RKI Instruments 65-2511RK, 65-2512RK Operating Manual

Page 1
65-2511RK/65-2512RK
Oxygen Detector
Operator’s Manual
www.rkiinstruments.com
Page 2
Product W arranty
RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge. This warranty does not apply to those items which by their nature are subject to deterioration or consumption in normal service, and which must be cleaned, repaired, or replaced on a routine basis. Examples of such items are:
a) Absorbent cartridges d) Batteries b) Pump diaphragms and valves e) Filter elements c) Fuses
Warranty is voided by abuse including mechanical damage, alteration, rough handling, or repair procedures not in accordance with the operator’s manual. This warranty indicates the full extent of our liability, and we are not responsible for removal or replacement costs, local repair costs, transportation costs, or contingent expenses incurred without our prior approval.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI INSTRUMENTS, INC., INCLUDING BUT NOT LIMITED TO, THE WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL RKI INSTRUMENTS, INC., BE LIABLE FOR INDIRECT, INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
This warranty covers instruments and parts sold to users by authorized distributors, dealers, and representatives as appointed by RKI Instruments, Inc.
We do not assume indemnication for any accident or damage caused by the operation of this gas monitor, and our warranty is limited to the replacement of parts or our complete goods.
65-2511RK/65-2512RK Oxygen Detector• 2
Page 3
Table of Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Specications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Oxygen Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Junction Box. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Mounting the Oxygen Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Wiring the Oxygen Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Startup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Introducing Incoming Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Setting the Fresh Air Reading. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Replacing the Oxygen Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Setting the Fresh Air Reading. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Setting the Zero Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
65-2502RK/65-2510RK Oxygen Detector• 3
Page 4
Overview
This manual describes the 65-2512RK oxygen detector . This manual also describes how to install, start up, maintain, and calibrate the detector when used with a gas monitoring controller. A parts list at the end of this manual lists replacement parts and accessories for the oxygen detector.
The 65-2512RK oxygen detector includes a junction box. This manual may also be used for the 65-2511RK oxygen detector which does not include a junction box and is normally mounted in one of a controller’s conduit hubs. If you are using a 65-2511RK oxygen detector, disregard all references to the junction box and junction box terminal strip.
Specications
Table 1 lists specications for the oxygen detector.
Table 1: 65-2511RK/65-2512RK Specications
Target Gas Oxygen Sampling Method Diffusion Detection Range 0 - 25.0% oxygen Response Time 90% in 30 seconds Operating Temperature -20° C to 45° C Output Millivolt (mV)
65-2511RK/65-2512RK Oxygen Detector• 4
Page 5
Description
This section describes the oxygen detector and the junction box.
Oxygen Detector
The detector’s sensing element along with signal conditioning components are encapsulated within a conduit mounting black anodized aluminum housing. The sensing element used is a capillary type that is not susceptible to output changes with changes in atmospheric pressure.Through a series of chemical and electronic reactions, the detector produces a millivolt output that is proportional to the detection range. 3/4” NPT mounting threads at the top of the detector allow you to mount the detector to the junction box or a 3/4” NPT conduit tting. Two color-coded leads extend from the top of the detector. The leads allow you to connect the detector to a controller.
Junction Box
The junction box allows you to install the oxygen detector at a mounting site that is remote from a controller, and it protects the detector wiring connections. Two conduit hubs allow you to mount the oxygen detector to the junction box and connect the wiring from the detector to a controller. Three spacers installed on the back of the junction box control the distance of the junction box from a mounting surface and insure that there is enough room to install a calibration cup on the detector during calibration.
Installation
A terminal block within the junction box facilitates the wiring process. A cover on the front of the junction box allows access to the interior of the junction box.
This section describes procedures to mount the oxygen detector in the monitoring environment and wire the detector to a controller.
Mounting the Oxygen Detector
NOTE: If you are mounting a 65-2511RK, it does not include a junction box and is
usually factory installed in one of a controller’s conduit hubs or may be eld installed using the 3/4” NPT threads on the end with the wires. The 65-2512RK includes a junction box as shown in Figure 1 below.
1. Select a mounting site that is representative of the monitoring environment. Consider the following when you select the mounting site.
Select a site where the detector is not likely to be bumped or disturbed. Make sure
there is sufcient room to perform start-up, maintenance, and calibration procedures.
Select a site that is at normal breathing level.
65-2502RK/65-2510RK Oxygen Detector• 5
Page 6
3.6
Threads For Calibration Cup Installation
2.7 .75
.38
6.7 MAX
Spacer, 3X
Figure 1: Mounting the Oxygen Detector
2. At the mounting site you select, hang or mount the junction box with the detector
facing down (see Figure 1).
Wiring the Oxygen Detector to a Controller
WARNING: Always verify that the power to the controller is off before you make
wiring connections.
1. Turn off the controller.
2. Turn off or unplug incoming power.
3. Remove the cover from the junction box. If the detector is already installed in the junction box, go to step 6.
4. Guide the detector leads through the bottom conduit hub of the junction box, then screw the mounting threads of the detector into the conduit hub.
5. Connect the detector leads to the terminal block in the junction box.
6. Guide a two-conductor, shielded cable or two wires in conduit through the top conduit hub of the junction box. Use appropriate conduit ttings.
7. Connect the wires to the terminals opposite the detector leads.
65-2511RK/65-2512RK Oxygen Detector• 6
Page 7
CAUTION: Leave the shield drain wire insulated and disconnected at the detector. You will
connect the opposite end of the cable’s drain wire at the controller.
8. Secure the junction box cover to the junction box.
9. Route the cable or wires in conduit leading from the detector through one of the conduit hubs at the controller. Use appropriate conduit ttings.
10. Connect the wires to the appropriate controller detector terminal strip. See the controller operator’s manual for wiring instructions.
CAUTION: Do not route power and detector wiring through the same conduit hub. The power
cable may disrupt the transmission of the detector signal to the controller.
Green
(-)
Controller Oxygen Detector Terminals
(+)
White
J-Box
Terminal Strip
White
Detector Wires
Green
Oxygen Detector
Figure 2: Wiring the Oxygen Detector to a Controller
65-2502RK/65-2510RK Oxygen Detector• 7
Page 8
Start Up
This section describes procedures to start up the oxygen detector and place the detector into normal operation.
Introducing Incoming Power
1. Complete the installation procedures described earlier in this manual.
2. Verify that the power wiring to the controller is correct and secure. Refer to the controller instruction manual.
3. Turn on or plug in the incoming power, then turn on the controller.
4. Verify that the controller is on and operating properly. Refer to the controller operator’s manual.
CAUTION: Allow the detector to warm up for 5 minutes before you continue with the next
section, “Setting the Fresh Air Reading.”
Setting the Fresh Air Reading
CAUTION: If you suspect that the monitoring environment is not a fresh air environment, use
the zero air calibration cylinder to introduce “fresh air” to the detector and verify an accurate fresh air reading.
1. Verify that the detector is in a fresh air environment (environment known to be of normal oxygen content and free of toxic and combustible gasses).
2. Verify a reading of 20.9% oxygen at the controller. If the display reading is 20.9% oxygen, start up is complete. The oxygen detector is in
normal operation. If the display reading is not 20.9% oxygen, continue with step 3.
3. Perform a fresh air adjustment operation at the controller. See the controller operator’s manual for instructions.
65-2511RK/65-2512RK Oxygen Detector• 8
Page 9
Maintenance
This section describes maintenance procedures. It includes preventive maintenance, troubleshooting, and component replacement procedures.
Preventive Maintenance
This section describes a preventive maintenance schedule to ensure the optimum performance of the oxygen detector. It includes daily, monthly, and quarterly procedures.
Daily
Verify a display reading of 20.9% oxygen at the controller. Investigate signicant changes in the reading.
Monthly
This procedure describes a test to verify that the oxygen detector responds properly to oxygen deciency.
WARNING: The controller is not an active gas monitoring device during the response
test procedure.
NOTE: Performing a response test on the oxygen detector may cause alarms. Be sure to
put the controller into its calibration program or disable external alarms before performing this test
1. Place the controller into its calibration program or disable external alarms.
2. Verify that the controller display reading is 20.9% oxygen. If the controller reading is not 20.9% oxygen, set the fresh air reading, then continue
this procedure. See the controller operator’s manual for instructions to set the fresh air reading.
3. Exhale into the bottom of the oxygen detector housing.
4. Stop exhaling into the bottom of the detector, then verify that the reading on the controller display decreased from the normal reading of 20.9% oxygen.
NOTE: If the reading does not decrease, calibrate the detector as described in the
Calibration section of this Manual.
5. When the display reading rises above the decreasing alarm setpoint, return the controller to normal operation.
Quarterly
Calibrate the detector as described in the Calibration section of this manual. See the calibration frequency discussion in the Calibration section to determine if a quarterly calibration schedule ts your needs.
65-2502RK/65-2510RK Oxygen Detector• 9
Page 10
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended action for problems you may encounter with the oxygen detector.
NOTE: This troubleshooting guide describes detector problems only. See the controller
operator’s manual for problems you may encounter with the controller.
Table 2:Troubleshooting the Oxygen Detector
Condition Symptom(s) Probable Causes Recommended Action
Fail Condition • Controller indicates a
Slow or No Response/ Difcult or Unable to Calibrate
fail condition.
• Detector responds slowly or does not respond to response test.
• Unable to accurately set the fresh air or zero reading during calibration.
• Detector requires frequent calibration.
Note: Under “normal”
circumstances, the detector requires calibration once every three months.
Some applications may require a more frequent calibration schedule.
Replacing the Oxygen Detector
• The detector wiring is disconnected or misconnected.
• The detector is malfunctioning.
• The calibration cylinder is low, out-dated, or defective.
• The detector is malfunctioning.
1. Verify that the detector wiring is correct and secure.
2. Calibrate the detector.
3. If the fail condition continues, replace the detector.
4. If the fail condition continues, contact RKI for further instruction.
1. Verify that the calibration cylinder contains an adequate supply of a fresh test sample.
2. If the calibration/response difculties continue, replace the detector.
3. If the calibration/response difculties continue, contact RKI for further instruction.
1. Turn off the controller.
2. Turn off power to the controller.
3. Remove the junction box cover.
4. Disconnect the detector leads from the terminal block inside the junction box. Note the position of the color-coded leads as you remove them.
5. Unscrew the detector from the junction box.
6. Guide the detector leads of the replacement detector through the bottom conduit hub of the junction box, then screw the mounting threads of the detector into the conduit hub and tighten rmly.
7. Connect the detector leads to the terminal block in the same position as the leads you removed in step 4.
8. Secure the junction box cover to the junction box.
9. Turn on power to the controller.
65-2511RK/65-2512RK Oxygen Detector• 10
Page 11
Calibration
10. Turn on the controller.
11. Calibrate the replacement detector as described in the Calibration section of this manual.
This section describes how to calibrate the oxygen detector. It includes procedures to assemble the calibration kit, set the fresh air reading, set the zero reading and return to normal operation. It describes calibration using a xed ow regulator which has no on/ off knob and allows sample to ow as soon as it is screwed into a cylinder. RKI Instruments, Inc. recommends using a 0.5 LPM (liters per minute) xed ow regulator.
Calibration Frequency
Although there is no particular calibration frequency that is correct for all applications, a calibration frequency of every 3 to 6 months is adequate for most applications. Unless experience in a particular application dictates otherwise, RKI Instruments, Inc. recommends a calibration frequency of every 3 months (quarterly).
If an application is not very demanding, for example detection in a clean, temperature controlled environment, and calibration adjustments are minimal at calibration, then a calibration frequency of every 6 months is adequate.
If an application is very demanding, for example if the environment is not well controlled, then more frequent calibration than every 3 months may be necessary.
WARNING: The controller is not an active gas monitoring device during the calibration
procedure.
Setting the Fresh Air Reading
CAUTION: If you suspect the monitoring environment is not of normal oxygen content, 20.9%,
use the calibration kit and a zero air calibration cylinder to introduce “fresh air” to the detector and verify an accurate fresh air setting.
1. Place the controller into its calibration program or disable external alarms.
NOTE: Calibrating the oxygen detector may cause alarms. Be sure to put the controller
into its calibration program or disable external alarms before continuing.
2. Verify that the detector is in a fresh air environment.
3. Follow the directions in the controller’s instruction manual for setting the fresh air reading to 20.9% oxygen.
Setting the Zero Reading
1. Screw the calibration cup onto the bottom of the oxygen detector.
2. Use the sample tubing to connect the regulator to the calibration cup.
NOTE : Do not screw the regulator into the calibration cylinder at this time.
65-2502RK/65-2510RK Oxygen Detector• 11
Page 12
3. Follow the instructions in the controller’s operator’s manual for setting the oxygen zero reading.
4. When the instructions call for exposing the detector to gas, normally 100% nitrogen, screw the regulator into the cylinder and allow the gas to ow to the detector for 2 minutes before continuing with the instructions. The detector signal should be stable after two minutes.
5. After setting the zero reading, unscrew the regulator from the cylinder and remove the calibration cup from the detector.
NOTE: For convenience, leave the components of the calibration kit connected by the
sample tubing.
6. When the display reading rises above the decreasing alarm setpoint, return the controller to normal operation.
NOTE : If you do not allow the oxygen reading to increase above the decreasing alarm
point, then unwanted alarms may occur.
7. Verify that the controller display reading increases and stabilizes at 20.9% oxygen.
8. Store the components of the calibration kit in a safe and convenient place.
Parts List
Table 3 lists replacement parts and accessories for the oxygen detector.
Table 3: Parts List
Part Number Description
06-1248RK Sample tubing (3/16 in. x 5/16 in.; specify length when ordering) 18-0400RK-01 Junction box with spacers 65-2511RK Oxygen detector, conduit-mounting, capillary type, not including
junction box 65-2512RK Oxygen detector including junction box. 71-0109RK
81-F301RK-LV Calibration kit (34 liter) 81-0076RK-01 Zero air calibration cylinder (34 liter) 81-0078RK Calibration cylinder (100% nitrogen, 17-liter) 81-0078RK-01 Calibration cylinder (100% nitrogen; 34-liter)
65-2511RK/65-2512RK Oxygen Detector Operator’s Manual (this
document)
81-1003RK Regulator, 0.5 liter/minute; continuous ow
81-1117RK Calibration cup
65-2511RK/65-2512RK Oxygen Detector• 12
(for 17- and 34-liter calibration cylinders)
Loading...