User’s Guide RIGOL
RIGOL OCXO-C08 Highly Stable Clock
Product Overview
The highly stable clock (i.g. OCXO module; with the order No. OCXO-C08) is an option for
RSA5000/RSA3000 series real-time spectrum analyzer and consists of the following parts.
connecting finger: makes electrical connection between the OCXO module and the
real-time spectrum an alyzer.
gilded rails: ensures good electrical connection between the OCXO module and the
OCXO rail grooves of the real-time spectru m an a lyzer.
OCXO: indicates the oven controlled crystal oscillator.
Disassembly&Assembly Instructions
The connecting method of the OCXO module and the real-time spectrum analyzer is as
shown in the figure below.
Connecting Finger
OCXO
Real-time Spectrum Analyzer
OCXO Plate
OCXO Slot
1. Assembly Procedures:
(1) Remove the OCXO plate on the rear panel of the spectrum analyzer by using the
T10 screwdriver.
(2) Insert the OCXO module into the OCXO slot, as shown in the above figure. Ensure
that the OCXO module is firmly connected.
(3) Fasten the OCXO plate with 3 screws.
2. Disassembly Procedures:
(1) Remove the OCXO plate on the rear panel of the spectrum analyzer by using the
T10 screwdriver.
Jan. 2018
User’s Guide RIGOL
Temperature: 0℃ to 50℃; humidity: 0 to 80%RH
Temperature: -20℃ to 60℃; humidity: 0 to 90%RH
34.5 mm × 15.1 mm × 77.3 mm(W × H × D)
(2) Pull out the OCXO module with the needle-nose pliers.
(3) Fasten the OCXO plate with 3 screws.
When disassembling or assembling the OCXO module, please wear antistatic
gloves or make other antistatic precautions to avoid direct contact with hands.
To avoid electric shock, make sure that the instrument is turned off and
disconnected from the power supply before disassembling or assembling the
OCXO module.
Specifications
Note: The specifications can be guaranteed only when the instrument has been warmed up for 1 hour.
Notices to Usage
After the OCXO module is installed, power it on and start up the spectrum analyzer. During
the start-up process, the spectrum analyzer recognizes the OCXO module automatically and
synchronizes with the OCXO module. Note that at this point, no external sync signal is
allowed to be connected to the s pectrum analyzer.
Press System About System Option Info. If the instrument interface shows that
the OCXO-C08 option is activated, the OCXO module and the spectrum analyzer are
successfully synchronized. If it is failed to be activated, the synchronization fails.
After the OCXO module is installed on the spectrum anal yzer, the OCXO
module will still be working if you only turn off the Power key on the front
panel without cutting off the AC power supply on the rear panel. Therefore,
Jan. 2018
please cut off the AC power if you do not use the spectrum analyzer for a long