Part Number: 71-0242RK
Revision: P1
Released: 9/12/11
RKI Instruments, Inc.
www.rkiinstruments.com
WARNING
Read and understand this instruction manual before operating
detector . Improper use of the dete ctor could result in bodily harm
or death.
Periodic calibration and maintenance of the detector is essential
for proper operation and correct readings. Please calibrate and
maintain this detector regularly! Frequency of calibration
depends upon the type of use you have and the sensor types.
T ypical calibration frequencies for most applications a re between
3 and 6 months, but can be required more often or less often
based on your usage.
65-2405RKSS Combustible Gas Transmitter
Product Warranty
RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in
materials, workmanship, and performance for a period of one year fr o m date of shipment
from RKI Instruments, Inc. Any parts found defective withi n tha t period will be repaired
or replaced, at our option, free of charge. This warranty does not apply to those items
which by their nature are subject to deterioration or consumption in normal ser v ice, and
which must be cleaned, repaired, or replaced on a routine basis. Examples of such items
are:
W arranty is voided by abuse including mechanical damage, alteration, rough handling, or
repair procedures not in accordance with the operator’s manual. This warranty indicates
the full extent of our liability , a nd we are not r esponsible for removal or r eplacement costs,
local repair costs, transportation costs, or contingent expenses incurred without our prior
approval.
a) Absorbent cartridgesd) Batteries
b) Pump diaphragms and valvese) Filter elements
c) Fuses
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER
WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED,
AND ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF
RKI INSTRUMENTS, INC. INCLUDING BUT NOT LIMITED TO, THE
WARRANTY OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE. IN NO EVENT SHALL RKI INSTRUMENTS,
INC. BE LIABLE FOR INDIRECT, INCIDENTAL, OR CONSEQUENTIAL
LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF
ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR
OPERATE PROPERLY.
This warranty covers instruments and parts sold to users by authorized distributors,
dealers, and representatives as appointed by RKI Instruments, Inc.
We do not assum e i ndemnification for any accident or damage caused by the op e r a t ion of
this gas monitor, and our warranty is limited to the replacement of parts or our complete
goods.
This manual describes the 65-2405RKSS combustible gas transmitter. This manual also
describes how to install, start up, con fi gure, mainta in, a nd ca librate the transmitter when
it is used with a gas monitoring controller. A parts list at the end of this manual lists
replacement parts and accessories for the combustible gas transmitter.
Specifications
Table 1 lists specifications for th e combustible gas transmitter.
Table 1:Specifications
DescriptionSpecification
Target Gas
Area ClassificationExplosionproof for Class I, Groups B, C, and D
Sampling MethodDiffusion
Signal Output4 to 20 mA
Detection Range0 to 100% LEL
Accuracy± 5% of reading or ± 2% LEL (whichever is greater)
Response Time90% in 45 seconds
1
If calibration to a combustible gas other than methane is required, specify the required
calibration gas at time of order.
Combustible gas, methane calibration standard
1
WARNING :When using the 65-2405RKSS, you must follow the instructions and
warnings in this manual to assure proper and safe opera tion of the
65-2405RKSS and to minimize the risk of personal injury. Be sure to
maintain and periodically calibrate the 65-2405RKSS as described in this
manual.
65-2405RKSS Combustible Ga s Tra nsm itter • 1
Description
This section describes the components of the combustible gas transmitter. The transmitter
is a 4 - 20 mA type detector head. It consists of the combustible gas detector, amplifier , and
junction box.
Factory A djust Pot
Span Pot
Zero Pot
Fact ory A djust Pot
Jumper Pins,
Factory Use Only
Red + Test P oint
Black - Test P oint
Reducing Bushing,
3/4NPTto1/2NPT
Controller Term inal Strip
Amplifier
SPAN
ZERO
Sensor
Current
148mA
S
PWR / SIG
G
LEL
WRB
Mounting Screw
Detector Terminal Strip
Detector (sensor) Current
Label
J-Box
Figure 1: Combustible Gas Transmitter Component Location
Combustible Gas Detector
The combustible gas detector is a catalytic type detector that produces an electrical output
that corresponds to the detection range. It is packaged in a 1/2 inch NPT nipple with a
sintered metal flame arrestor on one end allowing ambient air to diffuse into the detector.
The flame arrest or al so co ntai n s any s park s whi ch may occur within the detector. The 1/2
inch NPT mounting threads at the top of the detector allow you to mount it into the
bottom conduit hub of the junction box. A 3/4 in. x 1/2 in. NPT reducer is necessary to
install the combustible gas detector in the junction box. A rainshield screws onto the
bottom of the detector (flame arrestor end). The rainshield helps protect the detector from
rain and debris in the monitoring environment. Four color-coded leads extend from the
top of the detector. The leads allow you to connect the detector to the amplifier.
2 • 65-2405RKSS Combustible Gas Transmitter