RIGOL is a registered trademark of RIGOL Technologies, Inc.
Publication Number
UGE16104-1110
Notices
RIGOL products are protected by patent law in and outside
of P.R.C.
RIGOL reserves the right to modify or change parts of or all
the specifications and pricing policies at company’s sole
decision.
Information in this publication replaces all previously
corresponding material.
RIGOL shall not be liable for losses caused by either
incidental or consequential in connection with the furnishing,
use or performance of this manual as well as any information
contained.
Any part of this document is forbidden to be copied or
photocopied or rearranged without prior written approval of
RIGOL.
Product Certification
RIGOL guarantees this product conforms to the national and
industrial standards in China as well as the ISO9001:2008
standard and the ISO14001:2004 standard. Other international
standard conformance certification is in progress.
Contact Us
If you have any problem or requirement when using our products,
please contact RIGOL Technologies, Inc. or your local distributors,
or visit: www.rigol.com.
RP1001C User’s Guide I
RIGOL
General Safety Summary
Please review the following safety precautions carefully before
putting the instrument into operation so as to avoid any personal
injuries or damages to the instrument and any product connected
to it. To prevent potential hazar ds, pl eas e use the i nstrum ent o nly
specified by this manual.
Ground The Instrument.
The instrument is grounded through the Protective Earth lead of
the power cord. To avoid electric shock, it is essential to connect
the earth terminal of power cord to the Protective Earth terminal
before any inputs or outputs.
Observe All Terminal Ratings.
To avoid fire or shock hazard, observe all ratings and markers on
the instrument and check your manual for more inf ormation about
ratings before connecting.
Replace Fuse Properly
Make sure that no overvoltage (such as voltage caused by
thunderbolt) this product. Otherwise, the operator might be
exposed to electric shock.
Do Not Operate Without Covers.
Do not operate the instrument with covers or panels removed.
Avoid Circuit or Wire Exposure.
Do not touch exposed junctions and components when the unit is
powered.
II RP1001C User’s Guide
RIGOL
Do Not Operate With Suspected Failures.
If you suspect damage occurs to the instrument, have it inspected
by qualif ied servic e p er so n nel before further operations. Any
maintenance, adjustment or replacement especially to circuits or
accessories must be performed by RIGOL authorized personnel.
Keep Well Ventilation.
Inadequate ventilation may cause increasing of temperature or
damages to the device. So please keep well ventilated and inspect
the intake and fan regularly.
Do Not Operate in Wet Conditions.
In order to avoid short circuiting to the interior of the device or
electric shock, please do not operate in a humid environment.
Do Not Operate in an Explosive Atmosphere.
In order to avoid damages to the device or personal injuries, it is
important to operate the device away from an explosive
atmosphere.
Keep Product Surfaces Clean and Dry.
To avoid the influence of dust and/or moisture in air, please keep
the surface of device clean and dry.
Electrostatic Prevention.
Operate in an electrostatic discharge protective area environment
to avoid damages induced by static discharges. Always ground
both the internal and external conductors of the cable to release
static before connecting.
RP1001C User’s Guide III
RIGOL
DANGER
Double
Safety
Protective
Chassis
Test
Safety Terms and Symbols
Terms in this Manual. These terms may appear in this manual:
WARNING
Warning statements indicate the conditions or practices
that could result in injury or loss of life.
CAUTION
Caution statements indicate the conditions or practices
that could result in damage to this product or other
property.
Terms on the Product. These terms may appear on the
Product:
indicates an injury or hazard may immediately
happen.
WARNING indicates an injury or hazard may be accessible
potentially.
CAUTIONindicates a potential damage to the instrument or
other property might occur.
Symbols on the Product. These symbols may appear on the
product:
Insulation
Warning
IV RP1001C User’s Guide
Earth
Terminal
Ground
Ground
RIGOL
Contents
Guaranty and Declaration ................................................... I
General Safety Summary ................................................... II
Safety Terms and Symbols ............................................... IV
Appendix 1 Gain versus Frequency at 1 A Peak (Typical) ..... 13
Appendix 2 Maximum Current versus Frequency ................. 14
Appendix 3 DC Signal Linearity in 10mV/A Range (Typical) .. 14
Appendix 4 Phase versus Frequency at 1 A Peak (Typical) ... 15
RP1001C User’s Guide V
RIGOL
RP1001C Overview
The RP1001C current probe enables a general purpose
oscilloscope to display AC and DC current signals up to 100 A Peak
(70 A RMS). The RP1001C current probe can also make AC and
DC measurements with a multimeter by using the BNC-to-banana
plug adapter.
Test Clamp
Test Clamp Button
BNC Interface
Maximum Conductor Size: 11mm
Battery Indicator
Overload Indicator
Range Switch
Zero Adjustment Knob
AC Adapter Input
Figure 1 RP1001C Current Probe
RP1001C User’s Guide 1
RIGOL
Overload indicator:
Control/Indicator Description
Current flow symbol: The arrow shows
the positive direction of current flow
.
Zero adjustment: After connecting to
power and fixing the current probe to a
fixed position, rotate the knob to adjust
the probe output tothe pr oper DC off set. It
may also be used to offset a DC signal
component. Zeroing is not needed for AC
measurements unless your instrument
cannot isolate a DC component (if
present).
Range switch: Slide the switch from OFF
to either the 10 mV/A or 100mV/A range.
When 10 mV/A or 100mV/A is selected,
the probe is turned on, and the green
battery indicator (ON) lights..
Battery indicator: The green
battery/power indicator lights when the
probe is turned on. For more information,
please refer to To Install the Battery.
The red overload
indicator lights and flicks continuously if
the measured signal is greater than the
maximum range of the selected range
capacity.
2 RP1001C User’s Guide
RIGOL
Basic Operation
Before using the current probe, the battery or specified AC power
adapter must be installed. For the battery installation instructions,
refer to To Install the Battery.
CAUTION
Do not clamp the probe onto circuits with voltages
greater than 600 VAC. Otherwise, personal injury or
damage to the probe may result.
CAUTION
Always connect the RP1001C current probe output to the
instrument before clamping onto the circuit under test.
Connection between the Current Probe and
Oscilloscope
1 Set the coupling mode of the oscilloscope to DC and connect
the BNC connector of the RP1001C current probe to the
oscilloscope input via dual-BNC coaxial cable.
2 Move the Range switch to the 10 mV/A or 100 mV/A position
to turn on the RP1001C current probe. At this point, the green
battery indicator lights.
3 Use the zero adjustment knob to adjust the probe to the
proper DC offset.
RP1001C User’s Guide 1
RIGOL
Note: when measuring DC signals, accurate DC value is
possible only when the current probe is fixed at a certain
position, otherwise, the change of the position of the probe
will cause the DC drift. When measuring AC signals, users can
move the current probe.
4 Before connecting the current probe to circuit, open the test
clamp and clamp it onto the conductor as shown i n the figure
below.
Figure 2 Connection between the Test Clamp and Conductor
2 RP1001C User’s Guide
RIGOL
5 Adjust the probe channel and oscilloscope as necessary to get
a clear and stable view of the signal. At this point, you can see
both the current components of AC and DC. Set the coupli ng
mode of the oscilloscope to AC to see the current components
of AC. The operation method of the current components of AC
is the ratio of the voltage amplitude measured thr ough the
oscilloscope and the switch range currently selected of the
current probe (10 mV/A or 100 mV/A).
Note: The current drawn by different devices look much
different than that of others. While the RMS current can only
be used in low frequency current, the momentary peaks may
be quite high. The figure on the next page shows the
difference between the line current drawn by a resistive load
and a motor controller.
Figure 3 Typical Current Waveforms
RP1001C User’s Guide 3
RIGOL
Connection between the Current Probe and Digital
Meter
Connect the BNC connector of the RP1001C current probe and the
input terminal of the digital meter using the BNC-to-banana plug
adapter provided with the accessories.
To measure only AC current, select the AC volts of the meter. To
measure DC current, select the DC volts of the meter, and adjust
the probe output to the proper DC offset. When connecting the
current under test, please note the current convention arrow on
the test clamp to get the proper polarity reading.
Note: To increase the measurement sensitivity of the RP1001C
current probe, loop additional turns of the wire under test through
the test clamp. The sensitivity of the RP1001C current probe is
multiplied times the number of loops in the test clamp, for
example, 10 mV/A X 4 turns = 40 mV/A.
4 RP1001C User’s Guide
RIGOL
Maintenance
Battery
1 The RP1001C current probe uses a single square 9 V battery.
This instrument is a high power product. Please use the
specified alkaline battery.
2 As the battery in the RP1001C current probe is drained,
significant measurement errors may occur. The green battery
indicator will continue to light until a low battery voltage of
6.5 V is reached.
To Install the Battery
1 Remove the test clamp from the circuit.
2 Remove the three screws on the rear cover and take off the
rear cover . Y ou can see the location of the battery and remove
the battery.
3 While observing polarity, install the new
corresponding location.
4 Install the rear cover.
alkaline battery at the
Battery
RP1001C User’s Guide 5
RIGOL
AC Power Adapter
1 The specified AC power adapter can be used to avoid
measurement error due to poor battery durability.
2 When using an AC power adapter for an extended time, you
are recommended to remove the battery. This is because
heating will result in battery leakage, and battery electrolyte
will rust the circuit board, thus creating major damage.
Furthermore, as batteries are high pollution products, you are
recommended to avoid using them as far as possible.
Cleaning
To clean the probe exterior, use a soft cloth dampened in a
solution of mild detergent and water. To clean the core, open the
test clamp and clean t he exposed core surfaces with a cotton swap
dampened with isopropyl alcohol (isopropanol). Lubricate the
clamp mating surfaces with light oil.
Note: Do not clean with solvents or abr asives. Do not immerse the
probe.
6 RP1001C User’s Guide
RIGOL
Preparation for shipment
A special box is provided to be used for this product, convenient
for storage and shipment. Please do not discard it. If the original
packaging is unfit for use or not available, use the following
packaging guidelines.
1 Use a stur dy shipping carton ha ving inside dimensions at least
one inch greater than the probe dimensions.
2 Put the probe into a plastic bag or wrap to protect it from
dampness.
3 Place the probe into the box and stabilize it with light
packaging material.
4 Seal the carton with shipping tape.
RP1001C User’s Guide 7
RIGOL
Current Range
DC Accuracy,
Gain versus
Appendix 1 Gain versus
Max. Floating
Voltage and Current Ratings
Battery Type and
Specifications
These characteristics apply to an adjusted RP1001C current probe
installed on an oscilloscope of any brand. The oscilloscope must be
warmed up for at least 20 minutes and be in an environment with
the temperature at 10℃-30℃ and the humidity at 0-80%.
Electrical Characteristics
10mV/A, 100mV/A
100mV/A: ±3% ±50mA
typical
(50mA to 10A peak range)
10mV/A: ±4% ±50mA
(500mA to 40A peak range)
100mV/A: ±15% maximum
(40A peak to 100A peak range)
See
frequency, typical
Max. Working
Current
Max. Working
Voltage
Voltage
Frequency Range DC to 300kHz (-3dB)
Life, typical
8 RP1001C User’s Guide
Frequency at 1 A Peak (Typical)
See Voltage and Current Ratings
See Voltage and Current Ratings
See
9V NEDA 1604A, IEC 6LR61
13 hours (1 each)
RIGOL
Phase versus
Appendix 4 Phase versus
DC
[1]
[1]
AC Peak
AC
RMS CAT
DC Signal Linearity,
typical
See Appendix 3 DC Signal Linearity
in 10mV/A Range (Typical)
See
frequency, typical
Frequency at 1 A Peak (Typical)
Voltage and Current Ratings
Parameter Max. Working Current
(A)
DC+AC
10mV/A
Range
100
10 600 600
100
10 600 600
100mV/A
Range
Peak
100 10 600 600
200 20 1200 --
Peak-peak
70.7 7.07 300 300
III
RMS CAT II 70.7 7.07 600 600
RMS CAT I 70.7 7.07 600 600
[1]
Note
: For frequency derating, please refer to Appendix 2
Maximum Current versus Frequency.
Max.
Working
Voltage
(V)
Max.
Floating
Voltage
(V)
RP1001C User’s Guide 9
RIGOL
Dimensions
280mm x 70mm x 32mm
Max. Conductor Size
11mm
Dual-BNC Coaxial Cable Length
100cm
Weight
Working
Temperature
0℃ to +50℃
Storage
Temperature
-20℃ to +80℃
Humidity
0℃ to +40℃, 95% humidity
Pollution
Degree
2
Low Voltage Directive 73/23/EEC, as
clamps for electrical measurement and test.
IEC61010-1/A2:1995: Safety requirements
Physical Characteristics
Environmental Characteristics
+40℃ to +50℃, 45% humidity
Certifications and Compliances
EC Declaration
of Conformity
– Low Voltage
amended by 93/68/EEC
Low Voltage Directive 73/23/EEC, as
amended by 93/68/EEC
EN 61010-1/A2:1995: Safety requirements
for electrical equipment for measurement,
control and laboratory use - Part 1: Gener al
requirements.
EN 61010-2-032:1995: Safety requirements
for electrical equipment for measurement,
control and laboratory use - Part 2-032:
Particular requirements for hand-held current
260g (battery excluded)
Additional
10 RP1001C User’s Guide
Compliance
for electrical equipment for measurement,
clamps for electrical measurement and test.
CAT III: Equipment in a fixed industrial
operated circuits of electronic equipment.
Installation
(Overvoltage)
Category
RIGOL
control and laboratory use - Part 1: Gener al
requirements
.
IEC61010-2-032:1994: Safety requirements
for electrical equipment for measurement,
control, and laboratory use - Part 2-032:
Particular requirements for hand-held current
location (usually permanently connected)
CAT II: Equipment at this level includes
appliances, portable tools, and similar
products. Equipment is usually
cord-connected.
CAT I: Secondary (signal level) or battery
Pollution
Degree
A measure of the contaminates that could
occur in the environment around and within a
product. Products should be used only in the
environment for which they are rated.
Pollution Degree 1:
No pollution or only dry, nonconductive
pollution occurs. Products in this category are
generally encapsulated, hermetically sealed,
or located in clean rooms.
Pollution Degree 2:
Normally only dry, nonconductive pollution
occurs. Occasionally a temporary conductivity
that is caused by condensation must be
expected. This location is a typical
office/home environment. Temporary
condensation occurs only when the pro duct is
out of service.
RP1001C User’s Guide 11
RIGOL
RP1001C
Accessories
BNC-to-banana Plug Adapter
Dual-BNC Coaxial Cable
AC Power Adapter
1 An English User’s Guide
2 A dual-BNC coaxial cable
3 A BNC-to-banana plug adapter
4 An AC power adapter that accords with the standard of the
destination country
12 RP1001C User’s Guide
Appendix
Appendix 1 Gain versus Frequency at 1 A Peak
(Typical)
RIGOL
RP1001C User’s Guide 13
RIGOL
Appendix 2 Maximum Current versus Frequency
Appendix 3 DC Signal Linearity in 10mV/A Range
(Typical)
14 RP1001C User’s Guide
RIGOL
Appendix 4 Phase versus Frequency at 1 A Peak
(Typical)
RP1001C User’s Guide 15
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.