12 Supply the device with power supply (24 V DC) via the
system.
If the power LED illuminates green permanently,
ready for operation.
If the alarm and warning LED do no
ates trouble-free.
Section 3.1, "Light emitting diodes" on page 14
The USM-GT-400 is supplied without connection box to the North
American
through a flame arrester. The marking of the cable (numbers) is
(always) identical to the terminal assignment.
region, the connection is made to cables that are led
t flash, the device oper-
the device is
A Earthing screw M6B Earthing screw M6
C Earthing cable
Fig. 2-2:Connect to earth - Ultrasonic gas meters DN150 (6")
Covers of the transducers (A and B)The covers protect the connections and the lines of the transduc-
er (TD) against contamination and mech
anical damage.
Ultrasonic electronics (C)The ultrasonic electronics is in a pressur
housing mounted on the ultrasonic gas meter. The ultrasonic
electronics evaluates the data recorded by the transducers. In
addition to the display, the parameters can be shown and evaluated on a computer using the RMGView
Lifting eyes (D)The lifting eyes can be used to safely transport the device using
a
suitable lifting gear.
Connection flange (E)The device is bolted onto the g
flanges.
Transducer (F)The transducers are installed in the housing of the ultrasonic ga
eter and are not visible once installed.
m
Retaining bolts (G)The retaining bolts are mounted when delivering the device. The
retaining bolts
away. The bolts must be mounted to ensure for a safe installation
or de-installation.
In this chapter, you are provided with information as to how the
ultrasonic gas meter records the data. The necessary formulas
are listed for this purpose.
Figure 4.1 shows the general working principle. Transducer TD1
and TD2 ar
ment and form a measurement path wit
ultrasonic pulse travels along the measuring path from sensor
TD1 to transducer TD2 more quickly than the other way around.
This is caused physically by vector addition of the flow velocity to
the speed of sound, the arrow above the
of flow.
e positioned opposite to each other for the measure-
(For a center path. Outer paths have a corresponding
value.)
For this calculation, it is important that only the transit times and
the device parameters such as the transducer distance and the
angle of the measuring path to the flow direction are required. All
parameters that include a gas dependency are omitted.
Legend
In order to consider the ave
rage flow profile, in particular an
asymmetrical or swirl-affected flow, a total of 6 paths is measured
in 3 levels with the USM-GT-400 ultrasonic gas meter. The 3 levels can be derived ma
tically via an integration procedure,
thema
the so-called Gauss integration.
"Arrangement of the ultrasonic transducers" on page 15
The respective average path velocities (designated with
for the respective measuring path i) along these measuring paths
results analogically to the fo
Under certain conditions such as, e.g., sm
rmula above.
aller deviations from
the tolerances during the production, it may be necessary to correct the path velocities with a common factor:
Fig. 4-4:Formula, corrected path v elocity
v
k= Correction factor for the path velocities
= Corrected path velocity(m/s)
ki
(This factor is named vw factor d1for the forward
direction and vw factor d2 for back-flow; see
chapter parameter setting 10.5.4)
Thus, one obtains for the average flow velocity:
Fig. 4-5:Formula, flow velocity
Legend
v
w
= Average flow velocity(m/s)
w
= Weighting factor with regard to the flow profile
i
The summation and the weighting specified result from the mathematic Gaussian integration procedure.
The USM-GT-400 provides parameters that allow a rating of the
installation. If the values are within given ranges, then good measurement conditions can be assumed. If the
the accuracy of the reading may be affected by disturbed flow
conditions. Please contact in this case the RMG service.
"Manufacturer" on page I
Turbulence
Due to the actual flow, in particular the turbulence, there will be
characteristic variations (variance
locities (i = 1..6; number of ultra
allow an assessment of the installation. The ultrasonic path averaged turbulence (
) is calculated as follows:
values are outside
) of the individual path ve-
sonic measuring paths), which
Legend
Fig. 4-6:Variance
Fig. 4-7:Turbulence
=
Time averaged flow velocity along the ultrasonic
path
=
Flow velocity along the ultrasonic path
=
20; number of values for turbulence calculation
(named Turb./Perf. count; see chapter parameter)
Typical values at very good flow conditions for
at 2-3%, for the outer paths, the turbulence increases to 4%. If
these values are above 10%, then disturbed conditions can be
assumed affecting the measurement accuracy. At lowest velocities the turbulence calculation is
swit
ched off.
middle paths are
Profile and symmetry factor
For a fully developed flow the middle path (3 + 4) have the highest
velocity, the two outer paths (1 + 2; 5 + 6) are more or less equal.
The profile factor (
ues below 1.00 or above 1.50, the flow conditions should be
checked.
Fig. 4-8:Profile factor
The symmetry factor (SY) usually is between 0.9-1.10; at values
below 0.75 or above 1.25 the flow conditions should be checked.
Fig. 4-9:Symmetry factor
Meter performance
This value (
paths could be measured and involved in the flow c
is calculated on base of the last 20 measurements (same as
turbulence).
Fig. 4-10:Meter performance
The highest value is 100%; under normal conditions it is above
95%. Since 2 measurement paths can fail before the
USM-GT-400 loses its calibrated accuracy,
down to 66%; if the path failure is caused by a defect transducer
an immediate repair of the failed transducers should to be
initiated.
All of these values are given at the display of th
they are identical to the values in the RMGView
) indicates whether the velocity of all ultrasonic
There are several influences (eg. Reynolds number) resulting in
a not exactly proportional relation between the measured and according to the formula 4.5 calculated mean v
The constants const-Gx (x = m2, m1, 0, 1, 2) are calculated from
the measured value pairs of the deviation with the respective flow
velocity.
The calculated correction F
flow velocity for F in the formula above.
Fig. 4-15:Formula, corrected meter factor
The process volume flow and the corrected process volume flow
result, as listed above, from the multiplication of the corresponding velocities with the pipe cross-section. The correction formulas
above are ther
flows.
=Average flow velocity (m/s)
=Constants of the basic correction
(x = m2, m1, 0, 1, 2)
is used for the corrected average
1
efore accordingly easy to transfer to the volume
PolynomialThe characteristic curve correctio
Fig. 4-16:Formula, error equation
lso carried out via a 4th
n is a
degree polynomial that represents the error curve of the device.
F
2
Q
m
const-n
The constants Konst-n (n = m2 bis n = 2) are calculated from the
measured value pairs error F
curve correction K
process volume flow.
The correction of the base line with a polynomial described in
section 4.2.1 takes into account the typical, characteristic curve
of the USM-GT-400 in an ideal way. This correction is recommended for custody transf
MID is valid. Nevertheless, a comparable accuracy can be
achieved with a piecewise linearization, if a sufficient number of
interpolation points are used. Between the interpolation points, a
simple linear interpolation is used. The correction of the base line
with the piecewise linearization may also be used in all countries
where the MID is valid if the error curve of the raw data mets the
requirements of the ISO 17089.
In order to achiev
points should be placed in the relevant flow rate range. To take
into account the higher gradient of the curve at lower flow rates
the intervals should not be equidistant; recommended are more
points in this lower flow rate range.
a sufficient accuracy most of the measuring
e
etering in all countries where the
er m
Diagnostic function Speed of Sound
4.3.1
4.3.2
The USM-GT-400 can calculate the SoS in 3 different ways.
Standard method of SoS calculation
The first calculation is realized with help of the transit time
and
suring path with the length
speed of sound
Fig. 4-18:Calculation of SoS
This first option is pretty fast and is almost permanently "online"
available.
of the ultrasonic pulses (see above) along the mea-
. It is straight forward to result in the
or
to:
SoS calculation via gas components
The second version of SoS calculation uses pressure, temperature and composition of the gas to determine the SoS according
to the
specifications of the AGA 10 standard (AGA Report No. 10,
Speed of Sound in Natural Gas and Other Related Hydrocarbon
Gases; January, 2003; AGA - American Gas Association). The
calculation is based on statistical considerations of thermodynamics; since it is very complex
Knowing the gas composition precisely values such as density,
sound velocity and other gas properties can be calculated with
very high accuracy.
Depending on the type of gas analyzer it may take 5-10 minutes
to determine the volume fractions
nents accurately. Accordingly, the precise allocation of flow to
gas composition can be done in this time frame only.
, it will not be pres
of the individual gas compo-
ented here.
the
Diagnostic function SoS
The USM-GT-400 determines with highest accuracy the flow rate
of the gas flowing through it. For the payoff the gas quality, respectively the calorific value of the gas resulting from the gas
composition
lows a second billing of the volume flow rate with the "right" ga
composition, ie the "right" calorific value.
is of course of big interest, too. The USM-GT-400 al-
s
4.3.3
This temporal resolution can be achieved receiving permanently
gas composition data from a gas analyzer. A comparison of
the
the two differently calculated SoS‘s in the USM-GT-400 allows
the immediate detection of any deviation; in particular, another
gas composition results in a different SoS. A confirmation of another gas composition then provide
the data of the gas analysis instrument.
The temporal correlation of the actual gas composition (using the
calculated via the gas composition) to the SoS using method
SoS
1 results in the higher temporal resolution for the gas composition, respectively the
Extended SoS calculation
The third possibility SoS calculation is presented under the name
"Extended SoS measurement". This new method is introduced
as an additional determination.
Transducer TD2 receives at first the ultrasonic pulse (light blue)
coming from TD1 at the time (
ed and moves back to TD1 (dark blue). There this puls
reflected, too and reaches TD2 (red) again after the time
(
a new possibility to calculate
Fig. 4-20:Calculation of extended SoS
Due to a 10-times smaller variance of the SoS calculation this
method offers significantly more accurate result compared to the
standard method (version 1). There are 2 reasons for this result;
first, the transmitter / receiver error is eliminated (especially T
the transit time of the pulse in electronic and transducer is different in the individual transducers) and secondly, any
turbulence in medium has lowest influence to the transit time (the
time interval between 12TDt and 21TDt is as short as possible).
Having typical measuring conditions, this method can easily be
applied, but there are conditions at which this method may fail.
). The differences of
). This pulse is partially reflect-
e is
and
:
results in
W
flow
;
The SoS calculation according to method 1 and 3 run simultaneously and controlled using the same criteria. If correct, the result
o
the extended measurement is preferred due to its higher accu-
the measurement conditions, both methods are revalued again.
If correct the extended SoS calculation will be chosen again.
TW setting
Measurement tolerances and/or errors of the standard method
are permanently controlled using the comparison with the expanded method. Having both values
determined. When the calibration switch is open, the T
the standard measurement can be corrected to the value of the
extended measurement. This is an important adjustment help in
case of a transducer replaced, but also serves as an accurate
path lengths determination between the transducers during the
dry calibration.
a correction for TW can be
-value of
W
4.4
Import of gas composition data
To use the diagnostic function SoS, respectively to calculate it
from the gas composition the USM-GT-400 requires the volume
fractions of the individual gas components in the gas (up to 21
components), the pressure and the temperature. From these
data SoS is calculated using the guidelines of the AGA 10 standard. For the data transfer of the
are available:
the AGA-10 calculation these default values are used as fixed
gas shares (AX-20 - AX-44; in chapter 10 the matrix notation of
parameters, measured values and variables will be explained).
To change these values they have to be confirmed by selecting
"Accept new Comp." in parameters AX-11and to be confirmed in
"takeover gas components". Only then they will be taken over as
new values for the AGA-10 calculation.
1. Parameter AX-01 „AGA-10 Sourse“
Default data
2. Setting of the default values of the individual gas components
4.4.2
Parameter AX-20 to AX-
Methane default value ....
....
Propene default value
3. Takeover with parameter AX-1
„Taking over new components“
44
1
„AGA-10 Source“
Option 4: Data input on fixed defaults for air
In mode "default air" fixed values of air composition for the gas
analysis can be used. With the additional parameter "rel. humidity" in AX-06 the water co
mol-% and the remaining components of the air are normalized
to 100%. The unnormalized default values for air are:
Nitrogen: 78.105 mol-%
Oxygen: 20.946 mol-%
Argon: 0.916 mol-%
Carbon dioxide:
ntent and component is calculated in
033 mol-%
0.
Water: 0.0 .. mol-% (calculated)
.
The water content is calculated via the relative humidity
1. Parameter AX-01 "AGA-10 source":
"Default air"
2. Setting of the default values
Paramater AX-06 "relative
All other possibilities to transmit the volume fractions of the
vidual gas components on the USM-GT-400,
of the USM-GT-400.
relative humidity
humidity“
will use interface 2
indi-
Terminal connections
The following figure shows the terminal connections.
The SoS calculation depends in addition to the gas components
also on the gas pressure and temperature. How to measure the
pressure is described in chapter 8.4; temperature measurement
is given in chapter 7.4. Setting of the parameters AX-02 "SoS
Source Temp." and AX-03 "SoS Source Pressure" allows to select whether these measured values of temperature and pressure
ar
e u
sed for AGA-10 calculation or default values AX-04 and AX-
05.
Fig. 4-23:Pressure and Temperature input
The electric connection of pressure (p) and temperature (T) has
to be done a terminals 26 to 31; AUX1 = p; AUX2 = T.
The data of a gas chromatograph (eg. RMG GC9000 or GC9300)
can be delivered as RMGBus telegram via the RMGBus protocol.
Therefore, the coordinate AX-01 “SoS AGA-10 source data“ is
set to "Serial port 2" and the serial interface in the mode "RMGBus". Additionally the parameters of the interfaces USM-GT-400
a
the RMGBus master device have to be aligned to each other.
nd
Because the content of the telegram may have data from different
streams, the
AX-09 "Stream selection". Parameter AX-
fixes how many components are part of the telegram. When using a GC9000 this parameter has to be set to "RMGBus" to offer
a correct
1. Parameter J-
2. Setting the parameters of serial 2:
„right“ stream has to be set with the parameter
evaluation for older versions of the software GC9000.
"RMGBus"
J-26 "baud"
27 "bits"
J-
28 "parity"
J-
Opt. Ser2 mode"
25 "
08 "RMGBus mode"
4.4.4
atch RMGBus master:
M
"Serial port 2"
3. Parameter AX-01 "AGA-10 source":
"Serial port 2"
4. Parameter AX-07 "maximum tim
Time in minutes, within which a new telegram has to come
via RMGB
5. Parameter AX-08 "RM
GC9000: "RMGBus"
GC9300: "RMGBus 24 Komp."
6. Parameter AX-09 "Stre
Allows the setting of the
us
GBus mod
am selection":
desired streams.
eout":
e":
Data via Modbus (USM-GT-400 is SLAVE)
The gas data can be written to Modbus USM-GT-400 (USM-GT400 is Slave). Data source can be any field devices that operate
as a Modbus master on the bus. The individual gas components
will be written into the Modbus register of parameters AY- 20 to
AY- 44 . To accept these values for the AGA-10 calculation parameters AX-
will be set as:
USM-GT-400 gets the gas data via Modbus. It is Modbus master
and asks continuously if new data are available. In this case, all
the components are re-read and fed to the AGA-10 calculation.
Parameter AX-10 "Modbus Master Target" sets which device the
USM-GT-400 is addressing. If the GC9300 is chosen no Modbus
register needs to be set at AZ-01 to AZ-54 for status and part of
the gas component.
Time [minutes] in which a new telegram must
via RMGBus
2. Adjustment of the parame
J-26 "baud"
27 "bits"
J-
28 "parity"
J-
Modbus configuration:
3.
J-29 "Modbus protocol 2"
Mas
ter has to be set to "RTU" or "ASCII"
J-30 M
Hardware configuration can be se
RS485
J-31 "Modbus address 2"
Sla
4. Parameter AX-01 "AGA-10 source":
"Serial port 2"
5. Parameter AX-07 "maximu
odbus2 HW Fashion
ve address of the device with the gas data
r of serial port 2
te
lected as RS232 or
m tim
eout":
have come
Time [minutes] in which a new telegram must
Timeout: During the transfer of data, an ad
available, generating a status signal if no new data arrived within
the adjusted time.
6. Parameter AX-10
"Modbus Master Target": "GC9300"?
If yes, continue after 8, otherwise at 7
7. Parameter AZ-01 - AZ-54
Enter Modbus registers of the gas components and status
the slave device
of
justable time-out is
have come
Treatment of the gas data
The gas data are validated after transmission and optionally normalized. The AGA-10 gas equation
components; it might even accept up to 24 components adding
some (surplus) gas components to other components.
Hexane+: sum of n-hexane, n-heptane, n-octane, n-nonane and
n-decane. If there is only hexane+ in the samples and none of the
above mentioned components, then hexane+ is added to hexane. In case one of these components is > 0
nored.
ig
, then hexane is +
Normalization to 100 mol-%: If
100 mol-%, then the components are normalized to a total of
100 mol-% (can only be applied if the sum
< 110 mol-%). Otherwise, Bit 0 in AW-0
will be set and the calculation takes place with 100 mol-% methane instead.
The pressure measurement is disturbed. Calculation will
be done with the default value.
4
Simulation active
4.5
5
6
7
8
915
The message "188: AGA-10" appears, if the status code is not 0.
The Modbus master function is flexible in order to support PGC's
other manufacturers, too; for example a Siemens PGC. Activating the RS 485 interface as Mod
chapter 10.3.3.
Batch mode
Parameters E-01 "USE09 mode" is set t
no data (results) of the DSP will be accepted
no value
no value
no value
Error AGA 10 calculation
There is an error within the AGA-10
culation is on hold. Reason may be wrong pressure or
temperature values, ...
no value
bus master
o simulation =
calculation. The cal-
is described in
In general, the USM-GT-400 is set for an optimal operation without disturbance. The setting / changing of the batch mode allows
adaptation to disturbed c
at "high-turbulent" flow conditions as well as at "strong background noise" conditions. High turbu
velocity profiles and rapidly changing asymmetries. A "smallest"
batch mode should be chosen. At strong background noise, the
signal detection can be disturbed, too. A "longer" batch mode increases the signal stability significantly
permits a change of the signal duration.
• P1 Number of F-batches per
• ...
• P8 Number of F-batches per
The default values ar
activated. All larger values are squared; F-Batch 2 means there
are 4 signals superimposed. If the F-Batch is active the ring down
time should be chosen as short as possible, preferably to 0 ms.
The slow batch mode can be activated in coordinate AI-09; it is
to be squared for all paths, too.
In this chapter you will receive information on using the device in
a safe manner.
Contents5.
5.1Intended use ....................................................... 37
5.2Layout of instructions
5.3Qualification of the personnel
5.4Safety instructions
5.4.1Hazards during transporting ................................... 40
5.4.2Hazards during installation
5.4.3Hazards during start up
5.4.4Hazards during cleaning
5.4.5Hazards during maintenance and repairs
5.4.6Hazards during operation
5.4.7Hazards for operation in potentially
explosive environments
....................................... 38
.......................... 39
............................................ 39
..................................... 41
.......................................... 43
......................................... 43
.............. 44
....................................... 45
.......................................... 45
5.1
5.5Responsibilities of the operator ....................... 46
Intended use
The Ultrasonic Flowmeter USM GT400 device is used to measure the flow velocity of the gases in a pipeline and calculate th
operating flow during running operation.
The Ultrasonic Flowmeter USM GT40
as ultrasonic gas meter or device in the following.
When used for its intended purpose, the ultrasonic ga
suitable for use in potentially explosive areas classified as
zone 1.
The device complies with ignition protection class
II 2 G Ex de IIB+H2 T6 Gb, ambient temperature between -40°C
and +55°
or
Class 1, Division 1 Group B, C and
code T5/T6 is -40°C to +40/+55°C..
The ultrasonic gas meter complies with the standards, directives
and gu
Chapter 14.5, „Standards, directives a
C
idelines.
page 178
bient temperature for
D, am
uidelines“ on
nd g
5.2
These technical limits must be maintained for a safe us
ultrasonic gas meter:
Chapter 13, „Technical specifications“ on page 159
Layout of instructions
The following instructions are used:
This warning instruction informs you of potentially hazardous situations that can occur as a result of incorrect operation or human error. If
these situations are not avoided, they can lead to fatal or severest
injuries.
This warning instruction informs you of possible hazardous situations
that can occur as a result of incorrect operation or human error. If
these situations are not avoided, they can lead to fatal or severe inju
ries.
e of the
-
This warning instruction informs you of possible hazardous situations
that can occur as a result of incorrect operation or human error. If
these situations are not avoided, they can lead to slight or minor inju
ries.
This warning instruction informs you of potentially hazardous situations that can occur as a result of incorrect operation or human error.
these situations are not avoided, they can result in material damage to
the device or the vicinity.
This information gives you tips on how to simplify your work. With
this screen, you additionally receive further information on the
device or the work process.
Operating personnelThe operating personnel are to use and operate the device within
the scope of the intended purpose.
Maintenance personnelWork on the device must only be carried out by specialist person-
nel that can carry out the respective work assigned to them as a
result of their
applicable regulations. These specialist personnel are familiar
with the legal guidelines for accident prevention and can evaluate
and avoid possible risks by themselves.
•Mechanical installation must only be carrie
respectively qualified specialist personnel.
•Installation on electrical components m
out by qualified electricians.
The specialist personnel require a training especially for
work
ing in potentially explosive env
sonnel are persons that can ver
education according to DIN VDE 0105, IEC 364 or a similar
national standards..
•Initial start up
personal (training by RMG) or by service personal from RMG.
•Maintenance and cleaning mus
respectively qualified specialist personnel.
aining, knowledge and experience as well as the
tr
d out by the
ly be carried
ust on
ironment. Specialist per-
ify a training / further
only be carried out by especially trained
must
t only be c
arried out by the
5.4
Safety instructions
Observe the following safety instructions
Non-observance of these safety instructions can lead to a risk of life
and limb and health of the person as well as damage to the environment or property damage.
Note that the safety instructions in this operating instruction and
on the device cannot cover all possible hazardous situations as
the combination of different circumstances is impossible to predict. To simply follow the instr
be sufficient enough to ensure for correct operation. Always be
observant and also consider the following:
•Befor
•Always
e working with the device for the first time, read through
this ope
instructions carefully.
the plac
rating instruction and, in particular, follow the safety
keep the operating instructions within reach for use at
•The operating instruction warns against the residual risks for
users, third parties, devices or other material assets. The
safety instructions used refer to residual risks that cannot be
avoided due to the design.
•For safe operation, t
and followed.
•Oper
•Also obs
•Th
•Se
•Cha
•For safe operation, the technical spec
•For a safe operation, t
ate the device only in a sound state and when observing
operating instruction.
the
erve the local
and assembly guidelines.
e manufacturer is not responsible for any damage that
sult as a consequence of not observing the operating
re
instruction.
rvice and maintenance work or repairs that are not
scribed in the operating instruction must not be carried out
de
without previous consultation with the manufacturer.
nges to the device are forbidden.
observed and followed. Performance limits must not be
exceeded.
scope of its intended use.
he safety
he dev
instructions must be observed
legal accident prevention, installation
ifications must be
ice must only be used in the
5.4.1
The device is exposed to different life phases, su
stallation, start upstart up, operation, maintenance and cleaning.
T
he following sections must be sorte
the life phases.
Hazards during transporting
The device may be damaged when lifting and putting down, tipping over or falling down. By disregarding the load bearing
capacity of t
severe injuries for persons in the vicinity.
•Li
ft the device only on the inten
•Before lifti
•Never stand under suspended loads.
•Observe the weight spe
meter at hand.
he lifting gear, the device may fall. There is a risk of
valves must not be removed if the system is subject to pres-
and
sure. Components may dangerously spray
cause intoxication and burns. Risk of explosion!
•Fo
r the flange connection, use only the matching combina-
of screw bolts, nuts and sea
tion
tightening torque of the flange connection for this
combination.
•In do
•Use only genuin
ing so, observe the specific
facturer or system operator.
e spare parts from RMG.
It is forbidden to install spare pa
turers. It voids all guaran
explosion protection is no longer ensured.
if you have the respective qualifica-
only
ls. Select the appropriate
ations of the system manu-
rts from third-party manufac-
tees and claims for guarantee. The
. Escaping gas may
When working on live devices in potentially explosi
spheres, resulting sparks may lead to an explosion.
•Only
•Make sure that there is no potentially explosive atmosphere
•Afte
•Chec
Sp
ar
Under normal operating conditions, the transducer cannot be accessed from the outside, thus no sparks can
or friction of the transducer against hard materials.
•Th
•pr
work on de-energized devices when in potentially explo-
sive atmosph
before s
r working provide on pressurized components, leaks may
. Escaping gas may lead to intoxication, Risk of
occur
explosion!
k all components for leaks!
ecial requirements for a safe operation in potentially explosive
eas classified as zone 1:
e transducers are made from tit
rub against the transducers, this can generate a spark thus
leading to an explosion!
If the device is loaded with a pressure that is too high, components may leak and burst.
•Never
Flange joining elements, pressure t
valves must not be removed if the system is subject to pressure.
Components may dangerously spray. Escaping gas may cause
intoxication and burns. Risk of explosion!
•Obs
The device can be heated or cooled
gas. You may be subject to burns when making skin contact with
the device.
•W
Breakages or cracks may be caused to the device if the gas temperature or ambient temperature is o
temperature ranges. Gas escaping may cause intoxication and
burns. Risk of explosion!
•Never
exceed the maximum operating p
cations on the type plate).
erve the specifications of the system ma
system operator.
ear protective gloves that protect against heat and cold for
work.
this
exceed the maximum gas temp
ent temperature of 80°C.
ressure (see specifi-
apping screw connections and
nufacturer or
by the temperature of the
utside the specified
erature and / or ambi-
5.4.7
Hazards for operation in potentially explosive environments
If the device is operated with damaged or missing components
then gas may escape. In event of damaged threads, the ignition
penetration safe gap is no longer guaranteed. Escaping gas may
cause intoxication and burns. Risk of explosion!
•Op
erate the device only in a so
If you carry out technical changes to the
can no longer be guaranteed.
•You being the operator must ensure that only sufficiently
qualified personnel work on the device.
„Qualification of the personnel“ on page 39
•M
ake sure that all employees that are using the device have
r
ead and understood this operating instruction. Moreover,
you are also obliged to train the personnel at regular intervals
and inform them of the hazards.
•Mak
•The res
•Provide the personnel with the ne
•Using suitable meas
e sure that all work on the de
qualified persons and checked by responsible specialist
personnel.
ponsibilities for installation, operation troubl
ing, maintenance and cleaning must b
equipment.
„Qualification of the personnel“ on page 39
u
res, ensure that that constructive risks
are ruled out when using the device. Inform your personnel
about the risks when using the device.
The device will be packed to customer-specifics according to the
transport requirements. In this chapter you will receive information on the standard packaging
A Euro palletB Retaining bolts x 2
C Ultrasonic gas meterD Retaining strap x 2
E Timber wedge x 2
6.1.3
Fig. 6-1:Device secured on a Euro pallet
The device is as standard supplied on a Euro pallet (A). The device can be protected by a customer-specific outer packaging. To
secure the devic
device (C) is supplied with retaining straps (D) and bolted to timber wed
addition
The device can be transported on the pallet using a lifting cart or
a fork lift.
ges (E). The retaining bolts (B) of the device provide
al support.
Unpacking the device
Remove the outer packaging
The outer packaging of the device is suite
cifics in order to protect the de
influences from the environment during transport.
The device may be damaged when lifting and putting down, tipping
over or falling down. By disregarding the load bearing capacity of the
lifting gear, the device may fall. There is a risk of severe injuries for
persons in the vicinity.
tended
• Lift the device only on the in
• Before lifting, make sure that the load is safely secured.
• Never stand under suspended loads.
• Observe the weight specifications for the ultrasonic gas me
hand.
lifting eyes.
ter at
1 Attach a suitable lifting gear (B) to the lifting eyes (A) of the
Fig. 6-4:Remove the timber wedges and retaining straps
3 Undo and remove the tension straps (B).
Risk of injury during transport
The device may be damaged when lifting and putting down, tipping
over or falling down. By disregarding the load bearing capacity of the
lifting gear, the device may fall. There is a risk of severe injuries for
persons in the vicinity.
• Lift the device only on the intend
• Before lifting, make sure that the load is safely secured.
• Never stand under suspended loads.
• Observe the weight specifications for the ultrasonic ga
hand.
ed lifting
eyes.
s meter at
4 Carefully lift the device with a lifting gear until the Euro pallet
If the packaging material and the Euro pallet are no longer required, dispose of the material in an environmentally responsible
manne
r according to the country-specific standards and
guidelines.
Prior to installation
The transport locks must only be removed once the device has
been installed in the system and the device has been transported
to the installation location.
Damage to the device from contamination and humidity
If transport locks are removed too early, contamination and humidity
may enter the device. The device could be damaged.
• Disassemble the transport
device.
The following belong to the transport locks:
•Blind plug
•Protective sticker
•Retaining bolts
Re
mmendation: remove the retaining bolts
co
solutely necessary. This is the only way to ensure that the
ab
device does not tip over or roll away after being installed.
The inside of the device is protected using
tection mat. The corrosion protection mat must be
before installation.
A Position of the corrosion protection mat
Fig. 6-6:Corrosion protection mat in the device
a corrosion pro-
removed
6.2
1 Remove the corrosion protection mat (A) from the device.
Packing the device for transportation
The device must be packed to customer-specifics according to
the transport requirements. In this chapter you will receive information as to how the device is as standar
For packaging, use the original packaging mater
set that was supplied with the device.
If you no longer have the original packaging material and sealing
set, you ca
quired from RMG.
RMG service would be pleased to consult you as to how the device should be packed.
You need the following for standard packaging:
•Euro pallet with timber wedges (with the original packaging,
the timber
pallet).
•T
•Transport locks
•Sealing set (blind plugs)
n order the packaging material and sealing set re-
The device may be damaged when lifting and putting down, tipping
over or falling down. By disregarding the load bearing capacity of the
lifting gear, the device may fall. There is a risk of severe injuries for
persons in the vicinity.
tended
• Lift the device only on the in
• Before lifting, make sure that the load is safely secured.
• Never stand under suspended loads.
• Observe the weight specifications for the ultrasonic gas me
hand.
lifting eyes.
ter at
1 Attach a suitable lifting gear (B) to the lifting eyes (A) of the
device.
2 Tension the chain of the lifting gear slig
htly.
3 Undo the bolted connections from the system so that the
Without the retaining bolts (A) the flange must be guided
en the timber wedges (C).
betwe
Risk of injury during transport
The device may be damaged when lifting and putting down, tipping
over or falling down. By disregarding the load bearing capacity of the
lifting gear, the device may fall. There is a risk of severe injuries for
persons in the vicinity.
• Lift the device only on the intended lifting
• Before lifting, make sure that the load i
• Never stand under suspended loads.
• Observe the weight specifications for the ultrasonic ga
hand.
eyes.
s safely secured.
s meter at
5 Carefully place the device on the Euro pallet with the lifting
gear.
6 Secure the device using the tensioning straps (B).
Provide the connection box (ATEX / IECEx) with blind
plugs
A WrenchB Screwed cable gland
C Blind plug
Fig. 6-11:Mount the blind plugs
1 Insert the blind plugs (C) into the connection.
2 Tighten the screw connection (B) with a suitable wr
The following sealing bolts supplied must be used in countries
where CSA / FM guidelines apply. If only transport is taking place,
you can use 1/2" or 1" screws with appropriate length as an
alternative.
Damage caused to the device from incorrect storage / transport
If the device is not protected correctly by the packaging material, dirt
or humidity may enter the device and damage it.
• Pack the device according to the instruction.
Consider the special transport requirements on the packaging mate-
•
rial, e.g., for transpor
• Please contact RMG service in case of doubt.
Use the original packaging that was supplied along with the
device. Please contact RMG service if you have any
questions.
Options for the outer packaging can be, for example:
• sea-proof wooden crate
• cardboard packagings
t overseas.
6.3
Storage
1 Protect the device in the outer packaging against environ-
mental influences.
In this chapter you will receive information on the correct storage
of the device. You are also provided with information that must be
observed when storing for long periods.
Mortal danger from damage in the warehouse
If the device is stored for longer than one year, the device may be
damaged from incorrect outer packaging or securing of the device. In
potentially explosive environment, a defective device may lead to an
explosion. Danger of poisoning!
• Avoid long storage times.
ice if
• Have the device checked by RMG serv
ger than one year. For this purpose, send
Damage caused to the device from incorrect storage / transport
If the device is not protected correctly by the packaging material, dirt
or humidity may enter the device and damage it.
• Pack the device according to the in
• Consider the special transport requirements on the packaging material, e.g., for transport overseas.
• Please contact RMG service in case of dou
1 Packing the device.
„Packing the device for transportation“ on page 55
2 Observe the approved ambient temperature for storage.
Chapter 13.1, „Performance data“ on page 160
Checking the device after storage
Mortal danger from damage in the warehouse
If the device is stored for longer than one year, the device may be
damaged from incorrect outer packaging or securing of the device. In
potentially explosive environment, a defective device may lead to an
explosion. Danger of poisoning!
• Avoid longer storage times.
• Have the device checked by RMG service if the
ger than one year. For this purpose, send the device to RMG.
struction.
bt.
storage
time is lon-
Checking the device for any signs of damage
There is a high risk to life and limb if a dam
aged device is
used.
The following damage can compromise safety and the function of the device:
• notches on the flange sealing su
rfaces
• corrosion in the device or on the sealing surfaces
• cracked glass of the viewing window
• clouded glass of the viewing window
• cracks, flaking on the housing
he covers
or t
• flaking paint
1 Check that the device is intact by
If you discover that there is, e.g., any d
age to the device, the device may only b
If incorrect flange seals are used for several ultrasonic gas meters,
potentially explosive gas mixtures can escape due to leaks. Risk of
intoxication and explosion! Moreover, the pressure of the flange
increases with the improper tightening of the screw bolt.
• Make sure that the flat seal does
face into the pipeline.
Malfunctions from incorrect seals
If flange seals are used for ultrasonic gas meters that protrude into the
pipeline, the measuring accuracy may be influenced.
• Make sure that the flat seal does
face into the pipeline.
The durability of the flange connections has been verified for
seals with the following maximum material values according to
the AD2000 regulations.
Flat seals:k0 x KD = 20 x bD | k1 = 1.3 x bD (N/mm)
not protrude over the sealing sur-
not protrude over the sealing sur-
Grooved gaskets:k0 x KD = 15 x bD | k1 = 1.1 x bD (N/mm)
Spiral seals:k0 x KD = 50 x bD | k1 = 1.4 x bD (N/mm)
You have different possibilities when installing the device into
your system. Please verify a proper inner diameter of any pipe
that is connected to the meter.
–See „Inner diameter of connecting spool pieces“ on
page 168
Dependency on the gas flow direction
In order for the installation to correspond to the requirements of
the Measurement Instrument Directive 2004/22/EC (MID) or
Measurement Canada (MC), the device must be installed with an
inlet and outlet piping.
With this installation, the device can be used for
surements and for secondary measurements.
If the devices do not match one-another for these installation possibilities, incorrect measurements may occur.
• Please consult RMG if a Face-to-Face installation is possible with
the desired devices and number of devices.
A Tapered boreB Inner diameter
Fig. 7-9:Face-to-Face installation
With this installation option, two or several devices are connected
with one-another via the flanges. Third-party manufacturer devices can also be connected to RMG devices.
For this purpose, a tapered bore can be applied to the flanges on
the inlet and outlet piping.
The
flanges that are used to connect the devices with one-another do not require a tapering. For third-party manufacturer devices,
you have to check if a tape
For the device with the smaller inner diameter it is mandatory to use a
tapering.
If two RMG devices are connected with one-another, the inner diameter must be continuously the same. Different siz
cannot be connected to one-another.
If required, you can connect one or two flow computers to the
device.
Follow the installation g
Operating instructions of the flow computer
Th
e 2 interfaces RS485-1 and RS485-2 have the same features
and you ma
However, the RS 485-1 (in contrary
permit a parameterizable byte sequence for the data types Long
and Float. We therefore recommend that you use the RS 485-1
for the DZU protocol and the RS 485-2 for instance F communication. You may find more in
Flow computer from RMGThe device is compatible with the following flow computer series
from RMG
•ER
•ERZ 24
If you want to use the flow computer from RMG specified above,
do not have to carry out any configurations. The flow comput-
you
ers from RMG can directly process the
gas meter from RMG directly. For this purpose, the flow computer
has to be connected to the digital interface RS485-1 in order to
allow all diagnosis functions to be used. If you want to install an
additional flow computer for reasons of security, this must be connected via the interface RS485-2.
y change in the following1 to 2 (and 2 to 1) freely.
Flow computers from third-party manufacturers can be connected to the device. These can only be conn
interface RS485-2. This interface communicates via a Modbus
protocol. In order to be able to use all diagnosis functions, the
Modbus must be configured. You can also use the high-frequency outputs pulse 1 and 2. When par
the maximum possible gas flow rate correlates to a maximum frequency of 2 kHz. All diagnosis functions cann
interface.
If you are using a third-party manu
configure the flow computer.
ected to the digit
meterizing, take care that
a
ot be used via this
facturer device, you have to
al
7 Construction and Planning
Explosive atmospheres
Device group II zone 1
Maximum distance 500 m / 1640 ft
Safe area
or
Class1, Division1
Connecting a flow computer for example an
ERZ2000 / ERZ2400
Fig. 7-10:Connection diagram for connection of a flow computer fo r example an ERZ2000 / ERZ2400
The cable length must not exceed a length of 500 meters /
1640 feet.
In potentially explosive environments, dangerous voltages can still
remain as ignition sources for up to one minute after being switched
off.
• Disconnect the device from the power supply before star
maintenance work.
• Secure the device against being switched back on.
• Cordon-off the work area of the device, e.g., using a b
signs.
• After switching off the device, wait at least one minute before starting work. Ensure that the device is volt
earth and short-circuit.
Mortal danger from components under pressure
Flange joining elements, pressure tapping screw conn
valves must not be removed if pressure is applied to the device. Components may be flung-around. Gas escaping may cause intoxication
burns. Risk of explosion!
and
• Carry out work on the device only when it is in a voltage-free state,
is vented and without pressure.
Mortal danger from incorrect work
Hazards can only be recognized and avoid
sonnel. If work is carried out by persons who have not been trained
these special activities in potentially explosive areas, they may cause
an explosion.
• Have installations only carried out by especia
(expert according to DIN VDE 0105, IEC 364 or similar national
standards).
Mortal danger from damaged sealing surfaces
If sealing surfaces are damaged, e.g., from notches or scratch
leaks may occur. Risk of intoxication and explosion!
• Install only an undamaged device.
age-free. Then connect to
ed by specially trained per-
ting the
arrier and
ections and
for
lly trained personnel
es,
Risk of injury during transport
The device may be damaged when lifting and putting down, tipping
over or falling down. By disregarding the load bearing capacity of the
lifting gear, the device may fall. There is a risk of severe injuries for
persons in the vicinity.
• Lift the device only on the intend
• Before lifting, make sure that the load is safely secured.
• Never stand under suspended loads.
• Observe the weight specifications for the ultrasonic ga
hand.
Risk of injury from missing retaining bolts
If the device is put down without the retaining bolts, it can tip over or
roll away. Serious injuries may occur.
• Before starting work, make sure that the retaining bolts are screwed
in.
For NEC 500In countries where CSA / FM guidelines apply, the connec-
tions not required must be provided with gask
factory. Please leave these in the screw connection and only
connect the cable that leaves the conduit seal. When joining
the conduits to the flame block, ensure for a slight gradient
away from the flame block in order to avoid the accumulation
of water on the conduit seal. Moreover, also ensure that you
do not twist the conduit seal when securing the conduit as the
cable in the electrics housing may tear-off as a result. If necessary, use a respective bolted connection (union).
et screws ex-
8.2
8.2.1
For all devices6 Secure the device with r
Chapter 6.2, „Ensure for a safe position of the device“ on
page 56
7 Check the device for any signs of damage.
Chapter 6.3.2, „Checking the device for any signs of dam-
age“ on page 63
8 Clean the sealing surface of the flange from c
with a gentle cle
aning agent.
etainin
Installation of the device
Mounting the inlet and outlet piping
Escaping gas from incorrect seals
If flange seals are used for ultrasonic gas meters that protrude into the
pipeline, potentially explosive gas mixtures may escape due to leaks.
Risk of intoxication and explosion!
• Make sure that the flat seal does
face into the pipeline.
not protrude over the sealing sur-
g bolts for the installation.
ontamination
Observe the instructions for the dimensions!
Chapter 13.5,„Weights and dimensions“ on page 163
A Inlet pipingB Ultrasonic gas meter
C Outlet piping
Fig. 8-2:Installation of the Inlet and outlet piping
The bolted connection of Inlet (A) and outlet piping (C) must be
tightened according to the tightening torque of the plant manufacturer. The tightening torque must comply with the bolts and seals
used.
8.2.2
1 Clean
2 Tighten the bolts cross-wise in order to avoid tensioning.
In general, only the horizontal installation of the USM GT400 is
strongly recommended. Turning the meter by more than 2 flange
holes should not be used to avoid the collection of condensate in the
sensor pockets. Only in dry and clean gas, other installation positions
might be possible also, but it is not recommendable.
the sealing surface of the flang
with a gentle cleaning agent.
Installation of the connection box
The device can be ordered in different versions. Depending on
the version ordered, another procedure must be carried out for
the installation.
These are the order versions:
•conn
•conn
ection box according to ATEX / IECEx
„Installatio
page 82
ection according to NEC 500
no
box must be installed here, only connect the cable accord-
Cable specificationThe following cable specifications are according to a complete
assignment of the USM-GT-400 ATEX / IECEx version. The cable types listed are recomme
technically comparable cable types.
ndations that can be replaced by
Power supply
24 VDC
Interface:
RS485-0, RS485-1,
RS485-2
(can be layed in one
cable)
AUX1LIYCY
AUX2LIYCY
Analog out: 4..20 mALIYCY
Warning + alarmLIYCY
Pulse1 + pulse2 + I/
O1+ I/O2
ÖLFLEX® CLASSIC
3 x 1.5 mm²
3 x 2.5 mm
LIYCY (TP)
3 x 2 x 0.75 mm²
2 x 0.75 mm²
2 x 2 x 0.75 mm²
2 x 0.75 mm²
2 x 2 x 0.75 mm²
LIYCY (TP)
4 x 2 x 0.75 mm²
Cable
12.3 mm
13.5 mm
9,4 mm
6.0 mm
8.5 mm
6.0 mm
8.5 mm
10.7 mm
Twisted pair cable (TP) are only required in case of multiple circuits in one cable. Otherwise, LIYCY 2 x 0.75 mm² is sufficient for
all signal outputs.
In those countries where the standards ATEX and IECEx are valid, the device is supplied
The external connection box is connected
sonic electronics ex-factory and pr
electronics. The external connection box does not have to be
mounted.
with the connection box Ex-de (A).
electrically to the ultra-
e-assembled on the ultrasonic
Connection according to NEC 500
The number of lines that are permitted to be routed through the
cable gland (½" and ¾") on the electrics housing and flame block
is limited. Accordingly, this results in 4 different constellations that
reflect the possibilities for connection.
The numbers given below at each figure o
number of cables; all cables are labeled due to the number of the
basic upper terminal block.
The interfaces RS 485-0, RS 485-1 and RS 485-2 are basically
equal and can be set for all possible connections. However, there
are minor differences. These are taken into account in the recommended connections and make it
recommended devices or the PC if they are followed.
ier to connect the
eas
Please use the connections as recommended below.
If connections are used differently additional, extensive settings
may become necessary.
Recommended connections at the digital outputs.
RS 485-1RS 485-2
IGM-protocol,
DZU-protocol
ERZ2000,
ERZ2400,
Instanz-F,
nd
2
ERZ ...,
Flowcomputer
other suppliers
protocol,
device
RS 485-0
RMGView
(service)
USM
ERZ2000-NG,
ERZ2000-DI
features
no parameterizable byte
sequence for
data types Long
and Float
no parameterizable byte
sequence for
data types Long
and Float
Modbus-Master, can
handle IGM- and
DZU-protocol, too,
parameterizable
byte sequence for
data types Long and
Float
Connecting a flow computer from third-party manufac-
turers
A Connection Warning messagesB Connection Warning messagesC Connection Pulse 1
D Connection Pulse 2E Connection RS 485-1F Connection of S 485-2
Fig. 8-16:Connection assignment on the terminal strip
Flow computers from third-party manufacturers can be connected to RS 485-1 or RS 485-2 terminals. This interface
comm
unicates via a Modbus protocol.
All diagnosis functions can be made available via a configuration
e Modbus.
of th
Flow computers from third-party manufacturers can also be connected to the terminals Pulse 1 and Pulse 2. When
pa
rameterizing, take care that the maximum possible gas flow
rate is assigned to a maximum frequency of 2 kHz. All diagnosis
functions are not available.
1 Connect the flow computer to terminals RS 485-1 (E), RS
48
5-2 (F) or Pulse 1 (C) and 2 (D).
W
arning and alarm messages are also available. You also
have to
operation.
connect a direction contact for bi-directional
2 Connect the terminal Warn (A) fo
3 Connect the terminal Alarm (B) for alarm messages.
Even though DSfG Device-F (DSfG Instanz-F) is a German standard
the wish for a common connection for different ultrasonic gas flow rate
meters may arise in other countries as well. Therefore this connection
and its treatment is mentioned here, too.
The intension to connect Ultrasonic gas meters comparable to an
electronic evaluations unit via the sam e proto col ari ses due to the
wish to transfer "all" data determined by an ultrasonic gas meter,
i. e. measured values as well as status information or diagnostic
data. Therefore the connection via DSfG-device-F has become
the standard in Germany.
Since the USM GT400 does not have its own DSfG bus access,
its
DSfG-ins
computer, the ERZ 2000-NG, which has this access. To realize
the access the necessary data are transferred between the
ERZ 2000-NG and USM GT400 via Modbus,
ferred to as Instance-F, although it only provides the data
required for DSfG
tance-F protocol is implemented externally via a flow
Fig. 8-17:Data exchange between ERZ 2000 NG and USM-GT-400
The corresponding settings can be found in the VK Modbus
Master USM menu in the ERZ2000-NG. The corresponding
register expressions can be found in the VJ register expressions
menu. In the USM GT400, the Modbus registers of instance F
are listed in column BA.
Electrical connection
The following figure shows the rear panel of the ERZ2000-NG.
The USM GT400 is connected to the serial interface COM6.