35-3000RKA-LEL Sample Draw Combustible Gas
Detector
Specifications
Table 1 lists specifications for the sample draw combustible gas detector.
Table 1: Specifications
Target GasCombustible Gas
Input Power24 VDC
Construction (housing)Fiberglass/polyester (NEMA 4X)
Dimensions8.5 in. H x 6.5 in. W x 4.25 in. D
Weight4.5 lbs.
Sampling Methodsample draw
Sample Flow1.5 SCFH (nominal)
Detection Range0 to 100% LEL
Response Time90% in 30 seconds
Linearity
Repeatability
5% of detection range
2% of detection range
Sample Draw Combustible Gas Detector • 1
O
Description
This section describes the components of the sample draw combustible gas detector. The
sample draw detector consists of the housing, flow system, and detection system.
Mounting
Flow Adjust
Potentiometer
Pu mp /flo w sw itch
terminal block
Fail LED
Pilot LED
Flowmeter
Bypass valve
Particle filter
LE L Transmitter
Foot, 4X
Flow Alarm Setpoint
Adjustment Potentiometer
Se nso r Wiring
Terminal St rip
External Wiring
Terminal Strip
Exhaust
Inlet
Factory Set Pots
Test Points
100 - 500 mV range
JUMPER PINS FOR
FACTORY U SE ONLY
Span Pot
Zero Pot
Housing
The sample draw detector’s fiberglass housing is weather- and corrosion-resistant. It is
suitable for installation where general purpose equipment is in use.
LEL
sensor
P/N 57-1050RK
REV. 0
GND
24V
TP +
4-20
POWER/SIG
RED
WHT
SENSOR
GRN
BLK
ZER
SPAN
TP
Terminal Block
3/4" Conduit
Hub, 2X
Figure 1: Sample draw Combustible Gas Detector Component Location
The housing door is hinged on the left side and is secured by two latches on the right side.
The flowmeter and status lights are visible through a window in the housing door.
Four mounting feet are attached to the back of the housing (one at each corner). Use the
mounting feet to install the housing to a vertical surface. Use the two conduit hubs on the
bottom of the housing to make wiring connections.
An aluminum subpanel is mounted to the interior of the housing. The sample draw
detector’s internal components are mounted to the subpanel.
2 • Sample Draw Combustible Gas Detector
Flow System
The sample draw detector’s flow system consists of the INLET fitting, filter, pump,
flowmeter, bypass valve, status lights, pressure switch, and EXHAUST fitting (seeFigure 1). Figure 2 illustrates how the gas sample moves through the flow system.
To
Exhaust
Bypass
Valve
Flowmeter
InletFilterPumpSensor
Pressure Switch
Figure 2: Sample draw Combustible Gas Detector Flow Diagram
INLET fitting
The INLET fitting on the right side of the housing allows the gas sample to enter the
sample draw detector. The INLET fitting accepts 1/4 in. rigid tubing. See the Installation
section on page 7 to connect tubing to the INLET fitting.
Filter
The dust filter is below the main circuit board. The filter prevents particulates in the
incoming gas sample from damaging the flow and detection systems. Replace the filter
when it appears dirty, discolored, or clogged.
Pump
The pump is behind the main circuit board near the top of the sample draw detector.
The pump pulls the gas sample into the sample draw detector. The pump operates on
24 VAC, which is generated from the 24 VDC supplied to the sample draw detector.
Flowmeter
The flowmeter is attached to the main circuit board near the top left corner (see Figure 1.)
You can see it through the window in the door. A ball in the flowmeter column indicates
the flow rate of the sample draw detector. The flowmeter measures the flow in the range
0.2 to 2.0 SCFH (Standard Cubic Feet per Hour). The optimum flow rate is 1.5 SCFH.
Bypass valve
The bypass valve is to the left of the flowmeter. The bypass valve adjusts the flow rate to
the sensor. Use a flat-blade screwdriver to adjust the bypass valve.
NOTE:The bypass valve allows fine adjustments of the flow rate. For a wider range of
adjustment, use the flow adjust potentiometer (see Figure 1.)
Sample Draw Combustible Gas Detector • 3
Status lights
Two status lights are above the flowmeter. They are also visible through the window in
the housing door.
Pilot light
The green Pilot light is on when the sample draw detector is receiving power.
Fail light
The red Fail light is on when the sample flow rate is below the low flow level.
NOTE:The default low flow level is 0.6 SCFH (±0.2). See “Adjusting the Low Flow
Setting” on page 15 to adjust this setting.
Pressure switch
The pressure switch is mounted to the opposite side of the main circuit board. The
pressure switch monitors the flow rate of the incoming gas sample.
If the flow rate falls below the preset low flow level, the pressure switch causes the fail
relay to interrupt the signal in the 4-20 mA line. This causes a downscale reading at the
monitor. The low flow level is factory-set at 0.6 SCFH (±0.2 SCFH).
EXHAUST fitting
The EXHAUST fitting on the right side of the housing allows the gas sample to exit the
sample draw detector. The EXHAUST fitting accepts 1/4 in. rigid tubing. See the
Installation section on page 7 to connect tubing to the EXHAUST fitting.
Detection System
The detection system consists of the combustible gas sensor, the hydrogen transmitter,
and the main circuit board.
Combustible gas sensor
The combustible gas sensor is installed in a cavity block. The cavity block is mounted to
the aluminumsubpanel near the bottom of the sample draw detector. The combustible
gas sensor includes the sensing elements, flame arrestor, connector, and sensor leads.
NOTE:The cavity block includes a cavity for an oxygen sensor. This version of the
sample draw detector does not include the oxygen sensor.
Sensing elements
Two sensing elements are protected within the sensor assembly. Through a series of
thermal and electronic reactions, these elements produce an output that is pr oportional to
the detection range of the sample draw detector . The transmitter converts the output to a 4
- 20 mA signal which can be used by a recording or monitoring device.
The porous flame arrestor allows the gas sample to enter the sensor assembly and contact
the sensing element. The flame arrestor also contains sparks within the sensor.
4 • Sample Draw Combustible Gas Detector
Connector
The top of the sensor includes five pins that plug into the socket connector . This connector
allows you to replace the sensor without disconnecting the wiring. The sensor leads are
soldered to the connector.
Sensor leads
Four color-coded leads extend from the connector. The leads allow you to connect the
combustible gas sensor to the main circuit board.
Transmitter
The transmitter is mounted to the left of the combustible sensor. It consists of the span
pot, zero pot, transmitter interconnect terminal strip, sensor terminal strip and test points.
Span/zero pots
The span and zero pots are located at the bottom edge of the transmitter and are used for
calibration. Use the span pot to make adjustments to gas response readings and the zero
pot to make adjustments to the zero reading
Transmitter interconnect terminal strip
The transmitter interconnect terminal strip is the six-point terminal strip near the right
edge of the transmitter. The transmitter is factory wired to the sensor and main circuit
board.
Test points
The test points are located on the left side of the transmitter and are labeled TP+ and TP-.
A 100 mV - 500 mV output is available at these testpoints for use during calibration.
Main Circuit Board
The main circuit board includes the interconnect terminal strip, transmitter terminal strip,
pump terminal strip, and relay (see Figure 1).
NOTE:The flowmeter and status lights are mounted to the main circuit board but are
considered part of the flow system.
Interconnect terminal strip
The interconnect terminal strip is the nine-point terminal strip near the bottom edge of the
main circuit board. Use the interconnect terminal strip to connect the sample draw
detector to power and external devices.
Transmitter terminal strip
The transmitter terminal strip is the nine-point terminal strip near the right edge of the
circuit board. Use the transmitter terminal strip to connect the transmitter to the main
circuit board.
NOTE:The transmitter is factory-wired to the circuit board. See the “Installation” section
on page 7 for all wiring procedures related to the sample draw detector.
Sample Draw Combustible Gas Detector • 5
Pump terminal strip
The pump terminal strip is the four-point terminal strip near the top edge of the circuit
board. Use the pump terminal strip to connect the pump and pressure switch to the main
circuit board.
NOTE:The pump and pressure switch are factory-wired to the circuit board. See the
“Installation” on page 7 for all wiring procedures related to the sample draw
detector.
Relay
The relay is to the left of the detector terminal strip. The relay is single-pole, double-throw
(SPDT) and is rated for 2 amps at 25 VDC (resistive). If the pressure switch senses a low
flow condition, the relay interrupts the signal from the transmitter. The interrupted
transmitter signal causes a fail condition at the Pioneer.
6 • Sample Draw Combustible Gas Detector
Loading...
+ 12 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.