RKI Instruments 35-3001A-01 User Manual

Page 1
35-3001A-01
Combustible Gas Sample-Draw Detector
Operator’s Manual
Part Number: 71-0306RK Revision: P4 Released: 9/17/14
www.rkiinstruments.com
Page 2
WARNING
Read and understand this instruction manual before operating detector . Improper use of the dete ctor could result in bodily harm or death.
Periodic calibration and maintenance of the detector is essential for proper operation and correct readings. Please calibrate and maintain this detector regularly! Frequency of calibration depends upon the type of use you have and the sensor types. T ypical calibration frequencies for most applications a re between 3 and 6 months, but can be required more often or less often based on your usage.
Page 3
Product Warranty
RKI Instruments, Inc. warrants gas alarm equipment s old by us t o be free from def ects in material s, workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge. This warrant y does not apply to those items which by their nature are subject to deterioration or consumption in normal service, and which must b e cleaned, repaired, o r replaced on a routine basis. Examples of such items are:
Warranty is voided by abuse including mechanical damage, alteration, rough handling, or repair procedures not in accordance with the operator’s manual. This warranty indicates the full extent of our liability, and we are not responsible for removal or replacement costs, local repair costs, transportation costs, or contingent expenses incurred without our prior approval.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI INSTRUMENTS, INC. INCLUDING BUT N OT LIMITED T O, THE WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL RKI INSTRUMENTS, INC. BE LIABLE FOR INDIRECT, INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
a) Absorbent cartridges d) Batteries b) Pump diaphragms and valves e) Filter elements c) Fuses
This warranty covers instruments and parts sold to users by authorized distributors, dealers, and representatives as appointed by RKI Instruments, Inc.
We do not assume indemnification for any accident or damage caused by the operation of this gas monitor, and our warranty is limited to the replacement of parts or our complete goods.
35-3001A-01 Combustible Gas Sample-Draw Detector
Page 4
Table of Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
External Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Internal Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Mounting the Combustible Gas Sample-Draw Detector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Removing the Inlet Fitting Tubing Stub, if Necessary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Connecting the Sample Lines to the Sample-Draw Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Wiring the Combustible Gas Sample-Draw Detector to a Controller. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Start Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Introducing Incoming Power. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Setting the Zero Reading. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Preventive Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Replacing Components of the Combustible Gas Sample-Draw D etector. . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Adjusting the Low Flow Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Calibration Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Preparing for Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Setting the Zero Reading. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Setting the Response Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Returning to Normal Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Page 5
Overview
This operator’s manual describes the 35-3001A-01 combustible gas sample-draw detector. This manual also describes how to install, start up, maintain, and calibrate the sample-draw detector when using it with a gas monitoring controller. A parts list at the end of this manual lists replacement parts and accessories for the sample-draw detector.
Specifications
Table 1 lists specifications for the combustible gas sample-draw detector. See the controller Operator’s Manual for information specific to the controller.
Target Gas and Detection Range Combustible gas: 0 - 100% LEL Input Power 24 VDC Nominal (18.5 VDC - 30 VDC) Construction (housing) Fiberglass/polyester (NEMA 4X) Dimensions 8.5 in. H x 6.5 in. W x 4.25 in. D
Table 1: Specifications
W eight 4.5 lbs. Sampling Method Sample-draw Sample Flow 3.0 SCFH typical, with no inlet or exhaust line Flow to Sensor 1.0 SCFH (nominal) Maximum Recommended Inlet/
Exhaust Line Length for 1/4” O.D. x 1/8” I.D. Tubing
Maximum Recommended Inlet/ Exhaust Line Length for 1/4” O.D. x 0.170” I.D. Tubing
Response Time 90% in 30 seconds Accuracy ± 5% of reading or ± 2% LEL (whichever is greater)
WARNING: When using the 35-3001A-01, you must fo llow the instructions and warnings in
this manual to assure proper and safe operation of the 35-3001A-01 and to minimize the risk of personal injury. Be sure to maintain and periodically calibrate t h e 35-3001A- 01 as describe d in this manual.
Inlet
100 feet 0 feet
50 feet 50 feet
Inlet
100 feet 100 feet
Exhaust
Exhaust
35-3001A-01 Co m bu s t i b l e G a s Sample-Draw De t e c t o r • 1
Page 6
Description
SensorFlow
Control Valve
Fail LED
PilotLED
This section describes the components of the combustible gas sample-draw detector.
Zero Pot
Span P ot
Pum p Connector
O
W
X
+
Y G
B
E
_
N
Mounting Foo t, 4X
Flowmeter Circuit
Board Connector
Factory Set Pots
Test Points
100-500 mV Range
Sensor
Cur r ent
148mA
W
GBR
S
LEL
PWR / SIG
Controller
Terminal Strip
Oxy gen Sensor Term i nal Strip
Pum p
LEL Sensor
Pum p Reset
Switch
Flowm eter
Pressure
Switch
Adjustment
Screw
(behind PCB)
Hydrophobic
Filt er
This End To Inlet
Fit tin g
Particle Filter
Detector Term i nal
Strip
LEL Am plifier
+
S
+
AMP2AMP
+
S S R
+
AMP1
AMP2
+
W B
LEL/IR
OXY
W
WG
+ OXYGLEL/IR
GS R
B
3/4" Conduit Hub
LEL F low Block
Detector/Amp Term i nal Strip (Factory W ired)
Int erconnect Term i nal
Strip
INLET
Inlet
Fitting
EXHAUST
Exhaust
Fitting
Figure 1: Combusti bl e Gas Sa mp l e-Draw Detector Component Location
2 • 35-3001A-01 Combusti ble Gas Sample-Draw D etector
Page 7
External Components
This section describes the sample-draw detector’s external components.
Housing
The sample-draw detector’s fiberglass housing is weather- and corrosion-resistant. It is suitable fo r installation where general purpose equipment is in use.
The housing door is hinged on the left side and is secured by two latches on the right side. The flowmeter and status LEDs are visible through a window in the housing door.
Four mounting feet are attached to the back of the housing (one at each corner). Use the mounting feet to install the housing to a vertical surface.
Sample Fittings
The sample fittings are located on the left side of the bottom of the housing. The inlet fitting is near the front of the housing and the exhaust fitting is near the back of the housing. The inlet fitting has a short factory installed tubing stub for use with the particle filter. The sample fittings accept 1/4 in. rigid tubing. See the Installation section on page 7 to connect tubing to the sample fittings.
Inlet Fitting Tubing Stub
A short tubing stub comes factory installed in the inlet fittin g. It is used for connecting the particle filter. If the particle filter is not used, or if you install the particle filter in a different location, the tubing stub needs to be removed and replaced with tubing.
Conduit Hub
One 3/4” conduit hub is located on the right sid e of the bottom of the hous ing. It is used fo r routing wiring into the housing by using conduit or an appropriate cable bushing.
Internal Components
This section describes the sample-draw detector’s internal components (see Figure 1). Figure 2 illustrates how the gas sample moves through the flow system.
Exhaust
Inlet
Pressure Switch
Flowmeter PCB
Restrictor
Sensor
Flow
Control
Valve
Flowmeter
Sensor
LEL
ParticleFilter
Pump
HydrophobicFilter
Figure 2: Combustible Gas Sample-Draw Detector Flow Diagram
35-3001A-01 Co m bu s t i b l e G a s Sample-Draw De t e c t o r • 3
Page 8
Main Circuit Board
The main circuit board includes the detect or/amp terminal strip, intercon nect terminal st rip, oxygen sensor terminal strip, pump connector, and flowmet er circuit board connector (see Figure 1).
Detector/Amp Terminal Strip
The detector/amp terminal strip is the upper twelve-point terminal strip in the bot tom right corner of the main circuit board. Use the detector/amp terminal strip to connect the amplifier to the main circuit board.
NOTE: The amplifier is factory-wired to the main circuit board. See the “Installation” section on
page 7 for all wiring procedures related to the sample-draw detector.
Interconnect Terminal Strip
The interconnect terminal strip is the lower twelve-point terminal strip in the bottom right corner of the main circuit board. Use the interconnect termin al stri p to con n ect the sam ple-draw detector to a controller.
Oxygen Sensor Terminal Strip
The oxygen sensor terminal strip is a two-point terminal strip in the upper right corner of the main circuit board. This terminal strip is not used in this version of the sample-draw detector.
Pump Connector
The pump connector is the two-point connector below the oxygen terminal strip in the upper right corner of the main circuit board. Use the pump co nnector to connect the pump to the main circuit board.
NOTE: The pump is factory-wired to the main circuit board. See “Installation” on page 7 for all
wiring procedures related to the sample-draw detector.
Flowmeter Circuit Board Connector
The flowmeter circuit board connector is a six-position connector in the upper left corner of the main circuit board. Use the flowmeter circuit board connector to connect the flowmeter circuit board to the main circuit board.
NOTE: The flowmeter circuit board is factory wired to the main circuit board. See “Installation”
on page 7 for all wiring procedures rela ted to the sample-draw detector.
Flowmeter Circuit Board
The flowmeter circuit board is mounted to the left side of the main circuit board using standoffs. It includes the flowmeter, sensor flow control valve, status LEDs, pressure switch, and pump reset switch.
Flowmeter
The flowmeter is mounted to the right side of the flowmeter circuit board. You can see it through the window in the door. A ball in the flowmeter column indicates the flow rate to the sensor. The flowmeter measures the flow in the range 0.2 to 2.0 SCFH (Standard Cubic Feet per Hour). The optimum flow rate is 1.0 SCFH.
Sensor Flow Control Valve
The sensor flow control valve is mounted to the flowmeter circuit board above the flowmeter. The sensor flow control valve adjusts the flow rate to the detector. Turn the valve’s knob
4 • 35-3001A-01 Combusti ble Gas Sample-Draw D etector
Page 9
counterclockwise to increase the flow and clockwise to decrease the flow.
Status LEDs
T wo status LEDs are abov e the flowmeter. They are also visible through the windo w in t he hou sing door. The green Pilot LED is on when the sample-draw detector is receiving power from the controller. The red Fail LED is on when the sample flow rate is below the low flow level.
Pressure Switch
The pressure switch is mounted to the back of the flowmeter circuit board. The pressure switch monitors the flow rate of the incoming gas sample.
If the flow rate falls below the preset low flow level, the pressure switch causes the Fail LED to turn on and interrupts the signal from the detector. The interrupted detector signal causes a fail condition at the controller. The low flow level is factory-set at 0.6 SCFH (±0.1 SCFH). See “Adjusting the Low Flow Setting” on page 17 to adjust this setting.
Pump Reset Switch
The pump reset switch is located to the left of the status LEDs. When a low flow condition occurs, the pump will be shut off. To reset the low flow condition and start the pump again, press and hold the pump reset switch for about 2 seconds, then release.
Hydrophobic Filter
The hydrophobic filter is located toward the bottom left of the main circuit board. The filter prevents particulates and water in the incoming gas sample from damaging the flow and detection systems. Replace the filter when it appears dirty, discolored, or clogged.
Pump
The pump is mounted to the right side of the main circuit board. The pump pulls the gas sample into the sample-draw detector. The pump operates on 24 VAC, which is generated from the 24 VDC supplied by the controller.
Combustible Gas Sensor
The catalytic combustible gas sensor detects comb ustib le gas in the % LEL range. It us es a catalytic element for detection. The reaction of gas with oxygen on the catalyst causes a change in the resistance of the element which changes the curr ent flowing through it. The current is proportional to the detection range of the sample-draw detector.
The combustible gas sensor is installed in the flow block and the flow block is mounted to the middle of the main circuit board. The sensor housing includes a sintered metal flame arrestor on one end that allows gas to diffuse int o the sensor. On the other end, five pins extend from the sensor and plug into the socket connector. The connector allows you to replace the sensor without disconnecting the wiring. Four leads extending from the connector are factory wired to the amplifier.
Amplifier
The amplifier converts the electrical output from the detector to a 4 to 20 mA signal that corresponds to the detection range and transmits the signal to a gas monitoring controller. A label on the amplifier indicates the detector drive current. This d rive cu rrent is factory s et and is dictated by the combustible gas to which the detector is calibrated. Consult RKI Instruments, I nc. if you plan to change the calibration gas for the sample-draw detector, for example from a methane calibration to a hexane calibration. The amplifier is mounted on the top middle edge of the main circuit board. It consists of the zero pot, span pot, controller terminal strip, detector terminal strip, and test points.
35-3001A-01 Co m bu s t i b l e G a s Sample-Draw De t e c t o r • 5
Page 10
Zero Pot
The zero pot is located in the upper right corner of the amplifier (see Figure 1). Use a small flat blade screwdriver to turn the zero pot’s adjustment screw and adjust the amplifier’s zero (f res h air) output during the star t-up and cali bration procedures. Turn the adjustment screw clockwise to increase the fresh air output and counterclockwis e to decrease the fresh air output.
Span Pot
The span pot is located below the zero pot (see Figure 1). Use a small flat blade screwdriver to turn the span pot’s adjustment screw and adjust the amplifier’s gas response output during the calibration procedure. Turn the adjustment screw clockwise to increase the span output and counterclockwise to decrease the span output.
CAUTION: The amplifier includes additional pots. They are factory-set. Do not adjust them.
Controller and Detector Terminal Strips
The controller terminal strip and detector terminal strip are three- and four-position plug-in style terminal strips, respectively. The controller terminal strip is located on the right side of the amplifier and the detector terminal strip is to the left of it. Both terminal strips are factory wired to the sensor and main circuit board.
T est Points
The test points are on the left side of the amplifier (see Figure 1). The test points produce a 100 mV to 500 mV output that corresponds to the sample-draw detector’s 4 to 20 mA output. Use the test points and a voltmeter to measure the amplifier’s output during the start-up and calibration procedures. The black test point in the upper left corner is the negative (-) test point and the red test point to the left of the zero and span pots is the positive (+) test point.
6 • 35-3001A-01 Combusti ble Gas Sample-Draw D etector
Page 11
Installation
This section describes procedures to mount the sample-draw detector in the monitoring environment and wire the sample-draw detector to a controller.
Mounting the Sample-Draw Combustible Gas Detector
1. Select the mounting site. Consider the following when you select the mounting site:
Is there enough room to open the housing door and make wi ring and tubi ng connect io ns at the bottom of the housing?
Make sure there is sufficient room to perform start-up, maintenance, and calibration procedures.
Are the flowmeter and status LEDs visible?
6.50
1.13
Ø.30x.50(4X)MOUNTING
.78
.40
8.50
Pil o t Fail
Sa mple Drawing
Gas Dete ctor
ww w.rkiinstruments.com
4.00
35-3001
8.88
.80
.85
4.50
NOTE: Housing is4.3 inchesdeep
Figure 3: Outline and Mounting Dimensions
35-3001A-01 Co m bu s t i b l e G a s Sample-Draw De t e c t o r • 7
Page 12
2. Close and latch the housing door.
NOTE: The sample-draw detector is shipped with the mounting feet “tucked under” the housing
to protect the mounting feet during shipment.
3. Slightly loosen the screw that secures the mounting foot to the housing, then rotate the mounting foot 180 degrees (see Figure 3).
4. Tighten the screw that secures the mounting foot to the housing.
5. Repeat steps 3 and 4 for the remaining three mounting feet.
6. Position the sample-draw housing on a verti cal surface at eye level (4 1/2 to 5 feet from the floor).
7. Insert 1/4 inch screws through the slots in the mounting feet to secure the housing to the mounting surface.
Removing the Inlet Fitting Tubing Stub, if Necessary
A short tubing stub comes factory installed in the inlet fittin g. It is used for connecting the particle filter. If the particle filter is not used, or if you install the particle filter in a different location, the tubing stub needs to be removed and replaced with tubing.
O-ring
Tube Nut
1/4 " Tube
Inside of Case
Outsideof Case
Fitting Body
Back Ferrule
Front Ferrule
1/4 " Tube
Figure 4: Inlet Fitting with Tubing Stub
CAUTION: Do not pull the tubing stub downward to remove it.
1. Unscrew the outside inlet fitting tube nut from the fitting body. The tubing stub should come out with the tube nut. Be careful not to lose the O-ring that may come out with the tubing stub.
2. Push the tubing stub up and out of the front ferrul e being careful not to lose the ferrule set.
3. Push the new tubing up through the inlet fitting tube nut and replace the ferrule set and the O-
8 • 35-3001A-01 Combusti ble Gas Sample-Draw D etector
Page 13
ring, if it came out, in the orientation shown in Figure 4.
4. Screw the inlet fitting tube nut back onto the fitting body. See the next section for instructions to install a new piece of tubing into the fitting.
Connecting the Sample Lines to the Sample-Draw Detector
See Figure 4 for the fitting layout.
Installing the Inlet Line with Particle Filter
1. Connect the particle filter to the tubing stub on the inlet by pushing the flexible tubing on the filter onto the stub. Be sure the arrow on the particle filter is pointing toward the inlet fitting.
2. Connect a length of sample tubing to the other side of the particle filter and route it to the sampling area. 1/4” O.D. rigid polypropylene, Teflon, or flexible polyurethane tubing may be used. RKI Instruments , Inc. recomm ends using either 1/4” O.D. x 1/8” I.D. or 1/4” O.D. x
0.170” I.D. tubing based on your length requirements. See “Specifications” on page 1 for maximum tubing lengths based on tubing size.
CAUTION: If you use flexible sample tubing (polyurethane is acceptable), use an appropriate
insert if necessary to provide support to the tubing and ensure a good seal when connecting to the flexible stub on the filter.
CAUTION: Avoid lo ops or slumps in the incoming sample lin e. To reduce r esponse time, keep the
incoming sample line as short as possible.
Installing the Inlet Line without Particle Filter
1. Remove the tubing stub as described in “Removing the Inlet Fitting Tubing Stub, if Necessary” on page 8.
2. Loosen the nut on the inlet fitting until 3 threads are visible.
3. Push 1/4” O.D. rigid polypropylene or rigid Teflon sample tubing into the fitting until it stops. Flexible polyurethane tubing may be used with an appropriate insert. RKI Instruments, Inc. recommends using either 1/4” O.D. x 1/8” I.D . or 1/4” O.D. x 0. 170” I.D. t ubing based o n your length requirements. See “Specifications” on page 1 for maximum tubing lengths based on tubing size.
CAUTION: If you use flexible sample tubing (polyurethane is acceptable), use an appropriate
insert to seal the connection between the tubing and the inlet fitting.
4. Hand tighten the nut on the inlet fitting so the ferrules clamp on the tubing. If the tube nut is removed, see Figure 4 for the arrangement of the fitting components.
5. Route tubing from the inlet fitting to the sample area. See “Specifications” on page 1 for maximum tubing lengths based on tubing size.
CAUTION: Avoid lo ops or slumps in the incoming sample lin e. To reduce r esponse time, keep the
incoming sample line as short as possible.
35-3001A-01 Co m bu s t i b l e G a s Sample-Draw De t e c t o r • 9
Page 14
Installing the Exhaust Line
1. Loosen the nut on the exhaust fitting until 3 threads are visible.
2. Push 1/4” O.D. rigid polypropylene or rigid Teflon sample tubing into the fitting until it stops. Flexible polyurethane tubing may be used with an appropriate insert. RKI Instruments, Inc. recommends using either 1/4” O.D. x 1/8” I.D . or 1/4” O.D. x 0. 170” I.D. t ubing based o n your length requirements. See “Specifications” on page 1 for maximum tubing lengths based on tubing size.
CAUTION: If you use flexible sample tubing (polyurethane is acceptable), use an appropriate
insert to seal the connection between the tubing and the inlet fitting.
3. Hand tighten the nut on the exhaust fitting so the ferrules clamp on the tubing. If the tube nut is removed, see Figure 4 for the arrangement of the fitting components.
4. Route the opposite end of the tu bing t o an op en area where the sample can safely disp erse or to an exhaust duct. See “Specifications ” on page 1 for maximum tubing lengths based on tubing size.
Wiring the Sample-Draw Combustible Gas Detector to a Controller
WARNING : Always verify that the controller is off and that power to the controller is off
before you make wiring connections.
1. Turn off the controller.
2. Turn off power to the controller.
3. Unlatch and open the housing door of the sample-draw detector.
4. Guide a three-conductor , shielded cabl e or thr ee wires in co nduit thro ugh the cond uit hub at the bottom of the sample-d raw housing. A minimum of 18 AWG wire is recommen ded.
5. Connect the cable to the sample-draw detector’s interconnect terminal strip as shown in Figure
5.
6. Close and latch the housing door of the sample-draw detector.
CAUTION: If using shielded cable, leave the cable shield’s drain wire insulated and
disconnected at the sample-draw detector. You will connect the opposite end of the drain wire at the controller.
7. Route the cable or wires in conduit leading from the sample-draw detector through one of the conduit hubs at the controller.
10 • 35-3001A-01 Combustible Gas Sample-Draw Detecto r
Page 15
8. Connect the wires to the applicable detector/transmitter terminal strip at th e contr oller as shown in Figure 5. Refer to the controller operator’s manual and the controller detector head specification sheet for the 35-3001A-01 for detector/terminal strip connections specific to the controller.
Not Used on This Version
Red
White
Green
Black
To LEL Sensor
Sensor
Current
148mA
W
LEL
ZE RO
SP AN
GBR
S
PWR / SIG
AMP 1
AMP 2
OXY LEL /IR
Controller or
Recording
Dev ice
9. If shielded cable is used, connect the cable’s drain wire to an available chassis (earth) ground at
PATS201PATS201
PATS101PATS101
AMP 1
AMP 2
OXY
LEL/IR
+24VDC
4/20 Signal
- (DC G round)
Figure 5: Wiring the Sample-Draw Detector to a Controller
the controller. RKI controllers typically have a ground stud that can be used to ground the cable’s drain wire.
35-3 001A-01 Combustible Gas Sample- Draw Detector • 11
Page 16
Start Up
This section describes procedures to start up the sample-draw detector and place the sample-draw detector into normal operation.
Introducing Incoming Power
1. Complete the installation procedures described earlier in this m anual.
2. Verify that the wiring is correct and secure. Refer to the controller operator’s manual for connections at the controller.
3. Turn on or plug in the power to the controller, then turn on the controll er.
4. Verify that the sample-draw detector’s Pilot LED is on.
5. Verify that the controller is on and operating properly. Refer to the controller operator’s manual.
6. Verify that the flowmeter indicates a flow rate of approximately 1.0 SCFH. If necessary, use the sensor flow control valve to adjust the flow rate. Turn the valve’s knob counterclockwise to increase the flow and clockwise to decrease the flow.
NOTE: The following step tests for leaks in the sample line. This test will cause a low flow
condition at the sample-draw detector and a fail condition at the controller. Be sure to put the controller into its calibration program or disable external alarms before performing this test.
7. Verify that the incoming sample line is not leaking. To test the sample line, press and hold the reset switch and plug the open end of the sample line with your thumb. If the flowmeter ball drops to the bottom of the flowmeter, the incoming sample line is not leaking.
8. Remove your thumb from the sample line, release the reset switch, and verify the flowmeter returns to a normal flow rate.
9. Enable alarms or place the controller in normal operation.
CAUTION: Allow the sample-draw detector to warm up for 5 minutes before you continue with
the next section, “Setting the Zero Reading.”
Setting the Zero Reading
CAUTION: If you suspect the presence of combustible gas in the monitoring environment, use
the calibration kit and the zero air calibration cylinder to introduce “fresh air” to the sensor and verify an accurate zero setting. See “Calibration” on page 18 for instructions on using a zero air calibration cylinder for setting the zero reading.
1. Verify that the sample-draw detector is sampling a fresh air environmen t (env iron ment known to be free of combustible gas).
2. Open the housing door.
3. Set a voltmeter to measure in the millivolt (mV) range.
4. Plug the voltmeter leads into the test points on the amplifier. Plug the positive lead into the red (+) test point; plug the negative lead into the black (-) test point.
5. Verify a voltmeter reading of 100 mV (± 2 mV).
12 • 35-3001A-01 Combustible Gas Sample-Draw Detecto r
Page 17
6. If necessary, use a small flat-blade screwdriver to adjust the zero potentiometer until the
7. Remove the voltmeter leads from the test points.
8. Close the housing door.
Maintenance
This section describes maintenance procedures. It includes preventive maintenance procedures. This section also includes procedures to troubleshoot the sample-draw detector, replace components of the sample-draw detector, and adjust the low flow setting.
Preventive Maintenance
This section describes a preventive maintenance schedule to ensure the optimum performance of the sample-draw detector. It includes daily, monthly, and qu arterly procedures.
Daily Visual Checks
1. Verify that the Pilot LED is on.
2. Verify that the flowmeter indicates a flow rate of approximately 1.0 SCFH. If necessary use the
voltmeter reading is 100 mV (± 2 mV).
sensor flow control valve to adjust the flow rate to 1.0 SCFH. Turn the valve’s knob counterclockwise to increase the flow and clockwise to decrease the flow.
3. Verify a display reading of 0 %LEL at the controller or a reading of 100 mV at the amplifier test points. Investigate significant changes in the display reading.
Monthly Response Test
This procedure describes a test to verify that the sample-draw detector responds properly to the target gas.
NOTE: To reduce the response time of this test, use a short incoming sample line. If the sample-
draw detector’s sample line is long, connect a shorter line for this test. Make sure you reconnect the sample line after you complete this procedure.
NOTE: Performing a response test on the sample-draw detector may cause alarms. Be sure to put
the controller into its calibration program or disable external alarms before performing this test.
Preparing for the response test
NOTE: This procedure describes the RKI calibration kit that includes a demand flow re gulator.
1. Verify that the display reading at the controller is 0 % LEL. If the display reading is not 0 % LEL, set the zero reading as described in “Calibration” on
page 18, then continue this procedure.
2. Set a voltmeter to measure in the millivolt (mV) range.
3. Open the housing door, then plug the voltmeter leads into the test points on the amplifier. Plug the positive lead into the red (+) test point; plug the negative lead into the black (-) test
point.
35-3 001A-01 Combustible Gas Sample- Draw Detector • 13
Page 18
4. Use the following formula to determine the correct test points output for the test sample.
Output (mV) = (calibrating sample/fullscale) X 400 + 100
For example, with a test sample of 50 %LEL and a fullscale setting of 100 %LEL, the correct output is 300 mV.
300 (mV) = (50/100) X 400 +100
Performing the response test
1. Screw the demand flow regulator into the calibration cylinder.
2. Connect the calibration tubing from the demand flow regulator to the inlet fi tting. Gas will begin to flow.
3. After approximately one minute, verify that the reading at the controller stabilizes within ± 20% of the concentration of the test sample. If the reading is not within ± 20% of the test sample, calibrate the sample-draw detector as described in “Calibration” on page 18.
4. Remove the calibration tubing from the inlet fitting, then reconnect the sample tubing to th e inlet fitting.
5. Remove the voltmeter leads from the amplifier test points.
6. Close the housing door.
7. Store the calibration kit in a safe place.
Quarterly Calibration
Calibrate the sample-draw detector as described in “Calibration” on page 18.
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended action for problems you may encounter with the sample-draw detector.
NOTE: This troubleshooting guide describes sample-draw detector problems only. See the
controller Operator’s Manual if the control ler exhibits any problems.
Fail Condition
Symptoms
The sample-draw detector’s Fail LED is on.
The monitoring device is operating properly but indicates a reading well below zero or a failure alarm.
Probable causes
The sample-draw detector’s flow rate is too low because of an obstructed sample line, failed pump, etc.
The sample-draw detector is malfunctioning.
The sensor and/or amplifier wiring is disconnected or misconnected.
Recommended action
1. At the sample-draw detector, set the correct flow rate with the sensor flow control val ve.
2. If you cannot set the correct flow rate, check the sample lines for obstructions or kinks.
3. Verify that the sensor wiring and amplifier wiring is correct and secure. “Wiring the Sample­Draw Combustible Gas Detector to a Controller” on page 10 describes sensor and amplifier wiring connections.
14 • 35-3001A-01 Combustible Gas Sample-Draw Detecto r
Page 19
4. Calibrate the sample-draw detector as described in “Calibration” on page 18.
5. If the fail condition continues, replace the sensor as described in “Replacing the Combustible Gas Sensor” on page 15.
6. If the fail condition continues, contact R KI Instruments, Inc. for further instruction.
Slow or No Response/Difficult or Unable to Calibrate
Symptoms
The sensor responds slowly or does not respond during the monthly response test.
Unable to accurately set the zero or response reading during the calibration procedure.
The sensor requires fre quent calibration.
Probable causes
The calibration cylinder is low, out-dated, or defective.
If a demand flow regulator cal ibra tion kit i s used, t he deman d flow r egulator is not functi oning properly.
The sample-draw detector’s flow rate is too low because of an obstructed sample line, failed pump, etc.
The sample-draw detector is malfunctioning.
Recommended action
1. Verify that the calibration cylinder contains an adequate supply of a fresh test sample.
2. If a demand flow regulator calibration kit is used, use a different demand flow regulator to determine if the original one is functioning properly.
3. If necessary, set the correct flow rate with the sensor flow control valve.
4. If you cannot set the correct flow rate, check the sample line for obstructions or kinks.
5. If the calibration/response difficulties continue, replace the sensor as described later in this section.
6. If the calibration/response difficulties continue, contact RKI Instruments, Inc. for further instruction.
Replacing Components of the Combustible Gas Sample-Draw Detector
This section includes procedures to replace the sensor, hydrophobic filter, and particle filter.
Replacing the Combustible Gas Sensor
1. Turn off the controller.
2. Turn off power to the controller.
3. Open the housing door of the sample-draw detector.
4. Unscrew and remove the two screws that secure t he sensor retaining plate, then lift the plate, connector, and sensor out of the housing.
5. Unplug the connector from the sensor.
6. Verify that you are using the correct replacement sensor (NC-6245 is printed on the sensor), then plug the sensor into the connector.
7. Place the sensor in the combustible gas sensor cavity, then position the retaining plate on the two standoffs.
8. Secure the retaining plate to the standoffs with the two screws you removed in step 4.
35-3 001A-01 Combustible Gas Sample- Draw Detector • 15
Page 20
9. Turn on power to the controller and place it into normal operation.
10. Turn on the contr ol ler.
CAUTION: Allow the replacement sensor to warm up for 5 minutes before you continue.
11. Calibrate the replacement sensor as described in “Calibration” on page 18.
Replacing the Amplifier
1. Turn off the controller.
2. Turn off power to the controller.
3. Open the housing door of the sample-draw detector.
4. Remove the detector and controller plug-in terminal strips by grasping t he sides of the strips with your fingers. Let the terminal strips hang by their connected wires while you replace the amplifier.
5. Unscrew the screw in the lower left corner of the amplifier and remove the screw , lock washer , and flat washer. Be careful not to lose any of these parts.
6. Remove th e old amplifier from the main circu it board.
7. Install the new amplifier in the same orientation as the old amplifier. See Figure 1.
8. Reinstall the screw, lock washer, and flat washer you removed in step 5.
9. Install the detector and controller plug-in terminals strips into their sockets on the new amplifier. If controller leads or detector cable leads were removed from the plug-in terminal strips during this procedure, refer to Table 2 and Table 3 below.
Ta ble 2:Reconnecting the Amplifier to the Detector/Amp Terminal Strip
Amplifier Controller
Terminal Strip
PWR/SIG “-” ­PWR/SIG “S” S PWR/SIG “+” +
Table 3:Reconnecting the LEL
Detector to the Amplifier
Amplifier Detector
T erminal Strip
DETECTOR “R” RED
DETECTOR “W” WHT
Detector/Amp Terminal
Strip on Main PCB
Detector Lead
DETECTOR “G” GREEN DETECTOR “B” BLK
16 • 35-3001A-01 Combustible Gas Sample-Draw Detecto r
Page 21
NOTE: When the sample-draw detector is first powered up with a new amplifier, the initial
output may be either hi g h o r b elo w zero depending on the setting of the zero pot. Be s ur e to make arrangements so that this does not cause unwanted alarms.
10. Turn on power to the controll er.
11. Tur n on the controller and place it into normal operation.
12. Allow the sample-draw detector to warmup for 5 minutes.
13. Calibrate the sample-draw detector as described in “Calibration” on page 18 of this manual.
Replacing the Hydrophobic Filter
1. Open the housing door of the sample-draw detector.
2. Disconnect the filter from the rubber elbows on each end of the filter, then remove the filter from the sample-draw detector.
3. Install the new filter. Be sure the side of the filter marked “INLET” is connected to the elbow that is connected to the inlet fitting.
4. Verify that the flow rate is approximately 1.0 SCFH, then close the housing door.
Replacing the Particle Filter
1. Disconnect the tubing routed to the sampling area, if installed, from the particle filter.
2. Disconnect the particle filter from the inlet fitting’s tubing stub.
3. Install the new particle filter onto the inlet fitting’s tubing stub. Be sure the arrow on the particle filter is pointing toward the inlet fitting.
4. Reinstall the tubing routed to the sampling area.
Adjusting the Low Flow Setting
NOTE: Adjusting the low flow setting will cause a low flow alarm at the sample-draw detector
and a fail alarm at the controller. Be sure to put the controller into its calibration program or disable external alarms before performing this test.
The factory-set low flow setting is 0.6 SCFH (±0.1). To adjust the low flow setting:
1. Use the sensor flow control valve to set the flow to 0.6 SCFH. Turn the valve’s knob counterclockwise to increase the flow and clockwise to decrease the flow.
If the sample-draw detector goes into low flow alarm before you can adjust the flow down to
0.6 SCFH, adjust the pressure switch adjustment screw 1/4 turn clockwise, then attempt to set the flow again. Repeat this step until you are able to adjust the flow to 0.6 SCFH.
NOTE: The pressure switch adjustment screw is accessible through a circular cutout in the
flowmeter circuit board.
2. Slowly turn the pressure switch adjustment screw counterclockwise just until the sample-draw detector goes into low flow alarm.
3. Turn the sensor flow control valve’s knob counterclockwise to increase the flow until the unit is out of low flow alarm.
4. Decrease the flow very slowly by turning the sensor flow control valve’s knob clockwise and verify that the low flow alarm is 0.6 SCFH (±0.1).
35-3 001A-01 Combustible Gas Sample- Draw Detector • 17
Page 22
If the low flow alarm is set too low, turn the pressure switch adjustment screw slightly clockwise. Repeat steps 3 and 4 if necessary.
5. Use the sensor flow control valve to set the flow to 1.0 SCFH.
6. Make sure the sample-draw detector’s Fail LED is off.
Calibration Frequency
Although there is no particular calibration frequency that is correct for all sample draw detector applications, a calibration frequency of ever y 3 to 6 months is adequate for most applications. Unless experience in a particular application dictates otherwis e, RKI Instruments , Inc. recommends a calibration frequency of every 3 months.
If an application is not very demanding, for example detection in a clean, temperature controlled environment where combustible gases are not normally pr esent and calibration adjustments are minimal at calibration, then a calibration frequency of every 6 months is adequate.
If an application is very demanding, for example if combustible gases are present often and in significant concentrations or the environment is not well controlled, then more frequent calibration than every 3 months may be necessary.
Calibration
This section describes how to calibrate the combustible gas sample-draw detector. It includes procedures to prepare for calibration, set the zero reading, set the response reading, and return to normal operation.
The standard calibration gas for the sample-draw detector is methane. The sample-draw detector may be calibrated to other combustible gases such as hexane or hydrogen. Use the correct calibration gas for your installation.
NOTE: This procedure describes calibration using a demand flow regulator.
Prepari ng for Ca libration
1. Follow the instructions in the controller’s operator’s manual for entering calibration mode.
NOTE: If you can verify a fresh air environment, it is not necessary to us e the zero air calib ration
cylinder to set the zero reading.
2. Screw the regulator into a zero air calibration cylinder.
3. Open the housing door.
4. Set a voltmeter to measure in the millivolt (mV) range.
5. Plug the positive lead into the red (+) amplifier test point; plug the negative lead into the black (-) amplifier test point.
18 • 35-3001A-01 Combustible Gas Sample-Draw Detecto r
Page 23
6. Use the following formula to determine the correct test points output for the calibrating sample.
Output (mV) = (calibrating sample/fullscale) X 400 + 100
For example, with a calibrating sample of 50% LEL and a fullscale setting of 100% LEL, the correct output is 300 mV.
300 (mV) = (50/100) X 400 +100
Setting the Zero Reading
1. Connect the sample tubing from the demand flow regulator to the sample-draw detector’s inlet line. This step is not necessary if you verified a fresh air environment earlier in this
procedure.
2. Allow the sample-draw detector to draw sample for one minute.
3. Verify a voltmeter reading of 100 mV (± 2 mV).
4. If necessary, use a small flat-blade screwdriver to adjust the zero potentiometer until the voltmeter reading is 100 mV (± 2 mV). If you used a zero air calibration cylinder to set the zero reading, proceed to step 5. If you verified a fresh air environment, proceed to the next section, Setting the Response Reading.
5. Disconnect the sample tubing from the inlet line.
6. Unscrew the regulator from the zero air calibration cylinder.
Setting the Response Reading
1. Screw the regulator into the combustible gas calibration cylinder.
2. Connect the sample tubing from the demand flow regulator to the sample-draw detector’s inlet line.
3. Allow the sample-draw detector to draw sample for one minute.
4. When the reading on the voltmeter stabilizes, verify that the reading matches the response reading (±2 mV) you determined earlier.
5. If necessary, use the span potentiometer on the amplifier to adjust the reading to match the correct response reading.
6. Disconnect the sample tubing from the inlet line.
7. Unscrew the regulator from the combustible gas calibration cylinder.
Returning to Normal Operation
1. Wait appro ximately one minute to allow the combustible gas reading to stabilize.
2. Remove the voltmeter leads from the amplifier test points.
3. Close the housing door.
4. Follow the instructions in the controller’s operator’s manual to exit the calibration mode.
5. Store the components of the calibration kit in a safe and convenient place.
35-3 001A-01 Combustible Gas Sample- Draw Detector • 19
Page 24
Parts List
Table 2 lists replacement parts and accessories for the sample-draw combustible gas detector.
Table 4: Parts List
Part Number Description
06-1248RK Sample tubing, 3/16 in. x 5/16 in., specify length 06-1248RK-03 Sample tubing, 3/16 in. x 5/16 in., 3 feet (for calibration ki t) 30-1016RK Pump 33-0165RK Hydrophobic filter 33-0167RK Particle filter 57-1053RK-01 Amplifier (specify target gas when ordering) 71-0306RK Operator’s Manual, 35-3001A-01 Sample-Draw Detector 81-0002RK-01 Calibration cylinder, 50% LEL hydrogen in air 34 liter 81-0002RK-03 Calibration cylinder, 50% LEL hydrogen in air, 103 liter 81-0007RK-01 Calibration cylinder, 50% LEL hexane in air, 34 liter 81-0012RK-01 Calibration cylinder, 50% LEL methane in air, 34 liter 81-0012RK-03 Calibration cylinder, 50% LEL methane in air 103 liter 81-0076RK-01 Zero air calibration cylinder (34 liter) 81-0076RK-03 Zero air calibration cylinder (103 liter) 81-1054RK Regulator, demand flow, for 58- and 103-liter cylinders 81-1055RK Regulator, demand flow, for 17- and 34-liter steel cylinders NC-6245 Combustible gas sensor, plug-in type, LEL range
20 • 35-3001A-01 Combustible Gas Sample-Draw Detecto r