RIGOL VB1080 Users Guide

User’s Guide RIGOL
User’s Guide RIGOL
VB1080 VSWR Bridge
Product Overview
VB1080 is used in combination with the RIGOL DSA series spectrum analyzer to measure
S11-related parameters (such as return loss, reflection coefficient and VSWR). VB1080
provides three N (Female) connectors as shown in the figure below. IN: Signal input terminal. Here the signal generator or the output terminal of the
OUT: Signal output terminal. Here the
power meter or the RF input terminal of the spectrum analyzer is connected.
DUT: Here the device under test is
connected.
Measurement Connection
Connect VB1080 to the spectrum analyzer
as shown in the figure on the right.
Connect the spectrum analyzer
Use 2 Dual N (Male) adaptors to
connect the output terminal of the
tracking generator and the RF input
terminal of the spectrum analyzer to the IN terminal and OUT terminal of
the VSWR bridge respectively.
Connect the device under test
Do not use cables or adaptors as far as possible to avoid additional reflection.
Connect the device under test
Typical Applications
Measurement of the S11-related parameters of the filter, amplifier, mixer, etc.
Resonant frequency and VSWR tests of the antenna.
Spectrum Analyzer
Adapter
VSWR Bridge
Specifications
Frequency
Frequency range 2 GHz to 8 GHz
Connector
Connector type N (Female) type
Adaptor dual N (Male) type
Impedance 50
Insertion Loss
IN to DUT <1 dB (typical)
Directivity
Typ. 20 dB
Min. 15 dB
Input Power
Maximum input power +27 dBm (0.5 W)
General Specifications
Dimensions
Weight
Operation temperature
Storage temperature
115 mm×41 mm×17.5 mm
256 mm×190 mm×43 mm (with package)
0.2kg
0.9kg (with package)
-20 to 50
-40 ℃ to 100
Jun. 2014
Jun. 2014
Loading...