Read and understand this instruction manual before operating
detector. Improper use of the detector could result in bodily harm
or death.
Periodic calibration and maintenance of the detector is essential
for proper operation and correct readings. Please calibrate and
maintain this detector regularly! Frequency of calibration
depends upon the type of use you have and the sensor types.
Typical calibration frequencies for most applications are between
3 and 6 months, but can be required more often or less often
based on your usage.
35-3001-01H-DIL Hydrogen Sample-Draw Detector
Product Warranty
RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials,
workmanship, and performance for a period of one year from date of shipment from RKI
Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our
option, free of charge. This warranty does not apply to those items which by their nature are subject
to deterioration or consumption in normal service, and which must be cleaned, repaired, or replaced
on a routine basis. Examples of such items are:
Warranty is voided by abuse including mechanical damage, alteration, rough handling, or repair
procedures not in accordance with the operator’s manual. This warranty indicates the full extent of
our liability, and we are not responsible for removal or replacement costs, local repair costs,
transportation costs, or contingent expenses incurred without our prior approval.
THIS W ARRAN T Y IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER
WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND
ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI
INSTRUMENTS, INC. INCLUDING BUT N OT LIMITED TO, THE WARRANTY
OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN
NO EVENT SHALL RKI INSTRUMENTS, INC. BE LIABLE FOR INDIRECT,
INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND
CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS
PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
a) Absorbent cartridgesd) Batteries
b) Pump diaphragms and valvese) Filter elements
c) Fuses
This warranty covers instruments and parts sold to users by authorized distributors, dealers, and
representatives as appointed by RKI Instruments, Inc.
We do not assume indemnification for any accident or damage caused by the operation of this gas
monitor, and our warranty is limited to the replacement of parts or our complete goods.
This operator’s manual describes the 35-3001-01H-DIL hydrogen sample-draw detector. This
manual also describes how to install, start up, maintain, and calibrate the sample-draw detector
when using it with a gas monitoring controller. A parts list at the end of this manual lists
replacement parts and accessories for the sample-draw detector.
About the 35-3001-01H-DIL
The 35-3001-01H-DIL is designed to monitor hydrogen in an inerted area (an area that has no, or
almost no, oxygen present). The catalytic LEL sensor used to detect hydrogen must have at least
5% oxygen present to detect gas properly. A dilution fitting installed in the flow system before the
LEL sensor allows enough air to enter the gas sample so that the LEL sensor has enough oxygen to
function properly. The dilution ratio is approximately 1:1 but will vary slightly with increased
sample line lengths. The longer the sample line, the more dilution air will enter the gas sample
through the dilution fitting. Because of this, it is important when calibrating the detector head to
apply the calibration gas to the sample line inlet if a sample line is installed. In addition, because
the detector head includes internal dilution, it must be calibrated using a sample bag instead of a
demand flow regulator. See “Calibration” on page 18 for detailed calibration instructions.
Specifications
Table 1 lists specifications for the hydrogen sample-draw detector. See the controller Operator’s
Manual for information specific to the controller.
Table 1: Specifications
Target Gases and Detection RangeHydrogen: 0 - 100% LEL
Input Power24 VDC Nominal (18.5 VDC - 30 VDC)
Construction (housing)Fiberglass/polyester (NEMA 4X)
Dimensions8.5 in. H x 6.5 in. W x 4.25 in. D
Weight4.5 lbs.
Sampling MethodSample-draw
Sample Flow3.0 SCFH typical, with no inlet or exhaust line
Low Flow Setpoint1.0 ± 0.1 SCFH, measured in the sample inlet line
Sensor Flow0.8 SCFH (nominal)
Maximum Recommended Inlet/
Exhaust Line Length for
1/4” O.D. x 1/8” I.D. Tubing
Inlet
100 feet0 feet
50 feet50 feet
Exhaust
Maximum Recommended Inlet/
Exhaust Line Length for
1/4” O.D. x 0.170” I.D. Tubing
Response Time90% in 30 seconds
Inlet
100 feet100 feet
35-3001-01H-DIL Hydrogen Sample-Draw Detecto r • 1
Exhaust
Table 1: Specifications
Accuracy± 5% of reading or ± 2% LEL (whichever is greater)
WAR NING:When using the 35-3001-01H-DIL, you must follow the instructions and
warnings in this manual to assure proper and safe operation of the
35-3001-01H-DIL and to minimize the risk of personal injury. Be sure to
maintain and periodically calibrate the 35-3001-01H-DIL as described in this
manual.
2 • 35-3001-01H-DIL Hydrogen Sample-Draw Detector
Description
O
X
Y
G
E
N
B
+
W
Oxygen
Sensor
Terminal
Strip
Pump
LEL Sensor
Dilution
Fitting
WBG
LEL Flow
Block
_
AMP 2SOXY
+
S
+
+
AMP
Sensor Flow
Control Valve
Fail LED
Pilot LED
Pressure
Switch
Adjustment
Screw
(behind PCB)
Flowmeter
WBG
Interconnect
Terminal
Strip
Pump Reset
Switch
OXY
GW
+
+
AMP 1
S
+
S
AMP 2
Hydrophobic
Filter
LEL/ IR
R
R
Flowmeter Circuit
Board Connec tor
Pump Connector
LEL/ IR
3/4" Conduit Hub
Particle Filter
Exhaust
Fitting
Inlet
Fitting
This End To
Inlet Fitting
Mounting
Foot, 4X
EXHAUST
INLET
Detector/Amp
Terminal Strip
(Factory
Wired)
This section describes the components of the hydrogen sample-draw detector.
35-3001-01H-DIL Hydrogen Sample-Draw Detecto r • 3
External Components
LEL
Sensor
Flowmeter
Hydrophobic Filter
Pump
Exhaust
Pressure Sw itch
Flowmeter PCB
Sensor Flow
Contr ol Val v e
Restrictor
Inlet
Dilution Fitting,
appr ox. 1:1 ratio
Particle Filter
This section describes the sample-draw detector’s external components.
Housing
The sample-draw detector’s fiberglass housing is weather- and corrosion-resistant. It is suitable for
installation where general purpose equipment is in use.
The housing door is hinged on the left side and is secured by two latches on the right side. The
flowmeter and status LEDs are visible through a window in the housing door.
Four mounting feet are attached to the back of the housing (one at each corner). Use the mounting
feet to install the housing to a vertical surface.
Sample Fittings
The sample fittings are located on the left side of the bottom of the housing. The inlet fitting is near
the front of the housing and the exhaust fitting is near the back of the housing. The sample fittings
accept 1/4 in. rigid tubing. See the Installation section on page 7 to connect tubing to the sample
fittings.
Particle Filter
A particle filter with a tubing stub on one end is shipped with the instrument but it is not factory
installed. If the particle filter is installed directly to the inlet fitting, the tubing stub must be used. If
the particle filter is installed somewhere else, like at the end of the inlet line, the tubing stub can be
used or removed.
Conduit Hub
One 3/4” conduit hub is located on the right side of the bottom of the housing. It is used for routing
wiring into the housing by using conduit or an appropriate cable bushing.
Internal Components
This section describes the sample-draw detector’s internal components (see Figure 1). Figure 2
illustrates how the gas sample moves through the flow system.