RKI Instruments 35-3000RK-OXY User Manual

Page 1
www.rkiinstruments.com
Oxygen Sample-Draw Detector
Operator’s Manual
Part Number: 71-0071RK Revision: P1 Released: 7/18/02
RKI Instruments Inc.
1855 Whipple Road Hayward, CA 94544
PH: 800-754-5165 Fax: 510-441-5650
Page 2
Product W arranty
RKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge. This warranty does not apply to those items which by their nature are subject to deterioration or consumption in normal service, and which must be cleaned, repaired, or replaced on a routine basis. Examples of such items are:
a) Absorbent cartridges d) Batteries b) Pump diaphragms and valves e) Filter elements c) Fuses
W arranty is voided by abuse including mechanical damage, alteration, r ough handling, or repair procedures not in accordance with the operator’s manual. This warranty indicates the full extent of our liability, and we are not responsible for removal or replacement costs, local repair costs, transportation costs, or contingent expenses incurred without our prior approval.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND ALL OTHER OBLIGATIONS OR LIABILITIES ON THE PART OF RKI INSTRUMENTS, INC., INCLUDING BUT NOT LIMITED TO, THE WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL RKI INSTRUMENTS, INC., BE LIABLE FOR INDIRECT, INCIDENTAL, OR CONSEQUENTIAL LOSS OR DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
This warranty covers instruments and parts sold to users by authorized distributors, dealers, and representatives as appointed by RKI Instruments, Inc.
We do not assume indemnification for any accident or damage caused by the operation of this gas monitor, and our warranty is limited to the replacement of parts or our complete goods.
2 • 35-3000RK-OXY Sample-Draw Oxygen Detector
Page 3
Table of Contents
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Housing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Flow System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Detection System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Mounting the Oxygen Sample-Draw Detector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Connecting the Sample Lines to the Sample-Draw Detector. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Wiring the Oxygen Sample-Draw Detector to a Controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Start Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Introducing Incoming Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Setting the Fresh Air Reading. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Replacing Components of the Oxygen Sample-Draw Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Adjusting the Low Flow Setting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Assembling the Calibration Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Setting the Fresh Air Reading. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Setting the Zero Reading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
35-3000RK-OXY Sample-Draw Oxygen Detector • 3
Page 4
Overview
This instruction manual describes the 35-3000RK-OXY sample-draw oxygen detector . This manual also describes how to install, start up, maintain, and calibrate the sample-draw detector when using it with a gas monitoring controller. A parts list at the end of this manual lists replacement parts and accessories for the sample-draw oxygen detector. See the operator’s manual for the controller for information specific to the controller
Specifications
Table 1 lists specifications for the sample-draw oxygen detector.
Table 1: Specifications
Target Gas Oxygen (O Input Power 24 VDC Nominal (23 VDC - 28 VDC)
Construction (housing) Fiberglass/polyester (NEMA 4X) Dimensions 8.5 in. H x 6.5 in. W x 4.25 in. D Weight 4.5 lbs. Sampling Method Sample-draw Sample Flow 1.5 SCFH (nominal) Detection Range 0 to 25.0% (by volume) Response Time 90% in 30 seconds
)
2
4 • 35-3000RK-OXY Sample-Draw Oxygen Detector
Page 5
Description
Pilot light
This section describes the components of the sample-draw oxygen detector. The sample­draw detector consists of the housing, flow system, and detection system.
Mounting foot
(total of 4)
Flow adjust
potentiometer
Pump terminal strip
Fail light
Bypass valve
Flowmeter
Low flow
potentiometer
Relay
Detector terminal strip
EXHAUST fitting
INLET fitting
Interconnect terminal strip
Power terminal strip
(not used)
Conduit hub
(total of 2)
Filter
Combustible gas sensor
(not used in this version)
Oxygen sensor
Figure 1: Sample-draw Oxygen Detector Component Location
Housing
The sample-draw detector’s fiberglass housing is weather- and corrosion-resistant. It is suitable for installation where general purpose equipment is in use.
The housing door is hinged on the left side and is secured by two latches on the right side. The flowmeter and status lights are visible through a window in the housing door.
Four mounting feet are attached to the back of the housing (one at each corner). Use the mounting feet to install the housing to a vertical surface. Use the two conduit hubs on the bottom of the housing to make wiring connections.
An aluminum subpanel is mounted to the interior of the housing. The sample-draw detector’s internal components are mounted to the subpanel.
35-3000RK-OXY Sample-Draw Oxygen Detector • 5
Page 6
Flow System
The sample-draw detectors ow system consists of the inlet tting, lter, pump, owmeter, bypass valve, status lights, pressure switch, and exhaust fitting (see Figure 1). Figure 2 illustrates how the gas sample moves through the ow system.
Bypass
Valve
Flowmeter
Inlet Filter Pump Sensor
Pressure Switch
To
Exhaust
Figure 2: Sample-draw Oxygen Detector Flow Diagram
Inlet fitting
The inlet tting on the right side of the housing allows the gas sample to enter the sample­draw detector. The inlet tting accepts 1/4 in. rigid tubing. See the Installation section on page 9 to connect tubing to the inlet tting.
Filter
The dust lter is below the main circuit board. The lter prevents particulates in the incoming gas sample from damaging the ow and detection systems. Replace the lter when it appears dirty, discolored, or clogged.
Pump
The pump is behind the main circuit board near the top of the sample-draw detector. The pump pulls the gas sample into the sample-draw detector. The pump operates on 24 VAC, which is generated from the 24 VDC supplied to the sample draw detector.
Flowmeter
The owmeter is attached to the main circuit board near the top left corner (see Figure 1.) You can see it through the window in the door. A ball in the owmeter column indicates the ow rate of the sample-draw detector. The owmeter measures the ow in the range
0.2 to 2.0 SCFH (Standard Cubic Feet per Hour). The optimum ow rate is 1.5 SCFH.
Bypass valve
The bypass valve is to the left of the owmeter. The bypass valve adjusts the ow rate to the sensor. Use a at-blade screwdriver to adjust the bypass valve.
NOTE: The bypass valve allows ne adjustments of the ow rate. For a wider range of
adjustment, use the ow adjust potentiometer (see Figure 1.)
6 • 35-3000RK-OXY Sample-Draw Oxygen Detector
Page 7
Status lights
Two status lights are above the owmeter. They are also visible through the window in the housing door.
Pilot light
The green Pilot light is on when the sample-draw detector is receiving power from the Controller.
Fail light
The red fail light is on when the sample ow rate is below the low ow level
NOTE: The factory set low ow level is 0.6 SCFH (±0.2). See Adjusting the Low Flow
Setting on page 16 to adjust this setting.
Pressure switch
The pressure switch is mounted to the opposite side of the main circuit board. The pressure switch monitors the ow rate of the incoming gas sample.
If the ow rate falls below the preset low ow level, the pressure switch causes the fail relay to interrupt the signal from the sensor. The interrupted sensor signal causes a fail condition at the controller. The low ow level is factory-set at 0.6 SCFH (±0.2 SCFH).
Exhaust fitting
The exhaust tting on the right side of the housing allows the gas sample to exit the sample-draw detector . The exhaust tting accepts 1/4 in. rigid tubing. See the Installation section on page 9 to connect tubing to the exhaust tting.
Detection System
The detection system consists of the oxygen sensor and the main circuit board.
Oxygen sensor
The oxygen sensor is installed in a cavity block. The cavity block is mounted to the aluminum subpanel near the bottom of the sample-draw detector. The oxygen sensor includes the oxygen cell, connector, and sensor leads.
NOTE: The cavity block includes a cavity for a combustible gas sensor. This version of
the sample-draw detector does not include the combustible gas sensor.
Oxygen cell
The oxygen cell is protected within the sensor assembly. Through a series of chemical and electronic reactions, the cell produces a millivolt output that is proportional to the detection range of the sample-draw detector.
Connector
The cable that extends from the sensor terminates in a 7-position socket that connects to a mating 7-pin male connector. The socket and connector allow you to replace the sensor without disconnecting the wiring. The sensor leads are soldered to the male connector.
Sensor leads
Two color-coded leads extend from the connector. The leads allow you to connect the oxygen sensor to the main circuit board.
35-3000RK-OXY Sample-Draw Oxygen Detector • 7
Page 8
Main Circuit Board
The main circuit board includes the interconnect terminal strip, detector terminal strip, pump terminal strip, and relay (see Figure 1).
NOTE: The owmeter and status lights are mounted to the main circuit board but are
considered part of the ow system.
Interconnect terminal strip
The interconnect terminal strip is the nine-point terminal strip near the bottom edge of the main circuit board. Use the interconnect terminal strip to connect the sample-draw detector to a controller.
Detector terminal strip
The detector terminal strip is the nine-point terminal strip near the right edge of the circuit board. Use the detector terminal strip to connect the oxygen sensor to the main circuit board.
NOTE: The oxygen sensor is factory-wired to the circuit board. See the Installation
section on page 9 for all wiring procedures related to the sample-draw detector.
Pump terminal strip
The pump terminal strip is the four-point terminal strip near the top edge of the circuit board. Use the pump terminal strip to connect the pump and pressure switch to the main circuit board.
NOTE: The pump and pressure switch are factory-wired to the circuit board.
Relay
The relay is to the left of the detector terminal strip. The relay is single-pole, double-throw (SPDT) and is rated for 2 amps at 25 VDC (resistive). If the pressure switch senses a low ow condition, the relay interrupts the signal from the sensor. The interrupted sensor signal causes a fail condition at the controller.
NOTE: No relay contacts are available for use in controlling an external device.
8 • 35-3000RK-OXY Sample-Draw Oxygen Detector
Page 9
Installation
This section describes procedures to mount the sample-draw oxygen detector in the monitoring environment and wire the sample-draw detector to a controller.
Mounting the Sample-Draw Detector
1. Select the mounting site. Consider the following when you select the mounting site.
Is there enough room to open the housing door and make wiring connections at the bottom of the housing and tubing connections at the right of the housing?
Make sure there is sufcient room to perform start-up, maintenance, and calibration procedures.
Are the owmeter and status lights visible?
7.3 in.
6.5 in.
4.0 in.
8.5 in.
8.9 in.
9.3 in.
Note: The housing is 4.25 in. deep.
Figure 3: Mounting the Sample-Draw Oxygen Detector
3/4 in. conduit hub
(total of 2 hubs)
35-3000RK-OXY Sample-Draw Oxygen Detector • 9
Page 10
2. Close and latch the housing door.
NOTE: The sample-draw detector is shipped with the mounting feet tucked under the
housing to protect the mounting feet during shipment.
3. Slightly loosen the screw that secures the mounting foot to the housing, then rotate the mounting foot 180 degrees (see Figure 3).
4. Tighten the screw that secures the mounting foot to the housing.
5. Repeat steps 3 and 4 for the remaining three mounting feet.
6. Position the sample-draw housing on a vertical surface at eye level (4 1/2 to 5 feet from the oor).
7. Insert 1/4 in. or 5/16 screws through the slots in the mounting feet to secure the housing to the mounting surface.
Connecting the Sample Lines to the Sample-Draw Detector
1. Attach 1/4 in. metal or plastic rigid sample tubing to the inlet tting.
CAUTION: If you use flexible sample tubing (vinyl or polyurethane for example), use an
appropriate metal insert to seal the connection between the tubing and the inlet fitting. See the Parts List at the end of this manual for an example of an appropriate metal insert.
2. Place the opposite end of the tubing at the sampling area.
CAUTION: Avoid loops or slumps in the incoming sample line. To reduce response time, keep the
incoming sample line as short as possible.
3. Attach rigid sample tubing to the exhaust tting.
4. Route the opposite end of the tubing to an open area where the sample can safely disperse.
Wiring the Sample-Draw Detector to a Controller
WARNING: Always verify that the controller is off and that power to the controller is
off before you make wiring connections.
1. Turn off the controller
2. Turn off power to the controller.
3. Unlatch and open the housing door of the sample-draw detector.
4. Guide a four-conductor, shielded cable or four wires in conduit through one of the conduit hubs at the bottom of the sample-draw housing.
5. Connect the cable to the sample-draw detectors interconnect terminal strip as shown in Figure 4.
6. Close and latch the housing door of the sample-draw detector.
10 • 35-3000RK-OXY Sample-Draw Oxygen Detector
Page 11
CAUTION: If using shielded cable, leave the cable’s drain wir e insulated and disconnected at the
sample-draw detector. You will connect the opposite end of the drain wire at the controller.
7. Route the cable or wires in conduit leading from the sample-draw detector through one of the conduit hubs on the controller.
CAUTION: At the controller, do not route controller power and sample-draw detector wiring
through the same conduit hub. The controller power cable may disrupt the transmission of the sample-draw detector signal to the controller.
8. Connect the wires to the applicable detector terminal strip and power terminals at the controller as shown in Figure 4.
OxygenSample Draw Housing
Pump, Internally Wired
Pressure Switch, Internally Wired
PSW
PUMP
Controller Oxygen
Detector Terminals
ControllerTerminals
Or PowerSupply
PCB in Single Point
Sample- Drawing
Detector
GNDH
N
115VAC
NotUsedOn ThisVersion
RD
White (+) Green (-)
-(DC Ground)
+( 23 VDC - 28 VDC)
WHT GRN BLK GND
LEL/ O2 AMP
24V
4/20
+
P- AMP
ToOxygen
Sensor
Green White
_
Figure 4: Wiring the Sample-Draw Oxygen Detector to a Controller
5. Connect the cable shield to an available chassis ground. The grounding screw on one of the controllers grounded conduit hubs is an example of a chassis ground.
35-3000RK-OXY Sample-Draw Oxygen Detector • 11
Page 12
Start Up
This section describes procedures to start up the sample-draw oxygen detector and place the sample-draw detector into normal operation.
Introducing Incoming Power
1. Complete the installation procedures described earlier in this manual.
2. Verify that the power wiring is correct and secure. Refer to the controller instruction manual for connections at the controller.
3. Turn on or plug in the incoming power at the power source end, then turn on the controller.
4. Verify that the sample-draw detectors PILOT light is on.
5. Verify that the controller is on and operating properly. Refer to the controller instruction manual.
6. Verify that the owmeter indicates a ow rate of approximately 1.5 SCFH. If necessary, use the bypass valve or ow adjust potentiometer to adjust the ow rate.
NOTE: The following step tests for leaks in the sample line. This test will cause a low
ow condition at the sample-draw detector and a fail condition at the controller. Be sure to put the controller into its calibration program or disable external alarms before performing this test.
7. Verify that the incoming sample line is not leaking. To test the sample line, plug the open end of the sample line with your thumb. If the owmeter ball drops to the bottom of the owmeter, the incoming sample line is not leaking.
8. Remove your thumb from the sample line, verify that the owmeter returns to a normal ow rate.
9. If alarms were disabled at the controller, enable them.
Setting the Fresh Air Reading
CAUTION: If you suspect the monitoring environment is not of normal oxygen content (20.9%),
use the calibration kit and zero air calibration cylinder to introduce “fresh air” to the detector and verify an accurate normal setting.
1. Verify that the sample-draw detector is sampling a fresh air environment (environment known to be of normal oxygen content).
2. Verify a reading of 20.9% at the controller. If the display reading is 20.9%, start up is complete. The sample-draw oxygen detector
is in normal operation. If the display reading is not 20.9%, continue with step 3.
3. Perform a zeroing operation at the controller. See the controller instruction manual for directions.
12 • 35-3000RK-OXY Sample-Draw Oxygen Detector
Page 13
Maintenance
This section describes maintenance procedures. It includes preventive maintenance procedures. This section also includes procedures to troubleshoot the sample-draw detector, replace components of the sample-draw detector, and adjust the low ow setting.
Preventive Maintenance
This section describes a preventive maintenance schedule to ensure the optimum performance of the sample-draw detector. It includes daily, monthly, and quarterly procedures.
Daily
1. Verify that the pilot light is on.
2. Verify that the owmeter indicates a ow rate of approximately 1.5 SCFH.
3. Verify a reading of 20.9% at the controller. Investigate signicant changes in the
Monthly
If necessary use the bypass valve or ow adjust potentiometer to adjust the ow rate to 1.5 SCFH.
reading.
This procedure describes a test to verify that the sample-draw detector responds pr operly to oxygen deciency.
NOTE: To reduce the response time of this test, use a short incoming sample line. If the
sample-draw detectors sample line is long, connect a shorter line for this test. Make sure you reconnect the sample line after you complete this procedure.
NOTE: Performing a response test on the sample-draw detector may cause alarms. Be
sure to put the controller into its calibration program or disable external alarms before performing this test
Performing the response test
1. Exhale into the sample-draw detectors incoming sample line.
2. Stop exhaling into the sample line, then verify that the readings at the controller decreased.
NOTE: If the reading does not decrease, calibrate the sample draw detector as described
in the Calibration section on page 17.
Quarterly
Calibrate the sample-draw detector as described in the Calibration section on page 17.
35-3000RK-OXY Sample-Draw Oxygen Detector • 13
Page 14
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended action for problems you may encounter with the sample-draw detector.
NOTE: This troubleshooting guide describes sample-draw detector problems only. See
the controller instruction manual for problems you may encounter with the controller.
Fail condition
Symptoms
The sample-draw detectors FAIL light is on.
The controller is operating properly but indicates a reading well below zero.
Probable causes
The detector wiring is disconnected or misconnected.
The sample-draw detectors flow rate is too low because of an obstructed sample line,
failed pump, etc.
The sample-draw detector is malfunctioning.
Recommended action
1. At the sample-draw detector, set the correct ow rate with the bypass valve or ow adjust potentiometer.
2. If you cannot set the correct ow rate, check the sample lines for obstructions or kinks.
3. Verify that the detector wiring is correct and secure. The Installation section on page 9 describes detector wiring connections.
4. Calibrate the sample-draw detector as described in the Calibration section on page 17.
5. If the fail condition continues, replace the oxygen sensor as described later in this section.
6. If the fail condition continues, contact RKI Instruments, Inc., for further instruction.
Slow or no response/difficult or unable to calibrate
Symptoms
The detector responds slowly or does not respond during the monthly response test.
Unable to accurately set the fresh air or zero reading during the calibration pr ocedur e.
The detector requires frequent calibration.
NOTE: Under normal circumstances, the detector r equires calibration once every thr ee
months. Some applications may require a more frequent calibration schedule.
Probable causes
The calibration cylinder is low, out-dated, or defective.
The sample-draw detectors flow rate is too low because of an obstructed sample line,
failed pump, etc.
The sample-draw detector is malfunctioning.
14 • 35-3000RK-OXY Sample-Draw Oxygen Detector
Page 15
Recommended action
1. Verify that the calibration cylinder contains an adequate supply of a fresh test sample.
2. If necessary, set the correct ow rate with the bypass valve or ow adjust potentiometer.
3. If you cannot set the correct ow rate, check the sample line for obstructions or kinks.
4. If the calibration/response difculties continue, replace the oxygen sensor as described later in this section.
5. If the calibration/response difculties continue, contact RKI Instruments, Inc., for further instruction.
Replacing Components of the Oxygen Sample-draw Detector
This section includes procedures to replace the sensor, lter, and ferrules.
Replacing the oxygen sensor
1. Turn off the controller.
2. Turn off power to the controller.
3. Open the housing door of the sample-draw detector.
4. Unscrew and remove the two screws that secure the sensor retraining plate, then lift the plate, connector, and sensor out of the housing.
5. Unplug the connector from the senor cable socket.
6. Plug the socket of the replacement sensor into the connector.
7. Place the sensor in the oxygen sensor cavity, then position the retaining plate on the two standoffs.
8. Secure the retaining plate to the standoffs with the two screws you removed in step 4.
9. Turn on power to the controller.
10. Turn on the controller.
11. Calibrate the replacement sensor as described in the Calibration section on page 17.
Replacing the filter
1. Turn off the controller.
2. Turn off power to the controller.
3. Open the housing door of the sample-draw detector.
4. Note the direction of the arrow on the lter. The arrow indicates the direction of the sample ow.
5. Disconnect the lter from the elbows on each end of the lter, then remove the lter from the sample-draw detector.
6. Make sure the arrow is pointing in the same direction as the arrow on the lter you removed, then connect each end of the replacement lter to the elbows.
7. Verify that the ow rate is approximately 1.5 SCFH, then close the housing door.
8. Turn on power to the controller.
9. Turn on the controller.
35-3000RK-OXY Sample-Draw Oxygen Detector • 15
Page 16
Replacing the ferrules
Both the inlet and exhaust ttings include two ferrules that seal the inlet or exhaust tubing to the tting. Replace the ferrules if the seal is bad or if you replace the sample tubing. Always replace the ferrules as a pair.
1. Disconnect the sample tubing from the tting, then unscrew the nut from the tting.
2. Verify that the ferrules did not remain in the nut. If necessary, remove the ferrules from the nut.
3. Position the nut so the threaded end is facing you, then insert the bottom (smaller) ferrule into the nut. Insert the ferrule so the at side is facing down.
NOTE: Make sure the bottom ferrule is laying at in the nut.
4. Insert the cone-shaped front ferrule on top of the bottom ferrule. Insert the ferrule so the smaller end of the cone is facing up.
5. Screw the nut onto the tting, then connect the sample tubing to the tting. Make sure you rmly tighten the tubing to the tting.
Adjusting the Low Flow Setting
NOTE: Adjusting the low ow setting will cause a low ow alarm at the sample-draw
detector and a fail alarm at the controller. Be sure to put the controller into its calibration program or disable external alarms before performing this test
The factory-set low ow setting is 0.6 SCFH (±0.2). To adjust the low ow setting:
1. Use the ow adjust potentiometer (VR1) to set the ow to 0.6 SCFH. If the sample-draw detector goes into low ow alarm before you can adjust the ow
down to 0.6 SCFH, adjust the low ow potentiometer 1/4 turn clockwise, then attempt to set the ow again. Repeat this step until you are able to adjust the ow to
0.6 SCFH.
2. Slowly turn the low ow potentiometer counterclockwise just until the sample-draw detector goes into low ow alarm.
NOTE: The low ow potentiometer is accessible through a circular cutout in the main
circuit board. The cutout is labeled PS1.
3. Increase the ow using VR1 until the unit is out of low ow alarm.
4. Decrease the ow very slowly and verify that the low ow alarm is 0.6 SCFH (±0.2). If the low ow alarm is set too low, turn the low ow potentiometer slightly
clockwise. Repeat steps 3 and 4 if necessary.
5. Use the ow adjust potentiometer (VR1) to set the ow to 1.5 SCFH.
6. Make sure the sample draw detectors Fail light is off.
16 • 35-3000RK-OXY Sample-Draw Oxygen Detector
Page 17
Calibration
This section describes how to calibrate the sample-draw oxygen detector. It includes procedures to assemble the calibration kit, set the fresh air reading, set the zero reading, and return to normal operation.
NOTE: Calibrating the sample draw detector may cause alarms. Be sure to put the
controller into its calibration program or disable external alarms before continuing.
NOTE: This procedure describes calibration using a gas collection bag. A demand-ow
calibration kit is also available for calibrating the sample-draw detector.
Assembling the Calibration Kit
NOTE: If you can verify a fresh air environment, it is not necessary to use a zero air
calibrating sample to set the fresh air reading at the controller.
1. Screw the dispensing valve onto the zero air calibration cylinder.
2. Connect the calibration kit sample tubing to the tting on the gas collection bag.
3. Disconnect the incoming sample line from the sample-draw detectors inlet tting, then connect the sample tubing from the gas collection bag to the inlet tting.
Allow the sample-draw pump to draw out any residual gas in the gas collection bag.
4. Close the tubing clamp. The tubing clamp is attached to the calibration kit sample tubing.
5. Disconnect the calibration kit sample tubing from the inlet tting.
6. Connect the tubing from the gas collection bag to the dispensing valve, then open the clamp.
7. Open the dispensing valve. The gas collection bag begins to ll.
8. Close the dispensing valve when the gas collection bag appears full.
9. Close the tubing clamp, then disconnect the sample tubing from the dispensing valve.
Setting the Fresh Air Reading
1. Open the clamp, then connect the sample tubing from the gas collection bag to the sample-draw detectors inlet tting.
2. Allow the sample draw detector to draw sample for one minute.
3. Follow the directions in the controllers instruction manual for setting the fresh air reading.
4. Allow the sample-draw detector to draw out any residual gas in the gas collection bag.
5. Disconnect the gas bag sample tubing from the inlet tting, then close the tubing clamp.
6. Unscrew the dispensing valve from the zero air calibration cylinder.
35-3000RK-OXY Sample-Draw Oxygen Detector • 17
Page 18
Setting the Zero Reading
1. Screw the dispensing valve onto the 100% N2 calibration cylinder.
2. Connect the sample tubing from the gas collection bag to the dispensing valve, then open the tubing clamp.
3. Open the dispensing valve. The gas collection bag begins to ll.
4. Close the dispensing valve when the gas collection bag appears full.
5. Close the tubing clamp, then disconnect the sample tubing from the dispensing valve. Open the tubing clamp, then connect the sample tubing from the gas collection bag to the sample-draw detectors inlet tting.
6. Allow the sample-draw detector to draw the calibrating sample for 1 minute.
7. Follow the directions in the controllers instruction manual for setting the zero (oxygen free) reading.
8. Allow the sample-draw detector to draw out any residual gas in the gas collection bag.
9. Disconnect the sample tubing from the sample-draw detectors inlet tting.
10. Reconnect the incoming sample line to the inlet tting.
11. Unscrew the dispensing valve from the calibration cylinder.
12. Wait 1 to 2 minutes to allow the oxygen reading at the controller to return to normal, then return the controller to normal operation.
NOTE: If you do not allow the oxygen reading to return to normal, then unwanted
alarms may occur.
13. Verify that the controller display reading stabilizes at 20.9%.
14. Store the components of the calibration kit in a safe and convenient place.
18 • 35-3000RK-OXY Sample-Draw Oxygen Detector
Page 19
Parts List
Table 3 lists replacement parts and accessories for the sample-draw oxygen detector.
Table 2: Parts List
Part Number Description
06-1248RK Sample tubing, 3/16 x 5/16, specify length (for calibration kit) 17-2593RK Brass insert (for inlet and exhaust ttings) 17-2683RK Front ferrule (for inlet and exhaust ttings) 17-2688RK Back ferrule (for inlet and exhaust ttings) 30-0610RK Pump 33-0163RK Filter (Balston DFU9933-05-DQ) 65-0601RK Oxygen sensor (plug-in type) 71-0071RK Operator’s Manual, 35-3000RK-OXY Oxygen Sample-Draw Detector 81-0076RK-01 Zero air calibration cylinder (34 liter) 81-0076RK-03 Zero air calibration cylinder (103 liter) 81-0078RK Calibration cylinder (100% Nitrogen; 17 liter) 81-0078RK-01 Calibration cylinder (100% Nitrogen; 34 liter) 81-0078RK-03 Calibration cylinder (100% Nitrogen; 103 liter) 81-1001RK Dispensing valve with knob, for 17 liter & 34 liter cylinders 81-1054RK Demand ow regulator, for 103 liter cylinder 81-1126RK Gas collection bag (2 liter) 81-F301RKS-LV Calibration kit (includes gas collection bag, 100% Nitrogen and
zero air calibration cylinders)
81-F302RKS-LV Calibration kit (includes gas collection bag and 100% Nitrogen
calibration cylinder)
35-3000RK-OXY Sample-Draw Oxygen Detector • 19
Loading...