Rohde & Schwarz EB-500 User manual

Page 1
R&S® EB500 Monitoring Receiver
Operating Manual
(XÖâJ2)
4072.8426.02 ─ 08
Operating Manual
Page 2
This manual describes the following models and options:
R&S®EB500 Monitoring Receiver 4072.5004.02 (without front control panel)
R&S®EB500 Monitoring Receiver 4072.5004.03 (with front control panel)
The firmware of the R&S EB500 makes use of several valuable open source software packages. Please refer to the "Open Source Acknowledgement" document (4072.8561.02) for a summary of the packages and the verbatim license texts. Rohde & Schwarz would like to thank the open source community for their valuable contribution to embedded computing.
© 2012 Rohde & Schwarz GmbH & Co. KG Muehldorfstr. 15, 81671 Munich, Germany Phone: +49 89 41 29 - 0 Fax: +49 89 41 29 12 164 E-mail: info@rohde-schwarz.com Internet: http://www.rohde-schwarz.com Subject to change – Data without tolerance limits is not binding. R&S® is a registered trademark of Rohde & Schwarz GmbH & Co. KG. Trade names are trademarks of the owners.
The following abbreviations are used throughout this manual: R&S®EB500 is abbreviated to R&S EB500.
Page 3
R&S® EB500
1.4.1 Electrical Specification..................................................................................................13
1.4.2 Mechanical Specification...............................................................................................14
1.4.3 Environmental Specification..........................................................................................14
1.4.4 DF Specification............................................................................................................14

Contents

Contents
1 Characteristics.......................................................................................9
1.1 Use..................................................................................................................................9
1.2 Description....................................................................................................................9
1.3 Application Examples.................................................................................................11
1.4 Specifications..............................................................................................................13
1.5 Equipment Supplied...................................................................................................16
1.6 Ordering Information..................................................................................................16
2 Preparation for Use..............................................................................17
2.1 Specific Safety Instructions.......................................................................................17
2.2 Setup............................................................................................................................18
2.2.1 Bench Operation...........................................................................................................18
2.2.2 Rack Mounting..............................................................................................................19
2.2.3 In-vehicle Mounting.......................................................................................................20
2.2.4 EMI Protective Measures..............................................................................................20
2.2.5
2.2.6 Power On and Off.........................................................................................................21
2.2.7 STANDBY and READY.................................................................................................21
2.2.8 Connecting External Accessories.................................................................................22
Connecting the R&S EB500 to the Power Supply.........................................................20
3 Operation..............................................................................................24
3.1 Front-Panel Tour.........................................................................................................24
3.1.1
3.1.2
R&S EB500 Model 03 (with front control panel)............................................................24
R&S EB500 Model 02 (without front control panel).......................................................29
3.2 Rear-Panel Tour..........................................................................................................30
3.2.1 X1 Power Supply...........................................................................................................31
3.2.2 X13 / X14 Antennas......................................................................................................31
3.2.3 X7 LAN Ethernet...........................................................................................................32
3.2.4 Miscellaneous IO...........................................................................................................32
3Operating Manual 4072.8426.02 ─ 08
Page 4
R&S® EB500
3.2.5 AUX / AUX Audio..........................................................................................................33
3.2.6 Antenna Control Interface.............................................................................................34
3.2.7 Serial Interfaces............................................................................................................34
3.3.1 GUI Layout....................................................................................................................35
3.3.2 Description of Panels....................................................................................................37
3.3.3 Using Front-Panel Keys and Dialogs............................................................................88
4.2.1 SCPI Introduction........................................................................................................192
4.2.2 Structure of a Command.............................................................................................192
Contents
3.3 Graphical User Interface (GUI)...................................................................................35
4 Remote Control via LAN Interface....................................................191
4.1 General.......................................................................................................................191
4.2 Structure and Syntax of the Device Messages......................................................192
4.2.3 Structure of a Command Line.....................................................................................195
4.2.4 Responses to Queries.................................................................................................196
4.2.5 Parameters..................................................................................................................196
4.2.6 Overview of the Syntax Elements...............................................................................199
4.3 Notation......................................................................................................................199
4.3.1 Upper and Lower-Case Notation.................................................................................199
4.3.2 Vertical stroke |............................................................................................................200
4.3.3 Square Brackets [ ]......................................................................................................200
4.3.4 Braces { }.....................................................................................................................200
4.4 Common Commands................................................................................................200
4.5 Device-Specific Commands.....................................................................................202
4.5.1 ABORt Subsystem......................................................................................................202
4.5.2 CALCulate Subsystem................................................................................................202
4.5.3 CALibration Subsystem...............................................................................................211
4.5.4 DIAGnostic Subsystem...............................................................................................212
4.5.5 DISPlay Subsystem....................................................................................................220
4.5.6 FORMat Subsystem....................................................................................................221
4.5.7 INITiate Subsystem.....................................................................................................224
4.5.8 INPut Subsystem........................................................................................................225
4.5.9 MEASure Subsystem..................................................................................................227
4.5.10 MEMory Subsystem....................................................................................................235
4Operating Manual 4072.8426.02 ─ 08
Page 5
R&S® EB500
4.5.11 OUTPut Subsystem....................................................................................................240
4.5.12 ROUTe Subsystem.....................................................................................................248
4.5.13 SENSe Subsystem......................................................................................................253
4.5.14 STATus Subsystem....................................................................................................291
4.5.15 SYSTem Subsystem...................................................................................................301
4.5.16 TEST Subsystem........................................................................................................337
4.5.17 TRACe Subsystem......................................................................................................338
4.7.1 Structure of a SCPI Status Register............................................................................350
4.7.2 Overview of the Status Registers................................................................................353
4.7.3 Description of the Status Registers.............................................................................354
4.7.4 Use of the Status Reporting System...........................................................................362
Contents
4.6 Device Model and Command Processing...............................................................346
4.7 Status Reporting System.........................................................................................350
5 Service................................................................................................365
5.1 Maintenance..............................................................................................................365
5.1.1 General.......................................................................................................................365
5.1.2 Alignment of the 10-MHz Reference Oscillator Crystal...............................................366
5.1.3 Restore Default Settings ............................................................................................366
5.1.4 Built-in Self-Tests .......................................................................................................367
5.1.5 Troubleshooting..........................................................................................................368
5.2 Repair.........................................................................................................................369
5.2.1 Procedure in Case of Service and Ordering of Spare Parts.......................................369
5.2.2
5.2.3 Removing / Installing a Module...................................................................................373
5.2.4 Disassembling the Front Panel Unit............................................................................375
5.2.5 Replacing LCD or FPC Board.....................................................................................376
5.2.6 Replacing a NAND Flash Card on Front Panel Controller (IPS1 Board)....................378
5.2.7 Replacing the Batteries...............................................................................................379
5.2.8
Disassembling the R&S EB500...................................................................................371
Assembling the R&S EB500.......................................................................................380
5.3 Retrieve Firmware Versions.....................................................................................380
5.4 Receiver Firmware and GUI Update for Model 03..................................................382
5.4.1 Preparations................................................................................................................382
5.4.2 Installation / Update Procedure...................................................................................384
5Operating Manual 4072.8426.02 ─ 08
Page 6
R&S® EB500
5.5.1 Preparations................................................................................................................389
5.5.2 Firmware Update Using the Update32 Tool................................................................390
5.6.1 Preparations................................................................................................................393
5.6.2 Installation Procedure.................................................................................................394
5.7.1 Using Update32 Tool (Model 02/03)...........................................................................398
5.7.2 Using GUI (Model 03 or Remote Access GUI)............................................................400
5.9.1 Spare Parts.................................................................................................................400
5.9.2 Available Power Cables..............................................................................................401
Contents
5.5 Receiver Firmware Update for Model 02.................................................................388
5.6 Remote Access GUI Installation..............................................................................393
5.7 Change Device IP Address.......................................................................................398
5.8 Clearing User Data....................................................................................................400
5.9 Documents.................................................................................................................400
6 Mass Data Output...............................................................................403
6.1 General.......................................................................................................................403
6.2 Data Streaming via UDP...........................................................................................404
6.3 Data Streaming via TCP............................................................................................404
6.4 EB200 Protocol..........................................................................................................405
6.5 Remote Commands..................................................................................................423
7 LAN Configuration.............................................................................429
7.1 General and Default Settings...................................................................................429
7.2 Configuration of the LAN Interface with SERIAL Interface...................................429
7.3 Configuration of the LAN Parameters with the "Update32" Tool.........................435
8 LAN Programming Examples............................................................436
8.1 Setting up a Connection...........................................................................................436
8.2 Initializing the Unit....................................................................................................437
8.3 Transmitting Device-Setting Commands................................................................437
8.4 Reading Device Settings..........................................................................................438
8.5 Processing SRQs......................................................................................................438
8.6 Program Example......................................................................................................439
8.7 Program Example UDP.............................................................................................439
9 Measurement Functions....................................................................441
6Operating Manual 4072.8426.02 ─ 08
Page 7
R&S® EB500
10.1 General Information..................................................................................................451
10.2 Frequency Mode........................................................................................................451
10.3 FFT Selectivity, FFT Bandwidth (Read-only)..........................................................453
Contents
9.1 Detectors....................................................................................................................441
9.2 CONTINUOUS Measurement Mode.........................................................................446
9.3 Measuring Time.........................................................................................................446
9.4 "PERIODIC" Measuring Mode..................................................................................448
9.5 Measuring Time Different from DEFAULT..............................................................448
9.6 Simultaneous Execution of Measuring Functions.................................................449
9.7 Availability of Measuring Functions........................................................................449
9.8 Direction-Finding Measurement..............................................................................450
10 IF Panorama........................................................................................451
10.4 IF Panorama Mode, Measurement Time, Measurement Mode..............................453
10.5 Clear Panorama.........................................................................................................455
10.6 Demodulation Frequency.........................................................................................456
10.7 Peak Search, Squelch On/Off, Squelch Threshold................................................456
11 Direction Finding................................................................................457
11.1 Component Overview...............................................................................................457
11.2 Rack-Mounting, Grounding and Power Supply......................................................459
11.3 Antennas....................................................................................................................460
11.3.1 Operating Principle......................................................................................................460
11.3.2 Choosing the Antenna Site.........................................................................................460
11.3.3 Antenna Installation and Accessories.........................................................................460
11.3.4 Example of a Stationary Installation............................................................................462
11.3.5 DF Vehicles.................................................................................................................463
11.4 Electronic Compass..................................................................................................465
11.4.1 Use..............................................................................................................................465
11.4.2 Installation...................................................................................................................465
11.4.3 Calibration...................................................................................................................465
11.5 Cable Sets DDF 1C-x.................................................................................................468
11.5.1 Cable Coding..............................................................................................................468
11.5.2 Circuit Diagrams, Cable Numbers...............................................................................469
11.6 North Adjustment......................................................................................................471
7Operating Manual 4072.8426.02 ─ 08
Page 8
R&S® EB500
11.7 Installation Concepts................................................................................................472
11.8 Operation...................................................................................................................474
11.8.1 General Instructions....................................................................................................474
11.8.2 Most Frequently Used Settings...................................................................................475
11.8.3 Operation in a Moving Vehicle....................................................................................476
11.9 Corrrection Data........................................................................................................478
11.9.1 Antenna Factors..........................................................................................................478
11.9.2 Cable Attenuation........................................................................................................478
11.9.3 Azimuth Correction......................................................................................................478
11.9.4 Omniphase Correction................................................................................................478
12.1 General Information..................................................................................................479
Contents
12 Polychrome IF Panorama..................................................................479
12.2 Squelch On/Off, Squelch Threshold........................................................................481
12.3 Polychrome IF Panorama Mode...............................................................................481
12.4 100% Time, Observation Time.................................................................................481
12.5 Lower Limit, Upper Limit..........................................................................................482
12.6 Polychrome IF Panorama Clear...............................................................................482
13 Digital Down Converters...................................................................483
13.1 Introduction...............................................................................................................483
13.2 Device-Specific Commands for DDCs....................................................................483
13.2.1 CALCulate Subsystem................................................................................................483
13.2.2 OUTPut Subsystem....................................................................................................484
13.2.3 SENSe Subsystem......................................................................................................486
13.2.4 SYSTem Subsystem...................................................................................................491
13.2.5 TRACe Subsystem......................................................................................................496
List of Commands..............................................................................501
Index....................................................................................................512
8Operating Manual 4072.8426.02 ─ 08
Page 9
R&S® EB500
Characteristics
Use

1 Characteristics

1.1 Use

The R&S EB500 has been designed for tasks such as signal searches, radio monitoring or spectrum monitoring in line with ITU recommendations. It is suitable for both mobile and stationary use. The R&S EB500 is ready to meet future requirements in signal receiver and analysis technology. With the available options, the R&S EB500 can be optimally adapted to the tasks to be performed. There is sufficient performance capability in signal processing which will allow the device to be expanded for future tasks. The R&S EB500 can be remote-controlled via its LAN interface. Since the protocol on the remote-control interface is compatible with other products from Rohde & Schwarz, the R&S EB500 can easily be integrated into any new or existing system.

1.2 Description

The basic version of the R&S EB500 covers the wide frequency range from 20 MHz to
3.6 GHz. An optional plug-in HF Preselector module makes it possible to expand the frequency range down to 9 kHz and an optional software option can expand the frequency range up to 6 GHz. The R&S EB500 is outstanding, particularly due to its wide frequency range, excellent receive characteristics, a real-time bandwidth of 20 MHz, and a wealth of functions, all accessible through its intuitive front panel and 5.7” color display.
Although operation has been optimized for monitoring tasks, its adaptable functions allow the R&S EB500 to be employed in other areas as well, especially where real time signal analysis is required. Because the R&S EB500 is equipped with extensive preselection functions, the signal sum load is reduced in scenarios that involve a large number of powerful transmitters. All the functions of the R&S EB500 can be remote-controlled via the LAN interface. All measurement results, including the spectrum data, are output via this interface.
The R&S EB500 lends itself particularly well to the following tasks (see also chapter 1.3,
"Application Examples", on page 11):
Quick overview of the activities in a user-definable frequency range
Monitoring of predefined frequencies
Storage of up to 10 000 frequencies
Searches in the frequency range with user-definable start and stop frequency and
fixed step width Searches for and analysis of interfering signals
Detection of illegal transmitters
Monitoring of radio and navigation services that are relevant to safety
Monitoring of self-conducted radiocommunications training exercises in a specific
frequency band
9Operating Manual 4072.8426.02 ─ 08
Page 10
R&S® EB500
Characteristics
Description
Monitoring of selected emissions
Radiomonitoring for tactical and strategic intelligence
Radiomonitoring for internal and external security
Monitoring in line with ITU recommendations
Coverage measurements and monitoring of networks
Storage of spectra or digital baseband data, internal and external
Analysis of digital signals
Frontend
The frontend modules of the R&S EB500 consist of an RF converter module, a synthe­sizer module and an optional HF preselection module. There are two antenna inputs, HF/ V/UHF and HF are fed into the RF converter module.
With the HF option installed, signals from 9 kHz to 32 MHz can be obtained (which covers the full HF range from 3 MHz to 30 MHz). The desired antenna input (HF/V/UHF or HF) is selected and routed to the HF preselection module for preselection. The HF preselec­ted signal is then routed to the HF A/D Converter for further processing in the digital domain. The upper limit for the HF preselection module can be set in the antenna con­figuration menu.
Otherwise, for V/UHF signals above 20 MHz, the input is only obtained from the HF/V/ UHF antenna input. This signal is then preselected and down-converted into the IF signal which is fed into the IF A/D Converter for further processing in the digital domain.
Digital signal processing
All signal processing is performed by means of powerful signal processors and field pro­grammable gate arrays (FPGAs).
The incoming IF signal, which is approximately 57.4 MHz, or direct HF signal is fed to the respective 16 bit A/D converter and first converted to digital samples with a sampling rate of 76.8 MHz. These samples are then processed in two separate paths, namely, the IF Panorama path and the Demodulation path. The R&S EB500 features a real time band­width of 20 MHz. The IF Panorama path starts with a Digital Down Conversion block, digital low-pass filter and FFT block. The bandwidth of the spectrum can be configured by the user in the range of 1 kHz to 20 MHz. Before the IF spectrum is output on the display or via the LAN interface, results are post-processed by means of the AVERAGE, CLEARWRITE, MINHOLD or MAXHOLD function as selected by the user.
The Demodulation path starts with a second Digital Down Conversion block and low-pass filter, which are used to prepare the signal for level measurement, ITU measurement and demodulation. To process the different signals with the optimum signal-to-noise ratio, the receiver contains 29 IF filters with bandwidths from 100 Hz to 5 MHz, which can be selected independently of the IF panorama bandwidth. Prior to the level measurement, the magnitude of the level is determined and weighted by means of the AVERAGE, PEAK, RMS or FAST function, as selected by the user. Next, the measured level is output via the display or the LAN interface. For the demodulation of analog signals, the complex baseband data after the low-pass filter is fed to the AM, FM, USB, LSB, ISB, pulse or CW demodulation stage and then subjected to the automatic gain control (AGC) or manual gain control (MGC). The complex baseband data (I/Q data) representing the digitized
10Operating Manual 4072.8426.02 ─ 08
Page 11
R&S® EB500
Characteristics
Application Examples
signal or video data is output directly following the AGC/MGC block via the LAN interface. The results that are obtained are available in digital format and can be output as required via the LAN interface. The analog video/IF output is fed through the D/A converter and then output to the Video A/B output port. The digital audio data are output via the LAN and also reconverted to analog signals for output via the loudspeaker.

1.3 Application Examples

Quick search in the frequency range
The R&S EB500 "Panorama Scan" function is used to perform an ultra-quick scan of a user-definable frequency range. It thus provides a quick overview of the spectrum occu­pancy. Changes caused by illegal radio services, interferences, transient emissions etc can immediately be seen at a glance. The R&S EB500 can be switched to "listen" mode simply by pressing a key. The signal of interest can be selected, demodulated, and ana­lyzed by using the demodulation frequency. The resolution of the FFT calculation can be set according to the channel spacing of various radio services. Thus, this FFT scan pro­vides highest scan rates even with small resolution bandwidth, a big advantage when weak signals must be detected quickly.
Search using channel spacing
The "Frequency Scan" function is used to perform searches in the frequency range using fixed channel spacing. The receiver scans the frequency range and analyzes every channel for occupancy. If a signal is above the threshold, the R&S EB500 dwells at the signal's frequency for the predefined hold time, and the signal can be demodulated and processed. In the case of analog methods, the demodulated signal is audible on the loudspeaker or headphones.
The "Frequency Scan" function is specially designed for the monitoring of radio services that employ fixed frequency spacing (channel spacing).
Searches at specific frequencies
The "Memory Scan" function is used to scan a series of predefined channels and analyze them for occupancy. If a signal is above the threshold, the R&S EB500 dwells at the signal's frequency for the predefined hold time, and the signal can be demodulated and processed. In the case of analog methods, the demodulated signal is audible on the loudspeaker or headphones.
In this mode, parameters such as frequency, demodulation mode, bandwidth or squelch levels can be set individually for each channel. The content of each memory channel can be transferred to the receiver manually by using the MEM > "RCL" keys. The settings can also be changed and then stored in memory by means of the MEM > "SAVE" keys. The parameters for multiple channels can be set by calling up a display table.
This scan is especially useful when it is necessary to scan individual frequencies that do not have fixed channel spacing, or when frequency blocks need to be scanned at regular intervals (e.g. f1, f2, f3, f1, f2, f3, etc).
11Operating Manual 4072.8426.02 ─ 08
Page 12
R&S® EB500
Characteristics
Application Examples
Demodulation and analysis
Analog modulated signals are demodulated in the R&S EB500. The signal is audible on the built-in loudspeaker or via headphones. A PC-based software application is available for analyzing complex signals. This application provides a large number of functions such as classification (detection of modulation mode), vector analysis, bit-stream analysis or decoding.
For online analysis, the digital data is transferred from the receiver to the PC via LAN in real time. For offline analysis, the software accesses data that was recorded in the R&S EB500 itself or externally.
Interference search
Owing to its compact design and many special functions, the R&S EB500 is ideally suited for detecting all kinds of radio interference. Special functions such as adjustable mea­surement time and continuous (averaged) or periodic (maximum value after timeout of the measurement) level output have been integrated for these tasks. Since these func­tions are also applied in the RF spectrum, even non-periodic interferers can be easily detected, which are otherwise easy to miss in a quick spectrum because they occur at erratic intervals. As a result, the source of interference can be quickly detected and elim­inated – an aspect that is extremely important in security-critical radio communications (e.g. air-traffic communication).
Detection of frequency-agile signals, e.g. hoppers
Due to its wide real time bandwidth and high scan rates, the R&S EB500 can also reliably detect signals with quickly varying frequencies – even hoppers with hop rates up to 2000 hops/s. If the "MAXHOLD" function is selected, the occupied frequency range of the transmitter will be displayed very fast.
Detection of burst signals, radar emissions
Due to the wide real-time bandwidth of 20 MHz, even very narrow bursts such as emitted by radar equipment can be detected and analyzed. By means of the IF panorama, parameters such as pulse duration can be measured.
Data recording
The R&S EB500 provides the following capabilities for recording and storing data (only model 03 - front-panel control version):
3.8 gigabytes on internal flash memory
External storage via USB 2.0 The digital data can also be output via the LAN interface
in realtime, externally monitored on a PC and externally recorded (e.g. to a PC hard disk). The recording time primarily depends on the capacity of the storage medium. Recording can be done with various formats.
Civil monitoring in compliance with ITU guidelines
Owing to its performance, the R&S EB500 meets all requirements for measurements in line with the ITU-R recommendations and the ITU Spectrum Monitoring Handbook
12Operating Manual 4072.8426.02 ─ 08
Page 13
R&S® EB500
Characteristics
Specifications
(2002). If the R&S EB500-IM option is installed, the following measurements can be per- formed:
Frequency and offset in line with ITU-R SM.377
Field strength in line with ITU-R SM.378
Modulation in line with ITU-R SM.328
Spectrum occupancy on the control PC in line with ITU-R SM.182/SM.328
Bandwidth in line with ITU-R SM.443
Detection of mono and stereo transmissions in the case of FM broadcasting trans-
mitters
Radio data system (RDS) analysis
The signal content is demodulated and decoded. The results are output on the display or external PC and are also made available via LAN. Messages such as station name, frequency lists, traffic information etc, can be seen at a glance.
Selective call/Pager analysis
The R&S EB500-SL option can be used to decode selective-call methods and to demod- ulate pagers. The following selective-call methods are supported: CCIR1, CCIR7, CCITT, EEA, EIA, EURO, DCS, DTMF, CTCSS, NATEL, VDEW, ZVEI1, ZVEI2. Other methods are available on request. The results are output on the display or on the user interface of the PC.
Direction finding of emissions
In addition to detection, the direction from which the signal originates is also an important criterion in radiomonitoring. When the DF upgrade R&S EB500-DF is added, the R&S EB500 can be used as a single-channel direction finder. If two or more stations are networked, the emitter location can also be determined. The DF upgrade consists of an installation kit for the receiver. The corresponding direction-finding antenna must be selected separately.

1.4 Specifications

1.4.1 Electrical Specification

The R&S EB500 requires DC input power: DC supply should be in the range 10 - 32 V / 12 A – 4A. AC supply should be in the range 100V -240 V, 50-60Hz via an external AC/ DC adaptor.
13Operating Manual 4072.8426.02 ─ 08
Page 14
R&S® EB500
Characteristics
Specifications

1.4.2 Mechanical Specification

1.4.3 Environmental Specification

Dimensions (width x height x depth) 213 mm x 132 mm x 450 mm (w/o handles and feet)
Weight approximately 7.5 kg
The R&S EB500 without display is classified as Class C and R&S EB500 with display as Class C2.
Environmental conditions acc. to EN 60068-2-1, 60068-2-2
Nominal temperature range
Operating temperature range 0 °C to 50 °C
Storage temperature range
Humidity 25 °C / 55 °C, 95 % relative humidity (model 02), non-
½ 19” x 3 HU
-10 °C to 55 °C for R&S EB500 (model 02)
0 °C to 55 °C for R&S EB500 (model 03)
-40 °C to 70 °C for R&S EB500 (model 02)
-30 °C to 70 °C for R&S EB500 (model 03)
-20 °C to 70 °C with AC-DC adaptor
condensing
25 °C / 40 °C, 80 % relative humidity (model 03), non­condensing
Shock acc. to EN 60068-2-27
Vibration (sine) acc. to EN 60068-2-6 (5 Hz to 150 Hz)
Vibration (random) acc. to EN60068-2-64 (10 Hz to 500 Hz)
Electromagnetic compatibility (EMC) acc. to EN 300339, ETSI EN301489-1, ETSI
MTBF

1.4.4 DF Specification

DF upgrade
Direction finding functions are only accessible with the installed DF upgrade R&S EB500­DF (Direction Finding).
MIL-STD 810E, Method 516.4, Procedure I
EN301489-22
>= 35 000 h for R&S EB500 (model 02)
>= 17 000 h for R&S EB500 (model 03)
14Operating Manual 4072.8426.02 ─ 08
Page 15
R&S® EB500
Characteristics
Specifications
WDF option
Wideband Direction finding functions are only accessible with the installed option R&S EB500-WDF.
In some literature, the term WDF is referred to as WFFM (Wideband Fixed Frequency Mode).
DF method
Instrument DF accuracy Throughout the entire frequency
System DF accuracy In non-reflecting test environment,
Reflection resistance
DF sensitivity Depending on the DF antenna
Polarization
Realtime bandwidth for wideband direction finding
(Requires the R&S EB500-WDF option)
Channel spacing for wideband direction finding
In the VHF/UHF/SHF ranges
In the HF range
range
depending on the DF antenna
With R&S©ADD119 (300 kHz to 30 MHz) 20 MHz to 80 MHz
80 MHz to 1.3 GHz
DF error at 50 % reflections, for receive frequencies > 30 MHz
With R&S©ADD197
With R&S©
ADD119/196/295/071/075
Correlative interferometer
Wattson-Watt
0.5° RMS
≤ 2° RMS
≤ 2° RMS, 1° RMS typ.
≤ 1° RMS, 0.5° RMS
1.7° RMS
Vertical, horizontal and circu-
lar Vertical
Up to 20 MHz
2 MHz, 1 MHz, 500 kHz, 200 kHz, 100 kHz, 50 kHz, 25 kHz, 20 kHz,
12.5 kHz, 10 kHz, 8.33 kHz, 5 kHz, 2 kHz, 1 kHz, 500 Hz, 200 Hz
Minimum signal duration For a single burst signal 1 ms
Minimum burst duration For multiple burst signals 0.5 ms
Display resolution Selectable 0.1° or 1°
Operating modes
DF bandwidths
DF-capable modulation modes
FFM (Fixed Frequency Mode), FSCAN (Frequency Scan), MSCAN (Memory Scan)
1200 kHz, 480 kHz, 240 kHz, 120 kHz, 60 kHz, 30kHz, 15 kHz, 12 kHz, 7.5 kHz, 6 kHz, 5 kHz, 3.75 kHz 3 kHz, 1.875 kHz, 1.5 kHz, 1.2 kHz, 750 Hz, 600 Hz, 375 Hz, 300 Hz, 187 Hz, 150 Hz, 120 Hz, 75 Hz, 60 Hz, 37 Hz, 30 Hz, 18 Hz, 15 Hz, 7 Hz, 3 Hz, 1Hz, 0.75 Hz, 0.375 Hz
CW, AM, FM, SSB, FSK, PSK
15Operating Manual 4072.8426.02 ─ 08
Page 16
R&S® EB500
Characteristics
Equipment Supplied

1.5 Equipment Supplied

1.6 Ordering Information

Designation Order number
CD set R&S EB500
R&S EB500 Software and Documentation CD ROM
Option CD ROM (only where SW option is ordered) 4072.8610.00
R&S EB500 Getting Started (English)
1x Ethernet patch cable 2m 4055.6458.00
Power cable country-specific
Designation Note Type Order number
Monitoring Receiver with­out front control panel
Monitoring Receiver with front control panel
Model 02
Model 03
4072.8510.00
4072.8532.00
4072.8432.00
R&S EB500
R&S EB500
4072.5004.02
4072.5004.03
HF Frequency Range Extension (9 kHz to 32 MHz)
SHF Frequency Range Extension (3.6 GHz to 6 GHz)
Panorama Scan (RF Spectrum)
ITU Measurement Soft­ware
DF Upgrade (Direction Finding)
Digital Down Converter Option Key
DF Error Correction
(Prerequisite is DF upgrade available)
Wideband Direction Finder
(Prerequisite is DF upgrade available)
Selective Call / Pager Decoder
HW-Module
Option Key
Option Key
Option Key
Upgrade
Option Key
Option Key
Option Key
R&S EB500-HF
R&S EB500-FE
R&S EB500-PS
R&S EB500-IM
R&S EB500-DF
R&S EB500-DDC
R&S EB500-COR
R&S EB500-WDF
R&S EB500-SL
4072.8003.02
4072.9300.02
4072.9200.02
4072.9100.02
4072.9400.02
4072.9500.02
4072.9600.02
4072.9651.02
4072.9800.02
16Operating Manual 4072.8426.02 ─ 08
Page 17
R&S® EB500
Preparation for Use
Specific Safety Instructions

2 Preparation for Use

This section describes the basic steps to be taken when setting up the R&S EB500 for the first time.

2.1 Specific Safety Instructions

General safety instructions
Please make sure you observe the basic safety instructions included in this documenta­tion as well as the instructions for setup and connection to prevent personal injury or damage to the R&S EB500. This is of particular importance when you use the R&S EB500 for the first time.
The following safety instructions apply in particular:
IEC 364
VDE 0100
DIN 57100
These safety regulations deal with the following aspects:
Prevention of accidents
Protection against overvoltage
Insulation of equipment
Grounding
Characteristics and laying of lines and cables
Provisions for operational facilities and rooms and systems of a special nature
Setup
Before turning on the R&S EB500, please make sure that the following conditions are fulfilled:
Covers are in place and all fasteners are tightened.
Fan openings are unobstructed.
Signal levels at the input connectors are all within the specified ranges.
Signal outputs are correctly connected and not overloaded.
The R&S EB500 is dry and shows no condensation.
Non-observance may cause damage to the R&S EB500 or other devices in the setup.
17Operating Manual 4072.8426.02 ─ 08
Page 18
R&S® EB500
Preparation for Use
Setup
The R&S EB500 is supplied completely assembled except for the handles and mounting brackets, which must be attached by the user.

2.2 Setup

2.2.1 Bench Operation

Equipment cooling
Do not expose the R&S EB500 to humidity. Leave at least 50 mm of empty space along both side panels in order to ensure proper equipment cooling.
There are no special requirements for desktop use. To facilitate access to the front panel elements, you should raise the front of the R&S EB500 by folding out its standing feet.
18Operating Manual 4072.8426.02 ─ 08
Page 19
R&S® EB500
Preparation for Use
Setup
Risk of injury
The feet may fold in if they are not folded out completely or if the R&S EB500 is shifted. The feet may break if they are overloaded. Fold the feet completely in or completely out to ensure stability of the R&S EB500 and personal safety. To avoid injuries, never shift the R&S EB500 when its feet are folded out. The overall load (the device's own weight plus that of any devices stacked on top of it) on the folded-out feet must not exceed 500 N. Place the R&S EB500 on a stable surface. Secure any devices stacked on top of it against slipping (e.g. by locking their feet on the top front frame). When the R&S EB500 is standing on its folded-out feet, do not work under it and do not put anything under it as this would pose a risk of personal injury or material damage.
The R&S EB500 can be used in any of the positions shown here.

2.2.2 Rack Mounting

Ambient temperature
The R&S EB500 should be used in an area where the ambient temperature does not exceed –10 °C to +55 °C (model 02) or 0 °C to 55 °C (model 03), respectively. The R&S EB500 is fan-cooled and must be installed with sufficient space along the sides to ensure a free flow of air. Make sure that there is sufficient space for hot air to escape from the R&S EB500. To ensure sufficient cooling do not attach telescopic rails to the sides of the unit.
19Operating Manual 4072.8426.02 ─ 08
Page 20
R&S® EB500
Preparation for Use
Setup

2.2.3 In-vehicle Mounting

For use in vehicles, the receiver can be powered directly from the vehicle's battery via the DC input.

2.2.4 EMI Protective Measures

In order to avoid electromagnetic interference (EMI), the R&S EB500 may only be oper­ated when it is closed and all shielding covers are in place. Use only appropriate shielded signal and control cables with proper termination.
2.2.5

Connecting the R&S EB500 to the Power Supply

Connect the R&S EB500 observing the following sections and instructions for use. The R&S EB500 is suitable only for DC operation.
2.2.5.1 Connecting to the Power Adapter
The R&S EB500 is connected to the AC supply 100V to 240V via the supplied AC/DC adapter (R&S P/N: 3586.4426.00) and the socket X1 (10V to 32VDC) on the rear panel. The connection to X1 is via a Neutrik Speakon NL4FX connector (shown in figure 2-1 below).
Installing the connector:
1. Insert the Speakon® NL4FX connector into socket X1 on the rear panel.
2. Turn the connector clockwise until it is locked in place and secured by the safety latch.
Removing the connector:
1. Press and chuck back the safety latch of the Speakon® NL4FX connector.
2. Turn the connector counterclockwise and withdraw it.
Fig. 2-1: Speakon® NL4FX
20Operating Manual 4072.8426.02 ─ 08
Page 21
R&S® EB500
Preparation for Use
Setup
2.2.5.2 Connecting to the DC Source
The R&S EB500 is connected to an external 10 VDC to 32 VDC source (e.g. battery) via the supplied DC cable (R&S P/N: 4072.7036.00) to the connector X1 (10V to 32VDC) on the rear panel.
DC supply voltage
Make sure that the available supply voltage is between 10 V and 32 V. Observe correct voltage polarity when connecting. Incorrect polarity may blow the fuse on the DC con­verter inside the R&S EB500 or damage the R&S EB500.
The DC connector is intended for disconnection of the receiver from DC-Source. Preferable use the R&S delivered Power Supply TRG150A240. If an other power supply is used, please note: This power supply shall be in accordance
with IEC / EN / UL / CSA 60950-1 or IEC / EN / UL / CSA 61010-1

2.2.6 Power On and Off

The DC power connection X1 is located at the top left corner of the R&S EB500 rear panel. With the DC power connected, the R&S EB500 is in STANDBY or READY state, depending on the state of the STANDBY toggle switch at the front panel of the R&S EB500 when it was last switched off. The standby power is below 0.8W.

2.2.7 STANDBY and READY

The STANDBY key is located at the bottom left corner of the front panel. With the DC power connected, press the STANDBY key briefly to switch the R&S EB500 from the STANDBY to the READY state and vice versa. In STANDBY state, the amber LED on the right will turn on and only the power switch circuit is being powered. In this state, it is safe to remove the DC power and disconnect the R&S EB500 from the AC/DC power adapter. In READY state, the green LED on the left is on. The R&S EB500 is ready for operation. All modules are being powered and the receiver initiates its startup procedure.
Model 02
In the case of R&S EB500 Model 02 (without front control panel), the green Power LED doubles as a 'Fail' status LED. In conditions of failure, the LED will turn red instead of green, as shown in the figures below.
21Operating Manual 4072.8426.02 ─ 08
Page 22
R&S® EB500
Preparation for Use
Setup
R&S EB500 Model 02 Ready R&S EB500 Model 02 in Failure mode. Refer to chap-
ter 5.1.5, "Troubleshooting", on page 368 for possible
actions to take.

2.2.8 Connecting External Accessories

For connecting an external mouse or keyboard a USB hub is required.
2.2.8.1 Connecting a USB Hub
You can connect a USB hub to the USB port on the front panel of the R&S EB500. The USB hub is detected automatically when connected. It is safe to connect and disconnect the hub during the measurement.
2.2.8.2 Connecting a Mouse
You can connect a mouse to the USB port on the front panel of the R&S EB500. The mouse is detected automatically when connected. It is safe to connect and disconnect the mouse during the measurement.
2.2.8.3 Connecting a Keyboard
You can connect a keyboard to the USB port on the front panel of the R&S EB500. The keyboard is detected automatically when connected. It is safe to connect and disconnect the keyboard during the measurement.
2.2.8.4 Connecting a LAN Cable
You can connect a LAN cable to the LAN port (X7) on the rear panel of the R&S EB500.
To establish a LAN connection, proceed as described below.
1.
Refer to chapter 5.7, "Change Device IP Address", on page 398 to learn how to set the receiver’s IP address.
2.
Connect a LAN cable to the LAN port. The R&S EB500 has an internal switch which automatically detects the type of LAN cable connected so you can use any standard
22Operating Manual 4072.8426.02 ─ 08
Page 23
R&S® EB500
Preparation for Use
Setup
type of LAN cable to establish a network connection with the R&S EB500 (dedicated or non-dedicated).
Dedicated vs. non-dedicated network connections
There are two methods to establish a LAN connection with the R&S EB500:
A non-dedicated network (Ethernet) connection from the R&S EB500 to an existing
network. The R&S EB500 is assigned an IP address and can coexist with a computer and with other hosts on the same network.
A dedicated network connection between the R&S EB500 and a single computer. The computer must be equipped with a network adapter and is directly connected to the R&S EB500. The use of hubs, switches or gateways is not required; however, data transfer is still made using the TCP/IP protocol.
Please refer to "Network Configuration Dialog" on page 134 to learn how to retrieve the IP address.
23Operating Manual 4072.8426.02 ─ 08
Page 24
R&S® EB500
Operation
Front-Panel Tour

3 Operation

This chapter contains a detailed description of how to operate the R&S EB500. Following an overview of the device's front and rear panel, respectively, its graphical user interface (GUI) is explained in detail. This chapter contains all the information required to access the various configuration options and settings necessary to operate the R&S EB500.

3.1 Front-Panel Tour

3.1.1

R&S EB500 Model 03 (with front control panel)

This chapter provides an overview of the front-panel elements of the R&S EB500 with front control panel.
The front panel of the R&S EB500 with display consists of the VGA display with the softkey area (bottom side), utility keys (left side), and the hardkey area with the connec­tors and control interfaces (right side). The individual controls are described in the fol­lowing sections.
24Operating Manual 4072.8426.02 ─ 08
Page 25
R&S® EB500
Operation
Front-Panel Tour
3.1.1.1 Display
The R&S EB500 is equipped with a 5.7" color display (640x480 pixels) providing control elements for the measurements and various output elements for the results.
3.1.1.2 Softkeys
"Softkeys" are located below the display. The actual function of a softkey is indicated on the label just above the softkey. Their purpose is to provide quick, direct access to the main settings in the active (highlighted) panel on the display. On the other hand they can provide selection options for the utility keys at the left: when either one of the utility keys APPL, MODE, SETUP, MENU, MEM or HELP is pressed the softkeys will provide the operator with a number of options to select. There are a total of six softkeys and in most cases there are more options than softkeys: in those cases the rightmost softkey "More" will serve the purpose of browsing to the next set of softkeys. The next set again might have a "More" key.
3.1.1.3 Front-Panel Controls and Indicators
STANDBY Key
The STANDBY key toggles between STANDBY and READY state and shuts down the R&S EB500. See chapter 2.2.7, "STANDBY and READY", on page 21 for an explanation of the LED colors related to the state of the device.
25Operating Manual 4072.8426.02 ─ 08
Page 26
R&S® EB500
Operation
Front-Panel Tour
Utility Keys
At the left side of the display we find the utility keys: these are keys that provide the operator with menus and dialogs to configure settings, and shortcuts to change the R&S EB500 into different receiver application or operation modes. There are also "Help", "Escape" and "Panel focus" keys.
ESC terminates an ongoing operation, e.g. by closing a dialog.
APPL sets the R&S EB500 to a particular application mode: RECEIVER, VIDEO, ITU, DF (optional). See chapter 3.3.3.5, "APPL", on page 89.
The MODE key is required in case the operator wants to change the receiver from a Fixed Frequency Mode to one of the scanning modes. It also provides fast access to scanning­related control functions such as "Run+", "Run-" and "Stop":
"Run-" runs the scan in negative direction.
"Stop" stops the scan.
"Run+" runs the scan in positive direction.
"Restart" will start the scan at the configured "START" frequency and run in positive
direction (only for FScan and MScan)
SETUP provides access to the setup and configuration dialogs (see chapter 3.3.3.10,
"SETUP", on page 128).
The MENU key gives the operator fast access to a number of operational functions such as "Record", "Zoom" or "Lock.
The MEM key is basically a shortcut to a number of memory-related dialogs.
HELP starts the online help system (see chapter 3.3.3.9, "HELP", on page 123).
The PANEL key provides the operator with the option to set the focus on one of the panels on the display. A panel "in focus" will be open for active control; i.e. the softkeys will "act" upon the panel in focus.
Receiver Control Keys
At the right side of the display you will find a row of keys that directly change receiver parameters. At the lower half you will find the four keys BW, AFC, DET and DEMOD.
BW selects the demodulation bandwidth (between 100 Hz and 20 MHz).
AFC switches automatic frequency control (AFC) ON or OFF.
DET selects the level measurement mode (AVERAGE, PEAK, FAST or RMS).
DEMOD selects the demodulation mode (FM, AM, PULS, PM, IQ, ISB, CW, USB, LSB).
At the upper half you will find another set of receiver control keys. These are ATT (Attenu­ation), MGC (Gain Control), and SQU (Squelch). Other keys at the right side are MARK (which is a shortcut to set and control markers in the spectrum), the SELECT rollkey (no label) and the rollkey for audio volume control.
SELECT rollkey and controls
26Operating Manual 4072.8426.02 ─ 08
Page 27
R&S® EB500
Operation
Front-Panel Tour
After pressing either one of the keys MARK, ATT, MGC or SQU, the SELECT rollkey will act on behalf of the respective Marker, Attenuation, (Manual) Gain Control or Squelch function of the receiver.
Pressing ATT will set the SELECT key to control automatic or manual attenuation. When pressing the SELECT rollkey, manual attenuation is selected. After that, by turning the rollkey the desired attenuation (between 0 and 40 dB for HF/V/UHF) can be set.
Pressing MGC will set the SELECT key to work on behalf of automatic or manual gain control (AGC and MGC, respectively). When manual gain control is set, turning the rollkey sets the desired gain control.
Pressing SQU will set the SELECT key to work on behalf of squelch. Pressing the rollkey will set squelch ON or OFF. If it is ON, then turning the rollkey sets the level threshold.
When MARK is pressed, a number of "Marker" related softkeys will appear and the SELECT rollkey can be used for operating upon the different X and Y markers in the panels. The softkeys provide options to choose sets of X-markers (X1/X2 and X3/X4) or Y-markers (Y1/Y2) as well as softkeys to display measurement values and to clear one set of markers or all markers: MEAS VAL, CLEAR, CLEAR ALL, FREQ VAL. See chap-
ter 3.3.3.4, "MARK", on page 89.
Pressing the "X1/X2" softkey will activate a set of X markers with the X1 (left) marker active, and pressing again will make the X2 (right) marker active. The active marker can be moved by turning the SELECT rollkey. Similar behavior can be achieved with the X3/ X4 and Y1/Y2 markers. Pressing the rollkey will hide the activated markers and pressing again will display the markers at their last positions. Note that only one marker pair can be activated at a time and for this active pair only one marker (the green marker) is active for control by the rollkey.
Navigation Keys
At the right side of the receiver control keys you will find a large number of keys which can be categorized as multipurpose: the navigation keys at the lower half and the data­entry keys (alphanumeric keys) at the upper half.
The navigation keys consist of one rollkey: the "Main Rollkey" or ROTARY KNOB and two arrow keys: CURSOR LEFT and CURSOR RIGHT.
The CURSOR LEFT / CURSOR RIGHT keys move the cursor in input fields and scroll within lists, dialogs or tables.
The main function of the ROTARY KNOB is to control the receiver frequency. In addition, it provides access to the following functions:
Increasing or decreasing numeric values in editing mode
Scrolling within lists or tables
Confirming entries (pressing the ROTARY KNOB is equivalent to pressing ENTER)
27Operating Manual 4072.8426.02 ─ 08
Page 28
R&S® EB500
Operation
Front-Panel Tour
Data Entry Functions
The keys in the data entry keypad are used to enter numbers and letters. Data entry keys are only enabled if an editable field has been selected for entry mode. This can be ach­ieved in the following manner:
Key-in a number when no dialog is active: The key will be interpreted as a frequency
value. Key-in a number in a highlighted softkey: The parameter of the softkey will take the
value of the number entered. Key-in a numeric or alphanumeric value in an active dialog field: The context of the
field will determine if the entry is interpreted as numeric or alphanumeric.
The keys 0 to 9 enter the corresponding numbers (numeric input fields) or characters (character input fields).
In numeric input fields, the key enters the decimal point. Multiple decimal points are not allowed. In alphanumeric fields the key enters special characters like (? ~ @ ! _ etc).
ENTER activates the edit mode for the selected input field or confirms and terminates an entry.
Volume Control Rollkey
The volume of the speaker and headphones is adjusted with this knob. Pressing the knob will toggle the speaker mute. Headphones remain active when the speaker is in mute position.
USB Port and Headphone Jack
At the bottom, right we can find:
A USB port which provides a port for external keyboard, mouse or mass storage (of
snapshots, recordings etc) on USB flash drive. A headphone jack for stereo audio on headphone or external speaker.
The front panel USB port complies with USB standard 2.0.
28Operating Manual 4072.8426.02 ─ 08
Page 29
R&S® EB500
Operation
Front-Panel Tour
USB port
The maximum current per USB port is 500 mA.
3.1.2

R&S EB500 Model 02 (without front control panel)

An overview of the front-panel elements of the R&S EB500 Model 02 is given in the paragraph below.
Remote Access
Remote Access
A GUI for remote access is available for the Windows XP platform and is particularly useful for the R&S EB500 Model 02 because it gives functionality similar to the Model
03. Follow the instructions in chapter 5.6, "Remote Access GUI Installation",
on page 393 to install the external GUI software. A Windows XP machine connected to the same network as the R&S EB500 Model 02 is required.
29Operating Manual 4072.8426.02 ─ 08
Page 30
R&S® EB500
Operation
Rear-Panel Tour
3.1.2.1 Standby key with LED indicators
The only key on a R&S EB500 Model 02 is the standby key. The key is accompanied by two indicators:
Ready (green) or Fail (red) LED indicator. The Ready LED also serves as Fail LED,
as shown in "Model 02" on page 21. Standby (amber) LED indicator.
3.1.2.2 Headphones Jack
The headphones Jack allows stereo headphones with a 3.5 mm stereo plug to be con­nected.

3.2 Rear-Panel Tour

This chapter describes the various inputs and outputs located on the rear panel of the R&S EB500.
30Operating Manual 4072.8426.02 ─ 08
Page 31
R&S® EB500
Operation
Rear-Panel Tour

3.2.1 X1 Power Supply

Standard Speakon NL4FX socket for DC power input. Socket accepts 10~32 VDC on pin 1+ and ground on pin 1- of connector. Power requirement as indicated. Refer to chap-
ter 2.2.5.1, "Connecting to the Power Adapter", on page 20 for instructions on connecting
to the power adapter using the Speakon NL4FX connector.

3.2.2 X13 / X14 Antennas

N-type female
X13 Antenna input: 9 kHz to 6 GHz, suitable for HF/V/UHF bands.
X14 Antenna input: 9 kHz to 32 MHz, suitable for HF band.
Interfaces X13, and X14
All interfaces must be operated with double shielded RF cables.
31Operating Manual 4072.8426.02 ─ 08
Page 32
R&S® EB500
Operation
Rear-Panel Tour
Antenna inputs and frequency ranges
Depending on the option R&S EB500-HF, the antenna inputs can be used for different frequency ranges: X13 is a HF/VHF/UHF input and X14 serves as HF input only.
When the HF option is installed, the HF input can be chosen from either X13 or X14. For the default configuration (only V/UHF) the only available input is X13.

3.2.3 X7 LAN Ethernet

The LAN port (RJ-45) at the rear panel is a 10/100/1000Base-T port. Use this port to integrate the R&S EB500 in a local area network (LAN), primarily for remote-control pur- poses.
Ethernet Cable
At least a CAT 6 double-shielded Ethernet cable should be used.

3.2.4 Miscellaneous IO

X8 VIDEO A and X9 VIDEO B are SMA connector outputs. Use these outputs for an
analog video monitor or connect an oscilloscope to both connectors to see the AM and FM demodulated signals at the same time. Both connectors can also be switched to
32Operating Manual 4072.8426.02 ─ 08
Page 33
R&S® EB500
Operation
Rear-Panel Tour
output-controlled analog IF in two channels, with an adjustable center frequency of 0 Hz to 70 MHz.
X10 REF IN is a BNC input for an external reference frequency. Use this connector to synchronize the receiver with a frequency normal.
X11 REF OUT is a BNC output for other devices, e.g. other R&S EB500 units. Use this connector to synchronize the other devices.
X12 GPS/TR IN is a BNC connector for an external 1 sec trigger pulse, e.g. from a GPS device.
Interfaces X8, X9, X10, X11 and X12
All interfaces must be operated with double shielded coaxial cables.

3.2.5 AUX / AUX Audio

X3 AUX AUDIO is a 25-pin D-Sub female connector with the output of different audio
signals. There are also some special trigger inputs on this connector. Use this connector if you want to record the analog audio signal of the R&S EB500
33Operating Manual 4072.8426.02 ─ 08
Page 34
R&S® EB500
Operation
Rear-Panel Tour
X4 AUX is a 25-pin D-Sub female connector for controlling external antenna switching units. Use this connector if you have an antenna switching unit in your system.
Interfaces X3, and X4
All interfaces must be operated with double shielded data cables.

3.2.6 Antenna Control Interface

X2 ANTENNA DF is a 25-pin D-Sub connector. If DF upgrade R&S EB500-DF is installed
then this connector can be used for controlling direction-finder antennas.

3.2.7 Serial Interfaces

EIA 232D compatible
X5 COM2, 9-pin D-Sub male connector.
X6 COM1, 9-pin D-Sub male connector.
X5 and X6 can be used for:
Serial GPS devices according to the NMEA0183 standard. Use this connector for an
external GPS device to determine the location and the exact time of the device. Serial compass devices according to the NMEA0183 standard. Use this connector
for an external compass device to determine the direction of your vehicle. This will mostly be used in combination with direction finding (requires DF upgrade R&S EB500-DF).
SCPI
Besides control of devices according to the NMEA0183 standard, the COM2 port X5 also allows external control of the R&S EB500 by using SCPI syntax. See chapter 7.2, "Con-
figuration of the LAN Interface with SERIAL Interface", on page 429
34Operating Manual 4072.8426.02 ─ 08
Page 35
R&S® EB500
Operation
Graphical User Interface (GUI)

3.3 Graphical User Interface (GUI)

This chapter provides a detailed description of the device's graphical user interface.

3.3.1 GUI Layout

Following is an overview of the layout and individual elements of the device's graphical user interface.
Fig. 3-1: GUI Layout.
In the table below follows a short description of the different panels, according to the panel number indicated in the figure.
Panel No. Description
1 Basic receiver parameters
2 Icons
Shows one or several of the following icons to indicate a specific function or condition of the R&S EB500:
Compass connected.
GPS connected.
Configuration: using external reference.
35Operating Manual 4072.8426.02 ─ 08
Page 36
R&S® EB500
Operation
Graphical User Interface (GUI)
Panel No. Description
Receiver overflow.
Recording in progress.
Marker selected.
Speaker is off.
Speaker is on.
Audio signal is available in stereo at current fre­quency.
The R&S EB500 is not locked.
The R&S EB500 is locked by another client. Access is not possible.
3 Mode
Shows the current receiver mode, e.g. FFM, PScan, FScan or MScan. The current date and time is also shown.
4 Frequency
Shows the current receiver frequency. As this is the most important parameter, it is dis­played much larger than the remaining parameters.
5 Measurement parameters
Shows the basic measurement parameters (level, frequency offset; numerically and bar graph).
The R&S EB500 is locked. Only the current operator can make changes (exclusive write access).
A critical error occurred in the receiver. The "Test Points" dialog will provide more informa­tion.
36Operating Manual 4072.8426.02 ─ 08
Page 37
R&S® EB500
Panel No. Description
6 Panel area
The panel area is used for displaying the various panels of the R&S EB500 (spectrum and waterfall are shown).
The light cyan color of the scale bar indicates that the panel is in focus: softkey
operations affect the panel (see below). A marker to show the squelch level relative to the trace level is also displayed. The dark blue bar around the demodulation frequency marking denotes the band-
width being demodulated. On the right side, the rainbow colored bar indicates the color that is mapped to the
corresponding spectrum level for the waterfall display.
7 Softkeys
Softkeys change their function based on the panel that is in focus. They show panel­dependent parameters, e.g. "SPAN" and "STEP" for IF panel and IF Waterfall panel. Other softkeys are only related to e.g. RF panel, RF Waterfall panel etc.
Softkeys can be related to hardkeys: they can e.g. provide selection options in case one of the Utility keys is pressed or in case DET or MARK is pressed.
Softkeys can also provide the option for selecting a unit (kHz, MHz etc) e.g. when keying­in a frequency, they can provide additional functionality when editing input fields in dia­logs, e.g. backspace or caps lock. Finally they can serve as shortcut keys for navigating through tables in dialogs, e.g. in the "Memory Setup" dialog or through the "Help" pages.
Operation
Graphical User Interface (GUI)

3.3.2 Description of Panels

The individual panels are described in more detail in the following sections.
3.3.2.1 Frequency Panel
By default, the frequency panel displays the following parameters:
date and time
operating mode ("FFM", "FScan", "FScan Stop" etc.)
frequency field
level (numerically and bar graph)
frequency offset (numerically)
The frequency panel can be configured such that the bar graph frequency offset is shown instead of the bar graph level (see "Panels Configuration Dialog" on page 139) as shown in the figure below.
The default setting is "Level". If you select "Fieldstrength", the level, offset and field­strength are shown as numerical values.
37Operating Manual 4072.8426.02 ─ 08
Page 38
R&S® EB500
Operation
Graphical User Interface (GUI)
The unit displayed in the frequency field can be changed via the panel configuration, using SETUP > "Config" > "Panels"; see "Panels Configuration Dialog" on page 139.
3.3.2.2 Parameter Panel
The parameter panel shows the most important receiver parameters.
Table 3-1: The table below lists all the parameters of the parameter panel with a short description of
ATT Attenuation (manual, automatic)
MGC Gain control (manual, automatic)
SQU Squelch (manual, automatic)
BW Bandwidth (0.1 kHz - 20.0 MHz)
AFC Automatic frequency correction "ON"/"OFF"
DET Detection ("AVG", "PEAK", "FAST", "RMS")
DEMOD Demodulation ("CW", "LSB", "USB", "ISB", "AM", "FM", "PULS", "PM", "IQ")
each
You can change these parameters using the appropriate front-panel controls.
Either one of the parameters "ATT", "MGC" or "SQU" at the top panel will be highlighted by means of a yellow background. This will tell the operator which of the parameters can currently be controlled by the SELECT rollkey.
If none are highlighted then the SELECT key is set to control the markers. This will be indicated by the "Marker" icon at the top right (see chapter 3.3.1, "GUI Layout", on page 35).
Other parameters will only be highlighted for a very short duration to indicate which parameter is being changed.
3.3.2.3 Parameter Input
Use the parameter dialog to enter a value directly. The dialog is active for the current GUI element, i.e. a softkey. Normally it will be active for the center frequency: The dialog will appear once a user starts entering a number.
When a GUI element has the focus, it is shown in a lighter color.
38Operating Manual 4072.8426.02 ─ 08
Page 39
R&S® EB500
Operation
Graphical User Interface (GUI)
Example: "MEAS TIME" softkey has the focus. The related parameter dialog will be:
As long as the dialog is on the foreground the default softkey bar will be replaced by the softkey bar below. These softkeys allow you to change the unit used for the parameter shown, replace a wrongly typed digit or change the sign.
Most of the time the focus is not with any of the softkeys. Instead the focus is on one of the panels, e.g. the IF panorama, which are not meant for input. In that case the parameter input will directly work on the receiver center frequency, as shown below.
The dialog will open by:
pressing ENTER. The dialog opens for the currently focused GUI element or the
receiver frequency. The current value is shown. pressing a number key. The dialog opens for the currently focused GUI element or
the receiver frequency. The number associated with the number key pressed is shown.
The dialog will close by:
pressing ENTER. The receiver will change the center frequency to the number just
entered. pressing ESC. The number just entered will be discarded.
pressing any of the receiver control keys ATT, MGC, SQU, BW, AFC, DET or
DEMOD. The number just entered will be redirected. See below.
Parameter redirection
39Operating Manual 4072.8426.02 ─ 08
Page 40
R&S® EB500
Operation
Graphical User Interface (GUI)
Once open, a particular dialog is not limited to the parameter shown. Input can be redir­ected by pressing one of the following controls while the dialog is still open.
When pressing ATT the attenuation will change to "Manual" with the value taken from
the dialog. When pressing MGC the gain control will change to "Manual" with the value taken
from the dialog. When pressing SQU the squelch will be switched on and the value will be taken from
the dialog. For the BW, DET and DEMOD keys the value from the dialog will represent the index
in the selector box, e.g. the second DET is "FAST" and the eighth BW is "2.7 kHz". For AFC there are only two options: 0 is "off" and 1 is "on".
3.3.2.4 Markers
The main marker for the receive frequency is displayed in yellow at the center of the spectrum. Range markers (red or purple) are activated with the marker selector, which can be reached via the MARK key (see: "Receiver Control Keys" on page 26).
Fig. 3-2: Markers.
Once the MARK key is pressed two things will happen:
The SELECT knob will become available for manipulating the markers. Take note of
the Marker icon in the icon panel (see:figure 3-1 ). The default softkey bar changes into a dedicated softkey bar to provide options for
enabling / disabling and controlling the markers.
"X1/X2" activates and/ selects the first and second frequency range marker (red).
"X3/X4" activates and selects the third and fourth frequency range marker (purple).
40Operating Manual 4072.8426.02 ─ 08
Page 41
R&S® EB500
Operation
Graphical User Interface (GUI)
"Y1/Y2" activates and selects the first and second level range marker (red).
"Clear" deactivates the currently activated marker pair.
"Clear All" deactivates all markers, i.e. X1, X2, X3, X4, Y1 and Y2.
"Marker Val" activates / deactivates the output of the level of the currently selected fre­quency range marker. This is only possible for X1, X2, X3 and X4.
"Freq Val" activates / deactivates output of the frequency and level of the received signal.
When the "Real Freq" softkey is pressed, the frequency value in the marker will show the actual frequency of the level indicated by the marker. See the note below for an explan­ation.
Only two range markers are selected at any one time. Selected range markers are dis­played in solid colors (red and green or purple and green). Only the green color marker can be moved by the SELECT knob. By pressing the softkey of the activated marker pair, the green marker can be toggled, e.g. between X1 and X2 in case of X1/X2 markers or X3 and X4 in case of X3/X4 markers. This way the movable marker can be swapped.
Real frequency display at the markers
When the stepsize is set to average, there are usually more frequency levels calculated than what can be can be displayed on the screen. This is because the screen resolution is 640 pixels in horizontal direction whereas the number of frequencies calculated by FFT can be much higher (up to 4000 depending on the settings). For some settings up to eight frequencies can fall into one pixel.
The algorithm that is used in the R&S EB500 will display for each pixel the frequency with the highest level of all the frequencies that fall in that particular pixel. The frequency indicated by the marker however is the frequency, according to the pixel in the screen (the mean of all the frequencies that fall in that particular pixel).
In order to be able to find out what is the actual frequency represented by the level indi­cated, you should press the "Real Freq" softkey. When "Real Freq" is highlighted you will see the frequency indicated by the marker change over time, because in a real signal the frequency with the highest level will never be exactly the same all the time.
Pressing the SELECT knob (see: "Receiver Control Keys" on page 26) activates or deactivates a single range marker. Non-selected range markers are displayed in a trans­lucent color (red or purple). Measurement output fields are displayed along the upper edge in the respective range-marker color.
Averaging of the measurement output can be switched on under SETUP > Config > Panels Configuration > Averaging For Ruler's Digital Output. The frequency span between two activated markers is shown along the upper edge.
The measurement output for range markers can also be shown in the measurement out­put panel (see: "Panels Configuration Dialog" on page 139).
41Operating Manual 4072.8426.02 ─ 08
Page 42
R&S® EB500
Operation
Graphical User Interface (GUI)
3.3.2.5 IF Panorama
Fig. 3-3: IF Panorama.
In FFM the IF panorama shows signals centered around the current receive frequency. The IF panorama bandwidth is determined by the frequency span setting (softkey "SPAN"). The dark blue background indicates the receiver bandwidth. The level threshold is indicated along the left margin. The color set for color indexing in the IF waterfall panel is shown along the right margin.
The ROTARY KNOB, the CURSOR LEFT / CURSOR RIGHT keys and the numeric key­pad all can be used to control the center frequency. The ROTARY KNOB can be used for turning (clockwise for higher frequencies and counter-clockwise for lower frequencies) and clicking (the center frequency will be rounded to the nearest "STEP" value). The CURSOR LEFT / CURSOR RIGHT keys will change the center frequency with half the "SPAN" to lower or higher frequencies: after pressing the CURSOR LEFT key the center frequency will be reduced by half the "SPAN" and after pressing the CURSOR RIGHT key it will be increased by half the "SPAN". The number keys allow direct entry of a new center frequency: If the focus is not on any dialog or softkey then the effect of just keying a number using the number keys will cause the receiver to use that particular number for the center frequency, provided the number is within the range of the receiver.
If you connect a mouse and left click the yellow main marker on the X axis its ruler (the yellow line above the marker) will briefly appear. You can change the position of an active X marker by left dragging it. If you left drag the yellow main marker along the X axis, you can change the center frequency. By right dragging the yellow main marker along the X axis you can change the demodulation frequency.
If you left click the squelch marker (level threshold marker) on the Y axis its ruler will briefly appear. By left dragging the squelch marker along the Y axis, you can change its value. By left dragging a Y marker along the Y axis you can change its position.
You can move the center frequency by left clicking into the spectrum. By right clicking into the spectrum, you can move the demodulation frequency to the clicked position. You can change the position of a marker in the spectrum by left dragging it. You can change the center frequency by left dragging the yellow main marker in the spectrum. By right dragging the yellow main marker in the spectrum, you can change the demodulation frequency.
If you connect a keyboard then you can also change the center frequency using the cursor keys. The CURSOR RIGHT and CURSOR UP keys will move the center frequency by half the set "SPAN" to a higher frequency. The CURSOR LEFT and CURSOR DOWN keys will move the center frequency by half the set "SPAN" to a lower frequency.
42Operating Manual 4072.8426.02 ─ 08
Page 43
R&S® EB500
Operation
Graphical User Interface (GUI)
If you press the mouse wheel or ROTARY KNOB, the current center frequency will be rounded to the step width. If you connect an external keyboard then SHIFT+<NUMERIC KEY> will round the center frequency to the corresponding decimal power, e.g. SHIFT +3 will round to kHz.
Use the following softkeys to configure the IF panorama:
Press the "DEMOD FREQ" button to set the demodulation frequency, which is normally the same as the receive frequency. Any deviations will be temporary, which means that the ROTARY KNOB will act upon the demodulation frequency for as long as the focus is on the "DEMOD FREQ" button. The frequency limits of the demodulation frequency are determined by the receive frequency and the IF panorama span.
In the IF Panorama it can easily be seen by a change in the position of the blue bandwidth bar whether the demodulation frequency has been changed. The bar will not be posi­tioned at the centre anymore, as can be seen in the figure below.
Once the focus is not on the "DEMOD FREQ" softkey anymore, a change in frequency will again affect the main frequency and the blue bandwidth bar will return back to the centre of the IF Panorama.
See also [SENSe:]FREQuency:DEModulation on page 265.
Press the "SPAN" button to set the IF panorama span.
A related vertical selector will pop-up, showing the possible values, which range from 1 kHz to 20 MHz.
See also [SENSe:]FREQuency:SPAN on page 270.
Press the "STEP" button to set the IF panorama step width. Step widths range from 625 mHz to 2 MHz.
43Operating Manual 4072.8426.02 ─ 08
Page 44
R&S® EB500
Operation
Graphical User Interface (GUI)
When set to "AUTO", the step width is determined automatically depending on the IF panorama span. A related vertical selector will pop-up, showing the available options.
Press the "SEARCH PEAK" button to center the IF panorama to the next left or right level peak, or to the absolute level peak. You can choose either the center frequency or the demodulation frequency. Once the softkey is pressed, a new row of softkeys will appear, which show the available options.
See also:
CALCulate:IFPan:MARKer:MAXimum:LEFT on page 204
CALCulate:IFPan:MARKer:MAXimum:RIGHt on page 204
CALCulate:IFPan:MARKer:MAXimum[:PEAK] on page 204
CALCulate:IFPan:MARKer:DEModulation:MAXimum:LEFT on page 204
CALCulate:IFPan:MARKer:DEModulation:MAXimum:RIGHt on page 204
CALCulate:IFPan:MARKer:DEModulation:MAXimum[:PEAK] on page 204
Press the "FFT MODE" button to select the FFT mode for the IF panorama data.
Once the softkey is pressed, a vertical selector will appear, which shows the available options. Available options are "MINIMUM", "MAXIMUM", "AVERAGE" and "CLRWRITE"
See also CALCulate:IFPan:AVERage:TYPE on page 202.
Press the "SELECTIVITY" button to set the selectivity in the IF panorama. You can only set the desired selectivity. The actual selectivity that can be achieved depends on the combination of step width and span.
44Operating Manual 4072.8426.02 ─ 08
Page 45
R&S® EB500
Operation
Graphical User Interface (GUI)
A related vertical selector shows the available options. Available options are "AUTO", "NORMAL", "NARROW" and "SHARP."
See also CALCulate:IFPan:SELectivity on page 205.
Press the "MEAS MODE" button to select the measurement mode. The measurement mode can be set to "CONT" ("CONTinuous") or "PER" ("PERiodic").
See also chapter 4.5.9, "MEASure Subsystem", on page 227.
Press the "MEAS TIME" button to set the measurement time. You can set a concrete time for periodic measurements or you can set it to "AUTO", in which case it will be determined automatically.
See also chapter 4.5.9, "MEASure Subsystem", on page 227.
Press the "POLYCHROME" button to operate the IF panorama in polychrome mode. You can set it to "OFF" or "HISTOGRAM" (color histogram). A related selector shows the available options.
If you select "HISTOGRAM", the frequency of the levels of the received signal measured over a defined time ("PERSISTENCE") will be shown in addition to the levels. Red dots indicate high occupancy (more frequently occurring levels) whereas blue dots indicate low occupancy (less frequently occurring levels).
Black areas indicate that no levels are being measured.
Fig. 3-4: Polychrome IF panorama.
See also CALCulate:PIFPan:MODE on page 208 .
Press the "100% TIME" button to set the 100% time (the level occupancy required to indicate 100% in the color bar scale) for the polychrome IF panorama.
See also CALCulate:PIFPan:ACTTime on page 207 .
Press the "PERSISTENCE" button to set persistence, i.e. the time during which the occupancy of a signal is to be measured.
See also CALCulate:PIFPan:OBSTime on page 209.
Press the "CLEAR POLYCHROME" button to clear the polychrome IF panorama; the measurement will restart.
45Operating Manual 4072.8426.02 ─ 08
Page 46
R&S® EB500
Operation
Graphical User Interface (GUI)
See also CALCulate:PIFPan:CLEar on page 208 .
Press the "LOW LIMIT" button to set the lower time limit for 100% time. Occupancies less than the lower limit will not be indicated with a color in the IF panel. You can set the lower limit to a value between 0 ms and the upper limit.
Press the "UPPER LIMIT" button to set the upper time limit for 100% time. Occupancies higher than the upper limit will not be indicated with a color in the IF panel. You can set the upper limit to a value between the lower limit and 100% time.
Press the "FREQ INCR" button to set the frequency increment for changing the receive frequency. You can choose between manual and automatic mode. If you select manual mode, the specified value will be used as increment. If you select "AUTO" mode, the step width of the currently focused panel will be used as increment.
See also [SENSe:]FREQuency:STEP[:INCRement] on page 272.
Press the "DIFF MODE" button to turn on the "Differential" mode. When turned on, the current spectrum is used as the reference; only the changes with respect to the reference spectrum are displayed.
This mode is indicated by the words "Differential Mode" displayed at the top of the spec­trum. Note that the waterfall panel does not support differential mode.
If you turn the "Differential" mode off, the regular spectrum is displayed again.
Press the "CONFIG IFPAN" button to configure the IF panorama. You can configure the value range of the Y axis, the color set for the waterfall and the "Hold Maximum" function.
See also chapter 3.3.3.11, "IF Panorama Dialog", on page 170.
3.3.2.6 IF Waterfall
Fig. 3-5: IF Waterfall.
The IF waterfall, which by default is displayed below the IF panorama, shows signal levels from the IF panorama over time. The signal levels are color coded; the colors correspond to the color set shown along the right margin. They correspond with the level in the IF panorama. The time in seconds is indicated along the left margin.
You can change the position of a Y marker by turning the SELECT rollkey or by left dragging it along the Y axis (in case a mouse is connected).
46Operating Manual 4072.8426.02 ─ 08
Page 47
R&S® EB500
Operation
Graphical User Interface (GUI)
If you want to change the center frequency by using a mouse, left click into the waterfall. Likewise, if you want to change the demodulation frequency, right click into the waterfall. You can also left drag a marker in the waterfall to change its position. You can change the center frequency by left dragging the yellow main marker on the X axis. By right dragging the yellow main marker on the X axis, you can change the demodulation fre­quency. If the IF waterfall is displayed in a single-panel layout (SETUP > "Layout" > "IF WF ONLY") then there will be no X axis and no marker.
Press the mouse wheel or ROTARY KNOB to round the current center frequency to the step width.
Use the following softkeys to configure the IF waterfall:
Press the "DEMOD FREQ" button to set the demodulation frequency, which is normally the same as the receive frequency. Any deviations will be temporary, which means that the ROTARY KNOB will act upon the demodulation frequency for as long as the focus is on the "DEMOD FREQ" button. The frequency limits of the demodulation frequency are determined by the receive frequency and the IF panorama span.
See also [SENSe:]FREQuency:DEModulation on page 265.
Press the "SPAN" button to set the IF panorama span.
A related vertical selector will pop-up, showing the possible values, which range from 1 kHz to 20 MHz.
See also [SENSe:]FREQuency:SPAN on page 270.
Press "STEP" button to set the IF panorama step width.
Step widths range from 625 mHz to 2 MHz. When set to "AUTO", the step width is deter­mined automatically depending on the IF panorama span. A related vertical selector will pop-up, showing the available options.
47Operating Manual 4072.8426.02 ─ 08
Page 48
R&S® EB500
Operation
Graphical User Interface (GUI)
Press the "FFT MODE" button to select the FFT mode for the IF panorama data.
Once the softkey is pressed, a vertical selector will appear, which show the available options. Available options are "MINIMUM", "MAXIMUM", "AVERAGE" and "CLRWRITE"
See also CALCulate:IFPan:AVERage:TYPE on page 202.
Press the "SELECTIVITY" button to set the selectivity in the IF panorama.
You can only set the desired selectivity. The actual selectivity that can be achieved depends on the combination of step width and span. A related selector shows the avail­able options. Available options are "AUTO", "NORMAL", "NARROW" and "SHARP."
See also CALCulate:IFPan:SELectivity on page 205.
Press the "MEAS MODE" button to select the measurement mode. The measurement mode can be set to "CONT" ("CONTinuous") or "PER" ("PERiodic").
See also chapter 4.5.9, "MEASure Subsystem", on page 227.
Press the "MEAS TIME" button to set the measurement time. You can set a concrete time for periodic measurements or you can set it to "AUTO", in which case it will be determined automatically.
See also chapter 4.5.9, "MEASure Subsystem", on page 227.
Press the "POLYCHROME" button to operate the IF panorama in polychrome mode. You can set it to "OFF" or "HISTOGRAM" (color histogram). A related selector shows the available options.
This is explained in detail in the IF Panorama above.
See also CALCulate:PIFPan:MODE on page 208 .
Press the "100% TIME" button to set the 100% time (the level occupancy required to indicate 100% in the color bar scale) for the polychrome IF panorama.
See also CALCulate:PIFPan:ACTTime on page 207 .
Press the "PERSISTENCE" button to set persistence, i.e. the time during which the occupancy of a signal is to be measured.
See also CALCulate:PIFPan:OBSTime on page 209.
Press the "CLEAR POLYCHROME" button to clear the polychrome IF panorama; the measurement will restart.
See also CALCulate:PIFPan:CLEar on page 208.
48Operating Manual 4072.8426.02 ─ 08
Page 49
R&S® EB500
Operation
Graphical User Interface (GUI)
Press the "LOW LIMIT" button to set the lower time limit for 100% time. Occupancies less than the lower limit will not be indicated with a color in the IF panel. You can set the lower limit to a value between 0 ms and the upper limit.
Press the "UPPER LIMIT" button to set the upper time limit for 100% time. Occupancies higher than the upper limit will not be indicated with a color in the IF panel. You can set the upper limit to a value between the lower limit and 100% time.
Press the "FREQ INCR" button to set the frequency increment for changing the receive frequency. You can choose between manual and automatic mode. If you select manual mode, the specified value will be used as increment. If you select "AUTO" mode, the step width of the currently focused panel will be used as increment.
Press the "DIFF MODE" button to turn on the "Differential" mode. When turned on, the current spectrum is used as the reference; only the changes with respect to the reference spectrum are displayed.
This mode is indicated by the words "Differential Mode" displayed at the top of the spec­trum. Note that the waterfall panel does not support differential mode.
Press the "CONFIG IF WF" button to configure the IF waterfall. You can configure the waterfall speed, and the color set to be used for the waterfall.
See also chapter 3.3.3.12, "IF Waterfall Dialog", on page 174.
3.3.2.7 RF Spectrum
Fig. 3-6: RF Spectrum.
The RF spectrum shows signals in Scan mode. The RF spectrum span width is deter­mined by setting the start and stop frequency. The level threshold is indicated along the left margin. The color set for color indexing in the RF waterfall is displayed along the right margin.
Just like in the IF panorama (FFM mode, see chapter 3.3.2.5, "IF Panorama", on page 42) you can use the main rollkey, the arrow keys and the numeric keypad to control the center frequency. However, you won't see an effect for every scan mode. In fact, only in PScan mode the center frequency will actually change and demodulation will take place once the scan is stopped. For FScan and MScan, demodulation takes place all the while and the center frequency can only be changed once the scan has stopped.
If you connect a keyboard then you can change the center frequency using the cursor keys. The CURSOR RIGHT and CURSOR UP keys will move the center frequency by half the set "SPAN" to a higher frequency. The CURSOR LEFT and CURSOR DOWN keys will move the center frequency by half the set "SPAN" to a lower frequency. This can be done during the scan or at scan stop.
49Operating Manual 4072.8426.02 ─ 08
Page 50
R&S® EB500
Operation
Graphical User Interface (GUI)
Depending on the scan mode, use the following softkeys to configure the RF panel:
Panorama scan
The panorama scan requires option R&S EB500-PS. Check SETUP > Information > Options.
Press the "START" button to set the start frequency for the panorama scan.
See also [SENSe:]FREQuency:PSCan:STARt on page 269.
Press the "STOP" button to set the stop frequency for the panorama scan.
See also [SENSe:]FREQuency:PSCan:STOP on page 269.
Press the "STEP" button to select the step width for the panorama scan.
A related selector dialog will pop-up providing the available options.
See also [SENSe:]PSCan:STEP on page 284.
Press the button "CYCLE COUNT" to set the cycle count, i.e. how often the frequency band between the start and stop frequency is to be scanned. You can set the cycle count to a number between 1 and 1000. You can also set it to "INFinity".
See also [SENSe:]PSCan:COUNt on page 283.
Press the button "FFT MODE" to select the FFT mode for the IF panorama data.
A related selector shows the available options. Available options are "MINIMUM", "MAX­IMUM", "AVERAGE" and "CLRWRITE"
See also: CALCulate:IFPan:AVERage:TYPE on page 202.
50Operating Manual 4072.8426.02 ─ 08
Page 51
R&S® EB500
Operation
Graphical User Interface (GUI)
Press the "CENTER" button to set the center frequency for the panorama scan.
See also [SENSe:]FREQuency:PSCan:CENTer on page 267.
Press the "SPAN" button to set the span for the panorama scan.
See also [SENSe:]FREQuency:PSCan:SPAN on page 268.
Press the "MEAS MODE" button to select the measurement mode. The measurement mode can be set to "CONT" ("CONTinuous") or "PER" ("PERiodic").
See also chapter 4.5.9, "MEASure Subsystem", on page 227.
Press the "MEAS TIME" button to set the measurement time. You can set a concrete time for periodic measurements or you can set it to "AUTO", in which case it will be determined automatically.
See also chapter 4.5.9, "MEASure Subsystem", on page 227.
Press the "FREQ INCR" button to set the frequency increment for changing the receive frequency. You can choose between manual and automatic mode. If you select manual mode, the specified value will be used as increment. If you select "AUTO" mode, the step width of the currently focused panel will be used as increment.
See also [SENSe:]FREQuency:STEP[:INCRement]? on page 272.
Press the "DIFF MODE" button to turn on the "Differential" mode. When turned on, the current spectrum is used as the reference; only the changes with respect to the reference spectrum are displayed.
This mode is indicated by the words "Differential Mode" displayed at the top of the spec­trum. Note that the waterfall panel does not support differential mode.
If you turn the "Differential" mode off, the regular spectrum is displayed again.
Press the button "CONFIG" to configure the RF spectrum. You can configure the value range of the Y axis and the "Hold Max" function. You can also choose between a comb and line spectrum and specify the color set to be used for the waterfall.
See also chapter 3.3.3.13, "RF Panorama Dialog", on page 176.
Frequency scan
Press the "START" button to set the start frequency for the frequency scan.
See also [SENSe:]FREQuency:STARt on page 271.
Press the "STOP" button to set the stop frequency for the frequency scan.
See also [SENSe:]FREQuency:STOP on page 272.
Press the "STEP" button to select the step width for the frequency scan.
See also [SENSe:]SWEep:STEP on page 290
Press the button "CYCLE COUNT" to set the cycle count, i.e. how often the frequency band between the start and stop frequency is to be scanned. You can set the cycle count to a number between 1 and 1000. You can also set it to "INFinity".
See also [SENSe:]SWEep:COUNt on page 287.
51Operating Manual 4072.8426.02 ─ 08
Page 52
R&S® EB500
Operation
Graphical User Interface (GUI)
Press the button "FFT MODE" to select the FFT mode for the IF panorama data.
A related selector shows the available options. Available options are "MINIMUM", "MAX­IMUM", "AVERAGE" and "CLRWRITE".
See also CALCulate:IFPan:AVERage:TYPE on page 202.
Press the "CENTER" button to set the center frequency for the frequency scan.
Press the "SPAN" button to set the span for the frequency scan.
Press the "MEAS MODE" button to select the measurement mode. The measurement mode can be set to "CONT" ("CONTinuous") or "PER" ("PERiodic").
See also chapter 4.5.9, "MEASure Subsystem", on page 227.
Press the "MEAS TIME" button to set the measurement time. You can set a concrete time for periodic measurements or you can set it to "AUTO", in which case it will be determined automatically.
See also chapter 4.5.9, "MEASure Subsystem", on page 227.
Press the "DWELL TIME" button to set the dwell time in milliseconds for each frequency. You can also set dwell time to "INFinity". The system will then dwell on the current fre­quency until you press RUN+ or RUN-.
See also [SENSe:]SWEep:DWELl on page 288.
Press the button "NO SIG TIME" to set the time for signal-controlled operation, i.e. if the signal disappears during dwell time, the system will wait for the time specified and then continue with the scan. You can also turn signal-controlled operation off.
See also:
[SENSe:]SWEep:HOLD:TIME on page 289
[SENSe:]SWEep:CONTrol[:ON] on page 286
Press the button "FUNCTION SUPPRESS" to configure the suppress list. Once the "FUNCTION SUPPRESS" button is pressed a new row of softkeys appears with options as shown below.
"Suppress" adds the current frequency to the list.
52Operating Manual 4072.8426.02 ─ 08
Page 53
R&S® EB500
Operation
Graphical User Interface (GUI)
"Supp-" adds the current frequency to the list and the scan will continue in the reverse
direction once the dwell time ended "Supp+" is similar to "Supp-" except that scan will continue in the forward direction
once the dwell time ended. "Store-" works like "Supp-" except that the current frequency is also copied to mem-
ory. "Store+" works like "Supp+" except that the current frequency is also copied to mem-
ory. "Sort" will sort the suppress list in ascending order.
"Clear All" empties the suppress list; it clears the list of all frequency ranges.
"Edit" will open the "Suppress List Setup" dialog, which is used to edit the frequencies
to be suppressed during a frequency scan. You can define, edit, sort and delete these frequencies in the "Suppress List Setup" dialog. See also chapter 3.3.3.21, "Suppress List Setup Dialog", on page 189.
Press the "FREQ INCR" button to set the frequency increment for changing the receive frequency. You can choose between manual and automatic mode. If you select manual mode, the specified value will be used as increment. If you select "AUTO" mode, the step width of the currently focused panel will be used as increment.
See also [SENSe:]FREQuency:STEP[:INCRement]? on page 272.
Press the "DIFF MODE" button to turn on the "Differential" mode. When turned on, the current spectrum is used as the reference; only the changes with respect to the reference spectrum are displayed.
This mode is indicated by the words "Differential Mode" displayed at the top of the spec­trum. Note that the waterfall panel does not support differential mode.
If you turn the "Differential" mode off, the regular spectrum is displayed again.
Press the button "CONFIG" to configure the RF spectrum. You can configure the value range of the Y axis and the "Hold Max" function. You can also choose between a comb and line spectrum and specify the color set to be used for the waterfall.
See alsochapter 3.3.3.13, "RF Panorama Dialog", on page 176 .
Memory scan
Press the "MEM SELECT" button to set the number of the current memory channel. Depending on "SELECT MODE", the new memory channel will/will not be loaded.
During scanning the softkey will indicate which memory channel is currently loaded.
See also [SENSe:]MSCan:CHANnel on page 279.
Press the button "SELECT MODE" to set the mode for "MEM SELECT". Setting it to "PREVIEW" mode changes only the number of the current memory channel. Mode "AUTO RECALL" causes the receiver to assume the settings associated with the new memory channel.
Press the "MEM SQUELCH" button to use the squelch value from memory (turn it "ON") or the global squelch value (turn it "OFF").
53Operating Manual 4072.8426.02 ─ 08
Page 54
R&S® EB500
Operation
Graphical User Interface (GUI)
See also OUTPut:SQUelch:CONTrol on page 244.
Press the button "CYCLE COUNT" to set the cycle count, i.e. how many cycles the fre­quency band between the start and stop frequency is to be scanned. You can set the cycle count to a number between 1 and 1000. You can also set it to "INFinity".
See also [SENSe:]MSCan:COUNt on page 280.
Press the button "FFT MODE" to select the FFT mode for the IF panorama data.
A related selector shows the available options. Available options are "MINIMUM", "MAX­IMUM", "AVERAGE" and "CLRWRITE".
See also CALCulate:IFPan:AVERage:TYPE on page 202.
Press the "DWELL TIME" button to set the dwell time in milliseconds for each frequency. You can also set dwell time to "INFinity". The system will then dwell on the current fre­quency until you press RUN+ or RUN-.
See also [SENSe:]MSCan:DWELl on page 282.
Press the button "NO SIG TIME" to set the time for signal-controlled operation, i.e. if the signal disappears during dwell time, the system will wait for the time specified and then continue with the scan. You can also turn signal-controlled operation off.
See also:
[SENSe:]MSCan:HOLD:TIME on page 282
[SENSe:]MSCan:CONTrol[:ON] on page 280
Press the button "FUNCTION SUPPRESS" to configure the suppress list. The options below appear in a new row of softkeys that replace the default softkey bar once the "SUPPRESS" button is pressed.
"Suppress" adds the current frequency to the list.
"Supp-" adds the current frequency to the list and the scan will continue in the reverse
direction once the dwell time ended. "Supp+" is similar to "Supp-" except that the scan will continue in the forward direction
once the dwell time ended.
See also chapter 3.3.3.21, "Suppress List Setup Dialog", on page 189.
Press the "MEAS MODE" button to select the measurement mode. The measurement mode can be set to "CONT" ("CONTinuous") or "PER" ("PERiodic").
See also chapter 4.5.9, "MEASure Subsystem", on page 227.
54Operating Manual 4072.8426.02 ─ 08
Page 55
R&S® EB500
Operation
Graphical User Interface (GUI)
Press the "MEAS TIME" button to set the measurement time. You can set a concrete time for periodic measurements or you can set it to "AUTO", in which case it will be determined automatically.
See also chapter 4.5.9, "MEASure Subsystem", on page 227.
Press the "FREQ INCR" button to set the frequency increment for changing the receive frequency. You can choose between manual and automatic mode. If you select manual mode, the specified value will be used as increment. If you select "AUTO" mode, the step width of the currently focused panel will be used as increment.
See also [SENSe:]FREQuency:STEP[:INCRement]? on page 272.
Press the "DIFF MODE" button to turn on the "Differential" mode. When turned on, the current spectrum is used as the reference; only the changes with respect to the reference spectrum are displayed.
This mode is indicated by the words "Differential Mode" displayed at the top of the spec­trum. Note that the waterfall panel does not support differential mode.
Press the button "CONFIG" to configure the RF spectrum. You can configure the value range of the Y axis and the "Hold Max" function. You can also specify the color set to be used for the waterfall.
See also chapter 3.3.3.13, "RF Panorama Dialog", on page 176.
3.3.2.8 RF Waterfall
Fig. 3-7: RF Waterfall.
The RF waterfall below the RF spectrum shows signal levels from the RF spectrum over time. The signal levels are color coded; the colors correspond to the color set shown along the right margin. The time in seconds is indicated along the left margin.
You can change the position of a Y marker by left dragging it along the Y axis.
Just like in the RF panorama (FFM mode, see chapter 3.3.2.5, "IF Panorama", on page 42) you can use the main rollkey, the arrow keys and the numeric keypad to control the center frequency. However, you won't see an effect for every scan mode. In fact, only in PScan mode the center frequency will actually change and demodulation will take place once the scan is stopped. For FScan and MScan, demodulation takes place all the while and the center frequency can only be changed once the scan has stopped.
55Operating Manual 4072.8426.02 ─ 08
Page 56
R&S® EB500
Operation
Graphical User Interface (GUI)
If you connect a keyboard then you can change the center frequency using the cursor keys. The CURSOR RIGHT and CURSOR UP keys will move the center frequency by half the set SPAN to a higher frequency. The CURSOR LEFT and CURSOR DOWN keys will move the center frequency by half the set SPAN to a lower frequency. This can be done during the scan or at scan stop.
Use the following softkeys to configure the RF waterfall, depending on the scan mode:
Panorama scan
The panorama scan requires option R&S EB500-PS. Check SETUP > Information > Options.
Press the "START" button to set the start frequency for the panorama scan.
See also [SENSe:]FREQuency:PSCan:STARt on page 269.
Press the "STOP" button to set the stop frequency for the panorama scan.
See also [SENSe:]FREQuency:PSCan:STOP on page 269.
A related vertical selector will pop-up providing the available options.
See also [SENSe:]PSCan:STEP on page 284.
Press the button "CYCLE COUNT" to set the cycle count, i.e. how for how many cycles the frequency band between the start and stop frequency is to be scanned. You can set the cycle count to a number between 1 and 1000. You can also set it to "INFinity".
See also [SENSe:]PSCan:COUNt on page 283.
Press the "CENTER" button to set the center frequency for the panorama scan.
See also [SENSe:]FREQuency:PSCan:CENTer on page 267.
Press the "SPAN" button to set the span for the panorama scan.
See also [SENSe:]FREQuency:PSCan:SPAN on page 268.
Press the "STEP" button to select the step width for the panorama scan.
56Operating Manual 4072.8426.02 ─ 08
Page 57
R&S® EB500
Operation
Graphical User Interface (GUI)
Press the "MEAS MODE" button to select the measurement mode. The measurement mode can be set to "CONT" ("CONTinuous") or "PER" ("PERiodic").
See also chapter 4.5.9, "MEASure Subsystem", on page 227.
Press the "MEAS TIME" button to set the measurement time. You can set a concrete time for periodic measurements or you can set it to "AUTO", in which case it will be determined automatically.
See also chapter 4.5.9, "MEASure Subsystem", on page 227.
Press the "FREQ INCR" button to set the frequency increment for changing the receive frequency. You can choose between manual and automatic mode. If you select manual mode, the specified value will be used as increment. If you select "AUTO" mode, the step width of the currently focused panel will be used as increment.
See also [SENSe:]FREQuency:STEP[:INCRement]? on page 272.
Press the "DIFF MODE" button to turn on the "Differential" mode. When turned on, the current spectrum is used as the reference; only the changes with respect to the reference spectrum are displayed.
This mode is indicated by the words "Differential Mode" displayed at the top of the spec­trum. Note that the waterfall panel does not support differential mode.
Press the button "CONFIG" to configure the RF waterfall to set the speed, interpolation mode and color set for the waterfall. In case the RF panorama is displayed as a comb spectrum, the waterfall will show gaps at the same spacing. These can be interpolated by ticking the "Interpolation Mode" checkbox in this dialog.
See also chapter 3.3.3.14, "RF Waterfall Dialog", on page 178.
Frequency scan
Press the "START" button to set the start frequency for the frequency scan.
See also [SENSe:]FREQuency:STARt on page 271.
Press the "STOP" button to set the stop frequency for the frequency scan.
See also [SENSe:]FREQuency:STOP on page 272.
Press the "STEP" button to select the step width for the frequency scan.
See also [SENSe:]SWEep:STEP on page 290.
Press the button "CYCLE COUNT" to set the cycle count, i.e. how many cycles the fre­quency band between the start and stop frequency is to be scanned. You can set the cycle count to a number between 1 and 1000. You can also set it to "INFinity".
See also [SENSe:]SWEep:COUNt on page 287.
Press the "CENTER" button to set the center frequency for the frequency scan.
Press the "SPAN" button to set the span for the frequency scan.
57Operating Manual 4072.8426.02 ─ 08
Page 58
R&S® EB500
Operation
Graphical User Interface (GUI)
Press the "MEAS MODE" button to select the measurement mode. The measurement mode can be set to "CONT" ("CONTinuous") or "PER" ("PERiodic").
See also chapter 4.5.9, "MEASure Subsystem", on page 227.
Press the "MEAS TIME" button to set the measurement time. You can set a concrete time for periodic measurements or you can set it to "AUTO", in which case it will be determined automatically.
See also chapter 4.5.9, "MEASure Subsystem", on page 227.
Press the "DWELL TIME" button to set the dwell time in milliseconds for each frequency. You can also set dwell time to "INFinity". The system will then dwell on the current fre­quency until you press RUN+ or RUN-.
See also [SENSe:]SWEep:DWELl on page 288.
Press the button "NO SIG TIME" to set the time for signal-controlled operation, i.e. if the signal disappears during dwell time, the system will wait for the time specified and then continue with the scan. You can also turn signal-controlled operation off.
See also:
[SENSe:]SWEep:HOLD:TIME on page 289
[SENSe:]SWEep:CONTrol[:ON] on page 286
Press the button "FUNCTION SUPPRESS" to configure the suppress list. The options below appear in a new softkey bar that replaces the default softkey bar once the "FUNC­TION SUPPRESS" button is pressed.
"Suppress" adds the current frequency to the list.
"Supp-" adds the current frequency to the list and the scan will continue in the reverse
direction once the dwell time ended "Supp+" is similar to "Supp-" except that the scan will continue in the forward direction
once the dwell time ended. "Store-" works like "Supp-" except that the current frequency is also copied to mem-
ory. "Store+" works like "Supp+" except that the current frequency is also copied to mem-
ory. "Sort" will sort the suppress list in ascending order.
"Clear All" empties the suppress list; it clears the list of all frequency ranges.
"Edit" will open the "Suppress List Setup" dialog, which is used to edit the frequencies
to be suppressed during a frequency scan. You can define, edit, sort and delete these frequencies in the "Suppress List Setup" dialog. See also chapter 3.3.3.21, "Suppress List Setup Dialog", on page 189.
58Operating Manual 4072.8426.02 ─ 08
Page 59
R&S® EB500
Operation
Graphical User Interface (GUI)
Press the "FREQ INCR" button to set the frequency increment for changing the receive frequency. You can choose between manual and automatic mode. If you select manual mode, the specified value will be used as increment. If you select "AUTO" mode, the step width of the currently focused panel will be used as increment.
See also [SENSe:]FREQuency:STEP[:INCRement]? on page 272.
Press the "DIFF MODE" button to turn on the "Differential" mode. When turned on, the current spectrum is used as the reference; only the changes with respect to the reference spectrum are displayed.
This mode is indicated by the words "Differential Mode" displayed at the top of the spec­trum. Note that the waterfall panel does not support differential mode.
Press the button "CONFIG" to configure the RF waterfall to set the speed, interpolation mode and color set for the waterfall. In case the RF panorama is displayed as a comb spectrum, the waterfall will show gaps at the same spacing. These can be interpolated by ticking the "Interpolation Mode" checkbox in this dialog. Note that gaps caused by suppressed frequency ranges will not be interpolated.
See also chapter 3.3.3.14, "RF Waterfall Dialog", on page 178.
See also [SENSe:]SWEep:COUNt on page 287.
Memory scan
Press the "MEM SELECT" button to set the number of the current memory channel. Depending on "SELECT MODE", the new memory channel will/will not be loaded.
During scanning the softkey will indicate which memory channel is currently loaded.
See also [SENSe:]MSCan:CHANnel on page 279.
Press the button "SELECT MODE" to set the mode for "MEM SELECT". Setting it to "PREVIEW" mode changes only the number of the current memory channel. Mode "AUTO RECALL" causes the receiver to assume the settings associated with the new memory channel.
Press the "MEM SQUELCH" button to use the squelch value from memory ("ON") or the global squelch value ("OFF").
See also OUTPut:SQUelch:CONTrol on page 244.
Press the button "CYCLE COUNT" to set the cycle count, i.e. how many cycles the fre­quency band between the start and stop frequency is to be scanned. You can set the cycle count to a number between 1 and 1000. You can also set it to "INFinity".
See also [SENSe:]MSCan:COUNt on page 280.
Press the "DWELL TIME" button to set the dwell time in milliseconds for each frequency. You can also set dwell time to "INFinity". The system will then dwell on the current fre­quency until you press RUN+ or RUN-.
See also [SENSe:]MSCan:DWELl on page 282.
Press the button "NO SIG TIME" to set the time for signal-controlled operation, i.e. if the signal disappears during dwell time, the system will wait for the time specified and then continue with the scan. You can also turn signal-controlled operation off.
59Operating Manual 4072.8426.02 ─ 08
Page 60
R&S® EB500
Operation
Graphical User Interface (GUI)
See also:
[SENSe:]MSCan:HOLD:TIME on page 282
[SENSe:]MSCan:CONTrol[:ON] on page 280
Press the button "FUNCTION SUPPRESS" to configure the suppress list. The options below appear in a softkey bar that replaces the default softkey bar once the "FUNCTION SUPPRESS" button is pressed.
"Suppress" adds the current frequency to the list.
"Supp-" adds the current frequency to the list and the scan will continue in the reverse
direction once the dwell time ended. "Supp+" is similar to "Supp-" except that scan will continue in the forward direction
once the dwell time ended.
See also chapter 3.3.3.21, "Suppress List Setup Dialog", on page 189.
Press the "MEAS MODE" button to select the measurement mode. The measurement mode can be set to "CONT" ("CONTinuous") or "PER" ("PERiodic").
See also chapter 4.5.9, "MEASure Subsystem", on page 227.
Press the "MEAS TIME" button to set the measurement time. You can set a concrete time for periodic measurements or you can set it to "AUTO", in which case it will be determined automatically.
See also chapter 4.5.9, "MEASure Subsystem", on page 227.
Press the "FREQ INCR" button to set the frequency increment for changing the receive frequency. You can choose between manual and automatic mode. If you select manual mode, the specified value will be used as increment. If you select "AUTO" mode, the step width of the currently focused panel will be used as increment.
See also [SENSe:]FREQuency:STEP[:INCRement]? on page 272.
Press the "DIFF MODE" button to turn on the "Differential" mode. When turned on, the current spectrum is used as the reference; only the changes with respect to the reference spectrum are displayed.
This mode is indicated by the words "Differential Mode" displayed at the top of the spec­trum. Note that the waterfall panel does not support differential mode.
Press the button "CONFIG" to configure the RF waterfall, i.e. to set the speed and color set for the waterfall.
See also chapter 3.3.3.14, "RF Waterfall Dialog", on page 178.
60Operating Manual 4072.8426.02 ─ 08
Page 61
R&S® EB500
Operation
Graphical User Interface (GUI)
3.3.2.9 Video Panorama
Fig. 3-8: Video Panorama.
In FFM the video panorama shows the spectrum of the demodulated signal. The span width of the video panorama is calculated, based on the receiver bandwidth, normally rounded off to a number larger than the bandwidth sufficient to provide the required FFT length (4096). The color set for color indexing in the video waterfall is shown along the right margin.
The ROTARY KNOB, the CURSOR LEFT / CURSOR RIGHT keys and the numeric key­pad all can be used to control the center frequency. The ROTARY KNOB can be used for turning (clockwise for higher frequencies and counter-clockwise for lower frequencies) and clicking (the center frequency will be rounded to the nearest step value). The CURSOR LEFT / CURSOR RIGHT keys will change the center frequency by half the "SPAN" to lower or higher frequencies: after pressing the left arrow the center frequency will be reduced by half the "SPAN", and after pressing the right arrow it will be increased by half the "SPAN". The number keys allow direct entry of a new center frequency: If the focus is not on any dialog or softkey then the effect of just keying a number using the number keys will cause the receiver to use that particular number for the center fre­quency, provided the number is within the range of the receiver.
You can move the center frequency by left clicking into the spectrum. By right clicking into the spectrum, you can move the demodulation frequency to the clicked position. You can change the position of a marker in the spectrum by left dragging it.
Use the following softkeys to configure the video panorama:
Press the button "VIDEO TRACE" to select which demodulation is used for the video panorama. A related selector shows the available options.
See also DISPlay:MENU[:NAME] on page 220.
Press the button "FREQ INCR" to set the frequency increment for changing the receive frequency. You can choose between manual and automatic mode. If you select manual mode, the specified value will be used as increment. If you select "AUTO" mode, the step width of the currently focused panel will be used as increment.
See also [SENSe:]FREQuency:STEP[:INCRement]? on page 272.
Press the button "CONFIG" to configure the video panorama, e.g. to set the value range for the Y axis.
See also chapter 3.3.3.15, "Video Panorama Dialog", on page 179.
61Operating Manual 4072.8426.02 ─ 08
Page 62
R&S® EB500
Operation
Graphical User Interface (GUI)
3.3.2.10 Video Waterfall
Fig. 3-9: Video Waterfall.
The video waterfall below the video panorama shows signal levels from the video pan­orama over time. The signal levels are color coded; the colors correspond to the color set shown along the right margin. The time in seconds is indicated along the left margin.
You can change the position of a Y marker by left dragging it along the Y axis.
Just like in the video panorama (see chapter 3.3.2.9, "Video Panorama", on page 61 ) you can use the the ROTARY KNOB, the CURSOR LEFT / CURSOR RIGHT keys and the numeric keypad to control the center frequency.
Use the following softkeys to configure the video waterfall:
Press the button "VIDEO TRACE" to select which demodulation is used for the video panorama. A related selector shows the available options.
See also DISPlay:MENU[:NAME] on page 220.
Press the button "FREQ INCR" to set the frequency increment for changing the receive frequency. You can choose between manual and automatic mode. If you select manual mode, the specified value will be used as increment. If you select "AUTO" mode, the step width of the currently focused panel will be used as increment.
See also [SENSe:]FREQuency:STEP[:INCRement]? on page 272.
Press the button "CONFIG" to configure the video waterfall, i.e. to set the speed and color set for the waterfall.
See also chapter 3.3.3.16, "Video Waterfall Dialog", on page 179.
62Operating Manual 4072.8426.02 ─ 08
Page 63
R&S® EB500
Operation
Graphical User Interface (GUI)
3.3.2.11 ITU Measurement
Fig. 3-10: ITU Panel: Frequency deviation measurement.
Fig. 3-11: ITU Panel: Modulation depth measurement.
The ITU panel shows the following measurement output in accordance with the ITU guidelines:
level, field strength
modulation depth
frequency deviation
PM phase deviation, bandwidth
Press the button "DEMOD FREQ" to set the demodulation frequency, which is normally the same as the receive frequency. Any deviations will be temporary. The frequency limits of the demodulation frequency are determined by the receive frequency and the IF pan­orama span.
See also .[SENSe:]FREQuency:DEModulation on page 265.
Press the button "BW MEASMODE" to select the bandwidth measurement mode, i.e. "XDB" or "BETA%".
Press the "XDB" button to set the "XDB" measurement mode value. You can select a value between 0 dB and 100 dB.
Press the "BETA%" button to set the "BETA%" measurement mode value. You can select a value between 0.1% and 99.9%.
63Operating Manual 4072.8426.02 ─ 08
Page 64
R&S® EB500
Operation
Graphical User Interface (GUI)
Press the "RESET" button to reset all maximum ITU measurement values.
This button will toggle between AM measurement (modulation depth in percent) and FM measurement (frequency deviation in kHz).
Depending on the setting of the softkey "AM/FM MODE", the ITU measurement will dis­play either the Modulation Depth (AM) or the Frequency Deviation (FM). The modulation depth and frequency deviation are displayed digitally (values on the left-hand side) as well as in analog form (as bars). The related maxima are shown digitally (values on the right-hand side) and in analog form (as a vertical line within the bar representation). The PM phase deviation display is not active unless the PM modulation mode is switched on. The ITU panel requires option R&S EB500-IM.
Press the "MEAS MODE" button to select the measurement mode. The measurement mode can be set to "CONT" ("CONTinuous") or "PER" ("PERiodic").
See also chapter 4.5.9, "MEASure Subsystem", on page 227.
Press the "MEAS TIME" button to set the measurement time. You can set a concrete time for periodic measurements or you can set it to "AUTO", in which case it will be determined automatically.
See also chapter 4.5.9, "MEASure Subsystem", on page 227.
64Operating Manual 4072.8426.02 ─ 08
Page 65
R&S® EB500
Operation
Graphical User Interface (GUI)
3.3.2.12 Polar Panel
Fig. 3-12: Polar Panel.
The polar panel shows the bearing of a signal within a 360° azimuth dial. Invalid bearings are indicated by a gray bearing indicator. If no new valid bearing values arrive after a period of more than one minute, the polar panel will be stopped. In this case, a "Stopped" message will appear along with the time at which the panel was stopped. If persistence has been turned on, the bearing indicator will be shown as a gray dashed line throughout the persistence period. Once the persistence period has expired, it will be shown as a solid gray line.
The polar panel requires DF upgrade R&S EB500-DF.
Besides the bearing of the signal the following additional information is displayed:
The bearing value in digital form (large black indicator).
65Operating Manual 4072.8426.02 ─ 08
Page 66
R&S® EB500
Operation
Graphical User Interface (GUI)
If the "Azimuth" softkey is set to "Average" (see the softkey description below), the
small black indicator will show the standard deviation for the average from the last 10 measurements. If the "Azimuth" softkey is set to "Current", the small black indicator will not show any value, just three dashes (- - -).
If the "Azimuth" softkey is set to "Average", the blue indicator will show the average
from the last 10 measurements. If the "Azimuth" softkey is set to "Current", the blue indicator will show the previous valid bearing value.
The DF quality in digital and analog form with DF quality squelch.
The level in digital and analog form with DF level squelch.
The field strength in digital form.
The DF bandwidth (current panorama bandwidth, which depends on the step width
and selectivity).
Use the following softkeys to configure the polar panel:
66Operating Manual 4072.8426.02 ─ 08
Page 67
R&S® EB500
Operation
Graphical User Interface (GUI)
Press the "DF QUALITY" button to set the DF quality, which ranges from 0% to 100%.
Once this softkey has the focus you can change the value by using the ROTARY KNOB or numeric keys.
Press the "SPAN" button to set the IF panorama span.
A related vertical selector will pop-up, showing the possible values, which range from 1 kHz to 20 MHz.
See also [SENSe:]FREQuency:SPAN on page 270.
Press the "STEP" button to set the IF panorama step width. Step widths range from 625 mHz to 2 MHz.
When set to "AUTO", the step width is determined automatically depending on the IF panorama span. A related vertical selector will pop-up, showing the available options.
Press the "LEVEL SQU" button to set the DF squelch for DF averaging in "NORM" ("NORMal") or "GATE" mode.
See also MEASure:DFINder|DF:THReshold[:UPPer] on page 232 .
Press the "SQU MODE" button to select or turn off the squelch mode. If the DF squelch is turned off, it is not shown in the DF panorama or the polar panel.
A related selector shows the available modes.
Available modes are "OFF", "GATE" and "NORMAL".
See also MEASure:DFINder|DF:MODE on page 230 .
67Operating Manual 4072.8426.02 ─ 08
Page 68
R&S® EB500
Operation
Graphical User Interface (GUI)
Press the "AF DF MODE" button to set the application in FFM mode to direction finding ("DF") or measurement ("AF"). If you change the application mode from "Receiver" to "DF", the related mode is activated automatically. In AF mode you can turn on the poly­chrome IF panorama and use the PScan, provided the option R&S EB500-PS is installed.
See also MEASure:APPLication on page 228.
Press the "DF MEAS TIME" button to set the DF measurement time, which is independent of the global measurement time. You can use the ROLLKEY or numeric pad to set the measurement time to the desired value.
Press the "Selectivity" button to set the selectivity in the panorama. You can only set the desired selectivity. The actual selectivity that can be achieved depends on the combina­tion of step width and span. A related selector shows the available options. Available options are "AUTO", "NORMAL", "NARROW" and "SHARP".
See also CALCulate:IFPan:SELectivity on page 205 .
Press the "SHOW ON TOP" button to show either the geographic north or the vehicle axis on top of the wind rose.
If the "Azimuth" softkey is set to "Average", the average and standard deviation from the last 10 measurements will be calculated and shown in the panel.
If the "Azimuth" softkey is set to "Current", the previous valid bearing value will be shown instead of the average.
Press the "FREQ INCR" button to set the frequency increment for changing the receive frequency. You can choose between manual and automatic mode. If you select manual mode, the specified value will be used as increment. If you select "AUTO" mode, the step width of the currently focused panel will be used as increment.
Press the "CONFIG POLAR" button to configure the polar panel. You can configure the value range of the level Y axis. You can configure the following parameters: value range of the level Y axis, background of the wind rose, persistence of the DF beam and accuracy of the DF value.
See also chapter 3.3.3.19, "Polar Display Dialog", on page 185 .
68Operating Manual 4072.8426.02 ─ 08
Page 69
R&S® EB500
Operation
Graphical User Interface (GUI)
3.3.2.13 DF IF Panorama
Fig. 3-13: DF IF Panorama.
In FFM the DF IF panorama shows signals centered around the current receive fre­quency. The DF IF panorama bandwidth is determined by the frequency span setting (softkey "SPAN"). The dark green background indicates the DF bandwidth. The DF level threshold is indicated along the left margin. The color set for color indexing in an IF waterfall panel is shown along the right margin.
The DF IF panorama requires DF upgrade R&S EB500-DF.
The ROTARY KNOB, the CURSOR LEFT / CURSOR RIGHT keys and the numeric key­pad all can be used to control the center frequency. The ROTARY KNOB can be used for turning (clockwise for higher frequencies and counter-clockwise for lower frequencies) and clicking (the center frequency will be rounded to the nearest "STEP" value). The CURSOR LEFT / CURSOR RIGHT keys will change the center frequency with half the "SPAN" to lower or higher frequencies: after pressing the CURSOR LEFT key the center frequency will be reduced by half the "SPAN" and after pressing the CURSOR RIGHT key it will be increased by half the "SPAN". The number keys allow direct entry of a new center frequency: If the focus is not on any dialog or softkey then the effect of just keying a number using the number keys will cause the receiver to use that particular number for the center frequency, provided the number is within the range of the receiver.
If you connect a mouse and left click the yellow main marker on the X axis its ruler (the yellow line above the marker) will briefly appear. You can change the position of an active X marker by left dragging it. If you left drag the yellow main marker along the X axis, you can change the center frequency. By right dragging the yellow main marker along the X axis you can change the demodulation frequency.
If you left click the squelch marker (level threshold marker) on the Y axis its ruler will briefly appear. By left dragging the squelch marker along the Y axis, you can change its value. By left dragging a Y marker along the Y axis you can change its position.
You can move the center frequency by left clicking into the spectrum. By right clicking into the spectrum, you can move the demodulation frequency to the clicked position. You can change the position of a marker in the spectrum by left dragging it. You can change the center frequency by left dragging the yellow main marker in the spectrum. By right
69Operating Manual 4072.8426.02 ─ 08
Page 70
R&S® EB500
Operation
Graphical User Interface (GUI)
dragging the yellow main marker in the spectrum, you can change the demodulation frequency.
If you connect a keyboard then you can also change the center frequency using the cursor keys. The CURSOR RIGHT and CURSOR UP keys will move the center frequency by half the set "SPAN" to a higher frequency. The CURSOR LEFT and CURSOR DOWN keys will move the center frequency by half the set "SPAN" to a lower frequency.
If you press the mouse wheel or ROTARY KNOB, the current center frequency will be rounded to the step width. If you connect an external keyboard then SHIFT+<NUMERIC KEY> will round the center frequency to the corresponding decimal power, e.g. SHIFT +3 will round to kHz.
Use the following softkeys to configure the IF panorama:
Press the "DEMOD FREQ" button to set the demodulation frequency, which is normally the same as the receive frequency. Any deviations will be temporary, which means that the ROTARY KNOB will act upon the demodulation frequency for as long as the focus is on the "DEMOD FREQ" button. The frequency limits of the demodulation frequency are determined by the receive frequency and the IF panorama span.
See also [SENSe:]FREQuency:DEModulation on page 265.
Press the "AF DF MODE" button to set the application in FFM mode to direction finding ("DF") or measurement ("AF"). If you change the application mode from "Receiver" to "DF", the related mode is activated automatically. In AF mode you can turn on the poly­chrome IF panorama and use the PScan, provided the option R&S EB500-PS is installed.
See also MEASure:APPLication on page 228.
Press the "LEVEL SQU" button to set the DF squelch for DF averaging in "NORM" ("NORMal") or "GATE" mode.
See also MEASure:DFINder|DF:THReshold[:UPPer] on page 232 .
Press the "SQU MODE" button to select or turn off the squelch mode. If the DF squelch is turned off, it is not shown in the DF panorama or the polar panel.
A related selector shows the available modes.
Available modes are "OFF", "GATE" and "NORM" ("NORMal").
See also MEASure:DFINder|DF:MODE on page 230.
Press the button "DF MEAS TIME" to set the DF measurement time, which is independent of the global measurement time. You can use the ROLLKEY or numeric pad to set the measurement time to the desired value. In addition, there is a selector which allows you to select one of seven predefined time periods.
See also MEASure:DFINder|DF:TIME on page 231.
Press the "SELECTIVITY" button to set the selectivity in the IF panorama.
70Operating Manual 4072.8426.02 ─ 08
Page 71
R&S® EB500
Operation
Graphical User Interface (GUI)
You can only set the desired selectivity. The actual selectivity that can be achieved depends on the combination of step width and span.
A related vertical selector shows the available options. Available options are "AUTO", "NORMAL", "NARROW" and "SHARP."
Press the "SPAN" button to set the IF panorama span.
A related vertical selector will pop-up, showing the possible values, which range from 1 kHz to 20 MHz.
See also [SENSe:]FREQuency:SPAN on page 270.
Press the "STEP" button to set the IF panorama step width.
Step widths range from 625 mHz to 2 MHz.
When set to "AUTO", the step width is determined automatically depending on the IF panorama span. A related vertical selector will pop-up, showing the available options.
Press the "SEARCH PEAK" button to center the IF panorama to the next left or right level peak, or to the absolute level peak. You can choose either the center frequency or the demodulation frequency. Once the softkey is pressed, a new row of softkeys will appear, which show the available options.
See also:
CALCulate:IFPan:MARKer:MAXimum:LEFT on page 204
71Operating Manual 4072.8426.02 ─ 08
Page 72
R&S® EB500
Operation
Graphical User Interface (GUI)
CALCulate:IFPan:MARKer:MAXimum:RIGHt on page 204
CALCulate:IFPan:MARKer:MAXimum[:PEAK] on page 204
CALCulate:IFPan:MARKer:DEModulation:MAXimum:LEFT on page 204
CALCulate:IFPan:MARKer:DEModulation:MAXimum:RIGHt on page 204
CALCulate:IFPan:MARKer:DEModulation:MAXimum[:PEAK] on page 204
Press the "FFT MODE" button to select the FFT mode for the IF panorama data.
Once the softkey is pressed, a vertical selector will appear, which shows the available options. Available options are "MINIMUM", "MAXIMUM", "AVERAGE" and "CLRWRITE"
Press the "SELECTIVITY" button to set the selectivity in the IF panorama.
You can only set the desired selectivity. The actual selectivity that can be achieved depends on the combination of step width and span.
A related vertical selector shows the available options. Available options are "AUTO", "NORMAL", "NARROW" and "SHARP."
Press the "SPAN" button to set the IF panorama span.
A related vertical selector will pop-up, showing the possible values, which range from 1 kHz to 20 MHz.
See also [SENSe:]FREQuency:SPAN on page 270.
Press the "STEP" button to set the IF panorama step width.
72Operating Manual 4072.8426.02 ─ 08
Page 73
R&S® EB500
Operation
Graphical User Interface (GUI)
Step widths range from 625 mHz to 2 MHz.
When set to "AUTO", the step width is determined automatically depending on the IF panorama span. A related vertical selector will pop-up, showing the available options.
Press the "MEAS MODE" button to select the measurement mode. The measurement mode can be set to "CONT" ("CONTinuous") or "PER" ("PERiodic").
See also chapter 4.5.9, "MEASure Subsystem", on page 227.
Press the "MEAS TIME" button to set the measurement time. You can set a concrete time for periodic measurements or you can set it to "AUTO", in which case it will be determined automatically.
See also chapter 4.5.9, "MEASure Subsystem", on page 227.
Press the "POLYCHROME" button to operate the IF panorama in polychrome mode. You can set it to "OFF" or "HISTOGRAM" (color histogram). A related selector shows the available options.
If you select "HISTOGRAM", the frequency of the levels of the received signal measured over a defined time ("PERSISTENCE") will be shown in addition to the levels. Red dots indicate high occupancy (more frequently occurring levels) whereas blue dots indicate low occupancy (less frequently occurring levels).
Black areas indicate that no levels are being measured.
Fig. 3-14: Polychrome IF panorama.
See also CALCulate:PIFPan:MODE on page 208.
Press the "100% TIME" button to set the 100% time (the level occupancy required to indicate 100% in the color bar scale) for the polychrome IF panorama.
See also CALCulate:PIFPan:ACTTime on page 207.
73Operating Manual 4072.8426.02 ─ 08
Page 74
R&S® EB500
Operation
Graphical User Interface (GUI)
Press the "PERSISTENCE" button to set persistence of the polychrome panorama, i.e. the time during which the occupancy of a signal is to be measured.
See also CALCulate:PIFPan:OBSTime on page 209.
Press the "CLEAR POLYCHROME" button to clear the polychrome IF panorama; the measurement will restart.
See also CALCulate:PIFPan:CLEar on page 208.
Press the "LOW LIMIT" button to set the lower time limit for 100% time. Occupancies less than the lower limit will not be indicated with a color in the IF panel. You can set the lower limit to a value between 0 ms and the upper limit.
Press the "UPPER LIMIT" button to set the upper time limit for 100% time. Occupancies higher than the upper limit will not be indicated with a color in the IF panel. You can set the upper limit to a value between the lower limit and 100% time.
Press the "FREQ INCR" button to set the frequency increment for changing the receive frequency. You can choose between manual and automatic mode. If you select manual mode, the specified value will be used as increment. If you select "AUTO" mode, the step width of the currently focused panel will be used as increment.
Press the "DIFF MODE" button to turn on the "Differential" mode. When turned on, the current spectrum is used as the reference; only the changes with respect to the reference spectrum are displayed.
This mode is indicated by the words "Differential Mode" displayed at the top of the spec­trum. Note that the waterfall panel does not support differential mode.
Press the "CONFIG DFPAN" button to configure the DF IF panorama. You can configure the value range of the Y axis, the color set for the waterfall and the "Hold Maximum" function.
See also: chapter 3.3.3.17, "DF Panorama Dialog", on page 180.
3.3.2.14 DF IF Waterfall
Fig. 3-15: DF IF Waterfall.
The DF IF waterfall, which by default is displayed below the DF IF panorama, shows signal levels from the DF IF panorama over time. The signal levels are color coded; the colors correspond to the color set shown along the right margin. They correspond with the level in the DF IF panorama. The time in seconds is indicated along the left margin.
74Operating Manual 4072.8426.02 ─ 08
Page 75
R&S® EB500
Operation
Graphical User Interface (GUI)
The DF IF waterfall requires DF upgrade R&S EB500-DF.
Color Coding
With option R&S EB500-WDF the color coding of the DF IF Waterfall panel can be changed from Level to Azimuth. With Azimuth coding the panel shows the waterfall panel for the measured DF angles, the colors representing the azimuth values as indicated in the color bar at the left of the azimuth panel.
Fig. 3-16: DF Waterfall panel with Azimuth color coding.
Use the Color Coding softkey (see below) to switch between level and azimuth colors.
You can change the position of a Y marker by turning the SELECT rollkey or by left dragging it along the Y axis (in case a mouse is connected).
If you want to change the center frequency by using a mouse, left click into the waterfall. Likewise, if you want to change the demodulation frequency, right click into the waterfall. You can also left drag a marker in the waterfall to change its position. You can change the center frequency by left dragging the yellow main marker on the X axis. By right dragging the yellow main marker on the X axis, you can change the demodulation fre­quency. If the IF waterfall is displayed in a single-panel layout (SETUP > "Layout" > "IF WF ONLY") then there will be no X axis and no marker.
Press the mouse wheel or ROTARY KNOB to round the current center frequency to the step width.
Use the following softkeys to configure the IF waterfall:
75Operating Manual 4072.8426.02 ─ 08
Page 76
R&S® EB500
Operation
Graphical User Interface (GUI)
Press the COL CODING button to change the color coding of the waterfall. The color coding option only makes sense when the azimuth reading is available, which requires option R&S EB500-WDF. See "Color Coding" on page 75 for an example of azimuth color coding.
Press the "AF DF MODE" button to set the application in FFM mode to direction finding ("DF") or measurement ("AF"). If you change the application mode from "Receiver" to "DF", the related mode is activated automatically. In AF mode you can turn on the poly­chrome IF panorama and use the PScan, provided the option R&S EB500-PS is installed.
See also MEASure:APPLication on page 228.
Press the "LEVEL SQU" button to set the DF squelch for DF averaging in "NORM" ("NORMal") or "GATE" mode.
See also MEASure:DFINder|DF:THReshold[:UPPer] on page 232.
Press the "SQU MODE" button to select or turn off the squelch mode. If the DF squelch is turned off, it is not shown in the DF panorama or the polar panel.
A related selector shows the available modes.
Available modes are "OFF", "GATE" and "NORM" ("NORMal").
See also MEASure:DFINder|DF:MODE on page 230.
Press the button "DF MEAS TIME" to set the DF measurement time, which is independent of the global measurement time. You can use the ROLLKEY or numeric pad to set the measurement time to the desired value. In addition, there is a selector which allows you to select one of seven predefined time periods.
See also MEASure:DFINder|DF:TIME on page 231.
Press the "SELECTIVITY" button to set the selectivity in the IF panorama. You can only set the desired selectivity. The actual selectivity that can be achieved depends on the combination of step width and span.
A related vertical selector shows the available options. Available options are "AUTO", "NORMAL", "NARROW" and "SHARP."
76Operating Manual 4072.8426.02 ─ 08
Page 77
R&S® EB500
Operation
Graphical User Interface (GUI)
Press the "SPAN" button to set the IF panorama span.
A related vertical selector will pop-up, showing the possible values, which range from 1 kHz to 20 MHz.
See also [SENSe:]FREQuency:SPAN on page 270.
Press the "STEP" button to set the IF panorama step width. Step widths range from 625 mHz to 2 MHz.
When set to "AUTO", the step width is determined automatically depending on the IF panorama span. A related vertical selector will pop-up, showing the available options.
Press the "FFT MODE" button to select the FFT mode for the IF panorama data.
Once the softkey is pressed, a vertical selector will appear, which shows the available options. Available options are "MINIMUM", "MAXIMUM", "AVERAGE" and "CLRWRITE"
Press the "MEAS MODE" button to select the measurement mode. The measurement mode can be set to "CONT" ("CONTinuous") or "PER" ("PERiodic").
See also chapter 4.5.9, "MEASure Subsystem", on page 227.
Press the "MEAS TIME" button to set the measurement time. You can set a concrete time for periodic measurements or you can set it to "AUTO", in which case it will be determined automatically.
See also chapter 4.5.9, "MEASure Subsystem", on page 227.
77Operating Manual 4072.8426.02 ─ 08
Page 78
R&S® EB500
Operation
Graphical User Interface (GUI)
Press the "POLYCHROME" button to operate the IF panorama in polychrome mode. You can set it to "OFF" or "HISTOGRAM" (color histogram). A related selector shows the available options.
If you select "HISTOGRAM", the frequency of the levels of the received signal measured over a defined time ("PERSISTENCE") will be shown in addition to the levels. Red dots indicate high occupancy (more frequently occurring levels) whereas blue dots indicate low occupancy (less frequently occurring levels).
Black areas indicate that no levels are being measured.
Fig. 3-17: Polychrome IF panorama.
See also CALCulate:PIFPan:MODE on page 208.
Press the "100% TIME" button to set the 100% time (the level occupancy required to indicate 100% in the color bar scale) for the polychrome IF panorama.
See also CALCulate:PIFPan:ACTTime on page 207.
Press the "PERSISTENCE" button to set persistence of the polychrome panorama, i.e. the time during which the occupancy of a signal is to be measured.
See also CALCulate:PIFPan:OBSTime on page 209.
Press the "CLEAR POLYCHROME" button to clear the polychrome IF panorama; the measurement will restart.
See also CALCulate:PIFPan:CLEar on page 208.
Press the "LOW LIMIT" button to set the lower time limit for 100% time. Occupancies less than the lower limit will not be indicated with a color in the IF panel. You can set the lower limit to a value between 0 ms and the upper limit.
Press the "UPPER LIMIT" button to set the upper time limit for 100% time. Occupancies higher than the upper limit will not be indicated with a color in the IF panel. You can set the upper limit to a value between the lower limit and 100% time.
Press the "FREQ INCR" button to set the frequency increment for changing the receive frequency. You can choose between manual and automatic mode. If you select manual mode, the specified value will be used as increment. If you select "AUTO" mode, the step width of the currently focused panel will be used as increment.
Press the "DIFF MODE" button to turn on the "Differential" mode. When turned on, the current spectrum is used as the reference; only the changes with respect to the reference spectrum are displayed.
78Operating Manual 4072.8426.02 ─ 08
Page 79
R&S® EB500
Operation
Graphical User Interface (GUI)
This mode is indicated by the words "Differential Mode" displayed at the top of the spec­trum. Note that the waterfall panel does not support differential mode.
Press the "CONFIG DFPAN" button to configure the DF IF panorama. You can configure the value range of the Y axis, the color set for the waterfall and the "Hold Maximum" function.
See also: chapter 3.3.3.18, "DF Waterfall Dialog", on page 183.
3.3.2.15 Azimuth Panel
Fig. 3-18: Azimuth panel.
The azimuth panel shows bearing angles dependent on the frequency. An azimuth ruler and cross-hairs cursor allows you to select a signal manually or automatically. See "RULER MODE" softkey. The color set for color indexing in the DF IF waterfall panel (see
chapter 3.3.2.14, "DF IF Waterfall", on page 74) is shown along the left margin. The
waterfall does not show the bearing values unless the "COL CODING" softkey has been set to AZIMUTH. The signal age is shown along the right margin. White stands for current signals. The signal age is updated every second and indicated by a darker color. After four seconds, the signals are removed from the display.
The azimuth panel requires option R&S EB500-WDF.
The ROTARY KNOB, the CURSOR LEFT / CURSOR RIGHT keys and the numeric key­pad all can be used to control the center frequency. The ROTARY KNOB can be used for turning (clockwise for higher frequencies and counter-clockwise for lower frequencies) and clicking (the center frequency will be rounded to the nearest "STEP" value). The CURSOR LEFT / CURSOR RIGHT keys will change the center frequency with half the "SPAN" to lower or higher frequencies: after pressing the CURSOR LEFT key the center frequency will be reduced by half the "SPAN" and after pressing the CURSOR RIGHT
79Operating Manual 4072.8426.02 ─ 08
Page 80
R&S® EB500
Operation
Graphical User Interface (GUI)
key it will be increased by half the "SPAN". The number keys allow direct entry of a new center frequency: If the focus is not on any dialog or softkey then the effect of just keying a number using the number keys will cause the receiver to use that particular number for the center frequency, provided the number is within the range of the receiver.
If you connect a mouse and left click the azimuth marker on the Y axis its ruler will briefly appear. You can change the value of the azimuth marker by left dragging it. You can change the position of a Y marker by left dragging it along the Y axis.
You can move the center frequency by left clicking into the azimuth panel. By right clicking into the azimuth panel, you can move the demodulation frequency to the clicked position. You can change the position of a marker in the azimuth panel by left dragging it. You can change the center frequency by left dragging the yellow main marker in the azimuth panel. By right dragging the yellow main marker in the azimuth panel, you can change the demodulation frequency.
If you connect a keyboard then you can also change the center frequency using the cursor keys. The CURSOR RIGHT and CURSOR UP keys will move the center frequency by half the set "SPAN" to a higher frequency. The CURSOR LEFT and CURSOR DOWN keys will move the center frequency by half the set "SPAN" to a lower frequency.
Use the following softkeys to configure the azimuth panel:
The "Ruler Mode" softkey can be used to set the azimuth ruler either in Auto (automatic) or Manual mode. In Auto mode there will only be a cross-hairs cursor at the azimuth of the demod frequency.
In Manual mode the cross-hairs cursor can be set manually by left dragging the cursor in the y-direction. In Manual mode there is also a marker on the y-axis. Dragging the marker along the y-axis will move the cursor. A ruler will shortly appear when the cursor position has changed.
Press the "AF DF MODE" button to set the application in FFM mode to direction finding ("DF") or measurement ("AF"). If you change the application mode from "Receiver" to "DF", the related mode is activated automatically. In AF mode you can turn on the poly­chrome IF panorama and use the PScan, provided the option R&S EB500-PS is installed.
See also MEASure:APPLication on page 228.
Press the "LEVEL SQU" button to set the DF squelch for DF averaging in "NORM" ("NORMal") or "GATE" mode.
See also MEASure:DFINder|DF:THReshold[:UPPer] on page 232 .
Press the "SQU MODE" button to select or turn off the squelch mode. If the DF squelch is turned off, it is not shown in the DF panorama or the polar panel.
A related selector shows the available modes.
Available modes are "OFF", "GATE" and "NORM" ("NORMal").
See also MEASure:DFINder|DF:MODE on page 230.
80Operating Manual 4072.8426.02 ─ 08
Page 81
R&S® EB500
Operation
Graphical User Interface (GUI)
Press the button "DF MEAS TIME" to set the DF measurement time, which is independent of the global measurement time. You can use the ROLLKEY or numeric pad to set the measurement time to the desired value. In addition, there is a selector which allows you to select one of seven predefined time periods.
See also MEASure:DFINder|DF:TIME on page 231.
Press the "SELECTIVITY" button to set the selectivity in the IF panorama.
You can only set the desired selectivity. The actual selectivity that can be achieved depends on the combination of step width and span.
A related vertical selector shows the available options. Available options are "AUTO", "NORMAL", "NARROW" and "SHARP."
Press the "SPAN" button to set the IF panorama span.
A related vertical selector will pop-up, showing the possible values, which range from 1 kHz to 20 MHz.
See also [SENSe:]FREQuency:SPAN on page 270.
Press the "STEP" button to set the IF panorama step width.
Step widths range from 625 mHz to 2 MHz.
When set to "AUTO", the step width is determined automatically depending on the IF panorama span. A related vertical selector will pop-up, showing the available options.
81Operating Manual 4072.8426.02 ─ 08
Page 82
R&S® EB500
Operation
Graphical User Interface (GUI)
Press the "FREQ INCR" button to set the frequency increment for changing the receive frequency. You can choose between manual and automatic mode. If you select manual mode, the specified value will be used as increment. If you select "AUTO" mode, the step width of the currently focused panel will be used as increment.
The "Azimuth" softkey is basically an indicator for the azimuth at the marker position.
3.3.2.16 DDC Display
Fig. 3-19: DDC Display
The DDC display shows the configuration for each of the four digital down converters (DDCs) of the R&S EB500. The DDC currently in use is indicated by the yellow bar along the top. It contains specific icons showing the current configuration and state of the DDC:
The DDC is shown in the spectrum.
The DDC is not shown in the spectrum.
The DDC audio is turned on.
The DDC audio is turned off.
The demodulation of the DDC is coupled to the demodulation of the primary receive path.
The demodulation of the DDC is independent of the demodulation in the primary receive path.
The frequency of the DDC is coupled to the frequency of the primary receive path.
The frequency of the DDC is independent of the frequency in the primary receive path.
The DDC data is being recorded.
Note
Digital Down Converter functions are only accessible with the installed software option R&S EB500-DDC.
A configuration comprises the following parameters:
frequency
82Operating Manual 4072.8426.02 ─ 08
Page 83
R&S® EB500
Operation
Graphical User Interface (GUI)
demodulation bandwidth
demodulation mode
squelch
volume and balance
visibility in the IF panorama
indication of the DDC's coupling to the center frequency
indication of the DDC's demodulation coupling to the demodulation of the primary
receive path.
The DDC currently being configured is highlighted in yellow. Once you have activated the panel, you can select the desired DDC using the CURSOR LEFT / CURSOR RIGHT keys. Use the ROLLKEY to change the receive frequency of the selected DDC.
See also [SENSe:]FREQuency:DDC[<numeric_suffix>] on page 488.
Use the BANDWIDTH keys to set the demodulation bandwidth.
See also[SENSe:]BANDwidth|BWIDth[:RESolution]:
DDC[<numeric_suffix>] on page 486 .
If you want to change the squelch or turn it off or on, use the SQU key together with the SELECT knob.
See also OUTPut:SQUelch:DDC[<numeric_suffix>][:STATe] on page 484.
If you want to change the volume or mute / unmute it, use the VOLUME / MUTE knob.
See also:
SYSTem:AUDio:DDC[<numeric_suffix>]:VOLume on page 492
SYSTem:AUDio:DDC[<numeric_suffix>][:STATe] on page 493
Use the configuration dialog of the DDC display to configure the volume, balance and visibility for all the DDC channels. The volume and balance are visualized via two bars marked "L" and "R". They represent the volume of the left and right channel, respectively. If you have the sound turned off, the speaker icon in the upper bar is crossed out. The same applies to the visibility icon. If a DDC is shown in the IF panorama, an icon with an eye is displayed in the upper bar. If the DDC is not shown, the eye is crossed out.
If you press the MEM > "RCL" and MEM > "Save" keys while the DDC panel is selected, the keys will affect the frequency of the current DDC rather than the center frequency (see "Recall to DDC frequencies" on page 121 and "Save DDC frequencies" on page 123).
The measured level of the related DDC is shown digitally (by value) and in analog form (represented by a bar). If no valid measurement values are available, no value is dis­played and the bar is not filled.
83Operating Manual 4072.8426.02 ─ 08
Page 84
R&S® EB500
Operation
Graphical User Interface (GUI)
You can also use the mouse to control the DDC display. Select a DDC by clicking into the desired DDC's field. Once a DDC has been selected, you can change its state. If you click the visibility icon you can turn visibility on or off. The DDC channels are visible in the IF panorama by means of dark-red bars and numbered markers, representing the DDC channels with the bandwidth for each channel (see the figure below).
Fig. 3-20: IF Panorama with all three DDC channels visible.
Clicking the speaker icon turns the audio on or off. Clicking the frequency or demodulation coupling icon activates or deactivates the coupling. Clicking the line for the DDC's demodulation bandwidth sets the next possible bandwidth for this particular DDC. The DDC's demodulation mode is set likewise. Clicking the squelch line turns the squelch on or off. When a squelch has been turned on, it is shown as a white marker in the level indicator. You can change it via drag & drop using the left mouse key.
The DDC display can be configured with the following softkeys:
Press the "DEMOD FREQ" button to set the demodulation frequency, which is normally the same as the receive frequency. Any deviations will be temporary, which means that the ROTARY KNOB will act upon the demodulation frequency for as long as the focus is on the "DEMOD FREQ" button. The frequency limits of the demodulation frequency are determined by the receive frequency and the IF panorama span.
See also [SENSe:]FREQuency:DEModulation on page 265.
Press the "SPAN" button to set the panorama span.
A related vertical selector will pop-up, showing the possible values, which range from 1 kHz to 20 MHz.
See also [SENSe:]FREQuency:SPAN on page 270.
84Operating Manual 4072.8426.02 ─ 08
Page 85
R&S® EB500
Operation
Graphical User Interface (GUI)
Press the "STEP" button to set the step width. Step widths range from 625 mHz to 2 MHz.
When set to "AUTO", the step width is determined automatically depending on the span. A related vertical selector will pop-up, showing the available options.
After pressing the "SEARCH DDC PEAK" button the following row of softkeys will appear.
The three leftmost options allow you to set the currently selected DDC to the next left, right or absolute level peak, respectively.
The "LEFT NEXT" button will allocate the peak left of the absolute level peak in the span to the first DDC channel that is not within the span, starting with DDC1.
The "RIGHT NEXT" button will allocate the peak right of the absolute level peak in the span to the first DDC channel that is not within the span, starting with DDC1.
Once all DDC channels are within the span, the "LEFT NEXT" and "RIGHT NEXT" buttons have no more effect.
See also:
CALCulate:IFPan:MARKer:DDC[<numeric_suffix>]:MAXimum:LEFT
on page 484
CALCulate:IFPan:MARKer:DDC[<numeric_suffix>]:MAXimum:RIGHt
on page 484
CALCulate:IFPan:MARKer:DDC[<numeric_suffix>]:MAXimum[:PEAK]
on page 483
CALCulate:IFPan:MARKer:DDC:NEXT:MAXimum:LEFT on page 484
CALCulate:IFPan:MARKer:DDC:NEXT:MAXimum:RIGHt on page 484
Press the "SEARCH PEAK" button to center the IF panorama to the next left or right level peak, or to the absolute level peak. You can choose either the center frequency or the demodulation frequency. Once the softkey is pressed, a new row of softkeys will appear, which show the available options.
See also:
85Operating Manual 4072.8426.02 ─ 08
Page 86
R&S® EB500
Operation
Graphical User Interface (GUI)
CALCulate:IFPan:MARKer:MAXimum:LEFT on page 204
CALCulate:IFPan:MARKer:MAXimum:RIGHt on page 204
CALCulate:IFPan:MARKer:MAXimum[:PEAK] on page 204
CALCulate:IFPan:MARKer:DEModulation:MAXimum:LEFT on page 204
CALCulate:IFPan:MARKer:DEModulation:MAXimum:RIGHt on page 204
CALCulate:IFPan:MARKer:DEModulation:MAXimum[:PEAK] on page 204
Press the "FREQ UNCOUPLED" / "FREQ COUPLED" softkey to activate or deactivate the coupling of the current DDC's frequency to the center frequency. If "COUPLED", the DDC's frequency will change along with the center frequency. If "UNCOUPLED", the frequency of the current DDC will not be affected by the center frequency.
Press "DEMOD UNCOUPLED" / "DEMOD COUPLED" softkey to activate or deactivate the coupling of the current DDC's demodulation to the primary receive path. If "COU­PLED", the demodulation bandwidth, demodulation mode and squelch of the DDC will change along with the settings of the primary receive path. If "UNCOUPLED", the current DDC's demodulation will not be affected by changes in the primary receive path.
The "COPY DDC > RX" softkey copies the center frequency to the receive frequency of the current DDC.
The "COPY RX > DDC" softkey copies the receive frequency of the current DDC to the center frequency.
Press the "SELECTIVITY" button to set the selectivity in the panorama. You can only set the desired selectivity. The actual selectivity that can be achieved depends on the com­bination of step width and span.
A related vertical selector shows the available options. Available options are "AUTO", "NORMAL", "NARROW" and "SHARP".
See also CALCulate:IFPan:SELectivity on page 205.
Press the "FFT MODE" button to select the FFT mode for the panorama data.
Once the softkey is pressed, a vertical selector will appear, which shows the available options. Available options are "MINIMUM", "MAXIMUM", "AVERAGE" and "CLRWRITE"
See also CALCulate:IFPan:AVERage:TYPE on page 202.
Press the "FREQ INCR" button to set the frequency increment for changing the receive frequency. You can choose between manual and automatic mode. If you select manual mode, the specified value will be used as increment. If you select "AUTO" mode, the step width of the currently focused panel will be used as increment.
86Operating Manual 4072.8426.02 ─ 08
Page 87
R&S® EB500
Operation
Graphical User Interface (GUI)
See also [SENSe:]FREQuency:STEP[:INCRement] on page 272.
Press the "SHOW ON" / "SHOW OFF" softkey to show or hide the selected DDC in the IF panorama. Set it to "ON" if you want it to be shown. An icon with an eye is displayed in the related DDC field and the DDC channels are shown in the IF panorama provided they are within the span.
Press the "SELECT" button to select the DDC to be configured. A related selector shows the available DDCs.
Press "CONFIG DDC" to open the "DDC Display" dialog. You can specify whether or not the demodulation bandwidth of the individual DDCs should be displayed in the IF pan­orama. You can also make all DDCs visible simultaneously and change their coupling status. In addition, you can set the volume and balance of the individual DDCs.
3.3.2.17 Selective Call
Fig. 3-21: Selective Call panel.
The Selective Call panel shows the decoded selective call and demodulated pager data. While the list on the left shows all data, the list on the right shows only filtered data. The Selective Call panel requires option R&S EB500-SL.
Use the LEFT or RIGHT cursor key to select a particular list. Within a list, you can scroll through the list entries using the UP or DOWN cursor key. If the last entry in a list is selected, automatic scrolling is activated, i.e. every new selective call will be selected automatically and the list will be scrolled down. If the last entry is not selected, automatic scrolling is deactivated and a new selective call will not cause any scrolling.
Use the mouse wheel or ROLLKEY to scroll the selected list up or down.
If you press the mouse wheel or ROLLKEY, the selected list will be switched.
Use the following softkeys to configure the Selective Call panel:
Press the "DEMOD FREQ" button to set the demodulation frequency, which is normally the same as the receive frequency. Any deviations will be temporary, which means that the ROTARY KNOB will act upon the demodulation frequency for as long as the focus is on the "DEMOD FREQ" button. The frequency limits of the demodulation frequency are determined by the receive frequency and the IF panorama span.
87Operating Manual 4072.8426.02 ─ 08
Page 88
R&S® EB500
Operation
Graphical User Interface (GUI)
See also [SENSe:]FREQuency:DEModulation on page 265.
Press the "FILTER" button to select one or more selective call methods as filter criteria. Filtered selective calls are additionally shown in the right-hand list. Once the filter has been switched on, a new set of softkeys with the different filter methods provided.
Available methods are "CCIR 1", "CCIR 7", "CCITT", "EEA", "EIA", "EURO", "NATEL", "VDEW", "ZVEI-1", "ZVEI-2", "DTMF", "CTCSS" and "DCS".
Toggling the "SELECT" button between "LEFT" and "RIGHT" changes the highlight between the "LEFT" list (Selective Calls) and "RIGHT" list (Filtered Selective Calls). Scrolling with the ROLLKEY will only act upon the list with the highlight.
The "CLEAR LEFT" button will clear the left-hand list of all entries.
The "CLEAR RIGHT" button will clear the right-hand list of all entries.

3.3.3 Using Front-Panel Keys and Dialogs

This chapter contains a detailed description of the front-panel keys of the R&S EB500 and how to use the associated dialogs.
3.3.3.1 BANDWIDTH
This key is used to set the demodulation bandwidth. Pressing the left side of the key (BW-) will decrease the bandwidth, whereas pressing the right side of the key (BW+) will increase the bandwidth. At the same time, when the key is pressed, a vertical selector will open and the desired demodulation bandwidth can be chosen directly by moving up and down.
Available bandwidths are 0.1 kHz, 0.15 kHz, 0.3 kHz, 0.6 kHz, 1 kHz, 1.5 kHz, 2.1 kHz,
2.4 kHz, 2.7 kHz, 3.1 kHz, 4 kHz, 4.8 kHz, 6 kHz, 9 kHz, 12 kHz, 15 kHz, 30 kHz, 50 kHz,
120 kHz, 150 kHz, 250 kHz, 300 kHz, 500 kHz, 800 kHz, 1000 kHz, 1250 kHz, 1500 kHz, 2000 kHz, 5000 kHz, 8000 kHz, 10,000 kHz, 12,500 kHz, 15,000 kHz and 20,000 kHz.
The selector will close after it has been idle for more than 3 sec. While the dialog is open, the user can scroll through the available bandwidth settings and while scrolling the high­lighted value will be adopted by the receiver so the effect is immediate. Scrolling can be done by pressing BW+ or BW-. The main rollkey can also be used for scrolling. When an arrow at the top or bottom of the selector is highlighted, it indicates that more values can be selected for the demodulation bandwidth.
See also[SENSe:]BANDwidth|BWIDth[:RESolution] on page 253 .
88Operating Manual 4072.8426.02 ─ 08
Page 89
R&S® EB500
Operation
Graphical User Interface (GUI)
3.3.3.2 DEMODULATION
This key is used to set the demodulation mode. Pressing the left side (DEMOD-) or right side (DEMOD+) of the key will change the demodulation from a list of options in reverse or forward directions. At the same time, when the key is pressed, a vertical selector will open and the desired demodulation mode can be chosen directly by moving up and down.
The vertical selector will show the available demodulation options: CW, USB, LSB, ISB, AM, FM, PULS, PM and IQ.
The selector will close after it has been left idle for more than 3 sec. While the selector is open, the user can scroll through the available demodulation settings. The highlighted value will be adopted by the receiver so the effect is immediate. Demodulation modes which are not available for the current receiver setting are shown with a dark background. Scrolling can be done by the DEMOD- or DEMOD+ key. The main rollkey can also be used for scrolling. When an arrow at the top or bottom of the selector is highlighted, it indicates that more modes can be selected for demodulation. From the figure on the left, it can be seen that there are more demodulation modes available when the user scrolls down to below PULS and that the demodulation modes CW, LSB and USB are not avail­able with the current receiver settings.
See also [SENSe:]FREQuency:DEModulation? on page 266.
3.3.3.3 DETECT
This key is used to set the level measurement mode. Once the DET key is pressed the softkey bar will be replaced by a set of softkeys which show the modes that are available. This will you to set a mode directly. The available modes are: "AVG", "PEAK", "FAST" and "RMS".
3.3.3.4 MARK
The MARK key is used to open a new set of softkeys which allows you to activate or deactivate the markers in the panels. You can also configure the way the markers are displayed.
See chapter 3.3.2.4, "Markers", on page 40 for a detailed explanation on how to operate the markers.
3.3.3.5 APPL
This key is used to set the receiver to a particular application mode. The default softkey bar will be replaced by a set of softkeys which show the modes available. Available application modes are "RECEIVER", "VIDEO", "ITU", "DF""DDC", "WDF" and "Decode".
Application mode ITU requires option R&S EB500-IM.
Application mode DF requires option R&S EB500-DF.
89Operating Manual 4072.8426.02 ─ 08
Page 90
R&S® EB500
Operation
Graphical User Interface (GUI)
Application mode DDC requires option R&S EB500-DDC.
Application mode WDF requires option R&S EB500-WDF. A prerequisite is DF Upgrade R&S EB500-DF.
Application mode Decode requires option R&S EB500-SL.
Application Mode WDF
The DF application mode is a subset of the WDF application mode. With R&S EB500­WDF installed all the DF-related features are accessible through the WDF application mode.
Application modes in Basic configuration:
Application modes with ITU option R&S EB500-IM and DF upgrade R&S EB500-DF installed.
Application modes with DF upgrade R&S EB500-DF and options R&S EB500-WDF, R&S EB500-DDC and R&S EB500-SL installed.
3.3.3.6 MODE
This key is used to open a selector which allows a user to set the scan mode of the R&S EB500 and to provide the controls required to run and stop a scan. Available modes are "FFM" and "SCAN" as well as a number of options depending on the type of scan.
When the user chooses SCAN he will be provided with a vertical selector which gives the three scan options PScan, FScan and MScan.
In case of PScan, there are only two options for controlling the scan: "Run" and "Stop", so the softkeys look like the figure below.
In case of FScan, there are more ways to control the scan: "Run-" to scan in the direction of lower frequencies, "Stop", "Run+" to scan in the direction of higher frequencies and "Restart" to restart the scanning from the start frequency in the direction of higher fre­quencies. The softkeys look like the figure below.
90Operating Manual 4072.8426.02 ─ 08
Page 91
R&S® EB500
Operation
Graphical User Interface (GUI)
In case of MScan, the options are the same as for FScan, except for the "Restart" option; the softkeys therefore look like the figure below.
Panorama Scan
The Panorama Scan (PScan) mode requires option R&S EB500-PS.
See also .[SENSe:]FREQuency:MODE on page 266.
3.3.3.7 MENU
The MENU key is used to select a choice of menu options to provide fast access to a number of utility functions of the R&S EB500. A new set of softkeys provides access to the available configuration options.
Below follows a short description of all the MENU softkeys.
A key-press on the "SNAPSHOT" key will take a snapshot of the full screen, including softkeys, dialogs etc.
Keyboard shortcut
There are certain GUI elements which cannot be captured with the "Snapshot" key. These are the softkeys, dialogs etc. under the MENU key (because the "Snapshot" key is also under the MENU key). For those cases, if you want to take a snapshot, you can plug-in a keyboard and use the combination Alt + Shift + P.
A fixed duration audio recording will open a dialog for audio recording. With the option R&S EB500-DDC installed, the dialog also provides the option to record audio and IQ data for the individual DDC channels.
For a detailed explanation, see "RECORD" on page 92.
Opens the "PRESET" softkey bar, which allows you to choose between different preset configurations or to save a configuration as one of the PRESET options; see: "Preset
Softkeys" on page 95 for more details.
The "Zoom +" key zooms in into an RF spectrum. The actual zoom can be configured with the markers: If markers are active then the function will zoom-in between the active markers.
The "Zoom -" key will zoom out, after a zoom-in has taken place. If no zoom-in was performed than pressing the key will not have any effect.
91Operating Manual 4072.8426.02 ─ 08
Page 92
R&S® EB500
Operation
Graphical User Interface (GUI)
The "Tests" key opens the "TESTS" softkey bar, which allows you to run a long or short self-test of the receiver hardware, review the test results and generate a test report.
See: "Test Points Dialog" on page 96 for more details.
The "File" key opens the "FILE" softkey bar, which has an "Import" function for importing profiles, memory and suppress lists from a USB flash drive. It also has an "Export" func­tion for exporting profiles, memory / suppress lists, test reports, snapshots etc to a USB flash drive.
For a detailed explanation, see "FILE" on page 98
The "Sanitizing" key will start a sanitizing process that will clear all the user data.
The Sanitizing function will be explained in detail in "Sanitizing" on page 113.
RECORD
This key is used to open the "Recording" dialog. This dialog is used to record. The dura­tion of a recording during replay is visualized by a progress bar at the bottom of the dialog.
Fig. 3-22: Recording Dialog.
92Operating Manual 4072.8426.02 ─ 08
Page 93
R&S® EB500
Operation
Graphical User Interface (GUI)
Use this field to control the recording sum. Press "Record" when the field is highlighted and the recording will start. The audio data recorded is stored in the folder indicated above.
The recording duration is displayed next to the indicator.
Use the "STOP" softkey to stop the recording.
The audio data recorded is saved into a "wav" file in the selected destination folder.
Recording from DDC channels
In order to start recording from one of the DDC channels, highlight the particular channel and start recording in a similar manner as when recording Sum Audio. DDC channels also allow recording of IQ data.
As shown in the figure above, a different set of softkeys will appear when one of the DDC fields is selected. These softkeys allow you to select recording of audio data, IQ data or both. The audio / IQ data recorded is stored in the folder above in separate "wav" files. Recording for DDC2 and DDC3 operates likewise.
The DDC fields are not available unless option R&S EB500-DDC is installed.
93Operating Manual 4072.8426.02 ─ 08
Page 94
R&S® EB500
Operation
Graphical User Interface (GUI)
The file browser lets you specify the destination folder for all new recordings:
Activate the browser by pressing the ENTER key. All recordings are stored Recording. You can create subfolders in Recording, using the file management util­ities (see ). You can now navigate within the browser. The files are represented in a tree­like structure. If you select a collapsed ("+") node with the mouse wheel or ROLLKEY, the node will expand. If it is an expanded ("-") node, it will collapse.
Use the CURSOR UP or CURSOR DOWN key to select an entry in the browser. Use the CURSOR LEFT or CURSOR RIGHT key to collapse or expand a node, respectively, and to scroll left or right. Press the ESC key or ENTER key to deactivate the browser. To the right of the browser is a list which shows all the "WAV" files stored in the currently selected folder:
In addition to the file names, the list shows the recording parameters for each file, i.e. the frequency, bandwidth and demodulation mode.
94Operating Manual 4072.8426.02 ─ 08
Page 95
R&S® EB500
Operation
Graphical User Interface (GUI)
Preset Softkeys
Some menu items provide access to a layer of submenu items; "Preset" is one of them. The respective options for "Preset" are shown below.
A short description of the different softkeys will follow.
Press the "Reset" key and the R&S EB500 main processor will reset. The GUI will freeze until the R&S EB500 main processor is running again.
Up to four configurations can be stored as Preset configurations. The Preset configura­tions are stored at the client side so each GUI, whether internal or external, can have its own preset configurations.
Press the "Default" key and the receiver will reload all its default values.
When pressing the "Last" key the R&S EB500 will reload its last configuration, i.e. the configuration before loading a profile or before a reset.
The "Config" key will open up a dialog for configuring (adding or deleting) a profile; see
"Profile Management Dialog" on page 95.
Profile Management Dialog
This dialog is used to save certain configurations as a preset profile and to manage or delete one or more profiles.
The table gives an overview of the different profiles. With the forward / backward selector keys you can navigate through the table and highlight an item for editing.
95Operating Manual 4072.8426.02 ─ 08
Page 96
R&S® EB500
Operation
Graphical User Interface (GUI)
This edit field will allow you to directly key-in the line to edit. This line is highlighted in the table.
The name typed in this field will appear in the softkey label, after the index just as shown below. The maximum length is 8 characters.
The description will only appear in the dialog; it can be of any length however.
There are three types of profile that can be saved, selectable by means of a vertical selector which will pop-up once this field gets highlighted.
Client: This profile contains only GUI-related settings such as layouts, snapshot
directory and the local host IP address. Receiver: This profile holds receiver-related settings, e.g. the receive frequency,
bandwidth, demodulation mode and start and stop limits
Memory: This profile holds the receiver's 10 000 memory locations.
You can also choose to select all profiles: All three profile types will be saved into one profile.
Once you click "Save" the profile will be saved.
Use the "Delete" button to delete one profile entry and the "Delete All" button to delete the whole profile list.
Test Points Dialog
After pressing the "Tests" the following options are provided:
The "Tests Dialog" key opens the "Test Points" dialog together with another set of soft­keys (see below). The "Short Test" and "Long Test" keys will also be explained below.
The "Test Points" dialog shows a table containing the test point voltages of the various R&S EB500 modules. Each row in the table corresponds to a particular test point. For
96Operating Manual 4072.8426.02 ─ 08
Page 97
R&S® EB500
Operation
Graphical User Interface (GUI)
each test point, the table's columns show the related module, the name of the test point, the current voltage and the permissible upper and lower voltage limits. The following symbols are used in the last column of the table:
97Operating Manual 4072.8426.02 ─ 08
Page 98
R&S® EB500
Operation
Graphical User Interface (GUI)
Voltage is within limits.
Voltage limits are exceeded.
Voltage is too high.
Voltage is too low.
The "Report" key generates a status report which is stored along with the log files. The status report contains the current device configuration, the state of the test points and the error history.
The "Short Test" initiates a "Short Test". See also chapter 4.5.16, "TEST Subsystem", on page 337.
The "Long Test" initiates a "Long Test". See also chapter 4.5.16, "TEST Subsystem", on page 337.
The "Next Group" key shows the test points of the next module in the upper rows of the table.
The "Next Error" key scrolls to the next error where the test point limits are exceeded.
The "Error Group" key groups all errors and shows them at the top of the table. This is particularly useful if you want to see all errors at a glance. Press "Error Group" a second time to cancel the grouping of errors.
This key scrolls up.
This key scrolls down.
FILE
After pressing MENU > "FILE" a new softkey bar as shown below will pop-up with the options: "Import" for importing files from a USB flash drive, "Export" for exporting files to a USB flash drive and "Manage" for managing files within the device or USB flash drive.
File Browser
The file import / file export / file management dialogs will all incorporate a file browser. After activating the file browser in the dialog, by using the MAIN ROLLKEY and/or the ENTER key, you can browse into the subfolders of the internal storage or USB flash drive. The accessible folders within the device appear in the data file tree, as can be seen from the path just above the file browser.
For different purpose the file browser can come with or without checkbox.
98Operating Manual 4072.8426.02 ─ 08
Page 99
R&S® EB500
Operation
Graphical User Interface (GUI)
File browser without checkbox
Fig. 3-23: File browser without checkbox.
If the file browser is activated, you can navigate the folder contents. The folders are rep­resented in a tree-like structure. If you select a collapsed ("+") node with the mouse wheel or ROTARY KNOB, the node will expand. If the node is already expanded ("-"), it will collapse.
If you select a node with one of the cursor keys, use the CURSOR LEFT or CURSOR RIGHT key to collapse or expand a node, respectively. You can also use the CURSOR LEFT or CURSOR RIGHT key to scroll left or right, respectively.
A selected file/folder is indicated just above the file browser. Once you have selected the desired source file/folder, you can deactivate the file browser by pressing MAIN ROLLKEY or ENTER key. If you deactivate the file browser by pressing ESC, the selected file/folder will revert back to what it was before the file browser is activated.
File browser with checkbox
Fig. 3-24: File browser with checkbox.
As can be seen from the figure above, this file browser comes with a check box before each item. With check box, you can select individual files/folders or a group of files/folders and operate on them. A checkbox has 3 states, unchecked, checked and partially checked.
Unchecked: file/folder is not selected.
99Operating Manual 4072.8426.02 ─ 08
Page 100
R&S® EB500
Operation
Graphical User Interface (GUI)
Checked: file/folder is selected.
Partially checked: the contents within this folder are partially selected.
The browser with checkbox comes with two softkeys as shown below. Pressing "(Un)Select" will put a check in the box before the file/folder in focus if it is currently unchecked. This means that the file/folder will be added to the selections for further operation. If the file/folder is already selected, the check will be removed, meaning the file/folder will be removed from the selections. The available space in the current drive, together with the count and size of selected files are indicated below the file browser.
Pressing "Unselect All" will clear all selections.
Selecting folders
If a folder is selected, all the contents in the folder and subfolders are also selected. If all the files/subfolders in a folder are selected, the folder itself may not be selected. You
must select the folder manually.
100Operating Manual 4072.8426.02 ─ 08
Loading...