RKI Instruments 35-3000RKA-H2S User Manual

Sample-Draw Hydrogen Sulfide Gas Detector for Use With Generic Device
Table 1 lists specifications for the sample-draw hydrogen sulfide detector. .
Table 1: Specifications
Target Gas Hydrogen Sulfide Input Power 24 VDC Output Signal 4-20 mA Construction (housing) Fiberglass/polyester (NEMA 4X) Dimensions 8.5 in. H x 6.5 in. W x 4.25 in. D Weight 4.5 lbs. Sampling Method Sample-draw Sample Flow 1.5 SCFH (nominal) Detection Range 0 to 100 ppm Response Time 90% in 30 seconds Accuracy ±5% of detection range Repeatability ±2% of detection range
Sample-Draw Hydrogen Sulfide Gas Detection • 1
Description
This section describes the components of the sample-draw hydrogen sulfide gas detector. The sample-draw detector consists of the housing, flow system, and detection system.
Mounting Foot,
Flow Adjust Potentiometer
Pump/flow switch terminal block
4X
Main Circuit Board
Flow Alarm Setpoint Adjustment Potentiometer
Fail LED
Pilot LED
Flowmeter Bypass valve
Particle filter
Transmitter
Test Points 100 - 500 mV range
Factory Set Pot
Zero Pot Span Pot
JUMPER PINS FOR FACTORY USE ONLY
ZERO
SPAN
TOXIC
Transmitter Te rminal Strip
Interconnect Terminal Strip
Exhaust
Inlet
Sensor Connector
24VFB4/20BATT
Hydrogen Sulfide Sensor
OXY
TOXIC
OXY
G
BKRD W
Not Used
3/4" Conduit Hub , 2X
Figure 1: Sample-draw Hydrogen Sulfide Gas Detector Component Location
Housing
The sample-draw detector’s fiberglass housing is weather- and corrosion-resistant. It is suitable for installation where general purpose equipment is in use.
The housing door is hinged on the left side and is secured by two latches on the right side. The flowmeter and status lights are visible through a window in the housing door.
Four mounting feet are attached to the back of the housing (one at each corner). Use the mounting feet to install the housing to a vertical surface. Use the two conduit hubs on the bottom of the housing to make wiring connections.
An aluminum subpanel is mounted to the interior of the housing. The sample-draw detector’s internal components are mounted to the subpanel.
2 • Sample-Draw Hydrogen Sulfide Gas Detection
Flow System
The sample-draw detector’s flow system consists of the INLET fitting, filter, pump, flowmeter, bypass valve, status lights, pressure switch, and EXHAUST fitting (see Figure 1). Figure 2 illustrates how the gas sample moves through the flow system.
To
Exhaust
Bypass
Valve
Flowmeter
Inlet Filter Pump Sensor
Pressure Switch
Figure 2: Sample-draw Hydrogen Sulfide Gas Detector Flow Diagram
INLET fitting
The INLET fitting on the right side of the housing allows the gas sample to enter the sample-draw detector. The INLET fitting accepts 1/4 in. rigid tubing. See the Installation section on page 6 to connect tubing to the INLET fitting.
Filter
The dust filter is below the main circuit board. The filter prevents particulates in the incoming gas sample from damaging the flow and detection systems. Replace the filter when it appears dirty, discolored, or clogged.
Pump
The pump is behind the main circuit board near the top of the sample-draw detector. The pump pulls the gas sample into the sample-draw detector. The pump operates on 24 VAC, which is generated from the 24 VDC supplied to the sample draw detector.
Flowmeter
The flowmeter is attached to the main circuit board near the top left corner (see Figure 1.) You can see it through the window in the door. A ball in the flowmeter column indicates the flow rate of the sample-draw detector. The flowmeter measures the flow in the range
0.2 to 2.0 SCFH (Standard Cubic Feet per Hour). The optimum flow rate is 1.5 SCFH.
Bypass valve
The bypass valve is to the left of the flowmeter. The bypass valve adjusts the flow rate to the sensor. Use a flat-blade screwdriver to adjust the bypass valve.
NOTE: The bypass valve allows fine adjustments of the flow rate. For a wider range of
adjustment, use the flow adjust potentiometer (see Figure 1.)
Sample-Draw Hydrogen Sulfide Gas Detection • 3
Status lights
Two status lights are above the flowmeter. They are also visible through the window in the housing door.
Pilot light
The green Pilot light is on when the sample-draw detector is receiving power from the Pioneer.
Fail light
The red Fail light is on when the sample flow rate is below the low flow level.
NOTE: The default low flow level is 0.6 SCFH (±0.2). See “Adjusting the Low Flow
Setting” on page 13 to adjust this setting.
Pressure switch
The pressure switch is mounted to the opposite side of the main circuit board. The pressure switch monitors the flow rate of the incoming gas sample.
If the flow rate falls below the preset low flow level, the pressure switch causes the fail relay to interrupt the signal in the 4-20 mA line. This causes a downscale reading at the monitor. The low flow level is factory-set at 0.6 SCFH (±0.2 SCFH).
EXHAUST fitting
The EXHAUST fitting on the right side of the housing allows the gas sample to exit the sample-draw detector. The EXHAUST fitting accepts 1/4 in. rigid tubing. See the Installation section on page 6 to connect tubing to the EXHAUST fitting.
Detection System
The detection system consists of the hydrogen sulfide gas sensor, the hydrogen sulfide transmitter, and the main circuit board.
Hydrogen sulfide gas sensor
The hydrogen sulfide gas sensor is installed in a cavity block. The cavity block is mounted to the aluminum subpanel on the right of the sample-draw detector. The hydrogen sulfide gas sensor includes the sensor and the cable with connector.
NOTE: The cavity block includes a cavity for a second hydrogen sulfide or a carbon
monoxide sensor. This version of the sample-draw detector does not include the second sensor.
Sensor
The sensor is pushed into the cavity block and held in place by two o-rings. Through a series of chemical and electrical reactions, the sensor produces an electrical output that is proportional to the detector range of the transmitter.
Cable w/Connector
The sensor has a two wire cable with a 4-pin connector soldered to it. The connector is plugged into the sensor connector in the lower right of the Main Circuit Board.
4 • Sample-Draw Hydrogen Sulfide Gas Detection
Hydrogen Sulfide Transmitter
The hydrogen sulfide transmitter is mounted to the left of the hydrogen sensor. It consists of the span pot, zero pot, two internally wired terminal strips, and the test points.
Span/zero pots
The span and zero pots are located at the left edge of the transmitter and are used for cali­bration. Use the span pot to make adjustments to gas response readings and the zero pot to make adjustments to the zero reading
Termainal Strips
There are two terminal strips which are used for internal factory wiring.
Test points
The test points are located at the top of the transmitter and are labeled TP+ and TP- . A 100 mV - 500 mV output is available at these testpoints for use during calibration.
Main Circuit Board
The main circuit board includes the interconnect terminal strip, transmitter terminal strip, pump terminal strip, and relay (see Figure 1).
NOTE: The flowmeter and status lights are mounted to the main circuit board but are
considered part of the flow system.
Interconnect terminal strip
The interconnect terminal strip is the nine-point terminal strip near the bottom edge of the main circuit board. Use the interconnect terminal strip to connect the sample-draw detector to power and an external device.
Transmitter terminal strip
The transmitter terminal strip is the nine-point terminal strip near the right edge of the circuit board. Use the transmitter terminal strip to connect the transmitter to the main circuit board.
NOTE: The transmitter is factory-wired to the circuit board. See the “Installation” section
on page 6 for all wiring procedures related to the sample-draw detector.
Pump terminal strip
The pump terminal strip is the four-point terminal strip near the top edge of the circuit board. Use the pump terminal strip to connect the pump and pressure switch to the main circuit board.
NOTE: The pump and pressure switch are factory-wired to the circuit board. See the
“Installation” on page 6 for all wiring procedures related to the sample-draw detector.
Relay
The relay is to the left of the detector terminal strip. The relay is double pole, double­throw (DPDT) and is rated for 2 amps at 25 VDC (resistive). If the pressur e switch senses a low flow condition, the relay interrupts the signal from the sensor. The interrupted sensor signal causes a fail condition at the Pioneer.
Sample-Draw Hydrogen Sulfide Gas Detection • 5
Installation
This section describes procedures to mount the sample-draw hydrogen sulfide gas detector in the monitoring environment and wire the sample-draw detector to power and an external device.
Mounting the Sample-Draw Combustible Gas Detector
1. Select the mounting site. Consider the following when you select the mounting site.
Is there enough room to open the housing door and make wiring connections at the bottom of the housing and tubing connections at the right of the housing?
Make sure there is sufficient room to perform start-up, maintenance, and calibration procedures.
Are the flowmeter and status lights visible?
7.3 in.
6.5 in.
4.0 in.
8.5 in.
8.9 in.
9.3 in.
3/4 in. conduit hub
Note: The housing is 4.25 in. deep.
Figure 3: Mounting the Sample-Draw Hydrogen Sulde Gas Detector
(total of 2 hubs)
6 • Sample-Draw Hydrogen Sulfide Gas Detection
2. Close and latch the housing door.
NOTE: The sample-draw detector is shipped with the mounting feet tucked under the
housing to protect the mounting feet during shipment.
3. Slightly loosen the screw that secures the mounting foot to the housing, then rotate the mounting foot 180 degrees (see Figure 3).
4. Tighten the screw that secures the mounting foot to the housing.
5. Repeat steps 3 and 4 for the remaining three mounting feet.
6. Position the sample-draw housing on a vertical surface at eye level (4 1/2 to 5 feet from the oor).
7. Insert 1/4 in. or 5/16 screws through the slots in the mounting feet to secure the housing to the mounting surface.
Connecting the Sample Lines to the Sample-Draw Hydrogen Sulfide Gas Detector
1. Attach 1/4 in. O.D. stainless steel, rigid teon, or rigid polypropylene sample tubing to the INLET tting.
CAUTION: If you use flexible sample tubing (polyurethane is acceptable), use an appropriate
metal insert to seal the connection between the tubing and the INLET fitting. See Appendix A, Parts List, for an example of an appropriate metal insert.
2. Place the opposite end of the tubing at the sampling area.
CAUTION: Avoid loops or slumps in the incoming sample line. To r educe r esponse time, keep the
incoming sample line as short as possible.
3. Attach rigid sample tubing to the EXHAUST tting.
4. Route the opposite end of the tubing to an open area where the sample can safely disperse.
Wiring the Sample-Draw Hydrogen Sulfide Gas Detector
WARNING: Always verify that the power source is OFF before you make wiring
connections.
1. Unlatch and open the housing door of the sample-draw detector.
2. Guide a three-conductor, shielded cable or three wires in conduit through one of the conduit hubs at the bottom of the sample-draw housing.
3. Connect the cable to the sample-draw detectors interconnect terminal strip as shown in Figure 4.
4. Close and latch the housing door of the sample-draw detector.
CAUTION: Leave the shield cable insulated and disconnected at the sample-draw detector.
You will connect the opposite end of the shield cable at the device.
Sample-Draw Hydrogen Sulfide Gas Detection • 7
5. Route the cable or wires in conduit leading from the sample-draw detector to the monitoring device and power.
6. Connect the cable shield to an available chassis ground at the device end.
H2S Sample Draw Housing
PUMP ASSY INTERNALLY
WIRED.
ZERO
SPAN
PUMP
OXY
TOXIC
Not Used
BATT 4/20
TOXIC
PSW
FB
24V
OXY
WBKRD G
PRESSURE SWITCH INTERNALLY
WIRED.
PCBIN SINGLE POINT SAMPLE DRAWING
DETECTOR ASSY
Transmitter, Internally
wired
H2S Sensor, Internally Connected
NOT USED ONTHIS
VERSION
115VAC
RDHN
CGND
LEL/O2 AMPP-AMP
_
4/20 +GND 24VWHT GRN BLK
4/20 (Feedback)
+
24VDC
-
Figure 4: Wiring the Sample-Draw Hydrogen Sulde Gas Detector
8 • Sample-Draw Hydrogen Sulfide Gas Detection
Start Up
This section describes procedures to start up the sample-draw hydrogen sulde gas detector and place the sample-draw detector into normal operation.
Introducing Incoming Power
1. Complete the installation procedures described earlier in this manual.
2. Verify that the power /device wiring is correct and secure.
3. Turn on or plug in the incoming power at the power source end.
4. Verify that the Pilot light is on.
5. Verify that the owmeter indicates a ow rate of approximately 1.5 SCFH. If necessary, use the bypass valve or ow adjust potentiometer to adjust the ow rate.
NOTE: The following step tests for leaks in the sample line. This test may cause a low
ow condition at the sample-draw detector.
6. Verify that the incoming sample line is not leaking. To test the sample line, plug the open end of the sample line with your thumb. If the owmeter ball drops to the bottom of the owmeter, the incoming sample line is not leaking.
7. Remove your thumb from the sample line, verify the owmeter returns to a normal ow rate.
Setting the Zero Reading
CAUTION: If you suspect the presence of hydrogen sulfide gas in the monitoring environment,
use the calibration kit and the zero air calibration cylinder to introduce “fresh air” to the sensor and verify an accurate zero setting.
1. Verify that the sample-draw detector is sampling a fresh air environment (environment known to be free of combustible gas).
2. Open the housing door.
3. Set a voltmeter to measure in the milivolt (mV) range.
4. Plug the voltmeter leads into the test points on the amplier. Plug the positive lead into the test point labeled TP+ ; plug the negative lead into the test point labeled TP- .
5. Verify a voltmeter reading of 100 mV (± 2 mV).
6. If necessary, use a small at-blade screwdriver to adjust the zero potentiometer until the voltmeter reading is 100 mV (± 2 mV).
7. Close the housing door.
Sample-Draw Hydrogen Sulfide Gas Detection • 9
Maintenance
This section describes maintenance procedures. It includes preventive maintenance procedures. This section also includes procedures to troubleshoot the sample-draw detector, replace components of the sample-draw hydrogen sulde gas detector, and adjust the low ow setting.
Preventive Maintenance
This section describes a preventive maintenance schedule to ensure the optimum performance of the sample-draw detector. It includes daily, monthly, and quarterly procedures.
Daily
1. Verify that the pilot light is on.
2. Verify that the owmeter indicates a ow rate of approximately 1.5 SCFH.
3. Verify a reading of 0 ppm (4 mA) at the monitoring device or a reading of 100 mV at
Monthly
If necessary use the bypass valve or ow adjust potentiometer to adjust the ow rate to 1.5 SCFH.
the transmitter test points. Investigate signicant changes in the reading.
This procedure describes a test to verify that the sample-draw hydrogen sulde gas detector responds properly to the target gas.
Preparing for the response test
1. Verify that the monitoring device is reading 0 ppm (4 mA). If the reading is not 0, set the zero reading as described in the Start Up section on
page 9, then continue this procedure.
2. Assemble the calibration kit as described in the Calibration section of this insert.
Performing the response test
NOTE: This procedure describes the RKI calibration kit that includes a gas collection
bag. A calibration kit that uses a demand ow regulator is also available.
1. Connect the calibration tubing from the gas collection bag to the inlet line at or near the INLET tting.
The sample-draw detectors pump automatically begins pulling the test sample from the gas collection bag when you connect the tubing to the inlet line.
2. After approximately one minute, verify that the reading at the monitoring device stabalizes within ± 10% of the concentration of the test sample. If the reading is not within ± 10% of the test sample, calibrate the sample-draw detector as described in the Calibration section of this manual.
3. Remove the calibration tubing from the inlet line, then reconnect the inlet line.
Quarterly
Calibrate the sample-draw detector as described in the Calibration section on page 14 .
10 • Sample-Draw Hydrogen Sulfide Gas Detection
Troubleshooting
The troubleshooting guide describes symptoms, probable causes, and recommended action for problems you may encounter with the sample-draw hydrogen sulde gas detector.
NOTE: This troubleshooting guide describes sample-draw detector problems only. See
the instruction manual for the monitoring device if it exhibits any problems.
Fail condition
Symptoms
The sample-draw detectors Fail light is on.
The monitoring device is operating properly but indicates a reading well below zero.
Probable causes
The sample-draw detectors ow rate is too low because of an obstructed sample line, failed pump, etc.
The sample-draw detector is malfunctioning.
The sensor or transmitter wiring is disconnected or misconnected.
Recommended action
1. At the sample-draw detector, set the correct ow rate with the bypass valve or ow adjust potentiometer.
2. If you cannot set the correct ow rate, check the sample lines for obstructions or kinks.
3. Verify that the detector and transmitter wiring are correct and secure. The Installation section on page 6 describes detector wiring connections.
4. Calibrate the sample-draw detector as described in the Calibration section on page 14.
5. If the fail condition continues, replace the sensor as described later in this section.
6. If the fail condition continues, contact RKI Instruments, Inc., for further instruction.
Slow or no response/difficult or unable to calibrate
Symptoms
The detector responds slowly or does not respond during the monthly response test.
Unable to accurately set the zero or response reading during the calibration
procedure.
The detector requires frequent calibration.
NOTE: Under normal circumstances, the detector requires calibration once a quarter.
Some applications may require a more frequent calibration schedule.
Probable causes
The calibration cylinder is low, out-dated, or defective.
The sample-draw detectors ow rate is too low because of an obstructed sample line,
failed pump, etc.
The sample-draw detector is malfunctioning.
Sample-Draw Hydrogen Sulfide Gas Detection • 11
Recommended action
1. Verify that the calibration cylinder contains an adequate supply of a fresh test sample.
2. If necessary, set the correct ow rate with the bypass valve or ow adjust potentiometer.
3. If you cannot set the correct ow rate, check the sample line for obstructions or kinks.
4. If the calibration/response difculties continue, replace the sensor as described later in this section.
5. If the calibration/response difculties continue, contact RKI Instruments, Inc., for further instruction.
Replacing Components of the Hydrogen Sulfide Gas Sample-draw Detector
This section includes procedures to replace the sensor, lter, and ferrules.
Replacing the hydrogen sulfide gas sensor
1. Turn off incomming power.
2. Open the housing door of the sample-draw detector.
3. Unplug the sensor connector from the main circuit board.
4. Grasp the sensor rmly and pull it out of the ow chamber.
5. Install the new sensor in the ow chamber. Make sure it is pushed all the way down.
6. Connect the sensor connector to the main circuit board.
7. Turn on incomming power.
CAUTION: Allow the replacement sensor to warm up for 15 minutes before you continue.
8. Calibrate the replacement sensor as described in the Calibration section on page 14.
Replacing the filter
1. Open the housing door of the sample-draw detector.
2. Note the direction of the arrow on the lter. The arrow indicates the direction of the sample ow.
3. Disconnect the lter from the elbows on each end of the lter, then remove the lter from the sample-draw detector.
4. Make sure the arrow is pointing in the same direction as the arrow on the lter you removed, then connect each end of the replacement lter to the elbows.
5. Verify that the ow rate is approximately 1.5 SCFH, then close the housing door.
Replacing the ferrules
The INLET and EXHAUST ttings each includes two ferrules that seal the incoming or exhaust tubing to the tting. Replace the ferrules if the seal is bad or if you replace the sample tubing. Always replace the ferrules as a pair.
1. Disconnect the sample tubing from the tting, then unscrew the nut from the tting.
2. Verify that the ferrules did not remain in the nut. If necessary, remove the ferrules from the nut.
3. Position the nut so the threaded end is facing you, then insert the back (smaller)
12 • Sample-Draw Hydrogen Sulfide Gas Detection
ferrule into the nut. Insert the ferrule so the at side is facing down.
NOTE: Make sure the bottom ferrule is laying at in the nut.
4. Insert the cone-shaped front ferrule on top of the bottom ferrule. Insert the ferrule so the smaller end of the cone is facing up.
5. Screw the nut onto the tting, then connect the sample tubing to the tting. Make sure you rmly tighten the tubing to the tting.
Adjusting the Low Flow Setting
The factory-set low ow setting is 0.6 SCFH (±0.2). To adjust the low ow setting:
1. Use the ow adjust potentiometer (VR1) to set the ow to 0.6 SCFH. If the sample-draw detector goes into low ow alarm before you can adjust the ow
down to 0.6 SCFH, adjust the low ow potentiometer 1/4 turn clockwise, then attempt to set the ow again. Repeat this step until you are able to adjust the ow to
0.6 SCFH.
2. Slowly turn the low ow potentiometer counterclockwise just until the sample-draw detector goes into low ow alarm.
NOTE: The low ow potentiometer is accessible through a circular cutout in the main
circuit board. The cutout is labeled PS1.
3. Increase the ow using VR1 until the unit is out of low ow alarm.
4. Decrease the ow very slowly and verify that the low ow alarm is 0.6 SCFH (±0.2). If the low ow alarm is set too low, turn the low ow potentiometer slightly
clockwise. Repeat steps 3 and 4 if necessary.
5. Use the ow adjust potentiometer (VR1) to set the ow to 1.5 SCFH.
6. Make sure the sample draw detectors Fail light is off.
Sample-Draw Hydrogen Sulfide Gas Detection • 13
Calibration
This section describes how to calibrate the sample-draw combustible gas detector. It includes procedures to assemble the calibration kit, set the zero reading, set the response reading, and return to normal operation.
NOTE: This procedure describes calibration using a gas collection bag. A demand-ow
calibration kit is also available for calibrating the hydrogen sulde gas sample­draw detector.
Preparing for Calibration
1. Open the housing door.
2. Set a voltmeter to measure in the millivolt (mV) range.
3. Plug the positive lead into the transmitter test point labeled TP+; plug the negative lead into the transmitter test point labeled TP-.
4. Use the following formula to determine the correct test points output for the calibrating sample.
Output (mV) = (calibrating sample/fullscale) X 400 + 100
For example, with a calibrating sample of 25 ppm hydrogen and a fullscale setting of 100 ppm hydrogen, the correct output is 200 mV.
200 (mV) = (25/100) X 400 +100
Assembling the Calibration Kit
NOTE: If you can verify a fresh air environment, it is not necessary to use a zero air
calibration cylinder to set the zero reading. Go to the next section, Setting the Zero Reading.
1. Connect the calibration kit sample tubing to the tting on the gas collection bag.
2. Connect the sample tubing from the gas collection bag to the inlet line at or near the INLET tting.
Allow the sample-draw pump to draw out any residual gas in the gas collection bag.
3. Disconnect the calibration kit sample tubing from the inlet line.
4. Close the clamp right away. The clamp is attached to the calibration kit sample tubing.
5. Connect the tubing from the gas collection bag to the xed ow regulator, then open the clamp.
6. Screw the xed ow regulator onto the zero air calibration cylinder. The gas collection bag begins to ll.
7. When the bag is full, unscrew the xed ow regulator form the cylinder.
8. Close the clamp, then disconnect the sample tubing from the xed ow regulator.
14 • Sample-Draw Hydrogen Sulfide Gas Detection
Setting the Zero Reading
1. Open the clamp, then connect the sample tubing from the gas collection bag to the sample-draw detectors inlet line. This step is not necessary if you veried a fresh
air environment earlier in this procedure.
2. Allow the reading to stabilize for approximately 1 minute.
3. Verify a voltmeter reading of 100 mV (± 2 mV).
4. If necessary, use a small at-blade screwdriver to adjust the zero potentiometer until the voltmeter reading is 100 mV (± 2 mV).
5. Allow the sample-draw pump to draw out any residual gas in the gas collection bag.
Steps 6 through 8 are not necessary if you veried a fresh air environment earlier in this procedure. Go to step 9.
6. Disconnect the sample tubing from the inlet line, then close the clamp.
7. Connect the sample tubing from the gas collection bag to the xed ow regulator, then open the clamp.
8. Screw the xed ow regulator onto the calibration gas cylinder. The gas collection bag begins to ll.
9. Unscrew the xed ow regulator from the cylinder when the gas collection bag appears full.
10. Close the clamp, then disconnect the sample tubing from the xed ow regulator.
Setting the Response Reading
1. Open the clamp, then connect the sample tubing from the gas collection bag to the inlet line at or near the sample-draw detectors INLET tting.
2. Allow the sample-draw detector to respond to the calibrating sample for approximately 1 minute.
3. When the reading on the voltmeter stabilizes, verify that the reading matches the response reading (±2 mV) you determined earlier.
4. If necessary, use the span potentiometer on the amplier to adjust the reading to match the correct response reading.
5. Allow the sample-draw pump to draw out any residual gas in the gas collection bag.
6. Disconnect the sample tubing from the sample-draw detectors inlet line, then close the clamp.
7. Reconnect the incoming sample line.
8. Wait 1 to 2 minutes to allow the hydrogen gas reading to decrease and stabilize.
9. Store the components of the calibration kit in a safe and convenient place.
Sample-Draw Hydrogen Sulfide Gas Detection • 15
Parts List
Table 4 lists replacement parts and accessories for the sample-draw hydrogen sulde gas detector.
Table 2: Parts List
Part Number Description
06-1248RK Sample tubing, 3/16 x 5/16, specify length, (for calibration kit) 17-2593RK Brass insert (for INLET and EXHAUST ttings) 17-2683RK Front ferrule (for INLET and EXHAUST ttings) 17-2688RK Back ferrule (for INLET and EXHAUST ttings) 30-0610RK Pump 33-0163RK Filter (Boston DFU9933-05-DQ) 65-2037RK Hydrogen sulde sensor w/cable and 4-pin connector 81-0151RK-02 Calibration cylinder, 25 ppm H2S in nitrogen, 58 liters 81-0076RK-03 Zero air calibration cylinder (103 liter) 81-1005RK Regulator, xed ow, 6 LPM 81-1126RK Gas collection bag (2 liter)
16 • Sample-Draw Hydrogen Sulfide Gas Detection
Loading...