Rohde and Schwarz ZVB20 Data Sheet

Vector Network Analyzer ¸ZVB
Specifications
Version
06.00
August
2006
Specifications
Specifications apply under the following conditions:
90 minutes warm-up time at ambient temperature, specified environmental conditions met, calibration cycle adhered to, and all internal adjustments performed. Data designated „overrange“ and data without tolerances: typical values only. Unless stated otherwise, specifications apply to test ports and a nominal source power of –10 dBm.
CONTENTS
Measurement range ...................................................................................................................................................................................3
Measurement speed...................................................................................................................................................................................5
Sweep times of R&S®ZVB ......................................................................................................................................................................5
Measurement accuracy ..............................................................................................................................................................................6
Effective system data .................................................................................................................................................................................9
Test port output ..........................................................................................................................................................................................9
Test port input...........................................................................................................................................................................................12
Additional front panel connectors .............................................................................................................................................................12
Display......................................................................................................................................................................................................13
Rear panel connectors .............................................................................................................................................................................13
Options .....................................................................................................................................................................................................14
General data.............................................................................................................................................................................................14
Ordering information.................................................................................................................................................................................15
Vector Network Analyzer R&S®ZVB Version 06.00, August 2006
2
Measurement range
Impedance
R&S®ZVB4 and R&S®ZVB8 type N, female Test port connector
®
ZVB20 3.5 mm, male
R&S
50 Ω
Number of test ports 2, 3, or 4
Frequency range
Static frequency accuracy
R&S®ZVB4 300 kHz to 4 GHz
R&S®ZVB8 300 kHz to 8 GHz
®
R&S
ZVB20 10 MHz to 20 GHz
without optional oven quartz 8 × 10
with optional oven quartz 1 × 10
–6
–7
Frequency resolution 1 Hz
Number of measurement points per trace 2 to 20001
Measurement bandwidths 1/2/5 steps 1 Hz to 500 kHz
Dynamic range of R&S®ZVB4 two-port model and R&S
®
ZVB8 two-port model
(without optional step attenuators)
from PORT 1 to PORT 2
300 kHz to 5 MHz >80 dB, typ. 100 dB
5 MHz to 100 MHz >100 dB, typ. 120 dB
100 MHz to 4 GHz >120 dB, typ. 130 dB
4 GHz to 7 GHz (R&S®ZVB8 only) >117 dB, typ. 127 dB
Dynamic range of R&S®ZVB4 three-port and four-port models and R&S
®
ZVB8 three-port and four-port models (without optional step attenuators)
7 GHz to 7.5 GHz (R&S®ZVB8 only) >115 dB, typ. 120 dB
7.5 GHz to 8 GHz (R&S
from PORT 1 to PORT 2 and from PORT 3 to PORT 4 (for four-port models only)
300 kHz to 5 MHz >80 dB, typ. 100 dB
5 MHz to 100 MHz >100 dB, typ. 120 dB
®
ZVB8 only) >108 dB, typ. 118 dB
100 MHz to 500 MHz >120 dB, typ. 130 dB
500 MHz to 4 GHz >123 dB, typ. 130 dB
4 GHz to 7 GHz (R&S®ZVB8 only) >120 dB, typ. 130 dB
Dynamic range of R&S®ZVB20
7 GHz to 7.5 GHz (R&S®ZVB8 only) >115 dB, typ. 125 dB
7.5 GHz to 8 GHz (R&S
from PORT 1 to PORT 2 and from PORT 3 to PORT 4 (for four-port model only)
®
ZVB8 only) >108 dB, typ. 125 dB
10 MHz to 100 MHz >80 dB, typ. 110 dB
100 MHz to 700 MHz >100 dB, typ. 130 dB
700 MHz to 8 GHz >120 dB, typ. 133 dB
8 GHz to 20 GHz >110 dB, typ. 122 dB
The dynamic range is defined as the difference between the maximum source power and the rms value of the data trace of the transmission magnitude produced by noise and crosstalk with test ports short-circuited. The specification is valid without system error correction and at 10 Hz measurement bandwidth. The dynamic range can be increased by using a measurement bandwidth of 1 Hz.
Version 06.00, August 2006 Vector Network Analyzer R&S®ZVB
3
Dynamic range in dB versus frequency in GHz of R&S
®
ZVB4
Dynamic range in dB versus frequency in GHz range of R&S
®
ZVB8
®
Dynamic range in dB versus frequency in GHz range of R&S
ZVB20
Vector Network Analyzer R&S®ZVB Version 06.00, August 2006
4
Measurement speed
Measurement time
Measurement time per point CW mode, 500 kHz measurement
Data transfer time
Time for measurement and data transfer for 201 measurements points, with start
Switching time between channels with not more than 2001 points <1 ms
Switching time between two preloaded instrument settings
for 201 measurements points, with span 100 MHz, 500 kHz measurement bandwidth, ALC and display switched off
with center frequency 1.1 GHz <6 ms
with center frequency 5.1 GHz <4.5 ms
bandwidth
for 201 measurements points
via IEC/IEEE bus <2.9 ms
via VX11 over 100 Mbit/s LAN <1.3 ms
via RSIB over 100 Mbit/s LAN <0.7 ms
frequency 1 GHz, stop frequency 1.1 GHz, 500 kHz measurement bandwidth, and display switched off
(No additional time for data transfer is needed, as it is done simultaneously during the measurement.)
with not more than 2001 points <10 ms
<4.5 µs
<6 ms
Sweep times of R&S®ZVB
Number of measurement points 51 101 201 401 801 1601
Start frequency 5 GHz, stop frequency 5.2 GHz, ALC off, and measurement bandwidth of 100 kHz
with full one port calibration or with correction switched off
with TOSM calibration 4.7 ms 8.6 ms 16.4 ms 32.4 ms 65 ms 170 ms
Start frequency 6 GHz, stop frequency 8 GHz, ALC off, and measurement bandwidth of 100 kHz
with full one port calibration or with correction switched off
with TOSM calibration 5.3 ms 9.8 ms 18 ms 33 ms 63 ms 160 ms
Start frequency 10 MHz, stop frequency 4 GHz (R&S®ZVB4), 8 GHz (R&S®ZVB8), or 20 GHz (R&S®ZVB20),
with full one port calibration or with correction switched off
with TOSM calibration 10.3 ms 16.6 ms 28 ms 47 ms 81 ms 190 ms
2.4 ms 3.9 ms 6.3 ms 11 ms 20.4 ms 40.2 ms
3.4 ms 6.2 ms 11 ms 17.3 ms 28.2 ms 49 ms
ALC off, and measurement bandwidth of 100 kHz
8.4 ms 12.6 ms 19.5 ms 30.5 ms 53.2 ms 88.2 ms
Version 06.00, August 2006 Vector Network Analyzer R&S®ZVB
5
Loading...
+ 11 hidden pages