Read and understand this instruction manual before operating
instrument. Improper use of the gas monitor could result in bodily
harm or death.
Maintenance of the gas monitor is essential for proper operation and
correct readings.
Bump test the instrument before each day’s use with a known
concentration of the target gas. A bump test can be done in User
Mode’s BUMP item or by applying gas in Measuring Mode. The
instrument does not need to be calibrated unless it does not pass the
User Mode bump test or does not respond appropriately, as defined by
the user, in Measuring Mode. For more information about bump test
and calibration requirements, see IEC 60079-29-2.
WARNING:Understand manual before operating. This is an intrinsically safe product.
Substitution of components may impair intrinsic safety. To prevent ignition
of a hazardous atmosphere, batteries must only be changed or charged in
an area known to be nonhazardous. Not tested in oxygen enriched
atmospheres (above 21%).
AVERTISSEMENT:Comprendre le manuel avant de l'utiliser. Ceci est un produit
intrinsèquement sûr. La substitution de composants peut nuire à la sécurité
intrinsèque. Pour éviter l'inflammation d'une atmosphère dangereuse, les
batteries ne doivent être remplacées ou chargées que dans une zone non
dangereuse. Non testé dans des atmosphères enrichies en oxygène (plus de
21%).
This chapter briefly describes the SC-04 gas monitor. This chapter also describes the
SC-04 Operator’s Manual (this document). Table 2 at the end of this chapter lists the
specifications for the SC-04.
About the SC-04
Using an advanced detection system, the SC-04 personal gas monitor detects the presence of
super toxic gases like HCN, PH3, and SO2. The SC-04’s compact size and easy-to-use design
make it ideally suited for a wide range of applications, including sewage treatment plants, utility
manholes, tunnels, hazardous waste sites, power stations, petrochemical refineries, mines, paper
mills, drilling rigs, and fire fighting stations. The SC-04 offers a full range of features, including:
•Liquid crystal display (LCD) for complete and understandable information at a glance
•Ultrabright alarm LED
•Distinctive audible/vibrating alarms for dangerous gas conditions and audible alarms for
unit malfunction
•Microprocessor control for reliability, ease of use, and advanced capabilities
•Data logging functions
•Alarm trend data
•STEL, TWA, and over range alarms
•Peak reading
•Built-in time function
•Lunch break feature
•QPS “C/US” classification for Class I, Division I, Groups A, B, C, and D hazardous
atmospheres
WARNING: The SC-04 detects elevated levels of super toxic gases which can be dangerous or
life threatening. When using the SC-04, you must follow the instructions and
warnings in this manual to assure proper and safe operation of the unit and to
minimize the risk of personal injury. Be sure to maintain and periodically
calibrate the SC-04 as described in this manual.
AVERTISSEMENT:Le SC-04 détecte les niveaux élevés de gaz super toxiques qui peuvent
Specifications
Table 1: Standard Sensor Specifications/Alarm Points
être dangereux ou mettre la vie en danger. Lorsque vous utilisez le SC04, vous devez suivre les instructions et les avertissements de ce manuel
pour assurer un fonctionnement correct et en toute sécurité de l'appareil
et pour réduire les risques de blessures. Assurez-vous de maintenir et
d’étalonner périodiquement le SC-04 comme décrit dans ce manuel.
Detection Range
Lowest
Detectable Limit
(LDL)
Reading
Increment
Warning
Setpoint
Alarm Setpoint
Alarm H
Setpoint
STEL Setpoint
Ammonia
)
(NH
3
0 - 400.0 ppm0 - 20.00
4.0 ppm0.10 ppm1.0 ppm0.30 ppm0.02 ppm0.20 ppm
0.5 ppm0.05 ppm0.1 ppm0.05 ppm0.01 ppm0.05 ppm
25 ppm1.00 ppm5.0 ppm2.00 ppm0.30 ppm2.00 ppm
35 ppm2.00 ppm10.0 ppm4.00 ppm0.60 ppm5.00 ppm
300 ppm10.00
35 ppm1 ppm10.0 ppm1.00 ppm1.00 ppm5.00 ppm
Chlorine
(Cl2)
ppm
ppm
Hydrogen
Cyanide
(HCN)
0 - 30.0 ppm0 - 20.00
30.0 ppm20.00 ppm0.60 ppm100.00
Nitrogen
Dioxide
(NO
)
2
ppm
Phosphine
)
(PH
3
0 - 20.00
ppm
Sulfur
Dioxide
(SO
)
2
0 - 100.00
ppm
ppm
TWA Setpoint
Sampling
25 ppm0.5 ppm4.7 ppm0.50 ppm0.30 ppm2.00 ppm
Table 2: SC-04 Specifications
Diffusion
Method
Response
T90 Within 10 Seconds
Time
Display
Operating
Temperature
Graphics LCD Display
Continuous environment: -20°C to 50°C/10 to 90% RH
Temporary environment (up to 15 minutes): -40°C to 60
This chapter describes the SC-04 instrument and its accessories.
Instrument Description
Case
The SC-04’s sturdy, high-impact plastic case is radio frequency (RF) resistant and is suitable for
use in many environmental conditions, indoors and out. The case is dust proof and water resistant.
A clear plastic window is located on the front of the case for viewing the LCD. The sensor
retainer is located on the right side of the case and allows access to the filters and sensor. A feature
in the lower left corner of the rear case is used to install the optional wrist strap.
LCD
A digital LCD (liquid crystal display) is visible through a clear plastic window in the top case.
The LCD shows the gas reading. The LCD also shows information for each of the SC-04’s
operating modes.
Two control buttons, AIR and POWER MODE, are located below the LCD.
Table 3: SC-04 Control Button Functions
ButtonFunction(s)
AIR
POWER MODE
• turns on LCD backlight
• resets alarm condition if LATCH is set to ON in Maintenance Mode
• enters User Mode, Maintenance Mode, and Gas Select Mode when
used with POWER MODE
• activates the demand zero function (adjusts the SC-04’s fresh air
reading)
• changes the value of a parameter available for adjustment
• scrolls through parameter options
• turns the SC-04 on and off
• turns on LCD backlight
• enters and scrolls through Display Mode
• enters instructions into the SC-04’s microprocessor
• resets alarm condition if LATCH is set to ON in Maintenance Mode
• enters User Mode, Maintenance Mode, and Gas Select Mode when
used with AIR
Alarm LED
The alarm LED above the sensor and buzzer openings alerts you to gas, low battery, and failure
alarms.
Buzzer
One solid-state electronic buzzer is located inside the case. Sound exits the case through a hole in
the upper left corner of the front case. The buzzer sounds for gas alarms, malfunctions, and low
battery voltage. It also provides feedback for button presses and while in Display, User,
Maintenance, or Gas Select Mode.
Vibrator
A vibrating motor inside the SC-04 case vibrates for gas alarms and unit malfunctions.
NOTE: If STEALTH is set to ON, the vibrator only functions when VIB in the STEALTH Gas
The sensor is an electrochemical cell that consists of two precious metal electrodes in a dilute acid
electrolyte. A gas permeable membrane covers the sensor face and allows gas to diffuse into the
electrolyte. The gas reacts in the sensor and produces a current proportional to the concentration
of the target gas. The SC-04’s circuitry amplifies the current, converts the current to a gas
concentration, and displays the concentration on the LCD.
Each target gas has its own sensor.
Filters
NO2 and SO2 Sensors’ H2S Removal Filter Disk (Tan)
An H2S removal filter disk is placed into a recess in the sensor gasket over the NO2 or SO2 sensor.
The filter disk prolongs the life of the sensors by preventing H2S in the ambient air from reaching
the sensor. The filter should be replaced every 6 months.
HCN Sensor H2S Removal Filter Disk (Dark Gray)
An H2S removal filter disk is placed into a recess in the sensor gasket over the HCN sensor. The
filter disk prolongs the life of the sensors by preventing H
sensor. The filter should be replaced every 6 months.
PH3 Sensor Humidity Filter (White)
S in the ambient air from reaching the
2
A white humidity filter is placed into a recess in the sensor gasket over the PH3 sensor. The filter
absorbs humidity in the sampling environment to prevent unstable readings around 0 ppm. The
filter should be replaced every 6 months.
NH3 Sensor’s Humidity Filter (White)
A white humidity filter is placed into a recess in the filter gasket over the NH3 sensor. The filter
absorbs humidity in the sampling environment to prevent unstable readings around 0 ppm. The
filter should be replaced every 6 months.
Hydrophobic Filter
The white, circular hydrophobic filter fits into a recessed area in the front case and is held in place
by the sensor gasket. It prevents water and particulates from entering the instrument.
Infrared Communications Port
An infrared (IR) communications port is located at the top of the instrument, near the LED.
Logged data transmits through the port in standard IrDA protocol. A computer’s infrared port or
an IrDA/USB cable connected to a USB port can be used to download data to the 04 Series
Datalogging Program. See the 04 Series Datalogging Program operator’s manual for data logging
and downloading instructions.
Batteries
2 AAA batteries (alkaline or Ni-MH) power the SC-04. At 25°C alkaline batteries will last at least
3,000 hours (125 days) and Ni-MH batteries will last at least 2,000 hours (83 days). The battery
icon in the upper right of the LCD shows remaining battery life.
A low battery warning activates when the SC-04 detects a low battery voltage. The SC-04 sounds
Open
Alligator Clip
Closed
Figure 2: Alligator Clip
a dead battery alarm when battery voltage is too low for Measuring Mode.
WARNING: Use only Duracell MN2400 or PC2400 or Eneloop BK-4MCC batteries to
maintain the QPS classification of the SC-04. Use of other batteries will void the
QPS classification and may void the warranty. Do not mix old/new or different
types of batteries.
AVERTISSEMENT: Utiliser uniquement des piles Duracell MN 2400 ou PC 2400 ou
Eneloop BK-4MCC de maintenir la classification QPS de la SC-04.
L’utilisation d’autres piles annule la classification QPS et peut annuler
la garantie. Ne mélangez pas les anciennes/nouvelles ou différents types
de piles.
WARNING: To prevent ignition of a hazardous atmosphere, the batteries must only be
changed in an area known to be nonhazardous.
AVERTISSEMENT: Pour éviter l'inflammation d'une atmosphère dangereuse, la batterie ne
doit être remplacée que dans une zone non dangereuse.
Standard Accessories
Alligator Clip
An alligator clip installs to 2 spring bars on the rear case. Use the alligator clip to attach the SC-04
to clothing or a belt. Teeth in the alligator clip’s jaws prevent slipping. The alligator clip can be
rotated to change the instrument’s orientation.
The calibration cup installs over the sensor. You must use the calibration cup when performing a
bump test, calibration, or gas test.
IrDA/USB Cable
Unless your computer has a built-in IrDA port, a IrDA/USB cable is needed to establish
communication between the SC-04 and the Datalogging Program or the Setup Program.
This chapter explains how to use the SC-04 to perform confined space entry monitoring or
general area monitoring in Measuring Mode.
Start Up
This section explains how to start up the SC-04, get it ready for operation, and turn it off.
Turning On the SC-04
To illustrate certain functions, the following description of the SC-04 start up sequence assumes
that the following items in User Mode are turned on: LUNCH, CAL.RMDR, and BP.RMDR in
User Mode, and ID DISP and A.ZERO in Maintenance Mode. If any of these items are turned
off, then the corresponding screens do not appear.
1. Press and briefly hold down POWER MODE. Release the button when you hear a beep.
2. If LUNCH is set to ON (factory setting if OFF, see page 65), the Lunch Break Screen
appears. The unit counts down from 5 seconds.
a. Continue Accumulating: To continue accumulating peak and time-weighted average
(TWA) readings from the last time the SC-04 was used, press and release POWER
MODE or allow the countdown to reach 0. The short-term exposure limit (STEL) reading
is reset each time the SC-04 is turned on.
b. Reset Accumulation: To reset the accumulation of peak and time-weighted average
(TWA) readings, press and release AIR before the countdown reaches 0.
3. If CAL.RMDR is set to ON (factory setting) and a calibration is due, the screen that appears
next depends on how CAL.EXPD is set in User Mode (see page 58). The three possible
screens are described below. If a calibration is not due, the instrument displays the number of
days left until a calibration is due.
• Option A, Perform
calibration: Press and
release POWER MODE to
enter User Mode and
perform a calibration. The
instrument takes you
straight to the calibration
start screen in User Mode’s
GAS CAL\A-CAL(E-CAL)
item (if Password Protection is set to On
using the 04 Series Setup
Program, you must enter a
password first). See page 46
for calibration instructions.
If the calibration is
successful, the screen above
will not appear again until
the unit is due for
calibration. If the calibration
is not successful, the screen
above will again appear in
the startup sequence.
• Option B, Bypass message
To continue without
performing a calibration,
press and release AIR.
Buzzer sounds double puls-
ing tone
The SC-04 cannot be used
until a successful calibration
is performed. Press and
release POWER MODE to
enter User Mode and perform
a calibration. The instrument
takes you straight to the cali-
bration start screen in User
Mode’s GAS CAL/A-CAL
(E-CAL) item (if Password
Protection is set to On using
the 04 Series Setup Program,
you must enter a password
first). If you don’t press
POWER MODE, the instru-
ment automatically goes to
the calibration start screen
after 6 seconds (if Password
Protection is set to On using
the 04 Series Setup Program,
you must enter a password
first). See page 46 for calibra-
tion instructions.
If the calibration is successful,
:
the screen above will not
appear again until the unit is
due for calibration. If the cali-
bration is not successful, the
screen above will again
appear in the startup
sequence.
None
• Option A, Perform
calibration: If you want to
enter User Mode and
perform a calibration, press
and release POWER
MODE. The instrument
takes you straight to the
calibration start screen in
User Mode’s GAS CAL/ACAL (E-CAL) item (if
Password Protection is set
to On using the 04 Series
Setup Program, you must
enter a password first).
• Option B, Bypass message
To continue without
performing a calibration,
wait a few seconds for the
instrument to continue with
its startup sequence.
:
4. If BP.RMDR is set to ON (factory setting is OFF) and a bump test is due, the screen that
appears next depends on how BP.EXPD is set in User Mode (see page 62). The three possible
screens are described below. If a bump test is not due, the instrument displays the number of
days left until a bump test is due.
• Option A, Perform bump
test: Press and release
POWER MODE to enter
User Mode and perform a
bump test. The instrument
takes you straight to the
bump test start screen in
User Mode’s BUMP item (if
Password Protection is set
to On using the 04 Series
Setup Program, you must
enter a password first). See
page 42 for bump test
instructions.
If the bump test is
successful, the screen above
will not appear again until
the unit is due for bump
testing. If the bump test is
not successful, the screen
above will again appear in
the startup sequence.
• Option B, Bypass message
To continue without
performing a bump test,
press and release AIR.
Buzzer sounds double puls-
ing tone
The SC-04 cannot be used
until a successful bump test
is performed. Press and
release POWER MODE to
enter User Mode and perform
a bump test. The instrument
takes you straight to the
bump test start screen in User
Mode’s BUMP item (if Pass-
word Protection is set to On
using the 04 Series Setup
Program, you must enter a
password first). If you don’t
press POWER MODE, the
instrument automatically
goes to the bump test start
screen after 6 seconds (if
Password Protection is set to
On using the 04 Series Setup
Program, you must enter a
password first). See page 42
for bump test instructions.
If the bump test is successful,
:
the screen above will not
appear again until the unit is
due for bump testing. If the
bump test is not successful,
the screen above will again
appear in the startup
sequence.
None
• Option A, Perform bump
test: If you want to enter
User Mode and perform a
bump test, press and release
POWER MODE. The
instrument takes you
straight to the bump test
start screen in User Mode’s
BUMP item (if Password Protection is set to On
using the 04 Series Setup
Program, you must enter a
password first).
• Option B, Bypass message
To continue without
performing a bump test,
wait a few seconds for the
instrument to continue with
its startup sequence.
:
5. The Date/Time Screen appears for a few seconds.
6. The Battery Voltage Screen appears for a few seconds. An “AL-L” at the bottom of the screen
batt
2.8
AL -L
v
SO2
100.00
F . S .
ppm
SO2
2.00
WARNING
ppm
SO2
ppm
SO2
5.00
ALARM
ppm
SO2
100.00
ALARM H
ppm
SO2
5.00
STEL
ppm
SO2
2.00
TWA
ppm
USER
ID
U_ID_001
STN
ID
S_ID_001
SO2
FAIL
SENSOR
ppm
indicates that the alarms are set to latching. An “AL-A” at the bottom of the screen indicates
that the alarms are set to auto reset. See page 92 for a description of how to change this
parameter.
7. The following screens display for 1 second each: Gas Name, Full Scale, Warning Setpoint,
Alarm Setpoint, Alarm H Setpoint, STEL Setpoint, and TWA Setpoint.
8. If ID DISP is set to ON (factory setting is OFF, see page 93), the User ID Screen appears for
a few seconds, followed by the Station ID Screen.
9. If the SC-04 experiences a sensor failure during start up, a screen indicating that the sensor
failed appears and the buzzer sounds a double pulsing tone once per second.
You cannot acknowledge the failure and continue to Measuring Mode. Replace the failed
sensor as soon as possible.
10. If A.ZERO is set to ON (factory setting is OFF, see page 93), the instrument prompts you to
A .ZERO
YES .MODE
NO .AIR
0000
PASS- W
SO2
0
ppm
do an auto zero. An auto zero operation sets the reading to 0 ppm.
WARNING: Make sure that the instrument is in a known fresh air environment (an
environment free of combustible or toxic gases and of normal oxygen
content, 20.9%) before performing an auto zero operation. If you perform an
auto zero operation in an area with gases present, the adjustment will not be
accurate.
You must
you do not press any key, after 15 seconds, the instrument enters Measuring Mode without
performing an auto zero.
If Password Protection is turned On (factory setting is Off) using the 04 Series Setup
Program, a user-set password is required to perform an auto zero. When the password
screen appears, adjust each digit with the AIR button and press and release the POWER
MODE button to move on to the next digit. Once the password is entered, the instrument
performs the auto zero.
11. The SC-04 is now monitoring for gas in Measuring Mode. The Measuring Mode Screen
displays the current gas reading.
press and release the POWER MODE button to perform an auto zero function. If
Perform a demand zero before using the SC-04. This sets the reading to 0 ppm.
1. Find a fresh-air environment. This is an environment free of toxic or combustible gases and of
normal oxygen content (20.9%).
2. Turn on the unit as described above in Turning On the SC-04.
3. Press and hold AIR. The buzzer pulses and the LCD prompts you to continue holding AIR (if
KEY.TONE is set to ON in User Mode).
4. Continue to hold AIR until the LCD prompts you to release it. The SC-04 sets the fresh air
reading. Start up is complete and the unit is now ready for monitoring.
5. If Password Protection is turned On (factory setting is Off) using the 04 Series Setup
Program, a user-set password is required to perform a demand zero. When the password
screen appears, adjust each digit with the AIR button and press and release the POWER
MODE button to move on to the next digit. Once the password is entered, the instrument sets
the fresh air reading.
Turning Off the SC-04
1. Press and hold POWER MODE.
2. OFF appears on the display and the buzzer pulses for about five seconds (if KEY.TONE is set
to ON in User Mode).
3. Release the button when OFF disappears from the display.
4. If Password Protection is turned On (factory setting is Off) using the 04 Series Setup
Program, a user-set password is required to turn off the SC-04. When the password screen
appears, adjust each digit with the AIR button and press and release the POWER MODE
button to move on to the next digit. Once the password is entered, the instrument shuts off.
When the SC-04 completes its startup sequence, it is in Measuring Mode. In Measuring Mode the
SC-04 continuously monitors the sampled atmosphere and displays the gas concentration. The
SC-04 is in Normal Operation if there are no alarm indications.
Heart Symbol: The heart symbol in the top right corner of the LCD indicates the operation status
and flashes when normal. A microprocessor error causes the heart symbol to stop flashing or to
disappear.
Check Mark: If BP.RMDR is set to ON and if a bump test is not due, a check mark appears in the
lower left corner of the LCD.
“S”: If the instrument is operating in Stealth Mode, an “S.” appears at the bottom of the LCD.
Backlight: In a low-light environment, press and release either button to turn on the display
backlight. See page 67 to program backlight duration.
Confirmation/Non-Compliance Indicator: If the BEEP item in User Mode is set to anything other
than OFF, the SC-04 gives periodic indications to confirm that it’s operating or to indicate a noncompliance (see page 66).
Monitoring an Area
1. Start up the SC-04 as described above in “Start Up” on page 17. It is now in Measuring Mode.
2. The instrument displays the gas reading.
3. Take the SC-04 to the monitoring area.
4. Wait at least 15 seconds.
5. If a reading is observed, allow the reading to stabilize to determine the gas concentration
present.
6. If a gas alarm occurs, take appropriate action. See page 26.
Interference Information
Some gases interfere with the super toxic sensors. For a complete list of these gases, see page 102.
This section covers alarm indications in Measuring Mode. It also describes how to reset the SC-04
after an alarm occurs and how to respond to an alarm condition.
NOTE: False alarms may be caused by radio frequency (RF) or electromagnetic (EMI)
Alarm Indications
The SC-04 buzzer sounds an alarm, the LED flashes, and the vibrator pulses when any sort of
alarm condition or failure occurs. If the SC-04 is operating in Stealth Mode, the buzzer does not
sound and the vibrator’s operation depends on the VIB setting in Gas Select Mode’s STEALTH
item. See page 101 for more information.
NOTE: If an alarm condition occurs while you are in Display Mode, the SC-04 automatically
The table below summarizes the types of alarms produced by the SC-04 and their indications.
interference. Keep the SC-04 away from RF and EMI sources such as radio transmitters
or large motors.
This section describes response to gas, over range, battery, sensor failure, clock failure, and
system failure alarms.
Responding to Gas Alarms
1. Follow your established procedure for an increasing gas condition.
2. Reset the alarm as necessary or allowed.
a. If LATCH is set to ON (factory setting) in Maintenance Mode, the gas reading must fall
below an alarm setting before you can reset the alarm condition using POWER MODE or
AIR.
If Password Protection is turned On (factory setting is Off) using the 04 Series Setup
Program, you must press POWER MODE and AIR at the same time and then enter a
user-set password to reset an alarm condition. When the password screen appears, adjust
each digit with the AIR button and press and release the POWER MODE button to move
on to the next digit. Once the password is entered, the alarm condition resets.
b. If LATCH is set to OFF in Maintenance Mode, the alarm condition automatically resets
when gas reading falls below an alarm setpoint.
Responding to Over Range Alarms
WARNING: An over range condition may indicate an extreme carbon monoxide
concentration or an explosive concentration. Confirm the gas concentration with
a different SC-04 or with another gas detecting device.
AVERTISSEMENT: Un dépassement de la plage peut indiquer une concentration extrême en
monoxyde de carbone ou une concentration en explosif. Confirmez la
concentration de gaz avec un SC-04 différent ou avec un autre dispositif
de détection de gaz.
1. Follow your established procedure for an extreme gas condition.
2. Reset the alarm using POWER MODE or AIR once the alarm condition clears if LATCH is
0000
PASS- W
set to ON (factory setting) in Maintenance Mode.
If Password Protection is turned On (factory setting is Off) using the 04 Series Setup
Program, you must press POWER MODE and AIR at the same time and then enter a userset password to reset an alarm condition. When the password screen appears, adjust each
digit with the AIR button and press and release the POWER MODE button to move on to
the next digit. Once the password is entered, the alarm condition resets.
3. Calibrate the SC-04 as described on page 46.
4. If the over range condition continues or if you are not able to successfully calibrate the unit,
you may need to replace the sensor.
5. If the over range condition continues after you replace the sensor, contact RKI Instruments,
Inc. for further instructions.
Responding to Battery Alarms
WARNING: The SC-04 is not operational as a gas monitoring device during a dead battery
alarm. Take the SC-04 to a non-hazardous area and replace the batteries as
described in “Replacing the Batteries (Alkaline or Ni-MH)” on page 74.
The SC-04 is fully functional during a low battery warning. However, only a couple of days of
operating time remain. The amount of time depends on LCD backlight use and alarm frequency.
Replace the batteries as described on page 74 as soon as possible.
NOTE: Alarms and the LCD backlight consume battery power and reduce the amount of
operating time remaining.
Responding to Sensor Failure Alarms
1. Calibrate the sensor as described on page 46.
2. If the sensor failure continues, replace the sensor as described on page 81.
3. If the sensor failure condition continues after replacing the sensor, contact RKI Instruments,
The SC-04 logs Measuring Mode gas readings, alarm data, and calibration data to its internal
memory. Logged data can be download it to a computer via the infrared communications port on
the front of the unit.
The data logging capacity depends on how often the SC-04 stores data and how often the
SC-04 is turned on and off. The table below illustrates how much data logging time is available
for the various interval times. It assumes that the unit is only turned on once and there are no
alarms. The data logging interval time must be set using the 04 Series Datalogging Program.
Table 6: Data Logging Capacity
Interval
Time
10 seconds10 hours
20 seconds20 hours
30 seconds30 hours
1 minute60 hours
3 minutes180 hours
5 minutes300 hours
10 minutes600 hours
Data Logging
Capacity
To utilize the SC-04’s downloading capability, you need:
•PC with Windows 7, Windows 8, Windows 10, or Windows 11
•IrDA port or IrDA/USB cable (cable available from RKI Instruments, Inc.)
•04 Series Datalogging Program (available at www.rkiinstruments.com/04series).
For a complete description of the Datalogging Program and procedures for downloading data to a
computer, see the 04 Series Datalogging Program Operator’s Manual.
This section describes Display Mode which is accessible from Measuring Mode. See Table 7
below for a list of Display Mode’s menu items, a short description of each item, and the page
number for further description.
Table 7: Display Mode Menu Items
Display Mode
Menu Item
PEAK (page 31)Displays the Peak reading.
STEL (page 32)Displays the STEL reading.
TWA (page 32)Displays the TWA reading.
USER ID (page 32)
STN ID (page 33)
CAL.DATA (page
B
34)
BP.DATA (page 34)
DATE (page 35)Displays the current date and time.
TEMP (page 35)Displays the current temperature.
ALARM--P (page
35)
Description
A
View and/or change the User ID.
A
View and/or change the Station ID.
Displays the last calibration date.
C
Displays the last bump test date.
View alarm points.
BUZZ.VOL
(page 36)
A
Only appears if DISP.SET is set to ON in User Mode (factory setting) and if ID DISP is set to ON in
Maintenance Mode (factory setting is OFF).
B
Only appears if CAL.RMDR is set to ON in User Mode (factory setting).
C
Only appears if BP.RMDR is set to ON in User Mode (factory setting is OFF).
D
Only appears if DISP.SET is set to ON in User Mode (factory setting).
D
Set the buzzer volume to LO or HI (factory setting).
Tips for Using Display Mode
•To enter Display Mode and scroll from one item to the next or skip an item when a question is
•To change a flashing parameter, press and release AIR. To reverse the movement in a list (ie.
SO2
20.00
PEAK
ppm
CLR
HOLD
CLR
RELEASE
from down to up or vice versa):
a. Press and hold AIR.
b. Immediately press POWER MODE and then release both buttons.
•To exit from an entered-information screen and go back to Measuring Mode, press and release
POWER MODE until you get to the Measuring Mode screen.
NOTE: Each screen displays for 20 seconds. If you do not press a button within 20 seconds, the
SC-04 automatically returns to Measuring Mode.
Peak Screen (PEAK)
The peak screen displays the highest concentration detected since the SC-04 was turned on. The
peak reading is stored until a higher level is detected, the peak reading is cleared, or the SC-04 is
turned off.
The lunch break feature enables the SC-04 to save the peak reading when it is turned off so it can
continue with the same peak when it is turned on again (see page 17).
To clear the peak reading, do the following:
NOTE: If Password Protection is set to On using the 04 Series Setup Program, the peak
reading cannot be cleared.
1. After entering Display Mode, press and release POWER MODE until PEAK appears.
2. Press and hold AIR until the screen prompts you to release it.
3. The peak reading is reset and the unit returns to the Peak Screen.
If you do not want to clear the peak reading, release AIR before the above screen sequence
occurs. The unit returns to the Peak Screen.
The STEL Screen displays the short term exposure limit (STEL) reading. The STEL reading is the
average reading over the last 15 minutes.
TWA Screen (TWA)
The TWA Screen displays the time weighted average (TWA) reading.
The TWA reading is the average reading over the last 8 hours. If 8 hours have not elapsed since
the last time the TWA reading was cleared, the average is still calculated over 8 hours. The
missing readings are assigned a value of 0. If LUNCH is set to OFF (factory setting), the TWA is
cleared when the SC-04 is turned off.
If LUNCH is set to ON, the SC-04 remembers the TWA reading when it is turned off and can
continue accumulation when it is turned on again (see page 17).
Changing the User ID (USER ID)
This screen only appears if DISP.SET in User Mode is set to ON (factory setting) and if ID DISP
in Maintenance Mode is set to ON (factory setting is OFF).
Use this screen to select a user ID from the 128 user IDs that are stored in the SC-04’s memory.
Before a user ID is selected on a brand new instrument, the user ID is “----------”. The factoryinstalled user IDs have a “U_ID_XXX” format.
The user ID provides a way to identify the SC-04 user during a data logging session.
User IDs can only be selected
the 04 Series Datalogging Program or 04 Series Setup Program.
1. After entering Display Mode, press and release POWER MODE until the USER ID screen
sequence appears.
in this menu item. In order to edit the 128 user IDs, you must use
2. To change the User ID, press and release AIR. The current User ID flashes.
USER
ID
U_ID_001
STN
ID
S_ID_001
STN
ID
CHG / AIR
STN
ID
NO /MODE
STN
ID
S_ID_001
3. Use AIR to scroll to the desired User ID.
4. Press and release POWER MODE to save the User ID and return to the USER ID screen in
Display Mode.
Changing the Station ID (STN ID)
This screen only appears if DISP.SET in User Mode is set to ON (factory setting) and if ID DISP
in Maintenance Mode is set to ON (factory setting is OFF).
Use this screen to select a station ID from the 128 station IDs that are stored in the SC-04’s
memory. Before a station ID is selected on a brand new instrument, the station ID is
“----------”. The factory-installed station IDs have a “S_ID_XXX” format.
The station ID provides a way to identify the SC-04 location during a data logging session.
User IDs can only be selected in this menu item. In order to edit the 128 user IDs, you must use
the 04 Series Datalogging Program or 04 Series Setup Program.
1. After entering Display Mode, press and release POWER MODE until the STN ID screen
sequence appears.
2. To change the Station ID, press and release AIR. The current Station ID flashes.
3. Use AIR to scroll to the desired Station ID.
4. Press and release POWER MODE to save the Station ID and return to the STN ID screen in
3. Use AIR to scroll through the Warning, Alarm, Alarm H, STEL, and TWA settings.
SO2
100.00
F. S.
ppm
SO2
2.00
WARNING
ppm
SO2
5.00
ALARM
ppm
SO2
100.00
ALARM H
ppm
SO2
5.00
STEL
ppm
SO2
2.00
TWA
ppm
AIR AIR
AIR AIR
AIR
HI
BUZZ .VOL
HI
CHG / AIR
HI
NO / MODE
4. While viewing the alarm settings for a particular alarm point, press and release AIR and
POWER MODE at the same time to simulate the alarm conditions. The buzzer will sound, the
LED will flash, and the instrument will vibrate just as it would if the displayed condition was
actually happening.
5. Press and release POWER MODE to return to the Alarm Points Screen.
Adjusting the Buzzer Volume (BUZZ.VOL)
The BUZZ.VOL screen allows you to adjust the volume of the instrument’s buzzer. This screen
only appears if DISP.SET in User Mode is set to ON (factory setting).
HI (factory setting): Buzzer volume is high.
LO: Buzzer volume is low.
1. While in Display Mode, press and release POWER MODE until BUZZ.VOL appears. The
current setting displays on the top line.
2. Press and release AIR. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the BUZZ.VOL item in
This section describes the SC-04 in User Mode. See Table 8 below for a list of the items found in
User Mode, the page that the item’s instructions can be found on, and a short description of the
item.
Table 8: User Mode Menu Items
User Mode
Menu Item
BUMP (page 42)Perform a bump test.
GAS CAL (page 46)Perform a fresh air adjustment, perform a span adjustment, change the calibration gas
Description
BUMP Perform a bump test.
STARTBegin the warmup sequence and enter Measuring Mode.
ESCAPEReturn to the BUMP menu item.
concentration.
AIR (page 46)Perform a fresh air adjustment.
A-CAL (page
47) or E-CAL
(page 51)
depending on ECAL User
Mode setting
A-CAL (or
E-CAL)
STARTBegin the warmup sequence and enter Measuring
CAL-PSet the calibration gas concentration.
ESCAPEReturn to the A-CAL item in the GAS CAL menu.
Perform a span adjustment.
Mode.
ESCAPEReturn to the GAS CAL item in User Mode.
SC-0 4 Operator’s Manual Chapter 5: User Mode and Calibration • 37
ON: The instrument notifies the user upon startup when a bump
test is due. Notification type depends on BP.EXPD setting below.
BP.DATA screen appears in Display Mode.
OFF (factory setting)
test is due. BP.DATA screen does not appear in Display Mode.
How often the instrument needs to be bump tested.
Options: 0 - 30 days (factory setting is 30 days)
Defines what action must be taken if a bump test is due upon
startup.
CONFIRM (factory setting)
edge that bump test is due and continue to Measuring Mode.
CANT.USE
bump test is performed.
NONE
sequence continues.
: Cannot enter Measuring Mode until a successful
: A screen indicates that bump test is due but warmup
WARNING: The SC-04 is not in operation as a gas detector while in User Mode.
1. Take the SC-04 to a non-hazardous location and turn it off if it is on.
2. Press and hold AIR, then press and hold POWER MODE. When you hear a beep, release the
buttons.
3. The screen that appears depends on the setting of User Mode’s PASS-W item.
If PASS-W is set to OFF (factory setting), continue with Step 6.
If PASS-W is set to ON, continue with Step 4.
4. If PASS-W is set to ON in User Mode, a password screen appears and the first digit flashes.
The factory-set password is 0405 but it can be changed.
5. Use AIR to select each password number then press POWER MODE to save it and move on
to the next number. To go back a number, press and hold AIR and POWER MODE for a few
seconds. To reverse the direction of change (ie. from increasing to decreasing or vice versa):
a. Press and hold AIR.
b. Immediately press POWER MODE and then release both buttons.
6. The BUMP item displays.
7. Use AIR to move through the User Mode items.
SC-0 4 Operator’s Manual Chapter 5: User Mode and Calibration • 41
•To scroll from one item to the next, press and release AIR. To reverse the scrolling direction:
a. Press and hold AIR.
b. Immediately press POWER MODE and then release both buttons.
c. The scrolling direction returns to the original direction when you exit and reenter a menu.
•To skip an item when a question is asked, press and release AIR.
•To enter an item and to save any changes, press and release POWER MODE.
•To change a flashing parameter, press and release AIR. To reverse the direction of change (ie.
from increasing to decreasing or vice versa):
a. Press and hold AIR.
b. Immediately press POWER MODE and then release both buttons.
•To exit an entered item without saving a change, press and hold AIR and POWER MODE for a
few seconds.
Performing a Bump Test (BUMP)
Bump test the instrument before each day’s use with a known concentration of the target gas. The
instrument does not need to be calibrated unless it does not pass the bump test.
To bump test the SC-04, you need:
•Calibration gas cylinder
Table 9: Calibration Concentration Limits
Channel
Ammonia8 ppm400 ppm
Chlorine0.15 ppm10.00 ppm
Hydrogen Cyanide0.9 ppm30.0 ppm
Nitrogen Dioxide0.50 ppm20.00 ppm
Phosphine0.05 ppm20.00 ppm
Sulfur Dioxide0.50 ppm100.00 ppm
•0.25 LPM fixed flow regulator
Min. Cal. Gas
Concentration
Max. Cal. Gas
Concentration
•Non-absorbent tubing
•Calibration cup
1. Confirm that the SC-04’s calibration gas value matches the concentration listed on the
calibration gas cylinder as described on page 55.
42 • Chapter 5: User Mode and Calibration SC-04 Operator’s Manual
2. Install the calibration cup onto the SC-04. Be sure the calibration cup is installed in the correct
POWER
MODE
AIR
Front ViewSide View
BUMP
SO2
5.00
BUMP
ppm
SO2
0.00
BUMP 30
ppm
SO2
0.00
APLY 30
ppm
direction and that it is pushed on all the way.
3. Use the tubing to connect the regulator to the inlet of the calibration cup (labeled “IN”).
4. While in User Mode, press AIR to scroll to BUMP.
5. Press and release POWER MODE. The display shows the gas and its assigned calibration
value (see page 55 if the calibration value does not match the calibration gas cylinder’s
concentration).
6. For toxic gas cylinders, it is important to vent the regulator while installing it onto the
cylinder. Venting the regulator during installation helps prevent air from getting into the
cylinder and degrading the gas. Open the regulator by turning the knob counterclockwise and
install it onto the cylinder.
7. Press and release POWER MODE.
8. The gas reading flashes, the bottom of the screen alternates between “APLY” and “BUMP”,
and the bottom of the screen counts down from the time set in
BUMP.SET\SETTING\GAS.TIME.
To back out of the gas application screen without performing the bump test, press and
release AIR and POWER MODE together.
SC-0 4 Operator’s Manual Chapter 5: User Mode and Calibration • 43
9. At the end of the countdown, the instrument analyzes the results. Follow the flow chart to
Figure 6: Bump Test Flow Chart
Bump test passed?
Y N
A-CAL On?
Calibration passed?
Y N
(default is on)
B
C D
Y N
A
SO2
P
BMP / CAL
ppm
SO2
4.70
BMP /
ppm
AIR
SO2
F
BMP / CAL
ppm
SO2
0.90
BMP /
ppm
AIR
determine the bump test outcome.
Option A from Flow ChartOption B from Flow Chart
• Bump test passed• Bump test failed
• A-CAL set to OFF (factory setting is
ON)
1. The instrument indicates that the
channel passed the bump test. Use
AIR to scroll between the bump test
result and the bump test gas
reading.
2. Close the regulator.
3. Unscrew the regulator.
4. Remove the calibration cup.
5. Press and release POWER MODE
to return to the BUMP screen in the BUMP menu.
6. Use AIR to scroll to START and
press and release POWER MODE
to enter Measuring Mode.
1. The instrument shows which
channels passed or failed the bump
test. The LED flashes and the
buzzer sounds. Use AIR to scroll
between the result and the reading.
2. Close the regulator.
3. Unscrew the regulator.
4. Remove the calibration cup.
5. Press and release POWER MODE
to return to the BUMP screen in the
BUMP menu.
6. Use AIR to scroll to START and
press and release POWER MODE
to enter Measuring Mode.
44 • Chapter 5: User Mode and Calibration SC-04 Operator’s Manual
www.GlobalTestSupply.com
Option C from Flow ChartOption D from Flow Chart
SO2
0.00
CAL 60
ppm
SO2
0.00
APLY 60
ppm
SO2
F P
BMP / CAL
ppm
SO2
0.90
BMP /
ppm
SO2
4.70
/ CAL
ppm
AIR
AIR
SO2
0.00
CAL 60
ppm
SO2
0.00
APLY 60
ppm
SO2
F F
BMP / CAL
ppm
SO2
0.00
BMP /
ppm
SO2
0.90
/ CAL
ppm
AIR
AIR
• Bump test failed
• A-CAL set to ON (factory setting)
• Calibration passed
1. A calibration immediately and
automatically starts. Continue to
apply the calibration gas.
2. The calibration time is the
difference between the GAS.TIME
and the CAL.TIME values defined
in the BUMP.SET\SETTING item
in User Mode.
3. The instrument shows the pass/fail
results of the bump test/calibration.
Use AIR to scroll between the
results and the readings.
• Bump test failed
• A-CAL set to ON (factory setting)
• Calibration failed
1. A calibration immediately and
automatically starts. Continue to
apply the calibration gas.
2. The calibration time is the
difference between the GAS.TIME
and the CAL.TIME values defined
in the BUMP.SET\SETTING item
in User Mode.
3. The instrument shows the pass/fail
results of the bump test/calibration.
The LED flashes and the buzzer
sounds. Use AIR to scroll between
the results and the readings.
SC-0 4 Operator’s Manual Chapter 5: User Mode and Calibration • 45
6. Remove the calibration cup.
7. Press and release POWER MODE
to return to the BUMP screen in the BUMP menu.
8. Use AIR to scroll to START and
press and release POWER MODE
to enter Measuring Mode.
www.GlobalTestSupply.com
4. Close the regulator.
5. Unscrew the regulator.
6. Remove the calibration cup.
7. Press and release POWER MODE
to return to the BUMP screen in the
BUMP menu.
8. Use AIR to scroll to START and
press and release POWER MODE
to enter Measuring Mode.
Performing a Calibration (GAS CAL)
GAS CAL
AIR
SO2
0.00
AIR
ppm
•Bump test the instrument before each day’s use with a known concentration of the target gas. A
bump test can be done in User Mode’s BUMP item or by applying gas in Measuring Mode.
The instrument does not need to be calibrated unless it does not pass the User Mode bump test
or does not respond appropriately, as defined by the user, in Measuring Mode.
•To fully calibrate the sensor, you must do a fresh air adjustment (AIR CAL) and a span
adjustment (A-CAL or E-CAL).
•The SC-04 can be calibrated using either A-CAL or E-CAL depending on the setting of the E-CAL User Mode item.
•A-CAL (appears if E-CAL User Mode item is set to OFF): Apply gas for a period of
time, then press and release POWER MODE to perform the adjustment.
•E-CAL (appears if E-CAL User Mode item is set to something besides OFF): As soon as
gas is applied and the reading reaches 10% of the auto calibration value, the instrument
counts down from the number of seconds specified in the E-CAL User Mode item and
automatically performs the adjustment.
Performing a Fresh Air Adjustment (AIR)
1. Find a fresh air environment, an environment of normal oxygen content (20.9%) that is free of
toxic and combustible gases.
2. While in User Mode, press AIR to scroll to GAS CAL.
3. Press and release POWER MODE. The AIR item appears.
4. Press and release POWER MODE to enter the AIR menu.
5. To return to the AIR menu without performing a fresh air adjustment, press and release
POWER MODE.
46 • Chapter 5: User Mode and Calibration SC-04 Operator’s Manual
2. Install the calibration cup onto the SC-04. Be sure the calibration cup is installed in the correct
POWER
MODE
AIR
Front ViewSide View
GAS CAL
A-CAL
SO2
5.00
A-CAL
ppm
direction and that it is pushed on all the way.
3. Use the tubing to connect the regulator to the inlet of the calibration cup (labeled “IN”).
Performing a Span Adjustment
1. While in User Mode, press AIR to scroll to GAS CAL.
2. Press and release POWER MODE. The AIR item appears.
3. Use AIR to scroll to the A-CAL item.
4. Press and release POWER MODE. The display shows the gas and its assigned calibration
value (see page 55 if the calibration value does not match the calibration gas cylinder’s
concentration).
5. For toxic gas cylinders, it is important to vent the regulator while installing it onto the
cylinder. Venting the regulator during installation helps prevent air from getting into the
cylinder and degrading the gas. Open the regulator by turning the knob counterclockwise and
install it onto the cylinder.
6. Press and release POWER MODE.
48 • Chapter 5: User Mode and Calibration SC-04 Operator’s Manual
11. The table below describes the 2 calibration outcomes.
PASS
A-CAL
SO2
5.10
A-CAL
ppm
SO2
10.00
MAX .SPAN
ppm
SO2
FAIL
A-CAL
ppm
Calibration PassedCalibration Failed
1. The instrument indicates that the
calibration passed.
2. The instrument shows the current
gas reading.
3. If MAX.SPAN is set to ON
(default is OFF, see page 100), the
response reading’s maximum
adjustment is displayed. A
maximum span of 10.00 ppm
indicates that the reading could
have been adjusted up to 10.00
ppm. If the maximum span value is
close to the calibration gas value,
the sensor should be replaced soon.
1. The instrument indicates that the
calibration failed.
2. Close the regulator.
3. Unscrew the regulator.
4. Remove the calibration cup.
5. Press and release POWER MODE
to return to the A-CAL screen in
the A-CAL menu.
6. See “Troubleshooting” on page 72.
The maximum adjustment is either
twice the calibration value or full
scale, whichever is lower.
4. Close the regulator.
5. The instrument continues to the
START screen.
6. Remove the calibration cup.
7. Press and release POWER MODE
to enter Measuring Mode.
50 • Chapter 5: User Mode and Calibration SC-04 Operator’s Manual
www.GlobalTestSupply.com
Performing a Span Adjustment in E-CAL
POWER
MODE
AIR
Front ViewSide View
The E-CAL item only appears if E-CAL in User Mode is set to ON (factory setting is OFF). If
E-CAL is set to OFF, see page 47 for calibration instructions.
Preparing for Span Adjustment
To adjust the span on the SC-04, you need:
•Calibration gas cylinder
Table 11: Calibration Concentration Limits
Channel
Ammonia8 ppm400 ppm
Chlorine0.15 ppm10.00 ppm
Hydrogen Cyanide0.9 ppm30.0 ppm
Nitrogen Dioxide0.50 ppm20.00 ppm
Phosphine0.05 ppm20.00 ppm
Sulfur Dioxide0.50 ppm100.00 ppm
Min. Cal.
Concentration
Max. Cal.
Concentration
•0.25 LPM fixed flow regulator
•Non-absorbent tubing
•Calibration cup
1. Confirm that the SC-04’s calibration gas value matches the concentration listed on the
calibration gas cylinder as described on page 55.
2. Install the calibration cup onto the SC-04. Be sure the calibration cup is installed in the correct
direction and that it is pushed on all the way.
3. Use the tubing to connect the regulator to the inlet of the calibration cup (labeled “IN”).
SC-0 4 Operator’s Manual Chapter 5: User Mode and Calibration • 51
1. While in User Mode, press AIR to scroll to GAS CAL.
2. Press and release POWER MODE. The AIR item appears.
3. Use AIR to scroll to the E-CAL item.
4. Press and release POWER MODE. The display shows the gas and its assigned calibration
value (see page 55 if the calibration value does not match the calibration gas cylinder’s
concentration).
5. Press and release POWER MODE.
6. The gas reading flashes and the bottom of the screen says “GAS IN”.
7. For toxic gas cylinders, it is important to vent the regulator while installing it onto the
cylinder. Venting the regulator during installation helps prevent air from getting into the
cylinder and degrading the gas. Open the regulator by turning the knob counterclockwise and
install it onto the cylinder.
52 • Chapter 5: User Mode and Calibration SC-04 Operator’s Manual
10. The table below describes the 2 calibration outcomes.
PASS
E-CAL
SO2
5.10
E-CAL
ppm
SO2
10.00
MAX .SPAN
ppm
SO2
FAIL
E-CAL
ppm
Calibration PassedCalibration Failed
1. The instrument indicates that the
calibration passed.
2. The instrument shows the current
gas reading.
3. If MAX.SPAN is set to ON
(default is OFF, see page 100), the
response reading’s maximum
adjustment is displayed. A
maximum span of 10.00 ppm
indicates that the reading could
have been adjusted up to 10.00
ppm. If the maximum span value is
close to the calibration gas value,
the sensor should be replaced soon.
1. The instrument indicates that the
calibration failed.
2. Close the regulator.
3. Unscrew the regulator.
4. Remove the calibration cup.
5. Press and release POWER MODE
to return to the E-CAL screen in
the E-CAL menu.
6. See “Troubleshooting” on page 72.
The maximum adjustment is either
twice the calibration value or full
scale, whichever is lower.
4. Close the regulator.
5. The instrument continues to the
START screen.
6. Remove the calibration cup.
7. Press and release POWER MODE
to enter Measuring Mode.
8. Use AIR to adjust the calibration gas value. The calibration gas value in the instrument must
match the value listed on the calibration gas cylinder you are using. Limits on the calibration
gas value are shown in the table below.
Table 12: Calibration Concentration Limits
Channel
Ammonia8 ppm400 ppm
Chlorine0.15 ppm10.00 ppm
Hydrogen Cyanide0.9 ppm30.0 ppm
Nitrogen Dioxide0.50 ppm20.00 ppm
Phosphine0.05 ppm20.00 ppm
Sulfur Dioxide0.50 ppm100.00 ppm
Min. Cal.
Concentration
Max. Cal
Concentration
9. Press and release POWER MODE to save the change. The calibration gas value stops flashing
and the unit returns to the gas name screen.
10. Use AIR to scroll to ESCAPE.
11. Press and release POWER MODE to return to the CAL--P item in the A-CAL menu.
12. Press AIR to scroll to ESCAPE.
13. Press and release POWER MODE to return to the A-CAL item in the GAS CAL menu.
14. See “Exiting the GAS CAL Menu” on page 56 to return to User Mode.
Exiting the GAS CAL Menu
1. While in the GAS CAL menu, press AIR to scroll to ESCAPE.
2. Press and release POWER MODE to return to the GAS CAL item in User Mode.
3. See “Entering Measuring Mode (START)” on page 71 to enter Measuring Mode.
56 • Chapter 5: User Mode and Calibration SC-04 Operator’s Manual
The CAL SET menu has 4 items: CAL.RMDR, CAL.INT, CAL.EXPD, and ESCAPE.
1. While in User Mode, press AIR to scroll to CAL SET.
2. Press and release POWER MODE. The CAL.RMDR item appears.
CAL.RMDR
ON (factory setting): The SC-04 gives an indication at start up if it is due for calibration. The type
of indication depends on the CAL.EXPD setting (see page 58).
OFF: The SC-04 does not give an indication at start up if it is due for calibration.
1. After entering the CAL SET menu, press AIR to scroll to CAL.RMDR.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the CAL.RMDR item in
the CAL SET menu.
5. See “Exiting the CAL SET Menu” on page 58 to return to User Mode.
CAL.INT
This setting defines the amount of time between calibrations and can be set in 1 day increments.
The minimum setting is 1 day and the maximum setting is 1000 days. The factory setting is 90
days.
1. After entering the CAL SET menu, press AIR to scroll to CAL.INT.
SC-0 4 Operator’s Manual Chapter 5: User Mode and Calibration • 57
2. Press and release POWER MODE. The current setting flashes.
CAL .EXPD
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the CAL.INT item in the
CAL SET menu.
5. See “Exiting the CAL SET Menu” on page 58 to return to User Mode.
CAL.EXPD
This item defines what indication is given during start up when calibration is due and
CAL.RMDR is set to ON (factory setting).
CONFIRM (factory setting): The SC-04 gives an indication at start up if calibration is past due.
Press and release AIR to continue without calibrating or press and release POWER MODE to
enter User Mode and perform a calibration.
CANT.USE
release POWER MODE to enter User Mode and perform a calibration. Pressing AIR has no
effect. A successful calibration must be performed in order to use the instrument.
NONE
acknowledge the indication. If desired, press POWER MODE to enter User Mode and perform a
calibration. The warm-up sequence continues on its own.
1. After entering the CAL SET menu, press AIR to scroll to CAL.EXPD.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the CAL.EXPD item in the CAL SET menu.
5. See “Exiting the CAL SET Menu” on page 58 to return to User Mode.
: The SC-04 gives an indication at start up that calibration is past due. Press and
: The SC-04 gives an indication at startup that calibration is past due. It is not necessary to
Exiting the CAL SET Menu
1. While in the CAL SET menu, press AIR to scroll to ESCAPE.
2. Press and release POWER MODE to return to the CAL SET item in User Mode.
3. See “Entering Measuring Mode (START)” on page 71 to enter Measuring Mode.
58 • Chapter 5: User Mode and Calibration SC-04 Operator’s Manual
The BUMP.SET menu has 5 items: SETTING, BP.RMDR, BP.INT, BP.EXPD, and ESCAPE.
1. While in User Mode, press AIR to scroll to BUMP.SET.
2. Press and release POWER MODE. The SETTING item appears.
SETTING
The SETTING menu has 5 items: GAS.TIME, CHECK, CAL.TIME, A-CAL, and ESCAPE.
1. After entering the BUMP.SET menu, press AIR to scroll to SETTING.
2. Press and release POWER MODE. The GAS.TIME item appears.
GAS.TIME
The GAS.TIME is the amount of time that the instrument is exposed to gas during a bump test.
The available choices are 30 seconds (factory setting), 45 seconds, 60 seconds, and 90 seconds.
1. After entering the SETTING menu, press AIR to scroll to GAS.TIME.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the GAS TIME item in the
SETTING menu.
5. See “Exiting the SETTING Menu” on page 61 to return to BUMP.SET menu.
SC-0 4 Operator’s Manual Chapter 5: User Mode and Calibration • 59
CHECK is the bump test tolerance value and is represented as a percentage of the calibration gas
concentration. It is the percentage that the bump test reading can differ from the auto calibration
value and still be considered a passed bump test. If the bump test reading differs more, the bump
test fails. The available values are 10%, 20%, 30%, 40%, and 50% (factory setting).
1. After entering the SETTING menu, press AIR to scroll to CHECK.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the CHECK item in the
SETTING menu.
5. See “Exiting the SETTING Menu” on page 61 to return to BUMP.SET menu.
CAL.TIME
The CAL.TIME is the total time the instrument is exposed to calibration gas if A-CAL is set to
ON and a bump test fails. The bump test time is deducted from the calibration time. For example,
if the CAL.TIME is set to 90 seconds and the GAS.TIME is set to 30 seconds, if the bump test
fails, the SC-04 is only exposed to gas for an additional 60 seconds. The available values are 90
seconds (factory setting), and 120 seconds.
1. After entering the SETTING menu, press AIR to scroll to CAL.TIME.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the CAL.TIME item in the
SETTING menu.
5. See “Exiting the SETTING Menu” on page 61 to return to BUMP.SET menu.
60 • Chapter 5: User Mode and Calibration SC-04 Operator’s Manual
ON (factory setting): If a bump test fails, the unit automatically begins a calibration.
OFF: If a bump test fails, the unit does not automatically begin a calibration.
1. After entering the SETTING menu, press AIR to scroll to A-CAL.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the A-CAL item in the
SETTING menu.
5. See “Exiting the SETTING Menu” on page 61 to return to BUMP.SET menu.
Exiting the SETTING Menu
1. While in the SETTING menu, press AIR to scroll to ESCAPE.
2. Press and release POWER MODE to return to the SETTING item in the BUMP.SET menu.
3. See “Exiting the BUMP.SET Menu” on page 63 to return to User Mode.
BP.RMDR
ON: The SC-04 gives an indication at start up if it is due for bump testing. The type of indication
depends on the BP.EXPD setting (see page 62). If the instrument is not due for bump testing, a
check mark appears in the lower left corner of the LCD.
OFF (factory setting): The SC-04 does not give an indication at start up if it is due for bump
testing.
1. After entering the BUMP.SET menu, press AIR to scroll to BP.RMDR.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the BP.RMDR item in the BUMP.SET menu.
5. See “Exiting the BUMP.SET Menu” on page 63 to return to User Mode.
SC-0 4 Operator’s Manual Chapter 5: User Mode and Calibration • 61
This setting defines the amount of time between bump tests and can be set in 1 day increments.
The minimum setting is 0 days and the maximum setting is 30 days (factory setting).
1. After entering the BUMP.SET menu, press AIR to scroll to BP.INT.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the BP.INT item in the BUMP.SET menu.
5. See “Exiting the BUMP.SET Menu” on page 63 to return to User Mode.
BP.EXPD
This item defines what indication is given during start up when a bump test is due and BP.RMDR
is set to ON (factory setting is OFF).
CONFIRM (factory setting): The SC-04 gives an indication at start up if a bump test is past due.
Press and release AIR to continue without bump testing or press and release POWER MODE to
enter User Mode and perform a bump test.
CANT.USE: The SC-04 gives an indication at start up that a bump test is past due. Press and
release POWER MODE to enter User Mode and perform a bump test. Pressing AIR has no effect.
A successful bump test must be performed in order to use the instrument.
NONE: The SC-04 gives an indication at startup that a bump test is past due. It is not necessary to
acknowledge the indication. If desired, press POWER MODE to enter User Mode and perform a
bump test. The warm-up sequence continues on its own.
1. After entering the BUMP.SET menu, press AIR to scroll to BP.EXPD.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the BP.EXPD item in the BUMP.SET menu.
5. See “Exiting the BUMP.SET Menu” on page 63 to return to User Mode.
62 • Chapter 5: User Mode and Calibration SC-04 Operator’s Manual
4. Use AIR to adjust the Warning setpoint. The alarm setpoint limitations are shown in the table
below.
ChannelAlarm Point Limitations
Cl
0.15 ppm WARNING ALARM ALARM H 20.00 ppm
2
HCN0.9 ppm WARNING ALARM ALARM H 30.0 ppm
NH
3
NO
2
PH
3
SO
2
5. Press and release POWER MODE to save the setting.
6. Repeat Step 4 and Step 5 for the Alarm, Alarm H, STEL and TWA settings.
7. The instrument returns to the channel selection screen.
8. See “ESCAPE” on page 65 to return to User Mode.
8.0 ppm WARNING ALARM ALARM H 400.0 ppm
0.50 ppm WARNING ALARM ALARM H 20.00 ppm
0.05 ppm WARNING ALARM ALARM H 20.00 ppm
0.50 ppm WARNING ALARM ALARM H 100.00 ppm
Defaulting the Alarm Points
Defaulting the alarm points defaults them back to factory settings as outlined in Table 2 on page 8
or to the settings saved in the SAVE-AP item in Gas Select Mode if you performed a SAVE-AP
operation.
1. After entering the ALARM-P menu, press AIR to scroll to DEF.ALMP.
64 • Chapter 5: User Mode and Calibration SC-04 Operator’s Manual
3. Press POWER MODE to perform an alarm default. Press AIR to return to the DEF.ALMP
item in the ALARM-P menu.
4. The instrument asks if you’re sure you want to default the alarm points.
5. Press POWER MODE to default the alarm points. Press AIR to return to the DEF.ALMP
item in the ALARM-P menu.
6. See “ESCAPE” on page 65 to return to User Mode.
ESCAPE
1. While in the ALARM-P menu, press AIR to scroll to ESCAPE.
2. Press and release POWER MODE to return to the ALARM--P item in User Mode.
3. See “Entering Measuring Mode (START)” on page 71 to enter Measuring Mode.
Updating the Lunch Break Setting (LUNCH)
OFF (factory setting): The SC-04 automatically starts new TWA and PEAK reading collection
and resets the time in operation at startup.
ON: The Lunch Break Screen displays during startup. From this screen, you can choose to
continue accumulating TWA and PEAK readings and the time in operation from the last time the
SC-04 was used or start collecting new readings and reset the time in operation.
1. While in User Mode, press AIR to scroll to LUNCH.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the LUNCH item in User
Mode.
5. See “Entering Measuring Mode (START)” on page 71 to enter Measuring Mode.
SC-0 4 Operator’s Manual Chapter 5: User Mode and Calibration • 65
The BEEP menu has 3 items: BEEP.SEL, BEEP.INT, and ESCAPE.
1. While in User Mode, press AIR to scroll to BEEP.
2. Press and release POWER MODE. The BEEP.SEL item appears.
BEEP.SEL
BEEP.SEL defines what kind of confirmation or non-compliance indication you want to occur in
Measuring Mode. The available choices are:
OFF (factory setting): The SC-04 does not provide a confirmation alert or non-compliance
indicator.
LED: The SC-04’s LED double flashes as often as defined by the BEEP.INT parameter to verify
that the instrument is operating.
BUZZER: The SC-04’s buzzer double beeps as often as defined by the BEEP.INT parameter to
verify that the instrument is operating.
LED+BUZ: The SC-04’s LED double flashes and the buzzer double beeps as often as defined by
the BEEP.INT parameter to verify that the instrument is operating.
BMP/CAL
CONFIRM (factory setting) or NONE, the SC-04’s LED double flashes as often as defined by
the BEEP.INT parameter to indicate a non-compliance. Once a bump test or calibration
(depending on which is due) is done, the LED stop flashing.
1. While in the BEEP menu, press AIR to scroll to BEEP.SEL.
2. Press and release POWER MODE. The current setting flashes.
: If a bump test or a calibration is due and if BP.EXPD or CAL.EXPD is set to
3. Use AIR to display the desired setting.
66 • Chapter 5: User Mode and Calibration SC-04 Operator’s Manual
4. Press and release POWER MODE to save the setting and return to the BEEP.SEL item in the
BEEP .INT
BL TIME
BEEP menu.
5. See “ESCAPE” on page 67 to return to User Mode.
BEEP.INT
The BEEP.INT parameter defines how often the confirmation alert or non-compliance indicator
selected in BEEP.SEL occurs. This setting only applies if the BEEP.SEL parameter is set to
something other than OFF (factory setting). The available choices are 0.5 minutes and 1-99
minutes in 1 minute increments. The factory setting is 5 minutes.
1. While in the BEEP menu, press AIR to scroll to BEEP.INT.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the BEEP.INT item in the BEEP menu.
5. See “ESCAPE” on page 67 to return to User Mode.
ESCAPE
1. While in the BEEP menu, press AIR to scroll to ESCAPE.
2. Press and release POWER MODE to return to the BEEP item in User Mode.
3. See “Entering Measuring Mode (START)” on page 71 to enter Measuring Mode.
Updating the Backlight Time (BL TIME)
This setting defines how long the LCD backlight stays on when you press any button. The
minimum setting is OFF; the maximum setting is 255 seconds. The factory setting is 30 seconds.
1. While in User Mode, press AIR to scroll to BL TIME.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
SC-0 4 Operator’s Manual Chapter 5: User Mode and Calibration • 67
This item only appears if ZSUP.DSP is set to ON in Maintenance Mode (factory setting is OFF).
The ZERO.SUP setting is not intended for field adjustment. The default setting is ON.
Zero
Sensor
Suppression
Val ue
Cl
2
HCN0.5 ppm
NH
3
NO
2
PH
3
SO
2
0.10 ppm
4 ppm
0.30 ppm
0.02 ppm
0.20 ppm
Zero Follower (ZERO.FLW)
This item only appears if ZFLW.DSP is set to ON in Maintenance Mode (factory setting is OFF).
The ZERO.FLW setting is not intended for field adjustment. The default setting is ON.
Turning Easy Calibration On/Off (E-CAL)
OFF (factory setting): Auto Calibration (A-CAL) item appears in GAS CAL menu instead of
Easy Calibration (E-CAL).
XX seconds: Easy Calibration (E-CAL) item appears in GAS CAL menu instead of Auto
Calibration (A-CAL). During a calibration, the instrument counts down from the number of
seconds you select.
NOTE: RKI Instruments recommends an E-CAL setting of 120 seconds for Cl2 and NH3 and
60 seconds for all other gases.
1. While in User Mode, press AIR to scroll to E-CAL.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
SC-0 4 Operator’s Manual Chapter 5: User Mode and Calibration • 69
4. Press and release POWER MODE to save the setting and return to the E-CAL item in User
DATE
2020
4.21
10:40
PASS- W
Mode.
5. See “Entering Measuring Mode (START)” on page 71 to enter Measuring Mode.
Setting the Date/Time (DATE)
1. While in User Mode, place the cursor next to DATE.
2. Press and release POWER MODE. The date and time display with the year flashing.
3. Use AIR to display the desired year.
4. Press and release POWER MODE to save the setting. The month setting flashes.
5. Repeat Step 3 and Step 4 to enter the month, day, hours, and minutes settings. The date and
time are saved and the instrument returns to the DATE item in User Mode.
6. See “Entering Measuring Mode (START)” on page 71 to enter Measuring Mode.
Turning the Password On/Off (PASS-W)
ON: The SC-04 prompts you for a password when you enter User Mode. The factory-set
password is 0405 but it can be changed.
OFF (factory setting): No password is required to enter User Mode.
1. While in User Mode, press AIR to scroll to PASS-W.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. If you selected OFF, press and release POWER MODE to save the setting and return to the PASS- W item in User Mode.
If you selected ON, continue with Step 5.
70 • Chapter 5: User Mode and Calibration SC-04 Operator’s Manual
5. Press and release POWER MODE. The Set Password Screen appears. The current password
0000
PASS- W
ROM/SUM
6180
72F8
ROM/SUM
START
appears and the first digit flashes.
6. Use AIR to display a number from 0 to 9.
7. Press and release POWER MODE to enter the selection and advance to the next number. To
go back a number, press and hold AIR and POWER MODE for a few seconds.
8. Repeat Step 6 and Step 7 to select the remaining numbers. When you press and release
POWER MODE to enter the last number, the password is saved and the instrument returns to
the PASS-W item in User Mode.
9. See “Entering Measuring Mode (START)” on page 71 to enter Measuring Mode.
Viewing the ROM/SUM (ROM/SUM)
The ROM/SUM screen shows the instrument’s firmware version and firmware checksum.
1. While in User Mode, press AIR to scroll to ROM/SUM.
2. Press and release POWER MODE. The screen shows the ROM/SUM. The ROM is the top
value and the SUM is the bottom value.
3. Press and release POWER MODE to return to the ROM/SUM item in User Mode.
4. See “Entering Measuring Mode (START)” on page 71 to enter Measuring Mode.
Entering Measuring Mode (START)
1. While in User Mode, press AIR to scroll to START.
2. Press and release POWER MODE. The instrument begins its warmup sequence.
SC-0 4 Operator’s Manual Chapter 5: User Mode and Calibration • 71
This chapter describes troubleshooting procedures for the SC-04. It also includes procedures for
replacing the batteries and replacing various consumable parts.
WARNING: RKI Instruments recommends that service, calibration, and repair of RKI gas
detectors be performed by personnel properly trained for this work. Replacing the
sensor and other parts with original equipment does not affect the intrinsic safety
of the instrument.
AVERTISSEMENT: RKI Instruments recommande que le service, l'étalonnage et la
réparation des détecteurs de gaz RKI soient effectués par du personnel
dûment formé à ces travaux. Le remplacement du capteur et d'autres
pièces par l'équipement d'origine n'affecte pas la sécurité intrinsèque de
l'instrument.
Troubleshooting
The troubleshooting table describes error messages, symptoms, probable causes, and
recommended action for problems you may encounter with the SC-04.
• The LCD shows an
abnormally high
reading but other gas
detection instruments
• Sensor filter needs to be
replaced.
• The SC-04 needs to be
recalibrated.
do not.
• The sensor needs
replacement.
• Calibration fails. • The calibration value does
not match the cylinder gas
concentration.
• The sample gas is not
reaching the sensor because
of a bad connection.
• The calibration cylinder is
out of gas or is outdated.
• The sensor needs
replacement.
1. Replace the sensor filter.
2. Recalibrate the unit.
3. If the difficulties continue, replace
the sensor and calibrate.
4. Allow instruments with NH
3
sensors 2 hours of sensor warmup
prior to operation if the sensor or
batteries are newly installed or if the
unit’s batteries have been
completely discharged.
5. If the difficulties continue, contact
RKI Instruments, Inc. for further
instruction.
1. Make sure the SC-04 has been
properly set up for calibration.
2. Check all calibration tubing for
leaks or for any bad connections.
3. Verify that the calibration cylinder
contains an adequate supply of fresh
test sample.
4. If the fail condition continues,
replace the sensor.
5. If the difficulties continue, contact
RKI Instruments, Inc. for further
instruction.
• Heart symbol at the
top of the screen
becomes steadily on
or disappears.
• A microprocessor error has
occurred.
• Contact RKI Instruments, Inc. for
further instruction.
WARNING: To prevent ignition of a hazardous atmosphere, batteries must only be changed in
an area known to be nonhazardous.
AVERTISSEMENT: Pour éviter l'inflammation d'une atmosphère dangereuse, la batterie ne
doit être remplacée que dans une zone non dangereuse.
WARNING: Use only Duracell MN2400 or PC2400 or Eneloop BK-4MCC batteries to
maintain the QPS classification of the SC-04. Use of other batteries will void the
QPS classification and may void the warranty. Do not mix old/new or different
types of batteries.
AVERTISSEMENT: Utiliser uniquement des piles Duracell MN 2400 ou PC 2400 ou
Eneloop BK-4MCC de maintenir la classification QPS de la SC-04.
L’utilisation d’autres piles annule la classification QPS et peut annuler
la garantie. Ne mélangez pas les anciennes/nouvelles ou différents types
de piles.
NOTE: If old batteries are completely out of charge (the unit does not power on), instruments
with NH3 sensors require 2 hours for sensor warmup before turning the instrument on
for normal operation.
Replace the batteries when the last bar in the battery icon is flashing.
1. Make sure the SC-04 is off.
2. Release the side of the alligator or belt clip that is opposite the hinge. You may need to o a
screwdriver to pry it open.
7. Reinstall the battery cover and tighten the battery cover screw.
www.GlobalTestSupply.com
8. If the instrument has been without batteries for more than 5 minutes, the date and time are
reset and need to be set again. When the new batteries are installed, the instrument
automatically turns on and displays the Date/Time Screen. Set the date and time as described
on page 70. Once the date and time are set, the instrument begins its warmup sequence. If you
do not set the date and time within 30 seconds, the instrument automatically begins its
warmup sequence.
Recharging the Batteries (Ni-MH Batteries Only)
Any battery charger capable of charging AAA Ni-MH batteries can be used to recharge the SC04’s Ni-MH batteries. RKI Instruments, Inc. recommends using one of the chargers specified on
page 83.
WARNING: To prevent ignition of a hazardous atmosphere, batteries must only be charged in
an area known to be nonhazardous.
NOTE: If old batteries have been completely discharged (the unit does not power on),
instruments with NH3 sensors require 2 hours for sensor warmup before turning the
instrument on for normal operation.
1. Remove the batteries from the SC-04 as described in Step 1 - Step 5 on page 74.
2. Install the Ni-MH batteries in the charger. See the battery charger’s manual for charging
instructions.
3. Put the batteries back in the SC-04 and reinstall the battery cover.
Replacing the Sensor Filter
NO2 and SO2 Sensors’ Filter: The H2S filter disk is tan in color. The filter should be replaced
every 6 months.
HCN Sensor Filter: The H2S filter disk is dark gray in color. The filter should be replaced every 6
months.
Sensor Filter: The humidity filter disk is white in color. The filter should be replaced every 6
PH
3
months.
NH3 Sensor Filter: The humidity filter is white in color. The filter should be replaced every 6
months.
8. The sensor filter sits in the middle of a black gasket. Pry the sensor filter out of the gasket.
Figure 13: Sensor Filter Location
Hydrophobic Filter
Sensor Gasket
Sensor Gasket
Sensor Filter
Sensor Filter
Front Case
FrontCase Inside ViewInstrument Side View
Rear Case
9. Install a new sensor filter. If the gasket got removed, install it with the flat side facing the
white hydrophobic filter.
10. If necessary, put the sensor in the socket on the rear case. The colored side faces away from
the rear case and the slots in the sensor line up with tabs in the socket.
CAUTION:Forcing a sensor into a socket without lining the sensor slots up with the
socket tabs may damage the sensor or socket.
11. Reinstall the front case to the rear case.
12. Reinstall the 4 screws that were removed in Step 3.
11. Reinstall the sensor gasket with the flat side facing the white hydrophobic filter.
12. If the sensor filter came loose, reinstall it in the middle of the sensor gasket.
13. If necessary, put the sensor in the socket on the rear case. The colored side faces away from
the rear case and the slots in the sensor line up with tabs in the socket.
CAUTION:Forcing a sensor into a socket without lining the sensor slots up with the
socket tabs may damage the sensor or socket.
14. Reinstall the front case to the rear case.
15. Reinstall the 4 screws that were removed in Step 3.
This appendix describes the SC-04 in Maintenance Mode. The SC-04 is factory-set to suit most
applications. Update settings in Maintenance Mode only if required for your specific application.
Maintenance Mode items and their factory settings are listed in Table 15 below.
Table 15: Maintenance Mode Menu Items
Maintenance
Mode
Menu Item
Description
GAS CAL (page
89)
GAS.TEST
(page 89)
SEN.DATE
(page 91)
Perform an air adjust, perform a span adjustment, change the calibration values.
AIRPerform a fresh air adjustment.
A-CAL or
E-CAL
depending
on E-CAL
User Mode
setting
ESCAPEReturn to the GAS CAL item in Maintenance Mode.
Apply gas to test sensor response and observe alarm indications without an alarm
event being recorded.
View and/or set the replacement date for the sensor and the batteries.
Perform a span adjustment and set the calibration gas concentra-
tion.
A-CAL (or
E-CAL)
STARTBegin the warmup sequence and enter Measuring
CAL-PSet the calibration gas concentration.
ESCAPEReturn to the A-CAL item in the GAS CAL menu.
Press and release POWER MODE to begin the warmup sequence and enter Mea-
suring Mode.
Entering Maintenance Mode
WARNING: The SC-04 is not in operation as a gas detector while in Maintenance Mode.
1. Take the SC-04 to a non-hazardous location and turn it off if it is on.
2. Press and hold AIR, then press and hold POWER MODE. You will hear a beep after one
second. Continue to hold the buttons down.
3. When you hear a second beep, release the buttons.
4. The screen that appears depends on the setting of Maintenance Mode’s PA SS -W item.
If PASS-W is set to OFF, continue with Step 8.
If PASS-W is set to ON (factory setting), continue with Step 5.
5. If PAS S- W is set to ON in Maintenance Mode, a password screen appears and the first digit is
flashing. The factory-set password is 0400 but it can be changed.
6. Use AIR to select each password number then press POWER MODE to save it and move on
to the next number. To go back a number, press and hold AIR and POWER MODE for a few
seconds. To reverse the direction of change (ie. from increasing to decreasing or vice versa):
a. Press and hold AIR.
b. Immediately press POWER MODE and then release both buttons.
9. Use AIR to move through the Maintenance Mode items.
Tips for Using Maintenance Mode
•To scroll from one item to the next, press and release AIR. To reverse the scrolling direction:
a. Press and hold AIR.
b. Immediately press POWER MODE and then release both buttons.
c. The scrolling direction returns to the original direction when you exit and reenter a menu.
•To skip an item when a question is asked, press and release AIR.
•To enter an item and to save any changes, press and release POWER MODE.
•To change a flashing parameter, press and release AIR. To reverse the direction of change (ie.
from increasing to decreasing or vice versa):
a. Press and hold AIR.
b. Immediately press POWER MODE and then release both buttons.
•To exit an entered item without saving a change, press and hold AIR and POWER MODE for a
few seconds.
Performing a Calibration (GAS CAL)
See page 46 for a description of the GAS CAL item.
Performing a Gas Test (GAS.TEST)
The GAS.TEST item allows you to apply gas to the instrument and see all alarm indications
except
though the buzzer sounds in the event of a real gas alarm condition while in Measuring Mode.
A gas test is not saved in logged data.
Preparing for a Gas Test
To perform a gas test on the SC-04, you need:
for the buzzer indication. There is no buzzer indication in the GAS.TEST menu even
•A calibration gas cylinder. The concentrations should be above the alarm condition you
want to check. Standard alarm points are listed on page 8.
1. Install the calibration cup onto the SC-04. Be sure the calibration cup is installed in the correct
POWER
MODE
AIR
Front ViewSide View
GAS .TEST
SO2
0.00
GAS .TEST
ppm
SO2
0.00
BUZZ .OFF
ppm
direction and that it is pushed on all the way.
2. Use the tubing to connect the regulator to the inlet of the calibration cup (labeled “IN”).
Performing a Gas Test
1. While in Maintenance Mode, press AIR to scroll to GAS.TEST.
2. Press and release POWER MODE. The current gas reading displays. The bottom of the LCD
alternates between “GAS.TEST” and “BUZZ.OFF”.
3. For toxic gas cylinders, it is important to vent the regulator while installing it onto the
cylinder. Venting the regulator during installation helps prevent air from getting into the
cylinder and degrading the gas. Open the regulator by turning the knob counterclockwise and
install it onto the cylinder.
4. The instrument initiates alarm indications except
in the GAS.TEST menu even though the buzzer sounds in the event of a real gas alarm
condition.
5. Turn the regulator knob clockwise to close the regulator.
6. Unscrew the regulator from the calibration cylinder.
ON: The SC-04 asks if you want to perform a fresh air adjustment at the end of the startup
sequence.
OFF (factory setting): The SC-04 does not ask if you want to perform a fresh air adjustment at the
end of the startup sequence.
1. While in Maintenance Mode, press AIR to scroll to A.ZERO.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the A.ZERO item in
Maintenance Mode.
5. See “Entering Measuring Mode (START)” on page 97 to enter Measuring Mode.
Turning the ID Display Function On/Off (ID DISP)
ON: The User ID and Station ID screens appear in the startup sequence. If DISP.SET in User
Mode is also set to ON, then the IDs can be changed in Display Mode.
OFF (factory setting): The User ID and Station ID screens do not appear in the startup sequence
and the IDs cannot be changed in Display Mode.
1. While in Maintenance Mode, press AIR to scroll to ID DISP.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the ID DISP item in
Maintenance Mode.
5. See “Entering Measuring Mode (START)” on page 97 to enter Measuring Mode.
ON: CYL SEL item appears in User and Maintenance Modes’ GAS CAL item.
OFF (factory setting): CYL SEL item does not appear in User and Maintenance Modes’ GAS
CAL item.
Setting the Date/Time (DATE)
1. From the main menu, place the cursor next to DATE.
2. Press and release POWER MODE. The date and time display with the year flashing.
3. Use AIR to display the desired year.
4. Press and release POWER MODE to save the setting. The month setting flashes.
5. Repeat Step 3 and Step 4 to enter the month, day, hours, and minutes settings. The date and
time are saved and the instrument returns to the DATE item in Maintenance Mode.
6. See “Entering Measuring Mode (START)” on page 97 to enter Measuring Mode.
Turning the Password On/Off (PASS-W)
ON (factory setting): The SC-04 prompts you for a password when you enter Maintenance Mode.
The factory-set password is 0400 but it can be changed.
OFF: No password is required to enter Maintenance Mode.
1. While in Maintenance Mode, press AIR to scroll to PASS -W.
2. Press and release POWER MODE. The current setting flashes.
4. If you selected OFF, press and release POWER MODE to save the setting and return to the
0400
PASS- W
ROM/SUM
6180
72F8
ROM/SUM
PASS- W item in User Mode.
If you selected ON, continue with Step 5.
5. Press and release POWER MODE. The Set Password Screen appears. The current password
displays and the first digit flashes.
6. Use AIR to display a number from 0 to 9.
7. Press and release POWER MODE to enter the selection and advance to the next number. To
go back a number, press and hold AIR and POWER MODE for a few seconds.
8. Repeat Step 6 and Step 7 to select the remaining numbers. When you press and release
POWER MODE to enter the last number, the password is saved and the instrument returns to
the PASSWORD item in Maintenance Mode.
9. See “Entering Measuring Mode (START)” on page 97 to enter Measuring Mode.
Viewing the ROM/SUM (ROM/SUM)
The ROM/SUM screen shows the instrument’s firmware version and firmware checksum.
1. While in Maintenance Mode, press AIR to scroll to ROM/SUM.
2. Press and release POWER MODE. The screen shows the ROM/SUM. The ROM is the top
value and the SUM is the bottom value.
3. Press and release POWER MODE to return to the ROM/SUM item in Maintenance Mode.
4. See “Entering Measuring Mode (START)” on page 97 to enter Measuring Mode.
This appendix describes the SC-04 in Gas Select Mode. The SC-04 is factory-set to suit most
applications. Update settings in Gas Select Mode only if required for your specific application. A
description of the Gas Select Mode items is shown in Table 16 below.
Table 16: Gas Select Mode Menu Items
Menu Item (Page # of
Description)
SAVE-AP (page 99)Set the current alarm points as the default alarm points.
MAX.SPAN (page 100)ON: Maximum span screen appears after a successful calibration.
STEALTH (page 101)STEALTH ON
START (page 101)Enter Measuring Mode
Description
OFF (factory setting)
STEALTH OFF
normally.
This setting has no effect unless STEALTH is set to ON.
VIB ON
VIB OFF
: Vibrator activates for alarm conditions.
(factory setting): Vibrator does not activate in any situation.
: No maximum span screen appears.
: No backlight, LED, or buzzer operation.
(factory setting): Backlight, LED, and buzzer operate
Entering Gas Select Mode
WARNING: The SC-04 is not in operation as a gas detector while in Gas Select Mode.
1. Take the SC-04 to a non-hazardous location and turn it off if it is on.
2. Press and hold AIR, then press and hold POWER MODE. You will hear a beep after one
second. Continue to hold the buttons down.
3. You will hear a second beep. Continue to hold the buttons down.
4. When you hear a third beep, release the buttons.
5. A password screen appears and the first digit flashes. The password is 2014.
98 • Appendix B: Gas Select Mode S C- 04 Operator’s Manual
6. Use AIR to select each password number then press POWER MODE to save it and move on
SAVE - AP
SAVE - AP
to the next number. To go back a number, press and hold AIR and POWER MODE for a few
seconds.
7. The SAVE-AP item displays.
8. Use AIR to move through the Gas Select Mode items.
Tips for Using Gas Select Mode
•To scroll from one item to the next, press and release AIR. To reverse the scrolling direction:
a. Press and hold AIR.
b. Immediately press POWER MODE and then release both buttons.
c. The scrolling direction returns to the original direction when you exit and reenter a menu.
•To skip an item when a question is asked, press and release AIR.
•To enter an item and to save any changes, press and release POWER MODE.
•To change a flashing parameter, press and release AIR. To reverse the direction of change (ie.
from increasing to decreasing or vice versa):
a. Press and hold AIR.
b. Immediately press POWER MODE and then release both buttons.
•To exit an entered item without saving a change, press and hold AIR and POWER MODE for a
few seconds.
Saving the Alarm Points (SAVE-AP)
Performing a SAVE-AP operation saves the current alarm setpoints.
Performing a DEF.ALMP operation in the ALARM-P User Mode menu sets the instrument’s
alarm points to those saved during the SAVE-AP operation (if performed).
Performing a SAVE-AP has no effect on an M.DEF in Maintenance Mode. An M.DEF operation
returns all instrument settings to the RKI default regardless of if a SAVE-AP operation was
performed.
1. While in Gas Select Mode, press AIR to scroll to SAVE-AP.
2. Press and release POWER MODE. The display cycles through the following screens.
SEL
SAVE - AP
SEL
YES .MODE
SEL
NO .AIR
MAX .SPAN
3. Press and release POWER MODE to save the current alarm point settings as the default.
4. The instrument returns to the SAVE-AP item in Gas Select Mode.
5. See “Exiting Gas Select Mode (START)” on page 101 to enter Measuring Mode.
Turning Calibration Max Span On/Off (MAX.SPAN)
ON: After a passed calibration, the SC-04 displays the response reading’s maximum adjustment.
A maximum span of 100 ppm indicates that the reading could have been adjusted up to 100 ppm.
If the maximum span value is close to the calibration value, the sensor should be replaced soon.
The upper limit on the maximum adjustment indicated is either twice the calibration value or full
scale, whichever is lower.
OFF (factory setting): There is no maximum span indication at the end of a calibration.
1. While in Gas Select Mode, press AIR to scroll to MAX.SPAN.
2. Press and release POWER MODE. The current setting flashes.
3. Use AIR to display the desired setting.
4. Press and release POWER MODE to save the setting and return to the MAX.SPAN item in
Gas Select Mode.
5. See “Exiting Gas Select Mode (START)” on page 101 to enter Measuring Mode.