RIGOL RP1001C User Manual

RIGOL
User’s Guide
RP1001C Current Probe
Jul. 2014
RIGOL Technologies, Inc

Guaranty and Declaration

Copyright
© 2012 RIGOL Technologies, Inc. All Rights Reserved.
Trademark Information
RIGOL is a registered trademark of RIGOL Technologies, Inc.
Publication Number
UGE16106-1110
Notices
RIGOL products are covered by P.R.C. and foreign patents, issued and
pending.
RIGOL reserves the right to modify or change parts of or all the
specifications and pricing policies at company’s sole decision.
Information in this publication replaces all previously corresponding
material.
Information in this publication is subject to change without notice. RIGOL shall not be liable for losses caused by either incidental or
consequential in connection with the furnishing, use or performance of this manual as well as any information contained.
Any part of this document is forbidden to be copied or photocopied or
rearranged without prior written approval of RIGOL.
Product Certification
RIGOL guarantees this product conforms to the national and industrial standards in China as well as the ISO9001:2008 standard and the ISO14001:2004 standard. Other international standard conformance certification is in progress.
Contact Us
If you have any problem or requirement when using our products or this manual, please contact RIGOL. E-mail: service@rigol.com Websites: www.rigol.com
RP1001C User’s Guide I

General Safety Summary

Please review the following safety precautions carefully before putting
the instrument into operation so as to avoid any personal injuries or
damages to the instrument and any product connected to it. To
prevent potential hazards, please use the instrument only specified by
this manual.
Ground The Instrument.
The instrument is grounded through the Protective Earth lead of the
power cord. To avoid electric shock, it is essential to connect the earth
terminal of power cord to the Protective Earth terminal before any
inputs or outputs.
Observe All Terminal Ratings.
To avoid fire or shock hazard, observe all ratings and markers on the
instrument and check your manual for more information about ratings
before connecting.
Replace Fuse Properly.
Make sure that no overvoltage (such as voltage caused by
thunderbolt) this product. Otherwise, the operator might be exposed
to electric shock.
Do Not Operate Without Covers.
Do not operate the instrument with covers or panels removed.
Avoid Circuit or Wire Exposure.
Do not touch exposed junctions and components when the unit is
powered.
II RP1001C User’s Guide
Do Not Operate With Suspected Failures.
If you suspect damage occurs to the instrument, have it inspected by
qualified service personnel before further operations. Any maintenance,
adjustment or replacement especially to circuits or accessories must be
performed by RIGOL authorized personnel.
Keep Well Ventilation.
Inadequate ventilation may cause increasing of temperature or
damages to the device. So please keep well ventilated and inspect the
intake and fan regularly.
Do Not Operate in Wet Conditions.
In order to avoid short circuiting to the interior of the device or
electric shock, please do not operate in a humid environment.
Do Not Operate in an Explosive Atmosphere.
In order to avoid damages to the device or personal injuries, it is
important to operate the device away from an explosive atmosphere.
Keep Product Surfaces Clean and Dry.
To avoid the influence of dust and/or moisture in air, please keep the
surface of device clean and dry.
Electrostatic Prevention.
Operate in an electrostatic discharge protective area environment to
avoid damages induced by static discharges. Always ground both the
internal and external conductors of the cable to release static before
connecting.
RP1001C User’s Guide III

Double
Safety
Protective
Chassis
Test

Safety Terms and Symbols

Terms in this Manual. These terms may appear in this manual:
WARNING
Warning statements indicate the conditions or practices
that could result in injury or loss of life.
CAUTION
Caution statements indicate the conditions or practices that
could result in damage to this product or other property.
Terms on the Product. These terms may appear on the Product:
DANGER indicates an injury or hazard may immediately happen.
WARNING indicates an injury or hazard may be accessible
potentially.
CAUTION
Symbols on the Product. These symbols may appear on the
product:
indicates a potential damage to the instrument or other
property might occur.
Insulation
Warning
Earth
Terminal
Ground
Ground
IV RP1001C User’s Guide
Contents
Guaranty and Declaration ........................................................ I
General Safety Summary ........................................................ II
Safety Terms and Symbols .................................................... IV
RP1001C Overview .................................................................. 1
Basic Operation ....................................................................... 3
Connection between the Current Probe and Oscilloscope ............ 3
Connection between the Current Probe and Digital Meter ........... 5
Maintenance ............................................................................ 6
Battery .................................................................................... 6
To Install the Battery................................................................ 6
AC Power Adapter .................................................................... 7
Cleaning .................................................................................. 7
Preparation for shipment .......................................................... 8
Specifications .......................................................................... 9
Electrical Characteristics ........................................................... 9
Voltage and Current Ratings ................................................... 10
Physical Characteristics .......................................................... 10
Environmental Characteristics ................................................. 10
Certifications and Compliances ............................................... 11
Accessories ............................................................................ 12
Appendix ................................................................................ 13
Appendix 1 Gain versus Frequency at 1 A Peak (Typical) .......... 13
Appendix 2 Maximum Current versus Frequency ...................... 13
Appendix 3 DC Signal Linearity in 10mV/A Range (Typical) ....... 14
Appendix 4 Phase versus Frequency at 1 A Peak (Typical) ........ 14
RP1001C User’s Guide V
RIGOL

RP1001C Overview

The RP1001C current probe enables a general purpose oscilloscope to
display AC and DC current signals up to 100 A Peak (70 A R M S). The
RP1001C current probe can also make AC and DC measurements
with a multimeter by using the BNC-to-banana plug adapter.
Test C la m p
Test Clamp Button
BNC Interface
Maximum Conductor Size: 11mm
Battery Indicator
Overload Indicator
Range Switch
Zero Adjustment Knob
AC Adapter Input
Figure 1 RP1001C Current Probe
RP1001C User’s Guide 1
Control/Indicator Description
Current flow symbol: The arrow shows
the positive direction of current flow.
Zero adjustment: After connecting to
power and fixing the current probe to a fixed
position, rotate the knob to adjust the probe
output to the proper DC offset. It may also
be used to offset a DC signal component.
Zeroing is not needed for AC measurements
unless your instrument cannot isolate a DC
component (if present).
Range switch: Slide the switch from OFF to
either the 10 mV/A or 100mV/A range. When
10 mV/A or 100mV/A is selected, the probe
is turned on, and the green battery indicator
(ON) lights.
Battery indicator: The green
battery/power indicator lights when the
probe is turned on. For more information,
please refer to To Install the Battery.
Overload indicator: The red overload
indicator lights and flicks continuously if the
measured signal is greater than the
maximum range of the selected range
capacity.
2 RP1001C User ’s Guide

Basic Operation

Before using the current probe, the battery or specified AC power
adapter must be installed. For the battery installation instructions,
refer to To Install the Battery.
CAUTION
Do not clamp the probe onto circuits with voltages greater
than 600 VAC. Otherwise, personal injury or damage to the
probe may result.
CAUTION
Always connect the RP1001C current probe output to the
instrument before clamping onto the circuit under test.

Connection between the Current Probe and Oscilloscope

1 Set the coupling mode of the oscilloscope to DC and connect the
BNC connector of the RP1001C current probe to the oscilloscope
input via dual-BNC coaxial cable.
2 Move the Range switch to the 10 mV/A or 100 mV/A position to
turn on the RP1001C current probe. At this point, the green
battery indicator lights.
3 Use the zero adjustment knob to adjust the probe to the proper
DC offset.
Note: when measuring DC signals, accurate DC value is possible
only when the current probe is fixed at a certain position,
RP1001C User’s Guide 3
otherwise, the change of the position of the probe will cause the
DC drift. When measuring AC signals, users can move the current
probe.
4 Before connecting the current probe to circuit, open the test
clamp and clamp it onto the conductor as shown in the figure
below.
Figure 2 Connection between the Test Clamp and Conductor
5 Adjust the probe channel and oscilloscope as necessary to get a
clear and stable view of the signal. At this point, you can see both
the current components of AC and DC. Set the coupling mode of
the oscilloscope to AC to see the current components of AC. The
operation method of the current components of AC is the ratio of
the voltage amplitude measured through the oscilloscope and
the switch range currently selected of the current probe (10
mV/A or 100 mV/A).
Note: The current drawn by different devices look much different
than that of others. While the RMS current can only be used in
low frequency current, the momentary peaks may be quite high.
The figure on the next page shows the difference between the
line current drawn by a resistive load and a motor controller.
4 RP1001C User’s Guide
Figure 3 Typical Current Waveforms

Connection between the Current Probe and Digital Meter

Connect the BNC connector of the RP1001C current probe and the
input terminal of the digital meter using the BNC-to-banana plug
adapter provided with the accessories.
To measure only AC current, select the AC volts of the meter. To
measure DC current, select the DC volts of the meter, and adjust the
probe output to the proper DC offset. When connecting the current
under test, please note the current convention arrow on the test
clamp to get the proper polarity reading.
Note: To increase the measurement sensitivity of the RP1001C
current probe, loop additional turns of the wire under test through the
test clamp. The sensitivity of the RP1001C current probe is multiplied
times the number of loops in the test clamp, for example, 10 mV/A X
4 turns = 40 mV/A.
RP1001C User’s Guide 5

Maintenance

Battery

1 The RP1001C current probe uses a single square 9 V battery.
This instrument is a high power product. Please use the specified
alkaline battery.
2 As the battery in the RP1001C current probe is drained,
significant measurement errors may occur. The green battery
indicator will continue to light until a low battery voltage of 6.5 V
is reached.

To Install the Battery

1 Remove the test clamp from the circuit.
2 Remove the screw on the battery compartment cover at the rear
panel of the probe and then remove the battery compartment
cover.
3 While observing polarity, install the alkaline battery at the
corresponding location.
4 Install the battery compartment cover.
Battery Compartment
6 RP1001C User’s Guide

AC Power Adapter

1 The specified AC power adapter can be used to avoid
measurement error due to poor battery durability.
2 When using an AC power adapter for an extended time, you are
recommended to remove the battery. This is because heating will
result in battery leakage, and battery electrolyte will rust the
circuit board, thus creating major damage. Furthermore, as
batteries are high pollution products, you are recommended to
avoid using them as far as possible.

Cleaning

To clean the probe exterior, use a soft cloth dampened in a solution of
mild detergent and water. To clean the core, open the test clamp and
clean the exposed core surfaces with a cotton swap dampened with
isopropyl alcohol (isopropanol). Lubricate the clamp mating surfaces
with light oil.
Note: Do not clean with solvents or abrasives. Do not immerse the
probe.
RP1001C User’s Guide 7

Preparation for shipment

A special box is provided to be used for this product, convenient for
storage and shipment. Please do not discard it. If the original
packaging is unfit for use or not available, use the following
packaging guidelines.
1 Use a sturdy shipping carton having inside dimensions at least
one inch greater than the probe dimensions.
2 Put the probe into a plastic bag or wrap to protect it from
dampness.
3 Place the probe into the box and stabilize it with light packaging
material.
4 Seal the carton with shipping tape.
8 RP1001C User’s Guide

Specifications

These characteristics apply to an adjusted RP1001C current probe
installed on an oscilloscope of any brand. The oscilloscope must be
warmed up for at least 20 minutes and be in an environment with the temperature at 10-30 and the humidity at 0-80%.

Electrical Characteristics

Current Range 10mV/A, 100mV/A
DC Accuracy, typical 100mV/A: ±3% ±50mA
(50mA to 10A peak range)
10mV/A: ±4% ±50mA
(500mA to 40A peak range)
100mV/A: ±15% maximum
(40A peak to 100A peak range)
Gain versus
frequency, typical
Max. Working Current See Voltage and Current Ratings
Max. Working Voltage 300V CAT III, 600V CAT II, see Voltage and
Max. Floating Voltage See Voltage and Current Ratings
Frequency Range DC to 300kHz (-3dB)
Battery Type and Life,
typical
DC Signal Linearity,
typical
Phase versus
frequency, typical
See Appendix 1 Gain versus Frequency at
1 A Peak (Typical)
Current Ratings
9V NEDA 1604A, IEC 6LR61
13 hours (1 each)
See Appendix 3 DC Signal Linearity in
10mV/A Range (Typical)
See Appendix 4 Phase versus Frequency
at 1 A Peak (Typical)
RP1001C User’s Guide 9

Voltage and Current Ratings

Parameter Max. Working Current (A) Max.
10mV/A
Range
DC 100
DC+AC Peak 100
AC Peak 100 10 600 600
AC
Peak-peak
RMS CAT III 70.7 7.0 7 300 300
RMS CAT II 70.7 7.07 600 600
RMS CAT I 70.7 7.0 7 600 600
[1]
Note
Current versus Frequency.
: For frequency derating, please refer to Appendix 2 Maximum
[1]
10 600 600
[1]
10 600 600
200 20 1200 --
100mV/A
Range
Working
Voltage
(V)

Physical Characteristics

Dimensions 280mm x 70mm x 32mm
Max. Conductor Size 11mm
Dual-BNC Coaxial Cable
Length
Weight 260g (battery excluded)
100cm

Environmental Characteristics

Working Temperature
Storage Temperature
Humidity
Pollution Degree 2
0 to +50
-20 to +80
0 to +40, 95% humidity
+40 to +50, 45% humidity
Max.
Floating
Voltage
(V)
10 RP1001C User’s Guide

Certifications and Compliances

EC Declaration
Low Voltage Directive 73/23/EEC, as amended by
current clamps for electrical measurement and test.
Additional
IEC61010-1/A2:1995: Safety requirements for electrical
current clamps for electrical measurement and test.
Installation
CAT III: Equipment in a fixed industrial location (usually
circuits of electronic equipment.
Pollution
A measure of the contaminates that could occur in the
No pollution or only dry, nonconductive pollution occurs.
occurs only when the product is out of service.
of Conformity – Low Voltage
Compliance
(Overvoltage) Category
Degree
93/68/EEC Low Voltage Directive 73/23/EEC, as amended by 93/68/EEC EN 61010-1/A2:1995: Safety requirements for electrical
equipment for measurement, control and laboratory use
- Part 1: General requirements.
EN 61010-2-032:1995: Safety requirements for electrical
equipment for measurement, control and laboratory use
- Part 2-032: Particular requirements for hand-held
equipment for measurement, control and laboratory use
- Part 1: General requirements.
IEC61010-2-032:1994: Safety requirements for electrical
equipment for measurement, control, and laboratory use
- Part 2-032: Particular requirements for hand-held
permanently connected) CAT II: Equipment at this level includes appliances, portable tools, and similar products. Equipment is usually cord-connected. CAT I: Secondary (signal level) or battery operated
environment around and within a product. Products should be used only in the environment for which they are rated. Pollution Degree 1:
Products in this category are generally encapsulated, hermetically sealed, or located in clean rooms. Pollution Degree 2: Normally only dry, nonconductive pollution occurs. Occasionally a temporary conductivity that is caused by condensation must be expected. This location is a typical office/home environment. Temporary condensation
RP1001C User’s Guide 11
AC

Accessories

Power Adapter
Dual-BNC Coaxial Cable
RP1001C
BNC-to-banana Plug Adapter
1 A Chinese&English User’s Guide
2 A dual-BNC coaxial cable
3 A BNC-to-banana plug adapter
4 An AC power adapter that accords with the
standard of the destination country
Chinese&English User’s Guide
12 RP1001C User’s Guide

Appendix

Appendix 1 Gain versus Frequency at 1 A Peak (Typical)

Appendix 2 Maximum Current versus Frequency

RP1001C User’s Guide 13

Appendix 3 DC Signal Linearity in 10mV/A Range (Typical)

Appendix 4 Phase versus Frequency at 1 A Peak (Typical)

14 RP1001C User’s Guide
Loading...