RKI Instruments 65-2496RK User manual

Page 1
65-2496RK/65-2499RK
Carbon Monoxide Detector
Operator’s Manual
Part Number: 71-0156RK
Revision: 0
Released: 2/16/11
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 2
WARNING
Read and understand this instruction manual before operating detector . Improper use of the dete ctor could result in bodily harm or death.
Periodic calibration and maintenance of the detector is essential for proper operation and correct readings. Please calibrate and maintain this detector regularly! Frequency of calibration depends upon the type of use you have and the sensor types. T ypical calibration frequencies for most applications are b etween 3 and 6 months, but can be required more often or less often based on your usage.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 3
Product Warranty
RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year fr o m date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge. This warranty does not apply to those items which by their nature are subject to deterioration or consumption in no r mal service, and which must be cleaned, repaired, or replaced on a routine basis. Examples of such items are:
W arranty is voided by abuse including mechanical damage, alteration, rough handling, or repair procedures not in accordance with the operator’s manual. This warranty indicates the full extent of our liability , a nd we are not responsible for r emoval or replacement costs, local repair costs, transportation costs, or contingent expenses incurred without our prior approval.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI INSTRUMENTS, INC. INCLUDING BUT NOT LIMITED TO, THE WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL RKI INSTRUMENTS, INC. BE LIABLE FOR INDIRECT, INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
a) Absorbent cartridges d) Batteries b) Pump diaphragms and valves e) Filter elements c) Fuses
This warranty covers instruments and parts sold to users by authorized distributors, dealers, and representatives as appointed by RKI Instruments, Inc.
We do not as su m e i ndemnificatio n fo r any accident o r damage caused by the operat io n of this gas monitor, and our warranty is limited to the replacement of parts or our complete goods.
65-2496RK/65-2499RK CO Detector
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 4
Table of Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
65-2496RK CO Detector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Junction Box. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Mounting the CO Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Wiring the CO Detector to a Controller. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Startup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Introducing Incoming Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Setting the Zero (Fresh Air) Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Replacing Components of the CO Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Calibration Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Preparing for Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Setting the Zero (Fresh Air) Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Setting the Response Reading (Span) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Returning to Normal Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 5
Overview
This manual describes the 65-2499RK carbon monoxide (CO) detector. This manual also describes how to install, start up, maintain, and calibrate the CO detector when used with a gas monitoring controller. A parts list at the end of this manual lists replacement parts and accessories for the CO detector.
The 65-2499RK CO detector includes the 65-2496K CO detector and a junction box. This manual may also be used for the 65-2496RK CO detector which does not include a junction box and is norma lly mounted in one of a controller’s conduit hubs. If you are using a 65-2496RK CO detector, disregard all references to the junction box and junction box terminal strip.
Specifications
WARNING: Do not use this product in a manner not specified in this instruction
manual.
Table 1 lists specifications for the 65-2496RK and 65-2499RK CO detectors.
Table 1: Specifications
Target Gas Carbon Monoxide (CO) Sampling Method Diffusion Detection Range 0 to 300 PPM (parts per million) Response Time 90% in 30 seconds Accuracy ± 5% of reading or ± 5 ppm CO (whichever is greater) Operating Temperature -4°F to 113°F (-20°C to 45°C)
WARNING: When using the 65-2496RK/65-2499RK, you must follow the instructions
and warnings in this manual to assure proper and safe operation of the 65-2496RK/65-2499RK and to minimize the risk of personal injury. Be sure to maintain and periodically calibrate the 65-2496RK/65-2499RK as described in this manual.
65-2496RK/65-2499RK CO De t ec t or • 1
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 6
Description
This section describes the components of the 65-2496RK and 65-2499RK detectors. The 65­2499RK includes the 65-2496RK CO detector and a j u nction box. A two point terminal strip is provided inside the junction box for detector connections. The 65-2496RK does not include a junction box. Figure 1 below shows the components of the 65-2499RK.
3/4 NPT Conduit Hub
Junction Box
Rubber Spacer, 3X
CO Detector
Not e: There is a two point
ter m inal strip in si d e th e junction bo x for detector connections.
Figure 1: 65-2499RK Component Location
65-2496RK CO Detector
The 65-2496RK CO detector consists of the detector housing body, detector housing cap, cap gasket, plug-in sensor, and the charcoal filter with rubber boot.
Detector Housing Body
Cap Gasket
CO Plug-in Sensor
Charcoal Filter w/Rubber Boot
Detector Housing Cap
Hydrophobic Membrane
Figure 2: 65-2496RK CO Detector Component Location
2 • 65-2496RK/65-2499RK CO Det ector
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 7
Detector Housing Body
The detector housing body protects the electronic components within the housing. Use the mounting threads at the top of the housing to screw the CO detector into a 3/4” NPT hub. Two wires extend from the top of the detector housing body. Use these wires to connect the detector to a controller. One of the wires is black and one of the wires is red.
The housing body includes a four-socket pattern at the bottom of the housing body. The plug-in sensor mates to this socket pattern. A pre-amplifier located between the sockets and the two wires conditions the sensor’s signal before the signal reaches the controller.
Housing Cap & Cap Gasket
The housing cap screws onto the detector housing. It retains the sensor and the rubber boot with the charcoal filter and protects them from damage. A hydrophobic membrane on the outside of the cap face keeps water and particulates away from the charcoal filter and sensor face behind the cap. Unscrew the detector cap to access the charcoal filter and sensor for maintenance or replacement. A cap gasket seals the interface between the housing and cap.
Plug-In Sensor
The plug-in sensor is secured in the detector assembly by the housing ca p. Through a series of chemical and electrical reactions, the sensor produces an electrical output that corresponds to the detection range of the detector.
Charcoal Filter
The disc-shaped charcoal filter is secured to face of the CO sensor with a rubber boot. The charcoal filter prevents interference gases such as hydrogen sulfide (H
hydrocarbons from producing false CO readings.
S) and certain
2
Junction Box
The junction box allows you to install the CO detector at a mounting site that is remote from a controller and protects the detector wiring conn ections. Two conduit hubs allow you to mount the CO detector to the junction box and connect the wiring from the detector to a controller. Three spacers installed on the back of the junction box control the distance of the junction box from a mounting surface and ensure that there is enough room to install a calibration cup on the detector during calibration. A terminal blo ck within the junction box facilitates the wiring connections. A cover on the front of the junction box allows access to the interior of the junction box.
65-2496RK/65-2499RK CO De t ec t or • 3
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 8
Installation
This section describes procedures to mount the CO detector in the monitoring environment and wire the detector to a controller.
Mounting the CO Detector
NOTE: If you are installing a 65-2496RK, it does not include a junction box and is usually
factory installed in one of a controller’s conduit hubs or may be field installed using the 3/4” NPT threads on the end with the wires. The 65-2499RK includes a junction box as shown in Figure 3 below.
1. Select a mounting site that is representative of the monitoring environment. Consider the following when you select the mounting site.
Select a site where the detector is not likely to be bumped or disturbed. Make sur e
there is sufficient room to perform start-up, maintenance, and calibration procedures.
Select a site where the target gas is likely to be found first.
NOTE: If your application does not require a specific mounting site, mount the detector
at approximately breathing level.
+
7.30 .25
.75
1.32
1 1/2-20 Thread for
65-2496RK
3/4 NPT Conduit Hub
3/4" NPT
3.65
6.80 Max
Ø 1.75
Calibration Cup
1 1/2-20 T hread fo r
Calibration Cup
Junction
Box
65-2499RK
Figure 3: Outline & Mounting Dimensions, 65-2496RK & 65-2499RK
2.70
.75
.38
Rubber Spacer, 3X
CO Detector
2. At the mounting site you select, hang or mount the junction box wit h the detector facing down (see Figure 3).
4 • 65-2496RK/65-2499RK CO Det ector
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 9
Wiring the CO Detector to a Controller
WARNING: Always verify that power to the controller is OFF before you make wiring
connections.
1. Turn off the controller.
2. Turn off or unplug power to the controller.
3. If the detector is mou nted remotely from a controller using the ju nction box, proceed to step 4.
If the detector is mounted directly to a controller, it is normally factory wired. Confirm that the detector’s black and red wires are connected to the appropriate controller detector terminals and skip to “Start Up” on page 7. See Figure 4, the controller operator’s manual, and the controller’s detector head specification sheet for the 65­2496RK detector for the wiring connections.
4. Remove the junction box cover.
5. Guide a two-conductor, shielded cable or two wires in conduit through the unused conduit hub of the junction box. Use appropriate conduit fittings and construction technique for the environmental rating of the junction box. The junction box is rated NEMA 4X.
6. Connect the two wires to the detector using the terminal block.
CAUTION: If using shielded cable, leave the drain wire insulated and disconnected at the
detector. You will connect the opposite end of the cable’s drain wire at the controller.
7. Secure the junction box cover to the junction box.
8. Route the cable or wires leading from the CO detector through one of the conduit hubs at the controller housing. Use appropriate conduit fittings and construction technique for the environmental rating of the controller. RKI controllers are typically rated NEMA 4X.
CAUTION: Do not route power and detector wiring through the same conduit hub. The po wer
cable may disrupt the transmission of the detector si gnal to the controller.
9. Connect the wires to the applicable controller terminal strip. See the controller operator’s manual and the controller’s detector head specification sheet for the 65­2499RK detector.
65-2496RK/65-2499RK CO De t ec t or • 5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 10
Controller Detector Terminals,
Ty pica l Designati ons
Controller Ho u s ing
+ S
Shielded Cable
Te rminal Strip
Black
CO Detector
Red
Figure 4: Wiring the CO Detector to a Controller
10. If using shielded cable, connect the cable’s drain wire to an available chassis ground at the controller. RKI controllers typically have a ground stud that is a convenient grounding location.
6 • 65-2496RK/65-2499RK CO Det ector
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 11
Start Up
This section describes procedures to start up the CO detector and place the detector into normal operation.
Introducing Incoming Power
1. Complete the installation procedures described earlier in this manual.
2. Verify that the power wiring to the controller is correct and secure. Refer to the controller operator’s manual.
3. Turn on or plug in the incoming power, then turn on the controller.
4. Verify that the controller is on and operating properly. Refer to the controller operator’s manu al.
CAUTION: Allow the detector to warm up for 5 minutes before you continue with the next
section, “Setting the Zero (Fresh Air) Reading”.
Setting the Zero (Fresh Air) Reading
CAUTION: If you suspect the presence of CO in the monitoring environment, use the calibration
kit and a zero air calibration cylinder to introduce “fresh air” to the CO dete ctor and verify an accurate zero setting. See “Calibration” on page 13 for instructions to introduce zero air to the CO detector.
1. Verify that the CO detector is in a fresh air environment (environment known to be free of toxic and combustible gases, and of normal oxygen content, 20.9%).
2. Verify a reading of 0 p pm at the controller. If the display reading is 0 ppm, start up is complete. The CO detector is in normal
operation. If the display reading is not 0 ppm, continue with step 3.
3. Perform a zero (fresh air) adjustment operation at the controller. See the controller operator’s manual fo r instructions.
65-2496RK/65-2499RK CO De t ec t or • 7
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 12
Maintenance
This section describes maintenance procedures. It includes preventive maintenance, troubleshooting, and component replacement procedures.
Preventive Maintenance
This section describes a preventive maintenance schedule to ensure the optimum performance of the CO detector. It includes daily, monthly, and quarterly procedures.
Daily
Verify a display reading of 0 ppm at the controller. Investigate significant changes in the display reading.
Monthly
This procedure describes a test to verify that the CO detector responds properly to the target gas. It describes the test using a calibration kit that includes a calibration cup, calibration gas, sample tubing, and a fixed flow regulator with an on/off knob. RKI Instruments, Inc. recommends using a 0.5 LPM (liters per minute) fixed flow regulator.
WARNING: Failure to use the recommended calibration gas flow rate will result in an
NOTE: Performing a response test on the CO detector may cause alarms. Be sure to put
inaccurate r e ad in g.
the controller into its calibration program or disable external alarms before performing this test.
Preparing for the response test
1. Place the controller into its calibration program or disable external alarms.
2. Verify that the controller display reading for the channel you are testing is 0 ppm. If the display reading is not 0 ppm, set the zero (fresh air) reading of the detector as
described in the Start Up section of this manual, then continue this procedure.
3. Screw the regulator into the calibration cylinder. Make sure the regulator is off. It is off when the on/off knob is turned all the way clockwise.
4. Screw the calibration cup onto the bottom of the CO detector.
5. Use the sample tubing to connect the regulator to the calibration cup.
Performing the response test
1. Turn the regulator’s on /off knob counterclockwise to open the regulator. Gas will begin to flow.
2. Allow the gas to flow for two minutes, then verify that the reading is within ± 20% of the cylinder gas concentration.
NOTE: If the reading is not within ± 20% of the correct response reading, calibrate the
detector as described in “Calibration” on page 13.
3. Turn the regulator’s on/off knob clockwise to close the regulator.
4. Unscrew the regulator from the calibration cylinder.
8 • 65-2496RK/65-2499RK CO Det ector
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 13
5. Unscrew the calibration cup from the CO detector. Make sure that you do not loosen the detector housing cap when you unscrew the calibration cup.
6. When the controller display reading falls below the alarm setpoints, return the controller to normal operation.
Quarterly
Calibrate the CO detector as described in “Calibration” on page 13.
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended action for problems you may encounter with the CO detector.
NOTE: This troubleshooting guide describes detector problems only. See the controller
operator’s manual for problems you may encounter with the controller.
Fail Condition
Symptoms
The controller indicates a fail condition.
Probable causes
The detector wiring is disconnected or misconnected.
The plug-in sensor is not properly plugged into the four-socket pattern in the detector housing body.
The detector’s zero reading is low enough to cause a fail condition.
The detector is malfunctioning.
Recommended action
Verify that the detector wiring is correct and secure.
Confirm that the plug-in sensor is installed properly.
Perform a zero (fresh air) adjustment. A full calibration is recommended.
If the fail condition continues, replace the plug-in sensor as described later in this section.
If the fail condition continues, contact RKI Instruments, Inc. for further instruction.
Slow or No Response/Difficult or Unable to Calibrate
Symptoms
The detector responds slowly or does not respond during the monthly response test.
Unable to accurately set the zero or response reading during the calibration procedure.
The detector requires frequent calibration.
NOTE: Under “normal” circumstances, the detector requires calibration once every three
months. Some applications may require a more frequent calibration schedule.See “Calibration Frequency” on page 13 for a di scussion of the calibration frequenc y.
65-2496RK/65-2499RK CO De t ec t or • 9
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 14
Probable causes
The plug-in sensor has been replaced and the shorting jumper has not been removed.
The calibration cylinder is low, out-dated, or defective.
The incorrect calibration cup or regulator is being used.
The membrane on the detector h ousing cap is blocked with dirt or some other particulate contamination .
The detector is malfunctioning.
Recommended action
1. Confirm that the shorting jumper on the plug-in sensor pins has been removed.
2. Ve r if y that the cal i brat i on cylind er con tai ns an ade qu at e suppl y of a fresh test sample.
3. Confirm that you are using the correct calibration cup and regulator. See “Parts List” on page 15 for the required calibration cup and regulator.
4. Check the face of the detector housing cap and remove any particulate contamination from the hydrophobic membrane if necessary. If the membrane appears saturated with contamination or damaged, replace the membrane as described in “Replacing the Hydrophobic Membrane” on page 12.
5. If the calibration/response difficulties continue, replace the plug-in sensor as described later in this section.
6. If t he calibration/response difficulties continue, contact RKI Instruments, Inc. for further instruction.
Unexplained Gas Readings
Symptom
The display indicates a CO reading when CO is known no t to be present.
Probable causes
The detector needs calibration.
The charcoal filter on the plug-in sensor is saturated and is no longer scrubbing out H
S or hydrocarbons tha t cause a response on the CO sensor.
2
Recommended action
1. Ca librate the detector as described in “Calibration” on page 13.
2. If the difficulties contin ue, replace the charcoal filter as described later in this section.
3. If the difficulties continue, replace the plug-in sensor as described later in this section.
4. If the difficulties continue, contact RKI Instruments, Inc. for further instruction.
Replacing Components of the CO Detector
This section includes a procedure to replace the plug-in CO sensor, one to replace the charcoal filter , one to replace the hydrophobic membrane, and one to replace the entire CO detector assembly. In most cases, it is not necessary to replace the entire detector assembly.
Replacing The Plug-in CO Sensor
CAUTION: The plug-in sensor contains electrolyte which is a dilute acid. Do not disassemble the
sensor when replacing it with a new one. If sensor electrolyte comes in contact with your skin, wash affected area thoroughly with soap and water.
1. Turn off the controller.
10 • 65-2496RK/65-2499RK CO Det ector
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 15
2. Turn off or unplug power to the controller.
3. Unscrew the detector housing cap from the detector housing body.
4. Unplu g an d remove the old CO sensor with the boot and charcoal filter attached.
5. Remove the rubber boot and charcoal filter from the old CO sensor.
6. Remove the replacement CO sensor from its packaging and remove the wire jumper that is shorting the WE and RE sensor pins. This wire jumper is installed on the sensor pins for shipment or storage but must be removed for the sensor to operate properly when installed in a detector.
WARNING: The CO sensor will not operate properly if the wire jumper is not removed.
7. Instal l the rubber boot with charcoal filter onto the replacement sensor’s face.
8. Ca refully plug the replacement senso r into the four-socket pattern that is located in the detector housing body.
WARNING: You must replace the plug-in sensor with the same type of sensor that is
installed. A detector cannot be converted from one type of detector to another by using a different plug-in sensor. For example, if you are replacing an CO sensor, you must replace it with a CO sensor.
9. Make sure the cap gasket is in place and screw the detector housing cap back onto the detector housing body.
10. Turn on power to the controller.
11. Turn on the controller.
CAUTION: Allow the replacement sensor to warm up for 5 minutes before you continue with the
next step.
12. Calibrate the detector as described in “Calibration” on page 13.
Replacing the Charcoal Filter
1. Turn off the controller.
2. Turn off or unplug power to the controller.
3. Unscrew the detector housing cap from the detector housing body.
4. Unplug and remove the CO sensor with the boot and charcoal filter attached.
5. Remove the rubber boot that secures the charcoal filter to the CO sensor.
6. Remove the charcoal filter from the rubber boot.
7. Place the replacement filter in the rubber boot in the same position as the filter you removed in the previous step.
8. Reinstall the rubber boot with charcoal filter on the CO sensor.
9. Ca refully plug the replacement senso r into the socket pattern that is located in the detector housing body.
10. Make sure the cap gasket is in place and screw the detector housing cap back onto the detector housing body.
11. Turn on power to the controller.
65-2496RK/65-2499RK CO De te c t or • 1 1
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 16
12. Turn on the controller.
Replacing the Hydrophobic Membrane
1. Turn off the controller.
2. Turn off or unplug incoming power to the controller.
3. Unscrew the detector housing cap from the detector housing body.
4. Gently pry up the edge of the white hydrophobic membrane with a small flat blade screwdriver or a similar tool.
5. Peel off the hydrophobic membrane. It may be necessary to clean off the detector housing cap face to remove any residue left from the adhesive backed membrane.
6. Install the new membrane in the recess on the face of the detector housing cap.
7. Make sure the cap gasket is in place and screw the detector housing cap back onto the detector housing body.
8. Turn on power to the controller.
9. Turn on the controller.
Replacing the CO Detector
NOTE: In most cases, it is only necessary to replace the plug-in sens or.
1. Turn off the controller.
2. Turn off or unplug incoming power to the controller.
3. If the detector is installed directly on a controller, open the controller door. If the detector is installed remotely from a controller in a junction box, remove the
junction box cover.
4. If the detector is installed directly on a controller, disconnect the detector leads from the detector terminal strip in the controller. Note the position of the color-coded leads as you remove them.
If the detector is installed remotely from a controller in a junction box, disconnect the detector leads from the terminal block in the junction box. Note the position of the color-coded leads as you remove them.
5. Unscrew the detector from the controller conduit hub or junction box conduit hub.
6. Guide the detector leads of the replacement detector through the controller conduit hub or junction box conduit hub, then screw the mounting threads of the detector into the hub. If necessary for environmental conditions, apply thread sealant or teflon tape to the hub and/or detector threads to seal them.
7. If the detector is installed directly on a controller, connect the detector leads to the appropriate detector terminal strip terminals. See Figure 4 on page 6 for wiring to a generic controller. See the controller operator’s manual and the controller’s detecto r head specification sheet for the 65-2496RK detector for wiring specific to your controller.
If the detector is installed remotely from a controller in a junction box, connect the detector leads to the terminal block the same way the old detector was wired (see Figure 4 on page 6). See the controller operator’s manual and the controller’s detector head specification sheet for the 65-2499RK detector to verify the connections to the controller are correct.
12 • 65-2496RK/65-2499RK CO Det ector
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 17
8. If the detector is installed remotely from a controller in a junction box, reinstall the junction box cover.
9. Turn on or plug in power to the controller.
10. Turn on the controller.
CAUTION: Allow the replacement detector to warm up for 5 minutes before you continue with
the next step.
11. Calibrate the replacement detector as described in “Calibration” on page 13.
Calibration Frequency
Although there is no particular calibration frequency that is correct for all applications, a calibration frequency of every 3 months is adequate for most CO detector applications. Unless experience in a particular application dictates oth erwise, RKI Instruments, Inc. recommends a calibration frequency of every 3 months for the CO detector.
If an application is not very demanding, for example detection in a clean, temperature controlled environment where CO is not normally present, and calibration adjustments are minimal at calibration, then a calibration frequency of every 6 months is adequate.
If an application is very demanding, for example if the environment is not well controlled or if CO is often present, then more frequent calibration than every 3 months may be necessary.
Calibration
This section describes how to calibrate the CO detector. It includes procedures to prepare for calibration, set the fresh air reading, set the response reading, and return to normal operation. It describes calibration using a calibration kit that includes a calibration cup, calibration gas, sample tubing, and a fixed flow regulator with an on/off knob. RKI Instruments, Inc. recommends using a 0.5 LPM (liters per minute) fixed flow regulator.
WARNING: Not using the recommended calibration cup and sample flowrate will
result in an inaccurate calibration. See“Parts List” on page 15 for the required calibration cup and regulator.
Prepari ng for Ca libra t ion
1. Screw the calibration cup onto the bottom of the CO detector.
2. S c rew the regulator into the zero air calibration cylinder. Make sure the regulator is off. It is off when the on/off knob is turned all the way clockwise.
3. Use the sample tubing to connect the regulator to the calibration cup.
NOTE If you can verify that the CO detector is in a fresh air environment, you do not
need to apply zero air to the detector before adjusting the fresh air reading.
4. Put the controller into its calibration program. See the controller operator’s manual for instructions to enter the calibration program.
65-2496RK/65-2499RK CO De te c t or • 1 3
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 18
Setting the Zero (Fresh Air) Reading
1. Follow the instructions in the controller operator’s manual for setting the zero reading.
2. When the instructions call for applying zero air to the detector, turn the regulator’s on/off knob counterclockwise to open it. Gas will begin to flow.
3. Allow the gas to flow for two minutes.
4. Set the fresh air reading according to the controller operator’s manual.
5. Turn the regulator ’s on/off knob clockwise to close it.
6. Unscrew the regulator from the zero air calibration cylinder . Leave the sample tubing connected to the regulator and the calibration cup.
NOTE: Depending on the size of your zero air cylinder , it is possible that you will have a
different regulator for the zero air cylinder and the CO gas cylinder. If necessary to fit the CO calibration gas cylinder, change the regulator.
Setting the Response Reading (Sp an)
1. Screw the regulator into the calibration gas cylinder. Make sure the regulator is off. It is off when the on/off knob is turned all the way clockwise.
2. Follow the directions in the controller operator’s manual for setting the response (span) reading.
3. When the instructions call for exposing the detector to gas, turn the regulator’s on/off knob counterclockwise to open it. Gas will begin to flow.
4. Allow the gas to flow to the detector for 2 minutes.
5. S et the response reading according to the controller operator ’s manual.
6. After setting the response reading, turn the regulator’s on/off knob clockwise to turn it off.
7. Unscrew the regulator from the cylinder.
8. Unscrew the calibration cup from the CO detector. Make sure that you do not loosen the detector housing cap when you unscrew the calibration cup.
NOTE: For convenience, leave the regulator and calibration cup connected by the
sample tubing.
Returning to Normal Operation
1. Allow about 45 seconds for the gas reading to decrease below the alarm points and then return the controller to normal operation.
NOTE: If you do not allow the gas reading to decrease below the alarm points, then
unwanted alarms may occur.
2. Verify that the controller display reading decreases and stabilizes at 0 ppm.
3. Store the components of the calibration kit in a safe and convenient place.
14 • 65-2496RK/65-2499RK CO Det ector
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 19
Parts List
Table 2 lists replacement parts and accessories for the CO detector.
Table 2: Parts List
Part Number Description
06-1248RK-03 3 foot length of sample tubing for calibration kit 07-0039RK Detector housing cap gasket 07-0203RK Rubber retaining boot (for charcoal filter) 33-0157RK Hydrophobic membrane, adhesive backed, for detector housing cap 33-7101RK Charcoal Filter Disk 65-2496RK CO replacement detector assembly (includes plug-in sensor) 71-0156RK 65-2496RK/65-2499RK CO Detector Operator’s Manual (this document) 81-0064RK Calibration cylinder,50 PPM CO in air; 17 liter steel 81-0064RK-01 Calibration cylinder, 50 PPM CO in air; 34 liter steel 81-0064RK-03 Calibration cylinder, 50 PPM CO in air; 103 liter steel 81-0076RK Zero air calibration cylinder, 17 liter steel 81-0076RK-01 Zero air calibration cylinder, 34 liter steel 81-0076RK-03 Zero air calibration cyl i nd e r, 103 liter steel 81-1050RK Regulator with gauge and knob, 0.5 LPM, for 17 liter and 34 liter steel
calibration cylinders
81-1051RK Regulator with gauge and knob, 0.5 LPM, for 34AL/58/103 liter cali-
bration cylinders 81-1117RK Calibration c u p ES-1531-CO CO replacement sensor
65-2496RK/65-2499RK CO De te c t or • 1 5