11 Measurements
b) Abort by Instrument
Measurements are norm ally performed until the end. Only in the case of insuff icient carrier power (see
above) are measurements aborted after the pre-measurement and an error message is output.
If the maximum num ber of table entries (presently 50) is exceeded in a (spurious) measurement, the
measurement is continued but no new entry is made in the table. This does not apply to the CPW
measurement, where all possible single results (7 x 16 = 112) are entered in the table. In CPI
measurements, the last 40 single measurements are stored.
2.1.3 Results of Measurements
MeasurementSingle resultLimit lineTable
PFEPeak and RMS phase error,
CPW,
CPI
PVTUpper and
MOD
SG
MOD List
Rx/Tx
MOD
List
ARFCN ±1.8 MHz
TRA
Sgl
TRA
List
SPUFor each section:
frequency error in ppm
One entry per single measurem ent:
CPW: Ctrl Level , measured value, limit value
and evaluation
CPI: test parameter, measured value
and evaluation
lower limit line
Upper limit line
For each section:
One frequency-level entry for each event.
One summary entry for eac h section (see below).
One entry per pair of test points :
Frequency offset (±), level(-), level(+).
Test points defined by st andard.
Upper limit line
One entry per pair of test points :
Frequency offset (±), level(-), level(+).
Test points at ± 400, 600, 1200 and 1800 k Hz.
One frequency-level entry for each event.
One summary entry for eac h section (see below).
The result tables of the MOD-LIST Rx/Tx and SPU measurements are arranged in groups:
The whole measurement range is divided into frequency sections.
A summary entry is generated for each section. T his entry contains the section lim its and the m axim um
level of the worst quality class measured in this section. All "events" are then listed as frequency-level
pairs, an event being a measured value which lies above limit or at least in the tolerance range.
Each measured value is assigned one of the following quality classes:
• PASSED(best quality class)
• MARGIN
• EXCEPTION
• FAILED(worst quality class)
1057.3440.412-19E-3
Page 87
Sequence of Operations of FSE-K11 MeasurementsFSE-K11
PASSED values are measured values below the tolerance range.
MARGIN values are values below the limit value but within the (user-definable) tolerance range.
EXCEPTION values are values above the limit value but not exceeding the threshold for outliers.
FAILED values violate both the limit value and the exception threshold.
All measured values except PASSES are events in the above sense.
The section entries (and the table as well) are assigned either PASSED or FAILED as a summary
evaluation.
If the number of outliers measured in one section exceeds the value defined by the standard, the section
is assigned FAILED even if all events are assigned EXCEPTION in the worst case.
The tables may contain a maximum of 50 lines (except for Carrier Power, where the maximum is
100 lines).
In some measur ements, theref ore, it may happen that FAILED is indic ated as a summ ary evaluation in
the table header although all visible entries in the table are evaluated PASSED.
The reason for this is that the FAILED evaluation m ight be output in the 65th line only and is therefore
not visible.
1057.3440.412-20E-3
Page 88
FSE-K11Exiting FSE-K11
2.2 Exiting FSE-K11
When FSE-K11 is exited, the status las t active is stor ed so that it can be restor ed as f ar as possible on
re-opening FSE-K11. The FSE-K11 settings are largely transferred to the other operating modes to
allow FSE-K11-conformal measurements also in these modes as far as this is expedient.
FSE-K11 can be exited by:
•
Power off
•
Preset and mode selection
•
<Mode> key and mode selection
Power off
FSE-K11 is restarted (warm s tar t) on power-up of FSE/FSIQ. FSE-K11 is in this case not exited but only
interrupted.
Preset
The instrument is reset to a defined default status. This function is comparable to a cold start. For
FSE-K11 this means: no re-activation of the status las t ac tive when FSE-K11 is r e- opened but activation
of the default status. The target status of the instrument after a pr eset is c learly defined independently of
the previous FSE-K11 status (see paragraph "Preset" below).
Changing the mode
The target status assum ed results from the FSE-K11 st atus last active and the s elected target mode.
The parameters for the target mode are set in s uch a way that measurements in line with FSE-K11 are
possible. For changing from FSE-K11 to the analyzer and vector analyzer modes, the following is
defined:
Transformation of (FSE-K11) parameters:
Trigger
•
TrgSource compatible (if not available in target mode: FreeRun).
•
TrgLevel is taken over for all modes (if available).
•
TrgDelay corresponding to trigger last active.
•
TrgPolarity, TrgEdge/Level are taken over.
Level
•
ExtAtt is represented by RefLvl offset.
•
RefLevel is set directly.
• RF attenuation is calculated in line with FSE-K11 (AUTO or by formula).
Caution:For RX measurements, the input attenuation is set to AUTO ===> not to 0 dB !
FrequencyFor the zero span or the vector mode, the ARFCN frequency is set as the center
frequency.
For the frequency sweep mode, the start and the stop frequency are selected in
accordance with the band last selected:
for TX/RX the band in question (possibly incl. ±2 MHz),
for TX ±1.8 MHz the center frequency f (AFRCN) with a span of 3.6 MHz,
for SPU the section last selected (or more sim ply, the complete band of the non-T X
measurement) .
1057.3440.412-21E-3
Page 89
Exiting FSE-K11FSE-K11
ScreenFull screen is taken for the analyzer target mode, split screen for the vector target mode (PFE).
TraceTrace assignment in FSE-K11:
2for single-trace measurements in LIST mode (trace 1 used for table)
1for single-trace measurements in SINGLE/CONT mode
1,3,2for PVT measurements with MAX Hold, Average, MIN Hold
2for SPU measurements
1,3,4for PFE measurements with MAX Hold, ClearWrite, MIN Hold
2for CPW/CPI measurements (Average)
Trace assignment is maintained as far as possible.
For changing from FSE-K11 to modes other than the analyzer or vector analyzer modes, the following is
defined:
On exiting FSE-K11, the FSE-K11 operating status last active is stored in the newly selected mode
(eg RECEIVER mode, or another mode) and activated again by this mode on re-opening FSE-K11.
1057.3440.412-22E-3
Page 90
FSE-K11Preset
2.3 Preset
FSE-K11 is reset to the default status. Then the following setting is assumed:
Analyzer default setting: Frequ-Sweep, FullSpan, Sweep Time Auto, RBW/VBW Auto, Trigger FreeRun
2.3.1 Analyzer Mode
The target status assumed results from the FSE-K11 status last active (or the last measurement
performed with FSE-K11, respectively.
PFESame as PVT.
CPW/CPIZeroSpan; f = f (ARFCN); sweep time 540 µs.
Trace 2 is renamed trace 1 because of full-screen display.
PVTZeroSpan; f = f (ARFCN); sweep time 800 µs or 100 µs.
MOD
TRAFrequSweep; span 3.6 MHz, f = f (ARFCN); sweep time in accordance with T RA formula
SPUContinuous FrequSweep; start and stop in accordance with frequency range last selected
FrequSweep; start and stop of span in accordance with TX / ARFCN ±1.8 MHz / RX;
non-RX: trigger at FreeRun, plus gated sweep with gate length 170 µs. Otherwise trigger
is taken over (setting analogous to CONTINUOUS measurement).
(settings analogous to CONTINUOUS measurement).
or with band (TX/RX/non-TX); AUTO sweep time. Trigger is taken over. Trace 2 is
renamed trace 1 because of full-screen display.
2.3.2 Vector Mode
The PFE status of FSE-K11 (vector mode) is set irrespective of the FSE-K11 status:
split screen with result table at the top and trace at the bottom.
1057.3440.412-23E-3
Page 91
General RemarksFSE-K11
2.4 General Remarks
In the case of the TRA, MOD and PVT m easurements, a pre-measurem ent is performed prior to the
main measurement.
This pre-measurement is used for defining the reference point for the limit lines.
In the case of PVT, the pre-m easurement tak es place in two s tages. In the f irst pre-m easurement with
max. 2 bursts, the actual level applied is determined. In the second pre-measurement with max.
20 bursts, the exact reference point for the limit lines is determ ined with an operating point optimally set
by means of the result of the first pre-measurement.
In the PFE, CPW and SPU m easurements, on the other hand, f ixed level values are used and thus no
pre-measurements are required.
"ExtAtt" (external attenuation) is taken into account in all measur ements in the carrier or non-T X band,
"RxGain" for all measurement in the RX band. The values have basically the same effect but are us ed
separately. RxGain can be entered any time, not only for an activated RX band measurement.
In all measurements with Min, Max and Average dis play, three traces are ac tivated only if the number of
sweeps is higher than 1. Likewise, vector triggering operates c orrectly only if the number of sweeps is
set to a value higher than 1 (this too is a feature of the basic unit). If this is not the case, an error
message is output, for example "Sync not found" if the midam ble is not detected, in whic h case there is
no synchronization to midamble.
Explanation of abbreviations used in the formulas below:
SigPwr(Expected) signal power, defined by the current settings for the power c lass and power
control level (compare section 3.3.4)
MeasPwrMeasured signal power
CorrPwrMeasured signal power, corrected by a value depending on the bandwidth
(30 kHz RBW: +8 dB)
MaxPwrMaximum (nominal) output power
F(ARFCN)Operating frequency, resulting from frequency setting (compare section 3.3.3)
Tx Su p20 dB or 0 dB is taken into account in the f orm ula depending on whether TX suppression
is set to ON or OFF
ExtAttExternal attenuation
RxGainRX band gain
SFHSFH (slow frequency hopping) factor:1 for SFH OFF, 3 for SFH ON
NOBNumber of bursts – value set by the user for the number of sweeps
Examples giving numbers, if any, refer to the default settings on power-up (PGSM-II).
1057.3440.412-24E-3
Page 92
FSE-K11Standard
2.5 Standard
StandardP-GSM
PhaseII
Measurement: SPU-Tx
Type of measurement/band assignment for initial selection of measurement:
Table 2-4Exceptions and base line values for MOD measurements
StandardRangeExceptionsBase l i ne valueApplicable to
PGSM-I400 kHz to 1800 kHz-36 dBm
X-GSM600 kHz to 6. 0 MHz3 of max. -36 dBm each-65 dBm
-59 / -64 / -69 dBm
6.0 MHz to TX band12 of max. -36 dBm each-65 dBm
-59 / -64 / -69 dBm
DCS-I800 kHz to 6.0 MHz3 of max. -36 dBm each-57 dBmPwrClass 1 to 4 (all)
6.0 MHz to TX band12 of max. -36 dBm each-57 dBmPwrClass 1 to 4 (all)
DCS-II600 kHz to 6.0 MHz3 of max. -36 dBm each-57 dBm
-57 / -62 / -67 dBm
6.0 MHz to TX band12 of max. -36 dBm each-57 dBm
-57 / -62 / -67 dBm
PCS600 kHz to 6.0 MHz3 of max. -36 dBm each-57 dBm
-57 / -62 / -67 dBm
6.0 MHz to TX band12 of max. -36 dBm each-57 dBm / -57 / -62 / -67 dBmPwrClass 1 t o 4
PwrClass 1 to 8
PwrClass M1 / M2 / M3
PwrClass 1 to 8
PwrClass M1 / M2 / M3
PwrClass 1 to 4
PwrClass M1 / M2 / M3
PwrClass 1 to 4
PwrClass M1 / M2 / M3
PwrClass 1 to 4
PwrClass M1 / M2 / M3
PwrClass M1 / M2 / M3
For details on the individual types of measurement please refer to chapter 3.
1057.3440.412-26E-3
Page 94
FSE-K11Starting the Application
3 Measurements with Option FSE-K11
Note:The descriptions of some of the functions are provided with notes pointing out that some
functions of FSE-K11 are avai lable on ly if Vector Signa l Ana lysis Opti on FSE- B7 is inst alled.
These notes apply only to FSE, since FSIQ contains this option in the basic version.
3.1 Starting the Application
CONFIGURATION
MODE
SETUP
GSM BTS
ANALYZER
The application is called with the
MODE
key and the
GSM BTS ANALYZ ER
softkey. A selection menu is offered according to the specific standards
with a softkey assigned to each measurement. The presettings are
performed in the submenu of the
SETTINGS
softkey or in the right side
menu of each measurement.
All settings can generally be performed using the softkeys.
In addition, a few hardkeys are available for performing special actions:
Hardkey list, availabilities