RKI Instruments 65-2516RK User Manual

Page 1
65-2516RK
Oxygen Transmitter
Operator’s Manual
Part Number: 71-0116RK Revision: 0 Released: 3/1/11
www.rkiinstruments.com
Page 2
WARNING
Read and understand this instruction manual before operating detector . Improper use of the dete ctor could result in bodily harm or death.
Periodic calibration and maintenance of the detector is essential for proper operation and correct readings. Please calibrate and maintain this detector regularly! Frequency of calibration depends upon the type of use you have and the sensor types. T ypical calibration frequencies for most applications a re between 3 and 6 months, but can be required more often or less often based on your usage.
Page 3
Product Warranty
RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year fr o m date of shipment from RKI Instruments, Inc. Any parts found defective withi n tha t period will be repaired or replaced, at our option, free of charge. This warranty does not apply to those items which by their nature are subject to deterioration or consumption in normal ser v ice, and which must be cleaned, repaired, or replaced on a routine basis. Examples of such items are:
W arranty is voided by abuse including mechanical damage, alteration, rough handling, or repair procedures not in accordance with the operator’s manual. This warranty indicates the full extent of our liability , a nd we are not r esponsible for removal or r eplacement costs, local repair costs, transportation costs, or contingent expenses incurred without our prior approval.
a) Absorbent cartridges d) Batteries b) Pump diaphragms and valves e) Filter elements c) Fuses
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI INSTRUMENTS, INC. INCLUDING BUT NOT LIMITED TO, THE WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL RK I INSTRUMENTS, INC. BE LIABLE FOR INDIRECT, INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
This warranty covers instruments and parts sold to users by authorized distributors, dealers, and representatives as appointed by RKI Instruments, Inc.
We do not assume i ndemnification for a ny accident or damag e c au sed by the operatio n of this gas monitor, and our warranty is limited to the replacement of parts or our complete goods.
65-2516RK Oxygen Transmitter
Page 4
Table of Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Oxygen Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Junction Box. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Mounting the Oxygen Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Wiring the Oxygen Transmitter to a Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Startup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Introducing Incoming Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Setting the Fresh Air Signal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Replacing Components of the Oxygen Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Calibration Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Preparing for Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Setting the Fresh Air Reading. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Setting the Zero Reading. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Returning to Normal Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Page 5
Overview
!
This instruction manual describes the explosion proof capillary type oxygen transmitter. This manual also describes how to insta ll, start up, configure, maintain, and calibrate the transmitter when using it with a gas monito ring controller. A parts list at the end of this manual lists replacement parts and accessories for the oxygen transm itter. See the controller operator’s manual for informat ion specific to the controller.
Specifications
WARN ING: Do not use this product in a manner not specified in thi s instruction
Table 1 lists specifications for the oxygen transmitter.
manual.
Table 1: Specifications
Target Gas Oxygen (O Area Classification Explosionproof for Class I, Groups B, C, and D
Temperature Code T6 Installation Category Installation Category 1. Signal level, special equipm ent or parts of
equipment, telecommunication, electronic, etc., with smaller transient overvoltages than Installati on Category (Overvoltage
Category) II (ref. IEC 664). Input Voltage 11 VDC - 30 VDC Sampling Method Diffusion Signal Output 4 to 20 mA Detection Range 0 to 25% volume Accuracy ± 0.5% O
Oxygen Detector Signal Output 0 mV at 0% volume oxygen nominal
18 mV at 25% oxygen nominal Response Time 90% in 30 seconds Operating Temperature -20°C to 45°C
)
2
2
NOTE: The following symbol on the detector label is a caution to the user to refer to this
documentation for installation and operation instructions:
W ARNING: When using the 65-2516RK, you must follow the instructions and warnings
in this manual to assure proper and safe operation of the 65-2516RK and to minimize the risk of personal injury. Be sure to maintain and period ically calibrate the 65-2516RK as described in this manual.
65-2516RK Oxygen T r an s mi t te r • 1
Page 6
Description
g
I
nstalled On Oxygen
Detector Terminal
This section describes the components of the oxygen transmitter. The transmitter consists of the oxygen detector, amplifier, and junction box.
3/4" NPT Conduit Openin
Test Point (-)
Factory Set
Potentiometer
Zero
Potentiometer
Span
Potentiometer
Jumper Block
Select Header
Strip
ZERO
SPAN
OXYGEN
TP
-
FB
Not
Used
4/20
POWER/SIG
SENSOR
OXY
Not
Used
W
+
TP
+
24V
G
For Wire Entry
Interconnect Terminal Strip
Test Point (+)
Amplifier
Securing Screw (2X)
Detector Housing
Detector Housing Cap
Oxygen
Detector
Flame Arrestor Guard
Figure 1: Oxygen Transmitt er Comp on ent Location
Oxygen Detector
The oxygen detector includes the detector housing and sensor.
Detector Housing
The detector housing protects the sensing components within the housing. Use the mounting threads at the top of the housi ng to screw the oxygen detector into the bottom conduit hub of the junction box. Use the removable cap near the bottom of the housing to access the sensor for maintenance or replacement. The cap protects the sensor from damage and includes a flame arrestor which contains any sparks wh ich may occu r withi n the detector housing. A cap gasket seals the interface between the housing and cap. The flame arrestor guard is permanently bonded to the cap.
Two wires extend from the top of the detector housing. Use these wires to connect the oxygen detector to the amplifier . The housing includes sockets installed on a circuit board. These sockets accept the sensor’s two pins to secure the sensor within the detector housing. The circuit board with the sockets conditions the sensor’s signal before the signal reaches the amplifier.
2 • 65-2516RK Oxygen Transmitt er
Page 7
Sensor
The sensor is secured within the sensor housing by the two pins. Through a series of chemical and electrical reactions, the sensor produces an electrical output that is proportional to the detector range of the transmitter.
Amplifier
The amplifier converts the electrical output from the sensor to a 4 to 20 mA signal (that is proportional to the detection range) and transmits the signal to a gas monitoring controller. The amplifier includes the amplifier type selector, detector terminal strip, interconnect terminal strip, span pot, zero pot, and test points (see Figure 1).
Amplifier Type Selector
The amplifier type selector is near the bottom left corner of the amplifier. It is to the left of the detector terminal strip and below the span pot.
The amplifier included with the oxygen transmitter is designed for use with RKI’s oxygen and toxic gas transmitters. The amplifier type selector determines for which transmitter the amplifier is intended. For oxygen transmitters, a jumper blo c k is installed over the OXYGEN selector (see Figure 1).
Detector Terminal Strip
The detector terminal strip is the four-point terminal strip near the bottom of the amplifier. Use the detector terminal strip to connect the oxygen detector to the amplifier.
NOTE: The oxygen detector is factory-wired to the amplifier. See the Installation section
of this manual for all wiring procedures related to the transmitter.
Interconnect Terminal Strip
The interconnect terminal strip is the four-point terminal strip near the to p of the amplifier. Use the interconnect terminal strip to connect the amplifier to a controller.
Span Pot
The span pot is on the left side of the amplifier. Of the three pots, the span pot is bottom most. Use the span pot to adjust the transmitter’s fresh air output during the start-up and calibration procedures.
Zero Pot
The zero pot is above the span pot. Use the zero pot to adjust the transmitter’s oxygen-free output during the calibration procedures.
CAUTION: The third potentiometer is factory-set. Do not adjust it.
Test Points
The test points (labeled TP- and TP+) are on the left and right side of the interconnect terminal strip. The test points produce a 100 to 500 mV output that is proportional to the transmitter’s 4 to 20 mA output. Use the test points and a voltme ter to mea sure the transmitter’s output during the start-up and calibration procedures.
65-2516RK Oxygen T r an s mi t te r • 3
Page 8
Installation
1
g
Junction Box
Use the junction box to install the oxyg en transmitter at a mounting site that is remote from the controller. The junction box also protects the amplifier and wiring connections made to the amplifier. Use the two 3/4 in. conduit hubs to mount the detector to the junction box (bottom hub) and connect wiring from the amplifier to the controller (top hub).
NOTE: The oxygen detector and amplifier are factory-mounted to the junction box.
Use the junction box’s two mountin g holes to mount the oxygen transmitter to a vertical surface at the monitoring site. Use the cover on the front of the junction box to access the interior of the junct ion box.
This section describes procedures to mount the oxygen transmitter in the monitoring environment and wire the transmitter to a gas monitoring controller.
Mounting the Oxygen Transmitter
1. Select a mounting site that is representative of the monitoring environment. Consider the following when you select the mounting site.
Select a site where the transmitter is not likely to be bumped or disturbed. Make
sure there is sufficient room to perform start-up, maintenance, and calibration procedures.
Select a site that is at normal breathing level.
CAUTION: Mount the ox ygen transmitter with the detector facing down (see Figure 2).
.25 Dia. Mountin Hole, 2X
3/4 NPT Conduit Hub
8.5 max
3.2 max
6.10
3.94
5.20
2.75
5.46
1 1/2-20 For Calibration Cup
.10
Figure 2: Mounting the Oxygen Transmitter
4 • 65-2516RK Oxygen Transmitt er
Page 9
NOTE: The oxygen detector used in this transmitter is normally provided with a Killark
HKB junction box and an HFC lid rated explosion pr oo f for Cl as s I, Gr o ups B, C, and D. This combination is shown in Figure 2 above. Any junction box with an internal volume less than or equal to 69 cubic inches and rated explosion proof for Class I, Groups B, C, and D may be used for th is transmitter as long as the amplifier will fit in it.
2. At the monitoring site, use #10 screws through the junction box’s two mounting holes to secure the junction box to a vertical surface.
Wiring the Oxygen Transmitter to a Controller
WARN ING: Always verify that the power source is OFF bef ore you make wiring
connections.
1. Turn off the controller.
2. Turn off or unplug incoming power to the controller.
3. Remove the junction box cover.
4. Verify that the detector leads are wired to the amplifier’s detector terminal strip. If necessary, connect the detector leads to the detector terminal strip as shown in Figure 3.
5. V er ify that the jumper block is installed over the Oxygen selector of the am pl if ier ty pe selector as shown in Figure 3.
6. Guide a two-conductor, shielded cable or two wires in conduit through the top conduit hub of the junction box.
WARN ING: To maintain the explosion proof classification of the oxygen detector/
junction box combination, a conduit seal must be used within 18 inches of the junction box conduit hu b used for wiring to the controller.
7. Connect the two wires to the interconnect terminal strip as follows (see Figure 3).
Connect the positive wire to the terminal labeled 24V +.
Connect the feedback (or signal) wire to the terminal labeled 4/20 FB.
CAUTION: If using shielded cable, leave the drain wire insulated and disconne cted at the
transmitter. You will connect the opposite end of the cable’s drain wire at the controller.
8. Secure the junction box cover to the junction box.
9. Route the cable or wires leading from the oxygen transmitter through one of the conduit hubs at the controller housing.
CAUTION: Do not route power and transmitter wiring through the same conduit hub. The
power cable may disrupt the transmission of the tran smitter signal to t he controller.
65-2516RK Oxygen T r an s mi t te r • 5
Page 10
10. Connect the wires to the to the applicable controller transmitter terminal strip as
er
shown in Figure 3.
4 - 20 mA In (FB or S)
+ 24 VDC
Cable Shield
Controller Transmitt
FB
+
4/20
W
OXY
24V
TP
+
G
Terminals,Typical
ZERO
SPAN
OXYGEN
TP
-
Not
Used
POWER/SIG
SENSOR
Not
Used
Green
White
Figure 3: Wiring the Oxygen Transmitter to a Controller
11. If shielded cable is used, connect the cable’s drain wire to an available chassis (earth) ground at the controller. RKI controllers typically have a ground stud that can be used to ground the cable’s drain wire.
6 • 65-2516RK Oxygen Transmitt er
Page 11
Start Up
This section describes procedures to start up the oxyg en tra nsmitter and place the transmitter into normal operation.
Introducing Incoming Power
1. Complete the installation procedures described earlier in this manual.
2. Verify that the power wiring to the controller is correct and secure. Refer to the controller operator’s manual.
3. Turn on or plug in the incoming power to the controller, then turn on the control ler.
4. Verify that the controller is on and operating properly. Refer to the controller operator’s ma nual.
CAUTION: Allow the transmitter to warm up for 5 minutes before you continue with the next
section, “Setting the Fresh Air Signal”.
Setting the Fresh Air Signal
WARN ING: Do not remove the detector housing cap or junction box cover while the
circuits are energized unless the area is deter mined t o be non-hazardous. Keep the detector housing cap and junction box cover tightly closed during operation.
NOTE: If you can verify that the detector is in a fresh air environment (environment
known to be of normal oxygen content and free of toxic and combustible gases), it is not necessary to apply zero air when verifying or setting the fresh air reading.
The procedure below describes applying zero emission air, usually called zero air , using a calibration kit that includes a calibration cup, calibration gas, sample tubing, and a fixed flow regulator with an on/off knob. RKI Instruments, Inc. recommends using a 0.5 LPM (liters per minute) fixed flow regulator.
1. Verify that the transmitter area is non-hazardous.
2. Unscrew and remove the junction box cover from the junction box.
3. Set a voltmeter to measure in the millivolt (mV) range.
4. Plug the voltmeter leads into the test points on the amplifier. Plug the positive lead into the test point labeled TP+; plug the ne gative lead into the test point labeled TP-.
5. Screw the calibration cup onto the bottom of the detector.
6. Screw the regulator into the zero air calibration cylinder.
7. Use the sample tubing to connect the regulator to the calibration cup.
8. Turn the regulator’s on/off knob counterclockwise to open it. Gas will begin to flow.
9. Allow the gas to flow for 2 minutes.
10. Verify a voltmeter reading of 434 mV (±2 mV).
11. If necessary, use a flat-blade screwdriver to adjust the span pot until the voltmeter reading is 434 mV (±2 mV).
65-2516RK Oxygen T r an s mi t te r • 7
Page 12
12. Turn the regulator’s on/off knob clockwise to close it.
13. Unscrew the calibration cup from the detector.
14. Unscrew the regulator from the zero air calibration cylinder. For convenience, leave
15. Store the components of the calibration kit in a safe and conve nient place.
16. Remove the voltmeter leads from the test points.
17. Secure the junction box cover to the junction box.
Maintenance
This section describes maintenance procedures. It includes preventive maintenance, troubleshooting, and component replacement procedures.
Preventive Maintenance
This section describes a preventive maintenance schedule to ensure the optimum performance of the oxygen transmitter. It includes daily, monthly, and quarterly procedures.
Daily
the sample tubing connected to the regulator and the calibration cup.
Verify a display reading of 20.9% oxygen at the controller. Investigate significant changes in the reading.
Monthly
This procedure describes a test to verify that the oxygen transmitter responds properly to oxygen deficiency.
NOTE: Performing a response test on the oxygen transmitter may cause alarms. Be sure
to put the controller into its calibration program or disable external alarms befor e performing this test.
Preparing for the response test
1. Verify that the display reading for the channel you are testing is 20.9% oxygen. If the display reading is not 20.9% oxygen, set the fresh air reading of the transmitter
as described in “Start Up” on page 7 of this man ual , then conti n ue this pr o c edure.
2. Set a voltmeter to measure in the millivolt (mV) range.
WARN ING: Do not remove the detector housing cap or junction box cover while the
circuits are energized unless the area is deter mined t o be non-hazardous. Keep the detector housing cap and junction box cover tightly closed during operation.
3. Remove the junction box cover, then plug the voltmeter leads into the test points on the amplifier. Plug the positive lead into the test point labeled TP+; plug the negative lead into the test point labeled TP-.
Performing the response test
1. Exhale into the bottom of the oxygen detector.
2. Stop exhaling into the bottom of the detector, then verify that the reading on the
8 • 65-2516RK Oxygen Transmitt er
Page 13
voltmeter decreased from the normal reading (434 mV).
NOTE: If the reading does not decrease, calibrate the transmitter as described in
“Calibrati on” on page 13 of this manual.
3. Remove the voltmeter leads from the amplifier test points, then secure the junction box cover to the junction box.
4. When the display reading rises above the decreasing alarm setpoint, return the controller to normal operation.
Quarterly
Calibrate the detector as described in “Calibration” on page 13. See the calibration frequency discussion in the calibration section to d etermine if a quarterly calibration schedule fits your needs.
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended action for problems you may encounter with the oxygen transmitter.
NOTE: This troubleshooting guide describes transmitter problems only. See the
controller operator’s manual for problems you may encounter with the controller.
Table 2:Troubleshooting the Oxygen Transmitter
Condition Symptom(s) Probable Causes Recommended Action
Fail Condition • Controller indicates a
Slow or No Response/ Difficult or Unable to Calibrate
fail condition.
• Transmitter responds slowly or does not respond to response test.
• Unable to accurately set the fresh air or zero reading during calibration.
• Transmitter requires frequent calibration.
Note: Under “normal” circumstances, the transmitter requires calibration once every three mont hs.
Some applications may require a more freque nt calibration schedule.
• The transmitter wiring is disconnected or misconnected.
• The transmitter is malfunctioning.
• The calibration cylinder is low, out-dated, or defective.
• The flame arrestor in the detector housing cap is wet or clogged with dirt or other particulates.
• The transmitter is malfunctioning.
1. Verify that the transmitter wiring is correct and secure.
2. Calibrate the transm itt er.
3. If the fail condition continues, replace the oxygen sensor.
4. If the fail condition continues, contact RKI for further instruction.
1. Verify that the calibration cylinder contains an adequate supply of a fresh test sample.
2. Check the detector housing cap to determine if the flame arrestor is wet or dirty. Clean or replace if necessary.
3. If the calibration/response difficulties continue, replace the oxygen sensor.
4. If the calibration/response difficulties continue, contact RKI for further instructions.
65-2516RK Oxygen T r an s mi t te r • 9
Page 14
Replacing Components of the Oxygen Transmitter
This section includes procedures to replace the oxygen sensor and amplifier. A procedure to replace the entire detector assembly is at the end of this section. In most cases, it is not necessary to replace the entire detector assembly.
Replacing the Oxygen Sensor
1. Turn off the controller.
2. Turn off or unplug incoming power to the controller.
3. Unscrew the detector housing cap from the detector housing.
4. Unplug and remove the oxyge n sensor.
5. Carefully plug the replacement sensor into the socket pattern that is located in the top section of the detector housing.
NOTE: Match the sensor’s male pins with the two female sockets as you plug the sensor
into the socket.
6. Screw the detector housing cap onto the detector housing.
7. Turn on or plug in incoming power to the controller.
8. Turn on the controller.
CAUTION: Allow the replacement sensor to warm up for 5 minutes before you continue with the
next step.
9. Calibrate the replacement sensor as described in the Calibration section of this manual.
Replacing the Amplifier
1. Turn off the controller.
2. Turn off or unplug incoming power to the controller.
3. Remove the junction box cover.
4. Disconnect the detector leads from the detector terminal strip.
5. Unscrew and remove the two screws that secure the amplifier to the junction box. The screws are at the top left and bottom right of the amplifier.
6. Remove the amplifier.
7. Place the new amplifier in the same position as the old amplifier.
8. Use the two screws you removed in step 5 to secure the new amplifier to the junction box.
9. Verify that the jumper block is installed over the OXYGEN selector of the amplifie r type selector as shown in Figure 3.
10 • 65-2516RK Oxygen Transmit ter
Page 15
10. Reconnect the wiring from the controller to the interconnect terminal strip as shown in Table 3 and Figure 3.
Table 3: Reconnecting the Oxygen Amplifier to a Controller
Amplifier
Interconnect Terminal
Strip
Controller Transmitter
Terminal Strip (typical)
4/20 FB 4 -20 (FB)
24V + + V (11 - 30 VDC)
11. Reconnect the detector leads to the detector terminal strip as shown in Table 4 a nd Figure 3.
Table 4: Reconnecting the Oxygen Detector to the Amplifier
Amplifier
Oxygen Detector Lead
Interconnect Terminal
Strip
White OXY W Green OXY G
12. Reinstall the junction box cover.
13. Turn on or plug in incoming power to the controller.
14. Turn on the controller.
CAUTION: Allow the transmitter to warm up for 5 minutes before you continue with th e next
step.
15. Calibrate the oxygen transmitter as described in the Calibration section of this manual.
Replacing the Oxygen Detect or
NOTE: In most cases, it is only necessary to replace the oxygen sensor.
1. Turn off the controller.
2. Turn off or unplug incoming power to the controller.
3. Remove the junction box cover.
4. Disconnect the detector leads from the detector terminal strip. Note the position of the color-coded leads as you remove them.
5. Unscrew the detector from the junction box.
6. Guide the detector leads of the replacement detector through the bottom conduit hub of the junction box, then screw the mounting threads of the detector into the conduit hub.
65-2516RK Oxygen T ra n sm i tt er • 11
Page 16
7. Connect the detector leads to the detector terminal strip as shown in Table 4 and Figure 3.
Table 5: Connecting the Replacement Oxygen Detector to the Amplifier
Amplifier
Oxygen Detector Lead
White OXY W Green OXY G
8. Reinstall the junction box cover.
9. Turn on or plug in incoming power to the controller.
10. Turn on the controller.
CAUTION: Allow the replacement detector to warm up for 5 minutes before y o u c o ntinue with
the next step.
11. Calibrate the replacement detector as described in “Calibration” on page 13.
12. Secure the junction box cover to the junction box.
Interconnect Terminal
Strip
Calibration Frequency
Although there is no particular calibration frequency that is correct for all applications, a calibration frequency of every 3 to 6 months is adequate for mo st applications. Unless experience in a particular application dictates otherwise, RKI Instruments, Inc. recommends a calibration frequency of every 3 months (quarterly).
If an application is not very demanding, for example detection in a clean, temperature controlled environment, and calibration adjustments a re minima l at calibration, then a calibration frequency of every 6 months is adequate.
If an application is very demanding, for example if the environment is not well controlled, then more frequent calibration than every 3 months may be necessary.
12 • 65-2516RK Oxygen Transmit ter
Page 17
Calibration
This section describes how to calibrate the oxygen transmitter. It includes procedures to prepare for calibration, set the fr esh air r eading, set the zero r eading, and return to normal operation. It describes the test using a calibration kit that includes a calibration cup, calibration gas, sample tubing, and a fixed flow regulator with an on/off knob. RKI Instruments, Inc. recommends using a 0.5 LPM (liters per minute) fixed flow regulator.
Prepari ng for Ca libration
NOTE: Calibrating the oxygen transmitter may cause alarms. Be sure to put the
controller into its calibration program or disable external alarms before continuing.
1. Unscrew and remove the junction box cover.
2. Set a voltmeter to measure in the millivolt (mV) range.
WARN ING: Do not remove the detector housing cap or junction box cover while the
circuits are energized unless the area is deter mined t o be non-hazardous. Keep the detector housing cap and junction box cover tightly closed during operation.
3. Plug the positive lead into the test point labeled TP+; plug the negative lead into the test point labeled TP-.
4. Screw the calibration cup onto the detector housing.
Setting the Fresh Air Reading
NOTE: If you can verify that the oxygen transmitter is in a fres h air envir onment, you do
not need to apply zero air to the detector before adjusting the zero reading.
1. Screw the regulator into the zero air calibration cylinder.
2. Use the sample tubing to connect the regulator to the calibration cup.
3. Turn the regulator’s on/off knob counterclockwise to open it.
4. Allow the gas to flow for 2 minutes.
5. Verify a voltmeter reading of 434 mV (±2 mV).
6. If necessary, use a flat-bla de screwdriver to adjust the span pot on the amplifier unt il the voltmeter reading is 434 mV (±2 mV).
7. Turn the regulator on/off knob clockwise to close it.
8. Unscrew the regulator from the zero air calibration cylinder. Leave the sample tubing connected to the regulator and the calibration cup.
Setting the Zero Reading
1. Screw the regulator into the 100% nitrogen calibration cylinder.
2. Turn the regulator’s on/off knob counterclockwise to open it.
3. Allow the gas to flow for two minutes and verify a reading of 100 mV (±2 mV).
4. If necessary, use the zero pot on the amplifier to adjust the reading to 100 mV (±2 mV).
65-2516RK Oxygen T ra n sm i tt er • 13
Page 18
5. Turn the regulator’s on/off knob clockwise to close it.
6. Unscrew the regulator from the calibration cylind er.
7. Unscrew the calibration cup from the oxygen detector.
NOTE: For convenience, leave the components of the calibration kit connected by the
sample tubing.
Returning to Normal Operation
1. Remove the voltmeter leads from the am plif ier test points.
2. Secure the junction box cover to the junction box.
3. When the display reading rises above the decreasing alarm setpoint, return the controller to normal operation.
NOTE: If you do not allow the oxygen reading to increase above the decreasing alarm
point, then unwanted alarms may occur.
4. Verify that the display reading increases and stabilizes at 20.9% volume.
5. Store the components of the calibration kit in a safe and convenient place.
14 • 65-2516RK Oxygen Transmit ter
Page 19
Parts List
Table 5 lists replacement parts and accessories for the oxygen transmitter.
Table 6: Parts List
Part Number Description
06-1248RK Sample tubing (order by the foot) 07-0033RK Detector housing cap gasket 18-0405RK-01 Junction box (without cover; pre drilled for amplifier) 18-0406RK Junction box cover 57-1062RK Amplifier for oxy gen tra nsmitter 65-1025RK Oxygen replacement sensor, plug-in 65-2514RK Oxygen replacement detector assembly (includes sensor), CSA clas-
sified
65-2516RK Oxygen transmitter (includes detector and amplifier), CSA classi-
fied 71-0116RK Oxygen Transmitter Operator’s Manual (this document) 81-F301RK-LV Calibration kit, 34 liter 81-0076RK-01 Zero air calibration cylinder (34 liter) 81-0078RK Calibration cylinder (100% nitrogen, 17-liter) 81-0078RK-01 Calibration cylinder (100% nitroge n ; 34-liter) 81-1050RK Regulator with gauge and knob, 0.5 LPM, for 17- and 34-liter steel
81-1117RK Calibration cup
calibration cylinders
65-2516RK Oxygen T ra n sm i tt er • 15
Loading...