For FSE options K10 and K11 (GSM mobile station and GSM base stati on anal yzer ), many internal limit
lines are used which are deri ved from the applicable s tandards. To make these limit lines access ible to
the user, they are available as data sets. There are 6 data sets:
K10_XGSM.LIA Limit lines for mobiles, standards EGSM, PGSM and RGSM
K10_DCS.LIA Limit lines for mobiles, standards DCS1800 (GSM 1800)
K10_PCS.LIA Limit lines for mobiles, standards PCS1900 (GSM 1900)
K11_XGSM.LIA Limit lines for base stations, standards EGSM, PGSM und RGSM
K11_DCS.LIA Limit lines for base stations, standards DCS1800 (GSM 1800)
K11_PCS.LIA Limit lines for base stations, standards PCS1900 (GSM 1900)
To load the data sets, proceed as follows:
1. Copy the desired data sets (one or several) from the CD to disk or analyzer.
2. Press the hardkey RECALL, and enter path (for example A:\) via softkey EDIT PATH
(see operating manual, section 2.7.3).
3. Select menu item “Lines“ via softkey ITEMS TO RECALL
(this is the default setting; see operating manual, section 2.7.3.1).
4. Select the desired data set via softkey DATA SET LIST. This will cause all limit lines of the selected
data sets to be copied to the limit line directory of the instrument.
From the GSM software, the limit lines can be accessed and edited via the softkey EDIT in the
righthand side menus of the following measurements:
•
Power vs time
•
Modulation spectrum
•
Transient spectrum or
•
Spurious
Outside GSM measurements, the limit lines are accessible in the analyzer mode of the instrument.
Rohde & Schwarz makes every effort to keep the safety standard of its products up to date and to offer
its customers the highest possible degree of safety. Our products and the auxiliary equipment required
for them are designed and tested in accordance with the relevant safety standards. Compliance with
these standards is continuously monitored by our quality assurance system. This product has been
designed and tested in accordance with the EC Certificate of Conformity and has left the manufacturer’s
plant in a condition fully complying with safety standards. To maintain this condition and to ensure safe
operation, observe all instructions and warnings provided in this manual. If you have any questions
regarding these safety instructions, Rohde & Schwarz will be happy to answer them.
Furthermore, it is your responsibility to use the product in an appropriate manner. This product is
designed for use solely in industrial and laboratory environments or in the field and must not be used in
any way that may cause personal injury or property damage. You are responsible if the product is used
for an intention other than its designated purpose or in disregard of the manufacturer's instructions. The
manufacturer shall assume no responsibility for such use of the product.
The product is used for its designated purpose if it is used in accordance with its operating manual and
within its performance limits (see data sheet, documentation, the following safety instructions). Using
the products requires technical skills and knowledge of English. It is therefore essential that the
products be used exclusively by skilled and specialized staff or thoroughly trained personnel with the
required skills. If personal safety gear is required for using Rohde & Schwarz products, this will be
indicated at the appropriate place in the product documentation.
Observe
operating
instructions
Supply
voltage
ON/OFF
Weight
indication for
units >18 kg
Standby
indication
Symbols and safety labels
Danger of
electric
shock
Direct
current
(DC)
Warning!
Hot
surface
PE terminal Ground
Alternating
current (AC)
Direct/alternating
current (DC/AC)
Ground
terminal
Device fully
protected by
double/reinforced
insulation
Attention!
Electrostatic
sensitive
devices
1171.0000.42-02.00 Sheet 1
Page 10
Safety Instructions
Observing the safety instructions will help prevent personal injury or damage of any kind caused by
dangerous situations. Therefore, carefully read through and adhere to the following safety instructions
before putting the product into operation. It is also absolutely essential to observe the additional safety
instructions on personal safety that appear in other parts of the documentation. In these safety
instructions, the word "product" refers to all merchandise sold and distributed by Rohde & Schwarz,
including instruments, systems and all accessories.
Tags and their meaning
DANGER
WARNING
CAUTION This tag indicates a safety hazard with a low potential of risk for the user
ATTENTION
NOTE
These tags are in accordance with the standard definition for civil applications in the European
Economic Area. Definitions that deviate from the standard definition may also exist. It is therefore
essential to make sure that the tags described here are always used only in connection with the
associated documentation and the associated product. The use of tags in connection with unassociated
products or unassociated documentation can result in misinterpretations and thus contribute to personal
injury or material damage.
This tag indicates a safety hazard with a high potential of risk for the
user that can result in death or serious injuries.
This tag indicates a safety hazard with a medium potential of risk for the
user that can result in death or serious injuries.
that can result in slight or minor injuries.
This tag indicates the possibility of incorrect use that can cause damage
to the product.
This tag indicates a situation where the user should pay special attention
to operating the product but which does not lead to damage.
Basic safety instructions
1. The product may be operated only under
the operating conditions and in the
positions specified by the manufacturer. Its
ventilation must not be obstructed during
operation. Unless otherwise specified, the
following requirements apply to
Rohde & Schwarz products:
prescribed operating position is always with
the housing floor facing down, IP protection
2X, pollution severity 2, overvoltage
category 2, use only in enclosed spaces,
max. operation altitude max. 2000 m.
Unless specified otherwise in the data
sheet, a tolerance of ±10% shall apply to
the nominal voltage and of ±5% to the
nominal frequency.
2. Applicable local or national safety
regulations and rules for the prevention of
accidents must be observed in all work
performed. The product may be opened
only by authorized, specially trained
personnel. Prior to performing any work on
the product or opening the product, the
product must be disconnected from the
supply network. Any adjustments,
replacements of parts, maintenance or
repair must be carried out only by technical
personnel authorized by Rohde & Schwarz.
Only original parts may be used for
replacing parts relevant to safety (e.g.
power switches, power transformers,
fuses). A safety test must always be
performed after parts relevant to safety
have been replaced (visual inspection, PE
conductor test, insulation resistance
measurement, leakage current
measurement, functional test).
3. As with all industrially manufactured goods,
the use of substances that induce an
allergic reaction (allergens, e.g. nickel)
such as aluminum cannot be generally
excluded. If you develop an allergic
reaction (such as a skin rash, frequent
sneezing, red eyes or respiratory
difficulties), consult a physician immediately
to determine the cause.
1171.0000.42-02.00 Sheet 2
Page 11
Safety Instructions
4. If products/components are mechanically
and/or thermically processed in a manner
that goes beyond their intended use,
hazardous substances (heavy-metal dust
such as lead, beryllium, nickel) may be
released. For this reason, the product may
only be disassembled, e.g. for disposal
purposes, by specially trained personnel.
Improper disassembly may be hazardous to
your health. National waste disposal
regulations must be observed.
5. If handling the product yields hazardous
substances or fuels that must be disposed
of in a special way, e.g. coolants or engine
oils that must be replenished regularly, the
safety instructions of the manufacturer of
the hazardous substances or fuels and the
applicable regional waste disposal
regulations must be observed. Also
observe the relevant safety instructions in
the product documentation.
6. Depending on the function, certain products
such as RF radio equipment can produce
an elevated level of electromagnetic
radiation. Considering that unborn life
requires increased protection, pregnant
women should be protected by appropriate
measures. Persons with pacemakers may
also be endangered by electromagnetic
radiation. The employer is required to
assess workplaces where there is a special
risk of exposure to radiation and, if
necessary, take measures to avert the
danger.
7. Operating the products requires special
training and intense concentration. Make
certain that persons who use the products
are physically, mentally and emotionally fit
enough to handle operating the products;
otherwise injuries or material damage may
occur. It is the responsibility of the
employer to select suitable personnel for
operating the products.
8. Prior to switching on the product, it must be
ensured that the nominal voltage setting on
the product matches the nominal voltage of
the AC supply network. If a different voltage
is to be set, the power fuse of the product
may have to be changed accordingly.
9. In the case of products of safety class I with
movable power cord and connector,
operation is permitted only on sockets with
earthing contact and protective earth
connection.
10. Intentionally breaking the protective earth
connection either in the feed line or in the
product itself is not permitted. Doing so can
result in the danger of an electric shock
from the product. If extension cords or
connector strips are implemented, they
must be checked on a regular basis to
ensure that they are safe to use.
11. If the product has no power switch for
disconnection from the AC supply, the plug
of the connecting cable is regarded as the
disconnecting device. In such cases, it
must be ensured that the power plug is
easily reachable and accessible at all times
(length of connecting cable approx. 2 m).
Functional or electronic switches are not
suitable for providing disconnection from
the AC supply. If products without power
switches are integrated in racks or systems,
a disconnecting device must be provided at
the system level.
12. Never use the product if the power cable is
damaged. By taking appropriate safety
measures and carefully laying the power
cable, ensure that the cable cannot be
damaged and that no one can be hurt by
e.g. tripping over the cable or suffering an
electric shock.
13. The product may be operated only from
TN/TT supply networks fused with max.
16 A.
14. Do not insert the plug into sockets that are
dusty or dirty. Insert the plug firmly and all
the way into the socket. Otherwise this can
result in sparks, fire and/or injuries.
15. Do not overload any sockets, extension
cords or connector strips; doing so can
cause fire or electric shocks.
16. For measurements in circuits with voltages
V
> 30 V, suitable measures (e.g.
rms
appropriate measuring equipment, fusing,
current limiting, electrical separation,
insulation) should be taken to avoid any
hazards.
17. Ensure that the connections with
information technology equipment comply
with IEC 950/EN 60950.
18. Never remove the cover or part of the
housing while you are operating the
product. This will expose circuits and
components and can lead to injuries, fire or
damage to the product.
1171.0000.42-02.00 Sheet 3
Page 12
Safety Instructions
19. If a product is to be permanently installed,
the connection between the PE terminal on
site and the product's PE conductor must
be made first before any other connection
is made. The product may be installed and
connected only by a skilled electrician.
20. For permanently installed equipment
without built-in fuses, circuit breakers or
similar protective devices, the supply circuit
must be fused in such a way that suitable
protection is provided for users and
products.
21. Do not insert any objects into the openings
in the housing that are not designed for this
purpose. Never pour any liquids onto or into
the housing. This can cause short circuits
inside the product and/or electric shocks,
fire or injuries.
22. Use suitable overvoltage protection to
ensure that no overvoltage (such as that
caused by a thunderstorm) can reach the
product. Otherwise the operating personnel
will be endangered by electric shocks.
23. Rohde & Schwarz products are not
protected against penetration of water,
unless otherwise specified (see also safety
instruction 1.). If this is not taken into
account, there exists the danger of electric
shock or damage to the product, which can
also lead to personal injury.
24. Never use the product under conditions in
which condensation has formed or can form
in or on the product, e.g. if the product was
moved from a cold to a warm environment.
matching Rohde & Schwarz type (see
spare parts list). Batteries and storage
batteries are hazardous waste. Dispose of
them only in specially marked containers.
Observe local regulations regarding waste
disposal. Do not short-circuit batteries or
storage batteries.
28. Please be aware that in the event of a fire,
toxic substances (gases, liquids etc.) that
may be hazardous to your health may
escape from the product.
29. Please be aware of the weight of the
product. Be careful when moving it;
otherwise you may injure your back or other
parts of your body.
30. Do not place the product on surfaces,
vehicles, cabinets or tables that for reasons
of weight or stability are unsuitable for this
purpose. Always follow the manufacturer's
installation instructions when installing the
product and fastening it to objects or
structures (e.g. walls and shelves).
31. Handles on the products are designed
exclusively for personnel to hold or carry
the product. It is therefore not permissible
to use handles for fastening the product to
or on means of transport such as cranes,
fork lifts, wagons, etc. The user is
responsible for securely fastening the
products to or on the means of transport
and for observing the safety regulations of
the manufacturer of the means of transport.
Noncompliance can result in personal injury
or material damage.
25. Do not close any slots or openings on the
product, since they are necessary for
ventilation and prevent the product from
overheating. Do not place the product on
soft surfaces such as sofas or rugs or
inside a closed housing, unless this is well
ventilated.
26. Do not place the product on heatgenerating devices such as radiators or fan
heaters. The temperature of the
environment must not exceed the maximum
temperature specified in the data sheet.
27. Batteries and storage batteries must not be
exposed to high temperatures or fire. Keep
batteries and storage batteries away from
children. If batteries or storage batteries are
improperly replaced, this can cause an
explosion (warning: lithium cells). Replace
the battery or storage battery only with the
1171.0000.42-02.00 Sheet 4
32. If you use the product in a vehicle, it is the
sole responsibility of the driver to drive the
vehicle safely. Adequately secure the
product in the vehicle to prevent injuries or
other damage in the event of an accident.
Never use the product in a moving vehicle if
doing so could distract the driver of the
vehicle. The driver is always responsible for
the safety of the vehicle; the manufacturer
assumes no responsibility for accidents or
collisions.
33. If a laser product (e.g. a CD/DVD drive) is
integrated in a Rohde & Schwarz product,
do not use any other settings or functions
than those described in the documentation.
Otherwise this may be hazardous to your
health, since the laser beam can cause
irreversible damage to your eyes. Never try
to take such products apart, and never look
into the laser beam.
Page 13
Por favor lea imprescindiblemente antes de
la primera puesta en funcionamiento las
siguientes informaciones de seguridad
Informaciones de seguridad
Es el principio de Rohde & Schwarz de tener a sus productos siempre al día con los estandards de
seguridad y de ofrecer a sus clientes el máximo grado de seguridad. Nuestros productos y todos los
equipos adicionales son siempre fabricados y examinados según las normas de seguridad vigentes.
Nuestra sección de gestión de la seguridad de calidad controla constantemente que sean cumplidas
estas normas. Este producto ha sido fabricado y examinado según el comprobante de conformidad
adjunto según las normas de la CE y ha salido de nuestra planta en estado impecable según los
estandards técnicos de seguridad. Para poder preservar este estado y garantizar un funcionamiento
libre de peligros, deberá el usuario atenerse a todas las informaciones, informaciones de seguridad y
notas de alerta. Rohde&Schwarz está siempre a su disposición en caso de que tengan preguntas
referentes a estas informaciones de seguridad.
Además queda en la responsabilidad del usuario utilizar el producto en la forma debida. Este producto
solamente fue elaborado para ser utilizado en la indústria y el laboratorio o para fines de campo y de
ninguna manera deberá ser utilizado de modo que alguna persona/cosa pueda ser dañada. El uso del
producto fuera de sus fines definidos o despreciando las informaciones de seguridad del fabricante
queda en la responsabilidad del usuario. El fabricante no se hace en ninguna forma responsable de
consecuencias a causa del maluso del producto.
Se parte del uso correcto del producto para los fines definidos si el producto es utilizado dentro de las
instrucciones del correspondiente manual del uso y dentro del margen de rendimiento definido (ver
hoja de datos, documentación, informaciones de seguridad que siguen). El uso de los productos hace
necesarios conocimientos profundos y el conocimiento del idioma inglés. Por eso se deberá tener en
cuenta de exclusivamente autorizar para el uso de los productos a personas péritas o debidamente
minuciosamente instruidas con los conocimientos citados. Si fuera necesaria indumentaria de
seguridad para el uso de productos de R&S, encontrará la información debida en la documentación del
producto en el capítulo correspondiente.
Símbolos y definiciones de seguridad
Ver manual
de
instrucciones
del uso
Informaciones
para
maquinaria
con uns peso
de > 18kg
Peligro de
golpe de
corriente
¡Advertencia!
Superficie
caliente
Conexión a
conductor
protector
Conexión
a tierra
Conexión
a masa
conductora
¡Cuidado!
Elementos de
construción
con peligro de
carga
electroestática
El aparato está
protegido en su
totalidad por un
aislamiento de
doble refuerzo
potencia EN
MARCHA/PARADA
Indicación
Stand-by
Corriente
continua
DC
Corriente
alterna AC
Corriente
continua/alterna
DC/AC
1171.0000.42-02.00 página 1
Page 14
Informaciones de seguridad
Tener en cuenta las informaciones de seguridad sirve para tratar de evitar daños y peligros de toda
clase. Es necesario de que se lean las siguientes informaciones de seguridad concienzudamente y se
tengan en cuenta debidamente antes de la puesta en funcionamiento del producto. También deberán
ser tenidas en cuenta las informaciones para la protección de personas que encontrarán en otro
capítulo de esta documentación y que también son obligatorias de seguir. En las informaciones de
seguridad actuales hemos juntado todos los objetos vendidos por Rohde&Schwarz bajo la
denominación de „producto“, entre ellos también aparatos, instalaciones así como toda clase de
accesorios.
Palabras de señal y su significado
PELIGRO Indica un punto de peligro con gran potencial de riesgo para el
ususario.Punto de peligro que puede llevar hasta la muerte o graves
heridas.
ADVERTENCIA Indica un punto de peligro con un protencial de riesgo mediano para el
usuario. Punto de peligro que puede llevar hasta la muerte o graves
heridas .
ATENCIÓN Indica un punto de peligro con un protencial de riesgo pequeño para el
usuario. Punto de peligro que puede llevar hasta heridas leves o
pequeñas
CUIDADO Indica la posibilidad de utilizar mal el producto y a consecuencia
dañarlo.
INFORMACIÓN Indica una situación en la que deberían seguirse las instrucciones en el
uso del producto, pero que no consecuentemente deben de llevar a un
daño del mismo.
Las palabras de señal corresponden a la definición habitual para aplicaciones civiles en el ámbito de la
comunidad económica europea. Pueden existir definiciones diferentes a esta definición. Por eso se
debera tener en cuenta que las palabras de señal aquí descritas sean utilizadas siempre solamente en
combinación con la correspondiente documentación y solamente en combinación con el producto
correspondiente. La utilización de las palabras de señal en combinación con productos o
documentaciones que no les correspondan puede llevar a malinterpretaciones y tener por
consecuencia daños en personas u objetos.
Informaciones de seguridad elementales
1. El producto solamente debe ser utilizado
según lo indicado por el fabricante referente
a la situación y posición de funcionamiento
sin que se obstruya la ventilación. Si no se
convino de otra manera, es para los
productos R&S válido lo que sigue:
como posición de funcionamiento se define
principialmente la posición con el suelo de la
caja para abajo , modo de protección IP 2X,
grado de suciedad 2, categoría de
sobrecarga eléctrica 2, utilizar solamente en
estancias interiores, utilización hasta 2000 m
sobre el nivel del mar.
A menos que se especifique otra cosa en la
hoja de datos, se aplicará una tolerancia de
±10% sobre el voltaje nominal y de ±5%
sobre la frecuencia nominal.
2. En todos los trabajos deberán ser tenidas en
cuenta las normas locales de seguridad de
trabajo y de prevención de accidentes. El
producto solamente debe de ser abierto por
personal périto autorizado. Antes de efectuar
trabajos en el producto o abrirlo deberá este
ser desconectado de la corriente. El ajuste,
el cambio de partes, la manutención y la
reparación deberán ser solamente
efectuadas por electricistas autorizados por
R&S. Si se reponen partes con importancia
para los aspectos de seguridad (por ejemplo
el enchufe, los transformadores o los
fusibles), solamente podrán ser sustituidos
por partes originales. Despues de cada
recambio de partes elementales para la
seguridad deberá ser efectuado un control de
1171.0000.42-02.00 página 2
Page 15
Informaciones de seguridad
seguridad (control a primera vista, control de
conductor protector, medición de resistencia
de aislamiento, medición de medición de la
corriente conductora, control de
funcionamiento).
3. Como en todo producto de fabricación
industrial no puede ser excluido en general
de que se produzcan al usarlo elementos
que puedan generar alergias, los llamados
elementos alergénicos (por ejemplo el
níquel). Si se producieran en el trato con
productos R&S reacciones alérgicas, como
por ejemplo urticaria, estornudos frecuentes,
irritación de la conjuntiva o dificultades al
respirar, se deberá consultar inmediatamente
a un médico para averigurar los motivos de
estas reacciones.
4. Si productos / elementos de construcción son
tratados fuera del funcionamiento definido de
forma mecánica o térmica, pueden generarse
elementos peligrosos (polvos de sustancia
de metales pesados como por ejemplo
plomo, berilio, níquel). La partición elemental
del producto, como por ejemplo sucede en el
tratamiento de materias residuales, debe de
ser efectuada solamente por personal
especializado para estos tratamientos. La
partición elemental efectuada
inadecuadamente puede generar daños para
la salud. Se deben tener en cuenta las
directivas nacionales referentes al
tratamiento de materias residuales.
5. En el caso de que se produjeran agentes de
peligro o combustibles en la aplicación del
producto que debieran de ser transferidos a
un tratamiento de materias residuales, como
por ejemplo agentes refrigerantes que deben
ser repuestos en periodos definidos, o
aceites para motores, deberan ser tenidas en
cuenta las prescripciones de seguridad del
fabricante de estos agentes de peligro o
combustibles y las regulaciones regionales
para el tratamiento de materias residuales.
Cuiden también de tener en cuenta en caso
dado las prescripciones de seguridad
especiales en la descripción del producto.
6. Ciertos productos, como por ejemplo las
instalaciones de radiación HF, pueden a
causa de su función natural, emitir una
radiación electromagnética aumentada. En
vista a la protección de la vida en desarrollo
deberían ser protegidas personas
embarazadas debidamente. También las
personas con un bypass pueden correr
peligro a causa de la radiación
electromagnética. El empresario está
comprometido a valorar y señalar areas de
trabajo en las que se corra un riesgo de
exposición a radiaciones aumentadas de
riesgo aumentado para evitar riesgos.
7. La utilización de los productos requiere
instrucciones especiales y una alta
concentración en el manejo. Debe de
ponerse por seguro de que las personas que
manejen los productos estén a la altura de
los requerimientos necesarios referente a
sus aptitudes físicas, psíquicas y
emocionales, ya que de otra manera no se
pueden excluir lesiones o daños de objetos.
El empresario lleva la responsabilidad de
seleccionar el personal usuario apto para el
manejo de los productos.
8. Antes de la puesta en marcha del producto
se deberá tener por seguro de que la tensión
preseleccionada en el producto equivalga a
la del la red de distribución. Si es necesario
cambiar la preselección de la tensión
también se deberán en caso dabo cambiar
los fusibles correspondientes del prodcuto.
9. Productos de la clase de seguridad I con
alimentación móvil y enchufe individual de
producto solamente deberán ser conectados
para el funcionamiento a tomas de corriente
de contacto de seguridad y con conductor
protector conectado.
10. Queda prohibida toda clase de interrupción
intencionada del conductor protector, tanto
en la toma de corriente como en el mismo
producto ya que puede tener como
consecuencia el peligro de golpe de corriente
por el producto. Si se utilizaran cables o
enchufes de extensión se deberá poner al
seguro, que es controlado su estado técnico
de seguridad.
11. Si el producto no está equipado con un
interruptor para desconectarlo de la red, se
deberá considerar el enchufe del cable de
distribución como interruptor. En estos casos
deberá asegurar de que el enchufe sea de
fácil acceso y nabejo (medida del cable de
distribución aproximadamente 2 m). Los
interruptores de función o electrónicos no
son aptos para el corte de la red eléctrica. Si
los productos sin interruptor están integrados
en construciones o instalaciones, se deberá
instalar el interruptor al nivel de la
instalación.
1171.0000.42-02.00 página 3
Page 16
Informaciones de seguridad
12. No utilice nunca el producto si está dañado el
cable eléctrico. Asegure a través de las
medidas de protección y de instalación
adecuadas de que el cable de eléctrico no
pueda ser dañado o de que nadie pueda ser
dañado por él, por ejemplo al tropezar o por
un golpe de corriente.
13. Solamente está permitido el funcionamiento
en redes de distribución TN/TT aseguradas
con fusibles de como máximo 16 A.
14. Nunca conecte el enchufe en tomas de
corriente sucias o llenas de polvo. Introduzca
el enchufe por completo y fuertemente en la
toma de corriente. Si no tiene en
consideración estas indicaciones se arriesga
a que se originen chispas, fuego y/o heridas.
15. No sobrecargue las tomas de corriente, los
cables de extensión o los enchufes de
extensión ya que esto pudiera causar fuego
o golpes de corriente.
16. En las mediciones en circuitos de corriente
con una tensión de entrada de Ueff > 30 V se
deberá tomar las precauciones debidas para
impedir cualquier peligro (por ejemplo
medios de medición adecuados, seguros,
limitación de tensión, corte protector,
aislamiento etc.).
17. En caso de conexión con aparatos de la
técnica informática se deberá tener en
cuenta que estos cumplan los requisitos de
la EC950/EN60950.
18. Nunca abra la tapa o parte de ella si el
producto está en funcionamiento. Esto pone
a descubierto los cables y componentes
eléctricos y puede causar heridas, fuego o
daños en el producto.
19. Si un producto es instalado fijamente en un
lugar, se deberá primero conectar el
conductor protector fijo con el conductor
protector del aparato antes de hacer
cualquier otra conexión. La instalación y la
conexión deberán ser efecutadas por un
electricista especializado.
20. En caso de que los productos que son
instalados fijamente en un lugar sean sin
protector implementado, autointerruptor o
similares objetos de protección, deberá la
toma de corriente estar protegida de manera
que los productos o los usuarios estén
suficientemente protegidos.
21. Por favor, no introduzca ningún objeto que
no esté destinado a ello en los orificios de la
caja del aparato. No vierta nunca ninguna
clase de líquidos sobre o en la caja. Esto
puede producir corto circuitos en el producto
y/o puede causar golpes de corriente, fuego
o heridas.
22. Asegúrese con la protección adecuada de
que no pueda originarse en el producto una
sobrecarga por ejemplo a causa de una
tormenta. Si no se verá el personal que lo
utilice expuesto al peligro de un golpe de
corriente.
23. Los productos R&S no están protegidos
contra el agua si no es que exista otra
indicación, ver también punto 1. Si no se
tiene en cuenta esto se arriesga el peligro de
golpe de corriente o de daños en el producto
lo cual también puede llevar al peligro de
personas.
24. No utilice el producto bajo condiciones en las
que pueda producirse y se hayan producido
líquidos de condensación en o dentro del
producto como por ejemplo cuando se
desplaza el producto de un lugar frío a un
lugar caliente.
25. Por favor no cierre ninguna ranura u orificio
del producto, ya que estas son necesarias
para la ventilación e impiden que el producto
se caliente demasiado. No pongan el
producto encima de materiales blandos como
por ejemplo sofás o alfombras o dentro de
una caja cerrada, si esta no está
suficientemente ventilada.
26. No ponga el producto sobre aparatos que
produzcan calor, como por ejemplo
radiadores o calentadores. La temperatura
ambiental no debe superar la temperatura
máxima especificada en la hoja de datos.
1171.0000.42-02.00 página 4
Page 17
Informaciones de seguridad
27. Baterías y acumuladores no deben de ser
expuestos a temperaturas altas o al fuego.
Guardar baterías y acumuladores fuera del
alcance de los niños. Si las baterías o los
acumuladores no son cambiados con la
debida atención existirá peligro de explosión
(atención celulas de Litio). Cambiar las
baterías o los acumuladores solamente por
los del tipo R&S correspondiente (ver lista de
piezas de recambio). Baterías y
acumuladores son deshechos problemáticos.
Por favor tirenlos en los recipientes
especiales para este fín. Por favor tengan en
cuenta las prescripciones nacionales de cada
país referente al tratamiento de deshechos.
Nunca sometan las baterías o acumuladores
a un corto circuito.
28. Tengan en consideración de que en caso de
un incendio pueden escaparse gases tóxicos
del producto, que pueden causar daños a la
salud.
29. Por favor tengan en cuenta que en caso de
un incendio pueden desprenderse del
producto agentes venenosos (gases, líquidos
etc.) que pueden generar daños a la salud.
30. No sitúe el producto encima de superficies,
vehículos, estantes o mesas, que por sus
características de peso o de estabilidad no
sean aptas para él. Siga siempre las
instrucciones de instalación del fabricante
cuando instale y asegure el producto en
objetos o estructuras (por ejemplo paredes y
estantes).
31. Las asas instaladas en los productos sirven
solamente de ayuda para el manejo que
solamente está previsto para personas. Por
eso no está permitido utilizar las asas para la
sujecion en o sobre medios de transporte
como por ejemplo grúas, carretillas
elevadoras de horquilla, carros etc. El
usuario es responsable de que los productos
sean sujetados de forma segura a los medios
de transporte y de que las prescripciones de
seguridad del fabricante de los medios de
transporte sean tenidas en cuenta. En caso
de que no se tengan en cuenta pueden
causarse daños en personas y objetos.
32. Si llega a utilizar el producto dentro de un
vehículo, queda en la responsabilidad
absoluta del conductor que conducir el
vehículo de manera segura. Asegure el
producto dentro del vehículo debidamente
para evitar en caso de un accidente las
lesiones u otra clase de daños. No utilice
nunca el producto dentro de un vehículo en
movimiento si esto pudiera distraer al
conductor. Siempre queda en la
responsabilidad absoluta del conductor la
seguridad del vehículo y el fabricante no
asumirá ninguna clase de responsabilidad
por accidentes o colisiones.
33. Dado el caso de que esté integrado un
producto de laser en un producto R&S (por
ejemplo CD/DVD-ROM) no utilice otras
instalaciones o funciones que las descritas
en la documentación. De otra manera pondrá
en peligro su salud, ya que el rayo laser
puede dañar irreversiblemente sus ojos.
Nunca trate de descomponer estos
productos. Nunca mire dentro del rayo laser.
1171.0000.42-02.00 página 5
Page 18
Page 19
FSE-K10Getting Started
1 Getting Started
This chapter is mainly intended for users operating the GSM MS analyzer mode for the first tim e. User s
are guided step by step through the most frequent tests that are performed on mobiles.
User guidance starts with connecting FSE/FSIQ and making the required presettings before the
individual measurements are explained.
The layout and contents of this chapter are applic ation-oriented and provide important inform ation on
FSE/FSIQ as well as the GSM900/DCS1800/PCS1900 system.
It is recommended to read through this chapter and at the same time perform the operations on
FSE/FSIQ. This ensures a simple and rapid introduction to the GSM MS firmware functions so that
correct and useful measurement results are achieved.
Each individual step is provided with a reference number, eg , relating to a more detailed inform ation
on the opposite page.
General information on FSE-K10 is provided in chapter 2. Chapter 3 gives detailed information on
menus and softkeys and contains a graphical menu overview. Remote-control commands and their
assignment to softkeys is described in chapter 4.
Note:The descriptions of some of the functions are pr ovided with a note to the effect that thes e
functions are available only with Vector Signal Analysis Option FSE-B7 installed. This note
applies only to FSE, while FSIQ contains these functions already in the basic version.
1057.3140.421.1E-2
Page 20
Connecting the FSEFSE-K10
1.1 Connecting the FSE/FSIQ
1.1.1 Standard Test Setup
In general, a system simulator (eg CRTP or also CMD from R&S) is required to test the m obile. This
system simulator makes the necessary measurement-specific settings and provides the necessary
frequency accuracy of the mobile.
The mobile has to be equipped with a test SIM.
If the mobile has a special service mode, most of the measurements can be performed without a
system simulator (exception: frequency errors. To attain the required frequency accuracy, the mobile
needs a corresponding synchronization burst).
Ext. reference signal
CMD 55/65
mit Option CMD-B3
FSE
Coupler
1
2
DEF3GHI
ABC
564
8
7
STU
ÜVW
.-0
SCRCLM
Multifunction connector
Pin9
Mobile
9
XYZ
TX / RX
External trigger
Imax
30 dBm!
.
Fig. 1-1Measurement setup with CMD 55/65 as system simulator for all measurements except
modulation spectrum RX band
1057.3140.421.2E-2
Page 21
FSE-K10Connecting the FSE/FSIQ
The CMD must be fitted with Option CMD-B3 (external reference frequency input).
À Connect frame trigger output of CMD65 with rear BNC connector
or gate input) of FSE/FSIQ.
With the aid of this external trigger signal FSE/FSIQ is able to exactly trigger even in case of
measurements outside the carrier (eg spectrum due to modulation).
À Connect antenna output (or TX output) of MS to FSE/FSIQ RF input via a coupler of suitable
attenuation (eg 20 dB).
For measurements of the spectrum due to modulation in the RX band, the attenuation between the
mobile’s antenna output and the FSE/FSIQ input must not exceed 6 dB. The TX signal must be
suppressed by at least 25 dB by means of a bandstop filter.
EXT TRIG GATE
(external trigger
1057.3140.421.3E-2
Page 22
Presetting the FSE/FSIQFSE-K10
1.2 Presetting the FSE/FSIQ
CONFIG URATION
MODE
SETUP
EXT REF
FREQUENCY
REF
INT EXT
CONFIGURATION
MODE
SETUP
GSM MS
ANALYZER
SETTINGS
Step 1
Setting the reference frequency
À Press SETUP key.
À Press
EXT. REF FREQUENCY
and enter the frequency according to
the frequency standard.
À Switch
REF INT/EXT
softkey to
(external reference frequency).
Step 2
Selecting the standard
À Successively press the
the
GSM MS ANALYZER
SETTINGS
softkey and the key.
The following menu for selecting the
different standards is displayed.
MODE
softkey, the
key
EXT
key,
MS TEST
SETTINGS
P-GSM 900
E-GSM 900
GSM 1800
(DCS 1800)
GSM 1900
(PCS 1900)
R-GSM 900
PHASE I
PHASE II
PHASE II+
1057.3140.421.4E-2
Page 23
FSE-K10Presetting the FSE/FSIQ
9
Additional information
Step 1
To attain the required frequency accuracy during frequency error measurement
ERROR)
frequency n x 1 MHz , where n = 1,2, ... 16.
, FSE/FSIQ has to be supplied with a highly accurate (error< 1*10
(PHASE/FREQ
-
) external reference
1057.3140.421.5E-2
Page 24
Presetting the FSE/FSIQFSE-K10
P-GSM 900
E-GSM 900
R-GSM 900
GSM 1800
(DCS 1800)
GSM 1900
(PCS 1900)
For GSM900 and DCS1800:
PHASE I
PHASE II
PHASE II+
Step 3
Select the correct standard P-GSM900, EGSM900, DCS 1800, PCS 1900, or RGSM depending on the MS to be tested.
PHASE I/II/II+
The
sequence of measurements and the
associated limit values according to the
selected phase of the standard.
À Choose between phase I, II or II+. With
some standards, the softkeys are not
available (PCS 1900) or partly available
because no phases or not all of them
are defined.
Returns to main menu Setting after
pressing the key
softkeys changes the
.
EXTERNAL
ATTEN
Step 4
Entering external attenuation
À Press the
À Enter the attenuation of the external
attenuator used. (Recommended
attenuation see ô)
EXTERNAL ATTEN
softkey.
1057.3140.421.6E-2
Page 25
FSE-K10Presetting the FSE/FSIQ
Additional information
Step 4
ô
The following values are recomm ended for the external attenuator to ensure that the FSE/FSIQ
RF input is protected and the sensitivity of the FSE/FSIQ is not reduced too much.
The FSE/FSIQ level display is automatically corrected by the entered value.
20 dB - 25 dB
15 dB - 20 dB
10 dB - 15 dB
5 dB - 10 dB
0 dB - 5 dB
1057.3140.421.7E-2
Page 26
Presetting the FSE/FSIQFSE-K10
Step 5
Selecting the transmission channel
ARFCN /
FREQUENCY
À Press the
softkey.
ARFCN / FREQUENCY
ARFCN
ARFCN/
AUTOSELECT
FREQUENCY
À Press the
AUTOSELECT
ARFCN, ARFCN
or
FREQUENCY
softkey.
With
ARFCN
selected:
À Enter the transmit channel number of
your mobile. í
FSE/FSIQ will tune to the associated
frequency in compliance with the
selected standard.
ARFCN AUTOSELECT
With
selected,
FSE/FSIQ automatically searches for the
active transmit channel on condition that
frequency hopping is not active.
FREQUENCY
With
selected:
À Enter the transmission frequency of
your MS or any other frequency (eg the
IF of the mobile).
Permissible frequencies are:
1.8 MHz frequency (maximum
device frequency– 1.8 MHz).
À Press
to return to the Settings
menu.
1057.3140.421.8E-2
Page 27
FSE-K10Presetting the FSE/FSIQ
Additional information
Step 5
í
Enter one of the possible channel numbers 1 to 124 for the P-GSM frequency range. For the
extended GSM band, enter a number between 975 and 1023 or 0. Num bers available for PCN
(DCS1800) are 512 to 885, and for PCS1900 512 to 810. In addition to the number s of the P-GSM
band (1 to 124), channel numbers 955 to 1023 or 0 are available for R-GSM.
To protect the FSE/FSIQ, the settings m ade in the POWER m enu are automatically checked to
make sure that the input level at the FSE/FSIQ does not exceed 27 dBm.
To prevent the occurrence of an overload, a warning is displayed for those settings which would
lead to the threshold level being exceeded (
POWER)
GSM 900
phase I
R/E/P-GSM900
phases II & II+
and the set value is ignored.
DCS1800
phase I
DCS1800
phases II & II+
PCS1900
POWER CLASS, EXTERNAL ATTEN, OUTPUT MS
1057.3140.421.11E-2
Page 30
Presetting the FSE/FSIQFSE-K10
Step 7
POWER
CTRL LEVEL
or
With reduced level of the MS, enter the
power control level in addition.
À Enter the desired power control level N.
û.
LIMIT/PWR
COUPLED
SIGNAL
POWER
LIMIT LINE
REF POWER
SIGNAL POWER
only if
LIMIT/PWR COUPLED
and
LIMIT REF POWER
is inactive.
are available
The coupling between power and that of
the power class as well as between the
power actually applied and the limits
selected with respect to this power can be
suppressed (softkey
COUPLED
).
LIMIT/PWR
À In this case enter the actual output
power of MS (external attenuation is
taken into account) by pressing the
SIGNAL POWER
With the
LIMIT LINE REF POWER
softkey.
softkey, the power determining the
selection of limits can be chos en (only for
measurements with limit values that vary
with the MS power).
Step 8
Selecting the current midamble
MIDAMBLE
MIDAMBLE
TSC_0
TSC_1
TSC_2
TSC_3
TSC_4
TSC_5
TSC_6
TSC_7
À Press the
MIDAMBLE
softkey
.
A table indicating the training
sequences (midamble) is displayed on
the screen.
À Selection is performed by means of the
and keys and the ENTER key.
1057.3140.421.12E-2
Page 31
FSE-K10Presetting the FSE/FSIQ
Additional information
Step 7
û
The following power control level values may be set for the different st andards, provided power of
power class ≥ power of power control level.
The maximum selectable power control level depends on the power class of the mobile.
Example:
P_GSM900
Power class = 2 (≡ 39 dBm)
PWR CTRL Level = 2 ... 15
If an unpermissible Power Control Level is entered, FSE/FSIQ signals:
CHECK POWER CLASS
1057.3140.421.13E-2
Page 32
Presetting the FSE/FSIQFSE-K10
Step 9
TRIGGER
TRIGGER
ADJUST
TRIGGER
FREE RUN
VIDEO
EXTERN
RF POWER
SLOPE
POS NEG
VIDEO
TRIGGER
ADJUST
TRIGGER
ADJUST
COARSE
ADJUST
FINE
ADJUST
AUTO
ADJUST
Setting the trigger
À Press the
TRIGGER
softkey
.
The default setting for the trigger is
EXTERN. This default setting is
applicable to all measurements. A
corresponding trigger signal is required
(frame trigger).
The default setting for the external
trigger level ø is 1.4 V but, if required,
the level can be adapted by pressing
EXTERN
and then entering a new
value in the data entry field. The
polarity of the trigger signal can be set
using the
SLOPE POS NEG
softkey.
The default setting is the positive slope.
À Press the
TRIGGER ADJUST
softkey
to set the trigger reference. Automatic
(preferably) or manual adjustment can
be performed.
VIDEO
TRIGGER
LEVEL
SLOPE
POS NEG
1057.3140.421.14E-2
Page 33
FSE-K10Presetting the FSE/FSIQ
Additional information
Step 9
ø
Ext. frame trigger
1.4V
T1
0dB
-20dB
Burst
Transition bit 73/74 (bit 13/14 of midamble)
T1 = Reference time for delay to be adjusted with
AUTO ADJUST
1057.3140.421.15E-2
Page 34
Presetting the FSE/FSIQFSE-K10
Either automatic adjustment:
AUTO
ADJUST
COARSE
ADJUST
and
FINE
ADJUST
À Press the
AUTO ADJUST
softkey.
FSE/FSIQ sets the time reference
between the trigger signal and the
midamble and takes this reference into
account for all measurements.
or manual adjustment:
(if automatic adjustment is not possible,
eg for bursts without midam ble or if option
FSE-B7 is not installed.)
À Press the
COARSE ADJUST
softkey
and set the time reference between the
external trigger and the burst with the
aid of the spinwheel. (Set the –20 dB
point of the rising burst edge to the
display center.)
À Press th e
FINE ADJUST
softkey and
perform fine adjustment of the trigger.
ù
1057.3140.421.16E-2
Page 35
FSE-K10Presetting the FSE/FSIQ
Additional information
Step 9
ù
Fig. 1-2Correct trigger adjustment by means of FINE ADJUST
Caution:The manual adjustment to the -20 dB point of the trigger edge can deliver the correct
time reference by approximation only since the high-precision reference to the
midamble is missing and the setting thus depends on the slope of the bur st edge. The
automatic adjustment delivers the correc t time reference to the midamble (depending
on the burst edge slope the -20 dB point can be offset by up to approx. 7 µs from the
screen center).
1057.3140.421.17E-2
Page 36
Measurement of Phase/Frequency ErrorFSE-K10
1.3 Measurement of Phase/Frequency Error
Note:Measurement of phase and frequency error is possible only if option FSE-B7 (Vector Signal
Analyzer) is fitted. Otherwise, the softkey will not appear on the screen.
Step 1
PHASE/FRE
ERROR
NO. OF
BURSTS
Q
À Press the
PHASE/FREQ
softkey .
Step 2
NO. OF
BURSTS
SET
MANUAL
SET TO
STANDARD
À Press the
NO OF BURSTS
softkey and
set the number of measurements to be
carried out.
The number of bursts defined in the
selected standard is set with SET TO
STANDARD (in the phase/frequency error
measurement: 20 for all standards).
Press
to return to the SETTINGS
menu.
1057.3140.421.18E-2
Page 37
FSE-K10Measurement of Phase/Frequency Error
Additional information
MS setting:
The measurement can also be performed on MSs with active SFH (slow frequency hopping).
FSE/FSIQ measures at the set receive frequenc y (ARFCN) . If a number of burst s > 1 has been s et,
only measurement results will be considered at which FSE/FSIQ is able to synchronize. Thus even if
the carrier frequency of the slot to be measured changes cyclically in successive frames, only slots
with the set midamble at the selected ARFCN will be considered in the measurement. However, this
will considerably slow down the measurement sequence (approx. by the factor 3 against a nonhopping MS). This disadvantage can be avoided if a trigger signal is provided for the FSE/FSIQ
(triggering not on every frame but only if the selected slot is at the set ARFCN).
The MS needs to receive a suitable synchronization burst from the system simulator for the
frequency error to be measured correctly.
1057.3140.421.19E-2
Page 38
Measurement of Phase/Frequency ErrorFSE-K10
Step 3
SINGLE
and
CONTINUOUS
À Starting the measurement sequence
SINGLE
with
When the m easurement is com pleted, an
overview of the numeric m odulation errors
is displayed in result window A for the 147
useful bits. ô
or
CONTINUOUS
.
1057.3140.421.20E-2
Page 39
FSE-K10Measurement of Phase/Frequency Error
Additional information
Step 3
ô
Fig. 1-3Measurement window PFE
The information on instantaneous frequency and power and the following numerical results are
displayed in the window. It also contains the corresponding limit values and pass/fail information.
Max hold and average value of peak phase error
Max hold value and average value of rms phase error
Max hold and average value of frequency error
Window B displays the phase error versus time, ie over the 147 useful bits of the normal burst.
Three traces are displayed at the same time:
• Trace 2: Max Hold (maximum positive phase error of all measured bursts)
• Trace 3: Min Hold. (maximum negative phase error of all measured bursts)
1057.3140.421.21E-2
Page 40
Measurement of Average Carrier PowerFSE-K10
1.4 Measurement of Average Carrier Power
Step 1
CARRIER
POWER
À Press the
CARRIER POWER
softkey
Step 2
NO. OF
BURSTS
and
SYNC T O
MIDAMBLE
NO. OF
BURSTS
SET
MANUAL
SET TO
STANDARD
À Press the
NO OF BURSTS
softkey and
set the number of measurements to be
carried out.
The number of bursts defined in the
selected standard is set with
STANDARD
(in the carrier power
SET TO
measurement): 1 for all standards.
Step 3
Possible only if option FSE-B7 is
installed:
In the default setting, softkey
MIDAMBLE
is active and thus the
synchronization to the burst midamble is
set. This setting is of advantage if the burs t
to be measured contains a midamble and a
bit-accurate time reference is required.
(Softkey colour: green)
À Disable
SYNC TO MIDAMBLE
bursts without midamble or if maximum
measurement speed is required.
(Softkey colour: grey) ô
SYNC TO
for
Step 4
MEAS
BANDWITH
1057.3140.421.22E-2
Setting the measurement bandwidth
À Press the
MEAS BANDWITH
softkey
A table indicating the settable
bandwidths is displayed on the screen.
Selecting STANDARD (default setting)
sets the measurement bandwidth
specified in the standard.
Or alternatively:
Select the measurement bandwidth 300
kHz or 1 MHz using the up/down keys
and press the Enter key.
.
Page 41
FSE-K10Measurement of Average Carrier Power
Additional information
MS setting:
The measurement may also be performed for slow frequency hopping of the MS, provided the
option FSE-B7 is installed and SYNC TO MIDAMBLE is active on FSE/FSIQ.
For faster measurement, switch off SFH on the MS.
Step 3
ô
Explanation of SYNC TO MIDAMBLE
If option FSE-B7 is built in and the softkey is active, FSE/FSIQ uses the vector signal analyzer to
synchronize to the midamble. Power measurement is performed in the scalar spectrum analyzer
mode by means of the time reference obtained by the signal demodulation.
Thus an accurate and updated tim e reference is ensured (even with a jittered trigger puls e). If the
softkey is not active, the vector analyzer mode is off and FSE/FSIQ uses the time reference set
using ADJUST TRIGGER.
1057.3140.421.23E-2
Page 42
Measurement of Average Carrier PowerFSE-K10
Step 5
CONDITIONS
NORM EXTR
and
INC PWR
CTRL LEVEL
MEAS MAX
OUTPUT PWR
À The tolerance range specified in the
standards for the measurement is
selected with the
NORM/EXTR
softkey and thus adapted
CONDITIONS
to the measurement conditions.
Default setting is NORM.
Step 6
À Press the key.
À Press the
softkey
MEAS MAX OUTPUT PWR
.
The reference level is determined at
maximum mobile power and the
permissible tolerance against the level
entered with the
softkey in the
POWER CLASS
MS TEST SETTINGS
menu.
Step 7
À The average carrier power is measured
at incremented power control level and
check for compliance with limit values
(see step 8). To do so, the MS level
must be reduced by one step each time
in interactive mode and softkey
STAT PWR CTRL
has then to be
pressed for the measurement.í
INCR
INC PWR
CTRL LEVEL
À If this has been done for all power
control levels, switchover from splitscreen to full-screen table is made by
pressing the
INC PWR CTRL LEVEL
softkey again.
Or alternatively step 8 (see text in the
following)
1057.3140.421.24E-2
Page 43
FSE-K10Measurement of Average Carrier Power
Additional information
Step 7
í
Upon completion of the measurement, a results table similar to the one below is displayed.
1057.3140.421.25E-2
Page 44
Measurement of Average Carrier PowerFSE-K10
Step 8
SGL MEAS
ON OFF
MEAS SGL
PWR LEVEL
POWER CTRL
LEVEL
CLEAR SGL
RESULT TAB
With the procedure as described for
step 7, many power control levels have to
be gone through even if only one power
control level is to be measured.
If single power control levels are to be
measured, proceed as follows:
À Switch softkey
SGL MEAS
to
ON.
À Select the power control level to be
measured by pressing the
CTRL LEVEL
À Press the
softkey.
MEAS SGL PWR LEVEL
POWER
softkey.
The power measurement for this power
control level is started.
The results obtained can be cleared from
the table using softkey
RESULT TAB
.
CLEAR SGL
The entries to be examined can be
selected with the cursor keys and the
spinwheel.
1057.3140.421.26E-2
Page 45
FSE-K10Measurement of Average Carrier Power
This page is left intentionally empty to ensure that the user’s notes are always opposite to the additional
information.
1057.3140.421.27E-2
Page 46
Measurement of Carrier Power versus TimeFSE-K10
1.5 Measurement of Carrier Power versus Time
Step 1
POWER
VS TIME
À Press the
POWER VS TIME
softkey
Step 2
NO. OF
BURSTS
and
SYNC T O
MIDAMBLE
NO. OF
BURSTS
SET
MANUAL
SET TO
STANDARD
À Press the
NO OF BURSTS
softkey and
set the number of measurements to be
carried out.
The number of bursts defined in the
selected standard is set with SET TO
STANDARD
(1 for all standards).
Step 3
Possible only if option FSE-B7 is
installed:
In the default setting, softkey
MIDAMBLE
is active and thus the
synchronization to the burst midamble is
set. This setting is of advantage if the burs t
to be measured contains a midamble and a
bit-accurate time reference is required.
À Disable
SYNC TO MIDAMBLE
maximum for bursts without midamble
or if maximum measurement speed is
required. ô
SYNC TO
for
Step 4
MEAS
BANDWITH
1057.3140.421.28E-2
Setting the measurement bandwidth
À Press the
MEAS BANDWITH
softkey
A table indicating the settable
bandwidths is displayed on the screen.
Selecting STANDARD (default setting)
sets the measurement bandwidth
specified in the standard.
Or alternatively:
Select the measurement bandwidth 300
kHz or 1 MHz using the up/down keys
and press the Enter key.
.
Page 47
FSE-K10Measurement of Carrier Power versus Time
Additional information
MS setting:
The measurement may also be performed for slow frequency hopping of the MS, provided the
option FSE-B7 is installed and
For fast measurement, switch off SFH on the MS.
SYNC TO MIDAMBLE
is active on FSE/FSIQ.
Step 3
ô
Explanation of SYNC TO MIDAMBLE
If option FSE-B7 is built in and the softkey is active, FSE/FSIQ uses the vector signal analyzer to
synchronize to the midamble. Power measurement is performed in the scalar spectrum analyzer
mode by means of the time reference obtained by the signal demodulation.
Thus an accurate and updated tim e reference is ensured (even with a jittered trigger puls e). If the
softkey is not active, the vector analyzer mode is off and FSE/FSIQ uses the time reference set
using ADJUST TRIGGER.
1057.3140.421.29E-2
Page 48
Measurement of Carrier Power versus TimeFSE-K10
Step 5
REF MEAS
AUTO USER
START
REF MEAS
and
FULL BURST
RISING
EDGE
oror
or
HIGH
RESOLUTION
FALLING
EDGE
À Press
REF MEAS AUTO / USER
to
select if a preview is to be performed
prior to each measurement for the
correct level adjustment of FSE/FSIQ
or if the preview is to be started
exclusively by the user using
REF MEAS
Default setting is
.
REF MEAS AUTO
START
.
Step 6
À Select the resolution for the display of
bursts.
•
FULL BURST
displayed.
•
HIGH RESOLUTION
useful bits are displayed with 1 dB
level resolution.
•
RISING EDGE: Display of the ris ing
burst edge with high time resolution.
•
FALLING EDGE: Display of the
falling burst edge with high time
resolution.
: The complete burst is
: The 147
Step 7
À To start the measurement, press the
SINGLE
í
The burst is displayed as selected in
step 6.
or the
CONTINUOUS
softkey.
1057.3140.421.30E-2
Page 49
FSE-K10Measurement of Carrier Power versus Time
T
m
m
Additional information
Step 7
í
30
20
10
0
-10
-20
-30
-40
-50
-60
34 dBm
34 dBm
22.4 dBm
22.4 dBm
1MAX
2MIN
3AVG
XG_MN2
RBW 1 MHzMax/Ref Lvl
VBW 1 MHz
SWT 800 Îs
LIMIT CHECK : PASSED
RF Att 20 dB
Mixer-20 dB
UnitdB
A
LN
TRG
1SA
2SA
3SA
EX
Date: 6.JUL.1999 15:37:04
Fig. 1-4Display of a
FULL BURST
80 Îs/Center 890.2 MHz
1057.3140.421.31E-2
Page 50
Measurement of Spectrum due to ModulationFSE-K10
1.6 M easur ement of Spectrum due to Modulation
If SFH is activated, the center SFH channel prescribed by the standard is set automatically in FSE/FSIQ.
If another frequency is to be used for the measurement, it can be set in the frequency menu (see
chapter 3, Measurements with Option FSE-K10).
Note:The measurement setup is described in chapter 1.1 and, in more detail, in chapter 3
(Measurements with Option FSE-K10).
Step 1
MODULATION
SPECTRUM
NO. OF
BURSTS
ARFCN
±1.8MHz
ARFCN ±/
TX BAND
r
o
RX BAND
GSM 900
r
o
RX BAND
NO. OF
BURSTS
SET
MANUAL
SET TO
STANDARD
TX BAND
r
o
RX BAND
RX BAND
DCS1800
withPCS 1900
r
o
withGSM90
DCS1800
À Press the MODULATION SPECTRUM
softkey. ô
Step 2
À Press the
set the number of measurements to be
carried out.
The number of bursts defined in the
selected standard is set with
STANDARD
This depends on the selected frequency
band. í
NO OF BURSTS
.
softkey and
SET TO
Step 3
÷
À Select the frequency range for the
measurement:
ARFCN ±1.8 MHz, TX BAND, RX
BAND
or in the combined band
ARFCN ±1.8 MHz/TX band.
0/
For GSM900 phase I, the m easurement in
TX band
measurement
carried out
as well as the combined
ARFCN +/TX band
.
is not
Caution:Change of measurement
setup for measurements in
RX band.
1057.3140.421.32E-2
Page 51
FSE-K10Measurement of Spectrum due to Modulation
Additional information
MS setting:
SFH (slow frequency hopping) is to be switched off.
Step 1
ô
As prescribed, the measurement is performed (excepting the midamble) from 50 to 90 % of the
burst provided adjust trigger is set correctly in the Setting menu; step 9, ù.
If necessary, the setting can be checked and corrected by pressing the
The measurem ent can be perf orm ed as a s pectrum or , as desc ribed in the standar ds, as zero span
measurement with subsequent output of the numerical results in a table.
ADJUST TRIGGER
softkey.
Step 2
í
The preset number of bursts depends on the selected fr equency range: ARFCN±1.8 MHz, TX band
or RX band (for P-GSM900 phase 1, there is no measurement in the TX band).
No. of Bursts
ARFCN ±1.8 MHz
TX band50
RX band50
200
Step 3
÷
The stringent requirements for emission measurements in the RX band make it absolutely
necessary to provide sufficient suppress ion (max. carr ier level <-10 dBm at the FSE/FSIQ input) of
the carrier signal, eg by means of a bandstop filter.
With RX BAND selected, FSE/FSIQ is set to 0 dB attenuation to achieve maximum sensitivity for the
measurement.
RX BAND GAIN
The
negative values, the insertion loss of the diplexer and bandstop filter connected ahead of the
analyzer), which is then taken into account in the FSE/FSIQ display. A preamplifier is norm ally not
needed.
The attenuation of an external attenuator entered under
measurements.
Caution: The level of +20 dBm at the FSE/FSIQ input may under no circumstances be exceeded
with 0 dB attenuation at the input attenuator.
softkey is for defining the gain of the external preamplifier (or, in case of
SETTING
is ignored in RX band
A warning to this effect is output when the
RX BAND
softkey is activated:
Caution:
Ext. attenuation ignored, disconnect ext. attenuator.
Connect bandpass or bandstop to suppress the TX band.
1057.3140.421.33E-2
Page 52
Measurement of Spectrum due to ModulationFSE-K10
Step 4
SINGLE
FREQ SWEEP
START
LIST
or
CONT
FREQ SWEEP
À Start the spectral measurement either
SINGLE FREQ SWEEP
with
FREQ SWEEP
.û
or
CONT
Step 5
À Start the non-conforming measurement
in the zero span by means of
LIST
.
ø
START
1057.3140.421.34E-2
Page 53
FSE-K10Measurement of Spectrum due to Modulation
Additional information
Step 4
û
Spectral display is not possible in the combined band ARFCN ± 1.8 MHz/TX BAND (softkeys
SINGLE FREQ SWEEP, CONT FREQ SWEEP
40
30
20
10
0
-10
-20
-30
-40
are not active).
-50
-60
Fig. 1-5Spectral display of spectrum due to modulation (
ARFCN ± 1.8 MHz range
Step 5
ø
SINGLE FREQ SWEEP
) in the
Fig. 1-6 Displayed results for zero span measurem ent (START LIST) in the ARFCN ±1.8 MHz
range
1057.3140.421.35E-2
Page 54
Measurement of Spectrum due to TransientsFSE-K10
1.7 Measurement of Spectrum due to Transients
Note:The measurement setup is described in chapter 1.1 and, in more detail, in chapter 3
(Measurements with Option FSE-K10).
Step 1
TRANSIENT
SPECTRUM
NO. OF
BURSTS
MEAS
BANDWITH
NO. OF
BURSTS
SET
MANUAL
SET TO
STANDARD
À Press the
TRANSIENT SPECTRUM
softkey.
Step 2
À Press the
set the number of measurements to be
carried out.
The number of bursts defined in the
selected standard is set with
STANDARD
NO OF BURSTS
.
softkey and
SET TO
Step 3
Setting the measurement bandwidth
À Press the
MEAS BANDWITH
A table indicating the settable
bandwidths is displayed on the screen.
Selecting STANDARD (default setting)
sets the measurement bandwidth
specified in the standard.
Or alternatively:
Select the measurement bandwidth 300
kHz or 1 MHz using the up/down keys
and press the Enter key.
softkey
.
Step 4
SINGLE
FREQ SWEEP
or
or
START
LIST
1057.3140.421.36E-2
CONT
FREQ SWEEP
ô
À Press the
CONT FREQ SWEEP
SINGLE FREQ SWEEP
softkey. í
The spectral measurement will be
started.
Alternatively:
À Press the
START LIST
softkey to start
the measurement in zero span ÷.
or
Page 55
FSE-K10Measurement of Spectrum due to Transients
Additional information
MS setting:
The measurem ent can be performed with or without SFH (slow frequenc y hopping). The FSE/FSIQ
test routines are devised to suit both cases.
Step 4
ô
The measurem ent can be performed as an overview frequency sweep showing the spectrum or as
described in the standards, ie as zero span m easurement with s ubsequent output of the numerical
results in a table. In both cases, the maximum of fset frequency of ±1.8 MHz marks the end of the
measurement.
í
Fig. 1-7 Spectrum due to transients of a GSM mobile station measured in a frequency sweep. Limit
lines and pass/fail information is shown.
÷
Fig. 1-8 Output of the spectrum due to transient in the list m ode. Measurement results, lim its and
pass/fail information for the offset frequencies stipulated in the standards as well as a
general pass/fail information are shown.
1057.3140.421.37E-2
Page 56
Measurement of Spurious EmissionsFSE-K10
1.8 Measurement of Spurious Emissions
Note: The measurement setup is described in chapter 1.1 and, in more detail, in chapter 3
(Measurements with Option FSE-K10).
Step 1
SPURIOUS
SPURIOUS
START LIST
SGL STEP
CONT SGL
STEP LIST
START
LIST
SWEEP
COUNT
TX BAND
<>TX BAND
IDLE
MODE
TX BAND
±2.0 MHZ
SELECT
STEP
À Press the
SPURIOUS
softkey.
SWEEP
COUNT
Step 2
À Press
SWEEP COUNT
to define the
number of sweeps that is to serve as a
basis for averaging.
1057.3140.421.38E-2
Page 57
FSE-K10Measurement of Spurious Emissions
Additional information
MS setting:
SFH can be used without restriction. However, it is advisable to switch off SFH of MS if the hopping
frequencies are close to the band limits.
1057.3140.421.39E-2
Page 58
Measurement of Spurious EmissionsFSE-K10
Step 3
TX BAND
IDLE
MODE
or
or
<> TX BAND
TX BAND
±2.0 MHz
or
À Determine the measurement range with
TX BAND, <> TX BAND
2.0 MHz
.
À With the aid of the
or
IDLE
softkey, you
TX BAND
may also choose whether you wish to
perform the measurement in the MS
idle mode or in the "MS allocated a
channel" mode.
±
6(/(&7
67(3
TX SUPPR
ON OFF
À Other subranges can be selected or
excluded in the used band using the
SELECT STEP
softkey.
Measurements in <> TX band:
For harmonics measurements, a carrier
suppression of at least 25 dB is requir ed
to avoid the measurement of harmonics
generated by FSE/FSIQ. The suppression
can be performed using a highpass filter.
In this case, switch also to TX SUPPR
ON. FSE/FSIQ then increases its
sensitivity by 20 dB (reference level
reduced by 20 dB).
Note:For measurements in the
IDLE mode, the mobile has
to be set continuously to the
PAGING
REORGANIZATION mode
and BS_AG_BLKS to 0.
ô
Step 4
SWEEPTIME
STD AUTO
À Select between two settings for the
analyzer sweep time using
SWEEPTIME
. The sweep time (very
long in some cases) required by the
selected standard can be set with
STD
The minimum sweep time of the
analyzer is set with
AUTO
.
1057.3140.421.40E-2
.
Page 59
FSE-K10Measurement of Spurious Emissions
Additional information
Step 3
ô
The paging reorganization mode is to be activated.
Note:In the IDLE mode, sweep times are extremely long (up to 4700 s depending on the
frequency range) because in this mode the rec eption of a paging TDMA frame by the
mobile must be ensured for each measurement point (measurement frequency).
1057.3140.421.41E-2
Page 60
Measurement of Spurious EmissionsFSE-K10
Step 5
START
LIST
Alternatively:
START LI ST
SGL STEP
and
CONT SGL
STEP LIS T
À If a complete measurement is to be
performed in the selected band (TX,
<>TX, TX BAND ±2.0 MHz) or in the
IDLE mode, press the
START LIST
softkey.
À Alternatively:
À For a
step-by-step
measurement of
the individual frequency bands, press at
first the
softkey and then
START LIST SGL STEP
CONT SGL LIST
for
each further measurement.
After the measurement has been
completed, a tabulated overview of results
with limits and pass/fail information is
displayed. í
In addition to the general information at the top of the table including the general PASS/FAIL
information, all frequency ranges, the m aximum level measured in the range and the applicable
range limit plus status information are listed.
A frequency range is defined by an identical RBW/VBW setting of the FSE/FSIQ and the same
limit.
EXT
If the margin or the limit of a frequency range is exceeded, spurious frequencies with levels are
listed in addition (with numbers).
1057.3140.421.43E-2
Page 62
General InformationFSE-K10
2 General Information on Option FSE-K10, GSM
MS Analyzer
Option FSE-K10 allo ws th e following measurements to be performed on GSM90 0 ( P-G SM , E-GSM and
R-GSM), DCS1800 or PCS1900 mobiles in line with standards (GSM 11.10, GSM 11.10-DCS,
GSM11.10-1 and J-STD-007 Air Interface) in a very convenient way:
•
PFEPhase/frequency error
•
CPWCarrier power
•
CPICarrier power individual
•
PVTPower versus tim e
•
MODModulation spectrum
•
TRATransient spectrum
•
SPUSpurious emissions
This chapter gives basic information on the sequence of operations, settings and messages of the
instrument. Details on the individual measurem ents will be found in chapter 3. T he information given
under "General Rem arks" is not needed for normal test oper ation. It is intended to p rovide a better
understanding of instrument settings and operations.
The option is call ed with the
offered according to the specific standards with a softkey assigned to each measurement. The
presettings are perf ormed in the subm enu of the
measurement.
Cold Start
Warm Start
Entering FSE-K10
Changing within FSE-K11
MODE
key and the
A complete reset to default values is made.
The previous state is assumed to a large extent. This applies to
parameter values as well as to the selec ted meas urement. The settings
specific to the type of measurement and band are made. The default
measurement t ype (list/single) is se lected. The result hist ory is cleare d,
so the results of pre vious m easurements are no longer availab le after a
warm start !
Re-entering
Action: Exiting FSE- K 1 0 v ia t he
mode, re-entering FSE-K10.
The procedure is the same as for the warm start exc ept for the result
history which remains stored.
Changing between FSE-K10 and FSE-K11
Action: Switching between FSE-K10 and FSE-K11 via the
The response is the same as for a cold start.
Changing the standard
Each change of standard caus es a complete r eset to the def ault values
valid for the standard in question. The reset is performed without any
warning to the user.
GSM MS ANALYZER
SETTINGS
softkey or in the right side m enu of each
MODE
softkey. A selection menu is
menu, changin g to no n-FSE- K1 0
MODE
menu.
Changing the phase
Changes of phase are performed with as few modifications as possible.
1057.3140.422.1E-1
Page 63
General InformationFSE-K10
If problems due to incompatible outp ut data are d etected dur ing a phase
change, the associated group of par ameters is reset to default values . In
such case a mes sage is output for a lim ited time inform ing the user on
the modifications performed.
With IEC/IEEE-bus control, the response is the same as with a change
of standard !
Possible problems and resulting actions:
Changing the
measurement type/band
Change of:
frequency group:
Level group:If the power class or the power control level d oes
There are no problems to be expected.
not exist in the new phase, the parameters
assigned to the level group are set to default
values.
Warning:"OutputPower & ExtAtten set to
Default!"
Measurement type
and band
Measurement type
When the measurement t ype is changed (eg f rom spurious to s pectrum
due to modulation), initialization actions (basic settings) specific to the
selected measur ement type are performed onc e. A default band or the
band last active is selected. List m easurement (for spurious, spectrum
due to transients, s pectrum due to modulation) or single measurem ent
(phase-frequency error, power versus time) can be selected as a
measurement type. In the case of the carrier power menu being entered,
the measurement last active is selected as measurement type.
Change of band
If the band is changed, no visible responses take place as a rule.
No other settings or initializations are made.
The procedure is the same as for a change of
measurement type (see below).
1057.3140.422.2E-1
Page 64
Sequence of Measurement OperationsFSE-K10
2.1 Sequence of Operations of FSE-K10 Measurements
Measurements performed with FSE-K10 can be divided into 2 types according to their sequence of
operations:
Measurements with and without premeasurements.
a)
b)
In the premeasurement, the currently applied signal le ve l, whic h is es sential for the main measurement,
is determined as r equ ired for the selected measur em ent. T h is app lies t o a ll r e la ti ve measurements (see
below) which are related to a refer ence value. If the measured signal level is below a m inimum value
(value expected –20 dB), the measurement is interrupted at this point and a warning is output.
Otherwise the subsequent main measurement is performed.
On activating a measurement menu, presettings are first performed which set the FSE/FSIQ to the
default state requir ed for the m easur ement type in q uestion. On s tarting the m easurem ent, the settings
required for the selec ted meas urement are car ried out. The measurem ent is then perfor med in several
consecutive steps, if necessary. At the end of the measurement, post settings (such as activating
markers) are performed and the result is displayed.
Premeasurement===>Main measurement
PreSettings[PostSettings]PreSettingsPostSettings
[sequence data]
Main measurement
PreSettingsPostSettings
[sequence data]
2.1.1 General Description of FSE-K10 Measurements
The following diagram shows the sequence of operations of a m easurement perfor med with FSE-K10.
This diagram applies to all m easurem ents ex cept for the C PW/CPI m easur ement, which is shown on a
separate diagram (see also chapter 3).
The desired measurement is selected in the main menu. General settings may be made before the
selection of the meas urement. Mos t par am eters ar e also a vaila ble f or adjus tm ents via t he s ide menu of
each measurement. On opening th e meas urement m enu, general s ettings s pecific to the m easurem ent
concerned are performed once. In the basic menu for a given measurement, the results of the
measurement last perf ormed ( table) are displa yed in the f ull-screen m ode (MOD, TR A, SPU, CPW ) on
calling up the menu. For the CPI, PVT and PF E m easurem ents, on the other hand, the s creen settings
for the main m easurement are activat ed immediatel y (for CPI the tab le cannot be dis played in the fullscreen mode). For MOD and SPU the band can be s elected. T he table conte nts change dep ending on
the selected band.
The measurement start i nitiates the (first) premeas urement. After completion of the premeasurement,
the measured signa l level is evaluat ed. In the go case, a second premeasurement is perform ed (for
PVT) with optimum sensitivity or the main measurement with level setting related to the
premeasurement. The result is displayed at the end of the measurement.
1057.3140.422.3E-1
Page 65
Sequence of Measurement OperationsFSE-K10
0
Main menu
1
General
measurement
parameters
2
3
Selection of
measurement
4
Default setting of
parameters
5
Measurement menu
Setting of parameters relevant
for the measurement
(ARFCN, ...)
Higher
menu level
10
6
Table
7
MOD/SPU
1R
9
Start
8
Optional:
band selection
<HV
10.210.1
1057.3140.422.4E-1
Page 66
Sequence of Measurement OperationsFSE-K10
Explanation of Legend:
0Entering the main menu of FSE-K10/K11 – It is possible to select general settings
(SETTINGS) or to select a measurement directly .
1General settings for GSM measurement (ARFCN, attenuation, power, trigger, ...).
2Main menu – Poss ibility of selecti ng one of the FSE-K 10/K11 m easurem ents PFE- CPW /CPI-
PVT-MOD-TRA- SPU.
3Selection of desired measurement via main menu.
4The FSE assumes the default setting provided for the selected measurement.
5, 6,7,8In the bas ic menu for a given m easurement, the res ults of the measurement last perform ed
(table) are displayed in the full-screen mode (MOD, TRA, SPU, CPW) on calling up the
menu. For the CPI, PVT and PFE m easurem ents, on the other hand, the scr een settings for
the main measurem ent are activated immediatel y (for CPI the table cannot be displa yed in
the full-screen mode). For MOD and SPU the band can be selected. The table contents
change depending on the selected band.
1057.3140.422.5E-1
Page 67
Sequence of Measurement OperationsFSE-K10
10.110.2
9
Start
11
MOD/TRA/
PVT
ja
nein
12
13.1
No
carrier
detected
No
(First) pre-
measurement
13
Carrier power
sufficient?
Yes
s. detail
14
PVT
?
Yes
15
Second
premeasurement
16
Main measurement
17
Display of
measurement trace
incl. limit lines
No
s. detail
s. detail
1057.3140.422.6E-1
Page 68
Sequence of Measurement OperationsFSE-K10
Explanation of Legend:
9Start of measurement with ’Single’, ’Continuous’ or ’StartList’.
11For measurements TRA/PVT and MOD (not Rx band) the (first) pre-measurement is
performed; for measurements PFE, C PW/CPI, MOD-Rx and SPU the main m eas urem ent is
started immediately.
12The sequence of operations of a (pre-)m easurement is presented in detail in a separate
diagram.
13Suffic ient carrier po wer: the meas ured si gnal leve l (inc l. ba ndwidt h corr ecti on) m a y be max .
20 dB below the signal level to be expected from the settings made.
13.1An error message is displayed temporarily and the IEC/IEEE status bit is set.
14, 15Following the premeasurement, a second premeasurement is performed for PVT to
accurately determine the reference.
The second premeasurement is performed with optimum sensitivity.
16Main measurement with level control orientated to premeasurement. The sequence of
operations is presented in detail in a separate chart
19The result is displayed as a table or a trace.
1057.3140.422.7E-1
Page 69
Sequence of Measurement OperationsFSE-K10
10
From main menu
100
Start
101
From main menu
"AUTO"
Yes
102
(First)
premeasurement
No
107
Determining and
saving pre-
measurement
Yes
104
No
carrier
detected
106
Premeasure-
ment USER ?
No
No
103
Carrier power
sufficient?
ja
105
Second
premeasurement
108
Main measurement
using (previous)
premeasurement
settings
109
Display of
measurement trace
incl. limit lines
1057.3140.422.8E-1
Page 70
Sequence of Measurement OperationsFSE-K10
The above chart presents the PVT measurement in detail. The sequence of operations is determined by
the setting for the premeasurement –
With the AUTO setting, the PVT measurement is performed as follows:
1st premeasurement - 2nd premeasurement - main measurement
The settings for the f irst premeas urement (with max . 2 measured bursts) are made in accor dance with
the expected level. The settings for the second premeasurement are made in accordance with the
signal level determined in the first premeasurement, and level measurement is performed with max.
20 bursts. The sig nal level determined in the second prem easurement is used as a reference for the
limit lines of the subsequent main measurement.
This two-stage proc edure results in highl y accurate level determ ination even in cases where the sign al
level is muc h belo w th e ex pec t ed va lue s ince, tha nks to the follow-up level c on t rol, i t is ens ure d that the
measurement is always performed with optimum sensitivity.
In the USER setting, the prem easurements are skipped and the le vel resulting from the most recent
premeasurement is us ed as the operating poi nt setting and also as the limit l ine reference. If no level
has been determined, the nominal value (expected signal level) is used.
This setting allows fast switching between the various burst views (Rising, Falling, Full, HighRes).
Actuating the
main measurement following.
START REF MEA S
REF MEAS AUTO or REF MEAS USER
softkey will trigg er a seq uence of two pre-m easurem ents witho ut an y
:
1057.3140.422.9E-1
Page 71
Sequence of Measurement OperationsFSE-K10
19
ZeroSpan vs FrequSweep,
Trigger, Trace, Detector, ...
Yes
Settings:
InpAtt & RefLvl,
20
Sync to
midamble
32
Trigger
FreeRun
No
No
30
31
Gated sweep
required
Yes
33
200
s. detail:
Sync to midamble
PFE / PVT & CPW
39
Premeasurement
result or final result
Ye
34
s
34
Waiting for trigger
event
35
36
Sweep & data
acquisition
37
SweepEnd at
Trigger/Gate?
No
38
Yes
NofBursts attained
SingleSweep or
List?
No
1057.3140.422.10E-1
Page 72
Sequence of Measurement OperationsFSE-K10
Looked at in detail, each (pre-)measurement consists of the following operations:
19The settings needed for the intended measurement are made.
20Synchroni zation to midamble is optional f or PVT and CPW/CPI and mandatory for PF E. This
choice is available only if Option FSE-B7 is fitted.
30-39Measurement without digital demodulation (= option B7).
31,33MOD measurem ents m ay requir e a "Ga ted Sweep ", ie a fr equenc y sweep m ade only a t spec if ic
times. For this purpose, an external gate signal (comparable to trigger) is indispensable.
32,33Unless FreeRun was sel ected, the trig ger e vent (ex ternal; option all y, for FSE- K10 a lso video or
RF power) is required before continuing .
34FSE-K10 waits for trigger event, then starts or continues the sweep.
35The sweep is enabled.
36Collection of measured data until end of sweep or, with gated sweep, during a specific gate
time.
37Check if sweep is terminated.
38Test: Check if number of preset sweeps has been attained AND if meas urement is of Single
type or List type.
39 The measurement is completed if the preset number of sweeps has been attained and the
measurement is of the Single type.
With digital demodulation, there are some differences as compared to the above sequence as far as the
trigger logic is concerned. T he sequence of operations wi th digital demodula tion activated is sho wn in
the following for PFE and PVT/CPW:
1057.3140.422.11E-1
Page 73
Sequence of Measurement OperationsFSE-K10
200
Sync to midamble
201
202
Trigger
FreeRun
206
Burst found?
Yes
Yes
Sweep & data acquisition
No
No
203
204
205
207
Burst not
found
203
Waiting for
trigger event
208
NofBurst > 1
No
Yes
209
Sync found?
Yes
No
End of measurement
210
Sync not
found
212
Trace &
limit check
222
211
NofBurst > 1
No
213
Max bursts &&
single?
Yes
Yes
No
1057.3140.422.12E-1
Page 74
Sequence of Measurement OperationsFSE-K10
PFE sequence in detail:
200-204Gated sweep is not possible – waiting for trigger event or FreeRun (not advisable).
205Collection of measured data
206Searching for bursts in a predefined time window (cf settings for result buffer).
207, 208In case of an error, a mes sage is output and the IEC/IEEE s tatus bit 'Burst not F ound' is
set. In the specia l case of "N ofBur sts == 1 ", the er ror c onditio n is ignor ed ho wever during
further processing.
209Search for sync pattern in the selected burst
210, 211In cas e of an error, a m essage is output and th e IEC/IEEE status bi t 'Sync not Found' is
set. In the specia l case of "N ofBur sts == 1 ", the er ror c onditio n is ignor ed ho wever during
further processing.
212Evaluati on of trace with numeric al indic ation of the determ ined frequ ency error and phase
error.
213If the predefined number of bursts has been attained and the measurement has been
started with 'Single', it ends at this point.
1057.3140.422.13E-1
Page 75
Sequence of Measurement OperationsFSE-K10
200
Sync to midamble
251
252
Trigger
FreeRun
256
Sync f ound?
Yes
No
Yes
No
253
254
255
Sweep &
data acquisition
257
Sync not
found
253
Waiting for trigger
event
258
NofBurst > 1
Yes
No
259
Trace &
limit check
222
End of mea s urement
260
Max bursts &&
single?
Yes
No
1057.3140.422.14E-1
Page 76
Sequence of Measurement OperationsFSE-K10
PVT/CPW sequence in detai l:
The sequence of operations for PVT/CPW differs from that for PFE in that no burst search is performed.
Only the sync pattern sequence (item s 256 and 257) is searched in th e recorded data stre am within a
limited range (compare "Sett ings") to determ ine a hi gh-prec ision tim e ref erence. In t he cas e of an error ,
the procedure is the same as for PFE, including the special case of "NofBursts == 1".
1057.3140.422.15E-1
Page 77
Sequence of Measurement OperationsFSE-K10
13
Display
table
split screen
Main measurement
1
2
3
MEAS MAX
OUTPUT PWR
4
Sta:= Dyn:= 0
measurement(0,0)
5
6
All pwr ctrl
level?
Yes
14
Display table
full screen
12
measurement with
Ctrl Lvl (Sta,Dyn)
No
7
Max. Dyn Ctrl
Level?
Yes
10
INC STATIC
PWR LEVEL
11
Sta := Sta + 1
Dyn := 0
No
No
8
INC DYNAM
PWR LEVEL
Yes
9
Dyn := Dyn + 1
1057.3140.422.16E-1
Page 78
Sequence of Measurement OperationsFSE-K10
The sequence of operations of a CPW measurement (BTS/K11) is described on the previous page.
The sequence of the CPW measurement for MS (FSE-K10) differs from that of the BTS by the
replacement of the 2 power control stages by a single one which is swept in strict sequential mode.
The measurement is initiated b y pres s ing the M EA S MAX O UTPUT PWR softkey. Then either the static
or the dynamic power control le vel can be incremented as require d. Each incrementation of the static
power control level wil l reset the dynam ic power c ontrol leve l to 0 ( zero) . If the initi ali zing sof tk e y MEAS
MAX OUTPUT PWR is actuated during the current session, the session is aborted (measurement
status: Aborted).
Entries by the user are controlled by the availability of softkeys INC STATIC PWR LEVEL and INC
DYNAM PWR LEVEL. if the maximum value of the d ynamic power co ntrol level is attained, the INC
DYNAM PWR LEVEL softke y is inhibited so that onl y the static power contr ol level c an be i ncreased. If
the maximum value is attained here too, pressing the softkey again causes the session to be terminated
and the total result to be displayed on the full screen.
In the ’full-screen table’ status, both INC softkeys are inh ibited and a new session ca n be started only
by pressing the MEAS MAX OUTPUT PWR softke y.
Unlike the CPW measurement, the CPI m easurement allows an y single carrier-power m easurements.
In contrast to all other measurements of FSE-K10/K11, single measurements have no inherent
relationship. There is no total r esu lt of the sess ion. T he CPI menu can be exited and acti vated a ny tim e
without any effect on the table contents of the CPI measurement.
CPI measurements are mainl y intende d for development engineers in the l ab who modify parameters of
their workstation and want to investi gate the eff ects thereof on th e carrier po wer. CPW measurements ,
on the other hand, are aimed at complying with acceptance test specifications for the GSM standard.
CPI measurements, therefore, only evaluate absolute levels and make no relative comparisons
between the levels of different power control level settings. Operator control is described by the
additional hints following the description of the softkey menus.
The following overview gives a classification of the Kxx measurements according to the above diagram:
MeasurementModeWithout With
premeasurement
KXX PFE.....XX---
KXX CPW.....XXXX
KXX CPI.....XXXX
KXX PVT.... .XX---Po ssible without premeasurement
KXX MOD RXSingle, ContXX--KXX MOD RXListXXXX
KXX MOD Tx, ARFCN ±1.8 MHzSingle, ContXX---Bandwidth c orrection necessary (see
KXX MOD Tx, ARFCN ±1.8 MHzListXXXXBandwidth correction necessary (see
KXX TRASingle, ContXX---PwrCoupling OFF: without premeas.
KXX TRAListXXXXPwrCoupling OFF: without premeas.
KXX SPU xxx.....XXXX
Sequence
data
Remarks
depending on user settings
Normally with two premeasurements
below)
below)
The spectrum-due-to-modulation measurements provide a reference value for the limit lines, which
should be measur ed with a 3 0 kHz bandwidth. To evaluate th e ac tu al car r ier po wer (ie the signal power
measured with infinite bandwidth) a bandwidth-dependent correction value is added to the preview
value: it is now 8 dB (analogous to GSM 05.50) for 30 kHz. The corrected value is also used for
selecting the limit lines.
1057.3140.422.17E-1
Page 79
Sequence of Measurement OperationsFSE-K10
2.1.1.1 Errors and Failures During Measurements
There are three main sources of error in GSM measurements:
Carrier signal missing:
Causes:Wrong operating frequency (ARFCN), SFH active, test line defective or incorrect
attenuation, wrong power class
Effects:Premeasurement: insufficient power measurement (eg average measurement with
SFH), measurement is aborted.
Sweep stops (measurements with midamble synchronization)
Main measurement:sweep stops when midam ble synchroni zation (PFE, PVT , CPW)
is activated and number of bursts > 1. When signal is applied
again, the sweep continues.
Trigger missing:
In all triggered measurements: sweep stops (indicated by trigger LED).
The following is recommended for the trigger source:
•
External trigger:allows all measurements without restrictions.
•
Video trigger:only measurements directly on the carrier are possible (PFE, PVT, CPW/CPI)
•
RF power:TRA and MOD-ARFCN +/-1.8 MHz) are p ossible in addition to m easurements for
video.
•
Free Run:permissible for SPU and TRA if no other trigger source is available.
Burst missing
Causes:dummy burst, SFH active, wrong frame trigger
Effects:In triggered measurements, limit lines may be exceeded (PVT);
sweep stops (in measurements with midamble synchronization).
1057.3140.422.18E-1
Page 80
Sequence of Measurement OperationsFSE-K10
2.1.2 Abort of Measurements
2.1.2.1 Abort by the User
In manual operatio n, each run ning m easurem ent can be aborte d by pres sing the star t sof tke y again ( ie
the key with which the measurement was initiated).
Starting another measur ement lik ewise caus es an a bort of the r unning m easurem ent. This als o appli es
to switchover from Single to Continuous or List, for example.
Each measurement is als o aborted on quittin g the menu (new! As f rom 1.90 for all measur ements) or
modifying a test parameter (even in the side menu).
All aborts, except the spurious single step measurement, are without warning. Only if – during a running
spurious single-step meas urement – the user attempts to restart this m easurement or to start the Lis t
measurement, is he queried prior to the restart being made.
For remote control, a special ABORT command is available which likewise causes the current
measurement to be aborted (see manual for IEC/IEEE-bus control).
2.1.2.2 Abort by the Instrument
Measurements are nor mally perf ormed until the e nd. Onl y in the cas e of insufficient carrier p ower (see
above) are measurements aborted after the pre-measurement with an error message being output.
If the maximum number of table entries (50 at present) is ex ceeded in a spurious measurem ent, the
measurement is continued but no other entry is made in the table. In CPI measurements, the last
40 single measurements are stored.
2.1.3 Results of Measurements
MeasurementSingle resultLimit lineTable
PFEPeak and RMS phase
CPW,
CPI
PVTUpper & lower
MOD
Sgl
MOD
List
Rx/Tx
MOD
List
ARFCN ±1.8 MHz
TRA
Sgl
TRA
List
SPUFor each section:
error;
frequency error in ‘ppm’
One entry per single measurement:
CPW: CtrlLevel, measured value and evaluation
CPI: test parameter, measured value and evaluation
Limit line
Upper limit line
For each section:
One ‘Frequency’-’Level’ entry for each event.
One summary entry for each section (see below).
One entry per pair of measurement:
Frequency offset (+/-), level(-), l evel (+).
Test points defined by standard.
Upper limit line
One entry per pair of measurement:
Frequency offset (+/-), level(-), l evel (+).
Test points at +/ -400, 600, 1200 and 1800 kHz.
One ‘Frequency’-’Level’ entry for each event.
One summary entry for each section (see below).
1057.3140.422.19E-1
Page 81
Sequence of Measurement OperationsFSE-K10
The result tables of the MOD-LIST Rx/Tx and SPU measurements are arranged in groups:
The whole measurement range is divided into frequency sections.
A summary entry is generate d for each sec tion. T his e ntr y contains the sec tion lim its and the m axim um
level of the worst qu ality class measured in t his section. All events are then listed as frequency-level
pairs, an event being a measured value which lies above the limit or is at least in the tolerance range.
Measured values are assigned to one of the following quality classes:
•
PASSEDbest quality classResults below the tolerance
•
MARGIN:values below the limit but in the tolerance range (user-defined)
•
EXCEPTION:values above the limit but still below the limit provided for
outliers
•
FAILEDworst quality class values violate the limit and also the permissible exceptions
All measured values except PASSES are events in the above sense.
The section entries (and the table as well) are assigned either PASSED or FAILED as total evaluation .
If the number of measured outliers exceeds in one section the value defined by the standard, the
section is assigned FAILED even if all events are assigned EXCEPTION in the worst case.
The tables have 50 lines at maximum.
In some measurements, therefor e, it may happen that "fai led" is indicated as a summar y evaluation in
the table header although all visible entries in the table are evaluated "passed".
The reason for this is that the "failed" evaluation might not appear until line 50 and is therefore not
visible.
1057.3140.422.20E-1
Page 82
Exiting FSE-K10FSE-K10
2.2 Exiting FSE-K10
When FSE-K10 is exit ed, the status las t active is s tored s o that t he oper atin g status c an b e r estored as
far as possible on re-entering. The FSE-K10 settings are largely transferred to the other operating
modes to allow FSE-K10- conformal meas urements also in the oth er operating modes as far as this is
expedient.
FSE-K10 can be exited by:
•
Power Off
•
Preset (mit anschließender Mode-Auswahl)
•
Taste
Power off
FSE-K10 is restarted (warm start) on power-up of FSE/FSIQ. FSE-K10 in this case is not exited but
only interrupted.
Preset
The instrument is reset to a defined default state. This function is comparable to a cold start. For
FSE-K10 this m eans: no re-activation of the last operating status on re- entering but activatio n of the
default status. The targ et status of the instrument af ter a preset is c learly def ined independent ly of the
current FSE-K10 status:
Analyzer, Frequ-Sweep Cont, Full Span, ...
MODE
mit anschließender Mode-Auswahl
Change of operating mode
The last state is a function of the FSE-K10 output state and target mode. The parameters are set for the
target mode in such a way that similar measurements to FSE-K10 are possible. For changing from
FSE-K10 to the analyzer and vector analyzer modes, the following is defined:
Transformation of (FSE-K 10) par ameters:
Trigger
TrgSource compatible (if not available in target mode: FreeRun)
TrgLevel is taken over for all modes (if available).
TrgDelay corresponding to trigger last active.
TrgPolarity, TrgEdge/Level are taken over.
Level
ExtAtt is represented by RefLvl offset.
RefLevel is set directly.
RF attenuation is calculated in line with FSE-K10 (AUTO or by formula).
Caution:For RX measurements, the input attenuation is set to AUTO ===> not to 0 dB !
Frequency
For the zero span or the vector mode, the ARFCN frequency is set as the center frequency.
For the frequency sweep m ode, the start and the stop frequ ency are selected in accordanc e
with the band last selected:
for Tx/Rx band the corresponding band (possibly incl. +/- 2 MHz),
for TX ±1.8MHz the center frequency f(AFRCN) with a span of 3.6 MHz,
with SPU the last selected section (or more simply the whole band of the NTx
measurement).
Screen
Full screen is taken for the analyzer mode, split screen for the vector analyzer split-screen
mode (PFE).
1057.3140.422.21E-1
Page 83
Exiting FSE-K10FSE-K10
Trace
Trace assignment in FSE-K10:
2for single-trace measurements in LIST mode (trace 1 used for table)
1for single-trace measurements in SINGLE/CONT mode
1,3,2for PVT measurements with MAX Hold, Average, MIN Hold
2for SPU measurements
1,3,4 for PFE measurements with MAX Hold, ClearWrite, MIN Hold
2for CPW/CPI measurements (Average)
Trace assignment is maintained as far as possible.
For changing to other modes, the following is defined:
On exiting FSE-K10, the FSE-K10 operati ng status las t active is s tored b y the ne wl y selected m ode (eg
RECEIVER mode, or another mode) and activated again by this mode when FSE-K10 is re-entered.
1057.3140.422.22E-1
Page 84
PresetFSE-K10
2.3 Preset
FSE-K10 is reset to the default status. Then the following setting is assumed:
Analyzer default setting: Frequ-S we ep, Ful lSp an, S weepTime Auto, RBW/VBW Auto, Trigger Free Run,
...
2.3.1 Signal Analysis Mode
The target status assumed results from the FSE-K10 status last active (or the last measurement
performed with FSE-K10, respectively).
wg. Trace 2 is renamed trace 1 because of full-screen display.
PVTZeroSpan; f=f(ARFCN); SweepTime 800 µs or 100 µs
MODFrequ-Sweep; Span/Start-Stop acc. to Tx/ARFCN ±1.8 MHz/Rx; NON-Rx: trigger at
FreeRun, plus gated sweep with gate length 170 µs. Otherwise trigger is taken over
(setting analogous to CONTINUOUS measurement).
TRAFrequSweep; span 3.6 MHz, f = f (ARFCN); sweep time in accordance with TRA
formula (settings analogous to CONTINUOUS measurement).
SPUFrequ-Sweep Continuous; Start-Stop FullSpan; SweepTime AUTO. Trigger is taken
over. Trace 2 is renamed trace 1 because of full-screen display.
2.3.2 Vector Signal Analysis Mode
The PFE status of FSE- K10 (vector mode) is set irrespective of the FSE-K10 status : split screen with
result table at the top and trace at the bottom.
2.3.3 Receiver Mode etc.
In the RECEIVER mode, the status is saved upon exiting FSE-K10 and reinstated upon re-entry.
1057.3140.422.23E-1
Page 85
General RemarksFSE-K10
2.4 General Remarks
In the case of the T RA, MOD and PVT measurements, a pr emeasurement is perform ed prior to the
main measurement.
The premeasurement is used to define the reference point for the limit lines.
In the case of PVT , the premeasurement takes place in two stag es. In the first premeasurement with
max. 2 bursts, the actual level applied is determined. A second premeasurement with maximum 20
bursts is then perform ed, which serves to determ ine the exact reference point for the limit lines using
the operating point optimally set with the result of the first premeasurement.
In the PFE, CPW and SPU measur ements, on the ot her hand, f ixed level va lues are use d and t hus no
premeasurements are required.
"ExtAtt" (external attenuat ion) is taken into account in all m easurements in the carrier or non- TX band,
"RxGain" for all measurem ent in the RX band. The values have b asically the sam e effect but are used
separately. RxGain can be entered any time, not only for an activated RX band measurement.
In all measurements with Min, Max and Average display, three traces are activated only if the number of
sweeps is higher than 1. Likewise, vector tri ggering operates cor rectly only if the n umber of sweeps is
set to a value higher than 1 (this too is a feature of the basic unit). If this is not the case, an error
message is output, for exam ple "S ync not f ound" if the m idam ble is not detec ted, in whic h case ther e is
no synchronization to midamble.
Table 2-3Exceptions and base lines for MOD measurements
StandardRangeExceptionsBase lineconcerning
PGSM-I400 KHz to 1800 KHz-36 dBm
X-GSM600 KHz to 6.0 MHz 3 to max -36 dBm-65 dBm
-59 / -64 / -69 dBm
6.0 MHz to TX band12 t o max -36 dBm-65 dBm
-59 / -64 / -69 dBm
DCS-I800 KHz to 6.0 MHz 3 to max -36 dBm-57 dBmPwrCls 1 to 4 (all)
6.0 MHz to TX band12 zu max -36 dBm-57 dBmPwrCls 1 to 4 (all)
DCS-II600 KHz to 6.0 MHz 3 to max -36 dBm-57 dBm
-57 / -62 / -67 dBm
6.0 MHz to TX band12 zu max -36 dBm-57 dBm
-57 / -62 / -67 dBm
PCS600 KHz to 6.0 MHz 3 to max -36 dBm-57 dBm
-57 / -62 / -67 dBm
6.0 MHz to TX band12 t o max -36 dBm-57 dBm
-57 / -62 / -67 dBm
PwrCls 1 to 8
PwrCls M1 / M2 / M3
PwrCls 1 to 8
PwrCls M1 / M2 / M3
PwrCls 1 to 4
PwrCls M1 / M2 / M3
PwrCls 1 to 4
PwrCls M1 / M2 / M3
PwrCls 1 to 4
PwrCls M1 / M2 / M3
PwrCls 1 to 4
PwrCls M1 / M2 / M3
For details on the individual types of measurement please refer to chapter 3.
1057.3140.422.26E-1
Page 88
FSE-K10Starting the Application
3 Measurements with Option FSE-K10
3.1 Starting the Application
CONFIGURATION
MODE
SETUP
GSM MS
ANALYZER
The application is called with the
softkey . A selection menu is offered according to the specific standards
with a softkey assigned to each measurement. The presettings are
performed in the submenu of the
menu of each measurement.
All settings can generally be performed using the softkeys.
In addition, the following keys are available for performing special actions:
1) only with Option FSE-B7 (Vector Analyzer) installed
2) The "Small MS ON/OFF" softkey is available only for R-GSM (and DCS-II+); it serves
for controlling the exceptions from the standard specifications that are permitted
for "Small MS" (presently for SPU Tx limit only).
TRIGGER
ADJUST
TRIGGER
LEVEL
SLOPE
POS NEG
1057.3140.423.2E-3
Page 90
FSE-K10Menu Overview
CONFIGURATION
MODE
SETUP
ANALYZER
VECTOR
ANALYZER
GSM MS
ANALYZER
PHASE/FREQ
ERROR
SINGLE
CONTINUOUS
GSM MS
ANALYZER
SETTINGS
PHASE/FREQ
CARRIER
1)
VS TIME
MODULATION
SPECTRUM
TRANSIENT
SPECTRUM
SPURIOUS
ERROR
POWER
POWER
1)
NO. OF
BURSTS
PHASE/FREQ
ERROR
ARFCN /
FREQUENCY
POWER
SETTING
X UNIT
SYMB TIME
TRIGGER
EDIT
NO. OF
BURSTS
SET
MANUAL
SET TO
STANDARD
1) only with Option FSE-B7 (Vector Analyzer) installed
2) The "Small MS ON/OFF" softkey is available only for R-GSM (and DCS-II+); it serves
for controlling the exceptions from the standard specifications that are permitted
for "Small MS" (presently for SPU Tx limit only).
EDIT
PHASE
PEAK
PHASE
RMS
FREQUENCY
USER LIMIT
ONOFF
TRIGGER
FREE RUN
VIDEO
EXTERN
RF POWER
SLOPE
POS NEG
TRIGGER
ADJUST
TRIGGER
ADJUST
COARSE
ADJUST
FINE
ADJUST
AUTO
ADJUST
TRIGGER
LEVEL
SLOPE
POS NEG
1)
POWER
SETTING
EXTERNAL
ATTEN
OUTPUT
MS POWER
POWER
CLASS
POWER
CTRL LEVEL
SMALL MS
ON OFF
LIMIT/PWR
COUPLED
SIGNAL
POWER
LIMIT LINE
REF POWER
2)
ARFCN /
FREQUENCY
ARFCN
ARFCN
AUTOSELECT
FREQUENCY
1057.3140.42 3.3E-3
Page 91
Menu OverviewFSE-K10
CONFIGURATION
MODE
SETUP
ANALYZER
VECTOR
ANALYZER
GSM MS
ANALYZER
CARRIER
POWER
MEAS MAX
OUTPUT PWR
INC PWR
CTRL LEVEL
NO. OF
BURSTS
GSM MS
ANALYZER
SETTINGS
PHASE/FREQ
CARRIER
1)
VS TIME
MODULATION
SPECTRUM
TRANSIENT
SPECTRUM
SPURIOUS
ERROR
POWER
POWER
1)
NO. OF
BURSTS
SET
MANUAL
SET TO
STANDARD
SGL MEAS
ON OFF
MEAS SGL
PWR LEVEL
CTRL LEVEL
CLEAR SGL
RESULT TAB
TRIGGER
FREE RUN
VIDEO
EXTERN
RF POWER
SLOPE
POS NEG
POWER
SIGNAL
POWER
FREQUENCY
CONDITIONS
NORM EXTR
TRIGGER
ADJUST
COARSE
ADJUST
FINE
ADJUST
AUTO
ADJUST
CARRIER
POWER
ARFCN /
POWER
SETTING
MEAS
BANDWITH
SYNC TO
MIDAMBLE
TRIGGER
1)
1)
POWER
SETTING
EXTERNAL
ATTEN
OUTPUT
MS POWER
POWER
CLASS
POWER
CTRL LEVEL
SMALL MS
ON OFF
2)
ARFCN /
FREQUENCY
ARFCN
ARFCN
AUTOSELECT
FREQUENCY
TRIGGER
ADJUST
1) only with Option FSE-B7 (Vector Analyzer) installed
2) The "Small MS ON/OFF" softkey is available only for R-GSM (and DCS-II+); it serves
for controlling the exceptions from the standard specifications that are permitted
for "Small MS" (presently for SPU Tx limit only).
TRIGGER
POS NEG
LEVEL
SLOPE
LIMIT/PWR
COUPLED
SIGNAL
POWER
LIMIT LINE
REF POWER
1057.3140.423.4E-3
Page 92
FSE-K10Menu Overview
CONFIGURATION
MODE
SETUP
ANALYZER
VECTOR
ANALYZER
GSM MS
ANALYZER
POWER
VS TIME
SINGLE
CONTINUOUS
BURST HIGH
RESOLUTION
NO. OF
BURSTS
GSM MS
ANALYZER
SETTINGS
PHASE/FREQ
CARRIER
POWER
1)
POWER
VS TIME
MODULATION
SPECTRUM
TRANSIENT
SPECTRUM
SPURIOUS
ERROR
1)
FULL BURST
RISING
EDGE
FALLING
EDGE
RISING
& FALLING
START
REF MEAS
POWER
VS TIME
REF MEAS
AUTO USER
ARFCN /
FREQUENCY
POWER
SETTING
X UNIT
SYMB TIME
MEAS
BANDWITH
SYNC TO
MIDAMBLE
TRIGGER
EDIT
1)
NO. OF
BURSTS
SET
MANUAL
SET TO
STANDARD
1) only with Option FSE-B7 (Vector Analyzer) installed
2) The "Small MS ON/OFF" softkey is available only for R-GSM (and DCS-II+); it serves
for controlling the exceptions from the standard specifications that are permitted
for "Small MS" (presently for SPU Tx limit only).
EDIT
LIMIT LINE
FILTER
NEW LIMIT
LINE
EDIT LIMIT
LINE
COPY
LIMIT LINE
DELETE
LIMIT LINE
USER LIMIT
ONOFF
PAGE
UP
PAGE
DOWN
TRIGGER
FREE RUN
VIDEO
EXTERN
RF POWER
SLOPE
POS NEG
TRIGGER
ADJUST
TRIGGER
ADJUST
COARSE
ADJUST
FINE
ADJUST
AUTO
ADJUST
TRIGGER
LEVEL
SLOPE
POS NEG
1)
POWER
SETTING
EXTERNAL
ATTEN
OUTPUT
MS POWER
POWER
CLASS
POWER
CTRL LEVEL
SMALL MS
ONOFF
LIMIT/PWR
COUPLED
SIGNAL
POWER
LIMIT LINE
REF POWER
2)
ARFCN /
FREQUENCY
ARFCN
ARFCN
AUTOSELECT
FREQUENCY
1057.3140.42 3.5E-3
Page 93
Menu OverviewFSE-K10
CONFIGURATION
MODE
SETUP
ANALYZER
VECTOR
ANALYZER
GSM MS
ANALYZER
3)
3)
4)
4)
MODULATION
SPECTRUM
ARFCN /
FREQUENCY
POWER
SETTING
RX BAND
GAIN
TRIGGER
EDIT
MODULATION
SPECTRUM
SINGLE
FREQ SWEEP
CONTINUOUS
FREQ SWEEP
START
LIST
NO. OF
BURSTS
GSM MS
ANALYZER
SETTINGS
PHASE/FREQ
CARRIER
POWER
1)
VS TIME
MODULATION
SPECTRUM
TRANSIENT
SPECTRUM
SPURIOUS
ERROR
POWER
1)
ARFCN
±1.8 MHZ
TX BAND
±1.8 MHZ /
TX BAND
RX BAND
GSM 900
RX BAND
GSM 1800
NO. OF
BURSTS
SET
MANUAL
SET TO
STANDARD
1) only with Option FSE-B7 (Vector Analyzer) installed
2) The "Small MS ON/OFF" softkey is available only for R-GSM (and DCS-II+);
it serves for controlling the exceptions from the standard specifications
that are permitted for "Small MS" (presently for SPU Tx limit only).
3) not for GSM900 phase
4) not for PCS1900, in that case just a softkey RX BAND
EDIT
LIMIT LINE
FILTER
NEW LIMIT
LINE
EDIT LIMIT
LINE
COPY
LIMIT LINE
DELETE
LIMIT LINE
USER LIMIT
ONOFF
PAGE
UP
PAGE
DOWN
TRIGGER
FREE RUN
RF POWER
POS NEG
TRIGGER
ADJUST
VIDEO
EXTERN
SLOPE
TRIGGER
ADJUST
COARSE
ADJUST
FINE
ADJUST
AUTO
ADJUST
TRIGGER
LEVEL
SLOPE
POS NEG
1)
POWER
SETTING
EXTERNAL
ATTEN
OUTPUT
MS POWER
POWER
CLASS
POWER
CTRL LEVEL
SMALL MS
ONOFF
LIMIT/PWR
COUPLED
SIGNAL
POWER
LIMIT LINE
REF POWER
2)
ARFCN /
FREQUENCY
ARFCN
ARFCN
AUTOSELECT
FREQUENCY
1057.3140.423.6E-3
Page 94
FSE-K10Menu Overview
CONFIGURATION
MODE
SETUP
ANALYZER
VECTOR
ANALYZER
GSM MS
ANALYZER
TRANSIENT
SPECTRUM
SINGLE
FREQ SWEEP
CONTINUOUS
FREQ SWEEP
START
LIST
NO. OF
BURSTS
GSM MS
ANALYZER
SETTINGS
PHASE/FREQ
ERROR
1)
CARRIER
1)
POWER
POWER
VS TIME
MODULATION
SPECTRUM
TRANSIENT
SPECTRUM
SPURIOUS
NO. OF
BURSTS
SET
MANUAL
SET TO
STANDARD
EDIT
LIMIT LINE
FILTER
NEW LIMIT
LINE
EDIT LIMIT
LINE
COPY
LIMIT LINE
DELETE
LIMIT LINE
TRANSIENT
SPECTRUM
ARFCN /
FREQUENCY
POWER
SETTING
MS SFH
ONOFF
MEAS
BANDWITH
TRIGGER
EDIT
TRIGGER
FREE RUN
VIDEO
EXTERN
RF POWER
SLOPE
POS NEG
TRIGGER
ADJUST
COARSE
ADJUST
FINE
ADJUST
AUTO
ADJUST
1)
POWER
SETTING
EXTERNAL
ATTEN
OUTPUT
MS POWER
POWER
CLASS
POWER
CTRL LEVEL
SMALL MS
ONOFF
2)
ARFCN /
FREQUENCY
ARFCN
ARFCN
AUTOSELECT
FREQUENCY
LIMIT/PWR
COUPLED
USER LIMIT
ONOFF
PAGE
UP
PAGE
DOWN
1) only with Option FSE-B7 (Vector Analyzer) installed
2) The "Small MS ON/OFF" softkey is available only for R-GSM (and DCS-II+); it serves
for controlling the exceptions from the standard specifications that are permitted
for "Small MS" (presently for SPU Tx limit only).
TRIGGER
ADJUST
TRIGGER
LEVEL
SLOPE
POS NEG
SIGNAL
POWER
LIMIT LINE
REF POWER
1057.3140.42 3.7E-3
Page 95
Menu OverviewFSE-K10
CONFIGURATION
MODE
SETUP
ANALYZER
VECTOR
ANALYZER
GSM MS
ANALYZER
SPURIOUS
START LIST
SGL STEP
CONT LIST
SGL STEP
START
LIST
SWEEP
COUNT
GSM MS
ANALYZER
SETTINGS
PHASE/FREQ
ERROR
CARRIER
POWER
1)
POWER
VS TIME
MODULATION
SPECTRUM
TRANSIENT
SPECTRUM
SPURIOUS
1)
TX BAND
<>TX BAND
IDLE
MODE
TX BAND
±2.0 MHZ
SELECT
STEP
SPURIOUS
ARFCN /
FREQUENCY
POWER
SETTING
MS SFH
ONOFF
TX SUPPR
ONOFF
ANTENNA
COND RAD
SWEEPTIME
STD AUTO
TRIGGER
EDIT
SWEEP
COUNT
SWP COUNT
TX / <>TX
1) only with Option FSE-B7 (Vector Analyzer) installed
2) The "Small MS ON/OFF" softkey is available only for R-GSM (and DCS-II+); it serves
for controlling the exceptions from the standard specifications that are permitted
for "Small MS" (presently for SPU Tx limit only).
EDIT
LIMIT LINE
FILTER
NEW LIMIT
LINE
EDIT LIMIT
LINE
COPY
LIMIT LINE
DELETE
LIMIT LINE
USER LIMIT
ONOFF
PAGE
UP
PAGE
DOWN
TRIGGER
FREE RUN
VIDEO
EXTERN
RF POWER
SLOPE
POS NEG
TRIGGER
ADJUST
TRIGGER
ADJUST
COARSE
ADJUST
FINE
ADJUST
AUTO
ADJUST
TRIGGER
LEVEL
SLOPE
POS NEG
1)
POWER
SETTING
EXTERNAL
ATTEN
OUTPUT
MS POWER
POWER
CLASS
POWER
CTRL LEVEL
SMALL MS
ONOFF
LIMIT/PWR
COUPLED
SIGNAL
POWER
LIMIT LINE
REF POWER
2)
ARFCN /
FREQUENCY
ARFCN
ARFCN
AUTOSELECT
FREQUENCY
1057.3140.423.8E-3
Page 96
FSE-K10Selection of Default Settings
3.3 Selection of Default Settings
CONFIGURATION
MODE
SETUP
Setting reference frequency
Press SETUP key.
EXT REF
FREQUENCY
Press
EXT. REF FREQU ENCY
the frequency standard.
REF
INT EXT
Set softkey
REF INT/EXT
The
SETTINGS
SETTINGS
parameters. The associated secondary menu permits to select the
transmission system. The parameters to be varied frequently such as
channel, power, midamble, etc. are to be found in the main part of the
submenu for ease of access.
set the standard to be used for the measurement.
The 5 softkeys are selection switches, ie when one is active, the others are
automatically de-activate d. If a standard is selec ted, the FSE/FSIQ uses the
E-GSM 900
standard-specific settings and limit values for the measurement.
With the standard P-GSM 900/p hase 1 the meas urement is made ac cording
GSM 1800
(DCS 1800)
to GSM11.10, with DC S 1800/phase 1 to th e ETSI standard ET S 300,020-3
(GSM 11.10-DCS), with P-GSM 900/p hase 2 an d DCS 1800 /phase 2 as well
as E-GSM 900 to the ETSI s tandard ET S 300,60 7-1 (G SM 11 .10-1) and wit h
GSM 1900
(PCS 1900)
PCS 1900 to the American s tandard J-ST D-007 Air Interf ace. The stan dards
GSM 05.05 and GSM 11.10-1 are also the basis for the measurement of
phase II+ (with R-GSM).
R-GSM 900
PHASE I
PHASE II
PHASE I/II/II+
associated limit values according to the selected phase of the standard .
With some standards, the softkeys are not available (PCS 1900) or partly
softkeys changes the s e qu enc e of measurements and the
The
available because no phases or not all of them are defined.
PHASE II+
1057.3140.42 3.9E-3
Page 97
Selection of Default SettingsFSE-K10
3.3.2 Consideration of the External Attenuation
For measurements at the RF interf ace, RF cab les, atten uator p ads, po wer spl itters, d irectiona l coupl ers
or other devices ar e used to match the level of the RF signal or connect f urther test equipment. The
attenuation from the RF interface of the transmitter to the RF input of the FSE/FSIQ can be corrected by
a constant attenuation factor.
Settings which produce a leve l >27 dBm at the input of the FSE/FSIQ – tak ing into account the other
settings -are generally not possible with the
To prevent the occ ur renc e of an overload, a war nin g is displayed f or thos e s ett in gs whic h would lead to
the threshold level being exceeded (
POWER CTRL LEVEL)
CONFIGURATION MODE - GSM MS ANALYZER - SETTINGS
Menu:
EXTERNAL
ATTEN
and the set value is ignored.
The softkey
used externally. This value is used to correct the level displayed. All
measured levels and the inscription of the Y-axis of the measurement
diagram are shifted by the selected correction value. The effect of the
softkey is identical with the input of a level offset (
using the
Default setting is 20 dB attenuation. This value is to be corrected when
another attenuation is used.
REF
Note:For measureme nts in the RX band with modulation spectrum,
EXTERNAL ATTEN
softkey.
POWER CLASS, EXTERNAL ATTEN, OUTPUT MS POWER,
EXTERNAL ATTEN
opens a window for entering the attenuation
REF LEVEL OFFSET
key in the analyzer mode.
the attenuation enter ed with E XTERNA L ATTEN is irrelevant. In
this case the RX band gain softkey has basically the same
effect.
The FSE/FSIQ switches to a n inter n al atte nua tio n of 0 d B at the
beginning of the meas urement a nd, at th e end of it, to the v alue
set prior to starting the measurement. The carrier is to be
sufficiently suppressed by appropriate measures (eg by means
of a bandstop filter). If an external attenuation is used for a
measurement in the RX band, it can be taken into acc ount via
RX BAND GAIN (negative value of RX BAND GAIN =
attenuation in the RX BAND).
If the DUT to be measured has a very high output power, an
FSE/FSIQ-internal atten uation (RF ATT) should without fail be
set. The 10 dB steps may be excessive in some cases since the
noise level increases by approx. 10 dB when adding an
attenuation of 10 dB.
If a high dynamic range is required, it is expedient to use an
external attenuation (w hich can be varied i n steps smal ler than
10 dB in most cases) in order to remedy to the large steps of
the RF ATT.
For each dB step of the external attenuator, the RF ATT is
increased by 10 dB approx. one dB later. This extends the
dynamic range.
)
1057.3140.423.10E-3
Page 98
FSE-K10Selection of Default Settings
3.3.3 Setting the Transmission Channel
The operating frequency (ie the carrier frequency) is entered via the ARFCN (Absolute Radio
Frequency Channel Num ber: the transm ission band is di vided into cha nnels of 200 kHz each of which
is identified by means of a standard-dependent number) or directly as a frequency value.
In the case of ARFCN entry, the frequency is limited to the standard range of the transmission band.
Direct entry of the operating frequenc y, on the other h and, allo ws valu es from 1.8 MHz to f
MHz; beyond the transmission band, the band definitions Tx, NTx, Rx make no sense, so only
measurements related to the carrier f requency are p ossible: this inc ludes all meas urements perf ormed
directly on the carrier - PFE, PVT, CPW/CPI – as well as the relative measurements T RA and MODARFCN ±1.8MHz. If the frequenc y entered is within the Tx band, th e described lim itations need not be
taken into account.
The search range of the automatic ARFCN determination covers the whole Tx band. Due to the
features of the marker-to-peak search, an ARFCN is always determined even if no carrier signal is
active in the Tx band.
Menu:
CONFIGURATION MODE - GSM MS ANALYZER - SETTINGS
ARFCN /
FREQUENCY
The
ARFCN / FREQUENCY
softkey calls a submenu which offers three possibilities for setting the
transmission frequency of the MS:
Setting the ARFCN [as channel number]
Automatic searching for the carrier frequency in the frequency band
defined by the standard and automatic setting of the found channel or
frequency.
Entering a frequency in Hz, kHz ...
(Absolute Radi o Frequency Channel Number)
analyzer max
-1.8
On calling the menu
generally opens.
With the
without limitation imposed by the standard-defined transmission band.
Three cases are distinguished:
a) the entered frequency is exactly on the ARFCN
b) the entered frequency is in the TX band but off a permissible
c) the entered frequency is out of the TX band.
The displayed channe l number is r ounded if the enter ed frequenc y does not
exactly correspond to an ARFCN.
On entering the menu and quitt ing the input field, the colour of the sof tkey
indicates the selection (only with a colour display):
a)
b) ARFCN
c)
FREQUENCY
ARFCN
ARFCN
ARFCN
is green,
is green,
is grey,
ARFCN / FREQUENC Y,
softkey it is possible to set the frequency freely, ie
FREQUENCY
FREQUENCY
FREQUENCY
is grey
is green
is green
the input field f or the ARFCN
1057.3140.42 3.11E-3
Page 99
Selection of Default SettingsFSE-K10
The center frequencies are assigned to the channels as follows:
channel number can be entered.
In the FSE/FSIQ, the center frequency is then s et to the fr equency specif ied
in the standard to correspond to the selected channel.
If a frequency which lies outside the trans mission channel as given by the
standard has been selected by means of the
FREQUENCY
softkey, an
empty input field is disp layed with a message (ARFC N not defined) instea d
of the channel number.
The
ARFCN AUTOSELECT
softkey automatically searches for the
transmission channel and sets the center frequency to the channel of the
carrier found. The sear ch is perf ormed in a s weep over t he ent ire freque ncy
range of the transmission band ((TX BAND). The maximum level occurring in
the transmission channel is searched for with the aid of the marker. The
marker frequency is assigned to the respective channel.
The function is performed once after it has been called up.
Frequency hopping is ignored durin g automatic search (AUTO ADJ UST), ie
the entire TX band is examined once. The SFH of the MS should therefore
be switched off.
1057.3140.423.12E-3
Page 100
FSE-K10Selection of Default Settings
FREQUENCY
The FREQUENCY softkey opens an input field in which the desired MS
frequency can be entered without link to the channel spacing.
The frequency unit (Hz, kHz, MHz) is selected in the DATA ENTRY field with
the grey keys next to the numeric keypad.
It is also possible to enter frequenci es which ar e out of the range s pec ified in
the selected standard.
If the entered frequency is within the TX band of the selected standard, the
channel which is next to the selected frequency (arithmetic rounding) is
displayed after pressing the ARFCN softkey.
Permissible frequencies are:
1.8 MHz frequency (maximum instrument frequency – 1.8 MHz).
3.3.4 Setting the Transmit Power of the MS under Test
The mobile's nominal transmit power is to be reported to the FSE/FSIQ so that it can select correct
default settings such as input attenuation, reference l evel and l imit values. The mobile's transmit power
is determined by its power class and the power c ontrol level. The power class defines the maximum
possible output power of the base s tati on, wher eas the pow er c ontrol lev el determines the actual output
power.
The information on the power set in this menu enables the FSE/FSIQ to make optimal setti ngs for Ref
Lvl and attenuation in the active measurement. The formulae for determining the settings are explained
in the measurement chapters.
The limit values for a measurement, for instance, are selected with reference to the power level.
However, the coupling of limit values and the power c ontrol level may also be undone, which allows
limit values of any power control level to be used for measurements with freely selectable power.
To protect the FSE/FSIQ, the settings made in the PO WER menu are automatically checked to make
sure that the input level at the FSE/FSIQ does not exc eed 27 dBm.
(MS output power – attenuation ≤ 27 dBm).
To prevent the occurr ence of an overload, a warning is displayed for thos e s ettings whic h would lead to
the threshold level being exceeded (
POWER CTRL LEVEL)
Since the power measurement is subjected to very stringent accuracy requirements, the analyzer
should not be overdriven.
To prevent this the unit monitors the signal level determined in the preview to establish if the RefLevel is
exceeded by more than 1 dB. If this is the case, a warning is output.
In addition to the output power of the DUT the ex ternal attenuation or an external amplifier has to be
taken into account.
An attenuation value is usually expected. For measurements in the Rx band an amplifier is more
suitable so there are 2 possible values which are exclusively used for the same operation.
• External attenuation (negative value: gain) for all measurements outside the Rx band;
• Rx gain (gain) for all Rx-band measurements (negative value: attenuation).
and the set value is ignored.
POWER CLASS, EXTERNAL ATTEN, OUTPUT MS POWER,
1057.3140.42 3.13 E-3
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.