RKI Instruments 35-3001A-05-02 User Manual

35-3001A-05-02
Carbon Dioxide Sample-Draw Detector
Operator’s Manual
Part Number: 71-0340 Revision: P2 Released: 9/17/14
www.rkiinstruments.com
WARNING
Read and understand this instruction manual before operating detector . Improper use of the dete ctor could result in bodily harm or death.
Periodic calibration and maintenance of the detector is essential for proper operation and correct readings. Please calibrate and maintain this detector regularly! Frequency of calibration depends upon the type of use you have and the sensor types. T ypical calibration frequencies for most applications a re between 6 and 12 months, but can be required more often or less often based on your usage.
Product Warranty
RKI Instruments, Inc. warrants gas alarm equipment s old by us t o be free from def ects in material s, workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge. This warrant y does not apply to those items which by their nature are subject to deterioration or consumption in normal service, and which must b e cleaned, repaired, o r replaced on a routine basis. Examples of such items are:
Warranty is voided by abuse including mechanical damage, alteration, rough handling, or repair procedures not in accordance with the operator’s manual. This warranty indicates the full extent of our liability, and we are not responsible for removal or replacement costs, local repair costs, transportation costs, or contingent expenses incurred without our prior approval.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI INSTRUMENTS, INC. INCLUDING BUT N OT LIMITED T O, THE WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL RKI INSTRUMENTS, INC. BE LIABLE FOR INDIRECT, INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
a) Absorbent cartridges d) Batteries b) Pump diaphragms and valves e) Filter elements c) Fuses
This warranty covers instruments and parts sold to users by authorized distributors, dealers, and representatives as appointed by RKI Instruments, Inc.
We do not assume indemnification for any accident or damage caused by the operation of this gas monitor, and our warranty is limited to the replacement of parts or our complete goods.
35-3001A-05-02 Carbo n D iox ide Sam ple- Dra w Detector
Table of Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
External Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Internal Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Mounting the Carbon Dioxide Sample-Draw Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Removing the Inlet Fitting Tubing Stub, if Necessary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Connecting the Sample Lines to the Sample-Draw Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Wiring the Carbon Dioxide Sample-Draw Detector to a Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Start Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Introducing Incoming Power. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Setting the Zero Reading. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Preventive Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Replacing Components of the Carbon Dioxide Sample-Draw Detector. . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Adjusting the Low Flow Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Calibration Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Preparing for Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Setting the Zero Reading. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Setting the Response Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Returning to Normal Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Overview
This operator’s manual describes the 35-3001A-05-02 carbon dioxide sample-draw detector. This manual also describes how to install, start up, maintain, and calibrate the sample-draw detector when using it with a gas monitoring controller. A parts list at the end of this manual lists replacement parts and accessories for the sample-draw detector.
Specifications
Table 1 lists specifications for the carbon dioxide sample-draw detector. See the controller Operator’s Manual for information specific to the controller.
Table 1: Specifications
Target Gas and Detection Range Carbon Dioxide (CO Input Power 24 VDC Nominal (18.5 VDC - 30 VDC)
Construction (housing) Fiberglass/polyester (NEMA 4X) Dimensions 8.5 in. H x 6.5 in. W x 4.25 in. D W eight 4.5 lbs. Sampling Method Sample-draw Sample Flow 3.0 SCFH typical, with no inlet or exhaust line Flow to Sensor 1.0 SCFH (nominal) Maximum Recommended Inlet/
Exhaust Line Length for 1/4” O.D. x 1/8” I.D. Tubing
Maximum Recommended Inlet/ Exhaust Line Length for 1/4” O.D. x 0.170” I.D. Tubing
Response Time 90% in 30 seconds Accuracy ± 5% of reading or ± 2% of full scale (whichever is greater)
Inlet
100 feet 0 feet
50 feet 50 feet
Inlet
100 feet 100 feet
): 0 - 5,000 ppm
2
Exhaust
Exhaust
WARNING: When using the 35-3001A-05-02, you must follow the instructions and warnings
in this manual to assu r e pr oper an d s afe o per at ion of t h e 35 -30 01A -05 -02 and to minimize the risk of personal injury. Be sure to maintain and periodically calibrate t h e 35-3001A- 05-02 as described in this manual.
35-3001A-05-02 Carbon Diox ide Sa mple -D raw Detector • 1
Description
SensorF low
Con trol Valve
Fail LED
Pilot LED
Pum p Reset
Switch
This section describes the components of the carbon dioxide sample-draw detector.
Zero Pot
Span Pot
Pum p Connector
O
W
X
+
Y G
B
E
_
N
Mounting Foo t, 4X
Flowm eter C ircuit
Board Connector
Factory Set Pots
Test Point s
100-500 m V Range
Sens o r Cur r ent 148mA
GBR
W
S
LEL
PWR / SIG
Controller
Terminal Strip
Detector Terminal Strip
IR Amplif ier
Oxy gen Sensor Term ina l Strip
Pum p
IR CO2 Sensor (inside block )
IR CO2 FlowBlock
IR CO2 Sensor PCB
Flowmeter
Pressure
Switch
Adjustment
Screw
(behind PCB)
Hyd rophobic
Filter
This End To In l et Fitting
Particle Filter
Inlet
Fitting
INLET
EXHAUST
Detector/Amp Term ina l Strip (Factory
+
+
AMP2AMP
AMP2
+
W B
GS R
LEL/IR
OXY
R
LEL/IR
BGW
S
WG
+ OXY
S
+
S
+
AMP 1
Wired)
Int erconnec t Termi nal Strip
3/4" Conduit Hub
Exhaust
Fitting
Figure 1: Carbon Dioxide Sample-Draw Detector Component Location
2 • 35-3001A-05-02 Carbon D ioxide Sample-Draw D etector
External Components
This section describes the sample-draw detector’s external components.
Housing
The sample-draw detector’s fiberglass housing is weather- and corrosion-resistant. It is suitable fo r installation where general purpose equipment is in use.
The housing door is hinged on the left side and is secured by two latches on the right side. The flowmeter and status LEDs are visible through a window in the housing door.
Four mounting feet are attached to the back of the housing (one at each corner). Use the mounting feet to install the housing to a vertical surface.
Sample Fittings
The sample fittings are located on the left side of the bottom of the housing. The inlet fitting is near the front of the housing and the exhaust fitting is near the back of the housing. The inlet fitting has a short factory installed tubing stub for use with the particle filter. The sample fittings accept 1/4 in. rigid tubing. See the Installation section on page 7 to connect tubing to the sample fittings.
Inlet Fitting Tubing Stub
A short tubing stub comes factory installed in the inlet fittin g. It is used for connecting the particle filter. If the particle filter is not used, or if you install the particle filter in a different location, the tubing stub needs to be removed and replaced with tubing.
Conduit Hub
One 3/4” conduit hub is located on the right sid e of the bottom of the hous ing. It is used fo r routing wiring into the housing by using conduit or an appropriate cable bushing.
Internal Components
This section describes the sample-draw detector’s internal components (see Figure 1). Figure 2 illustrates how the gas sample moves through the flow system.
Exhaust
Inlet
Pressure Switch
Flowmeter PCB
Restrictor
Sensor
Flow
Control
Valve
Flowmeter
IR CO2 Sensor
ParticleFilter
Pump
HydrophobicFilter
Figure 2: Carbon Dioxide Sample-Draw Detector Flow Diagram
35-3001A-05-02 Carbon Diox ide Sa mple -D raw Detector • 3
Main Circuit Board
The main circuit board includes the detect or/amp terminal strip, intercon nect terminal st rip, oxygen sensor terminal strip, pump connector, and flowmet er circuit board connector (see Figure 1).
Detector/Amp Terminal Strip
The detector/amp terminal strip is the upper twelve-point terminal strip in the bot tom right corner of the main circuit board. Use the detector/amp terminal strip to connect the amplifier to the main circuit board.
NOTE: The amplifier is factory-wired to the main circuit board. See the “Installation” section on
page 7 for all wiring procedures related to the sample-draw detector.
Interconnect Terminal Strip
The interconnect terminal strip is the lower twelve-point terminal strip in the bottom right corner of the main circuit board. Use the interconnect termin al stri p to con n ect the sam ple-draw detector to a controller.
Oxygen Sensor Terminal Strip
The oxygen sensor terminal strip is a two-point terminal strip in the upper right corner of the main circuit board. This terminal strip is not used in this version of the sample-draw detector.
Pump Connector
The pump connector is the two-point connector below the oxygen terminal strip in the upper right corner of the main circuit board. Use the pump co nnector to connect the pump to the main circuit board.
NOTE: The pump is factory-wired to the main circuit board. See “Installation” on page 7 for all
wiring procedures related to the sample-draw detector.
Flowmeter Circuit Board Connector
The flowmeter circuit board connector is a six-position connector in the upper left corner of the main circuit board. Use the flowmeter circuit board connector to connect the flowmeter circuit board to the main circuit board.
NOTE: The flowmeter circuit board is factory wired to the main circuit board. See “Installation”
on page 7 for all wiring procedures related to the sample-dr aw detector.
Flowmeter Circuit Board
The flowmeter circuit board is mounted to the left side of the main circuit board using standoffs. It includes the flowmeter, sensor flow control valve, status LEDs, pressure switch, and pump reset switch.
Flowmeter
The flowmeter is mounted to the right side of the flowmeter circuit board. You can see it through the window in the door. A ball in the flowmeter column indicates the flow rate to the sensor. The flowmeter measures the flow in the range 0.2 to 2.0 SCFH (Standard Cubic Feet per Hour). The optimum flow rate is 1.0 SCFH.
Sensor Flow Control Valve
The sensor flow control valve is mounted to the flowmeter circuit board above the flowmeter. The sensor flow control valve adjusts the flow rate to the detector. Turn the valve’s knob
4 • 35-3001A-05-02 Carbon D ioxide Sample-Draw D etector