Read and understand this instruction manual before
operating instrument. Improper use of the gas
monitor could result in bodily harm or death.
Periodic calibration and maintenance of the gas
monitor is essential for proper operation and correct
readings. Please calibrate and maintain this
instrument regularly! Frequency of calibration
depends upon the type of use y ou have and th e sensor
types. Typical calibration frequencie s for most
applications are between 1 and 3 months, but can be
required more often or less often based on your usage.
WARNING:Understand manual before operating. Substitution of components may
impair intrinsic safety. To prevent ignition of a hazardous atmosphere,
batteries must only be changed or charged in an area known to be
nonhazardous. Not tested in oxygen enriched atmospheres (above
21%).
Ta ble of ContentsEAGLE 2 Operator’s Manual
NOTE:RKI Instruments, Inc. recommends that you refer to ISA-RP12.13, Part II-1987
or an equivalent international recommended practice for guidance in the use of
combustible gas detection instruments.
EAGLE 2 Operator’s ManualTable of Contents
Chapter 1: Introduction
Overview
This chapter briefly describes the EAGLE 2 gas monitor. This chapter also describes the
EAGLE 2 Operator’s Manual (this document). Table 1 at the end of this chapter lists the
specifications for the EAGLE 2.
About the EAGLE 2
Using an advanced detection system consisting of up to six gas sensors, the EAGLE 2
sample draw gas monitor is capable of detecting the presence of combustible gas, oxygen
(O2), carbon monoxide (CO), hydrogen sulfide (H2S), and various other toxic gases
simultaneously. The EAGLE 2’s rugged, reliable, and easy-to-use design makes it ideally
suited for a wide range of applications, including sewage treatment plants, utility
manholes, tunnels, hazardous waste sites, power stations, petrochemical refineries, mines,
paper mills, drilling rigs, and fire fighting stations. The EAGLE 2 offers a full range of
features including:
•Simultaneous monitoring of one to six gases. The standard configuration includes four
sensors for combustible gas (%LEL, ppm, and %volume), oxygen, carbon monoxide
(CO), and hydrogen sulfide (H2S).
•Choice of three operating modes:
•Normal Mode for typical confined space or area monitoring. Normal Mode is the
standard factory setting.
•Bar Hole Mode for checking of bar holes when searching for underground gas
leaks
•Leak Check Mode for locating leaks in valves and piping
•Sample-drawing pump with up to 125 foot range
•Liquid crystal display (LCD) for complete and understandable information at a glance
•Ultrabright alarm LEDs
•Distinctive audible alarm for dangerous gas conditions or unit malfunction
•Microprocessor control for reliability, ease of use, and advanced capabilities
•Data logging functions (when used in Normal Mode)
•Alarm trend data (when used in Normal Mode)
•STEL and TWA (when used in Normal Mode) and over range alarms
•Peak readings (when used in Normal Mode)
•Built-in time function
•Lunch break feature
•RF shielded high impact plastic case
•CSA classified for Class I, Division I, Groups A, B, C, and D hazardous atmospheres
10 • OverviewEAGLE 2 Operator’s Manual
WARNING:The Model EAGLE 2 detects oxygen deficiency, elevated levels of
oxygen, combustible gases, carbon monoxide, and hydrogen sulfide, all
of which can be dangerous or life threatening. When u sing the EAGLE
2, you must follow the instructions and warnings in this manual to
assure proper and safe operation of the unit and to mini mize t he risk of
personal injury. Be sure to maintain and periodically calibrate the
EAGLE 2 as described in this manual.
NOTE:ONLY THE COMBUSTIBLE GAS DETECTION PORTION OF THIS
* When calibrating the EAGLE 2 with the Auto Calibration or the Single Calibration method, the calibration gas
value must be equal to or higher than the alarm 1 setting. See “Updating the Alarm Point Settings” on page 106
for instructions to change the alarm points if necessary for the desired calibration gas value.
Table 2: EAGLE 2 Specifications
Sampling
Sample Draw
Method
Response
T90 Within 30 Seconds
Time
Display Graphics LCD Display
Operating
-20°C to 50
°
C/Below 85% RH (Without Condensation)
Temperature
& Humidity
Indication
Accuracy
Combustible Gas (LEL), Catalytic Type Sensor
• -10°C to 40
• -20°C to 50
°
C: 5% of full scale
°
C: 6% of full scale
Combustible Gas (ppm), Catalytic Type Sensor
• ± 25 ppm or ± 5% of reading (whichever is greater)
Oxygen
•± 0.5% O
2
Hydrogen Sulfide
• ± 5% of reading or ± 2 ppm H
S (whichever is greater)
2
Carbon Monoxide
• ± 5% of reading or ± 5 ppm CO (whichever is greater)
12 • SpecificationsEAGLE 2 Operator’s Manual
Safety/
Regulatory
186718
CSA classified as Intrinsically Safe. Exia. Class I, Groups A, B, C, & D.
Temperature Code T3C.
Power
Supply
Continuous
Operating
Hours
@ 25 °C
CaseHigh-impact Plastic, RF Shielded, Dust and Weather Proof
• Hoses of Various Lengths, See “General Parts List” on page 79.
• Dilution Fitting (1:1 and 3:1)
• Various Probes, See “General Parts List” on page 79
• Product CD, includes Data Logger Management Program and Maintenance
Data Loader Program
• IrDA/USB Cable for connecting to a computer when using the Data Logger
Management Program and Maintenance Data Loader Program (not needed
if computer has an infrared port)
Dimensions
and Weight
Approximately 171(H) x 65(W) x 39(D) mm (5.6”H x 2.5”W x 1.5”D)
About this Manual
The EAGLE 2 Operator’s Manual uses the following conventions for notes, cautions, and
warnings.
NOTE:Describes additional or critical information.
CAUTION:Describes potential damage to equipment.
WARNING:Describes potential danger that can result in injury or death.
Approximately 310 g (11 oz.)
EAGLE 2 Operator’s ManualAbout this Manual • 13
Chapter 2: Description
LCD Display
Control Buttons
Alarm LEDs
RANGE
SHIFT
DISPLAY
ADJUST
NO
AIR
YES
PO WER
ENTE R
RESET
Inlet Fitting
Overview
This chapter describes the EAGLE 2 instrument and accessories.
Instrument Description
The EAGLE 2 includes the case, sensors, LCD, control buttons, printed circuit boards,
alarm LEDs, infrared communication port, buzzer, battery case and batteries, and flow
system.
The EAGLE 2’s sturdy, high-impact plastic case is radio frequency (RF) resistant and is
suitable for use in many environmental conditions, indoors and out. The case is dust proof
and water resistant. It’s two main components, the top case and bottom case, are held
together with three screws located on the bottom case. The interface between the top case
and bottom case is gasketed. A sturdy, well balanced handle on the top case allows you to
hold the instrument comfortably. A clear plastic window is located on the top case to the
front of the handle for viewing the LCD.
A removable battery case is located at the rear of the bottom case. A thumbscrew secures
the battery case to the bottom case. The interface between the battery case and the bottom
case is gasketed. Six raised feet on the bottom of the case, four on the bottom case and two
on the battery compartment, raise the EAGLE 2 slightly from the surface on which it rests.
Sensors
The EAGLE 2 uses up to six sensors to monitor combustible gas, oxygen (O2), carbon
monoxide (CO), hydrogen sulfide (H2S), and various other toxic gases simultaneously.
The sensors are located inside the EAGLE 2 bottom case and are installed in the flow
chamber. The sensors described below are the four standard sensors. See “Appendix D:
PID Sensors” for a description of the PID sensors, “Appendix E: ESM-01 Toxic Sensors”
for a description of the ESM-01 toxic sensors, “Appendix F: TC Sensors” for a description
of the TC sensors, “Appendix G: Infrared Carbon Dioxide Sensors” for a description of
the IR CO2 sensors, “Appendix H: Infrared Methane Sensor” for a description of the IR
methane sensors, and “Appendix I: Infrared Hydrocarbon Sensor” for a description of the
IR hydrocarbon sensor. The standard sensors use different detection principles as
described below.
Catalytic Combustible Gas Sensor (LEL Sensor)
The catalytic combustible gas (LEL) sensor detects combustible gas in the %LEL range. It
uses a catalytic element for detection. The reaction of gas with oxygen on the catalyst
causes a change in the resistance of the element which changes the current flowing
through it. The current is amplified by the EAGLE 2’s circuitry, converted to a
measurement of combustible gas concentration, and displayed on the LCD.
The LEL sensor housing includes a sintered metal flame arrestor on one end that allows
gas to diffuse into the sensor. On the other end, five pins extend from the sensor. The
sensor cable connects to these pins on one end and terminates in a four-position connector
on the other end which plugs into the HC socket on the main PCB (see “Main PCB” on
page 19).
Oxygen Sensor
The O2 sensor is a galvanic type of sensor. A membrane behind the openings on the sensor
face allows gas to diffuse into the sensor at a rate proportional to the partial pressure of
oxygen. The oxygen reacts in the sensor and produces a voltage proportional to the
concentration of oxygen. The voltage is measured by the EAGLE 2’s circuitry, converted
to a measurement of gas concentration, and displayed on the LCD.
The sensor includes a short cable that terminates in a round 7-position connector. It mates
with the OXY pins on the main PCB (see “Main PCB” on page 19).
The CO and H2S sensors are electrochemical sensors that consist of three precious metal
electrodes in a dilute acid electrolyte. A gas permeable membrane covers the sensor face
and allows gas to diffuse into the electrolyte. The gas reacts in the sensor and produces a
current proportional to the concentration of the target gas. The current is amplified by the
EAGLE 2’s circuitry, converted to a measurement of gas concentration, and displayed on
the LCD.
The CO and H
S sensors are physically very similar. Except for their markings and wire
2
colors, they look almost identical. A three-position connector at the end of a 2-wire cable
from each sensor plugs into a socket on the main PCB. The sockets on the main PCB for
the CO and H
S sensors are labeled CO and H2S. Normally, the CO connector plugs into
2
the CO socket and the H2S plugs into the H2S socket. However, because of the way that
the main PCB circuitry is arranged, if the CO sensor is plugged into the H2S socket and
the H2S sensor is plugged into the CO socket, the sensors will still operate properly and
the CO and H2S readings will still appear on the channels that are programmed for those
gases.
LCD
A digital LCD (liquid crystal display) is visible through a clear plastic window in the top
case. The LCD simultaneously shows the gas reading for all installed sensors. The LCD
also shows information for each of the EAGLE 2’s operating modes.
Control Buttons
Four control buttons are located below the LCD. They are, from left to right, POWER
ENTER RESET, DISPLAY ADJUST NO, AIR
Table 3: EAGLE 2 Control Button Functions
ButtonFunction(s)
POWER ENTER
RESET
• turns the EAGLE 2 on and off
• silences and resets audible alarm if Alarm Latching is set to
Latching and Alarm Silence is set to ON
• enters instructions, values, and settings into the EAGLE 2’s
microprocessor
▲YES,and RANGE ▼ SHIFT.
DISPLAY ADJUST NO• activates Display Mode
• silences and resets audible alarm if Alarm Latching is set to
Latching and Alarm Silence is set to ON
• enters instructions into the EAGLE 2’s microprocessor
AIR ▲ YES• activates the demand zero function (adjusts the EAGLE 2’s
fresh air reading)
• silences and resets audible alarm if Alarm Latching is set to
Latching and Alarm Silence is set to ON
• enters instructions into the EAGLE 2’s microprocessor
• moves the cursor on the LCD up the screen
• increases the value of a parameter available for adjustment
RANGE ▼ SHIFT• changes the detection units of the combustible gas channel
(when Catalytic Units is set to CHANGE OK in Setup Mode)
• silences and resets audible alarm if Alarm Latching is set to
Latching and Alarm Silence is set to ON
• enters instructions into the EAGLE 2’s microprocessor
• moves the cursor on the LCD down the screen
• decreases the value of a parameter available for adjustment
• scrolls through parameter options
Printed Circuit Boards (PCBs)
Figure 4: EAGLE 2 PCBs
The EAGLE 2’s PCBs analyze, record, control, store, and display the information
collected. The main PCB and I.S. barrier PCB are located in the bottom case. The display
PCB is located in the top case. The display PCB and I.S. barrier PCB are not user
serviceable and are not involved in any user performed maintenance. The main PCB is not
user serviceable, but it is involved in the replacement of sensors, so it is described below.
The main PCB is located on the right side of the bottom case. It slides into guiding
grooves at the bottom, front, and rear of the bottom case. It is held in place by the top case.
Connectors for the sensors, differential pressure switch, pump, display PCB, optional float
probe, and I.S. barrier PCB are located on the main PCB. In addition, an IR transmitter/
receiver is located at the front of the PCB behind the IR port on the front of the bottom
case.
Alarm LEDs
Two sets of red alarm LEDs (light emitting diodes) are visible through two raised, frosted
plastic lenses in the top case. Each set has two LEDs. They are above the LCD, one above
the left corner and one above the right corner. The alarm LEDs alert you to gas, low
battery, and failure alarms.
Infrared Communications Port
An infrared (IR) communications port is located on the left front of the bottom case when
the instrument is viewed from the front. The data transmitted through the port is in
standard IrDA protocol. A computer’s infrared port or an IrDA/USB cable connected to a
USB port can be used to download data saved by the EAGLE 2 to a computer using the
Eagle 2 Data Logger Management Program. See the Data Logger Management Program
operator’s manual for data logging and downloading instructions.
A solid-state electronic buzzer is located on the front of the top case. It is a panel mounting
type of buzzer and is water resistant and sealed to the inside of the top case with an O-ring.
The buzzer sounds for gas alarms, malfunctions, low battery voltage, and as an indicator
during use of the EAGLE 2’s many display and adjustment options.
Battery Case & Batteries
Four C-size alkaline batteries (standard) or optional rechargeable
C-size Ni-MH batteries power the EAGLE 2. They are installed in the battery case which
is located at the rear of the bottom case. The battery case is secured to the bottom case
with a thumbscrew.
Instrument run time is dependent upon battery type. At 25°C, alkaline batteries power the
EAGLE 2 for 16 hours of non-alarm operation. Ni-MH batteries will power the EAGLE 2
for 18 hours of non-alarm operation. The current battery voltage is viewable in Display
Mode (see “Display Mode” on page 41).
When the EAGLE 2 detects low battery voltage, a low battery warning is activated. When
battery voltage is too low for operation, the EAGLE 2 sounds a dead battery alarm.
The alkaline or Ni-MH batteries can be accessed for replacement by unscrewing the
thumbscrew that secures the battery case to the bottom case and pulling the battery case
away from the bottom case. The Ni-MH batteries can be recharged by using the EAGLE 2
charger (see “Replacing or Recharging the Batteries” on page 69).
NOTE:Use of batteries or battery chargers not specified by RKI Instruments, Inc. will
void the CSA classification and may void the warranty.
WARNING:To prevent ignition of a hazardous atmosphere, batteries must only be
changed or charged in an area known to be nonhazardous.
Flow System
The EAGLE 2 flow system consists of the inlet fitting, hydrophobic filter, pump, internal
tubing, differential pressure (DP) switch, sensor chamber, charcoal filter, and exhaust
fitting.
Inlet Fitting
The inlet fitting is on the right front (when viewed from the front) of the bottom case. It is
a nickel plated brass quick connect fitting. It mates with either the sample hose or with the
hydrophobic probe.
Hydrophobic Filter
The hydrophobic filter is located in the bottom case above the sensors. Normally the
hydrophobic probe accessory (see “Hose and Probe” on page 21) will prevent water and
particulate contamination from entering the flow system, but if the probe is not used, the
hydrophobic filter will stop water and particulates from penetrating further into the flow
system. If it becomes dirty or water logged, replace it (see “Replacing the Hydrophobic
Filter” on page 73).
Pump
A diaphragm pump inside the rear of the bottom case draws the sample to the sensors. It
can draw sample from as far as 125 feet from the EAGLE 2.
CAUTION:Sample hose lengths of more than 125 feet are not recommended for the
EAGLE 2 because of flow rate reduction and increased response time.
Consult RKI Instruments, Inc. for sample hose lengths longer than 125 feet.
Internal Tubing
The flow system includes polyurethane tubing to route the sample between the various
components of the flow system. The internal sample tubing is not user serviceable.
Differential Pressure (DP) Switch
The DP switch is inside the front of the bottom case. It senses the EAGLE 2’s flowrate by
monitoring the pressure drop between points in the flow system. When the flowrate
becomes too low for safe operation of the EAGLE 2, a set of contacts inside it open and
the EAGLE 2 indicates a low flow alarm.
Sensor Chamber
A PVC block in the bottom case is configured to accept the four gas sensors. It routes the
sample to each sensor. The LEL sensor and the oxygen sensor are retained in the sensor
chamber by brackets. The CO and H2S sensors are each pushed past two sealing O-rings
into the chamber and are retained by the O-ring compression force.
Charcoal Filter
The charcoal filter is located in the front of the flow chamber next to the CO sensor. It
contains activated charcoal. The CO sensor will respond if exposed to H2S and certain
hydrocarbon gases. The charcoal filter scrubs these gases out of the sample to avoid false
CO readings. If false or elevated CO readings are noticed, especially in the presence of
S, change the charcoal filter. The charcoal inside the filter cannot be replaced; the entire
H
2
filter must be replaced.
Exhaust Fitting
The exhaust fitting is located below the inlet fitting. It routes the gas sample out of the
EAGLE 2. It includes a female 10-32 thread that can be used for the installation of a hose
barb or other type of fitting that has a male 10-32 thread so that the exhaust can be routed
to a particular location with flexible tubing if desired.
Standard Accessories
Standard accessories include the shoulder strap, the sample hose, and the hydrophobic
probe.
Shoulder Strap
A comfortable elastic shoulder strap clips to the EAGLE 2 at the front and rear of the top
case. It clips to stainless steel rings that are installed in features on the top case. It can be
removed from the EAGLE 2 by opening the clip at each end of the strap and removing it
from the strap ring at the front and rear of the top case.
Hose and Probe
A 5 foot polyurethane sample hose and a 10 inch hydrophobic probe are included as
standard. The hose has a male quick connect fitting on one end and a female quick connect
fitting on the other end. The probe has a male quick connect fitting. Normally, the male
end of sample hose is installed in the EAGLE 2 inlet fitting and the probe is installed in
the female end of the hose. However, if the sample hose is not needed for monitoring a
particular area, the probe may be installed directly to the inlet fitting. Sample hose lengths
are available from 5 feet (standard length) to 125 feet (see “General Parts List” on
Sample Hose
Probe
page 79). A Teflon lined hose is provided with all units that contain a PID sensor. This
hose must be used when operating a PID EAGLE 2 (see “Appendix D: PID Sensors” on
page 125).
CAUTION:Sample hose lengths of more than 125 feet are not recommended for the
EAGLE 2 because of flow rate reduction and increased response time.
Consult RKI Instruments, Inc. for hose lengths longer than 125 feet.
The probe includes a replaceable particle filter and hydrophobic filter disk that prevent
particulates and water from entering the EAGLE 2’s flow system. See “Replacing the
Hydrophobic Probe’s Particle Filter and Hydrophobic Filter Disk” on page 72 for
instructions to replace the particle filter and hydrophobic filter disk.
NOTE:When using the probe with a PID EAGLE 2, be sure that the particle filter is not
installed.
Figure 6: Sample Hose and Hydrophobic Probe
22 • Standard AccessoriesEAGLE 2 Operator’s Manual
Optional Accessories
AC Adapter
Charging Module
Several optional accessories are available for the EAGLE 2. They include rechargeable
Ni-MH batteries, battery chargers, various special probes, and dilution fittings. The most
commonly used optional accessories are described below. Detailed instructions regarding
the use of these and other available accessories are included in other parts of this manual.
Data logging accessories are briefly described in “Data Logging” on page 52.
Rechargeable Ni-MH Batteries
Rechargeable Ni-MH batteries are available for the EAGLE 2. A fully charged set of NiMH batteries will power the EAGLE 2 for 18 hours. The batteries will last for a minimum
of 500 charge cycles. See “General Parts List” on page 79 for ordering information.
Battery Chargers
Three battery chargers are available for the EAGLE 2 to charge the optional Ni-MH
batteries, the standard AC charger, a DC charger with a vehicle plug adapter, and an AC/
DC charger with a vehicle plug adapter.
AC Charger
The standard AC charger consists of the charging module, which includes all of the
charging circuitry, and an AC adapter. The charging module includes a five foot cable with
a connector on the end that mates with the EAGLE 2 charging socket. The AC adapter
plugs into a 115 VAC wall outlet and connects to the charging module with a jack on the
end of a five foot DC output cable. The AC adapter will also work for 100 VAC or 220
VAC if an appropriate plug adapter is provided. The AC charging station is shown below
in Figure 7.
An optional DC powered charger is available with a vehicle plug 12 VDC adapter. It uses
the same charging module as the standard AC charger.
Figure 8: EAGLE 2 DC Charger
AC/DC Charger
A charger is also available that includes both the AC adapter and the 12 VDC vehicle plug
adapter. The charging module is the same as the one used for the AC charger and the DC
charger.
Optional Probes
Various optional probes designed for specific applications are available for the EAGLE 2.
They include the following:
•30 inch aluminum probe
Figure 9: 30 Inch Aluminum Probe
This probe is designed for applications where it is necessary to put the probe tip in
areas that are out of reach with the standard probe. A small breather hole near the end
of the probe tube prevents interruption of sampling and a low flow alarm if the probe
tip is blocked.
This probe is physically the same as the 30 inch aluminum probe and is intended for
applications where a high level of corrosion resistance is required in the long probe
tube.
•4 foot stainless steel probe
Figure 11: 4 Foot Stainless Steel Probe
This probe is designed for areas where it is necessary to put the probe tip in areas that
are out of reach for even the 30 inch probes. A stainless steel probe tube is used
because the length of the probe tube requires a high degree of rigidity. Stainless steel is
more rigid than other normally used materials. A small breather hole near the end of
the probe tube prevents interruption of sampling and a low flow alarm if the probe tip
is blocked.
•Barhole probe
Figure 12: Barhole Probe
This probe is designed specifically for barhole testing. See “Appendix K: Using the
EAGLE 2 in Bar Hole Mode” for an in-depth discussion of using the EAGLE 2 in Bar
Hole Mode.
•10 inch probe with dust filter
Figure 13: 10 Inch Probe With Dust Filter
This probe is designed for use where drawing water or moisture into the EAGLE 2 is
not a concern. Instead of a hydrophobic filter, a cotton dust filter is used.
•32 inch telescoping probe with dust filter
Figure 14: 32 Inch Telescoping Probe with Dust Filter
This probe is designed for use where it is necessary to put the probe tip in areas not
accessible with the 10 inch probe with dust filter and applications where the probe tube
must be collapsible for storage.
Figure 15: 7 Foot Telescoping Probe with Dust Filter
This probe is designed for use where it is necessary to put the probe tip in areas not
accessible with the 32 inch telescoping probe with dust filter and applications where
the probe tube must be collapsible for storage.
See “General Parts List” on page 79 for probe ordering information.
External Dilution Fittings
Two external dilution fittings are available for the EAGLE 2, a 1:1 dilution fitting and a
3:1 dilution fitting. They are designed to mate with the inlet fitting and accept a sample
hose or probe. The fittings are made with brass and nickel plated brass and are appropriate
for use with the four standard gases. The 1:1 fitting is normally used when it is necessary
to introduce air into a sample that has no oxygen or a very low level of oxygen, such as a
nitrogen purged sample. Both the 1:1 and 3:1 fittings can also be used when one of the
target gas levels in the sample area will likely be present in a concentration above the
detection range for that gas. Since the fittings partially consist of unplated brass, they are
not appropriate for detection of elevated levels of H2S or of gases that are easily absorbed
This chapter explains how to use the EAGLE 2 to perform confined space entry
monitoring or general area monitoring in Normal Mode. There are three operational
modes in Normal Mode: Measuring Mode, Display Mode, and Calibration Mode. While
in Normal Mode, the unit is normally operating in Measuring Mode. Display Mode and
Calibration Mode are accessible from Measuring Mode. Display Mode is described in this
chapter. Calibration Mode is described in “Chapter 4: Calibration Mode” on page 54.
Special versions of the EAGLE 2 can also operate in Leak Check Mode and Bar Hole
Mode. See “Appendix K: Using the EAGLE 2 in Bar Hole Mode” and “Appendix L:
Using the EAGLE 2 in Leak Check Mode” for operating instructions for Bar Hole and
Leak Check Mode, respectively.
Start Up
This section explains how to start up the EAGLE 2, get it ready for operation, and turn it
off.
NOTE:The screens illustrated in this section are for a standard 4-gas unit. The screens
displayed by your EAGLE 2 may be slightly different.
Turning On the EAGLE 2
To illustrate certain functions, the following description of the EAGLE 2 start up sequence
assumes that the following menu items in Setup Mode are turned on: LUNCH BREAK,
CAL REMINDER, and USER/STATION ID. If any of these items are turned off, then the
corresponding screens will not appear.
The EAGLE 2 may be used with a sample hose or with the probe installed directly to the
inlet fitting. Determine which configuration works best for your application.
1.Connect the sample hose or probe to the EAGLE 2’s quick connect inlet fitting.
2.If using a sample hose, connect the probe to the sample hose’s quick connect fitting.
3.Press and briefly hold down the POWER ENTER RESET button. Release the button
when you hear a beep.
4.The LCD will show the following screen for about ten seconds.
EAGLE 2 Operator’s ManualOverview • 27
5.The Battery Voltage Screen appears for a few seconds.
BATTERY MIN:
4.3 VOLTS
BATTERY NOW:
5.2 VOLTS
ACTIVE GASES
CH4 OXY
H2S CO
LUNCH BREAK MODE ON
RESUME
PEAK AND TWA
MEASUREMENTS? 2
6.The Active Gases Screen appears for a few seconds indicating which channels are
active and their target gas.
7.If LUNCH BREAK is turned on (see “Updating the Lunch Break Setting” on
page 118), the Resume Measurement Screen appears. The unit counts down from 5
seconds in the lower right corner of the LCD to the right of “MEASUREMENTS”.
•To continue accumulating peak and time-weighted average (TWA) readings from
the last time the EAGLE 2 was used, press and release the AIR ▲ YES button
before the countdown reaches 0 or allow the countdown to reach 0. If you do not
press the AIR ▲ YES button within the 5 second countdown, the EAGLE 2
automatically resumes accumulating the peak and TWA readings.The EAGLE 2
will also continue to keep track of operating time including the operating time
from the last time the EAGLE 2 was used. See “Time in Operation Screen” on
page 50 for more information about how the EAGLE 2 tracks the operating time.
The short-term exposure limit (STEL) reading is reset each time the EAGLE 2 is
turned on.
•To reset the accumulation of these measurements, press and release the
DISPLAY ADJUST NO button before the countdown reaches 0.
28 • Start U pEAGLE 2 Operator’s Manual
8.The gas alarm setpoints are displayed by three screens in sequence: the Low Alarm
A CH4 10 %LEL
L L OXY 19.5 vol%
O A H2S 10.0 ppm
WR CO 25 ppm
M
S
CH4 10 vol%
A CH4 50 %LEL
HL OXY 23.5 vol%
I A H2S 30.0 ppm
GR CO 50 ppm
HM
S
CH4 50 vol%
ALARMS STEL & TWA
H2S(ppm) 15.0 10.0
CO (ppm) 200 25
CALIBRATION DATE
IS PAST DUE
PERFORM
CALIBRATION?
Screen, High Alarm Screen, and STEL/TWA Alarm Screen. Each screen remains on
the LCD for three seconds.
9.After the alarm screens, if CAL REMINDER is turned on, the screen that appears
next depends on how CAL PAST DUE ACT is set in the Setup Mode Menu (see
“Updating the Calibration Past Due Action Setting” on page 113).
•If the unit is due for calibration and CAL PAST DUE ACT is set to CONFIRM
TO CAL, then the following screen displays and the buzzer sounds in a double
pulsing pattern.
To perform a calibration, press and release the AIR ▲ YES button. The EAGLE
2 will enter Calibration Mode and the LCD will show the Calibration Mode main
menu. See “Chapter 4: Calibration Mode” on page 54 for instructions to calibrate
the EAGLE 2. When you are done with the calibration and exit Calibration
Mode, the unit will begin the startup sequence. If the calibration was successful,
the screen above will not appear again until the unit is due for calibration. If the
calibration was not successful, the screen above will again appear in the startup
sequence.
To continue without performing a calibration, press and release the DISPLAY
ADJUST NO button.
EAGLE 2 Operator’s ManualStart Up • 29
•If the unit is due for calibration and CAL PAST DUE ACT is set to MUST
CALIBRATION DATE
IS PAST DUE
ENTER TO PERFORM
CALIBRATION
CALIBRATION DATE
IS PAST DUE
9/12/2008
15:00:00
CALIBRATE, then the following screen displays and the buzzer sounds in a
double pulsing pattern.
The EAGLE 2 cannot be used until a successful calibration has been performed.
Press and release the ENTER button to enter Calibration Mode. See “Chapter 4:
Calibration Mode” on page 54 for instructions to calibrate the EAGLE 2.
NOTE:In this situation, even if the password function has been turned on, no password
is required to perform a calibration.
When you are done with the calibration and exit Calibration Mode, the unit will
begin the startup sequence. If the calibration was successful, the screen above
will not appear again until the unit is due for calibration. If the calibration was
not successful, the screen above will again appear in the startup sequence.
•If the unit is due for calibration and CAL PAST DUE ACT is set to
NOTIFICATION ONLY, then the following alert screen displays and the buzzer
sounds in a double pulsing pattern.
Press and release the POWER ENTER RESET button to acknowledge the alert
and continue with the startup sequence.
10. The Date/Time Screen appears for a few seconds.
30 • Start U pEAGLE 2 Operator’s Manual
Loading...
+ 239 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.