Warning
Do not attempt to take measurements from sources when
the chassis or return lead is not ground.
This ground connection is critical to the safety operation of
the probe. Failure to make this connection may result in
personal injuries or damages to the probe or voltmeter. This
connection must be made before the probe tip comes into
contact with the high voltage and must not be removed
until after the probe tip has been removed.
Do not connect the ground clip to the high voltage source or
the probe tip to the ground for any reason.
BEFORE turning the high voltage on, make sure that no
part of your body is in contact with the device.
Remembering that the voltage being measured is 1000
times greater than the voltmeter reading.
When the measuring voltage are DC 8 to 10 KV, or
AC rms 5 to 7 KV or AC peak 15 to 20 KV, do not
exceed 60 sec on each measurement. You can break off for
5 minutes at least during the measurement. Otherwise
overheating will affect the measurement accuracy even
damage the probe. Disconnect the probe tip from the high
voltage source BEFORE removing the ground clip lead.
Cleaning
最大工
作电压
信噪比
带宽
衰减比例
上升时间
输入阻抗
输入电容
补偿范围
线长度
温度系数
精确度
安全认证
操作温度
存储温度
重量/尺寸
!
当测量电压为下述情况时,每
次测量时间不得超过60秒。期
间需断开并等待散热至少5分
钟。
Max. Working Voltage
Signal/Noise
Bandwidth DC ~40 MHz (-3 dB)
Division Ratio 1:1000
Rise Time
Input Impedance
Input Capacitance
Compensation Range 5 p
Cable Length
Temperature Coefficient
Accuracy
Safety App. CE
Operating Temperature
Storage Temperature
Weight / Dimension
性能指标 (Specifications)
DC: 0 ~ 10 KV DC
AC: pulse ≤ 20 KV peak to peak
AC: sine wave ≤ 7
DC ≥ 60 dB (1 KHz),
≥ 50 dB (1 MHz)
≤ 8.8 ns
100 MΩ ± 5%
3.0 pF ±
2.0 meter (± 0.2 M)
≤ 200 PPM /℃
DC
0.5 p
F ~ 50 p
AC
± 3% (1 KHz/1 KV )
-3dB 40 MHz
0 ~ 50℃
-20 ~ +70℃
Φ
250 g /
When the measuring voltage is under
the followings. Each measurement
should be within 60 sec
off for 5 minutes at least during the
measurement.
75 x 340 L mm
KV rms
F
. You can break
User’s Guide
Clean only the exterior probe body and cables. Use a soft
cotton cloth lightly moistened with a mild solution of
detergent and water. Do not submerge any portion of the
probe at any time.
Dry the probe thoroughly before any voltage measurement.
Make sure the probe is free from any solvents or solvent
fumes as these can case deterioration to the probe body and
cables.
Find Quality Products Online at: sales@GlobalTestSupply.com
直流:8 ~ 10 KV
交流 有效值:5 ~ 7 KV
交流 峰峰值:15 ~ 20 KV
DC: 8 ~ 10 KV
AC RMS: 5 ~ 7 KV
AC PEAK: 15~ 20 KV
北京普源精电科技有限公司 RIGOL Technologies,Inc.
S M1 0 2 0 0 0 2 5 7 9
www.GlobalTestSupply.com
© 2013 RIGOL Technologies, Inc. All Rights Reserved.
RP1010H
High Voltage Probe for Oscilloscope