RKI Instruments 65-2405RK-05 User manual

Page 1
Combustible Gas Transmitter
Part Number: 71-0180RK Revision: 0
65-2405RK-05
Operator’s Manual
Released: 2/16/11
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 2
WARNING
Read and understand this instruction manual before operating detector . Improper use of the dete ctor could result in bodily harm or death.
Periodic calibration and maintenance of the detect or is essential for proper operation and correct readings. Please calibrate and maintain this detector regularly! Frequency of calibration depends upon the type of use you have and the sensor types. T ypical calibration frequencies for most applications are b etween 3 and 6 months, but can be required more often or less often based on your usage.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 3
Product Warranty
RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year fr o m date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge. This warranty does not apply to those items which by their nature are subject to deterioration or consumption in no r mal service, and which must be cleaned, repaired, or replaced on a routine basis. Examples of such items are:
W arranty is voided by abuse including mechanical damage, alteration, rough handling, or repair procedures not in accordance with th e operator’s manual. This warranty indicates the full extent of our liability , a nd we are not responsible for r emoval or replacement costs, local repair costs, transportation costs, or contingent expenses incurred without our prior approval.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI INSTRUMENTS, INC. INCLUDING BUT NOT LIMITED TO, THE WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL RKI INSTRUMENTS, INC. BE LIABLE FOR INDIRECT, INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
a) Absorbent cartridges d) Batteries b) Pump diaphragms and valves e) Filter elements c) Fuses
This warranty covers instruments and parts sold to users by authorized distributors, dealers, and representatives as appointed by RKI Instruments, Inc.
We do not assume indemnification for a ny accident or damage ca u s e d by the operation of this gas monitor, and our warranty is limited to the replacement of parts or our complete goods.
65-2405RK-05 Combus tible G a s Transmitter
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 4
Table of Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Combustible Gas Detector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Junction Box. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Mounting the Combustible Gas Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Wiring the Combustible Gas Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Startup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Introducing Incoming Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Setting the Zero Signal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Replacing Components of the Combustible Gas Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Calibration Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Preparing for Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Setting the Zero Reading. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Setting the Response Reading. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Returning to Normal Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 5
Overview
This manual describes the 65-2405RK-05 combustible gas transm itter. This manual also describes how to install, start up, configure, maintain, and calibrate the transmitter w hen it is used with a gas monitoring controller. A parts list at the end of this manual lists replacement parts and accessories for the combustible gas transmitter.
Specifications
WARNING: Do not use this product in a manner not specified in this manual.
Table 1 lists specification s for the combustible gas transmitter.
Table 1:Specifications
Description Specification
Target Gas
Area Classification Explosionproof for Class I, Groups B, C, and D Sampling Method Diffusion Signal Output 4 to 20 mA Detection Range 0 to 100% LEL Accuracy ± 5% of reading or ± 2% LEL (whichever is greater) Response Time 90% in 45 seconds Operating Temperature
1
If calibration to a combustible gas other than methane is required, specify the required
calibration gas at time of order.
Combustible gas, methane calibration standard
-40°F to 167°F (-40°C to 75°C)
1
NOTE: The following symbol on the detector label is a caution to the user to refer to this
documentation for installation instructions:
!
WARNING: When using the 65-2405RK-05, you must follow the instructions and
warnings in this manual to assure proper and safe operation of the 65-2405RK-05 and to minimize the risk of personal injury. Be sure to maintain and periodically calibrate the 65-2405RK-05 as described in this manual.
65-2405RK-05 Combustible Ga s Tr a nsm itter • 1
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 6
Description
This section describes the components of the combustible gas transmitter. The transmitter is a 4 - 20 mA type detector head. It consists of the combustible gas detector, amplifier, and junction box.
Factory Adjust Pot
Span Pot
Zero Pot
Factory Adjust Pot
Ju mp er Pins,
Factory Use Only
+ Test Point
- Test Point
Controller Terminal Strip
Amplifier
SPAN
ZERO
Sensor
Current
mA
SBR
PWR / SIG
S2 001
57-1053RK
G
LEL
W
Detector Terminal Strip
J-Box
Reducing Bushing,
3/4 NPT to 1/2 NPT
Figure 1: Combustible Gas Transmitter Component Location
Combustible Gas Detector
The combustible gas detector is a catalytic type detector that produces an electrical output that corresponds to the detection range. It is packaged in a 1/2 inch NPT nipple with a sintered metal flame arrestor on one end allowing ambient air to diffuse into the detector. The flame arrest or al so co ntai n s any s park s whi ch may occur within the detector. The 1/2 inch NPT mounting threads at the top of the detector allow you to mount it into the bottom conduit hub of the junction box. A flame arrestor guard is located at the bottom of the detector and is permanently bonded to the detector housing. The flame arrestor guard protects the flame arrestor from impact damage. Four color-coded leads extend from the top of the detector. The leads allow you to connect the detector to the amplifier.
2 • 65-2405RK-05 Combustible Gas Transmitter
LEL Detector
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 7
Amplifier
The amplifier converts the electrical output from the detector to a 4 to 20 mA signal that corresponds to the detection range and transmits the signal to a gas monitoring controller . A foam gasket that orients the amplifier and keeps it from rotating is installed on the bottom of the amplifier. A label on the amplifier indicates the detector drive current. This drive current is factory set and is dictated by the combustible gas to which the detector is calibrated. Consult RKI Instruments, Inc. if you plan to change the calibration gas for the transmitter, for e xample from a methane calibration to a hexane calibration. The amplifier includes the detector terminal strip, controller terminal strip, span pot, zero pot, and test points (see Figure 1).
Controller Terminal Strip
The controller terminal strip is a three position plug-in style terminal strip located near the top of the amplifier . U se the contro ller terminal strip to wire the transmitter to a controller.
Detector Terminal Strip
The detector terminal strip is a four position plug-in style terminal strip located below the controller terminal strip. Use the detector terminal strip to connect the detector to the amplifier.
NOTE: The detector is factory-wired to the detector terminal strip. See “Wiring the
Combustible Gas Transmitter to a Controller” on page 5 for all wiring procedures related to the transmitter.
Zero Pot
The zero pot is located in the upper left corner of the amplifier (see Figure 1). Use a small flat blade screwdriver to turn the zero pot’s adjustment screw and ad just the amplifier’s zero (fresh air) output during the start-up and calibration pro cedures.
Span Pot
The span pot is located to the right of the zero pot (see Figure 1). Use a small flat blade screwdriver to turn the span pot’s adjustment screw and adjust the amplifier’s gas response output during the calibration procedure.
CAUTION: The amplifier includes two additional potentiometers. They are factory-set. Do not
adjust them.
Test Points
The test points are on the left side of the amplifier (see Figure 1). The test points produce a 100 mV to 500 mV output that corresponds to the transmitter’s 4 to 20 mA output. Use the test points and a voltmeter to measure the amplifier’s output during the start-up and calibration procedures. The black test point in the lower left corner is the negative (-) test point and the red test point below the zero pot is the positive (+) test point.
Junction Box
Use the junction box to install the transm itter at a mounting site that is remote from the controller. The junctio n bo x protects the amplifier and wirin g connections made to the amplifier. Use the top 3/4’’ conduit hub to connect wiring from the amplifier to the controller. Use the cover on the front of the junction box to access the interior of the junction box. The detector and amplifier are factory installed in th e junc tion box. Three spacers installed on the back of the junction box control the distance of the junction box
65-2405RK-05 Combustible Ga s Tr a nsm itter • 3
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 8
Installation
from a mounting surface and ensure that there is enough room to install a calibration cup on the detector during calibration.
This section describes procedures to mount the combustible gas transmitter in the monitoring environment and wire the transmitter to a controller.
Mounting the Combustible Gas Transmitter
1. Select a mounting site that is representative of the monitoring environment. Consider the following when you select the mounting site.
Select a site where the transmitter is not likely to be bumped or disturbed. Make
sure there is sufficient room to perform start-up, maintenance, and calibration procedures.
Select a site where the target gas is likely to be found first. For lighter gases,
mount the detector near the ceiling; for heavier gases, mount the detector near the floor.
3/4 NPT Female
2.70 .38
3.65
7.5 MAX Rubber
Spacer, 3X
J-Box
3/4 x 1/2
Reduci ng Bus hing
LEL Detector
.75
Figure 2: Mounting the Combustible Gas Transmitter
2. At the monitoring site you select, hang or mount the junction box with the detector facing down (see Figure 2).
4 • 65-2405RK-05 Combustible Gas Transmitter
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 9
Wiring the Combustible Gas Transmitter to a Controller
WARNING: Always verify that the power to the controller is off before you make
wiring connections.
1. Turn off the controller.
2. Turn off or unplug power to the controller.
3. Remove the junction box co ver.
4. The detector leads are factory wired. Verify that the detector leads are wired to the amplifier’s detector terminal strip as shown in Figure 3.
5. To gain access to a plug-in terminal strip for wiring, pull it out of its socket by grasping the terminal strip and pulling. The detector strip is keyed so that the controller and detector terminal strips cannot be reversed inadvertently.
WARN IN G: To maintain the explosion proof classification of the detector/junction box
combination, a conduit seal must be used within 18 inches of the junction box conduit hub used for wiring to the controller.
6. Guide a three-conductor, shielded cable or three wires in conduit through the top conduit hub of the junction box.
7. Pull out the controller terminal strip, and connect the three wires to the terminal strip as follows (see Figure 3).
Connect the positive wire to the PWR/SIG “+”terminal.
Connect the signal wire to the PWR/SIG “S” terminal.
Connect the negative wire to the PWR/SIG “-” terminal.
CAUTION: If shielded cable is used, leave the cable shield’s drain wire disconnected and
insulated at the transmitter. You will connect the opposite end of the cable’s drain wire to the controller’s chassis (earth) ground.
8. Reinstall the controller terminal strip into its socket.
9. Secure the junction box cover to the junction box.
10. Route the cable or wires leading from the combustible gas transmitter through one of the conduit hubs at the controller housing.
CAUTION: Do not route power and transmitter wiring through the same controller conduit hub.
The power cabl e may disrupt the tra n smission of the transmitter signal to the controller.
11. Connect the wires to the applicable detector/transmitter terminal strip at the controller as shown in Figure 3 .
65-2405RK-05 Combustible Ga s Tr a nsm itter • 5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 10
Cable Shield
+ 24 VDC 4 - 20 mA In (S)
- (D C Ground)
J-Box
Controller or
Recordin g Device
Amplifier (Transmitter)
S
PWR / SIG
B
LEL
W GR
Black Green White Red
Detector Wires
LEL Detector
Figure 3: Wiring the Combustible Gas Transmitter to a Controller
12. If shielded cable is used, connect the cable’s drain wire to an available chassis (earth) ground at the controller. RKI controllers typically have a ground stud that can be used to ground the cable’s drain wire.
6 • 65-2405RK-05 Combustible Gas Transmitter
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 11
Start Up
This section describes procedures to start up the combustible gas transmitter and place the transmitter into normal operation.
Introducing Incoming Power
1. Complete the installation procedures described earlier in this manual.
2. Verify that the po wer wiring to the controller is correct and secure. Refer to the controller operator’s manual.
3. Turn on power to the controller.
4. Turn on the controller.
5. Ve rify that the controller is on and operating properly. Refer to the controller operator’s manu al.
CAUTION: Allow the detector to warm up for 5 minutes before you continue with the next
section, “Setting the Zero Signal”.
Setting the Zero Signal
WARNING: Do not remove the junction box cover while the circuits are energized
unless the area is determined to be non-hazardous. Keep the junction box cover tightly closed during operation.
CAUTION: If you suspect the presence of combustible ga s in the monitoring environm ent, use
the calibration kit and the zero air calibration cylinder to introduce “fresh air” to the detector and verify an accurate zero setting. See “Calibration” on page 13 for instructions to apply zero air when setting the zero signal.
1. V erify that the transmitter is in a fresh air environment (environment known to be free of combustible gas).
2. Unscrew and remove the junction box cover from the junction box.
3. Set a voltmeter to measure in the millivolt (mV) range.
4. Plug the voltmeter leads into the test points on the amplifier. P lug the positive lead into the red (+) test point; plug the negative lead into the black (-) test point.
5. Verify a voltmeter reading of 100 mV (±2 mV).
6. If necessary, use a small fla t-blade screwdriver to adjust the zero pot until the voltmeter reading is 100 mV (±2 mV).
7. Remove the voltmeter leads from the test points.
8. Secure the junction box cover to the junction box.
65-2405RK-05 Combustible Ga s Tr a nsm itter • 7
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 12
Maintenance
This section describes maintenance procedures. It includes preventive maintenance, troubleshooting, and component replacement procedures.
Preventive Maintenance
This section describes a preventive maintenance schedule to ensure the optimum performance of the combustible gas transmitter. It includes daily, monthly, and quarterly procedures.
Daily
Verify a display reading of 0 %LEL at the controller. Investigate significant changes in the display reading.
Monthly
This procedure describes a test to verify that the combustible gas transmitter responds properly to the target gas.
WARNING: The controller is not an active gas monitoring device during the response
NOTE: Performing a response test on the combustible gas transmitter may cause alarms.
test procedure.
Be sure to put the controller into its calibration mode or disable external alarms before performing this test.
NOTE: The following procedure assumes the use of a calibration kit which includes a
calibration gas cylinder, a 0.5 LPM fixed flow regulator with an on/off knob, a calibration cup for the detector, and a length of sample tubing.
Preparing for the response test
1. Place the controller into its calibration mode or disable external alarms.
2. Ve rify that the controller display reading for the channel you are testing is 0 %LEL. If the display reading is not 0 %LEL, set the zero reading of the transmitter as
described in “Start Up” on page 7, then continue this procedure.
WARNING: Do not remove the junction box cover while the circuits are energized
unless the area is determined to be non-hazardous. Keep the junction box cover tightly closed during operation.
3. Unscrew and remove the junction box cover.
4. Set a voltmeter to read in the milli volt (mV) range.
5. Plug the voltmeter leads into the test points on the amplifier. P lug the positive lead into the red (+) test point; plug the negative lead into the black (-) test point.
8 • 65-2405RK-05 Combustible Gas Transmitter
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 13
6. Use the following formula to determine the correct test points output for the test sample.
Output (mV) = (calibrating sample/fullscale) X 400 + 100
For example, with a test sample of 50 %LEL and a fullscale setting of 100 %LEL, the correct output is 300 mV.
300 (mV) = (50/100) X 400 +100
7. Screw the calibration cup onto the bottom of the detector.
8. Screw the regulator into the calibration cylinder.
9. Use the calibration kit sample tubing to connect the regulator to the calibration cup.
Performing the response test
1. Turn the regulator’s on/off knob counterclockwise to open the regulator. Gas will begin to flow.
2. Allow the gas to flow for one minu te, then verify that the reading is within ± 20% of the response reading you determined earlier.
NOTE: If the reading is not within ± 20% of the correct response reading, calibrate the
transmitter as described in “Calibration” on page 13 of this manual.
3. Turn the regulator’s on/off knob clockwise to close the regulator.
4. Unscrew the regulator from the calibration cylinder.
5. Unscrew the calibration cup from the detector.
6. Remove the voltmeter leads from the amplifier test poin ts.
7. Reinstall the junction box cover.
8. When the controller display reading falls below the alarm setpoints, return the controller to normal operation.
9. Store th e components of the calibration kit in a safe place.
Quarterly
Calibrate the combustible gas transmitter as described in “Calibration” on page 13 of this manual. See the calibration frequency discussion in “Calibration Frequency” on page 12 to determine if a quarterly calibration schedule fits your needs.
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended action for problems you may encounter with the combustible gas transmitter.
NOTE: This troubleshooting guide describes transmitter problems only. See the
controller operator’s manual for problems you m ay encounter with the controller.
65-2405RK-05 Combustible Ga s Tr a nsm itter • 9
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 14
Table 2:Troubleshooting the Combustible Gas Transmitter
Condition Symptom(s) Probable Causes Recommended Action
Fail Condition • Controller indicates a
Slow or No Response/ Difficult or Unable to Calibrate
fail condition.
• Transmitter responds slowly or does not respond to response test.
• Unable to accurately set the zero or response reading during cali bration.
• Transmitter requires frequent calibration.
Note: Under “normal” circumstances, the transmitter requires calibration once every 3 months.
Some applications may require a more frequent calibration schedule.
• The transmitter wiring is disconnected or misconnected.
• The transmitter’s zero reading is low enough to cause a fail condition.
• The transmitter is malfunctioning.
• The calibration cylinder is low, out-dated, or defective.
• The calibration gas flow rate is too low.
• The calibration gas is not an appr opriate concentration.
• The transmitter is malfunctioning.
1. Verify that the transmitter wiring is correct and secure.
2. Calibrate the transm itter.
3. If the fail condition continues, replace the detector.
4. If the fail condition continues, contact RKI for further instruction.
1. Verify that the calibration cylinder contains an adequate supply of a fresh test sample.
2. Verify that the regulator used for calibration is a 0.5 LPM regulator.
3. Verify that the calibration gas concentration is appropriate for the transmitter. The concentration should be in the detection range, preferably about half of the detection range.
4. If the calibration/response difficulties continue, replace the detector.
5. If the calibration/response difficulties continue, contact RKI for further instruction.
Replacing Components of the Combustible Gas Transmitter
This section includes procedures to replace the combustible gas detector and amplifier.
Replacing the Combustible Gas Detector
1. Turn off the controller.
2. Turn off or unplug power to the controller.
3. Remove the junction box co ver.
4. Remove the detector terminal strip from its socket.
5. Disconnect the detector leads from the detector terminal strip. Note the position of the color-coded leads as you remove them.
6. Unscrew the detector from the junction box.
7. Guide the detector leads of the replacement detector through the bottom conduit hub of the junction box, then screw the mounting threads of the detector into the conduit hub. If necessary for environmental conditions, apply thread sealant or teflon tape to the hub and/or detector threads to seal them.
10 • 65-2405RK-05 Combustible Gas Transmitter
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 15
8. Connect the detector leads to the detector terminal strip as shown in Table 3 below and Figure 3 on page 6 of th is manual.
Table 3:Reconnecting the Combustible Gas
Detector to the Amplifier
Detector Lead Detector Terminal Strip
Red LEL “R”
White LEL “W”
Green LEL “G”
Black LEL “B”
9. Reinstall the detector terminal strip into its socket.
10. Reinstall the junction box cover.
11. Turn on or plug in power to the controller.
12. Turn on the controller and place it into normal operation.
CAUTION: Allow the replacement detector to warm up for 5 minutes before you continue with
the next step.
13. Calibrate the transmitter as described in “Calibration” on page 13 of this manual.
Replacing the Amplifier
1. Turn off the controller.
2. Turn off or unplug power to the controller.
3. Remove the junction box co ver.
4. Unplug the detector terminal strip and controller terminal strip from their sockets. You may leave the wires connected to the terminal strips.
5. Unscrew and remove the screw with the flat and lock washers that secures the amplifier to the junction box.
6. Remove the old amplifier.
7. Place the new amplifier in the same position as the old amplifier. A foam gasket that orients the amplifier and keeps it from rotating is installed on the bottom of the amplifier. Make sure th e amplifier is seated flat in the junction box.
8. Install the amplifier into the junction box with the screw, lock washer, and flat washer you removed in Step 5 ab ove.
9. Install the detector and controller terminal strips into their sockets on the new amplifier as shown in Figure 3 on page 6 of this manual. If controller leads or detector leads were removed during this procedure, refer to Table 4 and Table 5 below.
65-2405RK-05 Combustible Gas Transmitter • 11
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 16
Table 4:Reconnecting the Amplifier
to the Controller
Amplifier Controller
Terminal Strip
PWR/SIG “-” - (DC -) PWR/SIG “S” S (4 - 20 mA In) PWR/SIG “+” + 24V
Table 5:Reconnecting the ppm
Detector to the Amplifier
Amplifier Detector
T erminal Strip
DETECTOR “R” RED DETECTOR “W” WHT DETECTOR “G” GREEN
DETECTOR “B” BLK
Controller
Transm itte r Terminal
Strip (typical)
Detector Lead
NOTE: When a transmitter is first powered up with a new amplifier, the initial output
may be either high or below zero depending on the setting of the zero pot. Be sure to make arrangements so that this does not cause unwanted alarms.
10. Turn on or plug in power to the controller.
11. Turn on the controller and place it into normal operation.
12. Allow the transmitter to warm-up for 5 minutes.
13. Calibrate the transmitter as described in “Calibration” on page 13 of this manual.
Calibration Frequency
Although there is no particular calibration frequency that is correct for all applications, a calibration frequency of every 3 to 6 months is adequate for most combustible gas transmitter applications. Unless experience in a particular application dictates otherwise, RKI Instruments, Inc. recommends a calibration frequency of every 3 months.
If an application is not very demanding, for example detection in a clean, temperature controlled environment where combustible gas is not normally present and calibration adjustments are minimal at calibration, then a calibration frequency of every 6 months is adequate.
If an application is very demanding, for example if combustible gas is present often and in significant concentrations or the environment is not well controlled, then more frequent calibration than every 3 months may be necessary. If potential catalyst poisons are known or likely to be present, more frequent calibration than every 3 months will be necessary.
12 • 65-2405RK-05 Combustible Gas Transmitter
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 17
Calibration
This section describes how to calibrate the combustible gas transmitter. It includes procedures to prepare for calibration, set the zero reading, set the response reading, and return to normal operation.
W ARNING: The controller is not an active gas monitoring device during the calibration
procedure.
NOTE: The following procedure assumes the use of a calibration kit which includes a
calibration gas cylinder, a 0.5 LPM fixed flow regulator with an on/off knob, a calibration cup for the detector, and a short piece of sample tubing to connect the regulator to the calibration cup.
Prepari ng for Ca libra t ion
NOTE: Calibrating the combustible gas transmitter may cause alarms. Be sure to put the
controller into its calibration program or disable external alarms before continuing.
WARNING: Do not remove the junction box cover while the circuits are energized
unless the area is determined to be non-hazardous. Keep the junction box cover tightly closed during operation.
1. Unscrew and remove the junction box cover.
2. Set a voltmeter to measure in the millivolt (mV) range.
3. Plug the voltmeter leads into the test points on the amplifier. P lug the positive lead into the red (+) test point; plug the negative lead into the black (-) test point.
4. Use the following formula to determine the correct test points o utput for the calibrating sample.
Output (mV) = (calibrating sample/fullscale) X 400 + 100
For example, with a calibrating sample of 50 %LEL and a fullscale setting of 100 %LEL, the correct output is 300 mV.
300(mV) = (50/100) X 400 +100
5. Screw the calibration cup on the detector housin g.
Setting the Zero Reading
NOTE: If you can verify that the combustible gas transmitter is in a fresh air
environment, you do not need to apply zero air to the detector before adjusting the zero reading.
1. Screw the regulator into the zero air calibration cylinder.
2. Use the calibration kit sample tubing to connect the regulator to the calibration cup.
3. Turn the regulator’s on/off knob counterclockwise to open the regulator.
65-2405RK-05 Combustible Gas Transmitter • 13
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 18
4. Allow the gas to flow for one minute and verify a reading of 100 mV (±2 mV). If necessary, use the zero pot on the amplifier to adjust the reading to 100 mV (±2 mV).
5. Turn the regulator’s on/off knob clockwise to close the regulator.
6. Unscrew the regulator from the zero air calibration cylinder.
7. Leave the sample tubing connected to the regulator and the calibra tion cup.
Setting the Response Reading
1. Screw the regulator into the calibration cylinder. Verify that the calibration gas is representative of the transmitter’s target gas.
2. Turn the regulator’s on/off knob counterclockwise to open the regulator.
3. Allow the calibration gas to flow for one minute and verify that the reading matches the response reading (±2 mV) you determined earlier. If necessary, use the span pot on the amplifier to adjust the reading to match the correct response reading.
4. Turn the regulator’s on /off knob clockwise to close the regulator.
5. Unscrew the regulator from the calibration cylinder.
Returning to Normal Operation
1. Remove the voltmeter leads from the amplifier test poin ts.
2. Unscrew the calibration cup from the detector.
NOTE: For convenience, leave the regulator and calibration cup connected by the
sample tubing.
3. Secure the junction box cover to the junction box.
4. When the controller display reading falls below the alarm setpoints, return the controller to normal operation.
5. Verify that the controller display reading decreases and stabilizes at 0 %LEL.
6. Store th e components of the calibration kit in a safe and convenient place .
14 • 65-2405RK-05 Combustible Gas Transmitter
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 19
Parts List
Table 6 lists replacement parts and accessories for the combustible gas transmitter.
Table 6:Parts List
Part Number Description
06-1248RK Sample tubing (3/16 in. x 5/16 in.; specify length when ordering) 18-0001RK Reducer, 3/4 in. NPT x 1/2 in. NPT 18-0400RK-01 Junction box with spacers 57-1053RK Amplifier with gasket (spe cify t a rget gas and th at the 61 -0140RK-05 detec tor is
61-0140RK-05 Combustible gas detector, CSA type 65-2405RK-05 Combustible gas transmitt er (inclu des dete ctor and amplifi er; spec ify ta rget gas
71-0180RK 65-2405RK-05 Combustib le Ga s Transmitter Operator’s Manual (this docu-
81-0002RK-01 Steel calibration cylinder, 50% LEL hydrogen in air, 34-liter 81-0007RK-01 Steel calibration cylinder, 15% LEL hexane in air, 34-liter 81-0012RK-01 Steel calibration cylinder, 50% LEL methane in air, 34-liter 81-0076RK-01 Steel calibration cylinder, zero emission air, 34-liter 81-1050RK Regulator, 0.5 liter/minute; with pressure gauge and flow control knob, for 17
81-1117RK Calibration cup 81-F001RK-LV Calibration kit, includes regulator, sample tubing, calibration cup, and 34 liter
81-F002RK-LV Calibration kit, includes regulator, sample tubing, calibration cup, and 34 liter
81-F004RK-LV Calibration kit, includes regulator, sample tubing, calibration cup, and 34 liter
being used w hen ordering)
when ordering)
ment)
liter and 34 liter steel calibration cylinders
15 %LEL hexane in air steel calibration cylinder
50 %LEL methane in air steel calibrati on cylin der
50 %LEL hydrogen in air steel calibration cylinder
65-2405RK-05 Combustible Gas Transmitter • 15
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Loading...