RKI Instruments 35-3001A-01H User manual

Page 1
35-3001A-01H
Hydrogen Sample-Draw Detector
Operator’s Manual
Part Number: 71-0307RK Revision: A Released: 8/7/17
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 2
WARNING
Read and understand this instruction manual before operating detector. Improper use of the detector could result in bodily harm or death.
Periodic calibration and maintenance of the detector is essential for proper operation and correct readings. Please calibrate and maintain this detector regularly! Frequency of calibration depends upon the type of use you have and the sensor types. Typical calibration frequencies for most applications are between 3 and 6 months, but can be required more often or less often based on your usage.
35-3001A-01H Hydrogen Sample-Draw Detector
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 3

Product Warranty

RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge. This warranty does not apply to those items which by their nature are subject to deterioration or consumption in normal service, and which must be cleaned, repaired, or replaced on a routine basis. Examples of such items are:
Warranty is voided by abuse including mechanical damage, alteration, rough handling, or repair procedures not in accordance with the operator’s manual. This warranty indicates the full extent of our liability, and we are not responsible for removal or replacement costs, local repair costs, transportation costs, or contingent expenses incurred without our prior approval.
THIS W ARRAN T Y IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI INSTRUMENTS, INC. INCLUDING BUT N OT LIMITED TO, THE WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL RKI INSTRUMENTS, INC. BE LIABLE FOR INDIRECT, INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
a) Absorbent cartridges d) Batteries
b) Pump diaphragms and valves e) Filter elements
c) Fuses
This warranty covers instruments and parts sold to users by authorized distributors, dealers, and representatives as appointed by RKI Instruments, Inc.
We do not assume indemnification for any accident or damage caused by the operation of this gas monitor, and our warranty is limited to the replacement of parts or our complete goods.
35-3001A-01H Hydro gen Sample-Draw Detecto r
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 4

Table of Contents

Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
External Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Internal Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Mounting the Hydrogen Sample-Draw Detector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Connecting the Sample Lines to the Sample-Draw Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Wiring the Hydrogen Sample-Draw Detector to a Controller. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Start Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Introducing Incoming Power. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Setting the Zero Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Preventive Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Replacing Components of the Hydrogen Sample-Draw Detector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Adjusting the Low Flow Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Removing the Particle Filter’s Tubing Stub, if Necessary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Calibration Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Preparing for Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Setting the Zero Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Setting the Response Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Returning to Normal Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
35-3001A-01H Hydrogen Sample-Draw Detector
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 5

Overview

This operator’s manual describes the 35-3001A-01H hydrogen sample-draw detector. This manual also describes how to install, start up, maintain, and calibrate the sample-draw detector when using it with a gas monitoring controller. A parts list at the end of this manual lists replacement parts and accessories for the sample-draw detector.

Specifications

Table 1 lists specifications for the hydrogen sample-draw detector. See the controller Operator’s Manual for information specific to the controller.
Target Gas and Detection Range Hydrogen: 0 - 100% LEL
Input Power 24 VDC Nominal (18.5 VDC - 30 VDC)
Construction (housing) Fiberglass/polyester (NEMA 4X)
Dimensions 8.5 in. H x 6.5 in. W x 4.25 in. D
Table 1: Specifications
Weight 4.5 lbs.
Sampling Method Sample-draw
Sample Flow 3.0 SCFH typical, with no inlet or exhaust line
Flow to Sensor 1.0 SCFH (nominal)
Low Flow Setpoint 0.6 ± 0.1 SCFH
Maximum Recommended Inlet/ Exhaust Line Length for 1/4” O.D. x 1/8” I.D. Tubing
Maximum Recommended Inlet/ Exhaust Line Length for 1/4” O.D. x 0.170” I.D. Tubing
Response Time 90% in 30 seconds
Accuracy ± 5% of reading or ± 2% LEL (whichever is greater)
W ARNING: When using the 35-3001A-01H, you must follow the instructions and warnings in
this manual to assure proper and safe operation of the 35-3001A-01H and to minimize the risk of personal injury. Be sure to maintain and periodically calibrate th e 35-3001A-01H as described in this manual.
Inlet
100 feet 0 feet
50 feet 50 feet
Inlet
100 feet 100 feet
Exhaust
Exhaust
35-3001A-01H Hydro gen Sample-Draw Det ector • 1
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 6

Description

Pump
Oxygen Sensor Terminal Strip
Mounting Foot, 4X
W
+
B
O X Y G E N
_
S
PWR/SIG
Hydrogen Flow Block
Pilot LED
Fail LED
Sensor Flow
Control Valve
Interconnect Terminal Strip
Pump R eset
Switch
Flowmeter
WS
+
AMP1
+
AMP2
OXY
S R
+
WG
G
Hydrophobic
Filter
Particle Filter
B
LEL/ IR
Flowmeter Circuit
Board Connector
Inlet
Fitting
Exhaust
Fitting
This End To Inlet Fitting
3/4" Conduit Hub
Hydrogen Sensor
INLET
EXHAUST
Pump C onnector
Hydrogen Amplifier
Sensor Current
115mA
W
LEL
GBR
AMP2AMP
+
+
W B
Pressure
Switch
Adjustment
Screw
(behind PCB)
Detector/Amp Terminal Strip (Factory Wired)
LEL/ IR
GS R
+
OXY
S
This section describes the components of the hydrogen sample-draw detector.
Figure 1: Hydrogen Sample-Draw Detector Component Location
2 • 35-3001A-01H Hydrogen Sample-Draw Detector
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 7

External Components

Flowmeter
Pump
Hydrogen
Sensor

ParticleFilter

Exhaust
Flowmeter PCB
Hydrophobic Filter
Pressure Switch
Restrictor
Inlet
Sensor
Flow
Control
Valve
This section describes the sample-draw detector’s external components.

Housing

The sample-draw detector’s fiberglass housing is weather- and corrosion-resistant. It is suitable for installation where general purpose equipment is in use.
The housing door is hinged on the left side and is secured by two latches on the right side. The flowmeter and status LEDs are visible through a window in the housing door.
Four mounting feet are attached to the back of the housing (one at each corner). Use the mounting feet to install the housing to a vertical surface.

Sample Fittings

The sample fittings are located on the left side of the bottom of the housing. The inlet fitting is near the front of the housing and the exhaust fitting is near the back of the housing. The sample fittings accept 1/4 in. rigid tubing. See the Installation section on page 7 to connect tubing to the sample fittings.
Particle Filter
A particle filter with a tubing stub on one end is shipped with the instrument but it is not factory installed. If the particle filter is installed directly to the inlet fitting, the tubing stub must be used. If the particle filter is installed somewhere else, like at the end of the inlet line, the tubing stub can be used or removed.

Conduit Hub

One 3/4” conduit hub is located on the right side of the bottom of the housing. It is used for routing wiring into the housing by using conduit or an appropriate cable bushing.

Internal Components

This section describes the sample-draw detector’s internal components (see Figure 1). Figure 2 illustrates how the gas sample moves through the flow system.
Figure 2: Hydrogen Sample-Draw Detector Flow Diagram
Find Quality Products Online at: sales@GlobalTestSupply.com
35-3001A-01H Hydro gen Sample-Draw Det ector • 3
Page 8

Main Circuit Board

The main circuit board includes the detector/amp terminal strip, interconnect terminal strip, oxygen sensor terminal strip, pump connector, and flowmeter circuit board connector (see Figure 1).
Detector/Amp Terminal Strip
The detector/amp terminal strip is the upper twelve-point terminal strip in the bottom right corner of the main circuit board. Use the detector/amp terminal strip to connect the amplifier to the main circuit board.
NOTE: The amplifier is factory-wired to the main circuit board. See the “Installation” section on
page 7 for all wiring procedures related to the sample-draw detector.
Interconnect Terminal Strip
The interconnect terminal strip is the lower twelve-point terminal strip in the bottom right corner of the main circuit board. Use the interconnect terminal strip to connect the sample-draw detector to a controller.
Oxygen Sensor Terminal Strip
The oxygen sensor terminal strip is a two-point terminal strip in the upper right corner of the main circuit board. This terminal strip is not used in this version of the sample-draw detector.
Pump Connector
The pump connector is the two-point connector below the oxygen terminal strip in the upper right corner of the main circuit board. Use the pump connector to connect the pump to the main circuit board.
NOTE: The pump is factory-wired to the main circuit board. See “Installation” on page 7 for all
wiring procedures related to the sample-draw detector.
Flowmeter Circuit Board Connector
The flowmeter circuit board connector is a six-position connector in the upper left corner of the main circuit board. Use the flowmeter circuit board connector to connect the flowmeter circuit board to the main circuit board.
NOTE: The flowmeter circuit board is factory wired to the main circuit board. See “Installation”
on page 7 for all wiring procedures related to the sample-draw detector.

Flowmeter Circuit Board

The flowmeter circuit board is mounted to the left side of the main circuit board using standoffs. It includes the flowmeter, sensor flow control valve, status LEDs, pressure switch, and pump reset switch.
Flowmeter
The flowmeter is mounted to the right side of the flowmeter circuit board. You can see it through the window in the door. A ball in the flowmeter column indicates the flow rate to the sensor. The flowmeter measures the flow in the range 0.2 to 2.0 SCFH (Standard Cubic Feet per Hour). The optimum flow rate is 1.0 SCFH.
Sensor Flow Control Valve
The sensor flow control valve is mounted to the flowmeter circuit board above the flowmeter. The sensor flow control valve adjusts the flow rate to the detector. Turn the valve’s knob
4 • 35-3001A-01H Hydrogen Sample-Draw Detector
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 9
counterclockwise to increase the flow and clockwise to decrease the flow.
Status LEDs
Two status LEDs are above the flowmeter. They are also visible through the window in the housing door. The green Pilot LED is on when the sample-draw detector is receiving power from the controller. The red Fail LED is on when the sample flow rate is below the low flow level.
Pressure Switch
The pressure switch is mounted to the back of the flowmeter circuit board. The pressure switch monitors the flow rate of the incoming gas sample.
If the flow rate falls below the preset low flow level, the pressure switch causes the Fail LED to turn on and interrupts the signal from the detector. The interrupted detector signal causes a fail condition at the controller. The low flow level is factory-set at 0.6 SCFH (±0.1 SCFH). See “Adjusting the Low Flow Setting” on page 16 to adjust this setting.
Pump Reset Switch
The pump reset switch is located to the left of the status LEDs. When a low flow condition occurs, the pump will be shut off. To reset the low flow condition and start the pump again, press and hold the pump reset switch for about 2 seconds, then release.

Hydrophobic Filter

The hydrophobic filter is located toward the bottom left of the main circuit board. The filter prevents particulates and water in the incoming gas sample from damaging the flow and detection systems. Replace the filter when it appears dirty, discolored, or clogged.

Pump

The pump is mounted to the right side of the main circuit board. The pump pulls the gas sample into the sample-draw detector. The pump operates on 24 VAC, which is generated from the 24 VDC supplied by the controller.

Hydrogen Sensor

The catalytic hydrogen sensor detects hydrogen in the %LEL range. It uses a catalytic element for detection. The reaction of gas with oxygen on the catalyst causes a change in the resistance of the element which changes the current flowing through it. The current is proportional to the detection range of the sample-draw detector.
The hydrogen sensor is installed in the flow block and the flow block is mounted to the middle of the main circuit board. The sensor housing includes a sintered metal flame arrestor on one end that allows gas to diffuse into the sensor. The sensor has a built in molecular sieve that only allows hydrogen to diffuse into the detector. On the other end, five pins extend from the sensor and plug into the socket connector. The connector allows you to replace the sensor without disconnecting the wiring. Four leads extending from the connector are factory wired to the amplifier.
35-3001A-01H Hydro gen Sample-Draw Det ector • 5
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 10

Amplifier

S
PWR / SIG
Zero Pot
Span Pot
Controller Terminal Strip
Detector
Terminal Strip
Test Points
100-500 mV Range
Factory Set Pots
W
LEL
G
BR
The amplifier converts the electrical output from the detector to a 4 to 20 mA signal that corresponds to the detection range and transmits the signal to a gas monitoring controller. A label on the amplifier indicates the detector drive current. This drive current is factory set and is dictated by the combustible gas to which the detector is calibrated. Consult RKI Instruments, Inc. if you plan to change the calibration gas for the sample-draw detector, for example from a hydrogen calibration to a methane calibration. The amplifier is mounted on the top middle edge of the main circuit board. It consists of the zero pot, span pot, controller terminal strip, detector terminal strip, and test points.
Figure 3: Amplifier Component Location
Zero Pot
The zero pot is located in the upper right corner of the amplifier (see Figure 3). Use a small flat blade screwdriver to turn the zero pot’s adjustment screw and adjust the amplifier’s zero (fresh air) output during the start-up and calibration procedures. Turn the adjustment screw clockwise to increase the fresh air output and counterclockwise to decrease the fresh air output.
Span Pot
The span pot is located below the zero pot (see Figure 3). Use a small flat blade screwdriver to turn the span pot’s adjustment screw and adjust the amplifier’s gas response output during the calibration procedure. Turn the adjustment screw clockwise to increase the span output and counterclockwise to decrease the span output.
CAUTION: The amplifier includes additional pots. They are factory-set. Do not adjust them.
Controller and Detector Terminal Strips
The controller terminal strip and detector terminal strip are three- and four-position plug-in style terminal strips, respectively. The controller terminal strip is located on the right side of the amplifier and the detector terminal strip is to the left of it. Both terminal strips are factory wired to the sensor and main circuit board.
Test Points
The test points are on the left side of the amplifier (see Figure 3). The test points produce a 100 mV to 500 mV output that corresponds to the sample-draw detector’s 4 to 20 mA output. Use the test points and a voltmeter to measure the amplifier’s output during the start-up and calibration procedures. The black test point in the upper left corner is the negative (-) test point and the red test point to the left of the zero and span pots is the positive (+) test point.
6 • 35-3001A-01H Hydrogen Sample-Draw Detector
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 11

Installation

This section describes procedures to mount the sample-draw detector in the monitoring environment and wire the sample-draw detector to a controller.

Mounting the Sample-Draw Hydrogen Detector

1. Select the mounting site. Consider the following when you select the mounting site:
Is there enough room to open the housing door and make wiring and tubing connections at the bottom of the housing?
Make sure there is sufficient room to perform start-up, maintenance, and calibration procedures.
Are the flowmeter and status LEDs visible?
6.50
1.13
Ø.30x.50(4X)MOUNTING
.78
.40
8.50
Pilot Fail
4.00
Sampl e Drawi ng
s Dete ct or
Ga
35-300 1
8.88
.80
.85
4.50
NOTE: Housing is 4 .3 i nches d eep
Figure 4: Outline and Mounting Dimensions
35-3001A-01H Hydro gen Sample-Draw Det ector • 7
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 12
2. Close and latch the housing door.
NOTE: The sample-draw detector is shipped with the mounting feet “tucked under” the housing
to protect the mounting feet during shipment.
3. Slightly loosen the screw that secures the mounting foot to the housing, then rotate the mounting foot 180 degrees (see Figure 4).
4. Tighten the screw that secures the mounting foot to the housing.
5. Repeat steps 3 and 4 for the remaining three mounting feet.
6. Position the sample-draw housing on a vertical surface at eye level (4 1/2 to 5 feet from the floor).
7. Insert 1/4 inch screws through the slots in the mounting feet to secure the housing to the mounting surface.

Connecting the Sample Lines to the Sample-Draw Detector

See Figure 6 for the fitting layout.

Installing the Inlet Line with Particle Filter Connected to the Inlet Fitting

1. Loosen the nut on the inlet fitting until 3 threads are visible.
2. Push the particle filter’s tubing stub into the inlet fitting until it stops. Be sure the arrow on the particle filter is pointing toward the inlet fitting.
3. Hand tighten the nut on the inlet fitting so the ferrules clamp on the tubing. If the tube nut is removed, see Figure 6 for the arrangement of the fitting components.
4. Connect a length of sample tubing to the other side of the particle filter and route it to the sampling area. 1/4” O.D. rigid polypropylene, Teflon, or flexible polyurethane tubing may be used. RKI Instruments, Inc. recommends using either 1/4” O.D. x 1/8” I.D. or 1/4” O.D. x
0.170” I.D. tubing based on your length requirements. See “Specifications” on page 1 for maximum tubing lengths based on tubing size.
CAUTION: If you use flexible sample tubing (polyurethane is acceptable), use an appropriate
insert if necessary to provide support to the tubing and ensure a good seal when connecting to the flexible stub on the filter.
CAUTION: Avoid lo ops or slumps in the incoming sa mple line. To reduce response time, keep the
incoming sample line as short as possible.

Installing the Inlet Line without Particle Filter Connected to the Inlet Fitting

1. Loosen the nut on the inlet fitting until 3 threads are visible.
2. Push 1/4” O.D. rigid polypropylene or rigid Teflon sample tubing into the fitting until it stops. Flexible polyurethane tubing may be used with an appropriate insert. RKI Instruments, Inc. recommends using either 1/4” O.D. x 1/8” I.D. or 1/4” O.D. x 0.170” I.D. tubing based on your length requirements. See “Specifications” on page 1 for maximum tubing lengths based on tubing size.
CAUTION: If you use flexible sample tubing (polyurethane is acceptable), use an appropriate
insert to seal the connection between the tubing and the inlet fitting.
8 • 35-3001A-01H Hydrogen Sample-Draw Detector
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 13
3. Hand tighten the nut on the inlet fitting so the ferrules clamp on the tubing. If the tube nut is removed, see Figure 6 for the arrangement of the fitting components.
4. Route tubing from the inlet fitting to the sample area. See “Specifications” on page 1 for maximum tubing lengths based on tubing size.
CAUTION: Avoid lo ops or slumps in the incoming sa mple line. To reduce response time, keep the
incoming sample line as short as possible.
5. If desired, install the particle filter onto the end of the inlet line. Be sure the arrow on the particle filter is pointed toward the inlet line.

Installing the Exhaust Line

1. Loosen the nut on the exhaust fitting until 3 threads are visible.
2. Push 1/4” O.D. rigid polypropylene or rigid Teflon sample tubing into the fitting until it stops. Flexible polyurethane tubing may be used with an appropriate insert. RKI Instruments, Inc. recommends using either 1/4” O.D. x 1/8” I.D. or 1/4” O.D. x 0.170” I.D. tubing based on your length requirements. See “Specifications” on page 1 for maximum tubing lengths based on tubing size.
CAUTION: If you use flexible sample tubing (polyurethane is acceptable), use an appropriate
insert to seal the connection between the tubing and the inlet fitting.
3. Hand tighten the nut on the exhaust fitting so the ferrules clamp on the tubing. If the tube nut is removed, see Figure 6 for the arrangement of the fitting components.
4. Route the opposite end of the tubing to an open area where the sample can safely disperse or to an exhaust duct. See “Specifications” on page 1 for maximum tubing lengths based on tubing size.

Wiring the Sample-Draw Hydrogen Detector to a Controller

WARNING: Always verify that the controller is off and that power to the controller is off
before you make wiring connections.
1. Turn off the controller.
2. Turn off power to the controller.
3. Unlatch and open the housing door of the sample-draw detector.
4. Guide a three-conductor, shielded cable or three wires in conduit through the conduit hub at the bottom of the sample-draw housing. A minimum of 18 AWG wire is recommended.
5. Connect the cable to the sample-draw detector’s interconnect terminal strip as shown in Figure
5.
6. Close and latch the housing door of the sample-draw detector.
CAUTION: If using shielded cable, leave the cable shield’s drain wire insulated and
disconnected at the sample-draw detector. You will connect the opposite end of the drain wire at the controller.
7. Route the cable or wires in conduit leading from the sample-draw detector through one of the conduit hubs at the controller.
35-3001A-01H Hydro gen Sample-Draw Det ector • 9
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 14
8. Connect the wires to the applicable detector/transmitter terminal strip at the controller as
Not Used on This Version
S
PWR / SIG
ZE RO
SP AN
OXY LEL/ IR
Red
White
AMP 1
PATS201PATS201
LEL/ IR
+24VDC
AMP 1
OXY
Controller or
Recording
Device
AMP 2
4/20 Signal
-(DCGround)
To
Hydrogen
Se nsor
AMP 2
PATS101PATS101
W
LEL
GBR
Green
Black
Sens or
Current
115m A
shown in Figure 5. Refer to the controller operator’s manual and the controller detector head specification sheet for the 35-3001A-01H for detector/terminal strip connections specific to the controller.
Figure 5: Wiring the Sample-Draw Detector to a Controller
9. If shielded cable is used, connect the cable’s drain wire to an available chassis (earth) ground at the controller. RKI controllers typically have a ground stud that can be used to ground the
10 • 35-3001A-01H Hydrogen Sample-Draw Detector
Find Quality Products Online at: sales@GlobalTestSupply.com
cable’s drain wire.
Page 15

Start Up

This section describes procedures to start up the sample-draw detector and place the sample-draw detector into normal operation.

Introducing Incoming Power

1. Complete the installation procedures described earlier in this manual.
2. Verify that the wiring is correct and secure. Refer to the controller operator’s manual for connections at the controller.
3. Turn on or plug in the power to the controller, then turn on the controller.
4. Verify that the sample-draw detector’s Pilot LED is on.
5. Verify that the controller is on and operating properly. Refer to the controller operator’s manual.
6. Verify that the flowmeter indicates a flow rate of approximately 1.0 SCFH. If necessary, use the sensor flow control valve to adjust the flow rate. Turn the valve’s knob counterclockwise to increase the flow and clockwise to decrease the flow.
NOTE: The following step tests for leaks in the sample line. This test will cause a low flow
condition at the sample-draw detector and a fail condition at the controller. Be sure to put the controller into its calibration program or disable external alarms before performing this test.
7. Verify that the incoming sample line is not leaking. To test the sample line, press and hold the reset switch and plug the open end of the sample line with your thumb. If the flowmeter ball drops to the bottom of the flowmeter, the incoming sample line is not leaking.
8. Remove your thumb from the sample line, release the reset switch, and verify the flowmeter returns to a normal flow rate.
9. Enable alarms or place the controller in normal operation.
CAUTION: Allow the sample-draw detector to warm up for 5 minutes before you continue with
the next section, “Setting the Zero Reading.”

Setting the Zero Reading

CAUTION: If you suspect the presence of hydrogen in the monitoring environment, use the
calibration kit and the zero air calibration cylinder to introduce “fresh air” to the sensor and verify an accurate zero setting. See “Calibration” on page 18 for instructions on using a zero air calibration cylinder for setting the zero reading.
1. Verify that the sample-draw detector is sampling a fresh air environment (environment known to be free of combustible gas).
2. Open the housing door.
3. Set a voltmeter to measure in the millivolt (mV) range.
4. Plug the voltmeter leads into the test points on the amplifier. Plug the positive lead into the red (+) test point; plug the negative lead into the black (-) test point.
5. Verify a voltmeter reading of 100 mV (± 2 mV).
35-3001A-01H Hydrog en Sample-Draw Det ect or • 11
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 16
6. If necessary, use a small flat-blade screwdriver to adjust the zero potentiometer until the
7. Remove the voltmeter leads from the test points.
8. Close the housing door.

Maintenance

This section describes maintenance procedures. It includes preventive maintenance procedures. This section also includes procedures to troubleshoot the sample-draw detector, replace components of the sample-draw detector, adjust the low flow setting, and remove the particle filter’s tubing stub.

Preventive Maintenance

This section describes a preventive maintenance schedule to ensure the optimum performance of the sample-draw detector. It includes daily, monthly, and quarterly procedures.

Daily Visual Checks

1. Verify that the Pilot LED is on.
2. Verify that the flowmeter indicates a flow rate of approximately 1.0 SCFH. If necessary use the
voltmeter reading is 100 mV (± 2 mV).
sensor flow control valve to adjust the flow rate to 1.0 SCFH. Turn the valve’s knob counterclockwise to increase the flow and clockwise to decrease the flow.
3. Verify a display reading of 0 %LEL at the controller or a reading of 100 mV at the amplifier test points. Investigate significant changes in the display reading.

Monthly Response Test

This procedure describes a test to verify that the sample-draw detector responds properly to the target gas.
NOTE: To reduce the response time of this test, use a short incoming sample line. If the sample-
draw detector’s sample line is long, connect a shorter line for this test. Make sure you reconnect the sample line after you complete this procedure.
NOTE: Performing a response test on the sample-draw detector may cause alarms. Be sure to put
the controller into its calibration program or disable external alarms before performing this test.
Preparing for the response test
NOTE: This procedure describes the RKI calibration kit that includes a demand flow regulator.
1. Verify that the display reading at the controller is 0 % LEL.
If the display reading is not 0 % LEL, set the zero reading as described in “Calibration” on page 18, then continue this procedure.
2. Set a voltmeter to measure in the millivolt (mV) range.
3. Open the housing door, then plug the voltmeter leads into the test points on the amplifier.
Plug the positive lead into the red (+) test point; plug the negative lead into the black (-) test point.
12 • 35-3001A-01H Hydrogen Sample-Draw Detector
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 17
4. Use the following formula to determine the correct test points output for the test sample.
Output (mV) = (calibrating sample/fullscale) X 400 + 100
For example, with a test sample of 50 %LEL and a fullscale setting of 100 %LEL, the correct output is 300 mV.
300 (mV) = (50/100) X 400 +100
Performing the response test
1. Screw the demand flow regulator into the calibration cylinder.
2. Connect the calibration tubing from the demand flow regulator to the inlet fitting. Gas will begin to flow.
3. After approximately one minute, verify that the millivolt output at the amplifier stabilizes within ± 20% of the reading determined above. If the reading is not within ± 20% of the determined value, calibrate the sample-draw detector as described in “Calibration” on page 18.
4. Remove the calibration tubing from the inlet fitting, then reconnect the sample tubing to the inlet fitting.
5. Remove the voltmeter leads from the amplifier test points.
6. Close the housing door.
7. Store the calibration kit in a safe place.

Quarterly Calibration

Calibrate the sample-draw detector as described in “Calibration” on page 18.

Troubleshooting

The troubleshooting guide describes symptoms, probable causes, and recommended action for problems you may encounter with the sample-draw detector.
NOTE: This troubleshooting guide describes sample-draw detector problems only. See the
controller Operator’s Manual if the controller exhibits any problems.

Fail Condition

Symptoms
The sample-draw detector’s Fail LED is on.
The monitoring device is operating properly but indicates a reading well below zero or a failure alarm.
Probable causes
The sample-draw detector’s flow rate is too low because of an obstructed sample line, failed pump, etc.
The sample-draw detector is malfunctioning.
The sensor and/or amplifier wiring is disconnected or misconnected.
Recommended action
1. At the sample-draw detector, set the correct flow rate with the sensor flow control valve.
2. If you cannot set the correct flow rate, check the sample lines for obstructions or kinks.
3. Verify that the sensor wiring and amplifier wiring is correct and secure. “Wiring the Sample­Draw Hydrogen Detector to a Controller” on page 9 describes sensor and amplifier wiring connections.
35-3001A-01H Hydrog en Sample-Draw Det ect or • 13
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 18
4. Calibrate the sample-draw detector as described in “Calibration” on page 18.
5. If the fail condition continues, replace the sensor as described in “Replacing the Hydrogen Sensor” on page 14.
6. If the fail condition continues, contact RKI Instruments, Inc. for further instruction.

Slow or No Response/Difficult or Unable to Calibrate

Symptoms
The sensor responds slowly or does not respond during the monthly response test.
Unable to accurately set the zero or response reading during the calibration procedure.
The sensor requires frequent calibration.
Probable causes
The calibration cylinder is low, out-dated, or defective.
If a demand flow regulator calibration kit is used, the demand flow regulator is not functioning properly.
The sample-draw detector’s flow rate is too low because of an obstructed sample line, failed pump, etc.
The sample-draw detector is malfunctioning.
Recommended action
1. Verify that the calibration cylinder contains an adequate supply of a fresh test sample.
2. If a demand flow regulator calibration kit is used, use a different demand flow regulator to determine if the original one is functioning properly.
3. If necessary, set the correct flow rate with the sensor flow control valve.
4. If you cannot set the correct flow rate, check the sample line for obstructions or kinks.
5. If the calibration/response difficulties continue, replace the sensor as described later in this section.
6. If the calibration/response difficulties continue, contact RKI Instruments, Inc. for further instruction.

Replacing Components of the Hydrogen Sample-Draw Detector

This section includes procedures to replace the sensor, hydrophobic filter, and particle filter.

Replacing the Hydrogen Sensor

1. Turn off the controller.
2. Turn off power to the controller.
3. Open the housing door of the sample-draw detector.
4. Unscrew and remove the two screws that secure the sensor retaining plate, then lift the plate, connector, and sensor out of the housing.
5. Unplug the connector from the sensor.
6. Verify that you are using the correct replacement sensor (NC-6244A is printed on the sensor), then plug the sensor into the connector.
7. Place the sensor in the hydrogen sensor cavity, then position the retaining plate on the two standoffs.
8. Secure the retaining plate to the standoffs with the two screws you removed in step 4.
14 • 35-3001A-01H Hydrogen Sample-Draw Detector
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 19
9. Turn on power to the controller and place it into normal operation.
10. Turn on the controller.
CAUTION: Allow the replacement sensor to warm up for 5 minutes before you continue.
11. Calibrate the replacement sensor as described in “Calibration” on page 18.

Replacing the Amplifier

1. Turn off the controller.
2. Turn off power to the controller.
3. Open the housing door of the sample-draw detector.
4. Remove the detector and controller plug-in terminal strips by grasping the sides of the strips with your fingers. Let the terminal strips hang by their connected wires while you replace the amplifier.
5. Unscrew the screw in the lower left corner of the amplifier and remove the screw, lock washer, and flat washer. Be careful not to lose any of these parts.
6. Remove the old amplifier from the main circuit board.
7. Install the new amplifier in the same orientation as the old amplifier. See Figure 1.
8. Reinstall the screw, lock washer, and flat washer you removed in step 5.
9. Install the detector and controller plug-in terminals strips into their sockets on the new amplifier. If controller leads or detector cable leads were removed from the plug-in terminal strips during this procedure, refer to Table 2 and Table 3 below.
Table 2:Reconnecting the Amplifier to the Detector/Amp Terminal Strip
Amplifier Controller
Terminal Strip
PWR/SIG “-” -
PWR/SIG “S” S
PWR/SIG “+” +
Table 3:Reconnecting the LEL
Detector to the Amplifier
Amplifier Detector
Terminal Strip
DETECTOR “R” RED
DETECTOR “W” WHT
Detector/Amp Terminal
Strip on Main PCB
Detector Lead
DETECTOR “G” GREEN
DETECTOR “B” BLK
35-3001A-01H Hydrog en Sample-Draw Det ect or • 15
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 20
NOTE: When the sample-draw detector is first powered up with a new amplifier, the initial
output may be either high or below zero depending on the setting of the zero pot. Be sure to make arrangements so that this does not cause unwanted alarms.
10. Turn on power to the controller.
11. Turn on the controller and place it into normal operation.
12. Allow the sample-draw detector to warmup for 5 minutes.
13. Calibrate the sample-draw detector as described in “Calibration” on page 18 of this manual.

Replacing the Hydrophobic Filter

1. Turn off the controller.
2. Turn off power to the controller.
3. Open the housing door of the sample-draw detector.
4. Disconnect the filter from the rubber elbows on each end of the filter, then remove the filter from the sample-draw detector.
5. Install the new filter. Be sure the side of the filter marked “INLET” is connected to the elbow that is connected to the inlet fitting.
6. Turn on power to the controller.
7. Turn on the controller.
8. Verify that the flow rate is approximately 1.0 SCFH, then close the housing door.

Replacing the Particle Filter

1. Turn off the controller.
2. Turn off power to the controller.
3. If the particle filter is installed at the inlet fitting:
1. Disconnect the tubing routed to the sampling area, if installed, from the particle filter.
2. Disconnect the particle filter from the tubing stub.
3. Install the new particle filter onto the tubing stub. Be sure the arrow on the particle filter is
pointing toward the inlet fitting.
4. Reinstall the tubing routed to the sampling area.
4. If the particle filter is installed at the end of the inlet line:
1. Remove the particle filter from the tubing stub.
2. Install the new particle filter onto the tubing stub. Be sure the arrow on the particle filter is
pointing toward the inlet line.
5. Turn on power to the controller.
6. Turn on the controller.

Adjusting the Low Flow Setting

NOTE: Adjusting the low flow setting will cause a low flow alarm at the sample-draw detector
and a fail alarm at the controller. Be sure to put the controller into its calibration program or disable external alarms before performing this test.
The factory-set low flow setting is 0.6 SCFH (±0.1). To adjust the low flow setting:
16 • 35-3001A-01H Hydrogen Sample-Draw Detector
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 21
1. Use the sensor flow control valve to set the flow to 0.6 SCFH. Turn the valve’s knob
Outside of Case
Fitting Body
Front Ferrule
Back Ferrule
1/4" Tube
1/4" Tube
O-ring
Tube Nut
Inside of Case
counterclockwise to increase the flow and clockwise to decrease the flow.
If the sample-draw detector goes into low flow alarm before you can adjust the flow down to
0.6 SCFH, adjust the pressure switch adjustment screw 1/4 turn clockwise, then attempt to set the flow again. Repeat this step until you are able to adjust the flow to 0.6 SCFH.
NOTE: The pressure switch adjustment screw is accessible through a circular cutout in the
flowmeter circuit board.
2. Slowly turn the pressure switch adjustment screw counterclockwise just until the sample-draw detector goes into low flow alarm.
3. Turn the sensor flow control valve’s knob counterclockwise to increase the flow until the unit is out of low flow alarm when the reset switch is pressed and released.
4. Decrease the flow very slowly by turning the sensor flow control valve’s knob clockwise and verify that the low flow alarm is 0.6 SCFH (±0.1).
If the low flow alarm is set too high, turn the pressure switch adjustment screw slightly clockwise. Repeat steps 3 and 4 if necessary.
5. Use the sensor flow control valve to set the flow to 1.0 SCFH.
6. Make sure the sample-draw detector’s Fail LED is off.

Removing the Particle Filter’s Tubing Stub, if Necessary

A short tubing stub comes factory installed in the particle filter. It is used for connecting the particle filter to the inlet fitting. If you have installed the particle filter and no longer want it installed, you will need to remove the particle filter’s tubing stub from the inlet fitting and replace it with tubing.
Figure 6: Inlet Fitting with Tubing Stub
35-3001A-01H Hydrog en Sample-Draw Det ect or • 17
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 22
CAUTION: Do not pull the tubing stub downward to remove it.
1. Remove the particle filter from its tubing stub.
2. Unscrew the outside inlet fitting tube nut from the fitting body. The tubing stub should come out with the tube nut. Be careful not to lose the O-ring that may come out with the tubing stub. If you do lose the O-ring or if it is damaged, see “Parts List” on page 20 for the spare part number.
3. Push the tubing stub up and out of the front ferrule being careful not to lose the ferrule set. If you do lose the ferrule set or if it is damaged, see “Parts List” on page 20 for the spare part numbers.
4. Push the new tubing up through the inlet fitting tube nut and replace the ferrule set and the O­ring, if it came out, in the orientation shown in Figure 6.
5. Screw the inlet fitting tube nut back onto the fitting body. See “Connecting the Sample Lines to the Sample-Draw Detector” on page 8 for instructions to install a new piece of tubing into the fitting.

Calibration Frequency

Calibration

Although there is no particular calibration frequency that is correct for all sample draw detector applications, a calibration frequency of every 3 to 6 months is adequate for most applications. Unless experience in a particular application dictates otherwise, RKI Instruments, Inc. recommends a calibration frequency of every 3 months.
If an application is not very demanding, for example detection in a clean, temperature controlled environment where combustible gases are not normally present and calibration adjustments are minimal at calibration, then a calibration frequency of every 6 months is adequate.
If an application is very demanding, for example if combustible gases are present often and in significant concentrations or the environment is not well controlled, then more frequent calibration than every 3 months may be necessary.
This section describes how to calibrate the hydrogen sample-draw detector. It includes procedures to prepare for calibration, set the zero reading, set the response reading, and return to normal operation.
NOTE: This procedure describes calibration using a demand flow regulator.

Preparing for Calibration

1. Follow the instructions in the controller’s operator’s manual for entering calibration mode.
NOTE: If you can verify a fresh air environment, it is not necessary to use the zero air calibration
cylinder to set the zero reading.
2. Screw the regulator into a zero air calibration cylinder.
3. Open the housing door.
4. Set a voltmeter to measure in the millivolt (mV) range.
18 • 35-3001A-01H Hydrogen Sample-Draw Detector
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 23
5. Plug the positive lead into the red (+) amplifier test point; plug the negative lead into the black (-) amplifier test point labeled.
6. Use the following formula to determine the correct test points output for the calibrating sample.
Output (mV) = (calibrating sample/fullscale) X 400 + 100
For example, with a calibrating sample of 50% LEL and a fullscale setting of 100% LEL, the correct output is 300 mV.
300 (mV) = (50/100) X 400 +100

Setting the Zero Reading

1. Connect the sample tubing from the demand flow regulator to the sample-draw detector’s inlet line. This step is not necessary if you verified a fresh air environment earlier in this
procedure.
2. Allow the sample-draw detector to draw sample for one minute.
3. Verify a voltmeter reading of 100 mV (± 2 mV).
4. If necessary, use a small flat-blade screwdriver to adjust the zero potentiometer until the voltmeter reading is 100 mV (± 2 mV). If you used a zero air calibration cylinder to set the zero reading, proceed to step 5. If you verified a fresh air environment, proceed to the next section, Setting the Response Reading.
5. Disconnect the sample tubing from the inlet line.
6. Unscrew the regulator from the zero air calibration cylinder.

Setting the Response Reading

1. Screw the regulator into the hydrogen calibration cylinder.
2. Connect the sample tubing from the demand flow regulator to the sample-draw detector’s inlet line.
3. Allow the sample-draw detector to draw sample for one minute.
4. When the reading on the voltmeter stabilizes, verify that the reading matches the response reading (±2 mV) you determined earlier.
5. If necessary, use the span potentiometer on the amplifier to adjust the reading to match the correct response reading.
6. Disconnect the sample tubing from the inlet line.
7. Unscrew the regulator from the hydrogen calibration cylinder.

Returning to Normal Operation

1. Wait approximately one minute to allow the hydrogen reading to stabilize.
2. Remove the voltmeter leads from the amplifier test points.
3. Close the housing door.
4. Follow the instructions in the controller’s operator’s manual to exit the calibration mode.
5. Store the components of the calibration kit in a safe and convenient place.
35-3001A-01H Hydrog en Sample-Draw Det ect or • 19
Find Quality Products Online at: sales@GlobalTestSupply.com
Page 24

Parts List

Table 4 lists replacement parts and accessories for the sample-draw hydrogen detector.
Table 4: Parts List
Part Number Description
06-1248RK Sample tubing, 3/16 in. ID x 5/16 in. OD, specify length
06-1248RK-03 Sample tubing, 3/16 in. ID x 5/16 in. OD, 3 feet (for calibration kit)
17-2670 Inlet fitting O-ring
17-2671 Inlet fitting front ferrule
17-2672 Inlet fitting back ferrule and tube nut
30-1016RK Pump
33-0165RK Hydrophobic filter
33-0167RK Particle filter
57-1053RK-01 Amplifier (specify target gas when ordering)
61-0155RK Hydrogen sensor, plug-in type, LEL range
71-0307RK Operator’s Manual, 35-3001A-01H Sample-Draw Detector
81-0002RK-01 Calibration cylinder, 50% LEL hydrogen in air, 34 liter steel
81-0002RK-03 Calibration cylinder, 50% LEL hydrogen in air, 103 liter
81-0076RK-01 Zero air calibration cylinder, 34 liter steel
81-0076RK-03 Zero air calibration cylinder, 103 liter
81-1054RK Regulator, demand flow, for 34 liter aluminum, 58 liter, and 103 liter
calibration cylinders (cylinders with internal threads)
81-1055RK Regulator, demand flow, for 17 liter and 34 liter steel calibration cylinders
(cylinders with external threads)
20 • 35-3001A-01H Hydrogen Sample-Draw Detector
Find Quality Products Online at: sales@GlobalTestSupply.com
Loading...