RKI Instruments FP-40 User Manual

Instruction Manual
for
FORMALDEHYDE GAS DETECTOR
MODEL FP-30/FP-40
This instrument is sold and serviced by RKI Instruments, Union City, CA For questions, parts, or service please contact RKI at (800) 754-5165.
Please read and understand this instruction manual before operating this instrument.
Part Number: 71-0085RK
Revision: P3
Released: 8/27/03
1
Table of Contents
1.1 Start-up Operation 2
1.2 Installation of the Detection TAB 3
1.3 Detection Method 4
1.4 High Density Gas Detection 4
1.5 Memory of Stored Detection Results 4
1.6 Battery Voltage Check 7
1.7 Preliminary Sample-Draw 7
1.8 Purge Cycle 8
1.9 Completion of Detection 8
2.1 Storage of Detection TAB 9
2.2 Handling of Detection TAB 9
4.1 Battery Replacement 12
4.2 Daily Check / Regular Check 12
4.3 Replacement of parts 12
4.4 Storage and Treatment After Long Term Storage 12
13
6
6.1 Specifications 14
6.2 Accessories 14
6.3 Spare Parts List 14
7
15
2
1 OPERATION
1.1 Start-Up Operation
Press ON/OFF switch for 2 seconds to turn instrument on and start warm-up and self-diagnosis .
The program and instrument display function one of the following two ways:
If the detection TAB is not installed before start-up:
Press ON/OFF switch for 2 seconds
BATT E--------l------F Shows the battery capacity
Approx 5 sec
STAND BY OK Standby Condition
Approx 5 sec
SET TAB
If the detection TAB is installed before start-up
Waiting for TAB to be installed and selected
Press ON/OFF switch for 2 seconds
BATT----------l----F Shows the battery capacity
Approx 5 sec
SELF-DIAGNOSIS Self-diagnosis : Sensor check
Approx 5 sec
STAND BY OK Standby Condition
Approx 5 sec
SELECT TAB No
Select the detection TAB for desired range using up/down button.
3
1.2 Installation of Detection Tab
L
D
I
C
I
The detection TAB comes packaged in a sealed foil wrapper. Remove it from the wrapper. The detection surface of the TAB is protected by a plastic cover that fits over the TAB. Remove this cover. The detection TAB must only be handled by its edges or backside. Do not touch the indented center section of the TAB.
1. Open the detection TAB cover.
ift up this way.
etection TAB cover
nstrument (Side view)
2. Place the TAB into the slot under the cover, indented face down, and then close the cover slowly.
lose the cover.
Detection TAB
nstrument (Side view)
3. Press the center of detection TAB cover with finger to assure proper TAB seating.
When closing the detection TAB cover, close it slowly to avoid pinching finger.
Do not allow the lid to slam shut since that may cause damage to the instrument.
When installing the detection TAB, do not allow water drops or dust to get in the TAB port area.
4
1.3 Detection Method
Gas detection
Detection waiting condition
The instrument waits in this mode until the detection TAB is installed.
Installation of the detection TAB
Install the detection TAB.
Selection of the detection range
Select the detection range by pressing up/down keys.
Detection starts
Press START minutes or 30 minutes (as selected).
Display of detection result
After the detection cycle is completed, the average gas reading during the detection cycle is shown as the detection result.
Display of memory number
The number recorded for this reading in the memory is displayed.
Completion of detection
The gas reading and message will alternately be shown until the used detection tab is removed. Buzzer sounds every one second until used tab is removed.
Display example
switch to start the pump and start the detection cycle. Cycle is either 15
SET 1 Detection waiting condition
2 Install the detection TAB
SELECT TAB No
008 : HCHO START
008 : HCHO PUSH
008 : HCHO 1800 4 Start
Counting down the detection time. (Sec.)
008 : HCHO 0.100 5 Display of detection result
MO5 REGISTERED 6 Memory number display
REMOVE TAB
beep beep 7 Completion of detection
2 Select the detection TAB (time/range
selection) using the up/down keys.
008 : HCHO 0.100
Loading...
+ 11 hidden pages