• Read and understand this operating manual before using the
detector.
• You must operate the detector in accordance with the operating
manual.
• Regardless of warranty period, we shall not make any
compensation for accidents and damage caused by using this
product. The compensation shall be made only under the warranty
policy of products or parts replacement.
• Because this is a safety device, a regular maintenance for every
six months and daily maintenance must be performed.
• If you find abnormalities in the detector, please notify them to our
local representative immediately.
RKI Instruments, Inc. 33248 Central Ave. Union City, CA 94587
Phone: 800-754-5165 Fax: 510-441-5650
Preface
Preface
Thank you for choosing our gas detector head GD-70D Series. Please verify that the model number of the
product you purchased is included in the specifications on this manual.
This manual explains how to use the detector and its specifications. It contains information required for
using the detector properly. Not only the first-time users but also the users who have already used the
product must read and understand the operating manual to enhance the knowledge and experience before
using the detector.
<Definition of DANGER, WARNING, CAUTION, and NOTE>
DANGER
WARNING
CAUTION
NOTE
Indicates an imminently hazardous situation which, if not avoided, will result in
death or serious injury or serious damage to the product.
The use of this symbol is to be limited to the most extreme situation.
Indicates a potentially hazardous situation which, if not avoided, could result in
death or serious injury on the human body or object.
Indicates a potentially hazardous situation which, if not avoided, may result in
minor or moderate injury or some damage on the human body or objects. It
may also be used to alert against unsafe practices.
This means "ADVICE" at operation.
<Danger cases>
DANGER
This is not an explosion-proof device. You must not use it to detect gases exceeding the lower limit of explosion.
<Warning cases>
WARNING
Power Supply
Before turning on the detector, always check whether the voltage is properly applied. Do not use an unstable
power supply because it may cause malfunctions.
Need of grounding circuit
Do not cut the grounding circuit or disconnect the wire from the grounding terminal.
Defects in protective functions
Before starting the detector, check the protective functions for defects. When seeming defects are found in the
protective functions, such as protective grounding, do not start the detector.
External connection
Before connecting the detector to external equipments or external control circuit, securely connect it to a
protective grounding circuit.
Tubing
The detector is designed to suck gases around it under the atmospheric pressure. If excessive pressure is applied
to the sampling inlet and outlet (GAS IN, GAS OUT) of the detector, detected gases which may be leaked from its
inside pose dangers. Be sure that excessive pressure is not applied to the detector when it is used. Detected
gases must be exhausted from the detected gas exhausting outlet (GAS OUT) on the bottom of the detector to
which an exhaust tube is connected, to a point regarded as a safe place.
Handling the sensor unit
Do not disassemble the electrochemical type sensor unit (ESU) or galvanic cell type sensor unit (OSU) because
they contain electrolyte. Electrolyte may cause severe skin burns if it contacts skin, while it may cause blindness if
its contacts eyes. If electrolyte is adhered on you clothes, that part on your clothes is discolored or its material is
decomposed. If contact occurs, rinse the area immediately with a large quantity of water.
Zero adjustment (AIR Adjustment) in the atmosphere
When zero adjustment (AIR Adjustment for oxygen deficiency alarm) is performed in the atmosphere, check the
atmosphere for freshness before beginning the adjustment. If other gases exist, the adjustment cannot be
performed properly, thus leading to dangers when the gas leaks.
Operation in a gas
Do not operate the detector in a place where flammable/explosive gases or vapors are present. Operating the
detector in such an environment will lead to extreme dangers.
Response to a gas alarm
Issuance of a gas alarm indicates that there are extreme dangers. Take proper actions based on your judgment.
GD-70 - 4 -
Important Notices on Safety
<Precautions>
CAUTION
Do not use a transceiver (walkie-talkie) near the detector.
Radio wave from a transceiver near the detector or its cables may disturb commands. when using a transceiver, it
must be used in a place where it disturbs nothing.
To restart the detector, you must wait five seconds more before doing it.
Restarting the detector within five seconds may cause errors.
Verify that the reading on the flow rate indicator corresponds to the specified flow rate before using the
detector.
If it does not correspond to the specified flow rate, gas detection cannot be performed properly. Check whether
the flow rate is unstable or lost.
Attach the dust filter before using the detector.
Before using the detector, attach the specified filter to prevent disturbances by possible gas adsorption or air dust.
A dust filter to be used is varies depending on the gas to be detected. For more information on dust filters, please
contact our sales department.
Observe the operating restrictions to prevent condensation inside the tube.
Condensation formed inside the tube causes clogging or gas adsorption, which disturb accurate gas detection.
Thus, condensation must be avoided. In addition to the installation environment, carefully monitor the
temperature/humidity of the sampling point to prevent condensation inside the tube. In particular, when detecting
a gas which is solved into water and corrodes contacted materials, such as a strong acid gas, the gas is
undetectable and furthermore may corrode internal parts. Please observe the operating restrictions.
Do not use the external output of the detector to control other devices.
This is not a control device. You are not allowed to use its external output to control other devices.
Do not disassemble/modify the detector, or change the settings if not necessary.
Disassembling/Modifying the detector will invalidate the guarantee of the performance. Changing the settings
without understanding the specifications may cause alarm malfunctions. Please use the detector properly in
accordance with the operating manual.
Do not forget to perform a regular maintenance.
Since this is a safety device, a regular maintenance must be performed to ensure its safety. Continuing to use the
detector without performing a maintenance will compromise the sensitivity of the sensor, thus resulting in
inaccurate gas detection.
- 5 - GD-70D
1 Overview 1-1. Product components
1
Overview
1-1. Product components
<Gas Detector (GD-70D Series)>
<Standard Accessories>
• Operating manual
• Protective rubber cap (to be removed when using the
detector)
• Dedicated handling lever (for wiring)
• Dust filter
• Interference gas removal filter (to be supplied with some
sensor units)
<Pyrolyzer Unit (PLU-70) (Option)>
* This is needed in "pyrolyzer + electrochemical type (ESU)"
and "pyrolyzer + pyrolysis-particle type (SSU)".
For more information on the pyrolyzer unit (PLU-70), see the
individual operating manual.
Pyrolyzer unit
1-2. Purpose of use
• This is a fixed type gas detector head which detects leak of gases used in semiconductor factories, etc.
• The gas detector is a safety device, not an analyzer or densitometer which performs quantitative/qualitative
analysis/measurement for gases. You must understand the features of the detector before using it, so that
you can use it properly.
• The detector detects abnormalities in the air caused by presence of gases or other reasons (leak and
oxygen deficiency) with the built-in gas sensor unit. The concentrations of detected gases are shown on the
Character LCD.
• The built-in pump in the detector sucks gas to perform gas detection.
• The detector has two-step gas alarm contact and fault alarm contact.
• The detector outputs gas concentration in 4 - 20 mA or digital data.
GD-70D Analog Transmission (4-20 mA specification)
GD-70D-NT 2-wire type DC power-line communication (NT specification)
GD-70 - 6 -
1 Overview 1-3. Product specifications
1-3. Product specifications
<Common Specifications>
Concentration
Display
Suction Flow Rate
Power Display POWER lamp on (green)
Displays Gas name display/Flow rate indicator/Mode display/Communication status
Gas Alarm Display First: ALM1 lamp on (Red)/Second: ALM2 lamp on (Red)
Gas Alarm
Activation
Gas Alarm Contact No-voltage contact 1a or 1b (2 step independent)
Tube Connecting
Hole
Initial Clear Approx. 25 seconds
Structure Box type/Wall mounted type
External
Dimensions
Weight Approx. 0.9 kg
Outer Color Gas detector: grey
Character LCD (Digital and Bar Meter Display)
0.5L/min ±10%
display/Pyrolyzer connection display
Auto-reset
de-energized (energized at a time of alarm) or energized (de-energized at a
time of alarm)
System abnormalities/Sensor abnormalities/Flow Rate
abnormalities/Communication abnormalities/Pyrolyzer abnormalities
FAULT lamp on (Yellow)/Detail display
Auto-reset
No-voltage contact 1a or 1b
de-energized (energized at a time of alarm) or energized (de-energized at a
time of alarm)
Cable of CVV, etc. (1.25sq) - max. 6-core
Flow control/Calibration history/Alarm trend history/Event history
Rc1/4 (O.D Φ6-1t half-union for Teflon tube<PP>supplied)
Approx. 70(W)x120(H)x145(D) mm (projection potions excluded)
Front door: white
<Specifications for Each Model>
Model GD-70D GD-70D-NT
Transmission
System
Transmission
Specification
Recommended
Transmission
Cable
Recommended
Power Cable
Power Supply
3-wire type analog transmission
(Common Power Supply <Power,
Signal, Common>)
or 2-wire type analog transmission
4 - 20 mA DC (no-insulation/load
resistance under 300Ω)
Shielded cable of CVVS, etc. (1.25sq)
3-core or 2-core
Cable of CVV, etc.(1.25sq) - 2-core
(Those for 3-wire type are common with
the transmission cable.)
24VDC ±10% 24VDC ±10%
- 7 - GD-70D
2-wire type DC power-line
communication
DC power-line communication
system
Shielded twisted-pair cable
(1.25sq) of KPEV-S, etc. - 1P
(Common with the transmission
cable)
(Dedicated line by blocking filter)
1 Overview 1-3. Product specifications
<Specifications for Each Principle>
Detection
principle
Gas to be
detected
Detection
range
*1 Pyrolyzer unit combination specification. For information on the specifications
*2 Oxygen deficiency specification. For information on other specifications (oxygen leak
*3 The operating temperatures/humidity may be different depending on the gas to be
*4 To the alarm setpoint
*5 By providing the gas 1.6 times the alarm setpoint (excluding delay in the tube and in the
*6 By letting the detector detect the gas of 10 - 11 vol% (excluding delay in the tube and in
ESU
Electrochemical
type
Toxic gas *3 NF3 TEOS Flammable gas
Depending on
the gas to be
detected
Pump suction
type
Depending on
the gas to be
detected
30-70%RH 30-80%RH Below 95%RH
of the pyrolyzer unit (PLU-70), see the individual operating manual.
specifications), please contact us.
detected.
communication)
the communication). Within 30 seconds for 90% response
ESU+PLU
*1
Electrochemical
type
0-30ppm 0-15ppm Depending on
Pump suction type/pyrolysis
type
10ppm(1st)
20ppm(2nd)
Within ±30% *4
60 seconds or less *5
(H-HH)
Approx. 1.5 W
(Max. 4 W)
SSU+PLU
*1
Pyrolysis-
particle type
10ppm(1st)
10ppm(2nd)
0-40°C
SGU OSU
Semiconductor
type
Toxic gas
the gas to be
detected
Pump suction type
Depending on
the gas to be
detected
Flammable:
within ±25%
Toxic: within
±30%
*4
Flammable: 30
seconds or less
Toxic: 60
seconds or less
*5
Approx. 2.5 W
(Max. 5 W)
Galvanic cell
18vol%(1st)
18vol%(2nd)
Within ±1 vol%
<accuracy of the
Within ±0.7 vol%
5 seconds or less
two-step alarm
Approx. 1.5 W
*2
type
Oxygen
0-25vol%
reading>
*6
(L-LL,L-H)
(Max. 4 W)
GD-70D- 8 -
1 Overview 1-4. Names and functions for each part
y
1-4. Names and functions for each part
<Appearance>
M5 mounting hole
GAS OUT
GAS IN
Protective cover for power switch
Screw for grounding
earth rod
External wire hole
Connector for PLU
Connector for PoE
Display
Operating
switch
Sensor unit
displa
Power switch
(Unit: mm)
- 9 - GD-70D
1 Overview 1-4. Names and functions for each part
y
<Front Panel and Character LCD>
c MODE key
d TEST/SET key
e ▲key
f ▼key
⑭Concentration
bar indicator
⑮Alarm setpoint
indicator
⑫Concentration
value displa
⑰Communication
indicator
⑯Flow rate
indicator
hSensor unit nameplate display window
gLock lever
①
②
③
④
⑤
⑥
⑦
⑧
⑨
⑩
⑪
⑫
⑬
⑭
⑮
⑯
⑰
⑱
⑲
⑳
MODE key
TEST/SET key
▲ key
▼ key
Lock lever
Sensor unit nameplate
display window
Power lamp (POWER)
First alarm lamp
(ALM1)
Second alarm lamp
(ALM2)
Fault lamp (FAULT)
Gas name display
Concentration value
display
Unit display
Concentration bar
indicator
Alarm setpoint
indicator
Flow rate indicator
Communication
indicator
Maintenance indicator
Inhibit indicator
Pyrolyzer unit
connection indicator
Used to enter the maintenance mode.
It is also used to cancel or skip in a specific mode.
Used to enter the test mode.
It is used for value confirmation and so on in a specific mode.
Used to switch screen or change a value (UP).
Used to switch screen or change a value (DOWN).
Lever to lock the main unit. Push it to attach or detach the main unit.
Window to look at the nameplate of the sensor unit.
You can identify the currently attached sensor unit.
Power lamp. It lights in green when the power is on.
First alarm lamp. It lights in red when the first alarm is reached.
Second alarm lamp. It lights in red when the second alarm is reached.
Fault lamp. It lights in yellow when an abnormality is detected in the detector.
Displays the gas name in chemical formula and so on (e.g. Silane = SIH4).
Displays the gas concentration and so on.
Displays the unit according to the specification (ppm, ppb, vol%, %, %LEL).
The detectable range (full scale = FS) is divided into 20 with bars. The
increase in concentration is displayed in proportion to the full scale.
The alarm setpoints (AL1 and AL2) are indicated on the concentration bar.
Displays the suction flow rate. The center of the bars means the normal
suction flow rate of 0.5 L/min.
For GD-70D-NT, this indicator is displayed while transmitting data with the
upper unit (TX, RX).
Displayed during the maintenance mode. When this indicator is displayed,
the alarm contact is disconnected to be disabled.
Displayed when the inhibition (point skip) is set.
Displayed when the dedicated pyrolyzer unit (PLU-70) is connected.
jFirst alarm lamp
iPower lamp
⑱Maintenance
indicator
kSecond alarm lamp
lFault lamp
⑳Pyrolyzer unit
connection indicator
⑬Unit display
⑪Gas name display
⑲Inhibit indicator
GD-70D - 10 -
1 Overview 1-5. Units description
1-5. Units description
The detector consists of the following units.
CAUTION
• Each unit consists of precision parts. When a unit is detached, be careful not to drop it. Dropping the
unit compromises its original performance or causes malfunctions.
Wall-mounted unitMain unit
Pump unit
Sensor unit
<Detaching and Attaching Sensor Unit>
1. Push the two buttons at the top of the main unit together to
open the front cover. (Pushing only either one button cannot
open the cover, but then pushing the other can open it.)
* The front cover opens about 90 degrees at first. You can
push it down to 180 degrees.
CAUTION
• Turn the power off before detaching or attaching the
sensor unit.
2. Hold the convexed parts on both sides of the sensor unit and
pull it out.
If the sensor unit is hard to detach, insert a finger into the gap
at the top right of the sensor unit and pull it out.
3. When attaching the sensor unit to the main unit, thrust it onto
the main unit and make sure that the sensor unit is securely
fixed, otherwise it might come off.
4. After replacing the sensor unit, close the front cover. Be sure
to check the click of the front cover to fix it properly, otherwise
it might open again.
- 11 - GD-70D
1 Overview 1-5. Units description
<How to Use Sensor Unit>
A sensor unit installed in the detector is the same regardless of
the detection principle, thus sensor units are interchangeable.
Each sensor unit has a different color in accordance with the
principle as shown below. How to handle the sensor unit is varies
depending on its principle.
CAUTION
• The sensor unit must be handled carefully to ensure quality as safety device. When the sensor unit is
stored, a dedicated warehouse and power equipment for the sensor unit are needed. In principal, the
sensor unit must not be detached from the detector when it is handled or stored. We will take care of
your sensor unit.
• Be sure that the sensor unit is not installed improperly.
If a sensor unit of different specification or principle from the one shipped from the manufacturer, a
message will be displayed on the LCD of the detector (C-02).
If the message is displayed, check the specifications of the sensor unit.
• After the sensor unit is replaced, always perform a gas calibration (zero adjustment and span
adjustment).
Electrochemical
type (ESU)
Pyrolysis-particle
type (SSU)
• Do not disassemble the sensor unit because it contains electrolyte. If contact
occurs, rinse the area immediately with a large quantity of water.
• The sensor unit identifies the direction. Put the sensor unit in the dedicated case
while handling it. Do not place it on its side or upside-down.
• When a new sensor unit is installed, it must be warmed up.
Although warm-up time is different depending on the type of the installed sensor,
it is recommended that warm-up should be performed for three hours or more.
Please contact our sales department for more
information.
• The sensor unit must be stored in a clean, cool and
dark place away from direct sunlight. Some types of
the sensor units cannot be stored together with other
units. Please contact our sales department for more
information.
Dedicated case
• Although the sensor unit contains radioactive materials, it is certified as a
specified designing certification device, which is regarded as a device having
no influences on health. Observe the "Safety Manual" which stipulates
conditions for the certification. To dispose of the sensor unit, you must return
it to us. You do not need to take any additional actions.
• The sensor unit contains a small amount of radioactive materials. Do not
disassemble it, or dispose of it like other wastes.
• The sensor unit must be put into the dedicated case specified by us, and
stored away from direct sunlight in a clean place
where the temperature and humidity are maintained
at a normal level and where appropriate measures
are introduced to prevent it from being taken out
easily.
• When the sensor unit is transported out of your
Dedicated case
factory, please use a transportation company which
can handle specified designing certification devices
(L-type packages).
• For more information, see the "Safety Manual".
GD-70D- 12 -
1 Overview 1-5. Units description
Semiconductor
type (SGU)
Galvanic cell type
(OSU)
• Before using the sensor unit, it needs to be warmed up for a specified time.
The sensor unit is warmed up sufficiently in our factory before it is delivered to
you. Therefore, after you receive the sensor unit, please use it as soon as
possible so that unpowered time is minimized.
• The warm-up (powered) time before using the sensor unit is related to the
unpowered time.
Unpowered time
10 minutes or less 10 minutes or more 2 hours or more
1 hour or less 30 minutes or more 2 hours or more
24 hours or less 1 hour or more 4 hours or more
72 hours or less 4 hours or more 24 hours or more
10 days or less 2 days or more 2 days or more
Less than 1 month 7 days or more 7 days or more
Less than 3 months 14 days or more 14 days or more
3 months or more 1 month or more 1 month or more
• To store the sensor unit under unpowered state, it must be stored under
normal temperature/humidity in a clean place away from direct sunlight.
• Do not disassemble the sensor unit because it contains electrolyte. If contact
occurs, rinse the area immediately with a large quantity of water.
• The sensor unit must be stored under normal temperature/humidity in a clean
place away from direct sunlight.
Suggested powered time
SGU-8541 (H2) SGU (Others)
<Detaching and Attaching Pump Unit>
After detaching the sensor unit, push down to open the pump stopper. Pinch the part labeled "PULL" at the
top of the pump unit, and pull it out.
When attaching the pump unit to the main unit, position the
grooves of the pump unit at the rails of the main unit and push the
pump unit.
Do not push the center diaphragm. Push the part labeled "PULL".
And as the connecting point (flow path) of the pump unit is
greased, be sure to check for any dust.
"PULL"
Do not push here.
groove rail
connection
(flow path)
- 13 - GD-70D
1 Overview 1-6. Block diagram
(
)
A
<Wall-mounted Unit>
Mounting hole: A hole for wall-mount.
Terminal plate: A terminal to which a power supply, signal
cables and other external wires are connected
Lock lever: A lever to lock the main unit.
10P
.
1-6. Block diagram
<Electric Diagram>
Sensor unit
Amplifier
Power supply
POWER
Alarm contact controller
Gas alarm contact (ALM1,
LM2)
Fault alarm contact (FAULT)
* Standard setting of contact
activation
• De-energized
(De-energized at a
normal environment)
• ON-ALARM
(Closed contact at an
alarm time)
4-20 mA specification: 4-20 mA
DC Transmission
NT specification: digital data
transmission
Transmitter
Controller
(CPU)
Controller
(CPU)
Display
(POWER)(ALM1)(ALM2)(FAULT)
LCD
Operating unit
(MODE)(▲)(▼)(TEST/SET)
Pump
control
circuit
Pump
Flow
sensor
GD-70D- 14 -
1 Overview 1-6. Block diagram
<Tubing Diagram>
- 15 - GD-70D
2 Installation 2-1. Requirements
2
Installation
2-1. Requirements
Not only the first-time users but also the users who have already used the product must follow the operating
precautions.
Ignoring the precautions may damage the detector, resulting in inaccurate gas detection.
CAUTION
• After you received the detector, start using the detector within the specified operation start limit of the
sensor unit.
<Precautions for installation sites>
CAUTION
• This is a precision instrument. Because the detector may not provide the specified performance in
some places (environments), check the environment in the installation site, and then take appropriate
actions if necessary.
• Because the detector plays an important role for safety and disaster prevention, you must install as
many units of the detector as needed in appropriate points.
• Because points where gases leak and remain easily are different depending on the types of gases
and the working areas, please decide carefully installation sites and the number of units to be
installed.
Do not install the detector in a place with
vibrations or shocks.
The detector consists of sensitive electronic parts. The detector
must be installed in a stable place without vibrations or shocks
and it cannot drop.
Do not install the detector in a place exposed to
water, oil or chemicals.
When you select installation sites, avoid a place where the
detector is exposed to water, oil or chemicals.
Do not install the detector in a place where the
temperature drops below 0ºC or rises over 40ºC.
The operating temperatures of the detector are 0 - 40ºC. The
detector must be installed in a stable place where the operating
temperatures are maintained and do not change suddenly.
GD-70D- 16 -
2 Installation 2-1. Requirements
Do not install the detector in a place exposed to
direct sunlight or sudden changes in the
temperature.
When you select installation sites, avoid a place where it is
exposed to direct sunlight or radiant heat (infrared rays emitted
from a high-temperature object), and where the temperature
changes suddenly. Condensation may be formed inside the
detector, or the detector cannot adjust to sudden changes in the
temperature.
Keep the detector (and its cables) away from noise
source devices.
When you select installation points, avoid a place where
high-frequency/high-voltage exist.
Do not install the detector in a place where maintenance of the detector cannot be
performed or where handling the detector involves dangers.
Regular maintenance of the detector must be performed.
Do not install the detector in a place where the equipment must be stopped when maintenance is performed
in its inside, where parts of the equipment must be removed to perform maintenance, or where the detector
cannot be removed because tubes or racks prevent access to it. Do not install the detector in a place where
maintenance involves dangers, for example, near a high-voltage cable.
Do not install the detector in equipment which is not properly grounded.
Before installing the detector in equipment, the equipment must be grounded properly.
Do not install the detector in a place where other gases exist around it.
The detector must not be installed in a place where other gases exist around it.
<Precautions for system designing>
CAUTION
• An unstable power supply and noise may cause malfunctions or false alarms.
• The descriptions in this section must be reflected on the designing of a system using the detector.
Using a stable power supply
The external output and alarm contact of the detector may be activated when the power is turned on, when
momentary blackout occurs, or when the system is being stabilized. In such cases, use a safety power
supply, or take appropriate actions on the receiving side.
The detector must be provided with the following power supply.
Power supply voltage
Allowed time of
momentary blackout
Others
24 VDC
Up to 10 millisecond (To
recover from the
momentary blackout for 10
millisecond or more,
restart the detector.)
Do not use it with a power
supply of large power load
or high-frequency noise.
±10% (the terminal voltage of the detector)
Example of actions
To ensure continuous operation and activation, install
a protective power supply outside the detector.
Example of actions
Use a line filter to avoid the noise source if necessary.
Heat radiation designing
When it is installed in the closed instrumentation board, attach ventilation fans above and below the board.
- 17 - GD-70D
2 Installation 2-1. Requirements
Introducing protective measures against lightning
If cables are installed outside the factory/plant, or if internal cables are installed in the same duct as the
cables coming from outside the factory/plant, "lightning" will cause problems. Because lightning acts as a
large emission source while cables act as a receiving antenna, devices connected to the cables may be
damaged.
Lightning cannot be prevented. Cables installed in a metal conduit or under the ground cannot be
completely protected from inductive lightning surge caused by lightning. Although complete elimination of
disasters caused by lightning is impossible, the following protective measures can be taken.
Protection against
lightning
Grounding In addition to lightning, there are more sources of surge noise. To protect devices from
* The lightning arrester has a circuit to remove a surge voltage which damages field devices, so that
signals may be attenuated. Before installing a lightning arrester, verify that it works properly.
Take appropriate measures in accordance with the importance of the facilities
and the environment.
• Connect the transmission signal route by using optical fiber.
• Provide protection by a lightning arrester (cable safety retainer).
(Although inductive lightning surge can be transmitted through the cable, it is
prevented by installing a lightning arrester before the field devices and central
processing equipment. For information on how to use a lightning arrester, please
contact the manufacturer.)
these noise sources, the devices must be grounded.
Proper use of alarm contact
The alarm contact of the detector is used to transmit signals to activate an external buzzer, alarm lamp or
rotating lamp. Do not use the detector for controlling purpose (e.g., controlling the shutdown valve).
CAUTION
• The b contact (break contact) under de-energized state may be opened momentarily by a physical
shock, such as external force.
• When the b contact is selected for the alarm contact, take appropriate actions to prepare for a
momentary activation, for example, add signal delay operation (approximately one second) to the
receiving side of the b contact.
The specifications for the alarm contact of the detector are based on the resistant load conditions. If
inductive load is used at the alarm contact, the following errors will occur easily because counter
electromotive force is generated at the contact.
• Deposition, defective insulation or defective contact at the relay contact
• Damage of any electric parts due to high-voltage generated inside the detector
• Abnormal operations by an out-of-control CPU
CAUTION
• In principal, do not activate inductive load at the alarm contact of the detector. (In particular, never use
the inductive load to activate a fluorescent lamp or motor.)
• If inductive load is activated, relay it with an external relay (contact amplification). However, because
the coil of an external relay also involves inductive load, select a relay at a lower voltage (100 VAC or
below), and then protect the contact of the detector with an appropriate surge absorbing part, such as
If load is to be activated, appropriate measures must be taken to stabilize the operation of the detector and
protect the alarm contact referring to the following information.
• Relay it with an external relay at a lower voltage of 100 VAC or below (contact amplification). At the same
time, the surge absorbing part SK1 suitable for the specifications must be attached to the external relay.
• In addition, the surge absorbing part SK2 must be attached to the loaded side of the external relay if
necessary.
• It may be recommended that the surge absorbing part should be attached to the contact for certain load
conditions. It must be attached to an appropriate position by checking how the load is activated.
GD-70D- 18 -
2 Installation 2-2. Installation of detector
GD-70D
SK1
Coil
SK2
Load
Power
supply
Power
supply
Alarm contact
External relay
(low voltage relay)
* SK1, SK2: Surge absorbing parts
2-2. Installation of detector
NOTE
To use the pyrolyzer unit (PLU-70), also refer to the individual operating manual.
CAUTION
• Before installing the detector, remove the protective rubber caps from GAS IN and GAS OUT. If the
detector turned on with the rubber caps remaining while the installation, applied overload may
damage the pump and sensor. Do not forget to remove the caps.
- 19 - GD-70D
2 Installation 2-2. Installation of detector
<Install Dimension and Maintenance Space>
(90)
When installing by using 2 screws When installing by using 3 screws
CAUTION
• It is recommended that installation points should be away from each other for 10 mm or more.
Intervals between installation points must be at least 5 mm.
• When you install more than one unit of the detector in a line, install them in a rack or wall that are not
influenced by vibrations.
When the detectors are installed side-by-side, if the rack or wall in which the detectors are installed do
not have enough strength, vibrations from the pumps inside the detectors cause resonance between
them. In this case, take preventive actions, for example, reinforcing the rack or wall.
Leave the diagonal line area so that the installation space is reserved.
(300)
(30)
(150)
(Unit: mm)
Maintenance spaceMaintenance spaceMaintenance space Maintenance space Maintenance spaceMaintenance space
GD-70D- 20 -
2 Installation 2-2. Installation of detector
CAUTION
• The detector must be installed in the right direction to ensure its performance. Install the detector as
shown on the following example, and adjust its position so that it is placed horizontally. (* The
direction of the detector must also be kept during transportation, including when it is removed
temporarily or relocated.)
<Right Installation Direction>
Wall
Ceiling
Wall
Floor
<Installation of Wall-mounted Unit>
Attach the wall-mounted unit in the installation surface using two or three M5 screws.
Recommended mounting screw (M5)
Length of 8 mm or more
Flat washer of φ10 mm or less (small round)
After the wall-mounted unit is attached to the wall, install the main unit in the wall-mounted unit.
CAUTION
• Install the unit so that its surface is in intimate contact with the wall-mounted unit. A space between
the unit and the wall-mounted unit invites unnecessary vibrations and noises.
- 21 - GD-70D
2 Installation 2-2. Installation of detector
k
<Detaching and Attaching Main Unit>
Attaching Main Unit
At the position of 10 mm above the wall hanger unit, press the
main unit onto the wall hanger unit. Be sure to fit both side hooks
of the wall hanger unit in the grooves of the main unit.
Then press down the main unit to fix it. The lock at the bottom of
the main unit clicks to fix it properly. Make sure that the top
center of the wall hanger unit is above the main unit as viewed
from front.
Detaching Main Unit
While pushing the sky blue lever toward the wall hanger unit,
hold up the main unit.
If you cannot move the main unit, insert a larger flathead
screwdriver while pushing the lever, and you can easily detach it
as shown below. Do not rotate or move up and down the flathead
screwdriver. Simply insert it into the wall hanger unit.
CAUTION
• Be careful not to drop the main unit when detaching it.
Also, check the secure installation of the main unit after
attaching it to the wall hanger unit. If the main unit is not
securely installed, it might fall, causing an unexpected
injury or a damage of the unit.
• Turn the power off before detaching or attaching the main
unit.
When pressing the main unit,
align the side mark with the hoo
of the wall hanger unit for
smooth installation.
Make sure that the top of the
wall hanger unit is above the
main unit.
The lock clicks.
10mm
GD-70D- 22 -
2 Installation 2-3. How to wire
2-3. How to wire
NOTE
To use the pyrolyzer unit (PLU-70), also refer to the individual operating manual.
CAUTION
• Be careful not to damage the internal electronic circuit when wiring. In addition, be careful not to apply
stresses on the detector when (overweight) cables are installed.
• The power cables and signal cables must not be installed together with the motor power cables, etc.
When these cables must be installed together for unavoidable reasons, put the power cables and
signal cables in a metal conduit. The conduit must be connected to a grounding circuit.
• When stranded wires are used, prevent wires from contacting each other.
• Use the dedicated handling lever to wire.
<Recommended Cable>
For 3-wire type 4 - 20 mA
(common power supply)
For 4-wire type 4 - 20 mA Power: Cable of CVV, etc. (1.25sq) - 2-core
For 2-wire type DC power-line
communication system (NT)
For contact Cable of CVV, etc. (1.25sq) - max. 6-core
Shielded cable of CVVS, etc. (1.25sq) - 3-core
Signal: Shielded cable of CVVS, etc. (1.25sq) - 2-core
Shielded twisted-pair cable of KPEV-S, etc. (1.25sq) - 1P
For the 3-wire type (4 - 20 mA), the terminal 2 is used for common, and the terminals 2(-) and 3(+) are used
to output 4 - 20 mA.
For the NT specification, the terminals 3 and 4 are not used.
<Specifications of Terminal Plate>
Specifications of terminal plate
• Rated voltage: 250 VDC
• Rated current: 16 A
Top handling slot
Cage
clamp/spring
Driver slot
Conductive
part
Wiring hole
- 23 - GD-70D
2 Installation 2-3. How to wire
Connection conditions
• Cable: 0.08 - 2.5mm2
• Bare wire length: 8 - 9 mm
• Connecting tool: Dedicated handling lever (accessory)
CAUTION
• The specified bare wire length must be observed when the wire insulation is peeled off.
• Improper clamping of the wire due to a shorter bare wire length may cause defective electrification or
heating.
• Damaging the wire insulation due to a shorter bare wire length may cause defective electrification or
heating.
• Exposing the wire due to a longer bare wire length may cause defective insulation or a short circuit.
• Be careful not to break up the wire. If the wire is broken up when inserted to the terminal, this may
cause defective insulation or heating.
8 – 9 mm
Dedicated
handling lever
Compatible bar terminal
For a bar terminal, the following items are available.
• Bar terminal (ferrule): Model 216 Series (manufactured by WAGO)
• Crimping tool: Model VarioCrimp 4 (206-204) (manufactured by WAGO)
CAUTION
• A bar terminal of the specified model must be used. Using other bar terminals invalidates the
guarantee of the performance.
GD-70D- 24 -
2 Installation 2-3. How to wire
g
<How to Connect to Terminal Plate>
When cables are connected to the connectors, use the dedicated
lever or a flathead screwdriver to do it as shown below.
CAUTION
• The right tools must be used.
• In principal, one wire can be connected to one wiring
hole.
• When the wire is inserted into the driver slot by mistake, it
does not contact the conductive part. This may cause
defective electrification or heating.
• When the wire is inserted under the spring by mistake, it
does not contact the conductive part. This may cause
defective electrification or heatin
.
NOTE
How to Use the Dedicated Handling Lever
Push the lever with your
finger to lower the spring
in its inside.
While holding down the
lever, insert the wire into the
(round) wiring hole until it
reaches the deepest point.
Once the lever is released,
the wire is secured.
To check whether the wire is connected securely, pull the wire
gently. (Do not pull the wire strongly.)
<How to Clamp Cables>
Insert the supplied fastening-band to the hole on the
wall-mounted unit as shown in the figure. Position the band so
that its coarse side faces the wire. After inserting the
fastening-band, fix the wire together.
NOTE
When an optional cable ground is used, it must
be attached to the external wiring hole.
Push the dedicated
handling lever to
open the terminals,
into which you can
insert wires (one wire
to one terminal).
Position the
band so that
its coarse
side faces the
wire.
- 25 - GD-70D
Loading...
+ 56 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.