Robin SmartPAT 5500 User Manual

52 Hurricane Way, Norwich, NR6 6JB
Telephone +44(0)1603 256601 (Switchboard)
+44(0)1603 256602 (Sales)
Facsimile +44(0)1603 256603
PORTABLE APPLIANCE TESTER
SmartPAT 5500
User Manual
Portable Appliance Tester
CONTENTS
INTRODUCTION 2 Warnings and Cautions 2 Technical Specifications 3 Test Summary 5 Controls and Accessories 6 Results 9 INSTALLATION 9 Power Requirements 9 Connections 9 Setting Up 10 OPERATING INSTRUCTIONS 12 TESTING 12
Visual Check 12 Auto Testing 12 Reviewing 16 Storing 16 Storage Type 16 Programming 20 IEC Lead Test 24 Manual Testing 24 FAULT FINDING AND REPAIR 25 Calibration 25 Basic Maintenance 25 GENERAL ENQUIRIES 25
INTRODUCTION
Thank you for purchasing Robin Portable Appliance Tester. Please ensure you read and understand these instructions fully before starting any testing.
The SmartPAT 5500 is a State of the Art Portable Appliance Tester (PAT) with automatic testing capability. It is capable of performing preset earth bonding, insulation, fuse continuity, run/leakage and flash voltage tests. The equipment is contained in a high impact carrying case, see Figure 1. When opened access is provided to the operating panel, secured into the base of the case, this panel carries all the operating switches, connectors, serial port, floppy disc drive and the results display. The lid of the case provides a storage space for the cables associated with the appliance tester.
WARNINGS AND CAUTIONS
Although the equipment has been built to meet the requirements of the IEC publication 1010, Safety Requirements for Electronic Measuring Apparatus, the following warnings and cautions must be adhered to:
WARNINGS LETHAL VOLTAGES. THE SmartPAT 5500 CONTAINS HAZARDOUS VOLTAGES, CORRECT HIGH VOLTAGE HANDLING PRECAUTIONS,AS LAID DOWN IN THE ELECTRICITY AT WORK REGULATIONS 1989, MUST BE OBSERVED AT ALL TIMES.
CAUTIONS
EQUIPMENT DAMAGE. The equipment has been set for a nominal 230V/110V operation, it must never be connected to a higher voltage. EQUIPMENT DAMAGE. The mains supply is never to be connected to the IEC lead test connection. ABRASIVE MATERIALS. Routine Maintenance. Do not use abrasive materials, metal polish or furniture polish during cleaning procedures. TEMPERATURE CONTROL. Operating and Storage Temperatures below -10°C may cause permanent damage to the equipment. DATA STORAGE. This equipment has the ability to store test records within an onboard memory.The data within the memory should not be left to accumulate over a period of time but should be downloaded frequently (daily during normal testing) to a secure storage medium e.g. PC hard drive or floppy disc. Robin Electronics accepts no responsibility for lost of stored data due to misuse of the appliance tester or non-compliance with the operating instructions.
2
Technical Specifications
The technical specifications of the SmartPAT 5500 are as follows:
Mains input supply
Voltage 100 V to 120 V rms at 50/60 Hz ±10%
207 V to 255 V rms at 50/60 Hz ±10% Power consumption 30 W maximum (quiescent) Earth leakage current 3 mA maximum
Display
Type 2 line, 20 characters per line Liquid Crystal Display (LCD) Character Height 5.5 mm Character width 3 mm Viewing Angle ‘6 0’clock’ Contrast Adjustment Fixed
Communications
Format Serial Data Rate 1200 to 9600 bps (user selectable) Interface RS232/V24, support for IRDA adapter,
IBM keyboard (through adapter), HP Smartwand
and a Panasonic scanner
Environmental
Storage Temperature -10 to +50°C Operating Temperature 0 to 40°C (results within calibrated limits)
-10 to 60°C (results may be outside calibrated limits) Humidity 0 to 90% R-H non condensing Pollution level II IP Rating IP51 case closed, IP30 case open
Figure 1 SmartPAT 5500 tester
3
The Technical Specifications of the Tests performed by the SmartPAT 5500 Portable Appliance Tester are as follows:
Earth Bond
The earth bonding specifications are taken at a nominal mains input supply of 110 V or 235 V rms, unless otherwise stated.
High Current
Open Circuit Voltage 3 V rms ±10% Test Current 5 A, 8 A or 26 A ±15% into 0.1 ohm load.
(in-built intelligent current limit) Display Resolution 0.01 ohm to 19.99 ohm, then shows >20 ohm. Display Accuracy ±(3% + 4 digits) Test Duration 10s or 15s to 1 minute, dependent on time test button held down
Business Equipment
Open Circuit Voltage 3 V rms ±10% Test Current 100 mA +20% -0% at 230 Vrms into 0.5 ohm load. Display Resolution 0.01 ohm to 19.99 ohm, then shows >20 ohm. Display Accuracy ±(3% + 4 digits) Test Duration 10s or 15s to 1 minute, dependent on time test button held down Pass Threshold 0.1 ohm steps up to 5 ohm then 0.5 ohm steps to 19.99 ohm
(user selectable)
Insulation Test
The earth bonding specifications are taken at a nominal mains input supply of 110 V or 235 V rms, unless otherwise stated.
Open Circuit Voltage 560 V dc +16%,-0% Short Circuit Current 3.5 mA ±10% Current into 500k 1.0 mA minimum Display Resolution 0 1 to 99.9 Mohm, then 1 to 299 Mohm, >300 Mohm Display Accuracy ±(5% + 4 digits) Test Duration 10s or 15s to 1 minute, dependent on time test button held down Pass Threshold 0.1 Mohm steps to 0.9 Mohm
1 Mohm steps from 1 to 9 Mohm
10 Mohm steps from 10 to 300 Mohm
(user selectable)
Fuse/Load/Leakage Test
Leak
Earth Leakage Current 0.1mA to 9.9mA Display resolution 0.1mA Display accuracy ±(5% + 4 digits), capacitive/resistive leakage Leakage threshold 0.5 to 10 mA in 0.5 mA steps Abort on leakage > 15 mA
4
Load
Load Measurement 0.1kVA to 3.99kVA Display Resolution 0.1kVA Display Accuracy ±(5% + 4 digits), resistive or reactive load Abort on Surge > 60 A exponential, time constant 50 ms Abort on Run Current > 13 A (235 V), > 16 A (110 V) Test Duration 5s to 60s,Terminating immediately
when STOP button pressed.
Flash Test
Open Circuit Voltage 1500Vrms (Class I), 3000vrms (Class II) Short Circuit Current 5mA at normal mains input Display Resolution 0.1mA Display Accuracy ±(5% + 4 digits)
Test Summary
The following statements provide a brief summary of the tests performed by the SmartPat 5500.
Earth bond
Checks the resistance between the Earth pin of the appliance under test and any exposed metalwork on the appliance.
Insulation
Checks the resistance of the insulation between the Earth pin of the appliance under test cable plug or the earth bond crocodile clip, to the Live and Neutral pins of the appliance (pins are connected together within the SmartPAT 5500 for this test).
Fuse/Load/Leakage
A low dc fuse test is performed first to check the continuity of the fuse in the appliance under test, by checking the resistance between the Live and Neutral pins of the appliance cable plug. On completion of the fuse test an earth leakage test and a VoltAmpere test is performed with the appliance under test being powered up.
Flash
A 1.5kV rms voltage is applied between the earth and live/neutral pins of a class I appliance and 3 kV rms voltage between the flash probe and the live/neutral pins of a class II appliance.
5
Controls and Accessories
The cables, connectors, controls and indicators of the SmartPAT 5500 are shown in Figure 2 and listed below,together with a brief functional description:
Cables
Mains Captive mains cable to connect to the mains supply with either a
13A, 230V or a 16A, 110V plug (via adaptor). (See Figure 2 (6))
Earth Bond Captive cable, fitted with heavy duty crocodile clip,to connect to
appliance under test. (See Figure 2 (5))
Connectors
110 V Appliance Provides a socket to connect a 110 V appliance to the test set for
testing. (See Figure 2 (8))
230 V Appliance Provides a socket to connect a 230 V appliance to the test set for
testing. (See Figure 2 (7))
IEC Lead Test Enables IEC leads to be tested, used in conjunction with the 230 V
Appliance Input socket. (See Figure 2 (3))
Serial Port Allows an external computer/printer or any serial interface data
collector device to be connected to the Portable Appliance Tester.
(See Figure 2 (1))
Flash Probe Enables a probe to be inserted into the Portable Appliance Tester, to
allow high voltage insulation tests to be carried out. (See Figure 2 (4))
Disk Drive
Disk Drive 1.44Mb A 31/2” floppy disk drive to provide a means of storing the records
so that later they can be downloaded to a computer. (See Figure 2 (2))
WARNING When transferring data to a floppy disk it is essential that the data is either transferred to a computer or printed before the data is deleted from the tester.
6
Controls
The controls are all in the form of keys, there are two distinct sections: Main executing keys Test function and setting keys
Main Executing Keys GO Initiates tests and confirms selections. STOP Terminates tests and selections, when pressed returns display to show main
menu, see setting up section.
YES Allows a positive decision to be made, the key is also the up arrow for use
when scrolling through messages or lists.
NO Allows a negative decision to be made, the key is also the down arrow for use
when scrolling through messages or lists.
HELP When pressed will display a help message on the subject the SmartPAT 5500 is
currently employed in carrying out.
Test function and setting keys
These keys perform various functions, the alpha-numeric keys, half black and half yellow, provide the keyboard function for the PAT together with the space key.The use of these keys is detailed in the Installation section under setting up. The other four keys provide a means of amending the test parameters, ie alter current used or time duration of test.
All the 16 keys have a black legend above them, these are the function which that key will control. A brief description of the action controlled by each key is as follows:
AUTO Enables any autotest programmed in the SmartPAT 5500 to be selected
and run.
SITE Allows a code for each site testing is carried out at to be entered into the PAT
memory for storing on the records.
DATE Enables the date and time to be set in the PAT, this is automatically updated by
the PAT.
STORE Allows a test result to be stored in the PAT. PROG Initiates the programming mode of the PAT,For some of the major modes the
access code is required. (For Access Code details see Programming section)
CLEAR Enables results to be erased from the memory. BOND When pressed this key will initiate an earth bond test. INSU When pressed this key will initiate an insulation test. LOAD When pressed this key will initiate a load/leakage test. FLASH When pressed this key will initiate a flash test. LEADS When pressed this key will initiate an IEC lead test. VISUAL When pressed will allow the results of a visual test to be entered into
the records.
REVIEW Enables the results of an auto test to be viewed before they are stored, or any
stored result may be viewed.
SEARCH Enables a search of the records to be carried out. REPEAT Allows a search to be repeated. PRINT/PC Enables the data on the display to be printed or data to be downloaded to PC
or disk. It also enables a disk to be formatted.
7
Indicators
Bleeper (internal) Sounds to indicate that a test has been completed or failed (double bleep). Display Dot matrix Liquid Crystal Display, displays up to 40 characters at a time,
in two lines, to show the following:
Test type selected Test in progress messages Test results Warning messages.
Help messages
Figure 2 Controls and Accessories
8
Loading...
+ 19 hidden pages