RKI Instruments 65-2615RK-05-04 User Manual

65-2615RK-05-04
M2 Transmitter
Operator’s Manual
Part Number: 71-0326
Revision: P1
RKI Instruments, Inc.
www.rkiinstruments.com
WARNING
Read and understand this instruction manual before operating instrument. Improper use of the gas monitor could result in bodily harm or death.
Periodic calibration and maintenance of the gas monitor is essential for proper operation and correct readings. Please calibrate and maintain this instrument regularly! Frequency of calibration depends upon the type of use you have and the sensor types. Typical calibration frequencies for most applications are between 3 and 6 months, but can be required more often or less often based on your usage.
65-2615RK-05-04 M2 Transmitter Operator’s Manual
Product Warranty
RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year fr o m date of shipment from RKI Instruments, Inc. Any parts found defective withi n tha t period will be repaired or replaced, at our option, free of charge. This warranty does not apply to those items which by their nature are subject to deterioration or consumption in normal ser v ice, and which must be cleaned, repaired, or replaced on a routine basis. Examples of such items are:
W arranty is voided by abuse including mechanical damage, alteration, rough handling, or repair procedures not in accordance with the operator’s manual. This warranty indicates the full extent of our liability , a nd we are not r esponsible for removal or r eplacement costs, local repair costs, transportation costs, or contingent expenses incurred without our prior approval.
a) Absorbent cartridges d) Batteries b) Pump diaphragms and valves e) Filter elements c) Fuses
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI INSTRUMENTS, INC. INCLUDING BUT NOT LIMITED TO, THE WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL RKI INSTRUMENTS, INC. BE LIABLE FOR INDIRECT, INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
This warranty covers instruments and parts sold to users by authorized distributors, dealers, and representatives as appointed by RKI Instruments, Inc.
We do not assume indemnificatio n f o r a n y a c c ident or damage caus e d b y t he op e ration of this gas monitor, and our warranty is limited to the replacement of parts or our complete goods.
65-2615RK-05-04 M2 Transmitter Op erator’s Manual
Table of Contents
Chapter 1: Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
About the M2 Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
About this Manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Chapter 2: Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
External Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Internal Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Chapter 3: Installation & Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Mounting the M2 Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Wiring the M2 Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Start Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Chapter 4: Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Normal Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Information Screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
4 - 20 mA Signal Output Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Alarm Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Chapter 5: Configuration Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Viewing & Changing M2 Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Chapter 6: Gas Type Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Selecting the Gas Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
65-2615RK-05-04 M2 Transmitter Operator’s Manual
Chapter 7: Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Calibration Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Replacing Components of the H2S Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Chapter 8: RS-485 Modbus Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Wiring the M2 in a Modbus System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Using the M2 in a 4-wire Modbus System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Modbus Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
Supported Modbus Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Appendix A: Control Button Quick Reference Guide. . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Appendix B: PLC and DCS Device Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
Appendix C: Function Code 16 Registers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
(Appendix C available from RKI Instruments, Inc. Not included in manual as normally provided with M2 detector head.)
65-2615RK-05-04 M2 Transmitter Op erator’s Manual
Chapter 1: Introduction
Overview
This chapter briefly describes the 65-2615RK-05-04 M2 Transmitter. This chapter also describes the 65-2615RK-05-04 M2 Transmitter Operator’s Manual (this document). Table 1 at the end of this chapter lists the specifications for the M2.
About the M2 Transmitte r
The M2 transmitter is a fixed mount, continuous-monitoring detector hea d. All user adjustable parameters may be accessed using push button switches. In addition, calibration may be performed non-intrusively by use of a magnetic wand accessory which activates magnetic switches through a window at the front face of the detector head.
The M2 displays the current gas reading on an LCD display wh ich is visible through the window in the cover (front face) and provides a 4 - 20 mA signa l which indicates the target gas reading for use by a gas monitoring controller, recording device, or programmable controller. The M2 also provides an RS-485 Modbus output. Three sets of relay contacts, two controlled by the gas alarms and one by the fail alarm, rated at 115 VAC 5 amps are available for controlling devices such as lights or horns or for controlling higher rated relays.
Four operating modes allow you to display and change setup and calibration settings and change the gas type. They are Calibration Mode, Configuration Mode, Gas Type Mode, and Modbus Mode.
About this Manual
The M2 Transmitter Operator’s Manual uses the following conventions for notes, cautions, and warnings.
NOTE: Describes additional or critical info rmation.
CAUTION: Describes pote ntial damage to equip ment.
WARNING: Describes potential danger that can result in injury or death.
1 65-2615RK-05-04 M2 Transmitter Operator’s Manual
Specifications
Table 1 lists specifications for the M2.
Table 1: M2 Specifications
Target Gas/Detection Range Hydrogen Sulfide:
0 - 100 ppm, 1 ppm increments Alarm Sett ings (Alarm 1/Alarm 2) 10 ppm/50 ppm* Construction (housing) Explosion-proof Junction Box, NEMA 4 Area Classification Explosion-proof for Class I, Groups B, C, and D Sampling Method Diffusion Input Power 19 - 30 VDC Controls • Three push button switches
• Three magnetic switches for non-intrusive calibration Weight 5.5 lbs. Signal Output • 4 to 20 mA, 500 ohms impedance max
• RS-485 Modbus Operating Temperature -20°C to 50°C Accuracy ± 5% of reading or ± 2 ppm H
*
These are the RKI factory settings. You can change the alarm settings in Configuration Mode. See
S (whichever is greater)
2
“Viewing & Changing M2 Parameters” on page 23.
WARNING: When using the M2, you must follow the instructions and warnings in
this manual to assure proper and safe operation of the M 2 and to minimize the risk of personal injury. Be sure to maintain and periodically calibrate the M2 as described in this manual.
WARNING: M2 det ector heads with firmware version 5.0 or later have a different
Modbus register assignment than M2s wi th previous versions of firmware. If you are adding M2s with firmware version 5.0 or l ater to an existing Modbus network installation, confirm what firmware version is installed in the M2s already in the network and if necessary, take all steps required to address the differences in Modbus register assignments. See “Supported Modbus Functions” on page 42 for information on the current Modbus register assignments. The firmware version i s shown on the Information Screen described on page 18.
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 2
Chapter 2: Description
Overview
This chapter describes the components of the M2 transmitter. The transmitter is a 4 - 20 mA type detector head. It consists of the H
terminal PCB, the M2 junction box, and the detector junction box. The two junction box configuration is intended for situations where the detector needs to be installed at an inaccessible location. The detector junction box can be installed at the inaccessible location and the M2 junction box can be installed in a more readily accessible area.
External Description
This section describes the junction boxes and all external components of the M2 transmitter.
M ounting Slot ( 2x)
Magnetic Wand
3/4" NPT Conduit Hub
S detector, calibration adapter/splash guard,
2
Term i nal Strip
Wi ndow
3/4" NPT
Conduit
M2
Junction
Box
Figure 1: M2 External Component s
3 65-2615RK-05-04 M2 Transmitter Operator’s Manual
Junction Box
Hydrogen Sulfide
Calibration Adapter/
Splash G u ard
Hub
Detector
Detector
M2 Junction Box
r
The M2’s cast aluminum junction box protects the terminal PCB and wiring connections made to the terminal PCB. Use the top 3/4’’ conduit hub to connect wiring from the terminal PCB to external power and/or to an external device, typically an RKI controller. Use the bottom 3/4” conduit hub to wire the remotely installed H
S detector. Use the
2
junction box’s two mounting slots to mount the M2 to a vertical surface at the monitoring site. The window in the cover on the front of the junction box allows you to view the LCD display and use the magnetic wand to actuate the magnetic control switches so you can perform non-intrusive calibration. Removing the cover allows you to access the interior of the junction box.
Detector Junction Box
The detector junction box is intended to be installed remotely from the M2 junction box. The H
S detector is factory installed in the bottom 3/4” NPT conduit hub. A terminal strip
2
is provided for wiring connections and the detector is factory wired to one side of this terminal strip. The other side of the terminal strip is for user-supplied wiring to the detector terminal strip located in the M2 junction box. This wiring should go through the top 3/4” conduit hub. Three spacers installed on the back of the junction box controls the distance of the junction box from a mounting surface.
H2S Detector
The H2S detector consists of the detector housing body, detector housing cap, cap gasket, rubber boot, spacer, and the plug-in sensor.
Detector Housing Body
Cap Gasket
H2S Plug-in Sensor
Spacer
Rubber Boot
Detector Housing Cap
Flame Arresto Guard
Figure 1: H
S Detector Component Location
2
Detector Housing Body
The detector housing body protects the electronic components within the housing. Use the mounting threads at the top of the housing to screw the H
hub on the bottom of the detector junction box. Two wires extend from the top of the
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 4
S detector into the 3/4” NPT
2
detector housing body. Use these wires to connect the H2S detector to the terminal strip. One of the wires is red and one of the wires is black.
The housing includes a four-socket pattern. This socket pattern accepts the sensor’s four pins to secure the sensor within the detector housing. A pre-amplifier, located between the sockets and two interconnect wires, conditions the sensor’s signal before the signal reaches the controller.
Housing Cap & Cap Gasket
The housing cap screws onto the detector housing. It retains the plug-in sensor and protects it from damage. The housing cap also includes a flame arrestor which contains any sparks that may occur within the detector and a flame arrestor guard which protects the flame arrestor fro m damage. Unscrew the detector cap to access the plug-in sensor for maintenance or replacement. A cap gasket seals the interface between the housing and cap. A flame arrestor guard is permanently bonded to the cap.
Rubber Boot and Spacer
A rubber boot and spacer are installed between the detector housing cap and the sensor. They help ensure that the sensor remains plugged into the detector housing body.
Plug-In H
The sensor is secured within the detector assembly by the housing cap. Through a series of chemical and electrical reactions, the sensor produces an electrical output that corresponds to the detection range of the transmitter.
S Sensor
2
Calibration Adapter/Splash Guard
A calibration adapter/splash guard is installed on the H2S detector . A fitting at the bottom of the calibration adapter/splash guard allows a length of tubing to be connected for
calibration and routed to a more easily accessible location ( nea r the M2 junction box is optimal for ease of calibration). The calibration adapter/splash guard also protects the detector from splashing water or direct water spray.
Magnetic Wand
The magnetic wand is a short plastic rod with a magnet in one end. It is used to actuate the magnetic control switches on the control PCB while the junction box cover is still installed so that non-intrusive calibration can be performed.
5 65-2615RK-05-04 M2 Transmitter Operator’s Manual
Internal Description
y
This section describes the interna l components of the M2. The internal components of the M2 include t he t e rminal PCB which pro v i de s f or all the wiring connections to the M2 and the control PCB which displays the gas reading and has the control buttons.
3/4" NPT Condu i t Open i ng
Power/Signal
Terminal St rip
Detector Terminal Strip
fo r Wire Entry
PWR/SIG
- S +
TOXIC OXY
+ - + -
RS 485
A B C
Ju nction Box
Relay Terminal Strips
Terminal PCB
Control PCB
Magnetic Switches
UP/YES
DOWN/NO
- S +
PWR/SIG
C NC NO
+
ALARM 1 FAIL
C NC NO
TOXIC OXY
+
ALARM 2
C NC NO
R S 485
A B C
ENTER
RKI INSTRUMENTS
M2 TRAN SM ITTER
Fail LED
Alarm 1 LE D
Alarm 2 LED
RX LED
TX LED
Programming
Connector
(Factory Use)
Modbus
Terminal St rip
Termination
Push Button
C ontro l S witch e s
UP/YES
Jumper
Ribbon Cable Connector
View WithCover and
Control PCB Removed
Display
3/4" NP T Co nd uit Openi ng for Detector J- Box Wiring
DOWN/NO
ENTER
View WithCover Removed
Figure 2: M2 Internal Component s
Terminal PCB
The terminal PCB is encapsulated in epoxy for protection against moisture and physical damage. It is mounted into the rear of the junction box with three standoffs and rests on a thin layer of foam. A banana jack is screwed into each of the standoffs and used for mounting the control PCB. The terminal PCB converts the electrical output from the detector to a signal which can be displayed by the LCD display, a 4 - 20 mA signal (that is proportional to the detection range), and an RS-485 Modbus output signal. The 4 - 20 mA signal may be used by a recording device, gas monitor controller, or programmable controller. The Modbus output may be used to connect the M2 to a Modbus network. The terminal PCB also controls three relays, one fail and two gas alarm relays.
Two columns of plug-in style terminal strips are used to make all wiring connections to the M2. The column on the left consists of the power/signal, detector, and Modbus terminal strips. The column on the right consists of the relay terminal strips. A 20 position connector at the bottom of the terminal PCB is used to connect the terminal PCB to the
LCD Displa
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 6
control PCB with a ribbon cable. A 5 position connector on the left side of the terminal PCB is used by factory or field service personnel to program the M2.
Power/Sign al Terminal Strip
The power/signal terminal strip is a three position plug-in style terminal strip located at the top of the left terminal column. It is used to connect 24 VDC power to the M2 and to connect the 4 - 20 mA output signal to a device.
The signal output, the S terminal, does not have to be connected for the M2 to function. The S terminal is used if you want to connect the 4 - 20 mA output signal to another device such as a gas monitoring controller, chart recorder, or programmable controller (PLC).
Detector Terminal Strip
The detector terminal strip is a four position plug-in style terminal strip and is the middle terminal strip in the left terminal column. The two terminals labeled “TOXIC” are used to connect the detector to the M2.
Detector
Terminal Strip
PWR/SIG
- S +
TOXI C OXY
+ - + -
RS 485
A B C
PWR/SIG
- S +
+
TOXIC OXY
+
RS 4 85
A B C
C NC NO
C NC NO
C NC NO
FAIL
ALARM 1
ALARM 2
Figure 3: Detect o r Term inal Strip
NOTE: The detector is factory-wired to the M2. See “Wiring the M2 Transmitter” on
page 12 for all wiring procedures related to M2.
Modbus Terminal Str ip
The Modbus terminal strip is a three position plug-in style terminal strip and is the bottom terminal strip in the left terminal column. It allows connection of the M2 into a Modbus network.
Relay Terminal Strips
The right column of terminal strips consists of, from top to bottom, the fail, alarm 1, and alarm 2 relay terminal strips. The y are three-positio n plug-in style terminal strips. The relay terminal strips are used to connect devices such as lights and horns that are controlled by the relay contacts. The relay contacts are rated at 115 VAC, 5 amps. The relay contacts may also be used to control higher rated relays.
7 65-2615RK-05-04 M2 Transmitter Operator’s Manual
Termination Jumper
A two pin header with a termination jumper installed is located below the Modbus terminal strip. The jumper has no function unless the M2 is wired into a Modbus installation. See “Chapter 8: RS-485 Modbus Output” on page 37 for a description of using the M2 in a Modbus system.
Control PCB
The LCD display and control switches are located on the control PCB. It is installed on top of the terminal PCB by lining up its three spacing standoffs with the banana jacks in the terminal PCB mounting standoffs and pushing it ont o the banana jacks. The jacks retain the control PCB.
LCD Display
The LCD display is located at the top of the control PCB. It indicates the current gas reading and displays messages and parameters in the M2’s programs.
Control Buttons
The M2 includes three push button switches that allow you to enter the M2’s operating modes, navigate through the modes, update settings, an d save changes to the settings. The push button switches are located along the bottom edge of the control PCB (see Figure 2). The UP/YES button is on the left, the DOWN/NO button is in the middle, and the ENTER button is on the right.
Table 2: M2 Control Button Functions
Switch Function
UP (YES) Saves settings
Changes the displayed set t ing
Enters th e Calibration Program
Enters Gas Type Mode (press with DOWN/NO button)
Enters Configuration Mode (pres s with EN TER button )
DOWN (NO) Cancels setting changes
Changes the displayed set t ing
Enters Gas Type Mode (press with UP/YES button)
Enters Modbus Mode (press with ENTER button )
Displays the Information Screen
ENTER Initiates operations
Enters Configuration Mode (press with UP/YES button)
Enters Modbus Mode (press w ith DOW N /NO button)
Functions as an alarm reset switc h
Just above each push button switch is a magnetic switch with the same function as th e push button switch below it. The magnetic switches are for use in non-intrusive calibration. They are actuated by bringing the magnetic wand close enough to them to actuate them. Although the magnetic switches have the same functions as the push button switches, it is not practical to use them for operations other than calibration because it is not possible to actuate two magnetic switches at once with only one magnetic wand. Since displaying the Information Screen only requires the use of one switch, the wand may be used to show the Information Screen (see “Information Screen” on page 18).
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 8
Status LEDs
The M2 includes five status LEDs that are located above the display (see Figure 2).
Fail LED The fail LED turns on when the M2 is experiencing a fa il condition. A fail condition
can be caused by a detector failure or low detector signal.
Alarm 1 LED The alarm 1 LED is on when the M2 is experiencing an alarm 1 condition.
Alarm 2 LED The alarm 2 LED is on when the M2 is experiencing an alarm 2 condition.
RX & TX LED’s These LED’s indicate data being received (RX) and transmitted (TX) when the M2’s
Modbus output is operating.
9 65-2615RK-05-04 M2 Transmitter Operator’s Manual
Chapter 3: Installation & Startup
5
Overview
This chapter describes procedures to mount the M2 Transmitter in the monitoring environment and wire it to input power and devices.
Mounting the M2 Transmitter
1. Select a mounting site that is representative of the monitoring environment. Consider the following when you select the mounting site.
For the M2 junction box, select a site where the junction box is not likely to be
bumped or disturbed. Make sure there is sufficient room to perform start-up, maintenance, and calibration procedures. The site should be easily a ccessible.
For the detector junction box, select a site where the target gas is likely to be found
first. For lighter gases, mount the detector near the ceiling; for heavier gases, mount the detector near the floor.
.86
The junction boxes should be mounted no further than 200 feet from each other.
2. Use #10 screws through the M2 junction box’s two mounting slots to secure the junction box to a vertical surface. Figure 4 shows the outline and mounting dimensions the M2 junction box. Figu re 5 shows the outline and mounting dimensions for the detector junction box. Mount the detector junction box with the detector facing down as shown in Figure 5.
4.59
2.41
5.23
5.00
3/4 NPT Conduit Hub
1.02 .4 7 (2 X)
Figure 4: Outline & Mounting Dimensions, M2 Junction Box
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 10
.3 0 (2X)
r
r,
3/4 Conduit Hub
5.25
Ø3.65
2.70 .38
Rubbe Space 3X
H2S Detector
11.40 Max
Calibration Adapter/ Splash Guard
1.13
Figure 5: Outline & Mounting Dimensions, Detector Junction Box
3. Install 3/16 inch I.D. flexible po lyurethane tubing to the fitting at the bottom of the calibration adapter/splash guard and route it to an accessible area that is close to the M2 junction box. See the “Parts List” on page 47 for available tubing.
11 65-2615RK-05-04 M2 Transmitter Operator’s Manual
Wiring the M2 Transmitter
WARNING : Always verify that the power to the M2 is off before you make wiring
connections.
1. Confirm that no power is being applied to the M2.
2. Remove the detector junction box’s cover from the junction box.
3. Remove the M2 junction box’s cover from the junction box.
4. Grasp the control PCB by its edges.
5. Gently pull until the control PCB is pulled away from the banana jacks. Take care not to pull too hard and damage the cable which connects the control and terminal PCBs.
6. Let the control PCB hang by the cable. The terminal strips are now visible on the terminal PCB. The control PCB may be left hanging while wiring is done. If desired, the control PCB may be disconnected from the cable and set aside while wiring.
7. Guide a two-conductor, shielded cable or two wires in conduit through the top conduit hub of the detector junction box. To make wiring more convenient, use wire colors that correspond to the detector wire colors: red and black.
NOTE: Consult RKI Instruments, Inc. for cable lengths longer than 200 feet.
8. Connect the detector leads to the terminal block in the detector junction box.
9. Secure the detector junction box’s cover to the junction box.
10. Guide the other end of the two-conductor, shielded cable or two wires in conduit through the bottom conduit hub of the M2 junction box.
11. To gain access to a plug-in terminal strip for wiring, pull it out of its socket by grasping the wall between two terminal positions with needle nose pliers and pulling. Be careful not to exert too much force on the wall to avoid damage to the terminal strip. The detector terminal strip and th e alarm 1 r elay t ermi nal strip ma y be r emo ved by grasping them with your fingers if the adjacent terminal strips have been removed.
WARNING: If the M 2 is installed in a hazardous location, use app ropriatel y rated
conduit, conduit fittings, and appropriate construction technique that complies with the local electrical code. In addition, use appropriately rated conduit, conduit fittings, and appropriate construction technique to maintain the NEMA 4 environmental rating of the junction box and avoid water entering the junction box at the conduit hubs.
12. Pull out the detector terminal strip and connect the two wires to the terminal strip a s follows (see Figure 6).
Connect the wire corresponding to the detector ’s red wire to the TOXIC “+”
terminal.
Connect the wire corresponding to the detector’s black wire to the TOXIC “-”
terminal.
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 12
Detector
Terminal
Strip
CAUTION: If shielded cable is used, leave the cable shield ’s drain wire disconnected and
insulated at the detector junction box. You will connect the opposite end of the cable’s drain wire to the M2 junction box’s chassis (earth) ground.
Cable
Shield
FAIL
- S +
PWR/SIG
- S +
TO XIC O X Y
+ - + -
R S 485
A B C
C NC NO
+
C NC NO
+
TOXIC OXY
ALARM 2 ALARM 1
C NC NO
R S 485 PWR/SIG
A B C
Black Red
M2 J-Box
Black Red
Detector J-Box
H2S Detector
Calibration Adapter/Splash Guard
Figure 6: Wiring the Detector to the M2
13. Reinstall the detector terminal strip into its socket.
14. If shielded cable is used between the detector junction box and the M2 junction box, connect the cable’s drain wire to an available chassis (earth) ground at the M2 junction box. Do not connect the shield drain wire at the detector junction box.
15. Guide multi conductor shielded cable or cables or wires in conduit through the top conduit hub of the M2 junction box. The number of cables or wires needed will depend on whether the M2 is wired to a gas monitoring controller or just to power, whether any relays are used, and whether the Modbus output is used.
13 65-2615RK-05-04 M2 Transmitter Operator’s Manual
NOTE: If the M2 is being wired to a PLC or DCS device, see “Appendix B: PLC and
DCS Device Wiring” on page 49, then continue with step 8.
Use the following recommendations to determine how to wire the M2:
If Modbus connections will not be used and only the PWR/SIG connections will
be used, use a two or three conductor shielded cable or two or three wires in conduit for connections to the power/signal terminal strip depending on whether or not the signal (S) terminal is used. The S terminal has a 4 - 20 mA output, but if you do not need to monitor this signal and do not connect to th e S termi na l to access this signal, the M2 will still function completely.
Table 3: Wire Size for PWR/SIG Connections
Max Distance to Controller
w/18 Gauge Wire
Max Distance to Controller
w/16 Gauge Wire
Max Distance to Controller
w/14 Gauge Wire
2,500 ft. 5,000 ft. 8,000 ft.
If the PWR/SIG connections and one or more relays are used, route the
connections to the M2 in conduit. Use shielded cable in the conduit for the PWR/ SIG connections and unshielded cable or individual wires for the relay connections. Make sure any wire or cable used for relay wiring is appropriately rated for the power that it will carry.
CAUTION: If shielded cable is used for the PWR/SIG connections, leave the cable shield’s drain
wire insulated and disconnected at the M2 junction box. You will connect the opposite end of the cable’s drain wire at the controller or device.
If the M2 will be wired into a Modbus network, see “Chapter 8: RS-485 Modbus
Output” on page 37.
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 14
See Figure 7 below for field wiring connections to the M2.
RKI Controller
Terminals
(24 VDC) +
4 -20 mAIn (S)
(24VDC) -
Alarm Device
CableShield
See
Detector
Wiring
See
Modbus
Wiring
S
PWR/SIG
+
TOXIC OXY
+
A B C
RS 485
C NC NO
C NC NO
C NC NO
FAIL
ALARM 1
ALARM 2
Power
Fail Alarm Device
Alarm 1 Alarm Device
Alarm 2 Alarm Device
Typical Alarm WiringShown
Figure 7: Wiring the M2 Junction Box to a Controller and Alarm Devices
16. Re-install the control PCB (and ribbon cable if necessary). B e sure the ribbon cable is routed down below the control PCB so it will not be damaged by the cover when it is screwed back on.
Start Up
17. Secure the M2 junction box cover to the junction box.
18. Make controller, device, and relay connections as appropriate. If shielded cable is used for the PWR/SIG connections, connect the cable shield’s drain wire to an available chassis ground at the gas mon itoring controller, recording device, or program mable controller.
Introducing Incoming Power
1. Complete the installation procedures described earlier in this manual.
2. Verify that all wiring connections are correct and secure.
3. Turn on the incomi ng power.
4. If necessar y, tu r n on th e cont roller or other monitoring device that is connected to the M2.
5. The LCD display will indicate the firmware version when the M2 is first powered up and will then count down a o ne minute warm-up period b efore normal ope ration begins. During normal operation, the display will indicate the target gas and current gas reading. V erify that the display is indicating the target gas and current gas reading after the warm-up period is complete and normal operation begins.
H2S
0 ppm
15 65-2615RK-05-04 M2 Transmitter Operator’s Manual
NOTE: Allow the M2’s detector to warm up for 15 minutes before you continue with
the next section, “Adjusting the Fresh Air Reading.”
Adjusting the Fresh Air Reading
When the M2 is shipped from RKI Instruments, Inc., it is factory calibrated. If a full calibration is desired at startup, see “Calibration” on page 33.
Verify that the M2 is in a fresh air environment (environment known to be free of combustible or toxic gas vapors and of normal oxygen content, 20.9%).
CAUTION: If you suspect the presence of hydrogen sulfide in the monitoring environment, use
the calibration kit and the zero air calibration cylinder to introduce “fresh air” to the detector and verify an accurate fresh air setting. See the “Calibration” on page 33 for instructions on how to use a zero air cylinder when performing a fresh air adjustment.
Non-Intrusive Fresh Air Signal Adjustment
If the M2 is installed in a classified area and non-intrusive zero adjustment is required, follow the instructions below, but do not remove the junction box cover. Use the magnetic wand accessory to actuate the magnetic switches instead of pushing the control switch buttons. To actuate a magnetic switch and perform the same operation as pressing a control switch button, touch the magnet in the end of the magnetic wa nd to the M2’s glass cover directly over the magnetic switch you wish to actuate. Touching the glass and removing the wand is the same as pressing and releasing a button. Touching the glass and keeping the wand in place is the same as pressing and holding a button.
WARNING : The M2 i s not an active gas monitoring device dur in g the fresh air
adjustment procedure. The 4-20 mA output signal will “freeze” at 3.5 mA and all relays will remain in their non-alarm state while the M2 is in Calibration Mode. The output signal will not indicate current readings and the relays will not resume operating normally until the M 2 is in normal operation again.
NOTE: While in the calibration program, if there is no switch activity for the
calibration time-out period the unit will return to normal operation. The factory set time-out is 15 minutes. If you want a different time-out period, see “Viewing & Changing M2 Parameters” on page 23 for instructions to change the calibration time-out.
1. While in normal operation, press and hold the UP/YES button for 5 seconds to enter Calibration Mode. Release the button when the following screen appears.
Calib? YES/NO
2. Press and release the UP/YES button to continue. The display will indicate the target gas and CAL Mode for a few seconds before showing FreshAir Adjust?.
3. Press and release the UP/YES button. ENTER will alternate with FreshAir on the top display line and the current gas reading will be on the bottom di splay line.
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 16
4. Press and release the ENTER button. The M2 will perform a zero operation and the display will indicate SPAN w/Cal Gas?.
5. Press and release the DOWN/NO button. The display will indicate Leaving CAL Mode and the M2 will return to normal operation.
17 65-2615RK-05-04 M2 Transmitter Operator’s Manual
Chapter 4: Operation
Overview
This chapter describes the M2 in normal operation. This chapter also describes the M2 in alarm 1, alarm 2, and fail co nditions and suggests responses to these conditions.
Normal Operation
Normal operation is defined as follows:
The start-up procedure is complete.
The M2 is not indicating an alarm 1, alarm 2, or fail condition.
The M2 is not in Calibration, Configuration, or Gas Type modes.
During normal operation, the M2 simultaneously displays the current gas reading, unit of measure, and target gas.
Information Screen
The Information Screen displays the M2’s operating voltage and firmware and hardware information. To display the Information Screen, do the following:
1. Press and hold the DOWN/NO button for 3 seconds. The Information Screen will appear.
2. Continue holding the DOWN/NO button to keep the Information Screen on the display. The top line indicates the operating voltage that is connected to the M2. The second line indicates the firmware that is running, version 5.0 in this example, and the M2’s hardware version, version 2A in this example. This information may be useful if you need to contact the factory with question s about the M2.
3. Release the DOWN/NO button when you are done viewing the Information Screen. The display will return to the normal operation screen within a couple of seconds.
H2S
0 ppm
24.2V v5.0L 2A
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 18
4 - 20 mA Signal Output Operation
The output at the S terminal of the power/signal terminal s t rip is a 4 - 20 mA signal t hat corresponds to the detection range of the M2. During normal operation, this signal tracks the gas concentration on the LCD.
There are several cir cumstances where th e signal output will not track the display reading but will behave as follows:
When the M2 is in its warm-up period, the signal output will be fixed at 3.5 mA (zero).
When the M2’s gas type is changed, the M2 will enter Configuration Mode for you to verify the parameter settings. When you exit Configuration Mode, the dis play will indicate NEEDS CALIBRATION and will con tinue to indicate this until Calibration Mode is entered and a calibration is performed. In this situation, the signal output will be fixed at 3.5 mA from the time Gas Type Mode is entered until the M2 is calibrated and returns to normal operat ion.
If you enter Calibration Mode, Configuration Mode, Gas Type Mode, or Modbus Mode, the signal output will be fixed at 3.5 mA until the M2 returns to normal operation.
If the M2’s input power decreases below 18.5 volts so that the M2 is in a low power alarm, the signal output is fixed below 2.4 mA until the low power alarm is cleared.
If the M2 goes into a fail condition, after a 30 second d e lay, the signal output is fixed below 2.4 mA until the fail alarm is cleared. During the 30 second delay, the signal output follows the detector output. In the case of a downscal e reading, the display and the signal output continue to track the reading down to -99% of full scale (1.15 mA).
Alarm Indications
NOTE: The M2 includes al arm o n an d a la rm off delay settings for alarm 1 an d alarm 2.
Condition Cause Visual Indication(s)
Alarm 1
Alarm 2
1
1
The alarm indications described in this section operate according to the factory set alarm settings. See Table 5 on page 24 for all the factory settings.
Table 4: Visual and Audible Alarm Indications
Increasing gas reading at or above the alarm 1 setpoint
Increasing gas reading at or above the alarm 2 setpoint
• A1 LED is on
• Gas reading alternates
with ALARM-1 message
• A2 LED is on
• Gas reading alternates
with ALARM-2 message
Fail • Disconnected or misconnected
detector wiring
• Display reading at -10% of full scale or lower
• Defective components
19 65-2615RK-05-04 M2 Transmitter Operator’s Manual
• F LED is on
FAIL message replaces
gas reading
NOTE: There is a 30 second delay on the fail condition.
Table 4: Visual and Audible Alarm Indications
Condition Cause Visual Indication(s)
Low Power DC power source less than 18.5 volts. • F LED is on
LowPower message and
actual voltage of incoming DC power
*1If the M2 is in both an ala r m 1 and an a l arm 2 c ondition, both alarm LEDs are on an d t he
display alternates between the gas reading and the AL MS 1&2 message.
NOTE: You can select normally energized (N. EN) or normally de-energized (N. DE-
EN) alarm 1 and alarm 2 relay set tings in Configuration Mo d e. The following sections describe the standard factory setting for these relays which is N. DE-EN.
The fail relay is factory set as N. EN and is not user-adjustable.
Alarm 1 Condition
Alarm 1 Condition Indications
When the gas reading reaches the alarm 1 setpoint, the M2 senses an alarm 1 condition. The M2 alerts you to an alarm 1 condition as follows:
The A1 LED turns on.
The gas reading alternates with the ALARM-1 messag e.
The alarm 1 relay energizes.
Responding to an Alarm 1 Condition
1. Follow your established procedure for a low level toxic gas con d ition.
2. After the gas reading falls below the alarm 1 setpoint, press the ENTER button to reset the alarm 1 circuit. Resetting the alarm 1 circuit turns off the A1 LED, resets the LCD, and de-energizes the alarm 1 relay.
NOTE: If the ENTER button is pressed while the M2 is in an alarm 1 cond ition, the A1
LED will flash but all other indica tions will remain unchanged.
You cannot de-energize the al arm 1 relay until the gas reading falls below the alarm 1 setpoint.
Alarm 2 Condition
Alarm 2 Condition Indications
When the gas reading reaches the alarm 2 setpoint, the M2 senses an alarm 2 condition. The M2 alerts you to an alarm 2 condition as follows:
The A2 LED turns on.
The gas reading alternates with the ALARM-2 messag e.
The alarm 2 relay energizes.
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 20
NOTE: If the M2 is in both an alar m 1 and al arm 2 cond ition , both the A1 and A2 LED s
will be on, the gas reading will alternate with the ALMS 1&2 message, and both alarm relays will energize.
Responding to an Alarm 2 Condition
1. Follow your established procedure for a high level toxic gas co nd ition.
2. After the gas reading falls below the alarm 2 setpoint, press the ENTER button to reset the alarm circuit. Resetting the alarm circuit turns off the A2 light, resets the LCD, and de-energizes the alarm 2 relay.
NOTE: If the ENTER button is pressed while the M2 is in an alarm 2 cond ition, the A2
LED will flash but all other indica tions will remain unchanged.
You cannot de-energize the al arm 2 relay until the gas reading falls below the alarm 2 setpoint.
Fail Condition
Fail Condition Indications
The M2 indicates a fail condition for any of the following:
The detector wiring is disconnected or incorrectly connected.
The display reading is -10% of full scale or lower.
When the M2 senses a fail condition , it alerts you as follows:
The F LED turns on.
The gas reading is replaced by the FAIL message.
The fail relay de-energizes.
NOTE: The fail alarm has a 30 second delay.
Responding to a Fail Condition
1. Verify that the detector wiring is correctly and securely connected.
Low Power Alarm
Low Power Alarm Indications
The M2 senses a low power condition when the DC power source is 18.5 volts or less.
WARNING: While in a low power condition, the M2 is not an active gas monitor.
When the M2 senses a low power condi tion, it alerts you as follows:
The F LED turns on.
•The message LowPower is indicated on the top line of the LCD and the input voltage is displayed on the bottom line of the LCD.
The fail relay de-energizes.
21 65-2615RK-05-04 M2 Transmitter Operator’s Manual
NOTE: The low power alarm canno t be cl eared using the ENTER button.
When the voltage increases to 19.0 volts, the low power alarm is cleared and the M2 will begin its warm-up sequence.
Responding to a Low Power Condition
1. Determine and correct the cause of the low power condition.
2. When the input power increases above 19.0 volts, the M2 will begin its warm-up sequence.
3. Verify that the M2 enters normal operation after its warm-up sequence. If necessary, perform a fresh air adjustment. See “Adjusting the Fresh Air Reading” on page 16.
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 22
Chapter 5: Configuration Mode
Overview
This chapter describes how to view and change M2 parameters using Configuration Mode. It is accessed using the program buttons.
Configuration Mode includes a 5-minute time-out feature. If you do not press a control button for 5 minutes, the M2 automatically returns to no rmal operation.
NOTE: If the M2 returns to normal operation because of a time-out, it enters a warm-
up period just as it does when it is first turned on.
If you are installing a new M2, it has been setup at the factory. Use Configuration Mode only if you want to change the M2’s setup. If you want to change the detector type see “Chapter 6: Gas Type Mode” on page 26.
Viewing & Changing M2 Parameters
1. While in normal operation , simultaneously press and hold the UP/YES and ENTER buttons for 5 seconds to enter Configuration Mode. Release the buttons wh en the following screen appears.
Enter Config?
2. If you want to exi t Config uration M ode, pr ess and r eleas e the DOWN/N O button an d the M2 will return to normal operation.
If you want to continue in Configuration Mode, press and release the UP/YES button. The target gas and the full scale will be displayed for a few seconds b efo re the first adjustable par ameter, the alarm 1 setpoint, is displayed.
3. If you want to change the currently displayed parameter, use the UP/YES and DOWN/NO buttons to adjust it to the desired setting, then press ENTER to continue to the next parameter.
If the currently displayed parameter is OK, press the ENTER button to proceed to the next parameter.
Table 5 lists the M2 parameters you can set. Table 5 also lists the factory set value for each parameter.
23 65-2615RK-05-04 M2 Transmitter Operator’s Manual
Table 5: Configuration Parameters
Parameter
(Factory Set Value)
ALARM-1 (level) (10 ppm)
ALARM-1 (activation) (Increase)
ALARM-1 (relay action) (N. DE-EN)
ALARM-1 (relay reset) (LATCH)
A1 OnDy (alarm 1 on delay) (1 secs)
A1 OffDy (alarm 1 off delay) (0 sec.)
Description
The gas readin g at which the M2 initiates an alarm 1 condition.
Indicates if the alarm 1 circuit is activated by gas readings increasing (
Increase) or decreasing (Decrease) to the ALARM-1 Level.
If set as and energizes wh en an alarm 1 condition is initiated. If set as de-energizes when an alarm 1 cond iti on is initia ted .
If set as circuit after the alarm 1 condition passes. If set as the alarm 1 condition passes.
The amount of time the M2 delays activation of the alarm 1 circuit once an alarm 1 condition is initiated. It can be set in 1 second increments from 0 - 60 seconds, in 1 minute increments from 1 - 15 minutes, and in 15 minute increments from 15 - 60 minutes.
The amount of time the M2 delays turning off the alarm 1 circuit once an alarm 1 condition passes. This parameter appears only if the alarm 1 relay reset setting is set to increments from 0 - 60 seconds, in 1 minute increments from 1 - 15 minutes, and in 15 minute increments from 15 - 60 minutes.
N. DE-EN, the alarm 1 relay is de-energized in normal operatio n
N. EN, the alarm 1 relay is energized in normal operation and
LATCH, you must press the ENTER button to reset the alarm 1
SELF-RST, the M2 automatically resets the alarm 1 circuit after
SELF-RST. It can be set in 1 second
ALARM-2 (level) (50 ppm)
ALARM-2 (activation) (Increase)
ALARM-2 (relay action) (N. DE-EN)
ALARM-2 (relay reset) (LATCH)
A2 OnDy (alarm 2 on delay) (1 secs)
A2 OffDy (alarm 2 off delay) (0 sec.)
The gas readin g at which the M2 initiates an alarm 2 condition.
Indicates if the alarm 2 circuit is activated by gas readings increasing (
Increase) or decreasing (Decrease) to the ALARM-1 Level.
If set as and energizes wh en an alarm 2 condition is initiated. If set as de-energizes when an alarm 2 cond iti on is initia ted .
If set as circuit after the alarm 2 condition passes. If set as the alarm 2 condition passes.
The amount of time the M2 delays activation of the alarm 2 circuit once an alarm 2 condition is initiated. It can be set in 1 second increments from 0 - 60 seconds, in 1 minute increments from 1 - 15 minutes, and in 15 minute increments from 15 - 60 minutes.
The amount of time the M2 delays turning off the alarm 2 circuit once an alarm 2 condition passes. This parameter appears only if the alarm 2 relay reset setting is set to increments from 0 - 60 seconds, in 1 minute increments from 1 - 15 minutes, and in 15 minute increments from 15 - 60 minutes.
N. DE-EN, the alarm 2 relay is de-energized in normal operatio n
N. EN, the alarm 2 relay is energized in normal operation and
LATCH, you must press the ENTER button to reset the alarm 2
SELF-RST, the M2 automatically resets the alarm 2 circuit after
SELF-RST. It can be set in 1 second
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 24
Table 5: Configuration Parameters (Continued)
ZeroSupp (2 ppm)
FILTER (5 secs)
CAL Time (15 mins)
Auto Zero (Enabled)
Parameter
(Factory Set Value)
Description
The zero suppression feature helps prevent “jumpy” readings near the fresh air reading. For example, if the zero suppression is reading of 0 ppm for gas readings from -2 ppm to 2 ppm. It is settable from 0 to 6% of the detection range.
The filter feature helps “smooth out” jumpy or noisy signals from the detector. You can set the filter from 0 seconds to 60 seconds in 5 seconds increments. The displayed gas reading is the average reading over the previous time period defined by the filter setting. So if the filter is set to 5 seconds, the displayed gas reading is the average over the past 5 seconds.
The calibration t ime-out set s the am ount of time after th e last bu tton push while in Calibratio n Mo de th at th e M2 will wait be fore return ing to no rma l operation. If the calibra tion tim e-out is set to to normal operation automatically 15 minutes after the last button push.
If set to reading and make automatic zero adjustments under precisely specified conditions if it determ ine s tha t a reading ch ang e is du e to no rm al sens or drift. If set to
Enabled, the M2’s firmware will monitor the zero (fresh air)
Disabled, the Auto Zero function is turned off.
2.0 ppm, the M2 will display a
15 mins, then M2 will return
4. When you have scrolled through all the adjustable parameters, SAVE IT? YES/NO appears on th e display.
5. If you do not wish to save the adjustments and want to exit Configuration Mode, press and release the DOWN/NO button. The DO OVER? YES/NO message will display. Press and release the DOWN/NO button. The ABORT? YES/NO message will display. Press the UP/YES button to return to normal operation.
If you wish to change some of the adjustments made, press and release the DOWN/ NO button. The DO OVER? YES/NO message will display. Press and release the UP/ YES button. The Re-do Config message will display and the M2 will return to the first adjustable pa rameter. Go back to step 3 and continue.
If you wish to save the adjustments made, press and release the UP/YES button. Config Saved is indicated on the display for a few seconds and the M2 returns to normal operation.
25 65-2615RK-05-04 M2 Transmitter Operator’s Manual
Chapter 6: Gas Type Mode
Overview
This chapter describes how to use Gas T y pe Mode to select the M2’s gas type. The gas type determines the target gas and detection range.
CAUTION: The target gas is factory set and does not normally need to be changed. The
standard setting for the 65-2615RK-05-04 is hydrogen sulfide.
Gas Type Mode includes a 5-minute time-out feature. If you do not press a button for 5 minutes, the M2 automatically returns to normal operation.
NOTE: If the M2 enters normal operation because of a program time-out, it enters a
warm-up period just as it does when it is first turned on.
Selecting the Gas Type
1. While in normal operation or d uring the warm-up sequence, press and hold the UP/ YES and DOWN/NO buttons for five seconds. Release them when the following screen appears.
2. If you want to exit Gas Type Mode, press and release the DOWN/NO button. The display will indicate NO CHANGE and the M2 will return to normal operation.
If you want to continue in Gas Type Mode, press and release the UP/YES button. The top display line will indicate GasType? and the bottom display line will indicate the target gas choice.
3. Use the UP/YES or DOWN/NO button to scroll through the choices of target gas. Table 6 below lists the choices of target gas for the M2.
H2S 0 - 100 ppm CO 0 - 300 ppm
Table 6: Gas Types
Gas Type
Choices
Select GasType?
Detection Range
OXYGEN 0 - 25.0% volume SIH4 0 - 15.0 ppm PH3 0 - 1.00 ppm O3 0 - 1.00 ppm
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 26
Table 6: Gas Types
Gas Type
Choices
Detection Range
NO 0 - 100 ppm HF 0 - 9.00 ppm HCN 0 - 15.0 ppm HCL 0 - 15.0 ppm H2SE 0 - 5.00 ppm GEH4 0 - 1.50 ppm F2 0 - 3.00 ppm CLO2 0 - 1.00 ppm B2H6 0 - 5.00 ppm ASH3 0 - 1.50 ppm SO2 0 - 6.00 ppm NH3 0 - 200 ppm NH3 0 - 75.0 ppm CL2 0 - 3.00 ppm CL2 0 - 10.0 ppm
4. When the desired target gas is on the display, press and release the ENTER button. The display will ask SAVE IT? YES/NO.
5. To discard the gas type change, press and r elease th e DOWN/NO button. The display will ask DO OVER? YES/NO. Press the DOWN/NO button. The display will ask ABOR T? YES/ NO . Pr ess the UP/YES button. The M2 will return to normal operation without changing the gas type.
To save the gas type, press and release the UP/YES button. The display will indicate Gas T ype Up dated, the display will indicate the target gas and the detection range for a few seconds, and then the M2 will enter Configuration Mode automat ical ly so that the parameter settings in Configuration Mode can be verified or changed.
6. While in Configuration Mode press and release the ENTER button to accept the displayed parameter setting and move to the next one. Use the UP/YES and DOWN/ NO buttons to adjust a s et tin g. See “Chapter 5: Configuration Mode” on pag e 23 for a complete description of Configuration Mode.
7. When you have reviewed and accepted the last item in Configuration Mode by pressing the ENTER button, display will ask SAVE IT? YES/NO.
8. To save the configuration settin gs, press and release the UP/YES button.The display will indicate Config Saved and the M2 will begin its warm- up sequence.
To discard the settings and review them again, press the DOW N/NO button. The display will ask DO OVER? YES/NO. Press the UP/YES button and the display will indicate Re-do Config, then display the target gas and the detection range for a few
27 65-2615RK-05-04 M2 Transmitter Operator’s Manual
seconds before returning to Configuration Mode. Repeat steps 6 - 8 until the settings are as desired.
9. After the M2 completes its warm-up sequence, the display will indicate CAL NEEDED. Since the gas type has been changed, a successful calibration must be performed before the M2 can enter normal operation.
10. Press and release the UP/YES button to enter Calibration Mode. Normally, you must hold the UP/YES button for 5 seconds to enter Calibration Mode, but since a calibration is required, only a momentary push is needed to enter Calibration Mode after changing the gas type.
11. See “Calibration” on page 33 for calibration instructions.
NOTE: When calibrating an M2 after changing the gas typ e, the M2 will not ask if you
want to calibrate, or whether you want to perform a fresh air adjustment, span adjustment or zero adjustment. Since a complete calibratio n is required, the calibration sequence will proceed without asking if you want to perform certain operat ions.
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 28
Chapter 7: Maintenance
Overview
This chapter describes procedures for performing preventive maintenance, troubleshooting, calibrating the M2, and replacing field replaceable parts.
Preventive Maintenance
This section describes a recommended preventive maintenance schedule to ensure the optimum performance of the M2. It includes daily, monthly, and quarterly procedures.
Daily
Verify a display reading of 0 ppm. In vestigate significant changes in the reading.
Monthly
This procedure describes a test to verify that the M2 responds properly to the target gas.
NOTE: Performing a response test on the M2 may cause alarms. If you do not want
external alarms to act u ate, disable any external alarms duri ng the response test to avoid unwanted external alarm indication s .
NOTE: The following procedure assumes the use of a calibration kit which includes a
calibration gas cylinder and a fixed flow regulator with an on/off knob.
1. Disable all external alarms if external alarm indications are not desired during the response test.
2. Screw the calibration gas cylinder into the fixed flow regulator.
3. Use the flexible tubing coming from the calibration adapter/splash guard to connect the regulator to the calibration adapter/splash guard.
4. Turn the regulator knob counterclockwise to open the regulator.
5. Allow calibration gas to flow for the length of time determined in “D et ermining Response Time” on page 32.
6. Verify that the display reading is within ± 20% of the gas concentration .
NOTE: If the reading is not within ± 20% of the gas concentration, calibrate the M2 as
described in “Calibration” on page 33.
7. Turn the regulator knob clockw ise to close the regulator.
8. Unscrew the calibration gas cylinder from the fixed flow regulator.
9. Disconnect the calibration adapter/splash guard’s flexible tubing from the regulator.
NOTE: Do not disconnect the flexible tubing from the calibration adapter/splash guard.
29 65-2615RK-05-04 M2 Transmitter Operator’s Manual
10. If external alarms have been disabled, enable them when the display returns to a normal fresh air reading.
11. Store the components of the calibration kit in a safe place.
Quarterly
Calibrate the M2 as described in “Calibration” on page 33. See “Calibration Frequency” on page 32 for a discussion of calibration frequency guidelines.
Troubleshoo ting
The troubleshooting guide describes symptoms, probable causes, and recommended action for problems you may encounter with the M2.
NOTE: This troubleshooting guide describes M2 problems only . If the M2 is connected to
a controller, see the controller operator’s man ua l f or problems you may encounter with the controller.
Table 7:Troubleshooting the Hydrogen Sulfide Detector
Condition Symptom(s) Probable Causes Recommended Action
No Power The display backlight
is off and the display screen is blank.
Frequent or Suspect Alarms
The M2 alerts you to frequent or suspect alarms while th e fresh air readings remain on zero.
• The power wiring is disconnected or misconnected.
• The display cable is disconnected or misconnected.
•The M2 is experiencing false readings due to RFI or EMI.
• The detector wiring is disconnected, misconnected, or intermittent.
1. Verify that the wiring to the power source or controller is correct and secure.
2. Verify that the ribbon cable connecting the terminal PCB to the control PCB is securely installed. This ribbon cable plugs into a connector on the top edge of the control PCB and on the bottom of the terminal PC B.
3. If the power difficulties continue, contact RKI for further instruction.
1. Verify that the M2 wiring is properly shielded. See “Wiring the M2 Transmitter” on page 12.
2. Verify that the detector wiring is correct and secure.
3. Increase the alarm on delay setting in Configuration Mode.
4. If the frequent or suspect alarm difficulties continue, contact RKI for further instruction.
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 30
Table 7:Troubleshooting the Hydrogen Sulfide Detector (Continued)
Condition Symptom(s) Probable Causes Recommended Action
Flickering Display
Fail Condition
The display reading flickers often.
• M2 indicates a fail condition.
• Controller indicates a fail condition
•The M2 is experiencing false readings due to RFI or EMI.
•The noise filter setting is too low.
•The zero suppression setting is too low.
• The display screen is malfunctioning.
• The detector wiring to the terminal PCB is disconnected or misconnected.
•The wiring from the M2 to the controller is disconnected or misconnected.
• The detector fresh air signal is low enough to cause a fail condition.
•The detector is malfunctioning.
•The terminal or display PCB is malfunctioning.
1. Verify that the M2 wiring is properly shielded. See “Wiring the M2 Transmitter” on page 12.
2. Verify that the detector wiring is correct and secure.
3. Increase the filter setting in Configuration Mode.
4. Increase the zero suppression setting in Configuration Mode.
5. If the display difficulties continue, contact RKI for further instruction.
1. Verify that the detector wiring is correct and secure.
2. Verify that the wiring between the M2 and the controller is correct and secure.
3. Calibrate the M2.
4. If the fail condition continues, replace the detector.
5. If the fail condition continues, contact RKI for further instruction.
Slow or No Response/ Difficult or Unable to Calibrate
•M2 responds slowly or does not respond to response test.
•Unable to successfu l ly set the fresh air or span reading during calibration.
•M2 requires frequent calibration.
Note: See “Calibration Frequency” o n page 32 for calibration frequenc y guidelines.
• The calibration cylinder is low, out­dated, or defective.
• The calibration gas flow rate is too low.
• The calibration gas is not an appropriate concentration.
•The detector is malfunctioning.
• The terminal PCB is malfunctioning.
• The flexible tubing to the calibration adapter/splash guard has become kinked, blocked, or disconnected.
1. V erify that the calibration cylinder contains an adequate supply of a fresh test sample.
2. Verify that the regulator used for calibration is a 0.5 LPM regulator.
3. Verify that the calibration gas concentration is appropriate for the transmitter. The concentration should be in the detection range, preferably about half of the detection range.
4. Verify that the flexible tubing to the calibration adapter/splash guard is clear and connected to the calibration adapter/splash guard.
5. If the calibration/response difficulties continue, replace the detector.
6. If the calibration/response difficulties continue, contact RKI for further instruction.
31 65-2615RK-05-04 M2 Transmitter Operator’s Manual
Calibration Frequency
Although there is no particular calibration frequency that is correct for all applications, a calibration frequency of every 3 to 6 months is adequate for most hydrogen sulfide transmitter applications. Unless experience in a particular application dictates otherwise, RKI Instruments, Inc. recommends a calibration frequency of every 3 months.
If an application is not very demanding, for example detection in a clean, temperature controlled environment where hydrogen sulfide is not normally present and calibration adjustments are minimal at calibration, then a calibration frequency of every 6 months is adequate.
If an application is very demanding, for example if hyd rogen sulf id e is present often and in significant concentrations or the environment is not well controlled, then more frequent calibration than every 3 months may be necessary.
Determining Response Time
Since the detector junction box is installed remotely from the M2, the response time of the detector will depend on the length of flexible tubing tha t is attached to the calibration adapter/splash guard. To determine the response time based on the length of tubing connected:
1. Place the M2 into its calibration mode or disable external alarms.
2. Note the current gas reading displayed at the M2.
3. Screw the regulator into the calibration cylinder. Do not use a zero air cylinder for this operation.
4. Use the flexible tubing coming from the calibration adapter/splash guard to connect the regulator to the calibration adapter/splash guard.
5. Turn the regulator’s on/off knob counterclockwise to open the regulator. Gas will begin to flow. Take note of the time or start a stopwatch.
6. When you first start to notice an increase in the gas reading, note how much time passed between starting the gas flow and noticing a reading response.
7. Turn the regulator’s on/off knob clockwise to close the regulator.
8. Unscrew the regulator from the calibration cylind er.
9. Disconnect the calibration adapter/splash guard’s flexible tubing from the regulator.
NOTE: Do not disconnect the flexible tubing from the calibration adapter/splash guard.
10. Add 1 minute to the length of time determined in step 6. This is the response time for the detector. When performing a response test, zero adjustment, or span adjustment, gas must be applied for this length of tim e .
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 32
Calibration
This section describes how to calibrate the M2 transmitter. It includes procedures to prepare for calibration, enter Calibration Mode, adjust the fresh air (zero) setting, adjust the span setting, and return to normal operation.
WARNING : The M2 i s not an active gas monitoring device dur ing the calibration
procedure. The 4-20 mA output signal will “freeze” at 3.5 mA and all relays will remain in their non-alarm state while the M2 is in Calibration Mode. The output signal will not indicate current readings and the relays will not resume operating normally until the M2 is in normal operation again.
NOTE: The following procedure assumes the use of a calibration kit which includes a
calibration gas cylinder, a zero air cylinder and a fixed flow regulator with an on/off knob.
NOTE: While in the calibration program, if there is no switch activity for the
calibration time-out period the unit will return to normal operation. See “Viewing & Changing M2 Parameters” on page 23 for instructions to set the calibration time-out.
NOTE: The following procedure assumes that the target gas is present in a high
enough concentration to affect the fresh air (zero) reading. If a fresh air environment can be verified, applying zero air to the detector is not necessary when setting the zero reading.
Non-Intrusive Calibration
If the M2 is installed in a classified area and non-intrusive calibration is required, follow the instructions below, but do not remove the junction box cover. Use the magnetic wand accessory to actuate the magnetic switches instead of pushing the control switch buttons. To actuate a magnetic switch and perform th e same operat ion as pres sing a contr ol s witch button, touch the magnet in the end of the magnetic wand to the M2’s glass cover directly above the magnetic switch you wish to actuate. Touching the glass and removing the wand is the same as pressing and releasing a button. Touching the glass and keeping the wand in place is the same as pressing and holding a button.
Calibration Gas Response Memory Feature
The M2 has the capability to “remember” the detector’s response to the calibration gas after the gas is removed from the detector during the fresh air and span adjustment procedure. This feature enables one person to perform calibration if the detector is mounted remotely from the M2. When zero air is applied to the M2 during a fresh air adjustment, the M2 will freeze the display reading at the lowest response and the M2 will continue to display this reading and retain it in its mem ory until the fresh air adjustment procedure is completed. When calibration gas is applied to the M2 during a span adjustment, the M2 will freeze the display reading at the highest response to the calibration gas. The calibration gas can then be removed and the M2 will continue to display this reading and retain it in its memory until the spa n a d justment procedure is completed.
33 65-2615RK-05-04 M2 Transmitter Operator’s Manual
Adjusting the Fresh Air Reading
1. While in normal operation, press and hold the UP/YES button for 5 seconds to enter Calibration Mode. Release the button when the following screen appears.
Calib?
YES/NO
2. If you want to continue with calibration, press and release the UP/YES button. The display will indicate the target gas and CAL Mode for a few seconds before showing FreshAir Adjust?.
If you want to exit Calibration Mode, press and release the DOWN/NO button. The M2 will indicate Leaving CAL Mode and the M2 will return to normal operation.
3. T o continue with adjusting the fresh air reading, press and release the UP/YES button. ENTER will alternate with FreshAir on the top display line and the current gas reading will be on the bottom display lin e.
To skip adjusting the fresh air reading, press and release the DOWN/NO button. The display will indicate SP A N w/ C a l Gas?. Skip to the next section, “Adjusting the Span Setting”.
4. Screw the zero air cylinder into the fixed flow regulator.
5. Use the flexible tubing coming from the calibration adapter/splash guard to connect the regulator to the calibration adapter.
6. Turn the regulator knob counterclockwise to open the regulator.
7. Allow zero air to flow for the length of time determined in “Determining Response Tim e” on page 32.
8. Turn the regulator knob clockw ise to close the regulator.
9. Unscrew the zero air cylinder from the fixed flow regulator. Leave the calibration adapter/splash guard’s flexible tubing connected to the regulator. The M2 will continue to display the minimum gas response on the display and retain the response level in its memory.
10. Press and release the ENTER button. The M2 will perform a fresh air adjustment and the display will indicate SPAN w/Cal Gas?.
NOTE: If the fresh air adjustment fai ls, see “Troubleshoot ing ” on page 30 for
recommended actions.
Adjusting the Span Setting
1. If you want t o con tin ue w ith adj usti ng t he sp an se ttin g, pr es s an d r ele ase the UP/YE S button. APPLY will alternate with SPAN Gas on the top display line and the current gas reading will be on the bottom display line.
If you want to skip adjusting the span reading, press and release the DOWN/NO button. The display will indicate Leaving Cal Mode and the M2 will return to normal operation.
2. Screw the calibration gas cylinder into the fixed flow regulator.
3. Turn the regulator knob counterclockwise to open the regulator.
4. Allow calibration gas to flow for the length of time determined in “D et ermining
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 34
Response Time” on page 32.
5. Turn the regulator knob clockwise to close it.
6. Unscrew the calibration gas cylinder from the fixed flow regulator. The M2 will continue to display the maximum gas response on the display and retain the response level in its memory.
7. Adjust the gas reading up or down to match the calibration gas cylinder concentration by using the UP/YES and DOWN/NO buttons, then press and release the ENTER button.
8. The M2 will perform a span operation. The display will indicate SPAN Gas PASS for a few second s, then in dicat e SPAN Gas SAVED before indicating Leaving CAL Mode for a few seconds.
NOTE: If the span adjustment fails, see “Troubleshooting” on page 30 for
recommended actions.
9. The display will now alternate between the normal operation screen and the message REMOVE CAL GAS for 1 minute. If the calibration gas has not been removed from the detector, remove it now to avoid unwanted alarms.
During this 1 minute period, the signal output will remain fixed at 3.5 mA and the relays will remain in their non-alarm state to avoid unwanted alarms while the calibration gas clears from the detector. At the end of the 1 minute period, the REMOVE CAL GAS message will stop appearing and the M2 will return to normal operation.
10. Disconnect the calibration adapter/splash guard’s flexible tubing from the regulator.
NOTE: Do not disconnect the flexible tubing from the calibration adapter/splash guard.
11. Store the components of the calibration kit in a safe place.
Replacing Components of the H2S Detector
This section includes a procedure to replace the H2S plug-in sensor and the entire detector assembly. In most case s, it is not necessary to replace the entire detector a sse mbly.
Replacing the Plug-In H
CAUTION: The sensor contai ns el ectr o lyt e whic h is a dil ute ac id. Do not disass em ble the se nso r
when replacing it with a new one. If sensor electrolyte comes in contact with your skin, wash affected area thoroughly with soap and water.
1. Turn off or unplug power to the M2.
2. Unscrew the calibration adapter/splash gua rd from the detector hou sing cap.
3. Unscrew the detector housing cap from the detector housing body. Make sure not to lose the cap gasket.
4. Unplug an d remove the H
5. Remove the rubber boot and spacer from the old sensor.
S Sensor
2
S sensor with the rubber boot and spacer attached.
2
35 65-2615RK-05-04 M2 Transmitter Operator’s Manual
6. Install the spacer and rubber boot onto the replacement sensor’s face.
7. Carefully plug the replacement sensor into the four-socket pattern that is located in the detector housing.
8. Make sure the cap gasket is in place and screw the detector housing cap back onto the detector housing body.
9. Screw the calibration adapter/splash guard back onto the detector housing cap.
10. Turn on or plug in power to the M2.
CAUTION: Allow the r ep la cem ent sens or to w arm up for 5 m inu tes before you continue with the
next step.
11. Calibrate the transmitter as described in the “Calibration” on page 33.
Replacing the H
S Detector
2
NOTE: In most cases, it is only necessary to replace the plug-in H2S sensor.
1. Turn off or unplug power to the M2.
2. Remove the detector junction box cover.
3. Unscrew the calibration adapter/splash gua rd from the detector hou sing cap.
4. Disconnect the detector leads from the terminal block in the detector junction box. Note the position of the color-coded leads as you remove them.
5. Unscrew the detector from the detector junction box.
6. Guide the detector leads of the replacement detector through the bottom conduit hub of the detector junction box, then screw the mounting threads of the detector into the conduit hub. If necessary for environmental conditions, apply thread s eal an t o r t eflon tape to the hub and/or detector threads to seal them.
7. Connect the detector leads to the terminal block the same way the old detector was wired (see Figure 6 on page 13).
8. Screw the calibration adapter/splash guard onto the new detector housing cap.
9. Reinstall t he detector junct ion box cover.
10. Turn on or plug in power to the M2.
CAUTION: Allow the replacement detector to warm up for 5 minutes before y o u c o ntinue with
the next step.
11. Calibrate the replacement detector as described in “Calibration” on page 33.
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 36
Chapter 8: RS-485 Modbus Output
Overview
This chapter describes the M2’s RS-485 Modbus output and how to configure the M2 to make use of it. It also discusses how to wire the M2 into a Modbus syst e m.
The M2 provides an RS-485 serial communications interface. It is a Modbus Slave Device, supporting 2-wire RS-485 Modbus RTU serial communicatio ns.
Wiring the M2 in a Modbus System
The M2 is a 2-wire Modbus RTU device. When wiring the M2 into a Modbus system, adhere to standard Modbus wiring practices per the Modbus Over Serial Line Specifica tio n a nd I mpl eme nt ati on Gu id e V1 .0 or la te r. This do cum ent c an b e foun d on lin e at www.modbus.org/specs.php.
The Modbus protocol supports a maximum of 247 unique slave addresses (1-247). The M2’s line driver provides for up to 128 M2s to be connected together without the need for a repeater . Figure 10 & Figure 1 1 below illustrate typical M2 wiring configurations. If more than 128 M2s need to be connected together, RS-485 repeater(s) should be used such that no more than 128 M2s reside on any given network segment.
CAUTION: The network segment can only support 128 units if RKI M2s are being used. The
use of any other instruments will lower the number of units that may reside on any given network segment.
NOTE: Only M2s produced after November 2010 have the capability to support a 128-
unit segment. M2s produced before November 2010 can only support a 32-unit segment.
The MODBUS terminal strip is located on the terminal PCB (see Figure 2). The following signals are available at the Modbus terminal strip:
Table 8: Modbus Terminal Strip Signals
Modbus T erminal Label (RS-485
Name)
AD0 BD1 CCommon
Modbus Signal Name
Recommended Modbus Wiring
The recommended Modbus wiring for the M2 is illustrated in Figure 8 below. In this configuration, 5 wires are used for wiring the M2 into a Modbus system. Figure 8 also illustrates typical alarm device wiring.
37 65-2615RK-05-04 M2 Transmitter Operator’s Manual
Power Supply
al
al
24 VDC
Input
Terminals
Modbus
Controller
Common
S
PWR/SIG
See
Detector
Wiring
D
1
D
0
TOXIC OXY
RS 485
A B C
ALARM 2 ALARM 1 FAIL
C NC NO C NC NO C NC NO
See
Detector
Wiring
S
PWR/SIG
TOXIC OXY
RS 485
A B C
ALARM 2 ALARM 1 FAIL
C NC NO C NC NO C NC NO
Figure 8: Recommended Modbus Wiring
Alternate Modbus Wiring For Existing Installations
Although the wiring shown in Figure 8 is recommended, it is possible to wire the M2 into a Modbus system with only 4 wires in situations where a pre-existing system is being replaced and wiring is already in place. This should only be done if wi ring for a system that is being replaced is already installed and it is not practical to run another wire. See Figure 9 below for this wiring configuration. In this configuration, the wire between the “C” terminal on the M2 and the Common terminal on the Modbus controller is omitted. Instead, the Common terminal on the Modbus controller is connected to the “-” of the 24 VDC power supply.
To Addition M2s
Power Supply
24 VDC
Common
Input
Terminals
Modbus
Controller
S
PWR/SIG
See
Detector
Wiring
D
1
D
0
TOXIC OXY
RS 485
A B C
ALARM 2 ALARM 1 FAIL
C NC NO C NC NO C NC NO
See
Detector
Wiring
S
PWR/SIG
TOXIC OXY
RS 485
A B C
ALARM 2 ALARM 1 FAIL
C NC NO C NC NO C NC NO
To Addition M2s
Figure 9: Alternate Modbus Wirin g
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 38
Termination Jumper
ID = 127 ID = 126 ID = 125 ID = 1
ID = 128
7
Up to 128 M2s can be connected without a repeater.
The M2 includes a 2-pin termination header (see Figure 2) that is used when the M2 is used in a Modbus system. Every M2 is supplied with a termination jumper (a jumper block) installed onto this header. If the M2 is not used in a Modbus system, this jumper has no function. When the M2 is installed in a Modbus system, this jumper must be installed in an M2 that is at the end of a Modbus line. Any M2 in a Modbus system that is not at the end of a line must have the termination jumper removed (see Figure 10 & Figure 11 below).
ID = 128
M2
Termination
Jumper
Installed
RS-485 RS-485
Up to 128 M2s can be connected without a repeater.
Modbus
Master
M2
Termination
Jumper
Not Installed
M2
Termination
Jumper
Not Installed
M2
Termination
Jumper
Not Installed
Figure 10: Multiple M2s in a Daisy Chain Configuration
ID = 1
M2
Termination
Jumper
Installed
RS-485 RS-485
M2
Termination
Jumper
Not Installed
M2
Termination
Jumper
Not Installed
Modbus
Master
M2
Termination
Jumper
Not Installed
M2
Termination
Jumper
Not Installed
M2
Termination
Jumper
Installed
M2
Termination
Jumper
Not Installed
ID = 2 ID = 3 ID = 4
Figure 11: Multiple M2s in a Two Branch Configuration
39 65-2615RK-05-04 M2 Transmitter Operator’s Manual
ID = 12
Using the M2 in a 4-wire Modbus System
Although the M2 is a 2-wire Modbus RTU device, it can be used with a 4-wire Modbus controller if the system wiring is modified as follows:
Connect the controller’s TxD0 and RxD0 wires together and use this connection as the 2-wire Modbus D0 signal.
Connect the controller’s TxD1 and RxD1 wires together and use this connection as the 2-wire Modbus D1 signal.
Connect these D0 and D1 signals and the common wire from the controller to the 2­wire Modbus Network of M2s.
Modbus Mode
Modbus Mode allows you to configure various parameters relating to the Modbus setup of the M2. Modbus Mode includes a 5-minute time-out feature. If you do not press a control button for 5 minutes, the M2 automatically returns to norma l operation.
NOTE: If the M2 returns to normal operation because of a time-out, it enters a warm-
up period just as it does when it is first turned on.
1. While in normal operation, simultaneously press and hold the DOW N /NO and ENTER buttons for 5 seconds to enter Modbus Mode. Release the buttons when the following screen appears.
Set Up Modbus?
2. To exit Modbus Mode, press and release the DOWN/ NO button. The M2 will return to normal o peration.
To continue in Modbus Mode, press and release the UP/YES button. The current ENABLED/DISABLED setting will appear.
3. If you want to change the currently displayed parameter, use the UP/YES and DOWN/NO buttons to adjust it to the desired setting, then press ENTER to continue to the next parameter.
If the currently displayed parameter setting is OK, press the ENTER button to proceed to the next parameter.
Table 9 lists the M2 Modbus Mode p ara me te rs you can set. Table 9 also lists the factory set value for each parameter.
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 40
Table 9: Configuration Parameters
Modbus Mode Parameter Available Settings & Description
Enabled/Disabled Can be set to ENABLED or DISABLED (factory setting).
Enables or disables the Modbus output.
Slave ID The Slave ID can be set to values from 1 (factory setting) to 247.
The M2 will only receive messages from the Master which are addressed to this Slave ID (except for broadcast messages which are received by all slaves).
Baud Rate Can be set to 1200, 2400, 4800, 9600 (factory setting), 14440, &
19200. Parity EVEN (factory setting), ODD, NONE Response Delay The response delay can be set from 0 (factory setting) to 20 mS.
This is an optional additional delay inserted by the M2 prior to returning a response message to the Master. It may be helpful in some installations where the Master’s preparedness to receive responses might be delayed.
4. When you have scrolled through all the adjustable parameters, SAVE IT? YES/NO
appears on th e display.
5. If you do not wish to save the adjustments, press and release the DOWN/NO button.
The DO OVER? YES/NO message will display. Press and release the DOWN/NO button. The ABORT? YES/NO message will display. Press the UP/YES button to return to normal operation.
If you wish to change some of the adjustments made, press and release the DOWN/ NO button. The DO OVER? YES/NO message will display. Press and release the UP/ YES button. The Re-do MB Setup message will display and the M2 will return to the first adjustable parameter. Go back to step 3 and continue.
If you want to save the adjustments made, press and release the UP/YES button. Modbus Saved is indicated on the display for a few seconds and the M2 returns to normal operation.
NOTE: All M2s on a Modbus network must be configured with the same baud rate and
parity, and each M2 must have a unique slave ID. For an M2 to participate on a network, the ENABLED/DISABLED setting must be set to ENABLE.
41 65-2615RK-05-04 M2 Transmitter Operator’s Manual
Supported Modbus Functions
The M2 supports Function Code 03: Read Holding Registers and Function Code 16: Write Registers. The register assignments detailed below were implemented in M2 firmware version 5.0. Please see revision B of the M2 manual if you have an M2 with a firmware version previous to 5.0.
WARNING: Do not attempt to use registers according to the instructions below with
units that have firmware versions previous to 5.0. For the Modbus register assignments of M2s with firmware previous to 5.0, see revision B of the M2 Operator’s Manual. The firmware version is shown on the Information Screen described on page 18.
Function Code 03: Read Holding Registers
There are a total of 40 registers available to be read via this Function Code.
Registers 1 - 2
The format (bit & field assignments) for registers 1 - 2 is shown below in Table 10 ­Table 11.
Table 10: Register 1, Numerical Reading with Sign & Decimal Point Bit & Field Assignments
Bit(s) Value & Field
[15] Sign of Reading
0 = Positive 1 = Negative
[14:13] Format of Reading
0 = XXX (no decimal point) 1 = XX.X 2 = X.XX 3 = .XXX
[12:0] Numerical value of display reading (8000 max)
Table 11: Register 2, Operating State, Alarms & Relays Bit & Field Assignments
Bit(s) Value
[15:12] Operating State
0 = Undefined 1 = Initialization 2 = Normal 3 = Warm -up 4 = Fail 5 = Calibration 6 = Post-Calibration 7 = Low Power 8 = Setup
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 42
Table 11: Register 2, Operating State, Alarms & Relays Bit & Field Assignments (Continued)
Bit(s) Value
[11] Gas Type Change
0=Not Changed 1=Changed
[10] Configuration Change
0=Not Changed 1=Changed
[9] Calibration Activity Flag
0=No Calibration Activity 1=Calibration Activity Has Occurred
[8] Fail Status
0 = Fail Not As serted 1 = Fail Assert ed
[7:6] Alarm 2 Status
0 = No Alarm 1 = Unacknowle dged Alarm 2 = Acknowledged Alarm 3 = Unused Code
[5:4] Alarm 1 Status
0 = No Alarm 1 = Unacknowle dged Alarm 2 = Acknowledged Alarm 3 = Unused Code
[3] Overscale Status
0 = Not Overscale 1 = Overscale
[2] Fail Relay Status
0 = Not Energized 1 = Energized
[1] Alarm 2 Relay St atus
0 = Not Energized 1 = Energized
[0] Alarm 1 Relay St atus
0 = Not Energized 1 = Energized
Register 3
Register 3 is the supply voltage (0.1 volt per count).
Registers 4 - 11
The 16 bytes in these registers contain the same ASCII text information that is displayed on the M2’s LCD (not NULL Terminated).
43 65-2615RK-05-04 M2 Transmitter Operator’s Manual
Register 12
Register 12 is the range (full scale readout).
Register 13
Register 13 is t he alarm 1 set point. The decimal point location is the same as specified in Register 1.
Register 14
Register 14 is t he alarm 2 set point. The decimal point location is the same as specified in Register 1.
Registers 15 - 19
Registers 15 - 19 are the Gas Name ASCII String (NULL Terminated).
Registers 20 - 22
Registers 20 - 22 are the Gas Units String (NULL Terminated).
Register 23
Register 23 is t he alarm 1 trigger. 0=Decreasing 1=Increasing
Register 24
Register 24 is the alarm 1 normal relay state. 0=Normally De-Energized 1=Normally Energized
Register 25
Register 25 is the alarm 1 relay reset. 0=Latching 1=Self-Resetting
Register 26
Register 26 is the alarm 1 ON delay. Integer values are in seconds.
Register 27
Register 27 is the alarm 1 OFF dela y. Integer values are in seconds.
Register 28
Register 28 is t he alarm 2 trigger. 0=Decreasing 1=Increasing
Register 29
Register 29 is the alarm 2 normal relay state. 0=Normally De-Energized 1=Normally Energized
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 44
Register 30
Register 30 is the alarm 2 relay reset. 0=Latching 1=Self-Resetting
Register 31
Register 31 is the alarm 2 ON delay. Integer values are in seconds.
Register 32
Register 32 is the alarm 2 OFF dela y. Integer values are in seconds.
Register 33
Register 33 is zero suppression. Toxics and LEL: 0%-6% of full scale Oxygen: 0%-0.7% Oxygen The decimal point location is the same as specified in Register 1.
Register 34
Register 34 is noise filter. Integer values in seconds. 0-60 in 5 second increments.
Register 35
Register 35 is calibration timeout in minutes (5, 10, 15, 20, 25, or 30).
Register 36
Register 36 is calibration month (upper byte) and day (lower byte). This is zero based (e.g. January=0 and the 1st=0).
Register 37
Register 37 is the calibration year.
Register 38
Register 38 is the calibration outcome. Upper=Gas Operation Lower=Air Operation
Calibration Outcome Codes:
1=Sensor Missing 2=Sensor Too Weak 3=Signal Too Strong 4=Not Enou gh Span 5=Negative Re ading 6=Span Gas Value Is Not OK 7=Calculated Zero Is Too Low 8=Calculated Zero Is Too High 9=Zero Failure 10=Span Failure 11=Calibration Is Overscale 12=LEL Sensor Offset Is Too High 13=LEL Null Operation Failed
45 65-2615RK-05-04 M2 Transmitter Operator’s Manual
14=Operat ion Not Performed
Register 39
Register 39 is the auto-zero disable. 0=AutoZero Enabled 1=AutoZero Disabled
Register 40
Register 40 is the remote configuration register access level 0=None 1=Alarm Reset
Function Code 16: Write Registers
There are 21 registers in Function Code 16 that can be used to write to the M2. This manual only describes Register 16 because it can be used to reset an al arm condition . For a complete description of Function Code 16, request Appendix C from RKI Instruments, Inc.
Register 16
Register 16 performs the same function as the ENTER button on the M2. It is useful for remotely silencing alarms. With the M2 in an alarm condition, bit [0] can be used to silence it.
Table 12: Register 16, Button Function
NOTE: Register 40 in Function Code 03, the remote configuration register, must be set
to 1 in order for Register 16 to work.
Bit Value & Field
[0] ENTER (used for “Alarm Reset”)
0=Not Pressed 1=Pressed
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 46
Parts List
06-1248RK Calibration kit sample tubing (3/16 in. x 5/16 in.; specify length
07-0033RK Detector housing cap gasket 07-0203RK R ubber re taining boot 14-2101RK Spacer between sensor and rubber boot 18-0400RK-01 Detector junction box, with spacers 18-0431RK-02 M2 junction box 45-6132RK Termination jumper
Table 13 lists replacement parts and accessories for the M2 Transmitter.
Table 13: Parts List
Part Number Description
when ordering)
65-2423RK-05 H
S replacement detector assembly (in c ludes sensor), CSA classified
2
71-0326 65-2615RK-05-04 M2 Transmitter Operator’s Manual (this document) 81-0076RK-01 Zero air calibration cylinder, 34 liter steel 81-0076RK-03 Zero air calibration cylinder, 103 liter steel 81-0151RK-02 Calibration cylinder, H
81-0151RK-04 Calibration cylinder, H
S, 25 ppm in nitrogen, 58 liter aluminum
2
S, 25 ppm in nitrogen, 34 liter aluminum
2
81-1050RK Regulator, with gauge and knob, 0.5 liter/minute continuous flow 81-1116RK Calibration adapter/splash guard ES-1537-H2S Replacement plug-in sensor, H
S
2
47 65-2615RK-05-04 M2 Transmitter Operator’s Manual
Appendix A: Control Button Quick Reference Guide
The M2’s control buttons allow access to operational modes, resetting of alarms, and display of the Information Screen. Table 14 shows which button combinations perform these functions and which parameters are available for adjustment while in the operational modes. While in these mo d es , di splay prompts showing a “?” require you to respond by pressing either the UP/YES (for yes) or DOWN/NO (for no) button. To change parameter settings, use the UP/YES and DOWN/NO buttons to get the desired setting, then press and release the ENTER button to accept the setting.
Table 14: Control Button Quick Reference Chart
Mode Entered or
Function Performed
Alarm Reset Press and
Show Information Screen
Calibration Mode Press and Hold
Configuration Mode Press and Hold
Gas Type Mode Press and Hold
Button
Combination
Release ENTER Press and Hold
DOWN/NO
UP/YES for 5 seconds
UP/YES & ENTER for 5 seconds
UP/YES and DOWN/NO for 5 seconds
Adjustable or Viewable Parameters
n/a
•Input Voltage
• M2 Version/Revision Information
• Fresh Air Adjust
• Span (or Zero for Oxygen) Adjust with Calibration Gas
• Alarm Levels
• Increasing or Decreasi ng Alarms
• Normally Energized or Normally De­energized Relays
• Latching or Self-resetting Alarms
• Alarm On Delay
• Alarm Off Delay (if self-resetting)
• Zero Suppression
• Noise Filter
• Calibration Time-out Period
Gas Type
Modbus Mode Press and Hold
DOWN/NO and ENTER for 5 seconds
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 48
• Enable or Disable Modbus
•Slave ID
•Baud Rate
•Parity
• Response Delay
Appendix B: PLC and DCS Device Wiring
The M2 can be wired to a PLC or DCS device if desired.
1. Guide multi conductor shielded cable or cables or wires in conduit through the top conduit hub of the junction box. The number of cables or wires needed will depend on whether any relays are used and whether the Modbus output is used. Use the following recommendations to determine how to wire the M2:
If Modbus connections will not be used and only the PWR/ SIG terminal strip
connections will be used, use four conductor shielded cable or four wires in conduit for connections to the power/signal terminal strip.
Table 15: Wire Size for PWR/SIG Connections
Max Distance to Controller
w/18 Gauge Wire
2,500 ft. 5,000 ft. 8,000 ft.
If the PWR/SIG terminal strip connections and one or more relays ar e used, route
the connections to the M2 in conduit. Use shielded cable in the conduit for the PWR/SIG connections and unshielded cabl e or individual wires for the relay connections. Make sure any wire or cable used for relay wiring is appropriately rated for the power that it will carry.
NOTE: If shielded cable is used for the PWR/SIG connections, leave the cable shield’s
drain wire insulated and disconnected at the M2. You will connect the opposite end of the cable’s drain wire at the controller or device.
If the M2 will be wired into a Modbus network, see “Chapter 8: RS-485 Modbus
Output” on page 37.
Max Distance to Controller
w/16 Gauge Wire
Max Distance to Controller
w/14 Gauge Wire
49 65-2615RK-05-04 M2 Transmitter Operator’s Manual
24 VDC
Power Suppl y
24 VDC +
See Figure 12 below for field wiring connections to the M2.
24 VDC -
4 - 20 mA In (S )
(24 VDC) -
Controller
(PLC, DCS)
Alarm Device
FAIL
C NC NO
See
S
PWR/SIG
+
Detector
Wiring
TOXIC OXY
+
C NC NO
ALARM 1
See
Modbus
Wiring
A B C
RS 485
C NC NO
ALARM 2
Figure 12: PLC and DCS Device Wiring
Power
Fai l Alarm Device
Ala rm 1 Ala rm Dev i ce
Ala rm 2 Ala rm Dev i ce
Typic a l Alarm W iring Shown
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 50
Appendix C: Function Code 16 Registers
The M2 supports Function Code 16 that allows writing to the M2. There are 21 registers in this Function Code. The register assignments detailed below were implemented in M2 firmware version 5.0. Please see revision B of the M2 manual if you have an M2 with a firmware version previous to 5.0.
WARNING: Do not attempt to use registers according to the instructions below with
units that have firmware versions previous to 5.0. For the Modbus register assignments of M2s with firmware previous to 5.0, see revision B of the M2 Operator’s Manual. The firmware version is shown on the Information Screen described on page 18.
Register 1
Register 1 is the alarm 1 set point. Value must be an integer. Any decimals are omitted.
Register 2
Register 2 is the alarm 1 trigger. 0=Decreasing 1=Increasing
Register 3
Register 3 is the alarm 1 relay state. 0=Normally De-Energized 1=Normally Energized
Register 4
Register 4 is the alarm 1 relay reset. 0=Latching 1=Self-Resetting
Register 5
Register 5 is the alarm 1 ON delay. Integer value is in seconds.
Register 6
Register 6 is the alarm 1 OFF delay. Integer value is in seconds.
Register 7
Register 7 is the alarm 2 set point. Value must be an integer. Any decimals are omitted. The decimal point location is the same as specified in Register 1 of Function Code 03.
Register 8
Register 8 is the alarm 2 trigger. 0=Decreasing 1=Increasing
51 65-2615RK-05-04 M2 Transmitter Operator’s Manual
Register 9
Register 9 is the alarm 2 relay state. 0=Normally De-Energized 1=Normally Energized
Register 10
Register 10 is the alarm 2 relay reset. 0=Latching 1=Self-Resetting
Register 11
Register 11 is the alarm 2 ON delay. Integer value is in seconds.
Register 12
Register 12 is the alarm 2 OFF dela y. Integer value is in seconds.
Register 13
Register 13 is the zero suppression. Toxics and LEL: 0%-6% Full Scale Oxygen: 0%-0.7% Oxygen Value must be an integer. Any decimals are omitted.
Register 14
Register 14 is the noise filter. Integer value is in seconds (0-60 in 5 second increments).
Register 15
Register 15 is the calibration timeout. Value is in minutes (5, 10, 15, 20, 25, or 30).
Register 16
Register 16 is explained in “Supported Modbus Functions” in the M2 manual.
Register 17
Register 17 is the calibration month (MSB) and day (LSB). This is zero-based (e.g. January=0 and the 1st=0).
Register 18
Register 18 is the calibration year.
65-2615RK-05-04 M2 Tran smitter Operator’s Manual 52
Register 19
Register 19 is for clearing change flags. When parameters are changed at the M2, a flag is raised at the controller. This register can be used to clear those flags.
Table 16: Register 19, Clear Change Flags
Bit(s) Value & Field
[15:2] Unused [1] Write “1” to this bit to clear “Gas Type Changed” flag [0] Write “1” to this bit to clear “Configuration Changed” flag
Register 20
Register 20 is the AutoZero Disable. 0=AutoZero Enabled 1=AutoZero Disabled
Register 21
Register 21 is the Remote Configuration Regi ster. 0=No Remote Configuration 1=Reset Alarm
53 65-2615RK-05-04 M2 Transmitter Operator’s Manual
Loading...