RKC Instrument Inc DP-700AE, DP-700BE User manual

Page 1
Handheld Digital
P
Thermometer
D
-70
Instruction Manual
RKC INSTRUMENT INC.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
IMR01X01-E5
Page 2
Microsoft , Windows , Excel are the registered trademark of Microsoft Corporation. Company names and product names used in this manual are the trademarks or registered trademarks of the respective companies.
All Rights Reserved, Copyright 2003, RKC INSTRUMENT INC.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 3
Thank you for purchasing this RKC product. In order to achieve maximum performance and
f
ensure proper operation of the instrument, carefully read all the instructions in this manual. Please place the manual in a convenient location for easy reference.
SYMBOLS
WARNING
CAUTION
!
: This mark indicates precautions that must be taken if there is danger o
electric shock, fire, etc., which could result in loss of life or injury.
: This mark indicates that if these precautions and operating procedures
are not taken, damage to the instrument may result.
: This mark indicates that all precautions should be taken for safe usage.
: This mark indicates important information on installation, handling and operating procedures.
: This mark indicates supplemental information on installation, handling and operating procedures.
: This mark indicates where additional information may be located.
WARNING
!
Never use this product for medical equipment.
This product must be used in accordance with the specifications
to prevent fire or damage to the instrument and the equipment.
This product is not intended for use in locations subject to
flammable or explosive gases.
If any abnormality such as smoke, smell of burning or noise was
found, do not use the product. If so, fire may occur.
RKC is not responsible if this product is repaired, modified or
disassembled by other than factory-approved personnel. Malfunction may occur and warranty is void under these conditions.
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
i-1
Page 4
CAUTION
••All precautions described in this manual should be taken to avoid damage to the instrument or equipment.•• Prevent metal fragments or lead wire scraps from falling inside instrument case to avoid electric shock, fire or malfunction.
Do not drop nor apply strong impact to the product. If so, failure may result. Do not use the product while the battery cover is being removed. If so, failure may
result.
Load a battery so that its polarity will meet the indication on the product. If not, the
battery electrolyte may leak or heat be generated to result in injury or failure.
If the product is not used for a long period of time, remove the battery from the
mainframe. If not, the battery electrolyte may leak to result in failure.
Battery service life may change depending on battery performance, operating
condition and measuring environment.
Use this product within the following ambient temperature and ambient humidity.
- Allowable ambient temperature: 20 to 50 C
- Allowable ambient humidity: 5 to 95 %RH
(Absolute humidity: MAX.W.C 29.3 g/m3 dry air at 101.3 kPa)
Do not use or store the product in places such as listed below:
- Rapid changes in ambient temperature which may cause condensation.
- Corrosive or inflammable gases.
- Direct vibration or shock to the mainframe.
- Direct water, oil, chemicals, vapor or steam splashes. *
- Excessive dust, salt or iron particles. *
- Excessive induction noise, static electricity, magnetic fields or noise.
- Direct air flow from an air conditioner.
- Exposure to direct sunlight.
- Excessive heat accumulation.
* DP-700A: Equal to IP67, DP-700B: Equal to IP54
Turn off the power supply before cleaning the instrument. Do not use a volatile solvent such as paint thinner to clean the instrument.
Deformation or discoloration may occur. Use a soft, dry cloth to remove stains from the instrument.
To avoid damage to the instrument display, do not rub with an abrasive material or
push the front panel with a hard object.
NOTICE
This manual assumes that the reader has a fundamental knowledge of the principles of
electricity, process control, computer technology and communications.
The figures, diagrams and numeric values used in this manual are only for explanation
purpose.
RKC is not responsible for any damage or injury that is caused as a result of using this
instrument, instrument failure or indirect damage.
RKC is not responsible for any damage and/or injury resulting from the use of instruments
made by imitating this instrument.
Periodic maintenance is required for safe and proper operation of this instrument. Some
components have a limited service life, or characteristics that change over time.
Every effort has been made to ensure accuracy of all information contained herein. RKC
makes no warranty, expressed or implied, with respect to the accuracy of the information. The information in this manual is subject to change without prior notice.
No portion of this document may be reprinted, modified, copied, transmitted, digitized,
stored, processed or retrieved through any mechanical, electronic, optical or other means without prior written approval from RKC.
i-2
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
IMR01X01-E5
Page 5
CONTENTS
Page
1. OUTLINE .............................................................................. 1
1.1 Checking the Product ...................................................................................... 1
1.2 Parts Description ............................................................................................. 2
1.3 LCD Displays ................................................................................................... 3
1.4 Function Menu Configuration ........................................................................... 4
2. PREPARATION BEFORE USE ............................................ 5
2.1 Loading Battery ............................................................................................... 5
2.2 Setting Date and Time..................................................................................... 6
2.3 Connecting the Temperature Sensor .............................................................. 7
3. BASIC OPERATION ............................................................. 8
3.1 Turn Power On ................................................................................................ 8
3.2 Measuring Temperatures................................................................................. 9
4. JUDGING MEASURED TEMPERATURES [SIMPLIFIED
JUDGING FUNCTION] ....................................................... 10
4.1 Registering Tag Names (Measured Objects) ................................................. 11
4.2 Setting the Judgment Range of High/Low Limit Temperatures ...................... 14
4.3 Judging Measured Temperatures .................................................................. 15
5. RECORDING MEASURED TEMPERATURES
[LOGGING FUNCTION] ..................................................... 17
5.1 Manual Recording ......................................................................................... 18
5.2 Automatic Recording ..................................................................................... 19
5.3 Retrieving Recorded Data ............................................................................. 21
5.3.1 Retrieving only recorded data with specified tag name (Tag Filter) ................... 21
5.3.2 Retrieving recorded data with specified log registered No. (Log Jump) ............. 22
5.4 Deleting Recorded Data ................................................................................ 23
5.4.1 Deleting only one recorded data item (Delete?) ................................................. 23
5.4.2 Deleting all recorded data items (All Clear?) ...................................................... 24
5.5 Defragmenting Recorded Data ...................................................................... 25
6. PEAK HOLD/BOTTOM HOLD ........................................... 26
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
i-3
Page 6
Page
7. REGISTERING USER NAMES [ONLY FOR DP-700B] ..... 28
8. INITIAL SETTING ............................................................... 32
To select the display language (English/Japanese) ....................................... 32
To select the date format ............................................................................... 32
To select the sensor type ............................................................................... 33
To select the display unit ( C/ F) .................................................................... 33
To select the decimal point position ............................................................... 34
To correct the temperature display value (PV Bias) ....................................... 34
To lessen the fluctuation of a display value caused by noise (Digital Filter) ... 35
Saving battery consumption (Auto Power OFF) ............................................. 35
Locking/Unlocking setting operation (Operation Lock) ................................... 36
9. MANAGING DATA BY PC [ONLY FOR DP-700B] ............ 37
9.1 Connecting PC to DP-700B ........................................................................... 37
9.2 Storing Data ................................................................................................... 38
9.3 Disconnecting DP-700B from PC .................................................................. 41
10. APPENDIX ........................................................................ 42
10.1 Replace the Battery ..................................................................................... 42
10.2 Cleaning ...................................................................................................... 44
10.3 Troubleshooting ........................................................................................... 44
10.3.1 Error displays ................................................................................................... 44
10.3.2 Problem and solution ....................................................................................... 46
10.4 Specifications .............................................................................................. 47
i-4
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
IMR01X01-E5
Page 7
1. OUTLINE
This Chapter describes features, package contents and model code, etc. The handheld digital thermometer DP-700 has the following features:
Wide temperature measurement is possible
Thermocouple K: 200 to 1372 C ( Thermocouple J: 200 to 1200 C ( Thermocouple T: 50 to 400 C ( RTD (Pt100): 200 to 850 C (
Names of measured objects can be registered.
Names of measured objects can be registered using 11 characters maximum. (DP-700A: maximum 5 tags DP-700B: maximum 99 tags)
Simplified Judging Function
Judgment can be immediately made for each measured object. (Temperatures to judge the high/low limits need to be set.)
Logging Function (Manual logging mode, Automatic logging mode)
Measured object, measured temperature, date, and user name can be manually or automatically recorded in the memory. (DP-700A: maximum 99 logs DP-700B: maximum 9999 logs)
Peak and Bottom Hold Function
The maximum and minimum measured-temperatures can always be stored.
328 to 2501 F 328 to 2192 F
58 to 752 F
328 to 1562 F
), 199.9 to 999.9 C (F) ), 199.9 to 999.9 C (F)
), 50.0 to 400.0 C (
), 199.9 to 850.0 C ( 199.9 to 999.9 F)
58.0 to 752.0 F
)
Can be operated for 400 hours continuously using one Type LR6 (based on IEC and JIS) alkaline battery.
Data can be controlled using personal computers (Only for the DP-700B)
Measured data can be stored in the file in the CSV format using the USB port of each personal computer.
1.1 Checking the Product
Before using this product, check each of the following:
Model code Check that there are no scratch or breakage in external appearance. Check that all of the items delivered are complete. (Refer to below)
Model code
DP-700 /
(1) (2)
Accessories:
Instruction Manual (IMR01X01-E5) ··· 1 LR6 (IEC and JIS) Alkaline battery ············ 1 Strap ·········································· 1 USB cable (Only for DP-700B) ··············· 1
(1) Type
A: Economy type
(No USB, Logging: 99 logs, Backed up by SRAM)
B: High performance type
(With USB, Logging: 9999 logs, Backed up by EEPROM)
(2) Display language
J: Japanese E: English
If any of the products are missing, damaged, or if your manual is incomplete, please contact RKC sales office or the agent.
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
1
Page 8
1. OUTLINE
MENU
1.2 Parts Description
DP-700
LCD display
REC
USB connecter
(Only for DP-700B)
Operation keys
S
E
T
/POWER
C
S
E
Sensor connection jack
LCD display
Displays various information on measured temperature, measured date, set details of each function and error number.
Operation keys
MENU/POWER key:
Used to turn on/off the power. Pressing this key for more than 2 seconds can turn off the power.
Can call up or select the function menu screen while the power is turned on.
Can display the list of characters in the message column while characters are being entered.
SET key:
Establishes operation during execution.
ESC key:
Suspends operation during execution. If suspended, returns to the displayed state or screen just before being suspended.
Hooks for carrying the sensor
Battery cover
Strap hole
Sensors that can be attached to the hook are sold separately.
UP
key
( )/DOWN
key
( ):
Can select measured objects or user names already registered while the measurement screen is being displayed.
Can select the list of characters while characters are being entered.
Left shift
key
( )/Right shift
n select display details in the message
Ca
key
( ):
column while the measurement screen is being displayed.
Can move to the preceding or next function menu screen while the function menu (Fn0 to Fn9) screen is being displayed.
Can move the cursor character by character while characters are being entered.
REC
key
:
When in manual logging mode (manual recording), records the measured temperature and date at the time when this key is pressed.
When in automatic logging mode (automatic recording), used as a recording START/STOP key.
2
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
IMR01X01-E5
Page 9
1.3 LCD Displays
• • • • • • • • • • • • • • • • •
The following details are displayed on the LCD screen.
(1) (2) (3) (4)
(7)
(8)
1. OUTLINE
(5)
(6)
• • • • • • • •
(9)
(1) (Number display)
If measured object or user name is displayed, displays the corresponding registered number.
When recorded data items are displayed, displays “Display memory number/Total memory number of recorded data items of those data items.
(2) (Log execution display)
When in manual logging mode (manual recording), displayed only when the REC key is pressed.
When in automatic logging mode (automatic recording), continues lighting during recording and continues flashing during recording stop.
(3) (Burnout display)
Displayed when the sensor is broken, shorted or imperfectly connected.
(4) (
Remaining battery service life display
Displays the remaining battery service life (reference only).
Service life remains long.
Service life is going to shorten.
Service life is going to expire. (Battery replacement is required.) Service life has expired. (Immediate battery replacement is required.)
(5) Measured temperature display
Displays the measured temperature. In addition, displays information on error number, etc. if an error occurs.
(6) C or F (Temperature unit display)
(7) (
Simplified judgment display
Symbols to judged the measured temperature. To judge the measured temperature using these symbols, it is necessary to set the judgment points of its high and low limits.
Measured temperature exceeds judgment point of high limit
NG
temperature. Measured temperature is within
OK
judgment points. Measured temperature is below
judgment point of low limit
NG
temperature.
(8) Message column
)
Displays function menu screen name, measured date1, measured object 3, user name
1
Set the measured date, if necessary. (Refer to P. 6)
2
Not displayed if no user name is registered.
3
Number of registered characters: 11 maximum
2, 3
etc.
(9) ,
Displays message selection.
)
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
3
Page 10
1. OUTLINE
r
1.4 Function Menu Configuration
The DP-700 is provided with a function menu used to select and then set any function necessary for measured-temperature judgment and recording (logging). There for, set the desired function as needed by the customer. For details, refer to the relevant pages.
Power ON
Measurement screen
01 / 01 00:0 0
ESC key
Memory
Menu used to check and retrieve recorded measured-temperatures.
MENU/POWER key
When user name is registered.
SET key
This symbol is always displayed.
Fn0 to Fn8 are menus which are not necessary to be used when only temperatures are measured such as by a thermometer.
(P. 17)
TAG Setup
Menu used to register measured-object names, and to set judging conditions of measured temperatures.
User No.
key
Dick
key, key or key
ESC key or key
(P. 10)
Menu used to set time intervals of automatically recording measured­temperatures.
key
Logging
(P. 19)
Tag No.
key
key
FRY1
Function Memu
(P. 26)
Peak/Bottom
Menu used to check maximum/minimum measured-temperatures.
(P. 6)
SetupUti l i t y
Menu used to set dominical year/month/ day/time and to check information on versions.
: Press the MENU/POWER key or key : Press the key
Menu used to set items relating to initicalizeing setting.
If unnecessary to be changed, use their factory set values.
Initial setting items (factory set value): P. 32
Display language: English PV bias: 0.0 C Date format: month/day Digital filter: OFF Sensor type: TC-K Auto power OFF: 3 min. Display unit: C Operation lock: 0000 (Unlock) Decimal point position: 0.1 (One dicimal place)
UserName
Menu used to registe user names.
4
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
(P. 28)(P. 32)
IMR01X01-E5
Page 11
2. PREPARATION BEFORE USE
This chapter describes preparations necessary for operating this instrument for the first time.
2.1 Loading Battery
Procedures
1. Loosen the battery cover.
Turn the cover with a coin.
2. Remove the battery cover.
3. Load a battery without mistaking its polarity.
4. Then, firmly tighten the battery cover.
Insufficient tightening cannot assure the effect of water resistance (DP-700A: Equal to IP67, DP-700B: Equal to IP54).
After the battery is loaded, the instrument is ready to be reset-started. The function menu “Utility (Fn0)” for setting the dominical year is displayed on the LCD screen. If no
dominical year is set, no calendar function can be used. Always set the year if the calendar function is used.
Pressing the ESC key twice can display the measurement screen.
Alkaline battery (LR6)
Year
Year setting screen
For the battery life or replace method, refer to 10.1
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Replace the Battery
(P. 42).
5
Page 12
2. PREPARATION BEFORE USE
2.2 Setting Date and Time
Set the date and time displayed on the measurement screen. This setting is made on the function menu “Utility (Fn0).” The setting procedure is described using an example where the instrument is reset-started (from start on the Fn0 screen).
Battery setting
Reset start
Automati­cally
Year
Date
Time
ROM version
Return to Year
Press the ESC key twice
Returns to Measurement screen
Setting year
1. Press the or key to move the flashing digit.
2. Press the or key to set the “year.”
3. Press the SET key to store the new set value.
Flashing
Data range: 0000 (Only for Reset start)
2003 to 9999
Year
Factory set value: 0000
Setting month and day
1. Press the or key to set the “day (month).”
2. Press the key to move the blinking cursor to “month (day).”
3. Press the or key to set the “month (day).”
4. Press the SET key to store the new set value.
(“Month/Day” display)
(“Day/Month” display)*
Data range:
Month: 01 to 12 Day: 01 to 31
Factory set value:
01/0 1 m m / d d
01/0 1 d d / m m
01/01 (Month/Day)
: Press the SET
key
Month
* To display “Day/Month,” refer to the date format screen in “Setup (Fn9)” of the Function
menu (P. 32).
Month (Cursor blinking)Day
Day (Cursor blinking)
Setting time
1. Press the or key to set the “minute.”
2. Press the key to move the blinking cursor to “time.”
3. Press the or key to set the “time.”
4. Press the SET key to store the new set value.
Data range: Hour: 00 to 23
Minute: 00 to 59
0 1 / 0 1 0 0 : 0 0
6
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Factory set value: 00:00
Cursor blinking
IMR01X01-E5
Page 13
2. PREPARATION BEFORE USE
MENU
2.3 Connecting the Temperature Sensor
Connect the temperature sensor (sold separately) to the DP-700 mainframe.
Firmly connect the sensor cable to the mainframe so as not to cause imperfect contact.
DP-700
REC
S
E
T
/POWER
C
S
E
Temperature sensor (sold separately)
Temperature sensors for DP-700 are sold separately.
Contact with RKC office or the agent for details of temperature sensors.
Temperature sensor (sold separately)
A-DP700 [Converter plug for “3C/2C to 6C”] (sold separately)
This is a converter plug for RKC’s temperature sensors with 3C or 2C plug for use with DP-700.
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
7
Page 14
3. BASIC OPERATION
j
MENU
This chapter describes the basic operation of the instrument for temperature measurement.
3.1 Turn Power On
Press the MENU/POWER key. The power is turned on to display the measurement screen.
DP-700
Measured temperature
REC
S
E
T
/POWER
C
S
E
Simplified
udgment
display
Date and Time *
11/11 09:00
* If the data and time are not set in the
function menu [Utility (Fn0)], the factory set value is displayed. [Factory set value: 01/01 00:00]
When the power is turned on for the first time after preparations before operation are finished, “Simplified judgment display” is as the above screen example.
For the simplified judgment display, refer to 4. JUDGING MEASURED
TEMPERATURES [SIMPLIFIED JUDGING FUNCTION] (P. 10).
The power is turned off if no key operation is performed for more than the Auto power OFF time (factory set value: 3 min.) with the power turned on.
For the Auto power OFF setting, refer to Saving battery consumption (P. 35).
How to power off DP-700?
Press and hold the MENU/POWER key for 2 seconds or more.
What screen is displayed when the power is turned off and then is turned on again?
If the user name is registered to the user number selected, the type of screen displayed first differs.
For the type of screen displayed, refer to 1.4 Function Menu Configuration (P. 4).
8
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
IMR01X01-E5
Page 15
3. BASIC OPERATION
3.2 Measuring Temperatures
To measure temperatures, insert the sensor into or contact it with the measured object. When measuring temperatures, handle the sensor as instructed in the following.
WARNING
!
In order to prevent injuries, never direct the sensor tip toward a
person.
In order to prevent burning or frostbite, do not touch the sensor
during or just after its use in the high or low temperature environment.
CAUTION
Use the sensor within its temperature measuring range. Firmly insert the sensor cable into the sensor connector of the mainframe. If the sensor gets stained, wipe it with a clean cloth.
Example of display screen during measurement:
When a measured object is at a surface temperature of 90.0 C
Measured temperature
11/1 1 09:05
Measured date
It is possible to record the measured temperature and date in the memory.
For the data recording (logging) method, refer to 5. RECORDING
MEASURED TEMPERATURES [LOGGING FUNCTION] (P. 17).
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
9
Page 16
4. JUDGING MEASURED TEMPERATURES [SIMPLIFIED JUDGING FUNCTION]
This chapter describes settings and procedures to judge measured temperatures using the simplified judging function.
Before using the simplified judging function, register tag names (measured object names) and also set judgment criteria (judgment points of high and low temperatures). The registration and setting are made on the function menu “TAG Setup (Fn2).”
What is the simplified judging function?
This function is for judging whether or not measured temperatures are within the judgment criteria preset (judgment points of high and low temperatures) and then check judgment results according to simplified judgment symbols.
Screen Configurations of TAG Setup (Fn2) Menu:
Press the
Measurement
screen
MENU/POWER key twice
ESC
key
(P. 4)
Tag name screen
SET key
The selected tag name
setting screen
(Example: TAG01)
Maximum number of registration DP-700A: 5 tags DP-700B: 99 tags
SET key
ESC key
key
TAG Setup
SET key
Tag No.
TAG01
Tag name *
Judgment point of
high limit temp.
setting screen
key
SET key
* Each tag name before factory
shipment can be used as it is. When using the tag name before factory shipment, display Tag No. in the message column on the measurement screen.
ESC key
Judgment point of
low limit temp. setting screen
TAG01
(P. 11) (P. 14)
Cursor blinking
MENU/POWER key
Hi Limit
Low Limit
(P. 14)
SET key
Character entry (P. 11)
10
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
IMR01X01-E5
Page 17
4. JUDGING MEASURED TEMPERATURES [SIMPLIFIED JUDGING FUNCTION]
N
r
t
4.1 Registering Tag Names (Measured Objects)
Any measured object mane (maximum number of character: 11 characters) can be registered for each number. The tag name thus registered can be displayed on the measurement screen.
Character entry screen
Example: When the tag name screen of Tag No. 1 is selected.
Tag name setting screen
TAG01
Cursor blinking
Press the ESC key three times
SET key ESC key
SET key ESC key
Returns to Measurement screen
MENU/ POWER key
Character selecting screen
STUVWXYZ
MENU/ POWER key
Delete check screen
Delete?
(P. 13)
MENU/ POWER key
A list of characters including the currently selected character (cursor blinking) is displayed in the message column.
Select the list of characters by the or key and then select the desired character by the or key.
SET key: To execute selection ESC key: To suspended operation
Screen used to check whether or not the currently selected character (curso blinking) is deleted.
SET key: To execute deletion ESC key: To suspend operation
Insert check screen
SET key ESC key
Insert?
(P. 13)
MENU/ POWER key
Characters list
Numerals 0 1 2 3 4 5 67 8 9 Symbols @ ! # $ % & ’ `
Alphabets AB C D E F G H I
J K L M N O P Q R S T U V W X Y Z a b cd e f g hi j k l m n o p q r s t uv w x y z
+ - .; = ^ _ ( ) [ ] { }
Screen used to check whether to inser space for entering a character in the position displayed with the currently selected character (cursor blinking).
SET key: To execute insertion ESC key: To suspend operation
o spaces (no characters)
can be used.
Select the list of characters by the or key.
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
11
Page 18
4. JUDGING MEASURED TEMPERATURES [SIMPLIFIED JUDGING FUNCTION]
Procedures
Example: Change the tag name of Tag No. 2 from “TAG02” to “PAN02”
1. Press the SET key at TAG Setup (Fn2)
screen to display the tag name screen.
TA G 01
2. Press the key.
The display goes to the tag name screen of Tag No. 2.
TA G 02
3. Press the SET key.
The cursor is displayed in the position displayed with the tag name “T.”
5. Press the key. The list of characters including “P” is displayed.
JKLMNOPQR
6. Press the key to move the cursor to the position displayed with “P.”
JKLMNOPQR
7. Press the SET key to store the new set value. The screen returns to the tag name screen of Tag No. 2 and the cursor moves to “A.”
Blinking
TAG0 2
4. Press the MENU/POWER key. The list of characters is displayed in the message column.
8. Press the key to move the cursor to the position displayed with “G.”
PAG0 2
STUVWXYZ
PA G02
12
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
IMR01X01-E5
Page 19
4. JUDGING MEASURED TEMPERATURES [SIMPLIFIED JUDGING FUNCTION]
A
9. Press the MENU/POWER key. The list of characters is displayed in the message column.
BCDEFGHI
10. Press the The list of characters including “N” is displayed.
key.
JKLMNOPQR
11. Press the key to move the cursor to the position displayed with “N.”
When setting the judgment criteria (judgment points of high and low temperatures) succeedingly, press the SET key. (P. 14)
To return to the measurement screen, press the ESC key three times.
How to change the character without displaying the list of characters?
1. Move the cursor to the position displayed with the character to be revised by using the or key.
2. Display the character to be revised by using the or key.
How to delete one character?
1. Move the cursor to the position displayed with the character to be deleted using the or key.
JKLMNOPQR
12. Press the SET key to store the new set value. The screen returns to the tag name screen of Tag No.2 and the cursor moves to “0.”
PA N 02
The setting is finished.
2. Press the MENU/POWER key twice to display the screen for checking the deletion.
3. Press the SET key to delete the character where the cursor blinked.
How to insert space for entering a character in any position?
1. Move the cursor to the position where a character needs to be entered using the or key.
2. Press the MENU/POWER key three times to display the screen for checking the entry.
Press the SET key to insert space for
3.
entering the character before the character position where the cursor blinks.
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
13
Page 20
4. JUDGING MEASURED TEMPERATURES [SIMPLIFIED JUDGING FUNCTION]
4.2 Setting the Judgment Range of High/Low Limit Temperatures
Set the judgment points of high and low temperatures for judging measured temperatures. This setting is made for each tag name. This judgment is conducted from the time when measurement is started.
Procedures
Example: When setting a high temperature of 90.0 C and a low temperature of 50.0 C
for Tag No. 2 (tag name: PAN02)
1. Press the SET key at the tag name screen of Tag No. 2 to display the judgment point of high limit temperature setting screen.
Flashing
Hi Limit
2. Press the key to flash the tens digits.
Setting range: Same as input range
Factory set value:
Hi Limit
Press the key to change the number to
3.
“9.”
0.0 C [32.0 F]
4. Press the SET key to store the new value. The display goes to the next screen (Judgment point of low limit temperature).
Flashing
Low Limit
Press the key to flash the tens digits.
5.
Setting range: Same as input range
Factory set value:
Low Limit
6. Press the key to change the number to “5.”
0.0 C [32.0 F]
Low Limit
Hi Limit
7. Press the SET key to store the new set value. The display returns to the tag name screen of Tag No. 2.
The setting is finished.
To return to the measurement screen after the setting is finished, press the ESC key three times.
14
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
IMR01X01-E5
Page 21
4. JUDGING MEASURED TEMPERATURES [SIMPLIFIED JUDGING FUNCTION]
4.3 Judging Measured Temperatures
The judgment criteria of a measured temperature is described by referring to the following example.
Measured object: Hot water Tag number: 2 (tag name: PAN02) Judgment criteria: Judgment point of high limit temperature 90.0 C
Judgment point of low limit temperature 50.0 C
Procedures
1. Turn on the power. The measurement screen appears on the LCD.
2. Press the key. Tag name is displayed in the message column.
First check whether or not a displayed screen is for the tag name of Tag No. 2. If not, press the or key to select the tag name of Tag No. 2.
11 / 11 10: 0 0
Measurement screen
TA G 01
Example: Tag name screen of Tag No. 1
PA N 02
The tag name displayed when the key is pressed is that with the tag number selected last, which continues to be displayed until the power is turned off.
For example, if the tag name of Tag No. 5 is selected last, the tag name of Tag No. 5 is displayed in the message column when selected to the tag name from the measured date.
When the power is turned off with the tag name of Tag No. 5 displayed.
Power ON
1 1 / 1 1 1 0 : 0 0
Measured date
T A G 0 5
Tag name
Continued on the next page.
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
15
Page 22
4. JUDGING MEASURED TEMPERATURES [SIMPLIFIED JUDGING FUNCTION]
3. Temperatures in a measured object are measured.
When exceeding the judgment point (90.0 C) of high limit temperature
High limit temp.
Measured temp. (95.0 C)
Judgment range
Low limit temp.
Judgment symbol
PA N 02
Tag name screen
When falling below the judgment point (50.0 C) of low limit temperature
High limit temp.
Judgment range
Measured temp. (45.0 C)
Low limit temp.
Judgment symbol
PA N 02
Tag name screen
When within the judgment points (50.0 to 90.0 C)
High limit temp.
Judgment range
Low limit temp.
Measured temp. (80.0 C)
Judgment symbol
PA N 02
Tag name screen
Thus, the judgment of measured temperatures has been finished.
When needing to record measured temperatures * in the memory, refer to 5. RECORDING MEASURED TEMPERATURES
[LOGGING FUNCTION] (P. 17).
* No result of judgment can be recorded in the memory.
16
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
IMR01X01-E5
Page 23
5. RECORDING MEASURED TEMPERATURES
y
A
[LOGGING FUNCTION]
This chapter describes the procedure for recording measured data in the memory using the logging function.
Both manual recording and automatic recording are available for recording measured data in the memory. The measured data is recorded on the function menu “Memory (Fn1)” display screen manually or automatically. The data is recorded in the memory in the order of arrival.
Screen Configurations of Memory (Fn1) Menu:
Maximum number of recording data: DP-700A: 99 logs DP-700B: 9999 logs The maximum number of recorded data
items may not be recorded depending on the occupied state in the memory.
Data hold: For the DP-700A, the recorded data may be lost if the battery service life has expired or the battery is replaced.
Display memory screen
Measurement
screen
ESC key
Press the MENU/POWER key
ESC
key
(P. 4)
Memory
SET key
key
TAG01
MENU/ POWER key
Tag Filter screen
TagFilter
MENU/ POWER key
key
Recorded tag name
ESC
MENU/ POWER key
(P. 21)
Defrag check screen
ke
Message column
transfer
key
SET
Log Jump screen
Log Jump
(P. 22)
MENU/ POWER key
C
key
key
key
(Only for the shaded screen)
1 1 / 1 1 1 0 : 0 0
Recorded date
Delete check screen
MENU/ POWER key
All Clear check screen
key
Delete?
(P. 23)
MENU/ POWER key
SET
All Clear?
key
(P. 24)
17
Defrag?
key
SET
fter defragmentation is finished, returns to the display memory screen.
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
(P. 25)
If the SET key is pressed after “Yes” is selected, returns to the display memory screen after defragmentation is finished.
Page 24
5. RECORDING MEASURED TEMPERATURES [LOGGING FUNCTION]
M
5.1 Manual Recording
Measured data * just when the REC key is pressed is recorded in the memory. Only when the REC key is pressed, is displayed.
* Recordable measured data:
Measured temperature, Measured date (excluding “Year”), Measured tag No. (tag name), Measured user No.
Recording
Press the REC key while measured. The measurement result displayed on the screen is recorded in the memory.
Example: When measured temperatures are recorded on the tag name screen of Tag No. 2
DP-700
Only when the REC key is pressed, is displayed.
REC
S
E
C
S
T
E
No judgment result is recorded.
PA N 02
Display on the measurement screen when the REC key is pressed.
Checking of recording
Press the MENU/POWER key while displaying the measurement screen to display the Memory (Fn1) screen.
Pressing the SET key displays the recorded measured-data memory screen to be able to check the recorded measured-data.
Example: When the memory screen in the above example is displayed
Display memory No.: Indicates in what memory number measured data is recorded.
Total number of recorded memory data
PA N 02
If two or more recorded data items need to be checked in order, press the or key.
It is possible to change the display (tag name or recorded date) in the message column. If necessary, press the or key.
18
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
IMR01X01-E5
Page 25
5. RECORDING MEASURED TEMPERATURES [LOGGING FUNCTION]
5.2 Automatic Recording
Measured data * is recorded in the memory at time intervals preset. For automatic measured-data recording, the time interval needs to be set. Set it according to the following procedure.
* Recordable measured data:
Measured temperature, Measured date (excluding “Year”), Measured tag No. (tag name), Measured user No.
Setting interval time
1. Press the MENU/POWER key three times at the measurement screen to display the function menu “Logging (Fn3)” screen.
Logging
2. Press the SET key. The display goes to the interval time setting screen.
Flashing
Intervals
Interval time:
Setting range:
OFF (Manual logging) 1 to 3600 seconds (Automatic logging)
Factory set value:
OFF (Manual logging)
3. Set the interval time by the key and
the shift key ( , ). (Example: Interval time 15 seconds)
Flashing
Intervals
4. Press the SET key to store the new set value. The display returns to the Logging (Fn3) screen. If the interval time is set, flashes (recording stop).
The setting finished.
To return to the measurement screen after the setting is finished, press the ESC key.
How to cancel the interval time setting?
Press the ESC key. The display returns to the Logging (Fn3) screen.
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
19
Page 26
5. RECORDING MEASURED TEMPERATURES [LOGGING FUNCTION]
M
Recording
Press the REC key at the measurement screen. Recording starts at time intervals preset. keeps lighting while being recorded.
Example: When automatic measured-temperature recording is started while the tag
name screen of Tag No. 2 is displayed.
DP-700
REC
S
E
C
S
T
No judgment result is
E
recorded.
PA N 02
Display on the measurement screen when the REC key is pressed.
Lighting (during recording)
If the power is turned off by pressing the MENU/POWER key or even if the power is turned off by the activation of the Auto power OFF function (P. 35), during recording ( being lit) data recording continues at time intervals preset.
Recording stop
Press the REC key. flashes to stop data recording.
How to return to manual recording?
First, stop data recording and then turn the time interval to “OFF.”
For the interval time setting, refer to Setting interval time (P. 19).
Checking of recording
To the checking of recording, refer to 5.1 Manual Recording (P. 18).
It is also possible to check recorded measured-data by opening the display memory screen while automatic recording is being made. As the measured data being checked is displayed on the display memory screen, no measured data being recorded is displayed. In this case, the total number of recorded memories is counted every time the data is recorded.
Total number of recorded memory data
PA N 02
Display memory screen
20
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 27
5. RECORDING MEASURED TEMPERATURES [LOGGING FUNCTION]
5.3 Retrieving Recorded Data
Measured data is recorded in the memory in the order of arrival. The following two types of retrieval are available for checking any tag name and recorded measured-data recorded in the memory. Use either one of them depending on your application.
Tag filter: Only measured data with any tag name is retrieved and then displayed. Log jump: Recorded measured-data is displayed by specifying the display memory number.
5.3.1 Retrieving only recorded data with specified tag name (Tag Filter)
As measured data is recorded in the order of arrival, there may be a case where no data is recorded for each tag name. It is possible to retrieve and then display only the measured data with the specified tag name bay setting this tag filter.
Procedures
Press the MENU/POWER and SET keys
1.
in this order while displaying the measurement screen to display the display memory screen.
Press the key to change the item to
4.
“ON.”
TA G 01
Press the or key while displaying
2.
the display memory screen to display the log name to be retrieved.
Log name to be
TA G 03
Press the MENU/POWER key to display
3.
the Tag Filter screen.
TagFilter
retrieved
Flashing
Flashing
TagFilter
5.
Press the SET key to store the new set value. Returns to the display memory screen with the log name to be retrieved.
TA G 03
Pressing the or key displays only
6.
the display memory screen with the log name retrieved. To return to the measurement screen, press the SET key twice.
How to release the log filter?
Change the log filter setting to “OFF” by pressing the key and then press the SET key.
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
21
Page 28
5. RECORDING MEASURED TEMPERATURES [LOGGING FUNCTION]
5.3.2 Retrieving recorded data with specified log registered No. (Log Jump)
It is possible for the DP-700A to record up to 99 logs and for the DP-700B, up to 9999 logs, respectively, For this reason, it is more difficult to select the display memory screen of data to be checked as the number of recording data items increases. It is possible to retrieve and then display any data with specified display memory number by setting this Log Jump.
Procedures
1.
Press the MENU/POWER key and SET key in this order while displaying the measurement screen to display the display memory screen.
TA G 01
Press the MENU/POWER key twice to
2.
display the Log Jump screen.
Current display memory No.
Flashing
Log Jump
4.
Press the SET key. Displays the screen with the display memory number specified. To return to the measurement screen, press the SET key twice.
TA G 05
How to cancel the log jump?
Press the ESC key. The display returns to the Logging (Fn3) screen.
Press the key, key, or the shift key
3.
( , ) to display the log name to be retrieved.
Display memory No.
Flashing
to be retrieved
Log Jump
22
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
IMR01X01-E5
Page 29
5. RECORDING MEASURED TEMPERATURES [LOGGING FUNCTION]
5.4 Deleting Recorded Data
The following two types of procedure are available for deleting measured data recorded in the memory.
Deleted?: Deletes only one recorded data item. All data deleted?: Deleted all recorded data items.
5.4.1 Deleting only one recorded data item (Delete?)
Can delete only one recorded data item selected. If the procedure for this deletion is executed, the display memory number and the total number of recorded data items decrease by one.
TA G 0 1 TA G 0 1 TA G 0 5TA G 0 1
To delete
Delete
Procedures
1. Press the MENU/POWER key and SET key in this order while displaying the measurement screen to display the display memory screen.
TA G 01
2. Press the or the display memory screen to
screen to be deleted.
key while displaying
display the
Display memory No. to be deleted
TA G 01
TA G 0 1TA G 0 1 TA G 0 5
3. Press the MENU/POWER key three times to display the delete check screen.
Delete?
Haw to cancel “Delete?”
Press the ESC key. The display returns to the display memory screen.
Pressing the SET key executes the
4.
procedure for this deletion. The display returns to the display memory screen. To return to the measurement screen, press the SET key twice.
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
23
Page 30
5. RECORDING MEASURED TEMPERATURES [LOGGING FUNCTION]
A
A
A
5.4.2 Deleting all recorded data items (All Clear?)
Can delete all recorded data items recorded.
Procedures
1. Press the MENU/POWER and SET keys in this order while displaying the measurement screen to display the display memory screen.
TA G 01
2. Press the MENU/POWER key four times to display the All Clear check screen.
Flashing
ll Cl ear ?
4. Pressing the SET key executes the procedure for this deletion (All Clear).
The following progress check screen is displayed during the execution of this deletion.
Do not press the key until this screen automatically disappears.
Returns to the measurement screen after the procedure for this deletion is finished.
No deletion is suspended even it the ESC key is pressed while the procedure for this deletion is in progress.
Deletion
3. Press the key to change the item to “YES.”
Flashing
ll Cl e ar!
This deletion is finished.
Returns to the measurement screen
progress ratio (%)
ll Cl ear ?
How to cancel “YES”?
Press the ESC key. The display returns to the display memory screen.
To return to the measurement screen, press the ESC key three times.
24
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 31
5. RECORDING MEASURED TEMPERATURES [LOGGING FUNCTION]
5.5 Defragmenting Recorded Data
If measured data recording and deleting operation are repeated, the recording memory capacity may become insufficient. However, it becomes possible to efficiently secure the recordable area by rearranging recorded data in the memory and defragmenting the memory area.
Procedures
Press the MENU/POWER and SET keys
1.
in this order while displaying the measurement screen to display the display memory screen.
TA G 01
2. Press the MENU/POWER key five times to display the defrag check screen.
Total number of recorded memory data
Memory usage
De f ra g ?
ratio (%)
3. Pressing the SET key executes the procedure for this defragmentation.
The following progress check screen is displayed while defragmentation is being executed. Time required for this defragmentation differs depending of the memory area being used.
Do not press the key until this screen automatically disappears.
Returns to the display memory screen after the procedure for this defragmentation is finished.
Defragmentation
De f ra g !
This defragmentation is finished.
progress ratio (%)
Returns to the
display
memory screen
How to cancel the procedure for this defragmentation?
Press the ESC key. The display returns to the display
To return to the measurement screen,
press the ESC key twice. memory screen. To return to the measurement screen, press the ESC key twice.
The defragmentation is possible
even during automatic recording. In
addition, automatic recording
continues after the defragmentation
is finished.
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
25
Page 32
6. PEAK HOLD/BOTTOM HOLD
A
The peak hold/bottom hold function is used to store (hold) the maximum (peak) and minimum (bottom) measured temperature. Each of these values is updated when the measured temperature becomes more (or less) than the value now being held.
However, if the following operation is performed, the value now being held is reset and as a result the measured temperature just when reset becomes the peak hold or bottom hold value.
When replaced the battery. When changed sensor type, display unit or decimal point position When reset (peak/bottom) by the key operation
• •
Power ON
uto-
matically
Measurement
To display the peak hold value/bottom hold value
1. Press the SET key at The display goes to
2. Press the SET key. The display goes to
Peak/Bottom (Fn4) screen.
Peak hold value screen.
Bottom hold value screen.
Memory (Fn1)
TAG Setup
(Fn2)
Logging (Fn3)
Peak/Bottom
(Fn4)
User Name
(Fn8)
Setup (Fn9)
Utility (Fn0)
Returns to
Memory (Fn1)
screen
ESC key
Returns to Measurement screen
Press the
:
MENU/POWER key or key
3. Press the SET key. The display returns to
Peak/Bottom (Fn4)
SET
Peak/Bottom
Measured temperature (PV)
Temperature display value
Temperature display
value
(Bottom hold value)
Power ON
Peak/bottom hold action description
Peak hold value
Peak val.
(Peak hold value)
Bottom hold
value
display
updating
Peak hold
value
display
updating
Peak hold value screen.
Bottom hold value
SET
SET
PV
Peak reset
(Refer to P. 27)
Bottom val.
ESCESC
Time
26
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
IMR01X01-E5
Page 33
6. PEAK HOLD/BOTTOM HOLD
A
Power ON
uto-
matically
Measurement
Memory (Fn1)
TAG Setup
(Fn2)
Logging (Fn3)
Peak/Bottom
(Fn4)
User Name
(Fn8)
Setup (Fn9)
Utility (Fn0)
Returns to
Memory (Fn1)
screen
To reset the peak hold value
1. Press the MENU/POWER key at
Peak hold value screen.
2. “Peak rst?” is displayed in the message column.
To execute reset: Press the SET key. To cancel reset: Press the ESC key.
3. The display returns to
Peak hold value
Peak val.
Peak hold value
Peak val.
Peak hold value screen.
Peak reset
MENU/
POWER
ESC
(To cancel
Peak rst?
reset)
SET
To execute reset
(
Peak hold value becomes measured temperature just when reset.
To reset the bottom hold value
)
ESC key
Returns to Measurement screen
Press the
:
MENU/POWER key or key
1. Press the MENU/POWER key at Bottom
hold value screen.
2. “Bottom rst?” is displayed in the message column.
To execute reset: Press the SET key. To cancel reset: Press the ESC key.
3. The display returns to Bottom
Bottom hold value
MENU/
POWER
Bottom val.
Bottom hold value
Bottom val.
ESC (To cancel reset)
Bottom hold value becomes measured temperature just when reset.
hold value screen.
Bottom reset
Bottom rst• •
SET
(
To execute reset
)
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
27
Page 34
7. REGISTERING USER NAMES
r
[ONLY FOR DP-700B]
This chapter describes the procedure for registering user name. Any user name can be registered (11 characters max.) for each user number. In addition, it is possible to set the user by selecting the user number. The user thus set is retained until the power is turned off. The registered user name is displayed when the power is turned on.
Screen configurations of User Name (Fn8) menu:
Measurement
screen
SET key
Press the MENU/POWER key five times.
ESC key
UserName
ESC key
User name screen
Key
User name *
Number of maximum registration:
99 names
User name setting screen with
selected user number
(Example: User No. 1)
(P. 4)
SET key
User numbe
Key
The user can be set by selecting the user number.
* Nothing is registered at the
time of factory shipment and therefore this space is vacant.
SET key (To execute setting)
ESC key
Cursor blinking
(P. 29)
MENU/POWER key
(To suspend operation)
Character entry (P. 29)
28
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
IMR01X01-E5
Page 35
7. REGISTERING USER NAMES [ONLY FOR DP-700B]
N
r
t
Character entry screen
Example: When the user name screen of user No. 1 is selected.
User name setting screen
Cursor blinking
Press the ESC key three times
Returns to Measurement screen
MENU/ POWER key
SET key ESC key
SET key ESC key
SET key ESC key
Character selecting screen
01234 5 6789
MENU/ POWER key
Delete check screen
Delete?
(P. 31)
Insert check screen
MENU/ POWER key
Insert?
(P. 31)
MENU/ POWER key
A list of characters including the currently selected character (cursor blinking) is displayed in the message column.
Select the list of characters by the or key and then select the desired character by the or key.
SET key: To execute selection ESC key: To suspended operation
Screen used to check whether or not the currently selected character (curso blinking) is deleted.
SET key: To execute deletion ESC key: To suspend operation
Screen used to check whether to inser space for entering a character in the position displayed with the currently selected character (cursor blinking).
SET key: To execute insertion ESC key: To suspend operation
Characters list
Numerals 0 1 2 3 4 5 6 7 8 9 Symbols @! # $ % & ’ ` Alphabets A B C D E F G H I
J K L M N O P Q R
+ - . ; = ^ _ ( ) [ ] { }
o spaces (no characters)
can be used.
S T U V WX Y Z a b c d e f g h i j k l m no p qr
Select the list of characters by the or key.
s t u v w x y z
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
29
Page 36
7. REGISTERING USER NAMES [ONLY FOR DP-700B]
Procedures
Example: Registers the user name “RKC” to user No. 1.
1. Press the SET key at User Name menu (Fn8) screen. The display goes to user name screen of user No. 1.
Press the SET key.
2.
The cursor is displayed
5.
6. Press the MENU/POWER key.
Press the SET key to store “R.” The screen returns to the user name screen of user No. 1 and the cursor moves to the right.
R
The list of characters is displayed in the message column.
Blinking
Press the MENU/POWER key.
3.
The list of characters is displayed in the message column.
012345678 9
4. Press the key twice. The list of characters including “R” is displayed.
JKLMNOPQR
7. Press the key to move the cursor to the position displayed with “K.”
JKLMNOPQR
Press the SET key to store “K.”
8.
The screen returns to the user name screen of user No. 1 and the cursor moves to the right.
JKLMNOPQR
RK
30
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
IMR01X01-E5
Page 37
7. REGISTERING USER NAMES [ONLY FOR DP-700B]
A
A
9. Press the MENU/POWER key. The list of characters is displayed in the message column.
JKLMNOPQR
10. Press the key. The list of characters including “C” is displayed.
BCDEFGHI
13. Press the SET key to store the “RKC.” The cursor disappears. The name setting is finished. To return to the measurement screen, press the ESC key twice.
RK C
How to change the character without displaying the list of characters?
1. Move the cursor to the position displayed
with the character to be revised by using the or key.
2. Display the character to be revised by
using the or key.
Press the key to move the cursor to the
11.
position displayed with “C.”
BCDEFGHI
Press the SET key to store “C.”
12.
The screen returns to the user name screen of user No. 1 and the cursor moves to the right.
RK C
How to delete one character?
1. Move the cursor to the position displayed
with the character to be deleted using the
or key.
2. Press the MENU/POWER key twice to
display the screen for checking the deletion.
3. Press the SET key to delete the character
where the cursor blinked.
How to insert space for entering a character in any position?
1. Move the cursor to the position where a
character needs to be entered using the or key. Press the MENU/POWER key three times
2.
to display the screen for checking the entry. Press the SET key to insert space for
3.
entering the character before the character position where the cursor blinks.
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
31
Page 38
8. INITIAL SETTING
A
This chapter introduces each function which can be set according to the customer applications.••
For items that are not necessary to be changed, use the same setting values as the factory set values.
Power ON
uto-
matically
Measurement
Press the MENU/POWER key or key six times
Setup (Fn9)
Display
language
Date format
Sensor type
Display unit
Decimal point
position
PV Bias
Digital filter
Auto power
OFF
To select the display language (English/Japanese)
Select the language (English/Japanese) which needs to be displayed on the screen.
1. Press the or key to select the “English or Japanese.”
Data range: English or Japanese Factory set value: English
English Japanese
2. Press the SET key to store the new set value.
To return to the measurement screen, press the ESC key twice.
To select the date format
Select the date format.
1. Press the or key to select the “Date format.”
Data range: month/day or day/month Factory set value: month/day
Operation lock
Press the
:
SET key
month/day day/month
2. Press the SET key to store the new set value.
To return to the measurement screen, press the ESC key twice.
32
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 39
8. INITIAL SETTING
A
Power ON
uto-
matically
Measurement
Press the MENU/POWER key or key six times
Setup (Fn9)
Display
language
Date format
Sensor type
Display unit
Decimal point
position
To select the sensor type
Select the sensor type used for measurement.
1. Press the or key to select the “Sensor type.”
Data range: TC-K: Thermocouple K TC-J: Thermocouple J Pt100: RTD Pt100 TC-T: Thermocouple T Factory set value: TC-K (Thermocouple K)
TC-K TC-J Pt100
TC-T
PV Bias
Digital filter
Auto power
OFF
Operation lock
Press the
:
SET key
2. Press the SET key to store the new set value.
To return to the measurement screen, press the ESC key twice.
To select the display unit ( C/ F)
Select the display unit of measured temperature.
1. Press the or key to select the “Display unit.”
Data range: C or F Factory set value: C
Unit Unit
2. Press the SET key to store the new set value.
To return to the measurement screen, press the ESC key twice.
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
33
Page 40
8. INITIAL SETTING
A
Power ON
uto-
matically
Measurement
Press the MENU/POWER key or key six times
Setup (Fn9)
Display
language
Date format
Sensor type
Display unit
Decimal point
position
PV Bias
Digital filter
To select the decimal point position
Select the decimal point position of input range.• •
1. Press the or key to select the “Decimal point position.”
Data range: 0.1: One decimal place 1: No decimal place Factory set value: 0.1 (One decimal place)
Flashing
DP DP
2. Press the SET key to store the new set value.
To return to the measurement screen, press the ESC key twice.
To correct the temperature display value (PV Bias)
Set the bias (correction value) added to the measured temperature for sensor correction. The PV bias is used to compensate the individual variations of the sensors or correct the difference between the measured temperatures of other instruments.
Auto power
OFF
Operation lock
Press the
:
SET key
1. Press the or key to move the flashing digit.
2. Press the or key to set the “Correction value.”
Data range: 99.9 to 99.9 C ( F) Factory set value: 0.0
Flashing
To return to the measurement screen,
PV Bias
press the ESC key twice.
3. Press the SET key to store the new set value.
Example:
If PV bias correction value of 1.0 C is added to the measured temperature.
When the measured temperature is
99.0 C, temperature display will become 100.0 C (= 99.0 C + 1.0 C).
Measured temperature
99.0 C
PV bias
1.0 C
Measurement screen
11/11 09:10
34
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
IMR01X01-E5
Page 41
8. INITIAL SETTING
A
Power ON
uto-
matically
Measurement
Press the MENU/POWER key or key six times
Setup (Fn9)
Display
language
Date format
Sensor type
Display unit
Decimal point
position
To lessen the fluctuation of a display value caused by noise (Digital Filter)
Set the time of the first-order lag filter eliminate noise against the measured input. The digital filter is used to lessen the fluctuation of a display value caused by noise.
1. Press the or key to move the flashing digit.
2. Press the or key to set the “filter time.”
Data range: OFF (Unused), 1 to 100 seconds Factory set value: OFF
Flashing
D Filter
3. Press the SET key to store the new set value.
To return to the measurement screen, press the ESC key twice.
PV Bias
Digital filter
Auto power
OFF
Operation lock
Press the
:
SET key
Saving battery consumption (Auto Power OFF)
Set the time until the power is automatically turned off after no key operation is performed.
1. Press the or key to move the flashing digit.
2. Press the or key to set the “Auto power OFF time.”
Data range: OFF (Unused), 1 to 60 minutes Factory set value: 3
Flashing
Auto Off
3. Press the SET key to store the new set value.
To return to the measurement screen, press the ESC key twice.
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
35
Page 42
8. INITIAL SETTING
A
Power ON
uto-
matically
Measurement
Press the MENU/POWER key or key six times
Setup (Fn9)
Display
language
Date format
Sensor type
Display unit
Decimal point
position
Locking/Unlocking setting operation (Operation Lock)
Select setting disabled (locked)/setting enabled (unlocked) for each item. Misoperation during usage can be prevented.
1. Press the or key to move the flashing digit.
2. Press the or key to set the “Disabled (locked)/setting enabled
(unlocked).”
(4)(3) (2)(1)
Factory set value: 0000 (unlocked)
Flashing
Lock
To return to the measurement screen, press the ESC key twice.
Setting items Data range
(1) Calendar timer (date and time)
setting
Setting items of Logging (Fn3),
(2)
0: Setting enabled
(unlocked).
1: Disabled (locked)
User Name (Fn8) and Setup (Fn9) menus.
PV Bias
Digital filter
Auto power
OFF
Operation lock
Press the
:
SET key
Setting items of TAG Setup
(3)
menu (Fn2)
Deletion of record data (1 or all
(4)
records data)
3. Press the SET key to store the new set value.
36
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
IMR01X01-E5
Page 43
9. MANAGING DATA BY PC [ONLY FOR DP-700B]
For the DP-700B, it is possible to manage recorded (logging) temperature data, tag and initial set data stored in the memory by a personal computer with the DP-700B connected to the USB port of the computer. This chapter describes procedures for connecting the DP-700B to a personal computer and also for managing stored data.
<System Requirements for a USB Connection>
Personal Computer: PC with USB port (USB 2.0/1.1) OS: Windows XP (32-bit and 64-bit versions),
Windows Vista (32-bit and 64-bit versions),
Windows 7 (32-bit and 64-bit versions), Windows 8.1 (32-bit and 64-bit versions), Windows 10 (32-bit and 64-bit versions)
9.1 Connecting PC to DP-700B
As the DP-700B is operated by the OS standard driver, it can be used only by connecting it to a PC.
1. The DP-700B is connected to the USB port of a PC using the USB cable attached. The DP-700B can be connected with the power turned on or off. *
2. The message “ ***USB*** ” is displayed on the bottom of the DP-700B screen to disable key operation. In addition, data logging stops when being logged.
If the DP-700B is used for the first time, the OS standard driver is automatically installed. Start the windows system with the DP-700B unconnected to log on as a user account with administrator privilege. First start the system, and then connect the PC to the DP-700B.
DP-700B
(Right view)
Remove the USB connector cap.
Connect to the USB port
Mini B type
USB cable
PC
Connect to the USB port
A type
USB cable length: 1 m
* The DP-700B is compatible with the USB bus power. The power is supplied from the USB port
on the PC side via the USB cable when connected with USB. It is possible to manage data by connecting USB even with the DP-700B power turned off.
3. Open “Devices and Printers” of Windows to newly add the “USB mass storage device” icon.
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
37
Page 44
9. MANAGING DATA BY PC [ONLY FOR DP-700B]
9.2 Storing Data
Open “Devices and Printers” of Windows to double click the “USB mass storage device” icon. Thus, three CSV files in the DP-700B memory area displayed.
LOG_ .CSV: Recorded (logging) temperature data TAG_ .CSV: Tag setting data SETUP_ .CSV: Initial setting data
The LOG_ .CSV, TAG_ .CSV and SETUP_ .CSV files are created when the DP-700B is connected to USB. As a file name of each file created, characters of identifying each entity (DP-700B) are entered in . Thus, two or more DP-700B files are identified.
If no temperature data is stored within the DP-700B memory, the LOG_ .CSV file is not displayed.
LOG_ .CSV (Recorded temperature data)
Recorded (logging) temperature data. Read only file. It can be opened and referred to by table calculation software such as Excel. In addition, it can be copied it stored at a location other than the DP-700B using a new name.
CAUTION
Deleting the LOG_ .CSV file also clears all recorded temperature data within the DP-700B memory. Take care not to delete necessary recorded temperature data.
If no temperature data is stored within the DP-700B memory, the LOG_ .CSV file is not displayed.
Contents of file: Memory No., Date, Time, Temperature, Tag No., User No.
< When opened by Excel >
38
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
IMR01X01-E5
Page 45
9. MANAGING DATA BY PC [ONLY FOR DP-700B]
TAG_ .CSV (Tag setting data)
DP-700B tag setting data. Writable/readable file. If it is opened by table calculation software such as Excel and then is edited and overwrite-stored, the setting can be reflected to the DP-700B.
Writes is inhibited if setting items in the TAG Setup menu (Fn2) is set to “Not be set (locked).” Refer to Locking/Unlocking setting operation (P. 36).
Contents of file:
Name Data range Factory set value
Tag No. 1 to 99
Tag Name 11 characters
Refer to List of usable characters
(P. 41)
Judgment point of
Same as temperature input range
high temperature
Judgment point of
Same as temperature input range
low temperature
< When opened by Excel >
TAG01 to 99
0.0 C or 32.0 F
0.0 C or 32.0 F
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
39
Page 46
9. MANAGING DATA BY PC [ONLY FOR DP-700B]
SETUP_ .CSV (Initial setting data)
DP-700B initializing setting data. Writable/readable file. If it is opened by table calculation software such as Excel and then is edited and overwrite-stored, the setting can be reflected to the DP-700B.
Writes is inhibited if setting items in the Logging (Fn3), User Name (Fn8) and Setup (Fn9) menus are set to “Not be set (locked).” Refer to Locking/Unlocking setting operation (P. 36).
Contents of file:
Name Command Data range Factory set value
User name
Display language
Date format
Sensor type
Display unit
Decimal point position
PV bias Digital filter Auto power OFF Interval time
< When opened by Excel >
User=
Language=
Date_format=
Input=
Unit=
TempDP=
TBias= TFilter= AutoOFF= Interval=
11 characters
Blank (No name)
Refer to List of usable
characters (P. 41)
0: Japanese
1 1: English 0: Month/Day
0 1: Day/Month 0: Thermocouple K
0 1: Thermocouple J 2: RTD (Pt100) 3: Thermocouple T
0: C
0 1: F 0: No decimal place
1 1: One decimal place
99.9 to 99.9 C ( F)
0.0 0 to 100 seconds (0: OFF) 0 0 to 60 minutes (0: OFF) 3 0 to 3600 seconds (0: OFF) 0
Command
User No., User name
Setting data
40
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
IMR01X01-E5
Page 47
List of usable characters
Numeral 01 2 34 5 67 8 9 Alphabets
Symbols
Spaces cannot be used.
AB C D E F G HI J K L MN O P QR ST U V W X Y Z ab c d e f g hi j k l m n o p q r s t u v wx y z
@ ! # $ % & ’ ` + - . ; =^ _ ( ) [ ] { }
9. MANAGING DATA BY PC [ONLY FOR DP-700B]
Storing data in hard disk of PC
Copy data to the hard disk of the PC using Windows Explorer.
9.3 Disconnecting DP-700B from PC
Disconnect the DP-700B from the PC after the PC power is turned off or according to the following steps.
Do not disconnect the DP-700B from the PC when the file is opened or while the file is being stored. Malfunction may result.
The DP-700B can be disconnected with the power turned on or off.
The disconnecting procedure is described in “Windows 7.” Icons and windows displayed differ depending on OS used. For details, refer to the manual for “Windows.”
1. Check that the file is not opened or it is being stored. Close when the file is opened.
2. Click “Show hidden icons” in the taskbar. The notification area is displayed.
3. Click “Safely Remove Hardware” in the notification area.
4. Click “Eject USB mass storage devices” in the displayed menu.
The displayed contents depend on the number of USB devices connected to your computer. Carefully click the icon not to select a wrong device.
5. The “Safely Remove Hardware” icon moves to the notification area. When you see the message “Safe to remove hardware - The USB mass storage device can now be safely removed from the computer,” remove the device from your computer.
6. Disconnect the DP-700B from the PC.
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
41
Page 48
10. APPENDIX
10.1 Replace the Battery
If (Battery service life is going to expire) is displayed, immediately replace the battery.
Replace battery: Type LR6 (based on IEC and JIS) alkaline battery, One battery Battery life: Can be operated for 400 hours continuously (Around the ambient
temperature 23 C). However, the battery service life varies depending on the operating condition and environment. Therefore, the above service life period is only for reference.
If the battery is replaced, the instrument is reset-started. Therefore, there are some data items to be initialized. In addition, for the DP-700A all recorded data items are deleted. The state of each data when the battery is replaced (when reset-started) is shown in the following.
Monitor items
Name DP-700A DP-700B
Peek hold Reset
(Peak hold value becomes measured temperature just when reset.)
Bottom hold Reset
(Bottom hold value becomes measured temperature just when reset.)
Recorded (logging) data All recorded data clear Hold
Setting items
Name DP-700A DP-700B
Tag Filter Initialize (OFF) Tag Name Hold Judgment point of high
temperature Judgment point of low
temperature Interval time Hold User No. selection Initialize (User No. 1)
User name Hold
Display language Hold Date format Hold Sensor type Hold Display unit Hold Decimal point position Hold PV bias Hold Digital filter Hold Auto power OFF Hold Operation lock Calendar timer setting: Can be set (Unlock)
Other setting items: Hold Year Initialize (0) Date Initialize (01/01) Time Initialize (00:00)
Hold
Hold
42
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 49
10. APPENDIX
Procedures
1. Press the MENU/POWER key for more than two seconds to turn off the DP-700 power.
2. Loosen the battery cover.
Turn the cover with a coin.
Battery cover
3. Remove the battery cover.
4. Remove the old battery. Load a new battery without mistaking its polarity.
Old battery
battery
New Alkaline battery (LR6)
5. Then, firmly tighten the battery cover.
Insufficient tightening cannot assure the effect of water resistance (DP-700A: Equal to IP67, DP-700B: Equal to IP54).
After the battery is loaded, the instrument is ready to be reset-started. The function menu “Utility (Fn0)” for setting the dominical year is displayed on the LCD screen. If no
dominical year is set, no calendar function can be used. Always set the year if the calendar function is used.
Year
Year setting screen
Pressing the ESC key twice can display the measurement screen.
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
43
Page 50
10. APPENDIX
10.2 Cleaning
Turn off the power supply before cleaning the instrument.
Moisten a soft cloth with a diluted neutral detergent, completely wring the cloth, then wipe the instrument with cloth. Do not use a volatile solvent such as paint thinner to clean the instrument. Deformation or discoloration will occur.
The DP-700A can be washed in flowing water as it has water and dust resistant characteristics equal to IP67. However, the DP-700B (Water/dust resistance equal to IP54) cannot be washed in flowing water.
10.3 Troubleshooting
This section describes probable causes and treatment procedures if any abnormality occurs in the instrument. For any inquiries, please contact RKC sales office or the agent, to confirm the specifications of the product.
Prior to replacing the sensor, always turn off the power.
10.3.1 Error displays
Displays and description of errors are described in the following table:
Display Description Solution
Service life is going to expire. Replace the new battery.
Refer to 10.1 Replace the
Service life has expired.
Burnout
The sensor is broken, shorted or imperfectly connected.
Over-scale
Measured temperature is above the 5 % of measuring range high limit value or display range limit
[Flashing]
high (9999).
Battery (P. 42).
Check sensor and sensor connection.
Check input type, input range, sensor and sensor connection.
Underscale
Measured temperature is below the 5 % of measuring range low limit value or display range limit
[Flashing]
44
Find Quality Products Online at: sales@GlobalTestSupply.com
low ( 1999).
Continued on the next page.
IMR01X01-E5
www.GlobalTestSupply.com
Page 51
Continued from the previous page.
Display Description Solution
10. APPENDIX
(Err 1)
(Err 2)
(Err 4)
(Err 64)
Adjusted data error
Adjusted data range is abnormal.
EEPROM error
Response signal from EEPROM is abnormal. Data write failure
A/D conversion error
Response signal from A/D converter is abnormal. A/D conversion count value is out of the specified range.
Stack overflow
Stack point overflows
First remove the battery, and then load it again after a lapse of a few seconds.
If not normally operated with the battery re-loaded, please contact RKC sales office or the agent.
Watchdog timer error
The part of an internal task stops the action.
(Err 128)
If two or more self-diagnostic errors occur (Err 1, Err 2, Err 4, Err 64 and Err 128), the value obtained by adding the error number to the Err display is displayed.
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
45
Page 52
10. APPENDIX
10.3.2 Problem and solution
Problem, possible cause and solution of trouble described in the following table:
Problem Possible cause Solution
No power is turned on. (No display appears)
Measured temperature display is abnormal
No battery is loaded. Load a battery.
Refer to 2.1 Loading Battery
(P. 5).
The battery with less service life is used.
Replace the new battery. Refer to 10.1 Replace the Battery
(P. 42).
The battery with less service life is used.
Replace the new battery. Refer to 10.1 Replace the Battery
(P. 42).
Noise source is present near the instrument
Separate the noise source from the instrument.
Set the digital filter. Refer to To lessen the fluctuation
of a display value caused by noise (P. 35).
Proper sensor is not
Use the specified sensor.
used
Sensor insertion depth is insufficient
Check whether sensor is inserted loosely. If yes, fully insert the sensor.
Sensor insertion
position is not
Insert the sensor at the specified location.
appropriate
Measured temperature display value differs from the actual value
PV bias is set Set the PV bias to “OFF.”
Refer to To correct the
temperature display value (P. 34).
However, this is limited only to when the PV bias setting can be changed.
The display does not change No key operation is accepted
CPU overrun First remove the battery, and then
load it again after a lapse of a few seconds.
If not normally operated with the battery re-loaded, please contact RKC sales office or the agent.
46
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
IMR01X01-E5
Page 53
10.4 Specifications
Input
Input type: Universal input
Thermocouple: K, J, T (JIS-C1602-1995) RTD: Pt100 (JIS-C1604-1997)
Measuring range: Thermocouple K: 200 to 1372 C (Resolution: 1 C)
328 to 2501 F (Resolution: 1 F)
199.9 to 999.9 C (Resolution: 0.1 C)
199.9 to 999.9 F (Resolution: 0.1 F)
Thermocouple J: 200 to 1200 C (Resolution: 1 C)
328 to 2192 F (Resolution: 1 F)
199.9 to 999.9 C (Resolution: 0.1 C)
199.9 to 999.9 F (Resolution: 0.1 F)
Thermocouple T: 50 to 400 C ( 58 to 752 F)
50.0 to 400.0 C ( 58.0 to 752.0 F)
RTD Pt100: 200 to 850 C (Resolution: 1 C)
328 to 1562 F (Resolution: 1 F)
199.9 to 850.0 C (Resolution: 0.1 C)
199.9 to 999.9 F (Resolution: 0.1 F)
Sampling cycle: 0.5 seconds PV bias: 99.9 to 99.9 C ( F) Digital filter: First order lag digital filter:
Time constant: 0 to 100 seconds (0: OFF)
Moving average: 4 times (fixed)
10. APPENDIX
Performance
Display accuracy (At the ambient temperature 23 C ± 2 C):
(0.1 % of display value 1 digit) or 0.3 C (0.6 F)
Less than the value whichever is greater
Cold-junction temperature compensation error:
0.5 C (Within 5 to 40 C)
1.0 C (Within 20 to 5 C and 40 to 50 C)
Ambient temperature influence:
0.01% / C of span (Within 5 to 40 C)
0.02% / C of span (Within 20 to 5 C and 40 to 50 C)
Display
Reflective FSTN LCD
Measured temperature: 4-digit Display memory No.: 4-digit Total number of recorded data: 4-digit Message: 68 8 dots (11 characters) Recorded (logging) execution: Lit during execution, flashing during stop for automatic recording. Burnout: Lit when the sensor is disconnected, shorted or imperfectly contacted. Remaining battery service life: The remaining battery service life is indicated in three steps. Temperature unit: 12 8 dots ( C/ F)
IMR01X01-E5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
47
Page 54
10. APPENDIX
Data recording (logging)
Recording method: Manual logging mode: Can be manually recorded in the memory.
Automatic logging mode: Can be automatically recorded (in a set interval) in the memory.
Interval time: 0 (Manual log mode), 1 to 3600 seconds Recording items: Measured temperature, Tag name, Date, Time Recording capacity: DP-700A: Maximum 99 logs
DP-700B: Maximum 9999 logs
Memory backup: DP-700A: Backed up by SRAM
The recorded data may be lost if the battery service life has expired or the battery is replaced. DP-700B: Backed up by non-volatile memory (EEPROM) Number of writing: Approx. 100,000 times Data retaining period: Approx. 10 years Depending on storage and operating conditions
Simplified Judging Function
The judgment points of high and low temperatures are set. When within the judgment points, “ ” is displayed while when out of the judgment point, “ ” is displayed. is displayed, if out of the high judgment point, while , if out of the low judgment point.
Tag number function
Number of maximum registration:
DP-700A: 5 tags DP-700B: 99 tags
Registration items: Tag name, judgment points of high and low temperatures Tag name: 11 characters (Numeral, Alphabets, Symbols) Memory backup: Backed up by non-volatile memory (EEPROM)
Number of writing: Approx. 100,000 times Data retaining period: Approx. 10 years
User name function (Only for DP-700B)
Number of maximum registration:
99 names
Number of character: 11 characters (Numerals, Alphabets, Symbols) Memory backup: Same as Memory backup of Data recording (logging)
USB function (Only for DP-700B)
Interface: Based on USB (Universal Serial Bus) Ver2.0 Communication speed (Theory value):
Full Speed compliant (12 Mbps)
Connector type: USB series Mini-B Connection: The DP-700B is connected to the USB port of a PC using the USB
cable attached.
Power supply: Bus power (The power is supplied from the USB port on the PC side.) Device class: Mass Storage Class (Bulk-Only Transport) Compatible OS: Windows XP (32-bit and 64-bit versions),
Windows Vista (32-bit and 64-bit versions), Windows 7 (32-bit and 64-bit versions), Windows 8.1 (32-bit and 64-bit versions), Windows 10 (32-bit and 64-bit versions)
48
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
IMR01X01-E5
Page 55
10. APPENDIX
Other functions
Peak and bottom hold, Calendar timer, Operation lock, Self-diagnostic
Power
Battery: Type LR6 (based on IEC and JIS) alkaline battery, One battery Battery life: Approx. 400 hours (Continuous operating time, Around the ambient
temperature 23 C). However, the battery service life varies depending on the operating condition and environment. Therefore, the above service life period is only for reference.
Power consumption: 10 mW (Mean value while operated)
Standards
Waterproof and dustproof: DP-700A: IP67
DP-700B: IP54
CE marking: EMC: EN61326-1, Annex A RCM: EN55011
General specifications
Memory backup: Backed up by non-volatile memory (EEPROM) Number of writing: Approx. 100,000 times Data retaining period: Approx. 10 years Depending on storage and operating conditions
Backup data: Interval time, Display language,
Sensor type, Display unit,
Decimal point position, PV bias, Digital filter, Auto power OFF, Operation lock, Date format
Allowable ambient temperature:
20 to 50 C
However, in an environment of less than 0 C, the LCD response speed remarkably decreases.
Allowable ambient humidity: 5 to 95 %RH
Absolute humidity: MAX.W.C 29.3 g/m3 dry air at 101.3 kPa
Dimensions: 57 (W) 152 (H) 46 (D) mm Weight: Approx. 150 g (Including battery)