RKI Instruments 65-2399RK-CH4 User Manual

Page 1
65-2399RK-CH4
Part Number: 71-0201RK Revision: A Released: 4/12/11
Methane Transmitter
Operator’s Manual
RKI Instruments, Inc.
www.rkiinstruments.com
Page 2
WARNING
Read and understand this instruction manual before operating transmitter. Improper use of the transmitter could result in bodily harm or death.
Periodic calibration and maintenance of the transmitter is essential for proper operation and correct readings. Please calibrate and maintain this tra nsmitt er regula rly! Frequency of calibration depends upon the type of use you have and the sensor types. Typical calibration frequencies for most applications are between 6 and 12 months, but can be required more often or less often based on your usage.
65-2399RK-CH4 Combustible Gas Transmitter
Page 3
Product Warranty
RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year fr o m date of shipment from RKI Instruments, Inc. Any parts found defective withi n tha t period will be repaired or replaced, at our option, free of charge. This warranty does not apply to those items which by their nature are subject to deterioration or consumption in normal ser v ice, and which must be cleaned, repaired, or replaced on a routine basis. Examples of such items are:
W arranty is voided by abuse including mechanical damage, alteration, rough handling, or repair procedures not in accordance with the operator’s manual. This warranty indicates the full extent of our liability , a nd we are not r esponsible for removal or r eplacement costs, local repair costs, transportation costs, or contingent expenses incurred without our prior approval.
a) Absorbent cartridges d) Batteries b) Pump diaphragms and valves e) Filter elements c) Fuses
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI INSTRUMENTS, INC. INCLUDING BUT NOT LIMITED TO, THE WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL RK I INSTRUMENTS, INC. BE LIABLE FOR INDIRECT, INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
This warranty covers instruments and parts sold to users by authorized distributors, dealers, and representatives as appointed by RKI Instruments, Inc.
We do not assume i ndemnification for a n y a c c ident or damage cause d b y the operation of this gas monitor, and our warranty is limited to the replacement of parts or our complete goods.
65-2399RK-CH4 Combustible Gas Trans mitte r
Page 4
Table of Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Methane Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Junction Box. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Mounting the Methane Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Wiring the Methane Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Startup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Introducing Incoming Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Setting the Zero Signal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Replacing Components of the Methane Transmitter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Calibration Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Preparing for Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Setting the Zero Reading. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Setting the Response Reading. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Returning to Normal Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
65-2399RK-CH4 Combustible Gas Transmitter
Page 5
Overview
This manual describes the 65-2399RK-CH4 methane transmitter. This manual also describes how to install, start up, con fi gure, mainta in, a nd ca librate the transmitter when it is used with a gas monitoring controller. A parts list at the end of this manual lists replacement parts and accessories for the methane transmitter.
Specifications
Table 1 lists specifications for the methane transmitter.
Table 1:Specifications
Description Specification
Target Gas Methane (CH Sampling Method Diffusion
Signal Output 4 to 20 mA Detection Range 0 to 100% Volume Response Time 90% in 45 seconds Operating Temperature Accuracy ± 5% of reading or ± 2% of full scale (whichever is gr eater)
-4°F to 122°F (-20°C to 50°C)
)
4
WARN IN G: When using the 65-2399RK-CH4, you must follow the instructions and
warnings in this manual to assure proper and safe opera tion of the 65-2399RK-CH4 and to mini mize the risk of personal injury. Be sure to maintain and periodically calibrate the 65-2399RK-CH4 as described in this manual.
65-2399RK-CH4 Combus tible Ga s Tr a nsm itter • 1
Page 6
Description
l
The 65-2399RK-CH4 transmitter is calibrated to methane. The transmitter utilizes an infrared type of detector which has some advantages over a catalytic type of combustible detector. The infrared de tector will generally have a longer service life than a catalytic detector, it will req uire calibration less often, and it can detect methane even if there is no oxygen in the sample which allows d etection of methane in an inert atmosphere.
This section describes the components of the methane transmitter. The transmitter is a 4 ­20 mA type detector head. It consists of the methane detector, amplifier, and junction box.
Factory Adjust Pot
Jumper Pins,
Factory Use Only
100mV-500mV
Range
Factory Adjust Pot
Span Pot
Zero Pot
+ Test Point (Red)
- Test Point (Black)
Controller Terminal Strip
Amplifier
SPAN
ZERO
Sensor
Current
mA
S
PWR / SIG
BWR G
LEL
Detector Termina Strip
J-Box
IR CH4%Vol ume Detector
Figure 1: Methane Transmitter Component Location
2 • 65-2399RK-CH4 Combustible Gas Transmitter
Page 7
Infrared CH4 %Volume Detector
The infrared CH4 %Vol. detector consists of the detector housing body, detector housing cap, cap gasket, and the plug-in sensor.
Detector Housing Body
Cap Gask et
Plug-In IR CH4 %Volume Sensor
Detector Housing Cap
Hydrophobic Membrane
Figure 2: Infrared CH4 %Vol. Detector Component Location
Detector Housing Body
The detector housing body protects the electronic components within the housing. Use the mounting threads at the top of the housing to screw the IR CH
NPT hub on the bottom of the junction box. Four color coded leads, red, white, green, and black, extend from the other end of the detector. The leads allow you to connect the detector to the amplifier.
The housing includes three sockets installed on a circuit board. These sockets accept the plug-in sensor’s three pins to provide an electrical connection for the sensor. The circuit board with the sockets conditions the sensor’s signal before the signal reaches the amplifier.
Housing Cap & Cap Gasket
The housing cap screws onto the detector housing body. It retains the plug-in sensor a nd protects it from damage. A foam gasket is installed inside the ho using cap that seals against the sensor face. A hydrophobic membrane on the outside of the cap face keeps water and particulates away from the sensor face behind the cap. Unscrew the detector cap to access the plug-in sensor for maintenance or replacement. A cap gasket seals the interface between the housing body and cap.
Plug-In CH
The plug-in se ns or is secured in the de t ect or assembly by the housing cap. The sensor pro­duces a millivolt output that corresponds to the detection range of the trans mitter.
Sensor
4
detector into the 3/4”
4
65-2399RK-CH4 Combus tible Ga s Tr a nsm itter • 3
Page 8
Amplifier
The amplifier converts the electrical output from the detector to a 4 to 20 mA signal that corresponds to the detection range and transmits the signal to a gas monitoring controller . A foam gasket that orients the amplifier and keeps it from rotating is installed on the bottom of the amplifier. A label on the amplifier indicates the detector drive current. The detector drive current is factory set to 148 mA for the 65-2399RK-CH4. The amplifier includes the detector terminal strip, controller terminal strip, span pot, zero pot, and test points (see Figure 1).
Controller Terminal Strip
The controller terminal strip is a three position plug-in style terminal strip located near the top of the amplifier. Use the controller terminal strip to wire the methane transmitter to a controller.
Detector Terminal Strip
The detector terminal strip is a four position plug-in style terminal strip located below the controller terminal strip. Use the detector terminal strip to connect the IR CH
detector to the amplifier.
NOTE: The IR CH4 %Vol. detector is factory-wired to the detector term in al strip. See
“Wiring the Methane Transmitter to a Controller” on page 6 for all wiring procedures related to the transmitter.
%Vol.
4
Zero Pot
The zero pot is located in the upper left corner of the amplifier (see Figure 1). Use a small flat blade screwdriver to turn the zero pot’s adjustment screw and adjust th e amplifier’s zero (fresh air) output during the start-up and calibration procedures. Turn the adjust­ment screw clockwise to increase the zero output and counterclockwise to decrease the zero output.
Span Pot
The span pot is located to the right of the zero pot (see Figure 1). Use a small flat blade screwdriver to turn the span pot’s adjustment screw and adjust the amplifier’s gas response output during the calibration procedure. Turn the adjustment screw clockwise to increase the gas response and counterclockwise to decrease the gas response.
CAUTION: The amplifier includes two additional potentiometers. They are factory-set. Do not
adjust them.
Test Points
The test points are on the left side of the amplifier (see Figure 1). The test points produce a 100 mV to 500 mV output that corresponds to the transmitter’s 4 to 20 mA output. Use the test points and a voltmeter to measure the amplifier’s output during the star t-up and calibration procedures. The black test point in the lower left corner is the negative (-) test point and the red test point below the zero pot is the positive (+) test point.
Junction Box
Use the junction box to install the transmitter at a mounting site that is remote from the controller. The junction box protects the amplifier and wiring connections made to the amplifier. Use the top 3/4’’ conduit hub to connect wiring from the amplifier to the
4 • 65-2399RK-CH4 Combustible Gas Transmitter
Page 9
Installation
controller. Use the cover on the f ront of the junction box to access the interior of the junction box. The detector and amplifier are factory installed in the junction box. Three spacers installed on the back of the junction box control the distance of the junction box from a mounting surface and ensure that there is enough room to install a calibration cup on the detector during calibration.
This section describes procedures to mount the methane trans mitter in the monitoring environment and wire the transmitter to a controller.
Mounting the Methane Transmitter
1. Select a mounting site that is representative of the monitoring environment. Consider the following when you select the mounting site.
Select a site where the transmitter is not likely to be bumped or disturbed. Make
sure there is sufficient room to perform start-up, maintenance, and calibration procedures.
Select a site where the target gas is likely to be found first. For lighter gases,
mount the detector near the ceiling; for heavier gases, mount the detector near the floor.
3/4 NPT
Female
3.10
.75
.38
3.65
6.80 MAX Rubber
Spacers, 3X
J-Box
IR CH4
%Volume
Detector
1 1/2-20Thread for Calibration Cup
Figure 3: Mounting the Methane Transmitter
65-2399RK-CH4 Combus tible Ga s Tr a nsm itter • 5
Page 10
2. At the monitoring site you select, hang or mount th e jun c tion box with the detector facing down (see Figure 3).
Wiring the Methane Transmitter to a Controller
WARN ING: Always verify that the power to the controller is off before you make
wiring connections.
1. Turn off the controller.
2. Turn off or unplug power to the controller.
3. Remove the junction box cover.
4. The detector leads are factory wired. Verify that the detector leads are wired to the amplifier’s detector terminal strip as shown in Figure 4.
5. To gain access to a plug-in terminal strip for wiring, pull it out of its sock et by grasping the terminal strip and pulling. The socket for the detector strip and the plug for the controller strip are keyed so that the controller and detector terminal strips cannot be reversed inadvertently.
6. Guide a three-conductor, shielded cable, or three wires in conduit through the top conduit hub of the junction box. Use appropriate conduit fittings and construction technique for the environmental rating of the junction box. The junction box is rated NEMA 4X.
7. Pull out the controller terminal strip, and connect the three wires to the terminal strip as follows (see Figure 4).
Connect the positive wire to the PWR/SIG “+” terminal.
Connect the signal wire to the PWR/SIG “S” terminal.
Connect the negative wire to the PWR/SIG “-” terminal.
CAUTION: If shielded cable is used, leave the cable shield ’s drain wire disconnected and
insulated at the transmitter. You will connect the opposite end of the cable’s drain wire to the controller’s chassis (earth) ground.
8. Reinstall the controller terminal strip into its socket.
9. Secure the junction box cover to the junction box.
10. Route the cable or wires leading from the methane transmitter through one of the conduit hubs at the controller housing. Use appropriate conduit fittings and construction technique for the environmental rating of the controller. RKI controllers are typically rated NEMA 4X.
CAUTION: Do not route power and transmitter wiring through the same controller conduit hub.
The power cabl e may disrupt the transmission o f the transmitte r signal to the controller.
11. Connect the wires to the applicable detector/transmitter terminal strip at the controller as shown in Figure 4.
6 • 65-2399RK-CH4 Combustible Gas Transmitter
Page 11
Cable sheild
Amplifier
Black
+ 24 VDC 4 - 20 mA In (S)
- (DC Ground)
Controller
SPAN
ZERO
S
SIG/PWR
BGW
LEL
R
Green
Detector Wires
White
Red
J-Box
IR CH4 %Volume Detector
Figure 4: Wiring the Methane Transmitter to a Controller
12. If shielded cable is used, connect the cable’s drain wire to an available chassis (earth) ground at the controller. RKI controllers typically have a ground stud that can be used to ground the cable’s drain wire.
65-2399RK-CH4 Combus tible Ga s Tr a nsm itter • 7
Page 12
Start Up
This section describes procedures to start up the methane tran smitter and place the transmitter into normal operation.
Introducing Incoming Power
1. Complete the installation procedures described earlier in this manual.
2. Verify that the power wiring to the controller is correct and secure. Refer to the controller operator’s manual.
3. Turn on pow er to the controller.
4. Turn on the controller.
5. Verify that the controller is on and operating properly. Refer to the controller operator’s ma nual.
NOTE: When first powered up, the transmitter will enter about a one minute period
when the 4-20 mA output is stab ilizing and may be above the controller alarm points or well below zero momentarily. RKI controllers have a one minute warmup period when the controller does not display any gas reading or give any alarm indication. The methane transmitter’s 4-20 mA signal should be stable by the time the controller’s warmup period is over.
CAUTION: Allow the detector to warm up for 5 minutes before you continue with the next
section, “Setting the Zero Signal”.
Setting the Zero Signal
NOTE: If you can verify that the methane transmitter is in a fresh air environment, you
do not need to apply zero air to the detector before adjusting the zero reading.
The procedure below describes applying zero emission air, usually called zero air , using a calibration kit that includes a calibration cup, calibration gas, sample tubing, and a fixed flow regulator with an on/off knob. RKI Instruments, Inc. recommends using a 0.5 LPM (liters per minute) fixed flow regulator.
1. Unscrew and remove the junction box cover from the junction box.
2. Set a voltmeter to measure in the millivolt (mV) range.
3. Plug the voltmeter leads into the test points on the amplifier. Plug the positive lead into the red (+) test point; plug the negative lead into the black (-) test point.
4. Screw the calibration cup onto the bottom of the IR CH
5. Screw the regulator into the zero air calibration cylinder.
6. Use the calibration kit sample tubing to connect the regulator to the calibration cup.
7. Turn the regulator’s on/off knob counterclockwise to open it. Gas will begin to flow.
%Vol. detector.
4
8. Allow the gas to flow for one minute.
9. Verify a voltmeter reading of 100 mV (±2 mV).
10. If necessary, use a small flat-blade screwdriver to adjust the zero pot until th e
8 • 65-2399RK-CH4 Combustible Gas Transmitter
Page 13
11. Turn the regulator’s on/off knob clockwise to close it.
12. Unscrew the calibration cup from the detector.
13. Unscrew the regulator from the zero air calibration cylinder. For convenience, leave
14. Store the components of the calibration kit in a safe and conve nient place.
15. Remove the voltmeter leads from the test points.
16. Secure the junction box cover to the junction box.
Maintenance
This section describes maintenance procedures. It includes preventive maintenance, troubleshooting, and component replacement procedures.
Preventive Maintenance
This section describes a preventive maintenance schedule to ensure the optimum performance of the methane transmitter. It includes daily, monthly, and biannual procedures.
voltmeter reading is 100 mV (±2 mV).
the sample tubing connected to the regulator and the calibration cup.
Daily
Verify a display reading of 0 %Vol. at the controller. Investiga te significant changes in the display reading.
Monthly
This procedure describes a test to verify that the methane transmitter responds properly to the target gas.
WARN IN G: The controller is not an active gas monitoring device during the response
test procedure.
NOTE: Performing a response test on the methane transmitter may cause alarms. Be sure
to put the controller into its calibration mode or disable external alarms before performing this test.
NOTE: The following procedure assumes the use of a calibration kit which includes a
calibration gas cylinder, a 0.5 LPM fixed flow regulator with an on/off knob, a calibration cup for the detector, and a length of sample tubing.
Preparing for the response test
1. Place the controller into its calibration mode or disable external alarms.
2. Verify that the controller display reading for the channel you are testing is 0 %Vol. If the display reading is not 0 %Vol., set the zero reading of the transmitter as
described in “Start Up” on page 8, then continue this procedure.
3. Screw the regulator into the calibration cylinder.
4. Screw the calibration cup onto the bottom of the detector.
65-2399RK-CH4 Combus tible Ga s Tr a nsm itter • 9
Page 14
5. Use the calibration kit sample tubing to connect the regulator to the calibration cup.
6. Set a voltmeter to measure in the millivolt (mV) range.
7. Plug the voltmeter leads into the test points on the amplifier. Plug the positive lead into the red (+) test point; plug the negative lead into the black (-) test point.
8. Use the following formula to determine the correct test points output for the test sample.
Output (mV) = (calibrating sample/fullscale) X 400 + 100
For example, with a test sample of 50 %Vol. and a fullscale setting of 100 %Vol., the correct output is 300 mV.
300 (mV) = (50/100) X 400 +100
Performing the response test
1. Turn the regulator’s on/off knob counterclockwise to open the regulator. Gas will begin to flow.
2. Allow the gas to flow for one minute.
3. Verify that the reading is within ± 20% of the response reading you determined earlier.
NOTE: If the reading is not within ± 20% of the correct response reading, calibrate the
transmitter as described in “Calibration” on page 15 of this manual.
4. Turn the regulator’s on/off knob clockwise to close the regulator.
5. Unscrew the regulator from the calibration cylind er.
6. Unscrew the calibration cup from the detector.
NOTE: For convenience, leave the regulator and calibration cup connected by the
sample tubing.
7. Remove the voltmeter leads from the amplifier test points.
8. Reinstall the junction box cover.
9. When the controller display reading falls below th e alarm setpoints, return the controller to normal operation.
10. Store the components of the calibration kit in a sa fe place.
Biannually
Calibrate the methane transmitter as described in “Calibration” on page 15 of this manual.
10 • 65-2399RK-CH4 Combustible Gas Transmitter
Page 15
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended action for problems you may encounter with the methane trans mitter.
NOTE: This troubleshooting guide describes transmitter problems only. See the
controller operator’s manual for problems you may encounter with the controller.
Table 2:Troubleshooting the Methane Transmitter
Condition Symptom(s) Probable Causes Recommended Action
Fail Condition • Controller indicates a
Slow or No Response/ Difficult or Unable to Calibrate
fail condition.
• Transmitter responds slowly or does not respond to response test.
• Unable to accurately set the zero or response reading during calibration.
• Transmitter requires frequent calibration.
Note: Under “normal” circumstances, the transmitter requires calibration once every 6 months.
Some applications may require a more frequent calibration schedule.
• The transmitter wiring is disconnected or misconnected.
• The wiring from the detector to the amplifier is disconnected or misconnected.
• The plug-in sensor is not properly plugged into the 3 socket pattern in the detector housing body.
• The transmitter’s zero reading is low enough to cause a fail condition.
• The transmitter is malfunctioning.
• The calibration cylinder is low, out-dated, or defective.
• The calibration gas is not an appropriate concentration.
• The membrane on the detector housing cap is blocked with dirt or some other p a rticulate matter.
• The transmitter is malfunctioning.
1. Verify that the transmitter wiring is correct and secure.
2. V erify that th e wiring from the detector to the amplifier is correct and secure.
3. Confirm th at the plug-in sensor is properly installed.
4. Perform a zero adjustment. A full calibration is recommended.
5. If the fail condition continues, replace the sensor as described later in this section.
6. If the fail condition continues, contact RKI for further instruction.
1. Verify that the calibration cylinder contains an adequate supply of a fresh test sample.
2. Verify that the calibration gas concentration is appropriate for the transmitter.
3. Check the face of the detector housing cap and remove any particulate contamination from the hydrophobic membrane if necessary. If membrane appears saturated with contamination or damaged, replace the membrane as described in the next section.
4. If the calibration/response difficulties continue, replace the sensor as described later in this section.
5. If the calibration/response difficulties continue, contact RKI for further instruction.
65-2399RK-CH4 Combustible Gas Transmitter • 11
Page 16
Replacing Components of the Methane Transmitter
This section includes a procedure to replace the hydrophobic membrane, a procedure to replace the plug-in IR CH
detector assembly, and one to replace the amplifier. In most cases, it is not necessary to replace the entire detector assembly.
Replacing the Hydrophobic Membrane
1. Turn off the controller.
2. Turn off or unplug incoming power to the controller.
3. Unscrew the detector housing cap from the detector housing body.
4. Gently pry up the edge of the white hydrophobic mem b rane with a small flat blade screwdriver or a similar tool.
5. Peel off the hydrophobic membrane. It may be necessary to clean off the detector housing cap face to remove any residue left from the adhesive backed membrane.
6. Install the new membrane in t he recess on the face of the detector housing cap.
7. Make sure the cap gasket is in place and screw the detector housing cap back onto the detector housing body.
8. Turn on the pow e r to the controller.
%Vol. sensor, a procedure to replace the entire methane
4
9. Turn on the controller and place it into normal operation.
NOTE: When first powered up, the transmitter will enter about a one minute period
when the 4-20 mA output is stab ilizing and may be above the controller alarm points or well below zero momentarily. RKI controllers have a one minute warmup period when the controller does not display any gas reading or give any alarm indication. The methane transmitter’s 4-20 mA signal should be stable by the time the controller’s warmup period is over.
Replacing the Plug-In IR CH
%Volume Sensor
4
1. Turn off the controller.
2. Turn off or unplug power to the controller.
3. Unscrew the detector housing cap from the detector housing body. Make sure not to lose the cap gasket.
4. Unplug and remove the plug-in IR CH
%Volume sensor.
4
5. Carefully plug the replacement sensor into the socket pattern that is located in the detector housing.
6. Make sure the cap gasket is in place and screw the detector housing cap back onto the detector housing body.
7. Turn on pow er to the controller.
8. Turn on the controller and place it into normal operation.
12 • 65-2399RK-CH4 Combustible Gas Transmitter
Page 17
NOTE: When first powered up, the transmitter will enter about a one minute period
when the 4-20 mA output is stab ilizing and may be above the controller alarm points or well below zero momentarily. RKI controllers have a one minute warmup period when the controller does not display any gas reading or give any alarm indication. The methane transmitter’s 4-20 mA signal should be stable by the time the controller’s warmup period is over.
CAUTION: Allow the replacement sensor to warm up for 5 minutes before you continue with the
next step.
9. Calibrate the transmitter as described in “Calibration” on page 15.
Replacing the IR CH
%Volume Detector
4
1. Turn off the controller.
2. Turn off or unplug power to the controller.
3. Remove the junction box cover.
4. Remove the detector terminal strip from its socket.
5. Disconnect the detector leads from the detector terminal strip. Note the position of the color-coded leads as you remove them.
6. Unscrew the detector from the junction box.
7. Guide the detector leads of the replacement detector through the bottom conduit hub of the junction box, then screw the mounting threads of the detector into the conduit hub. If necessary for en vironm e ntal conditions, apply thread sealant or teflon tape to the hub and/or detector threads to seal them.
8. Connect the detector leads to the detector terminal strip as shown in Table 3 below and Figure 4 on page 7 of this manual.
Table 3:Reconnecting the Methane
Detector to the Amplifier
Detector Lead Detector Terminal Strip
Red %Vol. “R”
White %Vol. “W”
Green %Vol. “G”
Black %Vol. “B”
9. Reinstall the detector terminal strip into its socket.
10. Tu rn on or plug in power to the controller.
11. Turn on the controller and place it into normal operation.
65-2399RK-CH4 Combustible Gas Transmitter • 13
Page 18
NOTE: When first powered up, the transmitter will enter about a one minute period
when the 4-20 mA output is stab ilizing and may be above the controller alarm points or well below zero momentarily. RKI controllers have a one minute warmup period when the controller does not display any gas reading or give any alarm indication. The methane transmitter’s 4-20 mA signal should be stable by the time the controller’s warmup period is over.
CAUTION: Allow the replacement detector to warm up for 5 minutes before you co ntinue with
the next step.
12. Calibrate the transmitter as described in “Calibration” on page 15 of this manual.
Replacing the Amplifier
1. Turn off the controller.
2. Turn off or unplug power to the controller.
3. Remove the junction box cover.
4. Unplug the detector terminal strip and controller terminal strip from their sockets. You may leave the wires connected to the terminal strips.
5. Unscrew and remove the screw with the flat and lock washers that secures the amplifier to the junction box.
6. Remove the old amplifier.
7. Place the new amplifier in the same position as the old amplifier. A foam gasket that orients the amplifier and keeps it from rotating is installed on the bottom of the amplifier. Make sure the amplifier is seated flat in the junction box.
8. Install the new amplifier in to the junction box with the screw, lock washer, and flat washer you removed in Step 5 above.
9. Install the detector and controller terminals strips into their sockets on the new amplifier as shown in Figure 4 on page 7 of this manual. If controller leads or detector leads were removed d uring this procedure, refer to Table 4 and Table 5 b elow.
Table 4:Reconnecting the Amplifier
to the Controller
Amplifier Controller
Te rminal Strip
PWR/SIG “-” - (DC -) PWR/SIG “S” S (4 - 20 mA In) PWR/SIG “+” + 24V
Table 5:Reconnecting the CH4 %Vol.
Detector to the Amplifier
Controller
Transm itte r Terminal
Strip (typical)
Amplifier Detector
T erminal Strip
DETECTOR “R” RED
14 • 65-2399RK-CH4 Combustible Gas Transmitter
Detector Lead
Page 19
Table 5:Reconnecting the CH4 %Vol.
Detector to the Amplifier
Amplifier Detector
T erminal Strip
DETECTOR “W” WHT
DETECTOR “G” GREEN DETECTOR “B” BLK
Detector Lead
NOTE: When a transmitter is first powered up with a new amplifier, the initial output
may be either high or below zero depending on the setting of the zero pot. Be sure to make arrangements so that this does not cause unwanted alarms.
10. Turn on pow er to the controller.
11. Turn on the controller and place it into normal operation.
NOTE: When first powered up, the transmitter will enter about a one minute period
when the 4-20 mA output is stab ilizing and may be above the controller alarm points or well below zero momentarily. RKI controllers have a one minute warmup period when the controller does not display any gas reading or give any alarm indication. The methane transmitter’s 4-20 mA signal should be stable by the time the controller’s warmup period is over.
12. Allow the transmitter to warm-up for 5 minutes.
13. Calibrate the transmitter as described in “Calibration” on page 15 of this manual.
Calibration Frequency
Although there is no particular calibration frequency that is correct for all applications, a calibration fre quency of every 6 months is adequat e for most infrared meth ane transmitte r applications. Unless experience in a particular application dictates otherwise, RKI Instruments, Inc. recommends a calibration frequency of every 6 months.
If an application is not very demanding, for example detection in a clean, temperature controlled environment where calibration adjustments are minimal at calibration, then a calibration frequency of every 9 to 12 months is adequate.
If an application is very demanding, for example if the environment is not well controlled, then more frequent calibration than every 6 months may be necessary.
Calibration
This section describes how to calibrate the methane transmitter. It includes procedures to prepare for calibration, set the zero reading, set the response rea d ing, and return to normal operation.
W ARNING: The controller is not an active gas monitoring device during the calibration
procedure.
65-2399RK-CH4 Combustible Gas Transmitter • 15
Page 20
NOTE: The following procedure assumes the use of a calibration kit which includes a
calibration gas cylinder, a 0.5 LPM fixed flow regulator with an on/off knob, a calibration cup for the detector, and a short piece of sample tubing to connect the regulator to the calibration cup.
Prepari ng for Ca libration
1. Unscrew and remove the junction box cover.
2. Set a voltmeter to measure in the millivolt (mV) range.
3. Plug the voltmeter leads into the test points on the amplifier. Plug the positive lead into the red (+) test point; plug the negative lead into the black (-) test point.
4. Use the following formula to determine the correct test points output for the calibrating sample.
Output (mV) = (calibrating sample/fullscale) X 400 + 100
For example, with a calibrating sample of 50 %Vol. and a fullscale setting of 100 %Vol., the correct output is 300 mV.
300(mV) = (50/100) X 400 +100
5. Screw the calibration cup onto the detector housing.
6. Place the controller into its calibration mode or disable external alarms.
NOTE: Calibrating the methane transmitter may cause alarms. Be sure to put the
controller into its calibration mode or disable external alarms before continuing.
Setting the Zero Reading
NOTE: If you can verify that the methane transmitter is in a fresh air environment, you
do not need to apply zero air to the detector before adjusting the zero reading.
1. Screw the regulator into the zero air calibration cylinder.
2. Use the calibration kit sample tubing to connect the regulator to the calibration cup.
3. Turn the regulator’s on/off knob counterclockwise to open the regulator.
4. Allow the gas to flow for one minute and verify a reading of 100 mV (±2 mV). If necessary, use the zero pot on the amplifier to adjust the reading to 100 mV (±2 mV).
5. Turn the regulator’s on/off knob clockwise to close the regulator.
6. Unscrew the regulator from the zero air calibration cylinder.
7. Leave the sample tubing connected to the regulator and the calibration cup.
Setting the Response Reading
1. Screw the regulator into the calibration cylinder. Verify that the calibration gas is representative of the transmitter’s target gas.
2. Turn the regulator’s on/off knob counterclockwise to open the regulator.
3. Allow the calibration gas to flow for one minute and verify that the reading matches the response reading (±2 mV) you determined ea rlier. If necessary , use the span pot on the amplifier to adjust the reading to match the correct response reading.
16 • 65-2399RK-CH4 Combustible Gas Transmitter
Page 21
Parts List
4. Turn the regulator’s on/off knob clockwise to close the regulator.
5. Unscrew the regulator from the calibration cylind er.
Returning to Normal Operation
1. Remove the voltmeter leads from the amplifier test points.
2. Unscrew the calibration cup from the detector.
NOTE: For convenience, leave the regulator and calibration cup connected by the
sample tubing.
3. Secure the junction box cover to the junction box.
4. When the controller display reading falls below th e alarm setpoints, return the controller to normal operation.
5. Verify that the co ntroller display reading decreases and stabilizes at 0 % Vol.
6. Store the components of the calibration kit in a safe and convenient place.
Table 6 lists replacement parts and accessories for the methane transmitter.
Table 6:Parts List
Part Number Description
06-1248RK Sample tubing (3/16 in. x 5/16 in.; specify length when ordering) 18-0400RK-01 Junction box with spacers 33-0157RK Hydrophobic disk membr ane for detect or cap 57-1053RK Amplifier with gasket (specify detector part number when ordering) 61-0199RK-CH4 Infrared CH
61-5041RK-CH4 Infrared CH 65-2399RK-CH4 Methane transmitter (includes detector, junction box, and amplifier)
71-0201RK 81-0013RK-01 Steel calibration cylinder, 50% Volume methane in nitrogen, 34-liter 81-0076RK-01 Steel calibration cylinder, zero emission air, 34-liter 81-1050RK Regulator, 0.5 liter/minute; with pressure gauge and flow control knob, for 17 liter and
81-1103RK Calibration cup
65-2399RK-CH4 Methane Transmitter Operator’s Manual (this documen t)
34 liter steel calibration cylinders
%Vol. detector
4
replacement plug-in sensor
4
82-0006RK Pot adjust screwdriver, for calibration
65-2399RK-CH4 Combustible Gas Transmitter • 17
Loading...