Arbitrary waveform function with
innovative SiFi technology
Multiple analog and digital modulations
Up to 160 built-in waveforms
Sweep function
Standard harmonic generator
Channels and system setting
Waveform summing function
In line with LXI Core 2011 DeviceFile Management Function
Standard 7 digits/s full function frequency
counter with 200MHz bandwidth
2
RIGOL
Page 4
Specications
All the specications can be guaranteed if the following two conditions are met unless where noted.
· The generator is within the calibration period and has performed self-calibration.
· The generator has been working continuously for at least 30 minutes under the specied temperature (18℃ ~ 28℃).
All the specications are guaranteed unless those marked with “typical”.
Frequency Characteristics
Sine1 μHz to 25 MHz 1 μHz to 30 MHz1 μHz to 60 MHz
Square1 μHz to 25 MHz 1 μHz to 25 MHz1 μHz to 25 MHz
Ramp1 μHz to 500 kHz
Pulse1 μHz to 15 MHz 1 μHz to 15 MHz1 μHz to 25 MHz
Harmonic1uHz to 10 MHz 1 μHz to 10 MHz1uHz to 20 MHz
Noise (-3dB)25 MHz bandwidth 30 MHz bandwidth60 MHz bandwidth
Arbitrary Waveform1 μHz to 10 MHz 1 μHz to 10 MHz1 μHz to 20 MHz
Resolution1 μHz
Accuracy±1 ppm of the setting value, 18℃to 28
160 kinds, including Sinc, Exponential Rise, Exponential Fall, ECG, Gauss, HaverSine, Lorentz,
Dual-Tone, etc.
30 MHz60 MHz
1 μHz to 500 kHz
℃
1 μHz to 1 MHz
Sine Wave Spectrum Purity
Typical (0 dBm)
Harmonic Distortion
Total Harmonic Distortion<0.075% (10 Hz to 20 kHz, 0 dBm)
Spurious (non-harmonic)
Phase Noise
Signal Characteristics
Square
Rise/Fall Time
Overshoot
Duty Cycle0.01% to 99.99% (limited by the current frequency setting)
Non-symmetry1% of the period + 5 ns
Jitter (rms)
Ramp
Linearity≤1% of peak output (typical, 1 kHz, 1 VPP, 100% symmetry)
Symmetry0% to 100%
Pulse
Pulse Width16ns to 999.999 982 118ks (limited by the current frequency setting)
Duty Cycle0.001% to 99.999% (limited by the current frequency setting)
Rising/Falling Edge≥10 ns (limited by the current frequency setting and pulse width setting)
Harmonic Order≤8
Harmonic TypeEven Harmonic, Odd harmonic, Order Harmonic, User
Harmonic AmplitudeThe amplitude of each order of harmonic can be set
Harmonic PhaseThe phase of each order of harmonic can be set
Output Characteristics
Amplitude (into 50 Ω)
Range
Accuracy
Flatness
UnitVpp, Vrms, dBm
Resolution0.1mVpp or 4 digits
Offset (into 50 Ω)
Range (Peak ac+dc)±5Vpk ac+dc
Accuracy±(1% of the setting value + 5mV + 0.5% of the amplitude)
Waveform Output
Output Impedance50 Ω (typical)
ProtectionShort-circuit protection, automatically disable the waveform output when overload occurs
Typical (1 Vpp)
<10 ns
Typical (1 Vpp)
≤5 MHz: 2 ppm + 200 ps
> 5 MHz: 200 ps
≤10 MHz: 1.0 mVpp to 10 Vpp
≤30 MHz: 1.0 mVpp to 5.0 Vpp
≤60 MHz: 1.0 mVpp to 2.5 Vpp
Typical (1 kHz sine, 0 V offset, >10 mVpp, auto)
±(1% of the setting value) ±1 mV
Modulating WaveformSquare with 50% duty cycle
Key Frequency2 mHz to 1 MHz
PWM
Carrier WaveformPulse
SourceInternal/External
Modulating WaveformSine, Square, Ramp, Noise, Arb
Width Deviation0% to 100% of the pulse width
Modulating Frequency2 mHz to 1 MHz
External Modulation Input
Input Range75 mVRMS to ±5 Vac + dc
Input Bandwidth50 kHz
Input Impedance10KΩ
Burst Characteristics
Carrier WaveformSine, Square, Ramp, Pulse, Noise, Arb (except DC)
Carrier Frequency2mHz to 25MHz 2mHz to 30MHz 2 mHz to 60 MHz
Burst Count1 to 1,000,000 or Innite
Start/Stop Phase0° to 360°, 0.1° resolution
Internal Period1 μs to 500 s
Gated SourceExternal Trigger
Trigger SourceInternal, External or Manual
Trigger Delay0 ns to 100 s
Sweep Characteristics
Carrier WaveformSine, Square, Ramp, Arb (except DC)
TypeLinear, Log or Step
DirectionUp or Down
Start/Stop FrequencyThe same with the upper/lower limit of the corresponding carrier frequency
Sweep Time1 ms to 500 s
Hold/Return Time0 ms to 500 s
Trigger SourceInternal, External or Manual
MarkerFalling edge of the sync signal (programmable)
Frequency Counter
FunctionFrequency, Period, Positive/Negative Pulse Width, Duty Cycle
Frequency Resolution7 digits/second (Gate Time = 1s)
Frequency Range1 μHz to 200 MHz
Period MeasurementMeasurement Range 5ns to 16 days
Voltage Range and Sensitivity (non-modulating signal)
DC Offset Range±1.5 Vdc
DC Coupling
AC Coupling
Pulse Width and Duty Cycle Measurement
Frequency and Amplitude
Ranges
Pulse Width
Duty CycleMeasurement Range (display)0% to 100%
Input Characteristics
Input Signal RangeBreakdown Voltage±7Vac+dcInput Impedance = 1 MΩ
Input Adjustment
Input Trigger
Gate Time
1μHz to 100 MHz50 mVRMS to ±2.5 Vac + dc
100 MHz to 200 MHz100 mVRMS to ±2.5 Vac + dc
1 μHz to 100 MHz50 mVRMS to ±2.5 Vpp
100 MHz to 200 MHz100 mVRMS to ±2.5 Vpp
Sync Output
LevelTTL-compatible
Impedance50 Ω, nominal value
Overvoltage Protection
Occurred when:
The instrument amplitude setting is greater than 2Vpp or the output offset is greater than |2VDC| and the input voltage is greater
than ±11.5 × (1 ± 5%)V (<10kHz).
The instrument amplitude setting is lower than or equal to 2Vpp or the output offset is lower than or equal to |2VDC| and the
input voltage is greater than ±3.5 × (1 ± 5%)V (<10kHz).
General Specications
Power Supply
Power Voltage100 V to 240 V (45 Hz to 440 Hz)
Power ConsumptionLower than 40 W
Fuse250 V, T3.15 A
Display
Type3.5-inch TFT LCD
Resolution320 horizontal × RGB × 240 vertical resolution
Color16 M color
Environment
Temperature Range
Cooling MethodFan cooling
Humidity Range
Altitude
Mechanical
Dimensions (W×H×D)261.5 mm × 112 mm × 318.4 mm
Weight
Interfaces USB Host, USB Device, LAN
IP ProtectionIP2X
Calibration Interval1 year recommended calibration interval
Operating: 0℃ to 50
Non-operating: -40℃ to 70
Lower than 30℃: ≤95% relative humidity
30℃to 40℃: ≤75% relative humidity
40℃ to 50℃: ≤45% relative humidity
3 V/m (1.4 GHz to 2 GHz)
1 V/m (2.0 GHz to 2.7 GHz)
IEC 61000-4-4:20041 kV power lines
0.5kV (Phase to Neutral)
IEC 61000-4-5:2001
0.5kV (Phase to PE)
1 kV (Neutral to PE)
IEC 61000-4-6:20033V,0.15MHz-80MHz
Voltage dip:
0 % UT during half cycle
IEC 61000-4-11:2004
0 % UT during 1 cycle
70 % UT during 25 cycles
Short interruption:
0 % UT during 1 cycle
Electrical Safety in line with
USA:UL 61010-1:2012,
Canada: CAN/CSA-C22.2 No. 61010-1-2012
EN 61010-1:2010
DescriptionOrder Number
DG1022Z (25MHz, Dual-channel)DG1022Z
DG1032Z (30MHz, Dual-channel)DG1032Z
DG1062Z (60MHz, Dual-channel)DG1062Z
Power CordUSB CableCB-USBA-USBB-FF-150
BNC CableCB-BNC-BNC-MM-100
Quick Guide-
Resource CD (including User’s Guide and etc.)-
16Mpts Memory for ArbArb16M-DG1000Z
Rack Mount Kit (for single instrument)RM-1-DG1000Z
Rack Mount Kit (for dual instruments)RM-2-DG1000Z
40dB AttenuatorRA5040K
10W Power AmplierPA1011
USB-GPIB ConverterUSB-GPIB
HEADQUARTER
RIGOL
TECHNOLOGIES, INC.
No.156,Cai He Village,
Sha He Town,
Chang Ping District, Beijing,
102206 P.R.China
Tel:+86-10-80706688
Fax:+86-10-80720067
Electronic Measurement
Instrument service and support
email:EMD_support@rigol.com