Part Number: 71-0145RK
Revision: P1
Released: 7/10/07
www.rkiinstruments.com
Page 2
Product Warranty
RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in
materials, workmanship, and performance for a period of one year fr o m date of shipment
from RKI Instruments, Inc. Any parts found defective withi n tha t period will be repaired
or replaced, at our option, free of charge. This warranty does not apply to those items
which by their nature are subject to deterioration or consumption in normal ser v ice, and
which must be cleaned, repaired, or replaced on a routine basis. Examples of such items
are:
W arranty is voided by abuse including mechanical damage, alteration, rough handling, or
repair procedures not in accordance with the operator’s manual. This warranty indicates
the full extent of our liability , a nd we are not r esponsible for removal or r eplacement costs,
local repair costs, transportation costs, or contingent expenses incurred without our prior
approval.
a) Absorbent cartridgesd) Batteries
b) Pump diaphragms and valvese) Filter elements
c) Fuses
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER
WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED,
AND ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF
RKI INSTRUMENTS, INC., INCLUDING BUT NOT LIMITED TO, THE
WARRANTY OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE. IN NO EVENT SHALL RKI INSTRUMENTS,
INC., BE LIABLE FOR INDIRECT, INCIDENTAL, OR CONSEQUENTIAL
LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF
ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR
OPERATE PROPERLY.
This warranty covers instruments and parts sold to users by authorized distributors,
dealers, and representatives as appointed by RKI Instruments, Inc.
We do n ot assume inde mn ification f o r any accide nt o r d amage caused by the ope ration of
this gas monitor, and our warranty is limited to the replacement of parts or our complete
goods.
61-1004RK/61-0191RK Comb us t i bl e G a s De t e ct or • 3
Page 4
Overview
This manual describes the 65-1004RK CO2 (carbon dioxide) detector. This manual also
describes how to install, start up, maintain, and calibrate the detector when used with a
gas monitoring controller. A parts list at the end of this manual lists replacement parts and
accessories for the CO
detector.
2
The 61-1004RK CO
the 61-0191RK CO
detector includes a junction box. This manual may also be used for
2
detector which does not include a junction box and is normally
2
mounted in one of a controller’s conduit hu bs and factory wired to the controller. If you
are using a 61-0191RK CO
detector, disregard all references to the junction box and
2
junction box terminal block.
Specifications
Table 1 lists specifications for the CO2 detector.
Table 1: 61-1004RK Specifications
Detection RangeWith Junction Box
61-1004RK-02: 0 - 5,000 ppm CO
61-1004R K --03: 0 - 5% volume CO
Without Junction Box
61-0191RK-02: 0 - 5,000 ppm CO
61-0191RK-03: 0 - 5% volume CO
Area ClassificationExplosionproof for Class I, Groups B, C, and D
Sampling MethodDiffusion
2
2
2
2
Response Time90% in 45 seconds
4 • 61-1004RK/61-0191RK Combust ible Gas Detector
Page 5
Description
The 61-1004RK CO2 detector has two versions, the 61-1004RK-02 which has a detection
range of 0 - 5,000 ppm and the 61-1004RK-03 which has a detection range of 0 - 5%
volume. The detector is an infrared type of detector. This section describes the
components of the 61-1004RK. They include the infrared CO
detector and the junction
2
box.
Infrared CO2 Detector
The infrared CO2 detector is made up of a miniature infrared CO2 detector housed and
encapsulated in a pipe nipple. The pipe nipple has 3/4 inch NP T threads on each end and
a 1-1/4 inch hex that allow s removal or in stallation of the detector with a wrench. A
porous flame arrestor coated with a hydrophobic film that repels liquids is on one end of
the detector and allows sample gas to enter the detector. Four color coded leads, red,
white, green, and black, extend from the other end of the detector. The leads allow you to
connect the detector to the terminal block in the junction box.
To distinguish the 5,000 ppm detector from the 5% detector (in case the replacement
detector label that is applied to one of the leads is lost), a short length of black shrink
tubing is applied to the white wire of the 5,000 ppm detector near where the wire comes
out of the nipple and a short length of green shrink tubing is applied to the white wire of
the 5% detector.
The output of the CO
detectors. For this reason, the CO2 detector’s wire colors are the same as RKI’s catalytic
LEL detectors, black, white, green, and red. When the CO
controller, it is wired the same way an LEL detector is wired, to the LEL detector
terminals. See “Wiring the CO2 Detector to a Controller” on page 6 for wiring
connections.
detector mimics the output of RKI’s standard catalytic LEL
2
detector is wired to an RKI
2
Junction Box
The junction box allows you to install the detector at a mounting site that is remote from a
controller and it protects the detector wiring connections. Two 3/4” NPT conduit hubs
allow you to mount the detector to the junction box and connect the wiring from the
detector to a controller. The terminal block within the junction box facilitates the wiring to
the detector. A cover on the front of the junction box allows access to the interior of the
junction b ox. Th ree spac ers in stall ed on t he ba ck of the junction box control the distance of
the junction box from a mounting surface and insure that there is enough room to install a
calibration cup on the detector during calibration.
61-1004RK/61-0191RK Comb us t i bl e G a s De t e ct or • 5
Page 6
Installation
This section describes procedures to mount the CO2 detector in the monitoring
environment and wire the it to a controller. If you purchased a 61-0191RK detector that is
factory installed in and factory wired to a controller, detector installation is not necessary.
Mounting the CO2 Detector
1.Select a mounting site that is representative of the monitoring environment. Consider
the following when you select the mounting site.
•Select a site where the detector is not likely to be bumped or disturbed. Ma ke sur e
there is sufficient room to perform start-up, maintenance, and calibration
procedures.
•Select a site where the target gas is likely to be found first.
•Select a site that minimizes the possibility of someone breathing on the detector.
The exhaled CO
3/4 NPT
Female
may cause an alarm.
2
2.70
0.38
3.65
7.1 Max
Junction
Box
Spacer, 3X
IR CO2
Detector
Figure 1: Mounting the CO2 Detector
2.At the mounting site you select, hang or mount the junction box with the detector
facing down (see Figure 1).
Wiring the CO2 Detector to a Controller
WARNING: Always verify that the power to t he controller is off before you make
wiring connections.
6 • 61-1004RK/61-0191RK Combust ible Gas Detector
Page 7
As mentioned in “Infrared CO2 Detector” on page 5, the IR CO2 detector wires to an RKI
controller the same way as an RKI LEL detector, to the LEL detector terminals. See
Figure 2 below for the wiring connections.
1.Turn off the controller.
2.Turn off or unplug po wer to the controlle r.
3.Remove the cover from the junction box.
4. Guide a four-conductor, shielded cable or four wires in conduit through the top
conduit hub of the junction box. Connect the wires to the terminals opposite the
detector leads.
CAUTION: Leave the shield drain wire insulated and disconnected at the 61-1004RK. You will
connect the opposite end of the cabl e’s drain wire at the controller.
5.Secure the junction box cover to the junction box.
6.Route the cable or wires in conduit leading from the detector through one of the
conduit hubs at the controller.
CAUTION: Do not route power and detector wiring through the same conduit hub. The power
cable may disru pt the transmission of the detect or signal to the controller.
Cable Shield
Terminal Strip
Black
Green
White
Red
IR CO2
Detector
Red
White
Green
Black
Detector
Wires
Controller
LEL Detector
Terminals
Controller Housing
Figure 2: Wiring the CO
Detector to a Controller
2
61-1004RK/61-0191RK Comb us t i bl e G a s De t e ct or • 7
Page 8
Start Up
This section describes procedures to start up the CO2 detector and place the detector into
normal operation.
Introducing Incoming Power
1.Complete the installation procedures described earlier in this manual.
2.Verify that the power wiring to the controller is correct and secure. Refer to the
controller operator’s manual.
3.Turn on or plug in power to the controller, then turn on the controller.
4.Verify that the controller is on and operating properly. Refer to the controller
operator’s ma nual.
CAUTION: Allow the detector to warm up for 5 minutes before you continue with the next
section, “Confirming the Zero Reading.”
Confirming the Zero Reading
There is typically between 200 ppm and 400 ppm CO2 in air depending on the geographic
location, so it is necessary to apply a CO
zero reading.
free gas to the CO2 detector when setting the
2
The procedure below describes applying 100% nitrogen to the detector using a calibration
kit that includes a calibration cup, a calibration cylinder of 100% nitrogen, sample tubing,
and a 0.5 LPM (lit ers per minute) fixed flow regulator with an on/off knob.
1.Screw the calibration cup onto the bottom of the CO
detector.
2
2.Screw the regulator into the 100% nitrogen calibration cylinder.
3.Use the sample tubing to connect the regulator to the calibration cup.
4.Turn the regulator’s on/off knob counterclockwise to open it. Gas will begin to flow.
5.Allow the gas to flow for one minute.
6.Verify a reading of 0% CO
or 0 ppm CO2 depending on the detector you are
2
calibrating at the controller.
If the display reading is 0, the CO
detector is in normal operation and start up is
2
complete. Proceed to step 7 to disassemble the calibration kit.
If the display reading is not 0, turn the regulator’s on/off knob clockwise to close it,
then proceed to “Setting the Zero Reading” on page 12 to set the zero reading.
7.Turn the regulator’s on/off knob clockwise to close it.
8.Unscrew the cylinder from the regulator and remove the sample tubing from the
regulator.
9.Remove the calibration cup from the detector. You may leave the sample tubing
connected to the calibration cup for convenience.
10. Store the components of the calibration kit in a safe and conven ient place.
8 • 61-1004RK/61-0191RK Combust ible Gas Detector
Page 9
Maintenance
This section describes maintenance procedures. It includes preventive maintenance,
troubleshooting, and component replacement procedures.
Preventive Maintenance
This section describes a preventive maintenance schedule to ensure the optimum
performance of the CO
Daily
detector. It includes daily, monthly, and biannual procedures.
2
Verify a display reading of around 200 ppm to 400 ppm (0.02% to 0.04%) CO
at the
2
controller. Investigate significant changes in the reading.
Monthly
This procedure describes a test to verify that the CO
detector responds properly to the
2
target gas.
WARNING: The controller is not an active gas monitoring device during the response
test procedure.
NOTE: Performing a response test on the CO2 detector may cause alarms. Be sure to put
the controller into its calibration program or disable external alarms before
performing this test
The procedure below describes applying CO
includes a calibration cup, a CO
calibration gas cylinder, sample tubing, and a 0.5 LPM
2
to the detector using a calibration kit that
2
(liters per minute) fixed flow regulator with an on/off knob.
1. Place the controller into its calibration program or disable external alarms.
2.Verify that the controller display reading is around 200 ppm to 400 ppm (0.02% to
0.04%) CO
.
2
If the reading is significantly out of this range, set the zero reading then continue this
procedure. See“Setting the Zero Reading” on page 12 to set the zero reading.
3.Screw the calibration cup onto the detector.
4.Use the sample tubing to connect the regulator to the calibration cup.
5.Screw the regulator into the calibration cylinder.
6.Turn the on/off knob on the regulator counterclockwise to open the regulator.
7.Apply gas to the detector for one minute.
8.Verify that the reading is within ± 20% of the gas concentratio n.
NOTE: If the reading is not within ± 20% of the gas concentration, calibrate the detector
as describe d in “Calibrat i on” on page 11.
9.Turn the on/off knob clockwise to close the regulator.
10. Unscrew the regulator from the calibration cylinder and remove the sample tubing
from the regulator.
61-1004RK/61-0191RK Comb us t i bl e G a s De t e ct or • 9
Page 10
11. Unscrew the calibration cup from the detector. You may leave the sample tubing
connected to the calibration cup for convenience.
12. When the display reading falls below the alarm se tpoints, return the controller to
normal operation or enable external alarms.
13. Store the components of the calibration kit in a safe place.
Biannually
Calibrate the CO
detector every six months as described in “Calibration” on page 11.
2
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended
action for problems you may encounter with the CO
NOTE: This troubleshooting guide describes detector problems only. See the controller
operator’s manual for problems you may encounte r with the controller.
Slow or No
Response/
Difficult or
Unable to
Calibrate
fail condition.
• Detector responds
slowly or does not
respond to response
test.
• Unable to accurately
set the zero or
response reading
during calibration.
• Detector requires
frequent calibration.
Note: Under “normal”
circumstances, the
detector requires
calibration once every
six months.
Some applications may
require a more freque nt
calibration schedule.
• The detector wiring is
disconnected or
misconnected.
• The detector zero
signal is low enough to
cause a fail condition.
• The detector is
malfunctioning.
• The calibration cylinder
is low, out-dated, or
defective.
• The regulator flow rate
is not 0.5 LPM
• The detector is
malfunctioning.
detector.
2
1. Verify that the detector wiring is
correct and secure.
2. Calibrate the det ect or.
3. If the fail condition continues, replace
the detector.
4. If the fail condition continues, contact
RKI for further instruction.
1. Verify that the calibration cylinder
contains an adequate supply of a
fresh test sample.
2. Confirm that the regu lat or bei ng us ed
is a 0.5 LPM regulator.
3. If the calibration/response difficulties
continue, replace the detector.
4. If the calibration/response difficulties
continue, contact RKI for further
instruction.
Replacing the CO2 Detector
1.Turn off the controller.
2.Turn off power to the controller.
3.Remove the junction box cover.
10 • 61-1004RK/61-0191RK Combus tible Gas Detector
Page 11
4.Disconnect the detector leads from the terminal block inside the junction bo x. Note
the position of the color-coded leads as you remove them.
5.Unscrew the detector from the junction box.
6.Guide the detector leads of the replacement detector through the bottom conduit hub
of the junction box, then screw the mounting threads of the detector into the conduit
hub.
7.Connect the detector leads to the terminal block in the same position as the leads you
removed in step 4.
8.Secure the junction box cover to the junction box.
9.Turn on power to the controller.
10. Turn on the controller.
11. Calibrate the replacement detector as described in“Calibration” on page 11.
Calibration Frequency
Although there is no particular calibration frequency that is correct for all applications, a
calibration frequency of every 6 months is adequate for most infrared CO
applications. Unless experience in a particular application dictates otherwise, RKI
Instruments, Inc. recommends a calibration frequency of every 6 months for the infrared
CO
detector.
2
detector
2
Calibration
If an application is not very demanding, for example detection in a clean, temperature
controlled environment where CO
is not normally present, and calibration adjustments
2
are minimal at calibration, then a calibration frequency of every 9 to 12 months is
adequate.
If an application is very demanding, for example if the environment is not well controlled,
then more frequent calibration than every 6 months may be necessary.
This section describes how to calibrate the CO2 detector. It includes procedures to
assemble the calibration kit, set the zero reading, set the response reading and return to
normal operation.
W ARNING: The controller is not an active gas monitoring device during the calibration
procedure.
The following procedure assumes the use of a calibration kit which includes a 100%
nitrogen calibration gas cylinder for setting the zero, a CO
calibration gas cylinder for
2
setting the response reading, a 0.5 LPM fixed flow regulator with an on/off knob, a
calibration cup for the detector, and a short piece of sample tubing to connect the
regulator to the calibration cup.
Assembling the Calibration Kit
1.Screw the calibration cup onto the detector.
2.Use the sample tubing to connect the regulator to the calibration cup.
3. Place the controller into its calibration program or disable external alarms.
61-1004RK/61-0191RK Comb u st i bl e G a s D et e ct or • 1 1
Page 12
NOTE: Calibrating the CO2 detector may cause alarms. Be sure to put the controller into
its calibration program or disable external alarms before continuing.
Setting the Zero Reading
1.Follow the directions in the controller’s operator’s manual for setting the zero reading
(sometimes called the fresh air reading).
2.When the directions call for exposing the detector to zeroing gas, screw the regulator
into the cylinder, turn the on/off knob counterclockwise to open the regulator, and
allow the gas to flow to the detector for 1 minute before continuing w ith the
directions.
3.After setting the zero reading, turn the regulator’s on/off knob clockwise to close the
regulator.
4.Unscrew the 100% nitrogen cylinder from the regulator.
Setting the Response Reading
1.Follow the directions in the controller’s operator’s manual for setting the response
reading (span).
2.When the directions call for exposing the detector to calibration gas, screw the CO
2
cylinder onto the regulator, turn the on/off knob counterclockwise to open the
regulator, and allow the gas to flow to the detector for 1 minute before continuing
with the directions.
3.After setting the response reading, turn the regulator’s on/off knob clockwise to close
the regulator.
4.Unscrew the CO
cylinder from the regulator.
2
5.Remove the sample tubing from the regulator.
6. Remove the calibration cup from the detector. Leave the sample tubing connected to
the calibration cup for convenience.
7.Allow the reading at the controller to decrease below the alarm points before
returning the controller to normal operation or enabling external alarms.
NOTE: If you do not allow the gas reading decrease below the alarm points, then
unwanted alarms may occur.
8.Verify that the controller display reading decreases and stabilizes at 0.
9.Store the components of the calibration kit in a safe and convenient place.
61-1004RK/61-0191RK Comb u st i bl e G a s D et e ct or • 1 2
Page 13
Parts List
Table 3 lists replacement parts and accessories for the 61-1004RK and 61-0191RK CO2
detectors.
Table 3: Parts List
Part NumberDescription
18-0400RK-01Junction box with spacers
61-0191RK-020 to 5,000 ppm CO
61-0191RK-030 to 5% volume CO
71-0145RK61-1004RK/61-0191RK CO
infrared detector
2
infrared detector
2
Detector Operator’s Manual (this
2
document)
81-0070RK-01Calibration cylinder (2,000 ppm CO2 in nitrogen, 34 liter)
81-0072RK-01Calibration cylinder (2.5% CO2 in nitrogen, 34 liter)
81-0078RK-01100% nitrogen calibration cylinder (34 liter)
81-1050RKRegulator, 0.5 liter/minute, with gauge and knob, for 17- and 34-
liter calibration cylinders
81-1 10 3R KCalibration cup
81-F403RK-LVCalibrati on kit, for 3/4” NPT detector, 2.5% CO2, 34 liter
81-F404RK-LVCalibration kit, for 3/4” NPT detector, 2,000 ppm CO2, 34 liter
61-1004RK/61-0191RK Comb u st i bl e G a s D et e ct or • 1 3
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.