Rigol RP1000D Series, RP1050D, RP1025D User Manual

Page 1
RIGOL
User’s Guide
RP1000D Series High Voltage
Differential Probe
Feb. 2013
Page 2
Page 3
RIGOL
RP1000D User’s Guide I
Guaranty and Declaration
Copyright
© 2012 RIGOL Technologies, Inc. All Rights Reserved.
Trademark Information
RIGOL is a registered trademark of RIGOL Tech nologies, Inc.
Publication Number
UGE18102-1110
Notices
RIGOL products are protected by patent law in and outside of
P.R.C.
RIGOL reserves the right to modify or change parts of or all the
specifications and pricing policies at company’s sole decision.
Information in this publication replaces all previously
corresponding material.
RIGOL shall not be liable for losses caused by either incidental or
consequential in connection with the fur ni shing, use or performance of this manual as well as any information contained.
Any part of this document is forbidden to be copied or photocopied
or rearranged without prior written approval of RIGOL.
Product Certification
RIGOL guarantees this pro duct conforms to t he national and indus trial standards in China as well as the ISO9001:2008 s tandard and the ISO14001:2004 standard. Other international standard conformance certification is in progress.
Contact Us
If you have any problem or requirement when using our products, please contact RIGOL Technologies, Inc. or your local distributors, or visit: www.rigol.com.
Page 4
RIGOL
II RP1000D User’s Guide
General Safety Summary
Please review the following safety precautions carefully before putting the instrument into operation so as to avoid any personal injuries or damages to the instrume nt and any prod uct connected to it. To prevent potential hazards, please use the instrument only specified by this manual.
Ground The Instrument.
The instrument is grounded through the Protective Earth lead of the power cord. To avoid electric shock, it is essential to connect the earth terminal of power cord to the Protective Earth terminal before any inputs or outputs.
Observe All Terminal Ratings.
To avoid fire or shock hazard, observe all ratings and markers on the instrument and check your manual for more informat ion about ratings before connecting.
Do Not Operate Without Covers.
Do not operate the instrument with covers or panels removed.
Avoid Circuit or Wire Exposure.
Do not touch exposed junct ions and components when the unit is powered.
Do Not Operate With Suspected Failures.
If you suspect damage occurs to the instrument , have it inspected by qualified service personnel before further operations. Any maintenance, adjustment or replacement especially to circuits or accessories must be performed by RIGOL authorized personnel.
Page 5
RIGOL
RP1000D User’s Guide III
Keep Well Ventilation.
Inadequate ventilation may cause increasing of temperature or damages to the device. So please keep well ventilated and inspect the intake and fan regularly.
Do Not Operate in Wet Conditions.
In order to avoid short circuiting to the interior of the device or electric shock, please do not operate in a humid environment.
Do Not Operate in an Explosive Atmosphere.
In order to avoid damages to the device or personal injuries, it is important to operate the device away from an explosive atmosphere.
Keep Product Surfaces Clean and Dry.
To avoid the influence of dust and/or moisture in air, please keep the surface of device clean and dry.
Electrostatic Prevention.
Operate in an electrostatic discharge protective area environment to avoid damages induced by static discharges. Always ground both the internal and external conductors of the cable to release static before connecting.
Page 6
RIGOL
IV RP1000D User’s Guide
Safety Terms and Symbols
Terms in this Manual. These terms may appear in this manual:
WARNING
Warning statements indicate the conditions or practices that could result in injury or loss of life.
CAUTION
Caution statements indicate the conditions or practices that could result in damage to this product or other property.
Terms on the Product. These terms may appear on the Product:
DANGER
indicates an injury or hazard may immediately happen.
WARNING indicates an injury or hazard may be accessible potentially. CAUTION indicates a potential damage to the instrument or other
property might occur.
Symbols on the Product. These symbols may appear on the product:
Double
Insulation
Safety
Warning
Protective
Earth
Terminal
Chassis
Ground
Test
Ground
Page 7
RIGOL
RP1000D User’s Guide V
Contents
Guaranty and Declaration .......................................................... I
General Safety Summary .......................................................... II
Safety Terms and Symbols ...................................................... IV
RP1000D Overview .................................................................... 1
Basic Operations ........................................................................ 4
Cleaning and General Care ........................................................ 6
Warranty .................................................................................... 6
Specifications ............................................................................ 7
Technical Specifications ............................................................. 7
Operation Environment ............................................................ 10
General Specifications ............................................................. 10
Accessories .............................................................................. 11
Page 8
Page 9
RIGOL
RP1000D User’s Guide 1
RP1000D Overview
RP1000D series high voltage differential probe can convert high differential input voltage to low voltage and display the waveform on oscilloscope. Its working frequency is up to 25MHz (RP1025D), 50MHz (RP1050D) and 100MHz (RP1100D) and it is rather suitable for large electricity test and R&D.
RP1000D series high voltage differential probe is applicable to general purpose oscilloscope and the labeled attenuation ratios are those when the input impedance of the oscilloscope is 1MΩ. The attenuation ratios will double when the input impedance of the oscilloscope is 50Ω.
Figure 1 RP1025D High Voltage Differential Probe
BNC Interface
Input Terminal
Power Adaptor Interface
Attenuation Ratio and Power Button
Overload Indicator
Power Indicator
Page 10
RIGOL
2 RP1000D User’s Guide
Figure 2 RP1050D High Voltage Differential Probe
BNC Interface
Input Terminal
Power Adaptor Interface
Attenuation Ratio and
Power Button
Overload Indicator
Power Indicator
Page 11
RIGOL
RP1000D User’s Guide 3
Figure 3 RP1100D High Voltage Differential Probe
BNC Interface
Input Terminal
Power Adaptor Interface
Attenuation Ratio and Power Button
Overload Indicator
Power Indicator
Page 12
RIGOL
4 RP1000D User’s Guide
Basic Operations
1. Connect the red safety IC clip with one end of the r ed dual-banana plug silicon cable and the black safety IC clip with one end of the black dual-banana plug silicon cable provided in the accessories. Then, connect the red dual-banana plug silicon cable with the red (+) input terminal of the high voltage probe and the black dual-banana plug silicon cable with the black (-) input terminal of the high voltage probe.
Note: a) The safety IC clip can be replaced by the high v o l t age
dedicated IC clip, safety alligator clip or safety contact probe prod;
b) The dual-banana plug silicon cable can be replaced by high
voltage dedicated dual-banana plug silicon cable.
2. Connect one end of the dual-BNC coaxial cable to the BNC interface of the high voltage differential probe and the other end to input terminal of the oscilloscope.
3. Turn on the channel switch on the os cilloscope and adjust the hi gh voltage probe and oscilloscope to make the attenuation ratios of the two match. If the attenuation r atio of the oscilloscope does not match that of the high voltage probe, the actual vertical scale equals the attenuation ratio of the high voltage probe divided by the attenuation ratio of the oscilloscope and then times the vertical scale of the oscilloscope.
For example, when the attenuation ratio of the oscilloscope is set to 1X, the attenuation ratio of the high voltage prob e is set to X200
Page 13
RIGOL
RP1000D User’s Guide 5
and the vertical scale of the oscilloscope is 0.5V/div, the actual vertical scale is 200X0.5V/div=100V/div. When the input impedance of the oscilloscope is 50Ω, the actual vertical scale is 2X200X0.5V/div =200V/div.
Note: When the attenuation ratio of the oscilloscope matches the attenuation ratio of the high voltage probe, the vertical scale displayed on the oscilloscope is the actual scale.
Page 14
RIGOL
6 RP1000D User’s Guide
Cleaning and General Care
Cleaning:
This product has no particular requirement for cleaning. To clean the probe, please wipe the probe surface using soft and clean cloth dampened with detergent.
General Care:
Please store this product in anti-humidity case if the product will not be used for more than 60 days.
Warranty
RIGOL warrants that its products mainframe and accessories will be free from defects in materials and workmanship withi n the warranty period.
If a product is proven to be defective within the respective period, RIGOL guarantees the free replacement or repair of products which are approved defective. To get repair service, please contact with your nearest RIGOL sales and service office.
RIGOL does not provide any other warranty items except the one being provided by this summary and the warranty statement. The warranty items include but not being subjected to the hint guarantee items related to tradable characteristic and any particular purpose. RIGOL will not take any responsibility in cases regarding to indirect, particular and ensuing damage.
Page 15
RIGOL
RP1000D User’s Guide 7
Specifications
Technical Specifications
RP1025D:
Bandwidth
X50 or X200 attenuation ratio: DC - 25MHz
(-3dB)
X20 attenuation ratio: DC - 15MHz
Attenuation Ratio
X20, X50, X200
Accuracy
±2%
Input Voltage
Range (DC + AC peak-peak value)
X20 attenuation ratio: ≤ 140Vpp, about
45Vrms or DC X50 attenuation ratio: ≤ 350Vpp, about 110Vrms or DC X200 attenuation ratio: ≤ 1400Vpp, about
450Vrms or DC
Maximum Input
Voltage
Maximum differential voltage: 1400V (DC+AC
peak-peak value) or 450Vrms Voltage to ground at the input terminal:
600Vrms
Input Impedance
Differential: 4MΩ/1.2pF Single-ended and to ground: 2MΩ/2.3pF
Output Voltage
±6.5V
Output Impedance
50Ω
Rise Time
X50 or X200 attenuation ratio: 14ns X20 attenuation ratio: 23.4ns
Common-mode
Rejection
60Hz: > 80dB
100Hz: > 60dB
1MHz: > 50dB
Power Supply
Specified external 9V DC power supply (must be specified products acknowled ged b y
RIGOL
)
Power Consumption
0.4 watt
Page 16
RIGOL
8 RP1000D User’s Guide
RP1050D:
Bandwidth
X200, X500 or X1000 attenuation ratio: DC -
50MHz (-3dB)
X100 attenuation ratio: DC - 25MHz
Attenuation Ratio
X100, X200, X500, X1000
Accuracy
±2%
Input Voltage
Range (DC + AC peak-peak value)
X100 attenuation ratio: ≤ 700Vpp, about
230Vrms or DC X200 attenuation ratio: ≤ 1400Vpp, about 460Vrms or DC X500 attenuation ratio: ≤ 3500Vpp, about 1140Vrms or DC X1000 attenuation ratio: ≤ 7000Vpp,
2300Vrms or DC
Maximum Input
Voltage
Maximum differential voltage: 7000V (DC+AC
peak-peak vale) or 450Vrms Voltage to ground at the input terminal:
6500Vrms
Input Impedance
Differential: 100MΩ/1.2pF Single-ended and to ground: 50MΩ/2.3pF
Output Voltage
±7.0V
Output Impedance
50Ω
Rise Time
X200, X500 or X1000 attenuation ratio: 7ns X100 attenuation ratio: 14ns
Common-mode
Rejection
60Hz: > 80dB
100Hz: > 60dB
1MHz: > 50dB
Power Supply
Specified external 9V DC power supply (must
be specified products acknowled ged b y
RIGOL)
Power Consumption
0.4 watt
Page 17
RIGOL
RP1000D User’s Guide 9
RP1100D:
Bandwidth
X200, X500 or X1000 attenuation ratio: DC -
100MHz (-3dB)
X100 attenuation ratio: DC - 50MHz
Attenuation Ratio
X100, X200, X500, X1000
Accuracy
±2%
Input Voltage
Range (DC + AC peak-peak value)
X100 attenuation ratio: ≤ 700Vpp, about
230Vrms or DC X200 attenuation ratio: ≤ 1400Vpp, about 460Vrms or DC X500 attenuation ratio: ≤ 3500Vpp, about 1140Vrms or DC X1000 attenuation ratio: ≤ 7000Vpp, about
2300Vrms or DC
Maximum Input
Voltage
Maximum differential voltage: 7000V (DC+AC
peak-peak value) or 450Vrms Voltage to ground at the input terminal:
6500Vrms
Input Impedance
Differential: 100MΩ/1.2pF Single-ended and to ground: 50MΩ/2.3pF
Output Voltage
±7.0V
Output Impedance
50Ω
Rise Time
X200, X500 or X1000 attenuation ratio: 3.5ns X100 attenuation ratio: 7ns
Common-mode
Rejection
60Hz: > 80dB
100Hz: > 60dB
1MHz: > 50dB
Power Supply
Specified external 9V DC power supply (must
be specified products acknowled ged b y
RIGOL)
Power Consumption
0.4 watt
Page 18
RIGOL
10 RP1000D User’s Guide
Operation Environment
General Operation Storage
Temperature
+20℃ to +30℃ 0℃ to +50 -30℃ to +70℃
Humidity 70%RH 10% to 85%RH 10% to 90%RH
General Specifications
Probe Dimensions
RP1025D: about 214mm x 60mm x 35mm RP1050D: about 240mm x 8 5mm x 36mm RP1100D: about 240mm x 8 5mm x 36mm
Weight RP1025D: 280g
RP1050D: 280g RP1100D: 280g
Safety IEC 1010-1, CAT III, pollution degree 2 Electromagnetic Compatibility
Conform to EN50081-1 and 50082-1 standards
Maximum Voltage to Ground
RP1025D: 600Vrms RP1050D: 6500Vrms RP1100D: 6500Vrms
Using Environment Indoor environment Insulation Category Double insulation
Page 19
RIGOL
RP1000D User’s Guide 11
Accessories
RP1025D
Accessories Explanation Quantity
1.
User’s Guide Chinese & English
1
2.
AC power adaptor
---- 1
3.
Dual-BNC coaxial cable
50Ω impedance RG58C UL 100cm length
1
4.
Dual-banana plug silicon cable
UL 6KV 18AWG 60cm length
Red: 1; Black: 1
5.
Safety IC clip UL 1000V CAT III Red: 1; Black: 1
6.
Safety alligator clip
UL 1000V CAT II10A
Red: 1; Black: 1
RP1050D
Accessories Explanation Quantity
1. User’s Guide Chinese & English 1
2. AC power adaptor ---- 1
3.
Dual-BNC coaxial cable
50Ω impedance RG58C UL 100cm length
1
4.
High voltage dedicated dual-banana plug silicon cable
UL 20KV 18AWG 60cm length
Red: 1; Black: 1
5.
High voltage dedicated IC
clip
maximum 6500V (DC+AC p-p)
Red: 1; Black: 1
6. Safety alligator clip
UL 1000V CAT II10A
Red: 1; Black: 1
Page 20
RIGOL
12 RP1000D User’s Guide
RP1100D
Accessories Explanation Quantity
1. User’s Guide Chinese & English 1
2. AC power adaptor ---- 1
3.
Dual-BNC coaxial cable
50Ω impedance RG58C UL 100cm length
1
4.
High voltage dedicated dual-banana plug silicon cable
UL 20KV 18AWG 60cm length
Red: 1; Black: 1
5.
High voltage dedicated IC
clip
maximum 6500V (DC+AC p-p)
Red: 1; Black: 1
6. Safety alligator clip
UL 1000V CAT II10A
Red: 1; Black: 1
7. Safety contact probe prod,
UL 1000VCAT III
Red: 1; Black: 1