Rigol RPL1116 User guide

RIGOL
User’s Guide
RPL1116
RPL1116 Active Logic Probe
© 2013 RIGOL Technologies, Inc. All R ights Reserved.

General Safe ty Summary

Connect and disconnect the probe properly. Observe all terminals ratings. Do not touch exposed connections and components after power on. Do not operate with suspected failures. Do not operate without covers. Do not operate in an explosive atmosphere. Do not operate in wet conditions. Keep product surface clean and dry. Pay attention to handling safety.
- 1 -

Product Overview

Being a high-performance active logic probe, RPL1116 connects the digital signals under test to the MSO1000Z series digital oscilloscope to realize the logic analyzer function.
The 16 digital channels (D0-D15) of RPL1116 are divided into two channel groups (D0-D7 and D8-D15) each of which includes signal interfaces and ground interfaces. All the channels are marked with different numbers on the label of the probe head to identify different channels. RPL1116 provides 16 signal leads and 16 ground leads to realize flexible connection of signals and reference ground. Users can calibrate the probe zero by using the auto-calibration function in the LA menu of the oscilloscope.
- 2 -

The Using Method of the Logic Probe

1 Connect RPL1116 to the oscilloscope: connect the active logic
head to the main cable input, and then connect the main cable outp ut to the digital signal input terminal at the front panel of the oscillos cope as shown in the figure below.
- 3 -
2 Connect the signals under test to RPL1116: users can connect
any number (≤16) of the signals under test to RPL1116 according to the test need. Note that the amplitude of the input signal should not exceed the maximum working voltage range of the probe. RPL1116 provides two connection methods to realize convenient and flexible detection.
Method one: users can connect the signals under test through
the probe leads separately. You can easily identify the corresponding channel of each signal by the channel label on the probe leads and the color information on the probe head as shown in Figure 1.
Note: If crosstalk or ground bounce occurs during use, it might
be caused when channels share a single ground lead. So you are recommended to add a ground wire to the signal lines of each channel and twist them.
Method two: on the basis of method one, you can connect a
grabber to each lead and connect it to the device under test as
- 4 -
shown in Figure 2.
Figure 1 Figure 2
3 Set the probe: press LA at the front panel of the oscilloscope to
enter the probe setting menu. Users can view and set the following parameters under this menu: threshold level (the threshold levels of D0-D7 and D8-D15 can be adjusted independently), waveform size (applicable to all the channels; wherein, item L is only available when the number of active channels is no more tha n 8), channel label, probe calibration and so on.
Note: When the probe is connected to the oscilloscope for the first
- 5 -
time or the temperature change is more than 5 degrees, you are recommended to calibrate the probe zero using the auto-calibration function in the LA menu and please disconnect all the connections to the RPL1116 input terminal during the calibration.
4 Function Check: after finishing the above operations, the signal
under test will be displayed on the corresponding digital channel on the oscilloscope screen. If no signal is displayed, please adjust the oscilloscope to select proper general settings (such as the trigger mode and timebase). If signal is still not displayed, please check the electric connection and parameter settings again or please try to use other probe (such as analog probe) to check the signal state of the test point.
- 6 -
Threshold range
±15 V
Input capacitance
About 8 pF

Probe Specifications

Input channels 16
Threshold accuracy ±100m V+3% of threshold setting Max input voltage ±40 V Max input dynamic range ±10 V + threshold setting Min voltage swing 500 mV Min detectable pulse width Input impedance 100 kΩ±2%
Cable length About 90 cm Lead length About 25 cm Operation environment Storage environment
10 ns
0~50, 0~80%RH
-20~60, 0~90%RH
- 7 -

Accessories

Item Description Quantity
1 Main Cable 1 2 Active Logic Head 1 3 Lead 32 (2×16) 4 Grabber 32 5 Chinese and English User’s Guide 1 6 RPL1116 Packing Box 1
- 8 -

Accessories Sketch Map

Main Cable Active Logic Head
Lead Grabber
- 9 -
ZN102 000 265 9

Contact Us

If you have any problem or requirement when using our products or this manual, please contact RIGOL Technologies, Inc. E-mailservice@rigol.com Websitewww.rigol.com
UGE26X02-1110 Aug. 2014