RKI Instruments 35-3000RKA-LEL User Manual

±
±
35-3000RKA-LEL Sample Draw Combustible Gas Detector
Table 1 lists specications for the sample draw combustible gas detector.
Table 1: Specications
Target Gas Combustible Gas Input Power 24 VDC Construction (housing) Fiberglass/polyester (NEMA 4X) Dimensions 8.5 in. H x 6.5 in. W x 4.25 in. D Weight 4.5 lbs. Sampling Method sample draw Sample Flow 1.5 SCFH (nominal) Detection Range 0 to 100% LEL Response Time 90% in 30 seconds Linearity Repeatability
5% of detection range 2% of detection range
Sample Draw Combustible Gas Detector • 1
O
Description
This section describes the components of the sample draw combustible gas detector. The sample draw detector consists of the housing, ow system, and detection system.
Mounting
Flow Adjust Potentiometer
Pu mp /flo w sw itch terminal block
Fail LED
Pilot LED
Flowmeter
Bypass valve Particle filter
LE L Transmitter
Foot, 4X
Flow Alarm Setpoint Adjustment Potentiometer
Se nso r Wiring Terminal St rip
External Wiring Terminal Strip
Exhaust
Inlet
Factory Set Pots
Test Points 100 - 500 mV range
JUMPER PINS FOR FACTORY U SE ONLY
Span Pot
Zero Pot
Housing
The sample draw detector’s berglass housing is weather- and corrosion-resistant. It is suitable for installation where general purpose equipment is in use.
LEL sensor
P/N 57-1050RK
REV. 0
GND
24V
TP +
4-20
POWER/SIG
RED
WHT
SENSOR
GRN
BLK
ZER
SPAN
TP
Terminal Block
3/4" Conduit Hub, 2X
Figure 1: Sample draw Combustible Gas Detector Component Location
The housing door is hinged on the left side and is secured by two latches on the right side. The owmeter and status lights are visible through a window in the housing door.
Four mounting feet are attached to the back of the housing (one at each corner). Use the mounting feet to install the housing to a vertical surface. Use the two conduit hubs on the bottom of the housing to make wiring connections.
An aluminum subpanel is mounted to the interior of the housing. The sample draw detector’s internal components are mounted to the subpanel.
2 • Sample Draw Combustible Gas Detector
Flow System
The sample draw detector’s ow system consists of the INLET tting, lter, pump, owmeter, bypass valve, status lights, pressure switch, and EXHAUST fitting (see Figure 1). Figure 2 illustrates how the gas sample moves through the ow system.
To
Exhaust
Bypass
Valve
Flowmeter
Inlet Filter Pump Sensor
Pressure Switch
Figure 2: Sample draw Combustible Gas Detector Flow Diagram
INLET fitting
The INLET tting on the right side of the housing allows the gas sample to enter the sample draw detector. The INLET tting accepts 1/4 in. rigid tubing. See the Installation section on page 7 to connect tubing to the INLET tting.
Filter
The dust lter is below the main circuit board. The lter prevents particulates in the incoming gas sample from damaging the ow and detection systems. Replace the lter when it appears dirty, discolored, or clogged.
Pump
The pump is behind the main circuit board near the top of the sample draw detector. The pump pulls the gas sample into the sample draw detector. The pump operates on 24 VAC, which is generated from the 24 VDC supplied to the sample draw detector.
Flowmeter
The owmeter is attached to the main circuit board near the top left corner (see Figure 1.) You can see it through the window in the door. A ball in the owmeter column indicates the ow rate of the sample draw detector. The owmeter measures the ow in the range
0.2 to 2.0 SCFH (Standard Cubic Feet per Hour). The optimum ow rate is 1.5 SCFH.
Bypass valve
The bypass valve is to the left of the owmeter. The bypass valve adjusts the ow rate to the sensor. Use a at-blade screwdriver to adjust the bypass valve.
NOTE: The bypass valve allows ne adjustments of the ow rate. For a wider range of
adjustment, use the ow adjust potentiometer (see Figure 1.)
Sample Draw Combustible Gas Detector • 3
Status lights
Two status lights are above the owmeter. They are also visible through the window in the housing door.
Pilot light
The green Pilot light is on when the sample draw detector is receiving power.
Fail light
The red Fail light is on when the sample ow rate is below the low ow level.
NOTE: The default low ow level is 0.6 SCFH ( ± 0.2). See “Adjusting the Low Flow
Setting” on page 15 to adjust this setting.
Pressure switch
The pressure switch is mounted to the opposite side of the main circuit board. The pressure switch monitors the ow rate of the incoming gas sample.
If the ow rate falls below the preset low ow level, the pressure switch causes the fail relay to interrupt the signal in the 4-20 mA line. This causes a downscale reading at the monitor. The low ow level is factory-set at 0.6 SCFH ( ± 0.2 SCFH).
EXHAUST fitting
The EXHAUST tting on the right side of the housing allows the gas sample to exit the sample draw detector. The EXHAUST tting accepts 1/4 in. rigid tubing. See the Installation section on page 7 to connect tubing to the EXHAUST tting.
Detection System
The detection system consists of the combustible gas sensor, the hydrogen transmitter, and the main circuit board.
Combustible gas sensor
The combustible gas sensor is installed in a cavity block. The cavity block is mounted to the aluminum subpanel near the bottom of the sample draw detector. The combustible gas sensor includes the sensing elements, ame arrestor, connector, and sensor leads.
NOTE: The cavity block includes a cavity for an oxygen sensor. This version of the
sample draw detector does not include the oxygen sensor.
Sensing elements
Two sensing elements are protected within the sensor assembly. Through a series of thermal and electronic reactions, these elements produce an output that is pr oportional to the detection range of the sample draw detector . The transmitter converts the output to a 4
- 20 mA signal which can be used by a recording or monitoring device. The porous ame arrestor allows the gas sample to enter the sensor assembly and contact
the sensing element. The ame arrestor also contains sparks within the sensor.
4 • Sample Draw Combustible Gas Detector
Connector
The top of the sensor includes ve pins that plug into the socket connector . This connector allows you to replace the sensor without disconnecting the wiring. The sensor leads are soldered to the connector.
Sensor leads
Four color-coded leads extend from the connector. The leads allow you to connect the combustible gas sensor to the main circuit board.
Transmitter
The transmitter is mounted to the left of the combustible sensor. It consists of the span pot, zero pot, transmitter interconnect terminal strip, sensor terminal strip and test points.
Span/zero pots
The span and zero pots are located at the bottom edge of the transmitter and are used for calibration. Use the span pot to make adjustments to gas response readings and the zero pot to make adjustments to the zero reading
Transmitter interconnect terminal strip
The transmitter interconnect terminal strip is the six-point terminal strip near the right edge of the transmitter. The transmitter is factory wired to the sensor and main circuit board.
Test points
The test points are located on the left side of the transmitter and are labeled TP+ and TP- . A 100 mV - 500 mV output is available at these testpoints for use during calibration.
Main Circuit Board
The main circuit board includes the interconnect terminal strip, transmitter terminal strip, pump terminal strip, and relay (see Figure 1).
NOTE: The owmeter and status lights are mounted to the main circuit board but are
considered part of the ow system.
Interconnect terminal strip
The interconnect terminal strip is the nine-point terminal strip near the bottom edge of the main circuit board. Use the interconnect terminal strip to connect the sample draw detector to power and external devices.
Transmitter terminal strip
The transmitter terminal strip is the nine-point terminal strip near the right edge of the circuit board. Use the transmitter terminal strip to connect the transmitter to the main circuit board.
NOTE: The transmitter is factory-wired to the circuit board. See the “Installation” section
on page 7 for all wiring procedures related to the sample draw detector.
Sample Draw Combustible Gas Detector • 5
Pump terminal strip
The pump terminal strip is the four-point terminal strip near the top edge of the circuit board. Use the pump terminal strip to connect the pump and pressure switch to the main circuit board.
NOTE: The pump and pressure switch are factory-wired to the circuit board. See the
“Installation” on page 7 for all wiring procedures related to the sample draw detector.
Relay
The relay is to the left of the detector terminal strip. The relay is single-pole, double-throw (SPDT) and is rated for 2 amps at 25 VDC (resistive). If the pressure switch senses a low ow condition, the relay interrupts the signal from the transmitter. The interrupted transmitter signal causes a fail condition at the Pioneer.
6 • Sample Draw Combustible Gas Detector
Installation
This section describes procedures to mount the sample draw combustible gas detector in the monitoring environment and wire the sample draw detector to power and an external device.
Mounting the Sample draw Combustible Gas Detector
1. Select the mounting site. Consider the following when you select the mounting site.
Is there enough room to open the housing door and make wiring connections at the bottom of the housing and tubing connections at the right of the housing?
Make sure there is sufcient room to perform start-up, maintenance, and calibration procedures.
•Are the owmeter and status lights visible?
7.3 in.
9.3 in.
6.5 in.
4.0 in.
8.5 in.
8.9 in.
3/4 in. conduit hub
(total of 2 hubs)
Note: The housing is 4.25 in. deep.
Figure 3: Mounting the Sample Draw Combustible Gas Detector
Sample Draw Combustible Gas Detector • 7
2. Close and latch the housing door.
NOTE: The sample draw detector is shipped with the mounting feet “tucked under” the
housing to protect the mounting feet during shipment.
3. Slightly loosen the screw that secures the mounting foot to the housing, then rotate the mounting foot 180 degrees (see Figure 3).
4. Tighten the screw that secures the mounting foot to the housing.
5. Repeat steps 3 and 4 for the remaining three mounting feet.
6. Position the sample draw housing on a vertical surface at eye level (4 1/2 to 5 feet from the oor).
7. Insert 1/4 in. screws through the slots in the mounting feet to secure the housing to the mounting surface.
Connecting the Sample Lines to the Sample Draw Combustible Gas Detector
1. Attach 1/4 in. metal or plastic rigid sample tubing to the INLET tting.
CAUTION: If you use flexible sample tubing (vinyl or polyurethane for example), use an
appropriate metal insert to seal the connection between the tubing and the INLET tting. See Appendix A, Parts List, for an example of an appropriate metal insert.
2. Place the opposite end of the tubing at the sampling area.
CAUTION: Avoid loops or slumps in the incoming sample line. To reduce response time, keep the
incoming sample line as short as possible.
3. Attach rigid sample tubing to the EXHAUST tting.
4. Route the opposite end of the tubing to an open area where the sample can safely disperse.
Wiring the Sample Draw Combustible Gas Detector
WARNING: Always verify that the power source is OFF before you make wiring
connections.
1. Unlatch and open the housing door of the sample draw detector.
2. Guide a three-conductor, shielded cable or three wires in conduit through one of the conduit hubs at the bottom of the sample draw housing.
3. Connect the cable to the sample draw detector’s interconnect terminal strip as shown in Figure 4.
4. Close and latch the housing door of the sample draw detector.
CAUTION: Leave the shield cable insulated and disconnected at the sample draw detector.
You will connect the opposite end of the shield cable at the device.
8 • Sample Draw Combustible Gas Detector
5. Route the cable or wires in conduit leading from the sample draw detector to the recording or monitoring device and power.
6. Connect the cable shield to an available chassis ground at the device end.
LEL Sample Draw Housing
Transmitter, Internally Wi red
P/N 57-1050RK
REV. 0
GND
24V
TP +
4-20
POWER/SIG
RED
WHT GRN
SENSOR
BLK
ZEROSPAN
TP
WHITE
BLAC K GREEN
RED
LEL DETECTOR IN SAMPLE CHAMBER (INTERNALLY WIRED)
PUMP ASSY INTERNALLY
WIRED
PUM P PS W
PRES SURE SWITCH INTERNALLY WIRED.
PCB IN SINGLE POINT SAMPLE DRAWING DETECTOR ASS Y
NOT USED ON THIS
VE RS ION
HN
115VAC
CG ND
WHT GR N
RD
LEL/O2 AMP P- AM P
GN D 2 4 V +
BLK
_
4/20
4/20 (Feedback)
24VDC
+
Gr ound
Figure 4: Wiring the Sample Draw Combustible Gas Detector
4/20
24V
GN D
Sample Draw Combustible Gas Detector • 9
Start Up
This section describes procedures to start up the sample draw combustible gas detector and place the sample draw detector into normal operation.
Introducing Incoming Power
1. Complete the installation procedures described earlier in this manual.
2. Verify that the power/device wiring is correct and secure.
3. Turn on or plug in the incoming power at the power source end.
4. Verify that the Pilot light is on.
5. Verify that the owmeter indicates a ow rate of approximately 1.5 SCFH. If necessary, use the bypass valve or ow adjust potentiometer to adjust the ow rate.
NOTE: The following step tests for leaks in the sample line. This test may cause a low
ow condition at the sample draw detector.
6. Verify that the incoming sample line is not leaking. To test the sample line, plug the open end of the sample line with your thumb. If the owmeter ball drops to the bottom of the owmeter, the incoming sample line is not leaking.
7. Remove your thumb from the sample line, verify the owmeter returns to a normal ow rate.
Setting the Zero Reading
CAUTION: If you suspect the presence of combustible gas in the monitoring environment, use
the calibration kit and the zero air calibration cylinder to introduce “fresh air” to the sensor and verify an accurate zero setting.
1. Verify that the sample draw detector is sampling a fresh air environment (environment known to be free of combustible gas).
2. Open the housing door.
3. Set a voltmeter to measure in the milivolt (mV) range.
4. Plug the voltmeter leads into the test points on the amplier. Plug the positive lead into the test point labeled TP+ ; plug the negative lead into the test point labeled TP- .
5. Verify a voltmeter reading of 100 mV ( ± 2 mV).
6. If necessary, use a small at-blade screwdriver to adjust the zero potentiometer until the voltmeter reading is 100 mV ( ± 2 mV).
7. Close the housing door.
10 • Sample Draw Combustible Gas Detector
Maintenance
This section describes maintenance procedures. It includes preventive maintenance procedures. This section also includes procedures to troubleshoot the sample draw detector , r eplace components of the sample draw combustible gas detector, and adjust the low ow setting.
Preventive Maintenance
This section describes a preventive maintenance schedule to ensure the optimum performance of the sample draw detector. It includes daily, monthly, and quarterly procedures.
Daily
1. Verify that the pilot light is on.
2. Verify that the owmeter indicates a ow rate of approximately 1.5 SCFH.
3. Verify a reading of 0 %LEL (4 mA) at the monitoring device or a reading of 100 mV at
Monthly
If necessary use the bypass valve or ow adjust potentiometer to adjust the ow rate to 1.5 SCFH.
the transmitter test points. Investigate signicant changes in the reading.
This procedure describes a test to verify that the sample draw combustible gas detector responds properly to the target gas.
Preparing for the response test
NOTE: This procedure describes the RKI calibration kit that includes a gas collection
bag. A calibration kit that uses a demand ow regulator is also available.
1. Verify that the monitoring device is reading 0 %LEL (4 mA). If the reading is not 0, set the zero reading as described in the “Start Up” section on
page 10, then continue this procedure.
2. Connect the calibration kit sample tubing along with the tubing clamp to the tting on the gas collection bag.
3. Connect the sample tubing from the gas collection bag to the inlet line at or near the INLET tting.
Allow the sample draw pump to draw out any residual gas in the gas collection bag.
NOTE: This will cause a low ow alarm when the bag is empty. Normal operation of the
sample draw adapter will resume when the tubing is disconnected from the inlet line.
4. Disconnect the calibration kit sample tubing from the inlet line.
5. Close the clamp right away. The clamp is attached to the calibration kit sample tubing.
6. Connect the tubing from the gas collection bag to the xed ow regulator, then open the clamp.
7. Screw the fixed flow regulator onto the calibration gas cylinder. The gas collection bag begins to ll.
Sample Draw Combustible Gas Detector • 11
8. When the bag is full, unscrew the xed ow regulator form the cylinder.
9. Close the clamp, then disconnect the sample tubing from the xed ow regulator.
Performing the response test
1. Connect the calibration tubing from the gas collection bag to the inlet line at or near the INLET tting.
The sample draw detector’s pump automatically begins pulling the test sample from the gas collection bag when you connect the tubing to the inlet line.
2. After approximately one minute, verify that the reading at the monitoring device stabilizes within ± 10% of the concentration of the test sample. If the reading is not within ± 10% of the test sample, calibrate the sample draw detector as described in the Calibration section of this manual.
3. Remove the calibration tubing from the inlet line, then reconnect the inlet line.
Quarterly
Calibrate the sample draw detector as described in the “Calibration” section on page 16.
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended action for problems you may encounter with the sample draw combustible gas detector.
NOTE: This troubleshooting guide describes sample draw detector problems only. See
the instruction manual for monitoring device if it exhibits any problems.
Fail condition
Symptoms
The sample draw detector’s Fail light is on.
The monitoring device is indicating a reading well below zero.
The monitoring device is indicating a Fail condition.
Probable causes
The sample draw detector’s ow rate is too low because of an obstructed sample line, failed pump, etc.
The sample draw detector is malfunctioning.
The sensor or transmitter wiring is disconnected or misconnected.
Recommended action
1. At the sample draw detector, set the correct ow rate with the bypass valve or ow adjust potentiometer.
2. If you cannot set the correct ow rate, check the sample lines for obstructions or kinks.
3. Verify that the detector and transmitter wiring are correct and secure. The Installation section on page 7 describes detector wiring connections.
4. Calibrate the sample draw detector as described in the Calibration section on page 16.
5. If the fail condition continues, replace the sensor as described later in this section.
6. If the fail condition continues, contact RKI Instruments, Inc., for further instruction.
12 • Sample Draw Combustible Gas Detector
Slow or no response/difficult or unable to calibrate
Symptoms
The detector responds slowly or does not respond during the monthly response test.
Unable to accurately set the zero or response reading during the calibration procedure.
The detector requires frequent calibration.
NOTE: Under “normal” circumstances, the detector requires calibration once a quarter.
Some applications may require a more frequent calibration schedule.
Probable causes
The calibration cylinder is low, out-dated, or defective.
The sample draw detector’s ow rate is too low because of an obstructed sample line, failed pump, etc.
The sample draw detector is malfunctioning.
The demand ow regulator is not operating properly (if one is used for calibration instead of a gas bag).
Recommended action
1. Verify that the calibration cylinder contains an adequate supply of a fresh test sample.
2. If necessary, set the correct ow rate with the bypass valve or ow adjust potentiometer.
3. If you cannot set the correct ow rate, check the sample line for obstructions or kinks.
4. If a demand ow regulator is used for calibration, try a different demand ow regulator or a gas bag setup and see if the calibration/response difculties continue.
5. If the calibration/response difculties continue, replace the sensor as described later in this section.
6. If the calibration/response difculties continue, contact RKI Instruments, Inc., for further instruction.
Sample Draw Combustible Gas Detector • 13
Replacing Components of the Combustible Gas Sample Draw Detector
This section includes procedures to replace the sensor, lter, and ferrules.
Replacing the combustible gas sensor
1. Turn off incomming power.
2. Open the housing door of the sample draw detector.
3. Unscrew and remove the two screws that secure the retraining plate, then lift the plate, connector, and sensor out of the housing.
4. Unplug the connector from the sensor.
5. Verify that you are using the correct replacement sensor (NC-6240 is printed on the sensor), then plug the sensor into the connector.
6. Place the sensor in the combustible gas sensor cavity, then position the retaining plate on the two standoffs.
7. Secure the retaining plate to the standoffs with the two screws you removed in step 3.
8. Turn on incomming power.
CAUTION: Allow the replacement sensor to warm up for 15 minutes before you continue.
9. Calibrate the replacement sensor as described in the “Calibration” section on page 16.
Replacing the filter
1. Open the housing door of the sample draw detector.
2. Note the direction of the arrow on the lter. The arrow indicates the direction of the sample ow.
3. Disconnect the lter from the elbows on each end of the lter, then remove the lter from the sample draw detector.
4. Make sure the arrow is pointing in the same direction as the arrow on the lter you removed, then connect each end of the replacement lter to the elbows.
5. Verify that the ow rate is approximately 1.5 SCFH, then close the housing door.
Replacing the ferrules
The INLET and EXHAUST ttings each includes two ferrules that seal the incoming or exhaust tubing to the tting. Replace the ferrules if the seal is bad or if you replace the sample tubing. Always replace the ferrules as a pair.
1. Disconnect the sample tubing from the tting, then unscrew the nut from the tting.
2. Verify that the ferrules did not remain in the nut. If necessary, remove the ferrules from the nut.
3. Position the nut so the threaded end is facing you, then insert the bottom (smaller) ferrule into the nut. Insert the ferrule so the at side is facing down.
NOTE: Make sure the bottom ferrule is laying at in the nut.
4. Insert the cone-shaped front ferrule on top of the bottom ferrule. Insert the ferrule so the smaller end of the cone is facing up.
5. Screw the nut onto the tting, then connect the sample tubing to the tting. Make sure you rmly tighten the nut with a wrench.
14 • Sample Draw Combustible Gas Detector
Adjusting the Low Flow Setting
The factory-set low ow setting is 0.6 SCFH (±0.2). To adjust the low ow setting:
1. Use the ow adjust potentiometer (VR1) to set the ow to 0.6 SCFH. If the sample draw detector goes into low ow alarm before you can adjust the ow
down to 0.6 SCFH, adjust the low ow potentiometer 1/4 turn clockwise, then attempt to set the ow again. Repeat this step until you are able to adjust the ow to
0.6 SCFH.
2. Slowly turn the low ow potentiometer counterclockwise just until the sample draw detector goes into low ow alarm.
NOTE: The low ow potentiometer is accessible through a circular cutout in the main
circuit board. The cutout is labeled PS1.
3. Increase the ow using VR1 until the unit is out of low ow alarm.
4. Decrease the ow very slowly and verify that the low ow alarm is 0.6 SCFH (±0.2). If the low ow alarm is set too low, turn the low ow potentiometer slightly
clockwise. Repeat steps 3 and 4 if necessary.
5. Use the ow adjust potentiometer (VR1) to set the ow to 1.5 SCFH.
6. Make sure the sample draw detector’s Fail light is off.
Sample Draw Combustible Gas Detector • 15
Calibration
This section describes how to calibrate the sample draw combustible gas detector. It includes procedures to assemble the calibration kit, set the zero reading, set the response reading, and return to normal operation.
NOTE: This procedure describes calibration using a gas collection bag. A demand ow
calibration kit is also available for calibrating the combustible gas sample draw detector.
Preparing for Calibration
NOTE: Calibrating the sample draw adapter may cause alarms. Be sure to put the
controller into calibration mode or disable external alarms before calibrating.
1. Open the housing door.
2. Set a voltmeter to measure in the millivolt (mV) range.
3. Plug the positive lead into the transmitter test point labeled TP+; plug the negative lead into the transmitter test point labeled TP-.
4. Use the following formula to determine the correct test points output for the calibrating sample.
Output (mV) = (calibrating sample/fullscale) X 400 + 100
For example, with a calibrating sample of 50% LEL and a fullscale setting of 100% LEL, the correct output is 300 mV.
300 (mV) = (50/100) X 400 +100
Assembling the Calibration Kit
1. Connect the calibration kit sample tubing along with the tubing clamp to the tting on the gas collection bag.
2. Connect the sample tubing from the gas collection bag to the inlet line at or near the INLET tting.
Allow the sample draw pump to draw out any residual gas in the gas collection bag.
NOTE: This will cause a low flow alarm when the bag is empty. Normal operation of the
sample draw adapter will resume when the tubing is disconnected from the inlet line.
3. Disconnect the calibration kit sample tubing from the inlet line.
4. Close the clamp right away. The clamp is attached to the calibration kit sample tubing.
NOTE: If you can verify a fresh air environment, it is not necessary to use a zero air
calibration cylinder to set the zero reading. Go to the next section, “Setting the Zero Reading.”
16 • Sample Draw Combustible Gas Detector
5. Connect the tubing from the gas collection bag to the xed ow regulator, then open the clamp.
6. Screw the fixed flow regulator onto the zer o air calibration cylinder. The gas collection bag begins to ll.
7. When the bag is full, unscrew the xed ow regulator from the cylinder.
8. Close the clamp, then disconnect the sample tubing from the xed ow regulator.
Setting the Zero Reading
1. Open the clamp, then connect the sample tubing from the gas collection bag to the sample draw detector’s inlet line. This step is not necessary if you veried a fresh air
environment earlier in this procedure.
2. Allow the reading to stabilize for approximately 1 minute.
3. Verify a voltmeter reading of 100 mV (± 2 mV).
4. If necessary, use a small at-blade screwdriver to adjust the zero potentiometer until the voltmeter reading is 100 mV (± 2 mV).
Steps 5 and 6 are not necessary if you veried a fresh air environment earlier in this procedure.
5. Allow the sample draw pump to draw out any residual gas in the gas collection bag.
6. Disconnect the sample tubing from the inlet line, then close the clamp.
Setting the Response Reading
1. Connect the sample tubing from the gas collection bag to the xed ow regulator, then open the clamp.
2. Screw the fixed flow regulator onto the calibration gas cylinder. The gas collection bag begins to ll.
3. Unscrew the xed ow regulator from the cylinder when the gas collection bag appears full.
4. Close the clamp, then disconnect the sample tubing from the xed ow regulator.
5. Open the clamp, then connect the sample tubing from the gas collection bag to the inlet line at or near the sample draw detector’s INLET tting.
6. Allow the sample draw detector to respond to the calibrating sample for approximately 1 minute.
7. When the reading on the voltmeter stabilizes, verify that the reading matches the response reading (±2 mV) you determined earlier.
8. If necessary, use the span potentiometer on the amplier to adjust the reading to match the correct response reading.
9. Allow the sample draw pump to draw out any residual gas in the gas collection bag.
10. Disconnect the sample tubing from the sample draw detector’s inlet line, then close the clamp.
11. Reconnect the incoming sample line.
12. Wait 1 to 2 minutes to allow the combustible gas reading to decrease and stabilize.
Sample Draw Combustible Gas Detector • 17
Parts List
Table 4 lists replacement parts and accessories for the sample draw combustible gas detector.
Table 2: Parts List
Part Number Description
06-1248RK Sample tubing, 3/16 x 5/16, specify length, (for calibration kit) 17-2593RK Brass insert (for INLET and EXHAUST ttings) 17-2683RK Front ferrule (for INLET and EXHAUST ttings) 17-2688RK Back ferrule (for INLET and EXHAUST ttings) 30-0610RK Pump 33-0163RK Filter (Boston DFU9933-05-DQ) 57-1050RK Transmitter (specify target gas when ordering) 61-0145RK Combustible gas sensor, NC-6240, plug-in type, LEL range 81-0002RK-01 Calibration cylinder (50% LEL Hydrogen; 34 liter) 81-0002RK-03 Calibration cylinder (50% LEL Hydrogen; 103 liter) 81-0007RK-01 Calibration cylinder (50% LEL Hexane; 34 liter) 81-0012RK-01 Calibration cylinder (50% LEL Methane; 34 liter) 81-0012RK-03 Calibration cylinder (50% LEL Methane; 103 liter) 81-0076RK-01 Zero air calibration cylinder (34 liter) 81-1001RK Dispensing valve (with knob), for 34 liter cylinders 81-1054RK Regulator (demand ow), for 103 liter cylinders 81-1126RK Gas collection bag (2 liter)
18 • Sample Draw Combustible Gas Detector
Loading...