RKI Instruments 65-2390RK-HC, 65-2390RK-CH4 User manual

Page 1
65-2390RK
Combustible Gas Transmitter
Operator’s Manual
Part Number: 71-0119RK Revision: A Released: 4/12/11
www.GlobalTestSupply.com
Page 2
WARNING
Read and understand this instruction manual before operating transmitter. Improper use of the transmitter could result in bodily harm or death.
Periodic calibration and maintenance of the transmitter is essential for proper operation and correct readings. Please calibrate and maintain this tra nsmitt er regula rly! Frequency of calibration depends upon the type of use you have and the sensor types. Typical calibration frequencies for most applications are between 6 and 12 months, but can be required more often or less often based on your usage.
65-2390RK Combustible Gas Transmitter
www.GlobalTestSupply.com
Page 3
Product Warranty
RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year fr o m date of shipment from RKI Instruments, Inc. Any parts found defective withi n tha t period will be repaired or replaced, at our option, free of charge. This warranty does not apply to those items which by their nature are subject to deterioration or consumption in normal service, and which must be cleaned, repaired, or replaced on a routine basis. Examples of such items are:
W arranty is voided by abuse including mechanical damage, alteration, rough handling, or repair procedures not in accordance with the operator’s manual. This warranty indicates the full extent of our liability , a nd we are not r esponsible for removal or r eplacement costs, local repair costs, transportation costs, or contingent expenses incurred without our prior approval.
a) Absorbent cartridges d) Batteries b) Pump diaphragms and valves e) Filter elements c) Fuses
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI INSTRUMENTS, INC. INCLUDING BUT NOT LIMITED TO, THE WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL RKI INSTRUMENTS, INC. BE LIABLE FOR INDIRECT, INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
This warranty covers instruments and parts sold to users by authorized distributors, dealers, and representatives as appointed by RKI Instruments, Inc.
We do not assum e i ndemnification for any accident or damage caused by the op e r a t ion of this gas monitor, and our warranty is limited to the replacement of parts or our complete goods.
65-2390RK Combustible Ga s Tr ans mitte r
www.GlobalTestSupply.com
Page 4
Table of Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
IR LEL Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Junction Box. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Mounting the Combustible Gas Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Wiring the Combustible Gas Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Startup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Introducing Incoming Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Setting the Zero Signal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Replacing Components of the Combustible Gas Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Calibration Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Preparing for Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Setting the Zero Reading. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Setting the Response Reading. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Returning to Normal Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
65-2390RK Combustible Gas Transmitter
www.GlobalTestSupply.com
Page 5
Overview
This manual describes the 65-2390RK combustible gas transmitter. This manual also describes how to install, start up, con fi gure, mainta in, a nd ca librate the transmitter when it is used with a gas monitoring controller. A parts list at the end of this manual lists replacement parts and accessories for the combustible gas transmitter.
Specifications
Table 1 lists specifications for th e combustible gas transmitter.
Table 1:Specifications
Target/Calibration Gas 65-2390RK-CH4: Methane (CH
65-2390RK-HC: Propane
Area Classification Explosionproof for Class I, Groups B, C, and D Sampling Method Diffusion Signal Output 4 to 20 mA Detection Range 0 to 100% LEL Response Time 90% in 45 seconds Accuracy ± 5% of reading or ± 2% of full scale (which ever is greater)
)
4
W ARNING: When using the 65-2390RK, you must follow the instructions and warnings
in this manual to assure proper and safe operation of the 65-2390RK and to minimize the risk of personal injury. Be sure to maintain and period ically calibrate the 65-2390RK as described in this manual.
65-2390RK Combustible Gas Transm itter • 1
www.GlobalTestSupply.com
Page 6
Description
Interconnect Terminal Strip
The 65-2390RK combustible gas transmitter has two versions, the 65-2390RK-CH4 which is calibrated to methane and the 65-2390RK-HC which is calibrated to propane. The transmitter utilizes an infrared type of detector which has some a dvantages over a catalytic type of combustible detector. The infrared detector will generally have a longer service life than a catalytic detector, it will require calibration less often, and it can detect combustible gas even if there is no oxygen in the sample which allows detection of combustible gas in an inert atmosphere.
This section describes the components of the combustible gas transmitter. The transmitter is a 4 - 20 mA type detector head. It consists of the infrared LEL detector, amplifier, and junction box.
FactorySet Pots
J-Box
100-500 mV Range
Infrared LEL Detector
The infrared LEL detector is made up of a miniature infrared combustible gas LEL detector housed and encapsulated in a pipe nipple. The pipe nipple has 3/4 inch NPT threads on each end and a 1 1/4 inch hex that allows removal or installation of the detector with a wrench. A porous flame arrestor that is coated with a hydrophobic film that repels liquids is on one end of the detector and allows sample gas to enter the detector . Four color coded leads, r ed, white, green , and black, extend fr om the other end of the detector. The leads allow you to connect the detector to the amplifier.
Amplif ier
GND 24V 4/20
POWER/SIGSENSOR
TP +
RED WHT GRN BLK
TP -
Test Points
Jumper Pins for
Factory Use Only
Span Pot
Zero Pot
IR LEL Detector
Figure 1: Combustible Gas Transmitter Component Location
To distinguish the propane detector from the methane detector (in case the replacement sensor label that is applied to one of the leads is lost), a short length of red shrink tubing is
2 • 65-2390RK Combustible Gas Transmitter
www.GlobalTestSupply.com
Page 7
applied to the white wire of the propane detector near where the wire comes out of the nipple.
Amplifier
The amplifier converts the electrical output from the detector to a 4 to 20 mA signal that corresponds to the detection range and transmits the signal to a gas monitoring controller. The amplifier includes the interconnect terminal strip, span potentiometer, zero potentiometer, and test points (see Figure 1).
Interconnect Terminal Strip
The interconnect terminal strip is a seven-point terminal strip. Use the interconnect terminal strip to connect the IR LEL detector to the amplifier and the amplifier to a controller.
NOTE: The IR LEL detector is factory-wired to the amplifier. See the Installation section
of this manual for all wiring procedures related to the transmitter.
Span Potentiomete r
The span potentiometer is near the bottom of the amplifier (see Figure 1). Use the span potentiometer to adjust the transmitter’s response output during the calibration procedure. Turn the adjustment screw clockwise to increase the gas response and counterclockwise to decrease the gas response.
Zero Potentiometer
The zero potentiometer is to the right of the span potentiometer (see Figure 1). Use the zero potentiometer to adjust the transmitter’s target gas-free output during the start-up and calibration procedures. Turn the adjustment screw clockwise to increase the zero output and counterclockwise to decrease the zero output.
CAUTION: The amplifier includes two additional potentiometers. They are factory-set. Do not
adjust them.
Test Points
The test points (labeled TP+ and TP-) are to the left of the interconnect terminal strip (see Figure 1). The test points produce a 100 to 500 mV output that corresponds to the transmitter’s 4 to 20 mA output. Use the test points and a voltmeter to measure the transmitter’s output during the start-up and calibration procedures.
Junction Box
Use the junction box to install the comb ustible gas transmitter at a mounting site that is remote from the controller. The junction box also protects th e amplifi er and wiring connections made to the amplifier. Use the two 3/4 in. conduit hubs to mount the detector to the junction box (bottom hub) and connect wiring from the amplifier to the controller (top hub).
NOTE: The detector and amplifier are factory-mounted to the junction box.
Use the junction box’s two mountin g holes to mount the combustible gas transmitter to a vertical surface at the monitoring site. Use the cover on the front of the junction box to access the interior of the junction box.
65-2390RK Combustible Gas Transm itter • 3
www.GlobalTestSupply.com
Page 8
Installation
)
This section describes procedures to mount the combustibl e gas transmitter in the monitoring environment and wire the transmitter to a controller.
Mounting the Combustible Gas Transmitter
1. Select a mounting site that is representative of the monitoring environment. Consider the following when you select the mounting site.
Select a site where the transmitter is not likely to be bumped or disturbed. Make
sure there is sufficient room to perform start-up, maintenance, and calibration procedures.
Select a site where the target gas is likely to be found first. For lighter gases,
mount the detector near the ceiling; for heavier gases, mount the detector near the floor.
6.1
1.1
3.94
5.2
2.75
5.46
IR LEL Detector
Figure 2: Mounting the Combustible Gas Transmitter
Ø.25MoutingHoles (2X
3/4" NPT Conduit Hub
1.9 max
7.3 max
If the detector is mounted to the junction box, skip to step 5. If not, continue with step 2.
2. Remove the junction box cover.
3. Guide the four wires that extend from the top of the detector through the bottom conduit hub of the junction box.
4. Screw the detector into the bottom conduit hub of the junction box.
5. At the monitoring site, use #10 screws through the junction box’s two mounting holes to secure the junction box to a vertical surface.
4 • 65-2390RK Combustible Gas Transmitter
www.GlobalTestSupply.com
Page 9
CAUTION: Mount the combustible gas trans mitte r wi th th e d etector f aci ng down (see F igure 2.)
Wiring the Combustible Gas Transmitter to a Controller
WARNING : Always verify that the power to the controller is off before you make
wiring connections.
1. Turn off power to the controller.
2. Place the controller’s power switch in the OFF position.
3. Remove the junction box cover.
4. Verify that the detector leads are wired to the amplifier’s interconnect terminal strip. If necessary, connect the detector leads to the interconnect terminal strip as shown in Figure 3.
5. Guide a three-conductor, shield ed cable or three wires in conduit through the top conduit hub of the junction box.
6. Connect the three wires to the interconnect terminal strip as follows (see Figure 3).
Connect the positive wire to the 24VDC terminal.
Connect the feedback wire to the 4-20 (FB) terminal.
Connect the negative wire to the GND (DC -) terminal.
CAUTION: If shielded cable is used, leave the cable s hie ld’s drain wire insulate d and
disconnected at the transmitter. You will connect the opposite end of the cable’s drain wire at the controller to chassis ground.
7. Secure the junction box cover to the junction box.
8. Route the cable or wires leading from the combustible gas transmitter through one of the conduit hubs at the controller housing.
CAUTION: Do not route power and transmitter wiring through the same conduit hub. The
power cable may disrupt the transmission of the transmitter signal to the controller.
9. Connect the wires to the applicable transmitter terminal strip at the controller as shown in Figure 3 below.
65-2390RK Combustible Gas Transm itter • 5
www.GlobalTestSupply.com
Page 10
J-Box
e
Cable S hi eld
- (DC Ground) + 24 V DC
4 - 20 mA In (FB)
Amplifi er
GND 24V 4/20
POWER/SIGSENSOR RED WHT GRN BLK
Controller or
Recor di ng Dev i c
Red White
IR LEL Detector
Black
Green
Figure 3: Wiring the Combustible Gas Transmitter to a Controller
10. If shielded cable is used, connect the cable’s drain wire to an available chassis ground at the controller.
6 • 65-2390RK Combustible Gas Transmitter
www.GlobalTestSupply.com
Page 11
Start Up
This section describes procedures to start up the combustible gas transmitter and place th e transmitter into normal operation.
Introducing Incoming Power
1. Complete the installation procedures described earlier in this manual.
2. Verify that the power wiring to the controller is correct and secure. Refer to the controller operator’s manual.
3. Turn on or plug in power to the controller, then place the contr oller’s power switch in the ON position.
4. Verify that the controller is on and operating properly. Refer to the controller operator’s manu al.
NOTE: When first powered up, the transmitter will enter about a one minute period
when the 4-20 mA output is stab ilizing and may be above the controller alarm points or well below zero momentarily. RKI controllers have a one minute warmup period when the controller does not display any gas reading or give any alarm indication. The combustible gas transmitter’s 4-20 mA signal should be stable by the time the controller’s warmup period is over.
CAUTION: Allow the detector to warm up for 5 minutes before you continue with the next
section, “Setting the Zero Signal”.
Setting the Zero Signal
CAUTION: If you suspect the presence of combustible gas in the monitoring en vironment, use
the calibration kit and the zero air calibration cylinder to introduce “fresh air” to the detector and verify an accurate zero setting.
1. Verify that the transmitter is in a fresh air environment (environment known to be free of combustible and toxic gases and of normal oxygen content, 20.9%).
2. Unscrew and remove the junction box cover from the junction box.
3. Set a voltmeter to measure in the millivolt (mV) range.
4. Plug the voltmeter leads into the test points on the amplifier. Plug the positive lead into the test point labeled TP+; plug the ne gative lead into the test point labeled TP-.
5. Verify a voltmeter reading of 100 mV (±2 mV).
6. If necessary, use a flat-blade screwdriver to adjust the zero potentiometer until the voltmeter reading is 100 mV (±2 mV).
7. Secure the junction box cover to the junction box.
65-2390RK Combustible Gas Transm itter • 7
www.GlobalTestSupply.com
Page 12
Maintenance
This section describes maintenance procedures. It includes preventive maintenance, troubleshooting, and component replacement procedures.
Preventive Maintenance
This section describes a preventive maintenance schedule to ensure the optimum performance of the combustible gas transmitter. It includes daily, monthly, and biannua l procedures.
Daily
Verify a display reading of 0 %LEL at the controller. Investigate significant changes in the display reading.
Monthly
This procedure describes a test to verify that the combustible gas transmitter responds properly to the target gas.
WARNING : The controller is not an active gas monitoring device during the response
test procedure.
NOTE: Performing a response test on the combustible transmitter may cause alarms. Be
sure to put the controller into its calibration program or disable external alarms before performing this test.
NOTE: The following procedure assumes the use of a calibration kit which includes a
calibration gas cylinder, a 0.5 LPM fixed flow regulator with an on/off knob, a calibration cup for the detector, and a short piece of sample tubing to connect the regulator to the calibration cup.
Preparing for the response test
1. Place the controller into its calibration program or disable external alarms.
2. Verify that the controller display reading for the cha nnel you are testing is 0 %LEL. If the display reading is not zero, set the zero reading of the transmitter as described
in “Start Up” on page 7 of this manual, then continue this procedure.
3. Screw the calibration cup onto the bottom of the combustible detector.
4. Use the calibration kit sample tubing to connect the regulator to the calibration cup.
5. Set a voltmeter to measure in the millivolt (mV) range.
6. Remove the junction box cover, then plug the voltmeter leads into the test points on the amplifier.
Plug the positive lead into the test point labeled TP+; plug the negative lead into the test point labeled TP-.
8 • 65-2390RK Combustible Gas Transmitter
www.GlobalTestSupply.com
Page 13
7. Use the following formula to determine the correct test points output for the test sample.
Output (mV) = (calibrating sample/fullscale) X 400 + 100
For example, with a test sample of 50% LEL and a fullscale setting of 100% LEL, the correct output is 300 mV.
300 (mV) = (50/100) X 400 +100
Performing the response test
1. Screw the regulator into the calibration cylinder.
2. Turn the regulator knob coun terclockwise to open the regulator.
3. Allow the gas to flow for one minute.
4. Verify that the reading is within ± 20% of the response reading you determined earlier.
NOTE: If the reading is not within ± 20% of the correct response reading, calibrate the
transmitter as described in “Calibration” on page 13 of this manual.
5. Turn the regulator knob clockwise to close the regulator.
6. Unscrew the regulator from the calibration cylind er.
7. Unscrew the calibration cup from the detector.
8. Remove the voltmeter leads from the amplifier test poin ts.
9. Reinstall the junction box cov er.
10. When the controller display reading falls below the alarm setpoints, return the controller to normal operation.
11. Store the components of the calibration kit in a safe place.
Biannually
Calibrate the combustible gas transmitter as described in “Calibration” on page 13 of this manual.
65-2390RK Combustible Gas Transm itter • 9
www.GlobalTestSupply.com
Page 14
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended action for problems you may encounter with the combustible gas transmitter.
NOTE: This troubleshooting guide describes transmitter problems only. See the
controller operator’s manual for problems you may encounter with the controller.
Table 2:Troubleshooting the Combustible Gas Transmitter
Condition Symptom(s) Probable Causes Recommended Action
Fail Condition • Controller indicates a
Slow or No Response/ Difficult or Unable to Calibrate
fail condition.
• Transmitter responds slowly or does not respond to response test.
• Unable to accurately set the zero or response reading during calibration.
• Transmitter requires frequent calibration.
Note: Under “normal” circumstances, the transmitter requires calibration once every six months.
Some applications may require a more frequent calibration schedule.
• The transmitter wiring is disconnected or misconnected.
• The transmitter’s zero reading is low enough to cause a fail condition.
• The transmitter is malfunctioning.
• The calibration cylinder is low, out-dated, or defective.
• The transmitter is malfunctioning.
1. Verify that the transmitter wiring is correct and secure.
2. Calibrate the transm itt er.
3. If the fail condition continues, replace the detector.
4. If the fail condition continues, contact RKI for further instruction.
1. Verify that the calibration cylinder contains an adequate supply of a fresh test sample.
2. Verify that the regulator used for calibration is a 0.5 LPM regulator.
3. If the calibration/response difficulties continue, replace the detector.
4. If the calibration/response difficulties continue, contact RKI for further instruction.
10 • 65-2390RK Combustible Gas Transmitter
www.GlobalTestSupply.com
Page 15
Replacing Components of the Combustible Gas Transmitter
This section includes procedures to replace the IR LEL detector and amplifier.
Replacing the IR LEL Detector
1. Turn off power to the controller.
2. Place the controller’s power switch in the OFF position.
3. Remove the junction box cover.
4. Disconnect the detector leads from the interconnect terminal strip. Note the position of the color-coded leads as you remove them.
5. Unscrew the detector from the junction box.
6. Guide the detector leads of the replacement detector through the bottom conduit hub of the junction box, then screw the mounting threads of the detector into the conduit hub.
7. Connect the detector leads to the interconnect terminal strip as shown in Table 3 below and Figure 3 on page 6 of this manual.
Table 3: Reconnecting the Combustible Gas
Detector to the Amplifier
Detector Lead
Red RED
White WHT
Green GRN
Black BLK
Amplifier Interconnect
Terminal Strip
8. Turn on power to the controller.
9. Place the controller’s power switch in the ON position .
NOTE: When first powered up, the transmitter will enter about a one minute period
when the 4-20 mA output is stab ilizing and may be above the controller alarm points or well below zero momentarily. RKI controllers have a one minute warmup period when the controller does not display any gas reading or give any alarm indication. The combustible gas transmitter’s 4-20 mA signal should be stable by the time the controller’s warmup period is over.
CAUTION: Allow the replacement detector to warm up for 5 minutes befor e y o u c ontinue with
the next step.
10. Calibrate the replacement detector as described in “Calibration” on page 13 of this manual.
Replacing the Amplifier
1. Turn off power to the controller
2. Place the controller’s power switch in the OFF position.
3. Remove the junction box cover.
65-2390RK Combustible Gas Transm itter • 11
www.GlobalTestSupply.com
Page 16
4. Disconnect the detector leads from the inte rconne c t terminal strip.
5. Disconnect the wiring that connects the combustible gas transmitter to the controller from the amplifier’s interconnect terminal strip.
6. Unscrew and remove the two screws that secure the amplifier to the junction box. The screws are at the top left and bottom right of the amplifier.
7. Remove the amplifier.
8. Place the new amplifier in the same position as the amplifier you removed in the previous step.
9. Use the two screws you removed in step 6 to secure the amplifier to the junction box.
10. Reconnect the wiring that connects the controller to the combustible gas transmitter at the amplifier’s interconnect terminal strip as shown in Table 4 below and Figure 3 o n page 6 of this manual.
Table 4: Reconnecting the Combustible Gas Amplifier
to the Controller
Amplifier Interconnect
T erminal Strip
GND - (DC -)
4-20 4 - 20 mA (FB or S) 24V + 24V
Controller
Transmitter Terminal
Strip (typical)
11. Reconnect the detector leads to the amplifier’s interconnect terminal strip as shown in Table 5 below and Figure 3 on page 6 of this manual.
Table 5: Reconnecting the Combustible Gas
Detector to the Amplifier
Detector Lead
Red RED
White WHT
Green GRN
Black BLK
Amplifier Interconnect
Terminal Strip
12. Turn on power to the controller.
13. Turn on the controller and place it into normal operation.
NOTE: When first powered up, the transmitter will enter about a one minute period
when the 4-20 mA output is stab ilizing and may be above the controller alarm points or well below zero momentarily. RKI controllers have a one minute warmup period when the controller does not display any gas reading or give any alarm indication. The combustible gas transmitter’s 4-20 mA signal should be stable by the time the controller’s warmup period is over.
12 • 65-2390RK Combustible Gas Transmitter
www.GlobalTestSupply.com
Page 17
CAUTION: Allow the detector to warm up for 5 minutes before you continue with the next step.
14. Calibrate the combustible gas transmitter as de scribed in “Calibration” on page 13 of this manual.
Calibration Frequency
Although there is no particular calibration frequency that is correct for all applications, a calibration frequency of every 6 months is adequate for most infrared combustible gas transmitter applications. Unless experience in a particular application dictates otherwise, RKI Instruments, Inc. recommends a calibration frequency of every 6 months for the infrared combustible gas transmitter.
If an application is not very demanding, for example detection in a clean, temperature controlled environment where the toxic or co mbustible tar get gas is not normally pr esen t, and calibrati on adjustments are mini mal at calibration, then a calibration frequ ency of every 9 to 12 months is adequate.
If an application is very demanding, for example if the environment is not well controlled, then more frequent calibration than every 6 months may be necessary.
Calibration
This section describes how to calibrate the combustible gas transmitter. It includes procedures to prepare for calibration, set the zero reading, set the response reading, and return to normal operation.
W ARNING: The controller is not an active gas monitoring device during the calibration
procedure.
NOTE: The following procedure assumes the use of a calibration kit which includes a
calibration gas cylinder, a 0.5 LPM fixed flow regulator with an on/off knob, a calibration cup for the detector, and a short piece of sample tubing to connect the regulator to the calibration cup.
Prepari ng for Ca libration
1. Screw the calibration cup onto the bottom of the combustible gas detector.
2. Use the sample tubing that comes with the calibration kit to connect the regulator to the calibration cup.
3. Set a voltmeter to measure in the millivolt (mV) range.
4. Remove the junction box cover, then plug the voltmeter leads into the test points on the amplifier.
Plug the positive lead into the test point labeled TP+; plug the negative lead into the test point labeled TP-.
65-2390RK Combustible Gas Transm itter • 13
www.GlobalTestSupply.com
Page 18
5. Use the following formula to determine the correct test points output for the calibrating sample.
Output (mV) = (calibrating sample/fullscale) X 400 + 100
For example, with a calibrating sample of 50% LEL and a fullscale setting of 100% LEL, the correct output is 300 mV.
300 (mV) = (50/100) X 400 +100
6. Place the controller into its calibration program or disable external alarms.
NOTE: Calibrating the combustible transmitter may cause alarms . Be sure to put the
controller into its calibration program or disable external alarms before continuing.
Setting the Zero Reading
CAUTION: If you can verify that the combustible transmitter is in a fresh air environment, you
do not need to apply zero air to the detector before adjusting the zero reading.
1. Screw the regulator into the zero air calibration cylinder.
2. Turn the regulator knob coun terclockwise to open the regulator.
3. Allow the gas to flow for one minute and verify a reading of 100 mV (±2 mV). If necessary, use the zero potentiometer on the amplifier to adjust the reading to 100 mV (±2 mV).
4. Turn the regulator knob clockwise to close the regulator.
5. Unscrew the regulator from the zero air calibration cylinder. Leave the sample tubing connected to the regulator and the calibration cup.
Setting the Response Reading
1. Screw the regulator into the calibration cylinder. Verify that the calibration gas is representative of the transmitter’s target gas.
2. Turn the regulator knob coun terclockwise to open the regulator.
3. Allow the calibration gas to flow for one minute and verify that the reading matches the response reading (±2 mV) you determined earlier. If necessary, use the span potentiometer on the amplifier to adjust the reading to match the correct response reading.
4. Turn the regulator knob clockwise to close the regulator.
5. Unscrew the regulator from the calibration cylind er.
Returning to Normal Operation
1. Remove the voltmeter leads from the amplifier test poin ts.
2. Unscrew the calibration cup from the detector.
NOTE: For convenience, leave the regulator and calibration cup connected by the
sample tubing.
3. Secure the junction box cover to the junction box.
4. When the display reading falls below the ala r m setpoints, return the controller to
14 • 65-2390RK Combustible Gas Transmitter
www.GlobalTestSupply.com
Page 19
Parts List
normal operation.
5. Verify that the controller display reading decreases and stabilizes at 0%LEL.
6. Store the components of the calibration kit in a safe and convenient place.
Table 6 lists replacement parts and accessories for the combustible gas transmitter.
Table 6:Parts List
Part Number Description
06-1248RK Sample tubing (3/16 in. x 5/16 in.; specify length when ordering) 18-0405RK-01 Junction box (without cover; pre-drilled for amplifier) 18-0406RK Junction box cover (cover only) 57-1050RK Amplifier (specify detector part number when ordering) 61-0190RK-CH4 Methane infrared LEL detector 61-0190RK-HC HC infrared LEL detector (calibrated to propane) 65-2390RK-CH4 Methane transmitter (includes detector, junction box, and amplifier) 65-2390RK-HC Propane transmitter (includes detector, junction box and amplifier) 71-0119RK 65-2390RK Combustible Gas Transmitter Operator’s Manual (this document) 81-0004RK-01 Calibration cylinder (50% LEL propane; 34-liter) 81-0012RK-01 Calibration cylinder (50% LEL Methane; 34-liter) 81-0076RK-01 Zero air calibration cylinder (34-lite r ) 81-1050RK Regulator , 0 .5 liter/minute, with gauge and kn ob, for 17- and 34-liter calibration
cylinders 81-1103RK Calibration cup 82-0006RK Pot adjust screwdriver, for calibration
65-2390RK Combustible Gas Transm itter • 15
www.GlobalTestSupply.com
Loading...