The manual provides you with the information that is designed for trouble-free and safe
operation.
The TRZ04 is state of the art and conceived and manufactured according to the
recognized safety standards and guidelines. However, risks may arise during use that can
be easily avoided by observing this manual. For this reason, you may only use the device
as intended and in technically sound condition.
1
If the ultrasonic gas meter is not used for its intended purpose, warranty claims
will be void.
Warning
1.1.1 Abbreviations
The following abbreviations are used:
ca. circa, approximately
as app. as applicable
max. maximum
MC Measurement Canada
MID Measurement Instruments Directive
min. minimum
e.g. for example
1.1.2 Symbols
The following symbols are used:
1, 2, … Marks steps within a work operation
..
Manual TRZ 04 · EN04 · February, 1st 2017
Introduction
Installation and Commissioning
1.1.3 Layout of instructions
The following instructions are used:
2
This warning instruction informs you of potentially hazardous situations that can occur
as a result of incorrect operation or human error. If these situations are not avoided,
they can lead to fatal or severest injuries.
Danger
This warning instruction informs you of possible hazardous situations that can occur
as a result of incorrect operation or human error. If these situations are not avoided,
they can lead to fatal or severe injuries.
Warning
Caution
This warning instruction informs you of possible hazardous situations that can
occur as a result of incorrect operation or human error. If these situations are
not avoided, they can lead to slight or minor injuries.
Notice
This warning instruction informs you of potentially hazardous situations that
can occur as a result of incorrect operation or human error. If these situations
are not avoided, they can result in material damage to the device or the vicinity.
This notice may also give to you tips on how to simplify your work. With this
screen, you additionally receive further information on the device or the work
process.
Manual TRZ 04 · EN04 · February, 1st 2017
Installation and Commissioning
Introduction
1.1.4 Using the turbine meter TRZ04
1.1.4.1 Safety instructions
Observe the following safety instructions !!
Non-observance of these safety instructions can lead to a risk of life and limb
and health of the person as well as damage to the environment or property
damage.
Note that the safety instructions in this operating instruction and on the device cannot
cover all possible hazardous situations as the combination of different circumstances is
impossible to predict. To simply follow the instructions specified may not normally be
sufficient enough to ensure for correct operation. Always be observant and also consider
the following:
Danger
3
Before working with the device for the first time, read through this operating
instruction and, in particular, follow the safety instructions carefully.
The operating instruction warns against the residual risks for users, third parties,
devices or other material assets. The safety instructions used refer to residual risks
that cannot be avoided due to the design.
Operate the device only in a sound state and when observing the operating
instruction.
Also observe the local legal accident prevention, installation and assembly
guidelines.
Notice
Always keep the operating instructions within reach for use at the place of
installation.
For safe operation, the safety instructions must be observed and followed.
The manufacturer is not responsible for any damage that result as a
consequence of not observing the operating instruction.
Manual TRZ 04 · EN04 · February, 1st 2017
Introduction
Installation and Commissioning
4
Danger
Service and maintenance work or repairs that are not described in the
operating instruction must not be carried out without previous consultation
with the manufacturer.
Changes to the device are forbidden.
For safe operation, the technical specifications must be observed and followed.
Performance limits must not be exceeded (chapter 5 Specifications).
For a safe operation, the device must only be used in the scope of its intended
use.
Only use the listed set screws, bolts, nuts and gaskets or parts with similar
characteristics for the installation of the meter in the pipeline (chapter 1.14
Installing the device in the pipeline)
The device should only be used under the intended use for safe operation
(Chapter 1.3 Overview, Function and Application)
1.1.4.2 Hazards during commissioning
Initial start up Initial start-up must only be carried out by especially
trained personal (training by RMG) or by service personal
from RMG.
Notice
After the initial start-up an inspection report shall be drawn up. This, the
operation manual and CE declaration of conformity must always be accessible.
As far as possible all the sharp edges have been eliminated on the device.
Nevertheless, when working on the personal protective equipment must be
used, which must provide the operator.
Manual TRZ 04 · EN04 · February, 1st 2017
Installation and Commissioning
Introduction
Danger
Install the device TRZ04 in accordance with operating instructions. If the TRZ04
is not installed according to the manual, there is possibly an insufficient
explosion protection.
The explosion protection expires !!
Please take care during the installation to the flow direction marked on the
housing by an arrow.
When staff carry out work without sufficient qualification, risks remain
understated when working. Explosion or fire may happen. Perform the work
only if you have the appropriate qualifications and you are an expert.
If you do not use the appropriate tools and materials, components may be
damaged. Use tools that are recommended to you for the job in the operating
instructions.
Mechanical installation Mechanical installation must only be carried out by the
respectively qualified specialist personnel.
5
Electrical installation Installation on components must only be carried out by
qualified electricians.
Mechanical and/or
electrical installation
The specialist personnel require a training especially for
working in potentially explosive environment. Specialist
personnel are persons that can verify a training / further
education according to DIN VDE 0105, IEC 364 or a
similar national standards.
Manual TRZ 04 · EN04 · February, 1st 2017
Introduction
Installation and Commissioning
6
Danger
The installation and removal of the TRZ04 or to the TRZ04 connected
convertor may only be carried out in an explosion-free, de-pressurized
atmosphere. It is important to pay attention to the descriptions of the
operation manual during the installation process.
It is generally recommended to ask the RMG Service for any installation or
removal.
After working on pressurized components, a control of leaks must be carried
out.
All points above have to be considered for any repair and maintenance and
in general, when an opening of the meter (convertor) is required.
Parts to lock the flanges, locking screws, fittings and check valves, oil
supply and the pressure tap fittings, valves, HF pulse device, protecting pipe
and rotating adapter may not be opened during operation.
1.1.4.3 Danger when using, servicing and maintaining the device TRZ04
Operating personnel The operating personnel are to use and operate the
device within the scope of the intended purpose.
Maintenance personnel Work on the device must only be carried out by specialist
personnel that can carry out the respective work assigned
to them as a result of their training, knowledge and
experience as well as the applicable regulations. These
specialist personnel are familiar with the legal guidelines
for accident prevention and can evaluate and avoid
possible risks by themselves.
Maintaining and cleaning Maintenance and cleaning must only be carried out by the
respectively qualified specialist personnel.
Danger
If staff carry out work without sufficient qualification, risks may be
underestimated when working. Explosion or fire may happen. If work is
carried out in hazardous areas on voltage supporting equipment, resulting
sparks may cause an explosion.
Manual TRZ 04 · EN04 · February, 1st 2017
Installation and Commissioning
Introduction
If the unit is not cleaned in accordance with the operating instructions, the
device may be damaged. Clean the unit only in accordance with the operating
instructions.
When not using appropriate tools, components can be damaged. The explosion
protection expires.
Danger
- Clean with a damp cloth!
The TRZ04 may only be used as intended !!
(Chapter 1.3 Overview, Function and Application).
Avoid any use of the TRZ04 as possible climbing aid or attachments of TRZ04
as possible handholds !!
Danger
7
1.1.4.4 Specialized knowledge required
Notice
Persons working with or on the device must have the following knowledge:
Training / education for working in potentially explosive environments
The ability to correctly assess dangers and risks when using the device.
Possible dangers are, e.g., components under pressure or the result of
incorrect installation.
Recognize dangers that could be caused by the used flow medium.
Training / education by RMG for working with gas measuring instruments.
Education / instruction in all country-specific standards and directives to be
observed for work that is to be carried out on the device.
Manual TRZ 04 · EN04 · February, 1st 2017
8
Introduction
Installation and Commissioning
1.1.5 Risk assessment and mitigation
The TRZ04 subjects to risks in its use, which were judged by qualified staff of the company
RMG. Risks may arise due to high pressures, more rarely by too low. Even work outside
the permissible temperature range can lead to danger. Invalid current and voltage values
can trigger explosions in hazardous areas.
The risk assessment assumes that when installing or removing a turbine a draining and
venting of the pipeline takes place. Thus, and only then no explosive gas mixture is in the
pipeline.
Of course (chapter 1.1.4.4 Specialized knowledge required) only work by trained
personnel are permitted, which is also trained to know proper tools and use only this.
These risks have been considered during the development phase and action were taken to
minimize these risks.
Risk mitigation measures:
- All pressurized parts are designed in accordance to AD 2000 regulations, PED
Annex 1
- The complete design being pressurized is verified by the TÜV Hessen
- All pressurized parts are made with material certificate; there is a traceability to the
batch tracking of all pressurized parts
- The mechanical properties of all relevant pressurized components are checked by
tensile testing, impact testing and hardness testing
- In addition, non-destructive testing were applied: X-ray and ultrasonic testing of the
meter housing for defects in material, surface crack detection by magnetic particle
and dye-penetration
- With pressure tests 1.5 times of the intended operating pressure the strength of the
components were verified; leak test were carried out at 1.1 times of the intended
operating pressure. Successful tests were marked
- The maximum operating pressure is stated on the nameplate of the device, as well
as the temperature range. Operation of the device is only permitted within the
indicated ranges.
Manual TRZ 04 · EN04 · February, 1st 2017
Installation and Commissioning
Introduction
For any work in an explosive atmosphere (all areas):
Danger
- The pulse output of the turbine meter has to be connected to
intrinsically safe circuits, only.
- For maintenance and repair work only tool must be used, which is
approved for Ex zone. 1
- Otherwise, work must be carried out only if there is no explosive
atmosphere
- Danger of ignition caused by impact or friction must be avoided.
- In hazardous areas, the wiring / installation must be performed by
trained personnel in accordance to EN 60079-14 and according to
national regulations
- Skilled persons are in accordance to DIN VDE 0105 or IEC 364, or
comparable standards
- Use only trained and qualified personnel. Work on the measuring
system may only be carried out by qualified personnel and must be
checked by a specialist supervisor.
- Qualified persons are able to carry out such work. Qualification is
valid due to their training, experience or through instruction and their
knowledge of relevant standards, regulations, accident prevention
regulations and plant conditions for the safety of people. It is essential
that these people are able to recognize and avoid possible dangers.
9
1.1.6 Validity of the manual
This manual describes the turbine meter TRZ04. The TRZ04 is only a part of a complete
on site system. Observe also the instructions of other components of the site system. If
you find contradicting instructions, please contact RMG.
Notice
Make sure that the performance of the power connector corresponds to the
data on the nameplate. Additionally observe any existing national
regulations in the country. Use cable suitable for the cable glands.
Manual TRZ 04 · EN04 · February, 1st 2017
Introduction
Installation and Commissioning
10
Danger
Perform the work only if you have the appropriate qualifications and you are a
trained specialist.
1.1.6.1 Hazards during Operation
Please take care of the information provided by the plant manufacturer and/or plant
manager.
1.1.6.2 Danger during operation in hazardous areas
Operate the device in perfect and complete condition.
When you perform technical modifications to the instrument, a safe operation can no
longer be guaranteed.
Danger
Use the device only in the original condition. The TRZ04 may be operated in
explosion zone 1, but only within the permitted temperatures (Chapter 1.10
Temperature ranges)
1.1.6.3 Operator responsibility
Take care as plant manager that only sufficiently skilled personnel will work on the device.
Ensure that all employees who handle the equipment have read and understood these
instructions. Additionally, you are obliged to train staff on a regular basis and to inform
them about any dangers. Ensure that any work carried out on the device is done by
qualified personnel and checked by responsible specialists. You must establish clearly the
responsibilities for installation, operation, troubleshooting, maintenance and cleaning.
Point out to your staff any risks involved when using the device.
Manual TRZ 04 · EN04 · February, 1st 2017
Installation and Commissioning
Introduction
1.1.7 Transportation
The device is packaged customized according to the transport requirements.
Warning
Risk of injury during transport
Any foot screws must be fitted, if they are used as transit protection against
rolling and tipping.
Take additional care that a rolling and tilting is safely prevented.
To lift the meter only the special lifting eyes / eye bolts may be used.
Pay attention to the maximal loads for the lifting devices (chapter 5
Specifications) – don’t exceed those.
Make sure that the load is fastened securely before lifting.
Do not stand under lifted loads.
During lifting and/or depositing the device may slip, tip over or fall.
Disregarding the maximum load capacity of the lifting device, the device may
fall down. For any people standing nearby, there is a significant risk of serious
injury.
11
If the unit is supplied on a Euro pallet, the unit can be transported using a lift
truck or a forklift.
During the transport the meter has to be protected against shock or vibration.
The meter or its additionally mounted spool pieces are equipped with flanges
that are protected with a sticker foil or blind plugs made of plastic. These must
be removed completely before installation in the pipeline. Remaining rests of
these protections may change the flow pattern and may result in a
measurement error !!
For any transport e.g. for recalibration we ask you to use these protections.
Manual TRZ 04 · EN04 · February, 1st 2017
12
Introduction
Installation and Commissioning
1.1.8 Delivery
Supplied number of parts may differ depending on the optional customer order. "Usually"
is following in the delivery:
Part Quantity
Turbine meter TRZ04 1
1 small bottle of oil for lubrication 1
Manual 1
Test report 1
Calibration certificate 1
Material test report 1
Certificate strength 3.1 ..optional
… …
1.1.9 Dispose of packaging material
Dispose the packing material in an environmentally friendly way in accordance to the
national standards.
1.1.10 Storage
Avoid long periods of storage. Check the device after any storage for damage and
functionality. Ask for a check from the RMG service after a storage period for the device of
over one year. In this case please send the device to RMG.
Notice
For storage, a dry and protected space has to be provided.
It is important to ensure that all open pipe pieces must be sealed.
Manual TRZ 04 · EN04 · February, 1st 2017
Installation and Commissioning
Introduction
1.2 Structure of the manual
The introduction of this manual contains two parts. In the first part general guidelines are
listed; the symbols used and structure of advices are given, but also issued is a risk
assessment. In addition, it contains guidelines for the transport and storage of the TRZ04.
The second part introduces the operation of TRZ04; there are basic standards listed;
pressure and temperature ranges are given limiting the use of the TRZ04.
The second chapter describes the use of TRZ04. It explains how a high accuracy can be
achieved, how data can be received by the encoder and tapped there. The third chapter
provides recommendations on how the current operation can be ensured. These include
the accuracy influencing flow disturbances as well as regular lubrication of the bearings of
the turbine meter.
The fourth chapter describes the marking of the TRZ04 and in the fifth the technical data of
TRZ04 are summarized. The appendix includes details to the previous chapters, a
summary of certificates and approvals.
13
1.3 Overview, Function and Application
The TRZ 04 and TRZ 04-K are used to measure volume flow rates of non-aggressive
gases and fuel gases, especially natural gas. Measurement of corrosive gases is only
possible with special designs. The measurement of liquids is not possible and will destroy
the device.
The TRZ 04 is a flow rate meter, which can be used for custody transfer flow rate
measurement. The TRZ 04 K, however, can only be used for operational volume flow rate
measurements. Both types measure the gas quantity, means the volume flowing through
the unit at the actual pressure and temperature. Both determine the volume flow rate at the
actual conditions. The volume flow rate of the gas flowing through the meter is indicated by
a mechanical totalizer operating in m³.
In addition, the turbine can be equipped to supply high or low frequency pulses and reed
contacts can be fitted, too. The measured number of pulses are proportional to the volume
that has passed the meter. These pulses can be further processed by volume correctors or
remote totalizers.
The most important differences between the versions TRZ 04 und TRZ 04-K are:
TRZ 04
Approved for custody transfer metering
Installation length: 3 x DN
Manual TRZ 04 · EN04 · February, 1st 2017
Introduction
Installation and Commissioning
Accuracy 0.5% (above 0.2 Q
Blade monitoring system
TRZ 04-K
For secondary metering.
max
)
14
Installation length: 1.5 x DN (DN 50: 3 DN)
Accuracy 1% (above 0.2 Q
max
)
The following applies to all meter types:
Maximum operating pressure: 100 bar (not for gas meters with a plastic turbine wheel)
Meter sizes from G 65 to G 1600
Measuring range 1:20 at operating pressure pm = 1 barg;
Measuring range up to 1:100 at pm >= 3 barg (see table: Q
Connections in compliance with DIN or ANSI are available
Special designs for aggressive gases are available
Available for low temperatures (< - 20°C)
All gas meters can be operated in any position up to the nominal diameter of
depending on p
min
min
DN 200.
)
Notice
The meters with nominal diameters from DN 50 to DN 200 are approved
for custody transfer metering (up to G 1600).
The nominal diameters from DN 250 are special versions without MID
approval for which RMG does not guarantee the availability (only on
request).
The TRZ 04 and TRZ 04-K are approved for the use in explosive areas; the indication is:
II 1G Ex ia IIC T4 (HF-Sensors)
II 2G Ex ib IIC T6 (Encoder)
The corresponding conformity certificates can be found in the appendix.
Manual TRZ 04 · EN04 · February, 1st 2017
Installation and Commissioning
Introduction
Danger
To operation the encoder you have to care of:
The encoder must be mounted and removed by a qualified electrician, as
sensitive electronic circuits are part of the encoder.
Attention: Risk of destruction due to static electric discharge, for example
caused by the friction of clothing.
Do not change anything at the encoder, otherwise the approvals and certificate
lose their validity.
Information about temperature limits are given in chapter 1.10 Temperature ranges.
The TRZ04 meets the standards, guidelines and regulations:
Chapter 1.6 Standards / Guidelines and the appendix
These technical limitations you must comply with the TRZ04 for safe operation:
Chapter 5 Specifications
15
The device may only be operated with the following types of gases. Only with these stated
gases a safe operation is ensured:
•
Gases of class 1
•
Gases of class 2
•
Gases of class 3
The friction of the gas components must be within the limits in accordance with the
EN 437: 2009 are for test gases.
The following instructions must be observed:
Notice
The RMG TRZ 04 and TRZ 04-K can be operated in any mounting orientation up
to DN200, from the DN 250 is a horizontal orientation is required.
If a mounting orientation is specified when ordering, the meter has to be
installed in this orientation.
It is also important to ensure that the filling port of the lubricator faces upwards.
Manual TRZ 04 · EN04 · February, 1st 2017
16
Introduction
Installation and Commissioning
1.4 Method of operation
The method of operation of the mechanical turbine meter is based on the measurement of
the gas velocity. The velocity of the gas flowing through the gas meter is increased in the
flow straightener and the gas strikes the turbine wheel in a defined flow cross section. In
the flow straightener, unwanted vortices, turbulences and asymmetries are removed or
reduced. The turbine wheel is mounted axially, while the blades of the turbine wheel are
arranged at a certain angle to the gas flow. Within the measuring range (Q
the rotational speed of the turbine wheel is almost proportional to the mean gas velocity
and, therefore, to the rate of flow. The number of rotations is a measure of the volume that
has flowed through. The rotary movement of the turbine wheel is transmitted by a
magnetic coupling to the unpressurized meter head. Downstream of the coupling, there is
a gearing which reduces the rotational speed of the turbine wheel to match the mechanical
totalizer. Two LF pulse transmitters (reed contacts) and a tamper contact are located on
the totalizer.
Turbine
wheel
Meter head with
mechanical index,
Encoder and LF
pulse transmitters
min
- Q
max
),
Flow
straightener
Figure 1: Sectional drawing of a turbine meter.
Downstream of the turbine wheel, a cam wheel (reference wheel) is located on the same
shaft. Two HF sensors (proximity switches) generate a signal if a blade of the turbine
wheel (HF 3-D) or a cam of the reference wheel (HF 2-D) passes them. In this way, two
pulse sequences are generated which are out of phase. The generated pulses can be
further processed for secondary volume measurements or flow measurements.
Reference
wheel
HF pulse
transmitters
Manual TRZ 04 · EN04 · February, 1st 2017
Installation and Commissioning
1.5 Approvals
Type TRZ 04 has been approved for custody transfer metering.
The following approvals have been obtained:
MID approval No. T10417
PED approval No. CE 0091
Introduction
Independent whether the turbine meter can be remotely read (via the encoder) it is fitted
with a metrological controlled mechanical index. The reading of this index is the
measurement result that serves as the basis for billing.
Type TRZ 04-K has not been approved for custody transfer metering. The following
approvals have been obtained:
All RMG turbine meters have passed the disturbance measurements in compliance with
OIML Recommendation IR-32/89, Annex A, with slight and heavy flow disturbances.
Therefore, this meter design meets the requirements for installation in compliance with
Technical Guideline G13, Sec. 1. Test specifications are as laid down in PTB Testing
Instructions, Vol. 4, Volume gas meters, 2nd revised edition of 1992.
The RMG turbine meters of type TRZ 04 comply with EN 12261 and the European
Measuring Instrument Directive 2004/22/EG (MID).
17
1.7 Validity of meter proving
The turbine meters of type TRZ 04 which are suitable for custody transfer metering must
be subjected to subsequent meter proving at regular intervals. The validity of meter
proving is determined in the German Metering and Calibration Rules (MessEV, Edition 11,
December 2014) with the following periods:
Turbine meters without lubricator 8 years
Turbine meters with flow range 65 m3/h to 4.000 m3/h 12 years
Turbine meters with flow range > 4.000 m3/h to < 16.000 m3/h 16 years
Turbine meters with flow range 16.000 m3/h unlimited
Manual TRZ 04 · EN04 · February, 1st 2017
18
Introduction
Installation and Commissioning
1.8 Measuring ranges
The measuring ranges are between 10 and 25.000 m3/h (flowing conditions). A measuring
range is specified for each meter size. It is limited by the minimum flow rate Q
maximum flow rate Q
(see table on page 35).
max
For types TRZ 04 this is the flow range where the gas meter must indicate correct values
within the error limits specified by the German Metering and Calibration Rules.
Turbine meters of type TRZ 04 have measuring ranges up to 1:30 even under atmospheric
pressure. If a high-pressure test is conducted in compliance with Technical Guideline G 7
(PTB), the measuring range can be extended to 1:100. Then the minimum flow rate Q
HP is the lowest test point during high-pressure testing. Type TRZ 04 may then be used
for billing purposes within the specified HP flow and density ranges.
The measuring range of type TRZ 04-K is 1:16.
1.8.1 Extension of the measuring range
and the
min
min
In the range of 0.2 Q
max
to Q
, the measuring behavior of turbine meters is
max
determined by the aerodynamic conditions in the flow channel and the measuring cross
section. By means of many series of tests conducted both under atmospheric pressure
and under higher pressures and with an appropriate rating for such ranges, it is possible to
achieve a deviation of the calibration curve under atmospheric conditions and under highpressure conditions of < 0.5% in the flow range of
0.2 Q
max
to Q
. In the lower flow range, the measuring behavior is determined by the
max
relationship between the gas flow driving the measuring wheel and the slowing-down
torques due to drags (bearings and totalizer). The driving torques increase linearly with the
density and quadratic with the velocity of the gas to be measured. Due to the physical
conditions, the measuring range is therefore enlarged in relation to the density. The lower
flow limit shifts in the direction of smaller loads (see also the table in chapter 5.3 Qmin
depending on the operating pressure in natural gas).
Use the formula below as approximate equation:
2,1
QQ
minmd
3
)h/m(
The density can be determined with the following approximate formula:
3
nm
)m/kg(x)1p(
Manual TRZ 04 · EN04 · February, 1st 2017
Installation and Commissioning
The influence of the temperature is not taken into account in this formula.
With:
Qmd: Minimum flow rate at flowing conditions
: Minimum flow rate of the gas meter
Q
min
pm: Operating pressure in bar
: Density in kg/m3 (Density of air at 20°C and 1.01325 bar 1.2 kg/m3)
n: Standard density of the gas (standard density of natural gas 0,8 kg/m3)
Introduction
19
Manual TRZ 04 · EN04 · February, 1st 2017
20
05010
0
Introduction
Installation and Commissioning
1.9 Accuracy of measurement
The following error limits apply within the permissible measuring range:
Measuring range: Q
Calibration limit1)
TRZ 04
TRZ 04-K
2 % 1 %
1 % 0.5 %
2 %
(DN 50, DN 80: 3 %)
1)
maximum error due to the German Metering and Calibration Rules
It will be checked whether these limits are observed. They also apply to the high-pressure
range.
+2
min
to 0.2 Q
0.2 Q
max
to Q
max
max
1 %
(DN 50: 1.5 %)
+1
%0
-1
-2
1
Figure 2: Calibration curve of a turbine meter
The reproducibility, i.e. the difference between the results of two measurements under
identical conditions, is as follows:
TRZ 04, TRZ 04-K: 0.1%
Manual TRZ 04 · EN04 · February, 1st 2017
Installation and Commissioning
Introduction
1.10 Temperature ranges
For the standard designs of the turbine meters of types TRZ 04 and TRZ 04-K, the
following temperature ranges are permitted:
MID:
-25°C to +55°C (TRZ 04 for custody transfer metering)
PED 2014/68/EU:
-20°C to +80°C (Spheroidal graphite iron)
-40°C to +80°C (cast steel)
ATEX:
-25°C to +70°C (T4)
-25°C to +55°C (T6)
21
1.11 Pressure loss
The pressure loss of RMG turbine meters has been reduced to a minimum thanks to
modifications with regard to design. The measuring points for pressure loss are located 1 x
DN upstream and downstream of the gas meter. The pressure loss is calculated using the
following formula:
2
Q
Zp
p
where:
p pressure loss [mbar]
Z
pressure loss coefficient
p
density [kg/m3]
Q
volume flow rate at measurement conditions [m3/h]
m
DN is the nominal diameter of the gas meter [mm]
Gerätetyp Device type
TRZ 04 / TRZ 04-K turbine meter 3000
L1 perforated-plate straightener as per ISO/DIN 3150
L2 perforated-plate straightener as per ISO/DIN 6300
L3 perforated-plate straightener as per ISO/DIN 9450
LP-35 perforated-plate straightener as per RMG standard 1260
RB 19 tube-bundle straightener as per ISO/DIN 1260
DN
m
4
Zp
Manual TRZ 04 · EN04 · February, 1st 2017
22
Introduction
Installation and Commissioning
The values for Zp are approximate mean values. The exact value is calculated from the
pressure loss, which is determined during testing the turbine meter.
Example of calculation of the pressure loss for a turbine meter with upstream perforatedplate straightener:
TRZ 04, size DN 150
Qm = 650 m3/h
= 1.3 kg/m3 (natural gas)
From the table on the previous page you may find:
Z
p(TRZ04)
= 3000
Z
p(LP-35)
= 1260
Calculation:
Z
p(ges)
= 3000 + 1260
= 4260
2
650
3.14260p
150
4
= 4.2 mbar
1.12 Pressure tap
A pressure tap has been provided to connect the pressure transducer of a volume
corrector or a pressure gauge to read the pressure at measurement conditions prevailing
inside the gas meter. This pressure tap is identified by “pm”.
Manual TRZ 04 · EN04 · February, 1st 2017
Installation and Commissioning
Introduction
1.13 Using gas meters with different types of gases
Gas type Symbol Density at
0°C
Meter
case
Comments
1.013 bar
Acid gas Special Special measuring element
Air Ar 1.29 Standard
Ammonia
NH3
0.77 Standard O-rings/lubrication
Argon AR 1.78 Standard
Biogas Special Special measuring element
Butane C4H10 2.70 Standard
Carbon dioxide
CO2
1.98 Standard Exception: food industry
Carbon monoxide CO 1.25 Standard
Ethan
Ethylene (gaseous)
Freon (gaseous)
C2H6
C2H4
CCI2F2
1.36 Standard
1.26 Standard Special design
5.66 Standard O-rings/lubrication
Helium HE 0.18 Standard Reduced measuring range
Hydrogen
Hydrogen sulphide (0.2 %)
Methane
H2
H2S
CH4
0.09 Special Reduced measuring range
1.54 Standard Special measuring element
0.72 Standard
Natural gas 0.8 Standard
Nitrogen
Oxygen (100%)
Pentane
Propane
Propylene (gaseous)
Sulphur dioxide
N2
O2
C5H12
C3H8
C3H6
SO2
1.25 Standard
1.43 Standard Special design
3.46 Standard
2.02 Standard
1.92 Standard Special measuring element
2.93 Special Special design
Town gas
23
Manual TRZ 04 · EN04 · February, 1st 2017
24
Introduction
Installation and Commissioning
1.14 Installing the device in the pipeline
The units of RMG are equipped with flanges.
The flange connections for the connected pipes must comply with the connecting
dimensions of the flanges of the unit:
•
ANSI-pressure levels: Flange in accordance with ASME B 16.5.
•
DIN-pressure levels: Flange in accordance with DIN EN 1092.
Danger
Gas leakage for the wrong gaskets
If wrong flange gaskets are used for connecting turbine gas meters, gas may
flow through leaks resulting in an explosive gas mixture.
Poisoning and explosion hazard !!
In addition, the stress of the flange is increased inadmissible when tightening
the bolts.
Make sure that no flange gaskets are protruding into the pipe.
Pay attention to a save mounting / hanging-up of the TRZ04 during assembly to
avoid the risk of bruising. Avoid to have your fingers (or other body parts)
between any openings and gaps that are moved together when tightening the
flanges !!
Notice
Malfunction due to incorrect gaskets
When wrong flange gaskets are used for turbine meters, which protrude into
the pipe, the measurement accuracy may be affected.
Make sure that the flange gaskets on the sealing surfaces are not protrude
into the pipe.
The stability of the flange connection for gaskets and bolts with certain material properties
(see chapter 2.2.3 Seals and chapter 2.2.4 Screws) was proved to AD2000 regulations.
Other screw / flange - variants were not verified.
Manual TRZ 04 · EN04 · February, 1st 2017
Installation and Commissioning
d
2 Installation and Commissioning
2.1 Installation
Notice
Any components affecting the gas flow must be avoided directly upstream of
1)
the turbine meter (see DVGW Guideline G 492 II
and PTB Guideline G 13 2).
An inlet pipe of a minimum of 2 x DN is required upstream of the RMG TRZ
04 turbine meter. The inlet pipe must be designed as a straight pipe section
of the same nominal diameter as the gas meter. If there is a heavy flow
disturbance, it is necessary to use flow straighteners (see table on next
page). A pipe or fitting (bend) of the same nominal diameter as the gas meter
and an overall length of 2 x DN must be installed downstream of the gas
meter.
Temperature measuring instruments may only be installed at a distance of 1
x DN or in the case of nominal diameters of DN 300 at a minimum distance
of 300 mm.
If there are flow disturbances (e.g. due to a gas pressure regulator) upstream
of the inlet pipe, it is also necessary to use a perforated-plate straightener.
You can use perforated-plate straighteners complying with ISO 5167-1 or of
type RMG LP-35, the latter resulting in a pressure loss which is 2.5 times
lower than that of the standardized flow straightener.
2 DN
LP-35 perforated-plate straightener
25
2
D
1 DN
Perforated-plate straightener
N
d
DN
= e = 0.13 DN
•
DN
d
e
The opening angle of reducers or expansion fittings installed upstream of a turbine
meter of type TRZ 04 or TRZ 04-K must not exceed 30°.
In order to obtain precise measurement results, the turbine meter must be installed in
the gas line in such a way that no seals protrude from the flanges into the pipeline.
1)
can be ordered at DVGW, Josef-Wirmer-Str. 1-3, D-53123 Bonn (www.dvgw.de)
2)
can be ordered at PTB, Bundesallee 100, D-38116 Braunschweig (www.ptb.de)
Manual TRZ 04 · EN04 · February, 1st 2017
26
Installation and Commissioning
Notice
A protective screen should be installed on the intake side of the gas meter to
protect the turbine meter against any foreign particles possibly contained in
the gas flow. The protective screen can be a perforated plate of sheet metal
with a hole diameter of 3 mm which is available from RMG.
The pm tap which is located on the RMG turbine meter is the pressure-measuring point
which was used for taking the relevant pressure at measurement conditions during
meter proving. This pressure-measuring point is used for connecting pressuremeasuring instruments such as flow computers or volume correctors. Other connection
options (e.g. for temperature measurement) can be supplied on a pipe section on the
output side of the gas meter.
Danger
Protect the turbine meter from damage caused by heavy flow variations, e.g. if
the downstream pipeline system must be filled or blown off.
Danger
If it is necessary to do welding work on the gas line, such work can only be
performed at a safe distance from the gas meter. Extreme temperatures
prevailing in the gas line in the proximity of the gas meter can result in a
permanent damage to the gas meter.
Danger
All electrical connections between the gas meter and the amplifiers or the flow
computer must be carried out in compliance with the installation instructions.
Make sure that these connections are intrinsically safe.
Manual TRZ 04 · EN04 · February, 1st 2017
Installation and Commissioning
Danger
Any liquids remaining in the gas line after hydrostatic testing can damage the
interior parts of the gas meter. If hydrostatic testing is necessary, the turbine
meter must be replaced by a pipe section. Make sure that no liquid remains in
the gas line upstream of the gas meter after hydrostatic testing.
2.2 Operating data
Recommended threshold values for maximum service life and maximum accuracy:
Notice
Maximum overload:
Maximum flow or shock
loads
Maximum pressure
< 20% above Q
< 0.01·Q
max
/sec
e.g. starting up 0 - 100%: > 100 sec
< 0.1 bar/sec
change:
Maximum flow pulsation:
Particle size in the gas
< 5%
< 5 m
flow:
Lubrication of bearings
See chapter on “Lubrication”.
Lubrication intervals depend on the condition
of the gas (condensate, rust or dust)
Vibration / mechanical
shock:
< 1 mm/sec (vibration velocity)
These specified data must be determined and checked during commissioning, prior to
filling and during the start-up and running-in phases of the gas meters. If more than one
threshold value occurs at the same time, appropriate action must be taken in the station to
improve the measuring conditions.
, for a short time (< 30 sec)
max
1% of Q
ˆ
max
/sec
27
Manual TRZ 04 · EN04 · February, 1st 2017
28
Installation and Commissioning
Notice
The operator mustrecordall measuring data (gas meter and operating data)
during the whole period of operation in order to detect the causes of a
possible destruction of the gas meter at an early stage and take corrective
action in good time.
Corrective action or reduction of the critical operating conditions can be
achieved by the following measures for example:
Start-up screens (mesh width < 0.15 mm)
Filters
Perforated plates (Ø 3 - 4 mm) protecting the gas meter
Valves with control drive mechanism (flow variation)
Non-return valves (pulsation and return flow)
2.2.1 Technical Guideline G 13
The installation conditions for new stations complying with TRG G 13 and the simpler
installation conditions for RMG turbine meters are compared in the table below.
Type of flow
disturbance
Installation cond.
As per TR G13
Installation cond.
for RMG gas
Comments
meters Type TRZ
04
The outlet pipe can also be
designed as a bend.
Flow disturbances upstream
of this inlet pipe need not be
None
Inlet pipe 5 DN
Outlet pipe 2 DN
Inlet pipe 10 DN
Inlet pipe 2 DN
Outlet pipe 2 DN
considered, if the requirement for an alternating
and pulsating flow is fulfilled
Bend
Space bend
Inlet pipe 5 DN Inlet pipe 2 DN
Inlet pipe 5 DN
Inlet pipe 2 DN
and additionally 2
perforated-plate
straighteners or one
tube-bundle straightener
Manual TRZ 04 · EN04 · February, 1st 2017
Installation and Commissioning
Gas pressure
regulator with sound
attenuator
Inlet pipe 5 DN
Inlet pipe 2 DN
and additionally
one perforated-plate
straightener
Gas pressure
regulator without
sound attenuator
Diffuser
Inlet pipe 5 DN
and additionally
two perforated-plate
straighteners
Inlet pipe 5 DN
Inlet pipe 2 DN
and additionally
one perforated-plate
straightener
Inlet pipe 2 DN
and additionally
one perforated-plate
straighteners
Diffuser with swirling
flow
Inlet pipe 5 DN
and additionally
Inlet pipe 2 DN
two perforated-plate
straighteners
2.2.2 Perforated-plate straighteners
29
There are the following options for flow straighteners:
RMG L1 - L3 perforated-plate
straighteners complying with ISO 5167-1
and DIN 1952
e
d
a
RMG LP-35 perforated-plate straightener
D
d
D
Features ISO / DIN L1 to L3 RMG LP-35
Hole diameter d
Plate thickness e
Plate clearance a
Opening ratio m
Pressure loss, dyn. p
d 0.05 D
e d
0.5 D a 1 D
0.2 m 0.4
0.04 D 0.13 D
e = d 0.13 D
0.5 D -
0.3 0.6
5 - 15 (c² / 2) 2 - 15 (c² / 2)
D
e
Manual TRZ 04 · EN04 · February, 1st 2017
30
Installation and Commissioning
In conjunction with RMG turbine meters, these flow straighteners fulfil the requirements of
Technical Guideline G 13 and are approved under EU approval No. D 81 / 7.211.10 for
turbine meters.
2.2.3 Seals
It must be guaranteed that flange seals of RMG turbine meters do not protrude from the
flange into the gas line.
The stability of the flange connection has been verified for seals with the following
maximum material properties of AD2000 guideline:
− gaskets: k0 x KD = 20 x bD | k1 = 1.3 x bD [N/mm]
− grooved seals: k0 x KD = 15 x bD | k1 = 1.1 x bD [N/mm]
− spiral seals: k0 x KD = 50 x bD | k1 = 1.4 x bD [N/mm]
The stability of the flange connection was verified using the screws listed in this section in
combination with the seals listed in the previous section. Other screw / flange variants
were not verified.
31
Temperature ranges for screws and nuts
-10°C to +80°C -40°C to +80°C
Pressure Variant 1 Variant 2 Variant 3
up to and
including
40 bar
from 40 bar Screw bolts complying
Screws complying with
DIN EN ISO 4014
made of material 5.6,
Nuts complying with
DIN EN ISO 4032
made of material 5-2
with ANSI B1.1
made of material
ASTM A193 Grade B7,
Nuts complying with
ANSI B1.1
made of material
ASTM A194 Grade 2H
Screws complying with
DIN EN ISO 4014
made of material
25CrMo4,
Nuts complying with
DIN EN ISO 4032
made of material
25CrMo4
Screw bolts complying
with ANSI B1.1
made of material
ASTM A320 Grade L7,
Nuts complying with
ANSI B1.1
made of material
ASTM A320 Grade L7
Screw bolts
complying with
ANSI B1.1
made of material
42CrMo4,
Nuts complying
with ANSI B1.1
made of material
42CrMo4
Anti-fatigue bolts
complying with
DIN 2510
made of material
25CrMo4,
Nuts complying
with DIN 2510
made of material
25CrMo4
2.3 Meter head
The RMG turbine meter TRZ 04 or TRZ 04-K is fitted with the meter head of type “G”.
Manual TRZ 04 · EN04 · February, 1st 2017
32
Installation and Commissioning
Figure 3: Type “G” meter head
2.4 Standard version
The new standard design is type “G” meter head without mechanical drive shaft couplings.
Therefore, this design provides no options for connecting additional mechanical
equipment. The totalizer has the following features:
2 LF pulse transmitters (reed contacts)
1 tamper contact
Encoder type ENCO 08 (option)
Plastic case (optionally aluminum)
IP 66 + IP 67 protection class
Universally readable
Rotatable by 355° without breaking a seal
Notice
The electrical connector must be covered by a cap closure or a connecting
plug! Otherwise moisture may ingress into the meter head.
Manual TRZ 04 · EN04 · February, 1st 2017
Installation and Commissioning
The tamper contact switches when a magnet is held outside the meter head
close to the reed contacts.
2.5 Type “G-D” meter head
33
Figure 4: Type “G-D” meter head
Design and features like in the case of the type “G” meter head but additional with a
mechanical drive shaft in compliance with EN 12261 on the top side of the case. This
version is only possible with aluminum cap.
The direction of rotation of the drive shaft (with view on the drive shaft) is clockwise.
For dimensions and instructions for the connection see the description “Mechanical drive
shaft coupling”. Especially note the maximum permissible torques!
Type “G-D” meter head provides a mechanical drive shaft where additional equipment,
e.g. slip-on pulse transmitters, volume correctors, etc., can be attached under official
supervision.
Before you connect additional mechanically driven equipment, you must make sure that
the direction of rotation and the rotation rate U
the data on the gas meter. The torque required for driving the additional equipment must
not exceed the value stated on the indicating plate of the drive shaft.
ø 25
ø 6
(see chapter 5 Specifications) comply with
a
H
35
9
M6
2
14,7
6
20,7
24,5
Figure 5: Dimensions of the drive shaft coupling
2.5.3 Connection of additional equipment
The drive shaft couplings of the meter head (type “G-D”) can be used for connecting
additional mechanical or electronic measuring instruments. In this case, the total driving
torque must not exceed the approved value. If no additional equipment is used, the drive
shaft couplings are sealed.
The rotation rates of the drive shaft coupling depend on the meter size (see table).
Manual TRZ 04 · EN04 · February, 1st 2017
36
Installation and Commissioning
Notice
Attention
Additional equipment or additional measuring instruments can only be
attached or removed under official supervision. Before you connect
additional equipment, make sure that the direction of rotation and the
rotation rate of mechanical transmission coincide. Depending on the
location for attaching additional equipment, you can turn the meter head into
the most favorable position for installation without damaging the locking
seals by loosening two hexagon socket screws.
General
You can turn the meter head through 355° without breaking a seal. It is not necessary to
loosen a locking screw, the meter head is held in position because the rotary mechanism
is stiff. The rotation thus requires a slight effort.
Encoder
Figure 6: Type “G” meter head
Notice
If work has to be done on the meter head, please make sure that the official
seals are not damaged.
Manual TRZ 04 · EN04 · February, 1st 2017
Installation and Commissioning
r
C
2.5.4 Pulse transmitters
The TRZ 04 and TRZ 04-K turbine meters provide different sensors which supply volume
pulses in a variety of frequency ranges. These pulses can be further processed by volume
correctors or remote totalizers, for example. All pulse transmitters of the TRZ 04 have
been approved for custody transfer metering. The connection options are shown in the
following illustration.
onnection options
Volume corrector
1
E
E
LF
4
ENCO
1
E
E
1
E
E
E
HF3-DHF2-D
3
E
12
E
E
1
non [Ex] [Ex]
D
A
Servo-actuator
Gas flow recorder
Remote totalizer
Printer
Maximum-value recorder
Volume corrector
Alarm system
Volume corrector
0-20 mA
(0-4 mA)
(4-20 mA)
Min./max.-contact
flow indicator
Volume corrector
1 Ex i isolating amplifier
2 Digital-to-analog converte
3 Blade monitoring system
4 Tamper contact
37
Danger
All pulse transmitters are intrinsically safe and must be connected
only to intrinsically safe circuits if they are used in areas subject to
explosion hazards.
The safety barriers must meet the requirements of the
Ex ia IIC
or
For the NF and tamper contact the cable shield should only be
connected on the supply side.
Manual TRZ 04 · EN04 · February, 1st 2017
Ex ib IIC
type of protection.
38
Installation and Commissioning
2.5.5 Pulse transmitters in the meter head (LF)
The meter heads of the RMG TRZ 04 and TRZ 04-K and turbine meters are fitted with two
reed contacts as standard. The pulse value corresponds to the rotation rate U
. See the
a
indicating plate on the meter head for exact frequency data. The guide values are given in
the tables on pages 33-34. The maximum pulse frequency is 0.3 Hz.
2.5.6 Pulse transmitters in the measuring element (HF 2-D and HF
3-D)
The higher-frequency flow signals from the HF 2-D and HF 3-D proximity sensors are used
for control purposes and in conjunction with electronic flow computers suitable for custody
transfer metering. In the case of the HF 3-D pulse transmitter, the pulses are picked off at
the turbine wheel, while they are picked off at the reference wheel for the HF 2-D pulse
transmitter. The two wheels are arranged in such a way that two pulse series of the same
frequency are generated which are out of phase by 180°. The exact frequency is
determined during meter proving and is stated on a supplementary data plate on the meter
case. The guide values are listed in the tables on pages 33-34. The maximum pulse
frequency is approx. 2100 Hz at Q
, but depends on the meter size.
max
2.5.7 Tamper contact
A tamper contact is located directly next to the two reed contacts in the meter head. If the
attempt is made to block the reed contacts with a magnet held on the meter head, this
contact switches.
In passive state this contact is closed.
2.5.8 Encoder
With the rotary encoder type ENCO 08 the absolute totalizer reading of the mechanical
index is acquired and transmitted as a data record to a volume corrector. The encoder is
energy self-sufficient, which means, no battery or external power supply is necessary for
the operation. The energy required for digital transmission of the totalizer reading is
provided by the connected auxiliary device (volume corrector).
Manual TRZ 04 · EN04 · February, 1st 2017
Installation and Commissioning
Danger
The encoder is approved for use in hazardous areas, the marking is
as follows:
II 2 G Ex ib IIC T6
Warning
Attention must be paid to the selection of cables to guaranteethat the
permitted limits, in accordance with EC-type Examination Certificate of the
explosion protected equipment, are not exceeded.
Danger
The Ex-signal circuit is to be laid in its own cable. For the installation of the
encoder, a shielded cable may only be used (for example LIYCY 2 x 0.75 mm2,
coat color blue).
The cable shield has to be grounded at both sides. The earthing of the cable
shield is on the non-explosion protected site. At the other cable end is the
shield on the isolated mounted metallic ENCO 08 – housing.
Compliance with the regulations of DIN EN60079-14 the ENCO 08 housing can
grounded also in explosive areas.
The meter body of turbine and rotary meter and the aluminum housing of ENCO
08-M have to be grounded.
The fixed installation of the intrinsically safe cables is absolutely required.
39
You may find the connector pin assignments in the appendix Connector pin assignments
on page 57.
Manual TRZ 04 · EN04 · February, 1st 2017
40
Installation and Commissioning
2.6 Specifications of the pulse transmitters
Electrical data:
Reed-contact – Contact as make-contact element
max. switching voltage
max. switching current
max. conduct load
f
max
HF 2/HF 3 – proximity switch
Inductive proximity switch complying with DIN 19234 (NAMUR)
Supply voltage
current if switch is not operated
current if switch is operated
f
max
power
explosion protection acc. to
ATEX marking
ENCO 08 – Absolute Encoder
Supply voltage
Current
Power
Explosion protection acc. to
ATEX marking
20 V/DC
40 mA
200 mW
0,3 Hz
7 to 15 V DC
(with internal resistance Ri = 1 k)
3 mA
1 mA
2100 Hz
120 mW
II 1G Ex ia IIC T6
7 – 13,5 V DC
15 mA
51 mW
II 2G Ex ib IIC T6
Notice
The power rating in terms of the intrinsic safety, refer to the added
certificates (s. Annex)
Manual TRZ 04 · EN04 · February, 1st 2017
Installation and Commissioning
2.7 Temperature measurement
In order to measure the gas temperature, a resistance thermometer can be used in a
thermowell in a fitting downstream of the gas meter. It is recommended that a second
thermowell, e.g. for a monitoring thermometer, be used. If such thermowells are not
provided for in the meter case, temperature measurement must be performed at a
distance of up to 3 x DN or a maximum of 600 mm downstream of the turbine meter.
All turbine meters of types TRZ 04 from the nominal diameter of DN 80 (3") upwards can
be fitted with a thermowell for a PT100 resistance thermometer.
Due to their short size, turbine meters of type TRZ 04-K cannot be fitted with a thermowell.
Operating temperatures prevailing at outdoor metering stations in winter or at natural gas
metering stations downstream of regulating stations are usually in the range of -5C to
+10C. By comparison, higher operating temperatures prevail at metering stations located
downstream of compressor stations. For this reason, the measuring elements of the
temperature sensors located outside the gas line must be appropriately insulated against
the weather. In order to achieve optimum thermal conduction, oil as a heat transfer liquid
should be filled in the thermowells in any case.
41
2.8 Commissioning
2.8.1 Filling with oil
Notice
When the turbine meters are delivered, the lubricators do not contain any oil.
The oil pumps must not be filled with oil until they are at the installation site
prior to commissioning! A small bottle of oil is supplied with each gas meter.
Then you have to perform an initial lubrication.
For this see the detailed lubrication instructions in chapter 3.2 Lubrication.
2.8.2 Connecting the gas flow
Do not fill any downstream pipelines or station sections through the turbine
meter. This may speed up the turbine wheel and lead to excessively high
loads with resultant damage.
Warning
Manual TRZ 04 · EN04 · February, 1st 2017
Installation and Commissioning
Notice
42
Short-time overload operation of 20% above the maximum flow rate Q
max
is
permissible. Such load conditions should be avoided, however, since in
most cases they cannot be controlled and exceed the officially approved
limits for custody transfer metering. Moreover, such overloads reduce the
service life of the gas meter.
Warning
The gas flow must be free of foreign particles, dust or liquids. Otherwise it is
recommended that filters or separators be installed.
Danger
With the start-up of this device it is to be observed that all cables were correctly
laid and connected.
The housing must be completely closed.
With the connection and the start-up of the encoder, observe the guidelines of
the appropriate norms
DIN EN 60079-0 and
DIN EN 60079-14
Manual TRZ 04 · EN04 · February, 1st 2017
Operation
3 Operation
3.1 Influences on the error of measurement due to
operation
3.1.1 Intermittent operation
Rapid changes of the gas flow should be avoided, since the turbine wheel
can follow these changes only with a certain delay due to its inertia. In this
way, errors of measurement occur, in particular, if the gas flow is shut off.
Since in this case the turbine wheel slows down after the flow has stopped, a
gas volume is measured which is always higher than that which has actually
flowed through the gas meter.
Warning
3.1.2 Influence of pulsations
43
The gas flow must be free of shocks or pulsations. A gas metering station can be affected
by flow pulsations, however, if the following equipment has been installed upstream or
downstream in the system:
reciprocating piston compressors,
rotary displacement meters,
gas pressure regulators lacking steadiness of operation,
pipes where no gas flows (siphons).
Volume flow pulsation is the decisive quantity for evaluating the performance of gas
meters under the influence of pulsations. Volume flow pulsation is physically always
associated wit
h
pressure variations
.
The following relation is established in a first
approximation:
DN
Q
2
K
pQ
relrel
With:
^
Q
: is the relative volume flow pulsation (peak-to-peak)
rel
Q¯: is the mean volumetric flow
^
p
: is the relative pressure pulsation (peak-to-peak)
rel
DN: is the nominal diameter of the gas meter
K: is a constant depending on standard density, velocity of sound, compressibility,
pressure at base conditions, temperature, and station-specific parameters.
Manual TRZ 04 · EN04 · February, 1st 2017
44
Operation
With this relation, it is possible to estimate volume flow pulsation on the basis of pressure
pulsation, which can be measured more easily.
Direct measurement of volume flow pulsation is preferable, however, since the results are
more reliable. The crucial factor is the pulsation
at the location of measurement.
3.1.3 Consequences
Notice
In the case of pulsating flow, the turbine meter shows a measured value
which is too high.
Due to the pulse applied by the flow on the impeller which increases quadratically with the
flow velocity, the resultant rotational speed is higher than the mean value of the flow
velocity.
The influence with high gas density is less, but increases with a high mass moment of
inertia (heavy impeller) or fast-running wheels.
Warning
Furthermore,
High pulsation amplitudes may result in premature wear of shaft bearings
due to increased load.
3.1.4 Limiting values
Frequency ranges
It is generally unlikely that measured values will be distorted in the frequency range
above 100 Hz.
In practice, it is hardly possible to initiate any significant flow variations at such
frequencies.
Disturbances are to be expected most frequently in the frequency range between 0.1 Hz
and 100 Hz, since with typical station dimensions, the gas column can be expected to
produce resonances. Flow variations with a high relative amplitude may occur.
In the frequency range below 0.1 Hz, there is a quasi-steady flow which will not cause
any distortion with the gas meters.
Manual TRZ 04 · EN04 · February, 1st 2017
Operation
Pulsation amplitudes
Studies have shown that no disturbances are to be expected in the case of relative flow
pulsations below 5% (peak-to-peak) and relative pressure pulsations of less than
0.1% to 0.5% (peak-to-peak).
These data should be regarded as approximate values depending on the flow rate and
pulsation frequency involved.
45
Manual TRZ 04 · EN04 · February, 1st 2017
46
Operation
3.2 Lubrication
3.2.1 Lubricator
For lubrication permanently fitted oil pressure pumps are used. The various designs are
given in the tables below.
KO
GO
DS
DN PN 10, 16
50 DS (KO) KO DS (KO) KO
80 DS (KO) KO DS (KO) KO
100 DS (KO) KO DS (KO) KO
150 DS (KO) KO DS (KO) KO
200 KO KO
Types in brackets are options
3.2.2 Specification for lubricating oils
Type of oil pump Container volume Delivered quantity
Pushbutton pump
Lever pump
Permanent lubrication -- --
TRZ 04-K TRZ 04
Pressure classes Pressure classes
PN 25, 40, 64, 100
ANSI 150
ANSI 300, 600
8 cm3 0.114 cm3/stroke
150 cm3 1.5 cm3/stroke
PN 10, 16
ANSI 150
PN 25, 40, 64, 100
ANSI 300, 600
Notice
For lubrication purposes, we recommend that you use only
Shell Tellus S2 MA 10 or another lubricating oil with 2 to 4°E at 25°C in order
to avoid damage to the shaft bearings. You can order your supply of
lubricating oil from us in 1-liter containers under Ref. No. 82.11.148.00.
Shelf life of oil:
The shelf life depends on the operating conditions (e.g. UV light, humidity ..).
In principle the oil suffers no quality loss during the first 3-4 years.
Manual TRZ 04 · EN04 · February, 1st 2017
Operation
3.2.3 Initial lubrication
When the turbine meters are delivered, the lubricators do not contain any oil. The oil pumps
must not be filled with oil until they are at the installation site prior to commissioning. A small
bottle of oil is supplied with each gas meter.
For the initial lubrication more pump strokes are necessary than for the relubrications because
first the oil pipes have to be filled with oil.
Pump DN 50 – DN 200 DN 250 – DN 600
KO (pushbutton pump) 40 strokes 45 strokes
GO (lever pump) - 10 strokes
3.2.4 Relubrication
Notice
The relubrication intervals are stated on an indicating plate on the meter
case. See also “Lubrication procedure”.
47
Example:
Relubrication!
Every 3 months 2 strokes
Lubricating oil: 2-4°E at 25°C
see operating instructions
Notice
In the case of unfavorable operating conditions, such as condensate forming
because of water or hydrocarbon, or dust-laden gas, or service temperatures
above 50°C, we recommend that lubrication be performed at shorter
intervals, even daily under extreme conditions (permanent formation of
condensate).
If the gas meter is operated under the above-mentioned conditions, it is
likely that its service life will be reduced. If you have questions to the
relubrication in such cases please contact RMG.
Manual TRZ 04 · EN04 · February, 1st 2017
48
Operation
3.2.5 Lubrication procedure
To fill the pumps and lubricate the turbine meters, proceed as follows:
Pushbutton pump (KO)
To fill with oil
Unscrew cover.
Fill with oil.
Screw on cover again.
To lubricate up to 50 bar
Unscrew knurled cover (screwed all the way to
the pump casing when delivered)
Press pushbutton, which is now visible, once
for each stroke (delivered quantity: 0,114
cm3/stroke)
Screw on knurled cover again. Make sure you
stop screwing as soon as you can feel a
resistance and leave a gap.
To lubricate over 50 bar
Loosen the knurled cover (screwed all the way
to the pump casing when delivered)
Screw on the pushbutton tightly, once for each
stroke (delivered quantity: 0,114 cm3/stroke)
Screw on knurled cover again. Make sure you
stop screwing as soon as you can feel a
resistance and leave a gap.
Every 3 months 6 strokes
Manual TRZ 04 · EN04 · February, 1st 2017
Operation
3.2.6 Maintenance
RMG turbine meters are maintenance-free except that they require lubrication at regular
intervals. Since all gas meters fitted with oil pumps are shipped with an empty oil storage
tank, it is absolutely essential to fill the storage tank of the oil pump with oil and perform
the initial lubrication before the gas meter is put into service (see section “Lubrication”).
The operator should nevertheless check the turbine meter at regular intervals. In this
connection, please refer also to DVGW code of practice G 495 (Gas pressure regulators
for monitoring and servicing large-scale gas metering systems).
The gas meter should be checked for accuracy approximately every two years depending
on whether it is possible or necessary to perform such checks. This can be done as
follows:
In the station itself by connecting two gas meters in series.
In an officially acknowledged testing laboratory for gas measuring instruments.
In the factory.
49
Manual TRZ 04 · EN04 · February, 1st 2017
50
Labelling
4 Labelling
All important data which are necessary for operating the gas meter are stated on the
plates which are affixed to the meter case, totalizer or pulse transmitters.
Main data plate of the TRZ 04 turbine
DN
Q
t
Q
max
Q
min
by Honeywell
PS
m³/h
p
m³/h
TS
G
m³/h
LF reed 1m =
EN 12261 / AC 1,0
3
MXX
bar
TRZ 04
bar
No.
°C
Year
pulses
0102, 0091
t =-25°C...+55°C (TC2)
am
T10417
m
³
4
LF
1. reed 1
2. reed 2
3. control
PIN 1+4
PIN 3+6
PIN 2+5
2
1
1
7
3
8
5
6
Encoder
2
1
4
1. EZD-prot.+
3
2. EZD-prot.-
3. ground
PIN 1
PIN 3
PIN 4
meter with type “G” meter head
TS: Permissible temperature range for meter case
(PED 2014/68/EU)
PS: Maximum permissible pressure for meter case
(PED 2014/68/EU)
t
: Permissible temperature range for custody
am
transfer metering (MID)
Built-in pulse transmitters,
type “G” meter head
Pin 7 are 8 are not connected.
Built-in encoder
with type “G” meter head
Pin 2 is not connected.
Re - Lubrication
At 3 month intervals, 2 strokes.
Lubricant: 2-4°E at 25°C
Supplementary information
see operating instructions.
Indicating plate for relubrication of the
main bearings
Manual TRZ 04 · EN04 · February, 1st 2017
Re- Lubrication
At 3 month intervals, 6 strokes.
Lubricant: 2 - 4°E at 25°C
The containervolume of 0,15 litre is
sufficient for 3 months. Supplementary
information see operating instructions.
Labelling
Indicating plate for relubrication of the
main bearings
p
m
Arrow indicating the direction of flow
Indication of the reference pressure tap
51
Manual TRZ 04 · EN04 · February, 1st 2017
52
Specifications
5 Specifications
5.1 Meas. ranges/Dimensions/Pressure classes of the
TRZ 04
U
DN Size Measuring range
mm Q
in
Q
min
m3/h m3/h m3/h m3
Qt mm
max
DN Pressure classes/ Approx. weight.Dimensions
a
in PN kg ANSI kg L H C
50
80
100
150
200
G 65 5 100
2"
G 100 8 160 32 1
G 160 13 250 50 1 25/40 25 300 25 240 23096
3"
G 250 20 400 80 1 600 36
G 160 13 250 50 1
G 250 20 400 80 1 25/40 32 300 35 300 260120
4"
G 400 32 650 130 1 600 55 270
G 400 32 650 130 1
G 650 50 1000 200 1 25/40 60 300 65 450 265180
6"
G 1000 80 1600 320 10 600 100 285
8"
G 1000
G 1600
80
130
1600
2500
20
320
500
0,1
10
10
50
2"
80
3"
100
4"
150
6"
200
8"
10/16 13 150 13
25/40 21 300 13 150 21060
600 21
10/16 20 150 20
10/16 25 150 30 240
10/16 50 150 50 265
10/16 75 150 100 300
25/40 95 300 120 600 320240
64 150 600 160 320
PN pressure classes complying with DIN EN 1092,
ANSI pressure classes complying with B16.5.
Turbine meters for all pressure ranges and pipe
sizes can be equipped with a thermowell to
accommodate a resistance thermometer.
Manual TRZ 04 · EN04 · February, 1st 2017
Specifications
5.2 Meas. ranges/Dimensions/Pressure classes of
TRZ 04-K
Nominal Meas. range
dia / DN Qmin-Qmax
mm in
50 2“ 6-100 0,1 PN 10, 16, 25, 40
80 3“ 13-160
100 4“ 25-400
150 6“ 40-650
200 8“ 100-1600
m3/h m3
16-250
(25-400)
(40-650)
65-1000
(100-1600)
160-2500
Ua
1 PN 10, 16, 25, 40
1 PN 10, 16, 25, 40
1 PN 10, 16, 25, 40
10 PN 10, 16, 25, 40
Pressure
classes
ANSI 150, 300
ANSI 150
ANSI 150
ANSI 150
ANSI 150
L H Weight Case design
mm mm App. kg
150 212 10
120 245 14
150 255 25
175 285 40
200 305 60
Flanschausführung
L
On request you may get meters for other pressure ranges and other flange types as well.
5.3 Q
depending on the operating pressure in
min
natural gas
53
H
Values for custody transfer metering, according to MID approval
Meter case Spheroidal graphite iron, cast steel or welded
steel
Flow straightener Delrin, aluminum or steel
Turbine wheel Delrin or aluminum
Measuring element case Aluminum
Measuring element bea-
rings
Ball bearings Stainless steel
Shafts Stainless steel
Gearing Stainless steel or plastic
Magnetic coupling Stainless steel
Meter head cap Plastic or aluminum
Totalizer Plastic
Meter head with mechanical
index, Encoder and LF pulse
transmitters
Turbine wheel
Reference wheel
HF pulse transmitters
Figure 7: Sectional drawing of a turbine meter.
Manual TRZ 04 · EN04 · February, 1st 2017
Annex
Connector pin assignments
On the rear side of the meter head there are two sockets for Binder plugs of series 713.
The left one (optional) for connection of the encoder, the right one for the LF pulses. The
figure shows the view on the plugging side of the sockets.
Contact ENCO 08 LF / tamper HF 2-D / HF 3-D
3
4
4
3
1
2
5
6
8
7
12
3
2
4
1
1 Vo + LF 1 HF +
2 - tamper contact 3 Vo - LF 2 HF 4 PE LF 1 not connected
5 tamper contact
6 LF 2
7 not connected
8 not connected
Cable diameter for HF 2-D, HF 3-D and ENCO 08: 6-8 mm.
57
Manual TRZ 04 · EN04 · February, 1st 2017
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.