RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in
materials, workmanship, and performance for a period of one year from date of shipment
from RKI Instruments, Inc. Any parts found defective within that period will be repaired or
replaced, at our option, free of charge. This warranty does not apply to those items, which
by their nature, are subject to deterioration or consumption in normal service, and which
must be cleaned, repaired, or replaced on a routine basis. Examples of such items are as
follows:
a) Absorbent cartridgesd) Batteries
b) Pump diaphragms and valvese) Filter elements
c) Fuses
Warranty is voided by abuse including mechanical damage, alteration, rough handling, or
repair procedures not in accordance with the operator’s manual. This warranty indicates
the full extent of our liability , and we are not responsible for removal or replacement costs,
local repair costs, transportation costs, or contingent expenses incurred without our prior
approval.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER
WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND
ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI
INSTRUMENTS, INC., INCLUDING BUT NOT LIMITED TO, THE WARRANTY
OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO
EVENT SHALL RKI INSTRUMENTS, INC., BE LIABLE FOR INDIRECT,
INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND
CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS
PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
This warranty covers instruments and parts sold to users by authorized distributors,
dealers, and representatives as appointed by RKI Instruments, Inc.
We do not assume indemnification for any accident or damage caused by the operation of
this gas monitor, and our warranty is limited to the replacement of parts or our complete
goods.
Beacon 410 Gas Monitor Operator’s ManualTable of Contents
Chapter 1: Introduction
Overview
This chapter briefly describes the Beacon 410 Gas Monitor. This chapter also describes
the Beacon 410 Gas Monitor Operator’s Manual (this document). Table 1 at the end of this
chapter lists the specifications for the Beacon 410.
About the Beacon 410 Gas Monitor
The Beacon 410 is a fixed-mounted, continuous-monitoring controller. This multiple
channel gas monitor is capable of detecting gas at up to four locations. The display screen
simultaneously displays the gas readings of all active channels. Both direct connect
(internal amplifier type) detector heads and 4 - 20 mA transmitter (remote amplifier type)
detector heads may be used with the Beacon 410.
The Beacon 410 includes audible and visual alarms that warn you of hazardous gas
conditions. The alarm circuits include up to three levels of gas alarms. The fail circuit alerts
you to failures in the detector heads or Beacon 410.
The Beacon 410 has four selection menus that allow you to configure various Beacon 410
channel and instrument parameters, the Global Menu, Configuration Menu, Input Setup
Menu, and Modbus Menu. It also has an operational mode, Calibration Mode, that enables
you to calibrate the Beacon 410’s active channels.
About this Manual
The Beacon 410 Gas Monitor Operator’s Manual is organized as follows:
•Chapter 1 is an introduction to the Beacon 410.
•Chapter 2 describes the components of the Beacon 410.
•Chapter 3 describes the installation and start-up procedures of the Beacon 410.
•Chapter 4 describes the operatio n of the Beaco n 410.
•Chapter 5 describes the Global Menu which allows you to set instrument parameters.
•Chapter 6 describes the Configuratio n Menu whic h allo ws you to set channel
parameters.
•Chapter 7 describes the Input Setup Menu which allows you to set the input type for
each channel.
•Chapter 8 describes Calibration Mode which allows you to calibrate the Beacon 410’s
active channels.
•Chapter 9 describes the Beacon 410’s RS-485 Modbus output.
•Chapter 10 describes the Beacon 410’s maintenance requirements and procedures.
The Beacon 410 Gas Monitor Operator’s Manual uses the following conventions for notes,
cautions, and warnings:
Beacon 410 Gas Monitor Operator’s ManualOverview • 1
NOTE: Describes additional or critical information.
!
CAUTION: Describes potential damage to equipment.
WARNING:Describes potential danger that can result in injury or death.
Caution: refer to accompanying documentation
~ Vac (AC voltage)
Vdc (DC voltage)
2 • About this ManualBeacon 410 Gas Monitor Operator’s Manual
Construction (housing)Fiberglass/polyester with lexan window (NEMA 4X)
Dimensions12.5 in. H x 11.0 in. W x 6.4 in. D
Weight10.4 lbs. (without AC line cord)
Table 1: Beacon 410 Specifications
DescriptionSpecification
or
24 V ± 10%, 2.5A VDC
(31.8 cm H x 27.9 cm W x 16.3 cm D)
Environmental Conditions
•For indoor or outdoor locations (Type 4X)
•2000m max. altitude
•-20°C to 50°C (-4°F to 122°F) max. ambient
•Maximum humidity of 80% relative
•Mains supply voltage fluctuations not exceeding ± 10%
of nominal
•DC supply voltag e flu ctu ati ons not ex ce edi ng +10 % -8%
of nominal
•Overvoltage Category II, Polluti on Degre e 2
Safety/Regulatory
C US
186718
User Controls
Relays
•Reset switch
•Program buttons: ESCAPE, UP/YES, DOWN /NO, and
ENTER)
•Relay contacts rated for 10A @ 115/220V~ resistive or
10A @ 30V resistive
•SPDT, Form C (common, normally open, and normally
closed contacts)
Standard AccessoryOperator’s manual (this document)
Beacon 410 Gas Monitor Operator’s ManualSpecifi cations • 3
Chapter 2: Description
Overview
This chapter describes the Beacon 410’s external and internal components.
External Description
This section describes the housing and all external components of the Beacon 410. For
the purposes of this description, the housing door is considered the front of the monitor.
Housing
The Beacon 410’s fiberglass housing is weather- and corrosion-resistant. It is suitable for
installation where general purpose equipment is in use. The housing door is hinged on the
left side and is secured by two latches on the right side. The LCD display and status LEDs
are visible through windows in the housing door. Four mounting feet are attached to the
back of the housing (one at each corner). The mounting feet allow you to install the
housing to a vertical surface. Four conduit hubs on the bottom of the housing are for
external wiring connections.
CAUTION: To avoid electrical interference, do not route detector head and power wiring
through the same conduit hub.
Reset Switch
The reset switch is on the bottom of the housing in front of the buzzer. The reset switch
serves four functions:
•Resets the alarm circuits for “latched” alarms after an alarm 1, alarm 2, or alarm 3
condition passes.
You can set each channel for latched or self-resetting alarms in the Configuration
Menu. See “Viewing and Changing Channel Parameters” on page 35 for more
information.
•Silences the buzzer during an alarm 1, alarm 2, or alarm 3 condition if the buzzer
silence parameter in the Global Menu is set to CAN SILENCE BUZZER. See “Viewing
and Changing Global Parameters” on page 32 for more information on setting the
silence feature.
•Resets the strobe during an alarm 1, alarm 2, or alarm 3 condition if the strobe alarm
setting parameters are set to Resettable STROBE in the Configuration Menu. See
“Viewing and Changing Channel Parameters” on page 35 for more information on
setting the strobe alarm setting parameters.
•Displays and resets the minimum and maximum gas concentration values detected.
NOTE: Fail alarms cannot be reset or acknowledged with the reset switch. When a fail
condition passes, the Beacon 410 will automatically reset the fail alarm circuit.
4 • OverviewBeacon 410 Gas Monitor Operator’s Manual
Buzzer
Relay (13X)
Ground Stud
Controller
Terminal Strip
AC Fuse (2X)
Display
Cable
Power Switch
Main PCB
Control PCB
Reset Switch
Buzzer
Common/
Channel Alarm
Terminal Strip
Channel Alarm
Terminal Strip
Detector/
Transmitter
Terminal
Strips
Termination
Jumper
3/4 " Conduit Hub (4X)
Housing
AC In
Terminal Strip
The buzzer is on the bottom of the housing, behind the reset switch. The buzzer sounds
an audible alarm to warn you of gas alarms and instrument failures.
Internal Description
This section describes the internal components of the Beacon 410.
Figure 1: Beacon 410 Gas Monitor Component Location
Control PCB
The control PCB (printed circuit board) is mounted to the power supply mounting plate
which is in turn mounted to the main PCB. The power supply mounting plate and main
PCB are described below. The control PCB includes the LCD display, the LCD contrast
adjust pot, the status LEDs, and the program buttons. It is connected to the main PCB by
the display cable which is a 20 conductor ribbon cable assembly. The display cable
connects to a rectangular connector on the top edge of the control PCB and to the same
type of connector labelled “FRONT PANEL” on the top edge of the main PCB.
Beacon 410 Gas Monitor Operator’s ManualInternal Description • 5
Status LEDs
LCD Contrast
Adjust Pot
LCD DisplayDisplay Cable
Connector
Status LEDs
Control
Switches
Figure 2: Control PCB Component Location
LCD Display
During normal operation, the four line display simultaneously indicates the target gas,
current gas reading, and measuring unit of each active channel. The display also shows
messages, settings, and other data when you are operating the various selection menus
and operating modes.
LCD Contrast Adjust Pot
The LCD contrast adjust pot is located to the left of the LCD. If the LCD contrast is too
dark or too light to read easily, use a small phillips screwdriver adjust it until you can easily
read the LCD.
Status LEDs
The Beacon 410 includes seven status LEDs that indicate the current status of the
monitor: the RX & TX LEDs, the pilot LED, the fail LED, the alarm 1 LED, the alarm 2 LED,
and the alarm 3 LED (see Figure 2).
•RX & TX LEDs
These LEDs indicate data being received (RX) and transmitted (TX) when the Beacon
410’s Modbus output is operating.
•Pilot LED
The PILOT LED is on when the Beacon 410 is receiving incoming power, either AC or
DC power.
•Fail LED
The fail LED turns on when the Beacon 410 is experiencing a fail condition. A fail
condition can be caused by a failure within the Beacon 410 or the detector heads
6 • Internal DescriptionBeacon 410 Gas Monitor Operator’s Manual
wired to the Beacon 410 (see “Chapter 10: Maintenance” on page 64).
•Alarm 1 LED
The alarm 1 LED is on when the Beacon 410 is experiencing an alarm 1 condition.
•ALARM 2 LED
The alarm 2 LED is on when the Beacon 410 is experiencing an alarm 2 condition.
•ALARM 3 LED
The alarm 3 LED is on when the Beacon 410 is experiencing an alarm 3 condition.
Control Buttons
The Beacon 410 includes four control buttons that allow you to enter the selection menus
and Calibration Mode, navigate through the menus and Calibration Mode, update
instrument and channel parameter settings, and save changes to the settings. The control
buttons, listed in Table 2, are to the right of the LCD display (see Figure 2).
Table 2: Beacon 410 Control Button Functions
ButtonFunction
ESCAPE
UP (YES)
DOWN (NO)
ENTER
•Moves backward through the menu and mode screens
•Aborts operations
•Cancels changes you make in the menus
•Enters the Configuration Menu (press with ENTER button)
•Enters the Global Menu (press with the UP/YES button)
•Initiates an operation or proceeds to the next screen when a yes/no
question is asked on a screen
•Changes the displayed set ting
•Enters the Global Menu (press with ESCAPE) button
•Enters the Calibration Mode (press with ENTER button)
•Cancels an oper atio n o r se que nc e w he n a y es /no question is asked on a
screen.
•Changes the displayed set ting
•Enters the Modbus Menu (press with ENTER button)
•Saves changes you make in the menu and mode screens
•Accepts the displayed parameter setting
•Enters the Configuration Menu (press with ESCAPE button)
•Enters Calibration Mode (press with UP/YES button)
•Enters the Modbus Menu (press with DOWN/NO button)
Main PCB
Terminal Strips
The Beacon 410 includes 9 terminal strips for wiring connections. See “Wiring the Beacon
410 Gas Monitor” on page 15 for detailed wiring procedures.
•Str ob e Terminal Strip
The strobe terminal strip is a 2-point terminal strip located in the upper left corner of
the main PCB. When the optional strobe is ordered with a Beacon 410, the strobe
terminal strip is used to factory wire the strobe.
Beacon 410 Gas Monitor Operator’s ManualInternal Description • 7
CAUTION: The strobe terminals are intended for use with the RKI supplied optional
strobe. Consult RKI Instruments, Inc. before attempting to use these
terminals for some other alarm device.
•Detector/Transmitter Terminal Strips
Four detector/transmitter terminal strips are located along the bottom left side of the
main PCB (see Figure 1 on page 5). These four 11-point terminal strips facilitate
wiring connections to the detector heads. They also provide terminals to connect a
recording device to a 4 to 20 mA output for each channel. The top terminal strip is for
channel 1 connections and each subsequent strip is used for the next channel with the
bottom terminal strip being for channel 4 connections.
•Channel Alarm Terminal S trip
A channel alarm terminal strip is located to the right of the channel 4
detector/transmitter terminal strip (see Figure 1 on page 5). This 24-point terminal
strip facilitates wiring external alarm devices (horn, light, etc.) to relay contacts that
are field configurable for alarm levels and operation and are controlled by individual
channels.The contacts are labelled NO (normally open), NC (normally closed),
and C (common). See “Viewing and Changing Global Parameters” on page 32 and
“Viewing and Changing Channel Parameters” on page 35 for instructions to configure
the operation of these contacts.
•Common/Channel Alarm Terminal Strip
The common/channel alarm terminal strip is located in approximately the middle of the
main PCB above the channel alarm terminal strip. This 15-point terminal strip
facilitates wiring external alarm devices (horn, strobe, etc.) to relay contacts that are
field configurable as individual alarm contacts that are controlled by individual
channels or as common alarm contacts which are controlled by all channels. The
contacts are labelled NO (normally open), NC (normally closed), and C (common).
See “Viewing and Changing Global Parameters” on page 32 and “Viewing and
Changing Channel Parameters” on page 35 for instructions to configure the operation
of these contacts.
•Controller terminal strip
The 9-point controller terminal strip is along the lower right side of the main PCB (see
Figure 1 on page 5). The controller terminal strip facilitates various internal and
external wiring connections. Table 3 lists the function of each terminal.
T able 3: Terminal Assignments for the Controller Terminal Strip
TerminalConnects to:
EXT DC/24V BATT
+ connection from 24 VDC power source
+
EXT DC/24V BATT
- connection from 24 VDC power source
-
1
(or 24 V backup battery)
1
(or 24 V backup battery)
1
1
RS-485 A
RS-485 BAllow connection of the Beacon 410 to a Modbus network
RS-485 C
8 • Internal DescriptionBeacon 410 Gas Monitor Operator’s Manual
T able 3: Terminal Assignments for the Controller Terminal Strip (Continued)
If 24 VDC is used as primary power source do not make wiring connections to the AC terminal strip.
•AC terminal strip
The 3-point AC terminal strip is located above the controller terminal strip (see
Figure 1 on page 5). The AC terminal strip facilitates wiring connections to the AC
power source.
Relays
The Beacon 410 includes eight channel relays and one common fail relay . It also includes
four common/channel relays that can be defined as a group as channel or common relays.
All the relays have single-pole double-throw (SPDT) contacts, also known as form C
contacts, and are rated for 10 amps at 115 VAC (resistive). The contacts are available at
the channel alarm and common/channel alarm terminal strips and are labelled NO
(normally open), NC (normally closed), and C (common).
Beacon 410 Gas Monitor Operator’s ManualInternal Description • 9
Figure 3: Beacon 410 Relay Allocation
Channel Relays & Fail Relay
The eight channel relays are above the channel alarm terminal strip (see Figure 1 on
page 5 and Figure 3 on page 9). These relays are dedicated to specific channels. Figure 3
illustrates the allocation of the channel relays.
The fail relay is located directly to the left of the channel relays. The fail relay is a common
relay.
Common/Channel Relays
The four common/channel relays are above the common/channel alarm terminal strip (see
Figure 1 on page 5 and Figure 3 on page 9). These relays can be configured as either all
channel relays or all common relays in the Global Menu.
Termination Jumper
A two pin header with a termination jumper installed is located near the top edge of the
main PCB to the right of the display cable connector. It is labelled “RS-485 Line
Terminator” on the PCB silkscreen. The jumper has no function unless the Beacon 410 is
wired into a Modbus installation. See “Chapter 9: RS-485 Modbus Output” on page 54 for
a instructions to use the Beacon 410 in a Modbus system.
Ground Stud
The threaded ground stud is used for making connections to earth ground. It is located in
the lower right corner of the Main PCB and is connected through the main PCB to the
(ground) terminal on the AC terminal strip. A kep nut on the stud may be removed for
installation of one or more user supplied lugs to make wiring connections to earth ground.
This stud is typically used to connect the shield drain wire of shielded cable to earth
ground at the Beacon 410.
G
Power Switch
The power switch is between the common/channel relays and the fuses (see Figure 1 on
page 5 and Figure 3 on page 9). The power switch turns the incoming AC power source
on and off at the Beacon 410. When the switch is up, the power is on.
Power Supply
The power supply is mounted to the power supply mounting plate which is located behind
the display PCB. The power supply mounting plate is mounted to the main PCB with four
standoffs. The power supply receives AC power from the external power source and
converts it to a DC voltage that is usable by the Beacon 410 circuitry. A polycarbonate
cover prevents accidental contact with the AC terminals on the power supply.
AC & DC Circuit Protection
Two AC fuses are used in the Beacon 410. The two fuses are located on the right side of
main PCB, between the power switch and the AC in terminal strip (see Figure 1 on page 5
and Figure 3 on page 9). They cut off the incoming AC power in the event of a short circuit
or other electrical fault which causes a high current draw in the Beacon 410. They are
housed in vertical fuse holders and are held in each holder by a quarter turn cover. They
are fast blow fuses rated at 6 A, 250 V.
A polyswitch is used to protect the DC power input. It is located on the right side of the
main PCB above the AC fuses and is labelled on the PCB silkscreen as PS9. In the event
of a short circuit or other electrical fault which causes a high current draw in the Beacon
410, the polyswitch will interrupt the DC power if the unit is powered from DC. When the
fault situation is corrected, the polyswitch resets and the unit will continue to operate. The
polyswitch is not user serviceable.
10 • Internal DescriptionBeacon 410 Gas Monitor Operator’s Manual
Optional Accessories
Strobe Terminal S trip
This section describes the optional accessories available for the Beacon 410. Both
optional accessories are wired to the Strobe Terminal Strip as shown below.
Figure 4: Strobe Terminal Strip Location
Beacon 410 Gas Monitor Operator’s ManualOptional Accessories • 11
Alarm Strobe
13.39
Strobe (O ptional)
.90
The Beacon 410 can be ordered with a red alarm strobe light installed on the top of the
housing. The Beacon 410 retains its NEMA 4X rating with the strobe installed. Strobe
operation can be programmed in Configuration Mode (see “Chapter 6: Configuration
Menu” on page 35). The outline and mounting dimensions with the alarm strobe are the
same as the standard Beacon 410 with the excpetion of the height. The difference is
shown below. See Figure 7 for all outline and mounting dimensions.
Figure 5: Alarm Strobe Dimensions
Horn/Strobe
The Beacon 410 can be ordered with a horn/strobe. This optional horn/strobe allows the
user to have both a strobe and a horn connected to the “Strobe” terminals of the Beacon
410. It also allows the user to mount the horn/strobe away from the Beacon 410 so that it
can be somewhere more visible/audible. Strobe operation can be programmed in
Configuration Mode (see “Chapter 6: Configuration Menu” on page 35).
12 • Optional AccessoriesBeacon 410 Gas Monitor Operator’s Manual
NOTE: See “Chapter 3: Installation and Start Up” on page 14 for complete Beacon
Strobe Terminal Strip
-
Dimensions Shown in Inches
4.83
6.81
5.97
6.17
+
410 installation instructions.
CAUTION:Do not adjust the strobe brightness or the horn volum e at the strobe/horn, as this
may overload the Beacon 410 strobe control circuit.
The horn/strobe does not come factory installed to the Beacon 410. To install the horn/
strobe:
1. Mount the horn/strobe in the desired location.
2. Install an appropriately rated cable bushing or conduit to the left conduit hub on the
bottom of the Beacon 410 housing.
3. Install an appropriately rated cable bushing or conduit at the horn/strobe.
4. Run a cable or wires in conduit from the horn/strobe to the Beacon 410 through the
left conduit hub.
5. Connect the wires to the strobe terminal strip as shown in the figure below.
Figure 6: Horn/Strobe Wiring/Oultine and Mounting Dimensions
Beacon 410 Gas Monitor Operator’s ManualOptional Accessories • 13
Chapter 3: Installation and Start Up
Overview
This chapter describes procedures to mount the Beacon 410 Gas Monitor, make wiring
connections to the monitor, and start up the monitor.
WARNING:Perform all installation and start-up procedures in a known fresh air
environment, an environment free of combustible and toxic gasses and of nor mal
oxygen content. The Beacon 410 is not in operation as a gas monitoring
controller until the start up procedure is complete.
Mounting the Beacon 410 Gas Monitor
Perform the following procedure to install the Beacon 410 at the mounting site.
1. Select the mounting site. When you select the mounting site, consider the following
factors:
•Is an AC or DC power source available?
•Is a vertical surface available to mount the Beacon 410?
•Is there enough room to open the housing door and make wiring connections
through the conduit hubs at the bottom of the housing?
•Are the display screen and status lights visible?
2. Close and latch the housing door.
3. The Beacon 410 is shipped with the mounting feet positioned under the housing.
Loosen the screws that secure the feet to the housing, rotate the feet to their mounting
position as shown in Figure 7, then tighten the screws.
4. Prepare the selected mounting site as required to mount the Beacon 410. It should be
mounted at eye level (4 1/2 to 5 feet from the floor). Refer to Figure 7 for the outline
and mounting dimensions.
5. Position the monitor on the vertical mounting surface.
6. Insert 1/4 in. screws through the slots in the mounting feet at each corner of the
housing to secure the housing to the mounting surface.
14 • OverviewBeacon 410 Gas Monitor Operator’s Manual
Ø .31 x .50 slot, 4X
Door Latches
13.39
12.94
12.50
3/4" Conduit Hubs, 4X
6.43
3.63
8.00
6.25
10.50
Figure 7: Beacon 410 Gas Monitor Outline and Mounting Dimensions
Wiring the Beacon 410 Gas Monitor
This section describes procedures to connect the AC power source, DC power source,
Modbus wiring (refer to “Wiring the Beacon 410 in a Modbus System” on page 54),
external alarms, recording devices, and detector heads. See Figure 8 on page 16 for a
general diagram of all external wiring to the Beacon 410.
WARNING:Make all connections to the Beacon 410 before you plug in or turn on the AC or
DC power source. Before you make any wiring adjustments, always verify that all
power sources are not live.
Beacon 410 Gas Monitor Operator’s ManualWiring the Beacon 410 Gas Monitor • 15
Neutral
Not
Use d
DP S W
L
N
AC
RS - 485
BAC
ALARM
BUZZER
CH 4
RELAY-1
NO N C C
Alarm Device
Power
NO N C C
RELAY-CFAIL
NO NC C NO NC C
RELAY-A
CH 3
RELAY-1
NO N C CNO N C C
CH 2
RELAY-2
CH 1
RELAY-2
NO N C C NO NC C
CH 1
RELAY-1
BG
4-20m A
OUTPUT
OXY
S
AMP
G
W
G
Ground
EXT DC
24V BATT
CH 4
RELAY-2
NO N C C
Buzzer
(Factory wired)
Alarm Device
Power
24V DC
See ModbusWiring
In Chapter 9
AC Power
100/115 VAC
50 / 60 Hz
Li n e ( Hot)
RELAY-B
NO N C C
NO N C C
CH 2
RELAY-1
Strobe Terminal Strip
(Factory wired)
Optional Strobe
Red
/ Transmitter Ter minal Strip
(typical 1 of 4)
R
LEL
W
to DetectorHeadManual
Be ac on 410 Det e c tor Head
Sheet for Specific
Detector/Transmitter Wiring
NOTE: Line and Neutral are L1 and L2 for
220 VAC Wiring
Black
Controller Terminal Strip
Recording Device
500 Ohms Impedance Maximum
ACIn Terminal Strip
Reset Switch
(Factory wired)
ALARM
RESET
Alarm D evices,T ypicalAlarmWiring Shown
RelayContactsRatedFor10AmpsAt250 VAC
Common / Channel Alarm T erminalStrip
NO N C C
RELAY-D
Alarm D evices,T ypicalAlarmWiring Shown
RelayContactsRatedFor10AmpsAt250 VAC
Channel Alarm TerminalStrip
NO N C C
CH 3
RELAY-2
16 • Wiring the Beacon 410 Gas MonitorBeacon 410 Gas Monitor Operator’s Manual
Figure 8: Beacon 410 Gas Monitor External Wiring Diagram
Connecting the AC Power Source
220 VAC Wiring
AC In Terminal Strip
Earth Ground
N
G
L2
G
N
AC Power
100/115 VAC
50 / 60 Hz
AC In Terminal Strip
100/ 1 15 V AC Wirin g
Neutral
Earth Ground
Line (Hot)
L
L1
AC Power
220 VAC
50 / 60 Hz
L
NOTE: If you are using DC power as the primary power source, go to the next section,
“Connecting the DC Power Source”.
The AC in terminal strip will accept 24 - 14 AWG wire. When selecting wire to connect the
AC power source to the Beacon 410, be sure to meet the local electrical code. Also be
sure to use an appropriate circuit breaker in the AC line close to the Beacon 410 that
meets the local electrical code.
Perform the following procedure to connect the AC power source to the Beacon 410.
WARNING:Verify that the power sou rce is unplugged or turned off before you continue with
this procedure.
1. Turn off or unplug all incoming power to the Beacon 410.
2. Open the housing door, then place the power switch in the OFF position.
CAUTION: The power switch does not control DC input power.
3. Locate the AC in terminal strip (see Figure 1 on page 5). The terminals are labelled L,
N, and G for line, neutral, and ground respec ti ve ly.
4. Install an appropriately rated cable bushing or conduit in the right-most conduit hub on
the bottom of the Beacon 410 housing.
5. Guide the AC power cord or wires in conduit through the right-most conduit hub on the
bottom of the Beacon 410 housing.
CAUTION: Do not route power and detector head wiring through the same conduit hub.
The power wiring may disrupt the transmission of the detector head signal to
the monitor.
6. Connect the AC wires to the AC in terminal strip as shown in Figure 9 below.
Figure 9: AC Power Wiring
Connecting the DC Power Source
Beacon 410 Gas Monitor Operator’s ManualWiring the Beacon 410 Gas Monitor • 17
WARNING:Verify that the power sou rce is unplugged or turned off before you continue with
this procedure.
DC power may be used as a primary power source. It may also be used as a backup
power source with a 24 VDC battery if AC power is the primary power source. If your
Beacon 410 does not include the battery charging feature, you may use a self contained
DC Power In Terminals on
Con troller Terminal Strip
EXT DC
24V BATT
DC Power
24 VDC
24 VDC backup that keeps its’ batteries charged while AC power is on and recharges the
batteries when AC power returns after a power failure. If your Beacon 410 includes the
battery charging feature, see “Battery Charging (Optional)” on page 25 for a complete
description of this feature and what type of battery to use. If DC power is the
power source,
1. Turn off or unplug all incoming power to the Beacon 410.
2. Open the housing door, then place the power switch in the OFF position.
CAUTION: The power switch does not control DC input power.
3. Locate the DC input power terminals on the controller terminal strip near the lower
right edge of the main PCB (see Figure 1 on page 5). They are labelled
BATT+
4. Install an appropriately rated cable bushing or conduit in an unused conduit hub on
the bottom of the Beacon 410 housing.
5. Guide a DC power cord or two wires in conduit through the selected conduit hub on
the bottom of the Beacon 410 housing.
DO NOT connect AC power.
and EXT DC/24V BATT -.
primary
EXT DC/24V
CAUTION: Do not route power and detector head wiring through the same conduit hub.
The power wiring may disrupt the transmission of the detector head signal to
the monitor.
6. Connect the DC power wires to the controller terminal strip as shown in Figure 10
below.
Figure 10: DC Power Wiring
NOTE: The Beacon 410 will operate from the DC input down to 18.5 volts. If a self
contained backup battery is used, see its operator’s manual for a description of
its recharging characteristics. If your Beacon 410 includes the battery charging
feature and a backup battery is used, the Beacon 410 will recharge the battery
when AC power has returned after a power failure. See “Battery Charging
(Optional)” on page 25 for a complete description of the battery charging feature.
RS-485 Modbus Wiring
18 • Wiring the Beacon 410 Gas MonitorBeacon 410 Gas Monitor Operator’s Manual
See “Wiring the Beacon 410 in a Modbus System” on page 54 for wiring connections to
the RS-485 Modbus terminals.
Connecting External Alarms
(+) H
(-) N
External
Power Source
Channel 1, Relay1 AlarmTerminals
From ChannelAlarmTerminal Strip
External Alarm Device
CH 1 RELAY 1
NO NC C
Before connecting any external alarm devices to the relay contacts, make sure you know
how you want the devices to operate. For example, confirm under what alarm condition
you want a device to turn on or turn off and what channel is going to control the device.
Also make sure that the parameter settings that apply to the relays in the Global Menu and
the Configuration Menu are set so that the desired alarm device operation is obtained.
See “Viewing and Changing Global Parameters” on page 32 and “Viewing and Changing
Channel Parameters” on page 35 for information about the relay parameters.
Perform the following procedure to connect external alarm devices to the Beacon 410.
1. Turn off or unplug all incoming power to the Beacon 410.
2. Open the housing door, then place the power switch in the OFF position.
CAUTION: The power switch does not control DC input power.
3. Locate the applicable alarm terminal strip (see Figure 1 on page 5).
4. Install an appropriately rated cable bushing or conduit in an unused conduit hub on
the bottom of the Beacon 410.
5. Guide the wiring of the external alarm device through the selected conduit hub on the
bottom of the Beacon 410 housing.
CAUTION: Do not route the external alarm wiring and detector head wiring through the
same conduit hub. The external alarm wiring may disrupt the transmission of
the detector head signal to the Beacon 410.
6. Connect the leads from the external alarm device and an external power source to the
selected channel alarm or common/channel alarm relay contact terminals as shown in
Figure 8.
Figure 11: External Alarm Wiring
7. Repeat step 5 and step 6 for additional external alarm devices.
Beacon 410 Gas Monitor Operator’s ManualWiring the Beacon 410 Gas Monitor • 19
Connecting Recorders
Recorder Output Termin als Fro m a
Detector/Transmitter Terminal Strip,
Typical of 4
4-20 mA
OUTPUT
4 - 20 mA Input
Recordin g De vice
500 Ohms Impedance
Maximum
Perform the following procedure to connect an analog signal recording device to the
Beacon 410. The output at the recorder output terminals for each channel is a 4 - 20 mA
signal that corresponds to the detection range of the detector head connected to that
Beacon 410 channel. Be sure to read the recording device’s operator’s manual before
installation and follow all wiring procedures and recommendations made by the recording
device’s manufacturer.
1. Turn off or unplug incoming power to the Beacon 410.
2. Open the housing door, then place the power switch in the OFF position.
CAUTION: The power switch does not control DC input power.
3. Locate the recorder output terminals on the right end of the detector/transmitter
terminal strips. See “Beacon 410 Gas Monitor Component Location” on page 5 to
assist you in locating the recorder output terminals. They are labelled
OUTPUT +
and 4 - 20 mA OUTPUT -.
4. Install an appropriately rated cable bushing or conduit in an unused conduit hub on
the bottom of the Beacon 410 housing.
5. Guide the wiring from the recording device through the selected conduit hub on the
bottom of the Beacon 410 housing.
6. Connect the leads from the recording device to the recorder output terminals of the
selected active channels as shown in Figure 12 below.
4 - 20 mA
Figure 12: Recorder Output Wiring
Connecting RKI Detector Heads
When a Beacon 410 is ordered from the factory, any detector heads that were ordered
with it are already setup on particular channels of the Beacon 410. If you are adding an
RKI detector head to your existing system, you will need to setup one of the unused
channels to operate the detector head. See “Selecting the Detector Head Input Type and
Gas Setup” on page 41 for instructions to setup a new channel.
Perform the following procedure to connect an RKI detector head to the Beacon 410.
1. Turn off or unplug power to the Beacon 410.
2. Open the Beacon 410 door and place the power switch in the off position.
20 • Wiring the Beacon 410 Gas MonitorBeacon 410 Gas Monitor Operator’s Manual
CAUTION: The power switch does not control DC input power.
3. See the detector head operator’s manual for instructions on how to connect the
detector head to a controller.
4. Install an appropriately rated cable bushing or conduit in an unused conduit hub on
the bottom of the Beacon 410 housing.
5. Route the wires in conduit or shielded cable from the detector head through the
selected conduit hub into the Beacon 410. See Table 4 below for wire size and
distance guidelines.
6. Unshielded twisted pair cable in conduit or shielded twisted pair cable is
recommended for all the direct connect detector heads. For the LEL detector, pair and
twist the
recommended for the 2-wire and 3-wire 4 - 20 mA transmitters.
7. Connect the wires from the detector head to the appropriate detector/transmitter
terminals on the appropriate channel. See the detector head operator’s manual and
the Beacon 410 Detector Head Specification Sheet for detector head connections to
the Beacon 410. If shielded cable is used, leave the cable shield’s drain wire
disconnected and insulated at the detector head and connect the cable shield’s drain
wire at the Beacon 410 to the ground stud on the main PCB.
CAUTION: Do not route power and detector head wiring through the same conduit hub.
R & B wires and the W & G wires. Shielded cable or wires in conduit are
The power wiring may disrupt the transmission of the detector head’s signal
to the Beacon 410.
Table 4: Wire Size Guidelines for RKI Detector Head Wiring
Number of
Detector Head Type
Direct Connect LEL4500 ft.1,000 ft.2,000 ft.
Direct Connect Oxygen2500 ft.1,000 ft.2,000 ft.
Direct Connect H2S2500 ft.1,000 ft.2,000 ft.
Direct Connect CO2500 ft.1,000 ft.2,000 ft.
Direct Connect ESM-01
type
2-Wire 4 - 20 mA
Transmitter
3-Wire 4 - 20 mA
Transmitter
Wires to
Controller
2500 ft.1,000 ft.2,000 ft.
22,500 ft.5,000 ft.8,000 ft.
32,500 ft.5,000 ft.8,000 ft.
Max Distance
to Controller
w/18 Gauge
Wire
Max Distance
to Controller
w/16 Gauge
Wire
Max Distance
to Controller
w/14 Gauge
Wire
Connecting User-Supplied 4 - 20 mA Transmitters
The Beacon 410 may be used with a user supplied 2-wire or 3-wire 4 - 20 mA transmitter
which runs on 24 VDC. When this is done, the Beacon 410 is normally setup at RKI
Instruments with the following channel parameters: unit of measure, item name, and full
scale. For example, “PSI AIR” with a full scale of 10 PSI.
If a user supplied 4 - 20 mA transmitter is added in the field, it will be necessary to setup
the additional channel. See “Selecting the Detector Head Input Type and Gas Setup” on
Beacon 410 Gas Monitor Operator’s ManualWiring the Beacon 410 Gas Monitor • 21
page 41 for instructions to setup a new channel.
2-Wire Connection3-Wire Connection
4 - 20 mA T ransmitter Terminals
From Detector/Transmitter
Terminal Strip, Typical of 4
4 - 20 mA T ransmitter Terminals
From Detector/T ran smitter
Terminal Strip, Typical of 4
S
AMP
4 - 20 m A
+ 24 VDC
2-Wire 4 - 20
mA Transm itter
3-Wi re 4 - 20
mA Transmitte r
- (DC Ground)
4 - 20 mA
+ 24 VDC
S
AMP
Figure 13: Generic 4 to 20 mA Transmitter Wiring
Perform the following procedure to connect a 4 - 20 mA transmitter which you supply to
the Beacon 410.
1. Turn off or unplug power to the Beacon 410.
2. Open the Beacon 410 door and turn off the power switch.
CAUTION: The power switch does not control DC input power.
3. See the transmitter’s operator’s manual for instructions on how to connect wires to the
transmitter.
4. Install an appropriately rated cable bushing or conduit in an unused conduit hub on
the bottom of the Beacon 410 housing.
5. Route the wires from the transmitter through the selected conduit hub into the Beacon
410.
6. Connect the wires from the transmitter to the appropriate channel’s
detector/transmitter terminal strip. See the transmitter operator’s manual for controller
terminal connections and wiring recommendations.
CAUTION: Do not route power and transmitter wiring through the same conduit hub. The
power wiring may disrupt the transmission of the transmitter’s signal to the
Beacon 410.
Perform the following procedure to place the Beacon 410 into normal operation.
1. Complete the mounting and wiring procedures described earlier in this chapter.
2. Complete all installation procedures described in the detector head or user supplied
4 - 20 mA transmitter operator’s manuals.
3. Verify that all wiring connections are correct and secure and that the Beacon 410’s
power switch is in the OFF position.
22 • Starting Up the Beacon 410 Gas MonitorBeacon 410 Gas Monitor Operator’s Manual
4. Plug in or turn on the incoming power source (AC or DC).
10
stnemurtsnIIKR
Beacon4
0100
NF
0000:
in:0001
10001000100
:
O
M
W
1000B
0
aM
H
VERSIONI
:
OL
LEL%0ENAHTEM
1
2
3
4
:
:
:
CO0ppm
mpp0S2H
OXYGEN20 . 9%V
5. Turn on the power switch if AC power is used as primary power.
6. The following screen appears for a few seconds.
7. The Version Screen then appears. It shows the instrument’s hardware and firmware
versions for a few seconds. HW is the hardware version. Main is the main firmware
version. MB is the Modbus firmware version. The version numbers on the bottom line
are the firmware versions loaded for each channel.
To keep the version information screen on the display longer than a few seconds,
press and hold the ENTER button while it is being displayed. When you release the
ENTER button, the startup sequence will continue.
8. After the Version Screen has been displayed for a few seconds,
appears for each active channel. The warm-up time is counted down in seconds from
60 seconds for each active channel on the far right.
NOTE: To prevent unwanted alarms during warm up, the alarm circuits are not active
while the WARMING UP message is displayed.
9. Any unused channels are configured as
factory. For any unused channels,
channel.
If any channels have been configured as
STANDBY is displayed on the line for that channel.
See “Viewing and Changing Channel Parameters” on page 35 for a description of the
NOT USED and STANDBY configurations.
10. When the warm-up period is complete, normal operation will begin.
operation, the display will indicate the current gas reading and target gas. Verify that
the display is indicating the current gas reading and target gas for all active channels
after the warm-up period is complete and normal operation begins, for example:
NOT USED in the Configuration Menu at the
NOT USED is displayed on the line for that
STANDBY in the Configuration Menu,
WARMING UP
During normal
Beacon 410 Gas Monitor Operator’s ManualStarting Up the Beacon 410 Gas Monitor • 23
11. Verify that the PILOT light is on. If the PILOT light is not on, see the troubleshooting
guide in “Troubleshooting” on page 64.
12. Perform the start-up procedure for each detector head or user supplied 4 - 20 mA
transmitter as described in the detector head or user supplied transmitter’s operator’s
manual.
Chapter 4: Operation
:
OL
LEL%0ENAHTEM
1
2
3
4
:
:
:
CO0ppm
mpp0S2H
OXYGEN20 . 9%V
Overview
This chapter describes the Beacon 410 Gas Monitor in normal operation. This chapter
also describes the Beacon 410 in alarm 1, alarm 2, alarm 3, and fail conditions, and
suggests responses to these conditions.
Normal Operation
Normal operation is defined as follows:
•the start-up procedure is complete.
•the Beacon 410 is not indicating an alarm 1, alarm 2, alarm 3, or fail condition.
•the Beacon 410 is not running in one of the selection menus or Calibration Mode.
During normal operation, the Beacon 410 simultaneously displays the target gas, unit of
measure, and current gas reading for all active channels. The example below illustrates a
typical Beacon 410 channel allocation.
The PILOT LED is on indicating that the Beacon 410 is receiving incoming power.
4 - 20 mA Signal Output Operation
The output at the recorder output terminals on the detector/transmitter terminal strip for
each channel is a 4 - 20 mA signal that corresponds to the detection range of the Beacon
410. During normal operation, this signal tracks the gas concentration on the LCD.
There are several circumstances where the signal output will not track the display reading
but will behave as follows:
•When the Beacon 410 is in its warm-up period, the signal output will be fixed at 4 mA
(zero) for all channel types except oxygen. For oxygen channels, the output will be
fixed at 17.4 mA (20.9% oxygen) while the Beacon 410 is in warm-up.
•When a channel’s input type is changed to a new direct connect type of detector head
in the Input Setup Menu, the display will indicate
channel when you exit the Input Setup Menu and enter normal operation and will
continue to indicate this until Calibration Mode is entered and a calibration is
performed. If a custom gas name and range was defined in the Input Setup Menu, the
Beacon 410 will enter the Configuration Menu for you to verify the parameter settings
before continuing to the normal operation and displaying the
message. In this situation, the signal output for the newly configured channel will be
fixed at 0.7 mA until a complete calibration is performed on that channel.
NEEDS AIR/GAS Cal for that
NEEDS AIR/GAS Cal
24 • OverviewBeacon 410 Gas Monitor Operator’s Manual
•If you enter the Global Menu, the Configuration Menu, the Input Setup Menu, the
:00
00
20.920.9
00
1
2
3
4
:
:
:
lamr
Mtn/MaxValuesei
o
s
Wai tToRet urnTo
noitarepOo
r
e
u
N
O
R
P
shResetAgainT
Modbus Menu, or Calibration Mode, the signal output will be fixed at 3.5 mA until the
Beacon 410 returns to normal operation.
•If the Beacon 410’s input power decreases below 18.5 volts so that the Beacon 410 is
in a low power alarm, the signal output is fixed at 0.7 mA until the low power alarm is
cleared.
•If the Beacon 410 goes into a fail condition, after a 30 second delay, the signal output
is fixed at 0.7 mA until the fail alarm is cleared. During the 30 second delay, the output
follows the detector output.
•If a channel is setup as
•If a channel is setup as
NOT USED, the signal output will be fixed at about 0.7 mA.
STANDBY, the signal output will be fixed at 3.5 mA.
Viewing and Resetting Min/Max Readings
The Reset switch may be used to view and reset the minimum and maximum gas
readings for all active channels.
1. While the Beacon 410 is in normal operation, press and hold the reset switch button
for 5 seconds.
2. The display will show the following screen for about ten seconds.
The minimum reading is on the left and the maximum is on the right for each channel.
3. After the minimum and maximum readings have been displayed for about ten
seconds, the following screen appears for about seven seconds.
•To reset the minimum and maximum readings, before the unit returns to normal
operation press and release the reset switch button. The display indicates
Max Values Have Been Reset
•To return to normal operation without resetting the minimum and maximum read-
ings, do not press the reset switch button and allow the unit to return to normal
operation.
Battery Charging (Optional)
The Beacon 410 has an optional backup battery charging feature. In order for this feature
to be included, the Beacon 410 must be ordered with this feature. Consult RKI
Instruments, Inc. for ordering information.
Beacon 410 Gas Monitor Operator’s ManualViewing and Resetting Min/Max Readings • 25
Min/
and the unit will then return to normal operation.
The battery charging circuit is designed to charge lead acid type batteries. If AC power is
used as primary power and a backup battery is connected to the Beacon 410 ‘s EXT DC/
24V BATT terminals as shown in Figure 10 on page 18, the battery charging feature will
charge the battery if it is depleted and keep it charged with a charge current of
approximately 100 mA.
CAUTION: When a battery is used as backup power and the charging feature is included in the
Alarm Indications
This section describes the Beacon 410 in alarm 1, alarm 2, alarm 3 and fail conditions,
and suggests responses to these conditions. Table 5 below lists the alarm indications for
each condition.
NOTE: The Beacon 410 allows configuration of various alarm and alarm relay
Beacon 410, do not use a non-rechargeable battery or a backup battery that has it’s
own charging feature. Use RKI backup battery 49-8104RK or an appr opriately rated
24 VDC rechargeable lead acid type batte ry to backup a Beac on 410 whe n the batte ry
charging feature is included in the Beacon 410.
parameters. The description of alarm indications below assumes that all
parameters are at their factory set value. It also assumes that the alarm setpoints
are set such that alarm 1< alarm 2 < alarm 3 and all alarms are increasing except
for an oxygen channel where alarm 1>alarm 3 and alarm 2 > alarm 3 because
alarm 1 and alarm 3 are decreasing alarms and alarm 2 is an increasing alarm.
See “Viewing and Changing Channel Parameters” on page 35 for detailed
information on displaying or changing various channel parameters including
alarm and alarm relay parameters. T able 7 on page 37 lists the adjustable
parameters and their factory settings.
Table 5: Visual and Audible Alarm Indications
ConditionCauseVisual Indication(s)
Alarm 1Incr easing (decreasing for O2) gas
reading at or above the alarm 1
setpoint
Alarm 2Increasing gas reading at or above
the alarm 2 setpoint
•Alarm 1 LED is on
•Gas reading alternates with
ALARM-1 message
•If installed and set to activate for
alarm 1, strobe flashes
•Alarm 1 and alarm 2 LEDs
(alarm 2 only for oxygen) are on
•Gas reading alternates with
ALM 1,2 message (ALARM-2
for oxygen)
•If installed, strobe con t in ues to
flash if set to activate for alarm 1.
If set to activate for alarm 2 only
or if channel in alarm is an
oxygen channel, it begins to
flash when an alarm 2 condition
begins.
Audible
Indication
Pulsing
Tone
Pulsing
Tone
26 • Alarm IndicationsBeacon 410 Gas Monitor Operator’s Manual
Table 5: Visual and Audible Alarm Indications
ConditionCauseVisual Indication(s)
Alarm 3Incr easing (decreasing for O
reading at or above (below for O
the alarm 3 setpoint
Fail
•Disconnected or misconnected
detector head wiring
•Display reading below -10% of full
scale or lower
•Malfunctioning components
) gas
2
•Alarm 1, alarm 2, and alarm 3
)
2
LEDs (alarm 1 and alarm 3 for
oxygen) are on
•Gas reading alternates with
ALM 1,2,3 (ALM 1,3 for oxygen)
message
•If installed, strobe con t in ues to
flash if set to activ ate fo r ala rm 1
and/or alarm 2 for non oxygen
type channels or alarm 1 for
oxygen channels. If set to
activate for alarm 3 only, it
begins to flash when an alarm 3
condition begins.
•Fail LED is on
•FAIL message replaces the gas
reading
•If installed and set to activate for
fail, strobe flashes
NOTE: There is a 30 second
delay on the fail indications.
Low
Battery
No AC power
and DC power source
(primary or backup) less than 18.5
volts.
•FAIL LED is on
•Display shows LOW POWER STANDBY message and the
input DC voltage
Audible
Indication
Pulsing
tone
Steady
tone
None
NOTE: You can set the channel alarm relays and the common/channel alarm relays
(relays A, B, C, and D) to be either all normally energized or all normally deenergized in the Global Menu. You can also set relays A, B, C, and D as common
alarm relays or channel alarm relays in the Global Menu. The following sections
describe the factory settings of normally de-energized for the channel and
common/channel relays (A, B, C, and D) and common alarm relays for relays A,
B, C, and D. The fail relay is factory-set as normally energized and is not user
adjustable. See “Viewing and Changing Global Parameters” on page 32 for
instructions to change the setup of relays A, B, C, and D.
Alarm 1 Condition
This section describes the indications for an alarm 1 condition and suggests responses to
an alarm 1 condition.
Alarm 1 Condition Indications
When the gas reading of an active channel reaches the alarm 1 setpoint, the Beacon 410
senses an alarm 1 condition. The Beacon 410 alerts you to an alarm 1 condition as
follows:
•the alarm 1 LED turns on
•the gas reading in alarm 1 condition alternates with the
•the buzzer sounds a pulsing tone
•the common alarm 1 relay (relay A) and relay D energize
ALARM-1 message
Beacon 410 Gas Monitor Operator’s ManualAlarm Indications • 27
NOTE: Relay D is factory set as a common any alarm relay.
•the applicable alarm 1 channel relay energizes
•if installed and set to activate in an alarm 1 condition, the strobe flashes
Responding to an Alarm 1 Condition
This section suggests the following responses to an alarm 1 condition:
1. Follow your established procedure for a low-level combustible or toxic gas condition or
a decreasing oxygen condition.
2. To acknowledge the alarm condition and silence the buzzer while in an alarm 1
condition, press and release the reset switch. The alarm 1 LED will begin to flash
indicating the alarm condition has been acknowledged. You cannot de-energize the
alarm 1 relays until the gas reading falls below (rises above for oxygen) the alarm 1
setpoint.
3. Oxygen alarms are factory set as self-resetting and will automatically clear when the
oxygen reading rises above the alarm 1 setpoint.
4. Alarms for all other gas types are factory set as latching. After the gas reading falls
below the alarm 1 setpoint, press and release the reset switch to reset the alarm 1
circuit. Resetting the alarm 1 circuit silences the buzzer (if the alarm has not been
acknowledged), turns off the alarm 1 LED, resets the display for the channel(s) in
alarm, turns off the strobe if it is installed and set to activate for alarm 1, and deenergizes the common and channel alarm 1 relays and relay D.
Alarm 2 Condition
This section describes the indications for an alarm 2 condition and suggests responses to
an alarm 2 condition.
Alarm 2 Condition Indications
When the gas reading of an active channel reaches the alarm 2 setpoint, the Beacon 410
senses an alarm 2 condition. The Beacon 410 alerts you to an alarm 2 condition as
follows:
•the ALARM 2 LED turns on
•the gas reading during an alarm 2 condition alternates with the
for oxygen) message
•the buzzer sounds a pulsing tone
•the common alarm 2 relay (relay B) energizes
•for an oxygen channel, relay D energizes (it is already energized for non-oxygen
channels)
•the applicable alarm 2 channel relay energizes
•if installed and set to activate for alarm 1, the strobe continues to flash for non-oxygen
channels
•if installed and set to activate for alarm 2 only or for oxygen channels, the strobe
begins to flash when an alarm 2 condition begins
ALM 1,2 (ALARM-2
Responding to an Alarm 2 Condition
This section suggests responses to an alarm 2 condition.
1. Follow your established procedure for a high-level combustible or toxic gas condition,
28 • Alarm IndicationsBeacon 410 Gas Monitor Operator’s Manual
or an increasing oxygen condition.
2. To acknowledge the alarm condition and silence the buzzer while in an alarm 2
condition, press and release the reset switch. The alarm 2 LED will begin to flash
indicating the alarm condition has been acknowledged. You cannot de-energize the
alarm 2 relays until the gas reading falls below the alarm 2 setpoint.
3. Oxygen alarms are factory set as self-resetting and will automatically clear when the
oxygen reading falls below the alarm 2 setpoint.
4. Alarms for all other gas types are factory set as latching. After the gas reading falls
below the alarm 2 setpoint, press and release the reset switch to reset the alarm
circuit. Resetting the alarm circuit silences the buzzer (if the alarm has not been
acknowledged) turns off the alarm 2 LED, resets the display for the channel(s) in
alarm, turns off the strobe if it is installed and set to activate for alarm 2 only or if a
channel is an oxygen channel, and de-energizes the common and channel alarm 2
relays. For oxygen channels, the reset switch also de-energizes
relay D.
Alarm 3 Condition
This section describes the indications for an alarm 3 condition and suggests responses to
an alarm 3 condition.
NOTE: The factory set configuration of the Beacon 410 does not assign channel relay
contacts to the alarm 3 condition. Relays A, B, C, and D are factory set as
common alarm relays, but can be configured as channel alarm 3 relays. See
“Viewing and Changing Global Parameters ” on page 32 and “Viewing and
Changing Channel Parameters” on page 35 for instructions to change relays A,
B, C, and D from their factory setting.
Alarm 3 Condition Indications
When the gas reading of an active channel reaches the alarm 3 setpoint, the Beacon 410
senses an alarm 3 condition. The Beacon 410 alerts you to an alarm 3 condition as
follows:
•the alarm 3 LED turns on;
•the gas reading during an alarm 3 condition continues to flash and alternates with the
ALM 1,2,3 (ALM 1,3 for oxygen) message;
•the buzzer sounds a pulsing tone;
•the common alarm 3 relay (relay C) energizes
•if installed and set to activate for alarm 1 and/or alarm 2 for non-oxygen channels or
alarm 1 for oxygen channels, the strobe continues to flash
•if set to activate for alarm 3 only, it begins to flash when an alarm 3 condition begins
Responding to an Alarm 3 Condition
This section suggests responses to an alarm 2 condition.
1. Follow your established procedure for a high-level combustible or toxic gas condition,
or an increasing oxygen condition.
2. To acknowledge the alarm condition and silence the buzzer while in an alarm 3
condition, press and release the reset switch. The ALARM 3 LED will begin to flash
indicating the alarm condition has been acknowledged. You cannot de-energize the
Beacon 410 Gas Monitor Operator’s ManualAlarm Indications • 29
common alarm 3 relay (relay C) until the gas reading falls below the alarm 3 setpoint.
3. Oxygen alarms are factory set as self-resetting and will automatically clear when the
oxygen reading rises above the alarm 3 setpoint.
4. Alarms for all other gas types are factory set as latching. After the gas reading falls
below the alarm 3 setpoint, press and release the reset switch to reset the alarm
circuit. Resetting the alarm circuit silences the buzzer (if the alarm has not been
acknowledged) turns off the ALARM 3 LED, and de-energizes the common alarm 3
relay.
Fail Condition
This section describes the indications for a fail condition and suggests responses to a fail
condition.
Fail Condition Indications
The Beacon 410 senses a fail condition for any of the following:
•the detector head wiring is disconnected or incorrectly connected;
•the display reading is -10% of full scale or lower;
•the Beacon 410 or one of it’s detector heads is malfunctioning.
When the Beacon 410 senses a fail condition, it alerts you as follows:
•the fail LED turns on;
•the gas reading for the failing channel is replaced by the
•the buzzer sounds a steady tone;
•the common fail relay de-energizes.
NOTE: The fail alarm has a 30 second delay.
NOTE: If you elect to use relays A, B, C, and D as channel fail relays in the Configuration
Menu, the appropriate relay de-energizes in a fail condition. Relays A, B, C, and
D correspond to channels 1, 2, 3, and 4. See “Viewing and Changing Channel
Parameters” on page 35 for a instructions to define channel fail relays.
Responding to a Fail Condition
This section suggests responses to a fail condition.
1. Verify that the detector head wiring is correctly and securely connected.
2. See the troubleshooting guide in “Troubleshooting” on page 64.
FAIL message;
Low DC Power Condition
This section describes the audible and visual indications for a low DC power condition and
suggests responses to a low DC power condition. This condition only occurs when DC
power is used as the primary or a backup power source
Low DC Power Condition Indications
The Beacon 410 senses a DC low power condition when the DC power source is 18.5
volts or less.
30 • Alarm IndicationsBeacon 410 Gas Monitor Operator’s Manual
WARNING:While in a low power condition, the Beacon 410 is not an active ga s m onitor.
When the Beacon 410 senses a low DC power condition, it alerts you as follows:
•The FAIL LED turns on.
•The fail relay and all other relays are de-energized.
•The message
LOW POWER S STANDBY is indicated on the top line of the LCD and
the input voltage is displayed on the bottom line of the LCD.
NOTE: The low DC power alarm cannot be cleared using the reset switch.
When the DC input voltage increases to 19.0 volts, the low DC power alarm is cleared and
the Beacon 410 will begin its warm-up sequence.
Responding to a Low DC Power Condition
•If DC power is the primary power source:
1. For a
temporary DC power source, disconnect primary DC power at the Beacon
410, then connect a 24 VDC battery.
2. Determine and correct the cause of primary DC power loss.
When the DC power source rises above 19.0 volts, the Beacon 410 begins the
warm up process.
3. Verify that the Beacon 410 enters normal operation after its warm-up sequence.
•If DC power is the
backup power source :
1. If a non-rechargeable battery is used for backup, replace the battery.
2. Determine and correct the cause of primary AC power loss. When backup DC or
primary AC power is restored, the Beacon 410 begins the warm up process.
If the Beacon 410 was purchased with the battery charging feature, when AC
power is restored, the Beacon 410 charges the backup battery with a charge
current of approximately 100 mA until it is fully recharged. Charge time varies
depending on the battery size and how much the battery was depleted. Once the
battery is fully charged, the Beacon 410 maintains the charge with the 100 mA
charge current.
3. Verify that the Beacon 410 enters normal operation after its warm-up sequence.
Beacon 410 Gas Monitor Operator’s ManualAlarm Indications • 31
Chapter 5: Global Menu
T
N
PNT /ESC
DELATSNIEBORSO
EBOR
U
LT
l:noitaInstalS
/DNthen E
Overview
This chapter describes how to view and change Beacon 410 instrument parameters using
the Global Menu. The global parameters are the optional strobe’s installation setting, the
relay normal state setting, the usage setting for relays A, B, C, and D, and the buzzer
silence setting. The Global Menu is accessed using the ESCAPE and UP/YES control
buttons.
The Global Menu includes a 5-minute timeout feature. If you do not press a control button
for 5 minutes, the Beacon 410 automatically returns to normal operation.
NOTE: If the Beacon 410 returns to normal operation because of a timeout, it enters a
warm-up period just as it does when it is first turned on.
If you are installing a new Beacon 410, it has been set up at the factory. Use the
Global Menu only if you want to change the instrument parameter factory
settings.
Viewing and Changing Global Parameters
W ARNING:The Beacon 410 is not an active gas monitoring device while in the Global Menu.
When you enter the Global Menu, the 4-20 mA output signal will “freeze” at 3.5
mA for each active channel, all relays and alarm LEDs will remain in their
current state at the time that you entered the Global Menu, and the alarm buzzer
and strobe will turn off if they were activated until you exit the Global Menu. The
4 - 20 mA output signal will not indicate current readings and the relays and
alarm LEDs will not resume operating normally until the Beacon 410 is in
normal operation again.
NOTE: You must set the instrument parameters to the desired settings using the Global
Menu before making any changes to the channel parameters in the Configuration
Menu. Settings in the Global Menu influence what channel parameters appear in
the Configuration Menu.
1. While in Normal Operation, press and hold the ESCAPE and UP/YES buttons for
approximately 5 seconds to enter the Global Menu. Release the buttons when the
Have Entered the GLOBAL Menu
2. Press and release the UP/YES button to continue in the Global Menu to the first global
parameter setting screen, the Strobe Installation Screen.
message appears.
You
32 • OverviewBeacon 410 Gas Monitor Operator’s Manual
3. In the parameter screens, use the ENTER button to accept a displayed parameter
]
ACS
Ssgnitt?
[ktobacup]EP
NY[
e
E
a
ve S
/
setting and proceed to the next parameter or the ESCAPE button to backup a screen.
4. If you want to change a displayed parameter setting, use the UP/YES or DOWN/NO
button to change the parameter to the desired setting, then press and release the
ENTER button to accept the change and proceed to the next screen.
5. Continue through the parameter setting screens changing any parameter settings as
needed until the following message appears on the display.
6. To save the global parameter settings, press and release the UP /YES button.The
screen will indicate that the Beacon 410 is saving the settings for a few seconds then
the Beacon 410 will return to normal operation.
If you do not want to save the global parameter settings, press and release the
DOWN/NO button. The display will indicate that no settings have been changed for a
few seconds then the Beacon 410 will return to normal operation.
Table 6 below lists the global parameters that can be configured in the Global Menu along
with the factory set value for each parameter.
Table 6: G lobal Menu Parameters
Global Parameter
(Factory-Set Value)
STROBE Installation
(Factory setting depends on
whether a strobe is supplied
with the Beacon 410.)
If set to STROBE IS INSTALLED, the strobe circuitry is
functional and items in the Configuration Menu that apply to
the strobe operation become active.
If set to
does not function and items in the Configuration Menu that
apply to the strobe operation do not appear.
NO STROBE INSTALLED, the strobe circuitry
Description
Beacon 410 Gas Monitor Operator’s ManualViewing and Changing Global Parameters • 33
Table 6: Global Menu Parameters (Continued)
Global Parameter
(Factory-Set Value)
NORMAL RELAY STATES
(NORM DE-ENERGIZED)
RELA YS ABCD used as
(COMMON ALARM
RELAYS)
Description
If set to NORM DE-ENERGIZED, the gas alarm relays are
de-energized in normal operation and energize when the
appropriate alarm circuit is activated. The NO (normally
open) relay contacts are open during non-alarm operation
and close when the appropriate alarm condition occurs.
The NC (normally closed) relay contacts are closed during
non-alarm operation and open when the appropriate alarm
condition occurs.
If set to
NORMALLY ENERGIZED, the gas alarm relays
are energized in normal operation and de- en er giz e when
the appropriate alarm circuit is activated. The NO (normally
open) relay contacts are closed during non-alarm operation
and open when the appropriate alarm condition occurs. The
NC (normally closed) relay contacts are open during nonalarm operation and close when the appropriate alarm
condition occurs.
NOTE: The fail relay is factory set as normally energized
and is not field adjustable.
If set to
COMMON ALAR M R EL AYS, relays A, B, C, and D
are defined as common alarms assigned to the gas alarms
as follows:
Relay A: Common Alarm 1
Relay B: Common Alarm 2
Relay C: Common Alarm 3
Relay D: Common Any alarm
If set to
CHANNEL ALARM RELA YS, rela ys A, B, C, and D
are defined as configurable channel alarm relays assigned
to the detection channels as follows:
Relay A: Channel 1
Relay B: Channel 2
Relay C: Channel 3
Relay D: Channel 4
Each relay is factory assigned to alarm 3, but this can be
changed in the Configuration Menu.
BUZZER SILENCE
(CAN SILENCE BUZZER)
If set to
CAN SILENCE BUZZER, the buzzer can by
silenced by pressing and releasing the reset switch when
the Beacon 410 is in a gas alarm condition.
If set to
CAN”T SILENCE BUZZEZR, the buzzer cannot be
silenced by the reset switch while the Beacon 410 is in a
gas alarm condition.
34 • Viewing and Changing Global ParametersBeacon 410 Gas Monitor Operator’s Manual
Chapter 6: Configuration Menu
el
UeP
1
CE
S
/DNthenNT/ES
lennahCtc
Overview
This chapter describes how to view and change Beacon 410 channel parameters
including alarm point settings and channel relay settings using the Configuration Menu.
The Configuration Menu is accessed using the ESCAPE and ENTER control buttons.
The Configuration Menu includes a 5-minute timeout feature. If you do not press a control
button for 5 minutes, the Beacon 410 automatically returns to normal operation.
NOTE: If the Beacon 410 returns to normal operation because of a timeout, it enters a
warm-up period just as it does when it is first turned on.
If you are installing a new Beacon 410, it has been set up at the factory. Use
Configuration Mode only if you want to change one or more channel’s parameter
factory settings. If you want to change a channel’s detector type see “Chapter 7:
Input Setup Menu” on page 41.
Viewing and Changing Channel
Parameters
WARNING:The Beacon 410 is not an active gas monitoring device while in the
Configuration Menu. When you enter the Configuration Menu, the 4-20 mA
output signal will “freeze” at 3.5 mA for each active channel, all relays and
alarm LEDs will remain in their current state at the time that you entered the
Configuration Menu, and the alarm buzzer and strobe will turn off if they were
activated until you exit the Configuration Menu. The 4 - 20 mA output signal will
not indicate current readings and the relays and alarm LEDs will not resume
operating normally until the Beacon 410 is in normal operation again.
1. While in normal operation, simultaneously press and hold the ESCAPE and ENTER
buttons for approximately 5 seconds to enter the Configuration Menu. Release the
buttons when the
the display.
2. Press and release the UP/YES button to continue in the Configuration Menu to the
Select Channel Screen.
You Have Entered the CONFIGURATION Menu message appears on
3. Use the UP/YES or DOWN/NO buttons to select the channel whose parameters you
Beacon 410 Gas Monitor Operator’s ManualOverview • 35
want to view and/or update, then press and release the ENTER button. The first
channel parameter screen appears and it allows you to select whether the channel is
set to
ACTIVE, STANDBY, or, NOT USED.
4. Press and release the UP /YES or DOWN/NO button until the setting you want to
]
ACS
Ssgnitt?
[ktobacup]EP
NY[
e
E
a
ve S
/
D
D
[
o
o
Y]/N
MoreChannel s?
oTtnaWuoY
select appears, then press and release the ENTER button to save it. The next channel
parameter screen appears.
NOTE: If you selected STANDBY or NOT USED, the display will ask you if you still want
to review the channel settings. If yes, press and release the UP/YES button and
continue with step 5. If no, press and release the DOWN/NO button and proceed
to step 9.
NOTE: Use the ESCAPE button to go backwards in the menu.
5. In the parameter screens, use the ENTER button to accept a displayed parameter
setting and proceed to the next parameter or the ESCAPE button to backup a screen.
6. If you want to change a displayed parameter setting, use the UP/YES or DOWN/NO
button to change the parameter to the desired setting, then press and release the
ENTER button to accept the change and proceed to the next screen.
7. Continue through the parameter setting screens changing any parameter settings as
needed until the following message appears on the display.
8. To save the channel parameter settings, press and release the UP /YES button.The
screen will indicate that the Beacon 410 is saving the settings for a few seconds.
If you do not want to save the channel parameter settings, press and release the
DOWN/NO button. The display will indicate that no settings have been changed for a
few seconds and the channel parameters will return to their original settings.
9. The following screen will appear asking if you want to configure more channels.
If yes, press and release the UP/YES button to return to the Select Channel Screen.
Begin at instruction step 3 above to configure another channel.
If no, then press the DOWN /NO button to return to normal operation.
36 • Viewing and Changing Channel ParametersBeacon 410 Gas Monitor Operator’s Manual
Table 7 below lists the channel parameters that can be viewed and updated in the
Configuration Menu along with the factory set value for each parameter.
Table 7: Configuration Menu Parameters
Channel Parameter
(Factory-Set Value)
Channel Status
(Factory setting depends on
whether a detector is
supplied for a channel.)
Strobe Setting
(USE w/ Alarms Only)
Description
If set to ACTIVE, the channel will be fully functional.
If set to
will be replaced by
STANDBY, the channel’s gas name and reading
STANDBY on the display. The channel’s
relays will be fixed in their non-alarm state and the 4 - 20
mA output for the channel will be fixed at 3.5 mA. This
setting is typically used when maintenance is being
performed on a channel.
If set to
gas name and reading field for that channel indicates
USED
NOT USED, the channel is not functional and the
NOT
. All the channel’s relays are de-energized and the 4 -
20 mA output is fixed at about 0.7 mA.
If set to USE w/ Alarms Only, the strobe will activate when
the appropriate gas alarm condition occurs as defined in
the alarm strobe settings (see below) in the Configuration
Menu. It will not operate when a fail alarm occurs.
If set to
USE w/ Alarms and Fail, the strobe will activate
when the appropriate gas alarm condition occurs and when
a fail alarm occurs.
If set to
DON”T USE, the none of the strobe alarm settings
appear in the Configuration menu and the strobe circuit
does not function.
*This parameter appears only if the STROBE Installation
parameter in the Global Menu is set to Strobe Installed.
Alarm-1 Level
(See the Beacon 410
Detector Head Specification
sheet for the detector head
installed on this channel.)
Alarm-1 Inc/Dec
(DECREASING for oxygen,
INCREASING for all other
channel types)
Alarm-1 Latching
(SELF-RESETTING for
oxygen, LATCHING for all
other channel types)
Alarm-1 ON Delay
(1 Second)
ALARM-1 OFF DELAY
(0 Seconds)
The gas reading at which an alarm 1 condition occurs and
the Beacon 410 activates the alarm 1 circuit for this
channel.
Indicates if the alarm 1 circuit is activated by gas reading
INCREASING or DECREASING to the alarm 1 level.
If set as
LATCHING, you must press the RESET button to
reset the alarm 1 circuit after the alarm 1 condition passes.
If set as
SELF-RESETTING, the Beacon 410 automatically
resets the alarm 1 circuit after the alarm 1 condition passes.
The amount of time the Beacon 410 delays activation of the
alarm 1 circuit after the gas reading reaches the alarm 1
setpoint.
The amount of time the Beacon 410 delays resetting the
alarm 1 circuit after the alarm 1 condition passes.
*This parameter appears only if the ALARM-1 Latching
parameter is set to SELF-RESETTING.
Beacon 410 Gas Monitor Operator’s ManualViewing and Changing Channel Parameters • 37
Table 7: Configuration Menu Parameters (Continued)
Channel Parameter
(Factory-Set Value)
ALM-1 STROBE Sett ing
(NonResettable STROBE)
Alarm-2 Level
(See the Beacon 410
Detector Head Specification
sheet for the detector head
installed on this channel.)
Alarm-2 Inc/Dec
(INCREASING)
Alarm-2 Latching
(SELF-RESETTING for
oxygen, LATCHING for all
other channel types)
Description
If set to Resettable STROBE, the strobe can be turned off
with the Reset Switch while an alarm 1 condition still exists.
If set to
NonResettable STROBE, the strobe cannot be
turned off with the Reset Switch while an alarm 1 condition
still exists.
If set to
NO STROBE, the strobe will not activate due to an
alarm 1 condition.
*This parameter appears only if the STROBE Installation
parameter in the Global Menu is set to Strobe Installed
and the Strobe Setting parameter in the Configuration
Menu is set to USE w/ Alarms Only or USE w/ Alarms and Fail.
The gas reading at which an alarm 2 condition occurs and
the Beacon 410 activates the alarm 2 circuit for this
channel.
Indicates if the alarm 2 condition is activated by gas
readings
INCREASING or DECREASING to the alarm 2
level.
If set as
LATCHING, you must press the RESET button to
reset the alarm 2 circuit after the alarm 2 condition passes.
If set as
SELF-RESETTING, the Beacon 410 automatically
resets the alarm 2 circuit after the alarm 2 condition passes.
Alarm-2 ON Delay
(120 Seconds for oxygen,
1 Second for all other
channel types)
Alarm-2 OFF DELAY
(0 Seconds)
ALM-2 STROBE Sett ing
(NonResettable STROBE)
The amount of time the Beacon 410 delays activation of the
alarm 2 circuit after the gas reading reaches the alarm 2
setpoint.
The amount of time the Beacon 410 delays resetting the
alarm 2 circuit after the alarm 2 condition passes.
*This parameter appears only if the ALARM-2 Latching
parameter is set to SELF-RESETTING.
If set to Resettable STROBE, the strobe can be turned off
with the Reset Switch while an alarm 2 condition still exists.
If set to
NonResettable STROBE, the strobe cannot be
turned off with the Reset Switch while an alarm 2 condition
still exists.
If set to
NO STROBE, the strobe will not activate due to
an alarm 2 condition.
*This parameter appears only if the STROBE Installation
parameter in the Global Menu is set to Strobe Installed
and the Strobe Setting parameter in the Configuration
Mode is set to USE w/ Alarms Only or USE w/ Alarms and
Fail.
38 • Viewing and Changing Channel ParametersBeacon 410 Gas Monitor Operator’s Manual
Table 7: Configuration Menu Parameters (Continued)
Channel Parameter
(Factory-Set Value)
Alarm-3 Level
(See the Beacon 410
Detector Head Specification
sheet for the detector head
installed on this channel.)
Alarm-3 Inc/Dec
(DECREASING for oxygen,
INCREASING for all other
channel types)
Alarm-3 Latching
(SELF-RESETTING for
oxygen, LATCHING for all
other channel types)
Alarm-3 ON Delay
(1 Second)
Alarm-3 OFF DELAY
(0 Seconds)
Description
The gas reading at which an alarm 3 condition occurs and
the Beacon 410 activates the alarm 3 circuit for this
channel.
Indicates if the alarm 3 circuit is activated by gas readings
INCREASING or DECREASING to the alarm 3 level.
If set as
LATCHING, you must press the RESET button to
reset the alarm 3 circuit after the alarm 3 condition passes.
If set as
SELF-RESETTING, the Beacon 410 automatically
resets the alarm 3 circuit after the alarm 3 condition passes.
The amount of time the Beacon 410 delays activation of the
alarm 3 circuit after the gas reading reaches the alarm 3
setpoint.
The amount of time the Beacon 410 delays resetting the
alarm 3 circuit after the alarm 3 condition passes.
*This parameter appears only if the ALARM-3 Latching
parameter is set to SELF-RESETTING.
ALM-3 STROBE Sett ing
(NonResettable STROBE)
Relay-1 Assignment
(ALARM-1)
If set to Resettable STROBE, the strobe can be turned off
with the Reset Switch while an alarm 3 condition still exists.
If set to
NonResettable STROBE, the strobe cannot be
turned off with the Reset Switch while an alarm 3 condition
still exists.
If set to
NO STROBE, the strobe will not activate due to
an alarm 3 condition.
*This parameter appears only if the STROBE Installation
parameter in the Global Menu is set to Strobe Installed
and the Strobe Setting parameter in the Configuration
Mode is set to USE w/ Alarms Only or USE w/ Alarms and Fail.
The alarm condition that activates relay 1. It can be set for
activation by the following alarm conditions:
•ALARM-1
•ALARM-2
•ALARM-3
•ALARMS 1 or 2
•ALARMS 1 or 3
•ALARMS 2 or 3
•ALARMS 1, 2, or 3
•Use As Channel FAIL (If you choose this setting, the relay is
normally energized)
Beacon 410 Gas Monitor Operator’s ManualViewing and Changing Channel Parameters • 39
Table 7: Configuration Menu Parameters (Continued)
Channel Parameter
(Factory-Set Value)
Relay-2 Assignment
(ALARM-2)
Relay-X Assignment
(ALARM-3)
*Relay-X will be either
Relay-A for channel 1,
Relay-B for channel 2,
Relay-C for channel 3, or
Relay-D for channel 4.
Description
The alarm condition that activates relay 2. It can be set for
activation by the following alarm conditions:
•ALARM-1
•ALARM-2
•ALARM-3
•ALARMS 1 or 2
•ALARMS 1 or 3
•ALARMS 2 or 3
•ALARMS 1, 2, or 3
•Use As Channel FAIL (If you choose this setting, the relay is
normally energized)
The alarm condition that activates relay X. It can be set for
activation by the following alarm conditions:
•ALARM-1
•ALARM-2
•ALARM-3
•ALARMS 1 or 2
•ALARMS 1 or 3
•ALARMS 2 or 3
•ALARMS 1, 2, or 3
•Use As Channel FAIL (If you choose this setting, the relay is
normally energized)
NOISE FILTER
(5 seconds)
Deadband
(0.5%
oxygen for oxygen
channels; 2.0% of
Full Scale
in terms of the channel’s
units for all other channel
types.)
*This parameter appears on ly if the R ELAYS ABCD used as
parameter in the Global Menu is set to CHANNEL ALARM RELAYS.
The noise filter feature helps “smooth out” jumpy or noisy
signals from the detector head. The filter setting indicates
the number of seconds over which the displayed reading is
being averaged. You can set the noise filter from1 second
to 10 seconds.
A setting of 10 seconds produces the greatest amount of
smoothing, but causes the displayed gas reading to
respond more slowly to changes in gas concentrations.
A setting of 1 second gives the fastest gas reading
response to a change in gas concentration, but gives the
least amount of smoothing.
The deadband setting defines the first gas concentration
that will be displayed when the gas reading increases or
decreases from the normal fresh air concentration (20.9 for
oxygen and 0 for all other channel types). This feature
helps prevent “jumpy” readings near the normal fresh air
reading.
For example, if the zero suppression setting is
2 %LEL for
a combustible channel whose range is 0 - 100% LEL, the
Beacon 410 will display a reading of 0 %LEL for gas
readings from
-1 %LEL to 1 %LEL.
40 • Viewing and Changing Channel ParametersBeacon 410 Gas Monitor Operator’s Manual
Chapter 7: Input Setup Menu
u
ehTderetnEevaHu
YCo
oInt i nue?[Y/N]
NPUTSETUPMen
el
UeP
1
CE
S
/DNthenNT/ES
lennahCtc
Overview
This chapter describes how to use the Input Setup Menu to select each channel’s detector
head input type and the gas setup. The detector head input type determines whether a
direct connect type or a 4 - 20 mA type of detector head will be used and the gas setup
determines the target gas, detection units, and the detection range full scale. The Input
Setup Menu is accessed using the ENTER control button when powering up the Beacon
410.
The Input Setup Menu includes a 5-minute timeout feature. If you do not press a button for
5 minutes, the Beacon 410 automatically returns to normal operation.
NOTE: The Beacon 410 will begin its warm-up sequence when it exits the Input Setup
Menu before it enters normal operation whether because of a program timeout or
because you are done with input setup and are exiting manually .
Selecting the Detector Head Input
Type and Gas Setup
NOTE: The Beacon 410 must be turned off before you can enter the Input Setup Menu. If
your Beacon 410 is already installed and operating, make sure that turning the
Beacon 410 off will not cause unwanted alarms before continuing.
1. While the Beacon 410 is off, press and hold the ENTER button, then turn on the
Beacon 410 with the ON/OFF switch.
2. Continue to press the ENTER button while the Beacon 410 turns on and the display
shows
screen appears on the display.
3. If you want to exit the Input Menu, press and release the DOWN/NO button. The
Beacon 410 will exit the Input Menu and begin the warm-up sequence.
If you want to continue in the Input Menu, press and release the UP/YES button. The
Select Channel(s) screen will appear.
RKI Instruments Beacon 410. Release the ENTER button when the following
Beacon 410 Gas Monitor Operator’s ManualOverview • 41
4. The setup that you perform will apply to all the channels that you select. Select the
TNEnehtND/
ANNEL (S) :1 , 2
eyTTUP
CSE/
RETTIMNARTAm02-4S
p
P
N
H
U
I
C
H
r
P
0METHAE1 0%LEL
T/ESC
e
2,1:)S(LENNAC
P
U
N
t
eDdenfiputeSaGs
/DNhenEN
channel or channels you want to setup by using the UP/YES or DOWN/NO buttons to
scroll through the choices and then press and release the ENTER button to accept the
selection. The INPUT Type screen will appear. In the example below, channels 1 and
2 have been selected for setup.
5. Use the UP/YES or DOWN/NO button to scroll through the choices of input types.
Table 8 below lists the choices of input type for the Beacon 410.
Table 8: Beacon 410 Detector Head Input Types
Detector Head Input TypeDescription
4-20 mA TRANSMITTERA 4-20 mA TRANSMITTER detector head is connected
to the Beacon 410 with 2 or 3 wires, depending on the
detector head model, using the AMP +, S, and - terminals
on the detector/transmitter terminal strips. All calibration
adjustments are made at the detector head.
PREAMP DIRECTA
Beacon 410 with 2 wires using the AMP + and S
terminals on the detector/transmitter terminal strips. All
calibration adjustments are made at the Beacon 410.
OXYGEN DIRECT (25%)An
detector head in which the oxygen detector is wired to the
Beacon 410 with 2 wires using the OXY W (white) and G
(green) terminals from on the detector/transmitter
terminal strips. All calibration adjustments are made at
the Beacon 410.
RWGB DIRECT (LEL)An
in which the detector is wired to the Beacon 410 with 4
wires using the LEL R (red), W (white), G (green) and B
(black) terminals on the detector/transmitter terminal
strips. All calibration adjustments are made at the
Beacon 410. This type of detector head is typically a
combustible gas LEL detector head, but may be another
type such as a CO
6. When the desired detector head input type is shown on the display, press and release
the ENTER button to accept it. The PreDefined Gas Setup screen will appear.
PREAMP DIRECT detector head is connected to the
OXYGEN DIRECT detector head is an oxygen
RWGB DIRECT (LEL) detector head is detector head
detector.
2
42 • Selecting the Detector Head Input Type and Gas SetupBeacon 410 Gas Monitor Operator’s Manual
7. Use UP/YES or DOWN/NO button to scroll through the gas setup choices. There are
several choices whose gas name, full scale, and detection units are predefined. For
these choices, the gas name, full scale, and detection units are displayed. There is
also one choice named
CUSTOM that allows you to enter the gas name, full scale,
and detection units.
8. When the desired setup choice is shown on the display, press and release the ENTER
button to accept it.
If you selected a predefined gas setup, the Save Settings? screen will appear.
Proceed to step 10.
If you selected the
CUSTOM choice, the Gas Name screen will appear. Proceed to
step 9.
NOTE: The CUSTOM choice is especially useful if you are connecting a 4 - 20 mA
transmitter to the Beacon 410 that is not a gas detector, such as a temperature or
pressure transducer.
9. If you selected the
CUSTOM choice, perform the following steps to define the gas
name, detection units, and full scale.
•With the Gas Name screen displayed, use the UP/YES or DOWN/NO button to
scroll through the gas name choices. The list includes one choice called
Spell It.
This choice allows you to enter a name that is not on the list.
•When the desired choice is displayed, press and release the ENTER button.
If you selected a gas name, the Gas Units screen appears.
If you select ed
Spelled As:
Spell It, a screen appears that indicates Currentl y, Gas Name Is
. If the displayed name is OK, press and release the UP/YES button
and the Gas Units screen appears
If you want to change the displayed name, perform the following steps to enter the
desired gas name.
•Press and release the DOWN/NO button. The top display line indicates
Out Gas Name
.
Spell
•An eight character wide text field appears on the left side of the third display
line enclosed by single quotes. A vertical cursor appears below the far left text
position indicating that this character position can be edited. Use the UP/YES
or DOWN/NO button to scroll through the character choices. When the
desired character is displayed, press and release the ENTER button to save
that character and move to the next position. Use the ESCAPE button to
return to a previous character if you want to change it.
•Repeat the previous step until all the text positions are saved. After the last
character is saved, the Units screen appears.
•Use the UP/YES or DOWN/NO button to scroll through the units choices. When
the desired units are displayed, press and release the ENTER button. The Full
Scale Readout screen appears.
•USE the UP/YES or DOWN/NO button to scroll through the available choices for
the full scale setting. When the desired full scale setting is displayed, press and
release the ENTER button. The Save Settings? screen appears.
Beacon 410 Gas Monitor Operator’s ManualSelecting the Detector Head Inpu t Type and Gas Setup • 43
10. If you want to save the setup, press and release the UP/YES button. The display will
indicate that the settings have been saved before asking if you want to setup more
channels.
NOTE: If you defined a custom setup by selecting the CUSTOM choice in the
PreDefined Gas Setup screen, you will be required to review the channel
parameters from the Configuration Menu. See “Viewing and Changing Channel
Parameters” on page 35 for a description of the channel parameters.
If you do not want to save the setup, press and release the DOWN/NO button. Y ou will
return to the Select Channel(s) screen. Follow the instructions from step 4 above.
11. If you want to setup more channels, press and release the UP/YES button. Y ou will
return to the Select Channel(s) screen. Follow the instructions from step 4 above.
If you do not want to setup more channels, press and release the DOWN/NO button
and the Beacon 410 will exit the Input Setup Menu and begin its warm-up sequence.
44 • Selecting the Detector Head Input Type and Gas SetupBeacon 410 Gas Monitor Operator’s Manual
Chapter 8: Calibration Mode
Overview
This chapter describes how to calibrate the Beacon 410’s active channels using
Calibration Mode. In Calibration Mode, you can:
•Set the calibration timeout.
•Perform a fresh air adjust on selected channels.
•Perform a gas adjust on selected channels.
•View Max Spans.
Table 9 below briefly describes the four operations that can be performed in Calibration
Mode. A more detailed description of each operation is given later in this chapter.
Table 9: Calibration Mode Functions
OperationDescription
SET CAL TIMEOUTThe calibration timeout sets the amount of time after the
PERFORM AIR ADJUSTThis operation allows you to set the fresh air reading of
PERFORM GAS ADJUSTThis operation allows you to set the response reading of
VIEW MAX SPANSThis operation gives a measure of how much gas
Calibration Frequency
Although there is no particular calibration frequency that is correct for all applications, a
calibration frequency of every 3 to 6 months is adequate for most Beacon 410
applications. Unless experience in a particular application dictates otherwise, RKI
Instruments, Inc. recommends a calibration frequency of every 3 months.
last button push while in Calibration Mode that the
Beacon 410 will wait before automatically returning to
normal operation.
the selected channel(s) to zero (20.9% for an oxygen
channel).
the selected channels with calibration gas.
response adjustment remains on a direct connect type
channel by allowing you to view the maximum possible
adjustment on gas during the most recent calibration for
each channel.
If an application is not very demanding, for example detection in a clean, temperature
controlled environment where a combustible or toxic gas is not normally present, and
calibration adjustments are minimal at calibration, then a calibration frequency of every 6
months may be adequate.
If an application is very demanding, for example if a combustible or toxic gas is present
often and in significant concentrations or the environment is not well controlled, then more
frequent calibration than every 3 months may be necessary. For combustible gas
Beacon 410 Gas Monitor Operator’s ManualOverview • 45
detection using catalytic combustion type detectors, if potential catalyst poisons are
known or likely to be present, more frequent calibration than every 3 months will be
necessary.
Detector Head Types
The Beacon 410 can support both direct connect (internal amp) and 4-20 mA transmitter
(remote amp) detector heads. Since the Beacon 410 can support both types of detector
heads, there are three possible detector head combinations when calibrating the active
channels:
•Direct connect detector heads only.
If all the active channels are direct connect detector heads, then all calibration
adjustments are made at the Beacon 410 after calibration gas is applied to the
detectors.
•4-20 mA transmitter detector heads only.
If all the active channels are 4 - 20 mA transmitter detector heads, then all calibration
adjustments are made at the detector head when calibration gas is applied to the
detectors.
•Direct connect and 4-20 mA transmitter detector heads.
If both direct connect and 4-20 mA transmitter detector heads are installed, then
calibration adjustments must be made at the Beacon 410 for the direct connect
detector heads after applying gas to the detectors, and at the detector heads when
gas is applied for the 4-20 mA transmitter detector heads.
Calibration Gas Response Memory
Feature
When a direct connect detector head is installed, the Beacon 410 has the capability to
“remember” the detector’s response to the calibration gas after the gas is removed from
the detector during the fresh air or span (zero for an oxygen detector) adjustment
procedure. This feature enables one person to perform calibration if the detector heads
are mounted remotely from the Beacon 410. When zero air is applied to a detector head
during a fresh air adjustment in Calibration Mode, the Beacon 410 will freeze the display
reading at the lowest (highest for an oxygen detector head) response and the Beacon 410
will continue to display this reading and retain it in its memory until the fresh air adjustment
procedure is completed. When calibration gas is applied to a detector head during a span
(zero for an oxygen detector) adjustment in Calibration Mode, the Beacon 410 will freeze
the display reading at the highest (lowest for an oxygen detector) response to the
calibration gas. The calibration gas can then be removed and the Beacon 410 will
continue to display this reading and retain it in its memory until the span (zero for an
oxygen detector) adjustment procedure is completed.
46 • Detector Head TypesBeacon 410 Gas Monitor Operator’s Manual
Entering Calibration Mode
ALMEOUT
SE/TNEnehtND/
ICT
CPeU
S
lect Operation
SET
WARNING:The Beacon 410 is not an active gas monitoring device during the ca libration
procedure. When you enter Calibration Mode, the 4-20 mA output signal will
“freeze” at 3.5 mA for each active channel, all relays and alarm LEDs will
remain in their current state at the time that you entered Calibration Mode, and
the alarm buzzer and strobe will turn off if they were activated until you exi t
Calibration Mode. The 4 - 20 mA output signal will not indicate current readings
and the relays and alarm LEDs will not resume oper ating normally until the
Beacon 410 is in normal operation again.
1. Assemble the calibration kit(s). See the instruction manual for each detector head for
procedures specific to that detector head.
2. While in normal operation, simultaneously press and hold the ENTER and UP/YES
buttons for approximately 3 seconds to enter Calibration Mode. Release the buttons
when the
asks if you want to continue in the Calibration Menu.
3. If you want to exit Calibration Mode, press and release the DOWN/NO button. The
Beacon 410 will return to normal operation.
You Have Entered The CALIBRATION Mode... message displays and
If you want to continue in Calibration Mode, press and release the UP/YES button to
continue and display the Select Operation screen. The SET CAL TIMEOUT item will
be displayed.
The next two sections describe how to set the calibration timeout and calibrate the
Beacon 410’s detector heads.
Calibration Timeout Setting
The calibration timeout setting defines the amount of time after the last button push that
the Beacon 410 will remain in Calibration Mode before automatically exiting Calibration
Mode and entering normal operation. If the Beacon 410 exits Calibration Mode because of
the automatic timeout, it will begin it’s warm-up sequence just as it does when it is turned
on.
To view or change the calibration timeout setting, do the following:
1. Navigate to the Select Operation Screen as described in the previous section.
Beacon 410 Gas Monitor Operator’s ManualEntering Calibration Mo de • 47
2. With SET CAL TIMEOUT displayed, press and release the ENTER button. The
1
emiTnoitrbil
CUa
PC
outa
S
n5ut
/DNthenENT/E
seim
/
S
EC
1
U
lectChane
SE/TNnehtND
sutEdAjIorFR SH A RF
s()enl
P
e
Calibration Timeout Screen appears.
3. Use the UP/YES or DOWN/NO button to adjust the calibration timeout to the desired
setting. The calibration timeout can be set from 5 minutes to 240 minutes (4 hours) in
5 minute increments. The factory set calibration timeout is 15 minutes.
4. If you want to save the displayed calibration timeout setting, press and release the
ENTER button. The display will indicate that the setting has been saved before
returning to the Select Operation Screen.
If you do not want to save the displayed calibration timeout setting, press and release
the ESCAPE button. The display will indicate that nothing has changed before
returning to the Select Operation Screen.
5. Press and release the ESCAPE button.The
Mode...
6. Press the DOWN/NO button to exit Calibration Mode and enter normal operation.
message displays and asks if you want to continue in Calibration Mode.
Performing a Calibration
It is possible to perform only an air adjustment or a gas response adjustment since it is not
always necessary to perform a complete ca libration. For example, during startup, it is
typically only necessary to perform an air adjustment since the Beacon 410 is shipped
with all the factory setup channels calibrated. However, for a complete calibration, you
must perform both a fresh air adjustment and a gas response adjustment.
Performing a Fresh Air Adjustment
Performing a fresh air adjustment sets the gas reading in fresh air for the selected
channels to zero for all channel types except oxygen. An oxygen channel is set to 20.9%
in fresh air. If you suspect that the environment at a detector head location is not a fresh
air environment, an environment of normal oxygen content, 20.9%, and free of
combustible or toxic gasses, you will need to apply zero-emission air, also known as zero
air, to the detector when making the fresh air adjustment. See the operator’s manual for
each detector head that you are air adjusting for instructions specific to that detector head.
To perform a fresh air adjustment for the selected channels, do the following:
1. Navigate to the Select Operation Screen as described in “Entering Calibration Mode”
on page 47.
You Have Entered The CALIBRATION
2. Use the UP/YES or DOWN/NO button to display
Select Operation Screen.
3. Press and release the ENTER button. The Select Channel(s) screen appears.
48 • Performing a CalibrationBeacon 410 Gas Monitor Operator’s Manual
PERFORM AIR ADJUST on the
4. Use the UP/YES or DOWN/NO button to scroll through the channel combination
AotEPACSE
H
E
tceteDsaG
ENTRwhenDone
trob>
>
o
p
-
-
T
A
plyFRESAIR
sro
:
OL
LEL%0ENAHTEM
1
2
3
4
:
:
:
CO0p pm
mpp0S2H
OXYGEN20 . 9%V
L
MAtLAC
H2S0ppm
mpp0OC
Pa
:
:
:
:
4
3
2
1
METHANE0%LE
choices and select the channel or channels that you will air adjust.
5. Press and release the ENTER button to accept the displayed choice and continue
with the air adjustment. The display will alternate between the top two screens shown
below. In the example below, channels 1, 2, 3, and 4 have been selected.for an air
adjust.
If any of the displayed channels is a 4 - 20 mA type detector head, the gas name and
concentration will alternate with
4 - 20 mA type channel.
6. Verify that the detector heads are in a fresh air environment. If you cannot confirm a
fresh air environment, use a zero air cylinder to introduce fresh-air to each detector
when adjusting the fresh air reading below. If you are air adjusting both 4 - 20 mA and
direct connect types of detector heads, air adjust the 4 - 20 mA type detector heads at
the detector heads first, then apply zero air to and air adjust the direct connect type
detector heads to avoid exiting the air adjustment screens prematurely before you
have adjusted the 4 - 20 mA type detector heads.
7. For a 4 - 20 mA type detector head, adjust the detector head’s fresh air reading
(sometimes referred to as the zero reading for non-oxygen detector heads) at the
detector head. See the detector head operator’s manual for instructions on how to
adjust the fresh air reading.
8. If you are only air adjusting 4 - 20 mA type detector heads, press the ENTER button
when you are done to return to the Select Operation Screen. Proceed to step 10.
If you are air adjusting both 4 - 20 mA type detector heads and direct connect detector
heads, proceed to step 9.
9. For direct connect type detector heads, if the detector heads are in a fresh air
environment, press and release the ENTER button. The Beacon 410 will perform a
fresh air adjustment and the display will indicate
direct connect type detector head if the air adjust is successful. If the air adjust was
CAL at AMP. In the example above, channel 2 is a
AIR ADJUST PASSED for each
Beacon 410 Gas Monitor Operator’s ManualPerforming a Calibration • 49
not successful for any channel, the display will indicate AIR ADJUST FAILED for that
roL
ahCtcel
UeP
lne)(s
FSGACjAdtusA
1
CE
S
/DNthenNT/ES
en
htND/
esUeBoTSAGL
THANE50%LEL
EneCSE/TNP
MCE
A
H
U
C
ANNEL1
d
channel. The Beacon 410 will then return to the Select Operation screen.
If you suspect the detector area is not a fresh air environment, apply zero air to each
selected detector before pressing the ENTER button to perform the fresh air
adjustment. See each detector head’s operator’s manual for instructions to apply zero
air to each detector. The Beacon 410 will freeze the display reading at the lowest level
reached while applying zero air (highest level for an oxygen detector head). After
applying zero air for the required amount of time, usually two minutes, remove the
zero air source from the detector. Press and release the ENTER button at the Beacon
410. The Beacon 410 will perform a fresh air adjustment and the display will indicate
AIR ADJUST PASSED for each direct connect type detector head if the air adjust is
successful. The Beacon 410 will then return to the Select Operation screen.
10. If you want to perform a gas adjustment to complete the calibration, proceed to the
next section, “Performing a Gas Adjustment”.
If you do not want to perform a gas adjustment, press and release the ESCAPE button
to return to the initial screen that says
and then press and release the DOWN/NO button to exit Calibration Mode and return
to normal operation.
You Have Entered CALIBRATION Mode...
Performing a Gas Adjustment
Performing a gas adjustment sets the response level to calibration gas for the selected
channels. This is also known as a span adjustment. See the operator’s manual for each
detector head that you are gas adjusting for instructions specific to that detector head.
To perform a gas adjustment for the selected channels, do the following:
1. Navigate to the Select Operation Screen as described in “Entering Calibration Mode”
on page 47 or proceed to the next step if you just completed an air adjustment and are
at the Select Operation Screen.
2. Use the UP/YES or DOWN/NO button to display
Select Operation Screen.
3. Press and release the ENTER button. The Select Channel(s) screen appears.
4. Use the UP/YES or DOWN/NO button to scroll through the channel combination
choices and select the channel or channels that you will air adjust. In the gas
adjustment example below, channels 1, 2, 3, and 4 have been selected.
5. Press and release the ENTER button to accept the displayed choice and continue
with the gas adjustment. The display will show the calibration gas concentration that
the Beacon 410 expects you to use for the gas adjustment for the first selected direct
connect type channel.
PERFORM GAS ADJUST on the
50 • Performing a CalibrationBeacon 410 Gas Monitor Operator’s Manual
6. Use the UP/YES and DOWN;/NO buttons to make the displayed concentration match
A
T
-
-
p
o
>
>
RTNE
GasDe t ec t or s
LACylpSA
E
G
ESCAPEt oAbor t
enoDnehw
:
OL
LEL%0ENAHTEM
1
2
3
4
:
:
:
CO0p pm
mpp0S2H
OXYGEN2 0 . 9%V
L
MAtLAC
H2S0ppm
mpp0OC
Pa
:
:
:
:
4
3
2
1
METHANE0%LE
the calibration gas concentration, then press and release the ENTER button to accept
it. The display will show the expected calibration gas concentration for the next
selected direct connect type channel.
7. Repeat step 6 for each selected direct connect channel. When the last selected direct
connect channel’s calibration gas concentration has been accepted, the top two
screens shown below will alternate.
If any of the displayed channels is a 4 - 20 mA type detector head, the gas name and
concentration will alternate with
4 - 20 mA type channel
8. For a 4 - 20 mA type detector head, adjust the detector head’s gas response at the
detector head. See the detector head operator’s manual for instructions on how to
adjust the gas response reading. This is sometimes referred to as adjusting the span
(or zero for an oxygen channel).
9. If you are only gas adjusting 4 - 20 mA type detector heads, press the ENTER button
when you are done to return to the Select Operation Screen. Proceed to step 12.
If you are air adjusting both 4 - 20 mA type detector heads and direct connect detector
heads, proceed to step 10.
10. For direct connect type detector heads, apply calibration gas to each selected
detector. See each detector head’s operator’s manual for instructions on how to apply
gas to each detector.
When calibration gas is applied to a detector, the Beacon 410 will freeze the display
gas reading at the highest level reached while the gas was applied (lowest for an
oxygen detector head). After applying calibration gas for the required amount of time,
usually two minutes, remove the gas from the detector. The Beacon 410 will continue
to display the maximum gas response (lowest for an oxygen detector head) on the
display and retain the response level in its memory.
CAL at AMP. In the example above, channel 2 is a
Beacon 410 Gas Monitor Operator’s ManualPerforming a Calibration • 51
11. After you have applied gas to each direct connect detector head, press and release
the ENTER button. The Beacon 410 will perform a gas adjustment on the direct
connect type channels and indicate
type channel. If the air adjust was not successful for any channel, the display will
indicate
Select Operation screen.
12. Press and release the ESCAPE button to return to the initial Calibration Mode Screen
that says
the DOWN/NO button to exit Calibration Mode and return to normal operation.
Be sure to allow enough time between the application of calibration gas and returning to
normal operation for the gas reading to return to normal levels to avoid false alarms. The
alarm circuits are active when the Beacon 410 returns to normal operation.
AIR ADJUS T F AIL ED for that channel. The Beacon 410 will then return to the
You Have Ent ered the CALIBRATION Menu... and then press and release
Viewing Maximum Spans
This operation allows you to see how much gas response adjustment remains on direct
connect type channels after a calibration is performed. This is a useful gauge for whether
or not a sensor is nearing the end of its life. When you enter the VIEW MAX SPANS
screen, the Beacon 410 displays the maximum possible adjustment that the Beacon 410
could have made to the response reading with the calibration gas that was used for each
direct connect type channel during the most recent successful calibration. So if a
0 - 100 %LEL channel is calibrated using 50 %LEL gas and the maximum indicated span
in the VIEW MAX SPANS screen is 95 %LEL, this means that there was enough
adjustment left on that channel to set the reading to 95 %LEL when the detector was
exposed to 50 %LEL gas. If the maximum span value is close to the calibration gas value,
for example if it is 53 %LEL for a 0 - 100 %LEL channel when 50 %LEL calibration gas is
used, the sensor should be replaced soon.
GAS ADJUST PASSED for each direct connect
To make the maximum span values as meaningful as possible, there is a limit to how
much remaining adjustment the screen will indicate. The maximum remaining adjustment
that will be shown for any channel is 3 times the calibration gas concentration even if the
Beacon 410 could have adjusted it higher. For example, if 50% LEL gas was used to
calibrate a 0 - 100 % LEL channel, the highest maximum span value that will be displayed
is 150 %LEL. If the maximum span adjustment possible for a channel is more than 3 times
the calibration gas value, a “+” sign will be displayed next to the maximum span value.
The maximum span values that are displayed are from the most recent successful
calibration for each channel. So they are most useful soon after a calibration is performed.
If an extended period of time, for example 2 or 3 months, has passed since the last
successful calibration on a particular channel, the maximum span value for that channel
may no longer be accurate.
The maximum span value only has meaning for direct connect type channels. For 4 - 20
mA transmitter type channels, instead of a maximum span value,
displayed for those channels.
To view the maximum span values, do the following:
1. Navigate to the Select Operation Screen as described in “Entering Calibration Mode”
on page 47 or navigate to the Select Operation Screen after performing a calibration.
2. Use the UP/YES or DOWN/NO button to display
Operation Screen.
VIEW MAX SPANS on the Select
4 - 20 mA Xmtr will be
52 • Viewing Maximum SpansBeacon 410 Gas Monitor Operator’s Manual
3. Press and hold the ENTER button to display the maximum spans. In the example
r
LEL%+01ENAHTEM
1
2
3::
:
5
3EOLO:V%5.7NGYX
4-20mA Xmt
below, channels 1, 2, and 3 are active and channel 2 is a 4 - 20 mA type channel.
4. Release the ENTER button to return to the Select Operation screen.
5. Press and release the ESCAPE button to return to the initial Calibration Mode Screen
that says
the DOWN/NO button to exit Calibration Mode and return to normal operation.
You Have Ent ered the CALIBRATION Menu... and then press and release
Beacon 410 Gas Monitor Operator’s ManualViewing Ma xim um Spans • 53
Chapter 9: RS-485 Modbus Output
Overview
This chapter describes the Beacon 410’s RS-485 Modbus output and how to configure the
Beacon 410 to make use of it. It also discusses how to wire the Beacon 410 into a Modbus
system.
The Beacon 410 provides an RS-485 serial communications interface. It is a Modbus
Slave Device, supporting 2-wire RS-485 Modbus RTU serial communications.
Wiring the Beacon 410 in a Modbus
System
The Beacon 410 is a 2-wire Modbus RTU device. When wiring the Beacon 410 into a
Modbus system, adhere to standard Modbus wiring practices per the Modbus Over Serial
Line Specification and Implementation Guide V1.0 or later. This document can be found
online at www.modbus.org/specs.php.
The Modbus protocol supports a maximum of 247 unique slave addresses (1-247). The
Beacon 410’s line driver provides for up to 32 Beacon 410s (or some combination of
Beacon 410s and other Modbus slave devices such as the RKI Instruments M2) to be
connected together without the need for a repeater. Figure 15 & Figure 16 below illustrate
typical Beacon 410 wiring configurations. If more than 32 Beacon 410s (or Beacon 410s
and other slave devices) need to be connected together, an RS-485 repeater(s) should be
used such that no more than 32 Beacon 410s reside on any given network segment.
The MODBUS terminals are located on the controller terminal strip (See Figure 1). The
following signals are available at the Modbus terminals:
Table 10: Modbus Terminal Signals
Modbus Terminal Label
(RS-485 Name)
AD0
BD1
CCommon
Modbus Signal Name
54 • OverviewBeacon 410 Gas Monitor Operator’s Manual
Recommended Modbus Wiring
D1
Common
Beacon 410
Controller Terminal Strip
24V BATT
EXT DC
DP SW
Beacon 410
Controller Terminal Strip
DP SW
A
B
Modbus
Controller
24V BATT
EXT DC
DP SW
DP SW
A
B
ALARM
RESET
ALARM
BUZZER
To Additional
Beacon 410s
ALARM
RESET
ALARM
BUZZER
RS-485
CD0
Input
Terminals
RS-485
C
ID = 32
The recommended Modbus wiring for the Beacon 410 is illustrated in Figure 14 below.
Figure 14: Recommended Modbus Wiring
Termination Jumper
The Beacon 410 includes a 2-pin termination header (see Figure 1) that may need to be
installed when the Beacon 410 is used in a Modbus system. Every Beacon 410 is supplied
with a termination jumper (a jumper block) installed onto this header. If the Beacon 410 is
not used in a Modbus system, this jumper has no function. When the Beacon 410 is
installed in a Modbus system, this jumper must be installed in a Beacon 410 that is at the
end of a Modbus line. Any Beacon 410 in a Modbus system that is not at the end of a line
must have the termination jumper removed (see Figure 15 & Figure 16 below).
Beacon 410
Termination
Jumper
Installed
RS-485RS-485
Up to 32 Beacon 410s can be connected without a repeater.
Beacon 410
Termination
Jumper
Not Installed
Beacon 410
Termination
Jumper
Not Installed
ID = 31ID = 30ID = 29ID = 1
Figure 15: Multiple Beacon 410s in a Daisy Chain Configuration
Beacon 410
Termination
Jumper
Not Installed
Modbus
Master
Beacon 410
Termination
Jumper
Not Installed
Beacon 410 Gas Monitor Operator’s ManualWiring the Beacon 410 in a Modbus System • 55
Modbus
Master
Beacon 410
Termination
Jumper
Installed
Beacon 410
Termination
Jumper
Not Installed
Beacon 410
Termination
Jumper
Not Installed
Beacon 410
Termination
Jumper
Not Installed
Beacon 410
Termination
Jumper
Not Installed
ID = 1
ID = 2ID = 3ID = 4
Beacon 410
Termination
Jumper
Installed
ID = 32
ID = 31
Up to 32 Beacon 410s can be connected without a repeater.
RS-485RS-485
Figure 16: Multiple Beacon 410s in a Two Branch Configuration
Using the Beacon 410 in a 4-wire Modbus System
Although the Beacon 410 is a 2-wire Modbus RTU device, it can be used with a 4-wire
Modbus controller if the system wiring is modified as follows:
•Connect the controller’s TxD0 and RxD0 wires together and use this connection as
the 2-wire Modbus D0 signal.
•Connect the controller’s TxD1 and RxD1 wires together and use this connection as
the 2-wire Modbus D1 signal.
•Connect these D0 and D1 signals and the common wire from the controller to the 2wire Modbus Network of Beacon 410s.
Modbus Menu
The Modbus Menu allows you to configure various parameters relating to the Modbus
setup of the Beacon 410. The Modbus Menu includes a 5-minute timeout feature. If you do
not press a control button for 5 minutes, the Beacon 410 automatically returns to normal
operation.
NOTE: If the Beacon 410 returns to normal operation because of a timeout, it enters a
1. While in normal operation, simultaneously press and hold the ENTER and DOWN/NO
buttons for approximately 5 seconds to enter the Modbus Menu. Release the buttons
when the
warm-up period just as it does when it is first turned on.
You Have Entered the MODBUS Menu message appears on the display.
56 • Using the Beacon 410 in a 4-wire Modbus SystemBeacon 410 Gas Monitor Operator’s Manual
2. Press and release the UP/YES button to continue in the Modbus Menu to the Modbus
SABLED
CSE/TNnehtND/E
B
POU
M
DUSFeature is:
DI
]
ACS
Ssgnitt?
[ktobacup]EP
NY[
e
E
a
ve S
/
Feature Enable/Disable screen.
3.
In the parameter screens, use the ENTER button to accept a displayed parameter
setting and proceed to the next parameter or the ESCAPE button to backup a screen.
4. If you want to change a displayed parameter setting, use the UP/YES or DOWN/NO
button to change the parameter to the desired setting, then press and release the
ENTER button to accept the change and proceed to the next screen.
5. Continue through the parameter setting screens changing any parameter settings as
needed until the following message appears on the display.
6. To save the channel parameter settings, press and release the UP /YES button.The
screen will indicate that the Beacon 410 is saving the settings for a few seconds then
the Beacon 410 will return to normal operation.
If you do not want to save the channel parameter settings, press and release the
DOWN/NO button. The display will indicate that no settings have been changed for a
few seconds then the Beacon 410 will return to normal operation.
Table 11 below lists the Modbus configuration parameters that can be configured in the
Modbus Menu along with the factory set value for each parameter.
Enabled/DisabledCan be set to ENABLED or DISABLED (factory setting).
Enables or disables the Modbus output.
Slave IDThe Slave ID can be set to values from 1 (factory setting) to 247.
The Beacon 410 will only receive messages from the Master
which are addressed to this Slave ID (except for broadcast
messages which are received by all slaves)
Baud RateCan be set to 1200, 2400, 4800, 9600 (factory setting), 14440, &
19200
ParityEVEN (factory setting), ODD, NONE
Beacon 410 Gas Monitor Operator’s ManualModbus Menu • 57
Response DelayThe response delay can be set from 0 (factory setting) to 20 mS.
This is an optional additional delay inserted by the Beacon 410
prior to returning a response message to the Master. It may be
helpful in some installations where the Master’s preparedness to
receive responses might be delayed.
NOTE: All Beacon 410s on a Modbus network must be configured with the same baud
rate and parity, and each Beacon 410 must have a unique slave ID. For a Beacon
410 to participate on a network, the ENABLED/DISABLED setting must be set to
ENABLED.
Supported Modbus Functions
The Beacon 410 supports Function Code 03, Read Holding Registers. There are a total of
57 registers available to be read via this Function Code.
Register 40001, Instrument Information
The format for the first register, register 40001, is shown below in Table 12. Register
40001 gives information about the instrument as a whole, and should be checked before
reading or interpreting the channel data in the registers that follow. When the Beacon 410
is in any operating menu or mode other than normal operation, the channel registers
simply hold their existing values. Once the Beacon 410 returns to normal operation, the
channel registers are updated each second..
Current Operating Mode
(MS-Byte)
•0 = Normal Operation
•1 = Factory (Reserved)
•2 = Input Setup Menu
•3 = Channel Configuration Menu
•4 = Global Menu
•5 = Modbus Menu
•6 = Calibration Mode
•7 = Min/Max Display
•8 = Low Power Standby
T able 12: Register 40001
Common Resource Status Bits
“1” = Energized “0” = De-Energized
(LS-Byte)
•Bit-7 (msb) = Common Fail Relay
•Bit-6 = Relay-A
•Bit-5 = Relay-B
•Bit-4 = Relay-C
•Bit-3 = Relay-D
•Bit-2 = Strobe (optional)
•Bit-1 = Audible Alarm (Buzzer)
•Bit-0 (lsb) = (this bit is not used)
58 • Supported Modbus FunctionsBeacon 410 Gas Monitor Operator’s Manual
Registers 40002 - 40015, Channel 1 Information
Register 40002
Table 13 below shows the information in register 40002.
T able 13: Register 40002
Channel Op erational Sta te
(MS-Byte)
•0 = (this code is not used)
•1 = Standby
•2 = Normal Operation
•3 = Warm Up
•4 = Fail
•5 = Post-Calibration
•6 = Low Power
Registers 40003 & 40004
The 32 bits in registers 40003 and 40004 contain the channel 1 numerical reading in
single precision floating point IEEE format.
Register 40005
The 16 bits in register 40005 contain the channel 1 numerical reading in integer form
omitting the sign (+ or -) and the decimal point.
Register 40006
Channel Status Bits
(LS-Byte)
•Bit-7 (m sb) = Relay-1 (1 = energized)
•Bit-6 = Relay-2 (1 = energized)
•Bit-5 = (this bit is not used)
•Bit-4 = Alarm-1 (1 = asserted)
•Bit-3 = Alarm-2 (1 = asserted)
•Bit-2 = Alarm-3 (1 = asserted)
•Bits[1:0] = Channel Usage Code
[00] = Not Used
[01] = Standby
[10] = Active
[1 1] = unus ed co de
Table 14 below shows the information in register 40006.
Channel Reading Sign (+/-)
•0 = Positive (+)
•1 = Negative (-)
Registers 40007 - 40010
The 8 bytes (characters) in these registers contain the channel gas name in ASCII
characters.
Registers 40011 - 40013
The 6 bytes (characters) in these registers contain the channel gas reading in ASCII
characters.
Registers 40014 - 40015
The 4 bytes (characters) in registers 40014 and 40015 contain the channel gas units in
This chapter describes preventive maintenance procedures for the Beacon 410 and how
to replace the AC fuses. It includes a troubleshooting guide for problems you may
encounter with the Beacon 410.
Preventive Maintenance
Preventive maintenance of the Beacon 410 consists of daily, monthly, and quarterly
procedures to ensure that the installed detector heads remain on zero (20.9 for oxygen) in
fresh air and are responsive to the target gas. See the detector head operator’s manuals
for preventive maintenance procedures for each installed detector head.
Troubleshooting
Table 21 describes symptoms, probable causes, and recommended actions for the most
common problems you may encounter with the Beacon 410.
NOTE: This troubleshooting guide describes Beacon 410 problems only. See the
detector head operator’s manuals for preventive maintenance procedures that
apply to the detector heads installed on your Beacon 410.
•The power wiring is
disconnected or
misconnected.
•The AC fuse is blown.
•The display cable is
disconnected or
misconnected.
1. Verify that the wiring to the power
source is correct and secure.
2. At the Beacon 410, verify that the
wiring to the AC in terminal strip is
correct and secure.
3. Check the continuity of the AC fuses.
4. Verify that the display cable is
connected.
The display (ribbon) cabl e plugs into a
rectangular co nnector on th e top edge
of the control PCB and on the top
edge of the main PCB (labeled Front Panel).
5. If the power difficulties continue,
contact RKI for further instruction.
64 • OverviewBeacon 410 Gas Monitor Operator’s Manual
Table 21: Troubleshooting the Beacon 410 (Continued)
•The Beacon 410
alerts you to
frequent or suspect
alarms.
•The detector heads’
fresh air readings
remain on zero
(20.9 for oxygen)
despite alarm
occurrences.
•The displ ay reading
for one or more
channels rises or
falls briefly and then
returns to the
normal fresh air
reading.
•The Beacon 410 is
experiencing false
readings due to RFI or
EMI.
•The detector head
wiring has one or
more intermittent
connections.
•The Beacon 410 is
experiencing false
readings due to RFI or
EMI.
•The detector head
wiring has one or
more intermittent
connections.
•The noise filter setting
is too low.
•The dead band setting
is too low.
•The displ ay screen is
malfunctioning.
1. Verify that the detector head wiring is
properly shielded.
2. Verify that the detector head wiring is
correct and secure.
3. Verify that power and detector head
wiring is routed through separate
conduit hubs on the bottom of the
Beacon 410 housing.
4. Increase the alarm on delay settin g in
the Configuration Menu.
5. Increase the noise filter in the
Configuration Menu.
6. If the frequent or suspect alarm
difficulties continue, contact RKI for
further instruction.
1. Verify that the detector head wiring is
properly shielded.
2. Verify that the detector head wiring is
correct and secure.
3. Verify that power and detector head
wiring is routed through separate
conduit hubs on the bottom of the
Beacon 410 housing.
4. Increase the noise filter setting in the
Configuration Menu.
5. Increase the dead band setting in the
Configuration Menu.
6. If the difficulties co ntinue, cont act RKI
for further instruction.
Buzzer not
Working
Reset
Switch not
Working
•The buzzer does not
sound an audible
alarm during alarm
conditions.
•The buzzer sounds
weak or broken.
•Depending on the
instrument setup,
the buzzer or strobe
does not turn off in
the appropriate
situation when the
reset switch is
pressed.
•The applicable
alarm circuit does
not reset when you
press the reset
switch after an
alarm condition
passes.
•The buzze r is
disconnected or
misconnected.
•The buzze r is
malfunctioning.
•The reset switch is
disconnected or
misconnected.
•The reset switch is
malfunctioning.
•The Beacon 410 is
not properly setup for
the desired reset
switch operation.
1. Verify that the wiring to the ALARM BUZZER - and ALARM BUZZER +
terminals of the controller terminal
strip is correct and secure.
2. If the buzzer difficulties continue,
contact RKI for further instruction.
1. Verify that the wiring to the ALARM RESET (2) terminals of the controller
terminal strip is correct and secure.
2. If the reset switch diff iculties cont inue,
contact RKI for further instruction.
3. Make sure that the Beacon 410 is
properly setup in the Global Menu
and Configuration Menus for the
desired reset switch operation.
Beacon 410 Gas Monitor Operator’s ManualTroubleshooting • 65
Replacing the AC Fuses
The Beacon 410 has two user serviceable AC fuses.
NOTE: To replace other components of the Beacon 410, contact RKI Instruments, Inc.,
for further information.
1. Turn off or unplug all incoming power to the Beacon 410.
2. Open the Beacon 410 housing door, then place the power switch in the OFF position.
3. The fuses are located to the left of the AC in terminal strip. Use a flat-blade
screwdriver to rotate the applicable fuse holder 1/4 turn counterclockwise. The fuse
holder releases from the socket.
4. Remove the fuse holder from the socket, then remove the fuse from the fuse holder.
CAUTION: Verify that the replacement fuses are the same type and rating as the fuses
you are replacing. The Beacon 410’s fuses are fast blow fuses rated for
6A, 250 V
Parts List
5. Install the replacement fuse in the fuse holder, then place the fuse holder in the
socket.
6. Push the fuse holder into the socket, then turn the holder 1/4 turn clockwise to secure
it in the socket.
7. Plug in or turn on all incoming power to the Beacon 410.
8. Place the Beacon 410’s power switch in the ON position, then verify that the PILOT
light is on.
9. Close and secure the housing door.
Table 22 lists the part numbers and descriptions for replacement parts and accessories
offered for the Beacon 410 Gas Monitor.
T able 22: Parts List, Beacon 410 Gas Monitor
Part No.Description
18-0107RKConduit Hub (3/4 in.)
43-0440RKReset Switch
43-4170RKFast Blow Fuse, 6A, 250 V
51-0096RKStrobe/horn, 10-33 VDC, NEMA 4X
52-1016RKBuzzer
71-0131RKBeacon 410 Gas Monitor Operator’s Manual (this document)
66 • Replacing the AC FusesBeacon 410 Gas Monitor Operator’s Manual
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.