1. Outline of the Product ..................................................................................................................................... 3
Purpose of use ................................................................................................................................................ 3
Definition of DANGER, WARNING, CAUTION and NOTE ............................................................................ 4
Method of confirmation for CE marking type .................................................................................................. 4
2. Important Notices on Safety ........................................................................................................................... 5
3-1. Main unit and standard accessories ........................................................................................................ 7
3-2. Names and functions for each part ......................................................................................................... 9
4. How to Install ................................................................................................................................................ 11
4-1. Precautions for installation points ......................................................................................................... 11
4-2. Precautions for system designing ......................................................................................................... 12
4-3. Installation of main unit .......................................................................................................................... 14
4-4. Precautions for wiring ............................................................................................................................ 16
5. How to Use ................................................................................................................................................... 19
5-1. Before using the monitor ....................................................................................................................... 19
5-2. Preparation for start-up ......................................................................................................................... 19
5-6. User mode ............................................................................................................................................. 23
6-3. Suppression function ............................................................................................................................. 27
10-1. List of specifications ............................................................................................................................ 47
Thank you for choosing our indoor carbon dinoxide monitor RI-600 (hereinafter referred to as the monitor).
Please check that the model number of the product you purchased is included in the specifications on this
manual.
This manual describes how to use the monitor properly and its specifications. Not only the first-time users
but also the users who have already used the monitor must read and understand the operating manual and
use this product as described in this manual.
Note that the contents of this manual are subject to change without notice for product improvement. It is
also prohibited to copy or reproduce this manual, in whole or in part, without permission.
Regardless of warranty period, we shall not make any indemnification for accidents and damage caused by
using the monitor.
Make sure to read the warranty policy specified on the warranty.
Purpose of use
This carbon dioxide monitor measures carbon dioxide in the air. While displaying measured carbon dioxide
concentration on the LCD, the monitor converts it to an analog signal of 4 - 20 mA to output and outputs a
gas contact to ventilate the room at a gas alarm state which carbon dioxide concentration rises over preset
concentration (alarm setpoint).
The monitor is a safety unit, not an analyzer which performs quantitative/qualitative analysis/measurement
for gas. Check the specifications before use and conduct measurement properly in accordance with
purposes.
Throughout this manual, the following indications are used to ensure safe and effective work.
This message indicates that improper handling may cause serious
damage on life, health or assets.
DANGER
WARNING
CAUTION
This message indicates that improper handling may cause serious
damage on health or assets.
This message indicates that improper handling may cause minor damage
on health or assets.
NOTE
This message indicates advice on handling.
Method of confirmation for CE marking type
(DC specification only)
The CE marking is labeled on the detector in case of comply with CE marking. Please confirm the
instrument specification before using. Please refer Declaration of Conformity that is at the end of this
manual if you have CE marking type.
You can confirm instrument specification to see the CE marking as follows.
Do not use the monitor with it attached to a control device, equipment, etc.
2-2. Warning cases
WARNING
Before turning on the monitor, always check that the voltage is compliant with the specifications.
Operating on an unstable power supply may cause malfunctions.
Do not operate this monitor in a place where combustible/explosive gases or vapors are present.
Operating the monitor in such an environment will lead to extreme dangers.
Perform span adjustment at fixed intervals.
Do not run the power cable and remote sensor cable of the monitor in parallel with cables of
high-frequency or high-voltage and other device's power cables. It may cause malfunctions.
If a cable of high-frequency or high-voltage and the power cable need to intersect with each other,
it should be orthogonally connected.
When wiring, be careful not to apply any stresses on the cables by pulling, tightening or twisting,
etc.
Do not disassemble/modify the monitor. It may invalidate the warranty of the performance.
Changing the settings without understanding them may cause alarm malfunctions. Please use the
monitor properly in accordance with the operating manual.
2-3. Precautions
CAUTION
Do not use a device, such as a transceiver, which transmits a radio wave near the monitor or its
cables. It may affect the measurement. If a transceiver or other radio wave transmitting device is
used, it must be used in a place away from the monitor where it disturbs nothing.
Restarting the monitor within five seconds after turning it off may cause errors.
This is not a control unit. It is not allowed to use the external output of the monitor to control other
units.
This is a safety unit. Never fail to perform a regular maintenance to ensure safety. Continuing to use
the monitor without performing maintenance will compromise the sensitivity of the sensor, thus
resulting in inaccurate gas detection.
Do not pick the sensor or buzzer opening with a sharp-pointed item. The unit may cause
malfunction or get damaged, possibly resulting in incorrect measurements.
Do not let the monitor draw in water. Do not also install the monitor in a place where the monitor
may get wet. Ignoring this may cause malfunction because the monitor is not water- and drip-proof.
This is a precision device. Do not give strong shock or vibration to the monitor.
When the case is opened for wiring or other operation, do not touch inner parts. When wiring, be
sure that excessive pressure is not applied to the power cable and remote sensor cable.
- 3.2 m AC power cable (1 pc.) *Supplied only with AC
specification
- Operating manual (1 pc.)
CAUTION
The main unit and sensor unit (remote type) are precision devices. Be careful not to drop the
monitor when installing or uninstalling the main unit or sensor unit. Dropping the monitor may
compromise its original performance or cause malfunctions.
When installing the monitor, never fail to observe the following precautions.
Ignoring the precautions may damage the monitor, resulting in inaccurate gas detection.
CAUTION
This is a precision device. Because the monitor may not provide the specified performance in some
places (environments), check the environment in the installation site, and then take appropriate
actions if necessary. Because the monitor plays an important role for safety and disaster
prevention, it must be installed in appropriate points.
Do not install this product in any of the following locations.
- Place exposed to direct sunlight or outside
- Place exposed to water
- Place exposed to ventilation from an air conditioner, etc.
- Place exposed to soot, smoke or steam
- Place where the temperature drops below 0ºC or rises over 40ºC or the temperature changes
<Do not install the monitor in a place with vibrations or shocks.>
suddenly
- Place with high humidity like a bathroom
- Place with bad ventilation such as behind a curtain or under the shadow.
The monitor consists of sensitive electronic parts. The monitor must be installed
in a stable place without vibrations or shocks, etc. and it cannot drop.
<Do not install the monitor in a place exposed to water, oil or
chemicals, etc.>
When selecting installation points, avoid a place where the monitor is exposed to
liquids such as water, oil, or chemicals.
<Do not install the monitor in a place where the temperature
drops below 0ºC or rises over 40ºC.>
The operating temperature of the monitor is 0 to +40ºC. Install the monitor in a
stable place not exceeding the operating temperature range.
<Do not install the monitor in a place exposed to direct sunlight
or sudden changes in the temperature.>
Avoid a place where the monitor is exposed to direct sunlight or radiant heat
(infrared rays emitted from a high-temperature object), and where the unit
temperature changes suddenly. Condensation may be formed inside the
monitor, or the monitor cannot adjust to sudden changes in the temperature.
<Keep the monitor (and its cables) away from noise source
devices.>
When selecting installation points, avoid a place where
high-frequency/high-voltage devices exist.
<Do not install the monitor in a place where maintenance of the monitor cannot
be performed or where handling the monitor involves dangers.>
Regular maintenance of the monitor must be performed.
Do not install the monitor in a place where the machinery must be stopped when maintenance is
performed in its inside, where parts of the machinery must be removed to perform maintenance, or
where the monitor cannot be removed because racks or other things prevent access to it. Do not
install the monitor in a place where maintenance involves dangers, for example, near a high-voltage
cable.
4-2. Precautions for system designing
Note the following precautions for system designing of the monitor.
CAUTION
An unstable power supply and noise may cause malfunctions or false alarms.
Using a stable power supply
The external output and alarm contact of the monitor may be activated when the power is turned on, when
momentary blackout occurs, or when the system is being stabilized. In such cases, use a UPS
(uninterruptible power system), or take appropriate actions on the receiving side.
The monitor must be provided with the following power supply.
Power supply
Allowed time of
momentary
blackout
Others
100 - 120 VAC ±10% (50/60 Hz) or 24 VDC±10%
Up to 10 milliseconds
(To recover from the momentary blackout for 10 milliseconds or more, restart the
monitor.)
Example of actions
To ensure continuous operation and activation, install a UPS (uninterruptible
power system), etc. outside the monitor.
Do not use it with a power supply of large power load or high-frequency noise.
Example of actions
Use a line filter, etc. to avoid the noise source if necessary.
The alarm contact of the monitor is used to transmit signals to activate an air-conditioning control.
CAUTION
The b contact (break contact) under de-energized state may be opened momentarily by a physical
shock, such as external force.
When the b contact is selected for the alarm contact, take appropriate actions to prepare for a
momentary activation, for example, add signal delay operation (approximately one second) to the
receiving side of the b contact.
The specifications for the external output gas alarm contact of the monitor are based on the resistance load
conditions. If inductive load is used at the alarm contact, the following errors will occur easily because
counter electromotive force is generated at the contact.
- Deposition, defective insulation or defective contact at the relay contact
- Damage of any electric parts due to high-voltage generated inside the monitor
- Abnormal operations by an out-of-control CPU
If load is to be activated, appropriate measures must be taken to stabilize the operation of the monitor and
protect the alarm contact referring to the following information.
- Relay it with an external relay at a lower voltage of 100 VAC or below (contact amplification). At the same
time, the surge absorbing part SK1 suitable for the specifications must be attached to the external relay.
- In addition, the surge absorbing part SK2 must be attached to the loaded side of the external relay if
necessary.
- It may be recommended that the surge absorbing part should be attached to the contact for certain load
conditions. It must be attached to an appropriate position by checking how the load is activated.
RI-600
Coil
Alarm contact
External relay
(Lower voltage relay)
Load
CAUTION
In principle, do not activate inductive load at the alarm contact of the monitor. In particular, never
use the inductive load to activate a fluorescent lamp or motor, etc.
If inductive load is activated, relay it with an external relay (contact amplification). However,
because the coil of an external relay also involves inductive load, select a relay at a lower voltage
(100 VAC or below), and then protect the contact of the monitor with an appropriate surge
absorbing part, such as a CR circuit.
Sudden changes in the temperature may cause the readings not to come back.
Mounting hole (for M4
box
4-3. Installation of main unit
Install the main unit on the wall 50 to 180 cm up from the floor.
If wall screws are available, remove the screw at the lower part of the main unit to open the cover and
install the unit using the mounting holes on the back of the unit.
CAUTION
Install the sensor in a place not directly exposed to ventilation from an air conditioner.
screw)
*Supporting the switch
NOTE
To install the main unit to the wall with screws, use the mounting holes on the back of the unit
according to the following procedure.
(1) Loosen the screw at the lower part of the main unit and open the surface cover of the case.
(2) Push the both sides of the monitor until it clicked.
(3) Open the surface cover.
(4) Fix the main unit with two screws (M4) through the mounting holes.
(5) Put the surface cover back on the case and tighten the screw at the lower part of the main unit.
tranded wires are used, prevent wires from contacting each other.
4-4. Precautions for wiring
If the monitor operates on AC or DC power, or inductive load is used at the alarm contact, wiring work is
required.
The following cables are recommended for wiring the monitor with the power supply, signal cable and
contact.
<Recommended Cables>
Solid wire/stranded wire: 0.2 - 1.5 mm2
CVVS: 0.2 - 1.5 mm2
CVVS: 0.2 - 1.5 mm2
Cable such as CVV (0.2 - 1.5 mm2) Up to 4 cores
CAUTION
Be careful not to damage the internal electronic circuit when wiring. In addition, be careful not to
apply stresses on the monitor when (overweight) cables are installed.
The power and signal cables must be wired separately from the motor power cables.
<Cable Connection Conditions>
Connectable cable, bare wire length and connection tools are as follows:
- Cable: 0.2 - 1.5 mm2
- Bare wire length: 10 - 11 mm
- Connecting tools: Dedicated screwdrivers manufactured by WAGO and equivalent (edge width 3.0 to 4.5
mm x 0.5 mm)
When cables (wires) are connected to the terminal plate inside the main unit, use the dedicated
screwdriver or a compatible flathead screwdriver.
When connecting a stranded wire, be sure to press the push button and open the spring while connecting
the wire.
1 Push the push button
straight downward using
the compatible screwdriver
or equivalent to open the
2 Insert a wire with a
specified bare wire length
(10 mm) until the end of it
reaches the deepest point.
3 Release the screwdriver.
The wire is connected.
CAUTION
Never fail to use the correct tool.
Do not insert more than one wire into one wiring hole. If the total size (mm2) of two or more wires is
within the maximum wire connection range of the terminal plate, it may cause reduced spring clamping
force, defective insulation due to clogged wire sheath, defective contact or coming off of wires.
NOTE
<Compatible Screwdriver>
When opening the spring, use the compatible screwdriver manufactured by WAGO or equivalent (a
screwdriver with an edge width of 3.0 to 4.5 mm x 0.5 mm which can fully open the spring: See the table
below). In doing this work, be careful not to apply excessive force. Ignoring this may damage the
housing/push buttons or cause dropping off of the push buttons.
Screwdriver (M) straight type 210-120J
Screwdriver (M) straight type (short shaft & grip) 210-350/01
210-657
Screwdriver (M) straight type (insulated shaft type) 210-720
Not only the first-time users but also the users who have already used the monitor must follow the
operating precautions.
Ignoring the precautions may damage the monitor, resulting in inaccurate gas detection.
CAUTION
After you received the monitor, start using the monitor within the specified operation start limit of
its sensor.
5-2. Preparation for start-up
Check the following points before starting carbon dioxide concentration monitoring.
- Before turning on the power, check that the monitor is installed properly and the external wiring is done
properly.
- Check that the power supply voltage is compliant with the specifications.
- Because the external contact may be activated during the adjustment, take measures to prevent an
activated contact from having influences on external circuits.
Show the typical settings.
1-20: First Alarm Setpoint (ALM1)
1-21: Second Alarm Setpoint (ALM2)
1-22: Alarm Delay Time (DELY)
1-23: Zero Suppression Value (SUPP)
1-24: Not used (----)
1-26: Not used (----)
1-27: Alarm Summary (AL-S)
2-00: Gas Test (GAS)
2-01: Alarm Test (ALM)
2-02: Fault Alarm Test (FALT)
2-03: Display Test (LCD)
2-04: Not used (----)
Maintenance
mode
(Regular
maintenance)
Environmental
setting
Display 2-5 DISP
Factory mode
User mode
2-4 SET
2-6 FMOD Not used.
2-7 UMOD Return to the user mode.
2-40: Not used (----)
2-41: Not used (----)
2-42: Alarm Value Setting (AL-P)
2-43: Alarm Delay Time Setting (AL-D)
2-44: Alarm Pattern Setting (AL-T)
2-45: Zero Suppression Type Setting (SP-T)
2-46: Zero Suppression Value Setting (SAPP)
2-47: Not used (----)
2-48: External Output Adjustment (MA20)
2-49: Not used (----)
2-4A: Date/Time Setting (DATE)
2-4b: Not used (----)
2-4C: Not used (----)
2-4d: Buzzer ON/OFF Setting (BZZR)
2-4E: Gas Alarm Contact ON/OFF (AL-R)
2-4F: External Output Signal ON/OFF (MA-O)
2-4H: LCD Backlight ON/OFF (LCD)
2-50: Not used (----)
2-51: Not used (----)
2-52: Calibration Curve Number Display (GSEL)
2-53: Fault Detail Display (FALT)
2-54: Not used (----)
Slide the power switch down to power off.
After turning off the power switch of the monitor, turn off the power supply (100 VAC or 24 VDC) of the
monitor.
CAUTION
When the monitor is turned off, an alarm may be triggered on the upper (central) system.
Before turning off the monitor, the inhibit (point skip) on the upper (central) system must be
activated.
Decide whether the power can be turned off by checking the operation of the devices connected to
the external output or external contact output terminal of the monitor.
A gas alarm is activated when detected gas concentration reaches the preset alarm setpoint.
NOTE
The alarm setpoint (first alarm and second alarm) is factory-set. The setting values can be changed in
the maintenance mode (P.29).
Although the alarm delay time (standard: 2 seconds) works in the monitor to prevent a false activation,
it can be cancelled in the maintenance mode (P.29) if not needed.
Display operation
<Concentration Display>
In case of over the detection range (Over Scale), "∩∩∩∩" is displayed on the LCD.
<During Power-on>
The LCD is continuously displayed.
<Alarm Display (ALM1: Yellow Backlight), (ALM2: Red Backlight)>
The alarm consists of two steps. The yellow or red backlight lights up when the respective alarm setpoint
value is reached to or exceeded.(Default: Backlight is off)
Contact activation
When gas concentration reach or is above the alarm setpoint value, the contact is activated and after the
gas concentration fall down below the alarm setpoint value, the contact activation is reset automatically.
In case of measuring range : 0-2000ppm/0-5000ppm/0-10000ppm specification, hysteresis of 50ppm is
A fault alarm is triggered when an abnormality is detected in the monitor. After a fault alarm is triggered,
FAULT is displayed and the backlight (red) lights up on the LCD. (An error message is displayed on the
LCD. Determine the causes and take appropriate actions.)
After the monitor is successfully returned from the fault, it restarts with the process normally performed
right after it is turned on (initial clear).
If the monitor has problems and is repeatedly malfunctioning, contact RIKEN KEIKI immediately.
NOTE
For information on malfunctions (error messages), see "Troubleshooting" (P.45).
6-3. External output operation
The monitor outputs an analog signal of 4 - 20 mA according to CO2 concentration display on the LCD
The monitor is an important instrument for the purpose of safety.
To maintain the performance of the monitor and improve the reliability of safety, perform a regular
maintenance.
7-1. Maintenance intervals and items
This is a safety unit. Never fail to perform daily and regular maintenance before use.
・ Daily maintenance: Perform maintenance before commencing each work.
・ Regular maintenance: Perform maintenance once or more for every six months to maintain the
performance as a safety unit.
Power supply
check
Concentration
display check
Alarm test Inspect the alarm circuit by using the alarm test
Span
Adjustment
Gas alarm
check
WARNING
If an abnormality is found in the monitor, contact RIKEN KEIKI immediately.
Check that the LCD display lights up.
Check that the concentration display value does not
become the abnormal concentration or abnormal
variation.
*The CO2 concentration is also present in the
hundeds of ppm under the general air environment.
(Verifies that relevant points can be identified on the system.)
Span adjustment
Unit operation
check
parts
About maintenance services
We provide services on regular maintenance including such as span adjustment, other adjustments and
maintenance.
To make the calibration gas, dedicated tools, such as a gas cylinder of the specified concentration and gas
sampling bag must be used.
Our qualified service engineers have expertise, knowledge, etc. on the dedicated tools used for services,
along with other products.
To maintain the safety operation of the monitor, please use our maintenance service.
Typical maintenance services are listed as follows. For details, please contact RIKEN KEIKI.
<Typical Maintenance Services>
Power supply
check
Concentration
display check
Alarm test
Gas alarm check
Cleaning and
repair of the unit
Replacement of
consumable
Checks the power supply voltage.
Checks that the LCD display lights up.
Check that the concentration display value does not become the abnormal
concentration or abnormal variation.
*The CO2 concentration is also present in the hundeds of ppm under the general air
environment.
Inspects the alarm circuit by using the alarm test function.
Checks the alarm display (Checks the activation each for ALM1 and ALM2.)
Performs span adjustment using a calibration gas.
Checks the gas alarm by using the calibration gas.
Checks the alarm. (Checks triggering of alarm when the alarm setpoint is reached.)
Checks dust or damage on the surface, cover or internal parts of the unit, and
cleans or repairs such parts as needed.
(Visual diagnosis) Replaces parts which are cracked or damaged.
Operates the buttons to check the operation of functions and parameters, etc.
Replaces consumable parts, such as a sensor.
7-2. Maintenance (regular maintenance) mode
The maintenance mode allows for checking the status of the monitor and adjusting and changing the
settings.
WARNING
When the adjustment is completed, press the MODE button to return to the measuring state.
The monitor automatically returns to the measuring state in 10 hours.
29
www.GlobalTestSupply.com
Mode
Item LCD display
Details
Zero adjustment
Span adjustment
Zero/span
initialization
switching
switching
Maintenance
mode
(Regular
maintenance)
Gas introduction
display
Environmental
setting
Display 2-5 DISP
Factory mode
2-0 GAS TEST 2-00: Gas Test
2-1 ZERO Perform the zero adjustment.
2-2 SPAN Perform the span adjustment.
2-3 SDEF Not used.
2-4 SET 2-40: Not used (----)
2-6 F MODE Not used.
2-01: Alarm Test
2-02: Fault Alarm Test
2-03: Display test
2-04: Not used (----)
2-41: Not used (----)
2-42: Alarm Value Setting (AL-P)
2-43: Alarm Delay Time Setting (AL-D)
2-44: Not used (----)
2-45: Zero Suppression Type Setting (SP-T)
2-46: Zero Suppression Value Setting (SAPP)
2-47: Not used (----)
2-48: External Output Adjustment (MA20)
2-49: Not used (----)
2-4A: Date/Time Setting (DATE)
2-4b: Not used (----)
2-4C: Not used (----)
2-4d: Buzzer ON/OFF Setting (BZZR)
2-4E: Gas Alarm Contact ON/OFF (AL-R)
2-4F: External Output Signal ON/OFF
2-4H: LCD Backlight ON/OFF (LCD)
2-50:Not used (----)
2-51:Not used (----)
2-52:Calibration Curve Number Display (GSEL)
2-53:Fault Detail Display (FALT)
2-54:Not used (----)
Then press the SET
button again for three
seconds.
2-0. GAS TEST
Perform a test with the
gas.
Similar to the detection
condition,the reading
changes and an alarm
is displayed after gas
is introduced, but the
“ALM1” and “ALM2” do
not flash and the
contact is not
activated.
Perform a test with the
gas.
Similar to the detection
mode, the reading
changes and an alarm
is displayed after the
gas is introduced, but
the contact is not
activated.
2-00. GAS
Perform the gas test.
2-01. ALM
Perform the alarm test.
Use the ˄ button to
increase the display
value to the alarm
setpoint to trigger an
alarm.
2-02. FALT
Conduct a fault alarm
test.
Use the ˄ or ˅ button
to enable the function
to trigger an alarm.
Set the date/time.
Use the ˄ or ˅ button
to select each value in
the order of year,
month, day, hour and
minute and press the
SET button to confirm
it.
Contact RIKEN KEIKI for information on readjustment including calibration.
8
8. Storage and Disposal
8-1. Procedures to store the monitor or leave it for a
long time
The monitor must be stored under the following environmental conditions.
• In a dark place under the normal temperature and humidity away from direct sunlight
• In a place where gases, solvents or vapors, etc. are not present
Store the monitor in a shipping carton, if any, in which the product was delivered.
Store the monitor away from dust, etc. if the shipping carton is not available.
8-2. Procedures to relocate the monitor or use it again
When the monitor is relocated, refer to "How to Install" for relocation sites and wiring work.
The unpowered time must be minimized when the monitor is relocated.
When the monitor is used again after a long-period storage, perform a calibration.
CAUTION
8-3. Disposal of products
When the monitor is disposed of, it must be treated properly as an industrial waste in accordance with the
local regulations, etc.
Never fail to return the used sensor to RIKEN KEIKI.
The Troubleshooting does not explain the causes of all the malfunctions which occur on the monitor. This
simply helps to find the causes of malfunctions which frequently occur.
If the monitor shows a symptom which is not explained in this manual, or still has malfunctions even though
remedial actions are taken, please contact RIKEN KEIKI.
The power cannot
be turned on.
Abnormal
operations
Sensor
abnormalities
E-1
System
abnormalities
E-9 SYSTEM
The reading drops
and it remains so.
The power switch is
turned off.
Abnormalities/momentary
blackout of power supply
system
11-1. Detection principle of Non-Dispersive Infrared Ray
Model RI-600 applies Non-Dispersive Infrared Ray Absorption(NDIR) technique to detect target gas.
The infrared beam emitted from the light source passes through the gas cell and reaches to IR sensor. The
target gas enters into gas cell from gas inlet.
When target gas is entered into gas cell, the amount of infrared ray to be received by IR sensor decreases
since the infrared ray emitted from IR source is absorbed by the target gas.
This decreased amount is detected by IR sensor and output as gas concentration. There is a optical band
pass filter in front of IR sensor which can pass the absorption wave of target gas.
Then, there is no sensitivity for the gas which has the different absorption wave from the target gas.
Furthermore, there is no sensitivity against N2 and H2 etc. which cannot absorb infrared ray.
Infrared light
Gas cell
CO2
in Air
Dust fuilter
Optical filterタ
Infrared sensor
Display
Analog output
Contact
11-2. Definition of terms
Gas concentration indicated in the unit of one-millionth of the volume
Gas concentration indicated in the unit of one-hundredth of the volume
Calibration Find relationship of the readings, display values or setpoints with the actual values
Maintenance
mode
Initial clear The reading is unstable for seconds after the power is turned on. To prevent
Zero
Alarm delay time A function which temporarily suspends activation to prevent a false alarm caused by
When maintenance is performed on the monitor, the alarm contact is disconnected,
and a signal to indicate the maintenance mode status is sent out to the external
output signal. As a result, maintenance can be performed on a single unit of the
monitor.
malfunctions for that period, the alarm contact is deactivated. In addition, a signal to
indicate the initial clear status is sent out to the external output.
A function to cut off the influences of environmental changes, interference gases,
etc.