Rohde&Schwarz FSE-K11 User Manual

Page 1
Test and Measurement Division
Software Manual
GSM/DCS/PCS Base Station Tests
Application Firmware Module FSE-K11 Measurement according to Transmission Standards
1057.3392.02
1057.3440.42-03 1
Page 2
Supplement to FSE-K10 and FSE-K11 Manuals
K10/K11 Limit Lines
For FSE options K10 and K11 (GSM mobile station and GSM base stati on anal yzer ), many internal limit lines are used which are deri ved from the applicable s tandards. To make these limit lines access ible to
the user, they are available as data sets. There are 6 data sets:
K10_XGSM.LIA Limit lines for mobiles, standards EGSM, PGSM and RGSM K10_DCS.LIA Limit lines for mobiles, standards DCS1800 (GSM 1800) K10_PCS.LIA Limit lines for mobiles, standards PCS1900 (GSM 1900)
K11_XGSM.LIA Limit lines for base stations, standards EGSM, PGSM und RGSM K11_DCS.LIA Limit lines for base stations, standards DCS1800 (GSM 1800) K11_PCS.LIA Limit lines for base stations, standards PCS1900 (GSM 1900)
To load the data sets, proceed as follows:
1. Copy the desired data sets (one or several) from the CD to disk or analyzer.
2. Press the hardkey RECALL, and enter path (for example A:\) via softkey EDIT PATH (see operating manual, section 2.7.3).
3. Select menu item “Lines“ via softkey ITEMS TO RECALL
(this is the default setting; see operating manual, section 2.7.3.1).
4. Select the desired data set via softkey DATA SET LIST. This will cause all limit lines of the selected data sets to be copied to the limit line directory of the instrument.
From the GSM software, the limit lines can be accessed and edited via the softkey EDIT in the righthand side menus of the following measurements:
Power vs time
Modulation spectrum
Transient spectrum or
Spurious
Outside GSM measurements, the limit lines are accessible in the analyzer mode of the instrument.
1057.3140.41 A E-1
Page 3
FSE-K11 Contents

Contents

Index
Page
1 Getting Started.....................................................................................................1.1
1.1 Connecting the FSE/FSIQ..................................................................................................... 1.2
1.1.1 Standard Test Setup...................................................................................................1.2
1.1.2 Test Setup for Measurements in the RX Band
(Spurious  and Modulation Spectrum).....................................................................1.4
1.1.3 Test Setup for Measuring Spurious in the TX and outside the TX/RX Band
(<>TX Band): .............................................................................................................. 1.4
1.2 Presetting the FSE/FSIQ ....................................................................................................... 1.6
1.3 Measuring the Phase/Frequency Error .........................................................................1.22
1.4 Measuring the Average Carrier Power .......................................................................... 1.26
1.5 Measuring the Carrier Power versus Time  ô...............................................................1.32
1.6 Measuring the Spectrum due to Modulation  ................................................................ 1.38
1.7 Measurement of Spectrum due to Transients  .............................................................. 1.42
1.8 Measurement of Spurious  .............................................................................................. 1.44
2 General Information on Option FSE-K11...........................................................2-1
2.1 Sequence of Operations of FSE-K11 Measurements.........................................................2-2
2.1.1 Errors and Failures During Measurements...............................................................2-18
2.1.2 Abort of Measurements ............................................................................................2-18
a) Abort by User.....................................................................................2-18
b) Abort by Instrument ...........................................................................2-19
2.1.3 Results of Measurements.........................................................................................2-19
2.2 Exiting FSE-K11...................................................................................................................2-21
2.3 Preset ..................................................................................................................................2-23
2.3.1 Analyzer Mode..........................................................................................................2-23
2.3.2 Vector Mode..............................................................................................................2-23
2.4 General Remarks.................................................................................................................2-24
2.5 Standard...............................................................................................................................2-25
1057.3440.42 3 E-3
Page 4
Contents FSE-K11
Page
3Measurements with Option FSE-K11.................................................................3.1
3.1 Starting the Application........................................................................................................3.1
3.2 Menu Overview......................................................................................................................3.2
3.3 Selection of Default Settings................................................................................................3.9
3.3.1 Selection of the Standard............................................................................................3.9
3.3.2 Consideration of the External Attenuation................................................................3.10
3.3.3 Setting the Transmission Channel............................................................................3.11
3.3.4 Setting the Transmit Power of the BTS under Test..................................................3.14
3.3.5 Setting the Limit Margin (limit tolerance range)........................................................3.20
3.3.6 Selection of the Midamble........................................................................................3.21
3.3.7 Selection of the Timeslot..........................................................................................3.22
3.3.8 Setting of the Trigger................................................................................................3.22
3.3.9 Default Settings.........................................................................................................3.27
3.3.10Selection and Edition of Limit Lines..........................................................................3.29
3.3.11Limit Values and Limit Lines.....................................................................................3.32
3.3.11.1 Types of Limit Values..............................................................................3.32
3.3.11.2 Calculation of Relative Limit Lines..........................................................3.33
3.3.11.3 Special Features with Relative Limit Lines.............................................3.33
3.3.11.4 Forms of Output......................................................................................3.34
3.3.11.5 Table Contents of the Different Measurements......................................3.35
a)General Structure..............................................................................3.35
b)Table Head........................................................................................3.35
c)Table Contents CPW........................................................................3.36
d)Table Contents CPI..........................................................................3.36
e)Table Contents TRA.........................................................................3.36
f)Table Contents MOD........................................................................3.37
g)Table Contents SPU.........................................................................3.39
3.4 Measurement of Phase and Frequency Error...................................................................3.40
3.5 Measurement of Carrier Power..........................................................................................3.44
3.6 Measurement of Carrier Power Versus Time (Burst Timing)..........................................3.50
3.7 Measurement of Spectrum due to Modulation.................................................................3.57
3.7.1 Result Table..............................................................................................................3.60
3.7.2 Test Setup................................................................................................................3.60
3.7.3 Additional Information...............................................................................................3.61
3.7.3.1 RX Band Measurements.........................................................................3.62
3.7.3.2 Pre-Measurement...................................................................................3.63
1057.3440.42 4 E-3
Page 5
FSE-K11 Contents
Page
3.7.3.3 Main Measurement.................................................................................3.63
a)ARFCN ±1.8 MHz Measurement – Range Close to Carrier..............3.63
b)TX Band Measurement –
TX Band Excluding Range Close to Carrier......................................3.64
c)TX Band Measurement – Complete TX Band...................................3.66
3.7.3.4 Limit Lines – Selection, Base Lines, Values...........................................3.66
3.7.3.5 Exceptions..............................................................................................3.66
3.8 Measurement of Spectrum due to Transients..................................................................3.67
3.8.1 Additional Information...............................................................................................3.70
3.9 Measurement of Spurious Emissions...............................................................................3.72
3.9.1 Additional Information...............................................................................................3.77
4Remote Control....................................................................................................4.1
4.1 Description of Commands....................................................................................................4.1
4.1.1 CALCulate Subsystem................................................................................................4.1
4.1.1.1 CALCulate:LIMit Subsystem.....................................................................4.1
4.1.2 CONFigure Subsystem...............................................................................................4.5
4.1.2.1 CONFigure:BTS Subsystem.....................................................................4.5
4.1.2.2 CONFigure:BURSt Subsystem...............................................................4.12
4.1.2.3 CONFigure:MS Subsystem.....................................................................4.15
4.1.2.4 CONFigure:SPECtrum Subsystem.........................................................4.20
4.1.2.5 CONFigure:SPURious Subsystem.........................................................4.22
4.1.3 FETCh Subsystem....................................................................................................4.24
4.1.3.1 FETCh:BURSt Subsystem......................................................................4.24
4.1.3.2 FETCh:SPECtrum Subsystem................................................................4.28
4.1.3.3 FETCh:SPURious Subsystem................................................................4.31
4.1.3.4 FETCh:PTEMplate Subsystem...............................................................4.33
4.1.4 INSTrument Subsystem............................................................................................4.34
4.1.5 READ Subsystem.....................................................................................................4.35
4.1.5.1 READ:BURSt Subsystem.......................................................................4.35
4.1.5.2 READ:SPECtrum Subsystem.................................................................4.43
4.1.5.3 READ:SPURious Subsystem..................................................................4.45
4.1.6 SENSe Subsystem...................................................................................................4.47
4.1.6.1 SENSe:BANDwidth Subsystem..............................................................4.47
4.1.6.2 SENSe:CORRection-Subsystem............................................................4.48
4.1.7 TRIGger Subsystem.................................................................................................4.49
4.2 Table of Softkeys with IEC/IEEE-Bus Command Assignment........................................4.51
4.2.1 GSM BTS Analysis Mode (Option FSE-K11)............................................................4.51
4.2.1.1 CONFIGURATION Key Group...............................................................4.51
4.2.2 GSM MS Analysis Mode (Option FSE-K10).............................................................4.59
4.2.2.1 CONFIGURATION Key Group...............................................................4.59
1057.3440.42 5 E-3
Page 6
FSE-K11 Index

Index

Note: All softkeys are listed alphabetically under keyword "Softkey" with their names.
A
ARFCN (Absolute Radio Frequency Channel Number).3.12
Attenuation (external) ...................................................3.10
Attenuation factor ......................................................... 3.10
Average carrier power ..................................................3.44
B
Burst
Display modes........................................................ 3.51
Number..................................................................3.46
Burst timing..................................................................3.50
C
Carrier power................................................................3.44
Carrier power versus time.............................................3.50
Coarse adjustment of the trigger..........................3.24,
Connect bandpass or bandstop to s uppress TX-band! .3.58
3.26
O
Output of measured values in tabular form................... 3.66
Output power of BTS.................................................... 3.44
Overview measurement in frequency domain............... 3.66
P
Peak power.................................................................. 3.16
Phase.............................................................................3.9
Phase error..................................................................3.40
Power class ................................................................. 3.16
Power Control Level (static).........................................3.18
Power(average)............................................................3.44
Powers of power control levels.....................................3.44
R
RMS phase error.......................................................... 3.40
RX band....................................................................... 3.74
E
Edge
TDMA burst............................................................3.51
Edge of a burst.............................................................3.51
Error message
Connect bandpass or bandstop to s uppress
TX-band! ................................................................3.58
External trigger signal...................................................3.22
F
Fine adjustment of the trigger.......................................3.25
Frequency error............................................................3.40
H
Hopping mode..............................................................3.75
N
Number of sweeps...............................................1.44,
3.72
S
SET TO STANDARD.........................1.26, 1.32, 1.38, 1.42
Slot number..................................................................3.22
Slow frequency hopping mode..................................... 3.75
Soft key
ANALOG DEMODB7.............................................4.34
ANALYZER............................................................ 4.34
DIGITAL DEMODB7..............................................4.34
EMI TEST RECEIVERESIFSET............................4.34
RES BW MANUAL.................................................4.47
TRACKING GENERATOR..................................... 4.34
TV DEMOD............................................................ 4.34
VECTOR ANALYZER............................................ 4.34
Softkey
±
1.8 MHz /TX BAND.............................................. 3.57
<> TX BAND..........................................................
ARFCN..................................................................
ARFCN ±1.8 MHz.................................................. 3.57
ARFCN / FREQUENCY................................
ARFCN AUTOSELECT..........................................
BTS SFH ON/OFF.................................................
BURST HIGH RESOLUTION.................................
CARRIER POWER................................................
CLEAR SGL RESULT TAB....................................
CONDITIONS NORM/EXTR..................................3.46
CONT SGL STEP LIST..........................................
CONTINUOUS......................................
CONTINUOUS FREQ SWEEP..............................
COSITING GSM/DCS............................................
DCS 1800................................................................
DEFAULT SETTINGS............................................
DYNAM PWR CTRL LEVEL.........................
3.41
3.12, 3.52
, 3.50,
3.19, 3.47
3.73
3.13
3.13
3.75
3.51
3.44
3.47
3.71
3.56
3.67
3.75
3.9
3.27
1057.3440.41 I.1 E-3
Page 7
Index FSE-K11
EDIT.......................................................................3.29
E-GSM 900 ..............................................................
EXTERN.................................................................
EXTERNAL ATTEN................................................
FALLING EDGE.....................................................
FRAME COARSE...................................................
FRAME FINE .........................................................
FREE RUN.............................................................
3.9
3.24
3.10
3.51
3.24
3.25
3.24
FREQUENCY.........................................................3.13
FREQUENCY.........................................................3.30
FULL BURST.........................................................
GSM BTS ANALYZER.............................................
INC DYNAM PWR CTRL........................................
INC STATIC PWR CTRL........................................
3.51
3.1
3.46
3.45
LIMIT /PWR COUPLED..........................................3.19
LIMIT REF POWER...............................................3.20
MEAS BANDWIDTH .....................................3.52,
3.68
MEAS MAX OUTPUT PWR...................................3.45
MEAS SGL PWR LEVEL........................................
MIDAMBLE ............................................................
MODULATION SPECTRUM...................................
NO. OF BURSTS.................................3.51,
NO. OF BURSTS..........................................
3.47
3.21
3.56
3.57, 3.67
3.41
, 3.46
NOMINAL OUTPUT PWR......................................3.16
PCS 1900.................................................................
3.9
P-GSM 900, .............................................................3.9
PHASE I/II/II+...........................................................
PHASE PEAK ........................................................
PHASE RMS..........................................................
PHASE/FREQ ERROR..........................................
POWER CLASS.....................................................
POWER SETTING.................................................
POWER VS TIME..................................................
REF MEAS AUTO/USER.......................................
3.9
3.30
3.30
3.40
3.16
3.15
3.50
3.52
R-GSM 900,.............................................................3.9
RISING EDGE........................................................
RX BAND......................................................
RX BAND GAIN......................................................
SETTINGS...............................................................
SGL MEAS ON/OFF ..............................................
SIGNAL POWER...........................................
SINGLE..................................................................
3.51
3.58, 3.74
3.75
3.9
3.46
3.20, 3.47
3.41
SINGLE..................................................................3.50
SINGLE FREQ SWEEP................................
SLOPE POS/NEG ..................................................
SLOT ADJUST.......................................................
SLOT NO...............................................................
SPURIOUS ............................................................
START LIST.........................................
START LIST SGL STEP.........................................
START REF MEAS................................................
STATIC PWR CRTL LEVEL...................................
STATIC PWR CTRL LEVEL...................................
SWEEP COUNT............................................
SWEEPTIME..........................................................
SWEEPTIME STD/AUTO.......................................
3.56, 3.66
3.24
3.26
3.22
3.71
3.57, 3.68, 3.72
3.71
3.52
3.18
3.47
1.44, 3.72
1.46
3.75
SWP COUNT RX BAND.........................................3.72
SWP COUNT TX /<>TX.........................................3.72
3.48
3.57
, 3.53
3.66
3.23
3.24
3.26
, 3.72
3.58
SYNC TO MIDAMBLE...................................
TRANSIENT SPECTRUM......................................
TRIGGER...............................................................
TRIGGER ADJUST................................................
TRIGGER LEVEL...................................................
TX BAND.......................................................
TX BAND
±
2 MHz...................................................3.75
TX GATE ON/OFF .................................................
TX SUPPR ON/OFF...............................................3.75
X UNIT SYMB/TIME...............................................
3.52
Softkeys
SLOT NO MANUAL...............................................
3.22
Spectrum due to modulation.........................................3.56
Spectrum due to transients .......................................... 3.66
Spurious emissions......................................................3.71
Static Power Control Level........................................... 3.18
T
TDMA burst......................................................... 3.22,
Test parameters.............................................................3.9
Test setup
Spectrum due to modulation.................................. 3.59
Spurious emissions................................................ 3.74
TX BAND...............................................................3.72
Timeslot....................................................................... 3.22
Training sequence........................................................ 3.21
Transmission channel.................................................. 3.13
Transmission system .....................................................3.9
Trigger signal (external)............................................... 3.22
Trigger source.............................................................. 3.23
TSC (Training Sequence Code)................................... 3.21
TX band....................................................................... 3.75
3.51
1057.3440.41 I.2 E-3
Page 8
Page 9
Before putting the product into operation for
the first time, make sure to read the following
Safety Instructions
Rohde & Schwarz makes every effort to keep the safety standard of its products up to date and to offer its customers the highest possible degree of safety. Our products and the auxiliary equipment required for them are designed and tested in accordance with the relevant safety standards. Compliance with these standards is continuously monitored by our quality assurance system. This product has been designed and tested in accordance with the EC Certificate of Conformity and has left the manufacturer’s plant in a condition fully complying with safety standards. To maintain this condition and to ensure safe operation, observe all instructions and warnings provided in this manual. If you have any questions regarding these safety instructions, Rohde & Schwarz will be happy to answer them.
Furthermore, it is your responsibility to use the product in an appropriate manner. This product is designed for use solely in industrial and laboratory environments or in the field and must not be used in any way that may cause personal injury or property damage. You are responsible if the product is used for an intention other than its designated purpose or in disregard of the manufacturer's instructions. The manufacturer shall assume no responsibility for such use of the product.
The product is used for its designated purpose if it is used in accordance with its operating manual and within its performance limits (see data sheet, documentation, the following safety instructions). Using the products requires technical skills and knowledge of English. It is therefore essential that the products be used exclusively by skilled and specialized staff or thoroughly trained personnel with the required skills. If personal safety gear is required for using Rohde & Schwarz products, this will be indicated at the appropriate place in the product documentation.
Observe operating instructions
Supply voltage ON/OFF
Weight indication for units >18 kg
Standby indication
Symbols and safety labels
Danger of electric shock
Direct current (DC)
Warning! Hot surface
PE terminal Ground
Alternating current (AC)
Direct/alternating current (DC/AC)
Ground terminal
Device fully protected by double/reinforced insulation
Attention! Electrostatic sensitive devices
1171.0000.42-02.00 Sheet 1
Page 10
Safety Instructions
Observing the safety instructions will help prevent personal injury or damage of any kind caused by dangerous situations. Therefore, carefully read through and adhere to the following safety instructions before putting the product into operation. It is also absolutely essential to observe the additional safety instructions on personal safety that appear in other parts of the documentation. In these safety instructions, the word "product" refers to all merchandise sold and distributed by Rohde & Schwarz, including instruments, systems and all accessories.
Tags and their meaning
DANGER
WARNING
CAUTION This tag indicates a safety hazard with a low potential of risk for the user
ATTENTION
NOTE
These tags are in accordance with the standard definition for civil applications in the European Economic Area. Definitions that deviate from the standard definition may also exist. It is therefore essential to make sure that the tags described here are always used only in connection with the associated documentation and the associated product. The use of tags in connection with unassociated products or unassociated documentation can result in misinterpretations and thus contribute to personal injury or material damage.
This tag indicates a safety hazard with a high potential of risk for the user that can result in death or serious injuries.
This tag indicates a safety hazard with a medium potential of risk for the user that can result in death or serious injuries.
that can result in slight or minor injuries.
This tag indicates the possibility of incorrect use that can cause damage to the product.
This tag indicates a situation where the user should pay special attention to operating the product but which does not lead to damage.
Basic safety instructions
1. The product may be operated only under the operating conditions and in the positions specified by the manufacturer. Its ventilation must not be obstructed during operation. Unless otherwise specified, the following requirements apply to Rohde & Schwarz products: prescribed operating position is always with the housing floor facing down, IP protection 2X, pollution severity 2, overvoltage category 2, use only in enclosed spaces, max. operation altitude max. 2000 m. Unless specified otherwise in the data sheet, a tolerance of ±10% shall apply to the nominal voltage and of ±5% to the nominal frequency.
2. Applicable local or national safety regulations and rules for the prevention of accidents must be observed in all work performed. The product may be opened only by authorized, specially trained personnel. Prior to performing any work on the product or opening the product, the
product must be disconnected from the supply network. Any adjustments, replacements of parts, maintenance or repair must be carried out only by technical personnel authorized by Rohde & Schwarz. Only original parts may be used for replacing parts relevant to safety (e.g. power switches, power transformers, fuses). A safety test must always be performed after parts relevant to safety have been replaced (visual inspection, PE conductor test, insulation resistance measurement, leakage current measurement, functional test).
3. As with all industrially manufactured goods, the use of substances that induce an allergic reaction (allergens, e.g. nickel) such as aluminum cannot be generally excluded. If you develop an allergic reaction (such as a skin rash, frequent sneezing, red eyes or respiratory difficulties), consult a physician immediately to determine the cause.
1171.0000.42-02.00 Sheet 2
Page 11
Safety Instructions
4. If products/components are mechanically and/or thermically processed in a manner that goes beyond their intended use, hazardous substances (heavy-metal dust such as lead, beryllium, nickel) may be released. For this reason, the product may only be disassembled, e.g. for disposal purposes, by specially trained personnel. Improper disassembly may be hazardous to your health. National waste disposal regulations must be observed.
5. If handling the product yields hazardous substances or fuels that must be disposed of in a special way, e.g. coolants or engine oils that must be replenished regularly, the safety instructions of the manufacturer of the hazardous substances or fuels and the applicable regional waste disposal regulations must be observed. Also observe the relevant safety instructions in the product documentation.
6. Depending on the function, certain products such as RF radio equipment can produce an elevated level of electromagnetic radiation. Considering that unborn life requires increased protection, pregnant women should be protected by appropriate measures. Persons with pacemakers may also be endangered by electromagnetic radiation. The employer is required to assess workplaces where there is a special risk of exposure to radiation and, if necessary, take measures to avert the danger.
7. Operating the products requires special training and intense concentration. Make certain that persons who use the products are physically, mentally and emotionally fit enough to handle operating the products; otherwise injuries or material damage may occur. It is the responsibility of the employer to select suitable personnel for operating the products.
8. Prior to switching on the product, it must be ensured that the nominal voltage setting on the product matches the nominal voltage of the AC supply network. If a different voltage is to be set, the power fuse of the product may have to be changed accordingly.
9. In the case of products of safety class I with movable power cord and connector, operation is permitted only on sockets with earthing contact and protective earth connection.
10. Intentionally breaking the protective earth connection either in the feed line or in the product itself is not permitted. Doing so can result in the danger of an electric shock from the product. If extension cords or connector strips are implemented, they must be checked on a regular basis to ensure that they are safe to use.
11. If the product has no power switch for disconnection from the AC supply, the plug of the connecting cable is regarded as the disconnecting device. In such cases, it must be ensured that the power plug is easily reachable and accessible at all times (length of connecting cable approx. 2 m). Functional or electronic switches are not suitable for providing disconnection from the AC supply. If products without power switches are integrated in racks or systems, a disconnecting device must be provided at the system level.
12. Never use the product if the power cable is damaged. By taking appropriate safety measures and carefully laying the power cable, ensure that the cable cannot be damaged and that no one can be hurt by e.g. tripping over the cable or suffering an electric shock.
13. The product may be operated only from TN/TT supply networks fused with max. 16 A.
14. Do not insert the plug into sockets that are dusty or dirty. Insert the plug firmly and all the way into the socket. Otherwise this can result in sparks, fire and/or injuries.
15. Do not overload any sockets, extension cords or connector strips; doing so can cause fire or electric shocks.
16. For measurements in circuits with voltages V
> 30 V, suitable measures (e.g.
rms
appropriate measuring equipment, fusing, current limiting, electrical separation, insulation) should be taken to avoid any hazards.
17. Ensure that the connections with information technology equipment comply with IEC 950/EN 60950.
18. Never remove the cover or part of the housing while you are operating the product. This will expose circuits and components and can lead to injuries, fire or damage to the product.
1171.0000.42-02.00 Sheet 3
Page 12
Safety Instructions
19. If a product is to be permanently installed, the connection between the PE terminal on site and the product's PE conductor must be made first before any other connection is made. The product may be installed and connected only by a skilled electrician.
20. For permanently installed equipment without built-in fuses, circuit breakers or similar protective devices, the supply circuit must be fused in such a way that suitable protection is provided for users and products.
21. Do not insert any objects into the openings in the housing that are not designed for this purpose. Never pour any liquids onto or into the housing. This can cause short circuits inside the product and/or electric shocks, fire or injuries.
22. Use suitable overvoltage protection to ensure that no overvoltage (such as that caused by a thunderstorm) can reach the product. Otherwise the operating personnel will be endangered by electric shocks.
23. Rohde & Schwarz products are not protected against penetration of water, unless otherwise specified (see also safety instruction 1.). If this is not taken into account, there exists the danger of electric shock or damage to the product, which can also lead to personal injury.
24. Never use the product under conditions in which condensation has formed or can form in or on the product, e.g. if the product was moved from a cold to a warm environment.
matching Rohde & Schwarz type (see spare parts list). Batteries and storage batteries are hazardous waste. Dispose of them only in specially marked containers. Observe local regulations regarding waste disposal. Do not short-circuit batteries or storage batteries.
28. Please be aware that in the event of a fire, toxic substances (gases, liquids etc.) that may be hazardous to your health may escape from the product.
29. Please be aware of the weight of the product. Be careful when moving it; otherwise you may injure your back or other parts of your body.
30. Do not place the product on surfaces, vehicles, cabinets or tables that for reasons of weight or stability are unsuitable for this purpose. Always follow the manufacturer's installation instructions when installing the product and fastening it to objects or structures (e.g. walls and shelves).
31. Handles on the products are designed exclusively for personnel to hold or carry the product. It is therefore not permissible to use handles for fastening the product to or on means of transport such as cranes, fork lifts, wagons, etc. The user is responsible for securely fastening the products to or on the means of transport and for observing the safety regulations of the manufacturer of the means of transport. Noncompliance can result in personal injury or material damage.
25. Do not close any slots or openings on the product, since they are necessary for ventilation and prevent the product from overheating. Do not place the product on soft surfaces such as sofas or rugs or inside a closed housing, unless this is well ventilated.
26. Do not place the product on heat­generating devices such as radiators or fan heaters. The temperature of the environment must not exceed the maximum temperature specified in the data sheet.
27. Batteries and storage batteries must not be exposed to high temperatures or fire. Keep batteries and storage batteries away from children. If batteries or storage batteries are improperly replaced, this can cause an explosion (warning: lithium cells). Replace the battery or storage battery only with the
1171.0000.42-02.00 Sheet 4
32. If you use the product in a vehicle, it is the sole responsibility of the driver to drive the vehicle safely. Adequately secure the product in the vehicle to prevent injuries or other damage in the event of an accident. Never use the product in a moving vehicle if doing so could distract the driver of the vehicle. The driver is always responsible for the safety of the vehicle; the manufacturer assumes no responsibility for accidents or collisions.
33. If a laser product (e.g. a CD/DVD drive) is integrated in a Rohde & Schwarz product, do not use any other settings or functions than those described in the documentation. Otherwise this may be hazardous to your health, since the laser beam can cause irreversible damage to your eyes. Never try to take such products apart, and never look into the laser beam.
Page 13
Por favor lea imprescindiblemente antes de la primera puesta en funcionamiento las siguientes informaciones de seguridad
Informaciones de seguridad
Es el principio de Rohde & Schwarz de tener a sus productos siempre al día con los estandards de seguridad y de ofrecer a sus clientes el máximo grado de seguridad. Nuestros productos y todos los equipos adicionales son siempre fabricados y examinados según las normas de seguridad vigentes. Nuestra sección de gestión de la seguridad de calidad controla constantemente que sean cumplidas estas normas. Este producto ha sido fabricado y examinado según el comprobante de conformidad adjunto según las normas de la CE y ha salido de nuestra planta en estado impecable según los estandards técnicos de seguridad. Para poder preservar este estado y garantizar un funcionamiento libre de peligros, deberá el usuario atenerse a todas las informaciones, informaciones de seguridad y notas de alerta. Rohde&Schwarz está siempre a su disposición en caso de que tengan preguntas referentes a estas informaciones de seguridad.
Además queda en la responsabilidad del usuario utilizar el producto en la forma debida. Este producto solamente fue elaborado para ser utilizado en la indústria y el laboratorio o para fines de campo y de ninguna manera deberá ser utilizado de modo que alguna persona/cosa pueda ser dañada. El uso del producto fuera de sus fines definidos o despreciando las informaciones de seguridad del fabricante queda en la responsabilidad del usuario. El fabricante no se hace en ninguna forma responsable de consecuencias a causa del maluso del producto.
Se parte del uso correcto del producto para los fines definidos si el producto es utilizado dentro de las instrucciones del correspondiente manual del uso y dentro del margen de rendimiento definido (ver hoja de datos, documentación, informaciones de seguridad que siguen). El uso de los productos hace necesarios conocimientos profundos y el conocimiento del idioma inglés. Por eso se deberá tener en cuenta de exclusivamente autorizar para el uso de los productos a personas péritas o debidamente minuciosamente instruidas con los conocimientos citados. Si fuera necesaria indumentaria de seguridad para el uso de productos de R&S, encontrará la información debida en la documentación del producto en el capítulo correspondiente.
Símbolos y definiciones de seguridad
Ver manual de instrucciones del uso
Informaciones para maquinaria con uns peso de > 18kg
Peligro de golpe de corriente
¡Advertencia! Superficie caliente
Conexión a conductor protector
Conexión a tierra
Conexión a masa conductora
¡Cuidado! Elementos de construción con peligro de carga electroestática
El aparato está protegido en su totalidad por un aislamiento de doble refuerzo
potencia EN MARCHA/PARADA
Indicación Stand-by
Corriente continua DC
Corriente alterna AC
Corriente continua/alterna DC/AC
1171.0000.42-02.00 página 1
Page 14
Informaciones de seguridad
Tener en cuenta las informaciones de seguridad sirve para tratar de evitar daños y peligros de toda clase. Es necesario de que se lean las siguientes informaciones de seguridad concienzudamente y se tengan en cuenta debidamente antes de la puesta en funcionamiento del producto. También deberán ser tenidas en cuenta las informaciones para la protección de personas que encontrarán en otro capítulo de esta documentación y que también son obligatorias de seguir. En las informaciones de seguridad actuales hemos juntado todos los objetos vendidos por Rohde&Schwarz bajo la denominación de „producto“, entre ellos también aparatos, instalaciones así como toda clase de accesorios.
Palabras de señal y su significado
PELIGRO Indica un punto de peligro con gran potencial de riesgo para el
ususario.Punto de peligro que puede llevar hasta la muerte o graves heridas.
ADVERTENCIA Indica un punto de peligro con un protencial de riesgo mediano para el
usuario. Punto de peligro que puede llevar hasta la muerte o graves heridas .
ATENCIÓN Indica un punto de peligro con un protencial de riesgo pequeño para el
usuario. Punto de peligro que puede llevar hasta heridas leves o pequeñas
CUIDADO Indica la posibilidad de utilizar mal el producto y a consecuencia
dañarlo.
INFORMACIÓN Indica una situación en la que deberían seguirse las instrucciones en el
uso del producto, pero que no consecuentemente deben de llevar a un daño del mismo.
Las palabras de señal corresponden a la definición habitual para aplicaciones civiles en el ámbito de la comunidad económica europea. Pueden existir definiciones diferentes a esta definición. Por eso se debera tener en cuenta que las palabras de señal aquí descritas sean utilizadas siempre solamente en combinación con la correspondiente documentación y solamente en combinación con el producto correspondiente. La utilización de las palabras de señal en combinación con productos o documentaciones que no les correspondan puede llevar a malinterpretaciones y tener por consecuencia daños en personas u objetos.
Informaciones de seguridad elementales
1. El producto solamente debe ser utilizado según lo indicado por el fabricante referente a la situación y posición de funcionamiento sin que se obstruya la ventilación. Si no se convino de otra manera, es para los productos R&S válido lo que sigue: como posición de funcionamiento se define principialmente la posición con el suelo de la caja para abajo , modo de protección IP 2X, grado de suciedad 2, categoría de sobrecarga eléctrica 2, utilizar solamente en estancias interiores, utilización hasta 2000 m sobre el nivel del mar. A menos que se especifique otra cosa en la hoja de datos, se aplicará una tolerancia de ±10% sobre el voltaje nominal y de ±5% sobre la frecuencia nominal.
2. En todos los trabajos deberán ser tenidas en cuenta las normas locales de seguridad de trabajo y de prevención de accidentes. El producto solamente debe de ser abierto por personal périto autorizado. Antes de efectuar trabajos en el producto o abrirlo deberá este ser desconectado de la corriente. El ajuste, el cambio de partes, la manutención y la reparación deberán ser solamente efectuadas por electricistas autorizados por R&S. Si se reponen partes con importancia para los aspectos de seguridad (por ejemplo el enchufe, los transformadores o los fusibles), solamente podrán ser sustituidos por partes originales. Despues de cada recambio de partes elementales para la seguridad deberá ser efectuado un control de
1171.0000.42-02.00 página 2
Page 15
Informaciones de seguridad
seguridad (control a primera vista, control de conductor protector, medición de resistencia de aislamiento, medición de medición de la corriente conductora, control de funcionamiento).
3. Como en todo producto de fabricación industrial no puede ser excluido en general de que se produzcan al usarlo elementos que puedan generar alergias, los llamados elementos alergénicos (por ejemplo el níquel). Si se producieran en el trato con productos R&S reacciones alérgicas, como por ejemplo urticaria, estornudos frecuentes, irritación de la conjuntiva o dificultades al respirar, se deberá consultar inmediatamente a un médico para averigurar los motivos de estas reacciones.
4. Si productos / elementos de construcción son tratados fuera del funcionamiento definido de forma mecánica o térmica, pueden generarse elementos peligrosos (polvos de sustancia de metales pesados como por ejemplo plomo, berilio, níquel). La partición elemental del producto, como por ejemplo sucede en el tratamiento de materias residuales, debe de ser efectuada solamente por personal especializado para estos tratamientos. La partición elemental efectuada inadecuadamente puede generar daños para la salud. Se deben tener en cuenta las directivas nacionales referentes al tratamiento de materias residuales.
5. En el caso de que se produjeran agentes de peligro o combustibles en la aplicación del producto que debieran de ser transferidos a un tratamiento de materias residuales, como por ejemplo agentes refrigerantes que deben ser repuestos en periodos definidos, o aceites para motores, deberan ser tenidas en cuenta las prescripciones de seguridad del fabricante de estos agentes de peligro o combustibles y las regulaciones regionales para el tratamiento de materias residuales. Cuiden también de tener en cuenta en caso dado las prescripciones de seguridad especiales en la descripción del producto.
6. Ciertos productos, como por ejemplo las instalaciones de radiación HF, pueden a causa de su función natural, emitir una radiación electromagnética aumentada. En vista a la protección de la vida en desarrollo deberían ser protegidas personas embarazadas debidamente. También las personas con un bypass pueden correr
peligro a causa de la radiación electromagnética. El empresario está comprometido a valorar y señalar areas de trabajo en las que se corra un riesgo de exposición a radiaciones aumentadas de riesgo aumentado para evitar riesgos.
7. La utilización de los productos requiere instrucciones especiales y una alta concentración en el manejo. Debe de ponerse por seguro de que las personas que manejen los productos estén a la altura de los requerimientos necesarios referente a sus aptitudes físicas, psíquicas y emocionales, ya que de otra manera no se pueden excluir lesiones o daños de objetos. El empresario lleva la responsabilidad de seleccionar el personal usuario apto para el manejo de los productos.
8. Antes de la puesta en marcha del producto se deberá tener por seguro de que la tensión preseleccionada en el producto equivalga a la del la red de distribución. Si es necesario cambiar la preselección de la tensión también se deberán en caso dabo cambiar los fusibles correspondientes del prodcuto.
9. Productos de la clase de seguridad I con alimentación móvil y enchufe individual de producto solamente deberán ser conectados para el funcionamiento a tomas de corriente de contacto de seguridad y con conductor protector conectado.
10. Queda prohibida toda clase de interrupción intencionada del conductor protector, tanto en la toma de corriente como en el mismo producto ya que puede tener como consecuencia el peligro de golpe de corriente por el producto. Si se utilizaran cables o enchufes de extensión se deberá poner al seguro, que es controlado su estado técnico de seguridad.
11. Si el producto no está equipado con un interruptor para desconectarlo de la red, se deberá considerar el enchufe del cable de distribución como interruptor. En estos casos deberá asegurar de que el enchufe sea de fácil acceso y nabejo (medida del cable de distribución aproximadamente 2 m). Los interruptores de función o electrónicos no son aptos para el corte de la red eléctrica. Si los productos sin interruptor están integrados en construciones o instalaciones, se deberá instalar el interruptor al nivel de la instalación.
1171.0000.42-02.00 página 3
Page 16
Informaciones de seguridad
12. No utilice nunca el producto si está dañado el cable eléctrico. Asegure a través de las medidas de protección y de instalación adecuadas de que el cable de eléctrico no pueda ser dañado o de que nadie pueda ser dañado por él, por ejemplo al tropezar o por un golpe de corriente.
13. Solamente está permitido el funcionamiento en redes de distribución TN/TT aseguradas con fusibles de como máximo 16 A.
14. Nunca conecte el enchufe en tomas de corriente sucias o llenas de polvo. Introduzca el enchufe por completo y fuertemente en la toma de corriente. Si no tiene en consideración estas indicaciones se arriesga a que se originen chispas, fuego y/o heridas.
15. No sobrecargue las tomas de corriente, los cables de extensión o los enchufes de extensión ya que esto pudiera causar fuego o golpes de corriente.
16. En las mediciones en circuitos de corriente con una tensión de entrada de Ueff > 30 V se deberá tomar las precauciones debidas para impedir cualquier peligro (por ejemplo medios de medición adecuados, seguros, limitación de tensión, corte protector, aislamiento etc.).
17. En caso de conexión con aparatos de la técnica informática se deberá tener en cuenta que estos cumplan los requisitos de la EC950/EN60950.
18. Nunca abra la tapa o parte de ella si el producto está en funcionamiento. Esto pone a descubierto los cables y componentes eléctricos y puede causar heridas, fuego o daños en el producto.
19. Si un producto es instalado fijamente en un lugar, se deberá primero conectar el conductor protector fijo con el conductor protector del aparato antes de hacer cualquier otra conexión. La instalación y la conexión deberán ser efecutadas por un electricista especializado.
20. En caso de que los productos que son instalados fijamente en un lugar sean sin protector implementado, autointerruptor o similares objetos de protección, deberá la toma de corriente estar protegida de manera que los productos o los usuarios estén suficientemente protegidos.
21. Por favor, no introduzca ningún objeto que no esté destinado a ello en los orificios de la caja del aparato. No vierta nunca ninguna clase de líquidos sobre o en la caja. Esto puede producir corto circuitos en el producto y/o puede causar golpes de corriente, fuego o heridas.
22. Asegúrese con la protección adecuada de que no pueda originarse en el producto una sobrecarga por ejemplo a causa de una tormenta. Si no se verá el personal que lo utilice expuesto al peligro de un golpe de corriente.
23. Los productos R&S no están protegidos contra el agua si no es que exista otra indicación, ver también punto 1. Si no se tiene en cuenta esto se arriesga el peligro de golpe de corriente o de daños en el producto lo cual también puede llevar al peligro de personas.
24. No utilice el producto bajo condiciones en las que pueda producirse y se hayan producido líquidos de condensación en o dentro del producto como por ejemplo cuando se desplaza el producto de un lugar frío a un lugar caliente.
25. Por favor no cierre ninguna ranura u orificio del producto, ya que estas son necesarias para la ventilación e impiden que el producto se caliente demasiado. No pongan el producto encima de materiales blandos como por ejemplo sofás o alfombras o dentro de una caja cerrada, si esta no está suficientemente ventilada.
26. No ponga el producto sobre aparatos que produzcan calor, como por ejemplo radiadores o calentadores. La temperatura ambiental no debe superar la temperatura máxima especificada en la hoja de datos.
1171.0000.42-02.00 página 4
Page 17
Informaciones de seguridad
27. Baterías y acumuladores no deben de ser expuestos a temperaturas altas o al fuego. Guardar baterías y acumuladores fuera del alcance de los niños. Si las baterías o los acumuladores no son cambiados con la debida atención existirá peligro de explosión (atención celulas de Litio). Cambiar las baterías o los acumuladores solamente por los del tipo R&S correspondiente (ver lista de piezas de recambio). Baterías y acumuladores son deshechos problemáticos. Por favor tirenlos en los recipientes especiales para este fín. Por favor tengan en cuenta las prescripciones nacionales de cada país referente al tratamiento de deshechos. Nunca sometan las baterías o acumuladores a un corto circuito.
28. Tengan en consideración de que en caso de un incendio pueden escaparse gases tóxicos del producto, que pueden causar daños a la salud.
29. Por favor tengan en cuenta que en caso de un incendio pueden desprenderse del producto agentes venenosos (gases, líquidos etc.) que pueden generar daños a la salud.
30. No sitúe el producto encima de superficies, vehículos, estantes o mesas, que por sus características de peso o de estabilidad no sean aptas para él. Siga siempre las instrucciones de instalación del fabricante cuando instale y asegure el producto en objetos o estructuras (por ejemplo paredes y estantes).
31. Las asas instaladas en los productos sirven solamente de ayuda para el manejo que solamente está previsto para personas. Por eso no está permitido utilizar las asas para la sujecion en o sobre medios de transporte como por ejemplo grúas, carretillas elevadoras de horquilla, carros etc. El usuario es responsable de que los productos sean sujetados de forma segura a los medios de transporte y de que las prescripciones de seguridad del fabricante de los medios de transporte sean tenidas en cuenta. En caso de que no se tengan en cuenta pueden causarse daños en personas y objetos.
32. Si llega a utilizar el producto dentro de un vehículo, queda en la responsabilidad absoluta del conductor que conducir el vehículo de manera segura. Asegure el producto dentro del vehículo debidamente para evitar en caso de un accidente las lesiones u otra clase de daños. No utilice nunca el producto dentro de un vehículo en movimiento si esto pudiera distraer al conductor. Siempre queda en la responsabilidad absoluta del conductor la seguridad del vehículo y el fabricante no asumirá ninguna clase de responsabilidad por accidentes o colisiones.
33. Dado el caso de que esté integrado un producto de laser en un producto R&S (por ejemplo CD/DVD-ROM) no utilice otras instalaciones o funciones que las descritas en la documentación. De otra manera pondrá en peligro su salud, ya que el rayo laser puede dañar irreversiblemente sus ojos. Nunca trate de descomponer estos productos. Nunca mire dentro del rayo laser.
1171.0000.42-02.00 página 5
Page 18
Page 19
FSE-K11 Presetting the FSE/FSIQ
1 Getting Started
This chapter is mainly intended for operators using the GSM BTS analyzer mode for the first time. Operators are guided step by step through the most frequent base station tests.
The chapter describes FSE/FSIQ setup, relevant settings and individual measurements. Layout and content are application-oriented and provide important information on FSE/FSIQ as well as
the GSM/DCS1800/PCS1900 system. It is recommended to read through this chapter and perform the operations on the FSE/FSIQ at the
same time. T his ensures a simple and rapid introduction to the GSM BTS firmware func tions so that correct and useful measurement results are achieved.
Some of the steps are provided with reference num bers, eg , relating to a m ore detailed information on the opposite page.
For more detailed information on the individual menus and softk eys kindly refer to the reference part in chapter 2.
Note: The descriptions of some of the functions are provided with notes pointing out that some
functions of FSE-K11 are available only if Vector Signal Analysis Option FSE-B7 is installed. These notes apply only to FSE, since FSIQ c ontains this option in the basic version.
1057.3992.02 1.1 E-2
Page 20

FSE-K11 Connecting the FSE/FSIQ

1.1 Connecting the FSE/FSIQ

1.1.1 Standard Test Setup

Ext. reference signal
Ext. trigger
TX
RX
Trigger
Power attenuator
((ext. Referenzsignal = Ext. reference signal; Trigger Ex tern = Ext. trigger; Leistungsdämpfungs glied =
Power attenuator)) Fig. 1-1 BTS with separate RX input and TX output (connections for measurem ents in the RX band
shown as dotted lines)
Ext. reference signal
Ext.
TX
trigger
/ RX
Power attenuator
Trigger
Fig. 1-2 Test setup for BTS with combined RX/TX input/output
Caution: Another test setup is required for measurements in the RX band (see sections 1.1.2 and
1.1.3)
À Connect the frame trigger output of the BTS to the rear-panel BNC connector:
EXT TRIG GATE
(external trigger or gate input) of the FSE/FSIQ. W ith the aid of this external trigger signal FSE/FSIQ is able to trigger exactly to any slot even if several slots of the frame are occupied.
1057.3392.02 1.2 E-2
Page 21
FSE-K11 Presetting the FSE/FSIQ
À Connect the antenna output (or TX output) of the BT S via a suitable power attenuator to the RF input of
FSE/FSIQ (except for measurements in the RX band; special test setup).
1057.3992.02 1.3 E-2
Page 22
FSE-K11 Connecting the FSE/FSIQ
1.1.2 Test Setup for Measurements in the RX Band (Spurious  and Modulation Spectrum)
RX-band test setup for BTSs with combined RX/TX input/output:
Low noise amplifier
BTS
TX
Diplexer Diplexer
RX
Antenna signal
RX
Bandstop (optional)
TX
50 Ohms Load
RX-band test setup for BTSs with separate RX input and TX output:
(optional)
RF
FSE
50 Ohms Load
BTS
RX
TX
Low noise
Amplifier (optional)
FSE
1.1.3 Test Setup for Measuring Spurious in the TX and outside the TX/RX Band (<>TX Band):
Antenna
BTS
Signal
Attenuator
(20dB)
Bandstop
(optional)
FSE
1057.3392.02 1.4 E-2
Page 23
FSE-K11 Presetting the FSE/FSIQ
Additional information
The stringent requirements for measurements of spurious in the RX band on BTSs using a combined RX input/TX output demand adequate suppression of the carrier signal(s) (max. carr ier level < -10 dBm at the FSE/FSIQ input), eg with the aid of bandstop filters and/or diplexers. Since the requirements for G SM phase 1 are more severe than those f or phase 2, an external low-noise preamplifier may be required in addition because of the higher noise floor of FSEB, FSEM and FSEK compared to FSEA.
With BTSs using a To obtain maximum sensitivity, 0 dB attenuation is selected when RX BAND is set on the
FSE/FSIQ. Softkey
values the insertion loss of the preceding diplex ers and bandstop filters), which is then taken into account in the FSE display.
The attenuation of the external attenuator entered under Settings is ignored when the RX band is measured.
RX BAND GAIN
separate
RX input, the TX channel is sufficiently suppressed at the RX input.
defines the gain of the external pream plifier (or in the case of negative
Caution: The level at the FSE input of +20 dBm with a 0 dB input attenuator must under no
circumstances be exceeded.
The following warning is output when the RX-BAND softkey is pressed on the screen.
Caution: Connect bandpass or bandstop to suppress TX-band!
Notes on amplifier:
The noise figure of the preamplifier should be as low as possible (LF < approx. 6 dB) and the gain in the range 15 to 25 dB so that the noise floor of FSE/FSIQ (approx. 15 to 17 dB with FSEA30, approx. 18 to 21 dB with FSEB/M/K30) can be neglected.
Make sure that the preamplifier is not overdriven by insufficiently suppressed TX signals (for this
reason the amplifier gain should not be too high).
1057.3992.02 1.5 E-2
Page 24
FSE-K11 Measuring the Phasen/Frequency Error

1.2 Presetting the FSE/FSIQ

Step 1
CONFIGURATION
MODE
SETUP
EXT REF
FREQUENCY
REF
INT EXT
CONFIGURATION
MODE
SETUP
GSM BTS
ANALYZER
SETTINGS
Setting reference frequency
À Press SETUP key.
À Press softkey
EXT. REF FREQUENCY
and enter the frequency depending on the available frequency standard.
À Switch softkey
REF INT/EXT
to EXT
(external reference frequency).
Step 2
Selecting the standard
À Press the
BTS ANALYZER
after the other and then The menu for selecting the various standards is displayed.
MODE
key and softkeys and
SETTINGS
.
one
GSM
BTS TEST SETTINGS
P-GSM 900
E-GSM 900
GSM 1800
(DCS 1800)
GSM 1900
(PCS 1900)
R-GSM 900
PHASE I
PHASE II
PHASE II+
1057.3392.02 1.6 E-2
Page 25
FSE-K11 Presetting the FSE/FSIQ
9
Additional information
Step 1
To attain the required frequency accuracy in the frequency error measurement
ERROR)
2,...16 has to be applied to the FSE/FSIQ.
a highly accurate (deviation < 1*10
-
) external reference frequency n x 1 MHz with n = 1,
(PHASE/FREQ
1057.3992.02 1.7 E-2
Page 26
FSE-K11 Measuring the Phasen/Frequency Error
Step 3
P-GSM 900
E-GSM 900
R-GSM 900
DCS 1800 GSM 1800
PCS 1900 GSM 1900
with GSM900 and DCS1800:
PHASE I
PHASE II
PHASE II+
À Select standar d P-GSM900, E-G SM900,
DCS 1800, PCS 1900 or R-GSM, depending on the BTS to be tested.
À Select between Phase I, II or II+. With
some standards, the softkeys are not available (PCS 1900) or partly available because no phases or not all of them are defined.
Press
to return to the SETTINGS
menu.
Step 4
Entering the external attenuation
EXTERNAL
ATTEN
À Press the
EXTERNAL ATTEN
softkey.
À Enter the external attenuation. (For
recommended attenuation see ô)
1057.3392.02 1.8 E-2
Page 27
FSE-K11 Presetting the FSE/FSIQ
Additional information
Step 4
ô
The following values are recomm ended for the external attenuator to ensure that the FSE/FSIQ RF input is protected and the sensitivity of the FSE/FSIQ is not reduced too much. The FSE/FSIQ level display is automatically corrected by the entered value.
Max. power Recommended ext. attenuation
55 to 60 dBm 50 to 55 dBm 45 to 50 dBm 40 to 45 dBm 35 to 40 dBm 30 to 35 dBm 25 to 30 dBm 20 to 25 dBm
< 20 dBm 0 dB
35 to 40 dB 30 to 35 dB 25 to 30 dB 20 to 25 dB 15 to 20 dB 10 to 15 dB 5 to 10 dB 0 to 5 dB
1057.3992.02 1.9 E-2
Page 28
FSE-K11 Measuring the Phasen/Frequency Error
Step 5
Selecting the transmission channel
ARFCN /
FREQUENCY
À Press the
softkey.
ARFCN / FREQUENCY
ARFCN
ARFCN/
AUTOSELECT
FREQUENCY
À Press softkey
AUTOSELECT
ARFCN
With
ARFCN, ARFCN
or
FREQUENCY
selected:
.
À Enter the transmission channel number
of your base station. í FSE/FSIQ will tune to this frequency in compliance with the selected standard.
With
ARFCN AUTOSELECT
, FSE/FSIQ automatically searches for the active transmission channel on condition that frequency hopping is not active.
FREQUENCY
With
selected:
À Enter the transmission frequency of your
base station or any other frequency (eg the IF of a BTS module). Frequencies of the following range are permissible:
1.8MHz frequency (maximum device frequency - 1.8MHz).
À Press
to return to the SETTINGS
menu.
1057.3392.02 1.10 E-2
Page 29
FSE-K11 Presetting the FSE/FSIQ
Additional information
Step 5
í
Enter one of the possible channel numbers 1 to 124 for the P-GSM frequency range. Enter a number between 975 and 1023 or 0 for the extended GSM band. For PCN (DCS1800) the numbers 512 to 885 are available, for PCS1900 numbers 512 to 810. For R-GSM channel numbers 955 to 1023 and 0 are available in addition to the numbers of the P-GSM band (1 to 124).
Range Channel numbernFrequency downlink ( BS-->MS) Duplex
P-GSM 1 to 124 f = 890 MHz + (n * 0.2 MHz) +45 MHz E-GSM 0 to 124 f = 890 MHz + (n * 0.2 MHz) +45 MHz
DCS1800 512 t o 885 f = 1710.2 MHz + ((n - 512) * 0.2 MHz) +95 MHz PCS1900 512 to 810 f = 1850.2 MHz + ((n - 512) * 0.2 MHz) +80 MHz R-GSM 0 to 124 f = 890 MHz + (n * 0.2 MHz) +45 MHz
P-GSM 900
E-GSM 900
R-GSM 900
DCS 1800
offset
975 to 1023 f = 890 MHz + ((n - 1024) * 0.2 MHz) +45 MHz
955 to 1023 f = 890 MHz + ((n - 1024) * 0.2 MHz) +45 MHz
935.2MHz 959.8MHz 1 124
925.2MHz 934.8 935 935.2 959.8MHz 975 1023 0 1 124
921.2MHz 934.8 935 935.2 959.8MHz 955 1023 0 1 124
1805.2MHz 1879.8MHz 512 885
FREQ ARFCN
FREQ ARFCN
FREQ ARFCN
FREQ ARFCN
PCS 1900
1930.2MHz 1989.8MHz 512 810
FREQ ARFCN
1057.3992.02 1.11 E-2
Page 30
FSE-K11 Measuring the Phasen/Frequency Error
Step 6
Setting FSE/FSIQ to output power of BTS
POWER
SETTINGS
or
or
POWER CLASS
NOMINAL
OUTPUT PWR
LIMIT/PWR
COUPLED
SIGNAL
POWER
À Press softkey
POWER CLASS
OUTPUT POWER
and enter the
or
maximum output power or power class of the BTS. ÷
À As an alternative enter the nominal
power of the BTS by pressing softkey
NOMINAL OUTPUT PWR
.
In addition, it is possible to enter the power linked neither to a power class nor to a power control level. For this select
LIMIT/PWR COUPLED OFF
power using softkey
SIGNAL POWER
and enter the
.
1057.3392.02 1.12 E-2
Page 31
FSE-K11 Presetting the FSE/FSIQ
Additional information
Step 6
÷
The power classes which can be allocated to the various standards and the associated peak power values are indicated in the following tables:
Power Peak power Class -GSM 900 -
Phase 1
1 320W (55 dBm) 320W (55 dBm) - < 640W 320W (55 dBm) - < 640W 20 W (43 dBm) 2 160W (52 dBm) 160W (52 dBm) - < 320W 160W (52 dBm) - < 320W 10 W (40 dBm) 3 80 W (49 dBm) 80 W (49 dBm) - < 160W 80 W (49 dBm) - < 160W 5 W (37 dBm) 4 40 W (46 dBm) 40 W (46 dBm) - < 80W 40 W (46 dBm) - < 80W 2.5 W (34 dBm) 5 20W (43 dBm) 20W (43 dBm) - < 40W 20W (43 dBm ) - < 40W -­6 10W (40 dBm) 10W (40 dBm) - < 20W 10W (40 dBm ) - < 20W -­7 5W (37 dBm) 5W (37 dBm) - < 10W 5W (37 dBm ) - < 10W -­8 2.5W (34 dBm ) 2.5W (34 dBm) - < 5W 2.5W (34 dBm ) - < 5W -­M1 -- (>0.08) - 0.25 W (> 19) - 24 dBm (>0.08) - 0.25 W (> 19) - 24 dBm -­M2 -- (>0.03) - 0.08 W (> 14) - 19 dBm (0.03) - 0.08 W (> 14) - 19 dBm -­M3 -- (>0.01) - 0.03W (> 9) - 14 dBm (>0.01) - 0.03W (> 9) - 14 dBm --
P-GSM 900 -
Phase 2
E-GSM 900 -
Phase 2 and R-GSM
DCS1800 -
Phase 1
Peak powerPower
Class DCS1800 - PCS1900 (Watt) PCS1900 (dBm)
1 20 - (<40 W) 43 - < 46 dBm 2 10 - (<20 W) 40 - < 43 dBm 3 5 - (< 10 W) 37 - < 40 dBm 4 2.5 -.(<5 W) 34 - < 37 dBm 5--
6-­7-­8-­M1 (>0.5) - 1.6 W (>27) - 32 dB m
M2 (>0.16) - 0.5 W (>22) - 27 dBm M3 (>0.05) - 0.16 W (>17) - 22 dBm
0.5P01.6 27P032
0.16P00.5 22P027
0.05P00.16 17P022
20≤P0≤40 43P046 10≤P0≤20 40P043 5P010 37P040 2P0533≤P0≤37
To protect the FSE/FSIQ, the settings made in the PO WER menu are automatically checked to make sure that the input level at the FSE/FSIQ does not exceed 27 dBm.
To prevent the occurrence of an overload, a warning is displayed for those settings which would lead to the threshold level being exceeded (
POWER, POWER CTRL LEVEL)
and the set value is ignored.
POWER CLASS, EXTERNAL ATTEN, OUT PUT MS
1057.3992.02 1.13 E-2
Page 32
FSE-K11 Measuring the Phasen/Frequency Error
or
STATIC PWR CTRL LEVEL
DYNAM PWR
CTRL LEVEL
LIMIT/PWR
COUPLED
SIGNAL
POWER
LIMIT LINE
REF POWER
Step 7
With reduced level of the BTS enter the static and/or the dynamic power control level in addition. (The default setting is 0)
û: À Enter the desired static power control
level N (integer, N = 0, 1, 2 to 6). û.
À Enter the desired dynamic power control
level N (integer, N = 0, 1, 2 to 15). ø
The coupling between power and that of the power class as well as between the power actually applied and the limits selected with respect to this power can be suppressed (softkey
COUPLED
).
À In this case press softkey
POWER
and specify the actual output power of the BTS (external attenuation is taken into account).
With softkey
LIMIT LINE REF POWER
the power determining the selection of limits can be selected.
LIMIT/PWR
SIGNAL
SLOT NO.
SLOT NO. SLOT NO.
MANUAL
SLOT NO.
AUTOSELECT
. . .
Step 8
À Press softkey À Select the desired slot number 0 to 7
using
SLOT NO. MANUAL
automatic search with
SELECT.
The
FSE/FSIQ thus knows in which slot
the measurement should be performed.
SLOT NO.
or start the
SLOT NO. AUTO
ù
1057.3392.02 1.14 E-2
Page 33
FSE-K11 Presetting the FSE/FSIQ
Additional information
Step 7
û
Static Power Control Level Relative power referred to
00 dB 1 -2 dB 2 -4 dB 3 -6 dB 4 -8 dB 5 -10 dB 6 -12 dB
ø
Max. 16 dynamic power control levels are specif ied in addition to the static power control levels.
maximum output power (GSM/DCS/PCS)
They reduce the base station level referred to the maximum output power of the base station and the static power control level in up to 16 steps of 2 dB (dynamic power control level N = 0 to 15, attenuation = N × 2 dB).
Step 8
ù
A precondition for automatic search is that only one slot is active. (For an automatic detection of the slot the fram e trigger of the base station must be c onnected to the external trigger input of the unit and the trigger source set to
EXTERN
, see above).
1057.3992.02 1.15 E-2
Page 34
FSE-K11 Measuring the Phasen/Frequency Error
Step 9
Selecting the current midamble
MIDAMBLE
MIDAMBLE
AUTO TSC_0 TSC_1 TSC_2 TSC_3 TSC_4 TSC_5 TSC_6 TSC_7
À Press softkey
MIDAMBLE .
The table of training sequences
(
m
idamble)
is displayed. î
In the position AUTO (default s etting) FSE uses the training sequence (midamble) according to the slot number , for instanc e TSC_2 for slot number 2.
À If you wish to use another training
sequence, select it accordingly.
1057.3392.02 1.16 E-2
Page 35
FSE-K11 Presetting the FSE/FSIQ
Additional information
Step 9
î
Table for the midamble used in the GSM system: NAME PATTERN (Bit No.: 61 - 86) TSC_0 0 0 1 0 0 1 0 1 1 1 0 0 0 0 1 0 0 0 1 0 0 1 0 1 1 1
TSC_1 0 0 1 0 1 1 0 1 1 1 0 1 1 1 1 0 0 0 1 0 1 1 0 1 1 1 TSC 2 0 1 0 0 0 0 1 1 1 0 1 1 1 0 1 0 0 1 0 0 0 0 1 1 1 0 TSC 3 0 1 0 0 0 1 1 1 1 0 1 1 0 1 0 0 0 1 0 0 0 1 1 1 1 0 TSC_4 0 0 0 1 1 0 1 0 1 1 1 0 0 1 0 0 0 0 0 1 1 0 1 0 1 1 TSC_5 0 1 0 0 1 1 1 0 1 0 1 1 0 0 0 0 0 1 0 0 1 1 1 0 1 0 TSC_6 1 0 1 0 0 1 1 1 1 1 0 1 1 0 0 0 1 0 1 0 0 1 1 1 1 1 TSC_7 1 1 1 0 1 1 1 1 0 0 0 1 0 0 1 0 1 1 1 0 1 1 1 1 0 0
1057.3992.02 1.17 E-2
Page 36
FSE-K11 Measuring the Phasen/Frequency Error
Step 10
TRIGGER
TRIGGER ADJUST
TRIGGER
FREE RUN
EXTERN
SLOPE POS NEG
VIDEO
TRIGGER
ADJUST
TRIGGER
ADJUST
FRAME
COARSE
FRAME FINE
AUTO FRAME
ADJUST
Setting the trigger
À Press softkey
TRIGGER.
An external trigger signal (frame trigger of BTS) is required for almost all measurements. The frame trigger pulse is output once for each frame. In the ideal case it comes with bit 0 in slot 0 of the frame. ü But this depends on the BTS to be tested. For this reason trigger adjustment is provided (described later). The default setting for the trigger is therefore EXTERN. The default value for the external trigger level is 1.4 V but, if required, the level can be adapted by pressing
EXTERN
and then entering a new value in the data entry field. The polarity of the trigger signal can be set using softkey
SLOPE POS NEG
. Default
setting: positive slope.
À Press softkey
TRIGGER ADJUST
the trigger reference. Automatic (preferably) or manual adjustment can be performed.
to set
SLOT
ADJUST
AUTO SLOT
VIDEO
ADJUST
TRIGGER
LEVEL
SLOPE
POS NEG
1057.3392.02 1.18 E-2
Page 37
FSE-K11 Presetting the FSE/FSIQ
Additional information
Step 10
External frame tr igger
ü
1.4V
T
0 dB
Slot No. 0 Slot No. 1 Slot No. 2
Transition from bit 73 to 74
Delay between burst edge and transitionbit 13/14 dof midamble of slot N (where N=0,1,2...7): T=<283 .23 us> +N *57 6.92 us
Slot No. 3
Slot No. 4
Slot No. 5 Slot No. 6
Slot No. 7
Slot No. 0
1057.3992.02 1.19 E-2
Page 38
FSE-K11 Measuring the Phasen/Frequency Error
Either automatic adjustment:
AUTO FRAME
ADJUST
AUTO SLOT
ADJUST
FRAME
FINE
Only one slot can be active. À Press softkey
AUTO FRAME ADJUST
FSE/FSIQ sets the time reference between trigger signal and midamble ü and takes this referenc e into account for all measurements .
AUTO ADJUST
acts
upon all slots.
If only the slot used at present is to be adjusted:
À Press softkey
AUTO SLOT ADJUST
and adjust the time referenc e of the slot used at present.
or manual adjustment:
(if automatic adjustment is not possible, eg with bursts without midamble or when option FSE-B7 is not installed.)
À Press softkey
FRAME COARSE
and set the time reference of the external trigger to the frame with the aid of the spinwheel (the -20 dB point of the rising burst edge should be in the display center. This point is marked by crosshairs made up of the vertical screen center line and the -20 dB line).
.
FRAME
FINE
or
SLOT
ADJUST
À Press softkey
FRAME FINE
and adjust the time reference of all slots in the same way.
10
If the slots have a different length (156 or 157 bits), a fine adjustment can be performed for each slot with the aid of
SLOT ADJUST
.
FSE/FSIQ is now set as required by the BTS to be measured and measurements can be started.
1057.3392.02 1.20 E-2
Page 39
FSE-K11 Presetting the FSE/FSIQ
Additional information
10
Fig 1-3 Correcting adjustment for FRAME FINE
Caution: The manual adjustment to the -20 dB point of the trigger edge can yield only an
approximate time reference since the exact r eference to the midamble is missing. The automatic adjustment produces the correct time reference to the midamble ( depending on the steepness of the burst edge the -20 dB point can be up to approx . 7 from the screen center).
µ
s away
1057.3992.02 1.21 E-2
Page 40
FSE-K11 Measuring the Phasen/Frequency Error

1.3 Measuring the Phase/Frequency Error

Note: The phase and frequency error can only be measured if the Vector Signal Analyzer option
FSE-B7 is integrated.
Step 1
PHASE/FREQ
ERROR
NO. OF
BURSTS
À Press the softkey
ERROR
.
PHASE/FREQ
Step 2
À Press softkey
the number of measurements to be carried out. Default setting:
The number of bursts defined in the selected standard is set with SET TO STANDARD (in the phase/frequency error measurement: 20 for all standards).
Press menu.
NO. OF BURSTS
NO. OF BURSTS
to return to the SETTINGS
and set
= 1.
1057.3392.02 1.22 E-2
Page 41
FSE-K11 Presetting the FSE/FSIQ
Additional information
BTS setting:
At least 1 slot on the selected ARFCN must be active. All slots may be occupied. The measurement can also be performed on BTSs with active SFH (slow frequency hopping).
FSE/FSIQ measures at the set r eceive frequency (ARFCN). If a number of bursts > 1 is set, only results are considered at which FSE/FSIQ is able to synchronize. Thus even if the carrier frequency of the slot to be measured changes cyclically in successive frames, only slots with the set slot number at the selected ARFCN will be considered in the measurement. However, this will considerably slow down the measurement sequence (approx. by the factor 8 against a non-hopping BTS). This disadvantage can be avoided if a trigger signal is pr ovided for the FSE/FSIQ (triggering not on every frame but only if the selected slot is at the set ARFCN).
1057.3992.02 1.23 E-2
Page 42
FSE-K11 Measuring the Phasen/Frequency Error
Step 3
SINGLE
CONTINUOUS
or
À Start the measurement sequence with
SINGLE
When the measurement is completed, an overview of the numeric modulation errors is displayed in result window A f or the 147 useful bits. ô
or
CONTINUOUS
.
1057.3392.02 1.24 E-2
Page 43
FSE-K11 Presetting the FSE/FSIQ
Additional information
Step 3
ô
The information on instantaneous frequency and power and the following numerical results are displayed in the window. It also contains the corresponding limit values and pass/fail information.
Max. hold and average value of peak phase error
Max. hold and average value of rms phase error
Max. hold and average value of frequency error The phase error versus time is displayed in window B over the 147 useful bits of the normal burst.
Three traces are simultaneously displayed.
Trace 1: Clear Write (instantaneous phase error)
Trace 2: Max Hold (maximum positive phase error of all measured bursts)
Trace 3: Min Hold (maximum negative phase error of all measured bursts)
1057.3992.02 1.25 E-2
Page 44
FSE-K11 Measuring the Carrier Power versus Time

1.4 Measuring the Average Carrier Power

Step 1
CARRIER
POWER
NO. OF BURSTS
SYNC TO
MIDAMBLE
NO. OF BURSTS
SET
MANUAL
SET TO
STANDARD
À Press softkey
CARRIER POWER.
Step 2
À Press softkey
the desired number of measurements.
The number of bursts defined in the selected standard is set with SET TO STANDARD (in the carrier power measurement: 1 for all standards).
NO. OF BURSTS
and set
Step 3
Possible only if option FSE-B7 is installed:
In the default setting, softkey
MIDAMBLE
is active and thus the synchronization to the burst midamble is set. This setting is of advantage if the burst to be measured contains a midamble and a bit-accurate time reference is required. (softkey colour: green)
SYNC TO
À Disable
SYNC TO MIDAMBLE
for bursts without midamble or if maximum measurement s peed is required (s oftk ey colour: grey) ô.
1057.3392.02 1.26 E-2
Page 45
FSE-K11 Measuring the Carrier Power versus Time
Additional information
BTS setting:
At least three neighbouring slots of a TDMA frame of equal power must be active. If the BTS supports SFH (slow frequency hopping), SFH should be selected.
FSE/FSIQ measures at the set r eceive fr equency (ARFCN). If a num ber of bursts > 1 has been set, only measurement results will be considered at which FSE/FSIQ is able to synchronize. Thus even if the carrier frequency of the slot to be measured changes cyclically in successive frames, only slots with the set slot number at the selected ARFCN will be considered in the measurement.
However, this will considerably slow down the measurement sequence (approx. by the factor 3 compared to a BTS without SFH). This disadvantage can be avoided if a trigger signal is provided for the FSE/FSIQ (triggering not on every frame but only if the selected slot is at the set ARFCN).
Step 3
ô
Explanation concerning SYNC TO MIDAMBLE:
If option FSE-B7 is built in and the softkey is active, FSE/FSIQ uses the vector signal analyzer to synchronize to the midamble. Power measurement is performed in the scalar spectrum analyzer mode by means of the time reference obtained by th signal demodulation.
Thus an accurate and updated tim e reference is ensured (even with a jittered trigger puls e). If the softkey is not active, the vector analyzer mode is off and FSE/FSIQ uses the time reference set using ADJUST TRIGGER.
1057.3992.02 1.27 E-2
Page 46
FSE-K11 Measuring the Carrier Power versus Time
Step 4
Setting the measurement bandwidth
MEAS
BANDWITH
CONDITIONS
NORM EXTR
MEAS MAX
OUTPUT PWR
À Press softkey
A table indicating the settable bandwidths is displayed on the screen. Selecting STANDARD (default setting) sets the measurement bandwith specified in the standard.
Or: Select the measurem ent bandwidth 3 kHz or 1 MHz using the up/down keys and press the Enter key.
MEAS BANDWITH.
Step 5
À The tolerance range specified in the
standards for the measurement is selected with adapted to the measurement conditions.
Default setting is NORM.
CONDITIONS
and thus
Step 6
À The reference level at power control
level 0 (static and dynamic) is determined with
PWR
and checked for the permissible tolerance against the level entered with softkey
CLASS
OUTPUT POWER
in menu
MEAS MAX OUTPUT
or
POWER
BTS TEST SETTINGS
.
Step 7
INC ST A T PWR CTRL
1057.3392.02 1.28 E-2
À The average carrier power is measured
at incremented static power control levels (1, 2, 3, 4, 5, 6) and checked for compliance with limit values. To do so the BTS level must be reduced each time by one step in interactive mode and softkey then to be pressed for the measurement.í
INCR STAT PWR CTRL
has
Page 47
FSE-K11 Measuring the Carrier Power versus Time
T
Additional information
Step 7
í
Upon completion of the measurement (after pressing seven times corresponding previous level reduction of BTS) a results table similar to the one below is displayed:
¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬§
E-GSM 900 CARRIER POWER
«¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬
ARFCN: 63 Status: PASSED
Frequency: 947.60000 MHz Sync:MIDAMBLE RBW: 1 MHz Nominal Power(max): 46.0 dBm Conditions: NORMAL
Ext Atten: 42.0 dB No of Bursts: 3
«¬¬¬ª¬¬¬¬ªª¬¬¬¬¬¬ª¬¬¬¬¬¬ª¬¬¬¬¬¬¬¬¬¬¬¬¬ªª¬¬¬¬¬¬¬ª¬¬¬¬¬¬¬¬¬¬¬¬¬ªª¬¬¬¬¬¬¬
NoCTRL NORM MEAS LIMIT PWR  MEAS LIMIT DELTA  STATUS
LVL  PWR PWR min.. max  DELTA min.. max  S D dBm dBm dBm  dB dB 
«¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬
774 12 14.0 14.1 11.0.. 17.0 -2.1 -3.5.. -0.5 PASSED
784 13 12.0 12.2 9.0.. 15.0 -1.9 -3.5.. -0.5 PASSED 794 14 10.0 10.1 7.0.. 13.0 -2.0 -3.5.. -0.5 PASSED
804 15 8.0 8.1 5.0.. 11.0 -2.0 -3.5.. -0.5 PASSED 815 0 36.0 36.1 33.0.. 39.0 -2.0 -3.0.. -1.0 PASSED
825 1 34.0 34.3 31.0.. 37.0 -1.9 -3.5.. -0.5 PASSED
835 2 32.0 32.3 29.0.. 35.0 -2.0 -3.5.. -0.5 PASSED 845 3 30.0 30.2 27.0.. 33.0 -2.0 -3.5.. -0.5 PASSED
855 4 28.0 28.2 25.0.. 31.0 -2.0 -3.5.. -0.5 PASSED
865 5 26.0 26.3 23.0.. 29.0 -1.9 -3.5.. -0.5 PASSED 875 6 24.0 24.2 21.0.. 27.0 -2.0 -3.5.. -0.5 PASSED
885 7 22.0 22.3 19.0.. 25.0 -1.9 -3.5.. -0.5 PASSED
895 8 20.0 20.2 17.0.. 23.0 -2.1 -3.5.. -0.5 PASSED 905 9 18.0 18.2 15.0.. 21.0 -2.0 -3.5.. -0.5 PASSED
915 10 16.0 16.3 13.0.. 19.0 -1.9 -3.5.. -0.5 PASSED 925 11 14.0 14.1 11.0.. 17.0 -2.1 -3.5.. -0.5 PASSED
935 12 12.0 12.2 9.0.. 15.0 -1.9 -3.5.. -0.5 PASSED
945 13 10.0 10.1 7.0.. 13.0 -2.0 -3.5.. -0.5 PASSED
955 14 8.0 8.1 5.0.. 11.0 -2.0 -3.5.. -0.5 PASSED 965 15 6.0 6.2 3.0.. 9.0 -1.9 -3.5.. -0.5 PASSED
976 0 34.0 34.3 31.0.. 37.0 -1.9 -3.0.. -1.0 PASSED 986 1 32.0 32.3 29.0.. 35.0 -2.0 -3.5.. -0.5 PASSED
996 2 30.0 30.2 27.0.. 33.0 -2.0 -3.5.. -0.5 PASSED
1006 3 28.0 28.2 25.0.. 31.0 -2.0 -3.5.. -0.5 PASSED 1016 4 26.0 26.3 23.0.. 29.0 -1.9 -3.5.. -0.5 PASSED
¨¬¬¬©¬¬¬¬©©¬¬¬¬¬¬©¬¬¬¬¬¬©¬¬¬¬¬¬¬¬¬¬¬¬¬©©¬¬¬¬¬¬¬©¬¬¬¬¬¬¬¬¬¬¬¬¬©©¬¬¬¬¬¬¬Á
Date: 31.AUG.1999 13:05:49
INCR STAT PWR CTRL
and
EX
1057.3992.02 1.29 E-2
Page 48
FSE-K11 Measuring the Carrier Power versus Time
Step 8
INC DYN
PWR CTRL
À In interactive mode, set the BTS
accordingly for the measurements at incremented dynamic power control levels (at a specific s tatic power control level) and press softkey
PWR CTRL
fifteen times. ÷. After this
INC DYNC
has been done for static power control level 6, the display is switched from split­screen to full-screen output of the table.
À The table can be scrolled using the
spinwheel or keys Ù and Ú.
or
Step 9
With the previous procedure m any power control levels have to be gone through even if only one power control level is to be measured.
SGL MEAS
ON OFF
STATIC PWR CTRL LEVEL
MEAS SGL
PWR LEVEL
CLEAR SGL
RESULT TAB
and
DYNAM PWR
CTRL LEVEL
If single power control levels are to be measured, proceed as follows:
À Switch softkey
SGL MEAS
to
ON.
À Select the power control levels to be
measured using sof tkeys
CTRL LEVEL
and
DYNAM PWR CTRL
STATIC PWR
LEVEL.
À Start the power measurement of this
power control level using sof tkey
SGL PWR LEVEL
.
MEAS
The results obtained can be cleared from the table using softkey
RESULT TAB
.
CLEAR SGL
The entries to be examined can be selected with the cursor keys and the spinwheel.
1057.3392.02 1.30 E-2
Page 49
FSE-K11 Measuring the Carrier Power versus Time
Additional information
Step 8
÷
An example of the table is given below in which the static power control level has been incremented once (to 1), then all dynamic power control levels (to 15) and again the static power control level (to 6) have been incremented.
Example: Measurement of the dynamic power control level for static power level 1. The dynamic power control level can be incremented for all static power control levels or only for the required ones. The measurement sequence is completed when the softkey is actuated again at static power control level 6 or when again at static power control level 6 and dynamic power control level 15.
1057.3992.02 1.31 E-2
INC STAT PWR CTRL
INC STAT PWR CTRL
is actuated
Page 50

FSE-K11 Measuring the Carrier Power versus Time

1.5 Measuring the Carrier Power versus Time  ô
Step 1
POWER
VS TIME
À Press softkey
POWER VS TIME.
Step 2
NO. OF BURSTS
and
SYNC TO
MIDAMBLE
NO. OF BURSTS
SET
MANUAL
SET TO STANDARD
À Press softkey
NO. OF BURSTS
and set
the number of bursts to be measured.
The number of bursts defined in the selected standard is set with SET TO STANDARD
(1 for all standards).
Step 3
Possible only if option FSE-B7 is installed
À Activate softk ey
and set synchronization to the burst midamble if the burst to be measured contains a midamble and a bit- accurate time reference is required.
À Switch off
bursts without midamble or if m aximum measurement speed is required. í
SYNC TO MIDAMBLE
SYNC TO MIDAMBLE
for
1057.3392.02 1.32 E-2
Page 51
FSE-K11 Measuring the Carrier Power versus Time
Additional information
BTS setting:
A single timeslot should be active on the transceiver (TRX) to be measured, all other timeslots of the TDMA frame should be set to IDLE. The measurement can also be performed if the BTS is in the SFH mode if option FSE-B7 is installed and To accelerate the measurement, switch off SFH on the BTS.
ô
Explanation concerning SYNC TO MIDAMBLE:
If option FSE-B7 is built in and the softkey is active, FSE/FSIQ uses the vector signal analyzer to synchronize to the midamble. Power measurement is performed in the scalar spectrum analyzer mode by means of the time reference obtained by th signal demodulation.
Thus an accurate and updated tim e reference is ensured (even with a jittered trigger puls e). If the softkey is not active, the vector analyzer mode is off and FSE/FSIQ uses the time reference set using ADJUST TRIGGER.
SYNC TO MIDAMBLE
is active in FSE/FSIQ.
Step 3
í
FSE/FSIQ measures at the set r eceive fr equency (ARFCN). If a num ber of bursts > 1 has been set, only measurement results will be considered at which FSE/FSIQ is able to synchronize (only if softkey measured changes cyclically in successive frames, only slots with the set slot number at the selected ARFCN will be considered in the measurement. However, this will considerably slow down the measurement sequence (approx. by the factor 3 compared to SFH). This disadvantage can be avoided if a trigger signal is provided for the FSE/FSIQ (triggering not on every frame but only if the selected slot is at the s et ARFCN). In this case, the measurem ent can be perf ormed at s low frequency hopping without synchronization to the midamble.
SYNC TO MIDAMBLE
is activated). Thus even if the carrier frequency of the slot to be
1057.3992.02 1.33 E-2
Page 52
FSE-K11 Measuring the Carrier Power versus Time
Step 4
MEAS
BANDWITH
REF MEAS
AUTO USER
START
REF MEAS
Setting the measurement bandwidth À Press softkey
MEAS BANDWITH.
A table indicating the settable bandwidths is displayed on the screen. Selecting STANDARD (default setting) sets the measurement bandwith specified in the standard.
Or:
À Select the measurement bandwidth 3
kHz or 1 MHz using the up/down keys and press the Enter key.
Step 5
REF MEAS AUTO / USER
select if a preview is to be perfor m ed prior to each measurement for the c orrect level adjustment of FSE/FSIQ or if the preview is to be started exclusively by the user
START REF MEAS
using Default setting is
REF MEAS AUTO
permits to
.
.
FULL BURST
RISING FALLING
or
or
RESOLUTION
HIGH
or
Step 6
À Select the resolution for the display of
bursts: ÷
FULL BURST
displayed
HIGH RESOLUTION
bits are displayed with 1 dB level resolution.
RISING:
edge with high time resolution.
FALLING
edge with high time resolution.
FALLING EDGE:
fallenden Burstflanke mit hoher
Zeitauflösung.
: The complete burst is
: The 147 useful
Display of the rising burst
: Display of the falling burst
Darstellung der
1057.3392.02 1.34 E-2
Page 53
FSE-K11 Measuring the Carrier Power versus Time
T
m
m
Additional information
Step 6
÷
30
20
10
0
-10
-20
-30
-40
34 dBm
34 dBm
21.7 dBm
21.7 dBm
1MAX 2MIN 3AVG
XG_BNU
RBW 1 MHzMax/Ref Lvl VBW 1 MHz SWT 800 Îs
LIMIT CHECK : PASSED
RF Att 20 dB Mixer -20 dB Unit dB
A
LN
TRG
1SA 2SA 3SA
EX
-50
-60
Date: 6.JUL.1999 13:11:02
Fig. 1-4 Display of burst for
FULL BURST
80 Îs/Center 935.2 MHz
1057.3992.02 1.35 E-2
Page 54
FSE-K11 Measuring the Carrier Power versus Time
Step 7
SINGLE
or
CONTINOUS
À Press softkey
measurement sequence or CONTINUOUS for a continuous measurement sequence. The bursts are displayed according to the selection made at step 4.
SINGLE
to start a single
1057.3392.02 1.36 E-2
Page 55
FSE-K11 Measuring the Carrier Power versus Time
This page is left intentionally empty to ensure that the user’s notes are always opposite to the additional information.
1057.3992.02 1.37 E-2
Page 56

FSE-K11 Measuring the Spectrum due to Modulation

/
1.6 M easur ing the Spectrum due to Modulation
Note: The test setup is described in section 1.1.
Step 1
MODULATION
SPECTRUM
À Press softkey
SPECTRUM.
ô
MODULATION
Step 2
NO. OF BURSTS
ARFCN
±1.8MHz
ARFCN ±/
TX BAND
or
RX BAND GSM 900
.
NO. OF BURSTS
SET
MANUAL
SET TO STANDARD
or or
or
TX BAND
RX BAND
RX BAND DCS1800
with GSM900 DCS1800
À Press softkey
NO. OF BURSTS
and determine the number of bursts to be measured.
The number of bursts defined in the selected standard is set with SET TO STANDARD
. This depends on the selected frequency band. í
Step 3
À Select the frequency range for the
measurement:
TX-BAND, RX-BAND ARFCN ± 1.8 MHz/TX BAND
possible only with GSM/DCS phase 1).
Caution: The test setup is changed for
ARFCN ± 1.8 MHz
or combined band
. (RX band
measurement in RX band.
,
RX BAND
with PCS 1900
Step 4
SINGLE
FREQ SWEEP
or
1057.3392.02 1.38 E-2
CONT
FREQ SWEEP
À Start the measurement in the frequency
range (spectral display) with
FREQ SWEEP FREQ SWEEP
(single sweep) or
(continuous sweep). í
SINGLE
CONT
Page 57
FSE-K11 Measuring the Spectrum due to Modulation
Additional information
BTS setting:
Set all timeslots of a T DMA f r ame to the same power and modulate them with pseudo-random data. (Exception: timeslot 0 may also be modulated with normal BCCH data). SFH (slow frequency hopping ) should be switched off.
Step 1
ô
The measurem ent is performed as prescribed by excluding the midamble of 50 to 90 % of the bur s t. (Precondition: correct setting of Adjust Trigger in the SETTINGS menu; step 1 If required, the setting can be checked and corrected using the
ADJUST TRIGGER
ù).
softkey. The measurement can be performed as a spectral display measurement or, depending on the standard, as a zero-span measurement with subsequent output of tabulated numerical results.
Step 4
í
A spectral display is not possible in the combined band
SINGLE FREQ SWEEP, CONT FREQ SWEEP
40
30
20
10
0
-10
-20
-30
-40
-50
-60
Fig. 1-5 Spectral display (
SINGLE FREQ SWEEP
ARFCN ± 1.8 MHz/TX-BAND
cannot be operated).
) in the range
ARFCN ± 1.8 MHz
(softkeys
1057.3992.02 1.39 E-2
Page 58
FSE-K11 Measuring the Spectrum due to Modulation
Step 5
START
LIST
À Start the strictly standard-conforming
zero-span measurement with
LIST:
÷
START
1057.3392.02 1.40 E-2
Page 59
FSE-K11 Measuring the Spectrum due to Modulation
Additional information
Step 5
÷
Fig. 1-6 Result display for zero-span measurement (
START LIST
) in the range
ARFCN ±1.8 MHz
1057.3992.02 1.41 E-2
Page 60
FSE-K11 Measuring the Spectrum due to Modulation

1.7 Measurement of Spectrum due to Transients

Note: The test setup is described in section 1.1.
Step 1
TRANSIENT
SPECTRUM
NO. OF BURSTS
MEAS
BANDWITH
NO. OF BURSTS
SET
MANUAL
SET TO STANDARD
À Press softkey
TRANSIENT
SPECTRUM.
Step 2
À Press softkey
the number of bursts to be measured.
The number of bursts defined in the selected standard is set with SET TO STANDARD
Default setting is
NO. OF BURSTS
.
NO OF BURSTS
and set
=1
Step 3
À Set the measurement bandwidth with
MEAS BANDWIDTH
the settable bandwidths is displayed on the screen. Selecting STANDARD sets the measurement bandwith specified in the standard.
Selection is performed by means of the up/down keys and the Enter key.
. A table indicating
Step 4
SINGLE
FREQ SWEEP
oder
CONT
FREQ SWEEP
or
START
LIST
1057.3392.02 1.42 E-2
ô
À Press softkey
or
CONT FREQ SWEEP
SINGLE FREQ SWEEP
The measurement is started in the
spectral range. Alternatively: À Press softkey
START LIST
A zero-span measurement is started ÷.
. í
.
Page 61
FSE-K11 Measuring the Spectrum due to Modulation
Additional information
BTS setting:
The measurement can be performed with or without SFH (slow frequency hopping) as required. The FSE/FSIQ measurement routine is able to handle the two cases. According to the standards, the slots of a TDMA frame can be measured at two different levels depending on whether the BTS supports dynamic power control or not.
Step 4
ô
The measurement can be performed as an overview measurement in the frequency range with spectrum display or as a zero span measurement and subsequent output of tabulated numerical results depending on the standard. In both cases a maximum off set frequency of ±1.8 MHz is used for the measurement.
í
Fig. 1-7 Display of spectrum due to transients of a GSM base station (meas urement in fr equency
range). Limit lines are inserted, the pass/fail info is displayed.
÷
Fig. 1-8 Display of transient spectrum list (zero-span measurement). Results , limits and pass/fail
information for the offs et frequencies prescribed by the standards and a general pas s/fail information are also displayed.
1057.3992.02 1.43 E-2
Page 62
FSE-K11 Measuring the Spurious

1.8 Measurement of Spurious

Note: The test setup is described in section 1.1.
Step 1
SPURIOUS
SPURIOUS
START LIST
SGL STEP
SHOW
CONT LIST
SPURIOUS
SGL STEP
START
LIST
SWEEP COUNT
TX BAND
<> TX BAND
RX BAND
TX BAND
±2.0 MHZ
SELECT
STEP
À Press softkey
SPURIOUS
.
SWEEP COUNT
Step 2
In the submenu which can be opened by means of soft key
SWEEP COUNT
number of sweeps for averaging in the RX band or the number of sweeps for determining the peak values within and outside the TX band. The SWEEP number required by the standard can be set with SET TO STANDARD in the submenu.
Press
to return to the SETTINGS
menu.
set the
1057.3392.02 1.44 E-2
Page 63
FSE-K11 Measuring the Spurious
Additional information
BTS setting:
The BTS is to be configured depending on the selected frequency band (TX, <>TX or RX band): GSM900, DCS1800 phase 1:
TX band and <>TX band If the BTS supports SFH (slow frequency hopping), SFH should be selected (change between B, M and T channel). Switch every second timeslot to full power.
RX band: Switch all slots to full power.
GSM900/DCS18000 phase 2: The BTS is to be configured depending on the selected frequency band (TX, <>TX or RX band): TX band:
Switch on a TRX with maximum power on all slots. Switch off SFH.
<>TX band: Switch every second timeslot to full power. If the BTS supports SFH, SFH should be selected.
RX-Band: Switch all timeslots of the BTS to full power. Switch off SFH (slow frequency hopping).
1057.3992.02 1.45 E-2
Page 64
FSE-K11 Measuring the Spurious
Step 3
TX BAND
6(/(&7
67(3
TX SUPPR
ON OFF
or
<> TX BAND
or
RX BAND
À Determine the measurem ent range with
TX BAND, <> TX BAND
RX Band) or
RX BAND
(outside TX and
. The test setup has to be modified depending on the frequency band used (see setion 1.1.3).
Other subranges can be selected or exluded in the used band using softkey
SELECT STEP
. Measurement in <>TX Band: À For BTSs with an output power of
approx. > 35 dBm, the carrier signal(s) must be suppressed by at least 25 dB to give FSE/FSIQ sufficient sensitivity for a large measurement bandwidth (≥ 300 kHz) and also to prevent harmonics generated by FSE/FSIQ from being measured. The suppression can be performed using a bandstop filter or a highpass filter if harmonics are measured. In this case T X SUPPR ON has to be activated in addition. FSE then increases its sensitivity by 20 dB (reference level reduced by 20 dB).
BTS SFH
ON OFF
SWEEPTIME
STD AUTO
or
BTS SFH
ON OFF
Step 4
À If the BTS is to be meas ured in the SFH
mode (slow frequency hopping), switch
BTS SHF
to ON. (Condition: ARFCN =
M channel). ô
Step 5
À Select between two settings for the
analyzer sweep time using
SWEEPTIME
long in some cases) required by the selected standard can be set with The minimum sweep time of the analyzer is set with
. The sweep time (very
STD
AUTO
.
.
1057.3392.02 1.46 E-2
Page 65
FSE-K11 Measuring the Spurious
Additional information
Step 4
ô
The measurement time is extended as required (by the factor 3 compared to SFH OFF). For measurements in the TX band the B and T channels are skipped in addition to the ARFCN channel.
1057.3992.02 1.47 E-2
Page 66
FSE-K11 Measuring the Spurious
Step 6
COSITING
À If a cositing BTS is to be measured for
GSM900/DCS1800 phase 2, activate the softkey
COSITING
í.
Step 7
START
LIST
Alternativ:
START LI ST
SGL STEP
and
CONT SGL
STEP LIS T
À If a complete measurement is to be
performed in the selected TX, <>TX or RX band, press softkey
START LIST
. Alternative: À For a
step-by-step
measurement of the individual frequency bands (TX and <> TX band), press softkeys START LIST SGL STEP and CONT SGL LIST:
After the measurement has been completed, a tabulated overview of results with limits and pass/fail information is displayed. ÷
1057.3392.02 1.48 E-2
Page 67
FSE-K11 Measuring the Spurious
T
Additional information
Step 6
í
A cositing BTS can send and receive in the GSM band and the DCS1800 band. If measurements are per formed in the RX band of the system used (defined under
SETTINGS), the RX band of the other system is also measured.
Step 7
÷
¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬§ 
P-GSM 900 II SPURIOUS LIST TX Band
«¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬ 
ARFCN: 1 Status: FAILED
Frequency: 935.20000 MHz Carrier Power: 20.0 dBm Ext Atten: 20.0 dB
 «¬¬¬¬ª¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬ª¬¬¬¬¬¬¬¬¬¬¬ª¬¬¬¬¬¬¬¬¬ª¬¬¬¬¬¬¬¬ 
No.
 «¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬         ¨¬¬¬¬©¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬¬©¬¬¬¬¬¬¬¬¬¬¬©¬¬¬¬¬¬¬¬¬©¬¬¬¬¬¬¬¬Á
937.00000 MHz --- 941.20000 MHz-60.3 dBm-36 dBmPASSED
941.20000 MHz --- 960.00000 MHz-19.1 dBm-36 dBm>FAILED<
1
2
3
4
5
6
Frequency
950.09138 MHz
949.82766 MHz
949.18717 MHz
948.54669 MHz
948.81042 MHz
947.75551 MHz
Level
-19.1 dBm-36 dBm>FAILED<
-20.2 dBm-36 dBm>FAILED<
-23.5 dBm-36 dBm>FAILED<
-28.0 dBm-36 dBm>FAILED<
-28.7 dBm-36 dBm>FAILED<
-30.6 dBm-36 dBm>FAILED<
Limit
Status
  
       
EX
Fig. 1-9 Example of a spurious list In addition to the general information at the top of the table including the general PASS/FAIL
information, all frequency ranges, the m aximum level measured in the range and the applicable range limit plus status information are listed.
A frequency range is defined by an identical RBW/VBW setting of the FSE/FSIQ and the same LIMIT.
If the margin or the limit of a frequency range is exceeded, spurious frequencies with levels are listed in addition (with numbers).
1057.3992.02 1.49 E-2
Page 68
FSE-K11 Sequence of Operations of FSE-K11 Measurements

2 General Information on Option FSE-K11

This application allows the following measurements to be performed on GSM900, DCS1800 or PCS1900 base stations in line with standards (GSM 11.20, GSM 11.20-DCS, GSM11.21 and J-STD­007 Air Interface) in a very convenient way:
PFE Phase/frequency error
CPW Carrier power
PVT Power versus time
MOD Spectrum due to modulation
TRA Spectrum due to transients
SPU Spurious emissions
This chapter gives basic information on the sequence of operations, settings and messages of the instrument. Details on the individual m easurements will be found in chapter 3. The information given under "General Remarks" is not needed for normal test operation. It is intended to provide a better understanding of instrument settings and operations.
Cold Start Warm Start
Opening FSE-K11
Changing within FSE-K11
A complete reset to default values is made. The previous status is assumed to a large extent. This applies to all
parameter values and to the selected measurement. The settings specific to the type of measurement and band are made. The default measurement type (LIST/SINGLE) is selected. The result history is cleared, so the results of previous measurements are no longer available after a warm start !
Re-opening FSE-K11 Action: Exiting FSE-K11 via the mode m enu, changing to non-FSE-K11 mode, re-opening FSE-K11. The response is the same as with a warm start except that the result history remains stored.
Changing between FSE-K10 and FSE-K11 Action: Switching between FSE-K10 and FSE-K11 via the mode menu. The response is the same as with a cold start.
Changing the standard Each change of standard causes a complete r eset to the default values valid for the standard in question. The reset is performed without any warning to the user.
Changing the phase Changes of phase are perform ed with as few modific ations as possible. If problems due to incom patibility of output data are detected during a phase change, the responsible group of parameters is reset to default values. In such case a tempor ary message is output inf orming the user of the modifications made. With IEC/IEEE-bus control, the response is the same as with a change of standard !
Possible problems and resulting actions: Change of
frequency group:
1057.3440.41 2-1 E-3
No problems are to be expected.
Page 69

Sequence of Operations of FSE-K11 Measurements FSE-K11

Level group: If the power class or the power control level of the
previous phase does not exist for the new phase, the parameters assigned to the level group are set to default values.
Warning: "OutputPower & ExtAtten set to Default!"
Changing the measurement type / band
Measurement type and band:
Changing the measurement type When the measur ement type is changed (eg from SPU to MOD), initial actions (basic settings) specific to the selected measurement type are performed once. A def ault band or the band last active is selected. LIST measurement (for SPU, TRA, MOD) or SINGLE measurement (PFE, PVT) can be selected as the measurement mode. In the case of the CPW m enu being opened, the measurement last active is selected as the measurement mode.
Changing the band If the band is changed, no visible responses take place as a rule. There are no other settings or initial actions performed.
The response is the same as with a change of measurement type (see below).
2.1 Sequence of Operations of FSE-K11 Measurements
Measurements performed with FSE-K11 can be divided into 2 types according to their sequence of operations:
Measurements with and without pre-measurement
a
b
In the pre-measurement, the currently applied signal level, which is essential for the main measurement, is determined if requir ed for the selected m easurement. T his applies to all relative meas urements ( see below) which are based on a reference value. If the measured s ignal level is below a minimum value
(expected value –20 dB), the measurement is interrupted and a warning is output. Otherwise the m ain measurement is performed.
On opening a measurement menu, presettings are first performed which set the FSE/FSIQ to the default state required for the m easurem ent type in question. On starting the m easurem ent, the settings required for the selected measurement are carried out. Then the measurement is performed, if necessary in several consecutive steps. At the end of the measurement, post-settings (such as the activation of markers) are made if necessary and results are displayed.
Pre-measurement ===> Main measurement
PreSettings [PostSettings] PreSettings PostSettings
[Sequence data]
Main measurement
PreSettings PostSettings
[Sequence data]
1057.3440.41 2-2 E-3
Page 70
FSE-K11 Sequence of Operations of FSE-K11 Measurements
The following overview lists the types of FSE-K11 measurement in accordance with the above description:
Measurement Mode Pre-
KXX PFE ..... XX ---
KXX CPW ..... XX XX
KXX CPI ..... XX XX
KXX PVT ..... XX --- Also without pre-measurement, user-selectable;
KXX MOD RX SINGLE,
KXX MOD RX LIST XX XX KXX MOD Tx,
ARFCN ±1.8 MHz KXX MOD Tx,
ARFCN ±1.8 MHz
KXX SPU xxx ..... XX XX
CONT
SINGLE, CONT
LIST XX XX Bandwidth correc tion necessary (see below)
measurement without with
XX ---
Sequence data Remarks
normally with 2 pre-measurem ents
XX --- Bandwidth correction necessary (see below)
Measurements of the spectrum due to modulation require a reference value for the limit lines to be measured at 30 kHz bandwidth. To evaluate the actual carrier power (ie the signal power measured at infinite bandwidth) a bandwidth-dependent correction value is added to the pre-m easured value: this is at present 8 dB for 30 kHz (following GSM 05.50). The corrected value is also used for selecting the limit lines.
The following diagram shows the sequenc e of operations of a m easurement perf ormed with FSE-K11. This diagram applies to all measurem ents except for the CPW /CPI meas urement, which is shown in a separate diagram (for sequence of operations see also chapter 3).
The measurement type is selected in the main menu. Prior to this, general settings are performed if necessary, with most parameters also being acc essible from the side menus of the measur ements. On opening a measurement menu, general settings specific to the measurement in question are made. These settings are made only once. In the basic menu for a given measurement, the results of the measurement last perfor med (table) are displayed in the full-screen mode (MOD , TRA, SPU, CPW ) on opening the menu. For the CPI, PVT and PFE measurements, on the other hand, the screen settings for the main measurement are activated immediately (for CPI the table cannot be displayed in the full­screen mode). For MOD and SPU the band can be selected. T he table contents change depending on the selected band.
The measurem ent start initiates the (f irst) pre-m easurem ent. After com pletion of the pre-m easurem ent, the measured signal level is evaluated. If the pre-measurement is succ essful, either the second pre­measurement is performed with optimum sensitivity (for PVT) or the m ain measurem ent is carried out with level control orientated to the pre-measurement. The result is displayed at the end of the measurement.
1057.3440.41 2-3 E-3
Page 71
Sequence of Operations of FSE-K11 Measurements FSE-K11
0
Main menu
1
General
test
parameters
2
3
Selection of
measurement
4
Default setting
for preset
parameters
5
Measurement
menu
Setting of measurement-specific parameters (ARFCN, ...
Menu up
10
6
Display of
table
8
Optional:
band selection
7
MOD/
SPU
nein
9
Start
ja
10.210.1
Fig. 2-1 Sequence of operations upon selecting a measurement
1057.3440.41 2-4 E-3
Page 72
FSE-K11 Sequence of Operations of FSE-K11 Measurements
Explanation of Legend:
0 T he m ain menu of FSE-K11 is called: general settings (SETT INGS) or a m easurem ent can be
selected. 1 General settings for GSM measurement (ARFCN, attenuation, power, trigger, ...). 2 Main m enu: one of the FSE-K11 measurements PFE-CPW/CPI-PVT-MOD-T RA-SPU can be
selected. 3 Selection of measurement type via main menu. 4 FSE/FSIQ assumes default setting for the selected measurement. 5,6,7,8 In the basic menu for a given measurem ent, the results of the measurement last perfor med
(table) are displayed in the full-screen mode ( MOD, TRA, SPU, CPW ) on opening the m enu.
For the CPI, PVT and PFE measurem ents, on the other hand, the s creen settings for the m ain
measurement are activated immediately (for CPI the table cannot be displayed in the full-
screen mode). For MOD and SPU the band can be selected. The table contents change
depending on the selected band.
1057.3440.41 2-5 E-3
Page 73
Sequence of Operations of FSE-K11 Measurements FSE-K11
10.1 10.2
9
Start
11
13.1
No
Carrier
detected
No
MOD/TRA/
PVT
ja
12
(First) pre-
measurement
13
Carrier power
sufficient
ja
NO
see detail
chart
14
PVT?
Yes
15
Second pre-
measurement
16
Hauptmessung
17
Display of results
(table / trace
No
see detail
chart
see detail
chart
Fig. 2-2 Sequence of operations upon starting a measurement (overview)
1057.3440.41 2-6 E-3
Page 74
FSE-K11 Sequence of Operations of FSE-K11 Measurements
Explanation of Legend:
9 Start of measurement with ’SINGLE’, ’CONTINUOUS’ or ’START LIST’. 11 For the measurements TRA/PVT and MOD (not RX band), the (first) pre-measurement is
performed; for the measurements PFE, CPW/CPI, MOD-Rx and SPU the main measurement is
started immediately. 12 The sequence of operations of a pre-measurement is presented in detail in a separate diagram
(see "General sequence of operations of a measurement").
This sequence chart applies to both pre-measurements and main measurements. 13 Sufficient carrier power: the measured signal level (incl. bandwidth correction) must not be
more than 20 dB below the signal level to be expected for the settings made.
13.1 An error message is output temporarily and the IEC/IEEE status bit set. 14, 15 After the first pre-measurement, a second pre-measurement is carried out for PVT (see
"Sequence of operations of PVT measurement in detail") to determine the exact reference
value.
The second pre-measurement is performed with optimum sensitivity. 16 Main measurement with level control orientated to pre-measurement. The sequence of
operations is presented in detail in a separate chart (see "Gener al sequence of oper ations of a
measurement") .
This sequence chart applies to both pre- and main measurements. 17 The result is displayed as a table or a trace.
1057.3440.41 2-7 E-3
Page 75
Sequence of Operations of FSE-K11 Measurements FSE-K11
10
From main menu
100
Start
101
AUTO / pre-
measurement
ja
102
(Frist)
pre-measzrement
No
107
Determination
and storage of
level last pre-
measurement
Yes
104
No carrier
detected
106
USER pre-
measurement?
No
No
103
Carrier power
sufficient
g
ja
105
Second pr e-
measurement
108
Main measurement
with settings of
(previos) pre-
measurement
109
Display of
trace including
limit lines
Fig. 2-3 Sequence of operations of PVT measurement in detail
1057.3440.41 2-8 E-3
Page 76
FSE-K11 Sequence of Operations of FSE-K11 Measurements
The above chart presents the PVT measurement in detail. The sequence of oper ations is deter mined by
the setting for the pre-measurement – With AUTO, the PVT measurement is performed as follows:
1st pre-measurement - 2nd pre-measurement - main measurement. The settings for the fir st pre-m easurem ent (with m ax. 2 measured burs ts) ar e m ade in acc ordance with
the expected level. The settings for the second pre-measurement are made in accordance with the signal level determined in the first pre- measurement, and level measurement is performed with max. 20 burs ts. The signal level determined in the second pre-m easurement is us ed as a reference for the limit lines used in the subsequent main measurement.
Two-stage pre-measurement results in highly accurate level determination even in cases where the signal level is considerably below the expected value since, thanks to the follow-up level control, it is ensured that the measurement is always performed with optimum sensitivity.
With the USER setting, the pre-measurem ents are skipped, and the level determined in the last pre­measurement is used both for setting the operating point and as a reference f or the limit lines. If no level has been determined, the nominal value (expected signal level) is used. This setting allows fast switching between the various burst views (Rising, Falling, Full, HighRes). Actuating the Start RefMeas key will trigger a sequence of two pre-measurements without any main measurement following.
REF MEAS AUTO or REF MEAS USER.
1057.3440.41 2-9 E-3
Page 77
Sequence of Operations of FSE-K11 Measurements FSE-K11
19
ZeroSpan vs FrequSweep,
Trigger, Trace, Detector, ...
Yes
Settings:
InpAtt & RefLvl,
20
Sync to
Midamble
nein
No
Time reference
30
31
required?
200
See detail
Sync to midamble
PFE / PVT & CPW
32
Trigger
FreeRun
Yes
ja
33
34
35
36
Sweep and data
acquisition
n
37
Sweep end at
tirgger/gate
34
Waiting for trigger
event
No
39
Pre-
measurements
result or final
result
Yes
38
No
Number of Bursts
attianed and
SingleSweep
or List?
Fig. 2-4 General sequence of operations of a measurement
1057.3440.41 2-10 E-3
Page 78
FSE-K11 Sequence of Operations of FSE-K11 Measurements
The sequence of operations is the same for each m easurement (the above sequence c hart applies to both pre-measurements and main measurements):
19 The settings needed for the intended measurement are made. 20 Synchronization to midamble is optional for PVT and CPW/CPI and mandatory for PFE.
This choice is possible only if option FSE-B7 is fitted. 30-39 Measurement without digital demodulator (ie without option FSE-B7). 31,33 All measurements (except transient and spurious) require a time reference. 32,33 If FreeRun is not selected, FSE-K11 waits for the (external) trigger event here too. 34 FSE-K11 waits for the trigger event, then starts or continues the sweep. 35 The sweep is enabled. 36 Measured data are collected until the end of the sweep or, with gated sweep, during a specific
gate time. 37 It is checked if the sweep is terminated. 38 It is checked if the number of preset s weeps has been attained AND if the measurement is of
the SINGLE or the LIST type. 39 The m easurement is completed if the preset number of sweeps has been attained and the
measurement is of the SINGLE or the LIST type. If digital demodulation is active, there are s ome dif ferenc es concerning tr igger logic as com pared to the
sequence of operations shown in the chart "General sequence of operations of a measurement". The sequence of operations with digital demodulation active is shown below for PFE and PVT/CPW:
1057.3440.41 2-11 E-3
Page 79
Sequence of Operations of FSE-K11 Measurements FSE-K11
200
Sync to midamble
201
202
Trigger
FreeRun
206
Burst found?
Yes
No
Yes
No
203
204
205
Sweep and data
aaquisition
207
Burst not
found
203
Waiting for
trigger event
208
No. of Burst > 1
No
Yes
209
Sync found?
Yes
No
Measurement completed
210
Sync not
found
212
Trace and
Limit check
222
211
No. of Burst > 1
No
213
Max. No. of
Bursts and
SINGLE
Yes
Yes
No
Fig. 2-5 PFE - Sequence of operations of synchronization in detail
1057.3440.41 2-12 E-3
Page 80
FSE-K11 Sequence of Operations of FSE-K11 Measurements
PFE – sequence of operations: 200-204 Waiting for trigger event or FreeRun (not meaningful !).
205 Collection of measured data. 206 Search for bursts in a predefined time window (compare settings for result buffer). 207, 208 In case of an error, a message is output and the IEC/IEEE st atus bit 'Burst not Found' is set.
In the special case of "NofBur sts == 1", the error condition is ignored however dur ing further
processing. 209 Search for sync pattern in the selected burst. 210, 211 In case of an error , a message is output and the IEC/IEEE status bit 'Sync not Found' is set.
In the special case of "NofBursts == 1", however, the err or condition is ignored during f urther
processing. 212 Evaluation of trace with numerical indication of the measured frequency error and phase
error. 213 If the predefined number of bur sts has been attained and the m easurem ent has been star ted
with 'SINGLE', it ends at this point.
1057.3440.41 2-13 E-3
Page 81
Sequence of Operations of FSE-K11 Measurements FSE-K11
200
Sync to midamble
251
252
Trigger
FreeRun
256
Sync f ound?
Yes
No
Yes
No
253
254
255
Sweep and data
acquisition
257
Sync not
found
253
Waiting for
trigger event
No. ofBurst > 1
258
Yes
No
259
Trace and
Limit check
222
Measurement compledet
260
Max No. of bursts
and SINGLE?
Yes
No
Fig. 2-6 PVT/CPW - sequence of operations of synchronization in detail
1057.3440.41 2-14 E-3
Page 82
FSE-K11 Sequence of Operations of FSE-K11 Measurements
PVT/CPW – sequence of operations: The sequence of operations for PVT/CPW differs from that for PFE in that no burst search is performed.
Only the sync pattern (items 256 and 257) is searched within a limited range of the recorded data stream (compar e "Settings") to establish a high-precision time ref erence. In the case of an error, the procedure is the same as for PFE, including the special case of "NofBursts == 1".
200-254 Waiting for trigger event or FreeRun (not meaningful !). 255 Collection of measured data. 256 Search for sync pattern in the selected burst. 257,258 In case of an error, a m essage is output and the IEC/IEEE st atus bit 'Sync not Found' is set.
In the special case of "NofBur sts == 1", the error condition is ignored however dur ing further
processing. 259 Evaluation of trace with limit check. 260 If the predefined number of bur sts has been attained and the m easurem ent has been star ted
with 'SINGLE', it ends at this point.
1057.3440.41 2-15 E-3
Page 83
Sequence of Operations of FSE-K11 Measurements FSE-K11
1
Main menu
2
3
MEAS MAX
OUTPUT PWR
4
Sta:= Dyn:= 0
Measurement (0,0)
5
13
Half-screen display
of table
12
Messung mit
Ctrl Lvl (Sta,Dyn)
6
All Pwr Ctrl
levels?
No
7
Max. Dyn Ctrl
Level?
ja
10
INC STATIC
PWR LEVEL
11
Sta := Sta + 1
Dyn := 0
Yes
No
14
Full-screen display
of table
No
8
INC DYNAM PWR LEVEL
Yes
9
Dyn := Dyn + 1
Fig. 2-7 Sequence of operations of CPW measurement in detail
1057.3440.41 2-16 E-3
Page 84
FSE-K11 Sequence of Operations of FSE-K11 Measurements
The sequence of operations of a CPW m easurement ( BTS/K11) is shown above. The m easurement is initiated by pressing the ’MEAS MAX OUTPUT PWR’ key. Then either the static or the dynamic power control level can be increm ented as required. Each increm entation of the static power control level will reset the dynamic power control level to 0 (zero). If the initializing k ey ’MEAS MAX OUTPUT PW R’ is actuated again during the current session, the session is aborted (measurement status: ABORTED). Entries by the user are dependent on the availability of keys ’INC STATIC PWR LEVEL’ and ’INC DYNAM PWR LEVEL’: if the maxim um value of the dynamic power control level is attained, the ’INC DYNAM PWR LEVEL’ key is inhibited so that only the static power control level can be increased. If the maximum value is attained here too, pressing the key again causes the session to be term inated and the total result to be displayed on the full screen.
With the ’full-screen table’ status, both INC keys are inhibited, and a new session can be started only by pressing the ’MEAS MAX OUTPUT PWR’ key.
Unlike the CPW measurement, the CPI m easurement allows any single carrier-power m easurements. In contrast to all other measurements performed with FSE-K11, single CPI measurements are not related to one another. There is no total result for the session. The CPI menu can be exited and activated any time without this having any effect on the table contents of the CPI measurement.
CPI measurements are intended m ainly for developm ent engineers in the lab who m odify parameter s of their workstation and want to investigate the effect s thereof on the carrier power. CPW m easurem ents, on the other hand, are aimed at verifying compliance with acceptance test specifications of the GSM standard.
CPI measurements, therefor e, are only used for evaluating absolute levels; no relative c ompar isons are made between the levels obtained with different power control level settings. Operator control is described in the additional information following the description of the softkey menus.
Explanation of Legend:
Selection of CPW measurement, activation of main menu (deactivation of CPI measurement if necessary) 3,4 Start of measurement with maximum power in line with standard (CTRL LEVEL 0.0).
5,13 End of sweep, result is entered into table. 6 Measurement completed? 7-12 If a power control level is incremented, the next sweep is started. 14 End of measurement, switchover to full screen.
1057.3440.41 2-17 E-3
Page 85
Sequence of Operations of FSE-K11 Measurements FSE-K11

2.1.1 Errors and Failures During Measurements

There are three main sources of error in GSM measurements:
Carrier signal missing: Causes: Wrong operating frequency (ARFCN), SFH active, test line defective or incorrect
attenuation, wrong power class
Effects: Pr e-m easurem ent: inadequate power meas urem ent (eg average meas urement with
SFH), measurement is aborted. Sweep stops (in measurements with midamble synchronization).
Main measurement: s weep stops when midamble synchronization (PFE, PVT, CPW)
is active and number of bursts > 1. When signal is applied again, the sweep continues.
Trigger missing: In all triggered measurements: sweep stops (indicated by trigger LED).
The following is recommended for the trigger source:
Optimal solution: External trigger. This allows all measurements without restrictions.
FreeRun: Allowed for SPU and TRA if not possible otherwise
(MOD-Tx without gating also possible). Burst missing: Causes: SFH active, wrong frame trigger Effects: In triggered measurements, limit lines may be exceeded (PVT);
sweep stops (in measurements with midamble synchronization).
2.1.2 Abort of Measurements a) Abort by User
In manual operation, each running measurem ent can be aborted by pressing the start k ey again (ie the key with which the measurement was started). Starting another measurement likewise caus es an abort of the running m easurem ent. This also applies to switchover from SINGLE to CONTINUOUS or LIST, for example.
all
Measurements are also aborted on exiting the menu in question (New! From 1.90 for or modifying a test parameter (even in side menus).
measurements)
All aborts, except for spurious single-step measurements, are without warning. Only if – during a running spurious single-step measurem ent – the user attempts to res tart this measurem ent or to start the LIST measurement, is he queried to confirm the (re)start before it is made.
For remote control, a special ABO RT comm and is available which causes the current measurem ent to be aborted (see manual for IEC/IEEE-bus control).
1057.3440.41 2-18 E-3
Page 86
FSE-K11 Sequence of Operations of FSE-K11 Measurements
b) Abort by Instrument
Measurements are norm ally performed until the end. Only in the case of insuff icient carrier power (see above) are measurements aborted after the pre-measurement and an error message is output.
If the maximum num ber of table entries (presently 50) is exceeded in a (spurious) measurement, the measurement is continued but no new entry is made in the table. This does not apply to the CPW measurement, where all possible single results (7 x 16 = 112) are entered in the table. In CPI measurements, the last 40 single measurements are stored.

2.1.3 Results of Measurements

Measurement Single result Limit line Table
PFE Peak and RMS phase error,
CPW, CPI
PVT Upper and
MOD SG
MOD List Rx/Tx
MOD List ARFCN ±1.8 MHz
TRA Sgl
TRA List
SPU For each section:
frequency error in ppm
One entry per single measurem ent: CPW: Ctrl Level , measured value, limit value and evaluation CPI: test parameter, measured value and evaluation
lower limit line Upper limit line
For each section: One frequency-level entry for each event. One summary entry for eac h section (see below).
One entry per pair of test points : Frequency offset (±), level(-), level(+). Test points defined by st andard.
Upper limit line
One entry per pair of test points : Frequency offset (±), level(-), level(+). Test points at ± 400, 600, 1200 and 1800 k Hz.
One frequency-level entry for each event. One summary entry for eac h section (see below).
The result tables of the MOD-LIST Rx/Tx and SPU measurements are arranged in groups: The whole measurement range is divided into frequency sections. A summary entry is generated for each section. T his entry contains the section lim its and the m axim um
level of the worst quality class measured in this section. All "events" are then listed as frequency-level pairs, an event being a measured value which lies above limit or at least in the tolerance range.
Each measured value is assigned one of the following quality classes:
PASSED (best quality class)
MARGIN
EXCEPTION
FAILED (worst quality class)
1057.3440.41 2-19 E-3
Page 87
Sequence of Operations of FSE-K11 Measurements FSE-K11
PASSED values are measured values below the tolerance range. MARGIN values are values below the limit value but within the (user-definable) tolerance range. EXCEPTION values are values above the limit value but not exceeding the threshold for outliers. FAILED values violate both the limit value and the exception threshold. All measured values except PASSES are events in the above sense.
The section entries (and the table as well) are assigned either PASSED or FAILED as a summary evaluation.
If the number of outliers measured in one section exceeds the value defined by the standard, the section is assigned FAILED even if all events are assigned EXCEPTION in the worst case.
The tables may contain a maximum of 50 lines (except for Carrier Power, where the maximum is 100 lines). In some measur ements, theref ore, it may happen that FAILED is indic ated as a summ ary evaluation in the table header although all visible entries in the table are evaluated PASSED. The reason for this is that the FAILED evaluation m ight be output in the 65th line only and is therefore not visible.
1057.3440.41 2-20 E-3
Page 88

FSE-K11 Exiting FSE-K11

2.2 Exiting FSE-K11
When FSE-K11 is exited, the status las t active is stor ed so that it can be restor ed as f ar as possible on re-opening FSE-K11. The FSE-K11 settings are largely transferred to the other operating modes to allow FSE-K11-conformal measurements also in these modes as far as this is expedient.
FSE-K11 can be exited by:
Power off
Preset and mode selection
<Mode> key and mode selection
Power off FSE-K11 is restarted (warm s tar t) on power-up of FSE/FSIQ. FSE-K11 is in this case not exited but only interrupted.
Preset The instrument is reset to a defined default status. This function is comparable to a cold start. For FSE-K11 this means: no re-activation of the status las t ac tive when FSE-K11 is r e- opened but activation of the default status. The target status of the instrument after a pr eset is c learly defined independently of the previous FSE-K11 status (see paragraph "Preset" below).
Changing the mode The target status assum ed results from the FSE-K11 st atus last active and the s elected target mode. The parameters for the target mode are set in s uch a way that measurements in line with FSE-K11 are possible. For changing from FSE-K11 to the analyzer and vector analyzer modes, the following is defined:
Transformation of (FSE-K11) parameters: Trigger
TrgSource compatible (if not available in target mode: FreeRun).
TrgLevel is taken over for all modes (if available).
TrgDelay corresponding to trigger last active.
TrgPolarity, TrgEdge/Level are taken over.
Level
ExtAtt is represented by RefLvl offset.
RefLevel is set directly.
RF attenuation is calculated in line with FSE-K11 (AUTO or by formula).
Caution: For RX measurements, the input attenuation is set to AUTO ===> not to 0 dB !
Frequency For the zero span or the vector mode, the ARFCN frequency is set as the center
frequency. For the frequency sweep mode, the start and the stop frequency are selected in accordance with the band last selected:
for TX/RX the band in question (possibly incl. ±2 MHz), for TX ±1.8 MHz the center frequency f (AFRCN) with a span of 3.6 MHz, for SPU the section last selected (or more sim ply, the complete band of the non-T X
measurement) .
1057.3440.41 2-21 E-3
Page 89
Exiting FSE-K11 FSE-K11
Screen Full screen is taken for the analyzer target mode, split screen for the vector target mode (PFE).
Trace Trace assignment in FSE-K11:
2 for single-trace measurements in LIST mode (trace 1 used for table) 1 for single-trace measurements in SINGLE/CONT mode 1,3,2 for PVT measurements with MAX Hold, Average, MIN Hold 2 for SPU measurements 1,3,4 for PFE measurements with MAX Hold, ClearWrite, MIN Hold 2 for CPW/CPI measurements (Average) Trace assignment is maintained as far as possible.
For changing from FSE-K11 to modes other than the analyzer or vector analyzer modes, the following is defined:
On exiting FSE-K11, the FSE-K11 operating status last active is stored in the newly selected mode (eg RECEIVER mode, or another mode) and activated again by this mode on re-opening FSE-K11.
1057.3440.41 2-22 E-3
Page 90

FSE-K11 Preset

2.3 Preset
FSE-K11 is reset to the default status. Then the following setting is assumed: Analyzer default setting: Frequ-Sweep, FullSpan, Sweep Time Auto, RBW/VBW Auto, Trigger FreeRun

2.3.1 Analyzer Mode

The target status assumed results from the FSE-K11 status last active (or the last measurement performed with FSE-K11, respectively.
PFE Same as PVT. CPW/CPI ZeroSpan; f = f (ARFCN); sweep time 540 µs.
Trace 2 is renamed trace 1 because of full-screen display. PVT ZeroSpan; f = f (ARFCN); sweep time 800 µs or 100 µs. MOD
TRA FrequSweep; span 3.6 MHz, f = f (ARFCN); sweep time in accordance with T RA formula SPU Continuous FrequSweep; start and stop in accordance with frequency range last selected
FrequSweep; start and stop of span in accordance with TX / ARFCN ±1.8 MHz / RX;
non-RX: trigger at FreeRun, plus gated sweep with gate length 170 µs. Otherwise trigger
is taken over (setting analogous to CONTINUOUS measurement).
(settings analogous to CONTINUOUS measurement).
or with band (TX/RX/non-TX); AUTO sweep time. Trigger is taken over. Trace 2 is
renamed trace 1 because of full-screen display.

2.3.2 Vector Mode

The PFE status of FSE-K11 (vector mode) is set irrespective of the FSE-K11 status: split screen with result table at the top and trace at the bottom.
1057.3440.41 2-23 E-3
Page 91

General Remarks FSE-K11

2.4 General Remarks
In the case of the TRA, MOD and PVT m easurements, a pre-measurem ent is performed prior to the main measurement. This pre-measurement is used for defining the reference point for the limit lines. In the case of PVT, the pre-m easurement tak es place in two s tages. In the f irst pre-m easurement with max. 2 bursts, the actual level applied is determined. In the second pre-measurement with max. 20 bursts, the exact reference point for the limit lines is determ ined with an operating point optimally set by means of the result of the first pre-measurement.
In the PFE, CPW and SPU m easurements, on the other hand, f ixed level values are used and thus no pre-measurements are required.
"ExtAtt" (external attenuation) is taken into account in all measur ements in the carrier or non-T X band, "RxGain" for all measurement in the RX band. The values have basically the same effect but are us ed separately. RxGain can be entered any time, not only for an activated RX band measurement.
In all measurements with Min, Max and Average dis play, three traces are ac tivated only if the number of sweeps is higher than 1. Likewise, vector triggering operates c orrectly only if the number of sweeps is set to a value higher than 1 (this too is a feature of the basic unit). If this is not the case, an error message is output, for example "Sync not found" if the midam ble is not detected, in whic h case there is no synchronization to midamble.
Explanation of abbreviations used in the formulas below: SigPwr (Expected) signal power, defined by the current settings for the power c lass and power
control level (compare section 3.3.4) MeasPwr Measured signal power CorrPwr Measured signal power, corrected by a value depending on the bandwidth
(30 kHz RBW: +8 dB) MaxPwr Maximum (nominal) output power F(ARFCN) Operating frequency, resulting from frequency setting (compare section 3.3.3) Tx Su p 20 dB or 0 dB is taken into account in the f orm ula depending on whether TX suppression
is set to ON or OFF ExtAtt External attenuation RxGain RX band gain SFH SFH (slow frequency hopping) factor: 1 for SFH OFF, 3 for SFH ON NOB Number of bursts – value set by the user for the number of sweeps
Examples giving numbers, if any, refer to the default settings on power-up (PGSM-II).
1057.3440.41 2-24 E-3
Page 92

FSE-K11 Standard

2.5 Standard
Standard P-GSM Phase II Measurement: SPU-Tx
Type of measurement/band assignment for initial selection of measurement:
PFE CPW PVT MOD TRA SPU
Single MeasMax Full Single
ARFCN ±1.8 MHz
Table 2-1 Parameter values (BTS)
Parameter PGSM-I PGSM-II DCS-I DCS-II PCS E-GSM R-GSM
ARFCN 1 1 512 512 512 0 0 Frequ [MHz] 935.2 935.2 1805.2 1805.2 1930.2 935.0 935.0 ARFCN-Min/Max 1-124 1-124 512-885 512-885 512-810 975-1023
ExtAtt [dB] 20 20 20 20 20 20 20 PwrClass 4 4 1 1 1 4 4 PwrCtrLvl 0.0 0.0 0.0 0.0 0.0 0.0 0.0 OutputPwr [dBm] 46 46 43 43 43 46 46 PwrCouplingONONONONONON ON SFH OFF OFF OFF OFF OFF OFF OFF Tx-Supression OFF OFF OFF OFF OFF OFF OFF Limit Margin [dB] 1 1 1 1 1 1 1 Bursts 1 1 1 1 1 1 1
List ListTx List
0-124
955-1023 0-124
ARFCN-SFH 62 62 573 573 573 13 13 Frequ-SFH ARFCN-Min/Max 62-63 62-63 573-824 573-824 573-749 13-63 0-63 ARFCN-B/T 1/124 1/124 512/885 512/885 512/810 975/124 955/124
Table 2-2 Output powers of BTS power ranges:
BTS M1 > 19-24 dBm to > 27-32 dBm BTS M2 > 14-19 dBm to > 22-27 dBm BTS M3 > 9-14 dBm to > 17-22 dBm
1057.3440.41 2-25 E-3
Page 93
Standard FSE-K11
Table 2-3 General standard specifications
Range Channel
P-GSM 1 to 124 f = 890 MHz + (n * 0.2 MHz) +45 MHz E-GSM 0 to 124 f = 890 MHz + (n * 0.2 MHz) +45 MHz
DCS1800 512 t o 885 f = 1710.2 MHz + ((n - 512) * 0.2 MHz) +95 MHz PCS1900 512 to 810 f = 1850.2 MHz + ((n - 512) * 0.2 MHz) +80 MHz R-GSM 0 to 124 f = 890 MHz + (n * 0.2 MHz) +45 MHz
P-GSM 900
E-GSM 900
R-GSM 900
DCS 1800
PCS 1900
numbers n
975 to 1023 f = 890 MHz + ((n - 1024) * 0.2 MHz) +45 MHz
955 to 1023 f = 890 MHz + ((n - 1024) * 0.2 MHz) +45 MHz
925.2MHz 934.8 935 935.2 959.8MHz 975 1023 0 1 124
921.2MHz 934.8 935 935.2 959.8MHz 955 1023 0 1 124
1805.2MHz 1879.8MHz 512 885
1930.2MHz 1989.8MHz 512 810
Downlink frequency ( BS-->MS)
935.2MHz 959.8MHz 1 124
Duplex offset
FREQ
ARFCN
FREQ
ARFCN
FREQ
ARFCN
FREQ
ARFCN
FREQ
ARFCN
Table 2-4 Exceptions and base line values for MOD measurements
Standard Range Exceptions Base l i ne value Applicable to
PGSM-I 400 kHz to 1800 kHz -36 dBm X-GSM 600 kHz to 6. 0 MHz 3 of max. -36 dBm each -65 dBm
-59 / -64 / -69 dBm
6.0 MHz to TX band 12 of max. -36 dBm each -65 dBm
-59 / -64 / -69 dBm
DCS-I 800 kHz to 6.0 MHz 3 of max. -36 dBm each -57 dBm PwrClass 1 to 4 (all)
6.0 MHz to TX band 12 of max. -36 dBm each -57 dBm PwrClass 1 to 4 (all)
DCS-II 600 kHz to 6.0 MHz 3 of max. -36 dBm each -57 dBm
-57 / -62 / -67 dBm
6.0 MHz to TX band 12 of max. -36 dBm each -57 dBm
-57 / -62 / -67 dBm
PCS 600 kHz to 6.0 MHz 3 of max. -36 dBm each -57 dBm
-57 / -62 / -67 dBm
6.0 MHz to TX band 12 of max. -36 dBm each -57 dBm / -57 / -62 / -67 dBm PwrClass 1 t o 4
PwrClass 1 to 8 PwrClass M1 / M2 / M3
PwrClass 1 to 8 PwrClass M1 / M2 / M3
PwrClass 1 to 4 PwrClass M1 / M2 / M3
PwrClass 1 to 4 PwrClass M1 / M2 / M3
PwrClass 1 to 4 PwrClass M1 / M2 / M3
PwrClass M1 / M2 / M3
For details on the individual types of measurement please refer to chapter 3.
1057.3440.41 2-26 E-3
Page 94

FSE-K11 Starting the Application

3 Measurements with Option FSE-K11

Note: The descriptions of some of the functions are provided with notes pointing out that some
functions of FSE-K11 are avai lable on ly if Vector Signa l Ana lysis Opti on FSE- B7 is inst alled. These notes apply only to FSE, since FSIQ contains this option in the basic version.
3.1 Starting the Application
CONFIGURATION
MODE
SETUP
GSM BTS
ANALYZER
The application is called with the
MODE
key and the
GSM BTS ANALYZ ER
softkey. A selection menu is offered according to the specific standards with a softkey assigned to each measurement. The presettings are performed in the submenu of the
SETTINGS
softkey or in the right side
menu of each measurement. All settings can generally be performed using the softkeys.
In addition, a few hardkeys are available for performing special actions: Hardkey list, availabilities
Preset
Print
Save/Recall
Marker group (normal, delta)
Marker-to-trace
Marker info and search
Display line
Time line
RefPoint (Frequency/Time, RefLevel/RefLevelOffset)
1057.3440.42 3.1 E-3
Page 95

Menu Overview FSE-K11

3.2 Menu Overview
CONFIGURATION
MODE
SETUP
GSM BTS ANALYZER
ANALYZER
VECTOR
ANALYZER
GSM BTS ANALYZER
SETTINGS
ERROR
POWER
POWER
1)
PHASE/FREQ
CARRIER
1) VS TIME
MODULATION
SPECTRUM
TRANSIENT
SPECTRUM
SPURIOUS
BTS TEST SETTINGS
EXTERNAL
ATTEN
ARFCN /
FREQUENCY
POWER
SETTING
LIMIT MARGIN
SLOT No.
MIDAMBLE
1)
BTS TEST SETTINGS
P-GSM 900
E-GSM 900
GSM 1800
(DCS 1800)
GSM 1900
(PCS 1900)
R-GSM 900
PHASE I
PHASE II
ARFCN /
FREQUENCY
ARFCN
ARFCN
AUTOSELECT
FREQUENCY
SETTING
EXTERNAL
NOMINAL
OUTPUT PWR
POWER CLASS
STATIC PWR CTRL LEVEL
DYNAM PWR CTRL LEVEL
LIMIT/PWR
COUPLED
SIGNAL
LIMIT REF
POWER
1) only if option FSE-B7 (vector analyzer) is installed
POWER
ATTEN
POWER
SLOT NO.
SLOT NO.
SLOT NO.
AUTOSELECT
TRIGGER
FREE RUN
EXTERN
SLOPE
POS NEG
TRIGGER ADJUST
TRIGGER
DEFAULT
SETTINGS
TRIGGER ADJUST
FRAME
COARSE
FRAME
FINE
AUTO FRAME
ADJUST
SLOT
ADJUST
AUTO SLOT
ADJUST
TRIGGER
LEVEL
SLOPE
POS NEG
PHASE II+
1)
1)
1057.3440.42 3.2 E-3
Page 96
FSE-K11 Menu Overview
CONFIGURATION
MODE
SETUP
GSM BTS
ANALYZER
VECTOR
ANALYZER
GSM BTS ANALYZER
ANALYZER
SETTINGS
PHASE/FREQ
CARRIER
1) VS TIME
MODULATION
SPECTRUM
TRANSIENT
SPECTRUM
SPURIOUS
1)
ERROR
POWER
POWER
PHASE/FREQ
ERROR
SINGLE
CONTINUOUS
NO. OF BURSTS
PHASE/FREQ
ERROR
ARFCN /
FREQUENCY
POWER
SETTING
X UNIT
SYMB TIME
TRIGGER
NO. OF BURSTS
SET
MANUAL
SET TO
STANDARD
EDIT
PHASE
PEAK
PHASE
RMS
FREQUENCY
USER LIMIT ON OFF
EDIT
TRIGGER
FREE RUN
EXTERN
SLOPE
POS NEG
TRIGGER ADJUST
TRIGGER ADJUST
FRAME
COARSE
FRAME
FINE
AUTO FRAME
ADJUST
SLOT
ADJUST
AUTO SLOT
ADJUST
TRIGGER
LEVEL
SLOPE
POS NEG
POWER
SETTING
EXTERNAL
ATTEN
NOMINAL
OUTPUT PWR
POWER CLASS
STATIC PWR CTRL LEVEL
DYNAM PWR CTRL LEVEL
LIMIT/PWR
COUPLED
SIGNAL
POWER
LIMIT REF
POWER
ARFCN /
FREQUENCY
ARFCN
ARFCN
AUTOSELECT
FREQUENCY
1) only if option FS E-B7 (vector analyzer) is installed
1057.3440.42 3.3 E-3
Page 97
Menu Overview FSE-K11
CONFIGURATION
MODE
SETUP
GSM BTS ANALYZER
ANALYZER
VECTOR
ANALYZER
GSM BTS ANALYZER
SETTINGS
PHASE/FREQ
ERROR
CARRIER
POWER
1)
POWER
VS TIME
MODULATION
SPECTRUM
TRANSIENT
SPECTRUM
SPURIOUS
1)
CARRIER
POWER
MEAS MAX
OUTPUT PWR
INC STATIC
PWR LEVEL
INC DYNAM PWR LEVEL
NO. OF BURSTS
SGL MEAS
ON OFF
MEAS SGL
PWR LEVEL
STATIC PWR CTRL LEVEL
DYNAM PWR CTRL LEVEL
CLEAR SGL
RESULT TAB
SIGNAL
POWER
CARRIER
POWER
ARFCN /
FREQUENCY
POWER
SETTING
CONDITIONS
NORM EXTR
MEAS
BANDWITH
SYNC TO
MIDAMBLE
TRIGGER
1)
NO. OF BURSTS
SET
MANUAL
SET TO
STANDARD
TRIGGER
FREE RUN
EXTERN
SLOPE
POS NEG
TRIGGER ADJUST
1) only if option FS E-B7 (vector analyzer) is installed
TRIGGER ADJUST
FRAME
COARSE
FRAME
FINE
AUTO FRAME
ADJUST
SLOT
ADJUST
AUTO SLOT
ADJUST
TRIGGER
LEVEL
SLOPE
POS NEG
POWER
SETTING
EXTERNAL
ATTEN
NOMINAL
OUTPUT PWR
1)
1)
POWER CLASS
STATIC PWR CTRL LEVEL
DYNAM PWR CTRL LEVEL
LIMIT/PWR
COUPLED
SIGNAL
POWER
LIMIT REF
POWER
ARFCN /
FREQUENCY
ARFCN
ARFCN
AUTOSELECT
FREQUENCY
1057.3440.42 3.4 E-3
Page 98
FSE-K11 Menu Overview
CONFIGURATION
MODE
SETUP
GSM BTS
ANALYZER
VECTOR
ANALYZER
GSM BTS ANALYZER
ANALYZER
SETTINGS
PHASE/FREQ
CARRIER
POWER
1)
VS TIME
MODULATION
SPECTRUM
TRANSIENT
SPECTRUM
SPURIOUS
ERROR
POWER
1)
POWER
VS TIME
SINGLE
CONTINUOUS
BURST HIGH RESOLUTION
NO. OF BURSTS
FULL BURST
RISING
EDGE
FALLING
EDGE
RISING
& FALLING
POWER
VS TIME
REF MEAS
AUTO USER
ARFCN /
FREQUENCY
POWER
SETTING
X UNIT
SYMB TIME
MEAS
BANDWITH
SYNC TO
MIDAMBLE
TRIGGER
1)
NO. OF BURSTS
SET
MANUAL
SET TO
STANDARD
START
REF MEAS
EDIT
LIMIT LINE
FILTER
NEW LIMIT
LINE
EDIT LIMIT
LINE
COPY
LIMIT LINE
DELETE
LIMIT LINE
USER LIMIT ON OFF
PAGE
UP
PAGE DOWN
EDIT
TRIGGER
FREE RUN
EXTERN
SLOPE
POS NEG
TRIGGER ADJUST
TRIGGER ADJUST
FRAME
COARSE
FRAME
FINE
AUTO FRAME
ADJUST
SLOT
ADJUST
AUTO SLOT
ADJUST
TRIGGER
LEVEL
SLOPE
POS NEG
1)
1)
POWER
SETTING
EXTERNAL
ATTEN
NOMINAL
OUTPUT PWR
POWER CLASS
STATIC PWR CTRL LEVEL
DYNAM PWR CTRL LEVEL
LIMIT/PWR
COUPLED
SIGNAL
POWER
LIMIT REF
POWER
ARFCN /
FREQUENCY
ARFCN
ARFCN
AUTOSELECT
FREQUENCY
1) only if option FSE-B7 (vector analyzer) is installed
1057.3440.42 3.5 E-3
Page 99
Menu Overview FSE-K11
CONFIGURATION
MODE
SETUP
ANALYZER
VECTOR
ANALYZER
MODULATION
SPECTRUM
GSM BTS ANALYZER
SETTINGS
PHASE/FREQ
CARRIER
1)
POWER
VS TIME
MODULATION
SPECTRUM
TRANSIENT
SPECTRUM
SPURIOUS
ERROR
POWER
1)
SINGLE
FREQ SWEEP
CONTINUOUS FREQ SWEEP
START
LIST
NO. OF BURSTS
ARFCN
±1.8 MHZ
TX BAND
±1.8 MHZ /
TX BAND
RX BAND
2)
2)
3)
MODULATION
SPECTRUM
ARFCN /
FREQUENCY
POWER
SETTING
RX BAND GAIN
TX GATE
ON OFF
TRIGGER
3)
4)
GSM BTS ANALYZER
NO. OF BURSTS
SET
MANUAL
SET TO
STANDARD
EDIT
LIMIT LINE
FILTER
NEW LIMIT
LINE
EDIT LIMIT
LINE
COPY
LIMIT LINE
DELETE
LIMIT LINE
USER LIMIT ON OFF
PAGE
UP
PAGE DOWN
TRIGGER
FREE RUN
EXTERN
SLOPE
POS NEG
TRIGGER ADJUST
EDIT
TRIGGER
ADJUST
FRAME
COARSE
FRAME
FINE
AUTO FRAME
ADJUST
SLOT
ADJUST
AUTO SLOT
ADJUST
TRIGGER
LEVEL
SLOPE
POS NEG
1)
1)
POWER
SETTING
EXTERNAL
ATTEN
NOMINAL
OUTPUT PWR
POWER CLASS
STATIC PWR CTRL LEVEL
DYNAM PWR CTRL LEVEL
LIMIT/PWR
COUPLED
SIGNAL
POWER
LIMIT REF
POWER
ARFCN /
FREQUENCY
ARFCN
ARFCN
AUTOSELECT
FREQUENCY
1) only if option FSE-B7 (vector analyzer) is installed
2) not with P-GSM Phase I
3) only with Phase I
4) available with activeTX band only
1057.3440.42 3.6 E-3
Page 100
FSE-K11 Menu Overview
CONFIGURATION
MODE
SETUP
ANALYZER
VECTOR
ANALYZER
GSM BTS ANALYZER
TRANSIENT
SPECTRUM
SINGLE
FREQ SWEEP
CONTINUOUS FREQ SWEEP
START
GSM BTS ANALYZER
SETTINGS
PHASE/FREQ
CARRIER
1)
VS TIME
MODULATION
SPECTRUM
TRANSIENT
SPECTRUM
SPURIOUS
ERROR
POWER
POWER
1)
LIST
NO. OF BURSTS
TRANSIENT
SPECTRUM
ARFCN /
FREQUENCY
POWER
SETTING
BTS SFH
ON OFF
MEAS
BANDWITH
TRIGGER
EDIT
NO. OF BURSTS
SET
MANUAL
SET TO
STANDARD
LIMIT LINE
NEW LIMIT
EDIT LIMIT
LIMIT LINE
LIMIT LINE
USER LIMIT ON OFF
1) only if option FSE-B7 (vector analyzer) is installed
EDIT
FILTER
LINE
LINE
COPY
DELETE
PAGE
UP
PAGE DOWN
TRIGGER
FREE RUN
EXTERN
SLOPE
POS NEG
TRIGGER ADJUST
TRIGGER ADJUST
FRAME
COARSE
FRAME
FINE
AUTO FRAME
ADJUST
SLOT
ADJUST
AUTO SLOT
ADJUST
TRIGGER
LEVEL
SLOPE
POS NEG
1)
1)
POWER
SETTING
EXTERNAL
ATTEN
NOMINAL
OUTPUT PWR
POWER CLASS
STATIC PWR CTRL LEVEL
DYNAM PWR CTRL LEVEL
LIMIT/PWR
COUPLED
SIGNAL
POWER
LIMIT REF
POWER
ARFCN /
FREQUENCY
ARFCN
ARFCN
AUTOSELECT
FREQUENCY
1057.3440.42 3.7 E-3
Loading...