This chapter provides safety-related information, an overview of the user documentation and the conventions used in the documentation.
This section provides an overview of the R&S FPL1000 user documentation. Unless
specified otherwise, you find the documents at:
www.rohde-schwarz.com/manual/FPL1000
Introduces the R&S FPL1000 and describes how to set up and start working with the
product. Includes basic operations, typical measurement examples, and general information, e.g. safety instructions, etc.
A printed version is delivered with the instrument. A PDF version is available for download on the Internet.
1.1.2User manuals and help
Separate user manuals are provided for the base unit and the firmware applications:
●
Base unit manual
Contains the description of all instrument modes and functions. It also provides an
introduction to remote control, a complete description of the remote control commands with programming examples, and information on maintenance, instrument
interfaces and error messages. Includes the contents of the getting started manual.
●
Firmware application manual
Contains the description of the specific functions of a firmware application, including remote control commands. Basic information on operating the R&S FPL1000 is
not included.
The contents of the user manuals are available as help in the R&S FPL1000. The help
offers quick, context-sensitive access to the complete information for the base unit and
the firmware applications.
All user manuals are also available for download or for immediate display on the Internet.
1.1.3Service manual
Describes the performance test for checking compliance with rated specifications, firmware update, troubleshooting, adjustments, installing options and maintenance.
7User Manual 1178.3392.02 ─ 10
Page 8
R&S®FPL1-K7
1.1.4Instrument security procedures
1.1.5Printed safety instructions
1.1.6Data sheets and brochures
Preface
Documentation overview
The service manual is available for registered users on the global Rohde & Schwarz
information system (GLORIS):
https://gloris.rohde-schwarz.com
Deals with security issues when working with the R&S FPL1000 in secure areas. It is
available for download on the Internet.
Provides safety information in many languages. The printed document is delivered with
the product.
The data sheet contains the technical specifications of the R&S FPL1000. It also lists
the firmware applications and their order numbers, and optional accessories.
The brochure provides an overview of the instrument and deals with the specific characteristics.
See www.rohde-schwarz.com/brochure-datasheet/FPL1000
1.1.7Release notes and open source acknowledgment (OSA)
The release notes list new features, improvements and known issues of the current
firmware version, and describe the firmware installation.
The software makes use of several valuable open source software packages. An opensource acknowledgment document provides verbatim license texts of the used open
source software.
See www.rohde-schwarz.com/firmware/FPL1000
1.1.8Application notes, application cards, white papers, etc.
These documents deal with special applications or background information on particular topics.
See www.rohde-schwarz.com/application/FPL1000
8User Manual 1178.3392.02 ─ 10
Page 9
R&S®FPL1-K7
1.1.9Calibration certificate
1.2Conventions used in the documentation
1.2.1Typographical conventions
Preface
Conventions used in the documentation
The document is available on https://gloris.rohde-schwarz.com/calcert. You need the
device ID of your instrument, which you can find on a label on the rear panel.
The following text markers are used throughout this documentation:
ConventionDescription
"Graphical user interface elements"
[Keys]Key and knob names are enclosed by square brackets.
Filenames, commands,
program code
InputInput to be entered by the user is displayed in italics.
LinksLinks that you can click are displayed in blue font.
"References"References to other parts of the documentation are enclosed by quota-
All names of graphical user interface elements on the screen, such as
dialog boxes, menus, options, buttons, and softkeys are enclosed by
quotation marks.
Filenames, commands, coding samples and screen output are distinguished by their font.
tion marks.
1.2.2Conventions for procedure descriptions
When operating the instrument, several alternative methods may be available to perform the same task. In this case, the procedure using the touchscreen is described.
Any elements that can be activated by touching can also be clicked using an additionally connected mouse. The alternative procedure using the keys on the instrument or
the on-screen keyboard is only described if it deviates from the standard operating procedures.
The term "select" may refer to any of the described methods, i.e. using a finger on the
touchscreen, a mouse pointer in the display, or a key on the instrument or on a keyboard.
1.2.3Notes on screenshots
When describing the functions of the product, we use sample screenshots. These
screenshots are meant to illustrate as many as possible of the provided functions and
possible interdependencies between parameters. The shown values may not represent
realistic usage scenarios.
9User Manual 1178.3392.02 ─ 10
Page 10
R&S®FPL1-K7
Preface
Conventions used in the documentation
The screenshots usually show a fully equipped product, that is: with all options installed. Thus, some functions shown in the screenshots may not be available in your particular product configuration.
10User Manual 1178.3392.02 ─ 10
Page 11
R&S®FPL1-K7
2Welcome to the R&S FPL1-K7 AM/FM/PM
Welcome to the R&S FPL1-K7 AM/FM/PM Modulation Analysis application
Starting AM/FM/PM Modulation Analysis
Modulation Analysis application
The R&S FPL1-K7 AM/FM/PM Modulation Analysis application option converts the
R&S FPL1000 into an analog modulation analyzer for amplitude-, frequency- or phasemodulated signals. It measures not only characteristics of the useful modulation, but
also factors such as residual FM or synchronous modulation.
The R&S FPL1-K7 AM/FM/PM Modulation Analysis application features:
●
AM, FM, and PM demodulation, with various result displays:
–Modulation signal versus time
–Spectrum of the modulation signal (FFT)
–RF signal power versus time
–Spectrum of the RF signal
●
Determining maximum, minimum and average or current values in parallel over a
selected number of measurements
●
Maximum accuracy and temperature stability due to sampling (digitization) already
at the IF and digital down-conversion to the baseband (I/Q)
●
Error-free AM to FM conversion and vice versa, without deviation errors, frequency
response or frequency drift at DC coupling
●
Relative demodulation, in relation to a user-defined or measured reference value
Availability of the R&S FPL1-K7 AM/FM/PM Modulation Analysis application
Using the R&S FPL1-K7 AM/FM/PM Modulation Analysis application requires the
optional AM/FM/PM Modulation Analysis firmware (R&S FPL1-K7).
This user manual contains a description of the functionality that the application provides, including remote control operation.
All functions not discussed in this manual are the same as in the base unit and are
described in the R&S FPL1000 User Manual. The latest version is available for download at the product homepage http://www.rohde-schwarz.com/product/FPL1000.
2.1Starting AM/FM/PM Modulation Analysis
AM/FM/PM Modulation Analysis is a separate application on the R&S FPL1000.
To activate AM/FM/PM Modulation Analysis
1. Select the [MODE] key.
A dialog box opens that contains all operating modes and applications currently
available on your R&S FPL1000.
2. Select the "AM FM PM Analog Demod" item.
11User Manual 1178.3392.02 ─ 10
Page 12
R&S®FPL1-K7
Welcome to the R&S FPL1-K7 AM/FM/PM Modulation Analysis application
Understanding the display information
The R&S FPL1000 opens a new channel setup for the application for analog modulation analysis.
The measurement is started immediately with the default settings. It can be configured
in the analog modulation analysis "Overview" dialog box, which is displayed when you
select the "Overview" softkey from any menu (see Chapter 5.1, "Configuration over-
view", on page 33).
Multiple Channel setups and Sequencer Function
When you activate an application, a new channel setup is created which determines
the measurement settings for that application ("Channel"). The same application can
be activated with different measurement settings by creating several "Channel"s for the
same application.
The number of channel setups that can be configured at the same time depends on the
available memory on the instrument.
Only one measurement can be performed at any time, namely the one in the currently
active channel setup. However, to perform the configured measurements consecutively, a Sequencer function is provided.
If activated, the measurements configured in the currently defined "Channel"s are performed one after the other in the order of the tabs. The currently active measurement is
indicated by a
The result displays of the individual channel setups are updated in the tabs (as well as
the "MultiView") as the measurements are performed. Sequential operation itself is
independent of the currently displayed tab.
For details on the Sequencer function, see the R&S FPL1000 User Manual.
symbol in the tab label.
2.2Understanding the display information
The following figure shows a measurement diagram during analog modulation analysis. All different information areas are labeled. They are explained in more detail in the
following sections.
12User Manual 1178.3392.02 ─ 10
Page 13
R&S®FPL1-K7
Welcome to the R&S FPL1-K7 AM/FM/PM Modulation Analysis application
Understanding the display information
1
2
3
4
5
1 = Channel setup bar for firmware and measurement settings
2 = Diagram area
3 = Window title bar with diagram-specific (trace) information
4 = Instrument status bar with error messages and date/time display
5 = Diagram footer with diagram-specific information, depending on result display
Channel setup bar information
In the Analog Modulation Analysis application, the R&S FPL1000 shows the following
settings:
Table 2-1: Information displayed in the channel setup bar in the application for analog modulation
Ref LevelReference level
AttRF attenuation applied to input
OffsetReference level offset
AQTMeasurement time for data acquisition.
RBWResolution bandwidth
DBWDemodulation bandwidth
FreqCenter frequency for the RF signal
analysis
Window title bar information
For each diagram, the header provides the following information:
13User Manual 1178.3392.02 ─ 10
Page 14
R&S®FPL1-K7
Welcome to the R&S FPL1-K7 AM/FM/PM Modulation Analysis application
Understanding the display information
123456789
Figure 2-1: Window title bar information in the application for analog modulation analysis
1 = Window number
2 = Window type
3 = Trace color
4 = Trace number
5 = Detector
6 = Trace mode
7 = Reference value (at the defined reference position)
8 = AF coupling (AC/DC), only in AF time domains, if applicable
9 = Results are selected for demodulation output
Diagram footer information
The diagram footer (beneath the diagram) contains the following information, depending on the evaluation:
"RF Spectrum"
CF: Center frequency
of input signal
RF Time domain
CF: Center frequency
of input signal
AF Spectrum
AF CF: center frequency of demodulated signal
AF Time domain
CF: Center frequency
of input signal
Sweep pointsSpan: measured span
Sweep pointsTime per division
Sweep pointsAF Span: evaluated span
Sweep pointsTime per division
For most modes, the number of sweep points shown in the display are indicated in the
diagram footer. In zoom mode, the (rounded) number of currently displayed points are
indicated.
Status bar information
Global instrument settings, the instrument status and any irregularities are indicated in
the status bar beneath the diagram.
14User Manual 1178.3392.02 ─ 10
Page 15
R&S®FPL1-K7
3Measurements and result displays
Measurements and result displays
Access: "Overview" > "Display Config"
Or: [MEAS] > "Display Config"
The data that was measured by the R&S FPL1000 can be evaluated using various different methods. In the Analog Modulation Analysis application, up to six evaluation
methods can be displayed simultaneously in separate windows. The results can be displayed as absolute deviations or relative to a reference value or level.
The abbreviation "AF" (for Audio Frequency) refers to the demodulated AM, FM or PM
signal.
Basis for evaluation
All evaluations are based on the I/Q data set acquired during the measurement. The
spectrum of the modulated signal to be evaluated is determined by the demodulation
bandwidth. However, it can be restricted to a limited span ("AF Span") if only part of the
signal is of interest. Furthermore, the time base for evaluations in the time domain can
be restricted to analyze a smaller extract in more detail, see Chapter 4.6, "Time
domain zoom", on page 31.
AM Time Domain...........................................................................................................15
FM Time Domain...........................................................................................................16
PM Time Domain...........................................................................................................17
AM Spectrum................................................................................................................ 18
FM Spectrum.................................................................................................................19
Displays the modulation depth of the demodulated AM signal (in %) versus time.
15User Manual 1178.3392.02 ─ 10
Page 16
R&S®FPL1-K7
Measurements and result displays
Optionally, the settling time can be evaluated and displayed, see Chapter 5.7.7, "Set-
tling time", on page 70.
Remote command:
LAY:ADD? '1',RIGH,'XTIM:AM:REL'
(See LAYout:ADD[:WINDow]?on page 180)
FM Time Domain
Displays the frequency spectrum of the demodulated FM signal versus time.
16User Manual 1178.3392.02 ─ 10
Page 17
R&S®FPL1-K7
Measurements and result displays
Optionally, the settling time can be evaluated and displayed, see Chapter 5.7.7, "Set-
tling time", on page 70.
Remote command:
LAY:ADD? '1',RIGH,'XTIM:FM'
(See LAYout:ADD[:WINDow]?on page 180)
PM Time Domain
Displays the phase deviations of the demodulated PM signal (in rad or °) versus time.
17User Manual 1178.3392.02 ─ 10
Page 18
R&S®FPL1-K7
Measurements and result displays
Optionally, the settling time can be evaluated and displayed, see Chapter 5.7.7, "Set-
tling time", on page 70.
Remote command:
LAY:ADD? '1',RIGH,'XTIM:PM'
(See LAYout:ADD[:WINDow]?on page 180)
AM Spectrum
Displays the modulation depth of the demodulated AM signal (in % or dB) versus AF
span. The spectrum is calculated from the demodulated AM signal in the time domain
via FFT.
18User Manual 1178.3392.02 ─ 10
Page 19
R&S®FPL1-K7
Measurements and result displays
Remote command:
LAY:ADD? '1',RIGH,'XTIMe:AM:REL:AFSPectrum1'
(see LAYout:ADD[:WINDow]?on page 180)
FM Spectrum
Displays the frequency deviations of the demodulated FM signal (in Hz or dB) versus
AF span. The spectrum is calculated from the demodulated AM signal in the time
domain via FFT.
19User Manual 1178.3392.02 ─ 10
Page 20
R&S®FPL1-K7
Measurements and result displays
Remote command:
LAY:ADD? '1',RIGH,'XTIMe:FM:AFSPectrum1'
(see LAYout:ADD[:WINDow]?on page 180)
PM Spectrum
Displays the phase deviations of the demodulated PM signal (in rad, ° or dB) versus
AF span. The spectrum is calculated from the demodulated AM signal in the time
domain via FFT.
20User Manual 1178.3392.02 ─ 10
Page 21
R&S®FPL1-K7
Measurements and result displays
Remote command:
LAY:ADD? '1',RIGH,'XTIMe:PM:AFSPectrum1'
(see LAYout:ADD[:WINDow]?on page 180)
RF Time Domain
Displays the RF power of the input signal versus time. The level values represent the
magnitude of the I/Q data set.
21User Manual 1178.3392.02 ─ 10
Page 22
R&S®FPL1-K7
Measurements and result displays
Optionally, the settling time can be evaluated and displayed, see Chapter 5.7.7, "Set-
tling time", on page 70.
Remote command:
LAY:ADD? '1',RIGH,'XTIM:AM'
(see LAYout:ADD[:WINDow]?on page 180)
RF Spectrum
Displays the spectrum of the input signal. In contrast to the Spectrum application, the
frequency values are determined using FFT from the recorded I/Q data set.
22User Manual 1178.3392.02 ─ 10
Page 23
R&S®FPL1-K7
Measurements and result displays
Remote command:
LAY:ADD? '1',RIGH,'XTIM:SPECTRUM'
(see LAYout:ADD[:WINDow]?on page 180)
Result Summary
The "result summary" displays the results of the demodulation functions for all windows
in a table.
The following general results are provided:
For each demodulation, the following results are provided:
LabelDescription
"Carr Power"Measured carrier power
"Carr Offset"Carrier offset to nominal center frequency
"Mod. Depth"Modulation depth
23User Manual 1178.3392.02 ─ 10
Page 24
R&S®FPL1-K7
Measurements and result displays
Table 3-1: Result summary description
LabelDescription
"Settling Time"Time after which signal remains in a specified value range.
Only evaluated and displayed if enabled, see Chapter 5.7.7, "Settling time",
on page 70.
"+Peak"Positive peak (maximum)
"-Peak"Negative peak (minimum)
"+/-Peak/2"Average of positive and negative peaks
"RMS"Root Mean Square value
"Mod Freq"Modulation frequency
"SINAD"Signal-to-noise-and-distortion
(Calculated only if AF Spectrum is displayed)
Measures the ratio of the total power to the power of noise and harmonic distortions.
The noise and harmonic power is calculated inside the AF spectrum span. The DC
offset is removed before the calculation.
"DISTORT"Modulation distortion in %
(Calculated only if "SINAD" is also calculated)
Measures the distortion of the modulation in relation to the total power of the signal
inside the AF spectrum span. Indicates the quality of the modulation.
"THD"Total harmonic distortion
The ratio of the harmonics to the fundamental and harmonics. All harmonics inside
the AF spectrum span are considered up to the tenth harmonic.
(Calculated only if AF Spectrum is displayed)
Note: Relative demodulation results. Optionally, the demodulation results in relation to
user-defined or measured reference values are determined. See Chapter 5.7.6, "Result
table settings", on page 67.
In addition, the following general information for the input signal is provided:
●
"Carrier Power": the power of the carrier without modulation
●
"Carrier Offset": the deviation of the calculated carrier frequency to the ideal carrier
frequency
●
"Modulation Depth" (AM or "RF Time Domain" only): the difference in amplitude the
carrier signal is modulated with
24User Manual 1178.3392.02 ─ 10
Page 25
R&S®FPL1-K7
Measurements and result displays
Remote command:
LAY:ADD? '1',RIGH,RSUM, see LAYout:ADD[:WINDow]? on page 180
Results:
Chapter 11.7.3, "Retrieving result summary values", on page 192
Marker Table
Displays a table with the current marker values for the active markers.
This table is displayed automatically if configured accordingly.
(See "Marker Table Display"on page 92).
Tip: To navigate within long marker tables, simply scroll through the entries with your
finger on the touchscreen.
Remote command:
LAY:ADD? '1',RIGH, MTAB, see LAYout:ADD[:WINDow]? on page 180
Results:
CALCulate<n>:MARKer<m>:X on page 216
CALCulate<n>:MARKer<m>:Y? on page 216
Marker Peak List
The marker peak list determines the frequencies and levels of peaks in the spectrum or
time domain. How many peaks are displayed can be defined, as well as the sort order.
In addition, the detected peaks can be indicated in the diagram. The peak list can also
be exported to a file for analysis in an external application.
Remote command:
LAY:ADD? '1',RIGH, PEAK, see LAYout:ADD[:WINDow]? on page 180
Results:
CALCulate<n>:MARKer<m>:X on page 216
CALCulate<n>:MARKer<m>:Y? on page 216
25User Manual 1178.3392.02 ─ 10
Page 26
R&S®FPL1-K7
4Measurement basics
4.1Demodulation process
Measurement basics
Demodulation process
Some background knowledge on basic terms and principles used in Analog Modulation
Analysis measurements is provided here for a better understanding of the required
configuration settings.
The demodulation process is shown in Figure 4-1. All calculations are performed simultaneously with the same I/Q data set. Magnitude (= amplitude) and phase of the complex I/Q pairs are determined. The frequency result is obtained from the differential
phase.
For details on general I/Q data processing in the R&S FPL1000, refer to the reference
part of the I/Q Analysis remote control description in the R&S FPL1000 User Manual.
26User Manual 1178.3392.02 ─ 10
Page 27
R&S®FPL1-K7
Measurement basics
Demodulation process
Figure 4-1: Block diagram of software demodulator
27User Manual 1178.3392.02 ─ 10
Page 28
R&S®FPL1-K7
Measurement basics
Demodulation bandwidth
The AM DC, FM DC and PM DC raw data of the demodulators is fed into the "Trace
Arithmetic" block that combines consecutive data sets. Possible trace modes are:
Clear Write, Max Hold, Min Hold and Average. The output data of the "Trace Arithmetic" block can be read via remote control ([SENS:]ADEM:<evaluation>:RES?,
see [SENSe:]ADEMod:AM[:ABSolute][:TDOMain]:RESult?on page 187.
The collected measured values are evaluated by the selected detector. The result is
displayed on the screen and can be read out via remote control.
In addition, important parameters are calculated:
●
A counter determines the modulation frequency for AM, FM, and PM.
●
average power = carrier power (RF power)
●
average frequency = carrier frequency offset (FM)
●
The modulation depth or the frequency or phase deviation; the deviations are
determined from the trace data
AC coupling is possible with FM and PM display.
4.2Demodulation bandwidth
The demodulation bandwidth determines the span of the signal that is demodulated. It
is not the 3-dB bandwidth of the filter, but the useful bandwidth which is distortion-free
regarding phase and amplitude.
Therefore the following formulas apply:
●
AM: demodulation bandwidth ≥ 2 x modulation frequency
PM: demodulation bandwidth ≥ 2 x modulation frequency x (1 + phase deviation)
If the center frequency of the analyzer is not set exactly to the signal frequency, the
demodulation bandwidth must be increased by the carrier offset, in addition to the
requirement described above. The bandwidth must also be increased if FM or PM AC
coupling is selected.
In general, select the demodulation bandwidth as narrow as possible to improve the
S/N ratio. The residual FM caused by noise floor and phase noise increases dramatically with the bandwidth, especially with FM.
For help on determining the adequate demodulation bandwidth, see "Determining the
demodulation bandwidth"on page 116.
A practical example is described in Chapter 9, "Measurement example: demodulating
an FM signal", on page 111.
28User Manual 1178.3392.02 ─ 10
Page 29
R&S®FPL1-K7
4.3Sample rate and demodulation bandwidth
Measurement basics
Sample rate and demodulation bandwidth
The maximum demodulation bandwidths that can be obtained during the measurement, depending on the sample rate, are listed in the tables below for different demodulation filter types. The allowed value range of the measurement time and trigger offset
depends on the selected demodulation bandwidth and demodulation filter. If the AF filter or the AF trigger are not active, the measurement time increases by 20 %.
A maximum of 24 million samples can be captured, assuming sufficient memory is
available; thus the maximum measurement time can be determined according to the
following formula:
Meas.time
The minimum trigger offset is (-Meas.time
Table 4-1: Available demodulation bandwidths and corresponding sample rates
* Gaussian filter curve is limited by I/Q bandwidth
Flat top filters require an I/Q bandwidth at least the size of the demodulation bandwidth
on the R&S FPL1000. Gauss filters require at least twice the size of the demodulation
bandwidth on the R&S FPL1000. If necessary, install optional bandwidth extensions on
the R&S FPL1000 to use the required demodulation filter.
Large numbers of samples
Principally, the R&S FPL1000 can handle up to 24 million samples. However, when
480001 samples are exceeded, all traces that are not currently being displayed in a
window are deactivated to improve performance. The traces can only be activated
again when the samples are reduced.
Effects of measurement time on the stability of measurement results
Despite amplitude and frequency modulation, the display of carrier power and carrier
frequency offset is stable.
Stability is achieved by a digital filter which sufficiently suppresses the modulation. As
a prerequisite, the measurement time must be ≥ 3 x 1 / modulation frequency, i.e. at
least three periods of the AF signal are recorded.
The mean carrier power for calculating the AM is also calculated with a digital filter.
The filter returns stable results after a measurement time of ≥ 3 x 1 / modulation frequency, i.e. at least three cycles of the AF signal must be recorded before a stable AM
can be shown.
4.4AF triggers
The Analog Modulation Analysis application allows triggering to the demodulated signal. The display is stable if a minimum of five modulation periods are within the recording time.
Triggering is always DC-coupled. Therefore triggering is possible directly to the point
where a specific carrier level, phase or frequency is exceeded or not attained.
4.5AF filters
Additional filters applied after demodulation help filter out unwanted signals, or correct
pre-emphasized input signals. A CCITT filter allows you to evaluate the signal by simulating the characteristics of human hearing.
30User Manual 1178.3392.02 ─ 10
Page 31
R&S®FPL1-K7
4.6Time domain zoom
Measurement basics
Time domain zoom
For evaluations in the time domain, the demodulated data for a particular time span
can be extracted and displayed in more detail using the "Time Domain Zoom" function.
Zooming is useful if the measurement time is very large and thus each sweep point
represents a large time span. The time domain zoom function distributes the available
sweep points only among the time span defined by the zoom area length. The time
span displayed per division of the diagram is decreased. Thus, the display of the
extracted time span becomes more precise.
Figure 4-2: FM time domain measurement with a very long measurement time (200 ms)
Figure 4-3: FM time domain measurement with time domain zoom (2.0 ms per division)
The time domain zoom area affects not only the diagram display, but the entire evaluation for the current window.
In contrast to the time domain zoom, the graphical zoom is available for all diagram
evaluations. However, the graphical zoom is useful only if more measured values than
trace points are available. The (time) span represented by each measurement point
remains the same.
31User Manual 1178.3392.02 ─ 10
Page 32
R&S®FPL1-K7
Measurement basics
Time domain zoom
Time domain zoomGraphical zoom
32User Manual 1178.3392.02 ─ 10
Page 33
R&S®FPL1-K7
5Configuration
Configuration
Configuration overview
Access: [MODE] > "AM FM PM Analog Demod"
Analog Modulation Analysis requires a special application on the R&S FPL1000.
When you activate an R&S FPL1-K7 AM/FM/PM Modulation Analysis application the
first time, a set of parameters is passed on from the currently active application. After
initial setup, the parameters for the channel setup are stored upon exiting and restored
upon re-entering the channel setup. Thus, you can switch between applications quickly
and easily.
When you activate the R&S FPL1-K7 AM/FM/PM Modulation Analysis application,
Analog Modulation Analysis for the input signal is started automatically with the default
configuration. The "AM FM PM Analog Demod" menu is displayed and provides
access to the most important configuration functions.
The remote commands required to perform these tasks are described in Chapter 11,
"Remote commands for AM/FM/PM Modulation Analysis", on page 117.
Predefined settings
For commonly performed measurements, standard setup files are provided for quick
and easy configuration. Simply load an existing standard settings file and, if necessary,
adapt the measurement settings to your specific requirements.
For an overview of predefined standards and settings see Chapter A, "Predefined
Using the R&S FPL1-K7 AM/FM/PM Modulation Analysis application you can perform
Analog Modulation Analysis using predefined standard setting files, or independently of
standards using user-defined measurement settings. Such settings can be stored for
recurrent use.
33User Manual 1178.3392.02 ─ 10
Page 34
R&S®FPL1-K7
Configuration
Configuration overview
Thus, configuring AM/FM/PM Modulation Analysis measurements requires one of the
following tasks:
●
Selecting an existing standard settings file and, if necessary, adapting the measurement settings to your specific requirements.
●
Configuring the measurement settings and, if necessary, storing the settings in a
file.
"Overview" window
Throughout the channel setup configuration, an overview of the most important currently defined settings is provided in the "Overview".
In addition to the main measurement settings, the "Overview" provides quick access to
the main settings dialog boxes. The individual configuration steps are displayed in the
order of the data flow. In particular, the "Overview" provides quick access to the following configuration dialog boxes (listed in the recommended order of processing):
1. Input/Frontend
See Chapter 5.3, "Input and frontend settings", on page 37
2. Trigger
See Chapter 5.4, "Trigger configuration", on page 45
3. Data Acquisition
See Chapter 5.5, "Data acquisition", on page 48
4. Demod/Display
See Chapter 5.6, "Demodulation display", on page 53
5. Demodulation Settings
See Chapter 5.7, "Demodulation", on page 53
6. Analysis
See Chapter 6, "Analysis", on page 76
34User Manual 1178.3392.02 ─ 10
Page 35
R&S®FPL1-K7
Configuration
Configuration according to standards
7. (Optionally:) Outputs
See Chapter 5.8, "Output settings", on page 72
To configure settings
► Select any button in the "Overview" to open the corresponding dialog box.
Select a setting in the channel setup bar (at the top of the channel setup tab) to
change a specific setting.
Preset Channel setup
Select the "Preset Channel" button in the lower left-hand corner of the "Overview" to
restore all measurement settings in the current channel setup to their default values.
Note: Do not confuse the "Preset Channel" button with the [Preset] key, which restores
the entire instrument to its default values and thus closes all channel setups on the
R&S FPL1000 (except for the default channel setup)!
Remote command:
SYSTem:PRESet:CHANnel[:EXEC] on page 126
Setup Standard
Opens a file selection dialog box to select a predefined setup file. See "Setup Stan-
dard"on page 36.
Specific Settings for
The channel setup can contain several windows for different results. Thus, the settings
indicated in the "Overview" and configured in the dialog boxes vary depending on the
selected window.
Select an active window from the "Specific Settings for" selection list that is displayed
in the "Overview" and in all window-specific configuration dialog boxes.
The "Overview" and dialog boxes are updated to indicate the settings for the selected
window.
5.2Configuration according to standards
Access: "Overview" > "Setup Standard"
Various predefined settings files for common standards are provided for use with the
R&S FPL1-K7 AM/FM/PM Modulation Analysis application. In addition, you can create
your own settings files for user-specific measurements.
For details on which settings are defined and an overview of predefined standards see
Chapter A, "Predefined standards and settings", on page 278.
└ Restore Standard Files................................................................................... 37
Setup Standard
Opens a file selection dialog box to select a predefined setup file. The predefined settings are configured in the R&S FPL1-K7 AM/FM/PM Modulation Analysis application,
which allows for quick and easy configuration for commonly performed measurements.
Selecting Storage Location - Drive/ Path/ Files ← Setup Standard
Select the storage location of the file on the instrument or an external drive.
The default storage location for the settings files is:
Contains the name of the data file without the path or extension.
File names must be compatible with the Windows conventions for file names. In partic-
ular, they must not contain special characters such as ":", "*", "?".
For details on the filename and location, see the "Data Management" topic in the
R&S FPL1000 User Manual.
36User Manual 1178.3392.02 ─ 10
Page 37
R&S®FPL1-K7
Configuration
Input and frontend settings
Load Standard ← Setup Standard
Loads the selected measurement settings file.
Remote command:
[SENSe:]ADEMod:PRESet[:STANdard] on page 127
Save Standard ← Setup Standard
Saves the current measurement settings for a specific standard as a file with the
defined name.
Remote command:
[SENSe:]ADEMod:PRESet:STORe on page 127
Delete Standard ← Setup Standard
Deletes the selected standard. Standards predefined by Rohde & Schwarz can also be
deleted. A confirmation query is displayed to avoid unintentional deletion of the standard.
Note: Restoring predefined standard files. The standards predefined by Rohde &
Schwarz available at the time of delivery can be restored using the "Restore Standard
Files" function (see "Restore Standard Files"on page 37).
Restore Standard Files ← Setup Standard
Restores the standards predefined by Rohde & Schwarz available at the time of delivery.
Note that this function overwrites customized standards that have the same name as
predefined standards.
Remote command:
[SENSe:]ADEMod:PRESet:RESTore on page 127
5.3Input and frontend settings
Access: "Overview" > "Input/Frontend"
The source and characteristics of the input signal to be demodulated are configured in
the "Input/Frontend Settings" dialog box.
Power sensor settings are described in the R&S FPL1000 User Manual.
Internal generator settings are described in the R&S FPL1000 User Manual.
●Radio frequency input............................................................................................. 37
●Settings for input from I/Q data files........................................................................39
The RF input connector of the R&S FPL1000 must be protected against signal levels
that exceed the ranges specified in the data sheet. Therefore, the R&S FPL1000 is
equipped with an overload protection mechanism. This mechanism becomes active as
soon as the power at the input mixer exceeds the specified limit. It ensures that the
connection between RF input and input mixer is cut off.
The RF input connector of the R&S FPL1000 must be protected against signal levels
that exceed the ranges specified in the data sheet. Therefore, the R&S FPL1000 is
equipped with an overload protection mechanism for DC and signal frequencies up to
30 MHz. This mechanism becomes active as soon as the power at the input mixer
exceeds the specified limit. It ensures that the connection between RF input and input
mixer is cut off.
When the overload protection is activated, an error message is displayed in the status
bar ("INPUT OVLD"), and a message box informs you that the RF input was disconnected. Furthermore, a status bit (bit 3) in the STAT:QUES:POW status register is set.
In this case, you must decrease the level at the RF input connector and then close the
message box. Then measurement is possible again. Reactivating the RF input is also
possible via the remote command INPut<ip>:ATTenuation:PROTection:RESet.
The power sensor functions are described in the R&S FPL1000 User Manual.
Radio Frequency State................................................................................................. 38
Activates input from the "RF Input" connector.
Remote command:
INPut<ip>:SELect on page 129
Impedance
For some measurements, the reference impedance for the measured levels of the
R&S FPL1000 can be set to 50 Ω or 75 Ω.
38User Manual 1178.3392.02 ─ 10
Page 39
R&S®FPL1-K7
Configuration
Input and frontend settings
Select 75 Ω if the 50 Ω input impedance is transformed to a higher impedance using a
75 Ω adapter of the RAZ type. (That corresponds to 25Ω in series to the input impedance of the instrument.) The correction value in this case is 1.76 dB = 10 log (75Ω/
50Ω).
This value also affects the unit conversion (see "Reference Level"on page 41).
This function is not available for input from the optional "Digital Baseband" interface or
from the optional "Analog Baseband" interface. For analog baseband input, an impedance of 50 Ω is always used.
Remote command:
INPut<ip>:IMPedance on page 129
YIG-Preselector
Enables or disables the YIG-preselector.
This setting requires the option R&S FPL1-B11 on the R&S FPL1000.
An internal YIG-preselector at the input of the R&S FPL1000 ensures that image fre-
quencies are rejected. However, image rejection is only possible for a restricted bandwidth. To use the maximum bandwidth for signal analysis you can disable the YIG-preselector at the input of the R&S FPL1000. However, disabling the YIG-preselector can
lead to image-frequency display.
Note: Note that the YIG-preselector is active only on frequencies greater than 6 GHz
(for models R&S FPL1014 and R&S FPL1026). Therefore, switching the YIG-preselector on or off has no effect if the frequency is below that value.
For measurements that require I/Q analysis in large bandwidths at frequencies higher
than 6 GHz, it is strongly recommended that you disable the YIG-preselector.
Remote command:
INPut<ip>:FILTer:YIG[:STATe] on page 129
SAW filter
The R&S FPL1000 hardware contains both a wide and a narrow IF path. Depending on
the used analysis bandwidth, the R&S FPL1000 determines which IF path to use automatically. The wide IF path allows for a smoother signal at the center frequency, while
the narrow IF path suppresses possibly distorting signals further away from the center
frequency. Using this setting, you can affect which IF path is used.
"Auto"
"Off"
Remote command:
INPut<ip>:FILTer:SAW on page 128
The R&S FPL1000 determines which IF path to use automatically,
depending on the used analysis bandwidth.
Select I/Q data file.........................................................................................................40
I/Q Input File State
Enables input from the selected I/Q input file.
If enabled, the application performs measurements on the data from this file. Thus,
most measurement settings related to data acquisition (attenuation, center frequency,
measurement bandwidth, sample rate) cannot be changed. The measurement time
can only be decreased to perform measurements on an extract of the available data
only.
Note: Even when the file input is disabled, the input file remains selected and can be
enabled again quickly by changing the state.
Remote command:
INPut<ip>:SELect on page 129
Select I/Q data file
Opens a file selection dialog box to select an input file that contains I/Q data.
The I/Q data must have a specific format (.iq.tar) as described in R&S FPL1000 I/Q
Analyzer and I/Q Input user manual.
The default storage location for I/Q data files is
Defines the expected maximum input signal level. Signal levels above this value are
possibly not measured correctly, which is indicated by the "IF Overload" status display
("OVLD" for analog baseband or digital baseband input).
Defines the expected maximum reference level. Signal levels above this value are possibly not measured correctly. Signals above the reference level are indicated by an "IF
Overload" status display.
The reference level can also be used to scale power diagrams; the reference level is
then used for the calculation of the maximum on the y-axis.
Since the hardware of the R&S FPL1000 is adapted according to this value, it is recommended that you set the reference level close above the expected maximum signal
level. Thus you ensure an optimal measurement (no compression, good signal-tonoise ratio).
Defines an arithmetic level offset. This offset is added to the measured level. In some
result displays, the scaling of the y-axis is changed accordingly.
Define an offset if the signal is attenuated or amplified before it is fed into the
R&S FPL1000 so the application shows correct power results. All displayed power
level results are shifted by this value.
The setting range is ±200 dB in 0.01 dB steps.
Note, however, that the internal reference level (used to adjust the hardware settings to
the expected signal) ignores any "Reference Level Offset". Thus, it is important to keep
in mind the actual power level the R&S FPL1000 must handle. Do not rely on the displayed reference level (internal reference level = displayed reference level - offset).
Remote command:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:RLEVel:
OFFSet on page 136
Unit ← Reference Level
The R&S FPL1000 measures the signal voltage at the RF input.
In the default state, the level is displayed at a power level of 1 mW (= dBm). Via the
known input impedance (50 Ω or 75 Ω, see "Impedance"on page 38), conversion to
other units is possible.
The following units are available and directly convertible:
●
dBm
●
dBmV
●
dBμV
●
dBμA
●
dBpW
●
Volt
●
Ampere
●
Watt
Remote command:
INPut<ip>:IMPedance on page 129
CALCulate<n>:UNIT:POWer on page 136
Setting the Reference Level Automatically (Auto Level) ← Reference Level
Automatically determines a reference level which ensures that no overload occurs at
the R&S FPL1000 for the current input data. At the same time, the internal attenuators
are adjusted. As a result, the signal-to-noise ratio is optimized, while signal compression and clipping are minimized.
To determine the required reference level, a level measurement is performed on the
R&S FPL1000.
If necessary, you can optimize the reference level further. Decrease the attenuation
level manually to the lowest possible value before an overload occurs, then decrease
the reference level in the same way.
You can change the measurement time for the level measurement if necessary (see
"Changing the Automatic Measurement Time (Meastime Manual)"on page 74).
42User Manual 1178.3392.02 ─ 10
Page 43
R&S®FPL1-K7
Configuration
Input and frontend settings
Remote command:
[SENSe:]ADJust:LEVel on page 174
Attenuation Mode / Value
The RF attenuation can be set automatically as a function of the selected reference
level (Auto mode). Automatic attenuation ensures that no overload occurs at the RF
Input connector for the current reference level. It is the default setting.
In "Manual" mode, you can set the RF attenuation in 5 dB steps down to 0 dB (with
option R&S FPL1-B22: in 1 dB steps). Other entries are rounded to the next integer
value. The range is specified in the data sheet. If the defined reference level cannot be
set for the defined RF attenuation, the reference level is adjusted accordingly and the
warning "limit reached" is displayed.
NOTICE! Risk of hardware damage due to high power levels. When decreasing the
attenuation manually, ensure that the power level does not exceed the maximum level
allowed at the RF input, as an overload can lead to hardware damage.
Remote command:
INPut<ip>:ATTenuation on page 137
INPut<ip>:ATTenuation:AUTO on page 137
Impedance
For some measurements, the reference impedance for the measured levels of the
R&S FPL1000 can be set to 50 Ω or 75 Ω.
Select 75 Ω if the 50 Ω input impedance is transformed to a higher impedance using a
75 Ω adapter of the RAZ type. (That corresponds to 25Ω in series to the input impedance of the instrument.) The correction value in this case is 1.76 dB = 10 log (75Ω/
50Ω).
This value also affects the unit conversion (see "Reference Level"on page 41).
This function is not available for input from the optional "Digital Baseband" interface or
from the optional "Analog Baseband" interface. For analog baseband input, an impedance of 50 Ω is always used.
Remote command:
INPut<ip>:IMPedance on page 129
Preamplifier
If the (optional) internal preamplifier hardware is installed, a preamplifier can be activated for the RF input signal.
You can use a preamplifier to analyze signals from DUTs with low output power.
Note: If an optional external preamplifier is activated, the internal preamplifier is auto-
matically disabled, and vice versa.
The input signal is amplified by 20 dB if the preamplifier option is activated.
Remote command:
Center Frequency......................................................................................................... 44
Center Frequency Stepsize...........................................................................................44
Center Frequency
Defines the center frequency of the signal in Hertz.
The allowed range of values for the center frequency depends on the frequency span.
span > 0: span
f
and span
max
/2 ≤ f
min
depend on the instrument and are specified in the data sheet.
min
center
≤ f
max
– span
min
/2
Remote command:
[SENSe:]FREQuency:CENTer on page 133
Center Frequency Stepsize
Defines the step size of the center frequency. The step size can be coupled to the
demodulation bandwidth, or you can set it to a fixed value manually.
"0.1 * Demod
BW"
"0.5 * Demod
BW"
"X * Demod
BW"
(default:) Sets the step size for the center frequency to 10 % of the
demodulation bandwidth.
Sets the step size for the center frequency to 50 % of the demodulation bandwidth.
Sets the step size for the center frequency to a manually defined factor of the demodulation bandwidth. The "X-Factor" defines the percentage of the demodulation bandwidth.
Values between 1 % and 100 % in steps of 1 % are allowed. The
default setting is 10 %.
"= Center"
Sets the step size to the value of the center frequency and removes
the coupling of the step size to the demodulation bandwidth. The
used value is indicated in the "Value" field.
"Manual"
Defines a fixed step size for the center frequency. Enter the step size
in the "Value" field.
44User Manual 1178.3392.02 ─ 10
Page 45
R&S®FPL1-K7
5.4Trigger configuration
Configuration
Trigger configuration
Remote command:
[SENSe:]FREQuency:CENTer:STEP:LINK on page 134
[SENSe:]FREQuency:CENTer:STEP:LINK:FACTor on page 134
[SENSe:]FREQuency:CENTer:STEP on page 134
Access: "Overview" > "Trigger"
Triggering means to capture the interesting part of the signal. Choosing the right trigger
type and configuring all trigger settings correctly allows you to detect various incidents
in your demodulated signals.
Optionally, the trigger signal used by the R&S FPL1000 can be output to a connected
device, and an external trigger signal from a connected device can be used by the
R&S FPL1000.
Trigger settings are identical to the base unit, except for the available trigger sources.
Gating is not available for Analog Modulation Analysis.
In the R&S FPL1-K7 AM/FM/PM Modulation Analysis application, the next measurement can be triggered if the selected input signal exceeds the threshold specified using
the "Trigger Level" setting (see "Trigger Level"on page 47). Thus, a periodic signal
modulated onto the carrier frequency can be displayed. It is recommended that the
measurement time covers at least five periods of the audio signal.
Remote command:
TRIGger[:SEQuence]:SOURce on page 150
Free Run ← Trigger Source
No trigger source is considered. Data acquisition is started manually or automatically
and continues until stopped explicitly.
Remote command:
TRIG:SOUR IMM, see TRIGger[:SEQuence]:SOURce on page 150
External Trigger 1 ← Trigger Source
Data acquisition starts when the TTL signal fed into the trigger input connector of the
R&S FPL1000 meets or exceeds the specified trigger level.
(See "Trigger Level"on page 47).
Remote command:
TRIG:SOUR EXT
See TRIGger[:SEQuence]:SOURceon page 150
IF Power ← Trigger Source
The R&S FPL1000 starts capturing data as soon as the trigger level is exceeded
around the third intermediate frequency.
For frequency sweeps, the third IF represents the start frequency. The trigger threshold
depends on the defined trigger level, as well as on the RF attenuation and preamplification. A reference level offset, if defined, is also considered. The trigger bandwidth at
the intermediate frequency depends on the RBW and sweep type. For details on available trigger levels and trigger bandwidths, see the instrument data sheet.
For measurements on a fixed frequency (e.g. zero span or I/Q measurements), the
third IF represents the center frequency.
This trigger source is only available for RF input.
For details on available trigger levels and trigger bandwidths, see the data sheet.
Remote command:
TRIG:SOUR IFP, see TRIGger[:SEQuence]:SOURce on page 150
FM (Offline) / AM (Offline) / PM (Offline) / RF (Offline) ← Trigger Source
Triggers when the demodulated input signal exceeds the trigger level.
46User Manual 1178.3392.02 ─ 10
Page 47
R&S®FPL1-K7
Configuration
Trigger configuration
Remote command:
TRIGger[:SEQuence]:SOURce on page 150
Time ← Trigger Source
Triggers in a specified repetition interval.
See "Repetition Interval"on page 47.
Remote command:
TRIG:SOUR TIME, see TRIGger[:SEQuence]:SOURce on page 150
Trigger Level
Defines the trigger level for the specified trigger source.
For details on supported trigger levels, see the instrument data sheet.
Remote command:
TRIGger[:SEQuence]:LEVel:IFPower on page 148
TRIGger[:SEQuence]:LEVel:IQPower on page 148
TRIGger[:SEQuence]:LEVel[:EXTernal<port>] on page 147
TRIGger[:SEQuence]:LEVel:AM:RELative on page 148
TRIGger[:SEQuence]:LEVel:AM[:ABSolute] on page 149
TRIGger[:SEQuence]:LEVel:FM on page 149
TRIGger[:SEQuence]:LEVel:PM on page 149
Repetition Interval
Defines the repetition interval for a time trigger.
The shortest interval is 2 ms.
Set the repetition interval to the exact pulse period, burst length, frame length or other
repetitive signal characteristic. If the required interval cannot be set with the available
granularity, configure a multiple of the interval that can be set. Thus, the trigger
remains synchronized to the signal.
Remote command:
TRIGger[:SEQuence]:TIME:RINTerval on page 151
Trigger Offset
Defines the time offset between the trigger event and the start of the sweep.
Offset > 0:Start of the sweep is delayed
Offset < 0:Sweep starts earlier (pretrigger)
For the "Time" trigger source, this function is not available.
Remote command:
TRIGger[:SEQuence]:HOLDoff[:TIME] on page 146
Hysteresis
Defines the distance in dB to the trigger level that the trigger source must exceed
before a trigger event occurs. Setting a hysteresis avoids unwanted trigger events
caused by noise oscillation around the trigger level.
47User Manual 1178.3392.02 ─ 10
Page 48
R&S®FPL1-K7
Configuration
Data acquisition
This setting is only available for "IF Power" trigger sources. The range of the value is
between 3 dB and 50 dB with a step width of 1 dB.
(For details see the R&S FPL1000 I/Q Analyzer and I/Q Input User Manual.)
Remote command:
TRIGger[:SEQuence]:IFPower:HYSTeresis on page 147
Drop-Out Time
Defines the time that the input signal must stay below the trigger level before triggering
again.
Remote command:
TRIGger[:SEQuence]:DTIMe on page 146
Slope
For all trigger sources except time, you can define whether triggering occurs when the
signal rises to the trigger level or falls down to it.
Remote command:
TRIGger[:SEQuence]:SLOPe on page 150
Trigger Holdoff
Defines the minimum time (in seconds) that must pass between two trigger events.
Trigger events that occur during the holdoff time are ignored.
Remote command:
TRIGger[:SEQuence]:IFPower:HOLDoff on page 147
5.5Data acquisition
Access: "Overview" > "Data Acquisition"
You configure how data is acquired and then demodulated in the "Data Acquisition"
dialog box.
Defines the demodulation bandwidth of the measurement. The demodulation bandwidth determines the sample rate with which the input signal is captured and analyzed.
For recommendations on finding the correct demodulation bandwidth, see Chapter 4.2,
"Demodulation bandwidth", on page 28.
Remote command:
[SENSe:]BWIDth:DEMod on page 144
Demodulation Filter
Defines the filter to be used for demodulation.
For details on sample rates, measurement times and trigger offsets for various demod-
ulation bandwidths when using a Gaussian filter, see Chapter 4.3, "Sample rate and
demodulation bandwidth", on page 29.
"Flat"
"Gauss"
Remote command:
[SENSe:]BWIDth:DEMod:TYPE on page 144
Measurement Time (AQT)
Defines how long data is acquired for demodulation.
Remote command:
[SENSe:]ADEMod:MTIMe on page 142
Default
Optimizes the settling behavior of the filter
Resolution Bandwidth
Defines the resolution bandwidth for data acquisition. The available range is specified
in the data sheet.
After triggering, starts the measurement and repeats it continuously until stopped.
While the measurement is running, the "Continuous Sweep" softkey and the [RUN
CONT] key are highlighted. The running measurement can be aborted by selecting the
highlighted softkey or key again. The results are not deleted until a new measurement
is started.
50User Manual 1178.3392.02 ─ 10
Page 51
R&S®FPL1-K7
Configuration
Data acquisition
Note: Sequencer. If the Sequencer is active, the "Continuous Sweep" softkey only controls the sweep mode for the currently selected channel setup. However, the sweep
mode only takes effect the next time the Sequencer activates that channel setup, and
only for a channel-defined sequence. In this case, a channel setup in continuous
sweep mode is swept repeatedly.
Furthermore, the [RUN CONT] key controls the Sequencer, not individual sweeps.
[RUN CONT] starts the Sequencer in continuous mode.
For details on the Sequencer, see the R&S FPL1000 User Manual.
Remote command:
INITiate<n>:CONTinuous on page 176
Single Sweep / Run Single
After triggering, starts the number of sweeps set in "Sweep Count". The measurement
stops after the defined number of sweeps has been performed.
While the measurement is running, the "Single Sweep" softkey and the [RUN SINGLE]
key are highlighted. The running measurement can be aborted by selecting the highlighted softkey or key again.
Note: Sequencer. If the Sequencer is active, the "Single Sweep" softkey only controls
the sweep mode for the currently selected channel setup. However, the sweep mode
only takes effect the next time the Sequencer activates that channel setup, and only for
a channel-defined sequence. In this case, the Sequencer sweeps a channel setup in
single sweep mode only once.
Furthermore, the [RUN SINGLE] key controls the Sequencer, not individual sweeps.
[RUN SINGLE] starts the Sequencer in single mode.
If the Sequencer is off, only the evaluation for the currently displayed channel setup is
updated.
For details on the Sequencer, see the R&S FPL1000 User Manual.
Remote command:
INITiate<n>[:IMMediate] on page 177
Continue Single Sweep
After triggering, repeats the number of sweeps set in "Sweep Count", without deleting
the trace of the last measurement.
While the measurement is running, the "Continue Single Sweep" softkey and the [RUN
SINGLE] key are highlighted. The running measurement can be aborted by selecting
the highlighted softkey or key again.
Remote command:
INITiate<n>:CONMeas on page 176
Measurement Time (AQT)
Defines how long data is acquired for demodulation.
Remote command:
[SENSe:]ADEMod:MTIMe on page 142
Sweep Points
Defines the number of measured values to be collected during one sweep.
51User Manual 1178.3392.02 ─ 10
Page 52
R&S®FPL1-K7
Configuration
Data acquisition
All values from 101 to 100001 can be set. The default value is 1001 sweep points.
Remote command:
[SENSe:]SWEep[:WINDow<n>]:POINts on page 145
Sweep/Average Count
Defines the number of sweeps to be performed in the single sweep mode. Values from
0 to 200000 are allowed. If the values 0 or 1 are set, one sweep is performed.
The sweep count is applied to all the traces in all diagrams.
If the trace modes "Average", "Max Hold" or "Min Hold" are set, this value also deter-
mines the number of averaging or maximum search procedures.
In continuous sweep mode, if "Sweep Count" = 0 (default), averaging is performed
over 10 sweeps. For "Sweep Count" =1, no averaging, maxhold or minhold operations
are performed.
Remote command:
[SENSe:]SWEep:COUNt on page 145
[SENSe:]AVERage<n>:COUNt on page 212
Select Frame
Selects a specific frame, loads the corresponding trace from the memory, and displays
it in the Spectrum window.
Note that activating a marker or changing the position of the active marker automatically selects the frame that belongs to that marker.
This function is only available in single sweep mode or if the sweep is stopped, and
only if a spectrogram is selected.
The most recent frame is number 0, all previous frames have a negative number.
For more details, see the R&S FPL1000 User Manual.
Remote command:
CALCulate<n>:SPECtrogram:FRAMe:SELect on page 202
Continue Frame
Determines whether the results of the previous sweeps are included in the analysis of
the next sweeps for trace modes "Max Hold", "Min Hold", and "Average".
This function is available in single sweep mode only.
●
On
When the average or peak values are determined for the new sweep, the results of
the previous sweeps in the spectrogram are also considered.
●
Off
The average or peak values are determined from the results of the newly swept
frames only.
Remote command:
CALCulate<n>:SPECtrogram:CONTinuous on page 201
Frame Count
Determines how many frames are plotted during a single sweep (as opposed to a continuous sweep). The maximum number of possible frames depends on the history
depth (see "History Depth"on page 82).
52User Manual 1178.3392.02 ─ 10
Page 53
R&S®FPL1-K7
5.6Demodulation display
Configuration
Demodulation
Remote command:
CALCulate<n>:SPECtrogram:FRAMe:COUNt on page 201
Clear Spectrogram
Resets the spectrogram result display and clears the history buffer.
This function is only available if a spectrogram is selected.
Remote command:
CALCulate<n>:SPECtrogram:CLEar[:IMMediate] on page 200
Access: "Overview" > "Demod/Display"
The demodulated signal can be displayed using various evaluation methods. All evaluation methods available for the Analog Modulation Analysis application are displayed in
the evaluation bar in SmartGrid mode when you do one of the following:
Up to six evaluation methods can be displayed simultaneously in separate windows.
The Analog Modulation Analysis evaluation methods are described in Chapter 3, "Mea-
surements and result displays", on page 15.
For details on working with the SmartGrid, see the R&S FPL1000 Getting Started manual.
Zero Phase Reference Position (PM Time Domain only)..............................................56
Phase Wrap On/Off (PM Time Domain only)................................................................ 56
Squelch State
Activates the squelch function, that is: if the signal falls below a defined threshold, the
demodulated data is automatically set to 0. This is useful, for example, to avoid
demodulation noise during transmission breaks.
This function is only available for FM demodulation.
Remote command:
[SENSe:]ADEMod:SQUelch[:STATe] on page 153
Squelch Level
Defines the level threshold below which the demodulated data is set to 0 if squelching
is enabled. The squelch level is an absolute value.
Remote command:
[SENSe:]ADEMod:SQUelch:LEVel on page 153
AF Coupling
Controls the automatic correction of the frequency offset and phase offset of the input
signal:
This function is only available for FM or PM time domain evaluations.
54User Manual 1178.3392.02 ─ 10
Page 55
R&S®FPL1-K7
Configuration
Demodulation
●
FM time evaluation
If DC is selected, the absolute frequency is displayed. That means, an input signal
with an offset relative to the center frequency is not displayed symmetrically to the
zero line.
If AC is selected, the frequency offset is automatically corrected, i.e. the trace is
always symmetric to the zero line.
●
PM time evaluation
If DC is selected, the phase runs according to the existing frequency offset. In addition, the DC signal contains a phase offset of ± π.
If AC is selected, the frequency offset and phase offset are automatically corrected,
i.e. the trace is always symmetric to the zero line.
Remote command:
[SENSe:]ADEMod<n>:AF:COUPling on page 152
Selected Trace
Defines the trace used to determine the results in the "Result Summary".
Time Domain Zoom
Using the time domain zoom, the demodulated data for a particular time span is extracted and displayed in more detail. Time domain zoom is useful if the measurement time
is very large and thus each sweep point represents a large time span. The time domain
zoom function distributes the available sweep points only among the time span defined
by the zoom area length. The time span displayed per division of the diagram is
decreased. Thus, the display of the extracted time span becomes more precise. Note
that the time domain zoom area affects not only the diagram display, but the entire
evaluation for the current window.
This function is only available for evaluations in the time domain.
Tip: In addition to the Time Domain Zoom, a graphical zoom is available for all diagram
evaluations. However, the graphical zoom is useful only if more measured values than
trace points are available. The (time) span represented by each measurement point
remains the same.
For details see the R&S FPL1000 User Manual.
State ← Time Domain Zoom
Activates or deactivates the time domain zoom mode.
"On"
"Off"
Remote command:
[SENSe:]ADEMod<n>:ZOOM[:STATe] on page 156
Start ← Time Domain Zoom
Defines the start time for the time domain zoom area. For spectrum evaluations, the
start time is always 0.
Activates the time domain zoom.
Deactivates the time domain zoom and restores the original display. If
more measured values than measurement points are available, several measured values are combined in one measurement point
according to the method of the selected trace detector.
55User Manual 1178.3392.02 ─ 10
Page 56
R&S®FPL1-K7
Configuration
Demodulation
Remote command:
[SENSe:]ADEMod<n>:ZOOM:STARt on page 155
Length ← Time Domain Zoom
Defines the length of the time domain zoom area. Enter the length as a time value
manually, or use the "Auto" setting to set the length to the current number of sweep
points automatically.
Remote command:
[SENSe:]ADEMod<n>:ZOOM:LENGth on page 155
[SENSe:]ADEMod<n>:ZOOM:LENGth:MODE on page 155
Zero Phase Reference Position (PM Time Domain only)
Defines the position in time at which the phase of the PM-demodulated signal is set to
0 rad. In the default setting, the time of the first measured value is set to 0 rad. You can
define a different position manually, or select the time of the last measured value as the
reference position. The time of the last measured value corresponds to the acquisition
time, considering the trigger event and trigger offset, if applicable. If the acquisition
time or the trigger values are changed, the reference position is automatically adapted.
This setting is only available for PM time domain displays with DC coupling.
Remote command:
[SENSe:]ADEMod:PM:RPOint[:X] on page 152
[SENSe:]ADEMod:PM:RPOint[:X]:MODE on page 153
Phase Wrap On/Off (PM Time Domain only)
Activates/deactivates the phase wrap.
OnThe phase is displayed in the range ±180° (± Π). For example, if the phase exceeds +180°,
360° is subtracted from the phase value, with the display thus showing >-180°.
OffThe phase is not wrapped.
This setting is only available for PM time domain displays with DC coupling.
Remote command:
CALC:FORM PHAS/CALC:FORM UPH, see CALCulate<n>:FORMat on page 154
These settings are only available for AF Spectrum evaluations, not in the time domain.
AF Center......................................................................................................................57
AF Start.........................................................................................................................57
AF Stop......................................................................................................................... 57
AF Span........................................................................................................................ 58
AF Full Span................................................................................................................. 58
AF Center
Defines the center frequency of the demodulated data to evaluate.
Remote command:
[SENSe:]ADEMod:AF:CENTer on page 157
AF Start
Defines the start frequency of the demodulated data to evaluate.
Remote command:
[SENSe:]ADEMod:AF:STARt on page 157
AF Stop
Defines the stop frequency of the demodulated data to evaluate.
The maximum AF stop frequency corresponds to half the demodulation bandwidth.
Remote command:
[SENSe:]ADEMod:AF:STOP on page 158
57User Manual 1178.3392.02 ─ 10
Page 58
R&S®FPL1-K7
5.7.2.2RF evaluation
Configuration
Demodulation
AF Span
Defines the span (around the center frequency) of the demodulated data to evaluate.
The maximum span is DBW/2.
Remote command:
[SENSe:]ADEMod:AF:SPAN on page 157
AF Full Span
Sets the span (around the center frequency) of the demodulated data to the maximum
of DBW/2.
These settings are only available for RF evaluation, both in time and frequency
domain. Note that for RF data the center frequency and demodulation bandwidth correspond to the settings defined in the "Input" and "Data Acquisition" configuration.
Center Frequency......................................................................................................... 58
RF Full Span................................................................................................................. 59
Center Frequency
Defines the center frequency of the signal in Hertz.
The allowed range of values for the center frequency depends on the frequency span.
span > 0: span
min
/2 ≤ f
center
≤ f
max
– span
min
/2
58User Manual 1178.3392.02 ─ 10
Page 59
R&S®FPL1-K7
Configuration
Demodulation
f
and span
max
depend on the instrument and are specified in the data sheet.
min
Remote command:
[SENSe:]FREQuency:CENTer on page 133
Span
Defines the frequency span. The center frequency is kept constant. The following
range is allowed:
span = 0: 0 Hz
span >0:
span
and f
f
min
max
and span
max
≤ f
span
=DBW/2
min
≤ f
max
are specified in the data sheet.
Remote command:
[SENSe:]ADEMod:SPECtrum:SPAN[:MAXimum] on page 159
[SENSe:]ADEMod:SPECtrum:SPAN:ZOOM on page 158
Demodulation Bandwidth
Defines the demodulation bandwidth of the measurement. The demodulation bandwidth determines the sample rate with which the input signal is captured and analyzed.
For recommendations on finding the correct demodulation bandwidth, see Chapter 4.2,
"Demodulation bandwidth", on page 28.
Remote command:
[SENSe:]BWIDth:DEMod on page 144
RF Full Span
Sets the span (around the center frequency) of the RF data to be evaluated to the
demodulation bandwidth.
Remote command:
[SENSe:]ADEMod:SPECtrum:SPAN[:MAXimum] on page 159
Deactivating all AF Filters............................................................................................. 62
High Pass
Defines a high pass filter with the given limit to separate the DC component. The filters
are indicated by the 3 dB cutoff frequency. The 50 Hz and 300 Hz filters are designed
as 2nd-order Butterworth filter (12 dB/octave). The 20 Hz filter is designed as 3rd-order
Butterworth filter (18 dB/octave).
The high pass filters are active in the following demodulation bandwidth range:
NoneNo AF Filter used (default)
20 Hz100 Hz ≤ demodulation bandwidth ≤ 1.6 MHz
50 Hz:200 Hz ≤ demodulation bandwidth ≤ 3 MHz
300 Hz:800 Hz ≤ demodulation bandwidth ≤ 8 MHz
Manual:A high pass filter with the manually defined frequency is used.
Remote command:
[SENSe:]FILTer<n>:HPASs[:STATe] on page 162
[SENSe:]FILTer<n>:HPASs:FREQuency[:ABSolute] on page 162
[SENSe:]FILTer<n>:HPASs:FREQuency:MANual on page 162
Low Pass
Defines a low pass filter type. Relative and absolute low pass filter are available.
●
Absolute low pass filters:
60User Manual 1178.3392.02 ─ 10
Page 61
R&S®FPL1-K7
Configuration
Demodulation
Absolute filters are indicated by the 3 dB cutoff frequency. The 3 kHz, 15 kHz and
23 kHz filters are designed as 5th-order Butterworth filters (30 dB/octave). The 150
kHz filter is designed as 8th-order Butterworth filter (48 dB/octave).
The absolute low pass filters are active in the following demodulation bandwidth
range:
Filter typeDemodulation bandwidth
3 kHz:6.4 kHz ≤ demodulation bandwidth ≤ 3 MHz
15 kHz:50 kHz ≤ demodulation bandwidth ≤ 8 MHz
23 kHz50 kHz ≤ demodulation bandwidth ≤ 18 MHz
150 kHz:400 kHz ≤ demodulation bandwidth ≤ 8 MHz
Manual:A low pass filter with the manually defined frequency is used.
●
Relative low pass filters:
Relative filters (3 dB) can be selected in % of the demodulation bandwidth. The filters are designed as 5th-order Butterworth filter (30 dB/octave) and active for all
demodulation bandwidths.
●
"NONE" deactivates the AF low pass filter (default).
Remote command:
[SENSe:]FILTer<n>:LPASs[:STATe] on page 164
[SENSe:]FILTer<n>:LPASs:FREQuency[:ABSolute] on page 163
[SENSe:]FILTer<n>:LPASs:FREQuency:RELative on page 163
[SENSe:]FILTer<n>:LPASs:FREQuency:MANual on page 163
Weighting
Selects a weighting AF filter. By default, no weighting filter is active.
"A weighted"
Switches on the A weighted filter. The weighting filter is active in the
following demodulation bandwidth range:
100 kHz ≤ demodulation bandwidth ≤ 800 kHz
"CCITT"
Switches on a CCITT P.53 weighting filter. The weighting filter is
active in the following demodulation bandwidth range:
20 kHz ≤ demodulation bandwidth ≤ 3 MHz
"CCIR weighted"
Switches on the CCIR weighted filter. The weighting filter is active in
the following demodulation bandwidth range:
100 kHz ≤ demodulation bandwidth ≤ 3.0 MHz
"CCIR
unweighted"
Switches on the CCIR unweighted filter, which is the combination of
the 20 Hz highpass and 23 kHz low pass filter. The weighting filter is
active in the following demodulation bandwidth range:
50 kHz ≤ demodulation bandwidth ≤ 1.6 MHz
Remote command:
[SENSe:]FILTer<n>:CCITt[:STATe] on page 161
[SENSe:]FILTer<n>:CCIR[:UNWeighted][:STATe] on page 160
[SENSe:]FILTer<n>:CCIR:WEIGhted[:STATe] on page 160
[SENSe:]FILTer<n>:AWEighted[:STATe] on page 159
61User Manual 1178.3392.02 ─ 10
Page 62
R&S®FPL1-K7
Configuration
Demodulation
Deemphasis
Activates a deemphasis filter with the given time constant.
Sometimes a modulated signal is extorted by a pre-emphasis filter before transmission,
for example to eliminate frequencies that are more prone to interferences. In this case,
the emphasis function must be reversed after demodulation, which is done by the
deemphasis filter.
The deemphasis filter is active in the following demodulation bandwidth range:
25 µs:25 kHz ≤ demodulation bandwidth ≤ 40 MHz
50 µs:6.4 kHz ≤ demodulation bandwidth ≤ 18 MHz
75 µs:6.4 kHz ≤ demodulation bandwidth ≤ 18 MHz
750 µs:800 Hz ≤ demodulation bandwidth ≤ 3 MHz
Depending on the deemphasis filter, a minimum demodulation bandwidth is required
for an error less than 0.5 dB, up to a maximum AF frequency. The following table
shows the dependencies.
AF Auto Scale............................................................................................................... 65
Dev per Division/ dB per Division
Defines the modulation depth or the phase deviation or frequency deviation per division (logarithmic: 0.1 dB to 20 dB):
AM display:0.0001 % to 1000 %
FM display:
PM display:0.0001 rad/div to 1000 rad/div
Note: The value defined per division refers to the default display of 10 divisions on the
y-axis. If fewer divisions are displayed (e.g. because the window is reduced in height),
the range per division is increased. Thus, the same result range is displayed in the
smaller window. In this case, the per division value does not correspond to the actual
display.
Determines the position of the reference value for the modulation depth or the phase
deviation or frequency deviation on the y-axis of the diagram.
63User Manual 1178.3392.02 ─ 10
Page 64
R&S®FPL1-K7
Configuration
Demodulation
The position is entered as a percentage of the diagram height with 100 % corresponding to the upper diagram border. The default setting is 50 % (diagram center) for the
AF time evaluations and 100 % (upper diagram border) for the AF spectrum evaluations.
Determines the modulation depth or the phase deviation or the frequency deviation at
the reference line of the y-axis. The reference value can be set specifically for each
evaluation.
●
AF time display
The trace display takes individual frequency/phase offsets into account (in contrast,
the AF Coupling setting permits automatic correction by the average frequency/
phase offset of the signal, and therefore cannot be activated simultaneously).
●
AF spectrum display
In the default setting, the reference value defines the modulation depth or the
FM/PM deviation at the upper diagram border.
Possible values:
●
AM: 0 and ± 10000 %
●
FM: 0 and ± 10 MHz
●
PM: 0 and ± 10000 rad
Note: The reference value for the AF range in the window title bar is displayed with
respect to the defined reference position. The position can vary for different windows.
For time domain and frequency domain windows, for example, a different reference
value can be displayed, although the same reference is actually used (but the positions
vary).
Controls the automatic correction of the frequency offset and phase offset of the input
signal:
This function is only available for FM or PM time domain evaluations.
●
FM time evaluation
If DC is selected, the absolute frequency is displayed. That means, an input signal
with an offset relative to the center frequency is not displayed symmetrically to the
zero line.
If AC is selected, the frequency offset is automatically corrected, i.e. the trace is
always symmetric to the zero line.
●
PM time evaluation
If DC is selected, the phase runs according to the existing frequency offset. In addition, the DC signal contains a phase offset of ± π.
If AC is selected, the frequency offset and phase offset are automatically corrected,
i.e. the trace is always symmetric to the zero line.
64User Manual 1178.3392.02 ─ 10
Page 65
R&S®FPL1-K7
5.7.4.2RF evaluation
Configuration
Demodulation
Remote command:
[SENSe:]ADEMod<n>:AF:COUPling on page 152
Deviation
Switches between logarithmic and linear display of the modulation depth or the phase
deviation or the frequency deviation.
Remote command:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing on page 141
AF Auto Scale
Activates automatic scaling of the y-axis for AF measurements. RF power and RF
spectrum measurements are not affected by the auto-scaling.
Remote command:
[SENSe:]ADJust:SCALe[:Y]:AUTO[:CONTinuous] on page 174
Access: "Overview" > "Demod Settings" > "Scaling"
Or: "Meas Setup" > "Demod" > "Scaling" tab
These settings are only available for RF evaluations and the "result summary".
Defines the displayed y-axis range in dB.
The default value is 100 dB.
65User Manual 1178.3392.02 ─ 10
Page 66
R&S®FPL1-K7
Configuration
Demodulation
For Analog Modulation Analysis measurements, time domain scaling is defined in Hz
(default: 500 kHz).
Remote command:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe] on page 139
Ref Level Position
Defines the reference level position, i.e. the position of the maximum AD converter
value on the level axis in %.
0 % corresponds to the lower and 100 % to the upper limit of the diagram.
Values from -120 % to +600 % are available. Larger values are useful for small scales,
such as a power range of 10 dB or 20 dB, and low signal levels, for example 60 dB
below the reference level. In this case, large reference level position values allow you
to see the trace again.
Only available for RF measurements.
Remote command:
Automatically determines the optimal range and reference level position to be displayed for the current measurement settings.
The display is only set once; it is not adapted further if the measurement settings are
changed again.
Remote command:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:AUTO ONCE
on page 139
Scaling
Defines the scaling method for the y-axis.
"Logarithmic"
"Linear with
Unit"
"Linear Percent"
"Absolute"
"Relative"
Remote command:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing on page 141
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:MODE
on page 139
Logarithmic scaling (only available for logarithmic units - dB..., and A,
V, Watt)
Linear scaling in the unit of the measured signal
Linear scaling in percentages from 0 to 100
The labeling of the level lines refers to the absolute value of the reference level (not available for "Linear Percent")
The scaling is in dB, relative to the reference level (only available for
logarithmic units - dB...). The upper line of the grid (reference level) is
always at 0 dB.
66User Manual 1178.3392.02 ─ 10
Page 67
R&S®FPL1-K7
5.7.5Units
Configuration
Demodulation
Access: "Overview" > "Demod Settings" > "Unit"
Or: "Meas Setup" > "Demod" > "Unit" tab
The units define how the demodulated data is displayed.
Phase Unit (Rad/Deg)...................................................................................................67
THD Unit (%/ DB)..........................................................................................................67
The demodulation results are displayed in the "Result Summary" table (see also
"Result Summary"on page 23). The detectors used to determine the results can be
configured.
In addition to common absolute demodulation, the R&S FPL1-K7 AM/FM/PM Modulation Analysis application also provides demodulation results relative to user-defined or
measured reference values in the "Result Summary".
The settings for the "Result Summary" can be defined individually for the different modulation types (FM, AM, PM). For each modulation, a separate tab is provided in the dialog box.
Detector type for demodulation results
"+ Peak"
"- Peak"
"+/- Peak"
"RMS"
Remote command:
The detector is specified by the DETector<det> suffix in
CONFigure:RELative:AM|FM|PM:DETector<det>... commands.
Positive peak
Negative peak
Autopeak
Root mean square
Mode
Defines the mode with which the demodulation result is determined.
68User Manual 1178.3392.02 ─ 10
Page 69
R&S®FPL1-K7
Configuration
Demodulation
The modes are similar to those for the entire trace (see "Trace Mode"on page 77).
"Clear Write"
"Max Hold"
"Average"
Remote command:
CONFigure:ADEMod:RESults:AM:DETector<det>:MODE on page 168
CONFigure:ADEMod:RESults:FM:DETector<det>:MODE on page 168
CONFigure:ADEMod:RESults:PM:DETector<det>:MODE on page 168
State
Activates relative demodulation for the selected detector. If activated, the demodulated
result is set in relation to the Reference Value.
Remote command:
CONFigure:ADEMod:RESults:AM:DETector<det>:STATe on page 167
CONFigure:ADEMod:RESults:FM:DETector<det>:STATe on page 167
CONFigure:ADEMod:RESults:PM:DETector<det>:STATe on page 167
Overwrite mode: the detector value is overwritten by each sweep.
This is the default setting.
The maximum value is determined over several sweeps and displayed. The R&S FPL1000 saves each result only if the new value is
greater than the previous one.
The average result is determined over all sweeps.
Reference Value
Defines the reference value to be used for relative demodulation results and recalculates the results. If necessary, the detector is activated.
Note: A reference value 0 would provide infinite results and is thus automatically cor-
rected to 0.1.
Remote command:
CONFigure:ADEMod:RESults:AM:DETector<det>:REFerence on page 167
CONFigure:ADEMod:RESults:FM:DETector<det>:REFerence on page 167
CONFigure:ADEMod:RESults:PM:DETector<det>:REFerence on page 167
Meas -> Reference
Sets the Reference Value to be used for relative demodulation results to the currently
measured value for all relative detectors.
Note: A reference value 0 would provide infinite results and is thus automatically cor-
rected to 0.1.
If necessary, the detectors are activated.
Remote command:
The settling time defines when the signal remains within a specified target corridor. The
corridor is defined by the upper and lower settling limit. The function is available for all
time domain displays.
If enabled, the time is determined by evaluating the signal values from the end of the
measurement. The last position at which the signal exceeds the corridor is the settling
time. The settling time is defined with reference to the start of the data acquisition or
the trigger event. A possible trigger offset is not considered.
The settling time is indicated by a temporary marker in the time domain diagram. The
result is also indicated in the "result summary". The settling time is evaluated for each
time domain window for the selected trace (see "Selected Trace"on page 55).
Defines the upper limit of the settling time corridor. The value is defined with reference
to the reference value, see also "Reference Value"on page 64 and "Zero Phase Refer-
ence Position (PM Time Domain only)"on page 56.
Remote command:
[SENSe:]ADEMod:SETTling:TIME:LIMit:UPPer on page 170
Lower Settling Limit
Defines the lower limit of the settling time corridor. The value is defined with reference
to the reference value, see also "Reference Value"on page 64 and "Zero Phase Refer-
ence Position (PM Time Domain only)"on page 56.
Remote command:
[SENSe:]ADEMod:SETTling:TIME:LIMit:LOWer on page 169
71User Manual 1178.3392.02 ─ 10
Page 72
R&S®FPL1-K7
5.8Output settings
Configuration
Adjusting settings automatically
Access: "Overview" > "Output"
The R&S FPL1000 can provide signals to different output connectors.
These connectors are only available if the R&S FPL1-B5 option is installed.
For details on connectors, refer to the R&S FPL1000 Getting Started manual, "Front /
Rear Panel View" chapters.
Enables or disables the 28 V voltage supply for an external noise source connected to
the "Noise source control / Power sensor") connector. By switching the supply voltage
for an external noise source on or off in the firmware, you can enable or disable the
device as required.
This connector is only available if the R&S FPL1-B5 option is installed.
External noise sources are useful when you are measuring power levels that fall below
the noise floor of the R&S FPL1000 itself, for example when measuring the noise level
of an amplifier.
In this case, you can first connect an external noise source (whose noise power level is
known in advance) to the R&S FPL1000 and measure the total noise power. From this
value, you can determine the noise power of the R&S FPL1000. Then when you measure the power level of the actual DUT, you can deduct the known noise level from the
total power to obtain the power level of the DUT.
Some settings can be adjusted by the R&S FPL1000 automatically according to the
current measurement settings. To do so, a measurement is performed. You can configure this measurement.
Adjusting settings automatically during triggered measurements
When you select an auto adjust function, a measurement is performed to determine
the optimal settings. If you select an auto adjust function for a triggered measurement,
you are asked how you want the R&S FPL1000 to behave:
●
(default:) The measurement for adjustment waits for the next trigger
●
The measurement for adjustment is performed without waiting for a trigger.
The trigger source is temporarily set to "Free Run". After the measurement is completed, the original trigger source is restored. The trigger level is adjusted as follows:
–For IF Power and RF Power triggers:
Trigger Level = Reference Level - 15 dB
–For Video trigger:
Trigger Level = 85 %
Remote command:
[SENSe:]ADJust:CONFigure:TRIGger on page 174
Adjusting all Determinable Settings Automatically (Auto All)........................................ 73
Adjusting the Center Frequency Automatically (Auto Frequency)................................ 74
AF Auto Scale............................................................................................................... 74
Setting the Reference Level Automatically (Auto Level)...............................................74
Resetting the Automatic Measurement Time (Meastime Auto).....................................74
Changing the Automatic Measurement Time (Meastime Manual)................................ 74
Adjusting all Determinable Settings Automatically (Auto All)
Activates all automatic adjustment functions for the current measurement settings,
including:
73User Manual 1178.3392.02 ─ 10
Page 74
R&S®FPL1-K7
Configuration
Adjusting settings automatically
●
Auto Frequency
●
Auto Level
●
"AF Auto Scale"on page 65
Remote command:
[SENSe:]ADJust:ALL on page 171
Adjusting the Center Frequency Automatically (Auto Frequency)
The R&S FPL1000 adjusts the center frequency automatically.
The optimum center frequency is the frequency with the highest S/N ratio in the fre-
quency span. As this function uses the signal counter, it is intended for use with sinusoidal signals.
To set the optimal reference level, see "Setting the Reference Level Automatically
(Auto Level)"on page 42).
Remote command:
[SENSe:]ADJust:FREQuency on page 174
AF Auto Scale
Activates automatic scaling of the y-axis for AF measurements. RF power and RF
spectrum measurements are not affected by the auto-scaling.
Remote command:
[SENSe:]ADJust:SCALe[:Y]:AUTO[:CONTinuous] on page 174
Setting the Reference Level Automatically (Auto Level)
Automatically determines a reference level which ensures that no overload occurs at
the R&S FPL1000 for the current input data. At the same time, the internal attenuators
are adjusted. As a result, the signal-to-noise ratio is optimized, while signal compression and clipping are minimized.
To determine the required reference level, a level measurement is performed on the
R&S FPL1000.
If necessary, you can optimize the reference level further. Decrease the attenuation
level manually to the lowest possible value before an overload occurs, then decrease
the reference level in the same way.
You can change the measurement time for the level measurement if necessary (see
"Changing the Automatic Measurement Time (Meastime Manual)"on page 74).
Remote command:
[SENSe:]ADJust:LEVel on page 174
Resetting the Automatic Measurement Time (Meastime Auto)
Resets the measurement duration for automatic settings to the default value.
Remote command:
[SENSe:]ADJust:CONFigure:LEVel:DURation:MODE on page 172
Changing the Automatic Measurement Time (Meastime Manual)
This function allows you to change the measurement duration for automatic setting
adjustments. Enter the value in seconds.
74User Manual 1178.3392.02 ─ 10
Page 75
R&S®FPL1-K7
Configuration
Adjusting settings automatically
Note: The maximum measurement duration depends on the currently selected measurement and the installed (optional) hardware. Thus, the measurement duration
actually used to determine the automatic settings can be shorter than the value you
define here.
Remote command:
[SENSe:]ADJust:CONFigure:LEVel:DURation:MODE on page 172
[SENSe:]ADJust:CONFigure:LEVel:DURation on page 172
Upper Level Hysteresis
When the reference level is adjusted automatically using the Auto Level function, the
internal attenuators and the preamplifier are also adjusted. To avoid frequent adaptation due to small changes in the input signal, you can define a hysteresis. This setting
defines an upper threshold the signal must exceed (compared to the last measurement) before the reference level is adapted automatically.
Remote command:
[SENSe:]ADJust:CONFigure:HYSTeresis:UPPer on page 173
Lower Level Hysteresis
When the reference level is adjusted automatically using the Auto Level function, the
internal attenuators and the preamplifier are also adjusted. To avoid frequent adaptation due to small changes in the input signal, you can define a hysteresis. This setting
defines a lower threshold the signal must fall below (compared to the last measurement) before the reference level is adapted automatically.
Remote command:
[SENSe:]ADJust:CONFigure:HYSTeresis:LOWer on page 173
75User Manual 1178.3392.02 ─ 10
Page 76
R&S®FPL1-K7
6Analysis
Analysis
Trace settings
Access: "Overview" > "Analysis"
General result analysis settings concerning the trace, markers, lines etc. are identical
to the analysis functions in the base unit except for the special marker functions.
For a description of the lines functionality, see the R&S FPL1000 User Manual.
The remote commands required to perform these tasks are described in Chapter 11,
"Remote commands for AM/FM/PM Modulation Analysis", on page 117.
You can configure the settings for up to 6 individual traces in the same result display.
Each trace is displayed in a different color, indicated in the window title bar and the
trace settings.
In the R&S FPL1-K7 AM/FM/PM Modulation Analysis application, when you configure
the traces for a window with a specific evaluation (e.g. AM time domain), the traces in
all windows with the same evaluation are configured identically.
Trace data can also be exported to an ASCII file for further analysis. For details see
Chapter 6.3, "Trace / data export configuration", on page 85.
Selects the corresponding trace for configuration. The currently selected trace is highlighted.
Remote command:
Selected via numeric suffix of:TRACe<1...6> commands
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>[:STATe] on page 210
Trace Mode
Defines the update mode for subsequent traces.
"Clear/ Write"
"Max Hold"
"Min Hold"
"Average"
Overwrite mode (default): the trace is overwritten by each sweep.
The maximum value is determined over several sweeps and dis-
played. The R&S FPL1000 saves each trace point in the trace memory only if the new value is greater than the previous one.
The minimum value is determined from several measurements and
displayed. The R&S FPL1000 saves each trace point in the trace
memory only if the new value is lower than the previous one.
The average is formed over several sweeps.
The Sweep/Average Count determines the number of averaging procedures.
77User Manual 1178.3392.02 ─ 10
Page 78
R&S®FPL1-K7
Analysis
Trace settings
"View"
"Blank"
Remote command:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:MODE on page 208
Detector
Defines the trace detector to be used for trace analysis.
"Auto"
"Type"
Remote command:
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion] on page 213
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]:AUTO on page 213
The current contents of the trace memory are frozen and displayed.
Removes the selected trace from the display.
(default:) Selects the optimum detector for the selected trace and filter mode
Defines the selected detector type.
Note: If the EMI (R&S FPL1-K54) measurement option is installed,
additional detectors are available, even if EMI measurement is not
active. If you select a CISPR trace detector, the RBW filter type is
automatically also set to CISPR.
CISPR detectors are only available under the following conditions:
●
Time domain measurements and frequency measurements in
sweep mode (not FFT mode, not power measurements, emission
measurements, or statistics measurements)
●
Trigger mode "Free Run" or "External" (trigger offset ≥0 only for
"External")
●
Gate mode: "Off"
Hold
If activated, traces in "Min Hold", "Max Hold" and "Average" mode are not reset after
specific parameter changes have been made.
Normally, the measurement is started again after parameter changes, before the measurement results are analyzed (e.g. using a marker). In all cases that require a new
measurement after parameter changes, the trace is reset automatically to avoid false
results (e.g. with span changes). For applications that require no reset after parameter
changes, the automatic reset can be switched off.
Defines the mode with which the trace is averaged over several sweeps.
This setting is generally applicable if trace mode "Average" is selected.
For FFT sweeps, the setting also affects the VBW (regardless of whether the trace is
averaged).
(See the chapter on ACLR power measurements in the R&S FPL1000 User Manual.)
78User Manual 1178.3392.02 ─ 10
Page 79
R&S®FPL1-K7
Analysis
Trace settings
"Linear"
The power level values are converted into linear units before averaging. After the averaging, the data is converted back into its original
unit.
"Logarithmic"
For logarithmic scaling, the values are averaged in dBm. For linear
scaling, the behavior is the same as with linear averaging.
"Power"
Activates linear power averaging.
The power level values are converted into unit Watt before averaging.
After the averaging, the data is converted back into its original unit.
Use this mode to average power values in Volts or Amperes correctly.
In particular, for small VBW values (smaller than the RBW), use
power averaging mode for correct power measurements in FFT
sweep mode.
Remote command:
[SENSe:]AVERage<n>:TYPE on page 212
Average Count
Determines the number of averaging or maximum search procedures If the trace
modes "Average", "Max Hold" or "Min Hold" are set.
In continuous sweep mode, if sweep count = 0 (default), averaging is performed over
10 sweeps. For sweep count =1, no averaging, Max Hold or Min Hold operations are
performed.
Remote command:
[SENSe:]AVERage<n>:COUNt on page 212
Predefined Trace Settings - Quick Config
Commonly required trace settings have been predefined and can be applied very
quickly by selecting the appropriate button.
FunctionTrace Settings
Preset All TracesTrace 1:Clear Write
Traces 2-6:Blank
Set Trace Mode
Max | Avg | Min
Set Trace Mode
Max | ClrWrite | Min
Trace 1:Max Hold
Trace 2:Average
Trace 3:Min Hold
Traces 4-6:Blank
Trace 1:Max Hold
Trace 2:Clear Write
Trace 3:Min Hold
Traces 4-6:Blank
Trace 1/ Trace 2/ Trace 3/ Trace 4 (Softkeys)
Displays the "Traces" settings and focuses the "Mode" list for the selected trace.
79User Manual 1178.3392.02 ─ 10
Page 80
R&S®FPL1-K7
6.2Spectrogram settings
Analysis
Spectrogram settings
Remote command:
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>[:STATe] on page 210
Access: [TRACE] > "Spectrogram Config"
The individual settings available for spectrogram display are described here. For settings on color mapping, see Chapter 6.2.2, "Color map settings", on page 84.
Settings concerning the frames and how they are handled during a sweep are provided
as additional sweep settings for spectrogram display.
See Chapter 5.5, "Data acquisition", on page 48.
Search functions for spectrogram markers are described in Chapter 6.4.3, "Marker
Activates and deactivates a Spectrogram subwindow.
"On"
"Off"
Remote command:
CALCulate<n>:SPECtrogram:LAYout on page 202
3D Spectrogram State
Activates and deactivates a 3-dimensional spectrogram. As opposed to the common 2dimensional spectrogram, the power is not only indicated by a color mapping, but also
in a third dimension, the z-axis.
Displays the Spectrogram as a subwindow in the original result display.
Closes the Spectrogram subwindow.
81User Manual 1178.3392.02 ─ 10
Page 82
R&S®FPL1-K7
Analysis
Spectrogram settings
For details see the R&S FPL1000 User Manual.
Remote command:
CALCulate<n>:SPECtrogram:THReedim[:STATe] on page 203
Select Frame
Selects a specific frame, loads the corresponding trace from the memory, and displays
it in the Spectrum window.
Note that activating a marker or changing the position of the active marker automatically selects the frame that belongs to that marker.
This function is only available in single sweep mode or if the sweep is stopped, and
only if a spectrogram is selected.
The most recent frame is number 0, all previous frames have a negative number.
For more details, see the R&S FPL1000 User Manual.
Remote command:
CALCulate<n>:SPECtrogram:FRAMe:SELect on page 202
History Depth
Sets the number of frames that the R&S FPL1000 stores in its memory.
The maximum number of frames depends on the Sweep Points.
For an overview of the maximum number of frames depending on the number of
sweep points, see the R&S FPL1000 User Manual.
If the memory is full, the R&S FPL1000 deletes the oldest frames stored in the memory
and replaces them with the new data.
Remote command:
CALCulate<n>:SPECtrogram:HDEPth on page 202
3-D Display Depth
Defines the number of frames displayed in a 3-dimensional spectrogram.
For details see the R&S FPL1000 User Manual.
Trace
Selects the diagram trace on which the spectrogram is based.
Remote command:
CALCulate<n>:SGRam:TRACe on page 204
Time Stamp
Activates and deactivates the timestamp. The timestamp shows the system time while
the measurement is running. In single sweep mode or if the sweep is stopped, the
timestamp shows the time and date of the end of the sweep.
When active, the timestamp replaces the display of the frame number.
Remote command:
CALCulate<n>:SPECtrogram:TSTamp[:STATe] on page 205
CALCulate<n>:SPECtrogram:TSTamp:DATA? on page 204
82User Manual 1178.3392.02 ─ 10
Page 83
R&S®FPL1-K7
Analysis
Spectrogram settings
Color Mapping
Opens the "Color Mapping" dialog.
For details see the R&S FPL1000 User Manual.
Continuous Sweep / Run Cont
After triggering, starts the measurement and repeats it continuously until stopped.
While the measurement is running, the "Continuous Sweep" softkey and the [RUN
CONT] key are highlighted. The running measurement can be aborted by selecting the
highlighted softkey or key again. The results are not deleted until a new measurement
is started.
Note: Sequencer. If the Sequencer is active, the "Continuous Sweep" softkey only controls the sweep mode for the currently selected channel setup. However, the sweep
mode only takes effect the next time the Sequencer activates that channel setup, and
only for a channel-defined sequence. In this case, a channel setup in continuous
sweep mode is swept repeatedly.
Furthermore, the [RUN CONT] key controls the Sequencer, not individual sweeps.
[RUN CONT] starts the Sequencer in continuous mode.
For details on the Sequencer, see the R&S FPL1000 User Manual.
Remote command:
INITiate<n>:CONTinuous on page 176
Single Sweep / Run Single
After triggering, starts the number of sweeps set in "Sweep Count". The measurement
stops after the defined number of sweeps has been performed.
While the measurement is running, the "Single Sweep" softkey and the [RUN SINGLE]
key are highlighted. The running measurement can be aborted by selecting the highlighted softkey or key again.
Note: Sequencer. If the Sequencer is active, the "Single Sweep" softkey only controls
the sweep mode for the currently selected channel setup. However, the sweep mode
only takes effect the next time the Sequencer activates that channel setup, and only for
a channel-defined sequence. In this case, the Sequencer sweeps a channel setup in
single sweep mode only once.
Furthermore, the [RUN SINGLE] key controls the Sequencer, not individual sweeps.
[RUN SINGLE] starts the Sequencer in single mode.
If the Sequencer is off, only the evaluation for the currently displayed channel setup is
updated.
For details on the Sequencer, see the R&S FPL1000 User Manual.
Remote command:
INITiate<n>[:IMMediate] on page 177
Clear Spectrogram
Resets the spectrogram result display and clears the history buffer.
This function is only available if a spectrogram is selected.
Remote command:
CALCulate<n>:SPECtrogram:CLEar[:IMMediate] on page 200
In addition to the available color settings, the dialog box displays the current color map
and provides a preview of the display with the current settings.
Figure 6-1: Color Mapping dialog box
1 = Color map: shows the current color distribution
2 = Preview pane: shows a preview of the spectrogram with any changes that you make to the color
scheme
3 = Color curve pane: graphical representation of all settings available to customize the color scheme
4/5 = Color range start and stop sliders: define the range of the color map or amplitudes for the spectrogram
6 = Color curve slider: adjusts the focus of the color curve
7 = Histogram: shows the distribution of measured values
8 = Scale of the horizontal axis (value range)
Defines the lower and upper boundaries of the value range of the spectrogram.
Remote command:
DISPlay[:WINDow<n>]:SPECtrogram:COLor:LOWer on page 206
DISPlay[:WINDow<n>]:SPECtrogram:COLor:UPPer on page 206
84User Manual 1178.3392.02 ─ 10
Page 85
R&S®FPL1-K7
Analysis
Trace / data export configuration
Shape
Defines the shape and focus of the color curve for the spectrogram result display.
"-1 to <0"
"0"
">0 to 1"
Remote command:
DISPlay[:WINDow<n>]:SPECtrogram:COLor:SHAPe on page 206
Hot/Cold/Radar/Grayscale
Sets the color scheme for the spectrogram.
Remote command:
DISPlay[:WINDow<n>]:SPECtrogram:COLor[:STYLe] on page 207
Auto
Defines the color range automatically according to the existing measured values for
optimized display.
More colors are distributed among the lower values
Colors are distributed linearly among the values
More colors are distributed among the higher values
Set to Default
Sets the color mapping to the default settings.
Remote command:
DISPlay[:WINDow<n>]:SPECtrogram:COLor:DEFault on page 205
Or: [TRACE] > "Trace Config" > "Trace / Data Export"
The standard data management functions (e.g. saving or loading instrument settings)
that are available for all R&S FPL1000 applications are not described here.
See the R&S FPL1000 base unit user manual for a description of the standard functions.
85User Manual 1178.3392.02 ─ 10
Page 86
R&S®FPL1-K7
Analysis
Trace / data export configuration
Export all Traces and all Table Results.........................................................................86
Include Instrument & Measurement Settings................................................................ 86
Trace to Export..............................................................................................................86
Selects all displayed traces and result tables (e.g. "Result Summary", marker table
etc.) in the current application for export to an ASCII file.
Alternatively, you can select one specific trace only for export (see Trace to Export).
The results are output in the same order as they are displayed on the screen: window
by window, trace by trace, and table row by table row.
Remote command:
FORMat:DEXPort:TRACes on page 191
Include Instrument & Measurement Settings
Includes additional instrument and measurement settings in the header of the export
file for result data.
See the R&S FPL1000 base unit user manual for details.
Remote command:
FORMat:DEXPort:HEADer on page 191
Trace to Export
Defines an individual trace to be exported to a file.
This setting is not available if Export all Traces and all Table Results is selected.
86User Manual 1178.3392.02 ─ 10
Page 87
R&S®FPL1-K7
Analysis
Trace / data export configuration
Decimal Separator
Defines the decimal separator for floating-point numerals for the data export/import
files. Evaluation programs require different separators in different languages.
Remote command:
FORMat:DEXPort:DSEParator on page 190
Export Trace to ASCII File
Saves the selected trace or all traces in the currently active result display to the specified file and directory in the selected ASCII format.
"File Explorer": Instead of using the file manager of the R&S FPL1000 firmware, you
can also use the Microsoft Windows File Explorer to manage files.
Remote command:
MMEMory:STORe<n>:TRACe on page 192
File Type ← Export Trace to ASCII File
Determines the format of the ASCII file to be imported or exported.
Depending on the external program in which the data file was created or is evaluated,
a comma-separated list (CSV) or a plain data format (DAT) file is required.
87User Manual 1178.3392.02 ─ 10
Page 88
R&S®FPL1-K7
6.4Working with markers in the R&S FPL1-K7 AM/FM/PM
Analysis
Working with markers in the R&S FPL1-K7 AM/FM/PM Modulation Analysis application
Remote command:
FORMat:DEXPort:FORMat on page 189
Decimal Separator ← Export Trace to ASCII File
Defines the decimal separator for floating-point numerals for the data export/import
files. Evaluation programs require different separators in different languages.
Remote command:
FORMat:DEXPort:DSEParator on page 190
File Explorer ← Export Trace to ASCII File
Opens the Microsoft Windows File Explorer.
Remote command:
not supported
Modulation Analysis application
Access: "Overview" > "Analysis"
Basically, markers in the R&S FPL1-K7 AM/FM/PM Modulation Analysis application
are very similar to those in the Spectrum application. However, some additional functions are available.
In the R&S FPL1-K7 AM/FM/PM Modulation Analysis application, up to 17 markers or
delta markers can be activated for each window simultaneously.
88User Manual 1178.3392.02 ─ 10
Page 89
R&S®FPL1-K7
Analysis
Working with markers in the R&S FPL1-K7 AM/FM/PM Modulation Analysis application
The markers are distributed among 3 tabs for a better overview. By default, the first
marker is defined as a normal marker, whereas all others are defined as delta markers
with reference to the first marker. All markers are assigned to trace 1, but only the first
marker is active.
Linking to Another Marker.............................................................................................90
Assigning the Marker to a Trace................................................................................... 91
All Markers Off...............................................................................................................91
Selected Marker
Marker name. The marker which is currently selected for editing is highlighted orange.
Remote command:
Marker selected via suffix <m> in remote commands.
Marker State
Activates or deactivates the marker in the diagram.
Remote command:
CALCulate<n>:MARKer<m>[:STATe] on page 215
CALCulate<n>:DELTamarker<m>[:STATe] on page 219
89User Manual 1178.3392.02 ─ 10
Page 90
R&S®FPL1-K7
Analysis
Working with markers in the R&S FPL1-K7 AM/FM/PM Modulation Analysis application
Marker Position X-value
Defines the position (x-value) of the marker in the diagram. For normal markers, the
absolute position is indicated. For delta markers, the position relative to the reference
marker is provided.
Remote command:
CALCulate<n>:MARKer<m>:X on page 216
CALCulate<n>:DELTamarker<m>:X on page 219
Marker Type
Toggles the marker type.
The type for marker 1 is always "Normal", the type for delta marker 1 is always "Delta".
These types cannot be changed.
Note: If normal marker 1 is the active marker, switching the "Mkr Type" activates an
additional delta marker 1. For any other marker, switching the marker type does not
activate an additional marker, it only switches the type of the selected marker.
"Normal"
"Delta"
Remote command:
CALCulate<n>:MARKer<m>[:STATe] on page 215
CALCulate<n>:DELTamarker<m>[:STATe] on page 219
A normal marker indicates the absolute value at the defined position
in the diagram.
A delta marker defines the value of the marker relative to the specified reference marker (marker 1 by default).
Reference Marker
Defines a marker as the reference marker which is used to determine relative analysis
results (delta marker values).
If the reference marker is deactivated, the delta marker referring to it is also deactivated.
If a fixed reference point is configured (see "Defining a Fixed Reference"on page 92),
the reference point ("FXD") can also be selected instead of another marker.
Remote command:
CALCulate<n>:DELTamarker<m>:MREFerence on page 218
Linking to Another Marker
Links the current marker to the marker selected from the list of active markers. If the xaxis value of the initial marker is changed, the linked marker follows to the same position on the x-axis. Linking is off by default.
Using this function you can set two markers on different traces to measure the difference (e.g. between a max hold trace and a min hold trace or between a measurement
and a reference trace).
Remote command:
CALCulate<n>:MARKer<ms>:LINK:TO:MARKer<md> on page 215
CALCulate<n>:DELTamarker<ms>:LINK:TO:MARKer<md> on page 217
CALCulate<n>:DELTamarker<m>:LINK on page 217
90User Manual 1178.3392.02 ─ 10
Page 91
R&S®FPL1-K7
6.4.1.2General marker settings
Analysis
Working with markers in the R&S FPL1-K7 AM/FM/PM Modulation Analysis application
Assigning the Marker to a Trace
The "Trace" setting assigns the selected marker to an active trace. The trace determines which value the marker shows at the marker position. If the marker was previously assigned to a different trace, the marker remains on the previous frequency or
time, but indicates the value of the new trace.
If a trace is turned off, the assigned markers and marker functions are also deactivated.
Remote command:
CALCulate<n>:MARKer<m>:TRACe on page 216
All Markers Off
Deactivates all markers in one step.
Remote command:
Defining a Fixed Reference...........................................................................................92
Link Time Marker...........................................................................................................93
Link AF Spectrum Marker............................................................................................. 93
91User Manual 1178.3392.02 ─ 10
Page 92
R&S®FPL1-K7
Analysis
Working with markers in the R&S FPL1-K7 AM/FM/PM Modulation Analysis application
Marker Table Display
Defines how the marker information is displayed.
"On"
"Off"
"Auto"
Remote command:
DISPlay[:WINDow<n>]:MTABle on page 222
Marker Info
Turns the marker information displayed in the diagram on and off.
Displays the marker information in a table in a separate area beneath
the diagram.
No separate marker table is displayed.
If Marker Info is active, the marker information is displayed within the
diagram area.
(Default) If more than two markers are active, the marker table is displayed automatically.
If Marker Info is active, the marker information for up to two markers
is displayed in the diagram area.
Remote command:
DISPlay[:WINDow<n>]:MINFo[:STATe] on page 222
Marker Stepsize
Defines the size of the steps that the marker position is moved using the rotary knob.
"Standard"
"Sweep
Points"
Remote command:
CALCulate<n>:MARKer<m>:X:SSIZe on page 221
Defining a Fixed Reference
Instead of using a reference marker whose position can vary depending on the measurement results, you can define a fixed reference marker for trace analysis.
Note that this function is not available in all result displays.
The marker position is moved in steps of (Span/1000), which corresponds approximately to the number of pixels for the default display
of 1001 sweep points. This setting is most suitable to move the
marker over a larger distance.
The marker position is moved from one sweep point to the next. This
setting is required for a very precise positioning if more sweep points
are collected than the number of pixels that can be displayed on the
screen. It is the default mode.
92User Manual 1178.3392.02 ─ 10
Page 93
R&S®FPL1-K7
Analysis
Working with markers in the R&S FPL1-K7 AM/FM/PM Modulation Analysis application
For "State" = "On", a vertical and a horizontal red display line are displayed, marked as
"FXD". The normal marker 1 is activated and set to the peak value of the trace
assigned to marker 1, and a delta marker to the next peak. The fixed reference marker
is set to the position of marker 1 at the peak value. The delta marker refers to the fixed
reference marker.
The "Level" and "Frequency" or "Time" settings define the position and value of the reference marker. To move the fixed reference, move the red display lines marked "FXD"
in the diagram, or change the position settings in the "Marker Settings" tab of the
"Marker" dialog box.
Peak Search sets the fixed reference marker to the current maximum value of the
trace assigned to marker 1.
If activated, the fixed reference marker ("FXD") can also be selected as a Reference
Marker instead of another marker.
Remote command:
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed[:STATe] on page 239
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:Y on page 238
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:X on page 238
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:MAXimum[:PEAK]
on page 237
Link Time Marker
Links the markers in all time domain diagrams.
Remote command:
CALCulate<n>:MARKer<m>:LINK on page 221
Link AF Spectrum Marker
Links the markers in all AF spectrum displays.
Remote command:
CALCulate<n>:MARKer<m>:LINK on page 221
6.4.2Marker search settings and positioning functions
Several functions are available to set the marker to a specific position very quickly and
easily, or to use the current marker position to define another characteristic value. To
determine the required marker position, searches are performed. You can influence the
search results using special settings.
The remote commands required to define these settings are described in Chap-
ter 11.8.3, "Working with markers remotely", on page 214.
Markers are commonly used to determine peak values, i.e. maximum or minimum values, in the measured signal. Configuration settings allow you to influence the peak
search results.
Defines the minimum level value by which a signal must rise or fall so that it is identified as a maximum or a minimum by the search functions.
Entries from 0 dB to 80 dB are allowed; the resolution is 0.1 dB. The default setting for
the peak excursion is 6 dB.
For Analog Modulation Analysis, the unit and value range depend on the selected
result display type.
Remote command:
CALCulate<n>:MARKer<m>:PEXCursion on page 232
6.4.2.2Positioning functions
The following functions set the currently selected marker to the result of a peak search
or set other characteristic values to the current marker value.
Search Next Minimum...................................................................................................96
Select Marker
The "Select Marker" function opens a dialog box to select and activate or deactivate
one or more markers quickly.
94User Manual 1178.3392.02 ─ 10
Page 95
R&S®FPL1-K7
Analysis
Working with markers in the R&S FPL1-K7 AM/FM/PM Modulation Analysis application
Remote command:
CALCulate<n>:MARKer<m>[:STATe] on page 215
CALCulate<n>:DELTamarker<m>[:STATe] on page 219
Peak Search
Sets the selected marker/delta marker to the maximum of the trace. If no marker is
active, marker 1 is activated.
Remote command:
CALCulate<n>:MARKer<m>:MAXimum[:PEAK] on page 233
CALCulate<n>:DELTamarker<m>:MAXimum[:PEAK] on page 236
Search Next Peak
Sets the selected marker/delta marker to the next (lower) maximum of the assigned
trace. If no marker is active, marker 1 is activated.
Remote command:
CALCulate<n>:MARKer<m>:MAXimum:NEXT on page 233
CALCulate<n>:MARKer<m>:MAXimum:RIGHt on page 234
CALCulate<n>:MARKer<m>:MAXimum:LEFT on page 233
CALCulate<n>:DELTamarker<m>:MAXimum:NEXT on page 235
CALCulate<n>:DELTamarker<m>:MAXimum:RIGHt on page 236
CALCulate<n>:DELTamarker<m>:MAXimum:LEFT on page 235
Search Minimum
Sets the selected marker/delta marker to the minimum of the trace. If no marker is
active, marker 1 is activated.
Remote command:
CALCulate<n>:MARKer<m>:MINimum[:PEAK] on page 234
CALCulate<n>:DELTamarker<m>:MINimum[:PEAK] on page 237
95User Manual 1178.3392.02 ─ 10
Page 96
R&S®FPL1-K7
6.4.3Marker search settings for spectrograms
Analysis
Working with markers in the R&S FPL1-K7 AM/FM/PM Modulation Analysis application
Search Next Minimum
Sets the selected marker/delta marker to the next (higher) minimum of the selected
trace. If no marker is active, marker 1 is activated.
Remote command:
CALCulate<n>:MARKer<m>:MINimum:NEXT on page 234
CALCulate<n>:MARKer<m>:MINimum:LEFT on page 234
CALCulate<n>:MARKer<m>:MINimum:RIGHt on page 235
CALCulate<n>:DELTamarker<m>:MINimum:NEXT on page 236
CALCulate<n>:DELTamarker<m>:MINimum:LEFT on page 236
CALCulate<n>:DELTamarker<m>:MINimum:RIGHt on page 237
Spectrograms show not only the current sweep results, but also the sweep history.
Thus, when searching for peaks, you must define the search settings within a single
time frame (x-direction) and within several time frames (y-direction).
These settings are only available for spectrogram displays.
Search Mode for Next Peak in X-Direction................................................................... 97
Search Mode for Next Peak in Y-Direction....................................................................97
Working with markers in the R&S FPL1-K7 AM/FM/PM Modulation Analysis application
Search Mode for Next Peak in X-Direction
Selects the search mode for the next peak search within the currently selected frame.
"Left"
"Absolute"
"Right"
Remote command:
Chapter 11.8.3.5, "Positioning the marker", on page 233
Search Mode for Next Peak in Y-Direction
Selects the search mode for the next peak search within all frames at the current
marker position.
"Up"
"Absolute"
"Down"
Remote command:
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MAXimum:ABOVe on page 225
CALCulate<n>:DELTamarker<m>:SPECtrogram:Y:MAXimum:ABOVe
on page 230
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MAXimum:BELow on page 225
CALCulate<n>:DELTamarker<m>:SPECtrogram:Y:MAXimum:BELow
on page 230
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MAXimum:NEXT on page 226
CALCulate<n>:DELTamarker<m>:SPECtrogram:Y:MAXimum:NEXT on page 230
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MINimum:ABOVe on page 226
CALCulate<n>:DELTamarker<m>:SPECtrogram:Y:MINimum:ABOVe
on page 231
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MINimum:BELow on page 226
CALCulate<n>:DELTamarker<m>:SPECtrogram:Y:MINimum:BELow
on page 231
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MINimum:NEXT on page 227
CALCulate<n>:DELTamarker<m>:SPECtrogram:Y:MINimum:NEXT on page 231
Determines the next maximum/minimum to the left of the current
peak.
Determines the next maximum/minimum to either side of the current
peak.
Determines the next maximum/minimum to the right of the current
peak.
Determines the next maximum/minimum above the current peak (in
more recent frames).
Determines the next maximum/minimum above or below the current
peak (in all frames).
Determines the next maximum/minimum below the current peak (in
older frames).
Marker Search Type
Defines the type of search to be performed in the spectrogram.
"X-Search"
"Y-Search"
"XY-Search"
Searches only within the currently selected frame.
Searches within all frames but only at the current frequency position.
Searches in all frames at all positions.
97User Manual 1178.3392.02 ─ 10
Page 98
R&S®FPL1-K7
Analysis
Working with markers in the R&S FPL1-K7 AM/FM/PM Modulation Analysis application
Remote command:
Defined by the search function, see Chapter 11.8.3.3, "Marker search (spectrograms)",
on page 223
Marker Search Area
Defines which frames the search is performed in.
"Visible"
"Memory"
Remote command:
CALCulate<n>:MARKer<m>:SPECtrogram:SARea on page 224
CALCulate<n>:DELTamarker<m>:SPECtrogram:SARea on page 229
Peak Excursion
Defines the minimum level value by which a signal must rise or fall so that it is identified as a maximum or a minimum by the search functions.
Entries from 0 dB to 80 dB are allowed; the resolution is 0.1 dB. The default setting for
the peak excursion is 6 dB.
Only the visible frames are searched.
All frames stored in the memory are searched.
For Analog Modulation Analysis, the unit and value range depend on the selected
result display type.
●Deactivating all marker functions.......................................................................... 107
99User Manual 1178.3392.02 ─ 10
Page 100
R&S®FPL1-K7
6.4.4.1Measuring characteristic bandwidths (n db down marker)
Analysis
Working with markers in the R&S FPL1-K7 AM/FM/PM Modulation Analysis application
Access: "Overview" > "Analysis" > "Marker Functions" > "n dB down" > "n dB Down
Config"
Or: [MKR] > "Select Marker Function" > "n dB down" > "n dB Down Config"
When characterizing the shape of a signal, the bandwidth at a specified offset from its
peak level is often of interest. The offset is specified as a relative decrease in amplitude of n dB. To measure this bandwidth, you could use several markers and delta
markers and determine the bandwidth manually. However, using the n dB down marker
function makes the task very simple and quick.
The n dB down marker function uses the current value of marker 1 as the reference
point. It activates two temporary markers T1 and T2 located on the signal, whose level
is n dB below the level of the reference point. Marker T1 is placed to the left and
marker T2 to the right of the reference marker. The default setting for n is 3 dB, but it
can be changed.
If a positive offset is entered, the markers T1 and T2 are placed below the active reference point. If a negative value is entered (for example for notch filter measurements),
the markers T1 and T2 are placed above the active reference point.
Figure 6-2: n dB down marker function
The following marker function results are displayed:
Table 6-1: n dB down marker function results
LabelDescription
M1Current position and level of marker 1
ndBOffset value (n dB down)
100User Manual 1178.3392.02 ─ 10