Review the following safety precautions carefully before operate the instrument to
avoid any personal injuries or damages to the instrument and any products
connected to it.
To avoid potential hazards use the instrument as specified by this user’s guide only.
The instrument should be serviced by qualified personnel only.
To Avoid Fire or Personal Injury.
Use Proper Power Cord. Use the power cord designed for the instrument and
authorized in your country only.
Ground The Instrument. The oscilloscope is grounded through the grounding
conductor of the power cord. To avoid electric shock the instrument grounding
conductor(s) must be grounded properly, before making connections to the input or
output terminals of the instrument.
Connect The Probe. The probes’ ground terminals are at the same voltage level of
the instrument ground. Do not connect the ground terminals to a high voltage.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings
and marks on the instrument. Follow the user’s guide for further ratings information
before making connections to the instrument.
Do Not Operate Without Covers. Do not operate the instrument with covers or
panels removed.
Use Proper Fuse. Use the fuse of the type, voltage and current ratings as specified
for the instrument.
Avoid Circuit or Wire Exposure. Do not touch exposed connections and
components when power is on.
Do Not Operate With Suspected Failures. If suspected damage occurs with the
instrument, have it inspected by qualified service personnel before further
operations.
Provide Proper Ventilation. Refer to the installation instructions for proper
Do not operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
The disturbance test of all the models meet the limit values of A in the
standard of EN 61326: 1997+A1+A2+A3, but can’t meet the limit values
of B.
The VS5000 series Digital Oscilloscope is intended to be used for measurements in
Measurement Category I.
Measurement Category Definitions
Measurement Category I is for measurements performed on circuits not directly
connected to MAINS. Examples are measurements on circuits not derived from MAINS,
and specially protected (internal) MAINS derived circuits. In the latter case, transient
stresses are variable; for that reason, the transient withstand capability of the equipment
is made known to the user.
WARNING
IEC Measurement Category I. The HI and LO input terminals may be connected to
circuit terminal in IEC Category I installations for voltages up to 300 VAC. To avoid
the danger of electric shock, do not connect the inputs to circuit’s voltages above 300
VAC.
Transient overvoltage is also present on circuits that are isolated from mains. The
VS5000 series Digital Oscilloscopes is designed to safely withstand occasional
transient overvoltage up to 500 Vpk. Do not use this equipment to measure circuits
where transient overvoltage could exceed this level.
RIGOLVS5000 Series Digital Oscilloscopes offer good waveform view and
measurements in a compact, lightweight package. VS5000 series are ideal for
production test, field service, research and design and all of the applications
involving analog circuits test and troubleshooting, as well as education and training.
z Adjustable waveform intensity, more effective waveform view
z One-touch automatic setup for ease of use (AUTO)
z 20 Automatic measurements
z Automatic cursor tracking measurements
z Waveform recorder, record and replay dynamic waveforms
z User selectable fast offset calibration
z Built-in FFT function, Frequency Counter
z Digital filters, includes LPF, HPF, BPF, BRF
z Add, Subtract and Multiply Mathematic Functions
z Advanced trigger types include: Edge, Video, Pulse width, Slope, Alternative,
After receiving a new VS5000 series oscilloscope, please inspect the instrument as
following steps:
1. Inspect the shipping container for damage.
Keep the damaged shipping container or cushioning material until the contents of
the shipment have been checked for completeness and the instrument has been
checked mechanically and electrically.
2. Check the accessories.
Accessories supplied with the instrument are listed in "Accessories" in this guide.
If the contents are incomplete or damaged,please notify the RIGOL Sales
Representative.
3. Inspect the instrument.
In case there is any mechanical damage or defect, or the instrument does not
operate properly or fails performance tests, notify the RIGOL Sales
Representative.
If the shipping container is damaged, or the cushioning materials show signs of
stress, notify the carrier as well as the RIGOL sales office. Keep the shipping
materials for the carrier’s inspection.
RIGOL offices will arrange repair or replacement at option of RIGOL without
The first thing to do with a new oscilloscope is to become acquainted with its front
panel. This chapter will help to be familiar with the layout of the connectors,
graphical representations, interfaces, and how to use them. Read the chapter
carefully before further operations.
Figure 1-1
Front panel of VS5000 Series Oscilloscope
Figure 1-2
Rear panel of VS5000 Series Oscilloscope
VS5000 series oscilloscopes provide an easy-to-use user interface. In the main menu,
access the function options from the menu and the drop-down menu to operate
VS5000 series oscilloscope. There’re shortcuts for some common functions.
You are suggested to install the software of VS5000 and connect the equipment
when you get a new RIGOL VS5000 Oscilloscope. For more details please refer to
the following text:
Software Setting
Please insert and drive the disk accessory of the RIGOL VS5000 oscilloscope. Then,
double click the file named”
according to the hints.
Note: If you have already installed VS5000 software before, please uninstall the old
Make sure the power supply was connected correctly before using the instrument.
First, please insert power cord into the faucet of its adapter. Then connect the
adapter with oscilloscope and power on.
USB Connection
When the software setting and power supply connecting are finished, you can
connect the instrument to PC by USB cable. After connecting, a dialog about setup
will pop up to clew users to finish USB connection.
LAN Connection
A LAN port is offered at the rear panel in order to realize Remote Control which acts
a same role with USB. Besides, two steps are required to connect LAN: network
setting and connection setting.
1.Network Setting: There are two ways to config Network Protocol Property:
Obtain an IP automatically、obtain an IP manually.
(1) Obtain an IP automatically:
Please set network according to the following steps:
zConnecting: On condition that the USB cable was being connected, you can
configure l Property of Network protocol. First, connect oscilloscope to PC via
USB, and then, connect oscilloscope to LAN.
zOpen control->Network Setting, then it will pop up a dialog about configuration.
Click “obtain an IP automatically” option and press Apply to finish. .
zRestart oscilloscope and software, you can view the Property of Network
protocol in Configuration dialog box if USB communication is active.
(2) Obtain an IP manually:
zConnecting: On condition that the USB cable has being connected, you could
configure l Property of Network protocol. First, connect oscilloscope to PC via
USB, and then, connect oscilloscope to LAN.
zOpen control->Network Setting, it will pop up a dialog about configuration. Click
“obtain an IP automatically” option again to cancel Automatic setting. Then
Setting IP address、IP mask、Gateway by manual . Finally, click Apply to
configure Network protocol Property.
zRestart oscilloscope and software, you can view the Property of Network
protocol in Configuration dialog box if USB communication is active.
NOTES:
zMake sure that the oscilloscope has been connected with network through
Router in” Obtain an IP automatically”.
z Make sure the validity of IP address in ‘Obtain an IP manually’.
z Restart the oscilloscope and PC after finishing configuration.
z The function of Configure Network is invalidation in Network Communication.
2. Connection Setting:
z Connect software and oscilloscope in Network Communications.
z Select Network Communication. Open Control->Connection menu, it will pop up
a dialog, click check option LAN and select Network Communication, then set
IP address and port value to be the oscilloscope’s. Finally, click Apply to finish
connecting.
zIf communication has been found, System will clew that network connection is
successful. If failed, please check hardware connection and settings of Network
Protocol property.
NOTES:
zPlease pay attention to check the settings of Safety level of firewall when using
Network Communication.
zIn network communication, you are suggested to use “Resume default setting”
carefulness as Default connection is USB connection. Meanwhile, related result
may be brought if you cancel all the settings about Network Protocol Property.
Perform once quick check to verify the running conditions of the instrument.
1. Prepare for using the instrument.
Use the power cords designed for the oscilloscope only. Connect to a power
source that delivers 100 to 240 VAC, 45Hz to 440Hz. Power up the system,
initiate the program. Click the Storage button, select Load Default in the menu
box.
Status indicators
Channel 1, 2 and Ext Trigger
Probe compensator
Figure 1-5
Instruction of the Front Panel
2. Input a signal to a channel of the oscilloscope
VS5000 Series are equipped with dual channels plus external trigger and 16
channels logic analyzer (only for VS5000D models).
Please input signal in the light of following steps:
1) Set the attenuation switch on the probe as 10X (Figure 1-6) and connect the
probe on the oscilloscope with CH1. Aim the slot in the probe connector at the
faucet on BNC of CH1 and insert, then, turn right to lock the probe. Finally,
attach the tip of probe and ground nip to the Connector of Probe compensator.
Perform this function to match the characteristics of the probe and the channel input.
This should be performed whenever attaching a probe to any input channel at the
first time.
1. From the Probe menu, select attenuation to 1:10. Set the switch to X10 on the
probe and connect it to CH 1 of the oscilloscope. When using the probe hook-tip,
insert the tip onto the probe firmly to ensure a proper connection.
Attach the probe tip to the Probe Compensator and the reference lead to the
ground connector, select CH 1, and then press the
2. Check the shape of the displayed waveform.
Over compensated Correctly Compensated Under Compensated
button.
Figure 1-9
Instruction Waveform of Probe Compensation
3. If necessary, use a non-metallic tool to adjust the trimmer capacitor of the probe
for the flattest square wave being displayed on the oscilloscope.
4. Repeat if necessary.
WARNNING: To avoid electric shock while using the probe, be sure the
!
perfection of the insulated cable, and do not touch the metallic portions of
the probe head while it is connected with a voltage source.
1. Switch off power supply of the Device under Test if necessary to avoid short
circuit. Since no voltage is applied to the leads at this step, the oscilloscope may
remain on.
2. Connect one end of the flex cable FC1868 to the Logic Analyzer Input; connect
the other end to Logic Head LH1116. It is unnecessary to switch off power supply
of the oscilloscope while connecting the cable.
1-12
Figure 1-10
Connect the Flex Cable
CAUTION:Use only FC1868, LH1116, TC1100 and LC1150 made by
The VS5000 series oscilloscope has an automated function to display the input signal
in best fit. This feature requires a 50 Hz or higher and a duty cycle greater than 1%
input signal.
Press the
HORIZINTAL and TRIGGER controls to display the input signal. Adjust the controls
manually to get the best results if necessary.
Connect a signal to the Channel 1 (CH1) input:
1. Connect a signal to the oscilloscope as described above.
button, the oscilloscope automatically sets up VERTICAL,
2. Click the
button.
The oscilloscope may change the current settings to display the signal, i.e. the
vertical and horizontal scale, the trigger coupling, type, position, slope, level, and
mode.
ertical menu, which leads to the setting of the vertical
Figure 1-14
Vertical Menu
1. Use the Analog function in the Vertical menu to adjust the functions of the
analog channels. A voltage value is displayed for a short time indicating how far
the ground reference is located from the center of the screen. Use the left
mouse button to dragging the CH1 symbol, and/or use the Scale and Offset
options to set the scale and offset precisely.
Measurement hints
If the channel is DC coupled, measuring the DC component of the signal by
simply noting its distance from the ground symbol.
If the channel is AC coupled, the DC component of the signal is blocked,
allowing using greater sensitivity to display the AC component of the signal.
Double click the trigger level sign to set the trigger level to 0 as a shortcut key.
3. Double click the trigger point, to set the horizontal delayed time to 0 as a
shortcut key.
4. Click the button to start an acquisition regardless of an adequate trigger
signal, usually used in “Normal” or ”Single” trigger mode. If the acquisition is
already stopped, click this button to start the acquisition again.
Key point:
Holdoff: A time interval before the oscilloscope response to next trigger signal.
During this Holdoff period, the trigger system becomes “blind” to trigger
signals. This function helps to view complex signals such as an AM waveform.
In order to set the Holdoff time, set the Holdoff Time option in the Trigger
The VERTICAL, HORIZONTAL, and TRIGGER control systems and know how to
determine the system setup from the status bar of a VS5000 Series oscilloscope were
introduced in the previous chapter.
This chapter will go through all groups of buttons, knobs, and menus, expanding the
knowledge of operations by hints in this guide.
It is recommended to perform all of the following exercises to get the most of the
powerful measurement capabilities of the oscilloscope.
Understand the vertical system (
, ,
[1]
, Vertical )
Understand the horizontal system ( Horizontal )
Understand the trigger system ( Trigger )
To set up the sampling system ( Acquire )
To set up the display system ( Display )
To l oa d、save and print ( Storage )
To set up utility ( Utility )
To measure the waveform ( Analyzer )
To use the run control buttons ( , and )
[1] NOTE: This function is for VS5000D model only.
In order to get more information about the various measurements and operations of
Select the attenuation factor for the probe. The attenuation factor changes the
vertical scale of the oscilloscope so that the measurement results reflect the actual
voltage levels at the probe tip.
To change (or check) the probe attenuation setting, press the CH1 or CH2 button
(according to which channel are in using), toggle the Probe soft button to match the
attenuation factor of the probe.
This setting remains in effect until changed again.
Figure 2-6 gives an example for using a 1:1000 probe and its attenuation factor.
Invert turns the displayed waveform 180 degrees, with respect to the ground level.
When the oscilloscope is triggered on the inverted signal, the trigger is also inverted.
Figure 2-8 and 2-9 show the changes before after inversion.
It allows to turn single channel or group channels ON or OFF, to set the Wave Size, to
change display location of digital channel on screen and to select Threshold Type.
Left click the D
igital button in the Vertical menu, or right click the button to
show the Logic Analyzer window.
Logic Analyzer Interface
Table 2-3 Logic Analyzer Settings
FunctionSettingDescription
Figure 2-10
2-10
D7-D0 Set up channel group D7-D0
D15-D8 Set up channel group D15-D8
All on ON/OFF
Turn all (D7-D0 or D15-D8) digital
channels ON/OFF
Wave Size Big/Small Select the waveform display size.
Reset Reset the waveform display position.
Threshold
Typ e
TTL
CMOS
ECL
USER
User’s Guide for VS5000 Series
Select mode of all digital channels.
The threshold voltage can be set by user
in user-defined style.
The mathematic functions include “add”, “subtract”, “multiply” and “FFT” for Channel
1 and Channel 2. The mathematic result can be measured by the grid and the cursor.
The Math function may turn on four channels, each channel could be operated
independently (Except Delayed display).
Left click the Math button in the Vertical menu. Select Filter function. In Filter Type,
select the type of filter: Low Pass, High Pass, Band Pass or Band Reject. Set
high and low limit of frequency.
Turn o f f
Digital Filter
Turn o n
Digital Filter
Figure 2-14
Digital Filter Interface
Table 2-5 Digital Filter Function
FunctionSettingDescription
Source
Digital Filter
CH1
CH2
Off
On
Low Pass
Filter Type
High Pass
Band Pass
Band Reject
Upper limit
Lower limit
Define CH1 or CH2 as source
Turn off the digital filter
Turn on the digital filter
Setup as LPF (Low Pass Filter)
Setup as HPF(High Pass Filter)
Setup as BPF (Band Pass Filter)
2. Addition, Subtraction and Multiplication Function
In Math function, use the addition, subtraction and multiplication function to operate
and analyze the waveform.
Figure 2-15
The Addition, Subtraction and Multiplication Function
Enter the Math window. Select the math function in the Operator menu. Select
Source A and B. Then adjust the vertical Scale and Offset to view the math channel
The FFT (Fast Fourier Transform) process converts a time-domain signal into its
frequency components mathematically. FFT waveforms are useful in the following
applications:
z Measuring harmonic content and distortion in systems
z Characterizing noise in DC power supplies
z Analyzing vibration
Table 2-6 FFT Function
FunctionSettingDescription
FFT
OFF
ON
A+B
Operation
A-B
A×B
FFT
Source
CH1
CH2
Rectangle
Window
Hanning
Hamming
Blackman
Split Window
Scale
Split Window
Full Window
Vrms
dBVrms
Tur n off the FFT.
Open the FFT.
Add source A and source B
Subtract source B from source A
Multiply source A by source B
Fast Fourier Transform
Digital filter
Define CH1 or CH2 as FFT source
Select window for FFT
Display split waveform
Display full waveform
Set “Vrms ” as vertical unit
Set “dBVrms ” as vertical unit
VS5000 series digital oscilloscopes provide four FFT window functions. Each window
is a trade-off between frequency resolution and amplitude accuracy. It depends on
the desired measurements and source signals characteristics to determine the
window to use. Use the following guidelines to select the best window.
Table 2-7 FFT Windows description
WindowFeatures Best for measuring
Transients or bursts, the signal
Best frequency resolution
Rectangle
and worst magnitude
resolution. This is essentially
the same as no window.
Hanning
Hamming
Better frequency, poorer
magnitude accuracy than
Rectangular.
Hamming has slightly
better frequency resolution
than Hanning.
Blackman Best magnitude, worst
frequency resolution.
levels before and after the event
are nearly equal.
Equal-amplitude sine waves with
fixed frequencies.
Broadband random noise with a
relatively slow varying spectrum.
Sine, periodic, and narrow-band
random noise.
Transients or bursts where the
signal levels before and after the
events are significantly different.
Single frequency waveforms, to
find higher order harmonics.
Key points:
FFT Resolution
With a fixed FFT points, the lower sampling rate results in better resolution.
Nyquist Frequency
The highest frequency that any real-time digital oscilloscope can acquire
without aliasing. It’s normally half of the sample rate. This frequency is called
the Nyquist frequency. Frequency above the Nyquist frequency will be under
Reference Waveforms are saved waveforms to be selected for display. The reference
function will be available after saving the selected waveform to non-volatile memory.
Click the Reference button in the Vertical menu, go to the Reference window.
Figure 2-16
Reference Interface
Table 2-8 Reference Function
功能菜单设定说明
Display
OFF
ON
CH1
CH2
Source
MATH1
MATH2
MATH3
MATH4
Close waveform reference
Open waveform reference
Select CH1 as Reference channel
Select CH2 as Reference channel
Select Math1 as Reference channel
Select Math2 as Reference channel
Select Math3 as Reference channel
Select Math4 as Reference channel
The CH1, CH2, Ext Trigger and LA (for VS5000D only) channels on VS5000 series are
input channels. All functionalities applied will be based on operating the instrument
with channels. So MATH and REF can be regarded as relatively isolated channels.
To turn on/off any one of the channels, press the corresponding button on the front
panel. The button turns coloured indicates the channel is currently active. Press the
button again or chose Off in Channel/Logic Analyzer window to turn the channel
off.
Table 2-11 Turn On/Off channels
Channel ModeSettingsStatus Indicator
Channel 1 (CH1)
Channel 2 (CH2)
ON
OFF
ON
OFF
ON
MATH
OFF
[1]
ON
OFF
ON
OFF
Reference
Logic Analyzer
NOTE:
The status indicators (CH1, CH2 and LA) of the VS5000 series in the above table are
displayed on the lower-left of the main window.
CH1 (yellow background)
CH1 (gray background)
CH2 (blue background)
CH2 (gray background)
M1 M2 M3 M4 (to distinguish
between the four colors)
No indicator
Ref 1/2/3/4 (white background)
No indicator
LA (green background)
LA (gray background)
The four MATH function channels are distinguished between the four colors.
The oscilloscope shows the time per division in the scale readout. Since all active
waveforms use the same time base, the oscilloscope only displays one value for all
the active channels, except when using Delayed Scan, or Alternative Trigger.
The horizontal control changes the horizontal scale and position of waveforms. The
horizontal center of the screen is the time reference for waveforms. Changing the
horizontal scale causes the waveform to expand or contract with respect to the
screen center.
Horizontal position changes the displayed waveform position, relative to the trigger
point.
Scale
: Set the horizontal time/div (scale factor) for the main or the Delayed Scan
time base. When Delayed Scan is enabled, it changes the width of the window zone
by changing the Delayed Scan time base.
Offset
: Adjusts the horizontal position of all channel (including Math)
waveforms. The resolution of this control varies with the time base. Click the
Mid-circular button clears trigger offset and moves the trigger point to the horizontal
center of the screen.
①. The current waveform position window in the memory.
②. The trigger position in the memory.
③. The trigger position in the current waveform windows.
④. The horizontal time base (the main time base).
⑤. The trigger’s horizontal offset according to the center of the window.
The conventional oscilloscope display format. It shows the voltage of a waveform
record (on the vertical axis) as it varies over time (on the horizontal axis).
X-Y:
X-Y format displays channel 1 in the horizontal axis and channel 2 in the vertical axis.
Roll Mode:
In this mode, the waveform display rolls from right to left. No trigger or horizontal
offset control of waveforms is available during Roll Mode, and it’s only available
when set to 500 ms/div or slower.
Slow Scan Mode:
his mode is available when the horizontal time base is set to 50ms or slower. In this
mode, the oscilloscope acquires sufficient data for the left part to the trigger point,
then wait for tri
er, when trigger occurs, it continue to draw the rest part from the
trigger point to the end of the right side. When choosing this mode to view low
frequency signals, it is recommended that the channel coupling be set as DC.
Time/Div:
Horizontal scale. If the waveform acquisition is stopped (using the
Time/Div control expands or compresses the waveform.
The Delayed Scan is a magnified portion of the main waveform window. Use Delayed
Scan to locate and horizontally expand part of the main waveform window for a more
detailed (higher horizontal resolution) analysis of signal. The Delayed Scan time base
setting cannot be set slower than the main time base setting.
Main time base
Waveform to be horizontally expanded
Time base of Delayed Scan
Expanded waveform in horizontal
Figure 2-21
Delayed Scan
The following describes the Delayed Scan display.
The screen is divided into two parts. The upper half displays the main waveform
window and the lower half displays an expanded portion of the main waveform
window. This expanded portion of the main window is called the Delayed Scan
window. Two blocks shadow the upper half; the un-shadowed portion is expanded in
This format is useful for studying phase relationships between two signals.
Channel 1 in the horizontal axis(X) and Channel 2 in the vertical axis(Y). The
oscilloscope uses a none-trigger acquisition mode, data is displayed as dots.
Figure 2-22
The X-Y Display
The following modes or functions will not work in X-Y mode.
z LA Function (for VS5000D only)
z Automatic Measurements
z Cursor Measurements
z REF and MATH Operations
z Delayed Scan Mode
z Vector Display Mode
The trigger determines when the oscilloscope starts to acquire data and display a
waveform. When a trigger is set up properly, it can convert unstable displays or blank
screens into meaningful waveforms.
When the oscilloscope starts to acquire a waveform, it collects enough data so that it
can draw the waveform to the left of the trigger point. The oscilloscope continues to
acquire data while waiting for the trigger condition to occur. After it detects a trigger,
the oscilloscope continues to acquire enough data so that it can draw the waveform
to the right of the trigger point.
In the H
Mode: Select the trigger mode.
Setting: Set the trigger Source channel, trigger Level and trigger Slope mode.
orizontal of the main menu, there is Trigger menu include:
Sweep: Set the sweep mode to Auto, Normal or Single.
Trigger setting: Set the trigger Coupling, adjust the trigger Sensitivity and
An edge trigger determines whether the oscilloscope finds the trigger point on the
rising or the falling edge of a signal. Select Edge trigger Mode to trigger on Rising
edge or falling edge.
Table 2-13 Edge Trigger Settings
FunctionSettingDescription
Mode
Source
Slope
Sweep
Set Up To go to the Set Up menu
CH1
CH2
EXT
EXT/5
AC Line
D15-D0
Rising
Falling
Auto
Normal
Single
Select the trigger mode to Edge
Select CH1 as trigger signal
Select CH2 as trigger signal
Select EXT TRIG as trigger signal
Select attenuated EXT TRIG/5 as trigger signal
Select power line as trigger signal
Select a digital channel in D15-D0 as trigger
source (for VS5000D only)
Trigger on rising edge
Trigger on falling edge
Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred.
When trigger occurs, acquire one waveform
then stop
Pulse trigger occurs according to the width of a pulse. The abnormal signals can be
detected through setting up the pulse width condition.
Table 2-14 Pulse Width Trigger Settings
FunctionSettingDescription
CH1
CH2
EXT
Source
Level To set the trigger level
EXT/5
D15-D0
Select CH1 as trigger signal
Select CH2 as trigger signal
Select EXT TRIG as trigger signal
Select attenuated EXT TRIG/5 as
trigger signal
Select a digital channel in D15-D0
as trigger source(for VS5000D only)
+More than
+Less than
When
Pulse Width Set required pulse width
Sweep
Set Up To go to the Set Up menu
NOTE: The Pulse width adjusts range is 20ns - 10s. When the rule is met, it will
trigger and acquire the waveform.
+Equal
-More than
-Less than
-Equal
Auto
Normal
Single
+Pulse width more than
+Pulse width less than
+Pulse width equal to
-Pulse width more than
-Pulse width less than
-Pulse width equal to
Acquire waveform even no trigger
occurred
Acquire waveform when trigger
occurred.
When trigger occurs, acquire one
waveform then stop
When alternative trigger is on, the trigger sources are from two vertical channels.
This mode can be used to observe two non-related signals. Choose two different
trigger modes for the two vertical channels. The options are as follows: Edge, Pulse,
Slope and Video. The information of the trigger level of the two channels will be
displayed on the upper-right of the screen.
Table 2-17 Alternative Trigger Settings
FunctionSettingDescription
Channel
Trigger
Typ e
Set Up
NOTE:
In Alternative Trigger, each channel is allowed to set the trigger mode independently.
The trigger setting of CH1 and CH2 are mutually independently. Please make
attention to the Channel options.
CH1
CH2
Edge/Pulse/
Slope/Video
Set trigger mode for Channel 1
Set trigger mode for Channel 2
Set Edge/Pulse/Slope/Video Trigger
as the trigger type
Pattern trigger identifies trigger terms by checking appointed code. The code is
logical relationship of all channels, with High (H), Low (L), and Ignore (X), Rising and
Falling.
Table 2-18 Pattern Trigger Settings
FunctionSettingDescription
Defining D15-D0 Choose digital channel for Pattern trigger
H
L
Pattern
Sweep
Set Up
Key Points:
H (High): Logic high: Voltage is higher than threshold setting.
L (Low): Logic low: Voltage is lower than threshold setting.
X (Ignore): Don’t care. If all the channels are ignored, the oscilloscope won’t be
triggered.
Edge of appointed code
X
Rising
Falling
Auto
Normal
Single
High
Low
Ignore
Rising
Falling
Acquire waveform even no trigger occurred
Acquire waveform when trigger occurred
When trigger occurs, acquire one waveform
and then stop
To go to the Set Up menu
Appoint one code as edge only. If an edge is appointed, then appointed another
edge in a different channel and the first appointed edge will be set to X (Ignore).
In the trigger set up menu, set up different trigger settings according to different
trigger modes. When in the mode of Edge and Pulse, the source is D15-D0 (for
VS5000D only), only Holdoff is adjustable. When source is non-digital channel and in
slope trigger, only trigger coupling, trigger sensitivity and Holdoff can be set. For
video trigger, Sensitivity and Holdoff can be set. For Pattern trigger and Duration
trigger (for VS5000D only), only Holdoff can be set. When in Alternative trigger,
make different settings according to selected different trigger modes.
Table 2-20 (Settings for trigger coupling, trigger sensitivity and holdoff)
FunctionSettingDescription
Coupling
DC
AC
HF Reject
LF Reject
Allow all signals pass
Block DC signals
Reject high frequency signals
Reject DC and low frequency signals
Sensitivity
Holdoff
Time
Set trigger sensitivity
Set time slot before another trigger
event
Table 2-21 (Settings for sensitivity and holdoff)
FunctionSettingDescription
Sensitivity
Holdoff
Set trigger sensitivity
Set time slot before another trigger
Use Trigger Holdoff to stabilize a complex waveform, such as a pulse range. Holdoff
time is the oscilloscope’s waiting period before starting a new trigger. During Holdoff,
oscilloscope will not trigger until Holdoff ends. For instance: To trigger on the first
pulse on a group of them, users can set the holdoff time to Pulse cluster width.
Trigger
Point
Time of
Holdoff
Figure 2-26
Trigger Holdoff
To use trigger Holdoff:
1. Turn on the Trigger window, and go to the Trigger functions Menu.
2. Click the corresponding menu to set the trigger settings.
3. Click the left and the right arrow (
) button to change Holdoff time until
waveform is stable.
4. Click the circular button (
) could reset the Holdoff time to its default value.
Trigger occurs from several sources: Input channels (CH1, CH2 and D15-D0 (for
VS5000D only)), AC Line, Ext, Ext/5.
z CH1 or CH2: It is the most commonly used trigger source. The channel will work
when selected as a trigger source whatever displayed or not.
zExt Trig: The instrument can be trigged from a third source while acquiring data
from CH1 and CH2. For example, to trigger from an external clock or with a signal
from another part of the test circuit. The Ext, Ext/5 trigger sources use an
external trigger signal connected to the Ext Trig connector. It has a trigger level
range of +1.2 V to -1.2 V. The Ext/5 trigger source attenuates the signal by 5X,
which extends the trigger level range to +6 V to -6 V, allowing the oscilloscope to
trigger on a larger signal.
2. Sweep Mode:
The sweep mode determines how the oscilloscope behaves in the absence of a
trigger event. The oscilloscope provides three trigger modes: Auto, Normal, and
Single.
zAuto: This sweep mode allows the oscilloscope to acquire waveforms even
when it does not detect a trigger condition. If no trigger condition occurs while
the oscilloscope is waiting for a specific period (as determined by the time-base
setting), it will force itself to trigger.
When forcing invalid triggers, the oscilloscope cannot synchronize the waveform,
and the waveform seems to roll across the display. If valid triggers occur, the
display becomes stable on the screen.
Any factor results in the un-stability of waveforms can be detected by Auto
Trigger, such as the output of power supply.
NOTE: When horizontal control is set more than 50 ms/div, Auto mode allows the
oscilloscope to capture without trigger signal.
zNormal: The Normal mode allows the oscilloscope to acquire a waveform only
when it is triggered. If no trigger occurs, the oscilloscope keeps waiting, and the
previous waveform, if any, will remain on the display.
zSingle: In Single mode, after pressing the RUN/STOP key, the oscilloscope waits
for trigger. While the trigger occurs, the oscilloscope acquires one waveform and
Trigger coupling determines which signal component passed to the trigger circuit.
Coupling types include AC, DC, LF Reject and HF Reject.
z AC: AC coupling blocks DC components.
z DC: DC coupling passes both AC and DC components.
z LF Reject: LF Reject coupling blocks DC component, and attenuates all signal
with a frequency lower than 8 kHz.
zHF Reject: HF Reject coupling attenuates all signals with a frequency higher
than 150 kHz.
4.Pre-trigger/delayed trigger:
The data collected before and after trigger.
The trigger position is typically set at the horizontal center of the screen. In the
full-screen display 6 divisions’ data of pre-trigger and delayed trigger can be
surveyed. More data (14div) of pre-trigger and 1s delayed trigger can be
surveyed by adjusting the horizontal scale and offset.
This feature is very useful to study the events that led up to the trigger point.
Everything to the right of the trigger point is called post-trigger information. The
amount of delay range (pre-trigger and post-trigger information) depends on the
sweep speed selected.
z Adjustable trigger sensitivity
To avoid the influence of noise from the environment and get the stable trigger,
the trigger circuit has adopted stickiness. In VS5000 series, the stickiness is
adjustable from 0.1div-1.0div, which means when it sets to 1.0div, the trigger
circuit will not care any signal with peak-peak amplitude less than 1.0div, so as to
avoid the influence of the noise.
z Select Real Time acquisition to observe the single-shot or pulse signals.
z Select Equal Time to observe high frequency repetitive signals.
z To reduce the displayed random noise, select the Average acquisition and this
mode would make the waveform update slower.
zTo avoid signal aliasing, select Peak Detect acquisition.
Stop Acquisition: When the scope is acquiring waveforms, the waveforms are in
live status; when acquisition is stopped, a frozen waveform will be displayed. The
position and scale can still be adjusted by vertical control and horizontal control.
Key Points
Real-time Sampling:
VS5000 has Real-time sampling rate up to 400MSa/s. At the time base 50ns or
faster, the oscilloscopes use the sine(x)/x interpolation to expand the horizontal time
base.
Equivalent sampling:
Known as Repetitive sampling to get up to 40 ps of horizontal resolution (equivalent
25Gsa/s). This mode is good for observing repetitive signals, and it is not
recommended for single-shot or pulse.
Normal:
Oscilloscope acquires signal by equal time interval.
Average Acquisition:
Apply averaging to signal to remove uncorrelated noise and improve measurement
accuracy. Reduces random or uncorrelated noise in the signal display. The averaged
waveform is a running average over a specified number of acquisitions from 2 to
256.
Peak Detect:
Peak Detect mode captures the maximum and minimum values of a signal. Finds
highest and lowest record points over many acquisitions.
torage button to get into the Save/Load and Print options.
Figure 2-32
The Storage Menu
Storage Menu
Click button Storage, then pop up a storage function menu. You can save and load
waveform in Chinese and English through the menu as follows:
Table 2-24 Storage Functions
FunctionSettingDescription
Load
Save
Print
Load Waveform
Load Setup
Load Default
Save Waveform
Save Setup
Save Screen
PrintPreview
Print
Load the waveform that had saved
Load the instrument that had saved
Load the instrument default settings
Save the current waveform
Save the current instrument settings
Save the PC software window as a .bmp
or .jpg file
Preview the waveform
Print setting and do the print operation
The Measure functions are in the Analyzer menu; to do Horizontal Measure and
Vertical Measure, or to use shortcut buttons to perform measurement operations.
Figure 2-38
Choose the Measurement Menu
Menu explanation:
The oscilloscopes provide 20 parametric auto measurements (10 voltage and 10 time
measurements):
NOTE: The results of the automatic measurements will be displayed on the bottom
of the screen. Maximum 3 results could be displayed at the same time. When there is
no room, the next new measurement result will make the previous results moving
left, out of screen.
Apply Automatic Measurement as the following steps:
1. Select the signal channel for measuring, select CH1 or CH2 according to the
signal of interest.
2. To see all the measurement values, click All Measure in the Analyze menu, 18
measurement parameters will be displayed on the screen.
3. Choose the measurement parameters: measure projects according to the
Horizontal measurement 、Vertical measurement under Analyze Also , you can
drag the left icons to the waveform interface directly.
Get the measured value on the screen; select the parameters of interest by clicking
the correspond button, and read the data on the bottom of the screen.
To Clear the measure values; click Clear Measure, and all auto measured values
disappear from the screen.
The VS5000 series oscilloscopes provide automatic voltage measurements: Vpp,
Vmax, Vmin, Vavg, Vamp, Vrms, Vtop, Vbase, Overshoot and Preshoot. Figure 2-41
below shows a pulse with some of the voltage measurement points.
Overshoot
Vmax
Vpp
Vtop
amp
Vbase
Vmin
Preshoot
Figure 2-42
Definition of Voltage Parameters
Vpp: Peak-to-Peak Voltage.
Vmax: The maximum amplitude. The most positive peak voltage measured over the
entire waveform.
Vmin: The minimum amplitude. The most negative peak voltage measured over the
entire waveform.
Vamp: Voltage between Vtop and Vbase of a waveform
Vtop: Voltage of the waveform’s flat top, useful for square/pulse waveforms.
Vbase: Voltage of the waveform’s flat base, useful for square/pulse waveforms.
Overshoot: Defined as (Vmax-Vtop)/Vamp, useful for square and pulse waveforms.
Preshoot: Defined as (Vmin-Vbase)/Vamp, useful for square and pulse waveforms.
Average: The arithmetic mean over the entire waveform.
Vrms: The true Root Mean Square voltage over the entire waveform.
Click the Cursor button in the Analyze menu to select the Cursor functions.
Cursor
measurement
Figure 2-44
Choose the Menu of Cursor Measurement
The Cursor measurement has three modes: Auto Measure, Manual and Track.
Auto Measure:
This mode will take effect with Automatic Measurements. The instruments will
display cursors while measuring parameters automatically. These cursors
demonstrate the physical meanings of these measurements.
Manual:
In this mode, the screen displays two parallel cursors. Move the cursors to make
custom voltage or time measurements of the signal. The values are displayed on the
boxes below the menu. Before using cursors, make sure that to set the Signal Source
as the channel for measuring.
Track:
In this mode, the screen displays two cross cursors. The cross cursor sets the
position on the waveform automatically. Adjust the horizontal position of cursor on
the waveform by click the
oscilloscope displays the values of the coordinates on the boxes below the menu.
button or drag the cursor on the screen. The
NOTE: The Auto Measure mode for Cursor measuring will take no effect without