Part Number: 71-0340
Revision: P2
Released: 9/17/14
www.rkiinstruments.com
WARNING
Read and understand this instruction manual before operating
detector . Improper use of the dete ctor could result in bodily harm
or death.
Periodic calibration and maintenance of the detector is essential
for proper operation and correct readings. Please calibrate and
maintain this detector regularly! Frequency of calibration
depends upon the type of use you have and the sensor types.
T ypical calibration frequencies for most applications a re between
6 and 12 months, but can be required more often or less often
based on your usage.
35-3001A-05-02 Carbon Dioxi de Sample-Draw Detecto r
Product Warranty
RKI Instruments, Inc. warrants gas alarm equipment s old by us t o be free from def ects in material s,
workmanship, and performance for a period of one year from date of shipment from RKI
Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our
option, free of charge. This warrant y does not apply to those items which by their nature are subject
to deterioration or consumption in normal service, and which must b e cleaned, repaired, o r replaced
on a routine basis. Examples of such items are:
Warranty is voided by abuse including mechanical damage, alteration, rough handling, or repair
procedures not in accordance with the operator’s manual. This warranty indicates the full extent of
our liability, and we are not responsible for removal or replacement costs, local repair costs,
transportation costs, or contingent expenses incurred without our prior approval.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER
WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND
ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI
INSTRUMENTS, INC. INCLUDING BUT N OT LIMITED T O, THE WARRANTY
OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN
NO EVENT SHALL RKI INSTRUMENTS, INC. BE LIABLE FOR INDIRECT,
INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND
CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS
PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
a) Absorbent cartridgesd) Batteries
b) Pump diaphragms and valvese) Filter elements
c) Fuses
This warranty covers instruments and parts sold to users by authorized distributors, dealers, and
representatives as appointed by RKI Instruments, Inc.
We do not assume indemnification for any accident or damage caused by the operation of this gas
monitor, and our warranty is limited to the replacement of parts or our complete goods.
35-3001A-05-02 Carbo n D iox ide Sam ple- Dra w Detector
35-3001A-05-02 Carbon Dioxi de Sample-Draw Detecto r
Overview
This operator’s manual describes the 35-3001A-05-02 carbon dioxide sample-draw detector. This
manual also describes how to install, start up, maintain, and calibrate the sample-draw detector
when using it with a gas monitoring controller. A parts list at the end of this manual lists
replacement parts and accessories for the sample-draw detector.
Specifications
Table 1 lists specifications for the carbon dioxide sample-draw detector. See the controller
Operator’s Manual for information specific to the controller.
Table 1: Specifications
Target Gas and Detection RangeCarbon Dioxide (CO
Input Power24 VDC Nominal (18.5 VDC - 30 VDC)
Construction (housing)Fiberglass/polyester (NEMA 4X)
Dimensions8.5 in. H x 6.5 in. W x 4.25 in. D
W eight4.5 lbs.
Sampling MethodSample-draw
Sample Flow3.0 SCFH typical, with no inlet or exhaust line
Flow to Sensor1.0 SCFH (nominal)
Maximum Recommended Inlet/
Exhaust Line Length for
1/4” O.D. x 1/8” I.D. Tubing
Maximum Recommended Inlet/
Exhaust Line Length for
1/4” O.D. x 0.170” I.D. Tubing
Response Time90% in 30 seconds
Accuracy± 5% of reading or ± 2% of full scale (whichever is greater)
Inlet
100 feet0 feet
50 feet50 feet
Inlet
100 feet100 feet
): 0 - 5,000 ppm
2
Exhaust
Exhaust
WARNING:When using the 35-3001A-05-02, you must follow the instructions and warnings
in this manual to assu r e pr oper an d s afe o per at ion of t h e 35 -30 01A -05 -02 and to
minimize the risk of personal injury. Be sure to maintain and periodically
calibrate t h e 35-3001A- 05-02 as described in this manual.
35-3001A-05-02 Carbon Diox ide Sa mple -D raw Detector • 1
Description
SensorF low
Con trol Valve
Fail LED
Pilot LED
Pum p Reset
Switch
This section describes the components of the carbon dioxide sample-draw detector.
2 • 35-3001A-05-02 Carbon D ioxide Sample-Draw D etector
External Components
This section describes the sample-draw detector’s external components.
Housing
The sample-draw detector’s fiberglass housing is weather- and corrosion-resistant. It is suitable fo r
installation where general purpose equipment is in use.
The housing door is hinged on the left side and is secured by two latches on the right side. The
flowmeter and status LEDs are visible through a window in the housing door.
Four mounting feet are attached to the back of the housing (one at each corner). Use the mounting
feet to install the housing to a vertical surface.
Sample Fittings
The sample fittings are located on the left side of the bottom of the housing. The inlet fitting is near
the front of the housing and the exhaust fitting is near the back of the housing. The inlet fitting has
a short factory installed tubing stub for use with the particle filter. The sample fittings accept 1/4 in.
rigid tubing. See the Installation section on page 7 to connect tubing to the sample fittings.
Inlet Fitting Tubing Stub
A short tubing stub comes factory installed in the inlet fittin g. It is used for connecting the particle
filter. If the particle filter is not used, or if you install the particle filter in a different location, the
tubing stub needs to be removed and replaced with tubing.
Conduit Hub
One 3/4” conduit hub is located on the right sid e of the bottom of the hous ing. It is used fo r routing
wiring into the housing by using conduit or an appropriate cable bushing.
Internal Components
This section describes the sample-draw detector’s internal components (see Figure 1). Figure 2
illustrates how the gas sample moves through the flow system.
35-3001A-05-02 Carbon Diox ide Sa mple -D raw Detector • 3
Main Circuit Board
The main circuit board includes the detect or/amp terminal strip, intercon nect terminal st rip, oxygen
sensor terminal strip, pump connector, and flowmet er circuit board connector (see Figure 1).
Detector/Amp Terminal Strip
The detector/amp terminal strip is the upper twelve-point terminal strip in the bot tom right corner
of the main circuit board. Use the detector/amp terminal strip to connect the amplifier to the main
circuit board.
NOTE: The amplifier is factory-wired to the main circuit board. See the “Installation” section on
page 7 for all wiring procedures related to the sample-draw detector.
Interconnect Terminal Strip
The interconnect terminal strip is the lower twelve-point terminal strip in the bottom right corner of
the main circuit board. Use the interconnect termin al stri p to con n ect the sam ple-draw detector to a
controller.
Oxygen Sensor Terminal Strip
The oxygen sensor terminal strip is a two-point terminal strip in the upper right corner of the main
circuit board. This terminal strip is not used in this version of the sample-draw detector.
Pump Connector
The pump connector is the two-point connector below the oxygen terminal strip in the upper right
corner of the main circuit board. Use the pump co nnector to connect the pump to the main circuit
board.
NOTE: The pump is factory-wired to the main circuit board. See “Installation” on page 7 for all
wiring procedures related to the sample-draw detector.
Flowmeter Circuit Board Connector
The flowmeter circuit board connector is a six-position connector in the upper left corner of the
main circuit board. Use the flowmeter circuit board connector to connect the flowmeter circuit
board to the main circuit board.
NOTE: The flowmeter circuit board is factory wired to the main circuit board. See “Installation”
on page 7 for all wiring procedures related to the sample-dr aw detector.
Flowmeter Circuit Board
The flowmeter circuit board is mounted to the left side of the main circuit board using standoffs. It
includes the flowmeter, sensor flow control valve, status LEDs, pressure switch, and pump reset
switch.
Flowmeter
The flowmeter is mounted to the right side of the flowmeter circuit board. You can see it through
the window in the door. A ball in the flowmeter column indicates the flow rate to the sensor. The
flowmeter measures the flow in the range 0.2 to 2.0 SCFH (Standard Cubic Feet per Hour). The
optimum flow rate is 1.0 SCFH.
Sensor Flow Control Valve
The sensor flow control valve is mounted to the flowmeter circuit board above the flowmeter. The
sensor flow control valve adjusts the flow rate to the detector. Turn the valve’s knob
4 • 35-3001A-05-02 Carbon D ioxide Sample-Draw D etector