RKI Instruments 35-3001-01H-DIL, 35-3001-06H-DIL Operator's Manual

35-3001-01H-DIL
Hydrogen Sample-Draw Detector
Operator’s Manual
Part Number: 71-0381 Revision: A Released: 8/3/17
www.rkiinstruments.com
WARNING
Read and understand this instruction manual before operating detector. Improper use of the detector could result in bodily harm or death.
Periodic calibration and maintenance of the detector is essential for proper operation and correct readings. Please calibrate and maintain this detector regularly! Frequency of calibration depends upon the type of use you have and the sensor types. Typical calibration frequencies for most applications are between 3 and 6 months, but can be required more often or less often based on your usage.
35-3001-01H-DIL Hydrogen Sample-Draw Detector
Product Warranty
RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge. This warranty does not apply to those items which by their nature are subject to deterioration or consumption in normal service, and which must be cleaned, repaired, or replaced on a routine basis. Examples of such items are:
Warranty is voided by abuse including mechanical damage, alteration, rough handling, or repair procedures not in accordance with the operator’s manual. This warranty indicates the full extent of our liability, and we are not responsible for removal or replacement costs, local repair costs, transportation costs, or contingent expenses incurred without our prior approval.
THIS W ARRAN T Y IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI INSTRUMENTS, INC. INCLUDING BUT N OT LIMITED TO, THE WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL RKI INSTRUMENTS, INC. BE LIABLE FOR INDIRECT, INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
a) Absorbent cartridges d) Batteries
b) Pump diaphragms and valves e) Filter elements
c) Fuses
This warranty covers instruments and parts sold to users by authorized distributors, dealers, and representatives as appointed by RKI Instruments, Inc.
We do not assume indemnification for any accident or damage caused by the operation of this gas monitor, and our warranty is limited to the replacement of parts or our complete goods.
35-3001-01H-DIL Hydrogen Sample-Draw Det ect or
Table of Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
External Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Internal Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Mounting the Sample-Draw Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Connecting the Sample Lines to the Sample-Draw Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Wiring the Sample-Draw Detector to a Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Start Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Introducing Incoming Power. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Setting the Zero Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Preventive Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Replacing Components of the Sample-Draw Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Adjusting the Low Flow Setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Removing the Particle Filter’s Tubing Stub, if Necessary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Calibration Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Preparing for Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Setting the Zero Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Setting the Response Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Returning to Normal Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
35-3001-01H-DIL Hydrogen Sample-Draw Detector
Overview
This operator’s manual describes the 35-3001-01H-DIL hydrogen sample-draw detector. This manual also describes how to install, start up, maintain, and calibrate the sample-draw detector when using it with a gas monitoring controller. A parts list at the end of this manual lists replacement parts and accessories for the sample-draw detector.
About the 35-3001-01H-DIL
The 35-3001-01H-DIL is designed to monitor hydrogen in an inerted area (an area that has no, or almost no, oxygen present). The catalytic LEL sensor used to detect hydrogen must have at least 5% oxygen present to detect gas properly. A dilution fitting installed in the flow system before the LEL sensor allows enough air to enter the gas sample so that the LEL sensor has enough oxygen to function properly. The dilution ratio is approximately 1:1 but will vary slightly with increased sample line lengths. The longer the sample line, the more dilution air will enter the gas sample through the dilution fitting. Because of this, it is important when calibrating the detector head to apply the calibration gas to the sample line inlet if a sample line is installed. In addition, because the detector head includes internal dilution, it must be calibrated using a sample bag instead of a demand flow regulator. See “Calibration” on page 18 for detailed calibration instructions.
Specifications
Table 1 lists specifications for the hydrogen sample-draw detector. See the controller Operator’s Manual for information specific to the controller.
Table 1: Specifications
Target Gases and Detection Range Hydrogen: 0 - 100% LEL
Input Power 24 VDC Nominal (18.5 VDC - 30 VDC)
Construction (housing) Fiberglass/polyester (NEMA 4X)
Dimensions 8.5 in. H x 6.5 in. W x 4.25 in. D
Weight 4.5 lbs.
Sampling Method Sample-draw
Sample Flow 3.0 SCFH typical, with no inlet or exhaust line
Low Flow Setpoint 1.0 ± 0.1 SCFH, measured in the sample inlet line
Sensor Flow 0.8 SCFH (nominal)
Maximum Recommended Inlet/ Exhaust Line Length for 1/4” O.D. x 1/8” I.D. Tubing
Inlet
100 feet 0 feet
50 feet 50 feet
Exhaust
Maximum Recommended Inlet/ Exhaust Line Length for 1/4” O.D. x 0.170” I.D. Tubing
Response Time 90% in 30 seconds
Inlet
100 feet 100 feet
35-3001-01H-DIL Hydrogen Sample-Draw Detecto r • 1
Exhaust
Table 1: Specifications
Accuracy ± 5% of reading or ± 2% LEL (whichever is greater)
WAR NING: When using the 35-3001-01H-DIL, you must follow the instructions and
warnings in this manual to assure proper and safe operation of the 35-3001-01H-DIL and to minimize the risk of personal injury. Be sure to maintain and periodically calibrate the 35-3001-01H-DIL as described in this manual.
2 • 35-3001-01H-DIL Hydrogen Sample-Draw Detector
Description
O X Y G E N
B
+
W
Oxygen Sensor Terminal Strip
Pump
LEL Sensor
Dilution Fitting
W BG
LEL Flow Block
_
AMP 2SOXY
+
S
+
+
AMP
Sensor Flow
Control Valve
Fail LED
Pilot LED
Pressure
Switch
Adjustment
Screw
(behind PCB)
Flowmeter
W BG
Interconnect Terminal Strip
Pump Reset
Switch
OXY
GW
+
+
AMP 1
S
+
S
AMP 2
Hydrophobic
Filter
LEL/ IR
R
R
Flowmeter Circuit
Board Connec tor
Pump Connector
LEL/ IR
3/4" Conduit Hub
Particle Filter
Exhaust
Fitting
Inlet
Fitting
This End To Inlet Fitting
Mounting
Foot, 4X
EXHAUST
INLET
Detector/Amp Terminal Strip (Factory Wired)
This section describes the components of the hydrogen sample-draw detector.
Figure 1: Hydrogen Sample-Draw Detector Component Location
35-3001-01H-DIL Hydrogen Sample-Draw Detecto r • 3
External Components
LEL
Sensor
Flowmeter
Hydrophobic Filter
Pump
Exhaust
Pressure Sw itch
Flowmeter PCB
Sensor Flow
Contr ol Val v e
Restrictor
Inlet
Dilution Fitting,
appr ox. 1:1 ratio
Particle Filter
This section describes the sample-draw detector’s external components.
Housing
The sample-draw detector’s fiberglass housing is weather- and corrosion-resistant. It is suitable for installation where general purpose equipment is in use.
The housing door is hinged on the left side and is secured by two latches on the right side. The flowmeter and status LEDs are visible through a window in the housing door.
Four mounting feet are attached to the back of the housing (one at each corner). Use the mounting feet to install the housing to a vertical surface.
Sample Fittings
The sample fittings are located on the left side of the bottom of the housing. The inlet fitting is near the front of the housing and the exhaust fitting is near the back of the housing. The sample fittings accept 1/4 in. rigid tubing. See the Installation section on page 7 to connect tubing to the sample fittings.
Particle Filter
A particle filter with a tubing stub on one end is shipped with the instrument but it is not factory installed. If the particle filter is installed directly to the inlet fitting, the tubing stub must be used. If the particle filter is installed somewhere else, like at the end of the inlet line, the tubing stub can be used or removed.
Conduit Hub
One 3/4” conduit hub is located on the right side of the bottom of the housing. It is used for routing wiring into the housing by using conduit or an appropriate cable bushing.
Internal Components
This section describes the sample-draw detector’s internal components (see Figure 1). Figure 2 illustrates how the gas sample moves through the flow system.
Figure 2: Hydrogen Sample-Draw Detector Flow Diagram
4 • 35-3001-01H-DIL Hydrogen Sample-Draw Detector
Loading...
+ 16 hidden pages