Rigol VS5042D, VS5042, VS5202, VS5202D User guide

User’s Guide RIGOL
Publication Number UGA05109-1110 Oct. 2008
VS5000 Series Digital Oscilloscope
© 2008 RIGOL Technologies, Inc. All Rights Reserved
z © 2008 RIGOL Technologies, Inc. All Rights Reserved. z RIGOL products are protected by patent laws in and outside of the P.R. China. z Information in this publication replaces all previous corresponding material. z RIGOL reserves the right to modify or change part of or al l the specif ications and
pricing policies at company’s sole decision.
NOTE: RIGOL is registered trademark of RIGOL TECHNOLOGIES, INC.
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
I
System Requirement
To operate the VS5000 software, the needs of computer configuration are as follows:
Hardware Configuration:
z Upwards of 1G processor z 200M hardware space z 800x600 display resolution z 256RAM z USB
Operating System:
Microsoft Windows 2000、2003、XP.
Recommended Configuration:
z Windows XP System z 1024*768 or 1280*1024 resolution z DPI setting: Normal Size (96DPI) z Non- Unicode program language: English.
II
User’s Guide for VS5000 Series
© 2008 RIGOL Technologies, Inc

Safety Notices

Review the following safety precautions carefully before operate the instrument to avoid any personal injuries or damages to the instrument and any products connected to it. To avoid potential hazards use the instrument as specified by this user’s guide only.
The instrument should be serviced by qualified personnel only. To Avoid Fire or Personal Injury. Use Proper Power Cord. Use the power cord designed for the instrument and
authorized in your country only.
Ground The Instrument. The oscilloscope is grounded through the grounding
conductor of the power cord. To avoid electric shock the instrument grounding conductor(s) must be grounded properly, before making connections to the input or output terminals of the instrument.
Connect The Probe. The probes’ ground terminals are at the same voltage level of
the instrument ground. Do not connect the ground terminals to a high voltage.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all ratings
and marks on the instrument. Follow the user’s guide for further ratings information before making connections to the instrument.
Do Not Operate Without Covers. Do not operate the instrument with covers or
panels removed.
Use Proper Fuse. Use the fuse of the type, voltage and current ratings as specified
for the instrument.
Avoid Circuit or Wire Exposure. Do not touch exposed connections and
components when power is on.
Do Not Operate With Suspected Failures. If suspected damage occurs with the
instrument, have it inspected by qualified service personnel before further operations.
Provide Proper Ventilation. Refer to the installation instructions for proper
ventilation of the instrument.
Do not operate in Wet/Damp Conditions.
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
III
Do not operate in an Explosive Atmosphere. Keep Product Surfaces Clean and Dry. The disturbance test of all the models meet the limit values of A in the
standard of EN 61326: 1997+A1+A2+A3, but can’t meet the limit values of B.
IV
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
V
Measurement Category
The VS5000 series Digital Oscilloscope is intended to be used for measurements in
Measurement Category I.
Measurement Category Definitions
Measurement Category I is for measurements performed on circuits not directly
connected to MAINS. Examples are measurements on circuits not derived from MAINS,
and specially protected (internal) MAINS derived circuits. In the latter case, transient
stresses are variable; for that reason, the transient withstand capability of the equipment
is made known to the user.
WARNING
IEC Measurement Category I. The HI and LO input terminals may be connected to circuit terminal in IEC Category I installations for voltages up to 300 VAC. To avoid the danger of electric shock, do not connect the inputs to circuit’s voltages above 300 VAC. Transient overvoltage is also present on circuits that are isolated from mains. The VS5000 series Digital Oscilloscopes is designed to safely withstand occasional transient overvoltage up to 500 Vpk. Do not use this equipment to measure circuits where transient overvoltage could exceed this level.
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
V
Safety Terms and Symbols
Terms in this manual. These terms may appear in this guide:
!
!
Terms on the Product: These terms may appear on the product:
DANGER indicates an injury hazard may be immediately accessible.
WARNING indicates an injury hazard may be not immediately accessible.
CAUTION indicates that a potential damage to the instrument or other property
might occur.
Symbols on the Product: These symbols may appear on the Instrument:
Hazardous Voltage
WARNING: Warning statements identify conditions or practices that could result in injury or loss of life.
CAUTION: Caution statements identify conditions or practices that could result in damage to this product or other property.
!
Refer to Instructions
Protective Earth Terminal
Grounding Terminal of Chassis
Test Grounding Terminal
I
User’s Guide for VS5000 Series
© 2008 RIGOL Technologies, Inc
V

VS5000 Oscilloscopes at a Glance

RIGOL VS5000 Series Digital Oscilloscopes offer good waveform view and
measurements in a compact, lightweight package. VS5000 series are ideal for production test, field service, research and design and all of the applications involving analog circuits test and troubleshooting, as well as education and training.
Product features: z Dual Channel, Bandwidth:
200MHz (VS5202, VS5202D) 100MHz (VS5102, VS5102D) 60MHz (VS5062, VS5062D) 40MHz (VS5042, VS5042D) 25MHz (VS5022, VS5022D)
z 400 MSa/s maximum real-time sample rate and 25 GSa/s maximum equivalent
sample rate
z A true Mixed Signal Oscilloscope with 16 channels Logic Analyzer (for some
models)
z Memory depth: 1M points (Single Channel), 512k points (Dual Channels), 512k
points (Logic Analyzer)
z Adjustable waveform intensity, more effective waveform view z One-touch automatic setup for ease of use (AUTO) z 20 Automatic measurements z Automatic cursor tracking measurements z Waveform recorder, record and replay dynamic waveforms z User selectable fast offset calibration z Built-in FFT function, Frequency Counter z Digital filters, includes LPF, HPF, BPF, BRF z Add, Subtract and Multiply Mathematic Functions z Advanced trigger types include: Edge, Video, Pulse width, Slope, Alternative,
Pattern and Duration
z Adjustable trigger sensitivity
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
II
V
Construction of This Document
z Chpter1 Quick Start
To introduce the panls and menus to help user to realize the base operations.
z Chapter 2 Operating Your Oscilloscope
More details to help users to understand the functions and measurements of instrument deeply.
z Chapter 3 Application Examples
To show the measurement functions directly by application examples.
z Chapter 4 Troubleshooting
Provide you general troubleshooting to solve the problems during operating.
z Chpter 5 Specifications
List common specifications and characteristics of VS5000 series oscilloscope.
z Chapter 6 Appendix
Information about accessories, warranties, services and supports and the like.
III
User’s Guide for VS5000 Series
© 2008 RIGOL Technologies, Inc
Content
Safety Notices............................................................................................ III
VS5000 Oscilloscopes at a Glance............................................................... VII
Chapter 1 Quick Start............................................................................1-1
General Inspection ....................................................................................1-2
Front and Rear panels At a Glance..............................................................1-3
Software Setting and Equipment Connecting ...............................................1-4
To Understand the User’s Interface ............................................................1-7
Functional Check.......................................................................................1-9
To Compensate Probes ............................................................................1-11
To Use Digital Leads (for VS5000D only)................................................... 1-12
To Display a Signal Automatically .............................................................1-14
To Understand the Vertical Window.......................................................... 1-15
To Understand the Horizontal System ....................................................... 1-17
To Understand the Trigger the Oscilloscope ..............................................1-19
Chapter 2 Operating Your Oscilloscope.................................................2-1
To Set the Vertical system .........................................................................2-2
To Set the Horizontal System ................................................................... 2-26
To Set the Trigger System .......................................................................2-33
To Set the Sampling System ....................................................................2-48
To Save/Load and Print Setup ..................................................................2-53
To Set the Measurement Function ............................................................ 2-59
To Set the Utility Function........................................................................ 2-73
To Use Run Control Buttons .....................................................................2-81
Chapter 3 Application Examples............................................................3-1
Example 1: Do Simple Measurement...........................................................3-1
Example 2: View a Signal Delay Caused by a Circuit.....................................3-3
Example 3: Capture a Single-Shot Signal.....................................................3-1
Example 4: To reduce the Random Noise on a signal ...................................3-2
Example 5: Making Cursor Measurements ...................................................3-4
Example 6: The Application of the X-Y Operation.........................................3-6
Example 7: Triggering on a Video Signal .....................................................3-8
Example 8: FFT Cursor Measurement........................................................ 3-10
Example 9: Pass/Fall Test ........................................................................3-12
Example 10: Triggering on a Digital Signal ................................................ 3-14
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
IX
Chapter 4 Troubleshooting ................................................................... 4-1
Chapter 5 Specifications .......................................................................5-1
Chapter 6 Appendix............................................................................... 6-1
Appendix A: Accessories ............................................................................ 6-1
Appendix B: Warranty................................................................................6-2
Appendix C: Maintenance ..........................................................................6-3
Appendix D: Contact RIGOL .......................................................................6-4
Index ...........................................................................................................I
X
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series

Chapter 1 Quick Start

This chapter covers the following topics:
z General Inspection z Front and Rear Panels At a Glance
z Software Setting and Equipment Connecting
z To Understand the User Interface
z To perform a functional check
z To compensate probes
z How to use digital probes (VS5000D)
z To display a signal automatically
z To Understand the Vertical System
z To Understand the Horizontal System
z To Understand the Trigger System
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
1-1

General Inspection

After receiving a new VS5000 series oscilloscope, please inspect the instrument as following steps:
1. Inspect the shipping container for damage.
Keep the damaged shipping container or cushioning material until the contents of the shipment have been checked for completeness and the instrument has been checked mechanically and electrically.
2. Check the accessories.
Accessories supplied with the instrument are listed in "Accessories" in this guide. If the contents are incomplete or damagedplease notify the RIGOL Sales Representative.
3. Inspect the instrument.
In case there is any mechanical damage or defect, or the instrument does not operate properly or fails performance tests, notify the RIGOL Sales Representative.
If the shipping container is damaged, or the cushioning materials show signs of stress, notify the carrier as well as the RIGOL sales office. Keep the shipping materials for the carrier’s inspection.
RIGOL offices will arrange repair or replacement at option of RIGOL without
waiting for claim settlement.
1-2
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series

Front and Rear panels At a Glance

The first thing to do with a new oscilloscope is to become acquainted with its front panel. This chapter will help to be familiar with the layout of the connectors, graphical representations, interfaces, and how to use them. Read the chapter carefully before further operations.
Figure 1-1
Front panel of VS5000 Series Oscilloscope
Figure 1-2
Rear panel of VS5000 Series Oscilloscope
VS5000 series oscilloscopes provide an easy-to-use user interface. In the main menu, access the function options from the menu and the drop-down menu to operate VS5000 series oscilloscope. There’re shortcuts for some common functions.
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
1-3

Software Setting and Equipment Connecting

You are suggested to install the software of VS5000 and connect the equipment when you get a new RIGOL VS5000 Oscilloscope. For more details please refer to the following text:
Software Setting
Please insert and drive the disk accessory of the RIGOL VS5000 oscilloscope. Then, double click the file named” according to the hints.
Note: If you have already installed VS5000 software before, please uninstall the old
one and then start.
1-4
VS5000_SETUP _CHS.exe” to install the software
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
Equipment Connection
Power Supply
Make sure the power supply was connected correctly before using the instrument. First, please insert power cord into the faucet of its adapter. Then connect the adapter with oscilloscope and power on.
USB Connection
When the software setting and power supply connecting are finished, you can connect the instrument to PC by USB cable. After connecting, a dialog about setup will pop up to clew users to finish USB connection.
LAN Connection
A LAN port is offered at the rear panel in order to realize Remote Control which acts a same role with USB. Besides, two steps are required to connect LAN: network setting and connection setting.
1.Network Setting: There are two ways to config Network Protocol Property: Obtain an IP automaticallyobtain an IP manually.
(1) Obtain an IP automatically:
Please set network according to the following steps: z Connecting: On condition that the USB cable was being connected, you can
configure l Property of Network protocol. First, connect oscilloscope to PC via USB, and then, connect oscilloscope to LAN.
z Open control->Network Setting, then it will pop up a dialog about configuration.
Click “obtain an IP automatically” option and press Apply to finish. .
z Restart oscilloscope and software, you can view the Property of Network
protocol in Configuration dialog box if USB communication is active.
(2) Obtain an IP manually:
z Connecting: On condition that the USB cable has being connected, you could
configure l Property of Network protocol. First, connect oscilloscope to PC via
USB, and then, connect oscilloscope to LAN.
z Open control->Network Setting, it will pop up a dialog about configuration. Click
“obtain an IP automatically” option again to cancel Automatic setting. Then Setting IP addressIP maskGateway by manual . Finally, click Apply to configure Network protocol Property.
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
1-5
z Restart oscilloscope and software, you can view the Property of Network
protocol in Configuration dialog box if USB communication is active.
NOTES:
z Make sure that the oscilloscope has been connected with network through
Router in” Obtain an IP automatically”.
z Make sure the validity of IP address in ‘Obtain an IP manually’. z Restart the oscilloscope and PC after finishing configuration. z The function of Configure Network is invalidation in Network Communication.
2. Connection Setting:
z Connect software and oscilloscope in Network Communications. z Select Network Communication. Open Control->Connection menu, it will pop up
a dialog, click check option LAN and select Network Communication, then set IP address and port value to be the oscilloscope’s. Finally, click Apply to finish connecting.
z If communication has been found, System will clew that network connection is
successful. If failed, please check hardware connection and settings of Network Protocol property.
NOTES:
z Please pay attention to check the settings of Safety level of firewall when using
Network Communication.
z In network communication, you are suggested to use “Resume default setting”
carefulness as Default connection is USB connection. Meanwhile, related result may be brought if you cancel all the settings about Network Protocol Property.
1-6
User’s Guide for VS5000 Series
© 2008 RIGOL Technologies, Inc

To Understand the User’s Interface

Click the VS5000 software icon after you finished the software setting and equipment connecting. Then a user interface will be popped up as follows:
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
Figure 1-3
VS5000 Operation Interface
1-7
Trigger level
Channel 1
Channel 2
Digital
Channels
Trigger point, Time reference
1-8
Figure 1-4
Display Interface (Analog and Digital channels)
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series

Functional Check

Perform once quick check to verify the running conditions of the instrument.
1. Prepare for using the instrument.
Use the power cords designed for the oscilloscope only. Connect to a power source that delivers 100 to 240 VAC, 45Hz to 440Hz. Power up the system, initiate the program. Click the Storage button, select Load Default in the menu box.
Status indicators
Channel 1, 2 and Ext Trigger
Probe compensator
Figure 1-5
Instruction of the Front Panel
2. Input a signal to a channel of the oscilloscope
VS5000 Series are equipped with dual channels plus external trigger and 16 channels logic analyzer (only for VS5000D models).
Please input signal in the light of following steps:
1Set the attenuation switch on the probe as 10X (Figure 1-6) and connect the
probe on the oscilloscope with CH1. Aim the slot in the probe connector at the faucet on BNC of CH1 and insert, then, turn right to lock the probe. Finally, attach the tip of probe and ground nip to the Connector of Probe compensator.
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
1-9
Probe compensator
Figure 1-6
Probe Compensation
2Set the probe attenuation of the oscilloscope to X10. (The default is X1).
Method of set the probe attenuation: Right click the button into the Channel menu. In the Probe menu, select the
probe attenuation. Now select the 1:10 option.
Probe attenuation
Figure 1-7 Figure1-8 To Set the Probe To Set the Probe Attenuation
3Attach the tip of probe and ground nip to the Connector of Probe compensator.
Click the
button. A square wave will be displayed within a few seconds.
(Approximately 1 kHz, 3 V, peak- to- peak).
4Inspect Then CH2 with the same method. Click the
off, click the
button to turn Channel 2 on. Repeat steps 2 and 3.
button to turn Channel 1
1-10
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series

To Compensate Probes

Perform this function to match the characteristics of the probe and the channel input. This should be performed whenever attaching a probe to any input channel at the first time.
1. From the Probe menu, select attenuation to 1:10. Set the switch to X10 on the probe and connect it to CH 1 of the oscilloscope. When using the probe hook-tip, insert the tip onto the probe firmly to ensure a proper connection. Attach the probe tip to the Probe Compensator and the reference lead to the
ground connector, select CH 1, and then press the
2. Check the shape of the displayed waveform.
Over compensated Correctly Compensated Under Compensated
button.
Figure 1-9
Instruction Waveform of Probe Compensation
3. If necessary, use a non-metallic tool to adjust the trimmer capacitor of the probe for the flattest square wave being displayed on the oscilloscope.
4. Repeat if necessary.
WARNNING: To avoid electric shock while using the probe, be sure the
!
perfection of the insulated cable, and do not touch the metallic portions of the probe head while it is connected with a voltage source.
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
1-11

To Use Digital Leads (for VS5000D only)

1. Switch off power supply of the Device under Test if necessary to avoid short circuit. Since no voltage is applied to the leads at this step, the oscilloscope may remain on.
2. Connect one end of the flex cable FC1868 to the Logic Analyzer Input; connect
the other end to Logic Head LH1116. It is unnecessary to switch off power supply of the oscilloscope while connecting the cable.
1-12
Figure 1-10
Connect the Flex Cable
CAUTIONUse only FC1868, LH1116, TC1100 and LC1150 made by
!
RIGOL for specified mixed signal oscilloscopes.
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
3. Connect a test clip to one lead wire; make sure it is connected well.
Test clip
Figure 1-11
Connect the Test Clip
4. Test the device with the clip.
Figure 1-12
Testing Instruction
5. Remember to connect Ground Channel to the ground terminal of the DUT.
GND
Figure 1-13
Ground Connection
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
1-13

To Display a Signal Automatically

The VS5000 series oscilloscope has an automated function to display the input signal in best fit. This feature requires a 50 Hz or higher and a duty cycle greater than 1% input signal.
Press the HORIZINTAL and TRIGGER controls to display the input signal. Adjust the controls manually to get the best results if necessary.
Connect a signal to the Channel 1 (CH1) input:
1Connect a signal to the oscilloscope as described above.
button, the oscilloscope automatically sets up VERTICAL,
2Click the
button.
The oscilloscope may change the current settings to display the signal, i.e. the vertical and horizontal scale, the trigger coupling, type, position, slope, level, and mode.
1-14
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series

To Understand the Vertical Window

Figure 1-13 shows the V parameters.
ertical menu, which leads to the setting of the vertical
Figure 1-14
Vertical Menu
1Use the Analog function in the Vertical menu to adjust the functions of the
analog channels. A voltage value is displayed for a short time indicating how far the ground reference is located from the center of the screen. Use the left
mouse button to dragging the CH1 symbol, and/or use the Scale and Offset
options to set the scale and offset precisely.
Measurement hints
If the channel is DC coupled, measuring the DC component of the signal by simply noting its distance from the ground symbol. If the channel is AC coupled, the DC component of the signal is blocked, allowing using greater sensitivity to display the AC component of the signal.
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
1-15
Vertical offset back to 0 shortcut
Double click the channel sign, as a shortcut, to set the vertical display position back to 0.
2Change the vertical setup and notice that each change affects the status bar
differently.
Left-click the vertical status and use the number keyboard to input the exact
Scale value (volt/div); and/or adjust the Scale quickly with the
Click A
nalog, Digital, Math and Reference in the Vertical menu, to set analog
channels, digital channels, math and waveform reference functions.
button.
1-16
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series

To Understand the Horizontal System

Figure 1-14 shows the H parameters.
orizontal menu, which leads the setting of the horizontal
Figure 1-15
Horizontal Menu
1 Using the Horizontal function in the Horizontal menu to set the waveform
horizontal options. In the Horizontal window, set the waveform display format (Y-T, X-Y and Roll) and horizontal Scale and Offset.
Click the
button, changes the sweep speed in a 1-2-5 step sequence and the
value is displayed in the status bar. The time base ranges of the VS5000 series
[1]
are listed as follows. The horizontal scan speed is from 5ns/div
to 50s/div.
Delayed Scan Shortcut key
Click the
button to enter Delayed Scan mode.
[1] NOTE: The speed of horizontal scan varies in different models.
2Click the Offset function in the Horizontal window to adjust the horizontal
offset.
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
1-17
Horizontal offset back to 0 shortcut key
Click the middle circular button “ ” to set the horizontal offset to 0 as a shortcut key
3 In the Format window of the Horizontal functions window, select the display
mode to Y-T, X-Y or Roll.
Horizontal position control
Trig-Offset: In this setting, the trigger position will be changed horizontally
when the offset value changed.
1-18
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series

To Understand the Trigger the Oscilloscope

Figure 1-15 shows the Trigger Control, use the functions in the Trigger window of
orizontal menu. The following exercise guides through the buttons and
the H settings.
Figure 1-16
Trigger Control
1To set the trigger function in the T
rigger menu.
Drag the Trigger level to adjust the trigger level; or input the trigger level directly
on the top right corner (Status window right side of the
2 Click the
button, turn the Trigger window on (Figure 1-16), to set the
button).
trigger options.
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
1-19
Trigger level back to 0 shortcut key
Figure 1-17
Trigger Window
Double click the trigger level sign to set the trigger level to 0 as a shortcut key.
3Double click the trigger point, to set the horizontal delayed time to 0 as a
shortcut key.
4Click the button to start an acquisition regardless of an adequate trigger
signal, usually used in “Normal” or ”Single” trigger mode. If the acquisition is already stopped, click this button to start the acquisition again.
Key point:
Holdoff: A time interval before the oscilloscope response to next trigger signal. During this Holdoff period, the trigger system becomes “blind” to trigger signals. This function helps to view complex signals such as an AM waveform.
In order to set the Holdoff time, set the Holdoff Time option in the Trigger
menu.
1-20
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series

Chapter 2 Operating Your Oscilloscope

The VERTICAL, HORIZONTAL, and TRIGGER control systems and know how to determine the system setup from the status bar of a VS5000 Series oscilloscope were introduced in the previous chapter.
This chapter will go through all groups of buttons, knobs, and menus, expanding the knowledge of operations by hints in this guide.
It is recommended to perform all of the following exercises to get the most of the powerful measurement capabilities of the oscilloscope.
Understand the vertical system (
, ,
[1]
, Vertical )
Understand the horizontal system ( Horizontal )
Understand the trigger system ( Trigger )
To set up the sampling system ( Acquire )
To set up the display system ( Display )
To l oa dsave and print ( Storage )
To set up utility ( Utility )
To measure the waveform ( Analyzer )
To use the run control buttons ( , and )
[1] NOTE: This function is for VS5000D model only.
In order to get more information about the various measurements and operations of
VS5000, please read this chapter carefully.
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-1

To Set the Vertical system

Analog Channel Settings
Each channel of VS5000 has an operation menu.
Right click
Turn the function On/Off:
or ; the “Channel” window will appear.
Figure 2-1
Channel Window
In front of the function option there is a
function is turned OFF,
CAUTION: 1. To aviod damaging the input part of the oscilloscope, when use 50
ohm settings, make sure that the input voltage do not exceed the limit voltage
!
(5Vrms, corresponding to the 1:1 probe)
2. For the probe provided for the oscilloscope, do not use 50 ohm settings directly.
2-2
means it is turned ON.
User’s Guide for VS5000 Series
or choice box, means this
© 2008 RIGOL Technologies, Inc
Table 2-1 Analog Channel Functions
Function Setting Description
Pass both AC and DC components of the input signal
Block the DC component of the input signal
Coupling
DC
AC
Input Impedance
BW Limit
[1]
50 ohm
ON
OFF
1M ohm
Set the Input Impedance to be 1M ohm Set the Input Impedance to be 50 ohm
Limit the channel bandwidth to 20MHz to reduce display noise.
Get full bandwidth.
1:1
Probe
1:10 1:100
Set this to match the probe attenuation factor to make the vertical scale read correct
1:1000
Scale
Invert
Coarse
Exactitude
ON OFF
Select the resolution of the 1-2-5 sequence. Input the scale value directly.
Turn on the invert function. Restore to original display of the waveform.
button def ines a
[1] NOTE: The Input Impedance option is only used for VS5202D or
VS5202.
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-3
1. Channel Coupling
To use Channel 2 as an example, input a sine wave signal with DC offset.
Right click the
enter the Channel window. Set the Coupling to AC. In this setting, it blocks the DC
component of the input signal.
button or select the Analog function in the Vertical menu, and
AC Coupling
Figure 2-2
AC Coupling Setting
Right click the
button or select the Analog function in the Vertical menu, to
enter the Channel window. Set the Coupling to DC. In this setting, it blocks the AC
component of the input signal.
2-4
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
DC Coupling
Figure 2-3
DC Coupling Setting
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-5
2. Set up the Channel Bandwidth Limit
To use Channel 1 as an example, input a signal that contains high frequency component.
Right-click the button, the Channel window will appear. Turn the BW Limit
function off. The oscilloscope is set to full bandwidth and will pass the high frequency component in the signal.
Turn o f f t h e BW Limit
Figure 2-4
Close the BW Limit
Right-click the button, the Channel window will appear. Turn the BW Limit
function on. It will reject the frequency component higher than 20MHz.
20MHz BW Limit
Figure 2-5
Open the BW Limit
2-6
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
3. Probe Attenuation Setting
Select the attenuation factor for the probe. The attenuation factor changes the vertical scale of the oscilloscope so that the measurement results reflect the actual voltage levels at the probe tip.
To change (or check) the probe attenuation setting, press the CH1 or CH2 button
(according to which channel are in using), toggle the Probe soft button to match the
attenuation factor of the probe.
This setting remains in effect until changed again.
Figure 2-6 gives an example for using a 1:1000 probe and its attenuation factor.
Vertical volt/div
Figure 2-6
Probe Attenuation Setting
Table 2-2 Probe setting
Probe attenuation factors Corresponding settings
1:1 1X
1:10 10X
1:100 100X
1:1000 1000X
Probe Attenuation
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-7
4. Volts/Div Settings
The Volts/Div control has Coarse or Fine configuration. The Vertical Sensitivity is
2mV-5V/div.
Coarse:
It is used for setting the Vertical Sensitivity in the form of 1-2-5 that is 2mV/div, 5mV/div, 10mV /div 20mV/div…...10 V/div.
Fine:
This setting changes the vertical scale with accurate value.
Fine adjustment setting value
Fine adjustment data Coarse adjustment
Fine adjustment data Coarse adjustment
Figure 2-7
Scale and Offset Window
Set the vertical scale and offset in the Scale and Offset function of the Channel
window.
2-8
User’s Guide for VS5000 Series
© 2008 RIGOL Technologies, Inc
5. To Invert a Waveform
Invert turns the displayed waveform 180 degrees, with respect to the ground level. When the oscilloscope is triggered on the inverted signal, the trigger is also inverted.
Figure 2-8 and 2-9 show the changes before after inversion.
Invert OFF
Figure 2-8
The Waveform before Inversion
Figure 2-9
The Waveform of Inversion
Invert ON
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-9
Digital Channel Settings (for VS5000D only)
It allows to turn single channel or group channels ON or OFF, to set the Wave Size, to change display location of digital channel on screen and to select Threshold Type.
Left click the D
igital button in the Vertical menu, or right click the button to
show the Logic Analyzer window.
Logic Analyzer Interface
Table 2-3 Logic Analyzer Settings
Function Setting Description
Figure 2-10
2-10
D7-D0 Set up channel group D7-D0
D15-D8 Set up channel group D15-D8
All on ON/OFF
Turn all (D7-D0 or D15-D8) digital channels ON/OFF
Wave Size Big/Small Select the waveform display size.
Reset Reset the waveform display position.
Threshold Typ e
TTL CMOS ECL USER
User’s Guide for VS5000 Series
Select mode of all digital channels.
The threshold voltage can be set by user in user-defined style.
© 2008 RIGOL Technologies, Inc
1Display and re-line up the digital channels
1Tu rn o n th e Logic Analyzer window. Use the options in this window to set the
Channel ON/OFF, Wave Size, Threshold Type and Reset the waveform display position.
2Drag the digital channel sign to move the digital waveform to the designated
location.
3Click the Reset button to reset the waveform position.
Figure 2-11
Open the Channels of D0~D7
2Set Threshold Type of Digital Channels
Select logic standard or User to define the threshold voltage.
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-11
Threshold type
Figure 2-12
To Set the Threshold Type
Setup the value of threshold by user
Threshold explanation
LOGIC STANDARD THRESHOULD VOLTAGE
TTL 1.4V CMOS 2.5V ECL -1.3V USER -8V to +8V
2-12
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
Math Functions
The mathematic functions include “add”, “subtract”, “multiply” and “FFT” for Channel 1 and Channel 2. The mathematic result can be measured by the grid and the cursor. The Math function may turn on four channels, each channel could be operated independently (Except Delayed display).
Mathematic Function Interface
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
Figure 2-13
2-13
Table 2-4 Mathematic Function
Function Setting Description
Math
On Off
A+B A- B
Operation
A×B FFT Filter
Source A
Source B
CH1 CH2
CH1 CH2
Turn this math channel on. Turn this math channel off. Add source A and source B
Subtract source B from source A Multiply source A by source B
Fast Fourier Transform Digital filter
Define CH1 or CH2 as source A
Define CH1 or CH2 as source B
2-14
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
1. Digital Filter
Left click the Math button in the Vertical menu. Select Filter function. In Filter Type, select the type of filter: Low Pass, High Pass, Band Pass or Band Reject. Set
high and low limit of frequency.
Turn o f f Digital Filter
Turn o n Digital Filter
Figure 2-14
Digital Filter Interface
Table 2-5 Digital Filter Function
Function Setting Description
Source
Digital Filter
CH1 CH2
Off On
Low Pass
Filter Type
High Pass Band Pass Band Reject
Upper limit
Lower limit
Define CH1 or CH2 as source
Turn off the digital filter Turn on the digital filter Setup as LPF (Low Pass Filter)
Setup as HPF(High Pass Filter) Setup as BPF (Band Pass Filter)
Setup as BRF(Band Reject Filter)
Set high limit
Set low limit
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-15
2. Addition, Subtraction and Multiplication Function
In Math function, use the addition, subtraction and multiplication function to operate and analyze the waveform.
Figure 2-15
The Addition, Subtraction and Multiplication Function
Enter the Math window. Select the math function in the Operator menu. Select Source A and B. Then adjust the vertical Scale and Offset to view the math channel
clearly.
2-16
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
3. Using the FFT
The FFT (Fast Fourier Transform) process converts a time-domain signal into its frequency components mathematically. FFT waveforms are useful in the following applications:
z Measuring harmonic content and distortion in systems z Characterizing noise in DC power supplies z Analyzing vibration
Table 2-6 FFT Function
Function Setting Description
FFT
OFF ON
A+B
Operation
A-B A×B FFT
Source
CH1 CH2
Rectangle
Window
Hanning Hamming Blackman
Split Window
Scale
Split Window Full Window
Vrms dBVrms
Tur n off the FFT. Open the FFT. Add source A and source B
Subtract source B from source A Multiply source A by source B
Fast Fourier Transform Digital filter
Define CH1 or CH2 as FFT source
Select window for FFT
Display split waveform Display full waveform
Set “Vrms ” as vertical unit Set “dBVrms ” as vertical unit
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-17
Key points for FFT
1Signals that have a DC component or offset can cause incorrect FFT
waveform component magnitude values. To minimize the DC component,
choose AC Coupling on the source signal.
2To reduce random noise and aliases components in repetitive or single-shot
events, set the oscilloscope acquisition mode to average.
3To display FFT waveforms with a large dynamic range, use the dBVrms scale.
The dBVrms scale displays component magnitudes using a log scale.
2-18
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
Selecting a FFT Window
VS5000 series digital oscilloscopes provide four FFT window functions. Each window is a trade-off between frequency resolution and amplitude accuracy. It depends on the desired measurements and source signals characteristics to determine the window to use. Use the following guidelines to select the best window.
Table 2-7 FFT Windows description
Window Features Best for measuring
Transients or bursts, the signal
Best frequency resolution
Rectangle
and worst magnitude resolution. This is essentially the same as no window.
Hanning
Hamming
Better frequency, poorer magnitude accuracy than Rectangular. Hamming has slightly better frequency resolution than Hanning.
Blackman Best magnitude, worst
frequency resolution.
levels before and after the event are nearly equal. Equal-amplitude sine waves with fixed frequencies. Broadband random noise with a relatively slow varying spectrum.
Sine, periodic, and narrow-band random noise.
Transients or bursts where the signal levels before and after the events are significantly different.
Single frequency waveforms, to find higher order harmonics.
Key points: FFT Resolution
With a fixed FFT points, the lower sampling rate results in better resolution.
Nyquist Frequency
The highest frequency that any real-time digital oscilloscope can acquire without aliasing. It’s normally half of the sample rate. This frequency is called the Nyquist frequency. Frequency above the Nyquist frequency will be under
sampled, causing a situation known as aliasing.
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-19
Set the Reference
Reference Waveforms are saved waveforms to be selected for display. The reference function will be available after saving the selected waveform to non-volatile memory.
Click the Reference button in the Vertical menu, go to the Reference window.
Figure 2-16
Reference Interface
Table 2-8 Reference Function
功能菜单 设定 说明
Display
OFF ON
CH1 CH2
Source
MATH1 MATH2 MATH3 MATH4
Close waveform reference Open waveform reference
Select CH1 as Reference channel Select CH2 as Reference channel Select Math1 as Reference channel Select Math2 as Reference channel Select Math3 as Reference channel Select Math4 as Reference channel
2-20
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
File
Vertical Setting
Horizontal Setting
Off On
Vertical displacement Vertical scale
Horizontal displacement Horizontal scale
Turn the file selection off Turn the file selection on
Vertical displacement setting Vertical scale setting
Ho ri zo n ta l displacement setting Horizontal scale setting
Save Save Reference waveform
Load Into the Open File window
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-21
1 Export
Click the Load button in the Reference window.
Table 2-9 Export Function
Function Setting Description
Driver
Open Load the file that was selected
Cancel Cancel the current operation
Keyboard
Figure 2-17
Open a Reference File
Select the driver location
Switch the driver
Use the soft-keyboard input the file name
2-22
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
2 Import
Click the Save button in the Reference window.
Figure 2-18
Save a Reference File
Table 2-10 Import Function.
Function Setting Description
Driver
Select the driver location
Switch the driver
Save Save the file that was named
Cancel Cancel the current operation
Keyboard
Use the soft-keyboard input the file name
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-23
NOTE:
The File function is default Off (
), to save the file in the hard driver, turn
the File function On ( ).
Operation Comments
1Click On (
) to turn the Reference function on.
2Click Source to select CH1, CH2, MATH1, MATH2, MATH3 and MATH4 channel. 3Adjust the Scale and Offset of the Vertical and Horizontal.
4Click the Save button to save the waveform being displayed on the screen as
Reference.
NOTE: The Reference function is not available in X-Y mode.
2-24
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
Turn on/off Channels
The CH1, CH2, Ext Trigger and LA (for VS5000D only) channels on VS5000 series are input channels. All functionalities applied will be based on operating the instrument with channels. So MATH and REF can be regarded as relatively isolated channels.
To turn on/off any one of the channels, press the corresponding button on the front panel. The button turns coloured indicates the channel is currently active. Press the
button again or chose Off in Channel/Logic Analyzer window to turn the channel
off.
Table 2-11 Turn On/Off channels
Channel Mode Settings Status Indicator
Channel 1 (CH1)
Channel 2 (CH2)
ON OFF
ON OFF
ON
MATH
OFF
[1]
ON OFF
ON OFF
Reference
Logic Analyzer
NOTE:
The status indicators (CH1, CH2 and LA) of the VS5000 series in the above table are displayed on the lower-left of the main window.
CH1 (yellow background) CH1 (gray background)
CH2 (blue background) CH2 (gray background)
M1 M2 M3 M4 (to distinguish between the four colors)
No indicator
Ref 1/2/3/4 (white background) No indicator
LA (green background) LA (gray background)
The four MATH function channels are distinguished between the four colors.
Click of
will turn all the digital channels on/off.
[1] Logic Analyzer for VS5000D model only.
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-25

To Set the Horizontal System

The oscilloscope shows the time per division in the scale readout. Since all active waveforms use the same time base, the oscilloscope only displays one value for all the active channels, except when using Delayed Scan, or Alternative Trigger.
The horizontal control changes the horizontal scale and position of waveforms. The horizontal center of the screen is the time reference for waveforms. Changing the horizontal scale causes the waveform to expand or contract with respect to the screen center.
Horizontal position changes the displayed waveform position, relative to the trigger point.
Scale
: Set the horizontal time/div (scale factor) for the main or the Delayed Scan
time base. When Delayed Scan is enabled, it changes the width of the window zone by changing the Delayed Scan time base.
Offset
: Adjusts the horizontal position of all channel (including Math) waveforms. The resolution of this control varies with the time base. Click the Mid-circular button clears trigger offset and moves the trigger point to the horizontal center of the screen.
Click the button
,open the horizontal dialog as follows:
2-26
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
Figure 2-19
Horizontal setting dialog
Table 2-12 Horizontal Setting Function
Function Setting Description
Show the relative relation between vertical voltage and horizontal time.
Show CH1 value at X axis; CH2 value at Y axis.
In Roll Mode, the waveform display updates from right to left.
Time Base
Y-T
X-Y
Roll
Scale Set the horizontal scale
Offset Set the horizontal offset
Delayed
Off On
Turn off Delayed Scan mode Turn on the Delayed Scan mode
All of the parameters will be observed on the soft interface directly.
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-27
①②
Figure 2-20
Horizontal Setting Interface
Marks
①. The current waveform position window in the memory. ②. The trigger position in the memory. ③. The trigger position in the current waveform windows. ④. The horizontal time base (the main time base). ⑤. The trigger’s horizontal offset according to the center of the window.
2-28
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
T
gg
Key Points Y-T:
The conventional oscilloscope display format. It shows the voltage of a waveform record (on the vertical axis) as it varies over time (on the horizontal axis).
X-Y:
X-Y format displays channel 1 in the horizontal axis and channel 2 in the vertical axis.
Roll Mode:
In this mode, the waveform display rolls from right to left. No trigger or horizontal offset control of waveforms is available during Roll Mode, and it’s only available when set to 500 ms/div or slower.
Slow Scan Mode:
his mode is available when the horizontal time base is set to 50ms or slower. In this mode, the oscilloscope acquires sufficient data for the left part to the trigger point, then wait for tri
er, when trigger occurs, it continue to draw the rest part from the trigger point to the end of the right side. When choosing this mode to view low frequency signals, it is recommended that the channel coupling be set as DC.
Time/Div:
Horizontal scale. If the waveform acquisition is stopped (using the Time/Div control expands or compresses the waveform.
button), the
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-29
Delayed Scan
The Delayed Scan is a magnified portion of the main waveform window. Use Delayed Scan to locate and horizontally expand part of the main waveform window for a more
detailed (higher horizontal resolution) analysis of signal. The Delayed Scan time base
setting cannot be set slower than the main time base setting.
Main time base
Waveform to be horizontally expanded
Time base of Delayed Scan
Expanded waveform in horizontal
Figure 2-21
Delayed Scan
The following describes the Delayed Scan display.
The screen is divided into two parts. The upper half displays the main waveform window and the lower half displays an expanded portion of the main waveform window. This expanded portion of the main window is called the Delayed Scan window. Two blocks shadow the upper half; the un-shadowed portion is expanded in
the lower half. The horizontal
2-30
and knobs control the size and position of the
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
Delayed Scan. The symbol at bottom of the screen means the main time base and the symbol on the center bottom means the Delayed Scan time.
Delayed Scan shortcut key:
Delayed Scan function can be activated not only by the menu but also by click
.
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-31
X-Y Mode
This format is useful for studying phase relationships between two signals.
Channel 1 in the horizontal axis(X) and Channel 2 in the vertical axis(Y). The oscilloscope uses a none-trigger acquisition mode, data is displayed as dots.
Figure 2-22
The X-Y Display
The following modes or functions will not work in X-Y mode.
z LA Function (for VS5000D only) z Automatic Measurements z Cursor Measurements z REF and MATH Operations z Delayed Scan Mode z Vector Display Mode
z Horizontal
z Trigger Controls
2-32
Button
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series

To Set the Trigger System

The trigger determines when the oscilloscope starts to acquire data and display a waveform. When a trigger is set up properly, it can convert unstable displays or blank screens into meaningful waveforms.
When the oscilloscope starts to acquire a waveform, it collects enough data so that it can draw the waveform to the left of the trigger point. The oscilloscope continues to acquire data while waiting for the trigger condition to occur. After it detects a trigger, the oscilloscope continues to acquire enough data so that it can draw the waveform to the right of the trigger point.
In the H
Mode: Select the trigger mode. Setting: Set the trigger Source channel, trigger Level and trigger Slope mode.
orizontal of the main menu, there is Trigger menu include:
Sweep: Set the sweep mode to Auto, Normal or Single. Trigger setting: Set the trigger Coupling, adjust the trigger Sensitivity and
adjust the Holdoff Time.
Figure 2-23
Trigger Settings dialog
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-33
Trigger Modes
The oscilloscope provides seven trigger modes: Edge, Pulse, Slope, Video, Alternative, Pattern and Duration trigger (for VS5000D only).
Edge:
An edge trigger occurs when the trigger input passes through a specified voltage level in the specified slope direction.
Pulse:
Use this trigger type to catch pulses with certain pulse width.
Video:
Use video trigger on fields or lines for standard video signals.
Slope:
The oscilloscope begins to trigger according to the signal rising or falling speed.
Alternative:
Trigger on non-synchronized signals
Pattern:
To trigger through detecting specified codes.
Duration:
To trigger within a specified time on the conditions of a specified code.
2-34
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
Settings for Edge Trigger
An edge trigger determines whether the oscilloscope finds the trigger point on the rising or the falling edge of a signal. Select Edge trigger Mode to trigger on Rising edge or falling edge.
Table 2-13 Edge Trigger Settings
Function Setting Description
Mode
Source
Slope
Sweep
Set Up To go to the Set Up menu
CH1 CH2 EXT EXT/5 AC Line D15-D0
Rising Falling
Auto
Normal Single
Select the trigger mode to Edge
Select CH1 as trigger signal Select CH2 as trigger signal Select EXT TRIG as trigger signal Select attenuated EXT TRIG/5 as trigger signal Select power line as trigger signal Select a digital channel in D15-D0 as trigger source (for VS5000D only) Trigger on rising edge Trigger on falling edge
Acquire waveform even no trigger occurred Acquire waveform when trigger occurred. When trigger occurs, acquire one waveform then stop
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-35
Settings for Pulse Width Trigger
Pulse trigger occurs according to the width of a pulse. The abnormal signals can be detected through setting up the pulse width condition.
Table 2-14 Pulse Width Trigger Settings
Function Setting Description
CH1
CH2
EXT
Source
Level To set the trigger level
EXT/5
D15-D0
Select CH1 as trigger signal Select CH2 as trigger signal Select EXT TRIG as trigger signal Select attenuated EXT TRIG/5 as trigger signal Select a digital channel in D15-D0 as trigger source(for VS5000D only)
+More than
+Less than
When
Pulse Width Set required pulse width
Sweep
Set Up To go to the Set Up menu
NOTE: The Pulse width adjusts range is 20ns - 10s. When the rule is met, it will
trigger and acquire the waveform.
+Equal
-More than
-Less than
-Equal
Auto
Normal
Single
+Pulse width more than +Pulse width less than +Pulse width equal to
-Pulse width more than
-Pulse width less than
-Pulse width equal to
Acquire waveform even no trigger occurred Acquire waveform when trigger occurred. When trigger occurs, acquire one waveform then stop
2-36
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
Settings for Video Trigger
Choose video trigger to trigger on fields or lines of NTSC, PAL, or SECAM standard video signals. Trigger coupling preset to DC.
Table 2-15 Video Trigger Settings
Function Setting Description
CH1
Source
CH2 EXT EXT/5
Polarity
Normal polarity
Inverted polarity
Standard
NTSC PAL/SECAM
Set up To go to the Set Up menu
Select CH1 as trigger source Select CH2 as trigger source Select EXT TRIG as trigger source Select EXT TRIG/5 as trigger source
Triggers on negative going sync pulses
Triggers on positive going sync pulses
Select Video standard
Sync
All Lines Line Num
Odd field Even field
Trigger on all lines Trigger on an specified line
Select to trigger on odd field Select to trigger on even field
Key points Sync Pulses: When Normal Polarity is selected, the trigger always occurs
on negative-going sync pulses. If the video signal has positive-going sync
pulses, use the inverted Polarity.
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-37
Figure 2-24
Video trigger: Line Synchronization
2-38
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
Figure 2-25
Video trigger: Field Synchronization
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-39
Settings for Slope Trigger
Slope trigger sets the oscilloscope as the positive/negative slope trigger within the specified time.
Table 2-16 Slope Trigger Settings
Function Setting Description
CH1
Source
When
Time To set slope time
Level A
Level B
Set Up
NOTE: Slope time can be set from 20ns to 10s. When a signal meets the trigger
condition, scope will execute the acquisition. Level A/ Level B is adjustable.
CH2
EXT
EXT/5 +More than
+Less than +Equal
-More than
-Less than
-Equal
Set channel 1 as trigger source Set channel 2 as trigger source Set EXT channel as trigger source Set EXT/5 as trigger source
+Slope width more than +Slope width less than +Slope width equal to
-Slope width more than
-Slope width less than
-Slope width equal to
To set the Level A
To set the Level B
To go to the Set Up menu
2-40
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
Settings for Alternative Trigger
When alternative trigger is on, the trigger sources are from two vertical channels. This mode can be used to observe two non-related signals. Choose two different trigger modes for the two vertical channels. The options are as follows: Edge, Pulse, Slope and Video. The information of the trigger level of the two channels will be displayed on the upper-right of the screen.
Table 2-17 Alternative Trigger Settings
Function Setting Description
Channel
Trigger Typ e
Set Up
NOTE:
In Alternative Trigger, each channel is allowed to set the trigger mode independently.
The trigger setting of CH1 and CH2 are mutually independently. Please make
attention to the Channel options.
CH1
CH2
Edge/Pulse/ Slope/Video
Set trigger mode for Channel 1 Set trigger mode for Channel 2
Set Edge/Pulse/Slope/Video Trigger as the trigger type
To go to the Set Up menu
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-41
Settings for Pattern Trigger (for VS5000D only)
Pattern trigger identifies trigger terms by checking appointed code. The code is logical relationship of all channels, with High (H), Low (L), and Ignore (X), Rising and Falling.
Table 2-18 Pattern Trigger Settings
Function Setting Description
Defining D15-D0 Choose digital channel for Pattern trigger
H L
Pattern
Sweep
Set Up
Key Points: H (High): Logic high: Voltage is higher than threshold setting. L (Low): Logic low: Voltage is lower than threshold setting. X (Ignore): Don’t care. If all the channels are ignored, the oscilloscope won’t be
triggered.
Edge of appointed code
X Rising Falling
Auto Normal Single
High Low Ignore Rising Falling
Acquire waveform even no trigger occurred Acquire waveform when trigger occurred When trigger occurs, acquire one waveform and then stop
To go to the Set Up menu
Appoint one code as edge only. If an edge is appointed, then appointed another edge in a different channel and the first appointed edge will be set to X (Ignore).
2-42
User’s Guide for VS5000 Series
© 2008 RIGOL Technologies, Inc
Settings for Duration Trigger (for VS5000D only)
Trigger in appointed time when code terms are satisfied.
Table 2-19 Duration Trigger Settings
Function Setting Description
More than
When
Time Set duration and limit symbol time
Key Points: Qualifier: A timer begins when code terms are satisfied. Duration
trigger occurs in the time set by the qualifier.
Less than Equal
Set time limit terms
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-43
Trigger Settings
In the trigger set up menu, set up different trigger settings according to different trigger modes. When in the mode of Edge and Pulse, the source is D15-D0 (for VS5000D only), only Holdoff is adjustable. When source is non-digital channel and in slope trigger, only trigger coupling, trigger sensitivity and Holdoff can be set. For video trigger, Sensitivity and Holdoff can be set. For Pattern trigger and Duration trigger (for VS5000D only), only Holdoff can be set. When in Alternative trigger, make different settings according to selected different trigger modes.
Table 2-20 (Settings for trigger coupling, trigger sensitivity and holdoff)
Function Setting Description
Coupling
DC AC HF Reject LF Reject
Allow all signals pass Block DC signals Reject high frequency signals Reject DC and low frequency signals
Sensitivity
Holdoff Time
Set trigger sensitivity
Set time slot before another trigger event
Table 2-21 (Settings for sensitivity and holdoff)
Function Setting Description
Sensitivity
Holdoff
Set trigger sensitivity Set time slot before another trigger
event
Table 2-22 (Settings only for holdoff)
Function Setting Description
Holdoff
Set time slot before another trigger event
2-44
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
Trigger Holdoff
Use Trigger Holdoff to stabilize a complex waveform, such as a pulse range. Holdoff time is the oscilloscope’s waiting period before starting a new trigger. During Holdoff, oscilloscope will not trigger until Holdoff ends. For instance: To trigger on the first pulse on a group of them, users can set the holdoff time to Pulse cluster width.
Trigger Point
Time of Holdoff
Figure 2-26
Trigger Holdoff
To use trigger Holdoff:
1. Turn on the Trigger window, and go to the Trigger functions Menu.
2. Click the corresponding menu to set the trigger settings.
3. Click the left and the right arrow (
) button to change Holdoff time until
waveform is stable.
4. Click the circular button (
) could reset the Holdoff time to its default value.
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-45
Trigger Key Points
1. Trigger Source:
Trigger occurs from several sources: Input channels (CH1, CH2 and D15-D0 (for VS5000D only)), AC Line, Ext, Ext/5.
z CH1 or CH2: It is the most commonly used trigger source. The channel will work
when selected as a trigger source whatever displayed or not.
z Ext Trig: The instrument can be trigged from a third source while acquiring data
from CH1 and CH2. For example, to trigger from an external clock or with a signal from another part of the test circuit. The Ext, Ext/5 trigger sources use an external trigger signal connected to the Ext Trig connector. It has a trigger level range of +1.2 V to -1.2 V. The Ext/5 trigger source attenuates the signal by 5X, which extends the trigger level range to +6 V to -6 V, allowing the oscilloscope to trigger on a larger signal.
2. Sweep Mode:
The sweep mode determines how the oscilloscope behaves in the absence of a trigger event. The oscilloscope provides three trigger modes: Auto, Normal, and
Single.
z Auto: This sweep mode allows the oscilloscope to acquire waveforms even
when it does not detect a trigger condition. If no trigger condition occurs while the oscilloscope is waiting for a specific period (as determined by the time-base setting), it will force itself to trigger. When forcing invalid triggers, the oscilloscope cannot synchronize the waveform, and the waveform seems to roll across the display. If valid triggers occur, the display becomes stable on the screen. Any factor results in the un-stability of waveforms can be detected by Auto Trigger, such as the output of power supply.
NOTE: When horizontal control is set more than 50 ms/div, Auto mode allows the
oscilloscope to capture without trigger signal.
z Normal: The Normal mode allows the oscilloscope to acquire a waveform only
when it is triggered. If no trigger occurs, the oscilloscope keeps waiting, and the previous waveform, if any, will remain on the display.
z Single: In Single mode, after pressing the RUN/STOP key, the oscilloscope waits
for trigger. While the trigger occurs, the oscilloscope acquires one waveform and
2-46
User’s Guide for VS5000 Series
© 2008 RIGOL Technologies, Inc
then stop.
3. Coupling:
Trigger coupling determines which signal component passed to the trigger circuit.
Coupling types include AC, DC, LF Reject and HF Reject.
z AC: AC coupling blocks DC components. z DC: DC coupling passes both AC and DC components. z LF Reject: LF Reject coupling blocks DC component, and attenuates all signal
with a frequency lower than 8 kHz.
z HF Reject: HF Reject coupling attenuates all signals with a frequency higher
than 150 kHz.
4. Pre-trigger/delayed trigger:
The data collected before and after trigger. The trigger position is typically set at the horizontal center of the screen. In the full-screen display 6 divisions’ data of pre-trigger and delayed trigger can be surveyed. More data (14div) of pre-trigger and 1s delayed trigger can be surveyed by adjusting the horizontal scale and offset. This feature is very useful to study the events that led up to the trigger point. Everything to the right of the trigger point is called post-trigger information. The amount of delay range (pre-trigger and post-trigger information) depends on the sweep speed selected.
z Adjustable trigger sensitivity
To avoid the influence of noise from the environment and get the stable trigger, the trigger circuit has adopted stickiness. In VS5000 series, the stickiness is adjustable from 0.1div-1.0div, which means when it sets to 1.0div, the trigger circuit will not care any signal with peak-peak amplitude less than 1.0div, so as to avoid the influence of the noise.
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-47

To Set the Sampling System

The Acquire button in the Horizontal menu is used for sampling system setup.
Acquire setup button
Figure 2-27
Acquire System option
Click the Acquire button in the Horizontal menu, go to the Acquire window.
2-48
Figure 2-28
Acquire System setting
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
Table 2-23 Acquire System Settings
Function Setting Description
Sampling
Real Time Equal Time
Normal
Acquisition
Average Peak Detect
Averages 2 to 256
Memory
Normal Long
Real-time sampling mode Equivalent sampling mode
Normal Acquisition mode Average Acquisition mode Peak Detect Acquisition mode
Step by multiple of two. Set average times from 2 to 256
Set up memory as 1k or 2k Set up memory as 512k or 1M
The waveform displayed on the screen will change in conjunction with the setting of the Acquire menu.
Signal that contains noise, and without average sampling
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
Figure2-29
2-49
Figure 2-30
Signal after average sampling
NOTE:
z Select Real Time acquisition to observe the single-shot or pulse signals. z Select Equal Time to observe high frequency repetitive signals. z To reduce the displayed random noise, select the Average acquisition and this
mode would make the waveform update slower.
z To avoid signal aliasing, select Peak Detect acquisition.
2-50
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
Figure 2-31
Peak Detect
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-51
Stop Acquisition: When the scope is acquiring waveforms, the waveforms are in
live status; when acquisition is stopped, a frozen waveform will be displayed. The position and scale can still be adjusted by vertical control and horizontal control.
Key Points
Real-time Sampling:
VS5000 has Real-time sampling rate up to 400MSa/s. At the time base 50ns or faster, the oscilloscopes use the sine(x)/x interpolation to expand the horizontal time
base.
Equivalent sampling:
Known as Repetitive sampling to get up to 40 ps of horizontal resolution (equivalent 25Gsa/s). This mode is good for observing repetitive signals, and it is not
recommended for single-shot or pulse.
Normal:
Oscilloscope acquires signal by equal time interval.
Average Acquisition:
Apply averaging to signal to remove uncorrelated noise and improve measurement accuracy. Reduces random or uncorrelated noise in the signal display. The averaged waveform is a running average over a specified number of acquisitions from 2 to
256.
Peak Detect:
Peak Detect mode captures the maximum and minimum values of a signal. Finds highest and lowest record points over many acquisitions.
2-52
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series

To Save/Load and Print Setup

Click the S
torage button to get into the Save/Load and Print options.
Figure 2-32
The Storage Menu
Storage Menu
Click button Storage, then pop up a storage function menu. You can save and load
waveform in Chinese and English through the menu as follows:
Table 2-24 Storage Functions
Function Setting Description
Load
Save
Print
Load Waveform Load Setup Load Default Save Waveform Save Setup Save Screen
PrintPreview Print
Load the waveform that had saved Load the instrument that had saved Load the instrument default settings Save the current waveform Save the current instrument settings Save the PC software window as a .bmp or .jpg file
Preview the waveform Print setting and do the print operation
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-53
Save/Load Window
Click the Load/Save button in the Storage menu.
Table 2-25 The Save/Load Functions
Function Setting Description
Open/Save Open/Save the file that is selected
Select folder in the Folder window
Window comments
Cancel Cancel the current operation
Select file in the File window Input the file name in the File Name window Select the file type in the File Type
Keyboard
Use the soft-keyboard input the file name
An example with the Open File window:
Current driver
Current folder
Input the file name
File type
Files
2-54
Figure 2-33
Open a Waveform File
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
Load Default
The oscilloscope has default settings and can be recalled at anytime by user.
Files location
Specify the files location to save/load the waveform and setup.
Load
Recall saved waveform, setup and default settings.
Save
Save waveform, setups and screen.
NOTE:
Select Save Waveform stores the waveforms, and the current settings of the
oscilloscope
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-55
Print Settings
Set the printer to print the current waveform.
Figure 2-34
The Print Menu
Print button
Table 2-26 The Print Functions
Function Setting Description
PrintPreview
Print
Preview the waveform which current will be print
Print the current waveform
2-56
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
1. PrintPreview
Click the PrintPreview button to get into the Preview window.
Figure 2-35
Print Preview
In this window, use the Zoom In and the Zoom Out to change the size of the waveform graph. Click the Close button to turn this window off.
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-57
2. Print
Click the Print button to get into the Print Setup window.
Figure 2-36
Colour Setting
Click OK, then pop up a print setting dialog as follows:
Figure 2-37
The Print Interface
In the Printer option select the document to print, and then click OK to execute
printing operation.
2-58
User’s Guide for VS5000 Series
© 2008 RIGOL Technologies, Inc

To Set the Measurement Function

The Measure functions are in the Analyzer menu; to do Horizontal Measure and Vertical Measure, or to use shortcut buttons to perform measurement operations.
Figure 2-38
Choose the Measurement Menu
Menu explanation:
The oscilloscopes provide 20 parametric auto measurements (10 voltage and 10 time measurements):
Vpp, Vmax, Vmin, Vtop, Vbase, Vamp, Vavg, Vrms, Overshoot, Preshoot.
Frequency, Period, Rise Time, Fall Time, +Width, -Width, +Duty, -Duty, Delay1-2 Delay1-2
.
,
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-59
1Shortcuts for Horizontal Measurements
Figure 2-39 Table 2-27 Horizontal Measurement Functions
Function Setting Description
Period Measure Period of a waveform
Frequency Measure Frequency of a waveform
Rise time Measure Rise Time of a rising edge
Fall time Measure Fall Time of a falling edge
+Width Measure +Pulse Width of a pulse wave
-Width Measure –Pulse Width of a pulse wave
+Duty Measure +Duty Cycle of a pulse wave
-Duty Measure –Duty Cycle of a pulse wave
Delay1→2
Delay1→2
Measure the signals delay between two channels at the rising edge
Measure the signals delay between two channels at the falling edge
2-60
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
Drives the automatic measurement shortcut enter the main display interface, starts this automatic measurement.
Figure 2-40
Quickly Measurement
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-61
2Shortcuts for Vertical measurements
Figure 2-41 Table 2-28 Vertical Measurement Functions
Function Setting Description
Vmax Measure maximum voltage of a waveform
Vmin Measure minimum voltage of a waveform
Vpp Measure Peak-to-Peak voltage
Vtop Measure a top voltage of a waveform
Vbase Measure a base voltage of a waveform
Vamp Measure voltage between Vtop and Vbase
Vavg Measure average voltage of a waveform
Vrms
Overshoot
Preshoot
Measure Root Mean Square Voltage of a waveform
Measure overshoot in percentage of an edge
Measure preshoot in percentage of an edge
NOTE: The results of the automatic measurements will be displayed on the bottom
of the screen. Maximum 3 results could be displayed at the same time. When there is no room, the next new measurement result will make the previous results moving left, out of screen.
2-62
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series
Apply Automatic Measurement as the following steps:
1. Select the signal channel for measuring, select CH1 or CH2 according to the
signal of interest.
2. To see all the measurement values, click All Measure in the Analyze menu, 18
measurement parameters will be displayed on the screen.
3. Choose the measurement parameters: measure projects according to the
Horizontal measurement Vertical measurement under Analyze Also , you can drag the left icons to the waveform interface directly.
Get the measured value on the screen; select the parameters of interest by clicking the correspond button, and read the data on the bottom of the screen.
To Clear the measure values; click Clear Measure, and all auto measured values disappear from the screen.
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-63
V
The Automatic Measurement of Voltage Parameters
The VS5000 series oscilloscopes provide automatic voltage measurements: Vpp, Vmax, Vmin, Vavg, Vamp, Vrms, Vtop, Vbase, Overshoot and Preshoot. Figure 2-41 below shows a pulse with some of the voltage measurement points.
Overshoot
Vmax
Vpp
Vtop
amp
Vbase
Vmin
Preshoot
Figure 2-42
Definition of Voltage Parameters
Vpp: Peak-to-Peak Voltage. Vmax: The maximum amplitude. The most positive peak voltage measured over the
entire waveform.
Vmin: The minimum amplitude. The most negative peak voltage measured over the
entire waveform.
Vamp: Voltage between Vtop and Vbase of a waveform Vtop: Voltage of the waveform’s flat top, useful for square/pulse waveforms. Vbase: Voltage of the waveform’s flat base, useful for square/pulse waveforms. Overshoot: Defined as (Vmax-Vtop)/Vamp, useful for square and pulse waveforms. Preshoot: Defined as (Vmin-Vbase)/Vamp, useful for square and pulse waveforms. Average: The arithmetic mean over the entire waveform. Vrms: The true Root Mean Square voltage over the entire waveform.
2-64
User’s Guide for VS5000 Series
© 2008 RIGOL Technologies, Inc
The Automatic Measurement of Time Parameters
VS5000 series oscilloscopes provide automatic timing parameters measurements; Frequency, Period, Rise Time, Fall Time, +Width, -Width, Delay 1→2
, Delay 1→2 ,
+Duty and -Duty.
Figure 2-42 shows a pulse with some of the time measurement points.
Rise Time
Fall Time
90%
50%
10%
-Width+Width
Figure 2-43
Definition of Time Parameters
Rise Time: Time that the rising edge of the first pulse in the waveform takes to rise
from 10% to 90% of its amplitude.
Fall Time: Time that the falling edge of the first pulse in the waveform takes to
fall from 90% to 10% of its amplitude.
+Width: The width of the first positive pulse in 50% amplitude points.
-Width: The width of the first negative pulse in the 50% amplitude points. Delay 1→2 Delay 1→2 +Duty: +Duty Cycle, defined as +Width/Period
-Duty: -Duty Cycle, defined as -Width/Period
: The delay between the two channels at the rising edge. : The delay between the two channels at the falling edge.
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-65
To Measure with Cursor Measurements
Click the Cursor button in the Analyze menu to select the Cursor functions.
Cursor measurement
Figure 2-44
Choose the Menu of Cursor Measurement
The Cursor measurement has three modes: Auto Measure, Manual and Track.
Auto Measure:
This mode will take effect with Automatic Measurements. The instruments will display cursors while measuring parameters automatically. These cursors demonstrate the physical meanings of these measurements.
Manual:
In this mode, the screen displays two parallel cursors. Move the cursors to make custom voltage or time measurements of the signal. The values are displayed on the boxes below the menu. Before using cursors, make sure that to set the Signal Source as the channel for measuring.
Track:
In this mode, the screen displays two cross cursors. The cross cursor sets the position on the waveform automatically. Adjust the horizontal position of cursor on the waveform by click the oscilloscope displays the values of the coordinates on the boxes below the menu.
button or drag the cursor on the screen. The
NOTE: The Auto Measure mode for Cursor measuring will take no effect without
automatic measurements.
2-66
User’s Guide for VS5000 Series
© 2008 RIGOL Technologies, Inc
Comments of Cursor Measurements
1. Auto Cursor Mode:
Table 2-29 Auto Cursor Measurement
Function Setting Description
Display the cursors for the current
Mode Auto
The cursors position
automatic measuring. (See the following figure)
Figure 2-45
Auto Measure Mode of Cursor Measurement
There will be no cursor display if no parameter is chosen in the M
easure menu. The
oscilloscope could move cursor automatically to measure 20 parameters in the
easure menu.
M
© 2008 RIGOL Technologies, Inc.
User’s Guide for VS5000 Series
2-67
2. Manual Cursor Mode
Figure 2-46
Manual Measure Mode of Cursor Measurement
Table 2-30 Manual Cursor Measurement
Function Setting Description
Shown as vertical line to measure the horizontal parameters Shown as horizontal line to measure the vertical parameters
Typ e
X
Y
CH1
Source
CH2
MATH (1, 2, 3 and 4)
Select the measurement signal source. (LA is applicable for VS5000D only)
LA
In this mode, the oscilloscope measures the Y or X coordinate values of the cursors, and the increments between the two cursors.
Please do as the following steps:
2-68
© 2008 RIGOL Technologies, Inc
User’s Guide for VS5000 Series