Opening the instrument and removing MOTHERBOARD2............................ 3.52
Installing the new MOTHERBOARD2 and completing the instrument........... 3.52
FRONTPANEL BOARD (at the front of the instrument frame) ...................................3.52
Opening the instrument and removing the FRONTPANEL BOARD ..............3.52
Installing the new FRONTPANEL BOARD and completing the instrument ... 3.52
REARPANEL BOARD1 (Interface board for Standard Connectors at the Rear of the
Module and Cable Exchange.......................................................................................................... 5.1
1100.4903.827E-4
TablesCMU
Tables
Table 1-1 Measuring instruments and auxiliary equipment for manual performance test.....................1.2
Table 1-2 Test report ...........................................................................................................................1.18
Table 2-1 Measuring instruments and auxiliary equipment for manual adjustment of the CMU ........... 2.1
Table 5-1 List of power cables available................................................................................................ 5.2
1100.4903.828E-4
CMUManuals
Contents of Manuals for Universal Radio Communication Tester CMU200/CMU300
Service Manual Instrument
This service manual for Universal Radio Communication Tester CMU provides information on
checking the generator for compliance with rated specific ations, as well as on adj ustm ent, repair and
troubleshooting. It further contains all the information neces sary for repairing the gener ator by the replacement of modules.
The service manual compris es five chapters and an annex (chapter 5) containing the CMU circuit doc umentation:
Chapter 1Provides all the information necessar y to c heck CMU for compliance with rated
specifications. The required test equipment is included, too.
Chapter 2Describes the adjustment of the +5 VDC reference source and of the 10-MHz
reference frequency source as well as the software- controlled adjustm ent of individual module data following module replacement.
Chapter 3Describes the design of CMU as well as simple measures for repair and fault
diagnosis, in particular, the replacem ent of m odules and ac cess to hardware s ettings by means of service commands.
Chapter 4Contains information on the extension and modification of CMU by installing
instrument software and retrofitting options.
Chapter 5Contains spare parts lists and exploded views of CMU.
Operating Manual
In the operating manual for CMU basic unit you will find information about the technical specif ic ations
of CMU, the controls and connectors on the f ront and rear panel, necess ary steps for putting the instrument into operation, the basic operating concept, manual and remote control.
For introduction typical measurem ent tasks are explained in detail us ing the f unctions of the user interface and program examples.
The operating manual contains all information on the charac teristic s, operation and r em ote control of
the CMU including RF and Audio measurements. It f urther provides hints on preventive m aintenance
and fault diagnosis by means of warnings and error messages output by the unit.
Service Manual Modules
The service manual modules is not delivered with the instrument but may be obtained from your R&S
service department using the order number 1100.4903.92.
Service manual modules contains inform ation about the individual modules of CMU. This com prises
the test and adjustment of the m odules, f ault detection within the modules and the inter face desc ription.
1100.4903.820.1E-4
CMUDocumentation
Operating Manuals for Digital and Analog Network Tests
The operating manuals listed in the f ollowing table describe the test of mobile phones supporting
different standards by means of the CMU200/CMU300 and the appropriate software and hardware
options. Except for hardware-spec ific chapters that ar e not relevant to the software options, the network test operating manuals are organized like the CMU operating manual.
For Options
ManualOrder NumberTypeDescriptionStock No.
Operating Manual
CMU-K21/-K22/-K23
Operating Manual
CMU-K27/-K28
Operating Manual
CMU-K29
Operating Manual
CMU-K30/-K31/K32/-K33
Operating Manual
CMU-K81/-K82
Operating Manual
CMU-K53
1115.6088.12CMU-K21
CMU-K22
CMU-K23
1115.6688.12CMU-K27
CMU-K28
1115.6888.12CMU-K29AMPS-MS for CMU-B211115.6807.02
1115.4185.12CMU-K30
CMU-K31
CMU-K32
CMU-K33
CMU-K39
CMU-K41
1115.5581.12CMU-K81
CMU-K82
1115.5081.12CMU-K53Bluetooth for CMU1115.5000.02
GSM900-MS for CMU-B21
GSM1800-MS for CMU-B21
GSM1900-MS for CMU-B21
TDMA800-MS for CMU-B21
TDMA1900-MS for CMU-B21
GSM400-BS for CMU-B21
GSM900-BS for CMU-B21
GSM1800-BS for CMU-B21
GSM1900-BS for CMU-B21
MOC/MTC
EDGE for CMU-K30/31/32/33
CDMA800-MS (IS95) for CMU-B81
CDMA1900-MS (IS95) for CMU-B81
1115.6007.02
1115.6107.02
1115.6207.02
1115.6607.02
1115.6707.02
1115.4004.02
1115.4104.02
1115.4104.02
1115.4104.02
1115.4791.02
1115.4604.02
1115.5500.02
1115.5600.02
The GSM base station tests described in operating manual CMU-K30/-K31/-K32 r equire a CMU300
(Universal Radio Comm unic ation Tester for BT S). All other r adio c ommunication equipment is tested
with model CMU200.
1100.4903.820.2E-4
CMUGeneral
1 Performance Test
This chapter provides the necessary information for checking the technical data of the CMU. Please
read the general notes on the test procedure on page 1.7 first. Then follows a list of the measuring
equipment required for the perf ormance test; a f orm for the tes t report is to be f ound at the end of this
chapter.
The adjustment of the instrument f or restoring the data integrity and the measuring equipment r equired
for this purpose will be described in chapter 2 of this service manual.
General
The technical data of a CMU can be checked in the following ways:
A) Calibration by an R&S Representative with an ACS Calibration System
Advantages
• Automatic procedure
• Small measurement uncertainties
• Calibration and readjustment
• In most cases, the instrument does not have to leave the country
B) Sending the Instrument to the Factory (Memmingen, Germany)
Advantages
• Automatic procedure at the final test setup
• Minimum measurement uncertainties
• Calibration and readjustment
Necessary Documents
• Operating manual CMU200/CMU300
• Data sheet CMU200
Note:It is recommended to read the following journal on the s ubjects "meas urement uncertainty "
and "tolerance analysis": ETSI Technical Report ETR 028, June 1997
1100.4903.821.1E–4
Measuring Instruments and Auxiliary EquipmentCMU
Measuring Instruments and Auxiliary Equipment
Table 1-1Measuring instruments and auxiliary equipment for manual performance test
Item Type of instrumentRequired characteristicsAppropriate deviceR & S order
1Signal generator100 kHz to 2.7 GHz,
2Spectrum analyzer100 kHz to 7 GHz,
3Power meterNRVD with sensors
4Power amplifier100 kHz to 2.7 GHz,
5Harmonics filterattenuate the harmonics of the
6Directional coupler50 MHz to 2.7 GHz,
7Network analyzer
or VSWR Bridge
Generation of a GSM signal
(dummy burst)
Demodulation of GSM signals
Pout = 100 W
power amplifiers to min 30 dBc
up to 100 W
100 kHz to 2.7 GHzZVR
SME03
SMIQ
FSE with FSE-B7
FSIQ
FSP–3
NRV-Z4
NRV-Z51
ZRC
number
1038.6002.03RX measurements
1066.30.10.20
1066.30.10.30
1066.4317.02
1093.4495.03
1043.0009.61Reflection coefficient/
Use
TX measurements
RX measurements,
TX measurements
RX measurements
RX measurements
RX measurements
VSWR
RF connectors
1100.4903.821.2E–4
CMUMeasuring Instruments and Auxiliary Equipment
Test Setups
The quality of the test setup has an effect on the measurement procedures.
Note:Make sure to use only high-quality coax cables and coax connectors as well as calibrated
measuring equipment.
Reference Frequencies
Test setup REF1:
REF1
Spectrum
Analyzer
(FSE
or
FSIQ)
GEN
REFIN
DUT
(CMU)
10 MHz Reference Freq.
RF3 OUT
Test setup REF2:
REF2
Spectrum
Analyzer
(FSE or
FSIQ)
DUT
(CMU)
REF IN
REF OUT 2
10 MHz Reference Freq.
TX Level Measurements
Test setup TX1, TX2, TX3, TX4 (depending on level range):
½ Normalize spectrum analyzer (FSIQ) to wideband power meter (NRVD) at Max. Level setting of the
CMU (test setups TX1, TX2, TX3).
The attenuator of the spectrum analyzer must be held at this position over 60 dB.
½ Normalize spectrum analyzer with preamplifier to spectrum analyzer at last level (test setups TX 3,
TX4).
The attenuator of the spectrum analyzer must be held at this position over 60 dB.
TX1
DUT
(CMU)
.. +5 dBm
RF3OUT
NRV-Z51
NRVD
1100.4903.821.3E–4
Measuring Instruments and Auxiliary EquipmentCMU
TX2
-33 dBm/
DUT
(CMU)
TX3
DUT
(CMU)
TX4
-16 dBm
RF1 /
RF2
-93 dBm to -33 dBm/
-76 dBm to -16 dBm/
-55 dBm to +5 dBm
RF1 /
RF2 /
RF3 OUT
10 MHz Reference Freq.
NRV-Z4
FSIQ or FSE
NRVD
-130 dBm to -93 dBm/
-117 dBm to -76 dBm/
DUT
(CMU)
-90 dBm to -55 dBm
RF1/
RF2/
RF3 OUT
10 MHz Reference Freq.
General TX measurements
Test setup TX5:
TX5
AUX3, pin2
DUT
(CMU)
RF1, RF2, RF3
10 MHz Reference Freq.
RF-PreAmp
36 dB
FSE
or FSIQ
ext. Trigger input
Spectrum
analyzer
(FSE or
FSIQ)
RX Level measurements
Test setup RX1, RX2 (depending on level range):
1100.4903.821.4E–4
CMUMeasuring Instruments and Auxiliary Equipment
RX1
RX2
GEN
GEN
Power
Ampl.
LPF
depending on max.
level of power splitter
Power
Ampl.
LPF
Power
Splitter
direc t . co upl
-30 dB
NRVZ4
NRV-
Z4
NRVD
DUT
(CMU)
NRVD
DUT
(CMU)
General RX Measurements
Test setup RX3:
RX3
GEN
RF1,RF2,RF4IN
10 MHz Reference Freq.
RX Demodulation Measurements
Test setup RX4:
RX4
GEN
RF4IN
DUT
(CMU)
DUT
(CMU)
IF3RX CH1
Demodulator
(FSE with
FSE-B7 or
FSIQ)
10 MHz Reference Freq.
10 MHz Reference Freq.
1100.4903.821.5E–4
Tolerance AnalysisCMU
Tolerance Analysis
Due to the small measur ement uncertainty of the CMU, the measuring equipm ent must m eet stringent
requirements. Since the measur ement uncer tainty of the measuring equipm ent to be achieved depends
on the test setup used, it is recommended to perform a tolerance analysis.
To be able to trace back errors in the measurement, the measurement uncertainty should also be
indicated in the test report.
The tolerances given in the test report refer to the values specified in the data sheet, ie the
measurement uncertainties of the test setup used are to be taken into account as well.
Unless otherwise specified, the specified tolerances are always to be observed.
Note:Please take into account the ETSI Technical Report ETR 028.
The given tolerances refer to CMU data sheet 04/99.
1100.4903.821.6E–4
CMUManual Test Procedure
Manual Test Procedure
Some additional measurem ents can only be performed us ing a mobile phone via the normal operating
menus of the CMU. These measurements are described in the section ’Function Test with MobileStations'.
The suggested frequencies and levels at which the measurements should be performed have been
selected according to the instrument concept. The user can of cour se also select other frequencies and
levels within the scope of values guaranteed in the data sheet.
Note:Before testing the rated specifications, allow the instrument to warm up for at least 15 min.
The ambient temperature should be 23 °C to 26 °C.
Selftest
The CMU offers various selftest options for checking the functioning and for troubleshooting. Before
carrying out the performance test, the Maintanance m enu in the BASE function gr oup should be called
up and the following selftests should be performed:
• Continuous Selftest
• 1Å4/3Å2 RF Loop Test
Preparation:
Test:½ Select the individual test in the BASE (MAINTENANCE) menu and chec k the res ults
Note:The Continuous Selftest combines the System Self test and Internal RF Loop Selftest. In
½ Start user SW of the CMU (switch on instrument).
(Continuous Selftest, 1Å4/3Å2 RF Loop Test).
this test, only a passed/failed message with error output is indicated. In case of err ors, an
error file ‘cst.err’ is created in addition.
1Å4/3Å2 RF Loop Test: Selftest RF Path RF 1ÅRF4 IN and RF3 OUTÅRF2 via external
N-coax cable by power measurements via internal generator and analyzer. In this tes t, all
measured values are indicated.
To obtain more detailed information start the following tests:
System selftest:
Internal RF Loop Selftest:
FE Selftest:
REF Selftest:
Selftest of the instrument for diagnostic voltages; only a
passed/failed message with error output is indicated.
Selftest of the RF path by means of power measur ements
via internal generator and analyzer. All measured values
are indicated.
Selftest of the RF FRONTEND module via diagnostic
voltages. All measured values are indicated.
Selftest of the REFERENCE BOARD module via diagnostic
voltages. All measured values are indicated.
DIG Selftest:
RXTX1 Selftest:
1100.4903.821.7E–4
Selftest of the DIGITAL BOARD module via diagnostic
voltages. All measured values are indicated
Selftest of the RXTX1 BOARD module via diagnostic
voltages. All measured values are indicated.
Manual Test ProcedureCMU
Interface Test
IEC-bus Interface
Preparation:
Test:½ Send the string ‘*IDN?<CR><NL>’ from the controller to the CMU and read the
½ Connect the IEC-bus interfaces of the CMU and the controller via IEC-bus
cables.
reply STRING of the CMU.
The reply STRING must contain the following message:
Test:½ Send the string ‘*IDN?<CR><NL>’ from the controller to the CMU and read the
½ Connect the RS–232 interfaces of the CMU and the controller via null-modem
cables (cf. page 5.1.1, "Measuring Instruments and Auxiliary Equipment").
Set the RS–232 interface of the controller to 8 data bits, 1 s tart bit, 1 stop bit, no
parity bit, XON/XOFF handshake and 19200 baud.
reply string of the CMU.
The reply string must contain the following message:
‘ROHDE & SCHWARZ,CMU<Var>,<Ser_Nr>,<Firmware_Vers._No>’
Reference Frequencies
For different range of adjustm ent of the internal r ef er enc e osc illator , the synchronization with an external
reference frequency is checked.
Note:The resolution of the frequency counter/analyzer should be max. 1/10 of the maximum
permissible deviation.
REF IN
The frequency and level ranges and the pull-in range are checked. The signal generator and the
frequency counter must be synchronized (test setup REF1).
Test:Use frequency counter to measure frequency deviation from 1 GHz.
½ Feed in at REF IN: 52 MHz sinewave, 0.5 V(rms)
½ Connect frequency counter to RF3 OUT.
GENERATOR.)
1100.4903.821.8E–4
CMUManual Test Procedure
Preparation:
Control:½ Set CMU to external reference 1 MHz, RF 1 GHz.
Test:½ Use frequency counter to measure frequency deviation from 1 GHz.
Preparation:
Control:½ Set CMU to external reference 10 MHz, RF 1 GHz.
Test:½ Use frequency counter to measure frequency deviation from 1.000 005 000 GHz.
Preparation:
Control:½ Set CMU to external reference 10 MHz, RF 1 GHz.
Test:½ Use frequency counter to measure frequency deviation from 0.999 995 GHz.
½ Feed in at REF IN: 1 MHz sinewave, 0.5 V(rms).
½ Connect frequency counter to RF3 OUT.
½ Feed in at REF IN: 10.000 050 MHz sinewave, 1.4 V(rms).
½ Connect frequency counter to RF3 OUT.
½ Feed in at REF IN: 9.999 950 MHz sinewave, 1.4 V(rms).
½ Connect frequency counter to RF3 OUT.
REF OUT 1
The level and frequency are checked.
Int. 10 MHz:
REF IN
signal:
½ Set CMU to internal reference.
½ Measure at REF OUT 1: 10 MHz, level > 1.4 V(pp).
½ Set CMU to external reference.
½ Feed in at REF IN: 52 MHz TTL, (as an alternative +16 dBm from signal)
½ Measure at REF OUT 1: 52 MHz, level > 1.4 V(pp).
½ Measure at REF OUT 1: frequency 52 MHz ± 1 Hz
½ Feed in at REF IN: 1 MHz TTL, (as an alternative +16 dBm from signal
generator).
½ Measure at REF OUT 1: 1 MHz, level > 0.5 V(rms) (1.4 V(pp)).
½ Measure at REF OUT 1: frequency 1 MHz ± 1 Hz.
REF OUT 2
The level and frequency are checked. T he CMU and the fr equency counter mu st be synchronized (test
setup REF2).
REF OUT 2
signal 13 MHz:
½ Set CMU to external reference.
½ Feed in at REF IN: 10 MHz sinewave, 0.5 V(rms).
½ Set CMU to REF OUT 2 13 MHz or 10 MHz (depending on SW; menu
Connection Control/Synch.).
½ Measure at REF OUT 2: 13 MHz or 10 MHz, level > 1.0 V(pp).
½ Measure at REF OUT 2: frequency 13 MHz or 10 MHz ± 1 Hz.
1100.4903.821.9E–4
Manual Test ProcedureCMU
TX Frequency Accuracy
Preparation:Test setup TX5, but CMU not synchronized with frequency counter/analyzer and
no external trigger.
CMU connector RF3 OUT.
Control:½ Set CMU to desired frequency, level 0 dBm.
Test:½ Determine frequency deviation from nominal frequency.
TX Frequency Settling Time
Preparation:Test setup TX5, in addition tr igger cable f rom CMU (D-s ub c onnector AUX3, pin2)
to analyzer.
CMU connector RF3 OUT.
CMU: Ramping off, hopping on, F1 = start freq., F2 = stop freq.
Analyzer: Sweep time 1 ms, Center = stop frequenc y, FM demodulation,
real time off, BW 50 kHz, 1 kHz/Div, external trigger, Slope
negative
Control:½ Set CMU to desired frequencies and hopping, level 0 dBm.
Test:Time from trigger point when the specified offset (< 1 kHz) from the stop
frequency is reached.
TX Level Error
Preparation:Test setup TX1 to TX4 (depending on level range).
Control:½ Set CMU to desired connector, frequency and level (RF Analyzer must be
OFF).
Test:½ Measure the TX level of the CMU.
Note:The given frequencies and levels are suggested values. Of course, it is also possible to use
other values for the measurement.
VSWR
Preparation:
½ Connect (scalar ) network analyzer to RF1, RF2, RF3 OUT, RF4 IN one after
CMU: Switch generator on and set level to minimum (–130 dBm or
–90 dBm), switch RF wideband analyzer on (RF1/RF2/RF4 IN).
CMUManual Test Procedure
TX Level Settling Time
Preparation:
Control:½ Set CMU to frequency = 1 GHz, specified level and ramping mode.
Test:½ The time period fr om the trigger point to the point in time when the nominal
Test setup TX5, in addition tr igger cable f rom CMU (D-s ub connec tor AUX3, pin2)
to analyzer.
CMU: Connector RF3 OUT 1GHz, Ramping On, Hopping Off.
Analyzer: Sweep time = 40 µs, Center = 1 GHz , Span = 0, RBW = 10
MHz, external trigger.
level < 0.5 dB has been reached is measured.
TX Harmonics
Preparation:Test setup TX5, no external trigger
Analyzer:Center = 2 x f
Control:½ Set CMU to connector RF1, specified frequency, level = –27 dBm.
Test:The suppression of the signal at twice or three times the nominal frequency is
measured relative to the nominal signal.
Control:½ Set CMU to connector RF2, specified frequency, level = –10 dBm.
or Center = 3 x f
nom
, Span = 1 MHz.
nom
Test:The suppression of the signal at twice or three times the nominal frequency is
measured relative to the nominal signal.
Control:½ Set CMU to connector RF3 OUT, specified frequency, level = +10 dBm.
Test:The suppression of the signal at twice and three times the nominal f requency is
measured relative to the nominal signal.
TX Non Harmonics
TX In-Band Spurious Signals
Spurious signals within the specified frequency bands are checked.
PreparationTest setup TX5, no external trigger
AnalyzerStart/Stop = specified frequency range, RBW = 100 kHz
Control:½ Set CMU to connector RF3 OUT, specified setting frequency, level = 0 dBm.
Test:The suppression of the signal is m easured at the tes t frequenc y relative to the set
signal.
1100.4903.821.11E–4
Manual Test ProcedureCMU
TX Fixed Spurious Signals
Fixed spurious signals are checked.
Preparation:Test setup TX5, no external trigger.
Analyzer:Center = specified test frequency, RBW = 100 kHz, Span = 1 MHz.
Control:½ Set CMU to connector RF3 OUT, specified setting frequency, specified level.
Test:The suppression of the signal is m easured at the tes t frequenc y relative to the set
signal.
TX SSB Phase Noise
PreparationTest setup TX5, no external trigger
½ Connect spectrum analyzer or modulation analyzer to RF3 OUT.
Control:½ Set CMU generator to specified RF frequency.
Output level at RF3 OUT 0 dBm,
Analyzer to specified center frequency,
Span = 50 kHz to 5 MHz, RBW = Span/500,
Noise measurement function.
Test½ Measure the phase noise at the specified spacing from the carrier.
TX Residual FM
Preparation:Test setup TX5, no external trigger.
½ Connect spectrum analyzer or modulation analyzer to RF1.
Control:½ Set CMU generator to the specified RF frequency.
Output level at RF1 –27 dBm, analyzer to specified center frequency, FM
demodulator.
Test:½ The residual FM with the specified weighting is measured.
TX Residual AM
Preparation:Test setup TX5, no external trigger.
½ Connect spectrum analyzer or modulation analyzer to RF1.
Control:½ Set CMU generator to specified RF frequency.
Output level at RF1 –27 dBm, analyzer to specified center frequency, AM
demodulator.
Test:½ The residual AM with the specified weighting is measured.
1100.4903.821.12E–4
CMUManual Test Procedure
TX Carrier/Sideband Suppression, Max. Distorti on
The modulation quality of the analog IQ modulator of the CMU is measured.
PreparationTest setup TX5, no external trigger.
½ Connect spectrum analyzer to RF3 OUT.
Control:½ Set CMU generator to specified RF frequency.
Output level at RF3 OUT, 0 dBm,
½ Switch on RF generator with offset modulation, 300- kHz baseband filter,
½ Set analyzer to center frequency f
= 1000 MHz, Span = 300 kHz / 3 MHz
c
Test:The suppression of the carrier at f
+ f
signal at f
.
c
mod
is measured relative to the useful sideband
c
RX Power Meter (Frequency-Selective)
Preparation:Test setup RX1, RX2 (depending on level range).
Control:½ Set CMU to desired RX frequency and level and Input in menu RF
ANALYZER/ GENERATOR.
Measuring Bandwidth = 1 kHz.
Test:½ Measure RX level measurement accuracy of CMU.
Note:The given frequencies and levels are s uggested v alues ; of c ours e, it is als o poss ible to us e
other values for the measurement.
RX Power Meter (Wideband)
PreparationTest setup RX1, RX2 (depending on level range).
Control:½ Set CMU to desired RX connector, frequency and level and Input in menu RF
Analyzer/ Generator.
Measuring Bandwidth =Wide
Test:½ Measure RX level measurement accuracy of CMU.
Note:T he given frequencies and levels are suggested values; of cours e, it is also possible to use
other values for the measurement.
RX Harmonics
Preparation:Test setup RX3,
Generator = f
Control:½ Set CMU to connector RF2, Max Level = 2 dBm.
1100.4903.821.13E–4
; level = 0 dBm.
in
Manual Test ProcedureCMU
Test:½ The suppress ion of the signal at twice and three times the input frequency is
measured relative to the input signal.
Preparation:
Test setup RX3
Generator = f
in
; level = –2 dBm
Control:½ Set CMU to connector RF4 IN, Max Level = 0 dBm.
Test:½ The suppression of the s ignal at twice and three times of input frequency is
measured relative to the input signal.
RX Spurious Response / Image Rejection
Preparation:Test setup RX3,
Generator = f
Control:½ Set CMU to connector RF2, Max Level = 2 dBm, Mode = Low Distortion
Test:½ The suppression of the spurious or image signal is measured relative to the
input signal.
Preparation:
Test setup RX3,
Generator = f
Control:½ Set CMU to connector RF4 IN, Max level = 0 dBm, Mode = Low Distortion
; level = 0 dBm.
in
; level = –2 dBm.
in
Test:The suppression of the spurious or im age signal is meas ured relative to the input
signal.
RX SSB Phase Noise
Preparation:Test setup RX3,
Generator = f
Control:½ Set CMU to connector RF2, Max Level = 10 dBm and to desired frequency.
½ Switch on frequency-selective power meter.
Test:The measurement is taken with a small test bandwidth at different carrier offsets.
Note: The input level is +10 dBm, RBW = 100 Hz
+ df ; level = 10 dBm.
iCMU
--> Phase noise = measured value –10 dB –21 dB.
1100.4903.821.14E–4
CMUManual Test Procedure
RX Residual FM/AM
Preparation:Test setup RX4,
Generator = f
Control:½ Set CMU to connector RF4 IN, Max Level = –20 dBm and to desired
frequency.
½ Switch on frequency-selective power meter.
Test:The measurement is taken with an external FM/AM demodulation instrument (FSE
with FSE-B7) via the IF3RXCH1 BNC connector at the rear panel of the CMU.
½ Set the FSE to desired frequency, ref. level, AF filters.
; level = –20 dBm.
iCMU
RX Dynamic/ Average Noise Level
Preparation:No signal is fed in, CMU generator is OFF.
Control:Set CMU to given receive frequency, Max. Level,Bandwidth, Mode = Low Noise.
Test:Use frequency-selective power meter, measurement bandwidth = 1 kHz / 500 kHz,
measured value in dB below reference level (Max. Level).
1100.4903.821.15E–4
Manual Test ProcedureCMU
Options for CMU200
The following tests can be carr ied out only if the GSM-MS (CMU- K2x) s of tware options are ins talled and
enabled by entering a key code.
TX GSM Modulation
Only with options CMU-K21, CMU-K22, CMU-K23 or CMU-K24:
The GSM phase/frequency error of a TX path is measured.
Preparation:Test setup TX5:
½ Connect spectrum analyzer FSIQ to RF3 OUT.
Control:½ Set CMU generator to specified RF frequency.
Output level at RF3 OUT10 dBm
Training Sequence -> GSM0;
Bit Mod. -> PRBS
Transmission -> BURST
Settings at spectrum analyzer FSIQ:
Mode -> Digital Standards -> GSM
Mode -> Meas Result -> Result_Length -> 146 (the useful part normally comprises
147 bits, however, the FSIQ can be set to an integer number of bits only and
therefore cuts off 0.5 bits at the beginning and at the end of the measurement
Test:Phase (rms and peak) and frequency errors are measured according to GSM
recommendation.
RX GSM Demodulation
Only with option CMU-K21, CMU-K22, CMU-K23 or CMU-K24:
The GSM phase/frequency error of a RX path is measured.
Preparation:½ Connect GSM signal generator to RF2 (test setup RX3).
The signal generator must be synchronized with the CMU via the 10 MHz
reference frequency.
Control:Signal generator SMIQ: GSM signal at given frequenc y, level according to table,
bursted with the following settings:
Digital Std -> GSM/EDGE -> State -> ON;
Digital Std -> GSM/EDGE -> Select Slot -> Burst type -> NORM
Digital Std -> GSM/EDGE -> Select Slot -> Slot Level -> FULL
Digital Std -> GSM/EDGE -> Select Slot -> Data -> PN9
Digital Std -> GSM/EDGE -> Select Slot -> TSC -> TSC0
CMU: GSM Non Signalling, training sequence = GSM 0, trigger source = IF
Power, Trigger Level = Medium
Test:
½ Measure GSM phase error (rms and peak) as well as frequency error.
1100.4903.821.16E–4
Loading...
+ 186 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.