ROHDE & SCHWARZ GmbH & Co. KG
Test and Measurement Division
Mühldorfstraße 15
81671 München, Germany
st
1
edition 08/2005
Printed in Germany.
Printed on FFC bleached paper.
Subject to alterations, Errors excepted.
Reprints, also in extracts, are only allowed with written permission of the manufacturer.
All rights reserved.
E-1147.1000.04 0-2 Operating manual, 08/2005
Page 3
R&S FS315 Chapter Overview
Chapter Overview
General
Chapter 1
Chapter 2
Chapter 3
Chapter 4
Chapter 5
Chapter 6
Chapter 7
Chapter 8
Chapter 9
Chapter 10
Content of the Manual for Spectrum Analyzer R&S FS315
Data Sheet
Safety Instructions
Certificate of Quality
EC Certificate of Conformity
Customer Support
List of Rohde & Schwarz Offices
Introduction
Control Elements
Putting the R&S FS315 into Operation
Getting Started
Manual Operating Concept
Using the R&S FS315
Remote Control/PC Software R&S FS315-K1
Instrument Interfaces
Error Messages
Index
Operating manual, 08/2005 0-3 E-1147.1000.04
Page 4
Content of the Manual R&S FS315
Content of the Manual
Operating Manual
Introduction
This operating manual provides information about:
Technical characteristics of the instrument
Putting into operation
Basic operating procedures and control elements
Operation via menus and remote control
By way of an introduction, a typical R&S FS315 measurement is described.
The operating manual also contains information about maintenance and troubleshooting based on the warnings and error messages issued by the instrument.
Setting range of reference level -110 dBm to +36 dBm
Resolution 0.1 dB
RF attenuation range manual selection or automati-
0 dB to 70 dB
cally coupled to reference level
Resolution 2 dB
Traces
1 active trace and 1 stored
trace
Trace detectors max peak, min peak, sample,
average, RMS
Trace functions clear/write, max hold, min hold,
average
Max. uncertainty of level measurement
Frequency response 9 kHz to 3 GHz,
< 1.0 dB
RF attenuation 0 dB to 70 dB
Reference level accuracy < 0.3 dB
Display nonlinearity
0 dB to -60 dB < 0.3 dB
-60 dB to -70 dB < 1 dB
Bandwidth switching uncertainty < 0.3 dB
Total measurement uncertainty
0 dB to -60 dB down from ref.
1.5 dB, typ. 0.7dB
level, RBW ≤ 5 MHz
Markers
Number of markers and delta mark-
1 marker and 1 delta marker
ers
Marker functions
peak, next peak left, next peak
right, center = marker fre-
quency, ref level = marker level
Marker displays
normal (level), noise marker,
frequency counter, n-dB down
(bandwidth)
Audio demodulation
zero-span only, RBW ≤ 1 MHz
AM and FM
E-1147.1000.04 0-14 Operating manual, 08/2005
Page 15
R&S FS315 Data Sheet
Trigger
Span ≥ 1 kHz
Trigger source free run, external
Trigger offset Sweep time > 100 ms
Span = 0 Hz
Trigger source free run, external, video
Trigger offset negative offset limited by
0 s ≤ Trigger offset ≤ 100 ms,
resolution 25 ns
-100 ms ≤ Trigger Offset ≤
sweep time
10 s, resolution 25 ns
Tracking Generator
Frequency
Frequency range 9 kHz to 3 GHz
Frequency offset
Setting range 0 Hz to 3 GHz
Resolution 0.1 Hz
Spectral purity
SSB phase noise 10 kHz carrier offset
< -90 dBc (1Hz)
9 kHz ≤ f ≤ 3 GHz
Level
Level setting range 0 dBm to -50 dBm
Resolution 0.1 dB
Max. deviation of output level 9 kHz to 3 GHz, 20° C to
Spurious
Harmonics output level -10 dBm < -20 dBc
Nonharmonics output level 0 dBm < -30 dBm
< 1 dB
30° C,
50 kHz ≤ RBW ≤ 3 MHz
Operating manual, 08/2005 0-15 E-1147.1000.04
Page 16
Data Sheet R&S FS315
Inputs
RF input
Connector N female (front panel)
Input impedance 50 Ω
VSWR RF attenuation 20 dB < 1.5
LO radiation RF attenuation 10 dB < -50 dBm
External trigger input
Connector N female (front panel)
Voltage level TTL
Reference frequency input
Connector BNC female (rear panel)
Reference frequency
Impedance 50 Ω
Input level 0 dBm to 20 dBm
10 MHz ± 50 Hz
Outputs
RF out (tracking generator)
Connector N female (front panel)
Output impedance 50 Ω
VSWR < 1.6
Reference frequency output
Connector BNC female (rear panel)
Reference frequency 10 MHz
Output impedance 50 Ω
Output level 7 dBm, nominal
AF output
Connector 3.5 mm mini jack for
Output impedance 15 Ω
headhhone (rear panel)
E-1147.1000.04 0-16 Operating manual, 08/2005
Page 17
R&S FS315 Data Sheet
Interfaces
USB host device-specific command set,
remote control via supplied
Windows driver (Windows XP/2000)
USB device B plug, protocol version 1.1
Connector for external monitor
(VGA)
Keyboard connector PS/2 female
15-pin D-SUB female
A plug, protocol version 1.1
General Data
Display
Type 5.4” active TFT color display
Resolution 320 x 240 pixels
Max. refresh rate 10 pictures/s, nominal
Power supply
Input voltage range autoranging 100 V to 240 V (AC)
Power consumption < 60 W
50 Hz to 60 Hz
Ambient conditions
Permissible temperature range meets DIN EN 60068-2-1/2 + 5° C to + 45° C
Storage temperature range -20° C to + 70° C
Relative humidity meets DIN EN 60068-2-3
Mechanical resistance
Sinusoidal vibration meets DIN EN 60068-2-6,
Random vibration meets DIN EN 60068-2-64 10 Hz to 500 Hz: 1.9g
Shock meets DIN EN 60068-2-27 and
Electromagnetic compatibility
EMI field strength
Safety
Dimensions (W x H x D)
Weight
95 % at + 40° C
(non-condensing)
5 Hz to 150 Hz: max. 2g at
DIN EN 61010-1 and
MIL-T-28800D class 5
MIL-STD-810
meets EN 55011 class B and
EN 61326 (EMC Directive of EU (89/336/EEC))
10 V/m
DIN EN 616010-1/IEC61010-1 UL3111-1; CSA22.2 No:1010.1
219 mm x 147 mm x 350 mm
approx. 9 kg
5 Hz,
55 Hz to 150 Hz: 0.5g constant
shock spectrum
Operating manual, 08/2005 0-17 E-1147.1000.04
Page 18
Safety Instructions R&S FS315
Safety Instructions
Note
Safety-related
symbols used on
equipment and
documentation
from R&S
This unit has been designed and tested in accordance with the EC Certificate
of Conformity and has left the manufacturer’s plant in a condition fully complying with safety standards.
To maintain this condition and to ensure safe operation, the user must observe all instructions and warnings given in this operating manual.
Observe operat-
ing instructions
Danger!
Shock hazard
Weight
indication for units
> 18 kg
Warning!
Hot surfaces
PE terminal Ground
Ground Attention!
Electrostatic sen-
require special
terminal
sitive devices
care
E-1147.1000.04 0-18 Operating manual, 08/2005
Page 19
R&S FS315 Safety Instructions
Safety 1.
instructions
The unit may be used only in the operating conditions and positions specified
by the manufacturer. Unless otherwise agreed, the following applies to R&S
products:
IP degree of protection 2X, pollution severity 2 overvoltage category 2, only
for indoor use, altitude max. 2000 m.
Unless specified otherwise in the data sheet, a tolerance of ±10 % shall apply
to the nominal voltage and of ±5 % to the nominal frequency.
For measurements in circuits with voltages V
2.
> 30 V, suitable measures
rms
should be taken to avoid any hazards (using, for example, appropriate measuring equipment, fusing, current limiting, electrical separation, insulation).
If the unit is to be permanently wired, the PE terminal of the unit must first be
3.
connected to the PE conductor on site before any other connections are
made. Installation and cabling of the unit to be performed only by qualified
technical personnel.
For permanently installed units without built-in fuses, circuit breakers or simi-
4.
lar protective devices, the supply circuit must be fused such as to provide
suitable protection for the users and equipment.
Prior to switching on the unit, it must be ensured that the nominal voltage set
5.
on the unit matches the nominal voltage of the AC supply network.
If a different voltage is to be set, the power fuse of the unit may have to be
changed accordingly.
Units of protection class I with disconnectible AC supply cable and appliance
6.
connector may be operated only from a power socket with earthing contact
and with the PE conductor connected.
It is not permissible to interrupt the PE conductor intentionally, neither in the
7.
incoming cable nor on the unit itself as this may cause the unit to become
electrically hazardous.
Any extension lines or multiple socket outlets used must be checked for compliance with relevant safety standards at regular intervals.
If the unit has no power switch for disconnection from the AC supply, the plug
8.
of the connecting cable is regarded as the disconnecting device.
In such cases it must be ensured that the power plug is easily reachable and
accessible at all times (length of connecting cable approx. 2 m). Functional or
electronic switches are not suitable for providing disconnection from the AC
supply.
If units without power switches are integrated in racks or systems, a disconnecting device must be provided at system level.
Applicable local or national safety regulations and rules for the prevention of
9.
accidents must be observed in all work performed.
Prior to performing any work on the unit or opening the unit, the latter must be
disconnected from the supply network.
Any adjustments, replacements of parts, maintenance or repair may be carried out only by authorized R&S technical personnel.
Only original parts may be used for replacing parts relevant to safety (e.
g.
power switches, power transformers, fuses). A safety test must be performed
after each replacement of parts relevant to safety (visual inspection, PE conductor test, insulation-resistance, leakage-current measurement, functional
test).
Ensure that the connections with information technology equipment comply
10.
with IEC950/EN60950.
Operating manual, 08/2005 0-19 E-1147.1000.04
Page 20
Safety Instructions R&S FS315
Lithium batteries must not be exposed to high temperatures or fire.
11.
Keep batteries away from children.
If the battery is replaced improperly, there is danger of explosion. Only replace the battery by R&S type (see spare part list).
Lithium batteries are suitable for environmentally-friendly disposal or specialized recycling. Dispose them into appropriate containers, only.
Do not short-circuit the battery.
Equipment returned or sent in for repair must be packed in the original pack-
12.
ing or in packing with electrostatic and mechanical protection.
Electrostatics via the connectors may damage the equipment. For the safe
13.
handling and operation of the equipment, appropriate measures against electrostatics should be implemented.
The outside of the instrument is suitably cleaned using a soft, lint-free dust-
14.
cloth. Never use solvents such as thinners, acetone and similar things, as
they may damage the front panel labeling or plastic parts.
Any additional safety instructions given in this manual are also to be ob-
15.
served.
E-1147.1000.04 0-20 Operating manual, 08/2005
Page 21
R&S FS315 Certificate of Quality
Certificate of Quality
Certificate of
quality
Dear Customer,
You have decided to buy a Rohde & Schwarz product. You are thus assured
of receiving a product that is manufactured using the most modern methods
available. This product was developed, manufactured and tested in compliance with our quality management system standards.
The Rohde & Schwarz quality management system is certified according to
ISO 9001.
Operating manual, 08/2005 0-21 E-1147.1000.04
Page 22
EC Certificate of Conformity R&S FS315
EC Certificate of Conformity
E-1147.1000.04 0-22 Operating manual, 08/2005
Page 23
R&S FS315 Customer Support
Customer Support
Technical support –
where and when you
need it
Up-to-date information and upgrades
Feedback
Customer support
center
For quick, expert help with any Rohde & Schwarz equipment, contact one of
our Customer Support Centers. A team of highly qualified engineers provides
telephone support and will work with you to find a solution to your query on
any aspect of the operation, programming or applications of Rohde &
Schwarz equipment.
To keep your Rohde & Schwarz equipment always up-to-date, please subscribe to an electronic newsletter at
ROHDE & SCHWARZ International GmbH
Liaison Office Istanbul
Bagdad Cad. 191/3, Arda Apt. B-Blok
81030 Selamicesme-Istanbul
ROHDE & SCHWARZ
Representative Office Kiev
4, Patris Loumoumba ul
252042 Kiev
ROHDE & SCHWARZ International GmbH
Liaison Office Middle East
P.O. Box 31156
Abu Dhabi
Professional Mobile Radio
ROHDE & SCHWARZ Bick Mobile Communication
P.O.Box 17466
Dubai
Service Center for the Middle East
ROHDE & SCHWARZ Emirates L.L.C.
ESNAAD Premises at Mussafah
P.O.Box 31156
Abu Dhabi
ROHDE & SCHWARZ UK Ltd.
Ancells Business Park
Fleet
Hampshire
GU 51 2UZ
AEROMARINE S.A.
Cerro Largo 1497
11200 Montevideo
Service and Calibrationa Center:
Rohde & Schwarz, Inc.
8661-A Robert Fulton Drive
Columbia, MD 21046-2265
R&D and Application Support
Rohde & Schwarz Inc.
8905 SW Nimbus Ave., Suite 240
Beaverton, OR 97008
Central Regional Office:
Rohde & Schwarz Inc.
8080 Tristar Drive
Suite 120
Irving, Texas 75063
ROHDE & SCHWARZ
Representative Office Vietnam, Unit 807, 8/F,
Schmidt Tower
239 Xuan Thuy Road
Cau Giay District Hanoi
+216 (71) 77 33 88
+216 (71) 77 05 53
+90 (216) 385 19 17
+90 (216) 385 19 18
rsturk@superonline.com
+38 (044) 268 83 64
+38 (044) 268 60 55
rsbkiev@public.ua.net
+971 (2) 633 56 70
+971 (4) 633 56 71
michael.rogler@rsd.rohde-schwarz.com
+971 (4) 883 71 35
+971 (4) 883 71 36
+971 (2) 55 49 411
+971 (2) 55 49 433
rsuaeam@emirates.net.ae
+44 (1252) 81 88 88
+44 (1252) 81 14 47
sales@rsuk.rohde-schwarz.com
+598 (2) 400 39 62
+598 (2) 401 85 97
mjn@aeromarine.com.uy
+1 (410) 910 50 02
+1 (410) 910 79 31
rsatv@rsa.rohde-schwarz.com
rsacomms@rsa.rohde-schwarz.com
+1 (503) 40 34 700
+1 (503) 40 34 701
info@rsa.rohde-schwarz.com
+1 (469) 713 53 00
+1 (469) 713 53 01
info@rsa.rohde-schwarz.com
+84 (4) 834 20 46
svnhn@schmidtgroup.com
E-1147.1000.04 0-32 Operating manual, 08/2005
Page 33
R&S FS315 Applications for the R&S FS315
1 Introduction
This chapter
Further
information
Chapter 1 describes the uses of the R&S FS315 and also provides information on the storage and transportation procedures that should be adopted for
the instrument. Warranty conditions are also explained.
Chapter 2 contains an overview of the R&S FS315’s control elements, indicators, etc.
Chapter 3 describes how to put the R&S FS315 into operation.
1.1 Applications for the R&S FS315
Uses
Performance
features
RF signals are used by telecommunication systems to transmit information.
The transmitted power must remain with strictly specified frequency bands,
otherwise interference with other services on adjacent channels would occur.
To ensure that the stringent requirements laid down for RF communication
systems are met, it is essential to have measuring equipment that can precisely measure and analyze the modulated and unmodulated signals involved.
The Spectrum Analyzer R&S FS315 has all the performance features required to perform precise level and frequency measurements.
The key features are:
High-Quality RF characteristics
Operation from
keypad
Remote control
via PC
Resolution bandwidth from 200 Hz to 1 MHz
Frequency counter with 1 Hz resolution
Maximum input level +33 dBm
Ergonomic user interface
All functions and measurement parameters can be set via menus using a
keypad and a rotary knob.
The current trace, along with the key parameters and status indicators
needed to evaluate the results, can be read at a glance from the TFT colour
screen.
The R&S FS315 is equipped as standard with a USB interface so that it can
communicate with a PC. All functions and parameters can be set.
Operating manual, 08/2005 1-33 E-1147.1000.04
Page 34
Supplied Accessories R&S FS315
1.2 Supplied Accessories
Content
1 power cord Europe
1 country specific power cord (if different from Europe)
1 manual German/English
1 USB cable
1 CD (Content: operating manual German/English, data sheet German/English PC software R&S FS315-K1,
Acrobat Reader™)
1.3 Warranty
Warranty
conditions
Returning a
defective R&S
FS315
Indicating claims
under the
warranty
The General Terms and Conditions of Rohde & Schwarz shall apply.
Please observe the packaging instructions ( 0-18). You will find the addresses of your nearest R&S representative and of the support center at the
front of the manual.
We would also ask you to state clearly if you are making a claim for repairs
under warranty, preferably by including your delivery note. Repair requests
that do not explicitly refer to the warranty will, in the first instance, incur
charges.
If your warranty has expired, we will, of course, repair your R&S FS315 in
accordance with our general installation and service conditions.
15 Input connector for external trigger
16 Connector for external USB host
17 Output connector for USB device
g. printer)
(e.
18 Power supply connector
19 AC line fuses
171615
18
20 AC line switch
21 Connector for external monitor
22 Connector for external keyboard
23 Output of internal/external reference
(10 MHz)
24 Input for external reference (10 MHz)
25 Headphone output
E-1147.1000.04 2-36 Operating manual, 08/2005
Page 37
R&S FS315 Unpacking the R&S FS315
3 Putting the R&S FS315 into Operation
This chapter
Further
information
Caution
Chapter 3 describes how to put the R&S FS315 into operation.
Chapter 2 contains an overview of the R&S FS315‘s control elements, indicators, etc.
Chapter 4 “Getting started”, takes you step-by-step through a number of simple measurements.
Chapter 8 is an in-depth description of the instrument’s interfaces.
Before putting the R&S FS315 into operation, make the following checks:
Ensure that the ventilation holes are free of obstructions.
Ensure that there are no unsuitable signal voltages connected to the
input.
The R&S FS315’s outputs may not be overloaded and correct polarity
must be ensured.
The instrument may be damaged if the above checks are not performed.
3.1 Unpacking the R&S FS315
Recommended
procedure
When you unpack the R&S FS315, proceed as follows:
1. Remove the R&S FS315 from its packaging and check that the delivery
is complete using the accessory list ( 1-34).
2. Carefully check the R&S FS315 for any damage.
3. If there is damage, immediately contact the carrier who delivered the
instrument. Under these circumstances, it is essential to keep the box in
which the R&S FS315 was transported and the packaging material ( 0-
18).
3.2 Setting up the Instrument
Caution
Setup instructions
Operating manual, 08/2005 3-37 E-1147.1000.04
There is a risk of injury from sharp edges and becoming wedged between the
setting lever and the handle.
The R&S FS315 must be assembled on a firm, level surface only. The instrument has a carrying handle which is also used for various setup options.
This handle can be moved into any position, depending on the particular field
of application.
Page 38
Setting up the Instrument R&S FS315
Setting the handle 1. Place the thumb and two fingers around the side-mounted setting lever
and loosen it with a turning action.
2. Slide the handle lengthwise while twisting it radially in steps of about 12°.
3. Close the setting lever by pressing on the outer surface.
Caution: There is a risk of injury from sharp edges and becoming
wedged between the setting lever and the handle.
4. Remove the protective film from the screen glass if necessary.
Caution: Use no pointed or sharp objects.
E-1147.1000.04 3-38 Operating manual, 08/2005
Page 39
R&S FS315 Connecting the R&S FS315 to the AC Line
3.3 Connecting the R&S FS315 to the AC Line
Caution
The R&S FS315 meets the requirements for Safety Class I to DIN EN 610101/IEC 61010-1, e.
g. all metal parts that can be touched or accessed without
removing the enclosure are connected to the protective ground of the power
supply network.
The connection to the AC line is made via a power cord and a connector with
a protective ground contact.
Automatic AC line
voltage
detection
Connecting the
AC line
When the R&S FS315 is connected to the AC line, it automatically sets itself
to the correct voltage (range: AC voltage 100 V to 240 V, AC frequency 50 Hz
to 60 Hz). There is no need to set the voltage manually or change the fuse.
1. Use the supplied power cord to connect the R&S FS315 to the AC line.
The power supply connector [18] is at the rear of the instrument.
2. Connect the power cord to the AC line.
3.4 Switching On the R&S FS315
Hazard
AC line switch on
the rear panel
The AC line is still connected to the R&S FS315 when the instrument is in the
standby mode.
The R&S FS315 is connected to the AC line via power supply connector [18].
AC line switch [20] which isolates the R&S FS315 from the AC line is located
next to the power supply connector.
ON/STANDBY
switch on the
front panel
ON operating state
After switching on the R&S FS315 by means of the AC line switch [20] at the
rear panel, it is in standby mode and the yellow LED [2] comes on. If you
press the ON/STANDBY switch [1], the instrument is switched on and the
green LED [2] comes on.
STANDBY operating state
To switch the R&S FS315 from the operating mode to standby mode, press
the ON/STANDBY switch [1] for approx. 2 seconds. After switching off the
ON/STANDBY switch [1] the yellow LED [2] comes on.
Switching on
the R&S FS315
1. Press the AC line switch [20] on the rear panel in the I position.
2. Press the ON/STANDBY switch [1] on the front panel.
The green ON LED [2] should come on.
Operating manual, 08/2005 3-39 E-1147.1000.04
Page 40
Function Test R&S FS315
3.5 Function Test
Caution
Function test
In error case
Note
When performing service procedures, follow the requirements of VDE 0701.
Only properly qualified technicians are allowed to repair the R&S FS315. The
instrument does not contain any parts the operator can repair.
After the R&S FS315 has been switched on ( 3-39), the green LED ON [2]
on the instrument’s front panel comes on. During booting, the "R&S Smart
Instruments" symbol appears on a blue screen background [14]. Booting the
R&S FS315 is completed when the measurement mask ( 5-50) and the
trace appear. If the measurement trace does not appear on the screen, the
sweep time ( 6-126) may have been too long.
If the measurement mask ( 5-50) does not appear and the red or the red or
green LED flash alternately, switch the R&S FS315 off and on. In case the
error continues, return the instrument to our service center for checking.
If the instrument shows an error message after booting, then follow the
instructions in the chapter “Error messages” ( 9-272).
If the measurement trace does not appear on the screen shortly after booting,
the sweep time may have been too long. In this case, reset the R&S FS315
by means of PRESET ( 6-169). If the trace still fails to appear, an error occurred and our service center must check the instrument.
3.6 EMC
EMC
requirements
The R&S FS315 meets the EMC Directive 89/336/EEC (applied standards
EN 55011 Class B and EN 61326).
To prevent EMI, the R&S FS315 may only be operated with its enclosure
closed. Only appropriately shielded signal and control cables may be used.
External units, such as keyboard, printer or monitor, that are to be connected
to the R&S FS315 must comply with EMC directives.
E-1147.1000.04 3-40 Operating manual, 08/2005
Page 41
R&S FS315 Connecting an External Keyboard
[
]
[
[
]
3.7 Connecting an External Keyboard
Caution
Use
Key assignment for
operating the
R&S FS315
Only connect the keyboard when the R&S FS315 is off or in the STANDBY
mode, otherwise malfunctions may occur at a later date.
You can connect an external PC keyboard via the 6-pin PS/2 KEYBOARD
connector [22] on the R&S FS315’s rear panel. The keyboard makes it easier
to enter file names. The keyboard allows data entry as well as operation of
the R&S FS315.
The keyboard is detected automatically when it is connected.
[ F10 ] ⇐ | ⇒ [ F11 ]
F1 ] ... [ F7
[ A ] ... [ D ]
[ F9 ]
[ Esc ]
Enter
[ ], [ ]
[ ], [ ]
0 ] ... [ 9 ], [ . ], [ - ]
Operating manual, 08/2005 3-41 E-1147.1000.04
Page 42
Connecting a USB Stick R&S FS315
3.8 Connecting a USB Stick
Use
Caution
To ensure that the USB stick is detected by the R&S FS315, the stick must
be formatted in the FAT32 file system.
You can connect an external USB stick to the USB device interface [17] at the
rear of the R&S FS315. The USB stick is an extension of the internal memory. You can use it to save or load instrument settings, to print into a file on
the USB stick or to transfer trace data to a PC. You can also use it to load
transducer lists ( 6-164) that have been created using the PC software, to
the instrument.
E-1147.1000.04 3-42 Operating manual, 08/2005
Page 43
R&S FS315 Level and Frequency Measurements
4 Getting Started
This chapter
Further
information
Chapter 4 uses a number of simple measurements to illustrate how to operate the R&S FS315.
For the following example, the initial setting for instrument is the default setting (factory). This is set by pressing the PRESET key in the menu ( 6-169).
The full default setup is described in chapter 6.
Chapter 5 contains an in-depth explanation of the basic operating steps, for
example selecting menus and setting parameters. The layout of the screen
and the information displayed on the screen are also described.
Chapter 6 describes all the R&S FS315‘s menus and the associated functions in detail.
4.1 Level and Frequency Measurements
4.1.1 Measurement Task
Caution
Measurement
problem
The input stage of the R&S FS315 can be destroyed by overloads or DC
components. If there is a possibility that the limits specified in the data sheet
may be violated, the input must be protected with an attenuator and/or a DC
block.
Determining the level and frequency of a signal is one of the most common
measurements a spectrum analyzer is used to make. Often, when making
these measurements on an unknown signal, the PRESET settings (factory)
are chosen as a suitable initial setup ( 6-169).
Solution
The center frequency (CENTER), the SPAN and the MARKER functions play
a key role in level and frequency measurements.
Operating manual, 08/2005 4-43 E-1147.1000.04
Page 44
Level and Frequency Measurements R&S FS315
4.1.2 Measurement Procedure
Introduction
Measurement
steps
In this example, a signal with a frequency of 200 MHz and a level of -30 dBm
is applied to the RF input [8]. The center frequency and the span are set
manually.
Perform the following steps:
1. Reset the R&S FS315.
• Press the
• Select
• Press the
key.
in the bottom menu bar using the
function key.
cursor keys.
2. Apply the signal.
3. Set the center frequency (CENTER) to 200 MHz.
• Select
in the bottom menu bar using the
• Press the
• Enter
using the numerical keys. Terminate the entry
by pressing the unit key
function key.
.
cursor keys.
4. Reduce the SPAN to 1 MHz.
• Stay in the
menu.
• Press the
• Enter
the unit key
function key.
from the numerical keys. Terminate the entry by pressing
.
5. Measure the level and frequency with the marker.
• Select
• Press the
• Press the
marker jumps to the signal peak. Turn the rotary knob
in the bottom menu bar using the
cursor keys.
function key.
function key in the submenu that appears. The
to
change the position of the marker.
6. Frequency measurements with the built-in frequency counter.
• Press the
function key in the submenu. You can
change the resolution of the frequency counter in the same submenu
with the
function key.
E-1147.1000.04 4-44 Operating manual, 08/2005
Page 45
R&S FS315 Measuring the Frequency Response of a Twoport
4.2 Measuring the Frequency Response of a Twoport
4.2.1 Measurement Task
Caution
Initial situation
Solution
If the DUT is an amplifier, the input stage of the R&S FS315 can be destroyed
by overloads or DC components. If there is a possibility that the limits specified in the data sheet could be violated, the input must be protected with a
power attenuator and/or a DC block.
It is important to ensure that the set output power of the tracking generator
does not exceed the maximum permissible input power of the DUT.
Determining the frequency response of a twoport (e.g. filter, amplifier) is a
common measurement task and requires the built-in tracking generator. The
measurement described below is performed based on the PRESET settings
(factory) of the R&S FS315 ( 6-169).
To measure the frequency response of a twoport, the DUT must be connected between the output of the tracking generator [9] and the input of the
spectrum analyzer [8]. With directional DUTs (e.g. amplifiers), make sure that
the input and output of the DUT are connected correctly:
DUT input Generator output [9]
DUT output Analyzer input [8]
You should select a span that will make the transmission characteristics of
the DUT visible. If you want the effect of the cables or an attenuator to be
omitted from the measurement result, you can perform calibration.
4.2.2 Measurement
Introduction
Operating manual, 08/2005 4-45 E-1147.1000.04
In the example below, a bandpass filter with a center frequency of 1 GHz and
a bandwidth of 80 MHz is to be measured.
Page 46
Measuring the Frequency Response of a Twoport R&S FS315
Measurement steps
Perform the following steps:
1. Reset the instrument.
• Press the
• Select
• Press the
key.
in the menu bar using the cursor keys.
function key.
2. Set the center frequency (CENTER) to 1 GHz.
• Select
• Enter
the unit key
in the menu bar using the cursor keys.
using the numerical keys. Complete the entry by pressing
.
3. Reduce the SPAN to 150 MHz.
• Stay in the
• Press the
• Enter
by pressing the unit key
menu.
function key.
using the numerical keys. Complete the entry
.
4. Switch on the tracking generator.
• Select
• Open the
• Press the
• Turn the rotary knob
in the menu bar using the cursor keys.
submenu.
function key.
to select the “Tracking Generator” set-
ting.
5. Calibrate the measurement.
• Use cables to connect the output of the tracking generator [9] to the
input of the spectrum analyzer [8].
• Press the
function key.
• Remove the connection between the input and output.
6. Connect the DUT between the N connectors on the front panel.
• Connect the DUT between the N connectors on the front panel. With
directional DUTs (e. g. amplifiers), make sure that the input and output of the DUT are connected correctly:
DUT input Generator output [9]
DUT output Analyzer input [8]
7. Change the position of the trace.
• Press the
function key and change the value using the key-
board, rotary knob or cursor keys.
E-1147.1000.04 4-46 Operating manual, 08/2005
Page 47
R&S FS315 Making Entries from the Keypad
5 Manual Operating Concept
This chapter
Further
information
Chapter 5 contains an overview of the R&S FS315’s basic manual operating
concept. This includes a description of the keypad, the screen layout, menu
operation and how to set parameters. There is an overview of the menus and
functions at the end of this chapter.
Chapter 4 contains a brief introduction that takes you step-by-step through
some simple measurements.
Chapter 6 contains an in-depth description of the menu functions.
Chapter 7 contains note for remote control the R&S FS315 via a USB interface.
5.1 Making Entries from the Keypad
Introduction
The R&S FS315 is operated using menus in conjunction with a keypad and a
rotary knob. The keypad comprises the following sections:
The numerical keys are used to enter numerical parameters.
− Inserts one of the digits “0” to “9” at the cursor position.
− Inserts a decimal point “.” at the cursor position.
− Inserts a minus sign “-” at the cursor position.
5.1.2 Unit Keys
Function
The unit keys are used to assign the appropriate unit to the number that has
been entered, simultaneously terminating the entry.
− Assigns GHz when a frequency setting is being made
− Assigns s when a time setting is being made
− Assigns MHz when a frequency setting is being made
− Assigns ms when a time setting is being made
− Assigns kHz when a frequency setting is being made
− Assigns µs when a time setting is being made
− Assigns Hz when a frequency setting is being made
Note
Operating manual, 08/2005 5-47 E-1147.1000.04
In the case of all other entries, the unit keys assume the same function as the
Enter key. ( 5-49).
Page 48
Making Entries from the Keypad R&S FS315
5.1.3 Rotary Knob
Function
As well as the numerical keys and the cursor keys, the rotary knob is also
used to set parameters.
The rotary knob has several functions:
− Incrementing (turn clockwise) or decrementing (turn counter-
clockwise) numerical instrument parameters using a specified step size.
− Positioning markers, limits, etc on the screen.
5.1.4 Cursor Keys
Function
As well as the numerical keys and the rotary knob, the cursor keys are also
used for entering parameters and to navigate through the menus.
The cursor keys have the following functions:
− Navigating through menus and selection fields
− The 3 or4 cursor keys move the cursor to the position you want within
the numerical editing line.
− The 6or 5cursor keys increment or decrement numerical parameter
entries.
5.1.5 Function Keys
Function
In the function area, various instrument functions are displayed depending on
which menu has been selected.
The displayed instrument functions are assigned to the seven function keys
down the right side of the screen. This means that each function key can
have a variety of functions ( 5-52).
When a function key is pressed, various responses can be elicited:
− Immediate activation of a function or toggling between settings
− Entry of a value or selection of a setting/function
− Confirmation of a new setting and opening of a new menu item
− Branching to a submenu
E-1147.1000.04 5-48 Operating manual, 08/2005
Page 49
R&S FS315 Making Entries from the Keypad
5.1.6 Action Keys
Function
The action keys are used to complete menu-guided settings.
− With numerical entries, pressing this key completes an entry without
closing the entry field.
− With a selection field, the setting is adopted and the selection window is
closed.
− Pressing this key assigns Hz as the unit of measurement when a fre-
quency setting is being made.
Note: Pressing a unit key will also complete the entry of setting data.
− Pressing this key closes the entry field or selection field after data has
been entered. The new value is set.
5.1.7 BACK/SYS Key
Function 1
SYS
If the entry field is closed the BACK/SYS [3] key will open or close the system menu (SYS).
− When you press the key, the measuring menu is blanked out and re-
placed by the SYS menu. Other functions are assigned to the function
keys [13] and the measurement diagram is replaced by the system parameters ( 6-167).
− By repeatedly pressing the key, you can quit the SYS menu and accept
the new settings.
Function 2
If the entry field is open the BACK/SYS [3] key can be used to correct numeri-
cal inputs.
BACK
− If the parameter entry field is open, the key functions as a BACK key (i.e.
a value entered using the keyboard can be deleted again one character
at a time). The key has no effect if the value in an entry field was
changed using the cursor keys or rotary knob.
Operating manual, 08/2005 5-49 E-1147.1000.04
Page 50
Screen Display R&S FS315
5.2 Screen Display
Introduction
Screen layout
The screen [14] provides on-going information about events and the parameters associated with the selected measurement functions.
The display mode for the measurement results, the lettering of the function
keys and the type of menu all depend on the current settings.
The screen is divided into three areas:
I Diagram area
II Menu area
III Function area
I
III
II
E-1147.1000.04 5-50 Operating manual, 08/2005
Page 51
R&S FS315 Screen Display
5.2.1 Diagram Area
Displays in the
diagram area
The diagram area contains:
Measurement diagrams and the associated scales (h) and traces (e)
Measured value displays, e.
g. display line (d) and markers (c)
Parameter field (b) and status displays (a)
Selection fields and entry fields that come up on the screen (g)
Error messages that come up on the screen (f)
b
a
c
d
h
e
f
g
Status display
Parameter field
Measurement
diagram
The status display provides the following information:
Offs - level offset is on
MIN - current trace display mode, e.
The following are displayed in the parameter field:
A 10 x 8 grid is superimposed on the diagram area to make it easier to analyze traces.
The following settings are displayed in the measurement diagram:
SAMPLE- selected detector (Min Peak, Max Peak, Sample,
RMS, Average)
M1: - marker 1 plus marker position and level
M2: - marker 2 plus marker position and level
CH Pwr: - channel power at set channel bandwidth
TG: - level, frequency and calibration of tracking generator
Operating manual, 08/2005 5-51 E-1147.1000.04
Page 52
Screen Display R&S FS315
5.2.2 Menu Area
Menu display
Menus for setting the measurement parameters and the measurement functions are displayed in the menu area. The selected menu is highlighted, e.
FREQ/SPAN menu:
5.2.3 Function Area
Displaying the
current assignment
for the function keys
When a menu is selected, the associated instrument functions are displayed
in the function area.
The displayed instrument functions are assigned to the seven function keys
down the right-hand side of the screen. If a key in the function area does not
have any lettering, the key has not been assigned a function in the menu in
question. If a key has lettering, but not in full brightness, the key has temporarily (current setting) not been assigned a function.
g.
Activated function key
De-activated function key
No function assigned
E-1147.1000.04 5-52 Operating manual, 08/2005
Page 53
R&S FS315 Calling and Changing Menus
5.3 Calling and Changing Menus
Introduction
Calling or
changing menus
Operating the R&S FS315 is menu-guided. All the menus used to set the
measurement parameters and measurement functions are displayed in the
menu area. The instrument functions associated with any menu you select
are displayed in the function area.
Pressing a function key has one of the following effects:
Direct execution of an instrument function
Opening of entry or selection windows
Opening of submenus
The 3 or 4 cursor keys [6] are used for menu navigation.
1. Select a menu, e.
g. , with the
3
or
[6] cursor keys.
4
The menu name is highlighted and the appropriate function is assigned
to the function keys [13].
Note
A double arrow ⇒ pointing to a function key, e.
pressing this key will call a submenu.
g. , tells you that
Operating manual, 08/2005 5-53 E-1147.1000.04
Page 54
Calling and Changing Menus R&S FS315
Calling/quitting
submenus
2. Press the
function key in the
menu.
The MARKER 1 submenu opens and the new functions are assigned to
the function keys [13].
3. Press the
function key in the submenu.
The submenu is closed and the previous functions remain assigned to
the function keys [13].
E-1147.1000.04 5-54 Operating manual, 08/2005
Page 55
R&S FS315 Setting Parameters
5.4 Setting Parameters
There is a choice
of methods
Parameters can be set in a number of ways:
Direct selection of an instrument function (function key)
Selecting settings from selection fields
Entering numerical parameters in entry fields
The numerical keys [12], the unit keys [9], the rotary knob [11], the cursor
keys [6, 7] and the action keys [4, 5] can all be used to select and enter instrument parameters.
5.4.1 Direct Selection of Instrument Functions
Introduction
e.
g. :
Displaying marker 1
( 6-95)
When you select a menu, various instrument functions are displayed in the
function area. Some instrument functions can be set directly by pressing a
function key.
The function key you select is highlighted.
1. Press the
function key in the submenu.
Marker 1 is turned on and placed on the trace maximum. The M1 marker
values are displayed in the parameter field.
2. Press the
function key in the submenu.
Marker 1 moves from the maximum value of the trace to the next lowest
value to the right of the maximum value.
3. Press the
function key in the submenu.
Marker 1 disappears from the diagram area and the M1 marker values
are blanked out.
Operating manual, 08/2005 5-55 E-1147.1000.04
Page 56
Setting Parameters R&S FS315
5.4.2 Selecting Settings
Introduction
e.
g. :
Setting the level
display range
( 6-86)
When you select a menu, a number of instrument functions are displayed in
the function area. If certain function keys are then pressed, a selection field is
displayed in the diagram area. You can then choose and activate any of the
settings offered for selection.
The function key you select is highlighted.
1. Press the
A selection field containing the available settings is displayed. The default setting is “80 dB”.
function key in the menu.
Note
2. Select a level display range with rotary knob
3. Press the ENTER key [5] to close the selection field.
The new setting is saved.
If you want to keep the old setting, close the entry field with the
ESC/CANCEL key [4].
If there are more than five options, a scroll bar is displayed on the right next
to the selection field. At any one time, only five options are displayed on the
screen.
[11].
E-1147.1000.04 5-56 Operating manual, 08/2005
Page 57
R&S FS315 Setting Parameters
5.4.3 Entering Numerical Parameters
Introduction
When you select a menu, a number of instrument functions will be displayed
in the function area. If you press certain function keys, an entry field will be
displayed in the diagram area.
The function key you select is highlighted.
There are two ways of entering numerical parameters:
Entry of a number with the numerical keys
Entry of a number with the cursor keys and rotary knob
The entry with the cursor keys and rotary knob is useful if you only have a
rough idea of what the measurement parameters should be. As the screen is
continuously updated as the value is varied, a search is possible.
5.4.3.1 Entry with the Numerical Keys
e.g. :
Setting the
center frequency
( 6-73)
1. Press the
An entry field containing the current setting is displayed in the top lefthand corner.
function key in the menu.
Entering
a new value
2. Overwrite the old value, e.
g. 1234 MHz, with the numerical keys [12].
The frequency is adopted immediately in the selected unit and set by the
R&S FS315. The entry window remains open and the value can be
overwritten by a new value at any time.
Note: If you press ENTER instead of a unit key, the smallest unit will be
assigned (e.g. Hz).
Operating manual, 08/2005 5-57 E-1147.1000.04
Page 58
Setting Parameters R&S FS315
Correcting an entry 3. Press the SYS/BACK key [3] repeatedly until you have deleted the incor-
rect digit (e.g. the second two in 1.224 MHz).
2 ×
Then enter the correct final digits.
Terminating
parameter entry
Note
entries
Invalid
4. Press the ESC/CANCEL key [4] to abort the entry.
The old value isretained. The entry window is closed.
• Invalid parameter entry (e.g. 1234 GHz): If the entered value is outside
the permissible range, the largest or smallest permissible value is set
and a message containing the value range appears below the entry window.
When numbers are displayed, only the digit sequences 1 to 999 appear before the decimal point. In other words, if the digit sequence <5000> and the
unit <kHz> are entered, <5 MHz> appears in the display. Trailing zeros will
be eliminated in the display area.
E-1147.1000.04 5-58 Operating manual, 08/2005
Page 59
R&S FS315 Setting Parameters
5.4.3.2 Entry using the Cursor Keys and the Rotary Knob
e.g. :
Setting the span
( 6-73)
Entering
a new value
1. Press the
An entry field containing the current setting appears in the top left-hand
corner.
The step width for the change is different for the cursor keys and rotary
knob. The cursor keys change the value in large steps, whereas the rotary knob allows a more precise setting. The span is changed in the following preset step widths:
Cursor keys: Steps 1-2-5-10
function key in the menu.
Note
Rotary knob: The third digit is changed.
2a. Press the cursor key
Pressing the
the 6 cursor key once decrements the value by one.
cursor key once increments the value by one; pressing
5
[7] once so that a span of 2 GHz is displayed.
6
1 ×
2b. Turn the rotary knob [11] to left until you obtain the value you want, e. g.
1.95 GHz.
Turning clockwise increases the value; turning counter-clockwise reduces the value.
When you enter the center frequency, you can specify the step width of the
rotary knob. ( 6-80). In this case, the value is incremented or decremented
in accordance with the specified step width, with the cursor always moving to
the highest value, irrespective of its earlier position.
Operating manual, 08/2005 5-59 E-1147.1000.04
Page 60
Setting Parameters R&S FS315
Terminating
entries
3. Press the ESC/CANCEL key [4] to abort the entry.
The old value is retained. The entry window is closed.
parameter entry
Note
Invalid
If the limit value is reached, the numeric value in the entry window remains
the same and is neither increased nor decreased. No error message is issued.
When numbers are displayed, only the digit sequences 1 to 999 appear before the decimal point. In other words, if the digit sequence <5000> and the
unit <kHz> are entered, <5 MHz> appears in the display. Trailing zeros will
be eliminated in the display area.
E-1147.1000.04 5-60 Operating manual, 08/2005
Page 61
R&S FS315 Overview of all Menus and Functions
5.5 Overview of all Menus and Functions
5.5.1 Spectrum Analyzer
5.5.1.1 FREQ/SPAN Menu
Function key
assignment
Entering the center frequency ( 6-73)
Open submenu:
Display modes for the frequency axis
Entering the start frequency ( 6-79)
Entering the stop frequency ( 6-79)
Open submenu:
Entering the step width of the center frequency
Quitting the submenu
Entering the span ( 6-75)
Displaying the whole frequency range ( 6-76)
Switching over to the ZERO SPAN ( 6-77)
Restoring the previous setting ( 6-77)
Reducing the span ( 6-78)
Increasing the span ( 6-78)
Signal tracking ( 6-82)
Operating manual, 08/2005 5-61 E-1147.1000.04
Quitting the submenu
Setting the step size manually ( 6-81)
Setting the step size automatically ( 6-81)
Setting the step size to the center frequency ( 6-81)
Setting the step size to the marker frequency ( 6-81)
Page 62
Overview of all Menus and Functions R&S FS315
5.5.1.2 AMPT Menu
Function key
assignment
Entering the reference level ( 6-84)
Entering a level offset ( 6-85)
Selecting the level display range ( 6-86)
Selecting a unit for the level display ( 6-87)
Setting the RF input attenuation manually ( 6-89)
Setting the RF input attenuation automatically ( 6-90)
Selecting the setting High Sensitivity ( 6-92)
5.5.1.3 MKR Menu
Function key
assignment
Open submenu:
Activate marker 1
Quitting the submenu
Place marker 1 on the trace maximum ( 6-95)
Place marker 1 on the next trace maximum to the left ( 6-95)
Place marker 1 on the next trace maximum to the right ( 6-95)
Measure the signal frequency: Start measurement ( 6-97)
Measure signal frequency: Set resolution ( 6-97)
De-activate marker 1 ( 6-95)
Open submenu:
Activate marker 2
Quitting the submenu
Place marker 2 on the trace maximum ( 6-99)
Place marker 2 on the next trace maximum to the left ( 6-99)
Place marker 2 on the next trace maximum to the right ( 6-99)
Set marker 2 as a normal marker (NORM) ( 6-100)
Set marker 2 as a delta marker (DELTA) ( 6-100)
De-activate marker 2 ( 6-99)
E-1147.1000.04 5-62 Operating manual, 08/2005
Page 63
R&S FS315 Overview of all Menus and Functions
Open submenu:
Accepting marker values as settings
Quitting the submenu
Set the center frequency to the marker frequency ( 6-102)
Set the marker level as a reference level ( 6-102)
Set the step size for entering the center frequency to the marker
frequency ( 6-104)
Open submenu:
Marker measurement functions
Quitting the submenu
Measuring noise power density ( 6-106)
Measuring signal bandwidth ( 6-108)
Bring display line up on screen ( 6-110)
Open submenu:
Search criterion of functions NEXT PEAK LEFT/RIGHT
Quitting the submenu
Entering the peak excursion ( 6-112)
Open submenu:
Reference points for measuring level differences
Quitting the submenu
Activating the entry of arbitrary reference points ( 6-115)
Using M1 marker values as a reference point ( 6-115)
Entering the reference-point level ( 6-115)
Entering the reference-point frequency ( 6-115)
Entering the reference-point time ( 6-115)
Operating manual, 08/2005 5-63 E-1147.1000.04
Page 64
Overview of all Menus and Functions R&S FS315
5.5.1.4 BW/SWEEP Menu
Function key
assignment
Setting the resolution bandwidth manually ( 6-119)
Activating automatic video bandwidth setting ( 6-120)
Open submenu:
Setting the RBW/VBW coupling ratio
Quitting the submenu
Setting the coupling ratio RBW/VBW manually
( 6-122)
Activating the default setting for the coupling ratio RBW/VBW
( 6-122)
Switching over the coupling RBW/SPAN to low noise
( 6-123)
Open submenu:
Setting the sweep time
Quitting the submenu
Starting a continuous frequency sweep ( 6-125)
Performing an n-times sweep ( 6-125)
Setting the number of sweeps ( 6-125)
Repeating n-times sweeps ( 6-126)
Setting the sweep time manually ( 6-126)
Open submenu:
Evaluating the individual pixels of a trace
Activating automatic sweep-time setting ( 6-126)
PEAK detector is active ( 6-129)
SAMPLE detector is active ( 6-129)
RMS detector is active ( 6-129)
AVERAGE detector is active ( 6-129)
E-1147.1000.04 5-64 Operating manual, 08/2005
Page 65
R&S FS315 Overview of all Menus and Functions
5.5.1.5 TRACE Menu
Function key
assignment
Open submenu:
Selecting the active trace
Quitting the submenu
Turning on and activating Trace 1 ( 6-133)
Turning on and activating Trace 2 ( 6-133)
Trace display mode: Overwrite mode ( 6-135)
Trace display mode: Min./Max. hold ( 6-135)
Trace display mode: Averaging ( 6-135)
Freezing the trace ( 6-135)
Blanking out the trace ( 6-135)
Open submenu:
Trace difference
Quitting the submenu
Turning on the trace-difference mode ( 6-139)
Repositioning Trace 1 (result) ( 6-139)
Turning off the trace-difference mode ( 6-139)
5.5.1.6 TRIG Menu
Function key
assignment
Operating manual, 08/2005 5-65 E-1147.1000.04
Free-running measurements ( 6-141)
Triggering by an external TTL signal ( 6-142)
Triggering by the display level ( 6-141)
Selecting the trigger edge ( 6-142)
Entering a trigger offset ( 6-142)
Page 66
Overview of all Menus and Functions R&S FS315
5.5.1.7 MEAS Menu
Function key
assignment
Open submenu:
Measure the power in the time domain (ZERO SPAN)
Open submenu:
Measuring the channel power
Quitting the submenu
Output the peak value within the section ( 6-145)
Output the mean value within the section ( 6-145)
Output the mean square within the time domain ( 6-145)
Insert a vertical line
to limit the lower (left) part of a section ( 6-145)
Insert a vertical line
to limit the upper (right) part of a section ( 6-145)
Switch off the power measurement ( 6-145)
Quitting the submenu
Measure the channel power ( 6-149)
Set the channel bandwidth ( 6-149)
Measure the occupied bandwidth ( 6-151)
Open submenu:
Configuring the built-in generator
Set the power bandwidth ( 6-151)
Quitting the submenu
Enter the output level ( 6-159)
Enter the frequency or frequency offset ( 6-159)
Calibrate the test setup and activate the calibration ( 6-154)
Enter the reference position ( 6-154)
Enter the level offset ( 6-154)
Select the tracking generator operating mode ( 6-154)
E-1147.1000.04 5-66 Operating manual, 08/2005
Page 67
R&S FS315 Overview of all Menus and Functions
Open submenu:
AM/FM demodulation
Quitting the submenu
Select the AM/FM demodulation mode ( 6-161)
Adjust the headphone volume ( 6-161)
Set the demodulation bandwidth ( 6-161)
Switching off the AM/FM demodulation ( 6-161)
Measure the third-order intercept point ( 6-162)
Open submenu:
Activating the transducer function
Quitting the submenu
Load a transducer set from a USB stick ( 6-165)
Load a transducer set from the internal memory ( 6-165)
Save a transducer set to the internal memory ( 6-165)
Delete a transducer set from the internal memory( 6-165)
Deactivate a transducer set ( 6-165)
Operating manual, 08/2005 5-67 E-1147.1000.04
Page 68
Overview of all Menus and Functions R&S FS315
5.5.2 SYSTEM Functions
5.5.2.1 PRESET Menu
Function key
assignment
Calls an instrument default setting ( 6-170)
Selects an instrument default setting ( 6-170)
5.5.2.2 STATUS Menu
Function key
assignment
Spectrum analyzer settings ( 6-171)
Settings for the channel power measurement ( 6-173)
Tracking generator settings ( 6-174)
Demodulation settings ( 6-175)
Settings for transducer sets ( 6-176)
5.5.2.3 FILE Menu
Function key
assignment
Saves a user-defined setting ( 6-178)
Loads a user-defined setting ( 6-178)
Prints out a screenshot ( 6-180)
5.5.2.4 CONFIG Menu
Function key
assignment
E-1147.1000.04 5-68 Operating manual, 08/2005
Sets the date and time ( 6-185)
Selects an internal or external reference source ( 6-187)
Configures the instrument interfaces ( 6-188)
Sets the screen saver mode ( 6-190)
Selects an internal or external monitor ( 6-193)
Enable new functions
Page 69
R&S FS315 Overview of all Menus and Functions
5.5.2.5 SERVICE Menu
Function key
assignment
Performs a selftest ( 6-194)
5.5.2.6 INFO Menu
Function key
assignment
Displays module data ( 6-196)
Displays instrument statistics ( 6-196)
Displays system messages ( 6-197)
Operating manual, 08/2005 5-69 E-1147.1000.04
Page 70
R&S FS315 Factory Settings R&S FS315
6 Using the R&S FS315
This chapter
Further
information
All the spectrum analyzer functions and their uses are fully described in chapter 6. The order in which the menus are described follows that of the procedures for configuring and starting measurements:
The operating concept is explained in chapter 5, which also contains an overview of the menus and functions. The index at the end of this manual will also
help you find the information you want.
6.1 R&S FS315 Factory Settings
Switching on for the
first time
Frequency axis
Level axis
When you switch on the R&S FS315 ( 3-39), the settings you were using
when the instrument was last switched off are restored. When you switch on
for the first time, the factory default settings are activated:
Parameter Setting
Center frequency (CENTER) 1.5 GHz
Frequency span (SPAN) 3 GHz
Step size for center frequency 150 MHz
Reference level (REF LEVEL) -20 dBm
RF input
Bandwidth
Frequency sweep
Trace
Trigger
Markers
Tracking Generator
(Frequency Fixed)
(Tracking Mode)
Level offset (REF LEVEL OFFSET) 0 dB
Level range (RANGE) 80 dB
Level unit (UNIT) dBm
Input attenuation (RF ATTEN) AUTO (NORMAL)
Sweep time (SWEEP TIME) AUTO
Resolution bandwidth (RES BW) AUTO (1 MHz)
Video bandwidth (VIDEO BW) AUTO (1 MHz)
Sweep CONT SWEEP
Trace CLEAR/WRITE
Trigger FREE RUN
Marker 1 and marker 2 OFF
Mode OFF
Output level -20 dBm
Output frequency 1.5 GHz
Output level -20 dBm
Referency position 100 %
Trace Offset 0 dB
Note
E-1147.1000.04 6-70 Operating manual, 08/2005
The factory default setting is stored in non-volatile memory in the R&S FS315
and can be reloaded at any time ( 6-169).
Page 71
R&S FS315 Spectrum Analyzer
6.2 Spectrum Analyzer
Introduction
User interface
Menus for configuring and starting
measurements
The R&S FS315 measures signal spectra over the selected frequency range
using the selected resolution bandwidth and sweep. The time characteristic
of the amplitude at the set frequency is displayed in zero span.
When the R&S FS315 is switched on and the selftest passed, the spectrum
analyzer’s user interface is activated. The following is displayed on the
screen [14]:
The menus used to set the spectrum analyzer are displayed in the menu
area. The order of the menus mirrors that of the procedure for configuring
and starting measurements:
Selecting the frequency span ( 6-72)
(setting the x axis in the diagram area)
Setting the level axis and the RF input ( 6-83)
(setting the y axis in the diagram area)
Signal analysis with marker functions ( 6-93)
Setting the bandwidths and the sweep time ( 6-118)
Displaying the trace ( 6-131)
Triggering the measurement ( 6-140)
Measurement functions ( 6-143)
Operating manual, 08/2005 6-71 E-1147.1000.04
Page 72
Spectrum Analyzer R&S FS315
6.2.1 Selecting the Frequency Span (FREQ/SPAN Menu)
What the settings
are for
Selecting the
FREQ/SPAN
menu
Function key
assignment
Use the FREQ/SPAN menu to specify the frequency range.
• Select the
menu using the
3
or
cursor key [6].
4
The menu name is highlighted and the appropriate functions are assigned to the function keys [13].
1st variant: Setting the frequency range
Enter the center frequency ( 6-73)
Open submenu:
Enter the SPAN ( 6-75)
2nd variant: Setting the frequency range
Enter the START frequencies ( 6-79)
Enter the STOP frequencies ( 6-79)
Open submenu:
Entering the step width of the center frequency ( 6-80)
Switching on/off signal tracking ( 6-82)
E-1147.1000.04 6-72 Operating manual, 08/2005
Page 73
R&S FS315 Spectrum Analyzer
6.2.1.1 Entering the Center Frequency
Use
Entering the
center frequency
If you know the frequency of the signal you want to measure, you can set the
x axis of the diagram area accordingly. Enter the signal frequency as the
center frequency (CENTER) and specify a SPAN ( 6-75).
1. Press the
function key in the menu.
An entry field containing the current setting is displayed in the top lefthand corner.
2. Enter the new value ( 5-57).
The entry range for the center frequency is:
500 Hz ≤ Center ≤ 2.9999995 GHz
The new setting is saved and displayed in the diagram area.
CENTER
Operating manual, 08/2005 6-73 E-1147.1000.04
Page 74
Spectrum Analyzer R&S FS315
6.2.1.2 Frequency-Axis Display Modes
What the settings
are for
Selecting the
SPAN SETTINGS
submenu
The frequency axis (x axis) can be displayed in a number of ways. Using the
SPAN SETTINGS submenu, you can display the whole frequency range,
zoom in on subranges of the frequency axis or project the diagram area onto
the whole of the screen, if and when you require.
If you switchover to the time domain (ZERO SPAN), you can display the
amplitude of the carrier signal as functions of time.
• Press the
function key in the menu.
The submenu opens and the appropriate functions are assigned to the
function keys [13].
Function key
assignment
Quitting the submenu
Enter the SPAN ( 6-75)
Displaying the whole frequency range ( 6-76)
Switching over to the ZERO SPAN ( 6-77)
Restoring the previous setting ( 6-77)
Zoom function: Reducing the span ( 6-78)
Zoom function: Increasing the span ( 6-78)
E-1147.1000.04 6-74 Operating manual, 08/2005
Page 75
R&S FS315 Spectrum Analyzer
6.2.1.2.1 Entering the Span
Use
Setting the span
If you know the frequency of the signal you want to measure, you can set the
SPAN.
1. Press the
function key in the menu.
An entry field containing the current setting is displayed in the top lefthand corner.
2. Enter the new value ( 5-57).
The span entry range is:
1 kHz ≤ Span ≤ 3 GHz
The new setting is saved and displayed in the diagram area. The grid
lines associated with the frequency axis represent 1/10 of the current
span.
SPAN
Operating manual, 08/2005 6-75 E-1147.1000.04
Page 76
Spectrum Analyzer R&S FS315
6.2.1.2.2 Displaying the Whole Frequency Range
Use
Displaying the
whole frequency
range
Displaying the
last span
If you do not know the frequency of the signal to be investigated, you can
display the R&S FS315’s whole frequency range along the x axis of the diagram area. You can then systematically reduce the span to the appropriate
size ( 6-75).
The whole frequency range display can also be used to swap backwards and
forwards between an overview measurement (FULL SPAN) and a detailed
measurement (LAST SPAN, e.
1. Press the
function key in the
g. center frequency and span set manually).
menu.
The old setting is saved and the R&S FS315’s whole frequency range is
displayed in the diagram area.
2. Press the
The last span setting is displayed in the diagram area, e.
function key in the
menu.
g. 1 MHz:
Note
The last span is restored if and only if the span is > 0, in other words, the time
domain (ZERO SPAN) will not be selected automatically.
E-1147.1000.04 6-76 Operating manual, 08/2005
Page 77
R&S FS315 Spectrum Analyzer
6.2.1.2.3 Switching over to the ZERO SPAN
Use
Specifying
the signal
component
Switching over to
the ZERO SPAN
The amplitude of a particular signal component in the frequency spectrum
can be displayed as a function of time. To activate this mode, enter the frequency of the signal component as the center frequency. You can then define
the x axis to be the time axis.
1. Press the
function key in the menu.
An entry field containing the current setting is displayed in the top lefthand corner.
2. Enter the frequency of the signal component you want to display in the
time domain ( 6-73).
3. Press the
function key in the submenu.
The x axis becomes the time axis and the signal amplitude at the current
center frequency is displayed as a function of time (current sweep time).
The grid line spacing represents 1/10 of the current sweep time ( 6-
124).
Switching over to
the frequency
domain
SWEEP TIME
4. Press the
function key in the
The last span is displayed in the diagram area.
submenu.
Operating manual, 08/2005 6-77 E-1147.1000.04
Page 78
Spectrum Analyzer R&S FS315
6.2.1.2.4 ZOOM Functions
Use
Reducing
the span
Increasing
the span
If you want to increase or decrease the size of the screen window to analyze
traces, you can zoom onto sections of the span or fill the whole screen with
the current screen window.
• Press the
function key in the submenu.
The span is reduced by a factor of two. The center frequency remains
the same.
• Press the
function key in the submenu.
The span is increased by a factor of two. The center frequency remains
the same. The center frequency will be adapted if the new display frequency range exceeds the maximum permissible start or stop frequency.
E-1147.1000.04 6-78 Operating manual, 08/2005
Page 79
R&S FS315 Spectrum Analyzer
6.2.1.3 Entering the Start Frequency and the Stop Frequency
Use
Entering a start
frequency
Entering a stop
frequency
If you want to analyze a specific frequency range, you can enter a start frequency and a stop frequency. The frequency range is shown along the x axis
of the diagram area.
1. Press the
An entry field containing the current setting is displayed in the top lefthand corner.
2. Enter the new value ( 5-57).
The entry range for the start frequency is:
0 ≤ Start ≤ 2.999999 GHz
3. Press the
An entry field containing the current setting is displayed in the top lefthand corner.
4. Enter the new value ( 5-57).
The entry range for the stop frequency is:
1 kHz ≤ Stop ≤ 3 GHz
function key in the . menu.
function key in the menu.
Note
The new setting is saved and displayed in the diagram area. The grid
lines associated with the frequency axis represent 1/10 of the current
span.
STARTSTOP
The effect of pressing the
lettering of the frequency axis in the diagram area to the display showing the
current start and stop frequencies.
or the function key is to switch the
Operating manual, 08/2005 6-79 E-1147.1000.04
Page 80
Spectrum Analyzer R&S FS315
6.2.1.4 Entering the Step Width of the Center Frequency
What the settings
are for
Selecting the
CF STEP SIZE
submenu
Function key
assignment
Using the CF STEP SIZE submenu, you can set any step size of the rotary
knob you want or have the R&S FS315 set a step size that is a function of the
measurement parameters.
• Press the
The submenu opens and the appropriate functions are assigned to the
function keys [13].
Quitting the submenu
function key in the menu.
Note
Setting the step size manually ( 6-81)
Setting the step size automatically ( 6-81)
Setting the step size to the center frequency ( 6-81)
Setting the step size to the marker frequency ( 6-81)
The
only one setting can be activated at any one time. The function key in ques-
tion is highlighted.
The AUTO function is the default setting.
The
and function keys are selection keys. In other words,
function key is available only when marker 1 is turned on.
E-1147.1000.04 6-80 Operating manual, 08/2005
Page 81
R&S FS315 Spectrum Analyzer
6.2.1.4.1 Setting the Step Size
Setting the step size
manually
Setting the
step size
automatically
Setting the
step size to the
center frequency
If you want to examine the frequency spectrum at certain intervals, you can
enter the value of your choice for the step size of the rotary knob.
1. Press the
function key in the
submenu.
An entry field containing the current setting is displayed in the top lefthand corner.
2. Enter the new value ( 5-57).
The entry range for the step size is:
1 Hz ≤ Step Size ≤ 2.999999 GHz
If you want to change the center frequency using steps that are small compared with the span, select the automatic setting mode.
• Press the
function key in the
submenu.
The step size is set to a tenth of the current span.
If you make the step size equal to the center frequency (or an integer multiple
of the center frequency), it is easy to find harmonics which are multiples of
the center frequency.
• Press the
function key in the submenu.
Setting the
step size to the
marker frequency
The step size is made equal to the current center frequency.
If you make the step size equal the marker frequency, you can search
through the frequency spectrum and quickly jump back to the frequency at
which you started.
• Press the
function key in the submenu.
The step size is made equal to the current marker frequency ( 6-101).
Operating manual, 08/2005 6-81 E-1147.1000.04
Page 82
Spectrum Analyzer R&S FS315
6.2.1.5 Activating Signal Tracking
Use
Activating signal
tracking
If the frequency of the signal being analyzed is continually varying, you can
use this function to obtain a display that appears to be stationary. The power
of signals of this kind can then be measured over long periods.
When signal tracking is selected, the center frequency of the R&S FS315 is
continually set to the frequency of the signal with most power. This means
that a signal whose frequency is changing slowly in comparison with the
sweep time can be kept in the center of the screen. The capture range of the
function is ± span/2 referring to the center frequency. The spectrum analyzer’s center frequency is adjusted after each sweep.
• Press the
function key in the submenu.
The center frequency is constantly adjusted so that the signal peak is always in the middle of the screen. The latest center frequency to be set is
displayed. The peak will continue to be tracked provided it remains within
the span referring to the center frequency.
De-activating
signal tracking
• Press the
function key in the submenu.
The center frequency ceases to be tracked and remains at its last value.
E-1147.1000.04 6-82 Operating manual, 08/2005
Page 83
R&S FS315 Spectrum Analyzer
6.2.2 Setting the Level Axis and the RF Input (AMPT Menu)
What the settings
are for
Selecting the
AMPT menu
Function key
assignment
From the AMPT menu, you can define the y axis of the measurement diagram as the level axis. You can also specify the attenuation of the RF input.
• Select the
The menu name is highlighted and the appropriate functions are assigned to the function keys [13].
Entering the reference level ( 6-84)
menu with the
3
or
cursor key [6].
4
Note
Entering a level offset ( 6-85)
Selecting the level display range ( 6-86)
Selecting the level display unit ( 6-87)
Setting the RF input attenuation manually ( 6-89)
Setting the RF input attenuation automatically ( 6-90)
Selecting the setting High Sensitivity ( 6-92)
The
only one setting can be activated at any one time. The function key in ques-
tion is highlighted.
The RF ATTEN AUTO function is the default setting.
The
the unit of the antenna factors is dB.
and function keys are selection keys. In other words,
function key is only available for activated transducer factors if
Operating manual, 08/2005 6-83 E-1147.1000.04
Page 84
Spectrum Analyzer R&S FS315
6.2.2.1 Entering the Reference Level
Use
Note
Entering the
reference level
The upper diagram limit is specified with the reference level.
REF LEVEL
The currently activated unit is used as the unit for reference level entries (
6-87). The default setting is “dBm”.
1. Press the
function key in the menu.
An entry field containing the current setting is displayed in the top lefthand corner. The default setting is -20 dBm.
2. Enter a new value in 0.1 dB steps ( 5-57).
The entry range for the reference level is:
-110 [dBm] ≤ Reference
≤ +36 [dBm]
The new setting is saved and displayed in the diagram area. The reference level defines the upper limit of the diagram.
E-1147.1000.04 6-84 Operating manual, 08/2005
Page 85
R&S FS315 Spectrum Analyzer
6.2.2.2 Entering a Level Offset
Use
Entering a
level offset
If you have connected external attenuators or amplifiers between the DUT
and the RF input, you can enter a level offset to adjust the level display. The
level offset is added to the reference level.
REF LEVEL + REF LEVEL OFFSET
1. Press the
function key in the menu.
An entry field containing the current setting is displayed in the top lefthand corner. The default setting is 0 dB.
2. Enter a new value in 0.1 dB steps ( 5-57).
The entry range for the level offset is:
-100 [dB] ≤ Ref Lvl Offset ≤ +50 [dB]
The new setting is saved and displayed in the diagram area. The text
“Offs” is also displayed in the top left of the diagram.
The level offset is always entered in dB no matter what unit is used for
the reference level. If the units mV or mW are used, the offset is automatically converted so that the reference level is displayed correctly:
REF LEVEL
e.
g. = -20 dBm + 40 dB = 20 dBm
= REF LEVEL + REF LEVEL OFFSET
Display
Operating manual, 08/2005 6-85 E-1147.1000.04
Page 86
Spectrum Analyzer R&S FS315
6.2.2.3 Selecting the Level Display Range
Use
Entering the level
display range
Use the level display range to define the “visible” level range and set the
measurement diagram scale.
Levels can be displayed using either logarithmic or linear scales. The lettering
of the grid lines is automatically adapted.
8 dB
1 dB/div.
16 dB
2 dB/div.
40 dB
5 dB/div.
80 dB
10 dB/div.
linear 0..100%
12.5 %/div., linear scale, 100 % corresponds to the reference level, 0 %
to 0 V or 0 W
1. Press the
A selection field containing the available settings is displayed. The default setting is 80 dB.
function key in the menu.
Note
2. Select a level display range with rotary knob
3. Press the ENTER key [5] to close the selection field.
The new setting is saved and displayed in inverse video in the diagram
area. The level display range is referred to the reference level. This determines how the grid lines are lettered, for example, for REF LEVEL =
-20 dBm and RANGE = 80 dB, the first grid line is at -20 dBm and the
second at -30 dBm, and so on.
RANGE
The level display range uses the currently activated unit ( 6-87).
[11].
E-1147.1000.04 6-86 Operating manual, 08/2005
Page 87
R&S FS315 Spectrum Analyzer
6.2.2.4 Selecting the Level Display Unit
Note
Use To make it easier to analyze results, you can display levels in dBm, dBmV,
Selecting the
unit for the
level display
How RANGE and
UNIT are related
The function key is only available for activated transducer factors if
the unit of the antenna factors is dB.
dBµV, dBµA, mV and mW.
1. Press the
A selection field containing the available settings is displayed. The default setting is dBm.
2. Select a unit for the level display with rotary knob
3. Press the ENTER key [5] to close the selection field.
The new setting is saved and displayed in the diagram area.
Log display/RANGE in dBx (The letter x is a wildcard character for the letters m, mV or µV.)
If a dB display has been selected for RANGE ( 6-86) and a dBx display for
UNIT, the vertical-scale unit is dB. The values indicated by the level markers
and the relevant level entries are in dBx.
function key in the menu.
[11].
Operating manual, 08/2005 6-87 E-1147.1000.04
Page 88
Spectrum Analyzer R&S FS315
Log display/RANGE in xV, xA or xW (The letter x is a wildcard character for
the letters m or µ.)
If a mV or mW has been selected for UNIT, only the reference level is displayed in the selected unit. There is no grid lettering. The values shown by
the level markers and the relevant level entries are in the selected unit.
Linear display/RANGE in xV, xA or xW (The letter x is a wildcard character
for the letters m or µ.)
If the linear scale is selected for RANGE, the scale lettering indicates percent,
e.
g. 0 %, 25 %, 50 %, 75 % and 100 %. The values indicated by the level
markers and the relevant level entries are in the selected unit.
E-1147.1000.04 6-88 Operating manual, 08/2005
Page 89
R&S FS315 Spectrum Analyzer
6.2.2.5 Setting the RF Input Attenuation Manually
Use
Setting the input
attenuation
manually
Note
In the default setting, the input attenuation is coupled to the reference level. In
most cases, this is the setting that will be used because it prevents the input
stages from being overdriven and so, consequently, any incorrect measurements.
In exceptional cases, it may be necessary to adjust the input attenuation
manually with input attenuations from 0 to 70 dB.
1. Press the
An entry field containing the current setting is displayed in the top lefthand corner. The default setting is 16 dB.
2. Enter a new value in 2 dB steps ( 5-57).
The entry range for the input attenuation is:
0 [dB] ≤ RF Atten Manual ≤ 70 [dB]
The new setting is saved and the RF input is re-configured.
If you enter an intermediate value (e.g. 3 dB) which the R&S FS315 cannot
set, the next largest value will be set instead (4 dB).
function key in the menu.
Operating manual, 08/2005 6-89 E-1147.1000.04
Page 90
Spectrum Analyzer R&S FS315
6.2.2.6 Setting the RF Input Attenuation Automatically
Use
The input attenuation should be set automatically to prevent the R&S FS315’s
input mixer from being overloaded. There are three RF input modes you can
choose from (coupling between reference level and input attenuation) to optimize measurements:
NORMAL
Normal setting for measurements
LOW NOISE
Settings for measurements with low displayed average noise level of the
analyzer
LOW DISTORTION
Setting for measurements with low inherent distortion of the analyzer
E-1147.1000.04 6-90 Operating manual, 08/2005
Page 91
R&S FS315 Spectrum Analyzer
Activating the automatic input
attenuation
setting mode
1. Press the
function key in the menu.
A selection field containing the available settings is displayed. The default setting is NORMAL.
2. Using the rotary knob
[11] select a criterion for autosetting the input
attenuation.
3. Press the ENTER key [5] to close the selection field.
The new setting is saved and the RF input is configured automatically
according to the setting.
Operating manual, 08/2005 6-91 E-1147.1000.04
Page 92
Spectrum Analyzer R&S FS315
6.2.2.7 Selecting the Setting High Sensitivity
Use
Selecting the Setting
High Sensitivity
By setting SET HIGH SENS, you can change several settings in the device in
order to increase the input and measurement sensitivity of the analyzer in
comparison to the normal setting. This function is helpful when you want to
measure weak antenna signals with the analyzer.
SET HIGH SENS influences the following settings in the R&S FS315:
Sets the reference level (REF LEVEL, 6-84) to -10 dBm.
Sets the input attenuation (RF ATTEN, 6-89) to 0 dB.
Activates the coupling (RBW/SPAN LOW NOISE, 6-123).
If RES BW was set to MANUAL, the analyzer switches to AUTO.
• Press the
The setting RBW/SPAN LOW NOISE is indicated by the letters LN next
to the RBW display.
function key in the menu.
Note
You can switch the input attenuation of the analyzer to 0 dB by means of SET
HIGH SENS. To ensure that the R&S FS315 is not overdriven, the level at the
analyzer input must not exceed the reference level that has been set minus
15 dB. ( 6-84).
E-1147.1000.04 6-92 Operating manual, 08/2005
Page 93
R&S FS315 Spectrum Analyzer
6.2.3 Signal Analysis using Marker Functions (MKR Menu)
What the settings
are for
Selecting the
MKR menu
Function key
assignment
Using the MKR menu, you can position markers on any point on the trace to
read off any levels or level differences of interest. The measured values indicated by the markers (marker values) are displayed in the parameter field (
5-51).
A scrollable display line can also be shown on-screen.
• Select the
The menu name is highlighted and the appropriate functions are assigned to the function keys [13].
Open submenu:
Activate marker 1 ( 6-94)
menu with the
3
or
cursor key [6].
4
Note
Open submenu:
Activate marker 2 ( 6-98)
Open submenu:
Accepting marker values as settings ( 6-101)
Open submenu:
Marker measurement functions ( 6-105)
Show display line
( 6-110)
Open submenu:
Search criterion of functions NEXT PEAK LEFT/RIGHT ( 6-111)
Open submenu:
Reference points for level-difference measurements ( 6-114)
The
The
The
marker 1 is turned on.
and function keys are selection keys.
function key is available only when marker 1 is turned on.
, and function keys are available only when
Operating manual, 08/2005 6-93 E-1147.1000.04
Page 94
Spectrum Analyzer R&S FS315
6.2.3.1 Activating Marker 1
What the settings
are for
Selecting the
MARKER 1
submenu
From the MARKER 1 submenu, you can activate a marker and position it on
the trace. All the current level and frequency values (marker values) are displayed.
The frequency indicated by the marker can be measured more accurately
with the FREQ COUNT function.
• Press the
If marker 1 is not yet activated, the marker is set to the strongest signal
(peak). An entry field containing the current marker frequency appears in
the top left-hand corner. The submenu is opened and the respective
functions are assigned to the function keys [13].
If the marker is already activated, the entry field opens without the position of the marker being changed.
function key in the menu.
Function key
assignment
Note
Quitting the submenu
Activate marker 1 and place it on the trace maximum
( 6-95)
Place marker 1 on the next peak to the left ( 6-95)
Place marker 1 on the next peak to the right ( 6-95)
Measure the signal frequency: Start the measurement ( 6-97)
Measure the signal frequency: Set the resolution ( 6-97)
De-activate marker 1 ( 6-95)
The
been activated.
and function keys are available only when marker 1 has
E-1147.1000.04 6-94 Operating manual, 08/2005
Page 95
R&S FS315 Spectrum Analyzer
6.2.3.1.1 Reading off Measured Values with Marker 1
Use
Activating marker 1
and placing it on the
trace maximum
Placing marker 1
on the next trace
maximum
You can position marker 1 on the trace to read off specific levels and frequencies (marker values).
1. Press the
function key in the submenu.
Marker 1 is activated and placed on the current trace maximum. M1
marker values are displayed in the parameter field.
2. Press the
or function key in the submenu.
Marker 1 is positioned on the next trace maximum to the right (or left).
The new M1 marker values are displayed in the parameter field.
Note
The search criterion of functions NEXT PEAK LEFT/RIGHT can be changed
by using the function key
( 6-112).
Operating manual, 08/2005 6-95 E-1147.1000.04
Page 96
Spectrum Analyzer R&S FS315
Moving marker 1
to any position
De-activating
marker 1
3. Enter a new value for the marker frequency using the numerical keys
[12], rotary knob [11] or cursor keys [7]. The permissible entry range is:
– Span/2 ≤ f
f
Center
where f
f
- Marker frequency
Marker
- current center frequency ( 6-73)
Center
Marker
≤ f
Center
+ Span/2
Span - current Span ( 6-75)
The current M1 marker values are displayed in the parameter field.
• Press the
function key in the submenu.
Marker 1 (and marker 2, if activated) disappears from the diagram area
and the M1 (M2) marker values are blanked out.
Note
The marker can only be moved if the entry window is open. If you closed the
entry window with ESC/CANCEL key [4], you must briefly leave the Marker 1
menu before you can open the entry window again. Then press the
function key in the
menu to open the entry window again.
If the center frequency setting is such that the set marker is outside the frequency span, the last value is frozen as the reference.
E-1147.1000.04 6-96 Operating manual, 08/2005
Page 97
R&S FS315 Spectrum Analyzer
6.2.3.1.2 Frequency Measurements using the Frequency Counter
Use
Setting the
resolution
The normal marker indicates the frequency represented by the pixel on which
it is placed. The measurement uncertainty is relatively large because the
number of pixels is limited to 250.
When a measurement is made with the frequency counter, the sweep is
stopped at the frequency indicated by marker 1, the frequency is counted
using the required resolution (10 Hz, 1 Hz) and then the sweep is restarted.
There must be a periodic signal at the marker position if the measurement is
to be successful.
1. Position marker 1 on the trace ( 6-94).
The current M1 marker values are displayed in the parameter field.
2. Press the
function key in the submenu.
A selection field containing the available settings is displayed. The default setting is 1 Hz.
3. Select a resolution for the internal frequency counter with rotary
knob
[11], e.
g. 10 Hz.
4. Press the ENTER key [5] to close the selection field.
The new setting is saved.
Measuring the
signal frequency
De-activating the
measurement
Note
5. Press the
function key in the submenu.
The frequency is counted with the internal counter. The result (C1) is
displayed in the parameter field instead of the marker frequency (M1).
• Press the
function key in the submenu.
The internal counter is turned off. The current M1 marker values are
again displayed in the parameter field.
The absolute accuracy of the frequency measurement depends on the
frequency stability of the reference-oscillator. When the internal reference is
used, the accuracy is given according to the values specified in the data
sheet. ( 0-12).
Operating manual, 08/2005 6-97 E-1147.1000.04
Page 98
Spectrum Analyzer R&S FS315
6.2.3.2 Activating Marker 2
What the settings
are for
Selecting the
MARKER 2
submenu
From the MARKER 2 submenu, you can activate a second marker and position it on the trace. Marker 2 can be used either as a normal marker (NORM)
or as a delta marker (DELTA). The delta marker function expands the R&S
FS315’s analysis capability and can be used as a basis for a variety of measurement techniques ( 6-143).
Marker 2 can only be activated if marker 1 has been activated.
• Press the
If marker 2 is not yet activated, the marker is set to the strongest signal
(peak). An entry field containing the current marker frequency appears in
the top left-hand corner. The submenu is opened and the respective
functions are assigned to the function keys [13].
If the marker is already activated, the entry field opens without the position of the marker being changed.
function key in the menu.
Function key
assignment
Note
Quitting the submenu
Activate marker 2 and position it on the trace maximum
( 6-99)
Place marker 2 on the next peak to the left ( 6-99)
Place marker 2 on the next peak to the right ( 6-99)
Set marker 2 as a normal marker (NORM) ( 6-100)
Set marker 2 as a delta marker (DELTA) ( 6-100)
De-activate marker 2 ( 6-99)
The
only one setting can be activated at any one time. The function key in ques-
tion is highlighted. The MARKER NORM function is the default setting.
and function keys are selection keys. In other words,
E-1147.1000.04 6-98 Operating manual, 08/2005
Page 99
R&S FS315 Spectrum Analyzer
6.2.3.2.1 Reading off Measured Values with Marker 2
Use
Activating marker 2
as a normal marker
(NORM)
De-activating
marker 2
To read off specific levels, you can activate marker 2 as a normal marker (in
addition to marker 1) and position it on the trace. The current level and frequency can be displayed in this way (M2 marker values).
1. Press the
function key in the submenu.
Marker 2 is defined as a normal marker. The M2 absolute marker values
are displayed in the parameter field.
2. Position marker 2 on the trace ( 6-95).
The current M2 marker values are displayed in the parameter field.
• Press the
function key in the submenu.
Marker 2 disappears from the diagram area and the M2 marker values
are blanked out.
Operating manual, 08/2005 6-99 E-1147.1000.04
Page 100
Spectrum Analyzer R&S FS315
6.2.3.2.2 Reading off Level Differences
Use
Defining marker 2
as a delta marker
(DELTA)
De-activating
marker 2
You can define marker 2 as a delta marker to read off level differences from
the trace. The level difference between marker 2 and a reference point is
displayed. Marker 1 ( 6-94) or reference values that have been set manually
( 6-114) can be used as reference points.
1. Press the
function key in the submenu.
Marker 2 is defined as a delta marker. The relative delta marker values
(M), which represent the level difference between marker 2 and a reference point, e.
g. marker 1, are displayed in the parameter field.
2. Press the
function key in the submenu.
Marker 2 disappears from the diagram area and the M2 marker values
are blanked out.
E-1147.1000.04 6-100 Operating manual, 08/2005
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.