RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in
materials, workmanship, and performance for a period of one year from date of shipment from RKI
Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our
option, free of charge. This warranty does not apply to those items which by their nature are subject
to deterioration or consumption in normal service, and which must be cleaned, repaired, or replaced
on a routine basis. Examples of such items are:
Warranty is voided by abuse including mechanical damage, alteration, rough handling, or repair
procedures not in accordance with the operator’s manual. This warranty indicates the full extent of
our liability, and we are not responsible for removal or replacement costs, local repair costs,
transportation costs, or contingent expenses incurred without our prior approval.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER
WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND
ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI
INSTRUMENTS, INC., INCLUDING BUT NOT LIMITED TO, THE WARRANTY
OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN
NO EVENT SHALL RKI INSTRUMENTS, INC., BE LIABLE FOR INDIRECT,
INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND
CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS
PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
a) Absorbent cartridgesd) Batteries
b) Pump diaphragms and valvese) Filter elements
c) Fuses
This warranty covers instruments and parts sold to users by authorized distributors, dealers, and
representatives as appointed by RKI Instruments, Inc.
We do not assume indemnification for any accident or damage caused by the operation of this gas
monitor, and our warranty is limited to the replacement of parts or our complete goods.
This manual describes the hydrogen compensated carbon monoxide (CO) transmitter. This manual
also describes how to install, start up, maintain, and calibrate the transmitter. A parts list at the end
of this manual lists replacement parts and accessories for the CO transmitter.
Specifications
Table 1 lists specifications for the CO transmitter.
Table 1: Specifications
Target GasCarbon Monoxide (CO)
Area ClassificationExplosionproof for Class I, Groups B, C, and D
Input Voltage19 - 30 VDC
Sampling MethodDiffusion
Signal Output4 to 20 mA
Detection Range0 to 300 PPM (parts per million)
Response Time90% in 30 seconds
4 • 65-2435RK CO Transmitter
Description
This section describes the components of the CO transmitter. The transmitter consists of the CO
detector, amplifier, and junction box.
Tes t Point (-)
Tes t Poin t (+)
Null
Pot enti om eter
Spa n Potentiom eter
Zero Potentiom eter
Carbon Mo nox i de D etector
TP+
NULL
4-20mATransmitter
57-1252RK
SPAN
ZERO
3/4" Conduit Hub
Terminal Strip
TP
FB
24V
PWR
BLK
GRN
BRN
WHT
SENSOR
RED
Carbon Monoxide
Am pli fi e r
Securing Screw (2)
Detector Hous i ng
Ca p Ga s k e t (x 2 )
Detector Hous i n g Cap
Flame Arrestor
Figure 1: CO Transmitter Component Location
CO Detector
The CO detector includes the detector housing and sensor.
Detector Housing
The detector housing protects the sensing components within the housing. Use the mounting
threads at the top of the housing to screw the CO detector into the bottom conduit hub of the
junction box. Use the removable cap near the bottom of the housing to access the sensor for
maintenance or replacement. The cap protects the sensor from damage and includes a flame
arrestor which contains any sparks which may occur within the detector housing. Two cap gaskets
seal the interface between the housing and cap.
Two wires extend from the top of the detector housing. Use these wires to connect the CO detector
to the amplifier. The housing includes a four-socket pattern. This socket pattern accepts the sensor’s
four pins to secure the sensor within the detector housing. A pre-amplifier, located between the
sockets and two interconnect wires, conditions the sensor’s signal before the signal reaches the
amplifier.
Sensor
The sensor is secured within the sensor housing by the four pins. Through a series of chemical and
electrical reactions, the sensor produces an electrical output that is proportional to the detector
range of the transmitter.
65-2435RK CO Transmit ter • 5
Charcoal Filter
The disc-shaped charcoal filter is secured to face of the CO sensor with a rubber boot. The charcoal
filter prevents interference gases (hydrogen sulfide [H
producing false CO readings.
S] and certain hydrocarbons) from
2
Amplifier
The amplifier converts the electrical output from the sensor to a 4 to 20 mA signal (that is
proportional to the detection range) and transmits the signal to a gas monitoring controller. The
amplifier includes the terminal strip, span potentiometer, zero potentiometer, and test points (see
Figure 1.)
Terminal Strip
The terminal strip is a seven-point terminal strip on the right side of the amplifier. Use the terminal
strip to connect the CO detector to the amplifier and the amplifier to a controller. The terminal strip
is a plug-in style which can be removed for wiring by pulling up on it.
NOTE: The CO detector is factory-wired to the amplifier. See the Installation section of this
manual for all wiring procedures related to the transmitter.
Span Potentiometer
The span potentiometer is near the bottom of the amplifier. It is to the left of the zero potentiometer.
Use the span potentiometer to adjust the transmitter’s response output during the calibration
procedure.
Zero Potentiometer
The zero potentiometer is to the right of the span potentiometer. Use the zero potentiometer to
adjust the transmitter’s target gas-free output during the start-up and calibration procedures.
Null Potentiometer
The null potentiometer is on the left side of the amplifier. It is used to adjust the transmitters
hydrogen compensation when a CO sensor is replaced.
CAUTION: The null potentiometer is factory set before shipment. It should only be adjusted in
the field when the CO sensor is replaced. This adjustment is made only once after
installing a new sensor. Do not adjust this pot when calibrating the CO transmitter.
Test Points
The test points (labeled TP- and TP+) are near the top of the amplifier. The test points produce a
100 to 500 mV output that is proportional to the transmitter’s 4 to 20 mA output. Use the test points
and a voltmeter to measure the transmitter’s output during the start-up and calibration procedures.
Junction Box
Use the junction box to install the CO transmitter at a mounting site that is remote from the
controller. The junction box also protects the amplifier and wiring connections made to the
amplifier. Use the two 3/4 in. conduit hubs to mount the detector to the junction box (bottom hub)
and connect wiring from the amplifier to the controller (top hub).
NOTE: The CO detector and amplifier are factory-mounted to the junction box.
Use the junction box’s two mounting holes to mount the CO transmitter to a vertical surface at the
6 • 65-2435RK CO Transmitter
Installation
monitoring site. Use the cover on the front of the junction box to access the interior of the junction
box.
This section describes procedures to mount the CO transmitter in the monitoring environment and
wire the transmitter to a controller.
Mounting the CO Transmitter
1.Select a mounting site that is representative of the monitoring environment. Consider the
following when you select the mounting site.
•Select a site where the transmitter is not likely to be bumped or disturbed. Make sure there
is sufficient room to perform start-up, maintenance, and calibration procedures.
•Select a site where the target gas is likely to be found first.
NOTE: If your application does not require a specific mounting site, mount the transmitter at
approximately breathing level.
6.1
1.1
3.94
5.2
2.75
5.46
Ø.25Mouting Holes (2X)
3/4 C on du it Hub
8.3 ma x
2.85 max
Figure 2: Mounting the CO Transmitter
If the CO detector is mounted to the junction box, skip to step 5. If not, continue with
step 2.
2.Remove the junction box cover.
3.Guide the two wires that extend from the top of the CO detector through the bottom conduit
hub of the junction box.
4.Screw the CO detector into the bottom conduit hub of the junction box.
5.At the monitoring site, use #10 screws through the junction box’s two mounting holes to secure
the junction box to a vertical surface.
CAUTION: Mount the CO transmitter with the detector facing down (see Figure 2.)
65-2435RK CO Transmit ter • 7
Wiring the CO Transmitter to a Controller
WARNING:Always verify that the power source is OFF before you make wiring connections.
1.Turn off the controller.
2.Turn off or unplug incoming power at the power source end.
3.Remove the junction box cover.
4.Verify that the detector leads are wired to the amplifier’s terminal strip. If necessary, connect
the detector leads to the terminal strip as shown in Figure 3.
5.Guide a two-conductor, shielded cable or two wires in conduit through the top conduit hub of
the junction box.
6.Connect the two wires to the terminal strip as follows (see Figure 3.)
•Connect the positive wire to the terminal labeled 24V.
•Connect the feedback wire to the terminal labeled FB.
CAUTION: If using shielded cable, leave the drain wire insulated and disconnected at the
transmitter. You will connect the opposite end of the cable’s drain wire at the
controller.
7.Secure the junction box cover to the junction box.
8.Route the cable or wires leading from the CO transmitter through one of the conduit hubs at the
controller housing.
CAUTION: Do not route power and transmitter wiring through the same conduit hub. The power
cable may disrupt the transmission of the transmitter signal to the controller.
8 • 65-2435RK CO Transmitter
9.Connect the wires to the to the applicable controller transmitter terminal strip as shown in
Figure 3.
CO Amplifier
TP+
NULL
Transmitter
4-20mA
57-1252RK
SPAN
ZERO
TP
FB
24V
BLK
GRN
BRN
WHT
RED
4 - 20 mA In (FB)
+24 VDC (19 - 30 VDC)
Controller or
Recording Device
RED
WHT
BRN
GRN
BLK
CO Detector/ H2 Compensated
Factory Wired
Figure 3: Wiring the CO Transmitter to a Controller
10. Connect the cable’s drain to an available chassis ground at the controller.
65-2435RK CO Transmit ter • 9
Start Up
This section describes procedures to start up the CO transmitter and place the transmitter into
normal operation.
Introducing Incoming Power
1.Complete the installation procedures described earlier in this manual.
2.Verify that the power wiring to the controller is correct and secure. Refer to the controller
instruction manual.
3.Turn on or plug in the incoming power at the power source end, then turn on the controller.
4.Verify that the controller is on and operating properly. Refer to the controller instruction
manual.
CAUTION: Allow the transmitter to warm up for 5 minutes before you continue with the next
section, “Setting the Zero Signal.”
Setting the Zero Signal
CAUTION: If you suspect the presence of the target gas in the monitoring environment, use the
calibration kit and the zero air calibration cylinder to introduce “fresh air” to the
sensor and verify an accurate zero setting.
1.Verify that the transmitter is in a fresh air environment (environment known to be free of
carbon monoxide, other toxic and combustible gases and of normal oxygen content, 20.9%).
2.Unscrew and remove the junction box cover from the junction box.
3.Set a voltmeter to measure in the millivolt (mV) range.
4.Plug the voltmeter leads into the test points on the amplifier. Plug the positive lead into the test
point labeled TP+; plug the negative lead into the test point labeled TP-.
5.Verify a voltmeter reading of 100 mV (±2 mV).
6.If necessary, use a flat-blade screwdriver to adjust the zero potentiometer until the voltmeter
reading is 100 mV (±2 mV).
7.Remove the voltmeter leads and secure the junction box cover to the junction box.
Operating With a Background of Hydrogen
The CO sensor in the CO transmitter does respond to hydrogen in addition to CO. The transmitter
has the ability to compensate for this response up to 10,000 ppm hydrogen. If up to 10,000 ppm of
hydrogen is present, the CO transmitter will indicate only the CO concentration.
10 • 65-2435RK CO Transmitter
Maintenance
This section describes maintenance procedures. It includes preventive maintenance,
troubleshooting, and component replacement procedures.
Preventive Maintenance
This section describes a preventive maintenance schedule to ensure the optimum performance of
the CO transmitter. It includes daily, monthly, and quarterly procedures.
Daily
1.Verify a display reading of 0 PPM CO at the controller. Investigate significant changes in the
Monthly
This procedure describes a test to verify that the CO transmitter responds properly to carbon
monoxide. It describes the test using a fixed flow regulator which has no on/off knob and allows
sample to flow as soon as it is screwed into a cylinder. RKI Instruments, Inc. recommends using a
0.5 LPM (liters per minute) fixed flow regulator.
NOTE: Performing a response test on the CO transmitter may cause alarms. Be sure to put the
display reading.
controller into its calibration program or disable external alarms before performing this
test.
Preparing for the response test
1.Place the controller into its calibration program or disable external alarms.
2.Verify that the controller display reading for the channel you are testing is 0.
If the display reading is not zero, set the zero reading of the transmitter as described in the Start
Up section of this manual, then continue this procedure.
3.Screw the calibration cup onto the bottom of the CO detector.
4.Use the sample tubing to connect the regulator to the calibration cup.
NOTE: Do not screw the regulator into the calibration cylinder at this time.
5.Set a voltmeter to measure in the millivolt (mV) range.
6.Remove the junction box cover, then plug the voltmeter leads into the test points on the
amplifier.
Plug the positive lead into the test point labeled TP+; plug the negative lead into the test point
labeled TP-.
7.Use the following formula to determine the correct test points output for the test sample.
Output (mV) = (calibrating sample/fullscale) X 400 + 100
For example, with a test sample of 50 PPM CO and a fullscale setting of
300 PPM, the correct output is 167 mV.
167 (mV) = (50/300) X 400 +100
65-2435RK CO Transmitter • 11
Performing the response test
1.Screw the regulator into the calibration cylinder. The sample will begin to flow.
2.Allow the gas to flow for two minutes, then verify that the reading is within ± 10% of the
response reading you determined earlier.
NOTE: If the reading is not within ± 10% of the correct response reading, calibrate the
transmitter as described in the Calibration section of this manual.
3.Unscrew the regulator from the calibration cylinder.
4.Unscrew the calibration cup from the CO detector.
5.Remove the voltmeter leads from the amplifier test points.
6.Reinstall the junction box cover.
7.When the controller display reading falls below the alarm setpoints, return the controller to
normal operation.
Quarterly
Calibrate the CO transmitter as described in the Calibration section of this manual.
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended action for
problems you may encounter with the CO transmitter.
NOTE: This troubleshooting guide describes transmitter problems only. See the controller
instruction manual for problems you may encounter with the controller.
Fail Condition
Symptoms
•The controller indicates a fail condition.
Probable causes
•The transmitter wiring is disconnected or misconnected.
•The transmitter’s zero reading is low enough to cause a fail condition.
•The transmitter is malfunctioning.
Recommended action
•Verify that the transmitter wiring is correct and secure.
•Calibrate the transmitter.
•If the fail condition continues, replace the CO sensor.
•If the fail condition continues, contact RKI for further instruction.
Slow or No Response/Difficult or Unable to Calibrate
Symptoms
•The transmitter responds slowly or does not respond during the monthly response test.
•Unable to accurately set the zero or response reading during the calibration procedure.
•The transmitter requires frequent calibration.
12 • 65-2435RK CO Transmitter
NOTE: Under “normal” circumstances, the transmitter requires calibration once every three
months. Some applications may require a more frequent calibration schedule.
Probable causes
•The calibration cylinder is low, out-dated, or defective.
•The transmitter is malfunctioning.
Recommended action
1.Verify that the calibration cylinder contains an adequate supply of a fresh test sample.
2.If the calibration/response difficulties continue, replace the CO sensor as described later in this
section.
3.If the calibration/response difficulties continue, contact RKI Instruments, Inc., for further
instruction.
Replacing Components of the CO Transmitter
This section includes procedures to replace the CO sensor and amplifier. A procedure to replace the
entire detector assembly is at the end of this section. In most cases, it is not necessary to replace the
entire detector assembly.
Replacing the Sensor
4.Turn off the controller.
5.Turn off or unplug incoming power at the power source end.
6.Unscrew the bottom section of the CO detector housing from the top section.
NOTE: Be sure the two cap gaskets remain seated and that they don’t fall out.
7.Unplug and remove the CO sensor with the boot and charcoal filter attached.
8.Remove the rubber boot and charcoal filter from old sensor.
9.Install the rubber boot with charcoal filter onto the replacement sensor’s face.
10. Carefully plug the replacement sensor into the socket pattern that is located in the top section
of the detector housing.
NOTE: Match the sensor’s male pins with the four female sockets as you plug the sensor into the
socket.
11. Screw the bottom section of the detector housing onto the top section.
12. Turn on or plug in incoming power at the power source end.
13. Turn on the controller.
CAUTION: Allow the replacement sensor to warm up for 5 minutes before you continue with the
next step.
14. Set the null potentiometer on the amplifier as described in the Setting the Null Potentiometer
section.
65-2435RK CO Transmitter • 13
CAUTION: The null potentiometer must be set whenever a sensor is changed or the hydrogen
compensation may not work properly.
15. Calibrate the replacement sensor as described in the Calibration section of this manual.
Replacing the Charcoal Filter
1.Turn off the controller.
2.Turn off or unplug incoming power at the power source end.
3.Unscrew the bottom section of the CO detector housing from the top section.
NOTE: Be sure the two cap gaskets remain seated and that they don’t fall out.
4.Unplug and remove the CO sensor with the boot and charcoal filter attached.
5.Remove the rubber boot that secures the charcoal filter to the CO sensor.
6.Remove the charcoal filter from the rubber boot.
7.Place the replacement filter in the rubber boot in the same position as the filter you removed in
the previous step.
8.Reinstall the rubber boot with charcoal filter to the CO sensor.
9.Carefully plug the replacement sensor into the socket pattern that is located in the top section
of the detector housing.
NOTE: Match the sensor’s male pins with the four female sockets as you plug the sensor into the
sockets.
10. Screw the bottom section of the detector housing onto the top section.
11. Turn on or plug in incoming power at the power source end.
12. Turn on the controller.
Replacing the Amplifier
1.Turn off the controller.
2.Turn off or unplug incoming power at the power source end.
3.Remove the junction box cover.
4.Disconnect the detector leads from the terminal strip.
5.Unscrew and remove the two screws that secure the amplifier to the junction box.
The screws are at the top left and bottom right of the amplifier.
6.Remove the amplifier.
7.Place the new amplifier in the same position as the old amplifier.
8.Use the two screws you removed in step 5 to secure the new amplifier to the junction box.
14 • 65-2435RK CO Transmitter
9.Reconnect the wiring from the controller to the interconnect terminal strip as shown in Table 2
and Figure 3, Wiring the CO Transmitter to a Controller.
Table 2: Reconnecting the CO Amplifier to a Controller
Amplifier Terminal Strip
Controller Transmitter
Terminal Strip (typical)
FB4 -20 (FB)
24V+ V (19 - 30 VDC)
10. Reconnect the detector leads to the detector terminal strip as shown in Table 3 and Figure 3,
Wiring the CO Transmitter to a Controller.
Table 3: Reconnecting the CO Detector to the Amplifier
CO Detector LeadAmplifier Terminal Strip
BlackBLK
GreenGRN
BrownBRN
WhiteWHT
RedRED
11. Reinstall the junction box cover.
12. Turn on or plug in incoming power at the power source end.
13. Turn on the controller.
CAUTION: Allow the sensor to warm up for 5 minutesbefore you continue with the next step.
14. Calibrate the CO transmitter as described in the Calibration section of this manual.
Replacing the CO Detector
NOTE: In most cases, it is only necessary to replace the CO sensor.
1.Turn off the controller.
2.Turn off or unplug incoming power at the power source end.
3.Remove the junction box cover.
4.Disconnect the detector leads from the detector terminal strip. Note the position of the colorcoded leads as you remove them.
5.Unscrew the detector from the junction box.
6.Guide the detector leads of the replacement detector through the bottom conduit hub of the
junction box, then screw the mounting threads of the detector into the conduit hub.
65-2435RK CO Transmitter • 15
7.Connect the detector leads to the detector terminal strip as shown in Table 4 and Figure 3,
Wiring the CO Transmitter to a Controller.
Table 4: Connecting the Replacement CO Detector to the Amplifier
CO Detector Lead
Interconnect Terminal Strip
Amplifier
BlackBLK
GreenGRN
BrownBRN
WhiteWHT
RedRED
8.Reinstall the junction box cover.
9.Turn on or plug in incoming power at the power source end.
10. Turn on the controller.
CAUTION: Allow the replacement detector to warm up for 5 minutes before you continue with
the next step.
11. Set the null potentiometer on the amplifier as described in the Setting the Null Potentiometer
section.
CAUTION: The null potentiometer must be set whenever the complete detector assembly is
changed or the hydrogen compensation may not work properly.
12. Calibrate the replacement detector as described in the Calibration section of this manual.
13. Secure the junction box cover to the junction box.
16 • 65-2435RK CO Transmitter
Setting the Null Potentiometer
When an old sensor is replaced, the null potentiometer must be adjusted so that the CO
transmitter’s hydrogen compensation will work properly with the new sensor. The procedure
basically involves applying hydrogen to the transmitter and using the null potentiometer to cancel
out the hydrogen response. This adjustment is made only once after the sensor is replaced and does
not need to be made again until the sensor is replaced again. A concentration of 4,000 ppm or 5,000
ppm hydrogen is recommended for setting the null. Perform the following procedure to adjust the
null pot after changing a sensor.
Preparing for the Null Adjustment
1.Screw the calibration cup onto the bottom of the CO detector.
2.Use the sample tubing to connect the fixed flow regulator to the calibration cup.
NOTE: Do not screw the regulator into the hydrogen cylinder at this time.
3.Set a voltmeter to measure in the millivolt (mV) range.
4.Remove the junction box cover, then plug the voltmeter leads into the test points on the
amplifier.
Plug the positive lead into the test point labeled TP+; plug the negative lead into the test point
labeled TP-.
NOTE: Adjusting the null potentiometer may cause alarms. Be sure to put the controller into its
calibration program or disable external alarms before continuing.
Adjusting the Null Potentiometer
1.Screw the regulator into the hydrogen calibration cylinder. Gas will automatically begin to
flow.
2.Allow the gas to flow for two minutes, then adjust the null potentiometer clockwise and
counterclockwise until you find the adjustment that results in the lowest mV reading at the voltmeter. This adjustment position will be somewhere in the middle of the adjustment range
and not at extreme clockwise or extreme counterclockwise.
3.Unscrew the regulator from the hydrogen cylinder. Wait two minutes before proceeding.
Leave the sample tubing connected to the regulator and the calibration cup.
Resetting the Zero Reading
The adjustment of the null potentiometer may result in a change in the zero reading, so the zero
reading may need to be adjusted.
NOTEIf you can verify that the CO transmitter is in a fresh air environment, you do not need to
apply zero air to the detector before adjusting the zero reading.
1.Screw the regulator into the zero air calibration cylinder. Gas will automatically begin to flow.
65-2435RK CO Transmitter • 17
2.Allow the gas to flow for two minutes, then verify a reading of 100 mV (± 2mV). If necessary,
use the zero potentiometer on the amplifier to adjust the reading to 100 mV (± 2mV).
3.Unscrew the regulator from the zero air calibration cylinder.
Returning to Normal Operation
1.Remove the voltmeter leads from the amplifier test points.
2.Unscrew the calibration cup from the detector.
NOTE: For convenience, leave the components of the calibration kit connected by the sample
tubing.
3.Secure the junction box cover to the junction box.
4.Return the controller to normal operation.
5.Verify that the controller display reading stabilizes at 0 ppm.
6.Store the components of the calibration kit in a safe and convenient place.
18 • 65-2435RK CO Transmitter
Calibration
This section describes how to calibrate the CO transmitter. It includes procedures to prepare for
calibration, set the zero reading, set the response reading, and return to normal operation. It
describes calibration using a fixed flow regulator which has no on/off knob and allows sample to
flow as soon as it is screwed into a cylinder. RKI Instruments, Inc. recommends using a 0.5 LPM
(liters per minute) fixed flow regulator.
CAUTION: Do not adjust the null potentiometer during calibration. This will cause the hydrogen
compensation in the transmitter to operate improperly.
Preparing for Calibration
1.Screw the calibration cup onto the bottom of the CO detector.
2.Use the sample tubing to connect the fixed flow regulator to the calibration cup.
NOTE: Do not screw the regulator into the zero air calibration cylinder at this time.
3.Set a voltmeter to measure in the millivolt (mV) range.
4.Remove the junction box cover, then plug the voltmeter leads into the test points on the
amplifier.
Plug the positive lead into the test point labeled TP+; plug the negative lead into the test point
labeled TP-.
5.Use the following formula to determine the correct test points output for the calibrating
sample.
Output (mV) = (calibrating sample/fullscale) X 400 + 100
For example, with a calibrating sample of 50 PPM CO and a fullscale setting of
300 PPM, the correct output is 167 mV.
167 (mV) = (50/300) X 400 +100
NOTE: Calibrating the CO transmitter may cause alarms. Be sure to put the controller into its
calibration program or disable external alarms before continuing.
Setting the Zero Reading
NOTEIf you can verify that the CO transmitter is in a fresh air environment, you do not need to
apply zero air to the detector before adjusting the zero reading.
1.Screw the regulator into the zero air calibration cylinder. Gas will automatically begin to flow.
2.Allow the gas to flow for two minutes, then verify a reading of 100 mV (± 2mV). If necessary,
use the zero potentiometer on the amplifier to adjust the reading to 100 mV (± 2mV).
3.Unscrew the regulator from the zero air calibration cylinder.
Leave the sample tubing connected to the regulator and the calibration cup.
Setting the Response Reading
1.Screw the regulator into the calibration cylinder. Gas will begin to flow.
65-2435RK CO Transmitter • 19
2.Allow the gas to flow for two minutes, then verify that the reading matches the response
reading (± 2mV) you determined earlier. If necessary, use the span potentiometer on the
amplifier to adjust the reading to match the correct response reading.
3.Unscrew the regulator from the calibration cylinder.
Returning to Normal Operation
1.Remove the voltmeter leads from the amplifier test points.
2.Unscrew the calibration cup from the detector.
NOTE: For convenience, leave the components of the calibration kit connected by the sample
tubing.
3.Secure the junction box cover to the junction box.
4.When the display reading falls below the alarm setpoints, return the controller to normal
operation.
5.Verify that the controller display reading decreases and stabilizes at 0 ppm.
6.Store the components of the calibration kit in a safe and convenient place.
20 • 65-2435RK CO Transmitter
Parts List
Table 5 lists replacement parts and accessories for the CO transmitter.
Table 5: Parts List
Part NumberDescription
06-1248RKSample tubing (order by the foot)
07-0033RKDetector housing cap gasket (2 needed per assembly)