Rohde and Schwarz ZVM Data Sheet

Vector Network Analyzers ZVM, ZVK
Measurement systems meeting the highest standards – from 10 MHz to 20 GHz and 10 MHz to 40 GHz
Excellent dynamic range >115 dB (ZVM) >110 dB (ZVK) (measurement bandwidth 10 Hz)
Low inherent noise <–110 dBm (measurement bandwidth 10 Hz)
High measurement speed <0.5 ms/point (ZVM) <0.7 ms/point (ZVK)
Fast data transfer via IEC/IEEE bus Transfer time <15 ms (200 points)
Accurate calibration in test fixtures
and on wafers
Modern calibration techniques TOM, TRM, TRL, TNA, TOM-X
• Swept frequency-conversion and multitone measurements on ampli­fiers and mixers
– Arbitrary configuration of genera-
tor and receiver
– Selective receiver with fundamen-
tal mixing
Easy integration into PC environ-
ment and networks
Internal PC with WindowsNT
Embedding of virtual matching
networks
Import of virtual networks using CAE-compatible file formats (
.S2P, *.S4P, *.flp)
*
.S1P,
*
Overview
Vector network analyzers of ZV family
The vector network analyzer family from Rohde&Schwarz comprises the units ZVRL, ZVRE, ZVCE, ZVR, ZVC, ZVM and ZVK. All units are equipped with genera­tor, test set, reference and receiver chan­nels. The analyzers differ in frequency range, unidirectional or bidirectional measurement capabilities, active or pas­sive test set, number of reference chan­nels and thus availability of calibration techniques.
ZVK (active couplers)
10 MHz ..................................................... 40 GHz
ZVM (active couplers)
For the units up to 4 GHz (ZVRL, ZVRE, ZVR) and 8 GHz (ZVCE, ZVC), a separate data sheet (PD757.1802) is available. In addition, a free-of-charge CD-ROM (Order No. 1007.9074.14-03) contains extensive information on the network analyzers – including manuals, application notes and startup information. Additional information can also be found on the Rohde& Schwarz web site (www.rohde-schwarz.com).
10 MHz .............................. 20 GHz
300 kHz ....... 4 GHz
9 kHz ..................... 4 GHz
10 Hz .......................... 4 GHz
20 kHz ........................ 8 GHz
300 kHz ................ 8 GHz
ZVR, ZVRE (active bridges)
ZVR, ZVRE, ZVRL (passive bridges)
ZVR, ZVRE, ZVRL (with Ext. Measurements option)
ZVC, ZVCE (active couplers, passive bridges)
ZVC, ZVCE (active bridges)
Network Analyzers ZVM and ZVK ...
Versatile test set for universal use
ZVM and ZVK are compact instruments with integrated generator, two reference and two receiver channels and a bidirec­tional test set. This can be extended by attenuators with integrated switches in the generator and receiver paths. With this configuration, ZVM and ZVK offer direct access to all reference and receiver channels. This concept makes ZVM and ZVK well equipped for complex test set­ups, for example for bidirectional mea­surements on power amplifiers.
Fundamental mixing concept
ZVM and ZVK have two independent syn­thesizers for the generator and the receiver. In the receiver sections, funda­mental mixing is used up to high frequen­cies to provide the excellent dynamic range and outstanding selectivity, enabling straightforward measurements on frequency-converting DUTs or DUTs with extremely high selectivity.
Embedding and de-embedding of virtual networks, CAE software
The Virtual Embedding Networks option enables virtual embedding of arbitrary linear two-port networks into the test setup. The required data (
.S4P, *.flp) are obtained from a measure-
*
ment of the existing network or gener­ated by CAE tools from the theoretical model.
In tests of components that have to be matched to a given impedance, the matching network can thus be taken into account through mathematical algo­rithms of ZVM and ZVK instead of using the physical network. This method guar­antees high accuracy, ideal reproducibil­ity and maximum reliability without any loss of speed – great advantages espe­cially in production.
Conversely, by de-embedding, the influ­ence of a known network can be elimi­nated. The S-parameters of a chip can be analyzed, compensating for the effects of its housing and bonding leads through de-embedding.
.S1P, *.S2P,
*
Powerful and highly precise
Special calibration techniques
ZVM and ZVK feature modern calibration techniques patented by Rohde&Schwarz that allow full two-port calibration using fewer or only partially known standards. This simplifies the design of calibration standards used for example in test fixtures or on wafers. Thus calibration in non-coaxial systems can be performed with a minimum of effort at maximum accuracy and dynamic range.
Time-domain measurements
By transforming measurement data from the frequency to the time domain, discon­tinuities or impedances along the DUT can be displayed as a function of DUT length. With a maximum number of 2001 points, ZVM and ZVK can measure even very long DUTs with high resolution. Five filters allow the location of a discontinuity and the sidelobe suppression to be deter­mined with optimum resolution. The S-parameters of a given discontinuity can be displayed in the time domain by set­ting a window (gating). An additional processor module included in the corre­sponding option accelerates data
... designed for the most stringent demands
processing and the display of results to provide even realtime display – a valuable aid, for example in the tuning of band­pass filters with time domain transforma­tion.
printer, or the use of software tools on ZVM or ZVK for result processing or con­trol of the network analyzers via the IEC/ IEEE bus or an internal RSIB*) data bus. ZVM and ZVK can thus act as controllers of their own or for a complete test or pro-
Internal PC and Ethernet
ZVM and ZVK are based on Windows NT. The user has complete access to the hard
duction system. Moreover, the internal PC enables control and data exchange
via Ethernet. disk, the floppy disk drive and all inter­faces of the internal PC. This allows, for example, the connection of an external monitor, the installation of any type of
Decoupled 4-channel display
In the decoupled mode, the frequency
grid, measurement bandwidth, calibra-
ZVM and ZVK extend the frequency range of the Rohde&Schwarz network analyz­ers to 20 GHz and 40 GHz. Their outstanding performance in terms of speed, dynamic range and accuracy shows already in standard applications such as S-parameter or group delay measurements. This is enhanced by a wealth of mea­surement, display and logging functions. In addition, ZVM and ZVK can be used for complex measurement tasks, for example measurements on frequency-con­verting DUTs (conversion loss, intermodulation, spurious) and nonlinear measure­ments (intercept point and compression point).
tion technique and measurement mode can be configured independently for each of the four display channels. In amplifier measurements, this allows the simulta­neous measurement and display of important parameters in quasi-realtime, such as gain, compression point (power
Highlights in brief
ZVM ZVK
Frequency range 10 MHz to 20 GHz 10 MHz to 40 GHz
Frequency resolution 100 µHz
Impedance 50
Te st p o rt s PC 3.5 male 2.92 mm male
Measurement time (normalized)
Output power +5 dBm/+2 dBm to 85 dBm 0 dBm/5 dBm to 85 dBm
Power uncertainty <1 dB to 2 dB
Dynamic range (IF bandwidth 10 Hz)
Measurement bandwidths 1 Hz to 10 kHz (in 9 steps) and 26 kHz
Calibration techniques
2 Vector Network Analyzers ZVM, ZVK
<0.5 ms/point <0.7 ms/point
>85 dB (<0.5 GHz) >115 dB (0.5 GHz to 8 GHz) >110 dB (8 GHz to 16 GHz) >100 dB (16 GHz to 20 GHz)
TOM, TRM, TNA, TOM-X, AutoKal (all Rohde& Schwarz patents),
TRL, TOSM, normalization techniques
>80 dB (<0.5 GHz) >110 dB (0.5 GHz to 8 GHz) >105 dB (8 GHz to 16 GHz) >90 dB (16 GHz to 20 GHz) >90 dB (20 GHz to 28 GHz) >80 dB (28 GHz to 40 GHz)
sweep) and harmonics versus power or frequency, or compression point versus frequency (see screen display below).
Time-optimized calibration, measure­ment and control
The Rohde&Schwarz two-port calibration techniques reduce the number of required
calibration standards to a minimum of 2. This significantly cuts the time required for manual calibration. The short measure­ment time of
<
500 µs or <700 µs per point guarantees minimum sweep times through to realtime display. The output of a marker value via the IEC/IEEE bus takes
less than 5 ms, the transfer of a complete trace (200 points) less than 15 ms. These featur es are the b asis for the exce llent pe r­formance of ZVM and ZVK both in manual operation and in automated test systems.
*) Remote control via an internal software interface using the same SCPI command syntax as for normal
Vector Network Analyzers ZVM, ZVK 3
Convincing concepts – features in detail
1
Patented calibration techniques
Besides various normalization tech­niques, ZVM and ZVK offer the classic 12-term TOSM technique (Through, Open, Short, Match). In addition, the analyzers feature as standard other cali­bration techniques mostly patented by Rohde&Schwarz: TOM, TRM, TRL,TNA, TOM-X (Fig. 4).
The standards are defined as follows *):
THROUGH Through-connection of known length
OPEN Open circuit of known length and phase response
Extremely short measurement times
A powerful microprocessor system com­bined with ultra-fast synthesizers makes for extremely short measurement times even with a large number of test points and small measurement bandwidths (Fig. 3). This in conjunction with short IEC/IEEE-bus access and transfer times considerably speeds up automated test and production sequences.
Ultra-wide dynamic range
The extremely low-noise front end, using fundamental mixing, yields a dynamic range that, with appropriate configura­tion, by far exceeds the specified values of 115 dB (Fig. 2) and 110 dB. This excep­tionally wide range makes it possible to measure RF components with high stop­band attenuation and achieve high accu­racy also at low power levels.
The Rohd e&Schwarz calibration techniques offer maximum convenience and accuracy also for on-wafer measurements
SHORT Short circuit of known length
MATC H Matched termination
Bandpass filter measurement with ZVM: the extremely
low-noise front end based on fundamental mixing yields a
dynamic range of up to >140 dB (not guaranteed)
2
4 Vector Network Analyzers ZVM, ZVK
*) For detailed description see specifications.
LINE Ideal matched line of known length
accurate, full two-port calibration in test fixtures or on wafers (Fig. 1).
Measurements on amplifiers and frequency-converting DUTs
REFLECT Reflecting one-port standard, identical for PORT 1 and PORT 2
NETWORK Symmetrically reflecting two-port stan­dard
ATTENUATOR Matched attenuator with unknown attenuation
The advantage of the 7-term calibration techniques (TRL, TNA, TRM, TOM) is the reduced number of calibration standards required and their simplified description. In particular, the use of REFLECT or NETWORK avoids the use of OPEN and thus the complex determination of the fringing capacitance. This allows the design and production of calibration st an dard s a t rea so nabl e c ost a nd enab le s
TRL is recommended where high direc­tivity is necessary.
TNA is recommended for applications with symmetrical test ports and where a well-matched two-port or double match (ATTENUATOR) can be provided in suffi­cient quality. If the calibration step is car­ried out with NETWORK, the test fixture can simply be left open. Thus a full two­port calibration can be performed with only two standards and the same accu­racy as TOSM.
TOM offers the advantage of implicit ver­ification: errors resulting from faulty cali­bration standards or operator errors are automatically detected with high proba­bility and thus avoided already during cal­ibration.
The system concept of ZVM and ZVK with two independent synthesizers for the generator and receiver sections enables versatile measurements with excellent accuracy, wide dynamic range and high measurement speed on frequency-con­verting and nonlinear DUTs such as amplifiers and mixers. Three generators (one internal, two external) can be config­ured and controlled independently of each other. The fundamental mixing con­cept of ZVM and ZVK and the resulting high selectivity make additional external filters superfluous. The receiver will even detect weak signals such as intermodula­tion products and spurious, since the full sensitivity and dynamic range of ZVM and ZVK are available also for selective, fre­quency-converting DUTs.
3
Measurement times
Frequency sweeps with 401 test points
IF bandwidth 10 kHz
ZVM ZVK
Frequency range 10 MHz to 20 GHz or 10 MHz to 40 GHz
Two-port calibrated, bidirectional Normalized, unidirectional
Frequency range 1 GHz to 2 GHz
Two-port calibrated, bidirectional Normalized, unidirectional
IEC/IEEE-bus data transfer times for real and imaginary parts
Time between sending of query and availability of data
Number of test points 51 201 401
ASCII 40 ms 90 ms 160 ms
IEEE-754 floating point format (setting data 64 bit, measurement data 32 bit)
340 ms 210 ms
260 ms 140 ms
10 ms 15 ms 25 ms
430 ms 260 ms
290 ms 130 ms
Left: measurement and IEC/IEEE-bus times of ZVM and ZVK
Below: comparison of two-port calibration techniques
Two - po rt calibration techniques
TOM 5 Implicit verification
TRM 5 Especially for test fixtures
TRL 4 High directivity
TNA 3 Especially for planar circuits
TOSM 7 Classical method
TOM-X 5 (9) Eliminates crosstalk
Number of calibration steps Special feature
Vector Network Analyzers ZVM, ZVK 5
4
5
ROHDE & SCHWARZ VECTOR NETWORK ANALYZER 10 MHz ... 20 GHz ZVM
Front end with internal LO
1127.8500.60
6
Test setup for front-end measurement
The features in detail (cont’d)
Frequency-converting and nonlinear DUTs of any type (e.g. front ends, see Fig. 5) can thus be measured with little effort. The decoupled measurement and display mode permits the simultaneous display of different parameters on ZVM or ZVK (Figs 6 and 7).
The user-friendly MIXER MODE menu makes it very easy to configure mixer measurements with constant or swept RF, IF or LO. For more complex measure­ment tasks, the ARBITRARY mode offers almost unlimited configurations of the internal and the external generators and the receiver of ZVM and ZVK.
Typical measurements on amplifiers, fre­quency converters, multipliers, dividers, synthesizers etc are:
• sidebands of mixers with fixed or tracking IF
• any harmonics versus frequency or power
• intermodulation products of amplifi­ers and mixers (e.g. IP3, IP5, IP7...)
•spurious
Front-end measurement: simultaneous display of both sidebands, LO crosstalk and RF input matching
Amplifier measurement: simultaneous display of gain, compression and harmonics versus power, and display of compression point versus fre­quency
• mixture products of DUTs with multi­ple frequency conversion, multipliers, dividers and combinations of such components
In am plif ier m ea sure ment s, ZVM a nd ZV K can display even nonlinear parameters versus frequency, e.g.:
• n dB compression point
• second-order intercept point (SOI)
• third-order intercept point (TOI)
7
The Rohde&Schwarz network analyzers offer system error and power correction, yielding high accuracy in the measure­ment of S-parameters and absolute power.
Two DC inputs at the rear enable the display of DC voltages versus frequency, and in amplifier measurements the power added efficiency (PAE) can be displayed.
6 Vector Network Analyzers ZVM, ZVK
Test set and system configuration
ZVM and ZVK are four-channel instru­ments with two measurement and two reference channels. The test sets are of fully symmetrical design in the forward and the reverse direction (Fig. 8).
During a bidirectional sweep, the elec­tronic RF switch applies the signal to the DUT at every frequency point in the for­ward and the reverse direction: ZVM and ZVK thus indicate the fully corrected measured values during the sweep a valuable aid in alignments at small mea­surement bandwidths.
Optional step attenuators (ZVM-B21 to
-B24, ZVK-B21 to -B24) with attenuation from 0 dB to 70 dB in steps of 10 dB can be inserted into the generator and receiver paths (Fig. 8).
The attenuators extend the output power range down to –90 dBm and the maxi­mum input power at PORT 1/2 to +27 dBm.
If ZVM or ZVK are fitted with an attenua­tor in a receiver path, an additional test por t – IN PUT b 1 o r I NP UT b2 – i s a vail ab le on the front panel. An internal switch in the attenuator enables direct access to the respective receiver channel, bypass­ing the coupler. Sensitivity and dynamic range are thus increased by typically 10 dB. The two reference channels can be accessed directly too, since the associ­ated paths are routed via the front panel as standard (R1 CHANNEL IN/OUT, R2 CHANNEL IN/OUT, Fig. 8). In the case of ZVM, the PORT 1, PORT 2, INPUT b1 and INPUT b2 test ports are PC3.5 male connectors, the inputs and outputs of the reference channels are SMA female con­nectors. In the case of ZVK, all ports are
2.92 mm male connectors, the inputs and
outputs of the reference channels are
2.92 mm female connectors.
A network analyzer equipped with receiver step attenuators not only offers sensitivity and dynamic range increased by 10 dB, but also the functionality of an instrument without a test set ("Delete Testset"), i.e. direct access to the refer­ence and measurement channels.
Active DUTs can be powered and driven via the inner conductors of PORT 1 and PORT 2 with DC voltage of up to 30 V or 200 mA. The required DC power is applied via rear-panel BNC connectors.
The flexible concept of the network ana­lyzers allows the configuration of complex external test sets for special measure­ment tasks such as:
• group delay of mixers with the aid of a reference mixer
• high-power measurements on power amplifiers using a test set with preamplifiers (Fig. 9)
•S
measurement on power amplifier
22
during operation
Test set of ZVM and ZVK Proposal of external test set for measurements on power amplifiers (DUTs)
8
Vector Network Analyzers ZVM, ZVK 7
9
Specifications
Unless otherwise stated, specifications apply to test ports PORT 1 and PORT 2, a nominal output power of –10 dBm at the source port and an IF bandwidth 10 kHz.
Especially important data are framed in blue
Measurement range
Characteristic impedance 50
Port connectors
ZVM 3.5 mm (male) ZVK 2.92 mm (male)
Frequency
Range ZVM 10 MHz to 20 GHz Range ZVK 10 MHz to 40 GHz Uncertainty 4 x 10
Resolution 100 µHz
Number of test points (selectable) 1 to 2001
Measurement time per point
with min. 400 points ZVM ZVK and IF bandwidth of 10 Hz 10 kHz 10 kHz
with system error correction <200 ms <0.9 ms <1.1 ms normalized <100 ms <0.5 ms <0.7 ms
Dynamic range (without system error correction, without optional attenuator) ZVM ZVM ZVK ZVK at IF bandwidth of 10 Hz 10 kHz 10 Hz 10 kHz
up to 500 MHz >75 dB >45 dB >70 dB >40 dB 500 MHz to 8 GHz >115 dB >85 dB >110 dB >80 dB 8 GHz to 16 GHz >110 dB >80 dB >105 dB >75 dB 16 GHz to 20 GHz >100 dB >70 dB >90 dB >60 dB
20 GHz to 28 GHz >90 dB >60 dB
28 GHz to 40 GHz >80 dB >50 dB (The dynamic range is defined as the difference between the maximum nominal source power and the peak value displayed after smoothing the measured trace for the transmission magnitude with an aperture of 1%, the trace being caused by inherent noise and crosstalk, with test ports short-circuited.)
Measurement bandwidths
(IF bandwidths) 1 Hz to 10 kHz (half-decade steps) and
6
+ 1 x 10−6 x operating time in
years
26 kHz (full)
Measurement accuracy
The following data are valid between 20°C and 26°C, provided the instrument has reached thermal equilibrium (about 1 h after switch-on) and the temperature has not varied by more than 1 K after calibration. Validity of the data is conditional on the use of a suitable calibration kit by which the effective system data speci­fied below are achieved.
ZVM uncertainty of transmission measurements
after system error correction Specifications are based on a matched DUT, an IF bandwidth of 10 Hz, and a nom­inal output power of –10 dBm at the source port.
10 MHz to 500 MHz
for +15 dB to 25 dB 0.2 dB or 2° for 25 dB to 35 dB 1 dB or 6°
500 MHz to 8 GHz
for +15 dB to +5 dB 0.2 dB or 2° for +5 dB to 50 dB 0.1 dB or 1° for 50 dB to 65 dB 0.2 dB or 2°
for 65 dB to 80 dB 1 dB or 6°
8GHz to 16GHz
for +15 dB to 55 dB 0.2 dB or 2° for 55 dB to 70 dB 1 dB or 6°
16 GHz to 20 GHz
for +12 dB to +5 dB 0.3 dB or 3° for +5 dB to 30 dB 0.2 dB or 2° for 30 dB to 45 dB 0.3 dB or 3° for 45 dB to 60 dB 1 dB or 6°
ZVM uncertainty of reflection measurements
after system error correction Specifications are based on an isolating DUT, an IF bandwidth of 10 Hz, and a nominal output power of –10 dBm at the source port.
10 MHz to 20 GHz
for +10 dB to +3 dB 0.6 dB or 4° for +3 dB to 15 dB 0.4 dB or 3° for 15 dB to 25 dB 1 dB or 6° for 25 dB to 35 dB 3 dB or 20°
Variation of data trace at 0 dB per Kelvin of temperature variation <0.2 dB or <2°
ZVK uncertainty of transmission measurements
after system error correction Sp eci fic atio ns a re b ase d on a mat che d DU T, an I F ba ndw idt h of 10 H z, a nd a nom ­inal output power of –10 dBm at the source port.
10MHz to 500MHz
for +10 dB to 15 dB 0.2 dB or 2° for 15 dB to 30 dB 1 dB or 6°
500 MHz to 8 GHz
for +10 dB to +5 dB 0.2 dB or 2° for +5 dB to 45 dB 0.1 dB or 1° for 45 dB to 60 dB 0.2 dB or 2° for 60 dB to 75 dB 1 dB or 6°
8GHz to 16GHz
for +10 dB to 50 dB 0.2 dB or 2° for 50 dB to 65 dB 1 dB or 6°
16 GHz to 28 GHz
for +5 dB to 20 dB 0.2 dB or 2° for 20 dB to 35 dB 0.3 dB or 3° for 35 dB to 50 dB 1 dB or 6°
28 GHz to 40 GHz
for +5 dB to 10 dB 0.2 dB or 2° for 10 dB to 25 dB 0.3 dB or 3° for 25 dB to 40 dB 1 dB or 6°
ZVK uncertainty of reflection measurements
after system error correction Specifications are based on an isolating DUT, an IF bandwidth of 10 Hz, and a nominal output power of –10 dBm at the source port.
10 MHz to 20 GHz
for +5 dB to 15 dB 1 dB or 6° for 15 dB to 30 dB 3 dB or 20°
20 GHz to 40 GHz
for +5 dB to 0 dB 2 dB or 15° for 0 dB to 10 dB 1 dB or 6° for 10 dB to 25 dB 3 dB or 20°
Variation of data trace at 0 dB per Kelvin of temperature variation <0.2 dB or <2°
8 Vector Network Analyzers ZVM, ZVK
Effective system data
Input level
The following data are valid between 20°C and 26°C, provided the instrument has reached thermal equilibrium (about 1 h after switch-on) and the temperature has not varied by more than 1 K after calibration. The data are based on an IF bandwidth of 10 Hz and system error calibration by means of a suitable calibra­tion kit.
Frequency range 50 MHz to 20 GHz above 20 GHz
ZVM ZVK ZVK
Directivity >46 dB >42 dB >38 dB
Source match >36 dB >36 dB >33 dB
Reflection tracking <0.1 dB <0.1 dB <0.1 dB
Load match >46 dB >42 dB >38 dB
Transmission tracking <0.1 dB <0.1 dB <0.2 dB
Output power
Range without optional generator step attenuator
ZVM ZVK
up to 16 GHz 20 dBm to +5 dBm 20 dBm to 0 dBm
above 16 GHz 20 dBm to +2 dBm 20 dBm to 5 dBm Uncertainty at 10 dBm without optional power calibration 2 dB 2 dB
150 MHz to 16 GHz in
temperature range 20°C to 26 °C 1 dB 1 dB Linearity (referred to 10 dBm) <1 dB <1 dB
above 150 MHz in
temperature range 20°C to 26 °C <0.4 dB <0.4 dB Resolution 0.1 dB 0.1 dB
Spectral purity
Harmonics
at maximum nominal source power ZVM ZVK
up to 10 GHz <23 dBc <20 dBc
10 GHz to 20 GHz <17 dBc <15 dBc
above 20 GHz <25 dBc at 10 dBm source power
up to 10 GHz <30 dBc <30 dBc
above 10 GHz <25 dBc <25 dBc
Spurious <35 dBc <35 dBc
Maximum nominal input level
without optional receiver step attenuator +5 dBm with receiver step attenuator set to 0 dB +5 dBm with receiver step attenuator set to 30 dB +27 dBm
Level measurement uncertainty (without optional power calibration) in temperature range 20 °C to 26 °C
up to 500 MHz for +5 dBm to –45 dBm 2 dB 500 MHz to 16 GHz for +5 dBm to –70 dBm 2 dB 16 GHz to 20 GHz for +5 dBm to –50 dBm 2 dB
20 GHz to 28 GHz for +5 dBm to –50 dBm (ZVK) 3 dB above 28 GHz for +5 dBm to –30 dBm (ZVK) 4 dB
Damage level
without optional receiver step attenuator +27 dBm with receiver step attenuator set to 0 dB +27 dBm with receiver step attenuator set to 30 dB +30 dBm
Damage DC current/voltage 0.5 A or 30 V
RMS noise level at IF bandwidth 10 Hz
up to 500 MHz <80 dBm 500 MHz to 8 GHz <110 dBm 8 GHz to 16 GHz <105 dBm 16 GHz to 20 GHz <95 dBm 20 GHz to 28 GHz (ZVK) <95 dBm above 28 GHz (ZVK) <85 dBm
Match (without system error correction)
up to 50 MHz >10 dB 50 MHz to 8 GHz >12 dB 8 GHz to 20 GHz >10 dB above 20 GHz (ZVK) >8 dB
Reference channel inputs
R CHANNEL IN
ZVM ZVK
Connectors SMA (female) 2.92 mm (female) Match >12 dB >8 dB Maximum nominal input level +5 dBm +5 dBm Damage level +20 dBm +20 dBm
SSB phase noise
1 Hz bandwidth,10 kHz from carrier
up to 150 MHz <100 dBc
150 MHz to 1 GHz <90 dBc
above 1 GHz <90 dBc + 20 x log (f/GHz)
<78 dBc at 4 GHz <72 dBc at 8 GHz <64 dBc at 20 GHz <58 dBc at 40 GHz (ZVK)
Residual FM
RMS weighting from 10 Hz to 3 kHz
up to 150 MHz <2 Hz
150 MHz to 1 GHz <5 Hz
1 GHz to 2 GHz <10 Hz
2 GHz to 4 GHz <20 Hz
4 GHz to 8 GHz <40 Hz
8 GHz to 20 GHz <80 Hz
20 GHz to 40 GHz (ZVK) <160 Hz
System error correction techniques
ZVM and ZVK offer normalizations for reflection and transmission measurements, full one-port calibration (3-term, OSM), one-path two-port calibration, and the classic 12-term two-port calibration (TOSM). In addition, the following full two­port calibration methods are available: TOM, TRM, TRL, TNA and TOM-X (15­term). TOM, TRM, TNA and TOM-X are calibration methods patented by Rohde& Schwarz.
The names of the methods indicate the standards used for calibration:
T = Through
The T standard is a two-port standard which establishes a direct low-loss connec­tion between the two test ports. A frequency-dependent attenuation can be tak­en into account by the analyzer. The standard has to be well-matched and may have any electrical length, which has to be exactly known (compare L standard).
O = Open
Th e O s ta nda rd is a on e-p ort st and ar d. I t re al ize s t ota l re fl ect io n wi th a ma gn itu de of 1 in the ideal case and a phase of approx. 0°. The phase response versus fre­quency must be accurately known to the analyzer (coefficients C dependent increase of the return loss can be taken into account by the analyzer. The electrical length of the O standard may differ from zero and must be exactly known.
Vector Network Analyzers ZVM, ZVK 9
). A frequency-
i
S = Short
Th e S s tan dar d is a on e-p ort sta nda rd. It real ize s to ta l re fle cti on w ith a ma gni tud e of 1 in the ideal case and a phase of approx. 180° at short-circuit plane (coeffi-
). A frequency-dependent increase of the return loss can be taken into ac-
cients L
i
cou nt by t he analyzer. The e lectr ical len gth of the S standa rd may dif fer fr om zero an d m us t be kn own . I t c aus es a l eng th -p rop or tio na l f req ue nc y de pe nd enc e o f t he phase.
M = Match
The M standard is a one-port standard which in the ideal case realizes a zero-re­flection termination for the reference impedance (mostly 50 ). A sliding match is often used at high frequencies because it yields higher effective directivities than fixed loads.
R = Reflect
The R standard is a one-port standard. In contrast to the M standard it features high reflection which may assume any unknown value. It must be known howev­er wh eth er the ref le ct a pp roa ch es a n op en or a sh ort ci rcu it . If li ne t ran sf orm at ion is to be expected from open to short because of the electrical length of the R standard, the electrical length has to be approximately known.
L = Line
The L standard is a two-port standard. It establishes an almost perfectly matched connection between the two test ports and defines the reference impedance. A frequency-dependent attenuation caused by the L standard can be taken into ac­count by the analyzer. The L standard has to have an electrical length different from that of the T standard, but the difference should not amount to an integer multiple of half the wavelength (singularity).
Screen formats (examples) overlay, dual channel split,
quad channel split
Markers 8 normal markers or 7 delta markers for
each display channel Marker resolution 4 significant digits Marker formatting selectable, independent of trace format-
ting Automatic marker functions marker tracking, marker search,
marker target, band filter functions (Q,
shape factor, etc) Trace mathematics all four arithmetical operations with up
to three operands Display lines horizontal lines, circles or radial lines Limit lines pairs of curves formed from line seg-
me nts in Ca rte sia n dia gra ms, a ny c ircl es
in polar diagrams
Further connectors (rear panel)
PORT BIAS 1/2
(DC bias inputs for PORT 1/2)
Max. nominal input current/voltage 200 mA/30 V Damage current/voltage 500 mA/30 V
EXT TRIGGER (input for external trigger signal)
Edge-triggered TTL signal, polarity (selectable) positive or negative Minimum pulse width 1 µs
N = Network
The N standard is a two-port standard featuring symmetrical reflection which may as sum e any valu e other tha n zer o but has to b e id enti cal at b oth port s. Sa me as with the R standard it must be known whether the reflection approaches an open or a short circuit. Transmission of the N standard is arbitrary, need not be known and may vary arbitrarily versus frequency. In the extreme case it may even be 1 or zero.
A = Attenuator
The A standard is a two-port standard. It has to be well-matched and may feature any unknown attenuation different from that of the T standard.
TO M-X ( X = c ros sta lk) is a n ex ten sio n of the TOM m eth od. It c ons ide rs a ll p oss ibl e crosstalk between the four receiver channels (full model). Since this technique does not use approximations, it is particularly effective in the elimination of cross­talk and thus in increasing the effective dynamic range of the system. This meth­od however needs a higher effort.
Display
Screen 26 cm colour LCD Resolution 640 x 480 x 256 Sweep modes frequency, power, and time Parameter formats (examples) S parameters and derived quantities like
SWR, impedance, admittance, group delay, etc, as well as nonlinear parame­ters (optional) like n dB compression point, SOI and TOI. Complex parameters are displayed either in a complex form or formatted to magnitude, phase, real or imaginary part
Diagrams (examples) Cartesian: linear, simple or double loga-
rithmic, segmented polar: linear, logarithmic or segmented, Smith (any zoom), inverted Smith, Char­ter
Scaling (examples) 0.001 dB/ to 50 dB/
1 m°/ to 200 k°/ 1 pU/ to 1 GU/ (automatically variable number of grid lines through MAX/MIN scaling)
Multichannel display up to 4 independent display channels
(CH1 to CH4)
LEVEL (input for external level control)
Frequency range 0 Hz to 100 kHz Voltage range 0 V to 10 V Input impedance >10 k
DC MEAS INPUTS DC 1/2 (DC measurement inputs)
Voltag e ra nge 10 V to +10 V Measurement uncertainty 0.1 V Input impedance >10 k
EXT FREQ REF IN (input for external reference frequency)
Frequency (in 1 MHz steps) 1 MHz to 15 MHz Max. permissible deviation 6 x 10 Input level (V Input impedance 1 k
) 0.1 V to 3 V
rms
EXT FREQ REF OUT (output of internal reference frequency)
Frequency 10 MHz Frequency uncertainty 4 x 10
6
+ 1 x 10−6 x operating time in years
Level (sine) 12 dBm ± 3 dB into 50
EXTERNAL GENERATOR
Connectors for high-speed control of an external generator from Rohde&Schwarz families SME, SMP, SMT, etc. The BLANK signal is low at each frequency point of the sweep and high during the transition from one point to the next. The network analyzer controls the external generator by means of the TRIGGER signal. To set the generator to the next frequency point, the TRIGGER signal goes high for a brief period.
BLANK (input) TTL signal
TR IG GE R (output) TTL signal
ANALYZER MONITOR IBM-PC-compatible VGA connector for
analyzer screen
PC MONITOR IBM-PC-compatible VGA connector for
PC screen
MOUSE IBM-PC-compatible PS/2 connector
KEYBOARD IBM-PC-compatible 5-contact
DIN connector
USER (input/output) 16 bit TTL, user-programmable,
25-contact sub-D
6
10 Vector Network Analyzers ZVM, ZVK
COM 1/ COM 2 IBM-PC-compatible serial interfaces,
RS232, 9-contact sub-D
IEC BUS remote-control interface IEEE488,
IEC625, 24-contact (for general applica­tions)
IEC SYSTEM BUS remote-control interface IEEE488,
IEC625, 24-contact (for control of gener­ators, eg as local oscillators in mixer measurements)
LPT IBM-PC-compatible printer interface,
Centronics, 25-contact sub-D
MULTIPORT control of optional three-port and four-
port adapters
Optional interfaces (eg LAN Ethernet) are available and specified separately.
Generator Step Attenuator PORT 1/2 options These options permit the power of the output signal at PORT 1/2 to be attenuated in 10 dB steps between 0 dB and 70 dB. The use of an attenuator reduces the dy­na mic rang e to >10 5 d B bet wee n 12 GH z an d 16 GH z at a n IF ban dwi dth o f 10 Hz .
ZVM ZVK
Frequency range 10 MHz to 20 GHz 10 MHz to 40 GHz Attenuation 0 dB to 70 dB 0 dB to 70 dB Attenuation steps 10 dB 10 dB Attenuation uncertainty
up to 30 dB 3 dB 3 dB above 40 dB 10 MHz to 20 GHz 3 dB 3 dB
Output power ZVM ZVK
up to 16 GHz 90 dBm to +2 dBm 90 dBm to 3 dBm 16 GHz to 20 GHz 90 dBm to 2 dBm 90 dBm to 9 dBm
above 20 GHz 90 dBm to 9 dBm with “Additional Power“ setting with reduced specifications up to 16 GHz 85 dBm to +5 dBm 85 dBm to 0 dBm 16 GHz to 20 GHz 85 dBm to +2 dBm 85 dBm to 5 dBm
above 20 GHz 85 dBm to 5 dBm
20 GHz to 33 GHz 5 dB
Options
Time Domain option Display and gating of measured values in the time domain and transformation back to the frequency domain.
Mixer Measurements option This option allows network analysis for frequency-converting DUTs (single and multiple conversion) and almost any kind of harmonics and spurious measure­ments to be performed.
Nonlinear measurements option For special measurements on nonlinear DUTs, such as the determination of the n dB compression point versus frequency and the SOI and TOI intermodulation products.
Power calibration option Th is o pti on i s n ece ssa ry for pre cis e po wer ca lib rat ion of the net wor k a nal yze r. Th e source power (additional power meter, e.g. NRVD, NRVS or NRV from Rohde& Schwarz required) as well as the absolute power measurement of the receiver in­put signals can be calibrated.
Virtual Embedding Networks option This option allows measured networks or simulated networks from a CAD pro­gram to be taken into account in the measurement results. Mismatched DUTs such as SAW filters can be matched virtually without any additional hardware be­ing required. Complementary to calibration procedures, the effect of real embed­ding networks like test fixtures can be eliminated by calculation.
Ethernet option With this option the analyzer can be networked (LAN).
Receiver Step Attenuator PORT 1/2 options These options permit the level of the input signal at PORT 1/2 to be attenuated in
10 dB steps between 0 dB and 70 dB. Moreover, with this option fitted, an addi­tional receiver input INPUT b1/b2 is available on the front panel. The use of an attenuator reduces the dynamic range to >105 dB between 12 GHz and 16 GHz at an IF bandwidth of 10 Hz.
ZVM ZVK
Frequency range 10 MHz to 20 GHz 10 MHz to 40 GHz Attenuation 0 dB to 70 dB 0 dB to 70 dB Attenuation steps 10 dB 10 dB Attenuation uncertainty
up to 30 dB 3 dB 3 dB
above 40 dB 10 MHz to 20 GHz 3 dB 3 dB
Receiver Inputs INPUT b1/b2
Connectors 3.5 mm (male) 2.92 mm (male) Match above 50 MHz >10 dB >8 dB Maximum nominal input level 5 dBm 5 dBm Damage level +20dBm +20dBm
20 GHz to 33 GHz 5 dB
IEC/IEEE-Bus Interface for Integrated PC option This option provides a third IEC/IEEE-bus interface to the integrated PC in addition to the two IEC/IEEE-bus interfaces provided as standard.
Certified Quality System
ISO 9001
DQS REG. NO 1954
Vector Network Analyzers ZVM, ZVK 11
Options
Option Type Features and benefits
AutoKal ZVR-B1 Device for automatic full two-port calibration for connection to PORT 1/2. Frequency range DC to 8 GHz.
Time Domain ZVR-B2 Measurement of step and impulse response, delay measurements, gating in time and frequency domain
Mixer Measurements ZVR-B4 Independent configuration and control of external generators, internal generator and receiver of ZVM and ZVK
Nonlinear Measurements ZVR-B5 Measurement of n dB compression point and 2nd- or 3rd-order intercept points (SOI/TOI)
Power Calibration ZVR-B7 Power calibration to improve absolute amplitude accuracy of generator and receiver
3-Port Adapter ZVR-B8 External device extending PORT 1 to 2 ports; frequency range 9 kHz to 4 GHz, type N (f) connectors at the DUT side. Con-
Virtual Embedding Networks
4-Port Adapter ZVR-B14 External device extending PORT 1 and PORT 2 to 2 ports each (var. 02), or PORT 1 to 3 ports (var. 03); frequency range 9 kHz
Ethernet Interface for internal PC
IEC/IEEE-Bus Interface for internal PC
Generator Step Attenuator PORT 1
Generator Step Attenuator PORT 2
Receiver Step Attenuator PORT 1
Receiver Step Attenuator PORT 2
ZVR-K9 Adding virtual or correcting real existing networks by mathematical algorithms
FSE-B16 Ethernet interface for internal PC
FSE-B17 IEC/IEEE interface for internal PC (in addition to two IEC/IEEE-bus interfaces provided as standard)
ZVM-B21, ZVK-B21
ZVM-B22, ZVK-B22
ZVM-B23, ZVK-B23
ZVM-B24, ZVK-B24
Type N (f) connectors at the DUT side. Connection to ZVM requires PC3.5 (f) to N (f) adapters, for connection to ZVK,
2.92 mm (f) to N (f) adapters are required
• Full two-port calibration within a few seconds
• Localization of discontinuities, determination of reflection coefficients of discontinuities as a function of length/delay, supplementary function for calibration, tuning of filters, optimization of connectors, etc
• Easy converter and mixer measurements (conversion gain)
• Convenient measurements of amplifier and mixer products vs. frequency (spurious, harmonics, intermodulation products, etc)
• Display of compression point and SOI/TOI versus frequency
• High absolute power accuracy of generators (internal and external) and receivers for amplifier and mixer measure­ments
nection to ZVM requires PC3.5 (f) to N (f) adapters, for connection to ZVK, 2.92 mm (f) to N (f) adapters are required
• Measurements of 3-port devices such as duplex filters
• Replacing various test fixtures with physical matching networks by one single standard fixture and virtual networks
• High accuracy and reproducibility, e.g. in SAW filter measurements
to 4 GHz, type N (f) connectors at the DUT side. Connection to ZVM requires PC3.5 (f) to N (f) adapters, for connection to ZVK, 2.92 mm (f) to N (f) adapters are required
• Simultaneous measurement of two 2-port devices
• Measurements on diplexers
• Control and data transfer of ZVM or ZVK via Ethernet
• Control of ZVM or ZVK and external test equipment by internal PC
Mechanical attenuator for generator path to PORT 1. Attenuation between 0 dB and 70 dB in 10 dB steps
• Decrease of minimum generator output power down to –90 dBm at PORT 1
Mechanical attenuator for generator path to PORT 2. Attenuation between 0 dB and 70 dB in 10 dB steps
• Decrease of minimum generator output power down to –90 dBm at PORT 2
Mechanical attenuator for receiver path from PORT 1 and Input b1. Attenuation between 0 dB and 70 dB in 10 dB steps. Includes additional test port Input b1
• Increase of maximum receiver input power at PORT 1 to +27 dBm
• Direct access to measurement channel b1
Mechanical attenuator for receiver path from PORT 2 and Input b2. Attenuation between 0 dB and 70 dB in 10 dB steps. Includes additional test port Input b2
• Increase of maximum receiver input power at PORT 2 to +27 dBm
• Direct access to measurement channel b22
Vector Network Analyzers ZVM, ZVK 12
Ordering information
Order designation
Type Frequency range Order No.
Analyzers
Vector Network Analyzer
4-channel, 50 , active test set
ZVM 10 MHz to 20 GHz 1127.8500.60
Vector Network Analyzer
4-channel, 50 , active test set
ZVK 10 MHz to 40 GHz 1127.8651.60
Options
Time Domain ZVR-B2 1044.1009.02 Mixer Measurements Nonlinear Measurements ZVR-B5 1044.1321.02 Power Calibration Virtual Embedding
3)
Networks Ethernet AUI for internal PC FSE-B16 1073.5973.02 Ethernet BNC for internal PC FSE-B16 1073.5973.03 Ethernet RJ45 for internal PC
IEC/IEEE-Bus Interface for internal PC
Generator Step Attenuator for ZVM, PORT 1
Generator Step Attenuator for ZVM, PORT 2
Receiver Step Attenuator for ZVM, PORT 1
Receiver Step Attenuator for ZVM, PORT 2
Generator Step Attenuator for ZVK, PORT 1
Generator Step Attenuator for ZVK, PORT 2
Receiver Step Attenuator for ZVK, PORT 1
Receiver Step Attenuator for ZVK, PORT 2
1)
ZVR-B4 1044.1215.02
2)
ZVR-B7 1044.1544.02 ZVR-K9 1106.8830.02
FSE-B16 1073.5973.04
FSE-B17 1066.4017.02
ZVM-B21 1128.1009.11
ZVM-B22 1128.1009.21
4)
5)
ZVM-B23 1128.1009.12
ZVM-B24 1128.1009.22
ZVK-B21 1128.1409.11
ZVK-B22 1128.1409.21
4)
5)
ZVK-B23 1128.1409.12
ZVK-B24 1128.1409.22
ZVM, ZVK accessories
Test Cables (pairs)
PC3.5 (f)/PC3.5 (m), 50 (for ZVM)
2.92 mm (f)/2.92 mm (m), 50 (for ZVK)
Calibration Kits
PC3.5 (for ZVM) ZV-Z32 0 GHz to 26.5 GHz 1128.3501.02 PC3.5 incl. Sliding Matches
(for ZVM)
2.92 mm (for ZVK) ZV-Z34 0 GHz to 40 GHz 1128.3530.02
2.92 mm incl. Sliding Matches (for ZVK)
N, 50 ZV-Z21 0 GHz to 18 GHz 1085.7099.02 TRL Supplementary Kit,
N, 50 TRL Supplementary Kit,
PC3.5, 50 TOM-X Supplementary Kit,
N, 50 TOM-X Supplementary Kit,
PC3.5, 50
6)
6)
ZV-Z14 0 GHz to 26.5 GHz 1134.4093.02
ZV-Z15 0 GHz to 40 GHz 1134.4193.02
ZV-Z33 0 GHz to 26.5 GHz 1128.3518.02
ZV-Z35 0 GHz to 40 GHz 1128.3547.02
ZV-Z26 0.4 GHz to 18 GHz 1085.7318.02
ZV-Z27 0.4 GHz to 26.5 GHz 1085.7401.02 ZV-Z28 0 GHz to 18 GHz 1085.7499.03
ZV-Z29 4 GHz to 26.5 GHz 1085.7647.03
Sliding Matches
N (m), 50 ZV-Z41 1.7 GHz to 18 GHz 1085.8095.02 N (f), 50 ZV-Z41 1.7 GHz to 18 GHz 1085.8095.03 PC3.5 pair m, f (for ZVM) ZV-Z42 0 GHz to 26.5 GHz 1128.3524.02
2.92 mm pair m, f (for ZVK) ZV-Z44 0 GHz to 40 GHz 1128.3553.02
General accessories
Hardware Options N, 50
AutoKal7) ZVR-B1 0 GHz to 8 GHz 1044.0625.02 3-Port Adapter 4-Port Adapter (2 x SPDT)7)ZVR-B14 0 GHz to 4 GHz 1106.7510.02 4-Port Adapter (SP3T)
Test Cables (pairs)
N (m)/N (m), 50 ZV-Z11 0 GHz to 18 GHz 1085.6505.03 N (m)/N (m), 75 ZV-Z12 0 GHz to 4 GHz 1085.6570.02 N (m)/PC3.5 (m), 50 ZV-Z13 0 GHz to 18 GHz 1134.3997.02
Calibration Kits
N, 50 ZCAN 0 GHz to 3 GHz 0800.8515.52 N, 75 ZCAN 0 GHz to 3 GHz 0800.8515.72
Attenuators
1 W DNF 0 GHz to 12.4 GHz 50 W RBU 50 0 GHz to 2 GHz 100 W RBU 100 0 GHz to 2 GHz
Matching Pads, N, 50 Ω → N, 75
Series Resistor RAZ 0 GHz to 2.7 GHz 0358.5714.02 L Section RAM 0 GHz to 2.7 GHz 0358.5414.02
Various Accessories, N, 50
T Check ZV-Z60 0 GHz to 4 GHz 1108.4990.50 Bias Network ZV-Z61 2 MHz to 4 GHz 1106.8130.02 DC Block FSE-Z3 5 MHz to 7 GHz 4010.3895.00 Power Splitter 2 x 50 RVZ 0 GHz to 2.7 GHz 0800.6612.52
External SWR-Bridges
N (f), 50 ZRA 40 kHz to 150 MHz 1052.3607.52 N (f), 50 ZRB 2 5 MHz to 3 GHz 0373.9017.52 N (f), 75 ZRB 2 5 MHz to 2 GHz 0802.1018.73 N (f), 50 ZRC 40 kHz to 4 GHz 1039.9492.52 N (f), 75 ZRC 40 kHz to 2.5 GHz 1039.9492.72
Miscellaneous
Transit Cas e ZZK-965 1013.9437.00 19“-Rack Adapter with front
handles
1)
Harmonics and arbitrary frequency conversion measurement included.
2)
Power meter and sensor required.
3)
Only for ZVR, ZVC, ZVM, ZVK.
4)
Com prise s test po rt ’Inp ut b1’, f or bypa ssing c ouple r at PORT 1 .
5)
Com prise s test po rt ’Inp ut b2’, f or bypa ssing c ouple r at PORT 2 .
6)
For ruggedized port.
7)
Two a dapte rs PC 3.5 (f) /N (f) or 2 .92 mm (f )/N (f) req ired.
8)
X = 0: 3 dB, X = 1: 6 dB, X = 2: 10 dB, X = 3: 20 dB, X = 4: 30 dB.
9)
XX = 03: 3 dB, XX = 06: 6 dB, XX = 10: 10 dB, XX = 20: 20 dB, XX = 30: 30 dB.
7)
ZVR-B8 0 GHz to 4 GHz 1086.0000.02
7)
ZVR-B14 0 GHz to 4 GHz 1106.7510.03
0272.4X10.50
1073.8695.XX
1073.8495.XX
ZZA-96 0396.4928.00
8)
9)
9)
Vector Network Analyzers ZVM, ZVK 13
General data
Temperature loading
Specs complied with 5 °C to 40°C Operational 0°C to 50°C Storage temperature range 40°C to +70 °C
meets IEC68-2-1, IEC 68-2-2
Damp heat 40 °C at 95% rel. humidity,
meets IEC68-2-3
Mechanical resistance
Vibration test, sinusoidal 10 Hz to 55 Hz, max. 2 g,
55 Hz to 150 Hz, 0.5 g constant, 12 min/axis, meets IEC68-2-6, IEC1010-1, MIL-T-28800D class 5
Vibration test, random 10 Hz to 300 Hz, 1.2 g rms,
5 min/axis, meets IEC68-2-36
Shock test 40 g shock spectrum, method 516.3,
meets MIL-STD-810D, MIL-T-28800D classes 3 and 5
Calibration interval 1 year
EMC, emission following the provisions of Directives
89/336/EEC, revised by 91/263/EEC, 92/31/EEC, 93/68/EEC and EN50081-1
EMC, immunity following the provisions of Directives
89/336/EEC, revised by 91/263/EWG, 92/31/EEC, 93/68/EEC and EN50082-1
Safety meets EN 61010-1, UL3111-1,
CSAC22.2 No. 1010-1, IEC 1010-1
Power supply 100 V to 120 V (AC) with tolerance
±10%, 6 A, 50 Hz to 400 Hz with toler­ance 6 % and +10% or 200 V to 240 V (AC) with tolerance ±10%, 3 A, 50 Hz to 60 Hz with toler­ance 6% and +10%
safety class I to VDE411 Power consumption 280 W (standby: 10 W) Test mark VDE, GS, CSA, CSA-NRTL/, c mark Dimensions (W x H x D) 435 mm x 281 mm x 584 mm Weight 30 kg
Printed in Germany 1100 (U ko)
ROHDE& SCHWARZ GmbH & Co. KG Mühldorfstraße 15 ⋅ 81671 München ⋅ Germany P.O.B. 8014 69 81614 München ⋅ Germany
Telephone +49894129-0 ⋅ www.rohde-schwarz.com ⋅ CustomerSupport: Tel. +49 1805124242, Fax +49 89 4129-13777, E-mail: CustomerSupport@rohde-schwarz.com
PD 757.5543.22Vector Network Analyzers ZVM, ZVKTrade names are trademarks of the ownersSubject to changeData without tolerances: typical values