Read and understand this instruction manual before
operating instrument. Improper use of the gas monitor
could result in bodily harm or death.
Periodic calibration and maintenance of the gas
monitor is essential for proper operation and correct
readings. Please calibrate and maintain this instrument
regularly! Frequency of calibration depends upon the
type of use you have and the sensor types. Typical
calibration frequencies for most applications are
between 3 and 6 months, but can be required more
often or less often based on your usage.
65-2640RK-03 M2A Transmitter Operator’s Manual
Product Warranty
RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in
materials, workmanship, and performance for a period of one year fr o m date of shipment
from RKI Instruments, Inc. Any parts found defective withi n tha t period will be repaired
or replaced, at our option, free of charge. This warranty does not apply to those items
which by their nature are subject to deterioration or consumption in normal ser v ice, and
which must be cleaned, repaired, or replaced on a routine basis. Examples of such items
are:
W arranty is voided by abuse including mechanical damage, alteration, rough handling, or
repair procedures not in accordance with the operator’s manual. This warranty indicates
the full extent of our liability , a nd we are not r esponsible for removal or r eplacement costs,
local repair costs, transportation costs, or contingent expenses incurred without our prior
approval.
a) Absorbent cartridgesd) Batteries
b) Pump diaphragms and valvese) Filter elements
c) Fuses
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER
WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED,
AND ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF
RKI INSTRUMENTS, INC. INCLUDING BUT NOT LIMITED TO, THE
WARRANTY OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE. IN NO EVENT SHALL RKI INSTRUMENTS,
INC. BE LIABLE FOR INDIRECT, INCIDENTAL, OR CONSEQUENTIAL
LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF
ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR
OPERATE PROPERLY.
This warranty covers instruments and parts sold to users by authorized distributors,
dealers, and representatives as appointed by RKI Instruments, Inc.
We do not assume indemnificatio n f o r any accident or da ma g e c au sed by the operat io n of
this gas monitor, and our warranty is limited to the replacement of parts or our complete
goods.
65-2640RK-03 M2A Transmitte r Oper ato r’ s Manual
(Appendix C available from RKI Instruments, Inc. Not included in manual as normally
provided with M2A detector head.)
65-2640RK-03 M2A Transmitte r Oper ato r’ s Manual
Chapter 1: Introduction
Overview
This chapter briefly describes the 65-2640RK-03 M2A Transmitter. This chapter also
describes the 65-2640RK-03 M2A Transmitter Oper ator’s Manual (this document). Table 1 at
the end of this chapter lists the specifications for the M2A.
About the M2A Transmi tter
The M2A transmitter is a fixed mount, continuous-monitoring detector head. All user
adjustable parameters may be accessed using push button switches. In addition,
calibration may be performed non-intrusively by use of a magnetic wand accessory which
activates magnetic switches through a window at the front face of the detector head.
The M2A displays the current gas reading on an OL E D d isplay which is visible through
the window in the cover (front face) and provides a 4 - 20 mA sign al which indicates the
target gas reading for use by a gas monitoring controller, recording device, or
programmable controller. The M2A also pro vides an RS- 485 Mo dbus output. Thr ee sets of
relay contacts, two controlled by the gas alarms and one by the fail alarm, rated at 115
VAC 5 amps are available for controlling devices such as lights or horns or for controlling
higher rated relays.
Four operating modes allow you to display and change setup and calibration settings and
change the gas type. They are Calibration Mode, Configuration Mode, Gas Type Mode,
and Modbus Mode.
About this Manual
The M2A Transmitter Operator’s Manual uses the following conventions for notes, cautions,
and warnings.
NOTE:Describes additional or critical info rmation.
CAUTION:Describes pote ntial damage to equip ment.
WARNING:Describes potential danger that can result in injury or death.
165-2640RK-03 M2A Transmitter Operator’s Manual
Specifications
Table 1 lists specifications for the M2A.
Table 1: M2A Specifications
Target Gas/Detection RangeCombustible Gas (methane calibration standard):
0 - 100% LEL, 1% LEL increments
Alarm Settings
10 %LEL/50 %LEL
(Alarm 1/Alarm 2)
Construction (housing)Explosion-proof Junction Box, NEMA 4
Area ClassificationExplosion-proof for Class I, Groups B, C, and D
Sampling MethodDiffusion
Input Power 10 - 30 VDC
Controls• Three push button switches
• Three magnetic switches for non-intrusive calibrati on
Weight5.5 lbs.
Signal Output
•4 to 20 mA with impedance maximums listed below
-12 VDC input: 500 ohms impedance max
-24 VDC input: 1000 ohms impedance max
• RS-485 Modbus
Operating Temperature-40°C to 50°C
Accuracy± 5% of reading or ± 2% LEL (whichever is greater)
*
These are the RKI factory settings. You can change the alarm settings in Configuration Mode. See
“Vi ew ing & Changi n g M2A Para mete rs” on page 23.
WARNING:When using the M2A, you must follow the instructions and war ni ngs in
this manual to assure proper and safe operation of the M 2A and to
minimize the risk of personal injury. Be sure to maintain and periodically
calibrate the M2A as described in this manual.
65-2640RK-03 M2A Trans mitter Operator’s Manual2
Chapter 2: Description
Overview
This chapter describes the components of the M2A transmitter. The transmitter is a 4 - 20
mA type detector head. It consists of the combustible gas detector, calibration adapter,
terminal PCB, the M2A junction box, and the detector junction box. The two junction box
configuration is intended for situations where the detector needs to be installed at an
inaccessible location. The detector junction box can be installed at the inaccessible location
and the M2A junction box can be installed in a more readily accessible area.
External Description
This section describes the junction boxes and all external components of the M2A
transmitter.
Mounting Slot (2x)
Magnetic
Wand
3/4" NPT Conduit
Hub, Plugged
3/4" NPT Conduit Hub
Terminal
Strip
Window
3/4" NPT
Conduit
M2A
Junction
Box
Combustible Gas
3/4" NPT
Conduit Hub
Calibration Adapter
Figure 1: M2A External Components
365-2640RK-03 M2A Transmitter Operator’s Manual
Hub
Detector
Junction Box
Detector
M2A Junction Box
The M2A’ s cast aluminum junction box protects the terminal PCB and wiring connections
made to the terminal PCB. Use the three 3/4 in. conduit hubs to wire the remotely installed
combustible gas detector and connect wiring from an external device, typically an RKI controller.
The top hub is shipped with a factory installed and sealed conduit plug to avoid leaks into the
junction box. If necessary, the conduit plug can be removed and the top hub can be used for wiring.
WARNING:If the top conduit hub is used for wiring, be sure to seal the threads to e nsure
water does not enter the junction box. See “Wiring the M2A Transmitter” on
page 11 for complete wiring instructions.
Use the junction box’s two mounting slots to mount the M2A to a vertical surface at the
monitoring site. The window in the cover on the front of the junction box allows you to
view the OLED display and use the magnetic wand to actuate the magnetic control
switches so you can perform non-intrusive calibration. Removing the cover allows you to
access the interior of the junction box.
Detector Junction Box
The detector junction box is intended to be installed remotely from the M2A junction box.
The combustible gas detector is factory installed in the bottom 3/4” NPT conduit hub
using a 3/4 in. x 1/2 in. NPT reducer. A terminal strip is provided for wiring connections
and the detector is factory wired to one side of this terminal strip. The other side of the
terminal strip is for user-supplied wiring to the detector terminal strip located in the M2A
junction box. This wiring should go through the top 3/4” conduit hub. Three spacers
installed on the back of the junction box controls the di stance of the junction box from a
mounting surface.
Combustible Gas Detector
The combustible gas detector is a catalytic type detector that produces an electrical output
that corresponds to the detection range. It is packaged in a 1/2 inch NPT nipple with a
sintered metal flame arrestor on one end allowing ambient air to diffuse into the detector.
The flame arrest or al so co ntai n s any s park s whi ch may occur within the detector. The 1/2
inch NPT mounting threads at the top of the detector allow you to mount it into the
bottom conduit hub of the junction box. The junction box conduit hub is 3/4” NPT, so a
3/4 in. x 1/2 in. NPT reducer is used to install the combustible gas detector. Four colorcoded leads extend from the top of the detector. The leads allow you to connect the
detector to the terminal strip.
Calibration Adapter
A calibration adapter is installed on the combusti ble gas detector. A fitting at th e botto m
of the calibration adapter allows a length of tubing to be connected for calibration and
routed to a more easily accessible location (near the M2A junction box is optimal for ease
of calibration).
Magnetic Wand
The magnetic wand is a short plastic rod with a magnet in one end. It is used to actuate
the magnetic control switches on the control PCB while the junction box cover is still
installed so that non-intrusive calibration can be performed.
65-2640RK-03 M2A Trans mitter Operator’s Manual4
Internal Description
This section describes the internal components of the M2A. The internal components of
the M2A include the terminal PCB which provides for all the wiring connections to the
M2A and the control PCB which displays the gas reading and has the control buttons.
3/4" NPT Conduit Opening, Plugged
Power/Signal
Terminal Strip
Detector Terminal Strip
(Comb/CO2 Version Shown)
Detector Current
(Comb/CO2 TypeOnly,
Adjust Pot
Factory Adjust)
3/4" NPT
Conduit
Opening
forWire
Entry
Ground Jumper
Programming
Connector
(Factory Use)
Modbus Terminal Strip
Termination Jumper
Ribbon Cable Connector
Fail LED
Alarm 1 LED
Alarm 2 LED
RX LED
TX LED
Relay
Terminal
Strips
Junction Box
Terminal PCB
Control PCB
Magnetic
Switches
ENTER
DOWN/NO
- S +
LEL
R W G B
RS 485PWR/SIG
A B C
FAIL
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
UP/YES
Push Button
Control
NO
RKI INSTRUMENTS
M2A TRANSMITTER
Switches
UP/YES
DOWN/NO
ENTER
OLEDDisplay
View With Cover and
Control PCB Remo ved
View With Cover Removed
Figure 2: M2A Internal Comp onents
Terminal PCB
The terminal PCB is encapsulated in epoxy for protection against moisture and physical
damage. It is mounted into the rear of the junction box with three standoffs and rests on a
thin layer of foam. A banana jack is screwed into each of the standoffs and used for
mounting the control PCB. The terminal PCB converts the electrical output from the
detector to a signal which can be displayed by the OLED display, a 4 - 20 mA signal (that
is proportional to the detection range), and an RS-485 Modbus output signal. The 4 - 20
mA signal may be used by a recording device, gas monitor controller, or programmable
controller. The Modbus output may be used to connect the M2A to a Modbus network.
The terminal PCB also controls three relays, one fail and two gas alarm relays.
Two columns of plug-in style terminal strips are used to make all wiring connections to
the M2A. The column on the left consists of the power/signal, detector, and Modbus
terminal strips. The column on the right consists of the relay terminal strips. A 20 position
connector at the bottom of the terminal PCB is used to connect the terminal PCB to the
control PCB with a ribbon cable. A 5 position connector on the left side of the terminal
PCB is used by factory or field service personnel to program the M2A. A factory adjust pot
just above the programming connector is used to set the detector current.
565-2640RK-03 M2A Transmitter Operator’s Manual
Power/Sign al Terminal Strip
The power/signal terminal strip is a three position plug-in style terminal strip located at
the top of the left terminal column. It is used to connect 24 VDC power to the M2A and to
connect the 4 - 20 mA output signal to a device.
The signal output, the S terminal, does not have to be connected for the M2A to function.
The S terminal is used if you want to connect the 4 - 20 mA output signal to another device
such as a gas monitoring controller, chart recorder, or programmable controller (PLC).
Detector Terminal Strip
The detector terminal strip is a four position plug-in style terminal strip and is the middle
terminal strip in the left terminal column. All four terminals are used to connect the
detector to the M2A.
Detector
Terminal
Strip
- S +
LEL
R W G B
RS 485PWR/SIG
A B C
C NC NO
C NC NO
C NC NO
FAIL
ALARM 1
ALARM 2
Figure 3: Detect o r Term inal Strip
NOTE:The detector is factory-wired to the M2A. See “Wiring the M2A T ransmitter” on
page 11 for all wiring procedures related to M2 A.
Modbus Terminal Str ip
The Modbus terminal strip is a three position plug-in style terminal strip and is the
bottom terminal strip in the left terminal column. It allows connection of the M2A in to a
Modbus network.
Relay Terminal Strips
The right column of terminal strips consists of, from top to bottom, the fail, alarm 1, and
alarm 2 relay terminal strips. The y are three-positio n plug-in style terminal strips. The
relay terminal strips are used to connect devices such as lights and horns that are
controlled by the relay contacts. The relay contacts are rated at 115 VAC, 5 amps. The relay
contacts may also be used to control higher rated relays.
Termination Jumper
A two pin header with a termination jumper installed is located below the Modbus
terminal strip. Leave this jumper installed unless directed to remove it for a Modbus
65-2640RK-03 M2A Trans mitter Operator’s Manual6
installation. See “Chapter 8: RS-485 Modbus Output” on page 37 for a description of using
the M2A in a Modbus system.
Ground Jumper
A two pin header with a jumper installed is located along the left side of the terminal PCB.
Leave this jumper installed unless directed to remove it for a Modbus installation. See
“Chapter 8: RS-485 Modbus Output” on page 37 for a description of using the M2A in a
Modbus system.
Control PCB
The OLED display and control switches are located on the control PCB. It is installed on
top of the terminal PCB by lining up its three spacing standoffs with th e ban an a j ack s in
the terminal PCB mounting standoffs and pushing it onto the banana jacks. The jacks
retain the control PCB.
OLED Display
The OLED display is located at the top of the control PCB. It indicates the current gas
reading and displays messages and parameters in the M2A’s programs.
Control Buttons
The M2A includes three push button switches that allow you to enter the M2A ’s operating
modes, navigate through the modes, update settings, an d save changes to the settings.
The push button switches are located along the bottom edge of the control PCB (see
Figure 2). The UP/YES button is on the left, the DOWN/NO button is in the middle, and
the ENTER button is on the right.
Table 2: M2A Control Button Functions
SwitchFunction
UP (YES)•Saves settings
•Changes the displayed set ting
•Enters the Calibration Program
•Enters Gas Type Mode (press with DOWN/NO button)
•Enters Configuration Mode (pres s with EN TER button )
DOWN (NO)•Cancels setting changes
•Changes the displayed set ting
•Enters Gas Type Mode (press with UP/YES button)
•Enters Modbus Mode (press with ENTER button )
•Displays the Information Screen
ENTER•Initiates operations
•Enters Configuration Mode (press with UP/YES button)
•Enters Modbus Mode (press w ith DOW N/N O butt on)
•Functions as an alarm reset switc h
Just above each push button switch is a magnetic switch with the same function as th e
push button switch below it. The magnetic switches are for use in non-intrusive
calibration. They are actuated by bringing the magnetic wand close enough to them to
actuate them. Although the magnetic switches have the same functions as the push button
switches, it is not practical to use them for operations other than calibration because it is
not possible to actuate two magnetic switches at once with only one magnetic wand. Since
displaying the Information Screen only requires the use of one switch, the wand may be
used to show the Information Screen (see “Information Screen” on page 18).
765-2640RK-03 M2A Transmitter Operator’s Manual
Status LEDs
The M2A includes five status LEDs that are located above the display (see Figure 2).
•Fail LED
The fail LED turns on when the M2A is experiencing a fail condition. A fail condition
can be caused by a detector failure or low detector signal.
•Alarm 1 LED
The alarm 1 LED is on when the M2A is experiencing an alarm 1 condition.
•Alarm 2 LED
The alarm 2 LED is on when the M2A is experiencing an alarm 2 condition.
•RX & TX LED’s
These LED’s indicate data being received (RX) and transmitted (TX) when the M2A’s
Modbus output is operating.
65-2640RK-03 M2A Trans mitter Operator’s Manual8
Chapter 3: Installation & Startup
b
Overview
This chapter describes procedures to mount the M2A Transmitter in the monitoring
environment and wire it to input power and devices.
Mounting the M2A Transmitter
1.Select a mounting site that is representative of the monitoring environment. Consider
the following when you select the mounting site.
•For the M2A junction box, select a site where the junction box is not likely to be
bumped or disturbed. Make sure there is sufficient room to perform start-up,
maintenance, and calibration procedures. The site should be easily a ccessible.
•For the detector junction box, select a site where the target gas is likely to be found
first. For lighter gases, mount the detector near the ceiling; for heavier gases,
mount the detector near the floor.
5.86
•The junction boxes should be mounted no further than 200 feet from each other.
2.Use #10 screws through the M2A junction box’s two mounting slots to secure the
junction box to a vertical surface. Figure 4 shows the outline and mountin g
dimensions the M2A junction box. Figure 5 shows the outline and mounting
dimensions for the detector junction box. Mount the detector junction box with the
detector facing down as shown in Figure 5.
3.Install 3/16 inch I.D. flexible po lyurethane tubing to the fitting at the bottom of the
calibration adapter and route it to an accessible area that is close to the M2A junction
box. See the “Parts List” on page 47 for available tubing.
65-2640RK-03 M2A Trans mitter Operator’s Manual10
Wiring the M2A Transmitter
WARNING:Always verify that the power to the M2A is off before you make wiring
connections.
1.Confirm that no power is being applied to the M2A.
2.Remove the detector junction box’s cover from the junction box.
3.Remove the M2A junction box’s cover from the junction box.
4.Grasp the control PCB by its edges.
5.Gently pull until the control PCB is pulled away from the banana jacks. Take care not
to pull too hard and damage the cable which connects the control and terminal PCBs.
6.Let the control PCB hang by the cable. The terminal strips are now visible on the
terminal PCB. The control PCB may be left hanging while wiring is done. If desired,
the control PCB may be disconnected from the cable and set aside while wiring.
7.Guide multi conductor shielded cable or cables or wires in conduit through the conduit hub on
the left side of the junction box. To make wiring more convenient, use wire colors that
correspond to the detector wire colors: red, white, green, and black.
WARNING:If the top conduit hub is used for wiring, be sure to seal the threads to e nsure
water does not enter the junction box. A conduit seal, which is normally
required to maintain the junction box’ s hazardous location rating, will also help
to prevent water from entering through the top port.
NOTE: Consult RKI Instruments, Inc. for cable lengths longer than 200 feet.
8.Connect the detector leads to the terminal block in the detector junction box.
9.Secure the detector junction box’s cover to the junction box.
10. Guide the other end of the four-conductor, shielded cable or four wires in conduit
through the bottom conduit hub of the M2A junction box.
11. To gain access to a plug-in termina l strip for wiring, pull it out of its socket by
grasping the wall between two terminal positions with needle nose pliers and pulling.
Be careful not to exert too much force on the wall to avoid damage to the terminal
strip. The detector terminal strip and th e alarm 1 r elay t ermi nal strip ma y be r emo ved
by grasping them with your fingers if the adjacent terminal strips have been removed.
WARNING:If the M 2A is installed in a hazardous location, use appropriately rated
conduit, conduit fittings, and appropriate construction technique that
complies with the local electrical code.
In addition, use appropriately rated conduit, conduit fittings, and
appropriate construction technique to maintain the NEMA 4
environmental rating of the junction box and avoid water entering the
junction box at the conduit hubs.
1165-2640RK-03 M2A Transmitter Operator’s Manual
12. Pull out the detector terminal strip and connect the four wires to the terminal strip as
follows (see Figure 6).
•Connect the wire corresponding to the detector ’s red wire to the LEL “R”
terminal.
•Connect the wire corresponding to the detector’s white wire to the LEL “W”
terminal.
•Connect the wire corresponding to the detector’s green wire to the LEL “G”
terminal.
•Connect the wire corresponding to the detector’s black wire to the LEL “B”
terminal.
CAUTION: If shielded cable is used, leave the cable shield ’s drain wire disconnected and
insulated at the detector junction box. You will connect the opposite end of the cable’s
drain wire to the M2A junction box’s chassis (earth) ground.
65-2640RK-03 M2A Trans mitter Operator’s Manual12
Detector
T
erminal
Strip
Cable
Shield
- S +
LEL
R W G B
RS 485PWR/SIG
A B C
FAIL
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
Black
Green
White
Red
Black
M2A
J-Bo x
Green
White
Red
Detect or J -Box
Combustible Gas
Detector
Calibration Adapter
Figure 6: Wiring the Detector to the M2A
13. Reinstall the detector terminal strip into its socket.
14. If shielded cable is used between the detector junction box and the M2A junction box,
connect the cable’s drain wire to an available chassis (earth) ground at the M2A
junction box. Do not connect the shield drain wire at the detector junction box.
15. Guide multi conductor shielded cable or cables or wires in conduit through the top
conduit hub of the M2A junction box. The number of cables or wires needed will
depend on whether the M2A is wired to a gas monitoring controller or just to power,
whether any relays are used, and whether the Modbus output is used.
1365-2640RK-03 M2A Transmitter Operator’s Manual
Loading...
+ 40 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.